         Contents
1.       Lo    be TERMS OF EACH PART AND FUNCTIONS    1  FRONT PART      START   STOP BUTTON  Press    Start    to start the measurement  The cuff starts winding up automatically and begins to  pressurize  Press the button to stop the measurement  Pressurization will stop and the air will  be exhausted from the cuff     2  INFORMATIOM BOARD  This signs explain the measuring method and notice  Please follow the signs before the  measurement     3  LCD DISPLAY  The measurement process is displayed in letters and animations  After the test  the result is  displayed  Systolic  Diastolic blood pressure  Pulse  Pulse wave pattern  Blood pressure     4  CLOCK SECTION  Date and Time   On the right upper corner of LCD screen  date and time are displayed     5  PRINTER COVER  It protects the printer     6  Printer outlet of the result sheet  Result sheet is automatically printed out  Automatic cutter is equipped inside and it cuts the  result sheet automatically    D CUFF  It automatically wraps and releases the arm for measurement    ARM REST  When the arm is placed on the cuff  it sustains the arm and makes the right position      HUMAN SENSOR  When a user approaches  power is automatically turned on and vice versa  When the user is  wearing black  it may not sense the movement  Therefore  the user in black clothes rolls up  the sleeves and measures    EMERGENCY STOP BUTTON  When a user feels the pain due to high pressurization or irregular operation  press the button  to stop the opera
2.    EX PLUS    Automatic Blood Pressure Monitor       e 7 inch wide color LCD  Touch Screen    e Safety design with soft curved line   e Simple one touch operation     Instructions by voice messages     High speed printer with automatic paper cut    Built in emergency safety button    e Cuff guide ring with 150mm size for  comfortable measurement       R  SYSTOLIC    DIASTOLIC              MJAWON        O  aD STP   STOP  lt      O       SR    The device bears the CE label in accordance with the provisions of Medical Device  Directive 93 42 EEC     THE PERSONS RESPONSIBLE FOR PLACING DEVICES ON THE EC MARKET  UNDER MDD 93 42 EEC    JAWON MEDICAL CO  LTD   29  Gongdan 4 ro  Jillyang eup  Gyeongsan si  Gyeongsangbuk do  Korea  TEL  82 53 856 0993  FAX  82 53 856 0995       VITAKO Sp  Z 0 0   EC RER UL  MALEJ SYRENKI 2 71 790 SZCZECIN  POLAND  TEL  91 8522900   901  FAX  91 8522910    TABLE OF CONTENTS       INTRODUCTION  E 5  LANDICATIONS Eed 5  Keier Ing LE D  3  CLASSIFICATION AND COMDUIANCE ccc cici iiien 6  4  GAFETY PRECAUTION Seci iieo raai E rA 6  5  SAFETY SYMBOLS AND INFORMATION    9  6  Guidance for Electromagnetic compatbitvEMC 2222 12   e TERMS OF EACH PART AND FUNCTIONS  icien 16  TFRONTPAR EE 16  2  REAR FAR E 18  Ee 20  A E ETC 20   s  lle EEN LE DEE 21  1  CONNECTING ADAPT ER  te EE 21  2  LOADING THE PRINT PAPER  E 21  3  CONNECTING  PORTS  EE 23  4  CONNECTING DRINTER ccc ce ce eeececeeeeeeceeeeeeeaeeeeceeeeeeeeseeeeaeaeseeeeaeesssaeeeeseeeeees 23  5  CONNECTING
3.  Lt  l   The difference of your Bact   inter arm pressure is         39    bd MAINTENANCE    1  Pay attention to allowable current value of power   2  Avoid direct sunlight  humidity  dust  thick oil and salty or extreme changes in temperature   3  Do not install or store this device in some space where any chemicals or gas is stored   4  Do not use this device in any unstable  vibrating  or impact giving area   5  Connect the earth placed on the backside of this device to terminal plate to prevent any  electric shock from leakage current or a potential difference   O6  Do not put or drop anything on this device and avoid strong impact    7  Do not disassemble or remodel this device   8  If this unit has not been used for a long time  use this after confirming by an expert if all  function and appearance are in good condition    9  Do not splash any fluid on this device or insert any foreign substances   10  In case of inserting foreign substances or exposing to particular environment  this device  must be examined by an expert before use   11  Use the power cable  plug  and fuse that are offered by our company   At this time  confirm the covering of cable  the state of plug connection  and other check  points to the things below   e RS 232C cable     USB port   Adapter  When pulling out the power cable  turn off the power switch first and then pull the plug out   Storage ambient  Temperature  10   60   C  Humidity lower than 95    non condensing   Operation ambient  Temperat
4.  PEDAL RE BE 24  6  CONNECTING HAND GWITCH  ce cccecceeeeceeeeeeeeeeeeeeese cesses es eeeeeeseeeseeseesaeeeeesaay 24    DYSTEMN ee EE 25  1  Entering SYSTEM SETUP               cccccccsecccccesseesanseeeceeeneececeseeseauseeecasseeeassesesaneeesseseeesoas 25  E ET D 25  3  Entering MENU    view    25  4  How to escape from SYSTEM GEIUR iiic 25  5  How to move into SYSTEM GEITUR  en 25  S  J 1 EE 26     MERS ORE VE 31  1  CAUTIONS FOR MEAGUREMENT iien nenen 31  2  MEA SURE MEN Too E eee eee ene eee eee 32  4  The measuring method if ID iS ON    eee e ce ceee cece eeeceeeeeeeeceeeeeceaaeseeueseeeseeeessanseeeeaeenes 37  5  The measuring method by using    Three measurement MOIE  38      MAINTENANCE 2225340900009642 20 944 Nee E oat arnt 050 ETE te G ASO AR LANE MR RA aE EEN EE Eei 40     ERROR  amp  REPAIR E 41      AFTER Eege 42    1  AFTER S    ERVI Erdre A t eee eee aaa    2  PACKING AND TRANSPORT WEE  ee e E IR WE    e WARRANTY    42  42  43  44    E INTRODUCTION    We highly appreciate that you chose our company s product    You are kindly requested to be familiar with these directions before using this product and  always keep it together with the product  In case you are not sure about any directions or  problems arising while using the product  please contact our service center    We will provide you with detailed instructions     1  INDICATIONS    Indications for use   The EX PLUS 1300 is intended to measure systolic and diastolic blood pressure and pulse rate  in adul
