Home

Invacare® Medley Ergo

image

Contents

1. Hand control HB73 PTT SCO case e PM Hand control with tilt HB74 WNIT OO cca yee eee sm enbe oe tee eae eee ee ee eee Hand control HL72 with integrated ACP function DUE EE EY PE YY YP Pi Hand control HL73 with integrated ACP function 21 e NEY ERE EU Ru Hand control with tilt HL74 with integrated ACP function 1502456 70 6 sci aciem pecie dae QU aine E PRAE edi External power supply transformer with wire gt eeeeeom aa a a am mm mm Rastofix bracket for leg section s At D e EE e PE E E EEE EE E a Lifting pole bonos CC Safety Floor Mattress 6 Cleaning and Maintenance 6 1 Cleaning The bed does not tolerate cleaning in automatic washing plants or the use of water jet based cleaning equipment Important Make sure the power plug is removed from the socket outlet before cleaning All plugs must be installed during washing of the bed The bed should be washed down using a wet sponge cloth or brush Use ordinary household cleaning agents Never use acids alkalines or solvents such as acetone or cellulose thinner The bed does not tolerate cleaning in automatic washing plants or the use of water jet based cleaning equipment The hand control motors and control unit may be washed with brush and water but not with pressurized water Dry the bed after cleaning EUR C9 N 22 6 2 Maintenance chart
2. M Ergo 33 73 40 80 160 201 166 207 15 55 22 62 Ergo Low p 102 z NN a 32 8 2 Environmental conditions Environment Conditions Relative Atmospheric Lighting T t 2 iti emperature humidity pressure conditions Storage 5 800 hPa 93 1060 hPa y Operating From To GC 15 800 hPa An To 40 C 93 060 hPa y Be aware that when a bed has been stored under low temperatures it must adjusted to operating conditions before use Dust and lint have no effect on operation of bed 33 Shipping brackets Transport brackets 34 8 3 Weights Medley Ergo standard bed MedleyErgobedend lpiece MedleyErgolowbedend ipiece Mattress support with wood slats mains cord mattress retainers amp hand control head end Mattress support with steel slats mains cord mattress retainers amp hand control head end Mattress support with wood slats amp mattress retainers footend Mattress support with steel slats amp mattress retainers footend Medley Ergo accessories Medley Ergo full length woodsiderails Iside Medley Ergo full length steelsiderails Iside Scala Basic 2 side rail L Ble66su cas vox eu RR KK KK RR RR ERECTAE ves Verso side rail piste e ei E RR KK RR dece dee KK je A e RNN E E A icis yi emk edemeyen rk Mattresssupporte
3. KK kk kk 27 8 Technical specifications kk kk kk kk KK KK KK KK KK KK kK kk kk kk kk de 32 8 1 Dimensions gara kK kK kK kk kK kk kk kk kk 32 8 2 Environmental conditions kk kk a RR IR Rh 33 muc 35 J Waste disposal x ceno recati RR TS RUE REP ERR ER eee pee AUR GS eee PUE REAL ain a6 a 4 35 Congratulations Congratulations on your choice of Invacare Medley Ergo nursing bed Medley Ergo is the dismantable nursing bed from Invacare designed especially for home nursing Medley Ergo bed provides ease of handling combined with good functionality Please read the entire user manual before using the bed If you have any questions concerning the use or maintenance please contact us All indications of right and left are based on that a person lying on his back in the bed Please note that there may be sections in this user manual which are not relevant to your bed since this manual applies to all on the date of printing existing modules General information I I Product use e The Medley Ergo bed has been developed for domestic care long term care and features a comfortable sitting and lying position for the patient Furthermore ergonomical operation for the carer is ensured e The Medley Erg
4. Service and maintenance of the Medley Ergo must only be performed by personnel who have received the neces sary instruction or training Bed ld no Date Initials mE Visual inspection of all parts of the bed plastic deformation and or wear and tear of welded joints ft ft Check that mains cable and plug are intact Check the running of the cable Check all remaining cables for damages Also check the running of the cables Check of siderails fixing and locking movement Check the wheels security braking and free rolling 23 6 3 Maintenance amp e Personnel who have received the necessary instructions or training may only perform Service and maintenance of the