1. N1533 TM 2655P INSTRUCTION MANUAL Automatic Blood Pressure Monitor 1WMPD40002061 2005 A amp D Company Ltd All rights reserved No part of
2. Fig 16 Pin assignment Cable connection between the device and a personal computer TM 2655 TM 2655P D sub connector Personal computer D sub connector Signal Pin No Signal Pin No E 1 7 1 RXD 2 RXD 2 oa nia DIR 4 DTR 4 GND 5 lt GND 5 DSR 6 DSR 6 RTS 7 RTS 7 cTs 8 gt a CTS 8 E 9 9 Fig 17 Cable connection diagram 20 Note The protocol depends on the equipment connected 9 MAINTENANCE 9 1 Replacing the Arm Cuff Cover Replace the arm cuff cover as follows F Front frame 1 Locate the cover at the bottom of the arm insertion section Slide the cover downward to open it A screw is exposed Arm insertion section Using the screwdriver loosen the screw and remove the front frame 2 Remove the four screws on the rear panel to remove the armrest and the rear panel CS Armrest N Pas 3 Remove the front vinyl ring of the arm cuff cover from the groove Remove the rear vinyl ring of the arm cuff cover from the groove Pull out the arm cuff cover Note The vinyl rings will be used to support the new arm cuff cover 4 Place a new arm cuff cover in the arm cuff area Secure the Back front and rear vinyl rings of the arm cuff cover in the groove pu a TMA Smooth the cover cloth near the grooves po New arm y cuff cover 5 Replace the rear pane
3. 5 Press the PAPER FEED switch to feed the paper by about 10 cm Confirm that the paper is fed straight If not re install the paper because it may cause a paper jam PAPER FEED switch 14 Lower the lever to the down position 1 Press the PAPER FEED switch again and confirm that the paper is cut automatically Note Lower the lever for cutting the paper Lower the lever for printing the measurement data PAPER FEED switch Remove the printer paper from the paper outlet and close the cover Press lightly on the center of the printer cover to secure it in position Printer paper outlet Note With high speed printing about 700 printings can be performed Fig 11 Paper installation procedure With four line printing about 500 printings Replace the printer paper when the red end is reached Use the specified thermal paper which is sold separately 15 7 SETTING THE FUNCTIONS 7 1 Function Setting Mode The TM 2655 TM 2655P is provided with various functions which allow the user to set the device to suit the user s requirements The functions are set in the function setting mode The function setting mode display and switches used are as shown below AUTOMATIC BLOOD PRESSURE MONITOR SYSTOLIC mmHg Function III number 14 bi DIASTOLIC mmHg SELECT switch A switch COUNTER switch ES Function _ parameter a uy
4. E WU O 6 U JJ J Y 245 mm A L 390 mm 138 mm Fig 20 External dimensions Note The illustration above indicates the TM 2655P with the printer unit The dimensions are the same for the TM 2655 and TM 2655P 26 MEMO MEMO
5. aiid Consumables vs 332g irradia tedio 24 10 TROUBLESHOOTING AAKALA 25 11 EXTERNAL DIMENSIONS ARKA TARA 26 1 INTRODUCTION 1 1 Features The TM 2655 TM 2655P is a blood pressure monitor that measures systolic and diastolic blood pressure and pulse The TM 2655 is a standard model the TM 2655P is equipped with a printer unit The features of the devices are as follows COMPACT DESIGN Due to its compactness the TM 2655 TM 2655P can be installed in an area where space is limited You can measure your blood pressure on either arm AUTOMATIC CUFF SYSTEM AUTOMATIC EXHAUST ADJUSTMENT SYSTEM The arm cuff is wrapped around the arm by pressing the START STOP switch and deflation speed is automatically controlled No special adjustment is required All you have to do is insert your arm into the arm insertion section to the shoulder and press the START STOP switch The rest is done automatically You can easily and quickly measure your blood pressure The TM 2655 TM 2655P is suitable for mass medical check ups at firms and sports facilities and in hospital waiting rooms when used for a pre check of the patient s health status BUILT IN PRINTER TM 2655P only The TM 2655P is equipped with a printer unit which allows measurement results to be printed out for record keeping The print format can be selected from results only results with pulse pressure graph or a list of the results BUILT IN CLOCK The date and time of measurement
