Home
AssayMaxTM Mouse Albumin ELISA Kit
Contents
1. use e _ MIX Diluent Concentrate 10x If crystals have formed in the concentrate mix gently until the crystals have completely dissolved Dilute the MIX Diluent Concentrate 1 10 with reagent grade water Store for up to 30 days at 2 8 C e Standard Curve Reconstitute the 100 ug of Mouse Albumin Standard with 1 ml of MIX Diluent to generate a 100 ug ml standard stock solution Allow the standard to sit for 10 minutes with gentle agitation prior to making dilutions Prepare duplicate or triplicate standard points by serially diluting the standard stock solution 100 ug ml 1 4 with MIX Diluent to generate 25 6 25 1 563 0 391 and 0 098 ug ml solutions MIX Diluent serves as the zero standard 0 pg ml Any remaining solution should be frozen at 20 C and used within 30 days Standard SEA Mouse Albumin A Dilution Point ug ml 1 part Standard 100 pg ml P1 25 00 3 parts MIX Diluent 1 part P1 3 parts MIX Diluent 6 250 Pa 1partP3 3 parts MIX Diluent 0301 Pe MXDiluent 000 e Biotinylated Mouse Albumin 2x Reconstitute Biotinylated Mouse Albumin with 4 ml MIX Diluent to produce a 2 fold solution Allow to sit for 10 minutes with gentle agitation prior to making dilution The stock solution should be further diluted 1 2 with MIX Diluent Any remaining solution should be frozen at 20 C and used within 30 days e Wash Buffer Concentrate 20x If crystals have formed in the concentrate mix gently until
2. aling Unexpectedly Low or High Signal Intensity Microplate was left unattended between steps e Each step of the procedure should be performed uninterrupted Omission of step e Consult the provided procedure for complete list of steps Steps performed in incorrect order e Consult the provided procedure for the correct order Insufficient amount of reagents added to wells e Check pipette calibration e Check pipette for proper performance Wash step was skipped e Consult the provided procedure for all wash steps Improper wash buffer e Check that the correct wash buffer is being used Improper reagent preparation e Consult reagent preparation section for the correct dilutions of all reagents Insufficient or prolonged incubation periods e Consult the provided procedure for correct incubation time e Sandwich ELISA If samples generate OD values higher than the highest standard point P1 dilute samples further and repeat the assay Non optimal sample e Competitive ELISA If samples generate OD values lower dilution than the highest standard point P1 dilute samples further and repeat the assay e User should determine the optimal dilution factor for samples Contamination of e A new tip must be used for each addition of different reagents samples or reagents during the assay procedure Contents of wells e Verify that the sealing film is firmly in
3. and set up the program in advance Wash the microplate as described above Add 50 ul of Chromogen Substrate per well and incubate for 15 minutes or till the optimal blue color density develops Gently tap plate to ensure thorough mixing and break the bubbles in the well with pipette tip Add 50 ul of Stop Solution to each well The color will change from blue to yellow Read the absorbance on a microplate reader at a wavelength of 450 nm immediately If wavelength correction is available subtract readings at 570 nm from those at 450 nm to correct optical imperfections Otherwise read the plate at 450 nm only Please note that some unstable black particles may be generated at low concentration points after stopping the reaction for about 10 minutes which will reduce the readings Data Analysis Calculate the mean value of the duplicate or triplicate readings for each standard and sample To generate a standard curve plot the graph using the standard concentrations on the x axis and the corresponding mean 450 nm absorbance OD on the y axis The best fit line can be determined by regression analysis using log log or four parameter logistic curve fit Determine the unknown sample concentration from the Standard Curve and multiply the value by the dilution factor Typical Data The typical data is provided for reference only Individual laboratory means may vary from the values listed Variations between laboratories may be caused by tec
4. ate 10x A 10 fold concentrated buffered protein base 30 ml e Wash Buffer Concentrate 20x A 20 fold concentrated buffered surfactant 30 ml e Streptavidin Peroxidase Conjugate SP Conjugate A 100 fold concentrate 80 ul e Chromogen Substrate A ready to use stabilized peroxidase chromogen substrate tetramethylbenzidine 8 ml e Stop Solution A 0 5 N hydrochloric acid to stop the chromogen substrate reaction 12 ml Storage Condition e Upon arrival immediately store components of the kit at recommended temperatures up to the expiration date e Store SP Conjugate at 20 C e Store Microplate Diluent Concentrate 10x Wash Buffer Stop Solution and Chromogen Substrate at 2 8 C e Unused microplate wells may be returned to the foil pouch with the desiccant packs and resealed May be stored for up to 30 days in a vacuum desiccator e Diluent 1x may be stored for up to 30 days at 2 8 C e _ Store Standard and Biotinylated Protein at 2 8 C before reconstituting with Diluent and at 20 C after reconstituting with Diluent Other Supplies Required e Microplate reader capable of measuring absorbance at 450 nm Pipettes 1 20 ul 20 200 ul 200 1000 ul and multiple channel e Deionized or distilled reagent grade water Sample Collection Preparation and Storage e Plasma Collect plasma using one tenth volume of 0 1 M sodium citrate as an anticoagulant Centrifuge samples at 3000 x g for 10 minutes Di