A 5 When relocating the bed service must be carried out according to the service schedule Before use Ensure that all manual and electrical parts functions correctly and are in a secure state After 3 months Ensure that all manual and electrical parts are functioning and tighten bolts screws nuts etc Every two year Service should be performed after the maintenance chart See chapter 6 2 Maintenance chart Modification of Medley Ergo is strictly forbidden without authorization of Invacare 6 4 Lubrication plan We recommend lubricating the bed according to the following instructions Points of rotation in mattress support and base frame Lubricate with oil a 2 Motor attachment
5. Ses Height adjustment of the bed Key for locking and unlocking the electrical functions Insert the key below each button to respectively lock and unlock the hand control functions HL 70 with integrated ACP function HB70 The beds can be equipped with a hand control that activate the tilt of the mattress support this hand control can be ordered from Invacare please refer to the section Order numbers for accessories Medically trained persons must only operate the tilt function otherwise there is a risk of injury to the patient s health To deactivate bed functions remove mains cord plug from power outlet 13 3 2 Operating the braking castors Marks from the castors might under special conditions appear on different types of absorbing floor covering 1 including untreated or badly treated floors In matters of doubt Invacare recommends placing a suitable kind of protection between the castors and the floor Do not release the brake with the fingers as there can be a risk of trapping cidents during entering or exiting the bed and during handling of the patient N Always apply the brakes when there is no need for moving the bed to avoid ac When the bed is positioned correctly at least one castor at the head end and one castor in the foot end must be braked Braking Step on the pedal pos Releasing the brake Push up on underside of brake pedal Each of the castors is provided with brakes for
6. accessories or any other moving part of the bed Make sure that there are nothing under over or near the bed that can limit the movement of the bed or the mattress support such as furniture window frames and storage boxes Always lower the bed to the lowest position before leaving the patient in the bed unattended There is a risk of entrapment of fingers in the bed moving parts Never lower deck height of bed more than 40 cm when using bed with a lifter Use ground clearance info in section 8 of this manual to determine lifter compatibility with bed before use It can be dangerous to roll over the mains cable e Do not bring mains cable into moving parts e Disconnect the plug from the mains before moving the bed e he cables must be mounted in such a way that they are kept clear of the floor and do not block the castors We recommend to mount the mains cable on the hook for this purpose see picture 7 1 5 Warranty The warranty covers all material and production defects for 2 years from the date of delivery provided it can be demonstrated that such defects were present before delivery All manufacturing faults or defects must be promptly reported Invacare may repair the fault or replace the component The warranty provided by Invacare does not cover additional costs trans port packaging labour sundry expenses etc are for the customer s account The warranty does not cover Damage caused during transport tha
7. both lengthwise and crosswise locking The brake is foot operated a Placement in the room a N The bed must be placed so that the height adjustment up down is not obstructed BE E i by for example lifts or furniture Otherwise there is a risk for injury on user and or equipment 3 3 Adjustment of leg section optional Operate the leg section by lifting the leg section Up Lift the leg section Down Lift the leg section right up and then lower the leg section 3 4 Emergency lowering of the backrest and or thigh section An emergency release of the mattress support may be necessary in case of e g a power or motor failure An emergency release of the height adjustment is NOT possible 14 Remove the plug from the mains before emergency release of the mattress support Hold the backrest 2 Remove the pipe pin from the backrest motor 3 Lower the backrest motor 4 Lower the backrest A minimum of 2 persons are required to emergency release a mattress section Both persons hold the mattress section One person pulls out the pipe pin Both slowly lower the mattress section until it is completely down 4 Accessories When 3 4 length using side rails you must always make sure that the distance between the bed end and the side rail s handle upper edge is between 3 lt 6 cm for the the head section and more than 32 cm for the foot section There is a risk of entrapment of fingers on assembly and operati
