Fluoroscopy Table
Contents
1. IMPORTANT To use the Carbon Fiber Single Arm Board Unfold the Arm Board just enough to enable it to slide under the table top pad on the Fluoroscopy Table at a 90 angle to the side of the table Ask the patient to lay down on the table and position them to suit the needs of the procedure The weight of the patient will hold the Arm Board base in place under the table top pad Move the upper Arm Board to the angle you need It will stay at that position until you change it DO NOT place undue weight or downward pressure on the Carbon Fiber Arm Board It is a positioning device for the arms and should not be used as leverage to get on or off the table Injury can occur When finished Ask the patient to raise up enough for you to move the Arm Board base from under the table top pad Fold it up and store it away until the next time you need to use it N CAUTION Children must be supervised when around this equipment by a responsible adult Prone Position at 135 supine Position at 180 3 OPTIONAL ACCESSORIES OPTIONAL ACCESSORIES SPINAL IMAGING PLATFORM Spinal Imaging Platform LL O lt lt H O 6 The OAKWORKS Spinal Imaging Platform can support patients in a prone or supine position during pre operative sedation or post operative recovery It is also very useful for anesthesia applications This product is portable and can be used
2. E 2 54 cm typical 73 66 cm typical EXTENDED FOR A P VIEW 2 3 8 25 13 97 cm 20 96cm 38 96 52 cm 26 25 66 68 cm 55 88 cm 23 75 60 33 cm 666 90 1 2 54 cm typical 73 66 cm typical OBLIQUE 30 VIEW 5 5 5 5 13 97 cm 13 97cm 14 5 36 83 cm 38 96 52 cm 14 5 36 83cm 13 07 33 20 cm 141 14 11 cm 11 O Z O 11 OQ 11 C gt LL C O 11 x o 78 Carbon Fiber Top DIRECTIONS FOR USE Usable Imaging Area USABLE IMAGING AREA CAUDAL 30 VIEW CENTERED 7 81 5 5 5 5 7 81 19 84 cm 13 97 cm 13 97cm 19 84cm 21 53 34 cm Pn 55 88 cm 23 75 60 33 cm 2 54 cm typical CAUDAL 30 VIEW FULLY EXTENDED 5 5 5 5 13 97 cm 13 97cm 21 53 34 em 15 62 39 67 cm 55 88 cm 23 75 60 33 cm 2 54 cm typical DIREGTIONS FOR USE Usable Imaging Area 90 Carbon Fiber Top amp Detector Shelf Table USABLE IMAGING AREA CENTERED FOR A P VIEW 5 5 5 5 13 97 cm 13 97cm 36 96 52 cm 20 5 57 07 cm 23 75 60 33 cm 1 2 54 cm typical 43 50cm typical EXTENDED FOR A P VIEW 5 5 5 5 13 97 cm 13 97cm
3. Fluoroscopy Table with Trendelenberg Polycarbonate Top Instruction Manual for all models Polycarbonate Top standard Carbon Fiber Top Carbon Fiber Top with integrated Face Rest Carbon Fiber Top with Fiat shelf Standard Carbon Fiber Carbon Fiber with Integrated Face Rest Carbon Fiber Detector Shelf Medical Equipment O Copyright 2006 OAKWORKS Medical Equipment a division of OAKWORKS Inc Manual Part MMMNST0010 Printed in U S A All rights are reserved No part of this docu ment may be photocopied reproduced or trans lated to another language without prior written consent of OAKWORKS Medical Equipment a division of OAKWORKS Inc OAKWORKS 15 a registered trademark of OAKWORKS Inc Notice The information contained in this document is subject to change without notice and should not be construed as a commitment OAKWORKS Inc OAKWORKS Inc assumes no responsibility for any errors that may appear in this document nor does it make expressed ot implied warranty of any kind with regard to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose OAKWORKS Inc shall not be liable for incidental or consequential damages in connection with or arising out of the furnishing performance or use of this document and the program material which it describes TABLE OF CONTENTS TABLE
4. 12 6 LL 3 5 55 88 cm 12 O 30 48cm OQ E E a 2 54 cm typical 73 66 cm typical EXTENDED FOR A P VIEW 5 5 8 25 13 97 cm 20 96cm 26 25 38 12 66 68 cm 96 52 cm 30 48cm 22 55 88 cm 12 80 48cm Sel 2 54 cm typical 73 66 cm typical OBLIQUE 30 VIEW 5 5 5 5 13 97 cm 13 97cm 14 5 36 14 5 12 36 83 cm 96 52 cm 30 48cm 13 07 33 20 em 7 13 18 11cm 1 2 54 cm typical 73 66 cm typical DIRECTIONS FOR USE Usable Imaging Area 90 Carbon Fiber Top w Integrated Face Rest USABLE IMAGING AREA CAUDAL 30 VIEW CENTERED 781 58 5 5 761 6 48 19 84 cm 13 97 cm 13 97em 19 84em 616 46 em e 53 34 cm m 55 88 cm 12 30 48cm 11 C LL O LU x OQ q e e e e e ON l 1 2 54 cm typical CAUDAL 30 VIEW FULLY EXTENDED 5 5 56 6 48 13 97 cm 13 97em 16 46 em cm 53 34 29 55 88 cm 12 30 48cm 666 1 2 54 em typical 20 DIRECTIONS FOR USE Usable Imaging Area MOVING FLUOROSCOPY TABLE Be sure to lower table completely before attempting to move it LL cp O LL C Z O LL 0 at ae v 4 Unlocked caster Lo
