S8 Lightweight™ System

image

Contents

1. S8 LIGHTWEIGHT MASKS HUMIDIFIERS RESSCAN DATA CARD MODULE ACCESSORIES How TO USE THE S8 LIGHTWEIGHT SET UP THE S8 LIGHTWEIGHT ASSEMBLE THE MASK START TREATMENT STOP TREATMENT HOW TO ATTACH A HUMIDIFIER ATTACH A HUMIDAIRE 31 HUMIDIFIER ATTACH A HUMIDAIRE HUMIDIFIER HOW TO USE THE How TO CHANGE THE SETTINGS ON YOUR S8 LIGHTWEIGHT SO MODULES suot ue eb ble Be RU b NUTUS NEL ES qe USING A BLANK MODULE USING A RESSCAN DATA CARD MODULE CLEANING AND DAILY CLEANING WEEKLY CLEANING PERIODIC CLEANING REPLACING THE AIR FILTER SERVICING FREQUENTLY ASKED QUESTIONS TROUBLESHOOTING sese ren SYSTEM SPECIFICATIONS ND nA co 4 MM CONTENTS S8 LIGHTWEIGHT ACCESSORIES INTRODUCTION Thank you for choosing the S8 LIGHTWEIGHT flow generator The S8 LIGHTWEIGHT is a compact all in one package with an inbuilt power supply and the option of integrated data management and humidification INTRODUCTION This user manual contains the information you need for the correct use of your S8 LIGHTWEIGHT YOUR RESPONSIBILITY The owner or user of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from
2. environment NOTE Ut is the AC mains voltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic immunity The S8 series of devices is intended for use in the electromagnetic environment specified below The customer or the user of the 58 device should assure that the device is used in such an environment SYSTEM SPECIFICATIONS IEC60601 1 2 Compliance Immunity test Electromagnetic environment guidance test level level Portable and mobile RF communications equipment should be used no closer to any part of the S8 device including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 Vrms d 1 17 IEC 61000 4 6 150 kHz to 80 MHz Radiated RF 10 V m 10 V m d 0 35 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz d 0 70 NP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol CD NOTE 1 At 80 MHz and 800MHz the higher frequency
3. operation which is not in accordance with the operating instructions supplied maintenance or modifications carried out unless in accordance with authorised instructions and by authorised persons Please read this manual carefully before use This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information Warning alerts you to possible injury Caution explains special measures for the safe and effective use of the device Note is an informative or helpful note MEDICAL INFORMATION ABOUT THE S8 LIGHTWEIGHT The S8 LIGHTWEIGHT continuous positive airway pressure CPAP system is indicated for the treatment of obstructive sleep apnoea OSA in adult patients The S8 LiGHTWEIGHT CPAP system is intended for home and hospital use CONTRAINDICATIONS The S8 LIGHTWEIGHT is not a life support device and may stop operating during power failure or certain fault conditions It should not be used by patients who are dependent on continuous therapy Positive airway pressure therapy may be contraindicated in some patients with the following pre existing conditions severe bullous lung disease pneumothorax pathologically low blood pressure dehydration cerebrospinal fluid leak recent cranial surgery or trauma WARNINGS Read the entire manual before using the S8 LIGHTWEIGHT Advice contained in this manual shou prescribing physician d
4. 4 2 8 kV air 8 kV air floors are covered with synthetic material the relative humidity should be at least 3096 Electrical fast 2 kV for power 2 Mains power quality should transient burst supply lines be that of a typical IEC 61000 4 4 commercial or hospital 1 kV for input Not Applicable environment output lines Surge 1 kV differential 1 kV differential Mains power quality should IEC 61000 4 5 mode mode be that of a typical 2 kV common mode 2 kV common mode commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 596 Ut 95 dip in Ut or 0 5 cycle 4096 Ut 6096 dip in Ut or 5 cycles 7096 Ut 3096 dip in Ut or 25 cycles 596 Ut gt 95 dip in Ut or 5 sec 12V 79596 dip in 240V or 0 5 cycle 96V 60 dip in 240V or 5 cycles 168V 3096 dip in 240V or 25 cycles 12V gt 95 dip in 240V or 5 sec Mains power quality should be that of a typical commercial or hospital environment If the user of the S8 device requires continued operation during power mains interruptions it is recommended that the S8 device be powered from an uninterruptible power source Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital
5. SYSTEM The S8 LIGHTWEIGHT device is part of a system that consists of six elements S8 LIGHTWEIGHT flow generator Mask e Humidifier Data management products Premium noise guard front cap Accessories S8 LIGHTWEIGHT Please refer to the illustrations in section of the illustration sheet FRONT VIEW 1 Shows the keypad and LCD screen and the air outlet REAR VIEW A 2 Shows where the module attaches and the location of the air filter and power sockets POWER CORD A 3 Power cord to attach the S8 LIGHTWEIGHT to the AC mains power CARRY BAG A 4 The S8 LIGHTWEIGHT carry bag contains two pouches one for storing the 58 device the other for storing the HUMIDAIRE 3i When storing the HUMIDAIRE 3i always put it in the larger mesh pouch AIR TUBING The 2 m air tubing A 5 connects the S8 LIGHTWEIGHT to your mask You will need medium 52 cm air tubing see C 3 to connect the 58 LIGHTWEIGHT to the HUMIDAIRE humidifier MASKS You will also need a ResMed mask system supplied separately The ResMed mask systems that are recommended for use with the S8 LIGHTWEIGHT are NASAL MASKS MODULAR NASAL MASK e MIRAGE VISTA NASAL MASK ULTRA MIRAGE NASAL MASK MIRAGE NASAL MASK e MIRAGE ACTIVA NASAL MASK e PAPILLON NASAL PILLOWS SYSTEMS e MIRAGE SwiFT NASAL PILLOWS SYSTEM THE S8 LIGHTWEIGHT SYSTEM FULL FACE MASKS e MIRAGE FULL FACE MASK e MIRAGE FULL FAC
