User Manual - Temperature & Process Instruments, Inc.
Contents
1. E INSTRUMENTS MP 200 Manometer Table of contents I Technical specifications Technical features Specifications ll Introduction Description Connections Ill Browsing IV Menus Probe menu Functions Pressure Airflow Air velocity Area Duct type Sizes K2 Factor Units Hold Min Max Calculation Point point average Automatic average Automatic point point average Configuration Recording Thermocouple type Display Units Integration Compensation Pressure system Solenoid valve Parameters Language Date Time Beep Extinction RF logging Contrast Backlit CO 0 009 0009 aa d d 2 2 d ooo o 0 0 06 u bh a nh Q Q N N N N y L Technical specifications Technical features Pressure module Housing ABS shock proof de IP54 Bias en oso Keypad Metal coated 5 keypads 1 joystick verpressure allowed a 250 mbar e Overpressure allowed 2 500 Pa 500 mbar Conformity Electromagnetic compatibility Overpressure allowed 10 000 Pa 1 200 mbar as per NF EN 61326 1 Overpressure allowed 500 mBar 2 bar Power supply 4 alcalines batteries 1 5V LR6 Overpressure allowed 2 000 mBar 6 bar Ambient Neutral gas Operating temp from 0 to 50 C MP200 Connection See p 6 Storage temp from 20 to 80 C Display Auto shut off adjustable from 0 to 120 min o Weight 3409 Dim 50 x 54 mm blue blacklit Display of
2. For others measurement modules self calibration is performed by disconnecting the two pressure inlets of the sensor then by pressing Autoz key AIRFLOW Access Airflow function by means of Airflow key With Airflow function you can access to following sub functions Hold see Air velocity Area Config Configuration see Air velocity Params Parameters see Air velocity Avg Average see Air velocity Rec Recording see Air velocity AIRFLOW displa Area o oJ o e Duct type To select vent Type press OK or P Rec Probe y Select Lx W or Diam or K factor with arrow buttons A and V Confirm with OK If K factor is CAlarms e Paramek VelocifPressfArea Avg selected you must enter value You can choose a K factor already registered by selecting with A and Y Confirm with OK This factor can be modified by selecting with A and V then confirm with OK or Select Modify with OK or gt Enter factor by means of arrow keys A and Y Confirm with OK or gt e Sizes Press or OK to enter into sizes sub function You can choose an air vent already registered by selecting it with arrow keys A and V Confirm with OK or This air vent can be modified by selecting it with arrows keys A and Y then Confirm with OK or Select Modify with OK or K Enter sizes by means of arrow keys A and Y Confirm with OK or P IV Menus e K2 factor Press
3. 6 measurements including 4 simultaneously Specificatio MI Measuring units Resolutions Measuring range Accuracy 100 Pa 0 2 of reading 0 8Pa 2 From 0 to 500 Pa beyond 0 2 of oe ae Ji lio w pS From 0 to 2 500 Pa 0 2 of reading 2Pa 1Pa e ci U s From 0 to 10 000 Pa 0 2 of reading 10Pa 1Pa nn From 0 to 500 mbar 0 2 of reading 0 5mbar 0 1mbar From 0 to 2 000 mbar 0 2 of reading 2mbar mbar PITOT TUBE Air velocity m s fpm Km h mph From 2 to 5 m s 0 3 m s 0 1 m s i From 5 1 to 100 m s 0 5 of reading 0 2m s 0 1 m s X Airflow m h cfm l s m s From 0 to 99 999m h 0 2 of reading 1 PE 1 m h DEBIMO BLADES s Air velocity m s fom Km h mph From 4 to 20 m s 0 3 m s 0 1 mis AO From 21 to 100 m s 1 of reading 0 1m s 0 1 m s Airflow m3 h cfm l s m3 s From 0 to 99 999m h 0 2 of reading 1 PE 1 mh CURRENT VOLTAGE V mA From 0 to 2 5 V 2mV 0 001 V From 0 to 10 V 10mV 0 01 V From 0 to 4 20 mA 0 01mA 0 01 mA THERMOCOUPLE See related datasheet Qu C F K From 200 to 1 300 C 1 1 C or 0 4 Reading value 0 1 C J From 100 to 750 C 0 8 C or 0 4 Reading value 0 1 C gt T From 200 to 400 C 0 5 C or 0 4 Reading value 0 1 C TT ooo E All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measure
4. save your dataset click on Save with the access key 1 2 Automatic dataset An automatic dataset is composed of measuring points with interval of time a Click on OK or K to enter sub function b Select Auto with A and W Confirm wih OK c Select Name with A and V Confirm wih OK or K Enter dataset name with the arrow keys K and A Y Confirm wih OK d Enter dataset time and interval of time between 2 measurements by selecting Period with access key Select Duration or Interval with A and V Confirm wih OK Enter minutes and seconds with arrow keys amp and Y from 1 minutes to 24 hours for the duration and from 5 seconds to 10 minutes for the interval Confirm with OK e Select Start