5.  Press    CLOSE    to get back to SYSTEM SETUP menu        CLOSE    29     lt DETECTING ARM gt     Select operating of detecting arm sensor When  pressurized       Pre set  ON    Select  Detecting arm  and enter the menu    ON   Auto detect arm     Press    SET    button on touch pad to save it     Press    CLOSE    to get back to SYSTEM SETUP menu      lt SELECT LANGUAGE gt   Select the language on device       Pre set  ENGLISH    Select    Select language    and enter the menu   Choose the language     Press    SET    button on touch pad to save it     Press    CLOSE    to get back to SYSTEM SETUP menu      lt QR DOMAIN gt     Input the server address of QR code      Pre set  m jawon com     Input server address for want to use      Storage by pressing    SAVE    button      If you dont want input  press    Initial View button and  return initial view      30       DETECTING ARM       CLOSE  AUTOMATIC BLOOD PRESSURE MONITOR    SELECT LANGUAGE    KOREA  ENGLISH  JAPAN    Jgd   inital View e     E MEASUREMENT    1  CAUTIONS FOR MEASUREMENT    D  D  3   ER            9    Take off heavy sweater or shirts    Have a rest before the measurement    Do not move or talk during the measurement    Do not measure in following positions  standing  half sitting  leg crossed   Put your arm properly into the arm support     When you roll up the sleeves  make sure that the rolled up sleeves are not pressing the  upper arms     When the pulse is running too week and hard to be heard from ste
6.  UT    30   dip in UT   for 25 cycles    lt 5   UT     gt 95   dip in UT   for 5 sec    Magnetic field of  commercial  frequency   50 60Hz    IEC 61000 4 8      gt 95   dip in  UT    for 0 5 cycle   40   UT    60   dip in UT   for 5 cycles   70   UT    30   dip in UT   for 25 cycles    lt 5   UT     gt 95   dip in  UT    for 5 sec    of a typical commercial or hospital  environment  If the user of the EX  PLUS 1300 requires continued  operation during power mains  interruptions  it is recommended that  the EX PLUS 1300 be powered from  an uninterruptible power supply or a  battery     Power frequency magnetic fields  should be at levels characteristic of  a typical location in a typical  commercial or hospital environment     UT is the a c  mains voltage prior to application of the test level        3  Guidance and manufacturer s declaration   electromagnetic immunity 2  The EX PLUS 1300 is intended for use in the electromagnetic environment specified below  The  customer or the user of the EX PLUS 1300 should assure that it is used in such an environment     ramunitiytest E 606014 e LEES Fleciomegne  environment   level guidance  RF    Conducted RF  IEC 61000 4 6    3 Vrms  150 kHz to 80 MHz    Radiated RF  IEC 61000 4 3    3 V m  80 MHz to 2 5 GHz    13    iC   Portable and mobile  communications equipment should be  used no closer to any part of the EX  PLUS 1300  including cables  than the  recommended separation distance  calculated from the   equation  applicable to the
7.  Where the atmospheric pressure falls below 7O0kPa  700mbar  or exceeds 106kPa   1060mbar      Where the humidity is over 95  non condensing     Where the unit is exposed to spray or splashing water       Where the unit is exposed to dust     Where the unit is exposed to water vapor     Where the unit is exposed to salty atmosphere     Where the unit is exposed to explosive gas     Where the unit is exposed to excessive shocks or vibrations     Where the angle of inclination of mounting surface exceeds 10 degrees     Where the unit is exposed to direct sunlight     N 8  This equipment has been tested and found to comply with the limits for medical devices  to the IEC 60601 1 2 2007  These limits are designed to provide reasonable protection  against harmful interference in a typical medical installation  This equipment  generates uses and can radiate radio frequency energy and  if not installed and used in  accordance with the instructions  may cause harmful interference to other devices in  the vicinity  However  there is no guarantee that interference will not occur in a  particular installation  If this equipment does cause harmful interference to other  devices  which can be determined by turning the equipment off and on  the user is  encouraged to try to correct the interference by one or more of the following measures      Reorient or relocate the receiving device      Increase the separation between the equipment      Connect the equipment into an outlet on a circuit d
8.  bpm mmHg UI  l bpm    The difference of your     inter arm pressure is         When the difference between the two arms is more than 20mmHg for systolic  blood pressure or 10mmHg for diastolic blood pressure  the difference is  indicated in red        7  The result is printed out automatically     At the system setup  there s PRINT ON OFF  If the print is off  the result sheet  is not printed out        8  Pull your arm out from the cuff     Press the  QR code on the device screen to enlarge the  icon and scan the    QR code    with a mobile device to save  the data in the server   After scanning the    QR code     you can check the    Scan the left QR code    measurement data with your mobile device and manage wih sa pore    It on the  website        it at will from the server in which it is saved     33    9  When you press D button on the result screen  L H RI  it indicates which category your blood pressure is Staga 2 Hypertension  classified into  Gray represents the blood pressure of  your left arm     Stage 1 Hypertension    Prehypertension    Normal       10  When you want to measure one more time  press D button   The whole procedure restarts again     2  Measuring method by using a hand switch    Measure blood pressure on both arms simultaneously       Plug the cable and turn on the power    2  As the power is on  LCD displays the Standby mode     Put both arms into the cuff      An assistant stands next to the measurer and press    START        Pressurization star
9.  frequency of the  transmitter     Recommended separation distance    d  1 2V P    d  1 2V P 80 MHz to 900 MHz       d  2  P 900 MHz to 2 5 GHz    where P is the maximum output power  rating of the transmitter in watts  W   according to the transmitter  manufacturer and d is the  recommended separation distance in  meters  m      Field strengths from fixed RF  transmitters  as determined by an  electromagnetic site survey     should  be less than the compliance level in  each frequency range       Interference may occur in the vicinity  of equipment marked with the  following symbol     Dol    1  At 80 MHz and 900 MHz  the higher frequency range applies     2 These guidelines may not apply in all situations  Electromagnetic   propagation is affected by absorption and reflection from structures  objects   and people    Field strengths from fixed transmitters  such as base stations for radio   cellular cordless  telephones and land mobile radios  amateur radio  AM  and FM radio broadcast and TV broadcast cannot be predicted theoretically  with accuracy  To assess the electromagnetic environment due to fixed RF  transmitters  an electromagnetic site survey should be considered  If the  measured field strength in the location in which the EX PLUS 1300 is used  exceeds the applicable RF compliance level above  the EX PLUS 1300  should be observed to verify normal operation  If abnormal performance is  observed  additional measures may be necessary  such as reorienting or  relocati