6. ULTRA MIRAGE mask are given in the following table Device Pressure Power Power cm H20 VA W S8 LIGHTWEIGHT 5 17 8 10 23 12 15 29 16 20 34 20 Power outlet for HumidAire 3i Outlet range 100 240V 50 6095 10V 400Hz 2 0A Air Filter Two layered powder bonded polyester non woven fibre Air Tubing Flexible plastic 2 m Air Outlet The 22 mm conical outlet complies with EN 1281 1 IEC 60601 1 Classifications Class I double insulation Type CF Noise Level Radiated sound pressure is measured at lt 30dBA at 10 cm H20 and m according to ISO 7510 1 ACCESSORIES DATA CARD Dimensions L x W x D 65 04 mm x 22 15 mm x 0 85 mm SYSTEM SPECIFICATIONS 24 DATA CARD MODULE Dimensions x W x D 49 7 mm x 42 2 mm x 12 mm ALL DEVICES WHERE APPLICABLE Housing Construction Flame retardant engineering thermoplastic Environmental Conditions Operating Te Storage and Operating Al Electromagne mperature 5 to 40 Operating Humidity 10 95 non condensing Transport Temperature 20 C to 60 Storage and ransport Humidity 1026 9576 non condensing titude sea level to 259 m Electromagnetic Compatibility Product complies with all applicable electromagnetic compatibility requirements EMC according to IEC60601 1 2 for residential commercial and light industry environments For further details see Guidance and Manufacturer s Declaratio
7. be performed by an authorised service agent Explosion hazard do not use in the vicinity of flammable anaesthetics CAUTIONS At low pressures the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing Some rebreathing may occur The airflow for breathing produced by this device can be as much as 6 C higher than the temperature of the room Caution should be exercised if the room temperature is warmer than 32 C When AC mains power 100 240V AC is not available always use a ResMed DC 12 converter The DC 12 converter is available as an optional accessory It is not supplied with all models f supplied the Data Card module is designed for use with specific ResMed devices Only use the Data Card module if supplied as specified in this manual Note The above are general warnings and cautions Specific wamings cautions and notes appear with the relevant instructions in the manual ADVERSE EFFECTS Patients should report unusual chest pain severe headache or increased breathlessness to their prescribing physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during the course of therapy with the CPAP device drying of the nose mouth or throat nosebleed bloating ear or sinus discomfort eye irritation skin rashes chest discomfort MEDICAL INFORMATION THE S8 LIGHTWEIGHT
8. 09m 610 1219 m e 1220 1828 m e 829 259 m For example if you are at an altitude of 947 m your altitude compensation setting would be 610 1219 m as 947 m falls within that range Press the Left key Apply to select the correct setting HOW TO CHANGE THE LANGUAGE SETTING To display menus and messages in a different language select the S8 LIGHTWEIGHT menu by pressing the Left key in the standby Ramp screen Scroll to LANG and select Change Left key Press the Up or Down keys until you see the language that you require OTHER VIEWABLE INFORMATION You can also view the total number of hours of treatment USED HRS the number of sessions that the S8 LIGHTWEIGHT was used in a usage period USAGE and the current software version installed on your unit SW Access to all of these screens is provided via the menu See How to enter the menu screens on page 12 HOw TO USE THE KEYPAD MESSAGES ON THE S8 LiGHTWEIGHT LCD Your clinician may have set your S8 LIGHTWEIGHT to remind you about important events such as when to replace your mask w Data Card enabled and so on device is The reminder message is displayed on the LC delivering therapy The backlight on the LCD Your clinician can set any of the following reminders on your LCD hen to insert your Data Card if your D and is visible if the device is not flashes when a message is displayed Message Descr
9. 37 634 NZ toll free or 64 9 239 0193 Email reception resmed co nz ResMed SA Parc de la Bandonni re 2 rue Maurice Audibert Saint Priest France Tel 33 0 4 37 251 251 Fax 33 0 4 37 251 260 Email reception resmed fr ResMed Singapore Pte Ltd 238A Thomson Road 12 03 04 Novena Square Tower A Singapore Tel 65 6284 7177 Fax 65 6284 7787 Email reception resmed com sg ResMed Spain SL C Arturo Soria 245 Madrid Espa a Tel 34 93 590 8154 Fax 34 93 590 8153 Email angelo resmed es ResMed Sweden AB Industrigatan 2 S 461 37 Trollhattan Sverige Tel 46 520 420 110 Fax 46 520 397 15 Email reception resmed se Labhardt AG Thannerstrasse 57 CH 4054 Basel Schweiz Tel 41 061 307 9711 Fax 41 061 307 9722 Email info labhardt ch US DESIGNATED AGENT ResMed Corp EU AUTHORIZED REPRESENTATIVE ResMed UK Ltd www resmed com Protected by patents US US US US 6705315 Other patents pending Design registrations pending 58 S8 Lightweight HumidAire HumidAire 3i Mirage Mirage Activa Mirage Vista Ultra Mirage Mirage Swift and ResScan are trademarks of ResMed Ltd Papillon is a trademark of MAP Medizin Technologie GmbH ResMed Ltd 2005 CONTENTS INTRODUCTION 0 cece ccc hh hh mh YOUR RESPONSIBILITY MEDICAL INFORMATION ABOUT THE S8 LIGHTWEIGHT CONTRAINDICATIONS WARNINGS CAUTIONS ADVERSE EFFECTS THE S8 LIGHTWEIGHT SYSTEM
