Home

MTC Series - CERT-TRAK

image

Contents

1. damage to the sensor under test caused by the application of excessive temperatures The feature also aids in reducing drift resulting from extended periods of exposures to high temperatures This feature can be locked with an access code ngn 7s l Specification Sheet AS ays oes us gt A Re calibration adjustments The MTC series has a very easy and straightforward procedure for re calibration adjustment There is no need for a screwdriver or PC software The only thing you need is a reliable reference thermometer Place the probe in the calibrator and follow the instructions on the display Third party labs and calibration facilities will be able to perform this function if a certificate from an independent source is necessary Of course AMETEK can provide you with a traceable calibration certificate from our labs when you require a higher level of confidence Simplified calibration documentation All MTC instruments are supplied with RS232 computer interface and the JOFRACAL calibration software This software allows the user to customize his or her calibration routines The software is easy to use so you do not have to be a programmer to configure your own calibration procedures After calibration you can print out certificates that contain all necessary information for your ISM or similar quality systems oe ee Sadia ol oe 2 kde ee fr ee ee ed re mj a at mE l Bem a m CEA T ae a m b m Th
2. temperature Stability indicator The bold checkmark V on the display indicates that the calibrator has reached the desired set temperature and is stable The operator may change the stability criteria and establish a greater sense of security in the calibration results A convenient countdown timer is activated five minutes before the unit reaches stability Slope rate Ramp Automatic switch test Operators can save a lot of time using the automatic thermoswitch test function to find values for the Open and Close temperatures Additionally this feature displays the hysteresis deadband between the two points The feature ensures a very high repeatability when testing thermo switches Simply press the SWITCH TEST key to activate the function Auto stepping This feature saves manpower The operator may stay in the control room or on the bridge monitoring the output from the sensor under test while the MTC series calibrator is placed in the engine room and automatically changes the temperature using a programmed step value and rate Up to 9 different temperature steps may be programmed including the hold time for each step This feature is also ideal for burning in new sensors prior to installation this minimizes initial drift and allows for initial testing Maximum temperature From the setup menu the user can select the maximum temperature limit for the calibrator This function prevents
3. C 650 Ay wstecsc since tacueisionscssaacd deen csesetsavesinsentec 14 1 Ib 6 4 kg Insert dimensions MTC 140 A Diameter x length 0 75 x 3 9 in 19 mm x 100 mm MTC 320 A MTC 650 A Diameter x length 0000 1 x 4 7 in 26 mm x 120 mm Weight of non drilled insert approximate MTC 140 A secre ieee cnceradatoneesecises counaseacscececedeanmcosence 2 6 oz 75 g WEG HS ZO A scannanu 5 8 oz 170 g IVT OD OP cess pan ani 17 8 oz 510 g Shipping with std accessories carrying case Weight MT C 140 A cseccccesseseeeeseeeeeeeeees 27 6 Ib 12 5 kg Weight MTC 320 Aicscssscsccccdeveceicesissecteesastelesseeiecs 24 Ib 11 kg Weight MTC 650 A ccccccseeeeeceeeeeeeseeeseeeeees 27 b 12 kg Size L x W XH 19 9 x 9 1 x 16 3 in 507 x 232 x 415 mm Shipping with std accessories but no carrying case Weight MTC 140 A ccccccccsseeeeeeesseeeenesseeeeees 22 b 10 kg Weight MTC 320 A ccccccsseeeeeceeseeeeneeeeeeeees 17 5 Ib 8 kg Weight MTC 650 A ccccsisrcetaccesdecetnecettrawetecsetedects 21 Ib 9 5 kg Size L x W x H 16 4 x 9 8 x 14 6 in 465 x 255 x 470 mm Shipping carrying case only WEINI sacccodsassensasnctensuplegandianonseasieenetuaatanaunttes 11 Ib 5 0 kg Size L x W XH 19 9 x 9 1 x 16 3 in 507 x 232 x 415 mm Miscellaneous Serial data interface cccceesseeeeeeees RS232 9 pin Male Operating temperature 0008
