1. process must include inspection to assure damage has not occurred Instructions for cleaning the EpiScope Clean EpiScope with a mild glass cleaning type solution labeled compatible for use on plastic Use of prepackaged glass cleaner wipes is preferred however cleaning solutions can also be lightly applied to a lint free cloth and then used on the EpiScope Apply pre moistened wipe cloth to surfaces of the EpiScope and wipe to clean visible debris Instructions for cleaning the EpiScope s lens or contact plate glass Clean debris from the lens or contact plate with a cotton tipped applicator moistened with 70 ethyl alcohol Instructions for disinfecting the EpiScope s contact plate glass After cleaning the EpiScope s contact plate can be disinfected with the following hospital type low or intermediate level disinfectant solutions e Quaternary ammonium based with alcohol lt 50 low or intermediate level disinfectant e Inorganic chlorine based low or intermediate level disinfectant Note Combination cleaner disinfectants that carry a low or intermediate level disinfectant claim based upon the disinfectant types noted above are also suitable for use on the contact plate 1 Remove the contact plate as shown 2 Apply disinfectant to the contact plate per disinfectant agent manufacturer s instructions 3 After disinfection the optical element may be recleaned if necessary 4 Re install contact plate on
2. Scope contact plates are guaranteed for one year from date of purchase Technical specifications Magnification 10x Focal Range 5D correction Focal Length 25mm Welch Allyn Inc 4341 State Street Road Skaneateles Falls NY 13153 USA Tel 1 800 535 6663 or 1 315 685 4560 Fax 1 315 685 3361 www welchallyn com For Welch Allyn Service Center telephone numbers visit www welchallyn com about company locations htm cc rep Regulatory Affairs Representative Welch Allyn Limited Navan Business Park Dublin Road Navan County Meath Republic of Ireland 2013 Welch Allyn Inc All rights reserved To support the intended use of the product described in this publication the purchaser of the product is permitted to copy this publication for internal distribution only from the media provided by Welch Allyn Welch Allyn assumes no responsibility for any injury to anyone or for any illegal or improper use of the product that may result from failure to use this product in accordance with the instructions cautions warnings or statement of intended use published in this manual Welch Allyn is a trademark of Welch Allyn Inc
3. Welch Allyn Welch Allyn EpiScope Skin Surface Microscope Directions for Use 47300 Mat No 719632 Ver B 4 2013 Welch Allyn Inc 0297 Intended use Allows visualization and measurement of skin surface lesions Symbols WARNING identifies information within the manual to avoid injury A CAUTION identifies information within the manual to avoid equipment failure Temperature Range Humidity Range Meets essential requirements of European Medical Device Directive 93 42 EEC q Do not dispose of this product as unsorted municipal waste Prepare this product for reuse or separate collection as specified by Directive 2002 96 EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment WEEE If this product is contaminated this directive does not apply For more specific disposal information see www welchallyn com weee or contact Welch Allyn Customer Service 44 207 365 6780 Warnings and cautions WARNING This cleaning and disinfection procedure and the equipment and materials described must be followed and conducted by persons trained and familiar with medical device cleaning and disinfection WARNING Consult your facility s procedures and the reprocessing agents manufacturer s instructions for the agents reuse recommended Personal Protective Equipment and other safety precautions WARNING Lamps may be hot Allow to cool before r
4. emoval Handle and dispose of the lamp with care to avoid shattering Use protective eyeglasses when handling WARNING The EpiScope is intended to be used ONLY with intact skin CAUTION Failure to follow these instructions may cause damage to this device CAUTION Do not immerse or allow excessive moisture to contact the EpiScope Doing so may cause damage to the device CAUTION Only use the cleaner disinfectant types identified within these instructions Do not use other cleaners or disinfectants CAUTION Always follow the cleaner or disinfectant manufacturers instructions for proper use CAUTION Repeated reprocessing may degrade elements of the device Follow inspection procedures to assure damage has not occurred CAUTION Do not use abrasive material to clean lens Doing so can scratch and permanently damage optics CAUTION The EpiScope is not for diagnostic purpose use aa a a EE gt gt Components Rubber Eye Guard Focal Ring Battery Handle Contact Plate Rheostat 1 Install EpiScope onto Welch Allyn 3 5v handle Instructions 2 Power ON 3 Center EpiScope over lesion to examine surface and adjust focal ring Care and maintenance Cleaning and disinfection The following instructions apply to the Welch Allyn EpiScope Limitations on cleaning and disinfection This device may be cleaned and disinfected multiple times in accordance with these instructions Cleaning and disinfection
5. to the EpiScope Maintenance inspection and testing 1 Inspect the EpiScope for signs of damage 2 Appropriately discard any component that shows evidence of damage or deterioration 3 Contact Welch Allyn for component replacement 4 Verify the proper operation of the EpiScope per this manual s directions for use Storage Store the EpiScope in a dry clean area A carrying case accessory is available which holds the 3 5v rechargeable handle and extra contact plate Lamp replacement 1 Remove EpiScope 2 Remove lamp ri 3 Remove new 03000 lamp from packaging Be careful not to touch the glass surface If lamp surface is smudged wipe clean with a soft cloth 4 Insert 03000 Lamp until contact pin is below base flange 5 Install EpiScope Power sources The EpiScope Skin Surface Microscope is compatible with all 3 5 v Welch Allyn power sources Accessories 47310 Reticle Contact Plate 47320 Plano Contact Plate 05232 Carrying Case Warranty Welch Allyn warrants the product to be free of defects in material and workmanship and to perform in accordance with the manufacturer s specifications for the period of one year from the date of purchase from Welch Allyn or its authorized distributors or agents Welch Allyn will repair or replace free of charge any parts of its own manufacture proven to be defective through causes other than misuse neglect damage in shipment or normal wear The Epi