10. E MASK SERIES 2 ULTRA MIRAGE FULL FACE MASK WARNING ResMed mask systems are compatible for use with the S8 Lightweight Please refer to Warnings on page 2 HUMIDIFIERS A humidifier may be required if you are experiencing dryness of the nose throat or mouth The 58 LIGHTWEIGHT is compatible for use with HUMIDAIRE 3i integrated heated humidifier Figure B on the illustrations sheet e HUMIDAIRE heated humidifier Figure I on the illustrations sheet RESSCAN DATA CARD MODULE The RESSCAN Data Card module and RESSCAN Data Card C 1 may be used with the S8 LIGHTWEIGHT either to help your clinician to monitor your treatment or to provide you with updates to your device settings See Using a ResScan Data Card Module on page 15 for information about how to use the Data Card module and Data Card ACCESSORIES Please refer to the illustrations in section C of the illustration sheet Other accessories for use with the S8 LIGHTWEIGHT Premium noise guard front cap standard with S8 LIGHTWEIGHT C 2 Airtubing 3 m C 3 Medium air tubing 52 cm for HUMIDAIRE connection C 4 DC 12 converter C 5 Blank module C 6 Note ResMed regularly releases new products Please check our website at lt www resmed com gt How TO USE THE S8 LIGHTWEIGHT Please refer to the illustrations in section D of the illustration sheet SET UP THE S8 LIGHTWEIGHT 1 Place the 58 LIGHTWEIGHT on a table near the head of yo
11. Fr S8 Lightweight System Reorder number 338239 1 05 06 USER S MAN UAL English www resmed com Global Leaders in Sleep and Respiratory Medicine INTEGRATED HUMIDIFIER RESMED D SETTING UP THE S8 S8 Lightweight System Component of ILLUSTRATIONS A SYSTEM COMPONENTS HumidAire 2005 ResMed Ltd Carry handle I LCD screen E OTHER ACCESSORIES Air outlet Connector plug Air filter Module cover DC input Air inlet AC input 2005 ResMed Ltd 2005 ResMed Ltd 2005 ResMed Ltd HUMIDAIRE 3i and S8 LIGHTWEIGHT BLANK MODULE Data communications port Note Some products not available in all regions 2005 ResMed Ltd 2005 ResMed Ltd 8 58 E Es g ra S a USING THE DATA CARD USING A HUMIDAIRE 2005 ResMed Ltd 2005 ResMed Ltd RESMED S8 Lightweight System Reorder number 338239 1 05 06 USER S MAN UAL English Manufactured by ResMed Ltd Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia Tel 61 2 8884 1000 or 800 658 189 toll free Fax 61 2 8883 3114 Email reception resmed com au ResMed Offices ResMed Corp Danielson Street Poway CA 92064 6857 Tel 1 858 746 2400 or 1 800 424 0737 toll free Fax 1 858 746 2900 Email reception resmed com ResMed UK Ltd 65 Milton Park Abingdon
12. HT IF MAINS AC POWER IS NOT AVAILABLE A DC 12 converter is available to connect the S8 LIGHTWEIGHT to a 12 or 24V DC power source Contact your clinician for details CAUTION The device should not be connected to both AC and DC power sources simultaneously TROUBLESHOOTING If there is a problem try the following suggest ions If the problem cannot be solved contact your equipment supplier or ResMed Do not attempt to open the device Problem No display Possible Cause Power not connected TROUBLESHOOTING Solution Ensure the appropriate power cord is connected to your S8 Lightweight and the power switch if available is on Insufficient air delivered from 58 Lightweight Ramp Time is in use Air filter is dirty Air tubing is kinked or punctured Air tubing not connected properly Mask and headgear not positioned correctly Cap s missing from access port s on mask Altitude compensation setting is incorrect Pressure required for treatment may have changed Wait for air pressure to build up or change ramp time Replace air filter Straighten or replace tubing Check air tubing Adjust position of mask and headgear Replace cap s Adjust altitude compensation setting to the correct altitude range See your clinician to adjust the pressure Excessive air delivered from Altitude compensation Adjust altitude compensation the 58 Lig
13. Oxfordshire OXI4 4RX UK Tel 44 1235 862 997 Fax 44 1235 831 336 Email reception resmed co uk ResMed Asia Pacific Ltd Danielson Street Poway CA 92064 6857 Tel 61 2 9886 5000 or 800 991 900 toll free Fax 61 2 9889 1471 Email reception resmed com au ResMed Austria PurtscherstraBe 18 9500 Villach sterreich Tel 43 4242 430 830 Fax 43 4242 430 830 4 Email reception resmed at ResMed BV Nieuwe Parklaan 86 2587 BV Den Haag Nederland T l 31 70 358 6263 Fax 31 70 358 4333 Email receptiondh resmed nl ResMed Finland Niittykatu 6 FIN Espoo Suomi Tel 358 9 Fax 358 9 86768222 Email reception resmed fi ResMed GmbH amp Co KG FraunhoferstraBe 16 82152 Martinsried Deutschland Tel 49 0 89 99 01 00 or 49 0 800 2 777 000 geb renfrei Fax 49 0 89 99 01 10 55 Email reception resmed de ResMed Hong Kong Ltd Room 1714 Miramar Tower 132 134 Nathan Road Tsim Sha Tsui Hong Kong Tel 852 2366 0707 Fax 852 2366 4546 Email reception resmed com hk ResMed Japan Nihonbashi Hisamatsu Bldg 4F 2 28 1 Nihonbashi Hamacho Chuo Ku Tokyo 103 0007 Japan Tel 81 3 3662 5056 Fax 81 3 3662 5040 ResMed Malaysia Sdn Bhd Suite E 10 20 Plaza Mon t Kiara No 2 Jalan 1 70C Mon t Kiara 50480 Kuala Lumpur Malaysia Tel 60 3 6201 7177 Fax 60 3 6201 2177 Email reception resmed com my ResMed NZ Ltd PO Box 51 048 Pakuranga Auckland New Zealand Tel 0800 737 633 NZ toll free or 64 25 737 633 Fax 0800 7
14. Oxygen supports combustion Oxygen should not be used while you are Explanation Wh unlikely event of certain fault conditions pressures up to 30 cm H thing of exhaled mask H20 In the 20 are possible oxygen is used with this device the oxygen flow must be turned off when the device is not operating oxygen has been left on turn off the flow generator then wait 30 minutes before turning on the flow generatory again en the CPAP device is not in operation and the oxygen flow is ft on oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire This applies to most types of smoking or in the presence of an open flame supply is turned on Always ensure airflow is being generated by the device before the oxygen Always turn the oxygen supply off before stopping the airflow from the device Note At a fixed rate of supplemental oxygen flow the inhaled oxygen concentration will vary depending on where the oxygen is introduced the pressure settings patient breathing pattem mask selection and leak rate Ports may be incorporated into the mask or in connectors that are near the mask Do not use the S8 LIGHTWEIGHT if there are obvious external defects or unexplained changes in performance Do not open the S8 LIGHTWEIGHT case There are no user serviceable parts inside Repairs and internal servicing should only
