GasCheck 5000is Manual V2.2
Contents
1. Unrivalled Detection Page 10 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Installation Remove all packaging materials then open the storage case and inspect the equipment If the storage case or any other item is damaged notify your supplier and the carrier in writing within three days state the ltem Number and the Serial Number stamped on rear case of the Gas Check refer to Figure6 together with your order number and your supplier s invoice number Do not attempt to use the Gas Check if any item is damaged The batteries if provided are shipped in the spare battery clip and not inside the GasCheck instrument itself Refer to Figure 3 and check that the storage case contains the items listed in Table 3 If any item is missing notify your supplier in writing within three days Description Gas Check with short probe and nozzle fitted Long probe Box spanner Special screwdriver for access to battery compartment Spare battery holder with 4 batteries Table 3 Checklist of items in the storage case 1 Long probe 4 Spare battery holder 2 Box spanner 5 Gas Check with short 3 Special screwdriver probe and nozzle fitted Figure 3 Items in the storage case Unrivalled Detection Page 11 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd
2. Operation Gas Check 5000is both have menu keys for user accessible settings However normal use of the Gas Check would not involve any menus because these are only needed to set up the instrument for your application after which the instrument stores the settings internally when the instrument is switched off The display of information on the Gas Check 5000is depends on whether the instrument has been keyed into the set up menus or is simply being used for leak detection The Gas Check 5000is menu displays utilise 3 lines of information while the normal display uses 2 or 3 lines in each case the lowest line is for status information while the upper line s show s the menu topics or the leak rate values You may like to gain some confidence in the operation by simply using the control keys although if it has not yet been used you should carefully fit the battery clip Please take note of the warnings and cautions when fitting batteries or when in potentially hazardous areas Controls In normal use there is only one key to press beyond the power ON OFF control and that is the ZERO key Refer to Figure 4 Use the control keys as described on next page Unrivalled Detection Page 12 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Control Keys POWER ga e E Press this key to switch the Gas Check POWER on for use When this key is pressed from normal display Gas Check will turn its POWE
3. Item Number A 21500 A 28170 A 31117 Specifying voltage A USA K UK E Euro U Universal and C Car Calibration lon Science offer a calibration service including issue of traceable certification to National Standards Refurbish program lon Science offer a refurbishment service for all customer repairs Unrivalled Detection Page 38 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Instrument warranty and service Warranty Standard Warranty can be extended to up to 2 years on the GasCheck 5000 when registering your instrument via our website http www ionscience com instrument registration To receive your Extended Warranty you need to register within one month of purchase Terms and Conditions apply You will then receive a confirmation email that your Extended Warranty Period has been activated and processed Full details along with a copy of our Warranty Statement can be found by visiting http www ionscience com instrument registration Service At lon Science we recommend that all of our gas detection instruments be returned for service and factory calibration once every 12 months Contact lon Science or your local distributor for service options in your area Find your local distributor by visiting www ionscience com Contact Details UK Head Office lon Science Ltd The Way Fowlmere Cambridge SG8 7UJ UK Tel 44 0 1763 207206 Fax 44 0 1763 208814
4. Email info ionscience com Web www ionscience com Unrivalled Detection USA Office lon Science LLC 33 Commercial Drive Waterbury VT 05676 USA Tel 1 802 244 5153 Fax 1 802 244 8942 Email info ionscienceusa com Web www ionscience com Page 39 of 40 German Office lon Science Messtechnik GMBH Laubach 30 Metmann Neandertal 40822 GERMANY Tel 49 2104 14480 Fax 49 2104 144825 Email info ism d de Web www ism d de www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Manual log Manual Version Amendment Date updated Instrument PC Firmware Software GasCheck 5000 Gas Table updated 16 7 08 V4 16 V3 26 V1 4 GasCheck 5000 Update A A added to back of V4 i e 26 V1 5 ee A GasCheck 5000 Warranty added on cover b hd 16 bs 26 Declaration of Conformity updated Contents updated Warranty and Service added page 40 V1 7 2 Gascheck 5000 AAA Rechargeable Battery 3 12 10 hon 16 hase 26 V1 8 information added Pages 11 12 37 amp 40 5000 oe i V4 A a 26 V1 GasCheck 5000 lero front cover updated diosa MN 3 dell 16 ba 26 V2 0 Instrument registration note added to inside front cover Font and layout also updated GasCheck 5000 Changed Baseefa No on A V4 A a 26 V2 1 Pages 3 and 5 GasCheck 5000 Certification report updated ali hall 16 alll 26 V2 2 Unrivalled Detection Page 40 of 40 www lonscience com
5. 24 A A o yg A eee 24 AN eRe a enero REO Ce strane ee eae ne nner en OE Sen eee ee ane een 24 A o o A dees eee eee 26 PP e ene eee een 2 Weve a ge Cowl 6 CANO ici 27 Apis wale io YEA 6 0 01 a isa 28 Audio ana VISUAL ILO NS rsa iii 29 Prepare ne GaS CNet a 29 PEO AMM O A 31 WIECH OM Ne GaS COG I anida 31 Recommended leak Search procedure cccccccecceccsseceeceeeeecceeceeceeseesuecseseeeseeseeseesseeseeseeseeseeeseeseeseesaeees 31 Leak election ANA SCANIS cies sets Side ac 32 Leak detection at pipe joints A a 32 Data logging interruption fOr MENU ACCESS cooocncoccconccnncnnocnnoncnnncncnnnonnnonncnnnnnnnnrnnnrnnnnnnnnrnnnnnonencnnnnnnnoneninass 32 Switche GaS ChCK sacs sachs ieceeeda aia 32 AUTO SNE OW 0 sc a ccc ed nga ctw ed ee 32 Inspectiand cleat ine Gas Check ncscci os 33 Malena CO sus UI E e ES 34 Replace the Dalene eis eases a 34 All NAJINO a nd Mae eee eee 36 A oe oC ee ee eee eer eee ee ere 37 PS A A See ne ae tet ae es Cade oee easements Heat Seen 37 PIES e a lic 38 ACCESO MES iia dable 38 Gala to Meta ciales 38 RETMIrISADIOOaN oo 38 INSTEUIMENE Warranty ANG SOV ICC aii 39 A on A E E E EE Tee EE eer 39 O O 39 Soale eg Detalla catre 39 E A O A RO 40 Unrivalled Detection Page 4 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Introduction Scope and definitions This manual provides installation operation and maintenance instructions for the Gas Check 5000is