for dataset launching Automatic dataset 2 Launch a planned dataset A planned dataset is composed of several locations For each location the operator can enter a theorical value and a tolerance for the parameter to be controlled Planification must be made via the software a Click on OK or to enter into sub function b Select Planned with A and Y Confirm wih OK c Choose dataset name with A and V Confirm wih OK d Select the location with A and Y Confirm wih OK IV Menus 3 Preview of tables of points of datasets Dataset summary You can display tables of points of datasets performed on the device a Go to Recording menu b Select Display Click on OK to validate c Select dataset name with ar
5. K or OK to enter into the K2 factor sub function Select respectively ON or OFF with A and Y in order to enable or disable this function Confirm with OK e Units To select the unit press OK or P Select mm or in with arrow buttons A and Y Confirm with OK Air velocity Access Air velocity function by means of key With Air velocity function you can access to following sub functions Hold Config Configuration Params Parameters Avg Average Rec Recording AIR VELOCITY Display Hold Min Meaxc_ IO Press 1x in order to select HOLD function measurement holding on display _Rec Probe Hold ConfigeAutoZ Press 2x in order to select Min Max function display of minimum and maximum values Alarmse ParamseAirflo 4 Press Avg Press 3x back to the continuous measurement W Average Press or OK to enter Average sub function With A and Y you can select point point average auto point point automatic Confirm with OK or P e Point point average This function allows to calculate the average value of various points that you can select Numbers of selected points and parameter for which calculation is carried out are displayed For adding a new measuring point to this calculation press OK to confirm If you click on average icon max and min values standard deviation average of each channel and e numbers of measuring points will
6. be displayed If you want to see all values select Visu and scroll with A and Y e Automatic average This function allows to calculate an average value that the device measured in an interval chosen time Timer is displayed Select Start with access key for launching measurement If you click on average icon max and min values standard deviation average of each channel and time chosen will be displayed IV Menus D e Automatic point point average This function allows to calculate the average value of various points calculated themselves on a duration beforehand defined You must enter duration click on the Period icon Select minutes or seconds with arrow buttons A and Y Scroll digits withA and Y Confirm with OK The numbers of points is displayed Press Ok for launching measurement If you click on average icon max and min values standard deviation average of each channel and numbers of measuring points will be displayed You can view each measuring points if you click on Visu on fi g u r a t iora gg gg gg lt lt A If you use thermocouple probes you must enter type into the Configuration sub function Configuration sub function allows to e Select thermocouple type Click on OK or to enter into sub function a list of thermocouple available K J or T type appears Select type with A and V Confirm with OK Select display Use the arrow key or click on OK to ent
7. en connected to the instrument CO temp probe is connected on min DIN connectors C1 and or C2 mini Din C2 mini Din C1 connector connector Secured Mini Din Connector Retractable cable lg 450 mm up to 2 4 m lll Browsing amp r Power up 7 __ Enter code 0000 Homepage display Infos Probe Params Probe connection Probe display Return to previous screen Infos Measure Params Measurement PRESSURE Displa AIRFLOW displa Enter key code with directional pad if the locking is activated lt gt ad AV Select a sub menu with access keys or with arrow keys lt gt Select a connection with right and left keys gt Connections can be activated or deactivated with A or V Select a sub function with access keys or with arrow keys lt gt AIR VELOCITY displa Select a sub function with access keys or with arrow keys lt gt on L A Hold lt c Coniac Au Alarms ParamstAirflow Velocisr Avg Alarms Params Veloci Press Are toZ ne Hold wm Alarms Params Airflow Presse Avg Communication interrupted Check probes connection IV Menus Wire probes and modules with Smart plus system are automatically recognized from fir
8. er into sub function Select channel using A and W then confirm with OK Use A or Y to select ON displayed or OFF not displayed then confirm with OK e Select units Click on OK or to enter into sub function a list of units available appears Select unit required with A and V Confirm with OK Click on Esc to return to previous screen e Select integration The coefficient of integration allows to smooth the measure to avoid variations Click on OK or to enter into sub function a list of coefficient From 0 to 9 appears Select coefficient required with A and V Confirm with OK Coefficient 0 no integration important fluctuation in the shown measure e Select compensation It is possible to modify the value of the compensation in temperature Indeed the velocity and the airflow with Pitot s tube and with Debimo blades are calculated from a temperature of use in 20 C It is thus necessary to enter the real temperature of use to obtain more precise results Click on OK or to enter into the sub function Select or signs with A and Y with A and Y then pass on the first digit with Enter the first digit then move to the next one with K Confirm with OK e Select pressure system only available for Air velocity and Airflow functions Click on OK or to enter into sub function a list of pressure systems available appears Pitot tube L S Debimo or Other Select your system with A and Y Confirm with OK If Other is selec
9. ments carried out in the same conditions or carried out with required compensation The accuracy is expressed either by a deviation in C or by a percentage of the value concerned Only the bigger value is considered ll Introduction Top view Module connection O 0 mini DinC2 mini Din C1 connector connector Power supply connection USB port for PC cable Elastomer grip Keyp ad Access keys for circular menu Browsing joystick e 4 directions e Validation Escape key On Off key ll Introduction D Interchangeable measurement involdiiglable modules with Smart plus system are automatically recognized when connected to the instrument 1 Current Voltage module 3 Air velocity with Pitot tube Pressure module Pitot tupe optionnal It allows current or voltage measurements on V A1 or VA 2 channels with current voltage input cables or ammeter clamps Static Static pressure pressure PS PT 2 Pressure module l Dynamic pressure PT PS It allows differential pressure air velocity or airflow measurements with Pitot tube or Debimo on two pressure inputs and and thermocouple temperature measurement on Tc channel with wire thermocouple probes equipped with a miniature male connector SD Wire probes with Smart plus system Wire probes with Smart plus system are automatically recognized wh
10. nded to change the batteries 1 Remove the front part at the back of the instrument 2 Remove batteries 3 Insert new batteries AA LR6 1 5V in accordance with proprer polarity drew inside the housing 4 Replace the front V General informations Maintenance E Instruments performs calibration adjustment and maintenance of all your instruments to guarantee a constant level of quality of your measurements In regards of Quality insurance norms we recommend that the instruments are checked once a year E Instruments have 1 year guarantee for any manufacturing defect return to our After Sales Service required for appraisal E Instruments International LLC Distributed by 172 Middletown Blvd Suite B201 Temperature amp Process Instruments Inc E Langhorne PA 19047 1767 Central Park Ave INSTRUMENTS o H Fax 215 750 1399 Phone 914 673 0333 E mail Info E Inst com Web Site www tnp instruments com Web www E Inst com
11. row keys A et V Click on OK to validate Summary screen of selected dataset is displayed From this screen you can e Select other dataset using arrow keys and P e Display data of other channels using arrow keys A and V d Click on Mesure to display values table of selected dataset Dataset table From this screen you can ar ET e Browse values table of points of the same channel pressing Prev or Next e Change of channel with arrow keys and gt e Back to dataset summary screen pressing Visu 4 Delete all datasets Select Delete with A and V Confirm wih OK Parameters e Language Click on OK or gt to enter and a list of languages available appears Select language with arrow keys A and Y and Confirm wih OK e Date time Click on OK or gt to enter into sub function Enter the day with A and Y then move to the next digit with Repeat this operation for the month year hour and minute Confirm wih OK e Beep This sub function allows to enable or disable the keypad beep Click on OK or to enter into the sub function Select respectively ON or OFF with A and Y inorder to enable or disable the beep Confirm wih OK e Extinction This sub function allows to enable the automatic shut off and to select the delay in minute Click on OK or K to enter into the sub function Select with A and Y OFF in order to disable the automatic shut off or enter the delay from 15 to 120 minu