15. TLY ASKED QUESTIONS HOW WILL IT FEEL WHEN I AM BREATHING WITH THE S8 LIGHTWEIGHT When you first use your S8 LIGHTWEIGHT you may feel uncomfortable breathing against the airflow This is normal You will continue to breathe normally while you sleep Take time to adjust to this new sensation WHAT DO DO WHEN WANT TO GET UP IN THE NIGHT If you need to get up during the night remove your mask and stop therapy Always remember to restart therapy and put your mask on when you return to bed WHAT DO I DO IF FEEL AIR LEAKING OUT OF THE MASK The flow generator delivers the most effective treatment when the mask is well fitted and comfortable Treatment can be affected by leaks so it is important to eliminate any leaks that may arise For example try repositioning the mask and adjusting the straps If you are using a nasal mask try to keep your mouth closed during treatment Air leaks from your mouth can decrease the effectiveness of your treatment If mouth leaks are a problem a full face mask or chin strap may help Contact your clinician or equipment supplier for further details If you are using a full face mask and you have problems trying to get a comfortable mask fit contact your sleep clinic or equipment supplier You may benefit from a different size or style of mask ANYTHING SPECIAL SHOULD DO BEFORE PUTTING THE MASK ON Before wearing your mask wash your face to remove excess facial oils This will allow a be
16. backlight is always on while the S8 LIGHTWEIGHT is on How TO CHANGE THE SETTINGS ON YOUR S8 LIGHTWEIGHT You can change settings such as Ramp Time Altitude on the S8 LIGHTWEIGHT by using the keypad and LCD When the S8 LIGHTWEIGHT is in standby mode you can display a series of screens a menu on the LCD screen in order to view and change the settings for a particular function 58 LIGHTWEIGHT PATIENT MENU RAMP 10min Menu ALT 0 610m Change 1 Exit USEDHRS 960 Exit USAGE 120 180 t Exit LANG ENGLISH Change i Exit SW SX123456789 1 Exit Use the menu screens to change the ramp time altitude compensation and language You can also view details about the usage and software version of your S8 LIGHTWEIGHT HOw TO ENTER THE MENU SCREENS When the S8 LIGHTWEIGHT standby screen is displayed you will see the word Menu in the left hand corner To select the S8 LIGHTWEIGHT menu press the Left key HOW TO SCROLL THROUGH THE MENU When you are in the menu scroll through items by pressing the Up or Down keys HOW TO CHANGE A SETTING OPTION When you have scrolled to the item you require Press the Left key Change 2 Press the Up or Down keys until you see the setting option that you require 3 Press the Left key Apply to select the setting option If you wish to exit without changing the option press the Right key Cancel 12 HOW TO EXIT OUT OF A MENU Press the Right key Exit This w
17. cording to the instructions supplied with the mask Air tubing Disconnect the air tubing from the S8 LIGHTWEIGHT and humidifier if used and hang the tubing and mask in a clean dry place until next use Humidifier If you are using a humidifier clean it according to the instructions in the humidifier user s manual WEEKLY CLEANING Remove the air tubing from the S8 LIGHTWEIGHT and the mask 2 Wash the air tubing in warm water using mild detergent Rinse thoroughly hang and allow to dry 3 Before next use assemble the mask and headgear according to the user instructions 4 Reconnect the air tubing to the air outlet and mask PERIODIC CLEANING Clean the exterior of the S8 LIGHTWEIGHT with a damp cloth and mild detergent 2 Check the air filter for blockage or holes see Replacing the Air Filter on page 18 WARNING Beware of electric shock Do not immerse the unit or power cord in water Always unplug the unit before cleaning and be sure it is dry before plugging back in The mask system and air tubing are subject to normal wear and tear Inspect them regularly for damage CLEANING AND MAINTENANCE 17 18 REPLACING THE AIR FILTER To replace the air filter Remove the air filter cover at the back of the S8 LIGHTWEIGHT Air filter cover 2 Remove and discard the old air filter 3 Insert a new filter with the blue tinted side facing out from the device 4 Replace the air filter cover In
18. e serial adapter To remove the blank module pull it off the back of the S8 LIGHTWEIGHT to show the data communications port F 1 To attach the blank module push the blank module onto the data communications port until the module clicks into place F 2 Note To protect the data communications port always ensure there is a module attached to the device USING A RESSCAN DATA CARD MODULE Please refer to the illustrations in section amp of the illustration sheet The RESSCAN Data Card module is an accessory to the 58 LIGHTWEIGHT Use a RESSCAN Data Card with the Data Card module to collect data and update settings in the S8 LIGHTWEIGHT The module and card give your clinician convenient access to data about your treatment and can provide you with new device settings from your clinician without you having to leave home ATTACHING THE MODULE To attach the Data Card module push it onto the data communications port at the back of your S8 LIGHTWEIGHT until it clicks into place G 1 REMOVING THE MODULE To remove the Data Card module pull it off the back of the 58 LIGHTWEIGHT G 2 Note To protect the data communications port always ensure there is a module attached to the device UsiNG THE DATA CARD TO COLLECT DATA Please refer to the illustrations in section H of the illustration sheet If your clinician needs to review your treatment they will ask you to use the Data Card to copy data from your S8 LIGHTWEIGHT and to ret
