Standard Power Pack P25 Manual

image

Contents

1. losses resulting from insufficient decontamination of the device e Please prepare written confirmation use the Equip ment Decontamination Declaration following on the next page that the device is free of biologically dangerous chemical or radioactive contaminants This confirmation must be attached to the outside of the packaging e Use the original packing or a similarly robust packing when returning the device If not available contact Biometra or your local distributor e Label the outside of the box with CAUTION SENSITIVE INSTRUMENT and the RAN number sticker Attach the Decontamination Declarartion e Please enclose a note which contains the following a Sender s name and address b Name of a contact person for further inquiries with telephone number c Precise description of the fault which also reveals during which procedures the fault occurred if possible 19 Biometra An Analytik Jena Company 9 0 Equipment Decontamination Certificate To enable us to comply with german law i e 71 StrISchV 17 GefStoffV and 19 ChemG and to avoid exposure to hazardous materials during handling or repair please complete this form prior to the equipment leaving your laboratory COMPANY INSTITUTE ADDRESS PHONE NO FAX NO E MAIL EQUIPMENT Model Serial No If on loan evaluation Start Date Finish Date Hazardous materials used with this equipment Meth
2. pP green No error Power on pP dark Error e No connection to power e Faulty fuse e Device failure The Power On LED remains dark although the device is switched on when there is an operation or hardware failure In such case please verify that e the power outlet supplies electricity e the Standard Power Pack P25 is correctly connected to the power source e the fuses in the power inlet socket are not defective Caution Disconnect the device from the power A N supply by unplugging the power cord before any fuse replacement 14 Biometra An Analytik Jena Company 6 2 Display Indication The actual and set values for the voltage and current outputs will alternatively be shown on the LCD display see chapter 5 4 Monitoring Voltage and Current The two LEDs on the right side of the display indicate to which parameter the value shown on the display belongs The following rules apply LED indication Display indication mA LED illuminated Output current in mA VOLT LED illuminated Output voltage in volt mA LED flashes Set value for the current control in mA VOLT LED flashes Set value for the voltage control in volts 15 Biometra An Analytik Jena Company 6 3 Operating Status Display The Standard Power Pack P25 is in one of the following three operating modes voltage limitation current limitation and
3. de sign The design and manufacture of parts have been conceived with one purpose to produce units which will give satisfactory service Biometra GmbH guarantees this unit to be free from de fects in materials or workmanship under normal use or service for 24 month from date of shipment If during this time this unit proves defective in materials or workmanship Biometra GmbH will repair or replace it free of charge if returned to us prepaid This guarantee does not cover damage in transit dam age caused by carelessness misuse or neglect or un satisfactory performance as a result of conditions beyond our control or consequential losses as a result of failure of our product 25 Biometra An Analytik Jena Company Biometra GmbH Service Department Rudolf Wissell Str 30 Rudolf Wissell Str 14 16 D 37079 Goettingen D 37079 Goettingen Phone 49 0 5 51 50 68 6 0 Phone 49 0 5 51 50 68 6 10 or 12 Fax 49 05551 50686 66 Fax 49 0 551 50 68 6 11 e mail Info biometra com e mail Service biometra com internet http www biometra de 26
4. power limitation depending on the selected voltage and current settings and the ohmic resistance of the electrophoresis chamber The operating mode is indicated by the two LEDs located below the display The following rules apply LED indication Display indication CURRENT LED Current limitation illuminated VOLTAGE LED illuminated Voltage limitation CURRENT LED and Power limitation VOLTAGE LED illuminated In the voltage control mode the output voltage will remain equal to the selected value see also chapter 5 5 Working with Constant Voltage Should the output current exceed the selected default value the current limitation function of the Standard Power Pack P25 is activated The output voltage is now adjusted to the ohmic resistance of the electrophoresis chamber The output current remains limited to the selected default value see also chapter 5 6 Working with Constant Current Should the user s selected voltage and current defaults exceed the maximum allowable output power of 200W the device will reduce the maximum allowable output current according to figure 1 in chapter 5 7 Both LED indicators are illuminated when the current limitation is accessed to indicate that the output current due to the power limitation is lower than the value originally set by the user 16 Biometra An Analytik Jena Company 7 0 Cleaning Disconnect the
