1. MAX W eas S lt at D 1 Continuous measurement indicator 2 Unit of measurement C F 3 Low battery indicator 4 Low alarm indicator 5 Temperature value MAX MIN DIF difference between MAX and MIN values AVG average HAL high alarm LAL low alarm TK K thermocouple probe temperature and LOG recorded value 6 High alarm indicator 7 EMS MAX MIN DIF AVG HAL LAL TK and LOG indicator 8 Temperature value 9 Current measurement indicator 10 HOLD indicator fixed measurement 11 Emissivity value 12 Laser in operation indicator Mi Settings before measurement Before measuring temperature it is recommended to make some settings Set technical unit C or F Set the continuous measurement On or Off Set the alarm On or Off To set these 3 parameters open the battery door by pushing on both sides of the trigger It is not necessary to disconnect the battery to make these settings Unit e Set technical unit C or F Set the selector unit to C or F with a screwdriver e Set the continuous measurement This setting allows to let the Kiray200 instrument in measurement It does not shut off after 7 seconds Set the selector on On continuous measurement is active or on Off continuous measurement is inactive with a screwdriver __Lock Lock e Set the alarm This setting allows to activate or deactivate high and low a
2. Do not leave the device with the lock on lock at the top right of the screen because in this configuration the instrument does not turn off automatically To prevent damage on your instrument or equipment please carefully respect these conditions To protect from electric and magnetic fields of static electricit Protect from thermal shock WM CE certification This device meets with following standards requirements e EN 50081 1 1992 Electromagnetic compatibility Part 1 e EN 50082 1 1992 Electromagnetic compatibility Part 2 Mi Maintenance To install or change the 9V battery open the part near the trigger and put it in the battery compartment Mi Accessories e Case holster with passer by belt e User manual e K thermocouple probe Infrared sensor Emitted energy by the Laser sighting object in the form of radiation X Once returned to KIMO required waste collection will be assured in the respect of the environment in accordance to 2002 96 CE guidelines relating to WEEE EXPORT DEPARTMENT Boulevard de Beaubourg Emerainville BP 48 77312 MARNE LA VALLEE CEDEX 2 Tel 33 1 60 06 69 25 Fax 33 1 60 06 69 29 Ref NTang KIRAY200 07 09 B RCS 24 P rigueux 349 282 095 Non contractual document We reserve the right to modify the characteristics of our products without prior notice
3. User Manual Supplied with thermocouple K probe E Distance from the target Distance 150 300 900 mm Diameter 5 10 30 mm pie Make sure that the target is larger than the size of the laser sighting Pressure Temperature Humidity Air Velocity Airflow Sound level KIRAY 200 Infrared thermometer Infrared thermometer KIRAY 200 is an infrared thermometer used to diagnose inspect and check any temperature Thanks to its elaborated optical system it allows an easy and accurate measurement of little distant targets KIRAY 200 instrument has an internal memory which can save up to 20 measurements WM Technical features Instrument features Spectral response 006 8 14 um Optical cceeeseeeees D S 30 1 50 mm at 1500 mm Response time ccccceees Less than 1 second Temperature rang 0006 From 50 to 850 C Accuracy From 50 to 20 C 5 C From 20 to 200 C 1 5 of reading 2 C From 200 to 538 C 2 of reading 2 C From 538 to 850 C 3 5 of reading 5 C Display resolution 000 0 1 C Emissivity aaeeei Adjustable from 0 10 to 1 00 pre set at 0 95 Over range indication Display indication OL for a negative over range OL for a positive over range Laser Sighting cccccees Wavelength 630 670 nm Output lt 1mW Clas
4. d temperature by the K thermocouple probe LOG when KIRAY 200 instrument is turned on press MODE button until LOG appears at the bottom left of the screen Next to LOG a number between 1 and 20 also appears it corresponds to LOG location If no temperature has been recorded in the shown LOG location 4 dashes will appear in the lower right corner To record a temperature you have to be on LOG mode then choose an empty LOG location visible and press SET button during the measurement or when the measurement is fixed HOLD From this mode you can also clear all the recorded temperatures press and keep the trigger pressed and press down button at the same time until reaching the zero recording then press SET button while keeping ENT pressed A beep is emitted by KIRAY200 instrument and the LOG location moves automatically to 1 signifying that all data have been cleared W Emissivity Emissivity is a term used to describe the energy emitting characteristics of materials Most 90 of typical applications organic materials and painted or oxidized surfaces have an emissivity of 0 95 pre set in the unit Inaccurate readings will result from measuring shiny or polished metal surfaces To compensate cover the surface to be measured with masking tape or flat black paint Allow time for the tape to reach the same temperature as the material underneath it Measure the temperature of the tape or painted surface See table below
