Home
        scanCONTROL Demo-Software 1.5 - Micro
         Contents
1.      Display settings    Scaling       Fig  3 17   Display settings  input field  Parameters in the  Display settings  input field     1   Xscaling  Use these parameters to configure the scaling of the X axis in the 2D display  The  entries in the  Min   and  Max   fields are values in mm     Min Max  Use these parameters to manually adjust the scaling for the X axis     2   Zscaling  Use these parameters to configure the scaling of the Z axis in the 2D display  The  entries in the  Min   and  Max   fields are values in mm     Min Max  Use these parameters to manually adjust the scaling for the Z axis   Note   Display settings  is available both in online mode and in offline mode     Note  The  Display image data  measuring program does not include the function  Display settings      3 12 Toolbar Options    In each of the individual measuring programs a toolbar is located top right to the 2D display  This provides  the possibilities of scaling the 2D display  activating various options for the mouse interaction and setting  range selections and thresholds directly with the mouse  You can also display extended information of a  selected profile point  Different selection options are available depending on the active measuring  program  A list of all the possible settings is shown in the following     Reset the 2D display to the complete measuring field  see Chap  3 12 1      Activate automatic scaling of the 2D display  see Chap  3 12 2    1 1  Keep aspect ratio  see Chap  3 1
2.     Page 42       Functions of the Measuring Programs       3 Result display  The currently measured profile and the calculated line fits are shown in this display     4 Result status display      Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power      Exposure time  ms   The currently used exposure time     Angle 1  degree   This shows the currently measured value for the angle between the line fit in  range 1 and the positive X axis      Angle 2  degree   This shows the currently measured value for the angle between the line fit in  range 2 and the positive X axis      Difference  degree   Here the program calculates the difference between the angles in range 2  and range 1     Page 43    Functions of the Measuring Programs    4 5 Circle    This measuring program is used to fit a circle to profile points in a range marked by two range lines  The  center point and the diameter of the circle in relation to the X axis and the Z axis are computed        1      4 ICONNECT   scanCONTROL Demo Software                 File Measuring programs Evaluation Default settings      Profile  Measuring field   huge      0 Range ttes EEOC ree rre nne ree rre nr nnn ren nre nn nnn ne nn nn      coves erates  ua 0  m nn Ss  ERAT    Sensor Laser   Exposure time
3.    In offline mode you will see the  File  box instead of the  Scanner settings  box  You can use this to open  and play back a saved profile sequence  The  General  part in the upper section of the box displays  information about the last played back profile sequence     General    Mame   Type   SH     profile  LLTZ800 25  004   19 D2  6050141    File Format  full set      changed        04 02 2008 17 35 26    Invert signa    in    direction   inz direction 4    Show scanner settings         Abb  3 13   File  input field    Parameters in the  File  input field     1    N    Oo    Name  The name of the profile sequence   Type  The type of measuring system used for recording the profile sequence   SN  The serial number of the measurement system used for recording the profile sequence   File format  This indicates which information is contained in the saved profile sequence     x z only  The X and Z coordinates of the individual measurement points are saved     x z   data  A complete reflection is saved in the profile sequence  In a reflection the  following information is contained for each measurement point  X Z coordinates  width   intensity  threshold  moment of Oth and 1st order     full set  All four reflections  the time stamp and  where applicable  the measurement  system parameters used during saving   Changed  The modification date of the profile sequence     Path  In this input field you specify the complete path of the file to be played back  Alternatively   you can
4.   Deactivate this parameter to recalculate the X  and Z coordinates only  Activate this parameter to recalculate all data  X   Z coordinates  width   intensity  threshold  moment O   and 1   order   This parameter has no influence on the 2D display     Confirm your settings with the  OK  button     Note  Activate the    Resample all info    parameter to store a profile sequence  see Chapter 4 2  and  analyze it later with the    scanCONTROL 3D View  software     Note  To be able to use the filter functions  scanCONTROL requires a firmware version  gt   17  Check the  firmware version on the status line if necessary  see Chap  3 4 and 5      Note  The  Filter settings  dialog box is only available in online mode     Page 23    Working with the scanCONTROL Demo Software       3 8 Advanced Scanner Settings    The  Advanced scanner settings  dialog box offers you extended settings for the scanCONTROL    measuring system     Advanced scanner settings E    Scanner    Type  LLT2810 100 004 719 D1 SH  7030022 Next scanner      Processing    vw Suppress backlight  v Video Filter  vr High resolution     Calibration    Head and tail    Sensor    Laser power   standard    Trigger   internal     Shutter alignment   center      Delay shutter    Switch off laser   i short circuit          Peak    Min  width  Max  width  1023  Min  intensity  Max  intensity  1023  Analog aut   Frequency  kHz   io      v Hold last valid point  Voltage range X    10      Invertx   Voltage range Z    10    W In
5.   Page 14    Working with the scanCONTROL Demo Software    3  Working with the scanCONTROL Demo Software  3 1 Starting the scanCONTROL Demo Software       ES Once the installation of the software and the driver is completed you can start the software  To do  this  use the program shortcut on your desktop or select  scanCONTROL Demo Software  in the  Startup menu     Once you have started the program  the main window for starting the measuring programs will be  displayed     3 2 Elements in the Main Window       MICRO EPSILON  2D 3D SYSTEMS    MICRO EPSILON MESSTECHNIK GmbH  amp  Co  KG    K  nigbacher Strasse 15  D 94496 Ortenburg  Tel   0 85 42 1 68 0  Fax 0 85 42 1 68 90   www  micro epsilon com    Change mode    Online help                    ary  Fig  3 1  Main window 2 3 4          Main window    1 Menu bar  Used to call up all the measuring programs and settings which are available in the software   2 Options display  Used to launch the individual measuring programs    3 Documentation  Used to call up individual documentation    4 Online offline mode  Used to change between online and offline mode  The current mode is displayed    in the field with the blue background     Note  The elements in the main window offer access to the same functions as the buttons in the menu  bar     Note  In offline mode  the functions  Save profile  and  Display image data  are deactivated     Page 15    Working with the scanCONTROL Demo Software       Note       Never disconnect the IEEE1
6.   Time diagram  Diagram   When the time diagram is activated you can use this display to see the  time sequence of a sequence of profiles  The vertical axis represents the index of the measurement  points along the profile and is scaled in accordance with the parameter  Points per profile  which is  set under  Scanner settings   The individual profiles are entered on the horizontal axis as a function  of time  You can adjust the scaling for this axis in the settings under  Time diagram   The distance Z  to the sensor is symbolized using the colors shown in the key on the right     Result status display      Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power      Exposure time  ms   The currently used exposure time     No  of valid points  The number of valid points which have been detected for the current profile      Saturation      The saturation of the signal which is selected with the area marked in red  This  can be adjusted as required by changing the values for  exposure time  and  threshold   We  recommend a saturation value between 60  and 80      Note  The program saves the profiles in the SYLK format    sik   This format can be imported into  Microsoft Excel  One profile is saved in each file  whereby the file names are automatica
7.   ms     200  m      auto  i   pem   No  of profiles  1 5     25        Threshold  56  zen   Reflections   first sf      Invert signal    in X direction  v in Z direction 870  Filter      Advanced         83 0  Display settings    Scaling    X  Min    18 321 ax   3 152 781   30 4  21 2  124 x  3 0 6 1 15 2  Zi  Min    79 104 ax   98 738 X position  mm     Scan rate  1 s   Exposure time  ms   X center  mm    Z center  mm        98 7        94 8    80 8       Diameter  mm              Abb  4 5  Measuring program  Circle   Input fields and associated parameters in the measuring program  Circle      A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information on each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      B 2D display  This display visualizes the profile which has just been measured     C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings      Status bar  The status bar contains status messages and error messages     E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   3 12    1 Range  Use the two sliders to adjust the red lines wh
8.   see the  scanCONTROL2800 2810 and scanCONTROL2700 2710 instruction manual   depending on the type of  sensor     scanCONTROL 2700   scanCONTROL 2710     small  Measuring field index 7  640 x 120 pixels    standard  Measuring field index 2  640 x 360 pixels     large  Measuring field index 0  640 x 480 pixels    custom     You can select the measuring field index  0   7     scanCONTROL 2800   scanCONTROL 2810     small  Measuring field index 95  256 x 256 pixels    standard  Measuring field index 50  512 x 768 pixels    large  Measuring field index 18  768 x 768 pixels    huge  Measuring field index 0  1024 x 1024 pixels    custom     You can select the measuring field index  0   95   With this entry you specify the resolution in the direction of the X and the Z axis  The value for the X axis is  also influenced by the  Points per profile  box     Page 20    Working with the scanCONTROL Demo Software       2 Points per profile  This parameter defines the number of points that make up a profile  Depending on  the type of sensor you can choose between the following values     scanCONTROL 2700   scanCONTROL 2710     80    160    320    640    scanCONTROL 2800   scanCONTROL 2810     64  optional     128  optional     256    512    1024  With this parameter you specify the resolution in the direction of the X axis  This value has priority over the  entry in the  Measuring field  box but does not have any influence on the resolution in the direction of the  Z axis     3 Exposure t
9.  Displaying scanCONTROL Parameters in Offline Mode    When playing back saved profile sequences  you have the possibility of displaying the scanCONTROL  settings used during saving      2 Press the button    Show scanner settings     in the    File    box  see Chap  3 9   You then access the     Scanner settings  dialog box     In the dialog box you can choose between the following tabs     Scanner settings  The standard settings of the measurement system are displayed  You will find a  precise description of the individual parameters in Chap  3 6     Advanced settings  The extended settings of scanCONTROL are displayed  You will find a precise  description of the individual parameters in Chap  3 8     Filter settings  The filter settings of scanCONTROL are displayed  You will find a precise description of  the individual parameters in Chap  3 7     Note  The following software packages save the scanCONTROL parameters     scanCONTROL Demo Software  version 1 2 and above     scanCONTROL2810 Sensor Edition    scanCONTROL Setup Software 2X10    DeveloperDemo  version 2 0 and above  option must be activated      Note  The information used during saving is only available in the file format  full set      Page 28    Working with the scanCONTROL Demo Software       Scanner settings      Scanner settings   advanced settings   Filter settings      General    Type  LLT2800 25  004 v17 C2      Profile       Measuring Field  large  18   Points per profile  256    Sensor Laser  Exposure t