5. should not exceed 300mA The following set values will be applied Output voltage 150V Output current 300mA 10 Biometra An Analytik Jena Company 5 6 Working with Constant Current Adjust the current controllers set value to the desired output current see chapter 5 3 Current Setting when working with constant current The voltage controller s set value will then have been selected so high that the device will switch to current limit mode see chapter 5 2 Voltage Setting When the current output has reached the desired set value the Standard Power Pack P25 will limit the current The voltage output will then adjust to the ohmic resistance of the electrophoresis chamber and the current output will remain constant The LED indicator CURRENT see chapter 6 3 Operating Status Display is illuminated throughout the current limitation mode Should the set value for the voltage controller be VAN selected too low the output voltage will therefore not suffice to reach the desired current The output power of the Standard Power Pack P25 is VAN limited to 200W The maximum allowable current output of the device is limited to 200V as shown in figure 1 of chapter 5 7 Application example A semi dry blot should be conducted with a constant current of 800 mA For safety reasons the output voltage should not exceed 50V The following set values will be applied Output voltage 50V Output current 800mA 11 Biometr
6. Biometra An Analytik Jena Company Standard Power Pack P25 Order No 040 800 Manual Attention VAN Please read this manual carefully VAN prior to operating the device Biometra GmbH Rudolf Wissell StraBe 30 D 37079 G ttingen P O Box 1544 D 37005 G ttingen Phone 49 0 5 51 50 68 6 0 Fax 49 0 5 51 50 68 6 66 E mail Info biometra com Internet http www biometra com Service Division Rudolf Wissell Stra e 14 16 D 37079 G ttingen Phone 49 0 5 51 50 68 6 10 or 12 Fax 49 0 5 51 50 68 6 11 E mail Service biometra com Biometra An Analytik Jena Company This document describes the state at the time of publishing It needs not necessarily agree with fu ture versions Subject to change Standard Power Pack P25 Manual 2010 11 Biometra An Analytik Jena Company 1 0 Intended Uses and Specification The Biometra Standard Power Pack P25 is suitable for PAGE SDS PAGE agarose gel electrophoresis blotting and electroelution of proteins and DNA RNA from gels The Standard Power Pack P25 is intended for use with any electrophoresis applications designed to operate below 400 volts and 500 milliamps DC electrophoresis or 200 volts and 1000 milliamps DC blotting Four sets of outputs operate in parallel to provide 0 to 400 volts up to 1000 milliamps and up to 200 watts in constant voltage or constant current mode Vo
7. Power Pack Use only mild non abrasive detergents and a slightly moistened soft cloth Do not use solvents soak while cleaning or immerse in water 17 Biometra An Analytik Jena Company 8 0 Service Should you have any problem with the device please contact our service division or your local Biometra dealer Biometra GmbH Service Division Rudolf Wissell StraBe 14 16 D 37079 G ttingen Phone 49 0 5 51 50 68 6 10 or 12 Fax 49 0 5 51 50 68 6 11 E mail Service biometra com Please follow the return instructions when returning the device Instructions for return shipment e Return only defective devices For technical problems which are not definitively recognisable as device faults please contact the Technical Service Department at Biometra Tel 49 0 5 51 50 88 1 10 or 12 Fax 49 0 5 51 50 88 1 11 e mail Service biometra com e Please contact our service department for providing a re turn authorization number RAN This number has to be applied clearly visible to the outer box Returns without the RAN will be not be accepted e Important Carefully clean all parts of the instrument from residues and of biologically dangerous chemical or radioactive contaminants If an VAN instrument is contaminated Biometra will be 18 Biometra An Analytik Jena Company forced to refuse to accept the device The sender of the re pairorder will be held liable for possible damages and