6. Check 5000is use the special screwdriver supplied to undo the captive screw on the batter compartment cover and remove the cover This is located at the end of the Gas Check at position 5 Remove the battery holder If required you can carefully unplug the battery holder so that you can detach the holder from the Gas Check Remove the old batteries 4 from the battery holder Dispose of the batteries Inspect the inside of the battery compartment and battery holder If the batteries have leaked use a cloth dampened with a weak solution of detergent to clean the battery compartment and battery holder CAUTION The Gas Check 5000is will blow a factory fitted internal protection fuse if the battery clip is inadvertently reverse connected The complete instrument will have to be returned to the supplier for fuse replacement Fit four new batteries in the battery holder Ensure that you fit the batteries in the correct orientation Refit the battery holder to the Gas Check On the Gas Check 5000is refit the battery compartment cover and use the special screwdriver to tighten the captive screw and secure the cover in place CAUTION Ensure that the screw is located squarely to the threaded insert before tightening else the battery compartment cover may not be properly secured due to cross threading The instrument will then have to be returned to the supplier for replacement of the Gas Check case and new serial number Switch on th
7. You must use the Gas Check as specified in this manual Read this manual before you install and operate the Gas Check Important safety information is highlighted as WARNING and CAUTION instructions you must obey these instructions The use of WARNINGS and CAUTIONS is defined below WARNING Warnings are given where failure to observe the instruction could result in injury or death to people CAUTION Cautions are given where failure to observe the instruction could result in damage to the equipment associated equipment and process The units of measurement used throughout this manual conform to the Sl international system of units of measurement Description Refer to Figure 1 The Gas Check is a portable battery powered hand held gas leak detector Indications of a leak are shown on the digital display 2 and are also given by a front panel mounted LED flasher 3 and as audible clicks from the rear panel loud speaker 6 The Gas Check 5000is is intrinsically safe and is suitable for leak detection of hydrogen and other flammable gases and for leak detection in hazardous areas of Ex ia IIC T4 rating The Gas Check has a convenient storage and carrying case together with a spare battery holder a long flexible probe for leak detection in areas where access is restricted and a box spanner for probe changing Principal of operation When the Gas Check is switched on a small internal fan draws gas samples through the probe and
8. down into Clock setting Display Clock A Back YT 12 24h dy f Existing Clock setting for Topic Note No X or y at this level because we are not at the end of the tree The full clock settings are shown Operation of the up or down keys give access to the other topics at this level Date format Set time Set date The following frames show the CLOCK sub menus accessed by the A key Clock 12 24hr Display 12 24 Back X F 12h Jy 24hr Y Note X because 12hr not selected but a Y always in the bottom row to show that this is currently selected Clock Date Format Display Date format Be rs T dd mm Y J dd mm Y Note Both v in this example because of the same setting This also denotes the end of the tree Other option within this topic is mm dd Unrivalled Detection Page 17 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Clock Set Time Display Set time Back TF Set hour Oa existing Clock setting for Time Other options within this topic are Set min Synch RTC Clock Set Time Set hour A press on the ay takes us down into this next level Use UP and DOWN keys to change hour value Press PROCEED when this is what you require and the clock will proceed to minutes then to seconds If you are only changing hours for Daylight Saving Time changes then once PROCEED has been actioned the Back key can be used to return towards the no
9. format This is just an information menu only but the Y serves as an indicator that it is an end tree situation Also notice how the ay has been removed for this item Data Control Single cont Enter this by a a press Display single cont ack ay T Single 4 J existing setting ye The choice is given to toggle the Single or Continuous mode of data taking using the Logging Settings CAUTION Please reset the Gas Check clock date and time settings required for data acquisition before proceeding to take data if this is to be used as a real time record rather than a logging sheet Display Style Gas Check 5000is allows the user to select a preferred style of presentation either a bar graph with an exponent range indicated or a digital display with a similar exponent range The style applies to both the normal display mode and the menu settings for alarm level Enter this by a J oress Display T Style lt q Back X RJ l Bar existing setting Unrivalled Detection Page 21 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation The option is given here as per clearing Last data or All data is for information purposes to aid selection The choice is given to toggle the Bar at this level or to scroll through to see the other option Numerals Display Size Gas Check 5000is allows the user to select a preferred size of the display style presentation as a larger or a
10. into the Gas Check The Gas Check measures the thermal conductivity of the gas samples When you first switch on the Gas Check using the power I O key or when you press the zero key The Gas y Check zeros the leak rate display If further samples have a different thermal conductivity from the zeroed sample the Gas Check calculates and displays the difference as a leak rate This means that you can use the Gas Check to detect any gas other than that in which the Gas Check is zeroed Unrivalled Detection Page 5 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Introduction Applications You can use the Gas Check on many types of applications including those listed below Quality assurance testing on manufactured component seals Laboratory applications such as the detection of leaks from gas chromatographs from mass spectrometers from gas cylinders and fittings Industrial applications such as the detection of leaks from gas installations in cylinder receiving rooms from pipeline assemblies the detection of leaks from stored gases and the detection of vapours released from stored chemicals Medical applications such as the detection of leaks from anaesthetic gas bottles and pipelines and leak testing of membrane materials glove boxes and so forth Valve emission evaluation to EPA Environmental Protection Agency method 2 1 Figure 1 The Gas Check 5000is Leak testing of pipeline