19. h nose or throat 19 E error message 21 F frequently asked questions 19 H headgear 21 HumidAire attaching 9 HumidAire 3i attaching 9 storage 5 humidifier 19 cleaning 17 humidifiers 6 INDEX I international use 19 K keypad 5 11 keypad keys functions 11 L language setting 13 LCD 5 7 11 error messages 22 messages 14 left key 11 M mailback envelope 16 mask 19 21 air leak 19 attaching 7 cleaning 17 removing 19 mask leak 19 mask port 21 masks 5 medical information 1 menu screens 12 access 12 change 12 exit 13 scrolling 12 module attaching 5 blank 15 Data Card 15 modules 15 blank 15 29 nasal irritation 19 P power cord 5 21 connecting 7 power sockets 5 power sources 20 power supply 23 R ramp time 8 13 21 reminder replace mask 14 reminders 14 ResScan Data Card module 6 right key 11 S S8 Lightweight dimensions 23 24 servicing 18 settings 12 set up 7 start treatment 8 stop treatment 8 travelling with 19 setting up 7 software version number 13 standby screen 7 start stop key 11 SW 13 system specifications 23 T troubleshooting 21 U up key 11 Usage 13 use on aircraft 20 Used hours 13 Ww weekly cleaning 17
20. htweight setting is incorrect setting to the correct altitude range Display error message Component failure Call service EXXXX Call Service where xxxx defines an error or MOTOR FAULT Call Service or UPLOAD LANGUAGES Call Service or DATA LOST Call Service 21 22 Problem The following message is displayed on the LCD after you try to update settings using the Data Card Card Error Remove Card Possible Cause The Data Card is not inserted correctly The Data Card is only partly inserted You may have removed the Data Card before settings were copied to the S8 Lightweight The card is not a ResScan Data Card Solution Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device Ensure that the device is switched on and in standby mode before you insert the Data Card Ensure that the Data Card is inserted into the slot on the Data Card module as far as it will go Reinsert the Data Card and wait for the Settings Success Remove Card message to appear on the LCD Only use a ResScan Data Card The following message is displayed on the LCD after you try to copy data to a Data Card Card Error Remove Card The Data Card is not inserted correctly The card is not a ResScan Data Card Ensure that the Data Card is inserted with the arrow facing up and pointing toward the device Use a ResScan Data Card as specified in this manua
21. ill take you back to the Patient Menu screen HOW TO EXIT OUT OF AN ITEM WITHIN THE MENU Press the Right key Exit How TO CHANGE THE RAMP TIME During ramp time the pressure increases from a low pressure to the prescribed treatment pressure You can set the ramp time in 5 minute intervals from minutes OFF up to a maximum ramp time set by your clinician The standby screen displays the current ramp time in the right hand corner To change the ramp time press the Up or Down key until the desired ramp time appears Once a ramp time is selected begin therapy by pressing the Start Stop key When ramping begins the LCD will display the word RAMPING with the current pressure indicated as dashes Each dash is approximately equivalent to 2 cm H5O For example six dashes represents approximately 12 cm HO pressure HOW TO CHANGE THE ALTITUDE SETTING If you move or travel with your S8 LIGHTWEIGHT to areas of different altitude you will need to adjust the altitude compensation setting Changes in altitude will affect the pressure delivered by the S8 LIGHTWEIGHT You can use the Altitude setting ALT to specify the altitude range at which you are using your S8 LIGHTWEIGHT Select the S8 LIGHTWEIGHT menu by pressing the Left key in the standby RAMP screen Scroll to ALT and select Change Left key Press the Up or Down keys until you see the setting option that you require There are four altitude settings to choose from 0 6
22. iption Action INSERT CARD May appear if nsert your Data Card and follow any your device is instructions that your clinician has given Data Card you When you have done this the enabled message will disappear from the LCD Pressing the Left Ok key will also remove the message REPLACE May appear when Press the Left Ok key to remove the MASK your mask is due message from your LCD and replace your for replacement mask with a new one CALL May appear as a Press the Left Ok key to remove the PROVIDER reminder to message from your LCD and contact your contact your clinician clinician for example to discuss how your therapy 15 going REPLACE ay appear as a Press the Left Ok key to remove the FILTER reminder to message from your LCD and replace the replace the air air filter filter on your device SERVICE DUE Press the Left Ok key to remove the ay appear as a reminder to return your device for service message from your LCD and clinician contact your 14 S8 MODULES Note Do not remove modules while power is connected to the S8 Lightweight Your S8 LIGHTWEIGHT will have one of two modules fitted to the back a blank module or a RESSCAN Data Card module S8 MODULES USING A BLANK MODULE Please refer to the illustrations in section F of the illustration sheet The blank module is a protective cover on the S8 LIGHTWEIGHT At times you may wish to remove or attach this module for example to use th