5. hnique differences Standard Point Average OD P1 25 00 P2 6 250 P3 1 563 P4 0 391 P5 0 098 P6 0 000 Sample Mouse Pooled Sodium Citrate Plasma 30000x Standard Curve e The curve is provided for illustration only A standard curve should be generated each time the assay is performed Mouse Albumin Standard Curve 10 00F g 1 00 o 2 al O 0 10 0 01 0 1 1 0 10 0 100 0 mAL ug ml Reference Value e The mouse plasma levels of albumin are 25 50 mg ml e Mouse plasma and serum samples were tested n 20 On average albumin level was 39 95 mg ml Sample Average Value mg ml Mouse Pooled Plasma 38 4 Mouse Pooled Serum Performance Characteristics 41 5 e The minimum detectable dose of mouse albumin as calculated by 2SD from the mean of a zero standard was established to be 0 05 ug ml e Intra assay precision was determined by testing replicates of three plasma samples in one assay e _ Inter assay precision was determined by testing three plasma samples in twenty assays Intra Assay Precision Inter Assay Precision Sample 1 2 3 1 2 3 n 20 20 20 20 20 20 CV Average CV Recovery Standard Added Value 0 5 15 ug ml Recovery 86 111 Average Recovery Linearity 96 e Plasma and serum samples were serially diluted to test for l
6. inearity Average Percentage of Expected Value Sample Dilution Plasma Serum 1 15000 94 93 1 30000 99 98 1 60000 105 104 Cross Reactivity Name Cross Reactivity Bovine None Human None Rat lt 10 Swine None Mouse 100 Canine None Rabbit Troubleshooting None Issue Causes Course of Action Use of expired e Check the expiration date listed before use components e Do not interchange components from different lots e Check that the correct wash buffer is being used e Check that all wells are dry after aspiration Improper wash step e Check that the microplate washer is dispensing properly e If washing by pipette check for proper pipetting c technique 2 Splashing of reagents e Pipette properly in a controlled and careful manner 8 while loading wells 2 2 e Pipette properly in a controlled and careful manner a Inconsistent volumes er i e Check pipette calibration 3 loaded into wells R o e Check pipette for proper performance Insufficient mixing of reagent dilutions e Thoroughly agitate the lyophilized components after reconstitution e Thoroughly mix dilutions Improperly sealed microplate e Check the microplate pouch for proper sealing e Check that the microplate pouch has no punctures e Check that three desiccants are inside the microplate pouch prior to se
7. lute samples 1 30000 into MIX Diluent and assay The undiluted samples can be stored at 20 C or below for up to 3 months Avoid repeated freeze thaw cycles EDTA or Heparin can also be used as an anticoagulant e Serum Samples should be collected into a serum separator tube After clot formation centrifuge samples at 3000 x g for 10 minutes and remove serum Dilute samples 1 30000 into MIX Diluent and assay The undiluted samples can be stored at 20 C or below for up to 3 months Avoid repeated freeze thaw cycles Refer to Sample Dilution Guidelines below for further instruction Guidelines for Dilutions of 1 100 or Greater for reference only please follow the insert for specific dilution suggested 1 100 1 10000 4 ul sample 396 ul buffer 100x 4 ul sample 396 ul buffer 100x 100 fold dilution 4 ul of A 396 ul buffer 100x 10000 fold dilution Assuming the needed volume is less than Assuming the needed volume is less than or equal to 400 ul or equal to 400 ul 1 1000 1 100000 4 ul sample 396 ul buffer 100x 4 ul sample 396 ul buffer 100x 24 ul of A 216 ul buffer 10x 4 ul of A 396 ul buffer 100x 1000 fold dilution 24 ul of B 216 ul buffer 10x 100000 fold dilution Assuming the needed volume is less than Assuming the needed volume is less than or equal to 240 ul or equal to 240 ul Reagent Preparation e Freshly dilute all reagents and bring all reagents to room temperature before
