IFU FJ-Hook
Contents
1. VJ 014 Warning Do not force or screw the filter holder too tightly onto the introduction sheath If necessary use another filter Warning To avoid puncturing the introduction sheath during filter progression ensure that the introduction sheath is handled by its base and is not bent Slowly and carefully push the filter fully using the piston such that it penetrates the conical part of the filter holder The filter is now in position inside the introduction sheath 15 19 15 Warning If one or more filter strands are deformed in the filter holder do not insist and use another filter Remove the empty filter holder by rotating it counter clockwise 16 7 Advancing the filter to the placement site Under medical imaging move the filter down using the pushing catheter Align the pushing catheter s black mark with the base of the introduction sheath 17 17 Warning If difficulties are encountered pushing the filter through the introduction sheath femoral route in particular do not insist Remove the assembly and select a different approach to insert a new filter If necessary adjust the position of the filter in the inferior vena cava the hook on the filter head below the lowest renal vein Position the radio opaque ring of the 7F introduction sheath approximately 6 5cm below the lowest renal vein if prior tests have confirmed possible placement beneath the lowest renal vein Figure 3
2. 14 In the event of treatment with anti vitamin K a heparin relay should be defined with the prescribing physician ten days before the procedure 15 In case of preventive indication for ALN Vena Cava Filter with hook implantation in particular prior to surgery with a high risk of pulmonary embolism it is essential to ensure that the surgical team in charge of the patient is aware of the presence of the ALN Vena Cava Filter with hook 16 ALN warns users of the increased risk of migration in the following cases major surgery thorax abdomen etc procedures potentially modifying vascular hemodynamics procedures that may cause contact with the inferior vena cava and or the filter placement site 17 Being aware of the above mentioned potential risks it is important to ensure that all elements pertaining to ALN Vena Cava Filter with hook implantation are logged in the patient record which is checked prior to each procedure 18 Before the procedure the elements of the kit J guide and introduction system dilator introduction sheath should be rinsed with a sterile physiological saline solution or similar isotonic solution 19 The entire procedure must be conducted under medical imaging 8 20 X rays performed without radio contrast medium which do not clearly show the anatomy of the inferior vena cava may be misinterpreted 21 Strict bed rest for 12h after ALN Vena Cava Filter with hook implantation is strongly recommended
3. 38 Gently and carefully push the filter using the piston such that it penetrates the conical part of the filter holder If one or more filter strands are deformed in the filter holder do not insist and use another filter ALN therefore recommends that at least two implantation kits be kept available 39 If difficulties are encountered pushing the filter through the introduction sheath femoral route in particular do not insist Remove the assembly and select a different approach to insert a new filter ALN therefore recommends that at least two implantation kits be kept available General warnings relative to deploying the ALN Vena Cava Filter with hook 40 Do not release the filter until it is in its optimum position 4 Take care to avoid deploying the ALN Vena Cava Filter with hook in a thrombus this could prevent the filter from opening fully and from correctly anchoring in the inferior vena cava 42 Once the base of the introduction sheath reaches the black pusher marker deploy the filter by immobilizing the base of the pushing catheter and slowly sliding the introduction sheath out This ensures optimum filter positioning 43 Do not attempt to remove or reposition the filter once the hooks have been engaged into the walls of a vessel except by means of the ALN extraction and or repositioning kit or of a lasso suitably sized for vena cava filter removal intended for this purpose The sharp hooks could cause severe tissue damage 44
4. 5 16 T Rinse 6 2 4 4 Advance the guide in the needle Place the end of the J guide Upper route subrenal Leave an appropriate length of guide exposed Hold the guide in place and remove the needle 6 6 Warning During J guide progression check for the absence of bends or loops monitoring the progress under medical imaging Always advance the J guide further than the placement site and ensure that 2 or 3 cm of J guide extends beyond the proximal end of the introduction system 7 7 Warning Do not attempt to advance or remove the guide in the event of resistance The cause of resistance should be determined by radioscopy If the cause of resistance cannot be determined remove the guide and catheter together To facilitate progress of the introduction system we recommend making a sufficiently large incision to the tissues surrounding the guide Rinse the introduction system with a sterile physiological saline solution or similar isotonic solution 4 Advancing the introduction system introduction sheath dilator and cavography Insert the outer part of the J guide into the distal end of the introduction system 17 Advance the introduction system along the vein by sliding over the J guide 8 8 Advance the end of the introduction system to the vicinity of the iliac bifurcation Remove the J guide 9 E wu 9 Remove the dilator by r