4. F MTC320A MTC 320 A 50 to 320 C 122 to 608 F MTC650A MTC 650 A 50 to 650 C 122 to 1202 F Power supply 8th thru 10th characters 115 115VAC 50 60Hz 230 230VAC 50 Hz Mains power cable type 11th characters A EUROPEAN 230V B USA CANADA 115V C UK 240V D SOUTH AFRICA 220V E ITALY 220V F AUSTRALIA 240V G DENMARK 230V H SWITZERLAND 220V l ISRAEL 230V Insert type and size 12th thru 14th characters 003 Metric pre drilled 3 mm 004 Metric pre drilled 4 mm 005 Metric pre drilled 5 mm 006 Metric pre drilled 6 mm 007 Metric pre drilled 7 mm 008 Metric pre drilled 8 mm 009 Metric pre drilled 9 mm 010 Metric pre drilled 10 mm 011 Metric pre drilled 11 mm 012 Metric pre drilled 12 mm 013 Metric pre drilled 13 mm 014 Metric pre drilled 14 mm Not available for MTC 140 015 Metric pre drilled 15 mm Not available for MTC 140 016 Metric pre drilled 16 mm Not available for MTC 140 018 Metric pre drilled 18 mm Not available for MTC 140 020 Metric pre drilled 20 mm Not available for MTC 140 125 Inch pre drilled 1 8 in 187 Inch pre drilled 3 16 in 250 Inch pre drilled 1 4 in 312 Inch pre drilled 5 16 in 375 Inch pre drilled 3 8 in 437 Inch pre drilled 7 16 in 500 Inch pre drilled 1 2 in 562 Inch pre drilled 9 16 in 625 Inch pre drilled 5 8 in 688 Inch pre drilled 11 16 in Not available for MTC 140 750 Inch pre drilled 3 4 in Not available for MTC 140 813 Inch pre d
5. Hg to 10 000 psi multiple choices of pres sure ranges pumps and accuracies fully temperature compensated for problem free and accurate field use JOFRA Signal calibration Process signal measurement and simulation for easy control loop cali bration and measurement tasks from handheld field instruments for multi or single signals to laboratory refe rence level bench top instruments a matter of confidence AMETEK is a leading global manufacturer of electrical and electromechanical products for niche markets AMETEK s annual sales exceed 1 billion NYSE AME since 1930 Operations are in US Europe and Asia with about 1 3 of sales to markets outside the US AMETEK Test amp Calibration Instruments AMETEK Denmark A S Distributor USA Florida Tel 1 727 536 7831 Denmark Tel 45 4816 8000 Tel 800 527 9999 ametek ametek dk calinfo us ametek com AMETEK GmbH AMETEK Singapore Pte Ltd Germany Tel 49 2159 9136 0 Singapore Tel 65 6 484 2388 info ametek de aspl ametek com sg Pub Code SS CP 2400 US Issue 0408 Information within this document is subject to change without notice 8 ISO 9001 Manufacturer Copyright 2004 by AMETEK Inc AMETEK is a registered trademark of AMETEK Inc
6. O to 40 C 32 to 104 F Storage temperature ccce 20 to 50 C 4 to 122 F aTe a E E E E E T O to 90 RH Protection class secesswadsgenes desta tslead cece doane chtaettenssecesteosesebans IP 10 CE Conformity EN61326 1997 A1 1998 corr 2000 E RE E EN61010 1 1993 A2 1995 Specification Sheet Saf a4o0 us Marine approved temperature sensors STANDARD DELIVERY Standard delivery MTC 140 320 650 A MTC dry block calibrator user specified Mains power cable user specified Traceable certificate temperature performance Insert user specified Tool for insertion tubes User manual Test cables 1 x red 1 x black 3 pcs insulation plugs for 1 4 3 8 1 2 in 6 10 13 mm sensors MTC 140 A only Aluminium carrying case RS232 cable JOFRACAL software Part no 105337 123408 122832 60F 174 122822 60F135 60F 139 60F 138 60F 137 60F 141 60F 140 60F 143 60F 142 60F 136 104203 104216 60F 170 123469 65 F 100 105173 105366 124915 ACCESSORIES Description MTC series user manual Carrying case Cleaning brush 4 mm 3 Pkg Cleaning brush 6 mm 3 Pkg Cleaning brush 8 mm 3 Pkg Mains cable 115V USA Type B Mains cable 220V Australia Type F Mains cable 220V Italy Type E Mains cable 220V South Africa Type D Mains cable 230V Denmark Type G Mains cable 230V Europe Type A Mains cable 230V Israel Type Mains cable 230V Swi