8. environment in which radiated RF disturbances are controlled The customer or the user of the medical bed Medley Ergo can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the medical bed Medley Ergo as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter Rated maximum output m of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d WP d WP Ei For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE I At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 3l 8 Technical specifications e n Further information available upon request of Invacare 8 1 Dimensions All measurements are stated in cm All angles are stated in degrees All measurements and angles are stated without tolerances Invacare reserves the right to change the stated measurements and angles
9. handle The distance between the lifting pole and the mattress Minimum at 16 cm mattress 36 59 cm Maximum at 10 cm mattress 42 65 cm 4 4 Fitting the mattress support extension Remove the foot bed end from bed 2 nsert bed extension into frame of bed and secure with nut and bolt 3 Reinstall foot bed end onto bed 4 Add mattress deck extender to foot end of mattress deck 4 5 Mattress Safety aspects regarding combination of side rails and mattresses In order to obtain the highest possible safety level when using the bed in combination with side rails it is important to respect the minimum and maximum measures for mattresses stated in the table To mitigate the possibility of the user crawling over the top of the side rail with the risk of severe injury due to e g falling a mini mum vertical distance of 22 cm between the top of the mattress and the top of the side rail are to be respected The maximum mattress height in combination with a given side rail is stated in the table The Medley Ergo is equipped with mattress support brackets These brackets must be properly installed on both sides of the bed in combination with the correct mattress size to prevent patient entrapment and risk of suffocation To prevent head entrapment between the inside surface of the side rail and the mattress with the risk of e g suffocation a safety maximum space has to be respected When especially very thick mattresses or soft
10. points to mattress platform Lubricate with oil uo 3 Upper attachment point of Rastofix fitting Lubricate with oil ad NB The full length side rails gliding system must not be lubricated with oil otherwise the side rail bars will move sluggishly Lubricate with medically clean oil e g KEN WO 50 order no 81 3239 NB Remove the mains plug from the socket before cleaning 24 7 Electrical system Voltage supply Max current input ntermittent periodic motor operation Degree of protection Sound level Protection class Class II Alternating current Direct current Type B Applied Part Applied Part complying with the specified requirements for protection against electrical shock according to IEC60601 1 Potential equalization Indoor use only Thermal fuse product 230V 50 Hz 2 A 10 max 2 min 18 min IPX4 45 50 dB A np 4 gt e r The bed is not provided with a mains switch so the mains plug is the only separation from the mains 25 7 1 Trouble shooting the electrical system amem e ey 7 Motor plug not pushed fully into 1 Push the motor plug right into the control box control unit The motor is not running 2 The motor is defective 2 Replace the motor The relay in the control box makes a clicking noise 3 Motor cable is damaged 3 Replace the cable 4 Control box defective 4 Replace the control box 5 Hand control defective 5 Replace the hand contro