8. a An Analytik Jena Company 5 7 Power Limitation The output power of the Standard Power Pack P25 is limited to 200W Should the user s selected voltage and current defaults exceed the maximum allowable output power the device will reduce the default set value of the output current as shown in figure 1 1A Power Par Limitation 5 a 0 5 A 5 O OA OV 200 V 400 V Output Voltage Figure 1 Automatic limitation of the output current at high output voltage values The CURRENT and VOLTAGE LED indicators are both illuminated when the current limitation is accessed see chapter 6 3 Operating Status Display to give a signal that the output current due to the power limitation is lower than the value originally set by the user 12 Biometra An Analytik Jena Compan 5 8 Switching Off The Standard Power Pack P25 can be turned on or off by pressing the on off switch The device must always be turned off before the supply line to the electrophoresis chamber is to be connected or removed for any given reason power switch on 0 and LED above not illuminated 13 Biometra An Analytik Jena Company 6 0 Status Notification 6 1 Power ON LED A green LED located above the on off switch shows along with the on off switch position if the Standard Power Pack P25 works properly on off switch LED Meaning position indication O dark No error Power off
9. bly to the operating instructions will dissolve any responsibility and liability Observe the relevant rules for accident prevention as well as those of approved safety and occupational medicine Do not attempt any repair yourself In need of repair consult your local qualified service representative Only qualified personnel may open the housing parts The inner device may still contain dangerous voltage several minutes after disconnection from power supply Make sure the indicated voltage and frequency range of the device corresponds with the system voltage and that the power switch is on 0 before it is connected to the power supply Connect the device to a properly grounded outlet only Pull the power cord out of the outlet to safely disconnect from the power supply gt gt gt PP gt gt gt gt gt Biometra An Analytik Jena Compan Use exclusively the provided power cord or one that meets the legal requirements Pay attention to the dimensioning of the fuses and replace them if required The fuse holder is located in the lower part of the power connector on the back side of the Power Pack The following fuses are required Micro fuses slow blow 2 x 5A 250 V Caution For continued protection against risk of fire replace only with same type and rating of fuse Caution Double pole neutral fusing disconnect power before changing fuse In order to ensure the safe electrical
10. disconnection the floating outputs must never be grounded Make sure that your test setup always complies to this condition Only safe electrical cords certified according to IEC EN 61010 2 031 may be connected to the power outputs Make sure the Standard Power Pack P25 is located in a safe and dry place in proximity to the electrophoresis devices powered Keep away from liquids Should a liquid penetrate the P25 disconnect the device immediately gt gt e b amp b D Biometra An Analytik Jena Compan Do not operate in a damp or humid environment where condensing moisture may lead to creeping current See chapter 2 Technical Data for ambient operating temperature range When moving from a cold to a warm room allow 8 hours before reusing the device in order for it to reach the ambient temperature and for possible condensation to evaporate Use only mild non abrasive detergents Carefully clean the device to avoid liquids penetrating the inner housing When positioning the P25 make sure the ventilation openings on the back side are not covered Should the Power Pack become extremely warm smoke or display unusual information while in operation discontinue use immediately and contact a qualified service representative Biometra An Analytik Jena Company 4 0 Set Up 4 1 Content e Standard Power Pack P25 e Manual e Power Cord 4 2 Unpack and Check Your Biometra Standard Power Pack P25 has bee