10.  to SYSTEM SETUP menu        AUTOMATIC BLOOD PRESSURE MONITOR    20     lt  PRINT POSITION  gt     It adjusts the printing position in the direction of U D  up     down  and L R  left right  to fit to the pre formatted result   sheet      Select    PRINT POSITION    and enter the menu      Pre set  00 for U D and 00 for L R     Range  99 for U D and 99 for L R     Choose U D by pressing    V  A    on touch pad  Choose  L R by pressing     gt       on touch pad        CLOSE      Pressing P button moves print position down or right     Pressing d button moves print position up or left     Every single press moves print position by about 0 2 mm       Press    SET    button on touch pad to save it     Press    CLOSE    to get back to SYSTEM SETUP menu     L R  left right     is moving to the left and   is moving to the right     U D  up down     is moving up and   is moving down         lt PROGRAM INFORMATION  gt   It is for check the program information of this device       Select    PROGRAM INFORMATION    and enter the menu     Check the version and information of this program     Press    CLOSE    to get back to SYSTEM SETUP menu      lt MEASUREMENT VIEW gt     It sets the result screen type  RESULT VIEW      Pre set   GENERAL TYPE     Select    RESULT VIEW    and enter the menu       Select the result type  M  GENERAL TYPE    GENERAL TYPE  Basal type      DIAGRAM TYPE    DIAGRAM TYPE  The Blood pressure displayed by gage     Press    SET    button on touch pad to save it     
11. able  B type      PEDAL BUTTON PORT  It connects the PEDAL switch   HAND BUTTON PORT  It connects the HAND switch   a1  INFORMATION BOARD FIXER  It fixes the Information Board     18    REAR PART    M INFORMATION  BOARD FIXER    USB PORT   A TYPE     Fa      HAND BUTTON  PORT    R CAL 6  EARTH    9  PEDAL BUTTON   4  L CAL BER    A POWER    USB PORT    POWER INPUT        B TYPE      COMMUNICATION  PORT    The operator should not touch both USB port and the patient s body  simultaneously        19    3  ACCESSORIES    Adapter Power cable User Manual    R MEASUREMENT    D Dei help fror an expert wien you measure bleed pressure    H Leave your arm 2e the seraor part of the arm sapporter       Sea   3   n case you feet hut the pesters in we high and want to come he reanuremari prem ore KN MWY  at bathers Step arion Poda baten Hand in Tearterat Wage botor SZ H SS  EI It s recommended by WHO that patient should consult with a doctor re  when the difference between the two arms i more than Z0mmitg for    systolic bloed pressure ce 1Ommitg for diastolic blood pressere  E 7 A    the range mentioned above may be in the ris   of moe    disease    L    Poson   _ a Kafe   OSZE Meeta w d      i     eeng      ees       Z                 4  OPTIONS    A4 RESULT SHEET    EX PLUS 1300                            20    be INSTALLATION    1  CONNECTING ADAPTER  Connect the adaptor to the power input on the back of the device   Turn on the POWER switch     In order to avoid the risk of electrical 
12. directly    x Our company or local distributor will make available on request circuit diagrams  component  part list  descriptions  calibration or other information which will assist your appropriately  qualified technical personnel to repair those parts of unit which are designated by our  company as repairable    x Calibration interval for this device is 2 years    To ensure the device   s proper and sage operation  please contact our company or distributor  periodically for calibration     How to contact our company  Write us at     JAWON MEDICAL CO   LTD   29  Gongdan 4 ro  Jillyang eup  Gyeongsan si  Gyeongsangbuk do  Korea  TEL  82 53 856 0993   FAX  82 53 856 0995    You can also contact the following representative or your local distributor     2  PACKING AND TRANSPORT   Our company wraps this device up with the most suitable method to protect it from any impact or  damage during shipping and transporting  This device can be damaged during delivery if it is  packed with other ways except the one our company uses  Please handle this device carefully  without any impact in packing and delivering it    If this device needs to be transported wrap this device up again and transport it as follows     1  Turn off the power of this device     2  If peripherals are connected  turn off the power of them and then disconnect each device     3  Disassemble this device in reverse order to assembly     4  Wrap up this device with original packing materials        Transport it carefu
13. eight height m  x22       Very thin thin normal overweight obese    section  80  90    90  110    110  120   gt 120     5  The measuring method by    THREE MEASUREMENT MODE      It measures blood pressure 3 times in a row  The average blood pressure from the three  measurements indicates the more accurate data   It is possible to measure either one arm or both arms   System setup   gt  Measurement Mode   gt Three measurement mode  You can set the measuring interval   15sec  30sec  45sec  60sec      1  Measurement  Put both arms or one arm to the cuff and press    START    button   Pressurization starts automatically from the cuff and LCD indicates the current status      gt  DATA   2012 11 13     rower iziii measure FS  mme EJ orcove 20122 2 32       38    When the first measurement is completed  device stops until the interval time the user sets     POWER       MEasuRE O Pe AN     PRFESSI IRF    Three measurement mode   Measure 3times sequentially        When the second measurement is completed  device stops until the interval time the user  sets       DATE   2011 04 13  Finish LOE  2    16 34 24    DESST IRE    Three measurement mode   Measure 3times sequentially        It performs the measurements three times in a row     Result  After the third measurements  the average blood pressure is indicates as follows         POWER MEASURE PRINTER OR arcope DATA   2012 11 13    TIME  16 34 24    mmHg mmHg    MEANPRESS DC PULSE Win MEANPRESS  e fm PULSE  mmHg Lt pa     bpm   Lt mmHg