6. al Fig 12 Switches for setting the function and the function setting mode display Table 4 Switches for setting the functions A switch and SELECT switch Hold down the A switch and press the SELECT switch to enter the function setting mode COUNTER switch Press to select the function to set A switch Press to select the function parameter 1 Hold down the A switch and press the SELECT switch FO I appears in the SYSTOLIC pressure display This indicates that the monitor is now in the function setting mode 2 Press the COUNTER switch to select the function number to set Each time the switch is pressed the function number changes from FO 1 to FO to FJ 3 Press the A switch to select the function parameter 4 When the settings are complete turn the power off and turn it on again The settings are saved 16 7 2 Description of the Functions Table 5 Functions No Function SYSTOLIC DIASTOLIC Description display display F01 0 oFF Not available F02 Printer setting Fae oF F No printing See 7 3 I High speed printing Printing e 4 line printing Samples for what each J 4 line printing method printing y Graph printing eae KON 5 Table printing F03 dM LO a OMY F03 oFF No bit pattern printing on Bit pattern printing F04 Displaying time FOY 999 Th
7. 222 NN 7 PAE EIR S E S ANAT A PEE VAE I E AERO T Aa a ANA ma a a tented aa 7 3 INSTALLATION ii A A A dd 9 A OPERATION conan aria dopado 10 5 SETTING THE CLOCK as 12 5 1 Glock Adjustment Mod 6 ada Ka na dd 12 5 2 A Setting Example c ccccccceceeeceeecccece cece A E TE TET EET N T 12 6 INSTALLING THE PRINTER PAPER 11111100000 nnmnnn 14 7 SETTING THE FUNCTIONS a AGANG RING NAKARAAN 16 11 Function Setting MOOG 7 12 nabaha aid e E EE deus bng 16 f 2 Description of the FUNCIONS mamamana Aa a Aaah 17 3 Priniting Sampl68 it BAN NANNING AEI ALAGA ABA AA aNG 18 8 COMMUNICATION SPECIFICATIONS cccccccceeeeeeeeeeeeeeeeeeeeeeneeeeeeeeeeeeeeeeeneeeeeeeeeees 19 8 1 Channel 1 Miniature 8 pin DIN 1101 7722700700000000000 0000000400400 EESEEREEREEEEE EG KENA AKASASEKEE ENNA AKA AKAS REEE KENA KK As EEEE EnEn EEEE 19 8 2 Ehanmel 2 D SUD IPIN zerrin NAG eae eles GN le ee eee lee 20 9 MAINTENANCE coin 21 9 1 Replacing the Arm Cuff Cover 0 0 cceccenere eee t nee teenie ee CNN nn nc enters cane iE re TANEN rre 21 9 2 Replacing ihe PUSCS norn GNG ANA BANANA E 22 9 3 Checking the Gounte bass cee ee inked es SL ei ee ei 22 Displaying the COUNT a aed eee nd BANA 22 Resetting the COUNIET2 L32 ag nA a 23 Printing To counter graph sts chia ate Aah nai alent hahahhahaha aes 23 9 4 ENTONCES iii Ana NA ste vd a BA KANAN An 23 A NA HAAN devs BAG DALIAN AA ALILA NANANA neh eee 23 9 6 CIGANING BP 23 927 Options
8. MAP Means arterial pressure in mmHg Exx Error code xx 00 99 What is An Irregular Heartbeat The TM 2655P provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs An irregular heartbeat is defined as a heartbeat that varies by 25 from the average of all heartbeats during the blood pressure measurement It is important that you are relaxed remain still and do not talk during measurements Note We recommend that a physician should examine the patient if the O logo appears on the printout or request a second reading lt is possible for an Irregular Heartbeat O to occur if a patient moves position or engages in conversation during a reading so care should be taken to eliminate the possibility of these events recurring 3 INSTALLATION Follow the procedure below to install the TM 2655 TM 2655P 1 To ensure that the TM 2655 TM 2655P works properly install the device in an environment where m The temperature range is from 10 C to 40 C 50 F to 104 F m The humidity is less than 85 non condensing E It is away from water dust chemicals and corrosive or explosive gases E It is not exposed to direct sunlight 2 Place the TM 2655 TM 2655P on a table solid enough to support its weight 3 Adjust the height of the chair and table so that the arm insertion section is at the user s heart level TM 2655 TM 2655P Fig 4 Proper installation 4 Use the power c