11. smaller size for the bar graph or numerals and the exponent range value The size only refers to the normal display settings not to the menu settings for the alarms Enter this by a Apress Display Size Back Y Large X Sd y existing Inst setting Y The option is given here as per clearing Last data or All data is for information purposes to aid selection The choice is given to toggle the Large at this level or to scroll through to see the other option Small Note Similar X and Y indications are followed in both Size and Style Unrivalled Detection Page 22 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Gas Group Enter this by a ay press Ar CH4 CO2 H2 He O2 R11 R12 R22 R134a R1301 SF6 Group 0 Group 1 Group 2 Group 3 Group 4 Named gases have a polarity feature that is NOT available for the Gas Groups on Gas Check 5000is the response after zeroing is ONLY allowed to be in the same sense as the polarity of the named gas Gas Group Back X Y Ar Sd y existing gas setting Y Note Similar X and Y convention is followed Language Enter this by a ay press Language Back J English 4 y existing setting Y The option is given here as per clearing Last data or All data is for information purposes to aid selection The choice is given to toggle the English at this level or to scroll through to see what the other options are which in the ins
12. GasCheck 5000is Instrument User Manual V2 2 Unrivalled Detection gt Register your instrument online to receive your extended warranty www lionscience com Register your instrument online for extended warranty Thank you for purchasing your lon Science product The standard warranty of your instrument can be extended to up to five years on the Tiger and two years on other lon Science instruments To receive your extended warranty you must register your instrument online within one month of purchase terms and conditions apply Visit www ionscience com instrument registration Part number 861265 GASCHECK 5000IS MANUAL lon Science Ltd Declaration of conformity Manufacturer lon Science Ltd The Way Fowlmere Cambridge England SG8 7UJ Declare that the products Gas Check 5000is comply with the requirements of Directives 89 336 EEC EMC Directive 76 117 EEC Equipment and protective systems intended for use in potentially explosive atmospheres in conformity with the following standards and specifications Standards BS EN 13908 2002 ATEX Application of quality systems BS EN 50014 1998 ATEX general requirements BS EN 50020 2002 ATEX Intrinsic safety i BS EN 61010 1 2010 Safety reqs for measurement control amp lab equipment use BS EN ISO 9001 2008 Quality management systems BS EN 61326 1 1997 EMC Equipment for measurement control and laboratory Product marking is as fo
13. HECK 5000IS MANUAL lon Science Ltd Maintenance Replace the batteries WARNING Replace the batteries in the Gas Check 5000is in a safe area where there are no flammable gases If you do not there may be a risk of fire or explosion WARNING Use only dry cell non rechargeable AA batteries or lon Science AAA Rechargeable Battery Pack part No A 31117 in the GasCheck 5000is The internal resistance of unpotted rechargeable batteries is too low to maintain a T4 safety rating The recommended batteries are specified in Battery Data CAUTION The Gas Check 5000is will blow a factory fitted internal protection fuse if the battery clip is reverse connected The complete instrument will have to be returned to the supplier for fuse replacement Ensure that the rubber sleeve on the female part of the battery holder is correctly fitted when you fit new batteries If you do not you could refit the holder in the Gas Check 5000is with incorrect polarity which will blow an internal fuse and make it inoperative Note A spare battery holder is supplied in the case Refer to Figure 6 Turn over the Gas Check so that the battery compartment is at the top see detail A Gas Check 500015 1 Catch 4 Battery compartment 2 Battery compartment cover 5 Serial number 3 Captive screw Figure 6 Replace the batteries Unrivalled Detection Page 34 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Maintenance On the Gas
14. R off remembering all last settings except back light reverts to OFF state When in menus pressing this key brings Gas Check straight back into the normal display WITHOUT any of the settings changes being implemented ZERO BACK Just press zero to adjust for background levels reset the peak hold memory or after changing a gas group In menus successive presses of this key provides a simple step by step route BACK to the normal display PROCEED And Press the proceed key to reach the menus that allow you to control special settings of the Gas Check 5000is DOWN na Press or hold the Down key to scroll up through menus or to raise a numerical value Release when desired UP T Press or hold the Up key to scroll down through menus or to lower a numerical value Release when desired INFO il Press the Info key to start data logging as per the parameters last set via the data control menu Unrivalled Detection Page 13 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Ooperation Menu topics and corresponding displays The first pressing of the PROCEED key switches the display into the Gas Check 5000is menus These menus are shown by the relevant menu topic on the top line and the available choice and the current settings on the middle and bottom display lines The UP and DOWN keys scroll round the menus choices Some of the menus show a tick or cross at the right hand side to indicate the current se