5. 38 96 52 em 25 25 64 14 cm 15 75 40 01cm ee A a eo re d 23 75 60 33 cm 1 17 18 2 54 cm typical 43 50cm typical OBLIQUE 30 VIEW 5 5 5 5 13 97 cm 13 97em 38 96 52 cm 20 5 57 07 cm 13 07 33 20 cm L R 1 3 I 14 11 IQ 35 84 cm 1 2 54 cm typical 43 50cm typical 11 C O LL cp Z O 11 OQ LL C c O LL C Z O LU OQ DIREC TIONG FOR USE Usable Imaging Area 90 Carbon Fiber Top amp Detector Shelf Table USABLE IMAGING AREA CAUDAL 30 VIEW CENTERED 11 04 5 5 5 3 11 04 28 04 cm 13 97 cm 13 97cm 28 04cm 23 75 60 33 cm 8 7 2 54 cm typical 43 50cm typical CAUDAL 30 VIEW FULLY EXTENDED 2 5 5 5 5 23 13 97 cm 13 97em 13 28cm 16 85 42 81cm 23 75 60 33 cm 17 13 43 50cm typical 1 2 54 cm typical 18 DIRECTIONS FOR USE Usable Imaging Area 90 Carbon Fiber Top w Integrated Face Rest USABLE IMAGING AREA CENTERED FOR A P VIEW 5 5 5 5 LU 13 97 cm 13 97em 14 5 38 14 5
6. Earth Ground Terminal This symbol when used 10 this manual or on product labels warns that during transport there should be no stacking of containers This symbol when used 10 this manual or on product labels indicates that the product should be protected from moisture The humidity specifications for Transport amp Storage listed on page 18 This symbol when used 10 this manual or on product labels indicates that information 5 given regarding the recommended temperature limits during transport and storing This symbol when used 10 this manual or on product labels indicates the date of manufacture of the device This symbol when used in this manual or on product labels indicates alternating current AC This symbol when used in this manual or on product labels indicates direct current DC O H gt O WL Ze 11 E 00 gt gt Zz O gt O LLI 27 QO lt lt WARNINGS PRECAUTIONS INDICATIONS gt P P P P P P The OAKWORKS Fluoroscopy Table is indicated for use with mobile or compact C arm Fluoroscopy Imaging Systems where the x ray generator is located below the tabletop It is suitable to use for diagnostic x ray imaging and imaging during therapeutic procedures such as spinal injections vertebroplasty procedures and other pain management procedures CONTRAINDICATIONS The OAKWOR
7. OF CONTENTS Symbol pg Indications amp Contraindications pg 2 Warnings amp PISCQUHONS 545220 ds Mass ECO CEA CNi LANEN 02 2 Z Section Product Description amp Photo Pohea Donite TA Rann n EE EE pg 4 Carbon Fiber Top pg 5 5 Carbon Fiber Top w Integrated Head Rest 9 95 956 pg 6 Carbon Fiber Top w Detectot Shelf pg 7 lt Section Il Directions for Use Adus ng Heicht e Or Table pg 9 Hand Control pg 10 Operating Lett Richt Lravel Extension dese pg 11 on sls Cable Pao po 12 Positioning of Integrated Face Rest Pad Configurations pg 12 Usable Imaoine Arca Das msn pg 13 Moving the 1 pg 21 Section Ill Troubleshooting If Table Will Not Change Height or Angle pg 22 TABTE Top y Nor iG as a pg 22 Hand Control Replacement Instructions pg 23 Poot Control Replacement Instructions essees pg 24 Control Dox Replacement Instructions Z a na pg 25 ale IESE ihe Says ni cds pg 29 Gronnad PONE TE pg 29 Section IV Product Care and Maintenance pg 30 Section V Optional Accessories Carbon Fiber Ann Bo
8. U S A INSTRUCTION MANUAL Medical Equipment
9. below 1 Remove all cables from the Control Box The cables are secured in place by a retainer clip Grasp the retainer clip at either end and pull upward Work the retainer clip completely free from the cable outlets and set aside 25 TROUBLESHOOTING e Control Module Replacement Removing Retainer clip from Control Box cables Retainer Clip Remove lt E Remove all cables from their plug outlets then plug them back in again 2 Restore power to the Fluoroscopy Table by plugging the power cable into the power outlet If a different power outlet is used be sure that it is a grounded outlet 3 Attempt to operate the table functions from the Foot Control and the the Hand Control If the table functions operate return the cable retainer clip to its original position If the table fails to operate then replace the Control Module Control Box Replacement Remove power from the Fluoroscopy Table and discharge the Control Box A before proceeding 1 Unplug the power cord from the power outlet 2 Discharge the Control Box by pressing and holding any actuator button on either the foot control or hand control for twenty seconds 3 Follow the steps outlined above to remove the cable retainer clip from the Control Box and unplug all cables from their outlets It is recommended that the cables are marked or labeled so that they may be pl