9. amp D Company Limited Fig 13 Printing samples Note Bit pattern printing is available only for the TM 2655P With this format some items such as the company name can be printed For details contact your nearest dealer When detecting an irregular heart beat I H B is printed 18 8 COMMUNICATION SPECIFICATIONS The TM 2655 TM 2655P is equipped with two RS 232C channels Various settings for each channel are available in the function setting mode Refer to 7 2 Description of the Functions for details 8 1 Channel 1 Miniature 8 pin DIN N CAUTION The personal computer and medical equipment connected to the device are not allowed to be in the patient area e The personal computer used must conform to IEC60950 Specifications Table 6 Channel 1 specifications Standard Conforms to EIA RS 232C Transmission Asynchronous half duplex Baud rate 1200 2400 4800 9600 bps Can be changed in the function setting mode F07 Data bits 8 bits Parity bit None Stop bits 1 bit Code ASCII Equipment available for connection Channel 1 Scales and automatic weight and height scales manufactured by A amp D Channel 2 Personal computer Pin assignment Pin No Signal name Description 1 TXD Send data 2 RXD Receive data 8 7 6 3 RTS Ready to send 5 4 4 Used internally 1 5
10. can be printed along with the measurement results COUNTER UP TO 999999 COUNTS The counter indicates how often the device is used It can be a helpful guideline for maintaining the device SAFETY SYSTEM The TM 2655 TM 2655P adopts an emergency stop system In case of emergency pressing the EMERGENCY STOP switch will release the air in the arm cuff so that the arm can be removed easily This system also functions if there is a power failure REPLACEABLE ARM CUFF COVER The arm cuff cover can be easily replaced 1 2 Unpacking and Inspection N CAUTION Handle this device carefully at all times Strong shock to the device may cause trouble during operation Note Save the packing material for later use Unpack the TM 2655 TM 2655P carefully and verify that the following items are contained Power cable cord set Arm cuff cover 2 pcs TM 2655 TM 2655P With one factory installed 1111299 111125997 Printer paper 2 rolls Instruction manual Fig 1 Items contained 1 3 Specifications Table 1 Specifications Performance specifications General Power supply 120VAC 60 Hz 230VAC 50 Hz factory preset For USA area For Europe area Power consumption Approx 40VA Protection against electrical shock Display Class I Type B UN LED EMC EN 60601 1 2 Blood pressure measurement Blood pressure measuring method Oscillometric method Press
11. troubleshooting instructions request service from your supplier or from the A amp D service group The A amp D service group will provide technical information spare parts and units to authorized suppliers The technical testing procedures which should be done at least every two years can be performed either by the manufacturer or by an authorized repair service in accordance with the regulations governing manufacturing of medical products 9 6 Cleaning N CAUTION Never immerse the TM 2655 TM 2655P in water for cleaning that may damage the electronic parts inside O Never use thinner or a strong detergent for cleaning that may discolor or deform the plastic case and display panel 23 Housing case Clean the case gently using a soft cloth moistened with water and mild detergent or alcohol Display panel Clean the display gently so as not to scratch the panel surface Use a soft cloth moistened with water Arm cuff cover When the arm cuff cover becomes dirty or it is worn out replace with a new one Refer to 9 1 Replacing the Arm Cuff Cover Note If the cover is not installed properly it may cause problems in operation 9 7 Options and Consumables Options and consumables available for the TM 2655 TM 2655P are shown below Order them from the nearest dealer Use the part numbers when ordering Oo Table AA YY TM 9325 e Chair without backrest Gas shock suspension TM 9312A e Chair with backre
12. 5P e The illustration above is not to scale 2 2 Rear Panel COUNTER switch SELECT switch A switch External extension terminal 1 External extension terminal 2 2 3 Symbols Rating plate Fig 3 TM 2655 TM 2655P rear panel The illustration above is not to scale AC INPUT connector The symbols used with the TM 2655 TM 2655P have the following functions or meanings Table 2 Symbol descriptions Symbols Function Meaning O Starts and stops a measurement Turns the device on O Turns the device off IN Alternating current SN Serial number EI Fuse o RS 232C serial interface Date of manufacture 2000 Type B Device cuff and tubing are designed to provide special protection against electrical shocks C Es The medical device label by the EC directive Q noz The certification trade mark registered to the Australian Communications Authority by the Trademarks office BISIOLIC Systolic blood pressure in mmHg SYS DIASTOLIC Diastolic blood pressure in mmHg PULSE PUL Pulse beats per minute A See the instruction manual A WEEE label Irregular Heartbeat indicator 1 H B The indicator that prints out when an O irregular heartbeat or any excessive body movement is detected during the measurement