11. that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the medical bed Medley Ergo requires continued operation during power mains interruptions it is recommended that the medical bed Medley Ergo be powered from an uninterruptible power supply or a battery Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Power frequency 50 60 Hz Power frequency magnetic fields should be at magnetic field 3 A m 3 A m levels characteristic of a typical location in a typical commercial or hospital environment IEC 61000 4 8 U_ is the a c mains voltage prior to application of the test level Portable and mobile RF communications equipment should be used no closer to any part of the EQUIPMENT medical bed Medley Ergo including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF IEC 61000 4 6 d WP 1 3 5 Radiated RF d NP 80 MHz to 800 MHz 1 IEC 6 1000 4 3 d Ap 800 MHz to 2 5 GHz Ei 29 IEC 60601 test level Compliance level Electromagnetic environment guidance where p is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation dis tance in metres m Field strengths from f
12. the motor cable of the foot end on the hook at the foot end Run the backrest to its top position Place the head end motor cables on the 2 hooks at the head end and through the pipe pin of the backrest motor 9 Fasten the safety clip to the control box 10 Verify that the mains cable strain relief is attached to bed frame For Medley Ergo with 24V external power supply verify that mains cable is attached to control unit power supply cable oo oS Improper routing kinking shearing or other me After completion of assembly always check that bed is fully functional before placing patient in bed chanical damages of cables could lead user injury 2 4 Dismantle the bed e Dismantle siderails and lifting pole e Bring the bed to its lowest position and adjust all mattress support sections to horizontal position e Disconnect the power card Roll the cable onto the hook of the head end of the bed e Disconnect all the motor plugs from the motors do not disconnect the backrest motor cable cable e Dismantle the safety comb from the control by means of a tool e g a flat headed screwdriver e Pull out the cables of the bed end motors and leg section motor Separate the mattress support from the bed ends e Divide the head and foot end of the mattress support 3 Operating the bed UP DOWN UP DOWN A y 3 1 Operating of the hand control unit Y Sitting position Thigh section optional Vos
13. agnetic immunity The medical bed Medley Ergo is intended for use in the electromagnetic environment specified below The customer or the user of the bed should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 Electrostatic transient burst IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 1 I 6 kV contact t 8 kV air t 2 kV for power supply lines kV for input output lines kV differential mode 2 kV common mode lt 5 U gt 95 76 dip in U for 0 5 cycle 40 U 60 76 dip in U_ for 5 cycles 70 U 30 76 dip in U for 25 cycles lt 5 U 795 dip in U for 5 sec 6 kV contact t 8 kV air 2 kV for power supply lines kV for input output lines kV differential mode 2 kV common mode lt 5 U gt 95 dip in U for 0 5 cycle 40 U 60 76 dip in U for 5 cycles 70 U 30 76 dip in U for 25 cycles lt 5 U 795 76 dip in U for 5 sec Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 76 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be
14. er side rail bar is then installed onto the lower pawl Again please ensure that pawls and gliders are completely engaged See illustration K 8 The locking finger screws are installed and tightened 9 On the Medley Ergo low version please set one warning sticker on each end of the lower pawls of full length side rails The 4 stickers are put in the bed ends packaging When lowering the side rails there is a risk of entrapment between side rails and mattress support There is a risk entrapment on assembly and operation of the side rails There is a risk of foot crushing the floor and the lower pawl of the side rails when the mastress support and the side rails are in lower position 17 4 3 Lifting pole Position the lifting pole in such a way that the handle extends inwards across the bed Always lock lifting pole with thumb screw If the lifting pole is used while the handle has been turned away from the bed the bed can tip when the handle is used Maximum load of the lifting pole is 80 kg Make sure that the lifting pole screw is properly tightened Mounting Insert the lifting pole into the lifting pole tube and fix with the thumb screw Adjusting the height of the lifting pole handle Loosen the cord as shown in A 2 Thelifting handle can now be adjusted to the desired height 3 Press the cord together as shown in B and check that the cord is locked in the cord lock by pulling the
