Park XE-70 User Manual

image

Contents

1. AFM tip Login To Badger Enable Device Hardware Setup 1 Unplug head and remove it by pulling flaps towards you and sliding head out to the right Never handle head from bottom to ensure safety of the AFM tip 2 Mount sample on stage 3 Mount a new tip if one is not already mounted If using your own tips CAREFULLY remove the existing tip and place it in the good tips box Ensure new tip is rested in place and that bearings fit into tip slots 4 Replace head and push flaps away from you until slightly tight This secures the head position Plug head in 5 Check that SPM Controller and Monitors are On should already be on 6 Turn on Light Bank and Laser via Laser Switch LEDs on head indicate general location of laser on photodetector 7 Open uEye Demo click on Open USB Camera icon in top left corner Drag window to right hand monitor Aim Laser on End of Cantilever Adjust Mirror 1 yl sex MM pp y ama z Find and focus on cantilever If you don t see the cantilever right away first check that the optical microscope is focused at the proper height to see the cantilever and check that enough light is coming from the light bank If you still don t see the cantilever you will need to adjust the position of the microscope knobs located above head on microscope column Look at where the light from the microscope is shining on the tip assembly as this will help you to center the mi
2. in the frequency sweep window 7 Drive Adjust the Drive by a few tenths of a percent at a time to see if your image quality improves This will increase the force with which the tip oscillates Only adjust this setting after exhausting the other options as increasing the drive can deteriorate the tip life Getting a good image can take time It is a matter of iterating between adjusting the gain set point and drive If you are unable to get a good image there are three additional options 1 Hit the NCM ASetup button at the bottom of the scan parameters window This will run an automatic routine to try and find a good set of scan parameters The frequency window will open automatically Hit OK to close this window 2 If you still don t see your surface retract the tip from the sample increase the initial drive and re approach the sample 3 If you still don t see your surface it may be that the tip you are using is damaged Raise the optical microscope then raise the Z stage well away from your surface Remove the head Put the tip in the Bad Probes box and select a new probe from the Good Probes box Collect Image 1 Click on Scan Parameters icon to select image resolution 128x128 256x256 etc la You may need to adjust the scan parameters slightly after making a change 2 Click Start button under Image label 3 Once complete double click on the color bars next to the images This will fl
3. the surface Be absolutely sure you have moved in the correct direction 6 Use the Z stage control window to lower tip until it comes into focus 7 Refocus the microscope on sample surface 8 Focus the microscope away just until sample is blurry this is 100 um above the surface 9 CAUTIOUSLY lower tip until in focus Be careful not to lower the tip past the focal plane as you may crash the tip into the sample which could damage it 10 When sample is in focus image of tip should be slightly blurry 11 Check laser position to ensure A B and C D are still less than 0 3 V 12 Shut acoustic enclosure 13 Click Approach button in Z stage control When the approach stops the light in the Z stage Control window will stop blinking Ensure that the Z scanner bar is half green This indicates that the system is tracking the surface Z scanner troubleshooting If the Z scanner bar is all white the system did not find the surface and will actually move UP not down if the Approach button is pressed 1 Raise the Z stage a small amount just until the tip becomes blurry in the optical microscope and the topography trace goes flat 2 Z scanner will turn green 2a If the Z scanner bar does not turn green after you have retracted the tip 4 open the frequency sweep window icon at right and hit the Refresh button Loft iJ 3 Increase the Drive by 5 e g 10 to 15 not 10 to 10 5 4 Approach again Re