23. l The following message is displayed on the LCD after you try to update the settings using the Data Card Settings Invalid Remove Card The identification details on the Data Card do not match the details on your device Contact your clinician immediately The following message is displayed on the LCD after you try to update the settings using the Data Card Settings Error Remove Card There is a data error on the Data Card Contact your clinician immediately The following message is NOT displayed on the LCD after you try to update the settings using the Data Card Settings Success Remove Card The settings were not updated Contact your clinician immediately SYSTEM SPECIFICATIONS S8 LIGHTWEIGHT Performance Operating pressure range 4 to 20 cm H5O Dimensions x W x D 112 mm x 164 mm x 145 mm Weight 3 kg Power Supply Input range for S8 LIGHTWEIGHT with HUMIDAIRE 3i 00 240V 50 60Hz OV 400Hz 2 5A lt 140VA 110W maximum power consumption Instantaneous peak power consumption 340VA Refer to the DC 12 converter instructions for DC ratings Only use the ResMed DC 12 converter for DC input Actual power consumption will vary depending on factors such as the use of accessories height above sea level and ambient temperature Power consumption values for typical treatment conditions using a breathing machine set to 0 5 L and 5 breaths min with 2 m air tubing and an
24. lback envelope Fold up the mailback envelope 2 Place the mailback envelope containing the Data Card in the bubble mailer envelope and seal it 3 Make sure that your clinicians name and address details are on the bubble mailer envelope and mail it back immediately USING THE DATA CARD TO UPDATE SETTINGS If your clinician has provided a Data Card with new device settings With the device in standby Ramp mode insert the Data Card into the slot on the Data Card module Updating will start automatically he Inserted Please Wait message is displayed on the LCD while updating is in progress Updating takes approximately five seconds he Settings Success Remove message is displayed on the LCD if the settings were updated successfully Note This message only appears once If you re insert the Data Card after you have updated your settings this message is not displayed Remove the Data Card from the Data Card module Store the Data Card in its mailback envelope when not in use WARNING If your clinician has told you to use this Data Card to update the settings on your device and the Settings Success message does not appear contact your clinician immediately CLEANING AND MAINTENANCE You should regularly carry out the cleaning and maintenance described in this section Refer to your mask manual for detailed instructions DAILY CLEANING Mask Clean the mask ac
25. m configurations except when a PC is attached to the 58 device via a serial adapter in which case the system is Class A compliant Class A is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document Warnings The S8 device should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the S8 device should be observed to verify normal operation in the configuration in which it will be used The use of accessories eg humidifiers other than those specified in this manual is not recommended They may result in increased emissions or decreased immunity of the S8 device SYSTEM SPECIFICATIONS 25 26 Guidance and manufacturer s declaration electromagnetic immunity The S8 series of devices is intended for use in the electromagnetic environment specified below The customer or the user of the S8 device should assure that the device is used in such an environment Immunity test IEC60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD 6 kV contact 6 kV contact Floors should be wood concrete or ceramic tile If IEC 61000
26. n tic Emissions and Immunity on page 25 Note The manufacturer reserves the right to change these specifications without notice SYMBOLS WHICH APPEAR ON THE DEVICE A v EI Ps 1 Attention consult accompanying documents Type CF equipment Class Il equipment Drip proof Dangerous voltage Start Stop GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS AND IMMUNITY Guidance and manufacturer s declaration electromagnetic emissions The S8 series of devices is intended for use in the electromagnetic environment specified below The customer or the user of the S8 device should assure that the device is used in such an environment Electromagnetic environment Emissions test Compliance guidance RF emissions CISPR11 Group 1 The S8 device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 with serial Class B The S8 device is suitable for use in all adapter establishments including domestic establishments and those directly RF emissions CISPR 11 with USB Class B connected to the public low voltage adapter network that supplies buildings used for domestic purposes Harmonic Emissions Class A IEC 61000 3 2 Voltage Fluctuations Flicker Complies Emissions IEC 61000 3 3 Class B for all syste
27. ndition and the equipment is not damaged The S8 LIGHTWEIGHT is now ready for use with the HUMIDAIRE HOw TO ATTACH A HUMIDIFIER 10 HOW TO USE THE KEYPAD The control panel of the S8 LIGHTWEIGHT includes an LCD screen and keypad for navigating through the menus and delivering treatment LCD screen key Y Right ke Left key Down key Start Stop key LCD SCREEN The LCD screen displays the menus and treatment screens CJ KEYPAD The S8 LiGHTWEIGHT keypad has the following keys Key Function Start Stop Starts or stops treatment Co Up Allows you to increase settings options and scroll through the S8 LIGHTWEIGHT menu Down Allows you to decrease settings options and scroll through the S8 LIGHTWEIGHT menu Left Performs the function indicated by the guiding text displayed above it on the LCD screen Guiding text includes menu gj change and apply Right Performs the function indicated by the guiding text displayed above it on the LCD screen Guiding text includes exit and 2 cancel BACKLIGHTS The keypad and LCD are equipped with backlights that come on when the device is powered The LCD backlight turns off after two minutes of inactivity and comes back HOw TO USE THE KEYPAD 11 on when you press a key Once the Start Stop key has been pressed to start treatment the LCD backlight stays on for two minutes The keypad