15. ixed RF transmitters as deter mined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol c NOTE At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the medical bed Medley Ergo is used exceeds the applicable RF compliance level above the medical bed Medley Ergo should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the medical bed Medley Ergo o Over the frequency range 150 kHz to 80 MHz field strengths should be less than V J V m 30 Recommended separation distances between portable and mobile RF communications eguipment and the medical bed Medley Ergo The medical bed Medley Ergo is intended for use in an electromagnetic
16. l Control box is in order except 2 as ali iPod is hand L one direction on one channel Control box defective Replace the control box Motor is eee but the spindle does not move The motor cannot The motor cannot lif full load full load nee damaged Replace the motor Motor noise but no movement of m rod Piston rod Piston rod operate inwards and not outwards inwards and not outwards 26 7 2 Electro Magnetic Compliance EMC Guidance and manufacturer s declaration electromagnetic emission The medical bed Medley Ergo is intended for use in the electromagnetic environment specified below The customer or the user of the bed should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance M ie The medical bed Medley Ergo uses RF energy only for its internal function CISPR Group Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment partly me nie The medical bed Medley Ergo is suitable for use in all establishments including Class B domestic establishments and those directly connected to the public low voltage CISPR L partly power supply network that supplies buildings used for domestic purposes Harmonic emissions Class A IEC 61000 3 2 Voltage fluctuations flicker emissions Complies IEC 61000 3 3 27 Guidance and manufacturer s declaration electrom
17. l side rail 165 cm x 46 cm mounted on the side tube Release button in the head section I pair 1530817 0152 Scala Decubi 2 3 4 length collapsible steel side rail 168 cm x 54 cm mounted on the side tube gg Release button in the head section I pair 1554518 0152 Verso ll 3 4 length collapsible steel side rail 156 cm x 40 cm mounted on the side tube UCET Release button in the foot section I pair irl o ees Support handle 25x80 cm pc z x gt e gt a o oggggzgggggggggggxzxzPz P m Support handle 40x30 cm pc wr x ae cPxcZm_ _ _ mmprrm Support handle 40x50 cm pc OZGA ya epp gt o gt or gt oe reoeoe_aeeeebbebbbbbbbbbmeakbmmmnmn Support handle 25x30 cm pc hij ee Support handle 40x95 cm pc rare lav 10 cen eee 2 Ku han bd YY EE ye Wood Cover for bed ends beech one pair 55007750152 5 e e EY ER A PEY RE e Extended mattress support 15 cm 42 3980 7085 e grek a D pP US edu dq pat omer aun Besa ge eri Re dr atu quede Mattress retainer Ric eens Shipping fittings Medley Ergo unpainted 156039950 1582 upare rar bU ere fot aut eripe qc Sen i diii HER P e ep e Red E Transport fittings Medley Ergo painted REEL M I Hand control HB72 EUPE P
18. ly hand control and motors are protected according to IPX4 The IP rating is a measure of protection against ingress of foreign matter into a product Remove the plug from the mains voltage before moving the bed The cable must be kept clear off the floor and the castors during transportation The adjustment area of the mattress support is 40 80 cm or alternatively 33 73 cm for Medley Ergo The adjustment area of the mattress support is 28 68 cm or alternatively 21 61 cm for the Medley Ergo Low The angle between the lower leg section and horizontal is adjustable from 0 to 5 Certification The Medley Ergo bed is marked with the CE mark in accordance with directive 93 42 EEC concerning Medical Apparatus Medley Ergo bed have been approved according to IEC 60601 2 52 Beds provided with tilt function can not be approved according to IEC 60601 2 52 The Medley Ergo bed has been approved and marked with the T V mark The Medley Ergo bed has undergone a risk analysis according to EN ISO 14971 2007 The Medley Ergo bed has undergone a usability analysis according to IEC 60601 1 6 Invacare is certified according to DS EN ISO 9001 and ISO 13485 which ensures that our customers are always supplied with products of uniform quality Throughout the entire production process our materials products are quality controlled by the operators Additionally a final test is carried out prior to packaging and shipping If the product does not cor