5. Warnings relative to use of the J guide 22 The J guide should be used by a physician with appropriate cardiac catheterisation training Never advance the J guide or introduction system without radioscopic control 23 During J guide progression check for the absence of bends or loops monitoring the progress under medical imaging Always advance the J guide further than the placement site and ensure that 2 or 3 cm of J guide extends beyond the proximal end of the introduction system 24 Do not attempt to advance or remove the guide in the event of resistance The cause of resistance should be determined by radioscopy If the cause of resistance cannot be determined remove the guide and catheter together 25 Make sure to remove the J guide and dilator before performing the control cavography in order to avoid causing any trauma to the vena cava Cavography related warnings 26 Prior frontal and lateral cavography must be performed The purpose of this examination is to check the permeability and anatomy of the inferior vena cava ipsilateral iliac veins and renal veins measure the length and diameter of the subrenal inferior vena cava and detect any possible variations in inferior vena cava anatomy double vena cava left inferior vena cava agenesis etc diagnose inferior vena cava disease partial or total vena cava thrombosis extrinsic adenopathy related compression tumor syndrome or aortic aneurysm Determine the most s
6. Placement of an ALN Vena Cava Filter with hook for other indications is performed under the physician s responsibility and should follow on from a benefit versus risk analysis for the patient The indications listed below are those acknowledged by the SFICV They are based on the SIR and CIRSE agreements Acknowledged indications Deep vein thrombosis popliteal femoral iliac vein or vena cava with or without pulmonary embolism in patients presenting with one or more of the following characteristics anticoagulant treatment complications requiring interruption of treatment symptomatic recurrence of acute pulmonary embolism under efficient anticoagulant treatment this recurrence must be objectively confirmed symptomatic extension of thrombosis under efficient anticoagulant treatment this extension must be objectively confirmed contraindications to high dose anticoagulant treatments temporary or definitive Severe trauma cranial or spinal causing local diffuse hemorrhages in patients for whom preventive mechanical intermittent venous compression means cannot be applied in the demonstrated absence of pulmonary embolism or deep vein thrombosis Prevention of pulmonary embolism during high thromboembolic risk surgical procedures in patients with recent history of deep vein thrombosis or pulmonary embolism Relevant indications Prevention of pulmonary embolism during high thromboembolic risk surgical procedures in pa
7. positioning the ALN Vena Cava Filter with hook under the lowest renal vein following jugular route introduction Warning do not release the filter until it is in its optimum position the most appropriate placement site being confirmed by cavography Warning Deliberate positioning of the filter above the renal veins should only be performed after checking compatibility of the suprarenal vena cava maximum 28 mm and if the available space between renal veins and cardiac atrium is of at least 60 mm filter height 8 Advancing the filter to the placement site Warning Do not deploy the filter if no prior diameter measurements have been made Patients whose inferior vena cava diameter is greater than 28 mm cannot be implanted with an ALN Vena Cava Filter with hook as this latter cannot be correctly anchored and there is an increased risk of filter migration Slide the introduction sheath over the pushing catheter by a slow upwards removal movement without jerking until the filter is released approximately 6 cm 18 19 18 19 1 Warning Once base of the introduction sheath reaches black pusher marker deploy the filter by immobilizing the base of the pushing catheter and slowly sliding the introduction sheath out This ensures optimum filter positioning Warning Do not attempt to remove or reposition the filter once the hooks have been engaged into the walls of a vessel except by means of the ALN extraction and o
8. A poorly positioned filter but that nevertheless confers adequate protection against pulmonary embolism should be left in place If the filter is not positioned in such a manner as to protect against pulmonary embolism it must either be repositioned or a second filter implanted Removal whether surgical using the ALN extraction and or repositioning kit or using a lasso sized for the removal of vena cava filters intended for this purpose is recommended for an incorrectly positioned filter that could have cardio respiratory consequences Other warnings 10 45 The ALN Vena Cava Filter with hook has been tested and shown not to present any additional risks for patients undergoing MRI examination with a static magnetic force of up to 3 Tesla examination lasting 15 minutes or less with a SAR of 3 W kg The quality of MRI images may however be affected if the area of interest is at the exact or approximate location of the ALN Vena Cava Filter with hook The ALN Vena Cava Filter with hook is MR conditional ASTM F2052 See the MRI equipment manufacturer s precautions for use for imaging an implanted metallic prosthesis 46 Any attempt to manipulate the ALN Vena Cava Filter with hook using accessories other than those provided is prohibited and under the practitioner s entire responsibility 47 After use the kit components should be discarded as per usual medical practice and according to applicable regulations 11 7 Complications