13. CTS Clear to send 6 GND Signal ground Note 7 Used internally Do not use pins 4 7 and 8 8 z Used internally They are used by the device Fig 14 Pin assignment Cable connection between the device and a personal computer TM 2655 TM 2655P Personal computer Mini DIN connector D sub connector Signal Pin No Signal Pin No TXD 1 1 RXD 2 i 14 RXD 2 RTS 3 TXD 3 4 DTR 4 CTS 5 GND 5 GND 6 DSR 6 7 RTS 7 8 CTS 8 9 Fig 15 Cable connection diagram 19 8 2 Channel 2 D sub 9 pin be in the patient area N CAUTION The personal computer and medical equipment connected to the device are not allowed to e The personal computer used must conform to IEC60950 Specifications Table 7 Channel 2 specifications Standard Conforms to EIA RS 232C Transmission Asynchronous half duplex Baud rate 1200 2400 4800 9600 bps Can be changed in the function setting mode F08 Data bits 8 bits Parity bit None Stop bits 1 bit Code ASCII Equipment available for connection Channel 1 Personal computer Channel 2 Card Pin assignment reader Pin No Signal name Direction Description 1 m 2 RXD In Receive data 3 TXD Out Send data 4 DTR Out Data terminal ready 5 GND Signal ground 6 DSR In Data set ready 7 RTS Out Ready to send 8 CTS In Clear to send 9 z
14. Follow the precautions described in the manual for each device E Do not use a cellular telephone near the device It could affect the device s operation E The personal computer and medical equipment connected to the device are not allowed to be in the patient area m The personal computer used must conform to IEC60950 After use E Follow the predetermined procedure to return the operation switches to their original positions and then turn off the power m Do not forcibly pull out the cables Hold the connector with your hand when disconnecting the cable E Keep the device clean and in proper operating condition so that it can be used without problem during the next operation Clean the accessories and arrange them before storage Environmental protection Em Remove the built in lithium battery from the device when the device is to be disposed of m This device any part of and the built in lithium battery are not treated as ordinary household waste and must be disposed of according to the applicable regulations CONTENTS WARNING DEFINITION 5 2NA TANUAN AA l COMPLAN CE a a is Il SAFETY PRECAUTIONS pa kodar ANAN a 111 GENERAL PRECAUTIONS icon da ia iia inde da a IV CONTENTS oran aran 1 INTRODUCTION da 3 BA oa 3 1 2 Unpacking and Inspection visite viene GANAN GA NA A ANA LG 4 153 SpeciiGalionss ic a ana naalagaan naaa a da he nadal aaa Ee o 5 2 PART NAMES curia ANAN AA 6 2 Mad Una ngn ane ea ti eae ee de NATN eee 6
15. able provided with the device to connect between the AC INPUT connector and an electrical outlet Using the anti theft opening secure the device to the table with a solid chain Anti theft opening Power cable Fig 5 Power cable connection Note Be sure to use the correct voltage for the power source 9 4 OPERATION Turn on the POWER switch located on the rear panel When the POWER switch is turned on all of the display symbols will appear for several seconds Then Bm TM 2655 0 zero appears in the SYSTOLIC pressure display indicating that the device is ready for measurement EB TM 2655P The human sensor functions for three SYSTOLICY mm minutes When it does not detect a person dot appears in the SYSTOLIC pressure display indicating that the device is ready for measurement Fig 6 POWER switch and Note SYSTOLIC display TM 2566 Please do not press the START STOP switch when you turn on the power switch Pressing and holding the START STOP switch while all of the LED indicators are illuminated will cause the device to enter the internal inspection mode Take off your jacket If you wear a thick jacket take it off for a better measurement _ Shoulder Arm insertion section p gt Note Wearing a thick jacket may cause a faint pulse and result in a measurement error Adjust the height of the chair and table Adjust the height of the chair and