8. or mouse albumin has been pre coated onto a 96 well microplate with removable strips Albumin in standards and samples is competed with a biotinylated albumin sandwiched by the immobilized antibody and streptavidin peroxidase conjugate All unbound material is washed away and a peroxidase enzyme substrate is added The color development is stopped and the intensity of the color is measured Caution and Warning e This product is for Research Use Only and is Not For Use In Diagnostic Procedures Prepare all reagents working diluent buffer wash buffer standard biotinylated protein and SP conjugate as instructed prior to running the assay e Prepare all samples prior to running the assay The dilution factors for the samples are suggested in this insert However the user should determine the optimal dilution factor e Spin down the SP conjugate vial before opening and using contents e The Stop Solution is an acidic solution e The kit should not be used beyond the expiration date Reagents e Mouse Albumin Microplate A 96 well polystyrene microplate 12 strips of 8 wells coated with a polyclonal antibody against mouse albumin e Sealing Tapes Each kit contains 3 precut pressure sensitive sealing tapes that can be cut to fit the format of the individual assay e Mouse Albumin Standard Mouse albumin in a buffered protein base 100 ug lyophilized e Biotinylated Mouse Albumin 1 vial lyophilized e _ MIX Diluent Concentr
9. place before placing evaporate the assay in the incubator or at room temperature e Pipette properly in a controlled and careful manner Improper pipetting e Check pipette calibration e Check pipette for proper performance e Thoroughly agitate the lyophilized components after reconstitution Deficient Standard Curve Fit Insufficient mixing of reagent dilutions e Thoroughly mix dilutions Reference 1 Gekle M 2004 Ann Rev Physiol Version 8 3R Related Products e EA2201 1 AssayMax Human Albumin ELISA Kit Plasma and Serum samples EA3201 1 AssayMax Human Albumin ELISA Kit Urine Milk Saliva and Cell Culture samples EKA2201 1 AssayMax Monkey Albumin ELISA Kit Plasma Serum Urine and Cell Culture samples e EMA3201 1 AssayMax Mouse Albumin ELISA Kit Urine and Cell Culture samples e ERA2201 1 AssayMax Rat Albumin ELISA Kit Plasma and Serum samples e ERA3201 1 AssayMax Rat Albumin ELISA Kit Urine and Cell Culture samples ETA2202 1 AssayMax Rabbit Albumin ELISA Kit Plasma Serum Urine and Cell Culture samples e _ EPA3201 1 AssayMax Swine Albumin ELISA Kit Urine and Cell Culture samples e _ EPA2201 1 AssayMax Swine Albumin ELISA Kit Plasma and Serum samples www assaypro com e mail Support assaypro com
10. the crystals have completely dissolved Dilute the Wash Buffer Concentrate 1 20 with reagent grade water e SP Conjugate 100x Spin down the SP Conjugate briefly and dilute the desired amount of the conjugate 1 100 with MIX Diluent Any remaining solution should be frozen at 20 C Assay Procedure e Prepare all reagents standard solutions and samples as instructed Bring all reagents to room temperature before use The assay is performed at room temperature 20 25 C e Remove excess microplate strips from the plate frame and return them immediately to the foil pouch with desiccants inside Reseal the pouch securely to minimize exposure to water vapor and store in a vacuum desiccator e Add 25 ul of Mouse Albumin Standard or sample per well and immediately add 25 ul of Biotinylated Mouse Albumin to each well on top of the standard or sample and tap plate to mix gently Cover wells with a sealing tape and incubate for 1 hour Start the timer after the last addition e Wash five times with 200 ul of Wash Buffer manually Invert the plate each time and decant the contents hit 4 5 times on absorbent material to completely remove the liquid If using a machine wash six times with 300 ul of Wash Buffer and then invert the plate decanting the contents hit 4 5 times on absorbent material to completely remove the liquid Add 50 ul of Streptavidin Peroxidase Conjugate to each well and incubate for 30 minutes Turn on the microplate reader
11. yssarpro AssayMax Mouse Albumin ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St Charles MO 63301 T 636 447 9175 F 636 395 7419 WWW assaypro com For any questions regarding troubleshooting or performing the assay please contact our support team at support assaypro com Thank you for choosing Assaypro Assay Summary Step 1 Add 25 ul of Standard or Sample and 25 ul of Biotinylated Protein per well Incubate 1 hour Step 2 Wash then add 50 ul of SP Conjugate per well Incubate 30 minutes Step 3 Wash then add 50 ul of Chromogen Substrate per well Incubate 15 minutes Step 4 Add 50 ul of Stop Solution per well Read at 450 nm immediately Symbol Key BE Consult instructions for use Assay Template 12 11 10 Mouse Albumin ELISA Kit Catalog No EMA2201 1 Sample insert for reference use only Introduction Albumin a serum hepatic protein is the most abundant protein in serum It contributes to the maintenance of oncotic pressure as well as the transport of hydrophobic molecules 1 Principle of the Assay The AssayMax Mouse Albumin ELISA Enzyme Linked Immunosorbent Assay kit is designed for detection of mouse albumin in plasma and serum samples This assay employs a quantitative competitive enzyme immunoassay technique that measures mouse albumin in less than 2 hours A polyclonal antibody specific f
Download Pdf Manuals
Related Search
Related Contents
Samsung Galaxy Tab S2 Наръчник за потребителя(Lollipop) pdf, 2,4MB, deutsch Yaourtière avec 4 pots en verre Mode d`emploi PAR64 CX-2 RGBAW LED PAR user manual 取扱説明書 - Zebra Technologies Corporation Samsung NP370R5E User Manual (Windows 8) PRIDE DX-300 Service Manual HP 493269-001 Copyright © All rights reserved.