9. Filter following an implantation period ranging from 6 to 722 days The mean age of patients who received an indication for removal of their vena cava filter was of 66 8 16 5 years interval 19 96 years 23 The ALN Vena Cava Filter placement indications for these patients were a contraindication to treatment by anticoagulants recent surgery hemorrhage or coagulopathy complications to or failure of anticoagulant treatment extension of deep vein thrombosis new deep vein thrombosis and or pulmonary embolism hemorrhage etc or other prophylactic indication Filters that were implanted due to temporary contraindication to anticoagulant treatment were removed on resumption of treatment Filters that were implanted due to extension of an existing deep vein thrombosis or a new deep vein thrombosis under anticoagulants were removed after 6 months without clinical signs of recurrence Filters implanted due to complications to anticoagulant treatment were removed 3 to 6 months post implantation corresponding to the time required to consider the deep vein thrombosis or pulmonary embolism event as being chronic with a low risk of recurrence In the event of prophylactic indication filters were removed after completion of surgery The ALN Vena Cava Filters were removed after a mean implantation time of 93 15 days interval 6 722 days The ALN Vena Cava Filters were removed using an ALN extraction and or repositioning kit introduced via the in
10. possess a level of expertise in accordance with SFICV1 recommendations This means that it is necessary to have implanted at least one permanent filter and at least one optional or temporary filter in order to implant the ALN Vena Cava Filter with hook under good conditions The SFICV recommends That a Junior Operator should have performed 6 placements in a role of second operator and 6 placements himself herself under the supervision of a senior to be considered as competent That a Senior Operator should perform at least 6 placements per year to maintain his her level of expertise That the electroradiology operator should possess basic vascular procedure training and that the nurses including operating theatre nurses have undergone specific training in invasive and interventional vascular imaging 2 Material resources ALN recommends that ALN Vena Cava Filter with hook placement be conducted as any other vascular interventional procedure under resource and organization conditions as defined by the SFICV Le inferior vena cava measurement should be performed in the angiography room in which the filter is placed During the procedure all patients must be placed under continuous cardiac and intermittent blood pressure monitoring Vital signs should be recorded An intravenous route should be set up for all patients for administering solutes and drugs as required The equipment and drugs required for emergency resuscitation must be immed
11. route placement Do not use for any other approaches femoral or brachial 2 The ALN Vena Cava Filter with hook jugular route is provided ready to use sterile ethylene oxide sterilized and non pyrogenic Do not use the product if its packaging is damaged or open 3 The product should be stored in a cool dry and dark place 4 Do not use the ALN Vena Cava Filter with hook after its use by date 5 The ALN Vena Cava Filter with hook is intended for a single use It must not be reused and the kits components must not be re sterilized 6 Always use the jugular introduction system for a jugular approach the femoral introduction system for a femoral approach and the brachial introduction system for a brachial approach 7 The brachial and jugular kits are distinguished from the femoral and popliteal kits by the orientation of the filter inside the filter holder The kits are also distinguished by colour coding yellow for the brachial and jugular kits green for the femoral and popliteal kits 8 The brachial and popliteal kits are distinguished from the femoral and jugular kits by the length of the introduction system elements and of the J guide 9 The ALN Vena Cava Filter with hook should only be implanted using the introduction kit provided for this purpose by ALN 10 The ALN Vena Cava Filter with hook can be removed using one of the extraction kits sold for this purpose by ALN It can also be removed using a lasso or gooseneck suitabl
12. the filter gently and carefully to ensure that it passes into the filter holder s conical section under the best possible conditions No specific prevention Do not force or screw the filter holder too tightly onto the introduction sheath Ensure before pushing the filter that the dilator and J guide are removed Ensure that once the dilator removed the sheath is handled by its base and not bent 13 Leave the introduction system in place and use a new filter At least 2 implantation kits should be kept ready Leave the introduction system in place and use a new filter At least 2 implantation kits should be kept ready Leave the introduction system in place and use a new filter At least 2 implantation kits should be kept ready Remove the damaged filter and adapter ensure that the introduction sheath is not damaged and use a new filter if necessary If the introduction system is no longer compliant repeat the procedure At least 2 implantation kits should be kept ready 1st case transfixion occurs on a part of the sheath not inserted into the patient Remove the entire introduction system and use a new vena cava filter At least 2 implantation kits should be kept ready 2nd case transfixion occurs on a part of the sheath already inserted into the patient Insert a 9F catheter over the 7F sheath and remove the entire assembly 8 Choice of insertion technique and location The ALN