11. electrophoresis chamber s before switching on the Standard Power Pack P25 Then switch on the device using the power switch A green light appears over the power switch when power is turned on When power is turned on the P25 sets the output voltage or current to the latest value stored Should you be unsure of the latest saved settings detach all connections to the electrophoresis chamber and switch on the Standard Power Pack P25 in order to check the voltage or current settings see chapter 5 2 Voltage Setting and chapter 5 3 Current Setting The device must always be turned off before the supply line to the electrophoresis chamber is to be connected or removed for any given reason power switch on 0 and LED not illuminated 5 2 Voltage Setting By using the SELECT button switch to VOLT on the display Setthe desired output voltage by turning the knob In normal mode the display shows the output voltage actual value By turning the knob the display changes from the actual value to the desired set value recognizable by the flashing VOLT LED indicator The desired value increases with every clockwise rotation and decreases with every counter clockwise rotation Depending on speed and direction of the rotation the value will augment or reduce in steps of 1V 7 Biometra An Analytik Jena Compan 10V or 100V A set value modified by the user will 3 seconds after the last operation of the
12. knob precede all other values and will be set as the desired output voltage The voltage value V should be set without any device connected 5 3 Current Setting By using the SELECT button switch to mA on the display Set the desired output current by turning the knob In normal mode the display shows the current conduction actual value By turning the knob the display changes from the actual value to the desired set value recognizable by the blinking mA LED indicator The desired value increases with every clockwise rotation and decreases with every counter clockwise rotation Depending on speed and direction of the rotation the value will augment or reduce in steps of 1mA 10mA or 100mA A set value modified by the user will 3 seconds after the last operation of the knob precede all other values and will be set as the desired output current The electric current value mA should be set without any device connected Biometra An Analytik Jena Company 5 4 Monitoring Voltage and Current 5 4 1 Monitoring Set Values Choose the desired measurement category by using the SELECT button in order to monitor the default voltage and current values Then turn the knob to the right or to the left by one detent and the LCD displays will show the desired set value of the corresponding measurement The value appearing on the display will 3 seconds after the last operation of the knob precede all other va
13. ltage and current setting is adjustable in steps of 1V and 1mA A single LCD display provides output voltage or current readings Biometra An Analytik Jena Company 2 0 Technical Data 2 0 Technical Data OL Type Output Max Voltage Max Current Max Power Output Display Accuracy Number of Output Terminals Input Power Allowable deviation Ambient Operating Temperature Range Humidity Dimensions Weight Const Voltage and Const Current 400 VDC 1000mA in the range 0 200VDC lt 1000mA in the range 200 400VDC 200 W continuous LCD display 3 5 digits selectable between display mA and V Resolution 1 mA and 1 V respectively 5 v M for U gt 0 05 Umax Imax 5 of 0 05 Umax Imax for U I lt 0 05 Umax Imax Arithmetic mean over 3 s Ts 0 4 s Four sets of 4 mm safety sockets 100 240VAC 50 60Hz max 300W 6 10 0 to 40 C 0 to 70 W x Lx H 26 5 x 25 0 x 9 5 cm 3 kg The Standard Power Pack P25 is designed for use with AC net supply voltages between 100 Volt and 240 Volt 50 60Hz It is a switched mode power supply providing voltage and current controlled DC power respectively The DC output is free of ground and short circuit proof See also chapter 5 7 Biometra An Analytik Jena Company 3 0 Safety Instructions The Standard Power Pack P25 must only be used in accordance with its designation The use of this device unconforma
14. lues The LCD display will then automatically be switched to the actual value 5 4 2 Monitoring Actual Values Choose the desired measurement category at any time by using the SELECT button in order to control the output voltage and current In doing so only the LCD display will be switched The supplied current or applied voltage will not be affected by this operation 5 5 Working with Constant Voltage Adjust the voltage controllers set value to the desired output voltage see chapter 5 2 Voltage Setting when working with constant voltage Then select the maximum possible or allowable value for the current controllers set value The current controllers set value will then be the maximum possible or allowable value chosen see chapter 5 3 Current Setting The Standard Power Pack P25 supplies a constant output voltage until reaching the selected current limit voltage control The output current then varies according to the ohmic resistance of the electrophoresis chambers The output of a 9 Biometra An Analytik Jena Company constant voltage will be shown by the LED indicator VOLTAGE see chapter 6 3 Operating Status Display Should the maximum allowable output current be exceeded the Standard Power Pack P25 will limit the current and the output voltage will drop Application example An agarose gel electrophoresis should be conducted with a constant voltage of 150V For safety reasons the output current