14. eously at once     Three Measurement mode  Measure blood pressure on   both arms 3 times in a row  The average blood pressure       is displayed as a result     CLOSE    Pre set  Simple measurement mode      Select the    MEASUREMENT MODE and enter the menu      Select either    SIMPLE MEASUREMENT MODE    or    Three Measurement MODE         Set the interval time when    Three Measurement MODE    is selected  The interval time is the  break time between three times of measurements  Select the time from 15sec  30sec  45sec  and 60sec  Press    SET    to save the change      Press    CLOSE    to get back to SYSTEM SETUP menu         lt  CONVERTING TO AUTO STANDBY MODE  gt  Automatic return to initialized screen    Finish the measuring and over steady time it is returned to initialized screen   After the measurement  the result screen automatically Ge   returns to the initial screen  riled screen from resu screen automaticaly    gt  e    Pre set  ON OFF  It is displayed the result screen continue until next measuring     Select the    Converting to auto standby mode    and enter  the menu  LA  Za ON C  DFF SET      Select either    ON    or    OFF            AUTOMATIC BLOOD PRESSURE MONITOR      Press    SET    to save the change     Press    CLOSE    to get back to SYSTEM SETUP menu     E     lt  LOGO  gt     i gi  Set LOGO    positi on  It can be printed on upper or lower part of thermal paper by using FT mate which    blood pressure program manager  Please make inquiry about 
15. he information  contained in the product for diagnosis or treatment of health problem or  prescription of medication by yourself    If you have or suspect that you have a medical problem  consult with your    physician promptly    Defective unit or accessories must be packed in the replacement cartons to  be shipped off from you to our company    Shipping and insurance costs for return of defective unit must be prepaid by  the users        5  SAFETY SYMBOLS AND INFORMATION  The International Electrotechnical Commission  IEC  has established a set of symbols for  medical electrical equipment which classifies a connection or warning of any potential hazard   The classifications and symbols are shown below  Save these instructions for your safety     Degree of protection against electric shock  TYPE BF       Please observe operating instructions    General warning sign    General prohibition sign    General mandatory action sign    Caution    Waste Electrical and Electronic Equipment  WEEE    The device could be sent back to the manufacturer for recycling or  proper disposal after their Useful lives  Alternatively the device shall be  disposed in accordance with national laws after their useful lives      OFF   only for a part of equipment         ON   only for a part of equipment     This symbol is Used inside system   Identifies the point where the safety ground of the system is fastened to  the chassis     Do not open  This is for factory only     Alternating current    Date 
16. ied person with our  company or our distributors   bg 2  This device is specified as Class 1 type BF unit under the standard of EN 60601   1 2006  Safety of Medical Electrical Equipment    Therefore  patients must not touch or handle inner side of the system at any time      3  Do not modify the unit  If any modification is needed  ask our company or its authorized  dealer for service   4  The unit has previously been adjusted in the factory for optimum performance   D Do not attempt to adjust switches or any other things except those specified in this  manual for operation      5  If you have experienced any trouble with the unit  switch it off immediately  and contact  our company or its authorized dealer for assistance   Ab  6  If you plan to connect any device of other manufacturers electrically or mechanically to  the unit  contact our company or its authorized dealer for instructions before doing so   When you connect computer or other system to the unit  RS 232C   the attached  systems should be those certified by IEC 950 or equivalent standards for data  processing equipment   Configurations shall comply with the system standard EN 60601 1 2006   Everybody who connects additional equipment to the signal input part or signal output  part configures a medical system standard EN 60601 1 2006   If in doubt  consult the A S department of local distributor    N 7  Avoid the following environments for storage     Where the ambient temperature falls  20 C or exceeds 60  C    
17. ifferent from that to which the other  device s  are connected      Consult the manufacturer or field service technician for help       9  Do not to touch signal input  signal output or other connectors  and the patient   simultaneously     N 10  a statement that MEDICAL ELECTRICAL EQUIPMENT needs special precautions  regarding EMC and needs to be installed and put into service according to the EMC  information provided in the ACCOMPANYING DOCUMENTS      11  a statement that portable and mobile RF communications equipment can affect  MEDICAL ELECTRICAL EQUIPMENT       12  Please consult a physician or a trained health professional for interpretation of   measurement results    A 13  No phthalates are used for this product and its container    A 14  The cuff is not made with natural rubber latex    1  Measurements may be impaired if this device is used near televisions   microwave ovens  X ray equipment or other devices with strong electrical  fields  To prevent such interference  use the meter at a sufficient distance    from such devices or turn them off    2  Incorrect operation or failure of user to maintain the unit spares the  manufacturer or his agent of the responsibility for system   s non compliance  with specifications or responsibility for any damage or injury        This manual is made for informational purpose and this manual and product  are not meant to be a substitute for the advice provided by your own  physician or other medical problem  You should not use t