10.  Filter    Median  3 taps m Average  Is taps E   3    Resulting partmeters  Resample range      5 000mm           Resample resplution  0 016mm  Median range 0 047mmi  Average range  O0  0 78mm    Fig  3 11   Filter settings  dialog box  Parameters in the  Filter settings  dialog box    1 Range  The range which is equidistantly resampled      mne   deactivated   deactivated         2 Median  With this parameter you activate a median filter     none  The median filter is not used     3 taps  You use a median filter with a filter size of three     5 taps  You use a median filter with a filter size of five     7 taps  You use a median filter with a filter size of seven     3 Average  With this parameter you activate an average filter     none  The average filter is not used     3 taps  You use an average filter with a filter size of three     5 taps  You use an average filter with a filter size of five     T taps  You use an average filter with a filter size of seven     Page 22    Working with the scanCONTROL Demo Software       4 Interpolate invalid points  If equidistantly resampling is active  this specifies whether invalid points will  be replaced by neighbouring valid points using linear interpolation  Gaps will be closed if this parameter  is active  In order to leave gaps open you must not activate this parameter     Resample all info  If equidistantly resampling is active  this parameter specifies which data will be  recalculated according to the equidistantly resampling
11.  Hardware Wizard    Welcome to the Found New  Hardware Wizard    Windows will search For current and updated software by  looking on your computer  on the hardware installation CD  or an  the Windows Update Web site  with pour permission      Head aur privacy police    Can Windows connect to Windows Update to search for  software      Yes  this time only    ad every time   connect a device    Click Next to continue     Fig  2 11  Windows XP SP2 driver installation    If the Service Pack 2 is installed on your PC  then the  Hardware Update Wizard  will appear        E32 Select  No  not this time  and click on  Next  to confirm  see Fig  2 11    Found Mew Hardware Wizard  Bees  S    This wizard helps vau install software Far     CMU 1334 Digital Camera Device    9 If pour hardware came with an installation CD   lt   nr floppy disk  insert it now     What do vou want the wizard to do     CO Install the software automatically  Becommended      Click Next to continue        Fig  2 12  Windows XP driver installation   step 1    243 Select  Install from a list or specific location  advanced     fa Click on  Next  to confirm this dialog  see Fig  2 12      Page 12    Installation and Preparations for Measuring       Found Mew Hardware Wizard    Please choose your search and installation options        CO Search for the best driver in these locations     Use the check boxes below to limit or expand the default search  which includes local  paths and removable media  The best driver fo
12.  counter is to count positive edges  negative edges or both types of edges     Enable CMM trigger  Select this box to activate the programmable trigger output of the  scanCONTROL  The programmable trigger output is only available as an option for the  scanCONTROL   Invert trigger output  With this parameter you invert the trigger output   Trigger output  The port at which the trigger signal is issued      Port 1   Port 3     Port 2   Port 4  Trigger divisor  This gives the division ratio between the trigger signals and the profiles  Select  a value  n  to issue one trigger signal for every nth profile      Min  1     Max  255  Mark space ratio  This gives the mark space ratio for the decreasing edge      1 4   2 1      1 3   3 1    1 2   4 1    1 1    Skew correction  ys   This parameter provides a time correction of the trigger signal in  increments of 0 5 us      Minimum   256 0 us     Maximum  255 5 us    Confirm your settings with the  OK  button     Note  The parameter  Switch off laser  and the input field  Analog out  are only available using the  scanCONTROL 2800 or scanCONTROL 2810 type of sensor     Note  For further information on the parameters for the measuring system described here  please refer to  the documents  Quick reference scanCONTROL    and  Manual scanCONTROL  Part B    see Chap  3 3      Note  The  Advanced scanner settings  dialog box is only available in online mode     Page 26    Working with the scanCONTROL Demo Software       3 9    File Settings 
13.  define the tolerance limits of your measurement in  relation to the determined reference line  Enter a value in each of the input fields  Upper limit  and     Lower limit   The program displays the tolerance limits as red lines in the results display     Result display  The results display shows the currently measured profile  the reference line  green  line  and the selected tolerance limits  red lines   The following color coding is used for detected  points      Green  points within the tolerance range      Red  points above the upper limit of the tolerance range      Blue  points below the lower limit of the tolerance range     Result status display      Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power      Exposure time  ms   The currently used exposure time     Prop  of points      At this point the program indicates the proportion of points as a percentage  which lie below  within and above the tolerance limits      No  of points  At this point the program indicates the number of points which lie below  within  and above the tolerance limits      Minimum  Here the program shows the X Z coordinates and the distance  height  of the point  with the greatest negative distance to the reference line      Maximum  Here the program shows t
14.  driver installation   step 6    E32 Click on  Finish  in the next dialog  see Fig  2 10    The scanCONTROL driver is now installed     Note  Only use version 6 3 of the CMU driver  If a different driver or a different version of the CMU driver  for the scanCONTROL is installed at a later date then it will not be possible to operate the scanner with  the scanCONTROL Demo Software     Note  To install the driver at a later date  call up the installation dialog via the Device Manager  Control  Panel gt  System  gt  Hardware  gt  Device Manager   scanCONTROL is located under    Imaging devices  or  under    Other devices        Note  If you are requested to choose the driver files manually you will find the files in the installation  directory of scanCONTROL Demo Software   Installation path  Program Driver CMU1394    Page 11    Installation and Preparations for Measuring       2 4 Installation of the Drivers under Windows XP  Preparing the installation    E23 Completely finish the installation of the scanCONTROL Demo Software  This procedure is described  in Chap  2 2    E32 Connect the scanCONTROL via the IEEE1394 interface to the PC    43 Switch on the scanCONTROL and the power supply     The installation dialog will then appear on the screen      If no dialog appears  you can also call up the installation dialog via the Device Manager  Control Panel  System  gt  Hardware  gt  Device Manager   scanCONTROL is located under  Imaging Devices  or  Other  Devices         Found New
15.  first three measured peaks  The  peak with the smallest X coordinate counts as the first peak    Height  mm   Here  the program calculates the distances between the first three peaks and the  reference line     Page 49    Functions of the Measuring Programs       4 8 Flatness    This measuring program evaluates the flatness of a profile  To do this  it calculates how many points in the  profile lie within a user specified tolerance  The program fits a line in an adjustable range to do this  You  can define an upper and lower tolerance limit in relation to this line  The measuring program then  determines how many points in the current profile lie above  below and within this tolerance  as well as the  positions of the highest and lowest points       4 ICONNECT   scanCONTROL Demo Software                File Measuring programs Evaluation Default settings      Flatness    Profile   Measuring field   huge     o Bano ee NDR BESTREBEN EISEN EEE EEE REST ER     Point File  4   eT al ef  E  afr    Sensor Laser  Exposure time  ms    1 00      auto            u  No  of profiles  1 5     25 d    Threshold    128    Reflections   largest area       Invert signal    in x direction   in 2 direction  55  Filter      Advanced       CiT    Display settings    X  Min    18 566 nr  5 366  Z  Min   74 585 xi  78 573    Flatness settings    Scanner settings       81 3              X position  mm        Scan rate  1 s   Exposure time  ms      lt  Min  Valid  gt  Max   Prop  of points  96   No  o
16.  original signal recorded on the image  matrix of the scanCONTROL  This measuring program is useful for detecting and eliminating sources of  errors and interference  such as overexposure or multiple reflections         4 ICONNECT   scanCONTROL Demo Software l    cil x        File Measuring programs Evaluation Default settings      Scanner settings scanCONTROL       Display image data UD MEROSESILON          Measuring field   arae       55    Points per profile  256                 Sensor Laser       Exposure time  ms    0 20      auto  i     25    No  of profiles  1 5     ES        20    Threshold   A  128    Reflections           r Invert signal       F in   direction   in z direction          Filter      Advanced       Max  no  of profiles  1 s             Scan rate  1 s     1 2          Fig  4 3  Measuring program  Display image data     Input fields and associated parameters in the measuring program  Display image data      A    Scanner settings  With these values you can change the settings of scanCONTROL  Detailed  information on each of the parameters can be found in Chap  3 6  Scanner settings   Only the    values  Measuring field    Points per profile  and  Exposure time  are available in the  Display image  data  measuring program     2D display  This display visualizes the image data which has just been measured by the CMOS  array  The measuring field selected with the  Measuring field  parameter  see Chapter 3 6  is shown  on the display as a red rectangle     St
17.  s  Exposure time  ms    Saturation            2 3 4    Fig  4 2  Measuring program  Save profile     Input fields and associated parameters in the measuring program  Save profile      A    B    Scanner settings  With these values you can change the settings of scanCONTROL  Detailed  information on each of the parameters can be found in Chap  3 6  Scanner settings      2D display  This display visualizes the profile which has just been measured     Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings      Status bar  The status bar contains status messages and error messages     Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   3 12     Freeze 2D display Trigger    Freeze 2D display  Deactivates the 2D display  When the display is deactivated the  performance of the system rises  allowing profiles to be saved at a higher scan rate  This  option is automatically deactivated when starting the measuring program     Trigger mode  Activates the trigger mode  see also section  Record      Page 38    Functions of the Measuring Programs       2    Record    Trigger mode  not active  Press the red Record button to start recording the profile sequences   Now the profiles are stored in the RAM  The program will 
18.  use the button               This opens a dialog box for selecting the profile file to be played back  After choosing the  file  the entry field  Path  is updated     Play  This starts the playback of the selected profile sequence    Stop  This stops the playback of the selected profile sequence    Current profile  While a profile sequence is being played you will see a run through of the  profile numbers in this box  If you stop the playback you can activate any individual profile  directly by entering the profile number     Invert signal in X direction  The signal is inverted on the Z axis   Invert signal in Z direction  The signal is inverted in the middle of the measurement range  parallel to the X axis     Show scanner settings  You press this button to display the scanCONTROL settings selected  when the profile sequence was saved  see Chap  3 10      Note  The following software packages save the measurement system parameters     scanCONTROL Demo Software  version 1 2 and above     scanCONTROL2810 Sensor Edition    scanCONTROL Setup Software 2X10    DeveloperDemo  version 2 0 and above  option must be activated     Page 27    Working with the scanCONTROL Demo Software       Note  You can only play back  avi  files which have been recorded using the following programs     scanCONTROLDemo Software    scanCONTROL2810 Sensor Edition    scanCONTROL Setup Software 2X10    DeveloperDemo  version 2 0 and above     Note  The  File  box is only available in offline mode     3 10
19. 2 3    al Activate    Zoom    mode  see Chap  3 12 4    E  Activate  Move profile  mode  see Chap  3 12 5    vA Activate    Display extended information of a profile point    mode  see Chap  3 12 6   H  Activate    Set vertical range  red     mode  see Chap  3 12 7    H  Activate    Set vertical range  blue   mode  see Chap  3 12 7    E Activate  Set range  mode  see Chap  3 12 7      Activate    Set horizontal range    mode  see Chap  3 12 7         Activate    Set horizontal threshold    mode  see Chap  3 12 8     You will find detailed descriptions of the individual functions in the following     Page 31    Working with the scanCONTROL Demo Software       3 12 1 Reset 2D Display    Eg     Press this button in order to reset the 2D display  Thus  the complete measuring field is shown and the  profile transferred by scanCONTROL is displayed     3 12 2 Automatic Scaling of the 2D Display          You activate or deactivate the automatic scaling of the 2D display by pressing this button  If automatic  scaling is activated  the 2D display will be automatically adapted to the smallest and largest measured  values of the scanner  If automatic scaling is deactivated  the specified values from the    Min     and    Max      input fields  see Chapter 3 11  will be applied to the scaling of the 2D display     3 12 3 Keep Aspect Ratio    aa    Press this button in order to activate or deactivate the    Keep aspect ratio    option  Activating this option the  display is scaled so that 