19. mattresses low density or a combination of these it is important to focus on this risk The minimum width and length of mattresses in combination with a given side rail is stated in the table Medley Ergo Max Mattress Min Max Min Min Max Height Mattress Mattress Mattress Mattress Mattress Height Width Width Length Length For beds equipped with Mattress Support Extension add 15 cm 4 6 Transport brackets See instructions at the last page in this User manual to reconfigure bed for transport or storage 20 5 Order numbers for accessorles OR n e a en Gn nn fab tels e A EA SENE AU eas Wood side rail full length 1 pair Isa 2gm nnnnRm Wood side rail full length 15 cm l pair 5600 76 0 182 aaa IEEE ndo weg PIRE RR REESE PER EIER e ese Steel side rail full length 1 pair N 177461 7042 nr Sliders side rail Extra full length 1 pair 1535 470 0157 keme epik ene ekime d n EE REA Wood side rail full length height extender pair 1528930 0152 Scala Basic 2 3 4 length collapsible steel side rail 168 cm x 40 cm mounted on the side tube r Release button in the head section I pair 1529745 0152 Scala Basic Plus 2 3 4 length collapsible steel side rail 168 cm x 40 cm mounted on the side tube T ee Release button in the head section I pair 1530436 0152 Scala Medium 2 3 4 length collapsible stee
20. o bed is defined to be used according to Application environment 3 Long term care in a medical area where medical supervision is required and monitoring is provided if necessary and medical electrical equipment used in medical procedures may be provided to help maintain or improve the condition of the patient Application environment 4 Care provided in a domestic area where medical electrical equipment is used ocd to alleviate or compensate for an injury disability or disease 145 kg e Ifthe patient height exceeds 2 meters it is recommended to use a mattress support extension e The max patient weight 145 kg provided that the weight of the mattress and the accessories 9 do not exceed 35 kg Safe working load weight max 180 kg 80 kg e The bed is not intended for children under 12 years and psychiatric patients N e Furthermore Be aware of the limitations of use mentioned in the warnings chapter 1 4 e The following parts may be in contact with user during normal use of bed bed ends mattress deck mains cord and hand control e The Medley Ergo is mobile within a room with a patient in it all castors are lockable e The Medley Ergo bed is suitable for disinfection with household disinfectants 3 The Medley Ergo is suitable for refurbishment The estimated product life cycle is 5 years for a Medley Ergo bed The bed is quipped with loose cables which have to be exchanged if damaged The control unit external power supp
21. on of the side rail 4 1 Mounting dismounting the Scala 2 series and Verso steel side rail Mounting The side rail must be mounted according to the assembly instruction side rail Tighten the side rail with the finger screws Dismounting Loosen the finger screws and remove the side rail 4 2 Fitting the full length side rails A Disassembled gliding system The installation of steel and wooden side rails are identical Pawl The gliding system consists of 5 different parts which all must be installed simultaneously x A disassembled gliding system is shown in illustration A wez L B Removal The 4 pawls are clipped onto the glider B and C 2 For an ergonomic working position and easier assembly raise the bed to 1 3 of full height position The gliding system of one bed end is guided into the bed end according to illustration D and E 3 The upper side rail bar is installed according to illustration G 4 The gliding system in the opposite end are installed onto the upper side rail bar according to illustration H 5 The side rail bar is raised until an audible click can be heard Please be sure that the gliding system has engaged properly into the first hole see illustration F and 6 The lower pawl is unfolded and the lower side rail bar is installed onto the pawl according to illustration E and J 7 The lower pawl in the opposite end unfolded see illustration E and the low