10. on any treatment examination or surgical table 32 PRODUCT IDENTIFICATION Description Fluoroscopy Table Polycarbonate Top 3 prong erounded hospital grade power cord plug North America Fluoroscopy Table Carbon Fiber Top 3 prong grounded hospital grade power cord plug North America Fluoroscopy Table Polycarbonate Top Continental Plug Europe Fluoroscopy Table Carbon Fiber Top Continental Plug Europe Fluoroscopy Table Polycarbonate Top Swiss plug Switzerland Fluoroscopy Table Carbon Fiber Top Swiss plug Switzerland Fluoroscopy Table Polycarbonate Top British plug United Kingdom Fluoroscopy Table Carbon Fiber Top British plug United Kingdom Fluoroscopy Table Polycarbonate Top 2 prong polarized plug Japan amp Korea PRODUCT IDENTIFICATION FLUOROSCOPY TABLE Model No North America FLRIXXXXXXXXHZ FLRTXXXXXXXXCF Europe FLRTEUXXXXXXHZ FLRTEUXXXXXXCF SWiIzerano FLRISWXXXXXXHZ FLRTSWXXXXXXCF United Kingdom FLRTUKXXXXXXHZ FLRTUKXXXXXXCF RR RENEE FLRTJPXXXXXXHZ FLRTJPXXXXXXCF Fluoroscopy Table Carbon Fiber Top 2 prong polarized plug Japan amp Korea lt gt O FE LU O O al 33 lt O EE Z 11 OQ Q PRODUCT IDENTIFICATION PRODUCT IDENTIFICATION Power Cord Variables Model No FLRIXXX
11. KS Fluoroscopy Table should not be used with Fluoroscopy systems having intensifier screens or film cassettes larger than 12 inches 30 cm when an oblique angle of view is being used The table is not designed for and should not be used with Magnetic Resonance Imaging procedures Although the table accommodates a number of procedures it is not intended to setve as a surgical procedure table WARNINGS Improper use of this device can cause injury Be sure to read all operating instruc tions prior to use Weight limit patient and accessories Polycarbonate Top 205 kg 450 lbs Carbon Fiber Top 250 kg 550 lbs A pinch hazard exists under the side rail on either side of the Fluoroscopy Table The tabletop slides from side to side on a roller track system Be sure to keep fingers and arms away from the table s slide mechanism Patient injury can result if the table s slide mechanism 18 not locked prior to tilting the table Be sure to lock the table with the locking knobs provided prior to operating the tilt feature whenever a patient is present The table utilizes four locking casters to permit movement of the table within the imaging suite Accidental movement of the table may occur Lock at least two casters prior to accomplishing imaging of the patient Be certain that the table is completely lowered without any tilt being present prior to discharging an ambulatory patient The patient may lose balance and fall Elect
12. V Control Module 4796 06 Integrated Face Rest Pad 2 H x 12750 1 4795 06 Integrated Face Rest Pad 4 H 12780 1 2410 06 Crescent Face Pad 4495 02 Tabletop Pad for Polycarbonate Top 24 x 78 1 6404 02 Tabletop Pad for Carbon Fiber Top 22 x 78 1 6405 02 Tabletop Pad for Carbon Fiber Top 22 x 90 1 6406 02 Tabletop Pad for Carbon Fiber Top 24 x 78 1 6407 02 Tabletop Pad for Carbon Fiber Top 24 x 90 1 4788 Table Top Pad for Carbon Fiber Top w Integrated Pace Rest 22 x 90 1 Optional Accessories 3395 01 Carbon Fiber Arm Board 1 SPCFXXXXXXXX Spinal Imaging Platform 1 4796 06 2 5em 12 31cm Square Support Pad 1 7381 Battery 36 WARRANTY WARRANTY Please visit www oakworks com for the most current warranty information gt ZA lt 2 lt 3 NOTES 38 NOTES 39 Fluoroscopy Table with Trendelenberg CONTACT INFORMATION OAKWORKS Medical Equipment 923 East Wellspring Road New Freedom PA 17349 Toll Free USA only 800 916 4612 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com European Representative for CE Mark Information amp Compliance Emergo Europe Molenstraat 15 The Hague 2513 BH Netherlands Tel 31 70 3458570 Fax 31 70 3467299 E mail info emergogroup com Manual Part MMMNST0010 11 14 08 Edition 9 November 2008 English Printed in