12. st connection The Probe menu only appears when probes or module are connected This menu allows to view probe information plugged to C2 Module C1 or wireless connections See Connections p 6 for more information about connections Probe display Available information are e Sensor type Serial number Date of last calibration or adjustement Probes Status enabled ou disabled On enabled mode the probe is connected the measurement is carried out and the value is displayed On disabled mode the probe is connected the measurement is not carried out and the value is not displayed nto a Measura Parems Pressure PRESSURE display oT Access Pressure function by means of key With Pressure function you can access to following sub functions Hold see Air velocity Config Configuration see Air velocity Params Parameters see Air velocity Avg Average see Air velocity Rec Recording see Air velocity Rec 4 Probes Hold S Alarms ParamsAirflo VelocisrAvg W AutoZ This sub function allows to compensate for any long term drifts of the sensing element by a manual adjustment of the zero For the 500 Pa measurement module self calibration is performed by the solenoid valve Once pressing Autoz key the zero is readjusted This function can also be automatically performed by using the solenoid valve function
13. ted you must enter a value Click on OK or to enter into sub function With A and Y enter the first digit then move to the next one with gt Confirm with OK e Solenoid valve available with the 500 Pa module Click on OK or to enter into the sub function Select respectively ON or OFF with A and Y in order to enable or disable the solenoid valve function Confirm wih OK or K When the solenoid valve is enabled it runs every minute IV Menus Recordi The Recording menu allows a measurement dataset You can choose between a planned or a continuous dataset Memory capacity of the instrument is up to 8 000 points or 50 datasets 1 Create or launch a continuous dataset Select dataset A continuous dataset can be carried out using MP200 and is composed of several dated measuring points The operator can choose an automatic or a manual dataset with an instant value or an average This datasets can t be set using Datalogger 10 Software 1 1 Manual dataset Select name A manual dataset is composed of measuring points selected by the operator a Click on OK or gt to enter into sub function b Select Manual with A and Y Confirm wih OK L c Select Name with A and V Confirm wih OK or Enter dataset name with arrow keys K and AB LIS A A Y Confirm wih OK vy 23 d For measurement launching click on OK with the access key The number of points selected and the 456789 Zt parameter are displayed e To
14. tes Confirm wih OK IV Menus e RF logging This sub function allows to enable or disable the RF Logging Click on OK or to enter into the sub function Select respectively ON or OFF with A and Y in order to enable or disable this function Confirm wih OK e Contrast This sub function allows to modify the contast Click on OK or P to enter Select your contrast level from 0 to 9 or AUTO with A and V Confirm wih OK e Backlit This sub function allows to modify the backlit Click on OK or to enter Select your backlit level from 0 to 9 or AUTO with A and V Confirm wih OK If you select AUTO the MP200 adjuts automatically the backlit according to the room brightness e Key locking This sub function allows to enable or disable the key lock Click on OK or to enter into sub function Select respectively ON or OFF with A and Y in order to enable or disable this function Confirm wih OK If the locking is enabled the code menu appears e Code This sub function allows to enter the security code Click on OK or and the code appears Enter the first digit of the code with A and Y then move to the next one with K Confirm wih OK Downloadinc see DataLogger 10 user manual chapter Ill Read device page 6 V General informations This menu allows to view the serial number of instrument and firmware version When battery indicator flashes it is recomme