7. csttecantinintasmonteaus Maximum 5 VDC Test current soracsciriiiiicrrinnraniissiiiisennnn Maximum 2 5 mA JOFRACAL software Minimum hardware requirements for JOFRACAL calibration software e INTEL 486 processor PENTIUM 200 MHz recommended e 16 MB RAM 32 MB recommended e 40 MB free disk space on hard disk prior to installation e Standard VGA 800 x 600 16 colors compatible screen 1024 x 786 256 colors recommended e CD ROM drive for installation of the program e 1 free RS232 serial port Engine rooms testin exhaust gas KEY FEATURES Automatic switch test Finds switching temp ceceeees Open close hysteresis Slope rate programable c cccccsseeeeeeees 0 1 to 9 9 C F Auto stepping LOGI AMIMGED IC ia coccrhscctciesmuaiacnenestperceacndaieseacnedei Up to 9 steps Dwell time on each Step eeeeseeeeseseeeeees Programmable Enhanced stability Unstable mains protection ccseeceeseeeeeeees MVI Circuitry Stability indication ec cencrenteiacssoresersccsencteeeticcddees Yes in display Multi information display Stability TIC AO Ks icecaciresias comesecisencceecasuznsecnandane Bold checkmark Countdown timer before stable ccceeceeeeeees 4 minutes Temperature c ceseeeeeeeee SET and READ simultaneously Alphanumeric M SSAQECG cccceeeeeeecesseeeeeeseereneeeeeeeees Yes Calibration status ICONS scicdcvcesecteccuiesascesdunnceensndarc
8. e JOFRACAL software supports automatic calibration for all JF Instruments dry block calibrators equipped with a RS232 serial data interface For semi automatic calibrations the software also supports liquid baths ice points or other dry block heating and cooling sources Using the software s SCENARIO function allows for combining instruments in virtually any configuration The JOFRACAL software is able to store all your results in a certificate database a sensor database and an instrument database and use a database function history and search Mains specifications Voltage MTC 140 320 650 A 115V 90 127 230V 180 254 PICOUCIICY seca xectantantawintbavtgntavacessdanesombadtnaicecetmbontet 45 65 Hz Power consumption max MTC 140 A cceeeee 150 VA Power consumption max MTC 320 A 650 A 1150 VA Temperature range MTC 140 A UCT otters race E E E A E EET 140 C 284 F Minimum ambient temp 0 C 82 F 30 C 22 F Minimum ambient temp 23 C 73 F 00 17 C 1 F Minimum ambient temp 40 C 104 F 2 C 28 F MTC 320 Aiivcicecsiccinortescnieonsetdiesaudes 33 to 320 C 91 to 608 F MTC 650 Ay cic etannooaneecaiaweanccioaawenavi 33 to 650 C 91 to 1202 F Resolution user selectable S lectable sssrini 1 or 0 1 C F Stability PTAA A ccotecucescectivedieennatenccenseitedsesetaaet 0 05 C 0 09 F MTC 320 A 650 Au ccccceccceeeeeeeeee
9. eating elements and other devices are periodically cycled on and off The cycling of supply power can cause the temperature regulator to perform inconsistently leading to both inaccurate readings and un stable temperatures The MTC series calibrators MTC 320 A and MTC 650 A employ the MVI thus avoiding such stability problems The MVI circuitry continuously monitors the supply voltage and ensures a constant energy flow to the heating elements The MTC 140 A does not require the MVI circuitry because the Peltier elements are energized with a stabilised DC voltage Easy to use intuitive operation All instrument controls may be performed from the front panel The heat source is positioned away from the panel This design helps to protect the operator The main functions on the MTC series are designed with one key one function logic This means that there are no sub menus or difficult to remember multiple keystrokes necessary to access primary functions The easy to read backlit display features dedicated icons which help in identifying instrument conditions and operational steps Set temperature The Up and Down arrow keys allow the user to set the exact temperature desired with a resolution of 0 1 C or F Instrument setups The MTC series stores the complete instrument setup including engineering units stability criteria resolution display contrast slope ramp rate auto step settings and maximum