13. XXXXX 1 Ri so FLRTEUXXXXXX FLRTSWXXXXXX 4 FLRTUKXXXXXX FLRTJPXXXXXX m 34 Voltage AC 120v 60 Hz 220v 50 Hz 220v 50 Hz 220v 50 Hz 100v 50 60 Hz Power Cord Plug 3 prong grounded hospital grade North America Continental Plug Europe Swiss plug Switzerland British plug United Kingdom 2 prong polarized plug Japan amp Korea SPECIFICATIONS SPECIFICATIONS Designed For North Americo Europe Japan Mode of Operation Continuous Operation Continuous Operation Continuous Operation O 10 Duty Cycle 10 Duty Cycle 10 Duty Cycle 2 Z lt O O LLI O 22 Fiber Fiber piper us 22 amp 24 79 em Height Ranges 31 43 Standard 79 cm 109 79 cm 109 cm 29 41 Custom Low 74 cm 104 cm 74 cm 104 cm Tabletop Travel Range 11 11 28 cm 28 or 13cm Weight 350 Ibs 415 Ibs 136 kg 242 kg 136 kg 242 kg Shipping Weight 400 Ibs 465 Ibs 173 kg 280 kg 178 kg 280 kg Tabletop Capacity 450 Ibs 550 Ibs 205 kg 250 kg 205 kg 250 kg Storage amp fk Temperature 10 C 60 C Humidity 60 relative humidity Pressure no limitations known Transport During transport DO NOT stack containers Materials of Construction This product contains no latex Frame Extruded aluminum with powder coated or anodized surfaces Tabletop Polycarbo
14. able When operating the table s controls be sure to observe all cautions and warnings DIRECTIONS FOR USE Adjusting Height and Tilt Angle of the Table HAND CONTROL OPERATIONS LU C O LL Z O LU Left Side Right Side Til Control Tilt Control activates tilt up to 15 activates tilt up to 15 Trendelenberg Trendelenberg position position Reverse Trendelenberg Reverse Trendelenberg positioning stops when positioning stops when table is level press again table 16 level press again for complete reverse for complete reverse Table UP Table DOWN Control Control elevates tabletop lowers tabletop tilt is preserved during elevation table is lowered from does not operate at full any position Trendelenberg position DIREC HONS FOR USE Operating the Left Right Travel Extension OPERATING THE LEFT RIGHT TRAVEL EXTENSION The table 16 designed to allow travel to either the right or the left depending on the length of the top see pg 35 for specifications on travel range Left Right Travel a 7 LL O LIL C Z O LL OQ A locking knob 1s provided on each side of the table just under the side rall Under normal operation when a patient 18 present at least one of these knobs should be secured firmly Rotate the locking knob several turns counter clockwise to loosen Slide the tabl
15. ard uuu geo mre pg 31 Spina 11111 0 pg 92 ude 11 lt O LIL 6 11 lt TABLE OF CONTENTS CONTINUED Section VI Product pg 33 Section VII SPECIFICATIONS pg 35 SECTION VIII List of Parts pg 6 WY 1 0 E 00 7 CONTACT 11 1111 11 1 ucun aos CS DU Back Cover gt Er 0 gt P The OAKWORKS Fluoroscopy Table with Trendelenberg Feature is a radiographic table intended for use with mobile or compact stationary C arm Fluoroscopy Systems It is ideally suited for pain management imaging and therapeutic procedures This symbol when used in this manual and on product labels representsa caution warning Be sure to read and comply with all precautions and warnings This symbol when used in this manual and on product labels warns against an electrical shock hazard Be sure to observe and comply with all warnings This symbol when used in this manual and on product labels indicates the potential of exposure to harmful x rays Be sure to read and comply with all warnings This symbol when used in this manual and on product labels indicates that the table and components are a Type B Applied Part pursuant to IEC 601 1 and EN 60601 1 1990 This symbol when used in this manual or on product labels indicates a Protective
16. ate the functions of tilt or elevation and the Foot Control is operational To Replace the Hand Control follow these steps Remove Plug lt O JE 1 6 1 Remove the plug from the socket by pulling firmly upward 2 Obtain the new Hand Control Part No 2048 3 Insert the plug into the socket applying pressure until it is firmly seated Firmly Seat Plug 4 Test the functions of the Hand Control Should the new Hand Control fail to operate contact the Customer Service Department 23 TROUBLESHOOTING Foot Control Replacement FOOT CONTROL REPLACEMENT INSTRUCTIONS Replacement of the Foot Control will be necessary if the Foot Control does not actuate the function of elevation and the Hand Control is operational To Replace the Foot Control follow these steps 1 The Foot Control is connected to a shielded plug connector adjacent to the Control Module located below the base of the Fluoroscopy Table 0 lt O O af C O Fluoroscopy Table Control Module Shielded Plug Connector AN 2 Before proceeding disconnect the power cord from the power outlet 3 Access the connection to the Foot Control by lifting the top entry panel on the base Foot Control Cable Connection 4 The Foot Control cable is secured to the shielded cable plug through a friction fit connection Grasp the shielded cable plug and