13. After localization the renal veins can be marked by placing a metallic instrument on the operative field 18 5 Rinsing the introduction sheath Rinse the introduction sheath with a sterile physiological saline solution or similar isotonic solution 11 11 Warning Insufficient equipment rinsing may cause clot formation in the introduction system leading to poor filter anchoring and thus increasing the risk of migration on placement If the implantation procedure is longer than planned do not hesitate to rinse the introduction sheath regularly 6 Inserting the vena cava filter into the introduction sheath Warning Ensure that the J guide and dilator have been removed before inserting the filter into the introduction sheath Remove the filter holder from its sterile bag Remove the protective cap then present the filter holder along the axis of the introduction sheath 12 13 12 13 Warning The filter is packaged in its filter holder by ALN and is ready to use Never use a filter that has been removed from its filter holder nor attempt to re insert it into this latter Do not attempt to modify the filter in any way In particular do not increase or reduce its diameter or change the orientation of the filter in its filter holder Insert the tip of the filter holder into the base of the introduction sheath Rotate the filter holder by one quarter of a turn clockwise onto the base of the introduction sheath 14
14. Aln implants chirurgicaux INSTRUCTIONS FOR USE THE ALN VENA CAVA FILTER WITH HOOK JUGULAR ROUTE Canada 9 69 Product reference FJHOOK 10 2014 1 Contents 1 General 3 Po T 4 o KIE COBUOIWHS 2i iris IR IEEE ERO SEE E E RESO e Noi 5 a 6 35 nicibireiti 7 A E E 7 T euni ii eee 12 8 Choice of insertion technique and location 14 9 Necessary EeSOUEGCGOS 0o eiii see obs Enos ov vasa SE VT ege avos le EE CIV Re quo aas 15 10 Insertion procedUre 000000000000000000000000000000000000000000000000000000000000 16 11 Patient 22 0 22 15 M 23 IET e 23 IS aud handle sete vessseeveosvesbvesvva ve eu e VES 24 16 25 CS T m 25 1 General information The ALN Vena Cava Filter with hook is an optional filter definitive and or temporary It provides effective protection against pulmonary embolism The ALN Vena Cava Filter with hook is made of nine 316 LVM stainless steel strands Six strands are fitted with hooks for anchoring in the vena cava The other three strands serve to centre the filter inside the vena
15. Maximum SAR of 3 W kg for a 15 minute examination 22 During the non clinical test the ALN Vena Cava Filter with hook generates a 0 7 C increase in temperature for a 3 W kg SAR during 15 minutes on a 3 Tesla MRI Excite Software G3 0 052B General Electric Healthcare Milwaukee WI The quality of the MRI image may be impaired if the point of interest is located at the exact or approximate filter placement site In this case it may be necessary to optimize the machine parameters in order to improve the image In all cases see the MRI machine s user manual when conducting an examination on a patient implanted with an ALN Vena Cava Filter with hook 13 Removal Note The ALN Optional Vena Cava Filter can be retrieved within three months after the implantation by using the ALN Extraction kit or using a lasso or gooseneck suitably sized for the removal of vena cava filters After three months the ALN Optional Vena Cava Filter is implanted in a permanent way The wire diameter of the lasso used to remove an ALN Vena Cava Filter must not exceed 0 6mm The removal sheath used with the lasso must have an internal diameter of at least OF The operating procedure for removal using the ALN extraction and or repositioning kit is described in the instructions for use of the ALN Retrieval kits To remove the ALN Vena Cava Filter with a lasso first check using medical imaging that The filter does not display an excessive tilt as this woul
16. There are a number of potential complications associated with the use of vena cava filters Consequently the insertion procedure must not be performed by practitioners unaware of these complications These complications must be explained to the patient prior to implanting the ALN Vena Cava Filter with hook Once aware of the possible complications the benefit to risk ratio for the patient must be assessed prior to implanting the ALN Vena Cava Filter with hook Clinical complications related to implantation include but are not limited to 1 Incorrect filter placement or release Premature or incomplete filter deployment Filter tilting Bruising or hemorrhage at the puncture point Infection of the approach route Inability to anchor the filter in the vena cava potential filter migration to the heart and lungs 7 Pulmonary embolism caused by insertion system manipulation that could displace a thrombus during filter placement 8 Air embolism during filter placement DEAL D Clinical complications include but are not limited to 9 Secondary displacement or migration of the filter 10 Filter migration caused by migration of a large thrombus 11 Thrombus formation in the filter causing caval thrombosis or occlusion 12 Perforation of the vena cava vessels or an adjacent organ by one or more hooks 13 Death caused by migration of a thrombus to the heart and lungs 14 Puncture point thrombosis 15 Breakage of a filter ele
17. Vena Cava Filter with hook can be implanted by percutaneous or surgical approach Upper route access in order of preference o Right internal jugular vein Left internal jugular vein o subclavian vein o Basilic vein Lower route access in order of preference o Right femoral vein o femoral vein o Right popliteal vein o Left popliteal vein Implantation by jugular route Internal jugular vein YW Superior vena cava a Inferior vena cava K wm sm ELEC Implantation by brachial route Position of the filter Implantation by femoral route Implantation by popliteal route Figure 1 Different approach routes available for ALN Vena Cava Filter with hook implantation The approach route is selected according to clinical presentation the chosen insertion technique and the practitioner s experience When employing a surgical approach the right internal jugular vein is the most commonly selected approach The practitioner s choice of approach route is based on clinical circumstances 14 Placement site In absolute terms the hook on the filter head should be placed under the lowest renal vein Lowest renal vein ALN Vena Cava Filter with hook Inferior Vena Cava Figure 2 ALN Vena Cava Filter with hook placement site 9 Necessary resources 1 Human resources ALN recommends that both the operating physicians and non physicians performing the procedures should