10. eeeeeeeeeees 0 1 C 0 18 F Measured after the stability indicator has been on for 10 minutes Measuring time is 30 minutes Time to stability approximate MICO A senas N 5 min MTC 320 A 650 A vccsiicimicannaiteswnitencnbiessienessitanesznaceraaedanceinne 8 min Accuracy MTO TAU A esince 0 4 C 0 7 F MTC 320 A aaisessesssrrernsrernsrrrrrnrrnrrnrrnrrerrerenne 0 5 C 0 9 F MTO 650 Accsin cicionedinteiteeoeaneannse 0 9 C 1 6 F Specification when using the internal reference Load 4 mm OD reference probe in the center of the insert Immersion depth MTC 140 A insulation included 4 5 in 115 mm MTC 320 A 650A vce csisnecndsncendtcntsaeonsicawerenk 4 3 in 110 mm is performed mo range could be re more economical co could be achieved er GY oKX lt Specification Sheet Sepp o s Heating time MTC 140 A 17 to 23 C 1 to 73 F 3 minutes MTC 140 A 23 to 140 C 73 to 284 F 15 minutes MTC 320 A 33 to 320 C 91 to 608 F 4 minutes MTC 650 A 33 to 650 C 91 to 1202 F 10 minutes Cooling time MTC 140 A 100 to 0 C 212 to 32 F 10 minutes MTC 140 A Oto 15 C 32 to 5 F ee 13 minutes MTC 140 A 140 to 100 C 284 to 212 F 2 minutes MTC 320 A 320 to 100 C 608 to 212 F 16 minutes MTC 650 A 650 to 100 C 1202 to 212 F 28 minutes Switch input dry contact Test voltage cgictinticiaviidernesa
11. etendedetaades Yes Training mode heating cooling block disabled Simulation of all functions cccsccee sends deesccsudscseecesesmsdcusescsecdsexe Yes Simulating heating and cooling Approx 100 per minute Service facilities Adjustment of the unit from the keypad c ssseeeeeee Yes Self explanatory guide in display ccccsseeeeeeseeeeeeeeees Yes Other information Displays serial number software revision level and last calibration date Setup facilities Stability criteria Extra time before stable indication is shown Display reSOlUtiOn ccccssseecceeeseeceeeeeeeeesees 0 1 or 1 C F Temperature units cscosscrsacreressieneieneladerennetecaaceeness C and F Slope rate i ociaesdennd icteedeasesecsesesicetasecete weeccsoences 0 1 to 9 9 minute Maximum temperature c 0eee Any value within range Type approval vertifcate All JF Instruments calibrators are type approved by DET NORSKE VERITAS Li if DT eee ST AS TFR APs AL CPR TAT im pi pra F aer E oe ae oe ay HERE Pe HT PD ee Pee E iri 5 Specification Sheet seayemus lt Ke PHYSICAL SPECIFICATIONS Instrument dimensions MTC 140 A MTC 320 A MTC 650 A LXWXH enoni 9 5 x 5 5 x 12 8 in 241 x 139 x 325 mm Instrument weight MICA OF tess sacececcececanaceedeescaseseeeesseaneaceeavetae 14 3 Ib 6 5 kg UH Soo 1 0 0 eee eee een ene oe ee ee ee eee ee 11 Ib 5 kg MT
12. in 60F456 3 8 in 60F458 7 16 in 60F460 1 2 in 60F462 9 16 in 60F464 5 8 in 60F466 11 16 in N A 3 4 in N A 13 16 in N A 7 8 in N A Set of above inserts 124680 Multi hole M02 123480 Note MTC 140 only All multi hole inserts are delivered with a matching insulation plug 320A part no 100176 100178 100180 100181 100184 100187 100189 60F344 100192 60F348 100193 60F352 60F354 124682 123477 650 A part no 123444 100196 123445 100198 122516 100201 100202 105188 100204 100206 105189 100208 100209 123446 122517 122518 124685 123476 650 A part no 100195 100197 100199 100200 100203 100205 100207 60F408 100210 60F412 100211 60F416 60F418 124686 123478 Specification Sheet AD Sta Inserts undrilled 140A 320A 650 A part no Inserts part no part no 5 pack undrilled insertion tubes 60F448 100175 100194 Undrilled inserts MTC 140 A Undrilled inserts MTC 320 A 3mm MTC 650 A 4mm 3 16 in 6mm 5mm Multi hole M01 MTC 140 A 1 8 in 1 4 in 9mm Multi hole M02 MTC 140 A 3mm 4mm 3 8 in 6 mm 5 mm Multi hole M01 MTC 320 A 1 8 in MTC 650 A 3 16 in 1 4 in Multi hole M02 MTC 320 A MTC 650 A g J 1 IF Q P Key V ORDERING INFORMATION Model MTC series dry block temperature calibrators Order number Description Base model number 1st thru 7th characters MTC140A MTC 140 A 17 to 140 C 1 to 284