17. cked Caster Table locked down 2 TROUBLESHOOTING If Table Top will not move TROUBLESHOOTING IF THE TABLE TOP WILL NOT CHANGE HEIGHT OR ANGLE Good Connection Poor Connection lt o O L 1 5 O Cv Check the outlet to be sure that it has power and that the power cable is plugged in Unplug the power cable Check all connections underneath the table by turning the table onto it s side and inspecting all of them Make sure that all male and female connectors are firmly pushed into each other Even a slight misconnection can cause possible difficulty A The Lexan Polycarbonate table weighs 159 kg 350 Ibs and the Carbon Fiber table weighs 198 ke 415 lbs Turning the table on its side is not required and not recommended IF THE TABLE TOP WILL STILL NOT MOVE With the table plugged in press the bottom right button down indicator button 4 shown on pg 10 on the hand control and hold for at least 30 seconds Release then press the bottom left button up indicator button 3 shown on pg 10 This should re activate the system Press any button to make the table top move IF THE TABLE TOP WILL NOT SLIDE Check to be sure that both sliding top locking knobs are loosened TROUBLESHOOTING Hand Control Replacement HAND CONTROL REPLACEMENT INSTRUCTIONS Replacement of the Hand Control will be necessary if the Hand Control does not actu
18. e in the desired direction and extension Rotate the locking knob clockwise to secure the table against traveling Travel can occur 1f the locking knob 1s not secured by inadvertently pushing on the end of the table or by tilting the table When operating the table s locking knob be sure to avoid pinching Keep your fingers and other materials clear from pinch point When operating the table s slide mechanism be sure to observe all cautions and warnings to prevent injury to both the operator and the patient 1 DIRECTIONS FOR USE Removal of Table Pad Use and Positioning of the Integrated Face Rest Pad Configurations Underside of Tabletop fastener LL C c O LIL Z lt LL OQ Remove Table Pad by pulling flaps at the underside of the table on either end until the hook and loop fastener becomes unanchored Replace pad by first centering on the table and then pressing flaps in place anchoring with the hook and loop fasteners Prone Positioning Use the Crescent Face Pad and adjust accordingly Pads are held in place by Pad can be placed close in Or further away for hook amp loop fasteners to the edge of the table for patients who are tall smaller patients Supine Positioning Use either the 2 or 4 square pad for proper placement 25 SOF Pads are provided that give Pads attach to the integrated face rest two levels of elevation extension with hook amp
19. fications Chart pg 35 PRODUCT DESCRIPTION Carbon Fiber Top standard Corbon Fiber Top Hand Control 1 ws Foot Control 1 Z A 0 LLI O A O al Locking Castors Electronic Liff Towers Table Specs e Electronic Height Range 31 43 custom option 29 41 for 22 width only e Top Sizes 22 x 90 24 x 78 24 x 90 e Patient Weight Capacity 550 Ibs Weight 415 Ibs shipping 465 Ibs e Trendelenberg amp Reverse Trendelenberg 111 15 e Manual Traveling Top 22 overall travel on 78 length top 10 overall travel on 90 length top e Controls foot control amp hand control Included Accessories Table Top Pad with TerraTouch Upholstery and 1 Medical Grade Foam Options Battery Backup Carbon Fiber Arm Board Fluoro Extender spinal Imaging Platform For electrical soecs see Specifications Chart pg 35 PRODUCT DESCRIPTION Carbon Fiber Top w Integrated Head Rest Carbon Fiber Top with Removable Table Top Pad Face Cushion Z A O cp LU O OQ O Foot Control 1 Integrated Face Rest Includes 3 pad Set 1 2 B 12 Pad 2 4 H x 12 50 Pad 3 Crescent Pace Pad Electronic Lift Towers i Locking Casters 2 sets Table Specs e Electronic Height Range 31 43 custom optio
20. ith TerraTouch Upholstery and 1 Medical Grade Foam Battery Backup Carbon Fiber Arm Board Fluoro Extender spinal Imaging Platform For electrical soecs see Specifications Chart pg 35 DIRECTIONS FOR USE Adjusting Height and Tilt Angle of the Table DIRECTIONS FOR USE The OAKWORKS Fluoroscopy Table with Trendelenberg Positioning offers a variety of positioning capabilities for diagnostic x ray imaging and imaging during therapeutic procedures Although it was designed principally for use during pain management x ray diagnostic and therapeutic procedures it may also be used for other diagnostic procedures involving x ray imaging pro vided the instructions in this manual are observed Should the use of this device create a circumstance under which the A patient could be over exposed to the x ray being used discontinue use immediately and determine an alternative radiology table or alternative x ray generating source to use ADJUSTING HEIGHT AND TILT ANGLE OF THE TABLE LL gt O LL C gt O 11 OQ All OAKWORKS Fluoroscopy Tables come complete with one foot control module and a hand control to operate the height and Trendelenberg tilt functions of the table Operate the controls as shown below to raise or lower the height of the table and to increase or decrease the angle of tilt up to a maximum of 15 on either 51 Foot Control used only to change height of t