18. cava The hook on the filter head allows this latter s removal with a lasso or gooseneck The ALN Vena Cava Filter with hook has been tested and presents no additional risk for patients undergoing MRI examinations with magnetic forces of up to 3 Tesla The quality of MRI images may however be affected 1f the area of interest 1s at the exact or close to location of the ALN Vena Cava Filter with hook See the MRI equipment manufacturer s precautions for use for imaging an implanted metallic prosthesis The ALN Vena Cava Filter with hook device is intended to be implanted in patients whose vena cava diameter is less than or equal to 28 mm Read the instructions carefully before use These instructions should be used only if you have opted for jugular route placement Do not use for any other approaches femoral popliteal or brachial The ALN Vena Cava Filter with hook is delivered sterile It is intended for single use Note The ALN Optional Vena Cava Filter can be retrieved within three months after the implantation by using the ALN Extraction kit After three months the ALN Optional Vena Cava Filter is implanted in a permanent way 2 Symbols This section presents the definitions of the symbols used on the product and its packaging LA Warning Sterilized with ethylene oxide Do not reuse Do not re sterilize Do not use if the packaging is damaged Keep dry 3 Kit contents Ensure that all kit contents ar
19. d hinder or even prevent its removal using a lasso The hook is not in contact with the vessel wall No thrombus is present in the filter If the filter presents significant tilt if the hook is touching the vessel wall or if there is a thrombus in the filter the filter should not be removed The lasso should be prepared as per the manufacturer s recommendations Next Position the lasso above the hook of the ALN Vena Cava Filter with hook Ensure that the sheath slides readily along the lasso Completely open the lasso s loop and slowly engage it onto the hook of the ALN Vena Cava Filter with Hook Slowly lower the sheath over the lasso in order to close it around the hook Continue to slowly lower the sheath until the filter is completely enclosed Never pull on the hook using the lasso Remove the lasso sheath and filter assembly The ALN Vena Cava Filter with hook can also be permanently implanted The decision to remove the filter should be taken individually for each patient according to his her clinical condition 14 Clinical data A clinical trial involving 123 patients was conducted at 3 centers Bellevue University Hospital in Saint Etienne Georges Pompidou European Hospital in Paris and Lille University Hospital between November 2003 and June 2006 in view of assessing removal of the ALN Vena Cava Filter The trial was conducted on 123 patients who had received an indication for the removal of their ALN Vena Cava
20. e present before starting the procedure If an item 1s missing use another kit The ALN Vena Cava Filter jugular with hook route device comprises the following elements The filter 1 located in its filter holder 2 The filter filter holder assembly is packaged in an individual bag The filter holder itself is comprised of a conical tube protective cap clip and piston 3 The piston serves to push the filter into the introduction sheath by means of the metal tube An introduction system comprising a dilator 4 red Luer lock tip screwed onto the introduction sheath 7F 5 A pushing catheter 6 This element serves to insert the filter into the introduction sheath A puncture needle 7 A Teflon coated J guide designed for use in the context of percutaneous catheter insertion The ALN Vena Cava Filter with hook jugular route kit is delivered sterile and is a single use device It is sterilized with ethylene oxide x 2 Filter holder i 1 Filter positioned made Ml the filter holder for jugular 7 mtroduction gt gt _ gt 3 Introduction sheath Insertion system en m 4 Dilator red Luer Lock tip E 6 Pushing catheter di lt _ gt b C X of to 7 Puncture needle 8 Teflon coated J guide 4 Indications The ALN Vena Cava Filter with hook should only be implanted for the indications listed in this section
21. er additional commitments or responsibilities in relation to this product and does not authorize any third parties to do so 17 References SFICV Soci t Frangaise d Imagerie Cardiaque et Vasculaire French Society for Cardiac and Vascular Imaging Interventional Vascular Radiology Recommendations and Best Practices March 2007 Pages 93 101 2 SIR Society of Interventional Radiology Guidelines for the use of retrievable and convertible vena cave filters report from the Society of Interventional Radiology multidisciplinary consensus conference J A Kaufman et al Journal of Vascular Interventional Radiology 2006 17 3 449 59 CIRSE 2007 guidelines Cardiovascular and Interventional Radiological Society of Europe Quality Improvement guidelines for percutaneous inferior vena cave filter placement for the prevention of pulmonary embolism 5 Value of frontal caval measurement in the placement of inferior vena cava filters Deepak R Kaura et al Canadian Association of Radiologists Journal 1999 50 301 305 Guide Wire Entrapment by Inferior Vena Cava filters an experimental study M J Rosen et al Journal of American College of Surgeons Vol 201 No3 2005 25 Manufacturer ALN C 0459 FDA Approval Head i Route acm 20240 Ghisonaccia France Offices 589 Chemin du Niel 53230 Bormes les Mimosas Tel 433 0 4 94 01 05 01 Fax 33 0 4 94 01 09 01 France Limited TU X with a capita