4. you find a local maximum of A B voltage then turning the other knob until you find the local maximum again Repeat as needed until the sum is at the appropriate value If after several tries you still can t get the sum high enough remove the head and check mirror 2 s position visually it should be roughly parallel to the base Adjust mirror 2 knobs as needed and then replace head 4 Adjust Mirror 2 knobs to position laser spot on center of photodetector reducing the A B and C D voltages to less than 0 3V The XEP software will indicate the position of the laser as ared dot once it is reasonably near the center of the photodetector Option A Adjust the knobs while watching the laser position in the software and turning in the correct direction to move toward the center of the diagram minimize the A B and C D voltages Option B Use the diagram below as a guide for which knob to turn and in which direction depending upon where the laser is shining Photodetector L1 L2 L2 T1 L2 Mirror 2 Knobs airs k a k h f E a L1 T1 T tighten L loosen T2 L1 T2 T2 T1 Mode Scanner and Frequency Selection 1 Turn off Head in software 2 Click on Parts Select 3 Select proper mode of operation and scanner ranges Head Mode NC AFM ACTA tips should only be used for NC AFM non contact mode XY Voltage Mode High SOum x 50um or Low Sumx5um Z Voltage Mode High 12 um or Low
5. 1 7 um Z Scanner Range Set to 1 00 Cantilever General 4 Click OK 5 Turn Head back on in software Frequency Sweep Window will open 6 The software will automatically detect the cantilever s resonance peak and select a driving frequency just above that peak If a very weak peak appears peak is on the order of the noise or if there is strong non ideal behavior like a double peak replace the tip An additional indication of a bad cantilever is a Drive at near 100 This is a rare occurrence under normal operation 7 Click OK Frequency Sweep Frequency Sweep Report Start Freq Hz Start Freq Hz 326 707 313 728 End Freq Hz End Freq Hz 331 707 318 528 Sel Freq Hz Sel Freq Hz we 323 207 316 093 Set Point Set Point 13 7 nm 1355 1 nm Control Contro Drive Drive 100 32 3 Zoom Zoom C Show Phase C Show Phase L Show Fitting L Show Fitting Amplitude Amplitude O Feedback O Feedback Bad low amplitude 100 drive Good Initial Surface Approach 1 Ensure sample is under the tip by adjusting the x y stage controls 2 Using Z stage control window approach surface to within a few mm Only go as far as you feel comfortable that the tip is still well away from surface 3 Adjust microscope to focus on tip Then focus on sample 4 Move sample to desired area using x y stage controls 5 Focus the microscope away from sample by turning fine adjust knob turn this is 1 mm above
6. Park XE 70 AFM User Manual Ed Fei and Ryan Brock Updated 10 30 13 NOTE This document is intended as a quick reference for basic operation of the Park XE 70 and is by no means a comprehensive manual It is strongly suggested that users look through the Park documentation The better you understand the instrument the more effectively you will be able to use it Basic AFM Principles When the tip is suspended in air the laser will hit the center of the quadrant photodetector If the tip encounters a feature on the surface the tip deflects and the laser will move on the quadrant photodetector This is relayed to a motor controlling the tip which adjusts the tip s vertical position until the laser is aimed in the center of the quadrant photodetector once again Effectively this keeps the tip at a constant distance from the sample surface By recording the position of the motor height we can now know the topography of the sample surface The tips provided in this lab are intended for non contact AFM which is similar in principle to what is described above Rather than relying on physical contact with the surface the cantilever tip is oscillated at a given frequency above the sample surface and interactive forces from the surface lead to changes in the amplitude phase of this oscillation Setting up the AFM is simple adjust the mirrors until the light path follows the diagram below Quadrant photodetector ft Mirror 2
7. atten the image You can send the image to the Scan Area Control by right clicking on the image and hit Send to Scan Area Control Click on the Scan Area tab at the bottom of the scan control window This will allow you to shift your next scan area relative to the image you just collected and automatically adjusts the software settings to the appropriate scan size x y offsets and rotation To take an image in a different region of the sample disengage from surface using the Z stage control adjust position with the x y stage controls and repeat the surface approach procedure Modify with XEI You can send images directly to the XEI data analysis software from the Buffer in XEP by right clicking on them Or you can open them from the spmdata or User Files folders Flattening your sample To flatten your sample go to the flatten area button Select all the areas that are supposed to be flat and on the same height Exclude any non flat features e g mesas trenches etc that will distort the flattening routine Select the line routine 1 order regression Select x axis if your scan was done in the x axis y axis if your scan was done in the y axis Click Execute Save your flattened image preferably as a separate file You can also try using the line whole and difference routines Read more about the different flattening algorithms in the XEI manual Another useful feature is the ability to save t