18. lly in order not to give a shock to this device     42    be SPECIFICATION  Measuring Method Oscillometric  Display mode Color LCD Touch display  7inch     Measuring one arm   Systolic Blood Pressure  Diastolic Blood Pressure  Mean Blood Pressure  Pulse  Pulse Wave Pattern   Evaluation of Blood Pressure   Result Contents    Measuring both arms    Systolic Blood Pressure  Diastolic Blood Pressure    Mean Blood Pressure  Pulse  Pulse Wave Pattern   Evaluation of Blood Pressure  Inter arm Pressure Difference     Pressure 30 300mmHg  Pulse 30 200beats minute  Pressure  3mmHg  Pulse With in  3     High speed thermal printer  Input AC 100 240V  50 60Hz  Output DC 12V  5A ADAPTER  Temperature  10 60 C  Humidity Less than 95   Data transmission RS 232C  USB  Sere          43    Ed WARRANTY    WARRANTY    Automatic Blood Pressure Monitor Warranty period      Model     EE 1year  main unit only   Serial NO     O      Date of purchase    Customer Name   Address   Name   Dealer    Address             When you receive this warranty  make sure that the name of the dealer and  the month  day and year of purchase are all completed       This warranty will not be reissued  please keep it in a safe place        44    Periodic Check List    Management No     kene Enclosure No scratch  crack  Pass Fail  deformation and rust    2   Labels and panels   No peeling and dust  Passa    alkes   Nodamege rop    vom   No scratch and damage  rem     2   User manual     Keptinproperplace  Passa     mecra L 
19. ng the  instructions mentioned above     Save the change with    SET    button     Press    CLOSE    to get back to SYSTEM SETUP menu        Press    CLOSE    to stop setting the date and time    For Date Time  all settings should be done at once  hour  minute  month  day  and year    If you press    CLOSE    and stop the setting  the time and date returns to the old    setting    The function of Time and Calendar keeps running even after the power is off   Calendar program is set for 100 years  and it adjusts automatically even at a  leap year        26     lt   gt     Set the volume of key sound on the top monitor       Pre set  2    Select    VOLUME    and enter the menu       Set the volume by pressing             V    button     Press    SET    to save the change  TT      Press    CLOSE    to get back to SYSTEM SETUP menu  l IIL i l   l     lt THERMAL PRINT gt   Choose the print option and set the result sheet form     Pre set  ON      Select    THERMAL PRINT    and enter the menu      ON  Result sheet is printed after the measurement   OFF  Result sheet is not printed after the measurement      Press    RESULT SHEET FORM    and enter the menu      Select the print form and press    SET    button      Press    CLOSE    to get back to    THERMAL PRINT      menu      Press    CLOSE    to get back to SYSTEM SETUP menu     CLOSE     lt  MEASUREMENT MODE gt   Select the measurement mode      Simple Measurement mode  Measure blood pressure    ement mode   on both arms simultan
20. ng the EX PLUS 1300    P Over the frequency range 150 kHz to 80 MHz  field strengths should be less   than 3 V m     a       14    4  Recommended separation distances between portable and mobile RF communications  equipment and the EX PLUS 1300  The EX PLUS 1300 is intended for use in an electromagnetic environment in which radiated RF  disturbances are controlled  The customer or the user of the EX PLUS 1300 can help prevent  electromagnetic interference by maintaining a minimum distance between portable and mobile  RF communications equipment  transmitters  and the EX PLUS 1300 as recommended below   according to the maximum output power of the communications equipment    Rated maximum   Separation distance according to frequency of transmitter   output power m    of transmitter 150 kHz to 80 MHz 80 MHz to 900 MHz 900 MHz to 2 5 GHz  W   d  1 2V P d  1 2  P d  1 2v P    023   0 73   23   73  12 12       For transmitters rated at a maximum output power not listed above  the recommended  separation distance d in meters  m  can be estimated using the equation applicable to the  frequency of the transmitter  where P is the maximum output power rating of the transmitter in  watts  W  according to the transmitter manufacturer     1  At 80 MHz and 900 MHz  the separation distance for the higher frequency  range applies    2  These guidelines may not apply in all situations  Electromagnetic  propagation is affected by absorption and reflection from structures  objects  and people  
21. ntering  MENU  view  M DATE TYPE O LOGO    Select the menu on the touch screen  D DATE TIME   PERSONAL DATA INPUT  O VOLUM O VOICE    O THERMAL PRINT O PRINT   O MEASUREMENT MODE  1 PRINT POSITION    1 CONVERTING TO AUTO O PROGRAM INFORMATION  STANDBY MODE O RESULT VIEW    4  How to escape from SYSTEM SETUP  Press the blue icon on the upper right corner  CN      The initial screen appears  M DATE TYPE   O LOGO  O DATE TIME O PERSONAL DATA INPUT    VOLUM O VOICE  O THERMAL PRINT O PRINT  O MEASUREMENT MODE O PRINT POSITION  O CONVERTING TO AUTO O PROGRAM INFORMATION  STANDBY MODE O RESULT VIEW    AUTOMATIC BLOOD PRESSURE MONITOR                ao    5  How to move into SYSTEM SETUP    After setting the system from the selected menu  press the blue icon on the upper right corner   The initial screen appears     6  Setup    DATE TYPE     lt DATE TYPE gt   Set the Date Format     Pre set  YY MM DD  Year Month Day       Select    Date type        Press    SET    to save the change     Press    CLOSE    and get back to SYSTEM SETUP DD MM YY    menu      lt  DATE TIME  gt     Set the current time and the date     Pre set  The date released from the factory     Select Date Time and enter the menu     Press    YEAR    and the color changes     Press         button to increase the number   Press    V   button to decrease the number     After setting the current year  move to Month   Set the current month by pressing         and    V    button   Set the date  hour  and minute by followi
22. nts  IEC 61000 3 2 and those directly connected to the public low voltage  Voltage power supply network that supplies buildings used for  fluctuations  domestic purposes    l S   Compliance  flicker emissions  IEC 61000 3 3       2  Guidance and manufacturer s declaration   electromagnetic immunity  The EX PLUS 1300 is intended for use in the electromagnetic environment specified below  The  customer or the user of the EX PLUS 1300 should assure that it is used in such an environment       IEC 60601 test   Compliance Electromagnetic environment   Immunity test i  level level guidance    Floors should be wood  concrete or   6kV  Contact  6kV  Contact ceramic tile  If floors are covered   8kV  Air  8kV  Air with synthetic material  the relative   humidity should be at least 30       Electrostatic  discharge ESD   IEC 61000 4 2    AO Bees  2kV  Power  Electrical   fast     supply lines Mains power quality should be that    SCH unge  1kV  of a typical commercial or hospital    transition burst  IEC 61000 4 4 Input output environment     Input output lines ines     1 kV   differential Mains power quality should be that  mode of a typical commercial or hospital   2 kV common   environment    mode    Voltage drops  Mains power quality should be that     1 kV differential  Surge mode  IEC 61000 4 5  2 kV common  mode       12    dips  and  fluctuations of  input power  supply line IEC  61000 4 11      gt 95   dip in UT   for 0 5 cycle   40   UT    60   dip in UT   for 5 cycles   70  