20. 394 connection between the scanCONTROL and the computer while  the software is running      Never switch off the power supply to the scanCONTROL while the software is running      Never press the reset user mode button on the controller while the software is running      Never activate the standby mode or hibernation of your computer when the measuring system is  connected to the computer     This may cause the operating system to shut down unintentionally     3 3 Menu Bar  The following functions can be called up from both the menu bar and via the main window       Starting a measuring program    Changing between online mode and offline mode    Calling up the entire documentation    The  Default settings  function which is used to reset the measuring programs back to the default settings  is only accessible via the menu bar     Measuring programs Default settings  1 2 3 4 5    Fig  3 2  Menu bar    Note  You can use keyboard shortcuts to speed up your work  The key combinations are displayed on the  menu bar next to the corresponding function  e g   lt Ctrl   Alt   F12 gt  for  Exit   A listing of all keyboard  shortcuts can be found in Chap  7    Menu bar functions     1 File        File   Measuring programs Evaluation Default settings          Exit Strgtaltt Fi2       Fig  3 3  Menu    File       This menu item only contains the  Exit  function  This function is used to close the scanCONTROL Demo   Software     2 Measuring programs       File   Measuring programs   Evaluat
21. 99 5 F k   V in X direction IV in 2 direction z u  Filter      Advanced      dud u i A      Display settings   settings 94 8    2 03 23 a 56 88 12 0       X position  mm     Scaling    x  Min    4 105 EC  12 032    x   106 465 465       2  Min   94 775   775    scan rate  1 s    Exp  time  ms   Number of points   Height  mm    Width  mm    Area  mm      Profile settings    Properties    Type   maximum m      Left position  mm   Maximum position  mm   Right position  mm           3 4 5  Fig  4 9  Measuring program  Profile     Input fields and associated parameters in the measuring program  Profile      Min  height  mm     1 0  Min  width  points     5  Reference line   range 1 2                 A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information on each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      B 2D display  This display visualizes the profile which has just been measured     C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings      Status bar  The status bar contains status messages and error messages     E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detail
22. Found New Hardware Wizard    Select a Device Driver  Which driver do you want to install for this device     Found New Hardware Wizard      Start Device Driver Installation  The device driver will be installed with the default settings        Fig  2 8  Windows 2000 driver installation   step 4  Z3 Click on  Next  to confirm the next dialog  see Fig  2 8      The dialog  Digital signature not found  will then appear     Page 10    Installation and Preparations for Measuring       Digital Signature Not Found   x        The Microsoft digital signature affirme that software has  been tested with Windows and that the software has not  been altered since it was tested     The software vou are about to install does not contain a  Microsoft digital signature  Therefore  there rs no  guarantee that this software works corectly with  Windows     CMU 1394 Digital Camera Device    IF pou want to search Far Microsoft digitally signed  software  visit the Windows Update Web site at  http  windowsupdate  microsoft com to see if one is  available        Do vou want to continue the installation       b Ho   More Info      Fig  2 9  Windows 2000 driver installation   step 5       CA Click on  Yes  to confirm this dialog  see Fig  2 9      Found New Hardware Wizard       Completing the Found New  Hardware Wizard    CMU 1334 Digital Camera Device       Windows has finished installing the software for this device     To close this wizard  click Finish        Fancel         Fig  2 10  Windows 2000
23. MECT       Online help Fl  Manual scant ONTROL Demo Software       Documentation    Fig  3 7  Menu            This function is used to call up individual documents  If you click on the menu option  Online help  while a  measuring program is open then the documentation for this measuring program will appear     Page 17    Working with the scanCONTROL Demo Software       3 4 Online Mode    In online mode  the scanCONTROL Demo Software evaluates data currently being measured by  scanCONTROL   1 2 5      4 YCONNECT   scanCON  ROL Demo Software    scanCONTROL Profile    Profile    Measuringfield  huge z   o manga ee zm    Pon pen proie   os     xl Z distance n AOR MC LR LE te EXT 11 Q m    Sensor Laser 79 4  Exposure time  ms    1 00      auto    j   I 75 7  No  of profiles  1 5     25        Threshold    128    Reflections   largest area       Invert signal 714     v in X direction  v in Z direction  Filter    Advanced      n  Display settings    Max   22 161  Zi Min   56 1 Max   79 386  scan rate  1 s    Exp  time  ms     Profile settings Number of points    Properties Height  mm    Width  mm   Area  mm                 scanner settings            74 1          Type   Min  height  mm   i i Left position  mm    TE Maximum position  mm   Min  width  points    22   13  4    i Right position  mm     Reference line     Fig  3 8  Online mode  Online mode     The illustration shows the typical structure of a measuring program in online mode  as well as the shared  input and output field
24. ONTROL is a synonym for scanCONTROL2800  scanCONTROL2810  scanCONTROL2700  and scanCONTROL2710 measuring systems     1 1 Use of the Manual    The user manual contains general information on installation and use of the scanCONTROL Demo   Software     Users of the scanCONTROL Demo Software should read Chap  2   Installation and preparations for  measuring    Chap  3   Working with the scanCONTROL Demo Software   explains the use of the basic  functions in the measuring programs  In Chap  4   Functions of the measuring programs   you will find  user information relating to each of the measuring programs  with detailed explanations of how to use  these programs     Note  Please refer to the contact address printed on the inside cover for any issues relating to queries and  support or any other technical information     1 2 Functional Scope of the scanCONTROL Demo Software    The scanCONTROL Demo Software offers a simple way of transferring profile data from the  scanCONTROL measurement system and presenting them in a graphical representation  These profile  data are then further processed and evaluated in measuring programs  All data are transferred via the  IEEE1394 interface  FireWire  iLink  and can be saved as required using the scanCONTROL Demo   Software     1 3 Definition of Terms  Profile  Profile Sequence    The terms  profile  and  profile sequence  are used repeatedly in this manual in the context of the  scanCONTROL Demo Software  and are defined as follows  A  profil
25. PC   C4 Switch on the scanCONTROL and the power supply     The installation dialog will then appear on the screen  see Fig  2 5       If no dialog appears  you can also call up the installation dialog via the Device Manager  Control Panel  System  gt  Hardware  gt  Device Manager   scanCONTROL is located under  Imaging Devices  or  Other  Devices       Found New Hardware Wizard           Welcome to the Found New  Hardware Wizard    This wizard helps vou install a device driver for a  hardware device     To continue  click Next     Fig  2 5  Windows 2000 driver installation   step 1  CA Click on  Next  to confirm this dialog  see Fig  2 5      Install Hardware Device Drivers  A device driver i a software program that enables a hardware device to work with  an operating system        This wizard will complete the installation For this device     CMU 1334 Digital Camera Device          device driver is a software program that makes a hardware device work  Windows  needs driver files For your new device  To locate driver files and complete the  installation click  M ext     What do vau want the wizard ta do        Search for a suitable dri        lt  Back d Nee D L ancel         Fig  2 6  Windows 2000 driver installation   step 2    E32 Inthe next dialog  select  Display a list of the known drivers for this device so that   can choose a  specific driver    CA Click on    Next    to confirm this dialog  see Fig  2 6      Page 9    Installation and Preparations for Measuring       
26. SENSORS  amp  SYSTEMS    By far the best solution       MICRO EPSILON    Evaluation Defauk settings 2            eg standard Zj ja       Exposure tine  ms    2 00  gt     auto     ta E 24                   scan rate  1 s     Edge 1 Edge 2 Edge 3    Data transmission  evaluation and  visualization    scanCONTROL  Demo Software 1 5    MICRO EPSILON  MESSTECHNIK  GmbH  amp  Co  KG   K  nigbacher Strasse 15    D 94496 Ortenburg    Tel   49 85 42 1 68 0  Fax  49 85 42 1 68 90  e mail  info micro epsilon de  www micro epsilon com    Certified in compliance with  DIN EN ISO 9001  2000       X9751130 A070108ALI       wN    Contents    INLTOGUC EI ON aa aaa a na ae na Le E E EREE 4  Use or tie Ma races sso eee LI el 4  Functional Scope of the scanCONTROL Demo Software                cccceececeseeceeeeeeeeeeeetaneeteneeneess 4  Definition of Terms  Profile  Profile Sequence                      2 20002220002sn0ennnnnnnennnenennennnnennnnnenennnenn 4  Functions of the Measuring Programs                    u00222420000000000nennnn nennen nennen nnn nnne nnn nnns nn nnn 5  Installation and Preparations for Measuring                  uuuusunu00nnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnn 6  L nstallauonRequirements   uice e tac vs uU MEM UU MOI a MIL IM IDA MIT Mee 6  Installation of the scanCONTROL Demo Software              ccccccccecseeeeeeceeeeeesaeeeeeeeeeeeeeesaneeeeaeeees 6  Installation of the Drivers under Windows 2000               cccccccseeeeeeeeeeeeeceeeeeesaeeeeeeeaeeeeesaaeeeee
27. Scan rate  1 s   Exposure time  ms   Angle 1  degree   Angle 2  degree   Difference  degree      D           Abb  4 4  Measuring program  Angle   Input fields and associated parameters in the measuring program  Angle      A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information on each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      B 2D display  This display visualizes the profile which has just been measured    C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings     D Status bar  The status bar contains status messages and error messages    E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   3 12    1 Range 1  Use the two sliders to adjust the red lines which delimit range 1 in the 2D display  or input    the values  You can also set the range limits in the 2D display directly using the mouse  see  Chapter 3 12      2 Range 2  Use the two sliders to adjust the blue lines which delimit range 2 in the 2D display  or  input the values  You can also set the range limits in the 2D display directly using the mouse  see  Chapter 3 12  
28. The gradient of an edge is calculated from the intersecting points between the profile and  the edges of the noise range     The result display indicates all detected edges with a red cross  The measured values for the first three  edges are continuously updated in the result display     1 2 Profile runs through noise range         4 ICONNECT   scanCONTROL Demo Software    OO   Oumlm o i  ARENA    Scanner settings scanCONTROL eec    r Profile  Measuring field   nue 7  Ranga  ttm nn nnn na d nn n nn nn VN I a  Points er rofile  hos   1 BET OS heat le ST ef T T UT o T T Wat DT d US T I  BEI IT T Jr 0 DOT BEL Bot T T Bir  UT  T IST bey Vb  UT ef I DET Y o Ha US e er    OL D AC JT 9 d T SL  Ber var JO Ce e d en Der lat le  E y  LECT  CT ter er Ta Der aa et GT  var Jap VES Bel S 8 Je 9 eg E Teh OE IST Ta    L_ s N N Z distance  mm       z     4      Sensor Laser 88 0    Exposure time  ms     2 00       auto  1   1  No  of profiles  1 5     25     874 Ja  66 9  i AN    85 8                    1  Threshold  128        Reflections   largest area       r Invert signal    in x direction  v in Z direction  Filter      Advanced         Display settings    Scaling    X  Min    12 881 ES   0 389  2  Min   85 281 x  37 967    T  Threshold Value   ss  Scan rate  1 s   N  Noise Range Exposure time  ms   Na  of edges                79  5 4  X position  mm        Edge   Edge2  Edge3  X position  mm   Gradient     D        Abb  4 6  Measuring program  Edge   Input fields and associated parameter