15. X changes to a VY and then the Gas Check changes to a blank display but with the words Data transfer for the duration of the transfer or time out Unrivalled Detection Page 19 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Data Control Logging settings Enter this by a aj press Display Logging settings do wt Reset settings y Only if you want to return to default existing logging settings Y No Samples x Interval s Y Use the up or down scroll keys for other Topic selections Interval No Samples Data Control Logging settings Interval Enter this by a ay press and use the UP and DOWN keys to select a new repeat interval for data logging Press PROCEED key to enter this value in seconds BACK key to move back towards the normal display Display Sample time Gack X T J existing setting Y Data Control Logging settings No Samples As per Interval above This sets up the total number of samples per record for each press of the INFO key When the Single mode of data logging is used this menu is NOT accessible as only ONE sample is taken Unrivalled Detection Page 20 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Data Control Memory Used Display Data Control lt Back Y Memory Used Y Display age of memory used yv Note This is an unusual case but we are still trying to follow the same
16. and O2 oxygen For ease of use the Gas Check groups gases having similar thermal conductivity into one of the 5 gas groups shown in Table 1 Each group has a calibration curve stored in memory Although the Gas Check will detect ALL gases with a different thermal conductivity to air Table 1 lists those 12 common tracer gases that have been tested and assigned gas groups If the gas or gas mixture that you seek is not included in these groups select group O which is the most sensitive The Gas Check will then still find leaks quickly even though the measured values will not be displayed accurately Some gases give a negative response Minimum response time Short probe fitted Less than 1 second Long probe fitted About 9 seconds Recovery time About 1 second Maximum detection level with the short probe and nozzle fitted 20 ml sec Data logging RTC This is a nominal device for convenience and can be reset by a user if more accurate timings are required Note If the probe has been changed then the instrument should be recalibrated for more accurate measurements The readings given by the long probe can only be considered as indicative Unrivalled Detection Page 8 of 40 www lonscience com GASCHECK 500015 MANUAL lon Science Ltd Technical data Minimum detection levels see Table 1 The nozzle when fitted provides automatic 10 fold dilution of the gas stream entering the Gas Check TCD except when the ppm display is
17. ant to use the short probe or the long probe and whether you want the nozzle fitted as follows Use the Gas Check with the short probe and without the nozzle to detect small leak rates of approximately 1 x 10 4 ml sec and lower Press the nozzle key to clear the nozzle fitted indicator For large leaks use the Gas Check with the short probe and the nozzle fitted The nozzle extends the detection range for more accurate measurement of large leaks Press the nozzle key to set the nozzle fitted indicator If the area you want to leak check has restricted access remove the nozzle and the short probe and fit the long probe The long probe is flexible and can be bent in a gentle arc so that it can reach areas that would otherwise be difficult to reach Press the nozzle button to tell the Gas Check that the nozzle is not fitted when you fit the long probe Note that response time increases significantly when the long probe is fitted The nozzle simply pulls off and pushes onto the end of the Gas Check Unrivalled Detection Page 29 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program Remove the Gas Check nozzle or change the probe if required In some applications the grey probe cover may restrict access to the area under test the grey probe cover can be removed by pulling it off way from the instrument When the grey cover is removed great care should be taken when using the lt E instrument and the following pre
18. autionary Note The instrument has been factory calibrated using the standard short capillary which sets a certain flow rate into the detector The long capillary by nature of its construction has a different flow to improve the time response in detection This will change the calibration of the instrument when the long probe is used instead of the short capillary Thus the long probe is only to be used in finding leaks in difficult places where the standard short probe cannot reach The readings given by the long probe are only qualitative and the reading given by the display is only to be taken as being relative to another value given by another leak site while using the long probe Unrivalled Detection Page 30 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program WARNING Ensure safety guidelines are adhered to when using compressed air lon Science Ltd takes no responsibility for injury or damage caused by misuse of compressed air equipment Switch on the Gas Check Press I O key and wait for Gas Check to bleep amp flash the LCD to light up and show lon Science Ltd Cambridge England Then show Gascheck 5000is SW Version 1 49 Text Version 1 00 Then show Zeroing Detector After a short while the LCD display will then revert to the display format previously set up for the leak rate which can be either the bar graph or the digital display the latter in normal or large numeric display format When not d
19. ays Units Leak rate units At this level the following key functions are available Select the subsequent topic from the main menu by the PROCEED key aJ Leave the menu and go back to normal operation by the Zero Back key A Scroll through the menu by use of the up and down keys Escape back to the normal display without implementing any menu changes by pressing the 1 0 key once Unrivalled Detection Page 14 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation ABC Display Main menu lt lt Back Je ABC X symbol space status line The X indicates that it hasn t been selected The status line doesn t have the lt symbol If the Proceed key is pressed then the X turns to a Yand the lt symbol appears on the status line Press Proceed to reverse Alarm Level The alarm level format is in the same style as the display type see below for display style options The alarm level is linked to gas group setting such that if the Gas Group changes the alarm is not set the status symbol is removed from the display although the last alarm level value is retained for reconfirmation This is to ensure the Gas Check avoids incorrect alarm levels on a Gas Group change Press 49 to jump down into Alarm Level setting Display Alarm Level Back wt 2x10 lt Y existing Alarm level setting Y Note When the alarm is set to maximum the alarm function is disabled and the Main menu Alarm level