21. loop fasteners that 2 or 4 hold the pad securely in place 12 DIRECTIONS FOR USE Usable Imaging Area 78 Polycarbonate lt ex lt O lt LLI tm lt CENTERED FOR A P VIEW SINUS 5 13 97cm 38 96 52 cm 5 5 13 97 em 12 56 31 91 em 23 75 60 33 cm EXTENDED FOR A P VIEW 7 31 18 57cm 38 96 52 cm 5 5 13 97 cm 25 13 63 82 cm 23 75 60 33 cm OBLIQUE 30 VIEW 55 5 6 N Or ES CON e tO O Ww N Or 12 56 31 91 cm 23 75 60 33 cm 13 DIRECTIONS FOR USE Usable Imaging Area 78 Polycarbonate USABLE IMAGING AREA CAUDAL 30 VIEW CENTERED 6 6 5 5 13 97 16 76cm 21 53 34 cm N LO N LO O o om E 2 O Ww O NO NY ASM CAUDAL 30 VIEW FULLY EXTENDED 5 5 13 97 em 7 31 18 57cm 21 53 34 cm 13 20 63 82 cm 23 75 60 33 cm I4 78 Carbon Fiber Top DIRECTIONS FOR USE Usable Imaging Area USABLE IMAGING AREA CENTERED FOR A P VIEW 55 5 5 13 97 cm 13 97cm 38 96 52 cm 14 5 36 83 cm 14 5 36 83cm 2 35 55 88 cm 23 75 60 33 cm
22. n 29 41 e Widths 22 Length 90 e Weight Capacity 550 Ibs Weight 415 Ibs shipping 465 Ibs e Trendelenberg amp Reverse Trendelenberg 111 15 Manual Traveling Top 22 overall travel e Controls foot control amp 1 hand control e Included Accessories Options Table Top Pad with TerralTouch Upholstery amp Medical Grade Foam 1 2 H 12 se Head Rest Pad 1 4 H x 12 se Head Rest Pad 1 Crescent Face Pad Battery Backup Carbon Fiber Arm Board Fluoro Extender For electrical specs see Specifications Chart pg 35 Carbon Fiber Top w Flat Detector Shelf Carbon Fiber Top PRODUCT DESCRIPTION Carbon Fiber Flat Detector Shelf EN Control 1 Z A cp LL OQ O A O al Electronic Lift Towers Foot Control 1 Table Specs The Carbon Fiber Fluoroscopy Table with Flat Detector Shelf 15 compatible with Canon CXDI 50G Portable DR System and Siemens Medical Solutions Mobilett XP Digital Imaging System Electronic Height Range Widths Length Patient Weight Capacity Weight Trendelenberg amp Reverse Trendelenberg Manual Traveling Top Controls Included Accessories Options Sy 24 78 550 Ibs 415 lbs shipping 465 Ibs 11 15 22 overall travel 78 length top 1 foot control amp 1 hand control Table Top Pad w
23. nate 5 24 x 78 1 27 cm 61 cm x 198 cm Carbon Fiber 65 22 24 x 78 90 1 65 cm 56 61 cm x 198 229 cm Patient Comfort 1 25 mm upholstered foam pad This product complies with United States Department of Health and Human Services radiation performance standards 21 CFR Subchapter J in effect at the time of manufacture for radiographic tables Aluminum Filtration Equivalent of Tabletop HVL 2 7 mm Typical 1 39 mm Polycarbonate top 1 27 mm Carbon Fiber top Maximum at 1 50 mm HVL 3 6 mm Typical 1 62 mm Polycarbonate top Maximum at 1 74 mm HVL 3 6 mm Typical 2 12 mm Carbon Fiber Top Maximum at 2 5 mm ETL Listed Complies with UL 2601 1 version 2 CSA C22 2 No 601 1 M90 IEC 601 1 1 2 2201 9 35 LIST OF PARIS LIST OF PARTS Part No Description 0781 Locking Caster 4 2046 Lift Tower with Mounts 2 lt 2048 Hand Control with Coil Cord amp Plug 1 O 5 2009 Foot Control with Cord amp Plug 2 7 21 Traverse Support Wheel 3616 Traverse Locking Knob 3870 Cable T Connector 4026 4 120 Volt Control Module 4026 8 220 Volt Control Module 4026 3 100 Volt Control Module 5097 1 Hospital Grade Power Cord North America 5097 2 Power Cord United Kingdom 5097 3 Power Cord Continental 5097 4 Power Cord Swiss 3381 T3 15 AL 250V Fuse for 100V Control Module 3382 T2 5 AL 250V Fuse for 120V Control Module 3380 11 25 AL 250 Fuse for 220