8. croscope Position laser on cantilever If you don t see the laser reflected off the cantilever and the laser is on use the Mirror 1 knobs to move the laser along the vertical axis until you see it reflected on the edge of the probe then move horizontally as needed Tightening the vertical position knob will raise the laser loosening it will lower it Aim the laser near the end of the cantilever example above where the deflection will be the greatest NOTE Make sure that you position the primary beam spot you may also see a reflected laser spot that will be dimmer To make sure you are aiming the primary beam spot on the end of the cantilever move the laser to the base of the cantilever then loosen the vertical position knob This will lower the laser position to the end of the cantilever Aim Laser on Center of Photodetector Adjust Mirror 2 1 Open XEP data acquisition software 2 If frequency sweep window opens just click OK to close 3 Check A B voltage value measures total intensity of light hitting the photodetector which should be 2 5 3 V if using ACTA tips provided by the lab The correct value will be different if you use your own tips If the A B sum is significantly less than 3V this indicates that either the mirror 2 alignment is way off or the laser spot on the cantilever may actually be the reflected spot as mentioned above Adjust the mirror 2 knobs first turning one knob until
9. der of Operations Turn on system mount tip 1f needed Focus microscope on cantilever Position laser spot on end of cantilever Turn on XEP software Position laser reflection on center of photodetector A B 3V A B and C D 0 3V Select appropriate mode scanner and frequency Approach surface 1 Get optically close several mm li Focus microscope 1mm from surface move Z stage til cantilever in focus i Repeat at 100um from surface iv Click Approach Adjust scan parameters Z servo gain scan rate set point drive Collect images 10 Send to XEI to process save 11 Follow shutdown procedure
10. he numerical data from any image graph or table as a txt file by right clicking on the image graph table Save file to User Files folder Include tiff extension in filename Screenshots can also be exported as jpg png of bmp files by selecting Export from the File menu Shut down CAUTION ENSURE HEAD WILL NOT CONTACT MICROSCOPE 1 DO NOT hit Lift Z right away The stage will hit the microscope on the way up and damage one or both of them Use Z stage control to lift tip away from sample a small distance so the tip is no longer touching the sample 2 Open the acoustic enclosure 3 Raise microscope to full height 4 Hit Lift Z button in Z stage control 5 Refocus microscope on cantilever 6 Turn off laser switch unplug head and remove head 6a If the tip you used was good leave it on the AFM Otherwise remove head and place tip in the Bad Probes box 7 Remove sample from stage and replace head plug head in 8 Close XEP XEI software 9 Turn off Light Bank 10 Leave the SPM Controller on Log tip usage clean up after yourself and disable in Badger Note We recycle used sample pucks and also used tips Please put used items in the proper places Do not throw them in the trash If you notice any problems while running the instrument please make a comment on Badger and email the lab managers at nano spmstaff lists stanford edu 1 2 3 4 5 6 T 8 9 XE 70 AFM Or
11. peat until the system is able to engage the sample Optimize Scan Parameters a 1 Click on Input configuration icon E 2 Select channels of interest show at least trace and retrace of topography 3 Change the scan size to appropriate value Sum or 50um max depending on XY scanner A good image is achieved when trace and retrace show the same topography and are stable repeatedly show the same topography and there are no random spikes or other instabilities Trace and retrace may not overlap perfectly but if the shapes of the curves are identical then the image will generally be good To improve the tracking adjust the following 4 Z servo gain Increasing the gain will help the tip track the sample better However if the gain is too high you will see high frequency noise Increase decrease the gain until the best image is obtained In most cases the optimal value will be from 1 5 5 Scan rate Decrease the scan rate to 0 5 Hz or 0 25 Hz This effectively reduces the speed at which the tip moves allowing the system more time to adjust to sudden changes in height Be careful as this will also proportionately increase the time needed to take images 6 Set Point Adjusting the Set Point up down will change the distance the tip hovers above the sample surface when scanning and can improve resolution of features surface tracking This can be done manually by changing the number in the Set Point box or by adjusting the red bar