16. ars Is your posture proper Place your arm at your heart level relax and remain still during the measurement Do you remain still Do not move your arm during the measurement Relax and remain still Wearing a thick jacket may cause a faint pulse and result in a measurement error Take off the jacket This device bases its measurements on the heartbeat If you have a very weak or irregular heartbeat the device may have difficulty determining your blood pressure The printer does not print Is the printer paper installed PE appears Refer to 6 INSTALLING THE PRINTER PAPER to install the printer paper Is the printer head up HU appears Lower the lever and lock the printer paper in position Is the printer paper inserted straight or is the printer paper jammed Refer to 6 INSTALLING THE PRINTER PAPER to re install the printer paper Is the printer setting correct Refer to 7 2 Description of the Functions to select the correct printer setting Note If the actions described above do not solve the problem contact the dealer Do not attempt to repair the device yourself 25 11 EXTERNAL DIMENSIONS Jo em GO 0 3888 i E gt l T E N D D N E
17. asurements Only a trained medical professional is qualified to interpret the blood pressure measurements No device can replace regular medical checkups by a doctor A doctor should verify the blood pressure measurements before making adjustments to medication If the device gets wet because of a liquid spill If the device gets wet promptly stop using it Have the patient remove their arm from the device and secure the patient s safety Turn the power off and disconnect the power cable from the electrical outlet Affix a notice such as Out of Order and Do Not Use on the device Ask the nearest dealer for service N CAUTION In case of emergency Press the EMERGENCY STOP switch located on the front of the device This will release the air in the arm cuff so that the arm can be removed easily It also operates if there is a power failure To stop the operation at any time Press the START STOP switch located on the display panel The quick exhaust will release the air in the arm cuff and the arm cuff will become loose Maintenance Inspection Check all functions periodically Ask the nearest dealer for this inspection Do not disassemble or modify the device If using the device for the first time after an extended period of storage check the device for proper operation GENERAL PRECAUTIONS Obey the following precautions for safe and correct usage Precautions unique to the TM 2655 TM 2655P appear on the relevant pages in th
18. ch to display 4 for April Press the SELECT switch to make the PULSE display flash Press the A switch to display 20 Press the SELECT switch to make the hours section of the TIME display flash 12 Pressthe A switch to display 14 Press the SELECT switch to make the minutes section of the TIME display flash 10 11 Press the A switch to display 07 Press the SELECT switch to return to the measurement mode Note e f no operation is performed for one minute while setting the minutes section and for five seconds while setting the others the settings performed so far will be confirmed and the device will return to the measurement mode e The clock can be set up to December 31 2091 13 6 INSTALLING THE PRINTER PAPER This chapter is applicable only to the TM 2655P With the power turned on install the printer paper as follows 1 Cut the top end of the printer paper with scissors If the end is not smooth it may cause a paper jam or damage the printer head 2 Press lightly on the center of the printer cover to open the cover Printer cover 3 Raise the lever located on the right side of the printer Set the lever to the up position 2 of the figure The insertion of the printer paper is enabled Raise the lever to the up position 2 4 Insert the printer paper into the paper feed slot as shown The paper is fed automatically Paper feed slot