24. nto any other surfaces or mechanisms Disinfectants other than recommended above will harm the table s surface particularly the polycarbonate tabletop Wipe off excess with a lint free cloth Remove residual with a damp not wet cloth TABLE PAD Clean with a mild detergent solution Use of gluteraldehydes for disinfecting is not recommended 10 sodium hypochlorite solution phenol based surface disinfectants and quartenary ammonium compounds may be used Wipe off excess liquid and remove any residual solution with a damp not wet lint free cloth Be sure underside of table pad is completely dry prior to placing back onto tabletop The operator should ascertain the disinfecting properties of the agent being used prior to cleaning OPTIONAL ACCESSORIES OPTIONAL ACCESSORIES CARBON FIBER ARM BOARD Carbon Fiber Arm Board Upper Arm OAKWORK Board rotates as needed q full 180 Corbon Fiber Arm Board base slides CAUTION under rable top pad When making adjustments in positioning be sure to observe all cautions and warnings given in the manu alto prevent injury to both opetator and patient C 3 ma O 11 3 O lt lt lt 0 O O The OAKWORKS Carbon Fiber Single Arm Board conveniently slides under the table top pad and easily supports the patient s arm in a positionary range of 180 The unit can easily fold for compact storage when not in use
25. ons still do not operate contact the Customer Service Department 9 Install the Control Box onto the base assembly using the four 4 Phillips head screws removed previously Reconnect the cable retainer clip 28 TROUBLESHOOTING e Initialize and reset the system INITIALIZE THE SYSTEM 1 Lower both columns completely down 2 Press the down button on the hand control or foot control and hold down for 30 seconds RESET THE SYSTEM QO O 6 1 Lower both columns completely 2 Unplug the table from power source for 2 3 seconds 3 Plug table back into outlet 4 Hold down button on the hand control or foot control down for 30 seconds GROUND POINT TESTING 1 Pop the base top off 2 Connect the tester to one of the silver or gold bolts found under the top 3 Replace the top when finished Lu O lt TT lt lt gt 2 11 m lt O O D OQ O e Q PRODUCT CARE amp MAINTENACE PRODUCT CARE amp MAINTENANCE CLEANING IMPORTANT IMPORTANT CAUTION Before cleaning with any liquid cleaner be sure to unplug the power cord from the power outlet Be sure to read all cautions warnings and instructions given in the manual to prevent injury to both operator and client 30 TABLE Use only a mild detergent solution or 10 sodium hypochlorite solution on surfaces Be sure excess liquid does not drip o
26. rical Shock Hazard The power supply control module is located below the base of the table No user serviceable parts are inside Refer servicing to qualified personnel Unplug wall connector prior to contact with any cables connected to the power supply WARNINGS PRECAUTIONS WARNINGS cont A The potential of exposure to harmful x rays exists when this table is in use The use of adequate x ray barrier devices is necessary to provide protection to both the operator and the patient X ray barrier devices are recommended for the patient outside of the intended target area to prevent exposure to scattered radiation from the x ray generating source The Polycarbonate Tabletop of the Fluoroscopy Table has a typical aluminum filtration equivalence of 1 39 mm as measured at 100 kVp and a half value layer HVL of 2 7 mm or 1 62 mm as measured at 100 kVp and a half value layer HVL of 3 6 mm The Carbon Fiber tabletop has an aluminum filtration equivalence of 1 27 mm as measured at 100 kVp and a half value layer HVL of 2 7 mm or 2 12 mm as measured at 100 kVp and a half value layer HVL of 3 6 mm C O gt gt 5 A QO Z 27 lt The OAKWORKS Fluoroscopy Table may be used for x ray imaging where the x ray generator 15 located below the tabletop and the image receptor is located above the tabletop Using the Fluoroscopy table for x ray imaging where the x ray generator is located above
27. sed Side of Control Module 2 TROUBLESHOOTING e Control Module Replacement 2 Remove the six 6 star head screws that secure the upper and lower cases of the Control Module and open the Control Module 3 Locate the power supply board It will be connected to the power cable socket and contain a mounting for the fuses Remove the cable connector from the Power Supply Board and gently lift the Power Supply Board from the Control Module housing O lt ele Power Cable Supply Board Connector 4 Remove and replace both fuses Replace with OAKWORKS Part IMPORTANT No 3383 2 5 amp slo blow fuses or equivalent Use only 2 5 amp slo blow fuses rated for 250 volts For 220 volt versions of the Fluoroscopy Table use OAKWORKS Part No 3380 1 25 amp slo blow fuses or equivalent for 100 volt versions use Part No 3381 3 15 amp slo blow fuses or equivalent 5 Replace the Power Supply Board to the housing after re connecting the cable removed above 6 Replace the cover and secure with the six 6 star head screws Replace label onto the housing 7 Before installing the Control Box temporarily reconnect all the cables and the power cable 8 Test the Control Box to make sure it operates correctly by pressing any of the switches on either the Foot Control or the Hand Control If the table operates correctly proceed to the next step If the table functi