29. am  to select another scanner     Open the dialog box  Advanced scanner  The currently operated IEEE settings  and press the button  Next  to  Invalid device 1394 device is not a select a scanCONTROL or close  scanCONTROL  scanCONTROL Demo Software and  connect the scanCONTROL to the PC   Check the IEEE1394 connection between  The data transfer between scanCONTROL and the PC  Close          Data transfer interrupted   scanCONTROL and the PC has   scanCONTROL Demo Software   been interrupted  reconnect scanCONTROL to the PC and  restart the software     The selected value for the  parameter  No  of profiles  is too  large  Please refer to Quick  reference scanCONTROL to get  more informations     l The selected scanner settings Type in consisting values for the  Inconsistent  Exposure     RAE   M due  Exposure time  and  No  of parameters  Exposure time  and  No  of  time   No  of profiles Ple a  profiles  are mutually exclusive    profiles      Inconsistent  No  of  profiles    Decrease the value for the parameter  No   of profiles        Page 54    Status and Error messages       The selected scanner settings     Measuring field      Points per  Inconsistent  Meas  field   profile  and    No  of profiles  are   Type in consisting values for the    Points p  profile   No  mutually exclusive  Please refer   parameters    Measuring field      Points per    of profiles to Quick reference profile  and    No  of profiles    scanCONTROL to get more  informations     Comuptoroile Th
30. ase check that the scanCONTROL  measuring system is not connected to the computer      a Click on the  Installation  button  see Fig  2 2  to start installing the software     An input field opens in which you are required to select the software packages to be installed     Installation scanCONTROL software packages x     Please choose the components you want to install  an Your system                     Provides comfortable visualization and several analyses of  profile data        scanCONTROL 30 Tools     scanCONTROL 3D View 1 2    A demo program for comfortable 3D visualization of profile  data      scanCONTROL 3D Data 1 2  Provides comfortable 3D yisualization of profile data and    several standard formats to export 3D data  licensing  required        scanCONTROL Development 1 5  Develop your own signalgraphs Far scan MN TRUL     Vv Disable default windows driver For IEEE 1334 imaging devices   useful Far easy installation of the scanL  NTROL driver     caen      Fig  2 3  Installation scanCONTROL software packages  dialog box    E Activate the option  scanCONTROL Demo Software 1 5  in the  Installation scanCONTROL  software packages  dialog box  see Fig  2 3      4 Activate the    Disable default windows driver for IEEE1394 imaging devices  option to make the  scanCONTROL driver installation easier  Windows XP only  refer to Chapter 2 4         Optionally you can install scanCONTROL Development  This tool enables you to develop your own signal  graphs for scanCONTROL a
31. atus bar  The status bar contains status messages and error messages     Max  no  of profiles  1 s      With this button you open the dialog  Max  no of profiles  1 s   where  you can determine the maximum possible profile frequency depending on the currently selected  scanner settings  see below      Result status display     Scan rate  1 s   This value shows the effective scan rate being used by the program to evaluate  the scanner data     Page 40    Functions of the Measuring Programs             Max  no  of profiles  1 5  X     The maximum number of profiles per second depends on  several scanner settings and on Ehe number of sensors   operating simultaneously     Mo  of sensors  1        Setting Max  no of profiles  1 5   Measuring Field  BO0Hz   Processing  1000Hz   Number of Sensors  1600Hz   Exposure time  1000Hz   Result  BnnHz    IE          Fig  4 1  Dialog    Max  no  of profiles  1 s        The maximum possible profile frequency which scanCONTROL can be operated with depends on various  scanner settings and on the number of sensors  operating simultaneously  You can analyze the maximum  possible profile frequency in this dialog       No  of sensors  With this setting you determine the effect of the number of scanCONTROL  sensors to be operated simultaneously on the Fire Wire controller  see below       Measuring field  The maximum profile frequency resulting from the setting of the  Measuring  field  parameter      Processing  The maximum profile frequency resul
32. e   1024   Z distance  mm      EX E Zl       i    Sensor Laser 88 0    Exposure time  ms     1 00      auto  No  of profiles Lus     25    Threshold  128 rm  Reflections  flargest area        gt      Invert signal    in x direction  v in Z direction  Filter      Advanced                Display settings       Scaling  X  Min    27 419 C    25 65   2  Min   61 679 cm  37 96 bs    i   iust m Scan rate  1 5    No  of profiles  100 rm  v active     Exposure time  ms   Log settings No  of valid points    Saturation           Properties    Path         Data m   Profile index  Name    profile Trigger      active    0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100       2 3 4 5  Abb  4 1  Measuring program Display profile     Input fields and associated parameters in the measuring program  Display profile      A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information on each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      In the  Display profile  measuring program you have the possibility to transfer all measured  reflections from scanCONTROL and display a selected reflection as a profile  For this purpose  select the  all reflections  option for the  Reflections  parameter  Now select the reflection to be  displayed  Reflection    1    is the reflection with the smallest distance to the sensor and reflec
33. e  consists of a maximum of 1024  individual measurement points  each of which is defined by a X coordinate and a Z coordinate  The  software synchronously analyzes these measured profile data according to geometric criteria such as  angle  edge  circle or area  A time sequence of profiles is referred to as a  profile sequence      Note  More information about the measuring principle can be found in the scanCONTROL2800 2810 and  scanCONTROL2700 2710 instruction manual    Page 4    Introduction       1 4    Functions of the Measuring Programs    This list of functions is merely intended to provide a short overview  Detailed descriptions of all measuring  programs can be found in Chap  4     Display profile visualizes individual profiles and profile sequences and saves profiles in a format  which is compatible with Microsoft Excel    Save profile saves profile sequences for offline analysis    Display image data visualizes image data of the CMOS matrix of scanCONTROL    Angle measures the angle between two segments of a profile    Circle measures a segment of a circle in a profile    Edge detects edges in a profile    Peak detects local minimum maximum values in a profile    Flatness measures the flatness of a profile    Profile determines characteristic data for a significant section of a profile  e g  glue bead  weld  seam      Note  All measuring programs are available in online and offline versions  with the exception of  Save  profile  and  Display image data   The online 
34. e 2D display directly using the mouse  see  Chapter 3 12      Range  Use the two sliders to adjust the red lines which delimit the range in the 2D display  or input  the values  The program will then only detect edges within this range  You can also set the range  limits in the 2D display directly using the mouse  see Chapter 3 12      Result display  The results display shows the currently measured profile and the edges detected in  it  Edges are identified with a red cross     Result status display     Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power    Exposure time  ms   The currently used exposure time   No  of edges  At this point the program indicates the number of edges detected in the currently  measured profile    X position  mm   This display shows the position of the first three measured edges  The edge  with the smallest X coordinate counts as the first edge    Gradient  This display shows the gradient of the first three measured edges  The edge with the  smallest X coordinate counts as the first edge     Page 47    Functions of the Measuring Programs    4 7 Peak    This measuring program is used to detect local minimum maximum values within a profile  In the process   a reference line is fitted first  A minimum height in rela
35. e transmitted profile is Deactivate the filter or reset the  d damaged parameters to the default settings  5 2 In Offline Mode    Message             Descripion          TA    There is no fault present  You can use the  Play button to play back a new profile  The system is ready  sequence  If a profile sequence has  already been played back then the  individual profiles can now be displayed        Pa cae c e bena EI There is no fault present  Wait until the  Play p q 9 Play playback has finished or interrupt playback  back  i  by pressing the Stop button   Couldn t open file The specified name or path was Select the correct directory containing your  incorrect  recorded profile sequence     No points in region of No data were received from the  interest j scanCONTROL during the Repeat the recording of the profile   recording        Page 55    Notes       6  Notes    Please note the following when operating the scanCONTROL Demo Software     1    Never disconnect the IEEE1394 connection between the scanCONTROL and the computer while  the software is running    Never switch off the power supply to the scanCONTROL while the software is running    Never press the reset user mode button on the controller while the software is running    Never activate the standby mode or hibernation of your computer when the measuring system is  connected to the computer     This may cause the operating system to shut down unintentionally     Only use version 6 3 of the CMU driver  If a different d
36. easurements  This setting  minimizes the file size     x z   data  Stores the data of the selected reflection  X and Z coordinate  width   maximum  threshold  moment  order  0 and 1       full set  Stores all four reflections  the time stamp and the settings for scanCONTROL   only if  Loop back parameters  is active  see Chap  3 8   This setting enlarges the file  size     Result status display    Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power    Exposure time  ms   The currently used exposure time   Saturation  9o   The saturation of the current signal  This can be adjusted as required by  changing the values for  exposure time  and  threshold   We recommend a saturation value  between 60  and 80      Note  To store the scanCONTROL settings in the profile file  select the file format  full set  and activate     Loop back parameters  in the advanced scanner settings  see Chap  3 8      Note  There may be a discrepancy between the setpoint value and the actual value for the scan rate used    for saving if the number of scans is set too high or the processing power of the PC is insufficient     Page 39    Functions of the Measuring Programs       4 3    Display Image Data    The measuring program  Display image data  is used to display the
37. ed information can be found in Chap     3 12     Page 52    Functions of the Measuring Programs       1    Range 1  Use the two sliders to adjust the red lines which delimit range 1 in the 2D display  or input  the values  You can also set the range limits in the 2D display directly using the mouse  see  Chapter 3 12      Range 2  Use the two sliders to adjust the red lines which delimit range 2 in the 2D display  or input  the values  You can also set the range limits in the 2D display directly using the mouse  see  Chapter 3 12      Profile settings      Type  Use this parameter to define whether the section of the profile should be defined by a  local minimum value or a local maximum value      Min  height  mm   Use this setting to specify the minimum distance from the points to the  reference line for them to be recognized as a significant section of the profile      Min  width  points   Use this setting to specify the minimum number of points which must lie  next to each other and satisfy the criterion  Min  height  mm   to be recognized as a significant  section of the profile      Reference line  This indicates how the reference line is determined      Range 1   2  The program generates a reference line which corresponds to the line fit  through all points in ranges 1 and 2      Range 1   2 step  The program calculates the direction vector of the reference line from  the points in ranges 1 and 2  The position of the reference line is only formed from the  points in ran
38. enters are entered in the first reflection  the head points in the second  reflection and the tail points in the third reflection  This parameter has no effect on the display     Laser power  mW   You can operate the scanner laser at two different power levels or switch it  off     Off    reduced  scanCONTROL 2700 2710  2 mW  scanCONTROL 2800 2810  1 mW  if not  specified otherwise     standard  scanCONTROL 2700 2710  10 mW  scanCONTROL 2800 2810  15 mW  if  not specified otherwise   Trigger  This function enables you to control the scanCONTROL measurement system via the  synchronous input on the controller  scanCONTROL 2800 2810  and via the RS422 interface   scanCONTROL 2700 2710  respectively     internal  scanCONTROL transfers continuously according to the setting  No  of profiles   1 s    Please use the following trigger settings only when you control scanCONTROL via the  synchronous input on the controller and via the RS422 interface respectively     pos  edge  scanCONTROL is triggered with positive edges whereas the exposure time  corresponds to the parameter  Exposure time      neg  edge  scanCONTROL is triggered with negative edges whereas the exposure time  corresponds to the parameter  Exposure time      pos  pulse  scanCONTROL is triggered with positive edges whereas the exposure time  corresponds to the length of the positive trigger pulse     neg  pulse  scanCONTROL is triggered with negative edges whereas the exposure  corresponds to the length of the negative 
39. f points    Properties    Upper limit  rnm     0 5  Lower limit  rnm     1 5    X pos  Z pos  Height      Minimum  mm   Maximum  mm      D        2  Fig  4 8  Measuring programs  Flatness   Input fields and associated parameters in the measuring program  Flatness      A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information on each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      B 2D display  This display visualizes the profile which has just been measured     C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings      Status bar  The status bar contains status messages and error messages     E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   3 12    1 Range  Use the two sliders to adjust the red lines which delimit the range in the 2D display  or input    the values  The program will then calculate the flatness within this range  You can also set the  range limits in the 2D display directly using the mouse  see Chapter 3 12      Page 50    Functions of the Measuring Programs       2    Flatness settings  You use these parameters to
40. f this software may impair  or destabilize the corect operation of your system  either immediately or in the future  Microsoft strongly  recommends that you stop this installation now and  contact the hardware vendor for software that has  passed Windows Logo testing     C   Continue Anyway   STOP Installation     u       Fig  2 15  Windows XP driver installation   step 4  CA Click on  Continue anyway    to confirm this dialog  see Fig  2 15    Found New Hardware Wizard    Completing the Found New  Hardware Wizard    The wizard has finished installing the software Far     339  CMU 1394 Digital Camera Device    Click Finish to close the wizard     Cancel       Fig  2 16  Windows XP driver installation   step 5   4 Click on  Finish  in the next dialog  see Fig  2 16    The scanCONTROL driver is now installed    Note  Only use version 6 3 of the CMU driver  If a different driver or a different version of the CMU driver    for the scanCONTROL is installed at a later date then it will not be possible to operate the scanner with  the scanCONTROL Demo Software     Note  To install the driver at a later date  call up the installation dialog via the Device Manager  Control  Panel System gt  Hardware gt  Device Manager   scanCONTROL is located under  Imaging devices  or  under  Other devices      Note  If you are requested to choose the driver files manually you will find the files in the installation  directory of scanCONTROL Demo Software   Installation path  Program Driver CMU1394  
41. g program  Save  profile        1 2 5       4YCONNECT   scanCON    ROL Demo Software             File Measuring prografms Evaluation Default settings      Profile         profile  Type  LLT2800 25  004 v19 D2  SN  6050141   File Format  full set   Changed  04 02 2008 17 35 26    Path    A  ADatalprofile  avi x   76 7    74 1                Z distance  mm   794    re SEI OEE    714       58 8    Display settings 66 1    Scaling    X  Min    22 737 Max   22 16  Z  Min   66 096 Max   79 38  scan rate  1 s    Exp  time  ms     Profile Number of points  Properties Height  mm    Width  mm   Area  mm              Type    Min  height  mm       Left position  mm   Maximum position  mm    Min  width  points    22   13  4        Right position  mm     Reference line     Fig  3 9  Offline Mode  Offline mode     The illustration shows the typical structure of a measuring program in offline mode  as well as the joint  input and output fields of the measuring programs  A detailed description for each field is provided further  below in this user manual     1 File  In this input field you can select a previously saved profile sequence and start or stop the playback  of that profile sequence     2 2D display  This display is used to visualize a previously loaded profile  The display depends on the  parameters in the input fields  Scanner settings  and  Display settings      3 Display settings  You can use this input field to configure the scaling of the 2D display     4 Status bar  Status mes
42. ge 1      Range 1  The program only calculates the reference line from the points in range 1     Result display  The result display shows the currently measured profile  the reference line   horizontal green line   the minimum height  horizontal red line   the significant section of the profile   red points  and the left hand and right hand limits of the detected section  vertical green lines      Result status display       Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power      Exposure time  ms   The currently used exposure time     Number of points  The number of points in the profile section      Height  mm   The maximum height of the profile section      Area  mmv   The area of the profile section      Left position  mm  X Z  The point which delimits the section on the left      Maximum position  mm  X Z  The highest point in the profile section      Right position  mm  X Z  The point which delimits the profile section on the right     Page 53    Status and Error messages       5  Status and Error messages    Status messages and error messages are displayed in each measuring program  The possible messages  are shown below     5 1 In Online Mode    Active  LLT Type    ero OL There is no fault present  The  l measurement is active     Ty
43. he X Z coordinates and the distance  height  of the point  with the greatest positive distance to the reference line     Page 51    Functions of the Measuring Programs       4 9 Profile    The measuring program  Profile  is used to determine characteristic data for a significant section of a  profile  Examples of significant sections of a profile include a glue bead or a weld seam     First  the measuring program determines the points belonging to the significant section of the profile  A  reference line is fitted to do this  A minimum height in relation to the reference line is specified by the user   All points with a distance to the line which is greater than the minimum height are potential candidates for  the profile section being looked for  If a definable number  Min  width  of these points lie next to each other  then the program detects a  significant profile section      The results display contains information about the height  width and area of the currently detected section     1 2            4 YCONNECT   scanCONTROL Demo Software      File Measuring programs Evaluation Default settings      Scanner settings scanCONTROL  Profile  Measuring Field   huge fo     Ras SITE IR ER NIE I DER LET eee eee enemies  Points per profile    1024    EEA   m   a on        Sensor Laser    Exposure time  ms   E 10 0      auto    Profile CLE MRSN   NS       106 5         No  of profiles  1 s     104 1  Threshold    r   B        101 8 N  Reflections   first v   1       Invert signal 
44. ich delimit the range in the 2D display  or input    the values  The program then calculates a circle through the points in this range  You can also set  the range limits in the 2D display directly using the mouse  see Chapter 3 12      2 Result display  The currently measured profile and the calculated circle are shown in this display     Page 44    Functions of the Measuring Programs       3 Result status display      Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power      Exposure time  ms   The currently used exposure time     Xcenter  mm   This shows the X axis coordinate of the center point      Zcenter  mm   This shows the Z axis coordinate of the center point      Diameter  mm   This shows the diameter of the overall circle     Page 45    Functions of the Measuring Programs       4 6 Edge    This measuring program is used to measure the position and gradient of edges in a profile  A noise range   blue horizontal lines  and a threshold value  red horizontal line  are defined for this purpose  whereby the  threshold value must lie within the noise range  An edge is detected if the profile runs right through the  noise range  The position of an edge is calculated from the intersecting point between the profile and the  threshold value  
45. ilter Contains filter files for including DirectX     CD   Development Contains an environment for developing own  signal graphs for scanCONTROL and a SDK  for integrating scanCONTROL in C and C       CD   Development lconnect My Signalgraphs   Contains examples of signal graphs       Fig  2 1  Content of the scanCONTROL 5 1 CD  2 2 Installation of the scanCONTROL Demo Software    When you insert the scanCONTROL 5 1 CD the following installation dialog box appears which offers you  a number of options  Alternatively you can start the dialog box with the  CD   setup exe file     Page 6    Installation and Preparations for Measuring          eal scanCONTROL CD 5 1 E x        E Installation options for scanCONTROL    Installation quide   IMPORTANT  Please read the installation  guide first     Installs scanCONTROL software packages on  your system  Provides comfortable visualization  and several analyses of profile data        simple dema program for visualization of  start Developer D ema   profiles  Directly accessable from CD     IEEE 1394 driver has to be installed         MICRO EPSILON  nen   Displays documents and help files MICRO EPSILON Messtechnik GmbH  i K  nigbacher Str  15    D 34496 Ortenburg  Tel    49 8542  1 68 0    atone nM Ee   Opens the media contents with the Fax  49 95421 62 90  windows explorer     wWiwWwwW micra epsilon com    Exit   Close this dialog        Fig  2 2  Installation dialog box    Before you start to install the scanCONTROL Demo Software  ple
46. ime  ms   1 0  manual   Mo  of profiles  1  amp    75 0  Threshald  136    Reflections  largest area    Invert signal    ae ee NES    in X direction  no in 2 direction  yes       Fig  3 14  Tab sheet    Scanner settings  in the  Scanner settings  dialog    Scanner settings      Scanner settings X Advanced settings   Fiter settings              Processing          Suppress backlight  yes viden Filter      High resolution  yes Calibration     Head and tail  no    Sensor  Laser power  standard Trigger  internal  Shutter alignment  center Delay shutter  no    Switch off laser  if short circuit    Peak  Min  width  Max  width     Min  intensity  Max  inEensiby     Analog aut  Frequency  kHz   110 Hold last valid point   voltage range X  T  1U0V Invert X     voltage range 2       10   Invert 2     Miscellaneous  Loop back parameters  yes    Counter input  DIm     Count positive edges  no Count negative edges  no      CMM Trigger  Enabled  no Invert trigger output  no    Trigger output  Port 1 Trigger divisor  1  Mark space ratio  1 1 Skew correction  ps   0 0    Fig  3 15  Tab sheet    Advanced settings    in the    Scanner settings    dialog    Page 29    Working with the scanCONTROL Demo Software       Scanner settings       Fig  3 16  Tab sheet    Filter settings    in the    Scanner settings    dialog    Page 30    Working with the scanCONTROL Demo Software       3 11 Display Settings    The parameters in this input field influence the way in which the 2D display is represented
47. ime  ms   With this parameter you select the exposure time for the scanner  The measuring  program  Display profile  shows you the saturation in the resultant display  This is influenced mainly by the  exposure time and should be between 60  and 80  for a measurement     Minimum  0 01 ms     Maximum  40 0 ms     auto  Select this box to activate the automatic exposure time control of the scanCONTROL  measuring system     4 No  of profiles  1 s   This parameter regulates the number of measurements per second  This value  has priority over the entry in the  Exposure time  box      Minimum  25      Maximum  100  scanCONTROL 2700 2710  and 4000  scanCONTROL 2800 2810  respectively     5 Threshold  This value specifies the level of intensity at which the scanner recognizes a reflection  The  measuring program  Display profile  shows you the saturation in the resultant display  The saturation is  influenced mainly by the threshold and should be between 60  and 80  for a measurement      Minimum  O      Maximum  1023     6 Reflections  This box is for specifying which reflection is recognized as a profile point  This parameter is  only of importance for multiple reflections      first  The reflection closest to the scanner      last  The reflection furthest away from the scanner      largest area  The reflection with the largest area      highest intensity  The reflection with the highest intensity     7 Invert signal     inXdirection  The signal is inverted on the Z axis     in Z d
