1. 2 Unit Button Press and hold then release when desired unit is displayed g grams custom Watts 3 7 Segment 6 digit display Displays readings 4 Stability Indicator Displays once the scale is stabilized and ready to take a measurement TESTING 1 Place the USP 50SP on a stable and level surface Avoid air currents and mechanical vibrations Level the unit 2 Loosen the transducer positioning clamp and move it out of the way Position the tank on the rubber circle 3 Fill the test tank to 1 4 inch 6 4 mm below the top of the tank liner with fresh degassed water at a temperature of 24 0 3 0 C 75 2 5 4 F 4 Plug the AC Adapter into a power outlet and plug the cord into the power jack at the rear of the unit unless operating from battery power 5 Lower the cone target into the concentric target support sleeve located behind the test tank small tube inside of larger tube while simultaneously placing the cone target into the tank If the cone can swing in an arc it is not down far enough Tip the rod back and forth slightly to fully engage the rod Press the ON Zero Off button to zero the unit 6 To avoid introducing air into the degassed water insert the transducer at a 45 angle Then position so it is facing the target Vs to 1 4 inch 3 2 to 6 4 mm below the water level parallel to the water surface and directly above the center of the cone Check the transducer surface for un
2. A UNIT OF GROUP INTERNATIONAL INC DIGITAL ULTRASOUND WATTMETER 5 US ULTRASOUND WATTMETER USP 50SP USER MANUAL BC BIOMEDICAL USP 50SP ULTRASOUND WATTMETER TABLE OF CONTENTS WARNINGS CAUTIONS ii INTRODUCTION cursan 1 TESTSE UP 2 een ee et 4 TESTING zu e 5 MANUAL en set laa ea lern 7 EIMITEBWARRANTrE ee a sn 7 SPERIFISATIONS zer 2 een ran en a Nenn 8 NOTES se le a au N 9 WARNING USERS The USP 50SP is for use by skilled technical personnel only WARNING USE The USP 50SP is intended for testing only and should never be used in diagnostics treatment or any other capacity where it would come in contact with a patient CAUTION MODIFICATIONS The USP 50SP is intended for use within the published specifications Any application beyond these specifications or any unauthorized user modifications may result in hazards or improper operation CAUTION SERVICE The USP 50SP is intended to be serviced only by authorized service personnel Troubleshooting and service procedures should only be performed by qualified technica
3. AL OR CONSEQUENTIAL DAMAGES NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR ASSUME ANY LIABILITY REMEDIES THE PURCHASER S SOLE AND EXCLUSIVE REMEDY SHALL 1 THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS WITHOUT CHARGE 2 AT THE OPTION OF GROUP INTERNATIONAL INC THE REFUND OF THE PURCHASE PRICE P MANUALS BCGroup USP 50SP USP 50SP_UM_Rev03 doc SPECIFICATIONS ULTRASONIC POWER MEASUREMENT POWER RANGE 0 to 30 W RESOLUTION 50 mW MINIMUM DETECTABLE 50 mW POWER LEVEL ACCURACY 3 One Count Reading DIGITAL FILTER 2 5 seconds Integration DUT OPERATING DUT MAXIMUM TRANSDUCER SIZE 4 00 Inches 101 6 mm TEST MEDIUM Degassed Water 24 0 3 0 C 75 2 5 4 F ELECTRICAL BATTERY Quantity 4 AA IEC LR6 Alkaline BATTERY ELIMINATOR 12 VAC 500 mA BC20 40337 USA BC20 40341 EURO PHYSICAL amp ENVIRONMENTAL DISPLAY 6 digit Backlit LCD 0 5 Inch 12 7 mm thick TANK LINING CONSTRUCTION Neoprene CARRYING CASE Molded Polyethylene OVERALL Including case SIZE 17 00 x 13 00 x 6 00 Inches 431 8 x 330 2 x 152 4 OVERALL Including case wel lt 10 0 Lbs 4 5 kg 15 to 30 C OPERATING RANGE 59 to 86 F 40 to 60 C STORAGE RANGE 40 to 140 F NOTES 5 5 NOTES BC Group YOUR ONE STOP 0 6 BC GROUP
4. INTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com salesObcgroupintl com USP 50SP User Manual 08 12 Rev 03 Copyright 2012 Made in the USA
5. UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS NOTICE DISCLAIMER BC GROUP INTERNATIONAL INC RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME WITHOUT NOTICE IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DR SAINT CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com USP 50SP User Manual Copyright 2012 www bcgroupintl com Made in the USA 08 12 Rev 03 INTRODUCTION The Model USP 50SP Ultrasound Power Meter is designed to measure the ultrasonic power output of diagnostic or therapeutic transducers up to 30 Watts The principle of measurement is the radiant force method The USP 50SP uses a positioning clamp to hold the transducer in degassed water above a conical target The ultrasonic energy passes through the water to reflect off the target and is then absorbed by the rubber lining The radiant power is directly proportional to the total downward force weight on the target This force is then transferred through the target support assembly to the electro mechanical load cell inside the sc