22. iately available and staff trained in their use Necessary equipment not provided by ALN Sterile physiological saline solution or similar isotonic solution for system rinsing Sterile syringe for perfusing the sterile physiological saline solution Contrast medium used for cavography Local anesthetic Basic catheterization materials and miscellaneous consumables 15 10 Insertion procedure 1 Patient preparation Disinfect and position sterile tissues 1 2 2 1 Warning All forms of heparin treatment should normally be stopped before starting the procedure All decisions of a therapeutic nature such as interruption or resumption of anticoagulant treatment shall be assessed on a per patient basis Warning In the event of treatment with anti vitamin K a heparin relay should be defined with the prescribing physician ten days before the procedure 2 Anesthesia and puncture After local anaesthesia puncture the selected internal jugular vein 3 Inserting the J guide and rinsing the introduction system Warning The J guide should be used by a physician with appropriate cardiac catheterization training Never advance the J guide or introduction system without radiological control Flush the J guide with sterile physiological saline or similar isotonic solution fixing a syringe to the Luer base of the guide s distributor 4 Insert the flexible end of the guide into the needle
23. ior vena cava diameter is greater than 28 mm cannot be implanted with an ALN Vena Cava Filter with hook as this latter cannot be correctly anchored and there is a risk of probable filter migration 2 Presence of thrombi In the upper chest region In the right atrium At the puncture point In the ipsilateral iliac vein In the inferior vena cava In all of these cases the thrombi can potentially be displaced causing an embolism while manipulating the insertion system 3 Percutaneous insertion in patients presenting with abnormal coagulation times 4 In all cases in which radioscopic monitoring is contraindicated in the event of pregnancy in particular the advantages and disadvantages of filter placement must be carefully assessed 5 Known hypersensitivity to one of the components constitutes a contraindication to the placement of an ALN Vena Cava Filter with hook 6 If there is a risk of septic embolism placement of the ALN Vena Cava Filter with hook is contraindicated 7 Placement of an ALN Vena Cava Filter with hook is contraindicated in children and young adults 8 ALN Vena Cava Filters with hook should not be placed outside of the areas stated in the Indications section Inferior Vena Cava Superior Vena Cava in rare cases 6 Warnings General warnings relative to the use of an ALN Vena Cava Filter with hook 1 Read the instructions for use carefully These instructions should be used only if you have opted for jugular
24. l of 1 500 000 Euros BASTIA Trade Register B 411 237 704 aer 37 704 00019 PE 3250 A VAT no FR 00 411 237 704 Website http www aln2b com Email address contact aln2b com
25. ment 16 Recurrent pulmonary embolism 17 Long term deep vein thrombosis There are also a certain number of potential J guide related complications after filter implantation in particular 18 Risk of migration 19 Risk of tilting 20 Risk of the J guide becoming trapped in the filter Consequently any endovascular procedures after ALN Vena Cava Filter with hook insertion should only be conducted by practitioners who are fully aware of these complications 12 Known but rare equipment related technical complications that users may encounter include When pushing the filter through the introduction sheath one or more or the legs bend inside the filter holder When pushing the filter into the introduction sheath the filter refuses to enter remaining in its initial position Breakage of the filter holder s Luer lock tip upon or after connection to the 7F sheath Cannot insert the filter into the introduction sheath and filter deformation Introduction sheath perforated when pushing the filter with the pushing catheter and filter leg transfixion Incorrect filter positioning inside the filter holder Overly powerful push By femoral route stem incorrectly positioned under the head By jugular route filter head incorrectly positioned in piston Luer lock tip weakened by manufacturing defect The dilator was not removed Sheath folded during the procedure Push
26. nditions The product should be stored in a cool dry and dark place Do not use the ALN Vena Cava Filter with hook after its use by date The ALN Vena Cava Filter with hook is intended for a single use It must not be reused and the kits components must not be re sterilized After use the kit components should be discarded as per usual medical practice and according to applicable regulations 24 16 Guarantee ALN declares to have manufactured this product with all necessary care This guarantee replaces and excludes all other formal or tacit guarantees granted by virtue of any other provision not expressly stated in the present in particular all tacit guarantees concerning the product s suitability for sale and for specific use as inappropriate handling storage repeat cleaning and repeat sterilization of this product along with certain factors related to the patient diagnosis treatment and surgical procedures etc beyond the control of ALN have an impact on this product and on the results achieved with it ALN cannot be held responsible for any fortuitous or indirect damage nor for losses or costs resulting either directly or indirectly from the use of this product It shall however be obliged to replace it Certain countries do not allow the exclusion of implicit guarantees or accessory or indirect damage You may have access to other legal action in accordance with the laws of your country Furthermore ALN accepts no oth