20. b menu has the following topics and are accessed by pressing the y key Language Menu Zone Program Back ay T Language X y existing 1st line of Lang Text v Note Similar X and Y convention is followed Pressing the PROCEED key without an IR Link communication will result in a delayed reset of the unit The other two selections will also be displayed because of different information content on the third line Program Inst qe Sd Y Menu Y Menu status Menu status will only have a S W test to indicate whether it is in the non programmed normal default settings or has any of these settings changed THUS the bottom line will have one of two messages Factory Menu Y Custom Menu Y Pressing the PROCEED key without an IR Link communication will result in a delayed reset of the unit Note Similar X and Y convention is followed Program Inst Back X T Zone Sd y Zone status Zone status will only have a S W test to indicate whether it is in the non programmed zone or has the zone option programmed within it THUS the bottom line will have one of two messages Zone Set No zone Y Pressing the PROCEED key without an IR Link communication will result in a timed our rest of the unit Note Similar X and Y convention is followed Unrivalled Detection Page 25 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program Range Press y to jump down into the range topic Display Rang
21. cautions noted Do not remove the semi transparent probe sleeve this sleeve ensure the metal inner probe remains 1 mm from the surface of the test area avoiding accidental dirt and moisture ingress Do not touch the probe and especially the brass sensor housing Heat from fingers can result in significant changes in signal causing false readings Avoid placing the probe on wet or dirty surfaces blockage of the probe will result in instrument failure 2 INP Some application may require a longer probe to gain access to pneumatic joints and seams to be tested the GasCheck is also supplied with a 300mm probe that can be changed by the user To change probes carefully follow the following steps Switch the instrument OFF Remove the outer grey probe cover O Place the box spanner supplied with the instrument over the existing probe so the spanner fits over the brass nut unscrew the probe counter clockwise direction lt lt Withdraw the spanner and probe assembly O To refit the short or the long probe follow the steps above but in reverse order CAUTION When using the box spanner to tighten the probe nut ensure the nut is firmly tightened only by the hand Do not use additional tools as the tread may become damaged Should a probe become blocked use dry clean compressed air to blow out the blockage from the instrument end of the probe when the probe has been first removed off the instrument C
22. display will show Not set and no bell will shown once the Gas Check is put back in the normal display Unrivalled Detection Page 15 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Audio When the audio feature is selected the tick frequency of the Audio and Visual indicators is linked to the size of the leak rate displayed when Audio is not selected there is no link Display Main Menu Back ay Audio lt symbol space status line X The X indicates that it hasn t been selected The status bar doesn t have the lt symbol If Proceed key is pressed then the X turns to a Y and the loudspeaker symbol is placed down onto the third row Backlight The backlight control allows the user to switch on the backlight or select a TIME setting Press ay to jump down into Backlight setting and then the PROCEED key or the UP and DOWN keys Display Backlight Gack Y at On Off Status shown V will put the Backlight ON existing backlight TIME setting If the scroll up or down keys are used from this start point then the following options are shown 20s min 3min 10min 30min Note If time displayed the existing backlight time then a V is shown at the end of the 2nd row else a X Backlight ack Y 3mins lt Oy Existing backlight setting Unrivalled Detection Page 16 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Clock Press ay to jump
23. e Back Y JT Auto Range Existing status shown Y existing range status Y This is Auto Range for Y above If the scroll up or down keys are used from this start point then the following options are shown Manual x103 Manual x102 Manual x101 Note The exponents are only an example for the units of ppm selected Units Press ay to jump down into the units topic Display Units ack 2 ayt ppm Existing status shown Y y existing units status This is ppm for Y above If the scroll up or down keys are used from this start point then the following options are shown ml min ml s ft3 yr ppm sccm mm3 s mm3 min Some Gas Check versions have a language selection where the units of ml are shown as cc instead because this cc unit is more familiar than the Sl based ml unit of identical volume Unrivalled Detection Page 26 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program Display General The Gas Check LCD matrix display is arranged to show the leak rate and other relevant status information in a consistent fashion during the normal use and also during user changes to the settings of the application In normal use the upper portion of the LCD matrix is used for a large size digital representation of the leak rate and status information is shown as symbols appearing in the lower portion In response to Gas Check key presses written text will appear across the middle of the matrix display
24. e Gas Check to ensure that the batteries are correctly fitted If the display remains blank when you switch on the Gas Check Unrivalled Detection Page 35 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Maintenance Fault finding Fault symptom Action o Unit will not switch on No dead batteries fitted Try another battery set pen ey Unit will not switch off Software locked out Remove amp replace battery clip safely Battery symbol appears Battery state low Replace batteries safely Back light drains battery Excessive use of backlight Avoid or reduce use of backlight life Unit will not zero Software locked out by Remove amp replace batteries safely hardware LCD appears blank dim Contrast control setting wrong Turn off amp on press amp hold contrast Bar graph number do not Peak hold selected Press zero or change Peak Hold change state Range does not change Range not set in Auto range Press amp hold Range until Auto range He range will not show 5 Unit with nozzle ON and or Reset via Nozzle and Range keys range Gas group is wrong Gas Group not set correctly Press amp hold Gases until shown OK correctly correct cal recalibration Unit will not zero correctly Sensor contaminated Leave to stabilise Blocked probe Probe blocked by solid or Remove probe clean check now OK liquid Stalled motor Motor failed amp probe not Return to supplier for repair blocked Instrument read