48. indicate the progress status in the field   Buffered  profile number    The profiles are stored in the selected file if the maximum number of  profiles is reached  Press the Stop button to abort buffering and start saving  The length of the  process will vary according to the size of the profile sequence  The program will indicate the  progress status in the field  saved  profile number      Trigger mode  active  Press the red Record button  The program is ready to buffer profile  sequences in the RAM  Press the trigger button to buffer a single profile  The profiles are stored  in the selected file if the maximum number of profiles is reached  Press the Stop button to abort  buffering and start saving  The length of the process will vary according to the size of the profile  sequence  The program will indicate the progress status in the field  saved  profile number       File Buffer     Path  The directory in which the file is saved   Name  Enter a name for the profile sequence you wish to save here   No  of profiles  The maximum number of profiles to be saved in the file  This number should not  be chosen higher than required  as the proportion of the RAM which is reserved for the profiles  depends on it  You can record up to a maximum of 4000 profiles   Data  Use this field to select the informations which are stored in the profile file  This setting is  crucial for the resulting file size of the profile file     x z only  Stores the X and Z coordinates of the individual m
49. ion Default settings                   Display profile  Save profile  Display image data  Anale  Circle F    Edge F   Peak F8  Flatness    Profile    Fig  3 4  Menu    Measuring programs     This menu item contains all of the available measuring programs and the  Close  function   Close  is used  to terminate the active measuring program and return to the main window  It is not possible to select a  new measuring program from an active measuring program  First close the active measuring program   then start the desired measuring program     Page 16    Working with the scanCONTROL Demo Software       3 Evaluation              File Evaluation   Default settings           Measuring programs       Offline Strg F2    Fig  3 5  Menu    Evaluation       This menu item is used to switch between online mode and offline mode  It is not available while a  measuring program is active     4 Default settings         Evaluation   Default settings    Offline   Save profile      File       Measuring programs         Reset all Display image data    Angle        Circle  Edge  Peak         Flatness    Profile       Fig  3 6  Menu    Default settings       You can use this function to restore the individual measuring programs back to their default settings  It is  separately available for the online and offline versions of the measuring programs  Use the  Reset all   option to reset all of the measuring programs     5       File Measuring programs Evaluation Default settings    QW Info about ICOM
50. irection  The signal is inverted in the middle of the measurement range parallel to the X axis   This function is enabled in the default settings     8Filter     With this button you access the  Filter settings  dialog box  see Chap  3 7     9 Advanced     With this button you access the  Advanced scanner settings  dialog box  see Chap  3 8     Note  To be able to use the  Invert signal  function  scanCONTROL requires a firmware version  gt   13   Check the firmware version on the status line if necessary  see Chap  3 4 and 5      Note  The  Scanner settings  box is only available in online mode     Note  For further information on the configuration of the measuring system  please refer to the  scanCONTROL2800 2810 and scanCONTROL2700 2710 instruction manual and in the documents   Quick reference scanCONTROL  and  Manual scanCONTROL  Part B    see Chap  3 3      Page 21    Working with the scanCONTROL Demo Software       3 7 Filter Settings    In combination with the scanCONTROL measuring system the scanCONTROL Demo Software offers you  the possibility of equidistantly resampling the profile data in the X direction and of filtering the profile data   If you want to use the filter functions  it is recommended to activate the equidistantly resampling     4 Use the  Filter     button in the  Scanner settings  box  see Chap  3 6  to open the  Filter settings   dialog box     Filter settings    Resample  v Interpalate invalid points   Range  large E a j i 4  Jw Resample all infa  
51. l now be applied     Page 34    Working with the scanCONTROL Demo Software          Range     a E ELE LEE ELLE ELLE EE ELLE UELLE EE EE rere      EUM Mig Loi d vEp a a EA oe Ej  a al b m        1150       T         45 0    u  45 4 5  151 45 4 75 7  45 258   112 47 X position  mm     15 03  Range ooo  15 28   Z distance  mm  E em aj e  Al     115 0  101 0  u PN  F3  59 0      45 0  en  45 4 151 15 1 45 4 Dan  EEZ  34521 A position  mm     Fig  3 3  Set range  3 12 8 Set horizontal Threshold       Press this button in order to adjust horizontal thresholds in the 2D display directly with the mouse   Proceed as follows     E23 Move the mouse pointer in the 2D display so that it is located at the desired height of the threshold   E23 Now click the left mouse button in order to set the threshold     Page 35    Functions of the Measuring Programs       4  Functions of the Measuring Programs  4 1 Display Profile    Use the    Display profile    measuring program in order to display individual profiles and a sequence of  profiles  The measuring program also calculates the saturation of the signal in a selected range   Additionally you can save the profiles in a format which is compatible with EXCEL     1           4 ICONNECT   scanCONTROL Demo Software        15  x       File Measuring programs Evaluation Default setting ING             scanner settings    Display profile Cig  sapere        Profile    Measuring Field   huge v   e    rre nennen nene nennen nra en nnn  Points per profil
52. lly incremented   If  for example   profile  is specified as the profile name  the files  profile 0001 sIk    profile 0002 slk  etc   are generated  The profile counter is reset after 4000 profiles     Page 37    Functions of the Measuring Programs       4 2    Save Profile    The measuring program  Save profile  is used to create recordings of profile sequences  The  scanCONTROL Demo Software creates a file which can then subsequently be loaded again whith a  measuring program in offline mode  Use this function to analyze fast processes offline  In the process the  profiles are first stored in the computer s RAM and then saved in a file     TJ ICONNECT   DispMgri    1       ci x       File Measuring programs Evdluation Default settings      Profile    Measuring Field   Points per profile  1024    Sensor Laser    Exposure time  ms     1 00    No  of profiles  1 5     25    Threshold  200  Reflections  largest are    scanner settin js         Save profile  Ub WOES       Z distance  mm  alfe  E  e Zl    87 8        huge       81 8       75 7    Invert signal    in x direction Ivi 2 drecton    did  Filter      Advinced    N is    Scaling    X  Min    22 157 ax   2  Min   66 108 x        69 7  Display settinijs     22 74   79 34         63 7            Freeze 2D display   Trigger mode       No  of profiles    1000 Data   rut set          Bo     Trigger    buffered  0 saved  0      A Data  ea    profile           57 6     13 2  4 2 48 13 8 DOT  X position  mm          Scan rate  1
53. me effective until the  next profile is displayed  To do this  select a new profile under  Current profile      Proceed as follows to print out or save a profile  Generate a screenshot by pressing the  Print  key  on your PC keyboard  Switch to image editing software like Microsoft Paint  under Accessories  or  Word  Press   Ctrl   V   to paste the screenshot  and then print out the screenshot or save it as a  file in the relevant program     Proceed as follows if you cannot see a profile in the  2D display  and no error messages are  displayed in the  status bar   Close the active measuring program  Restore the factory settings for  the selected measuring program via the option  Default settings  in the menu bar  Restart the  selected measuring program  Valid parameters are then loaded into the  Scanner settings   and the  complete measuring range is displayed in the  2D display      Page 56    Keyboard Shortcuts       7  Keyboard Shortcuts    You can use the keyboard shortcuts below to speed up your work  The following combinations are  available     Key shortcut    Opens the complete documentation for the  scanCONTROL Demo Software  from the main menu  onl  Opens the Help file for the active measuring program   only from within a measuring program   LIEN Starts the measuring program  Display profile   from  the main menu only       Starts the measuring program  Save profile   from the  main menu in online mode only   Starts the measuring program  Display image data   md  f
54. n each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      B 2D display  This display visualizes the profile which has just been measured     C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings      Status bar  The status bar contains status messages and error messages     E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   9 14    1 Range 1  Use the two sliders to adjust the red lines which delimit range 1 in the 2D display  or input    the values  You can also set the range limits in the 2D display directly using the mouse  see  Chapter 3 12      Page 48    Functions of the Measuring Programs       2    Range 2  Use the two sliders to adjust the red lines which delimit range 2 in the 2D display  or input  the values  You can also set the range limits in the 2D display directly using the mouse  see  Chapter 3 12      Peak settings     Type  Use this parameter to define whether the program is looking for local minimum values or  local maximum values   Min  height  mm   Use this setting to specify the minimum distance from the points to the  reference line for them to be recognized as a peak   Min  width  points   Use this setting to specify the minimum 
55. naeess 9  Installation  of the Drivers  Under Windows XP esiaine nannten 12  Working with the scanCONTROL Demo Software              ccccccsseeeeeeeseeeeeeeeeeeeeeneeeeeeneeseaenees 15  Starting the scanCONTROL Demo Software                   sseessssssssssesesene enne nennen nnne nnn 15  Elements in the Main Window                 cccccccccseccceeeceececeueecauecaueesaueeeueeseaeeceeesaueeseneesaeessaeesaasenas 15  MENU Bar en omen eran ee ets aC RO  mM ML NM MI bm e Mu aiu M cU 16  OnliRe Mede as  TEM UN 18  OmMNE Moden    ee ccc CE xm 19  SCANNER SOLBITigS  een aba testet ues  20  miserias c                                           22  Advanced Scanner Settings    cod opto do ere ee 24  Ile SENNO S225 iia cT c ccn HMM MR 27  Displaying scanCONTROL Parameters in Offline Mode                         240002224000nnnnneennennn nennen 28  Di5play Setings  ne ee ee 31  TOO AN Sene nu ee ee ee Mexicana ete os 31  Reset 2D DISDIayu ues oai edel 32  Automatic Scaling of the 2D Display                       sseeessssssssseseeeneeenn nennen mnes 32  REED ASDECERAl0 RN PL                                              ee 32  ZOOM In  albspispldy iid ta etaed Lat aav dtu ps ae 32  MOV ed ro i  eet M c E RE 33  Display Extended Information of a Profile Point                          esseeeesseeeeeeseeeeereeeennn 33  SEL RAIN Cases ceca P                                                  len 34  Set horizontal  ThEeSholQ   uic ste ofr a e ee onte teddies i deos 35  Functions of the Measuri
56. nd provides examples of signal graphs      2 Confirm the input field selection with  Continue   gt       Page      Installation and Preparations for Measuring       The installation routine then checks your system environment and installs the additional components  if  necessary  These components are displayed in a dialog box before they are installed  see Fig  2 4      Additional components x     The following additional components are necessary and will  be installed on your system       Adobe Reader 3    Cancel         Fig  2 4     Additional components    dialog box  E32 Confirm the dialog box with  Continue  gt       The selected software packages will then be installed  Once all the software packages have been installed  you may be requested to restart the computer  When the installation process has been completed you can  install the driver for the scanCONTROL measuring system  see Chap  2 3 and 2 4      Note  Only use version 6 3 of the CMU driver  If a different driver or a different version of the CMU driver    for the scanCONTROL is installed at a later date then it will not be possible to operate the scanner with  the scanCONTROL Demo Software     Page 8    Installation and Preparations for Measuring       2 3 Installation of the Drivers under Windows 2000  Preparing the installation    C43 Completely finish the installation of the scanCONTROL Demo Software  This procedure is described  in Chap  2 2     C3 Connect the scanCONTROL via the IEEE1394 interface to the 