28. not supersede instructions given by the e patient should not connect a device to the data communication port unless C b Explanation The S8 LIGHTWEIGHT is intended to be used with s connectors which have vent holes to allow continuous flow of ai mask When the device is turned on and functioning properly new air from the device flushes the exhaled air out through the mask vent holes cient fresh air wi ai n nstructed to do so by their health ca products are designed to be connected to the data communication port t in injury or damage to the S8 LIGHTWEIGHT through the mask and the exhaled air may be rebreathed Rebrea r for longer than several minutes can in some circumstances lead to suffocation This applies to most models of CPAP devices he S8 LIGHTWEIGHT can be set to deliver pressures up to 20 cm onnecting other devices could resul device he reco Ocked owever when the device is not operating insuff S8 LiGHTWEIGHT should only be used with masks and connectors mmended by ResMed or by a physician or respiratory therapist A mask should not be used unless the S8 LIGHTWEIGHT device is turned on and operating properly The vent hole or holes associated with the mask should re provider or physician Only ResMed never be pecial masks or r out of the be provided the event of power failure or machine malfunction remove the e CPAP machines
29. of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output power of transmitter 150kHz to 80MHz 80 MHz to 800 MHz 800MHz to 2 5 GHz W 1 17 NP d 0 35 d 0 7 0 01 0 17 0 04 0 07 0 1 0 37 0 11 0 22 1 1 17 0 35 0 7 10 3 69 1 11 2 21 100 11 70 3 50 7 0 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the transmitter where P is the ing of the transmitter in watts W according to the transmitter manufacturer maximum output power ra NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people INDEX A accessories 6 air filter 5 21 replacing 18 air pressure 21 air tubing 5 cleaning 17 connect to air outlet 7 kinked 21 positioning 7 aircraft 20 airflow breathing with 19 altitude setting 13 20 21 B backlight LCD and keypad 11 carry bag 5 cleaning daily cleaning 17 periodic 17 contraindications 1 CPAP 1 D Data Card 15 copying data 16 data card 22 inserting 15 mailback envelope 16 removing 16 Data Card module 15 DC 12 converter 20 down key 11 dryness mout
30. range applies NOTE 2 These guidelines may not apply in all situations Elec and reflection from structures objects and people romagnetic propagation is affected by absorption Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the S8 device is used exceeds the applicable RF compliance level above the S8 device should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the S8 device P Over the frequency range 150 kHz to 80 MHz field strengths should be less than 10 V m 27 28 Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the 58 device as recommended below according to the maximum output power
31. sories are required for its use Notes a Make sure the water chamber of the HumidAire 3i is empty before you attach or detach the humidifier b Remove the connector plug on the S8 Lightweight see section A of the illustrations sheet before you attach the HumidAire 3i Reattach the connector plug firmly after you detach the HumidAire 3i STORING THE HUMIDAIRE 3i Store the HUMIDAIRE 3i in the larger mesh top pouch ofthe S8 LIGHTWEIGHT carry bag when not in use ATTACH A HUMIDAIRE HUMIDIFIER Please refer to the illustrations in section I of the illustration sheet To connect your 58 LIGHTWEIGHT to a HUMIDAIRE humidifier you need medium sized 52 cm air tubing Fill the HUMIDAIRE water chamber and place inside the HUMIDAIRE Connect the medium 52 cm and long 2 m air tubing 1 1 and close the humidifier lid Place the 58 LIGHTWEIGHT on top of the HUMIDAIRE 1 2 and connect the free end of the medium air tubing to the air outlet 1 3 Note To avoid water spilling into the S8 LIGHTWEIGHT do not put it underneath the humidifier Assemble your mask system and connect it to the free end of the long air tubing 1 4 Plug the HUMIDAIRE power cord into a power outlet Connect the S8 LIGHTWEIGHT power cord to the socket at the rear of the S8 LIGHTWEIGHT and plug the other end into a power outlet Please refer to the HumidAire User s Manual for further details WARNING Make sure that the power cord and plug are in good co
32. spect the air filter every month to check if it is blocked by dirt or contains holes With normal use of an S8 LIGHTWEIGHT the air filter needs to be replaced every six months or more often if your device is in a dusty environment WARNING Do not wash the air filter The air filter is not washable or reusable The air filter cover protects the device in the event of accidental liquid spillage onto the device Ensure the air filter and air filter cover are fitted at all times SERVICING CAUTION Do notattempt to open the S8 Lightweight case There are no user serviceable parts inside Inspection and repair should only be performed by an authorised agent Under no circumstances should you attempt to service or repair the flow generator yourself This product S8 LiGHTWEIGHT should be inspected by an authorised ResMed Service Centre five years from the date of manufacture Prior to this the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed Applicable ResMed warranty details are provided with the device at the time of original supply Of course as with all electrical devices if any irregularity becomes apparent you should exercise caution and have the device inspected by an Authorised ResMed Service Centre If you feel that your flow generator is not performing properly see Troubleshooting on page 21 FREQUEN