22. respond to the quality demands of Invacare it will be discarded If contrary to our expectations a problem should arise in connection with the delivered product please contact your Invacare sup plier An address list is Shown on the back side at this manual Invacare reserves the right to alter product specifications without prior notice 4 1 3 Labels and symbols WARNING This symbol warns of dangers If the necessary measurings not are taking it can result in injury or damage to the bed NOTE This symbol inform about important information and guidelines Instruction Reference to the user manual n bi r internetionsl Seri 1 i i Aca Partaga a aa C Electrical information Tel 81 225 183 380 Fax 351725 105 738 3 See chapter 7 Type no of the bed npe w 1557962 0182 ui ca 750 Hz MA rm g Date of production tev xe nn 2 a Mm isin Ae A P Medley Ergo Serial no e Symbols See chapter I and 7 1 4 Safety and Warnings Invacare accepts no liability for any use change or assembly of the product other than as stated in this user manual Accessories not mentioned in this manual must not be used The Medley Ergo bed fulfils all requirements regarding maximum distances However if the bed is used for the care of patients with a small body dimensions it must be especially noted that there is a risk of such a patient slipping through the openings between the side rails or th
23. rough the opening between the side rail and the mattress support The bed must not be used by patients under 12 years of age or by patients with body size equivalent to an average l2 years old or smaller The bed in combination with side rails must not be used for persons under 45 kg or under 150 cm height Either for persons that are restless spasms or confused unless a professional risk assessment has taken place and been accepted When using side rails it is essential to ensure correct fitting otherwise there is danger of entrapment suffocation between mattress support side rail and bed end Do not place any objects underneath bed Electromagnetic interference between the bed and other electrical products can occur To reduce or eliminate such electromagnetic interference increase the distance between the bed and the products or switch them off This medical bed can be used together with medical electrical equipment connected to the heart intracardially or the blood vessels intravascularly provided that following points are respected The medical bed should be equipped with means for potential equalization connection marked out by a symbol shown in the back of this manual Medical electrical equipment as describe above should not be fixed on the bed s metallic accessories such as side rails lifting pole drip rod bed ends ect In addition the medical electrical equipment power supply cord should be kept clear of the
24. s tOE Soa roa rmx exa kk kK KK kK KK KK KK KK KK da 14 3 3 Adjustment of leg section KK KK KK KK KK 14 3 4 Emergencyloweringofthebackrestandorthighsection 14 Mr GSO Db n o_wrm meee eemmmm yz 15 4 1 Mounting dismounting the steel side rail Scala2 amp Versoll 15 4 2 Fittingthefulllengthsiderails KI KI KK KK KK KK KK KK K 16 k _eessee e e e eo o e e e eW mmmbbrebbbbb 18 4 4 Fittingthemattresssupportextension 0 0000 19 45 ene a AE a EE NEN 20 46 Transport bi Meis senemi opema Ak EE aa a op am mean medi 20 5 Ordernumbersforaccessories KK kk kk ees 21 6 Cleaning and Maintenance 2 usce kk kk kK KK KK KK KK KK KK kk kK kk kk we ERR e 22 OEC un e ek e e e Tro 22 6 2 Mal tenanc chart acus s 344 amp l ERE amana 23 6 3 Mena 44 4 4 s z na n dn S RR PO RCRCR B9 dr CORR He de qi hike d 4 Oh 4h9 kinasi 24 OF Lubrication DIAN 4220 aaie PE kk d n ak EE Ye 24 7 Electrical system kk kk kK kk KK KK KK KK KK KK KK KK KK KK KK KK KK K RUP NE SUPE 25 7 Trouble shooting the electrical system kk kK kK KK KEK KK KK KK KK KK KK KK KK K 26 7 2 Electro Magnetic Compliance EMC
25. t is not directly reported to the forwarder at the moment of delivery Repairs performed by unauthorized centers and personnel Parts subject to normal wear Malicious damage or damaged caused by improper use of the bed 2 Receiving the bed 2 1 Parts of the bed e 1 Check the packaging if the bed shows any signs of damage upon delivery please see the terms of delivery Follow part must be provided Pair of bed ends l 2 Mattress support head end 3 4 pcs of mattress support bracket 4 Motor for head end 5 Mattress support foot end 6 Motors for bed end 7 Castors 8 Hand control Optional Transport fitting Support handle G Lifting pole Safety floor mattress Rastofix for leg