25. etecting leaking gas Gas Check still flashes and bleeps periodically to indicate that it is ON Gas Check remembers the settings on instrument power off so that the operator rarely if ever has to alter any of them The settings for gas groups range audio bleep peak hold and nozzle are shown on the LCD The response of Gas Check can readily be checked against an lon Science CalCheck Alternatively the tracer gas supply can be suitably fitted with an lon Science Ltd Standard Calibrated Leak Recommended leak search procedure General procedure Note When you use the Gas Check do not allow dirt or debris to enter the end of the probe or nozzle If the probe or nozzle is blocked the Gas Check will not operate correctly The following test procedures assume that you have fitted the short probe The response time of the Gas Check increases from about 1 second or less when the short probe is fitted to about 9 seconds when the long probe is fitted If you use the long probe for leak detection ensure that you move the end of the probe much more slowly along the area you leak test Ensure the unit under test contains a reasonable percentage of a trace gas that is different from normal air Unrivalled Detection Page 31 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program Move the end of the probe or nozzle around the area of a suspected leak and look at the display or listen to the audible clicks or note the visual LED
26. f the probe 2 or 7 through the middle of the box spanner Push the box spanner onto the nut 3 and turn the box spanner anticlockwise to undo and remove the probe Inspect the Gas Check If the case of the Gas Check is cracked is cracked or dented we recommend that you do not continue maintenance but return the Gas Check to your supplier for inspection and calibration Wipe the Gas Check with a clean lint free cloth moistened with a weak solution of detergent Carefully wipe the outer faces of the probe nozzle fitting to remove any small particles of dirt or debris Use a Suitable tool to push any blockage out of the end of the nozzle then wipe the outside of and the inner face of the wide part of the nozzle Inspect the probe The hole in the short probe is approximately 0 47mm 0 0185inches in diameter We recommend that you push stiff wire of 0 25mm 0 01inches diameter or less into the hole in the probe to remove any blockage The long probe is too long to be effectively cleaned If the probe is blocked we recommend that you obtain a new probe available as a spare Refit the required probe to the Gas Check Place a light wipe of thin petroleum jelly around the inner face of the wide part of the nozzle then fit the nozzle to the Gas Check Place the Gas Check and nozzle and the long probe if just removed in its storage case Store the case in suitable conditions Unrivalled Detection Page 33 of 40 www lonscience com GASC
27. flash rate on the Gas Check to identify the location of a leak Specific procedures for leak detection along seams and around pipe joints are described in the following sections When required Use the peak hold and Auto Ranging modes to continue to check for leaks but to display the largest leak detected If you want to zero the Gas Check move the end of the probe or nozzle away from any possible leak into a clean air zone then press the manual zero key Note that leak sensitivity depends on the ambient conditions If you want to accurately measure a leak compare the indicated leak with a calibration standard such as a portable reference leak When the Gas Check indicates that a leak has been found move the probe or nozzle back to determine the precise location of the leak The leak is located where the largest leak size is indicated on the display and where highest frequency clicks and visual flashing rate are emitted by the Gas Check on the current range Leak detection along seams Move the end of the Gas Check probe or nozzle along the length of the seam so that it just rests on the seam The speed at which you move along the seam depends upon the size of the leak that you search for If the leak is large you can move fairly fast along the seam at approximately 25 mm per second If the leak is small move slowly along the seam at approximately 10 mm per second Leak detection at pipe joints Use the same procedure as f
28. giving user confirmation of the changes to settings while status symbols will also change consistently Examples are Zeroing detector Nozzle On Peak hold On Leak rate indication Leakage is displayed as a single signed digit x 10 signed range for ease of interpretation and calibration Note that some gases eg CO2 may cause a negative leak indication after Gas Check zeroing in clean air Examples are Leak rate for a positive gas 5 x 10 5 ml s Leak rate for a negative gas 3x10 4ml s Positive Positive Negative Positive Negative Negative Negative Negative Positive Negative Positive Positive Note that for the named gases the response of the Gas Check 5000is series after zeroing will be only permitted in the same sense as for that named gas e positive for helium negative for R134a Unrivalled Detection Page 27 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program Other display symbols Small display indicators show zeroing nozzle fitted peak hold and low battery voltage Zeroing This symbol appears whilst the Gas Check is zeroing then it disappears Nozzle This symbol indicates that the Gas Check is being used with the Nozzle on Peak hold This symbol indicates that the Gas Check display is held to the maximum A Low battery This symbol indicates that the battery pack is lower than required for use gt Alarm bell This symbol shows that the alarm level has been se