15. n carefully packed to ensure its safety Please unpack it carefully control the integrity of its content and inspect the device Report any damage immediately to BIOMETRA Do not operate a device which is in no proper condition Should you have a complaint return the complete system in its original packing by following the return instructions see chapter 8 0 N Attention AN Please send back the enclosed postcard along with the filled out Warranty Certificate to Biometra in order to claim the full warranty Biometra An Analytik Jena Compan 4 3 Power Supply Before connecting make sure the actual voltage complies with the voltage range indicated on the type label The Standard Power Pack P25 must be grounded 4 4 Location The Standard Power Pack P25 must be located in a dry and safe place Do not place an electrophoresis chamber on the power supply device The ventilation openings on the back side must not be covered by other devices furniture or fixtures A sufficient air flow for convection cooling must be provided 4 5 Electrophoresis Chamber Connection The electrophoresis chamber is connected to the Standard Power Pack P25 by safety cords Use exclusively electrophoresis devices with safety connectors Switch off the Standard Power Pack P25 while working on the test equipment setup Biometra An Analytik Jena Company 5 0 Operation 5 1 Power On Establish the connection to the
16. od of cleaning decontamination The equipment has been cleaned and decontaminated NAME POSITION HEAD OF DIV DEP INSTITUTE COMPANY SIGNED DATE PLEASE RETURN THIS FORM TO BIOMETRA GMBH OR YOUR LOCAL BIOMETRA DISTRIBUTOR TOGETHER WITH THE EQUIPMENT PLEASE ATTACH THIS CERTIFICATE OUTSIDE THE PACKAGING INSTRUMENTS WITHOUT THIS CERTIFICATE ATTACHED WILL BE RETURNED TO SENDER 20 Biometra An Analytik Jena Company General Information for Decontamination Please contact your responsible health amp safety officer for details Use of radioactive substances Please contact your responsible person for details Use of genetically change organism or parts of those Please contact your responsible person for details 21 Biometra An Analytik Jena Company 10 0 Note for Disposal of Eletric Electronic Waste This symbol the crossed out wheelie bin means that this product should be brought to the return systems and or separate systems available to end users according to yours country regulations when this product has reached the end of its lifetime For details please contact your local distributor This symbol applies only to the countries within the EEA EEA European Economics Area comprising all EU members plus Norway Iceland and Liechtenstein 22 Biometra An Analytik Jena Company 11 0 Notes 23 Biometra An Analytik Jena Compan
17. y 12 0 EU Konformit tserkl rung EC Declaration of Conformity G ttingen 2010 im Sinne der EG Richtlinie ber elektrische Betriebsmittel zur Verwendung innerhalb bestimmter Spannungsgrenzen 73 23 EWG following the EC directive about electrical equipment for use within certain limits of voltage 73 23 EEC und and im Sinne der EG Richtlinie f r die elektromagnetische Vertr glichkeit 89 336 EWG following the EC directive about the electromagnetic compatibility 89 336 EEC Hiermit erkl ren wir da das folgende Power Pack Herewith we declare that the following Power Pack Typ type Standard Power Pack P25 Rev b Best Nr Order No 040 800 den grundlegenden Anforderungen der corresponds to the basic requirements of EG Niederspannungsrichtlinie 73 23 EWG und der EC low voltage directive 73 23 EEC and the EG Richtlinie ber die elektromagnetische Vertr glichkeit 89 336 EWG entsprechen EC directive about the electromagnetic compatibility 89 336 EEC Folgende harmonisierte Normen wurden angewandt The following harmonized standards have been used EN 60601 1 2 2001 EN 55011 1998 A1 1999 A2 2002 EN 61000 3 2 2000 EN 61000 3 3 1995 A1 2001 EN 61000 6 1 2001 EN 61010 1 2001 Dr J rgen Otte Quality Manager 24 Biometra An Analytik Jena Company 13 0 Warranty This laboratory instrument is produced with the highest practical standards of materials workmanship and

Download Pdf Manuals

image

Related Search

Related Contents

Bienvenue à Sorel-Tracy - Magazine Plaisirs d`été  Olympia P324ERBD User's Manual  CONCASSEURS GIRATOIRES  École de Sculpture Architecturale. Pilules Cuivreuses Anti  PBX-PFX-USER-N Rev C.indd  ロボシリンダ RCP5/RCP5CR アクチュエータ スライダタイプ 取扱説明書      SERVICE TOOL MAP110 USER MANUAL  Table of contents Contents of packaging  