57. ng Programs                              1lceeeeeeeseeee eene nennen nnn nnne nnn nnn nnn 36  BY 19  2 Al d  0 17      NNMERO AUR HERI ee PRICES ME YN Me 36  Sys NVM tad  2      or pee ee ne ee nee nda MED peel Mod IM EINE LE  38  DIsplayImage Data    Sect te cies ar D 40  PUN ee ee se cans es seg amen ea d mE EE eee MU LM 42  CIRCITER TEES 44  EOS ee Ye P ee tae ee 46  Peak Ep D DEDE 48  LATE SS acters sear ui M ia IRE Itu ee 50  ON MA EIE ee een 52  5tatus and Error Messages    in ee ee an ee re ee 54  Inline MOGG ans is ei ae ee dmt le m E el 54  IN  Offline Mode  kassel ensie elle 55  NOLES et ae pace eee eee ete es ee uei que Cet EL dr re 56  Keyboard Shortcuts 2 22 xdi ee a eee 57    Page 3    Introduction       1  Introduction    This user manual is designed to provide you with an overview of the functions available with the  scanCONTROL Demo Software  Together with the scanCONTROL measurement system  this software  enables you to analyze typical measurement tasks  The scanCONTROL measurement system is a high  speed sensor which is designed for use in industrial applications     The package is characterized by high speed and accuracy of measurement  The scanCONTROL Demo   Software will show you ways of finding solutions to various applications  all combined with the ultimate in  terms of accuracy  You can create your own measurement routines using the software ICONNECT     As a basis  the supplied CD offers a simple development environment with examples     Note  scanC
58. number of points which must lie  next to each other and satisfy the criterion  Min  height  mm   to be recognized as a peak   Reference line  This indicates how the reference line is determined     Range 1   2  The program generates a reference line which corresponds to the line fit  through all points in ranges 1 and 2     Range 1   2 step  The program calculates the direction vector of the reference line from  the points in ranges 1 and 2  The position of the reference line is only formed from the  points in range 1     Range 1  The program only calculates the reference line from the points in range 1     Result display  The result display shows the currently measured profile  the reference line  green  line   the minimum height  red line  and the measured peaks  red crosses      Result status display     Scan rate  1 s   This display shows the effective scan rate being used by the scanCONTROL  Demo Software to evaluate the scanner data  The scan rate may deviate from the parameter   No  of profiles  1 s   set in  Scanner settings  if the computer does not have sufficient  processing power    Exposure time  ms   The currently used exposure time   No  of peaks  At this point the program indicates the number of peaks detected in the currently  measured profile    X position  mm   This display shows the X coordinate of the first three measured peaks  The  peak with the smallest X coordinate counts as the first peak    Z position  mm   This display shows the Z coordinate of the
59. pe   Option      Version     Vary the distance between the scanner  and the measured object  or increase the  selected exposure time             No points in region of No data are being received from  interest the scanCONTROL         scanCONTROL is not   connected via IEEE1394 to the   Check the IEEE1394 connection and   No LLT found on PC  or the power supply to the check whether the scanner is switched on  IEEE1394 Bus scanCONTROL has been and connected to a power supply  Also  switched off  or no CMU driver is  check whether the CMU driver is installed   installed     scanCONTROL Demo Software   Open the    Advanced scanner settings     LLT S N not found on   ji   g  IEEE1394 Bus was last used with a different dialog box and press the button    Next    to  model of scanCONTROL  select the current scanCONTROL   No free LLT found on iei a nn  x Close other programs which are accessing  IEEE1394 Bus A  scanCONTROL   by other programs        Close scanCONTROL Demo Software    LLT parameters invalid Knkr  un TAU need  and restore the basic settings on the      controller by pressing the Reset button on  Error while data transfer ihecantrallac     Wrong driver version An incorrect version of the CMU    Reinstall the driver as described in section  CMU driver has been installed  2   Installation of the drivers     The currently used scanner is Close the other program accessing the  LLT is already in use already being used by another   scanCONTROL or press the button  Next   progr
60. river or a different version of the CMU driver  for the scanCONTROL is installed at a later date then it will not be possible to operate the  scanner with the scanCONTROL Demo Software     Select the    gt  Help menu item or press the  lt F1 gt  key directly in a measuring program to display  help information about that measuring program     The measuring program  Display profile  shows the saturation in the result display  This is greatly  influenced by the exposure time and the threshold value and affects the way in which the profile  data are calculated  The saturation should be between 60  and 80  during a measurement     There may be a discrepancy between the setpoint value and the actual value for the no  of scans  per second if the number of scans is set too high or the processing power of the PC is insufficient     Use the parameter    Measuring field  in the    Scanner settings  to define the resolution in the  directions of the X axis and the Z axis  The value of the X axis is also influenced by the input field     Number of points per profile   and takes priority over the  Measuring field  parameter  Choosing a  measuring field of  for example  512 x 256 pixels and a resolution of 1024 points per profile is not  consistent  In this case a measuring field of 1024 x 256 is selected automatically     If you have played back a profile sequence in offline mode and then wish to make changes to the  measurement settings or the display settings  these changes will not beco
61. rom the main menu in online mode only   Starts the measuring program  Angle   from the main  m   Starts the measuring program  Circle   from the main  menu only   Starts the measuring program  Edge   from the main  menu only   m Starts the measuring program  Peak   from the main  menu only   Starts the measuring program  Flatness   from the          main menu only     F Starts the measuring program  Profile   from the main  menu only   Ctrl F 1 Switches to online mode  from the main menu only     Ctrl  F2 owitches to offline mode  from the main menu only     Ctrl  Alt  F 12 Closes the scanCONTROL Demo Software  from the  main menu only   Closes the current measuring program and switches to  the main menu  from an active measuring program  only     F1  F2  F3  F4  F5  F6  F7  F8  F9  10       Page 57    MICRO EPSILON    www micro epsilon de    MICRO EPSILON    MICRO EPSILON  MESSTECHNIK   GmbH und Co  KG  K  nigbacher Str  15   D 94496 Ortenburg   Tel   49 85 42 1 68 0   Fax  49 85 42 1 68 90  e mail info micro epsilon de    X9751130 A070108ALI    
62. s in the measuring program  Edge      A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information on each of the  parameters can be found in Chap  3 6  Scanner settings  and Chap  3 9  File settings      B 2D display  This display visualizes the profile which has just been measured     C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings      Status bar  The status bar contains status messages and error messages     E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   3 12     Page 46    Functions of the Measuring Programs       1    Threshold value noise range     Noise range  blue   The size of the blue noise range in the 2D display can be adjusted by  moving the two sliders marked by  N  or by entering changing the values in the according input  fields  You can also set the range limits in the 2D display directly using the mouse  see Chapter  3 12     Threshold value  red   The position of the red threshold value in the 2D display can be  adjusted by moving the slider marked by  T  or by entering changing the value in the according  input field  You can also set the threshold in th
63. s of the measuring programs  A detailed description for each field is provided further  below in this user manual     1 Scanner settings  These values are used to change the scanCONTROL settings     2 2D display  This display is used to visualize the profile which has just been measured  The display  depends on the parameters in the input fields  Scanner settings  and  Display settings      3 Display settings  You can use this input field to configure the scaling of the 2D display     4 Status bar  Status messages and error messages are displayed in the status bar  For more detailed  information refer to Chap  5   Status and error messages       5 Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse interaction  and to set ranges directly with the mouse     Note  Depending on the measuring program  the display in online mode may also include additional input  fields     Note  The settings for a measuring program are saved and reused later when the measuring program is  started again  To restore the default settings  use the  Default settings  function from the menu bar     Note  The  Display image data  measuring program does not include the function  Display settings    Page 18    Working with the scanCONTROL Demo Software       3 5 Offline Mode    In offline mode the software evaluates data which have already been recorded  Use this mode to analyze  profile sequences later on in a measuring program  To record the data  use the measurin
64. sages and error messages are displayed in the status bar  For more detailed  information refer to Chap  5   Status and error messages       5 Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse interaction  and to set ranges directly with the mouse     Note  The measuring programs  Save profile  and  Display image data  are not available in offline mode     Page 19    Working with the scanCONTROL Demo Software       3 6 Scanner Settings    Use the parameters in this box to configure the scanCONTROL for your measuring task  Please note that  the parameters  Measuring field   Points per profile  and the parameters  Exposure time   No  of profiles   are dependent on each other  This section tells you about the following functions       Configurable parameters in the input fields     Impact of the parameters on the measurement     Restrictions on the resolution and exposure time     scanner settings    Profile                 Measuring Field  standard     U               N    Points per profile  256      r                                     P                     ua E El     Sensor Laser     Eeoa  7 ST a  SSS eas   veo wees isk     ee ees    Mens pe ay   Bes Tasse       Invert signal      in direction  v in 2 direction   7  cec    8 9  Fig  3 10     Scanner settings  input field            oO a fh Q       Parameters in the  Scanner settings  input field    1 Measuring field  Select one of the values to set the measuring field of the scanner
65. t   Peak Max  intensity  Maximum intensity of a reflection to be detected as a valid profile point     Page 25    Working with the scanCONTROL Demo Software       5    Frequency  kHz   With this parameter you specify the frequency of the analog outputs of the  scanCONTROL  You can deactivate the analog outputs by selecting O kHz      Minimum  0 kHz     Maximum  150 kHz  Hold last valid point  Activate this parameter to substitute invalid values at the analog outputs  with the respective last valid values   Voltage range X  With this parameter you can select the voltage range of the analog output for  the X coordinates       10V    10   0V      5V    0   10V  Voltage range Z  With this parameter you can select the voltage range of the analog output for  the Z coordinates       10V    10   0V       5V    0   10V  Invert X  With this parameter you can invert the analog output for the X coordinates   Invert Z  With this parameter you can invert the analog output for the Z coordinates     Loop back parameters  With this parameter you activate the parameter loop back  The  scanCONTROL settings are saved in the profile file in the measuring program  Save profile    see Chap  4 2  if  Loop back parameters  is activated and can then be displayed again later in  offline mode  see Chap  3 5      Count positive negative edges  Use this parameter to activate the internal counter of  scanCONTROL  The counter input is only available as an option for scanCONTROL  Select  whether the internal