23. o  2   Recorder Smooth operation with no   Pass Fail  alams   mm rop      Accessories   1   Power cord Smooth operation and Pass Fail  removal    Electrical Cheek   ll     Performance Power supply Screen display upon Pass Fail  e SI  2   Display No abnormality and Pass Fail  TT SC IST      3  Printing   reen Iren     a  Measurement   Proper measurement  re    ee rem  mos  ews ao    emeng     tures    zem    onenean   mees _      Copy this sheet for use                If repair is required  write down so in the Remarks column     Daily Check List    Management No     Tem     Inspection Subjeci Judgment   Remarks    Visual Check hd  Enclosure SR Scien Crack  Pass Fail  deformation and rust    Labels and panels  Caf     Smooth operation rer      Mainframe           Accessories    AJOIN   gt   A  D   lt   GC       Keys Smooth operation Pass Fail    Maint    ainframe Been Smooth operation with no Pass Fail  abnormal sound    Smooth operation and Pass Fail    removal  Screen display   upon    EE Pass Fail      No abnormality and    2   Display flickering Pass Fail    Other Clock  General Judgment  Mode Serial No  nsialatin place   Dale of purchases      Copy this sheet for use  If repair is required  write down so in the Remarks column     Accessories   1   Power cord       2    Electrical Check    Power supply    Performance                    Service center  TEL  82   2   587   4056  82   53   856   0993          If the problems continue  call the service center  When you ask  fo
24. of manufacture  Manufacturer     Co  Non ionizing radiation                   5  seme  gt  gt     Foot switch   Start or stop the measurement    And  when your arm is oppressed due to high pressurizing or  irregular operation is done  press this button then the cuff will be  exhausted rapidly        10    Hand held switch   Start or stop the measurement    And  when your arm is oppressed due to high pressurizing or  irregular operation is done  press this button then the cuff will be  exhausted rapidly     Authorized representative in the European community        11    6  Guidance for Electromagnetic compatibility  EMC     Details about the electromagnetic compatibility  EMC  of the EX PLUS 1300 are given below   Before using the EX PLUS 1300  be sure to read and understand the following information     1  Guidance and manufacturer   s declaration   electromagnetic emissions  The EX PLUS 1300 is intended for use in the electromagnetic environment specified below  The  customer or the user of the EX PLUS 1300 should assure that it is used in such an environment       Emissions test   Compliance Electromagnetic environment     guidance    The EX PLUS 1300 uses RF energy only for its  RF emissions internal function  Therefore  its RF emissions are very  CISPR 11 low and are not likely to cause any interference in  nearby electronic equipment     RF emissions    Harmonic The EX PLUS 1300 is suitable for use in all  emissions Class A establishments  including domestic establishme
25. r service  the manufacturer s label  serial number  date of  original purchase and explanation of malfunction will be required        x For purpose of improvement  specifications and subject to change  without notice     Vi JAWON MEDICAL    www jawon com    x Our company name will be changed to JANEX MEDICAL     HEAD OFFICE  29  Gongdan 4 ro  Jillyang eup  Gyeongsan si   Gyeongsangbuk do  Korea  TEL  82 53 856 0993  FAX  82 53 856 0995    EUROPEAN REPRESENTATIVE  VITAKO Sp  z 0 0     UL  MALEJ SYRENKI 2 71 790  SZCZECIN  POLAND TEL  91  8522900   901 FAX  91 8522910       JW EX1300E 20151001 Rev 4 0    
26. sert USB cable to the communication port and connect it to the computer as shown in the  picture below                                                                                            4  CONNECTING PRINTER  Option   Connect A4 printer for the output in A4 size   Insert USB cable to the communication port and connect it to the printer as shown in the picture  below        23    5  CONNECTING PEDAL SWITCH  Connect the PEDAL SWITCH as shown in the picture           zh    a  So    UIN         FI   ei       IT  l  N    6  CONNECTING HAND SWITCH  Connect the PEDAL SWITCH as shown in the picture        24    E SYSTEM SETUP    Set the operating conditions for EX PLUS 1300     1  Entering SYSTEM SETUP    y    At the initial screen  press  Power  button on the left upper MM O G Gei  corner for 4 5seconds  Popup screen appears as shown in the           picture  Press    1  gt 2  gt 3  gt 4    to enter SYSTEM SETUP    r 7 AG  2  Menu   1  DATE TYPE SYSTEM SETUP      2  DATE TIME   3  VOLUME    DATE TYPE O Loco   4  PRINT O DATE TIME O PERSONAL DATA INPUT   5  MEASUREMENT MODE aan ait   6  Automatic return to initialized screen D MEASUREMENT MODE    PRINT POSITION   7 LOGO O CONVERTING TO AUTO O PROGRAM INFORMATION   8 PERSONAL DATA INPUT STANDBY MODE O RESULT VIEW   9       D VO   C E AUTOMATIC BLOOD PRESSURE MONITOR    10  PRINT   11  PRINT POSITION   12  PROGRAM INFORMATION  13  MEASUREMENT VIEW   14  DETECTING ARM   15  SELECT LANGUAGE   16  QR DOMAIN    SYSTEM SETUP    3  E