13. in only 10 minutes The fast performance of the heating block is due to the special profile that minimizes mass and yet still accepts an insertion tube with a 1 in 26 mm outer diameter This design is a balanced compromise between temperature stability homogeneity and rapid heating and cooling MTC 140 A heating cooling block The model MTC 140 A features Peltier elements In 1834 Jean Peltier a French physicist found that an opposite thermocouple effect could be observed when an electric current was connected to a thermocouple Heat would be absorbed at one of the junctions and discharged at the other junction This effect is called the gt PELTIER EFFECT Specification Sheet SS C SETO US The practical Peltier element electronic heating pump consists of many elements of semiconductor material that are connected electrically in series and thermally in parallel These thermoelectric elements and their electrical interconnections are mounted between two ceramic plates The plates serve to mechanically hold the overall structure together and to electrically insulate the individual elements from one another MVI Improved temperature stability MVI stands for Mains power Variance Immunity Unstable mains power supplies are a major contributor to on site calibration inaccuracies Traditional temperature calibrators often become unstable in shipboard environments where large electrical motors h
14. n optimized A drilling guide is included if you buy undrilled inserts Use the cleaning brushes to clean the bores in your inserts when necessary compensation Insulation tube and plates part no 65F100 and 105173 Improve your calibration uncertainty by insulating the sensor under test Minimize the heat dissipation from the top of the block and through the sensor under test INSERTS FOR MTC SERIES General inserts description Inserts for MTC 140 A and MTC 320 A are made of aluminium Inserts for MTC 650 A are made of brass All specifications on hole sizes refer to the outer diameter of the sensor under test The correct clearance size is applied in all predrilled inserts Specially drilled inserts are available on request Inserts predrilled metric Probe 140A diameter part no 3mm 123428 4mm 60F451 5mm 123429 6mm 60F453 7mm 123430 8mm 105185 9mm 105186 10 mm 105187 11 mm 123431 12 mm 123432 13 mm 123433 14 mm N A 15 mm N A 16 mm N A 18 mm N A 20 mm N A Set of above inserts 124679 Multi hole M01 123479 Note MTC 140 only All multi hole inserts are delivered with a matching insulation plug 320 A part no 123436 100177 123437 100179 123438 100182 100183 100185 100188 100186 60F339 100190 100191 123439 123440 123441 124681 123475 Inserts predrilled imperial inch Probe 140 A diameter part no 1 8 in 60F450 3 16 in 60F452 1 4 in 60F454 5 16
15. rilled 13 16 in Not available for MTC 140 875 Inch pre drilled 7 8 in Not available for MTC 140 M01 Multi hole insert type 1 M02 Multi hole insert type 2 Options 15th thru 18th characters C Carrying case standard F Traceable certificate standard for Europe Asia Australia and Africa G NIST traceable certificate standard for Western Hemisphere H DANAK accredited certificate X Placeholder character for unused option MTC650A 230 A M01 CFXX Sample order number all 18 characters AMETEK www ametekcalibration com CALIBRATION INSTRUMENTS WWW jfinstruments com JOFRA MTC 650 A series dry block 230VAC power with European power cord pre drilled multi hole type 1 1 x 3mm 1x4 mm 1x 5mm 1x 6mm 1 x 9 mm including carrying case and traceable certificate 0ecause calibration is temperature software AMETEK Calibration Instruments offers a complete range of calibration equipment for pressure temperature and signal including software JOFRA Temperature standards Portable precision thermometer Dry block calibrators 4 series more than 20 models featuring speed portability accuracy and advanced documenting functions M amp G Primary pressure standards Pneumatic floating ball or hydraulic piston deadweight testers easy to use with accuracies up to 0 015 of reading JOFRA Pressure standards Convenient electronic systems ranging from 1 to 700 bar 25 in