66. ter into the 2D display    2 Press the left mouse button and keep it pressed   E32 Now move the mouse pointer in order to move the profile     3 12 6 Display Extended Information of a Profile Point    A    Press this button to activate or deactivate the  Display extended information of a profile point  mode   Proceed as follows     E23 Move the mouse pointer onto the desired profile point   E32 Click on the profile point with the left mouse button     The information for the point will now be displayed left to the toolbar for 5 seconds     Page 33    Working with the scanCONTROL Demo Software          eee m    A  0 603    77 152 Intensity  635 Width  11 E      1 1 Bir   l    UEM    M     a    ae  E i    Fig  3 2  Extended information of a profile point  The following information is displayed       X The X coordinate of the point     Z  The Z coordinate of the point     Intensity  The maximum intensity of the reflection from which the point has been calculated    Width  The width of the reflection from which the point has been calculated    3 12 7 Set Range    Hj s sj      Press the respective button in order to adjust the associated range in the 2D display directly with the  mouse  Proceed as follows     C4 Move the mouse pointer to any corner or side of the range to be set    E32 Now mark the range to be set  Hold the left mouse button pressed  Move the mouse pointer to the  opposite corner or side of the range to be set    Ea Release the mouse button  The selected range wil
67. the real aspect ratio of the displayed profile is retained  Deactivating this option  the scaling is adapted separately for the X and Y axes to the settings made in the    Min     and    Max     input  fields  see Chapter 3 11      3 12 4 Zoom in 2D Display    al    You activate or deactivate the    Zoom    mode by pressing this button  In the    Zoom    mode you can zoom in  parts of the 2D display and so directly set the scaling of the 2D display with your mouse  The values in the     Min     and    Max     input fields  see Chapter 3 11  are updated automatically  Proceed as follows     E32 Move the mouse pointer to any corner of the desired range to be zoomed in    C4 Now mark the range to be zoomed in  Hold the left mouse button pressed and move the mouse  pointer to the opposite corner of the range to be zoomed in     2 Release the mouse button  The marked range will be zoomed in     Page 32    Working with the scanCONTROL Demo Software       Z distance  mm  Elea    Zl H     115 0    101 0    ar    73 0 R    59 0        81 2  45 7  16 2 162 48 7 81 2  X position  mm     Z distance  mm  E3  e  n  a e  Zl H     103 3    95 2       83 1    8 0       82 8        28 4  16 5  4 7 71 19 0 30 8  X position  mm     Fig  3 1  Zooming in the 2D display  3 12 5 Move Profile    i     You activate or deactivate the  Move profile  mode by pressing this button  You can move the profile on  the 2D display with an active  Move profile  mode  Proceed as follows     E32 Move the mouse poin
68. ting from the setting in the  Processing   see  Chapter 3 8  input field      Number of sensors  The maximum profile frequency resulting from the setting of the  No  of  simultaneous operating sensors  parameter  see above       Exposure time  The maximum profile frequency resulting from the setting of the  Exposure  time  parameter      Result  The maximum profile frequency resulting from the above restrictions     Note  Further information on the maximum profile frequency can be found in the document  Quick  reference scanCONTROL      Page 41    Functions of the Measuring Programs       4 4 Angle    This measuring program calculates the angle between two selected ranges of a profile  The program  calculates the best line fit through the points in the selected range and measures the angles between  these lines and the X axis  as well as the angle between the two lines  The measured angles are  continuously updated in the results display               4 ICONNECT   scanCONTROL Demo Software      File Measuring programs Evaluation Default settings      Scanner settings scanCONTROL  Profile  Measuring Field   huge fo  Points per profile    1024       Sensor Laser    Exposure time  ms   i 1 00      auto    No  of profiles  1 5        Threshold       Reflections   largest area 3    Invert signal     v in X direction  v in Z direction  Filter      Advanced         Display settings    Scaling    x  Min    19 843 m  21 496  2  Min  132 205 ax   96 811           X position  mm        
69. tion  4   is the reflection with the largest distance to the sensor     B 2D display  This display visualizes the profile which has just been measured    C Display settings  You can use this input field to configure the representation of the 2D display   Detailed information on each of the parameters can be found in Chap  3 11  Display settings     D Status bar  The status bar contains status messages and error messages    E Toolbar  Use the toolbar to scale the 2D display  to activate various options for the mouse  interaction and to set ranges directly with the mouse  Detailed information can be found in Chap   3 12     Page 36    Functions of the Measuring Programs       1    Range  Use the sliders to adjust the red lines which delimit the range in the 2D display  or input the  values  You can also set the range limits in the 2D display directly using the mouse  see Chapter  3 12      Log settings  This defines the settings for the profile logs     Path  The directory in which the files are saved     Name  The name of the profile     Trigger  Press this button to record a single profile     Active  Use this function to start the log  The log starts and ends with the selection of this  checkbox     Time diagram  Settings   These parameters are used to configure the display of the time diagram   4  for displaying profile sequences      No  of profiles  This indicates the number of profiles to be displayed      Active  Activates deactivates the display of profile sequences   
70. tion to the reference line is specified then  All points  which have a greater distance to the line than the minimum height are potential candidates for peaks  If a  definable number  Min  width  of these points lie next to each other the program will detect a peak   whereby the peak position is defined by the point with the maximum distance to the line     1               4 ICONNECT   scanCONTROL Demo Software         File Measuring programs Evaluation Default settings      Scanner settings scanCONTROL  Profile  Measuring Field   huge fo Range 1  Points per profile    1024    Pas    Sensor Laser 88 2    Exposure time  ms   i      auto                 No  of profiles  1 5     86 8  Threshold  zu    E     Reflections    area       Invert signal    in X direction  v in Z direction  Filter      Advanced       Display settings    Scaling    X  Min    14  132 ECCE  4 871  a    Z  Min   83 233   x   86 319    85 5          Scan rate  1 s   Exposure time  ms   No  of peaks    Peak settings    Properties    Peak 1 Peak  Peak3    Type   maximum     Min  height  mm   fo  Min  width  points              Reference line   range 1 2        X position  mm     Z position  mm   Height  mm      D        3  Fig  4 7  Measuring program  Peak     Input fields and associated parameters in the measuring program  Peak      A Scanner settings  With these values you can change the settings of scanCONTROL  In offline  mode this input field is replaced with the  File  input field  Detailed information o
71. trigger pulse     Note  More information about the trigger function of scanCONTROL can be found in the  scanCONTROL2800 2810 and scanCONTROL2700 2710 instruction manual     Shutter alignment  With this parameter you determine how the specified scan rate is  maintained with active automatic exposure time control      center  The intervals are in each case aligned to the center of the interval       right  The intervals are in each case aligned to the end of the interval       left  The intervals are in each case aligned to the start of the interval      none  The intervals are not aligned   Delay shutter  With this parameter you determine whether the automatic exposure time control  is computed based on the original data  deactivated  or based on the filtered data  activated    Switch off laser  With this parameter you configure the laser safety interlock of  scanCONTROL depending on the protective circuit of pins 3 and 13 of the synchronous input   This parameter is only available using scanCONTROL 2800 2810       if left open  The laser is switched off with the pins open      if short circuit  The laser is switched off with the pins connected  This parameter is only   effective for laser class 2M  IIM      Peak Min  width  Minimum width of a reflection to be detected as a valid profile point    Peak Max  width  Maximum width of a reflection to be detected as a valid profile point    Peak Min  intensity  Minimum intensity of a reflection to be detected as a valid profile poin
72. und will be installed     Search removable media  floppy  CD ROM        Include this location in the search     Choose this option to select the driver from a lish  Windows does not guarantee that  the driver you choose will be the best match for your hardware     ETI a  ETE       Fig  2 13  Windows XP driver installation   step 2    5    CA Inthe next dialog  select  Don t search    will choose the driver to install         a Click on  Next  to confirm this dialog  see Fig  2 13      Found New Hardware Wizard    Select the device driver you want to install for this hardware                 Select the manufacturer and model of your hardware device and then click Mest  IF vau  have a disk that contains the driver you want to install  click Have Disk     Show compatible hardware      C U 1394 Digital Camera Device Je  e    9 This driver is not digitally signed     Tell me why driver signing is important      lt Back  fL Ne TL Cae     Fig  2 14  Windows XP driver installation   step 3       Ea Select  CMU 1394 Digital Camera Device  from the list and click on  Next  to confirm  see Fig  2 14      The  Hardware installation  dialog will then appear     Page 13    Installation and Preparations for Measuring       Hardware Installation    The software you are installing Far this hardware     CMU 1394 Digital Camera Device    has not passed Windows Logo testing to verify its compatibility    with Windows sP   Tell me why this testing is important       Continuing your installation o
73. version of a measuring program evaluates data directly from  the scanCONTROL and displays them on a PC  The offline version loads a previously saved profile  sequence and then evaluates it     Page 5    Installation and Preparations for Measuring       2  Installation and Preparations for Measuring  2 1 Installation Requirements    The following minimum system configuration is required for the scanCONTROL Demo Software to be able  to work properly       Windows 2000  Service Pack 4  or Windows XP    Internet Explorer 6 SP1     Pentium Ill 2 800 MHz     256 MB RAM     Screen resolution  1024x768    Proceed as follows to run the scanCONTROL Demo Software     1  Install the IEEE 1394 interface hardware  if not already equipped    2  Install the scanCONTROL Demo Software    3  Connect the scanCONTROL measuring system via the IEEE1394 to the computer   4  Install the driver for the scanCONTROL measuring system      CD ADocumentation Contains manuals  installation instructions and  Quick Reference     CD   Program scanCONTROL Contains the following software packages   scanCONTROL Demo Software 1 5   scanCONTROL 3D View 1 2   scanCONTROL 3D Data 1 2     CD   Program DeveloperDemo Demo program for scanner setting and  visualization of oo              data      CD   Support DirectX8 1 Contains DirectX8 1       DirectX8 1     CD   Support Driver CMU1394 Contains the driver files for scan CONTROL  version 6 3     CD   Support Adobe Reader 9 Contains Adobe Reader 9   CD   Support Flmager
74. vert 2    Miscellaneous       Y Loop back parameters    Counter input  DIn     Count positive edges   Count negative edges    CMM trigger    Enable CMM trigger    Invert trigger output    Trigger output   Port 1    Trigger divisor    1      Mark space ratio  ise    Skew correction  Lis    o      OK   Cancel      Fig  3 12   Advanced scanner settings  dialog box       dialog box     Parameters in the  Advanced scanner settings  dialog box    1    Note  Only experienced users of the scanCONTROL measuring system should adjust any settings in this      Type  The currently used type of measurement system and the associated firmware version are    displayed       SN  The serial number of the currently used measurement system is displayed       Next scanner  If a number of measurement systems are connected to the PC  you can change    between the individual measurement systems with this button     Page 24    Working with the scanCONTROL Demo Software       2    Suppress backlight  With this entry field you activate the automatic backlight compensation   Video filter  You activate this parameter to improve the image quality    High resolution  With this parameter you activate the subpixel resolution for the Z values   Calibration  You use this parameter to activate the calibration    Head and tail  You can only use this parameter with the subpixel resolution deactivated  With  the parameter activated the head  center and tail of a reflection are calculated for each  coordinate  The c
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
BETRIEBSANLEITUNG OPERATING INSTRUCTIONS SPECK  MANUEL D`UTILISATION DU  Samsung 760BF Εγχειρίδιο χρήσης    補聴器の適正広告・表示ガイドライン集について [PDFファイル/1.12MB]  Akku-Netzrasierer  Mode d`emploi    Soplete mecanizado para cortar de arco de plasma  RACK-200 CHASSIS USER`S MANUAL    Copyright © All rights reserved. 
   Failed to retrieve file