19. e results remain in the display unless another operation is performed 5 Displays the results for 5 seconds 10 Displays the results for 10 seconds ea Displays the results for 20 seconds FO5 External FOS l Terminals 1 and 2 to a PC connection Pa Terminal 1 to a PC Terminal 2 to a card reader 3 Terminal 1 to a scale Terminal 2 to a PC y N A FO6 Pressurization FOG Au Automatic pressurization 160 Pressurizes the cuff up to 160 mmHg 180 Pressurizes the cuff up to 180 mmHg c00 Pressurizes the cuff up to 200 mmHg FO7 Channel 1 baud F07 ied 1200 bps rate 240 2400 bps 480 4800 bps 960 9600 bps FO8 Channel 2 baud FOB leb 1200 bps rate 240 2400 bps 480 4800 bps 960 9600 bps FO9 F09 H9 Not available F10 Human sensor F 10 oFF Does not detect a human presence on Detects a human presence F11 Automatic Fii oFF Does not output the results external output automatically on Outputs the results automatically 17 7 3 Printing Samples High speed printing Graph printing o Jan 1 2006 10 29 o Dec 30 2010 15 29 SYS DIA PUL 133 59 78 sys 104 mmHg bpm MAP 8 Dein DIA Tl anne PUL 1 8 bon 4 line printing PULSE ENVELOPE gt Jan 1 2006 10 33 SYS 1 3 anie MAP 8 Bi 6 Omni 78 pm Table printing Bit pattern printing May 26 2000 11 41 ier ae as TINE ss DIA PUL SYS DIA PUL 65 78 00002 10 50 135 612 78 1 10 15 60 00003 10 51 132 61 77 mmHg bpm A
20. evice was designed for use by adults only Compliance with the Australian EMC Frame Work This device conforms to the following requirements EMC Emission Standard for Industrial Scientific and Medical Equipment AS NZS 2064 1997 EMC Generic Immunity standard AS NZS 4252 1 1994 This is evidenced by the C Tick label Environment for use The device is for use in offices sports facilities and hospitals waiting rooms etc NOTE This device does not have the function of automatic cycling measurement Therefore a part of this device does not conform to IEC60601 2 30 SAFETY PRECAUTIONS N WARNING Grounding To avoid electrical shock connect the power cable to an electrical outlet having a ground terminal for grounding the device Fuse To avoid a fire hazard use only a fuse of the proper type voltage and current rating as specified on the rear panel Never bypass a fuse by shorting across the fuse holder and connectors Service Internal service or adjustment to this device should be performed by a qualified person only Do not disassemble or modify the device Device failure If a device failure occurs promptly stop using it Have the patient remove their arm from the device and secure the patient s safety Turn the power off and disconnect the power cable from the electrical outlet Affix a notice such as Out of Order and Do Not Use on the device Ask the nearest dealer for services To interpret blood pressure me
21. is manual Read the manual thoroughly before use When installing and storing the device E Install or store the device away from moisture m Do not install or store the device where the device may be badly affected by extreme temperature humidity direct sunlight draft dust salinity or sulfur content in the air E Do not install or store the device where chemicals and corrosive or explosive gases are stored or present Install or store the device in a secure level and stable location Install or store the device where adequate power is provided Where a medical 3P electrical outlet is provided Before use m Check to make sure the device operates safely and accurately m Check all cables for proper connection E When other electrical equipment is used nearby at the same time a diagnostic error or dangerous situation may occur Check all connections to make sure they do not interfere with each other E When telemeters are used check that mutual interference will not cause a problem During use m Check the patient and the device during use E f medical or operational problems are found in the device or the patient stop using the device immediately check the status of the patient and take proper action m Do not use the device during MRI scanning Do not use the device on a patient using a heart lung machine E High frequency interference by electrosurgery or energy discharged by a defibrillator may damage the device
22. l the armrest and the front frame in the Front reversed order of removal Slide the cover upward to secure it a VA in position Note Nout COVEN The arm cuff cover is a consumable Purchase it separately y Remove the vinyl rings from the old arm cuff and install them in the new arm cuff cover The vinyl rings are not included with the optional arm cuff cover Fig 18 Arm cuff replacing procedure 21 9 2 Replacing the Fuses Replace the fuses as follows 1 Remove the caps of both fuse holders located on the rear UGE si panel o 2 Replace both fuses with new ones P KE E 3 Re install the fuse holder caps Fig 19 Fuse holders Note Use only fuses of the proper type voltage and current rating as specified on the rear panel and that conform to IEC60127 9 3 Checking the Counter The TM 2655 TM 2655P is equipped with a counter function which indicates how many times the device has measured blood pressure The counter data remains in memory even after the power is turned off Use the counter function to check the frequency of the device use or as the guideline for a periodic cleaning Displaying the counter Press the COUNTER switch The counter data appears in the SYSTOLIC and DIASTOLIC displays for about 5 seconds COUNTER switch Note Do not keep the COUNTER switch pressed It will reset the counter data to zero Ex The