27. shock  connect this device only to the  power supply equipped with the protective grounding     When connecting adaptor  place the arrow mark of adaptor  connection part up and correctly stick it in the socket on the rear  of the main body    Wrong connection could be a fire hazard        sech    2  LOADING THE PRINT PAPER    Pull the Top button up  Then press the Side button  Open the upper printer cover   Put the roller into the center hole of the thermal paper    Insert the thermal paper with the roller into the holder as shown in the picture   Take the edge of the paper out    Close the cover    It automatically cuts the paper     E oO eo oO    Thermal paper doesn t need the ink    One side is smooth and another is rough  It prints the letter only on the  smooth side    Always check the remainder of the paper and replace it at any time    When the red line appears on the right corner of the paper  replace the paper   Use only the exclusive paper  57mm     Keep paper rolls in a dark and ventilated place    Avoid the dust on the paper    Do not pull the paper during printing  It may cause a jam    When the paper is not loaded correctly  the paper will not be printed out  properly and it may cause the breakdown    When the printer cover is not properly closed  the paper can be jammed    inside  When the problem occurs  check the printer cover        EN    3  CONNECTING COMMUNICATION PORT  To transmit the data  connect EX PLUS 1300 to a computer or other external devices   In
28. sure is Normal   _    ema  mn  pm  Em  Ge  ae   ch  dem  de  dort  eee dead geg emm geng  gz  raal       V Classification of the blood pressure    National High Blood Pressure Education Program  National Heart  Lung and Blood institute   NIH  JNC7  2003     36    Diastolic nmHg        100  90  80  120 140 160 Systolic   mmHg     4  Measuring method with ID  Enter SYSTEM SETUP  From    PERSONAL DATA INPUT     activate ID use     JCT INTE OAMATION    _   _               ID  WEIGHT       HEIGHT    Ki       1  Input data  ID  On the initial display  press    ID      Enter ID using the number buttons     ID Input Range is from 000000001 to 999999999     Weight  After entering ID  press    WEIGHT      Enter weight     Weight Input Range is 10 0 248 0kg     Height  After entering weight  press    HEIGHT      Enter height     Height Input Range is 80 0 238 0cm     37    D DATA   2012 11 13       KEE MEASURE   Bogle Eu QR CODE TIME     16 34 24    2  Measurement  After entering the personal data  press    START      Pressurization starts automatically from the cuff and  LCD indicates the current status        3  Result  When the measurement is completed  the user s ID   weight  height  Body Mass Index  B M I  and fatness  are displayed as shown in the picture     Your fatness is    K       4  Criterion for judging result    e Fatness  Based on the standard weight  it calculates how fat the user is in percentage      current weight standard weight  standard weight X100  100  standard w
29. the using in detail of    FT   mate to the near branch       Select    LOGO    and enter the menu  THE UPPER LOGO  tis displayed the transfered logo on upper ofthe thermal pap  r      Select either    THE UPPER LOGO    or    THE UNDER THE UNDER LOGO  It is displayed the transfered logo on under of the thermal paper   LOGO         If the user wants to delete LOGO press    DELETE         Press    SET    to save the change      Press    CLOSE    to get back to SYSTEM SETUP menu         lt  PERSONAL DATA INPUT  gt     It allows the user to input personal data  PERSONAL DATA INPUT      Select    PERSONAL DATA INPUT    and enter the menu  w ON     OFF    Select either ON or    OFF    for ID      Select either    ON    or    OFF    for HEIGHT WEIGHT    INPUT     ON WV  OFF      Press  SET  to save the change  SET    Press    CLOSE    to get back to SYSTEM SETUP menu      lt  VOICE  gt     It sets the voice guidance     Pre set  ON    Select    Voice    and enter the menu     Select either    ON    or    OFF        Press    SET    to save the change  LJ  ON 1  OFF    Press    CLOSE    to get back to SYSTEM SETUP menu     CLOSE          idee  It selects the printing mode of the A4 printer      The printer and pre formatted result sheet is  optional      Select    PRINT    and enter the menu     Pre set  AUTO    AUTO  It is printed automatically   Manual  It is printed manually   Off  It is not printed     Press    SET    button on touch pad to save it     Press    CLOSE    to get back
30. thoscope  EX PLUS 1300  can not make a measurement     During the measurement  straighten your back and relax your arm     This device is only for adult        31    2  MEASUREMENT   There are 3 ways to measure blood pressure     1  Measuring method by using a foot switch    Measure blood pressure on both arms simultaneously      1  Plug the cable and turn on the power     2  As the power is on  the message appears as shown in  the picture     This machines evaluates the systolic  0  01010 ol CH Te Delles  010 0 I ol  T  F  mean blood pressure  pulse and inter arm    pressure difference     3  Put both arms into the cuff     4  There   s Pedal Switch placed at the bottom of the cart   Step down the pedal to start the measurement           Place your arm on the arm supporter with  the palm facing up    Adjust the height of the chair so that the arm  stays on the same height of the heart     When the arm is placed lower than the  heart  blood pressure will be higher than  actual value  and vice versa        32    DATA   2012 11 13    5  Pressurization starts automatically from the cuff KEE OR E A  and LCD indicates the current status        6  When the measurement is completed  LCD screen KESE AM   ETV A GR E E ATENA  displays the results  uq     er ec ears me  At the same time  voice guidance speaks out      Measurement completed  pull your arm out please   Thank you       The cuff deflates and turns to the initial state     mmHg mmHg    E PULSE MEANPRESS   Zi PULSE  mmHg Lt a
31. tion  The air will be rapidly exhausted from the cuff    a1  HAND SWITCH REST  When the cable switch is used  switch rest can be placed in the right side of device     2  HAND SWITCH  When the user measures blood pressure on both arms  push the hand switch to START and  STOP the device     3  PEDAL SWITCH  It is placed under the cart  It works as    START STOP        16    FRONT PART    USER S GUIDE FOR MEASUREMENT    Ra Ahora  qat help    om an cepert when you measure double arm blood  pressure        BA teava your arm on tha sansar part of the arm supporter       INFORMATION BOARD          ARM BEST  A    Ze                   T         LCD DISPLAY    HUMAN SENSOR    inter arm essure is    FF  G   GPRINTER COVER    EMERGENCY  re         STOP BUTTON      EP       AUTOMATIC CUTTER        START STOP BUTTON        2  HAND SWITCH    i  HAND SWITCH REST  3  PEDAL SWITCH    The side part of cart The under part of cart     E     2  REAR PART       POWER  It turns the power on and off     POWER INPUT  It connects the adapter     R CAL  It is only for the inspection  Do not open it     L CAL  It is only for the inspection  Do not open it     EARTH  POTENTIAL EQUALIZATION TERMINAL   For safety  the device should be placed on the flat surface     COMMUNICATION PORTS  RS 232C   It connects the main to other equipments with cable  RS 232C  to transfer the data      USB PORT  A TYPE   It connects the main body with USB cable  A type    USB PORT  B TYPE   It connects the main body with USB c