section Wood cover for bed ends Mattress support extension 15 cm Eguipotentiality plug Motor for leg end Side rail wood Side rail steel Side rail Verso 1 Side rail Scala 2 series Side rail height extender 2 2 Assembly of the bed For questions concerning the assembly of your bed please contact your dealer or Invacare customer service Mattress Support Push the foot end of the mattress support onto the head frame Tighten with two thumb screws Turn the four mattress handles of the backrest and footrest up Turn the locking ring on the bed ends to the position open Latch the mattress support to the bed ends and press firmly into position 6 Make sure that the bed ends are in
26. taylordolman Invacare Medley Ergo Bed Bett Cama Lit Letto User manual EN CE This manual must be given to the user of the product Before using this product read this manual and save for future reference Yes you can Invacare Medley Ergo User manual EN 3 Contents l General IDIOT SHOE su eo ane Henao cee n a b b J 0k EREE W8 AWA ER ERES P 3 BRE Product USC m MM J DM MD J JM JJJ J N aIAIEI NIMJJiMii Z BHNJMMDTuZ_MAMAMNAMMWMDEDMRnRMsrmDDRMMe 3 eZ CCPEMCAION 26 eons E E YE ERRET e es 4 leo Labels and SYIDO S lt sace s aah dak Rb H k aque Bara PRU n GR n qi kd di k A a Gan d l d 5 I 4 Safety and Warnings s sas acq arx KK KK KK KK KK kK kK kK dup Eun RR Rd e e e e RE ES 6 La YYY eeror __rr ll n ncs zxpbm_ r r EE e PE 8 2 Receiving te DE ca gg sun saa xank 002804 2 00 0 EEEE ADE Ki eo UCM W i i Qir DN idik a v led 9 ME Wares OP Ne De Lou soo ah ea ke seer ered pease beac eee eee aes RR EN E Ee 9 2 2 Assemb Ol tie DEG cheers cue Kale e020 n o DERE ERE PLU QALA QA n BAK een a 10 2 9 WVIFIBE ame kadem oem e mm 2 4 Dismantling the bed same yunus kK KK KK EG KK kk EU kk e kk e e e le e 12 2 Operatinie the Ded i c dep k k n dne l la x mamak ent inka 13 3 Operatingofthehandcontrolunit eee 13 3 2 Operating the braking Ca
27. the same level It is possible to level the bed in high position A and low position B 7 Turn the locking ring to the position lock Always put locking ring in the upper slot as shown in pictures The locking rings locks ensure that the mattress support is not accidentally lifted clear of the bed ends t is normal to have a small space between the top of the slot on the mattress hook and the bracket on the bed end VU d 0 No Rastofix j BD i gt li gt L 1 1 jj Ft d amp 2 Press bottom of Rastofix at the foot end of the bed 3 Squeeze upper end of Rastofix together o 4 Rotate Rastofix between tabs and release 10 Control Box The control box is latched to the backrest motor The control box is provided with a label with symbols showing where to connect the motors plugs Backrest motor Thigh section s motor Bed end motors Hand control 2 3 Wiring In order to prevent the cables from being torn apart when activating the motors it is very important to follow the instructions below Connect the cables of the bed end motors head and foot end directly to the control box 2 Wire the cable of the bed end motor foot through the pipe pin of the thigh motor 3 The cable of the thigh motor must be directed through the pipe pin of the thigh motor Connect the mains cable to the 230V plug Run the motors of the bed ends to their top position Place
28. xtension 15cm Jransport brackets uas oat eaae p ine be KK oe ei e ei Woodcoverforbedends lIpiece 9 Waste disposal 14 7 kg 13 7 kg 16 8 kg 15 9 kg 16 0 kg 15 1 kg 9 2 kg 9 1 kg 7 4 kg 8 3 kg 4 2 kg 0 3 kg 1 8 kg 2 1 kg 5 2 kg This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment WEEE Directive 2002 96 CE This product may contain substances that could be harmful to the environment if disposed of in places landfills that are not appropriate according to legislation The crossed out wheelie bin symbol is placed on this product to encourage you to recycle wherever possible Please be environmentally responsible and recycle this product through your recycling facility at its end of life All wooden parts must be dismantled and sent for incineration All electric parts must be dismantled and be disposed of as electric components Plastic parts must be sent for incineration or recycling Steel parts and castors must be disposed of as waste metals Waste disposal must comply with the laws and regulations pertaining waste handling in each country 35 taylorolman tel 0845 838 0600 email info taylordolman com www taylordolman com