23. measurement count at the figure is 2382 times AUTOMATIC BLOOD PRESSURE MONITO SYSTOLIC mmHg pa m The number of more than 1000 is displayed DIASTOLIC mmHg a H a The number of less than 1000 is displayed ling Ca The symbol is displayed at checking the counter TIME 14 14 22 Resetting the counter Hold down the COUNTER switch for 4 seconds or more The counter data will be reset to zero Fig 20 COUNTER switch 2010 12 13 Printing the counter graph Total 204 Weekly Count Press and hold the COUNTER switch press the PAPER FEED 20 switch to display the current measurement count 110 ank a a o Total The total measurement count of the product Tu We Th Fr Sa Su No Monthly Count Weekly Count The count for last one week Monthly Count The count for last one year Note1 7he Total can not be reset by pressing the COUNTER switch Note2 When the print selection is OFF the counter graph is not printed 9 4 Error Codes Table 9 Error codes Error code Description PE The printer paper has run out Install a new roll of printer paper HU The printer head is up Lower the lever and lock the printer paper in position Err A measurement error has occurred Refer to 10 Troubleshooting 9 5 Maintenance Do not open the device It uses delicate electronic components and an intricate air unit that could be damaged If you can not fix the problem using the
24. s the safety lever located on the front lower part of the device It disconnects the clutch to loosen the cuff O Please wait for about 10 minutes before repeating the measurement 5 SETTING THE CLOCK 5 1 Clock Adjustment Mode The clock is set in the clock adjustment mode The clock adjustment mode display and switches used are as shown below SELECT switch A switch SYSTOLIC mmHg Year n EN ARA DIASTOLIC mmHg a Month Hours Minutes AUTOMATIC BLOOD PRESSURE MONITOR Date o PO Imin The selected digit s flashes Fig 10 Switches for setting the clock and the clock adjustment mode display SELECT switch Table 3 Switches for setting the clock 1 Press to enter the clock adjustment mode 2 In the clock adjustment mode Press to select the unit to be adjusted The selected unit flashes Each time the switch is pressed the flashing unit changes from Year to Month Date Hours and Minutes 3 Press to quit the clock adjustment mode Aswitch Press to change the flashing digit s 5 2 A Setting Example The following is the procedure to adjust the clock to 14 07 April 20 2001 1 N O nO fF W N Press the SELECT switch to make the SYSTOLIC display flash Press the A switch to display 01 for 2001 Press the SELECT switch to make the DIASTOLIC display flash Press the A swit
25. st Gas shock suspension TM 9315A Oo Printer paper 5 rolls AA AA AX PP147 S Oo Arm cuff COVEY na AX 1 33003442 S Power cable cord set eee eee eee rere rere rere eer NAA ANAN AX KO243 Type C Oo Power cable cord set rn AX KO242 Type BF amp Power cable cord set AA AX KO1 1 5 Type A 24 10 TROUBLESHOOTING If the TM 2655 TM 2655P does not function properly or an error code appears try the following corrective actions Table 10 Troubleshooting Problem Check this Corrective action Nothing appears in the display even when the power is turned on Is the power cable connected properly Connect the power cable properly Has a fuse burned out Replace both fuses with new ones Size 5x20mm Rated voltage and current are specified on the rear panel E00 appears in the display E98 appears in the display Does air remain in the cuff After releasing the air from the cuff wait a few seconds and turn the power switch on again The cuff motor did not function or Cuff motor malfunction Contact the local A amp D dealer to repair the device E99 appears in the display The cuff will not inflate Is the arm cuff cover installed with high tension too tight The home position of the cuff cannot be detected Turn on the power switch again Refer to 9 1 Replacing the Arm Cuff Cover to re install the cover The device will not measure Err appe