6. ale The cell is in a computer controlled feedback loop and produces a digital readout Watts of power custom units or grams of force The choice of units Watts or grams is selected by front panel pushbuttons TEST SETUP Water as a Measurement Medium The measurements are to be performed in water because ultrasound propagation in water closely approximates that in tissues The ultrasonic attenuation in water can be taken as a lower limit on the attenuation that will be encountered in the body Large areas in the body can consist of low attenuating material such as urine and amniotic fluid The use of water prevents measurements in a more highly attenuating material such as liver equivalent gels from representing the highest possible intensities that might be encountered in the body Ultrasound propagation in water closely approximates human tissue and degassed water is the generally accepted test medium for ultrasound transducers see AIUM NEMA Standards Publication UL 1 1981 SAFETY STANDARD FOR DIAGNOSTIC ULTRASOUND EQUIPMENT Degassed Water Ultrasound Power measurement accuracy is affected if the water contains more than 5 10 PPM of air To degas boil distilled water for 30 minutes then seal the container tightly and place it in a refrigerator An alternate method of degassing water is to heat the water to the boiling point then pull a vacuum on it for five to ten minutes The degassed water storage container should be mad
7. e of glass or plastic Polystyrene containers should not be used since they allow oxygen to permeate and degrade the water quality Before testing pour water into tilted test tank with minimum amount of turbulence The test tank water surface will absorb oxygen therefore a change of degassed water is recommended before each test Water temperature affects accuracy so it is 2 recommended to use an ambient testing temperature of 24 0 3 0 C 75 2 5 4 F Sonic energy agitates the water surface through heating and scattering The time duration of each test should be limited to a few minutes Prolonged testing particularly at higher power levels will show visible signs of air bubbles on the transducer target and the absorbing rubber surface Transducer Wetting and Placement To avoid introducing air into the degassed water insert the transducer at a 45 angle then position so it is facing the target Verify that the transducer surface is uniformly wetted if not wipe the surface clean using your finger The transducer should be pointed toward and center positioned directly above the cone target Small measurement variations will occur due to placement of the transducer Try various positions above the target to minimize the magnitude of error due to positioning LAYOUT A Unit of BC Group International 1 On Zero Off Button Press to turn unit on or zero press and hold until OFF is displayed then release to turn off
8. iform wetting no air pocket or bubbles should be on its surface 7 Allow 5 minutes for the scale to stabilize With no ultrasonic power applied to the transducer press the ON Zero Off button to zero the unit 8 Check response by placing the 1 gram weight on the arm of the cone target the flat part that is out of the water The USP 50SP should read 1 00 0 10 grams 9 Remove the 1 gram weight 10 After the display s stability indicator is shown activate the DUT and record the measurement It is a good practice to take three readings and average them If measurement conditions are not stable use the grams mode and multiply the readings by 14 65 to obtain Watts 12 When finished unplug the USP 50SP empty the tank and place the dry target cone in the tank for protection MANUAL REVISIONS Revision Revisions Made Rev 01 Origination Rev 02 Misc Edits Format Updated Rev 03 Misc Edits Format Updated Pictures Updated LIMITED WARRANTY WARRANTY BC GROUP INTERNATIONAL INC WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE BC GROUP INTERNATIONAL INC IS NOT LIABLE FOR ANY INCIDENT
9. l personnel CAUTION ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance and accuracy of the USP 50SP If the unit is outside the Operating Specifications allow it to acclimate to specified conditions for at least 30 minutes before attempting to operate it CAUTION INSPECTION Inspect the USP 50SP before each use for wear It should be serviced if any parts are in question CAUTION LIQUIDS Do not submerge or spill liquids on the USP 50SP Do not operate the USP 50SP if internal components not intended for use with fluids may have been exposed to fluid as the internal leakage may have caused corrosion and be a potential hazard CAUTION CLEANING Do not immerse The USP 50SP should be cleaned by wiping gently with a damp lint free cloth A mild detergent can be used if desired NOTICE SYMBOLS Symbol Description OO Center Negative Per European Council Directive 2002 95 EC do not dispose of this product as unsorted municipal waste NOTICE ABBREVIATIONS Celsius degree Device Under Test Fahrenheit gram s International Electrotechnical Commission kilogram s Megahertz millimeter s milliwatt s Parts Per Million pounds United States of America Volt s Volt s Alternating Current Volt s Direct Current Watt s NOTICE DISCLAIMER BC GROUP INTERNATIONAL INC WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO