Home
CURING LIGHT LED.C USER`S MANUAL
Contents
1. composed mainly of high power LED optical fiber and main unit Picture 1 4 Technical specifications 4 1 Power supply Rechargeable Lithium battery Battery model ICR18490 Battery voltage and capacity 3 7V 1400mAh Input of adapter 100V to 240V 50Hz 60Hz Built in fuse of adapter 1A 250V 4 2 Applied part optical fiber 4 3 Light source Blue light Wave length 420nm 480nm Light intensity 1000mW cm 1200mW cm 4 4 Working condition Environment temperature 5 C to 40 C Relative humidity lt 80 Atmosphere pressure 70kPa to 106kPa 4 5 Dimensions 23mm 263mm 4 6 Net weight 124g 4 7 Consumption power lt 8W 4 8 Protection type against electrical shock class II 4 9 Protection against electrical shock type B 4 10 Protection against harmful ingress of water or particular matter ordinary equipment IPX0 4 11 Safety in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide not suitable under this condition 5 Instruction of recharging 5 1 Load the battery into the compartment of the pedestal charger Be sure not to opposite the electrode 5 2 Connect adapter to the AC100V 240V Machine standby when indicator lighting green Put the battery into the pedestal when indicator turns yellow starts recharging After recharging indicator turns green and shut off the electricity and take out the battery from the compartment 5 3 The battery has no memory and can be recharged any time 5 4 The m
2. for domestic purposes Voltage fluctuations flicker emissions Complies IEC 61000 3 3 Guidance amp Declaration electromagnetic immunity The models LED B amp LED C are intended for use in the electromagnetic environment specified below The customer or the user of the models LED B amp LED C should assure that It is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD 6 kV contact 6 kV contact Floors should be wood concrete or ceramic tile If floors are covered 8 kV air 8 kV air with synthetic material the relative IEC 61000 4 2 humidity should be at least 30 Electrical fast 2kV for power supply Mains power quality should be that lines 2kV for power of a typical commercial or hospital transient burst supply lines environment IEC 61000 4 4 1 kV for Input output pply lines Surge 1 kV line to line 2 kV line to ae ac quality moa be a i of a typical commercial or hospita IEC 61000 4 5 2 kV line to earth earth environment Voltage dips short lt 5 Ur Mains power quality should be that interruptions and lt 5 Ur gt 95 dip in of a typical commercial or hospital voltage variations on 295 dip in Ur Ur environment If the user of the power supply input for 0 5 cycle for 0 5 cycle models LED B amp LED C require lines 40 Ur 40 Ur continued operation durin
3. to 2 5 GHz 9 Be 2nk BO MHz to 200 MHZ d 2 3xP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter In watts W according to the transmitter manufacturer and d Is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur In the vicinity of equipment marked with the following symbol GY NOTE At 80 MHz end 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the models LED B amp LED C are used exceeds the applicable RF compliance level above the models LED B amp LED C should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the models LED B
4. GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner structure etc have claimed for several patents by WOODPECKER any copy or fake product must take legal responsibilities 19 Declaration of conformity 19 1 Product conforms to the following standards EN 61000 3 2 2006 EN ISO 14971 2009 EN 61000 3 3 2008 EN ISO 7405 2008 EN 60601 1 4 1996 EN ISO 17664 2004 EN 60825 1 2007 EN ISO 17665 1 2006 EN 980 2008 EN ISO 10993 1 2009 ISO 9687 1993 EN ISO 10993 5 2009 EN 1041 2008 EN ISO 10993 10 2010 19 2 EMC Declaration of conformity Guidance and manufacturer s declaration electromagnetic emissions The models LED B amp LED C are intended for use in the electromagnetic environment specified below The customer or the user of the models LED B amp LED C should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions The models LED B amp LED C use RF energy only for its internal CISPR 11 Group 1 function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment BE emissions Class B The models LED B amp LED C are suitable for use in all CISPR11 5 i i 7 a establishments including domestic establishments and those Harmonic emissions Class A directly connected to the public low voltage power supply network IEC 61000 3 2 4 i that supplies buildings used
5. WOODPECKER CURING LIGHT LED C USER S MANUAL Please read this manual before operating Industrial design patent No CN ene FDACE GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD www glwoodpecker com Certified Management System ENISO 9001 EN ISO 13485 A ous P 2 i gt ONDNBRWN eK Introduction Principle and usage Structure and components Technical specifications Instruction of recharging Operation Precaution Contraindication 9 10 After service 11 Packing List 12 Troubleshooting 13 Storage and transportation 14 Environmental protection 15 Manufacturer s right Content Maintenance 16 For technical data please contact 17 Symbol instruction 18 Statement 19 Declaration of conformity DPOnmnwmwmDmDvaAaandan amp WN RRR 1 Introduction Guilin Woodpecker Medical Instrument Co Ltd is a high tech enterprise in researching developing and producing dental equipment and has a perfect quality assurance system main products including ultrasonic scaler curing light micro motor apex locator and ultrasurgery etc 2 Principle and usage 2 1 LED C adopts the principle of ray radiation to solidify the light sensitive resin by shooting at it in a short time 2 2 This product is used to restore teeth and solidify material for whitening teeth 3 Structure and components LED C dental is
6. ain unit should be in a full charge when it s used for the first time the ordinary charge time for a full charge is 4 to 6 hours 5 5 After recharging shut off the electricity and take out the battery from the compartment Picture 1 Battery compartment Charging indicator Light hood ran Charger Optical fiber A 100V to 240V Adapter plu A pter plug Digital screen On Off button Time button ______ Full mode gt 9 Ramping mode Pulse mode x Adapter A Mode switch Battery compartment 6 Operation 6 1 Let the cathode of the battery face down to the battery compartment and tighten it to the main unit See Picture 1 6 2 Take off the red cap from the optical fiber and then insert the metal part into the front of LED C make sure to screw the fiber to the end 6 3 To install the light hood on as show in picture 1 6 4 Press the time button to choose the solidification time 4 working time modes are available 5 10 15 20 seconds 6 5 Lightly press the mode key Following three modes are available 6 5 1 Full power mode blue light shine in full power 6 5 2 Ramping mode The power of the blue light turns from weak to stronger and reaches the highest power in 5 seconds 6 5 3 Pulse mode blue light work in the mode of pulse 6 6 During the operation aim blue light at the position needing solidification Press the ON OFF switch the LED starts to work und
7. amp LED C Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3V m 12 Recommended separation distances between portable and mobile RF communications equipment and the models LED B amp LED C The models LED B amp LED C are intended for use in electromagnetic environment in which radiated RF disturbances is controlled The customer or the user of the models LED B amp LED C can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the models LED B amp LED C as recommended below according to the maximum output power of the communications equipment Rated maximum output Separation distance according to frequency of transmitter power of transmitter 150kHz to 80MHz Ati 800MHz to 2 5GHz w d 1 2xP d 1 2xP d 2 3xP 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 4 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W accordable to the transmitter manufacturer NOTE At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electro
8. cracked 3 Wipe off the resin 3 There is resin remained on 4 Charge the LED C the surface of optical fiber 4 Low battery Show Er 1 The battery s voltage is 1 Charge the LED C low 2 Contact the local distributor 2 The LED C is damaged or us Battery lasting The battery s capacity Change a new battery time becomes becomes smaller shorter Indicator light 1 Power supply cord is not 1 Check the power supply cord of pedestal connected well and connect it well doesn t shine in thewpioeessnt 2 Battery is fixed upside 2 Refix the battery pay attention recharging down to the ends of the battery If all the above solutions have been completed the machine still can not work normally Please contact our special repair shop or us 13 Storage and transportation 13 1 This equipment should be handled carefully kept away from shaking point installed or stored at shadowy dry cool and ventilated places 13 2 Don t store it together with articles that are combustible poisonous caustic and explosive 13 3 This equipment should be stored in the environment where the relative humidity is lt 80 the atmosphere pressure is 70kPa to 106kPa and the temperature is 10 C to 55 C 13 4 Excess impact or shake should be avoided during transportation 13 5 Don t mix it with dangerous articles during transportation 13 6 Keep it away from sun or snow or rain during transportation 14 Environmental protectio
9. er the selected mode Then it counts down to 0 second to end the solidification 6 7 After the operation please clean the fiber with calico in order not to affect the light intensity 6 8 The depth of solidification of composite is no less than 4mm per 10 seconds 6 9 The optical fiber can be spinned off by 360 and autoclaved under high temperature of 135 C and pressure of 0 22MPa 6 10 During operation if the indicator light of mode twinkles it means low volume recharge it at once 6 11 Please take out the battery and store it carefully when not operating regularly 7 Precaution 7 1 Please recharge the battery at least 4 hours before first time usage 7 2 Connect adapter to the AC100V 240V Machine standby when indicator lighting green Put the battery into the pedestal when indicator turns yellow starts recharging After recharging indicator turns green and shut off the electricity and take out the battery from the compartment 7 3 Avoid aiming at eyes directly 7 4 It is advised to use the original pedestal charger adapter and Lithium battery because other brand pedestal charger adapter and Lithium battery are likely to damage the circuit 7 5 Please recharge the battery in cool and ventilated room 7 6 It is forbidden to self taking apart the battery in order not to result in short circuit or leakage 7 7 It is forbidden to extrude shake or rock the battery The Lithium battery is forbidden to be in short circ
10. g power IEC 61000 4 11 60 dip in Uz 60 dip in Uz Mains interruptions it is for 5 cycles for 5 cycles recommended that the models 70 Ur 70 Ur LED B amp LED C be powered from 30 dip in U for 25 cycles 30 dip in U7 for 25 cycles an uninterruptible power supply or a battery lt 5 Ur lt 5 Ur gt 95 dip in U gt 95 dip in for 5 sec Uy for 5 sec Power frequency magnetic fields are ioe should be at levels characteristic of magnetic field 3 Alm 3 Alm a typical location in a typical commercial or hospital IEC 61000 4 8 environment NOTE Ur is the a c mains voltage prior to application of the test level 11 Guidance amp Declaration Electromagnetic immunity The models LED B amp LED C are intended for use in the electromagnetic environment specified below The customer or the user of the models LED B amp LED C should assure that it is used in such an environment IEC 60601 test Compliance Immunity test level level Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the models LED B amp LED C including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 8 Vrms 1 2 IEC 61000 4 6 150 kHz to 80 MHz 3V d L2kP Radiated RF BV m pa 1 2 IEC 61000 4 3 80 MHz
11. magnetic propagation is affected by absorption and reflection from structures objects and people The device has been tested and homologated in accordance with EN 60601 1 2 for EMC This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment 13 Scan and Login website for more information opyo Guilin Woodpecker Medical Instrument Co Ltd tal Information Industrial Park National High Tech Zone Guilin Guangxi 541004 P R China Tel Europe Sales Dept 86 773 5873196 86 773 2125222 North America South America amp Oceania Sales Dept 86 773 5873198 86 773 2125123 Asia amp Africa Sales Dept 86 773 5855350 86 773 2125896 Fax 86 773 5822450 E mail woodpecker mailgl cn sales glwoodpecker cn Website http www glwoodpecker com Wellkang Ltd www CE Marking eu 29 Harley St LONDON W1G 9QR UK ZMN WI 04 559 2 6 Edition
12. n There is not any harmful element in our product It can be disposed according to the local laws 15 Manufacturer s right We reserve the rights to change the design of the equipment the technique fittings the instruction manual and the content of the original packing list at any time without notice If there are some differences between blueprint and real equipment take the real equipment as the norm 16 For technical data please contact Wellkang Ltd www CE Marking eu 29 Harley St LONDON W1G 9QR UK 17 Symbol instruction WOODPECKER fi Trademark o Class equipment P AN Type B applied part amp e Recovery IPXO Ordinary equipment A Keep dry als Screw inside outside 5 Handle with care 55 C Date of manufacture joc Temperature limitation 80 Manufacturer RE Humidity limitation 106 Used indoor only Atmospheric pressure 70 for storage CE marked product R Appliance compliance WEFE directive FDA marked product HIDCEE Consult the accompanying documents Authorised Representative in the EUROPEAN COMMUNITY m O vs m P Consult the accompanying documents Got the quality management system certification and Certified Management ENIS08001 Bones CE certification issued by T V Rheinland 18 Statement All rights of modifying the product are reserved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to
13. uit situation and it is forbidden to put the battery with metal or other conductors WARNING If the curing light works for 40s continously the temperature of the top of optical fiber may reach 56 C WARNING Do not modify this equipment without authorization of the manufacturer 8 Contraindication The heart disease patients pregnant women and children should be cautious to use this equipment 9 Maintenance 9 1 Only the optical fiber can be autoclaved under high temperature and pressure 9 2 After operation each time please shut off the power source and clean the optical fiber 10 After service From the date this equipment has been sold base on the warranty card we will repair this equipment free of charge if it has quality problems please refer to the warranty card for the warranty period for units and parts 11 Packing List The components of the equipment are listed in the packing list 12 Troubleshooting Solution Non indication 1 Battery has run down 1 Charge the LED C Non act 2 Battery is fixed upside 2 Refix the battery pay attention down to the ends of the battery 3 The LED C works 3 Stop the operation for several continuously too long and the minutes heat protection system works Solution Light intensity 1 The optical fiber is not 1 Insert the optical fiber again insufficient inserted till the botton correctly 2 The optical fiber has 2 Change the optical fiber
Download Pdf Manuals
Related Search
Related Contents
SMARTS code, version 2.9.5 USER`S MANUAL Curtis RCD672 User's Manual 取扱説明書 アクセサリー SATELLITE PRO L770 Manual de instrucciones Serie BL-VP100 - psn 大型外部表示器 取扱説明書 Samsung SGH-C160 Manuel de l'utilisateur for Reference and Installation Manual - Security Help Desk TalkSwitch User Guide 10th ED v2 - FortiVoice 取扱説明書/寸法図 - セイホープロダクツ Copyright © All rights reserved.