28. the Foot Control plug and work the two apart until fully disconnected 24 IMPORTANT TROUBLESHOOTING e Foot Control Replacement e Control Module Replacement 5 It will be necessary to cut and remove the cable tie that holds the Foot Control cable in place Be sure to secure the new Foot Control cable with a new cable tie 6 Obtain the new Foot Control Part No 2049 and insert the cable plug into the shielded cable plug until firmly seated Z Firmly Seat Plug ER 5225 6 7 Replace top entry panel on base Restore power to the table by plugging the power into the power outlet If a different power outlet is used be sure that it is a grounded power outlet 8 Attempt to operate the table elevation functions from the Foot Control 9 If the table elevation functions fail to operate follow the Control Module replacement instructions to replace the Control Module Should this fail to return the Fluoroscopy Table to an operating condition contact the Customer Service Department CONTROL BOX REPLACEMENT INSTRUCTIONS Replacement of the Control Box may be necessary if the troubleshooting steps fail to remedy a non operating condition The Control Box is located below the base of the Fluoroscopy Table An offset screwdriver set is required Before proceeding disconnect the power cord from the power outlet and follow the additional troubleshooting steps
29. the tabletop and the image intensifier receptor is located below the tabletop is not recommended due to the risk of excessive patient exposure to x ray The Fluoroscopy Table may be used with the x ray generator above the tabletop and a film cassette also located on the tabletop The x ray generator should never be located above the tabletop when the OAKWORKS Fluoroscopy Table and the OAKWORKS Spinal Imaging Platform are used together This type of use requires that the x ray generator is located below the tabletop and the imaging intensifier ot film cassette located above the tabletop Disconnect the power supply plug prior to cleaning any surfaces below or inside of the base of the table PRODUCT DESCRIPTION Polycarbonate Top Polycarbonate Top Z A O cp LU O OQ O Electronic Lift Towers Locking Casters Foot Control 1 Table Specs e Electronic Height Range 31 43 e Widths 24 e Length FO e Patient Weight Capacity 450 5 Weight 350 Ibs shipping 430 Ibs e Trendelenberg amp Reverse Trendelenberg 111 15 e Manual Traveling Top 22 overall travel e Controls foot control amp hand control e Included Accessories Table Top Pad with Terralouch Upholstery and 1 Medical Grade Foam e Options Battery Backup Carbon Fiber Arm Board Fluoro Extender spinal Imaging Platform For electrical specs see Speci
30. ugged into the correct outlets on the new Control Box 4 Remove the four 4 Phillips head screws holding the Control Box to the base frame of the Fluoroscopy Table and remove the Control Box 26 TROUBLESHOOTING e Control Module Replacement Remove screws holding Control Box to base 5 Before installing the new Control Box temporarily reconnect all the cables and the power cable QO lt als z 6 6 Test the Control Box to make sure it operates correctly by pressing any of the switches on either the Foot Control or the Hand Control If the table operates correctly proceed to the next step If the table functions still do not operate contact the Customer Service Department 7 Install the Control Box onto the base assembly using the four 4 Phillips head screws removed previously Reconnect the cable retainer clip If it is not possible to obtain a replacement Control Box for example in an emergency or when contact with our Customer Service Department is not possible the person providing service may attempt to replace the fuses in the Control Box In some cases this may return the Fluoroscopy Table to an operable condition To replace the fuses first follow the steps above to remove the Control Box from the Fluoroscopy Table and then proceed to the steps below 1 Gently peel back the label from the side of the Control Module until the parting seam 16 fully expo