16. se protection Charge air Turbine and gear bearings Uptake gas e Steam Feed water Servo oil pitch propeller e Cooling water outlet Purifiers Refrigeration systems Starting air e Gland steam Hydraulic systems Condensing system Oil burner e Main steam Generators Condensers Safety valves DNV approval All JF Instruments calibrators are marine type approved by Det Norske Veritas Complete marine program The MTC series is part of a complete program of marine approved temperature pressure and signal calibrators including temperature sensors See more at www jfinstruments com AMETE K ISO 9001 Manufacturer CALIBRATION INSTRUMENTS 5 INSTRUMENTS M KN Marine Temperature Calibrator F The MTC series features a large backlit display that is easy to read even in well lit areas Units feature an informative display that provides icons and information regarding the status of the MTC and the calibration in progress The MTC series also features an auto step function Using this function you may stay in the control room or on the bridge and monitor the temperature reading while the calibrator located in the engine room by the sensor automatically steps through a number of pre programmed temperatures Fast heating and cooling The MTC 320 A and the MTC 650 A contain an innovative heating block profile This design heats up the MTC 320 A to maximum temperature in just 4 minutes and the MTC 650 A
17. temperature Marine Temperature Calibrator Type approved temperature calibrators to optimize the performance of your vessel i Wide temperature range The MTC series of portable temperature calibrators from Maer gen ge anor JF Instruments facilitates correct readings on all of your a Meche ane eo CUO temperature monitoring devices The MTC dry block MTC 650 A 33 to 650 C 91 to 1202 F i me N am design does not use any hazardous hot liquid and heats Fast calibration is timesaving up and cools down much faster than traditional ways of The specially designed heating block calibrating temperature sensors on a vessel You can profile heats up to 320 C 608 F in just 4 reach 320 C 608 F in ju ites and do it safely minutes and to 650 C 1202 F in only 10 minutes High flexibility You are not limited by fixed holes Inter changeable insertion tubes are used to match the diameter of your sensor under test Enhanced stability MVI circuitry ensures stability despite mains supply variations in the engine room environment Timesaving features Fast one key one function access to the automatic switch test and auto stepping Documentation made easy Mnitoring devica ley gt MO gt IOT RS232 communication interface and Maintain ane SS ea ate ee piston fuel valve cooling media Fuel oil inlet e Scavenge air e Exhaust gas Sea water cooling Thrust shaft plane stern tube bearings Crankca
18. tzerland Type H Mains cable 240V UK Type C Test cable set Thermal protection shield Insert removal tool Insulation plug MTC 140 A only 3 pcs for 1 4 in 6 mm 3 8 in 10 mm 1 2 in 13 mm Insulation tube 4 in 100 mm 10 insulation plates RS232 cable JOFRACAL software JF instruments also has a series of temperature sensors which are type approved by classification societies for marine applications such as measuring exhaust gases cooling water for diesel engines and product or room temperatures in refrigeration stores These sensors are specially designed with high enduance for vibration according to IEC 68 2 6 Approvals Lloyds Register of Shipping Germanischer Lloyd Bureau Veritas and Det Norske Veritas For further information please visit our web page at www jfinstruments com Carrying case part no 123408 The protective carrying case ensures safe transportation and storage of the instrument and all associated equipment The carrying case is included in the standard delivery Thermal Protection Shield part no 104216 An external heat shield is available and may be placed on top of the calibrator to reduce the hot air stream around the sensor under test This is especially important for testing thermocouples having head mounted transmitters with cold junction Inserts heat shield and cleaning brushes Always use the original inserts where material and physical dimensions have bee

Download Pdf Manuals

image

Related Search

Related Contents

  ESPAÑOL  Owner`s Manual  367KB  取扱説明書  

Copyright © All rights reserved.