32. tions by reading this manual  before operating your unit   Particular attention must be paid to all warnings  cautions and notes incorporated herein   The following conventions are used throughout the manual to denote information of special  emphasis     3  CLASSIFICATION AND COMPLIANCE  1  This device is classified as     Class 1 type BF against electric shock    Ordinary equipment without protection against ingress of water    Equipment not suitable for use in presence of a flammable anesthetic mixture by standard of  EN 60601 1 2006 Safety of Medical Electrical Equipment   2  This device is complied with Class A for Noise Emission  Level B for Noise immunity  by  standard of IEC 60601 1 2 2007 Electromagnetic Compatibility Requirements    3  This device is complies with the EN 1060 1  1995 A2 2009 Non invasive  Sphygmomanometers general requirements as well as EN 1060 3  1997 A2 2009  supplementary requirements for electro mechanical blood pressure measuring systems     4  SAFETY PRECAUTIONS  This device is designed and manufactured with consideration of safety of the operator and  subject and also to the reliability of the unit   The following precautions must be observed for additional safety   This device is designed and manufactured with consideration of safety of the operator and  subject and also to the reliability of the unit   The following precautions must be observed for additional safety   Ah 1  The unit must be operated only by  or under supervision of a qualif
33. ts automatically from the cuff and LCD indicates  the current status        The next step is same as the instruction                       00 from   1  Measuring method by using a foot switch           3  Measuring method by one arm    Measure blood pressure on one arm         Plug the cable and turn on the power    2  As the power is on  LCD displays the Standby mode        Put right or left arm to the cuff        For measuring right arm  put right arm into the cuff on the right side     For measuring left arm  put left arm into the cuff on the left side        34      Press the start rower weem GS  mme F orcooe 22    50  8  Pressurization starts automatically from the cuff  and LCD indicates the current status           When the measurement is completed  LCD screen LTE OR ECH  displays the results   At the same time  voice guidance speaks out      Measurement completed  pull your arm out please   Thank you      The cuff deflates and turns to the initial state        Sie    3  Result sheet    V Result sheet format  both arms     2011 09 15 15 07  Lt    V Result sheet format  one arm     2011 09 15 15 07  Lt    He oe SYSTOLIC 130 mts    100 mmt DIASTOLIC           85 mmo  MEANPRESS       80 mmHg  9 3 mmHg    2011 09 15 15 07    130 mmHg             1 00 mmHg  120 80 mmHg   bem    Inter arm Difference    SYSTOLIC   10  mmuo  DIASTOLIC   __5  mmHg    Refer to the results and consult to  physician about specific medical  information     PULSE WAVE PATTE RM        Your Blood Pres
34. ts with an arm circumference of 22 33 cm    The EX PLUS 1300 is intended for use in physicians    offices  hospitals  clinics and other medical  facilities where non invasive blood pressure is performed on patients and invasive measurement  is contraindicated    The blood pressure can be measured using 2 different methods as follow    1  one measurement with left or right arm   2  two measurements with both arms BP simultaneous dual arm measurement     Contraindications for use   As with any non invasive measurement device  there are clinical conditions which can influence  the accuracy of the results  Also  the subject   s position  physiological condition and other  environmental factors can affect the measurement calculation     The EX PLUS 1300 Non Invasive Blood Pressure Monitor should not be used with patients who  have the following conditions    1  Patients with a known arrhythmia    2  Patients with insufficient peripheral circulation  acute cases of low blood pressure or low  temperature      Patients who use a pacemaker      Patients experiencing a seizure      Children younger than 18 years old      Patients who should not have blood pressure measurements taken from their arms      Patients with an artificial heart      Patients whose artery cannot be found by palpation     CON    01          2  WORD DEFINITIONS  To ensure safe operation and long term performance stability  it is essential that you fully  understand the functions  operating and maintenance instruc
35. ure 10   40   C  Humidity 30   75    non condensing   Do not store or use this device under 70 kPa  700 mbar  or over 106 kPa  1060 mbar  of  atmospheric pressure   16  Cleaning  amp  Disinfection   1  Cleaning  When cleaning  use a soft cloth but do not use volatile solvent like benzene and  alcohol or a wet cloth  Wipe out minute dust once per 2   3 days with a dry cloth    2  Disinfection  Spray alcoholic water of glutaraldehyde disinfect solution  Then  wipe the  enclosure with a soft lint   17  Refer to    SAFETY PRECAUTIONS        12  13  14    O15    N      SS S    ON     40    E ERROR  amp  REPAIR    ERROR PRESSURE   pressure is high with the   When the message is    C SIE DOSE repeated  Call for    maintenance service     ERROR CUFF pressure is low as air leaks When the message is    repeated  Call for  maintenance service   ERROR MEASURE subject moves or speaks     Don   t move or speak   EE   When the message is  repeated  call for    maintenance service           41    be AFTER SERVICE    1  AFTER SERVICE  T If there is any problem with the unit  please follow the steps below    x Contact our company   s Overseas Service Department immediately    After gathering the model name  Serial Number  date of purchase and description of the  problem  contact our company with information shown below    x Try to solve the problem over the phone with the personnel of local service department    If the problem cannot be solved over the phone  just return to service department 
    