26. table so that the arm insertion section is at your heart level Insert your arm into the arm insertion section Note Fig 7 Proper posture Be sure to insert your arm to the shoulder Press the START STOP switch to start measurement 2 SYSTOUC mmHg The cuff will be pressurized automatically up to the predetermined pressure Refer to 7 2 Description of the Functions for details Pressurizing SYSTOLIC mmHg Measuring while reducing the pressure When pressurization is complete the automatic exhaust mechanism will gradually reduce the pressure in the cuff Just relax and remain still To next page Fig 8 Measuring procedure 10 10 When the measurement is complete the air is automatically From last page released from the cuff and the cuff will become loose po Results The measurement results appear in the display SYSTOLIC Amia I TM 2655P The measurement results are output to the T PUN A printer DIASTOLIC mmHg AL TM 2655P only Fig 9 Measurement results Note e To stop the measurement at any time press the START STOP switch The quick exhaust will reduce the pressure in the cuff and the cuff will become loose e f the air is not exhausted even when the START STOP switch is pressed press the EMERGENCY STOP switch e f your arm can not be removed from the cuff due to the power failure pres
27. this publication may be reproduced transmitted transcribed or translated into any language in any form by any means without the written permission of A amp D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice WARNING DEFINITIONS The warnings described in this manual have the following meanings N WARNING Important information to alert you to a situation that might cause injury and or damage to your property if instructions are not followed N CAUTION Important information to alert you to a situation that might cause minor injury and or damage to your property if instructions are not followed Note Important information that helps users operate the device COMPLIANCE Compliance with the European Directive 93 42 EEC for Medical Products This device conforms to the following requirements European Directive 93 42 EEC for Medical Products act Medical Products Act European Standards for Electrical Equipment EN 60601 1 General Safety Provisions EN 60601 1 2 and EN 55011 Electromagnetic Compatibility European Standards pertaining to Non Invasive Blood Pressure Instruments EN 1060 1 General Requirements EN 1060 3 Supplementary Requirements for Electromechanical Blood Pressure Measuring Systems This is evidenced by the C Ens mark of conformity accompanied by the reference number of a designated authority This d
28. ure detection method Capacitance type pressure transducer Pressure range 0 300 mmHg Accuracy Pressure 3 mmHg Pulse rate 5 Measurement range Blood pressure 10 280 mmHg Pulse rate 30 200 bpm Pressurizing method Micro pump Air pressure control method Rubber valve ceramic valve Rapid air exhaust system Electromagnetic valve Safety mechanism The electromagnetic valve is released when approx 320 mmHg or greater is detected Communications functions Environment specifications Serial output RS232C level Operating temperature and humidity 10 to 40 C 85 RH or less non condensing Storage transporting temperature and humidity 20 to 60 C 95 RH or less non condensing Physical specifications Overall dimensions 245 W X 322 H X 390 D mm Weight Approx 9 kg 2 PART NAMES 2 1 Main Unit SYSTOLIC pressure display DIASTOLIC pressure display PULSE display TIME display START STOP switch Printer unit TM 2655P only Arm insertion section Arm cuff Armrest Human sensor Functions only for TM 2655P Safety lever EMERGENCY STOP switch Fig 2 TM 2655 TM 2655P whole view Note e The printer unit is available only for the TM 2655P e The human sensor area appears the same for both models but it functions only for the TM 265

Download Pdf Manuals

image

Related Search

Related Contents

Philips RQ1050  NETGEAR N300 WiF Range Extender Model EX2700 User Manual  Samsung SMX-F50SP Priručnik za korisnike  ソーラーシリーズ 施工例集・商品特徴  0086 FR MODE D`EMPLOI AnaConDa  IP Camera CGI User Manual  Q-Logic 2310F Network Card User Manual  Full Product Brochure  Magnavox DP100MW8B A DVD Player User Manual  Ask Proxima C410/C420 User's Manual  