27. otating the red Luer lock counter clockwise 10 10 Carefully advance the introduction system few centimeters and perform a cavogram to view the renal veins During all of these steps monitor the procedure closely by medical imaging Warning frontal and lateral cavography must be performed before placing an ALN Vena Cava Filter with hook The purpose of this examination is to e check the permeability and anatomy of the inferior vena cava ipsilateral iliac veins and renal veins e measure the length and diameter of the subrenal inferior vena cava and detect any possible variations in IVC anatomy double vena cava left inferior vena cava agenesis etc e diagnose IVC disease partial or total vena cava thrombosis extrinsic adenopathy related compression tumor syndrome or aortic aneurysm e Determine the most suitable placement site in absolute terms the proximal end of the ALN Vena Cava Filter with hook should be placed at the lower level of the joining of the lowest renal vein Warning If there is a doubt concerning the morphology or size of the vena cava new shots should be taken from different angles Warning Do not deploy the filter if no prior diameter measurements have been made Warning Patients whose inferior vena cava diameter is greater than 28 mm cannot be implanted with an ALN Vena Cava Filter with hook as this latter cannot be correctly anchored and there is an increased risk of filter migration
28. r repositioning kit or of a lasso suitably sized for vena cava filter removal intended for this purpose The sharp hooks could cause severe tissue damage Warning A poorly positioned filter but that nevertheless confers adequate protection against pulmonary embolism should be left in place If the filter is not positioned in such a manner as to protect against pulmonary embolism it must either be repositioned or a second filter implanted Removal whether surgical using the ALN extraction and or repositioning kit or using a lasso sized for the removal of vena cava filters intended for this purpose is recommended for an incorrectly positioned filter that could have cardio respiratory consequences 21 9 End of procedure Remove the pushing catheter and perform a post operative control cavography with subtracted and non subtracted images frontal and lateral All images should be archived and available on film or electronic medium appended to the patient record Remove the introduction sheath Check hemostasis at the puncture point Warning After use the kit components should be discarded as per usual medical practice and according to applicable regulations A PA Plain Abdomen X ray image should be taken 24 hours after filter placement This reference document relating to filter stability integrity and positioning should be made available for subsequent viewing included in the patient record 11 Patient follow up Con
29. rarenal vena cava maximum 28 mm with the filter and having checked that the available space between renal veins and cardiac atrium is of at least 60 mm filter height Filter insertion related warnings 33 The filter is packaged in its filter holder ALN and is ready to use Never use a filter that has been removed from its filter holder nor attempt to re insert it into this latter Do not attempt to modify the filter in any way In particular do not increase or reduce its diameter or change the orientation of the filter in its filter holder 34 Ensure that the J guide and dilator have been removed before inserting the filter into the introduction sheath 35 To avoid puncturing the introduction sheath during filter progression ensure that the introduction sheath is handled by its base and is not bent 36 Always make sure that the introduction sheath is adequately rinsed with a sterile heparinized physiological saline solutions or similar isotonic solution Insufficient equipment rinsing may cause clot formation in the introduction system leading to poor filter anchoring and thus increasing the risk of migration on placement If the implantation procedure is longer than planned do not hesitate to rinse the introduction sheath regularly 37 Do not force or screw the filter holder too tightly onto the introduction sheath If necessary use another filter ALN therefore recommends that at least two implantation kits be kept available
30. sidering the above mentioned risks patient follow up should include the following points Examine the catheter insertion point to ensure it is healing correctly A frontal and lateral PA image should be taken before patient release then each year thereafter The patient should attend a follow up consultation after one month by the radiologist who placed the filter by this latter s team or failing this by the clinician who referred the patient for the procedure A written document stating 1 the type of filter filter type and batch number 2 placement date 3 hospital and unit in which the filter was implanted should be provided to the patient and his her attending physician These latter should be informed of the necessary follow up and possible complications In this respect the patient must be issued with the implantation card provided by ALN along with the Guide for persons with ALN Vena Cava Filters Inferior vena cava permeability control imaging should be performed if there is the slightest suspicion of recurrent pulmonary embolism or phlebitis 12 MRI A non clinical test conducted in accordance with the ASTM F2052 standard has demonstrated that the ALN Vena Cava Filter with hook is MR conditional A patient implanted with this filter can safely undergo MRI examination under the following conditions Static magnetic field of 3 Tesla or less Spatial magnetic field gradient of 720 Gauss cm or less