29. ings noisy Tighten nut slightly Instrument readings noisy Test environment Move test into clean area contaminated 5000is will not switch on Internal fuse blown Return to supplier for repair Table 5 Fault finding Unrivalled Detection Page 36 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Storage and Disposal Storage If you will store the Gas Check for a long time we recommend that you remove the batteries Refit the Gas Check in its storage case and store in dry cool conditions When required for use unpack Disposal Dispose of the Gas Check components and used batteries safely in accordance with all local and national safety and environmental requirements Some of the Gas Check carry case plastic material is recyclable Unrivalled Detection Page 37 of 40 www lonscience com GASCHECK 5000IS MANUAL Spares and accessories lon Science Ltd Order spare parts from your local agent or distributor When you order please state for each part required Serial Number on the back of your Gas Check refer to Figure 6 Item Number and description of part Spares Spare Long probe Short probe Replacement battery holder Nozzle 5000is Box spanner 5000is screwdriver Case Manual Accessories Accessories CalCheck Specifying Gas Leak rate and pressure IR Data Link AAA Rechargeable Battery Pack Item Number A 8045 A 8043 A 08139 08024CPCO A 08029 4 TS 01 31164 28053
30. joints gaskets chamber windows and so forth A Gas Check with short probe and nozzle fitted B Long probe 1 Nozzle 4 Tactile keys 2 Display 5 Battery compartment 3 LED flasher 6 Loudspeaker Unrivalled Detection Page 6 of 40 www lonscience com GASCHECK 5000IS MANUAL Technical Data lon Science Ltd General Operating temperature range Gas Check 5000is Storage temperature range Materials of construction Storage case O to 40 C 32 to 104 F 25 to 70 C 13 to 158 F Polypropylene with polyester foam insert Gas Check Polyurethane casing with a metal bracket Dimensions Storage case 420 x 320 x 97 mm 16 0 X120 Xalo Gas Check see Figure 2 Mass Complete storage case 1 6 kg 3 5 Ib Gas Check 0 5 kg 1 0 Ib Ingress ratings Minimum IP20 i i q A Length of short probe and nozzle Length of long probe 300 mm 12 Figure 2 Dimensions mm Unrivalled Detection Page 7 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Technical Data Performance Detector type Dual micro volume thermal conductivity cell Gases Detected All gases and vapours having a different thermal conductivity to the ambient air on which it was zeroed The larger the difference the greater the sensitivity so that Gas Check is not highly sensitive to the gases normally found in large concentrations in the ambient air e g N2 nitrogen
31. k 5000is Do not modify the GasCheck 5000is equipment to allow any recharging in the instrument AAA Rechargeable Battery Pack part No A 31117 This battery pack has been specially designed to be used with lon Science s intrinsically safe gas detection instruments for example PhoCheck s GasCheck s FirstCheck s and Hydrosteel 6000 is ATEX approved Please refer to the AAA Rechargeable Battery Pack User Instructions supplied with the batteries For Intrinsic Safety T4 Rating in the Gas Check 5000is AA rechargeable batteries must not be used in the battery holder There is no provision for the recharging of batteries within the GasCheck 5000is therefore please use the lon Science charger supplied with the AAA Rechargeable Battery Pack part No A 31117 For alkaline types MN1500 or LR6 type use only Duracell Standard Procell or Ultra or Ever Ready Energiser batteries Alternatively use any zinc carbon types R6PP or R6S Number of batteries required 4 Battery type Alkaline size AA Recommended batteries LR6 R6 or MN1500 Average operating life see Table 2 AAA Rechargeable Battery Pack 1 Average operating life see Table 2 Battery life hours Gas Check 5000is LR6 R6 Battery type MN1500 R6PP AAA Rechargeable Battery 12 Pack i Preferred battery types Table 2 Average battery life The use of the back light will dramatically shorten the battery life Do not change batteries in a hazardous area
32. llows Gas Check 5000is carries the Intrinsic Safety Marking to class E 11 2 G Ex ia IIC T4 20 C lt Ta lt 60 C Certificate Number Baseefa O2ATEX0093 Gas Check 5000is carries the CE mark 1180 On behalf of lon Science Ltd declare that on the date this product accompanied by this declaration is placed on the market the product conforms to all technical and regulatory requirements of the above listed directives Name Mark Stockdale Position Technical Director Signature Date 20th January 2010 Unrivalled Detection Page 3 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Contents Declaration OF CONTO ius li is 3 o A A A 5 Scope ana Ce MUONS cx das 5 APPICAMONS A ie E a E biaieacutetameencines 6 Technical Data nia dido da 7 Sener taa 7 Peron ace curo 8 A cated oo a eccteseabohinc aa autwe E bane Gsan Minamata umes 9 OPEN a 12 SON TO IS sidad 12 COMOROS oe aos bee sb a 13 Menu topics and corresponding dISPlayS oocooccoccocnconconcocnocnconcononnnnnnnononnonnnonnnnnnnnnnnnnnnnnnonnonnnnnnnoninnnnones 14 MAMEN Bee ea teen thee tee o O othe he ee ee eee eee 14 O EE EAE E AT A AES 15 O A o A E AE A ee eee ens AAE AE E E A ETN 15 A Oeae a ae ee ee a eT te eee eee 16 A O e geen Eye gee ere ee eae et eee ee ete 16 COCK 20 OLEO ee men In eer Ce ee eR ee re ee ae eer eee 17 DARC ONTO ere elo arn oR Sc o A E O 19 EF EE 0 PP o A 23 A O 0 II E 23 a yera AAR N ROO nn A A ee cle em
33. or leak detection along seams however move the probe or nozzle around the joint at approximately half the speed Light gases such as hydrogen and helium diffuse quickly into air so an apparent leak on the side or top of a joint may actually be a leak at the bottom of the joint Data logging interruption for menu access Gas Check 5000is settings are protected during data logging Press the INFO key to enter a menu dialogue Switch off the Gas Check When the Gas Check 5000is has been left for a few seconds without any key press then press the I O key The Gas Check will ignore I O presses during the instrument power up sequence and during any zeroing Auto Shut Down The unit will auto shut down after approximately 10 minutes if the keys have not been pressed and the signal has been less than half full scale of the range selected Unrivalled Detection Page 32 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Maintenance Inspect and clean the Gas Check CAUTION Do not clean the probe while it is attached to the Gas Check and do nat try to clean the sample hole inside the probe nozzle fitting which leads to the detector If you do you can damage the Gas Check Switch off the Gas Check Refer to Figure 5 If the nozzle is fitted hold the nozzle 1 by its wide base and gently twist and pull it to remove it from the Gas Check With the small end of the box spanner towards the Gas Check guide the end o