33. tter fit and prolong the life of the mask cushion Do not use moisturiser when wearing a mask MY MOUTH FEELS DRY DURING TREATMENT WHAT CAN 1 DO You may experience dryness of the nose mouth and or throat during the course of treatment especially during winter In many cases a humidifier may resolve this discomfort Contact your clinician for advice MY NOSE IS IRRITATED DURING TREATMENT WHAT CAN 1 DO You may experience sneezing and or a runny or blocked nose during the first few weeks of treatment In many cases nasal irritation can be resolved with a humidifier Consult your clinician or equipment supplier for advice CAN TRAVEL WITH THE S8 LIGHTWEIGHT International use Your S8 LIGHTWEIGHT flow generator has an internal power supply that enables it to operate in other countries It will operate on power supplies of 100 240 V and 50 60 Hz No special adjustment is necessary but you may need a plug adapter for the power outlet FREQUENTLY ASKED QUESTIONS 19 20 Altitude changes If you are moving or travelling to a higher or lower altitude see How to change the Altitude setting on page 13 On an aircraft Please consult the medical services department of your carrier if you intend to use your S8 LIGHTWEIGHT on an aircraft The S8 LIGHTWEIGHT complies with US FCC Part 15 Class B requirements Note Do not use your S8 Lightweight while the aircraft is taking off or landing CAN I USE MY S8 LIGHTWEIG
34. ur bed CAUTION Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord Note You can also place your S8 Lightweight on the floor beside or underneath your bed Ensure that the area is free from dust and clear of bedding clothes or any other objects that could block the air inlet 2 Connect the power cord to the socket at the rear of the flow generator D 1 Plug the other end of the power cord into the power outlet WARNING Make sure the power cord and plug are in good condition and the equipment is not damaged 3 Connect one end of the air tubing firmly onto the air outlet D 2 WARNING Only ResMed air tubing should be used with your flow generator different type of air tubing may alter the pressure you actually receive reducing the effectiveness of your treatment Blocking the hose while in operation could lead to overheating of the device n the clinical environment any personal computer that is used with your CPAP system must be at least 5 m away from or at least 2 5 m above the patient It must also comply with the relevant test standard For personal computers the international standard is IEC or equivalent ASSEMBLE THE MASK 1 Assemble your mask system according to the mask user instructions 2 Connect your mask system to the free end of the air tubing D 3 The S8 LIGHTWEIGHT is now ready for use D 4 START TREATMENT 1 Make sure the po
35. urn the card to them The Data Card will be provided in a convenient mailback envelope Data that is copied to a Data Card is still stored and available on the S8 LIGHTWEIGHT It takes approximately five seconds to copy data onto the Data Card 1 INSERT THE DATA CARD Switch on the S8 LIGHTWEIGHT and wait until you see the standby Ramp screen 15 16 Hold the Data Card with the arrow facing up and insert it into the slot in the Data Card module H 1 Push the card in smoothly until it stops moving H 2 and messages start appearing on the LCD 2 Copy DATA ONTO THE DATA CARD Data copying starts automatically when the Data Card is inserted into the Data Card module The Card Inserted Please Wait message is displayed on the 58 LIGHTWEIGHT LCD while data is being copied Copying takes approximately five seconds he Copy Complete Remove Card message is displayed on the LCD when copying has finished Remove the Data Card from the Data Card module 3 REMOVE THE DATA CARD Grip the end of the Data Card and pull it out of the Data Card module H 3 4 STORE THE DATA CARD Your Data Card should be stored in its mailback envelope when not in use 5 MAIL THE DATA CARD TO YOUR CLINICIAN A bubble mailer envelope is provided to ensure safe transportation of the Data Card When mailing the Data Card to your clinician Insert the Data Card into the pocket provided on the mai
36. wer is on The product name is displayed briefly on the LCD screen then the standby Ramp screen appears The key and LCD backlights also turn on 2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep How TO USt THE S8 LIGHTWEIGHT CAUTION Do not leave long lengths of air tubing around the top of your bed They could twist around your head or neck while you are sleeping Make sure the area around the flow generator is dry and clean It should also be clear of bedding clothes and other potential blockages Alter the ramp time if required see How to Change the Ramp Time on page 13 To start treatment press the Start Stop key Fit your mask as described in the mask user instructions WARNING A mask should not be used unless the S8 Lightweight device is turned on and operating properly When ramping begins the LCD will display the word RAMPING with the current pressure indicated as dashes Each dash is approximately equivalent to 2 cm H5O For example 6 dashes represents approximately 12 cm pressure STOP TREATMENT To stop treatment at any time remove your mask and press the Start Stop key HOW TO ATTACH A HUMIDIFIER ATTACH A HUMIDAIRE 3i HUMIDIFIER Please refer to the illustration in section E of the illustration sheet and your HumidAire 3i manudl The HUMIDAIRE 3i attaches to the front of the S8 LIGHTWEIGHT to provide heated humidification No other acces