31. ternal jugular vein 122 of the 123 filters were successfully removed success rate 99 The case of failed removal concerns a filter that had been implanted for 45 days and that was severely tilted gt 15 The filter head was in contact with the vessel wall and entirely encased in fibrous tissue For 4 patients a thrombus was present in their filter on removal These patients were administered anticoagulant therapy for 3 weeks before their filter could be removed Moderate tilting 15 was observed in 37 patients and severe tilting 215 in 6 patients In two cases of severe tilting implantation period between 38 and 81 days the filter head was in contact with the vessel wall Both of these filters however were successfully removed During the removal procedure two patients presented with bruising at the puncture site but with no other major complications No migration breakage or filter strand penetration was detected in these patients The removal procedures did not cause any vessel wall damage Early and Late Retrieval of the ALN Removable Vena Cava Filter Results from a Multicenter Study Pellerin O et al Cardiovascular and Interventional Radiology 2008 31 5 889 896 15 Packaging and handling The ALN Vena Cava Filter with hook is provided ready to use sterile ethylene oxide sterilized and non pyrogenic Do not use the product if its packaging is damaged or open The product must be handled under aseptic co
32. tients with recent history of deep vein thrombosis or pulmonary embolism Important note The validation of preventive placement of a vena cava filter prior to a surgical procedure must be performed on a per patient basis and the various practitioners must be aware of the associated risks The decision must be validated in collaboration with all concerned units Possible indications for suprarenal placement The specific indications for suprarenal filter placement are those defined by the CIRSE and SFICV agreements i e renal vein thrombosis and demonstrated pulmonary embolism inferior vena cava thrombosis extending to above the renal veins extension of thrombosis above a previously implanted vena cava filter pulmonary embolism related to isolated thrombosis of an enlarged gonadal vein anatomic variant with renal vein connecting to a very low section of the inferior vena cava or double vena cava Pregnancy is a very special case subrenal placement is not possible and would probably be hazardous In this case an optional suprarenal filter could be considered though there is insufficient experimental evidence in the literature to recommend such a placement Each case must be discussed individually in collaboration with the obstetricians In all cases the suprarenal placement of a vena cava filter should only be proposed after careful per case analysis of risk versus benefit 5 Contraindications Patients whose infer
33. uitable placement site in absolute terms the hook on the head of the ALN Vena Cava Filter with hook should be placed underneath the lowest renal vein 27 In case of pre surgery preventive placement we strongly recommend checking the filter by imaging both before and after surgery These images should be archived and appended to the patient record for viewing whenever desired 28 Ensure that the cavography can be performed without risk for the patient take into account any problems associated with use of contrast media allergy renal failure 29 If there is a doubt concerning the morphology or size of the vena cava new shots should be taken from different angles 30 Do not deploy the filter if no prior diameter measurements have been made The diameter of the inferior vena cava can be measured by frontal or lateral cavography It is important however to note that the cavography may over or underestimate the actual diameter of the inferior vena cava If megacava is suspected the diameter of the inferior vena cava should be measured by ultrasound or scanner 3 Patients whose inferior vena cava diameter is greater than 28 mm cannot be implanted with an ALN Vena Cava Filter with hook as this latter cannot be correctly anchored and there is an increased risk of filter migration 32 Deliberate positioning of the filter above the renal veins see placement indications page 7 should only be performed after checking compatibility of the sup
34. y sized for the removal of vena cava filters see removal section General procedure related warnings 11 Before placing the filter in the inferior vena cava it is important that the emboli originate from the lower limbs the filter being ineffective if the emboli are of cardiac or upper limb origin 12 ALN recommends that both the operating physicians and non physicians performing the procedures should possess a level of expertise in accordance with SFICV recommendations This means that it is necessary to have implanted at least one permanent filter and at least one optional or temporary filter in order to implant the ALN Vena Cava Filter with hook under good conditions The SFICV recommends That a Junior Operator should have performed 6 placements in a role of second operator and 6 placements himself herself under the supervision of a senior to be considered as competent That a Senior Operator should perform at least 6 placements per year to maintain his her level of expertise That the electroradiology operator should possess basic vascular procedure training and that the nurses including operating theatre nurses have undergone specific training in invasive and interventional vascular imaging 13 All forms of heparin treatment should normally be stopped before starting the procedure All decisions of a therapeutic nature such as interruption or resumption of anticoagulant treatment shall be assessed on a per patient basis
Download Pdf Manuals
Related Search
Related Contents
ISTRUZIONI PER L'USO OPERATING INSTRUCTIONS Aborder la formation linguistique en contexte professionnel Guide Oregon Scientific BAR608HGA Radar Detector User Manual Creative UP-7000 Patient Monitor User Manual HOSE CATCHER VE.90000 Kicker 2014 DXA Stereo Amplifier Owner's Manual MTT Raportti 6 Lincoln IMPINGER CONVEYOR OVENS 1600 User's Manual operation - Whaley Food Service KDC-C711/661(K,E)(En,F,Es,P) Copyright © All rights reserved.