34. rmal display mode with the required changes stored Display Set hour Back T SH 09a existing Clock setting for Time Clock Set Time Set min As per Set hour Clock Set Time Synch RTC This allows users to confirm that the minutes setting conforms to expectations and allows set to O seconds When the synchronisation time is shown and the minutes are incorrect press Back to revert to set minutes Pressing PROCEED will action the synchronisation to the displayed time and then revert to set time frame Unrivalled Detection Page 18 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Data Control This sub menu has the following topics and are accessed by pressing the J key Clear data Data transfer Loggingsettings only accessible in continuous logging mode Memory used single cont In most of these menus the third line shows current settings or values and also VY or X on the middle line Data Control Clear Data Enter this by a ay press Display Clear Data Gack y All data X J Note An X or y will be required following usual convention on the 2nd line Use the up or down scroll keys for other end tree selections Last data X Data Control Data Transfer Display Data control Back rs Data Transfer X display No of files Note An X or VY will be required following usual convention on the 2nd line When Data is being transferred the
35. selected The Gas Check nozzle should always be removed for assessing gas concentration levels in volume spaces smaller than about 1 litre when you are using the ppm display Note1 the minimum detection levels given in Table 1 apply when the Gas Check is used in clean laboratory air conditions When used in conditions where the ambient atmosphere contains clouds of the gas for which you leak test the minimum detection levels may be obscured Note2 An offset of up to 10 times the detection levels given in Table 1 may be experienced in the presence of electromagnetic interference 320 30 MHz at the levels specified in EN 50082 1 Table 1 Smallest Leak Detection Levels Positive 1 5x 10 ml sec Positive 2 0x 10 ml sec Negative 5 4 x 10 ml sec Positive 4 9x 10 ml sec Negative 2 1 x 10 ml sec Negative 5 1x 10 ml sec Negative 4 4x 10 ml sec Negative 8 0 x 10 ml sec Positive 5 8 x 10 ml sec Negative 7 0 x 10 ml sec Positive 6 3 x 10 ml sec Positive 5 7 x 10 ml sec Battery data Note1 the energy storage capacity of batteries and even their T4 intrinsic safety rating is affected by ambient temperature The data in Table 2 is for an ambient temperature of 20 C 68 F Note2 the use of the back light and the IR Data Port will dramatically shorten the battery life Unrivalled Detection Page 9 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Technical data WARNING For GasChec
36. t up for use Ji A ABC This symbol shows that the Automatic Background Compensation is applied x Auto Range This symbol shows when the Gas Check leak rate is not in a fixed range Hp Data Logging This symbol shows continuous Data Logging is in progress Data Logged This symbol shows continuous Data Logging was completed each dash 10 Audio This shows that the audible visual leak rate indication is ON co Exponent This symbol lies below the leak range exponent in the bar graph mode Negative This shows the leak rate is negative EA Bar graph The leak reading shown below is 2 x 10 Exponent range shown Unrivalled Detection Page 28 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program Audio and visual outputs The Gas Check 5000is series have an internal audio sounder and a flashing LED indicator They both operate together when the detected leak rate climbs from the zero indication towards the maximum on the leak rate range They also operate once to confirm a key press Prepare the Gas Check Select which probe to use Note Whenever possible use the Gas Check with the nozzle fitted The nozzle helps to protect the short probe from blockage or damage You cannot use the nozzle with the long probe Open the storage case and remove the Gas Check The Gas Check is supplied with the short probe fitted and the nozzle fitted a long probe is supplied in the storage case Determine whether you w
37. tance is one 2nd Programmed Language English this is the normal factory alternative Note Similar X and Y convention is followed Unrivalled Detection Page 23 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Operation Nozzle Display Main Menu ay Nozzle X symbol space status line The X indicates that it hasn t been selected The status line doesn t have the nozzle symbol If the PROCEED key pressed then the X turns to a and the nozzle symbol is placed down onto the third row Peak Hold Display Main Menu ay Peak Hold X symbol space status bar The X indicates that it hasn t been selected The status bar doesn t have the Peak Hold symbol If PROCEED key pressed then the X turns to a and the Peak Hold symbol is displayed in the third row Power off This sub menu has the following topics accessed by pressing the a key Manual Automatic Power off lt q green Y existing setting Y Manual will ensure the user has to press the I O key to turn off Gas Check Automatic waits approximately 10 minutes from last key press with signal below 50 of selected range before switching off Gas Check Manual will be useful for those users requiring continual operation for data logging purposes or leak ups Similar X and Y convention is followed as for other menus Unrivalled Detection Page 24 of 40 www lonscience com GASCHECK 5000IS MANUAL lon Science Ltd Program This su
38. ttings or choices The PROCEED key selects the current topic and in some menus pressing this will change the tick for a cross or a cross for a tick and also alter the instrument running status symbols displayed on the lowest line This has been found to help customers readily understand the menu selection for new instrument settings Basic Display Details Top Line Name of Menu or Topic lt Back Middle Line Operation Keys Currently user adjustable Bottom Line Instrument running status or Current Setting of Topic Where the menu structure has an underlying sub menu structure for settings the bottom line is blanked Main menu ABC Automatic Background Compensation not normally used Alarm Level Setting for the Audible Visual Alarm Audio Mute the leak rate Audible Visual Backlight Controls for the Back light display Clock Data logging clock functions Data Control Functions for data control including download via IR Link Display Bar graph or numeric display functions Gas Group Choice of Gas Group or named Gases Language Specialist option instead of English etc Nozzle Switch for whether nozzle is fitted or not for leak rate calibration purposes Peak Hold Switch for storing maximum reading on display between each zero key press Power off Controls whether or not Gas Check shuts down automatically after fixed time Program Specialist functions to set up Gas Check 5000is via IR Link and software Range Autoranging or fixed range displ