5. ement period keep ENT pressed the KIRAY200 instrument displays the temperature of the area sighted by the laser Press MODE button until MAX or MIN is displayed at the bottom of the screen These values relate to the temperatures taken by the instrument and the thermocouple probe DIF during a measurement period press MODE button until DIF appears at the bottom left of the screen The displayed value corresponds to the difference between MAX value and MIN value AVG during a measurement period press MODE button until AVG appears at the bottom left of the screen The displayed value corresponds to the average temperature calculated during a measurement period HAL when KIRAY 200 instrument is turned on press MODE button until HAL appears at the bottom left of the screen The displayed value corresponds to the alarm of high temperature Set this alarm by incrementing it with up button or by decrementing it with down button LAL when KIRAY 200 instrument is turned on press MODE button until LAL appears at the bottom left of the screen The displayed value corresponds to the alarm of low temperature Set this alarm by incrementing it with up button or by decrementing it with down button Alarms must be activated see paragraph Settings before measurement A TK when KIRAY 200 instrument is turned on press MODE button until TK appears at the bottom left of the screen The displayed value corresponds to the measure
6. for values of emissivity of specific materials Aluminium 0 30 Ice 0 98 Asbestos 0 95 lron 0 70 Asphalt 0 95 Lead 0 50 Basalt 0 70 Limestone 0 98 Brass 0 50 Oil 0 94 Brick 0 90 Paint 0 93 Carbon 0 85 Paper 0 95 Ceramic 0 95 Plastic 0 95 Concrete 0 95 Rubber 0 95 Copper 0 95 Sand 0 90 Dirt 0 94 Skin 0 98 Frozen food 0 90 Snow 0 90 Hot food 0 93 Steel 0 80 Glass 0 85 Textile 0 94 Water 0 93 Wood 0 94 E mportant information For correct measurements Do not take any measurement on metal or shiny or reflective surfaces Do not measure through transparent surfaces such as glass for example e Water vapor dust smoke etc may prevent correct measurements because they obstruct the optic of the instrument e Make sure that the target is larger than the size of the laser sighting Infrared thermometer how it works Infrared thermometers can measure the surface temperature of an object Its optic lens catches the energy emitted and reflected by the a eA object This energy is collected and focused onto a detector This ae panda information is displayed as temperature The laser pointer is only used to A aim at the target To avoid any inconvenience Do not aim directly or indirectly reflection on reflective surfaces the laser in the eyes e Change the batteries when the indicator blinks Do not use the thermometer around explosive gas vapor or dust
7. larm Set the selector on On alarms are active or on Off alarms are inactive with a screwdriver Alarm Set Alarm OFF EE on Mi Operating mode Press ENT trigger to turn on the instrument The backlighted screen indicating the temperature and the laser turn on Keep ENT pressed Place the laser sighting at the center of the area to be measured Release ENT Read the displayed temperature The display stays on for 7 seconds after the last manipulation HOLD appears at the top left of the screen measurement stays displayed The KIRAY 200 instrument keeps in memory the last function used E Command buttons _ ENT Trigger Turning on the device ENT pressed activation of the laser sighting and temperature measurement ENT released display is on HOLD HOLD fixed and give the last measurement Display stays on for 7 seconds If no buttons are activated and continuous measurement is inactive the instrument turns off after 7 seconds OFF ON Mode button Allows to define the required measurement Max Min AVG DIF etc pressing as many times on this button EMS when KIRAY200 instrument is turned on press MODE button until EMS appears at the bottom left of the screen Set emissivity by pressing on UP button to increment it or DOWN button to decrement it By default the emissivity is set to 0 95 MIN ou MAX select the Min or Max temperature During a measur
8. rigger ENT Battery compartment Set technical unit C F Set continuous measurement On O Set alarm On Off lt Probe input E KIRAY 200 buttons emee KIMO 1 Up button It allows to increment emissivity and the thresholds of high and low alarm and to move to the next recorded value 2 Set button It allows to activate or deactivate laser and backlight of the screen It also allows to record a temperature 3 Mode button It allows to navigate through the modes emissivity max value min value difference average high alarm low alarm TK value and recorded values 4 Down button It allows to decrement emissivity and the thresholds of high and low alarm and to move to the previous recorded value E Modes flow chart SCAN exh Ag Jac G ba om ae MODE mode button HAL 030 SET set button ENT trigger Set emissivity EMS with up and down buttons Display of maximum 35 2 MAX temperature Display of the difference of IF 17 9 temperature between max and min temperatures Display of minimum 17 3 IN temperature Vv MODE MODE A 20 3 Display of average temperature AVG calculated over a measurement MODE period v Display and set of the alarm of high temperature Display and set of the alarm of low temperature Display of the temperature given by the K thermocouple probe Display and recording of temperatures with SET button
9. s 2 Il Positive or negative temperature indication Automatic no indication for a positive temperature sign for a negative temperature Display cccccccccsceecseeeees 4 digits with LCD backlighted display Auto extinction c Automatic after 7 seconds of inactivity High low alarm ccceeee Flashing signal on display and beep signal with adjustable thresholds Power supply Alkaline 9V battery Autonomy 38 h inactive laser and backlight 15 h active laser and backlight Use temperature 008 From 0 to 50 C Storage temperature From 20 C to 60 C Relative humidity 0 From 10 to 90 RH in operating mode and gt 80 RH in storage DIMENSIONS ccecceeeeeereeees 175 x 110 x 45 mm Weight 230 g included battery Memory errereen 20 temperature values with unit of measurement C or F Accuracy for an ambient temperature from 18 to 28 C with a relative humidity lower than 80 RH e K thermocouple probe features Temperature rang 008 From 40 to 400 C Display range ccceeeeees From 50 to 1370 C Resolution 0 1 C Accuracy 1 5 of reading 3 C Cable lengfth cccceeeeee 1m E Description LCD backlightec display Up button Backlight and Down button laser button Mode button Laser sighting output Infrared sensor KI Oy e T