        GasCheck G3 Leak Detector Manual
         Contents
1.    GASCHECK G SERIES    CONTACT US      ACS                   PHONE FAX    Toll Free  800 465 1004  Phone  801 486 1004  Fax  801 486 1007       ADDRESS   LACO Technologies  Inc   3085 West Directors Row  Salt Lake City  UT 84104  WEB    www lacotech com  info lacotech com    SMT 07 1015 rev A2  ii    O 2010 LACO TECHNOLOGIES  INC     CONTENTS     2 69   m                      M un s  1  PE  icq dM 1  Bi OVE                                           2                                                  12  Ds NIAINTENANCE               16       2010 LACO TECHNOLOGIES  INC     1  SCOPE    2  SAFETY    LACO USER MANUAL   GASCHECK G SERIES    This manual contains safety  operation  and maintenance information  for GasCheck G Series handheld sniffer leak detector  Our leak detectors  are designed to ensure safety when used properly  It is the responsibility  of the user to follow safety related warnings  cautions  notes  and other  requirements described in this manual     Should a question  comment or problem arise  the information below   is provided as a guide  When returning a unit for factory service or  calibration  contact LACO for a return material authorization number   RMA   The device should be boxed securely and contain all contact  information  including return shipping information  If the device is being  sent to the factory for service  a written statement of the problem or  symptoms should then be included  The RMA number must be on the  outside of the package or ind
2.  Thermal conductivity    DETECTION TIME TO T90 1 sec  short probe  9 sec  long probe     CLEAR DOWN TIME 1 sec  short probe  9 sec  long probe   BATTERY TYPE Alkaline or NiMH AA  qty 4   BATTERY LIFE AT 20  C  68  F    40 hours  20 hours when using the backlight    2  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    Table 2  GasCheck Replacement Parts    PART NUMBER DESCRIPTION    A 08045 Long probe  A 08043 Short probe    Replacement battery holder    CM515 0 4101FAG 1 Leak Standard with Refillable Helium Reservoir   5x10 4 cc sec       515 0 4105    0 1 Open Style Leak Standard for use with helium  regulator and tank  5x10  4 cc sec    3 1   APPLICATIONS       Common applications include     e Quality assurance  Testing seal integrity after product manufacture    Laboratory applications  Detection of leaks from mass spectrometers  and chromatograph equipment    Industrial  Leaks from gas cylinders  pipe work and process equipment  e Medical  Testing of membrane materials and sealing of glove boxes    e Pneumatic  Valve seal testing     3 2   UNITS OF MEASUREMENT    The GasCheck is calibrated against a 5 E 4 cc s helium leak to allow  volumetric readings and also a 5000ppm helium to allow measurement of  concentrations     Table 3  Selectable Units of Measurement    UNIT DESCRIPTION    CC sec Cubic Centimeters per Second offers a reading that indicates  the volume of gas escaping into atmosphere from a single  point  i e   leakage from a hold in a ga
3.  When the battery symbol is filled in  the battery is full  When   the battery symbol is outlined  the battery is exhausted   BACKLIGHT The backlight symbol shows the status of the backlight  even  INDICATOR in bright daylight     GAS The GasCheck G2 and G2 measure the leak rate of gas  MEASUREMENT leakage in whichever units are selected     SOUND The symbol shows if the sounder is on or off  If the semicircle   INDICATOR on the right of the symbol is present  then the sound is  switched on  If not  the sounder is switched off  The beep  that occurs with each key press can not be switched off     BAR GRAPH The bar graph increases as gas is detected  This graph is not  scaled and should be used for indication only  The GasCheck  G1 has no measured units  so it relies on this bar graph to  indicate leakage  When gas is detected and the bar graph  has reached its maximum level  a different sensitivity level  can be selected to give a wider detection range     PEAK HOLD When selected  peak hold displays the maximum  INDICATOR measurement on the display  Pressing the ENTER key resets  the reading but also logs the peak reading in memory        8  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    3 8 FUNCTIONS  The GasCheck instruments have a variation of functionality  Each of the  functions listed below has a number to the right of the heading  This    number indicates which function is included with each model     Table 6  GasCheck Function Comparison    
4.  any used batteries  shall be in accordance with local and national safety and environmental  requirements     2 4 CALIBRATION    Calibration service is recommended on a yearly basis  LACO Technologies  has an A2LA Accredited calibration lab which complies to ISO IEC 17025   2005 and ANSI NCSL Z540 1 1994  Also complies with ISO 9001 quality  standard  Call LACO for more information on calibration services     The GasCheck is a portable handheld gas detector instrument  predominantly used for detecting gas leaks and can detect almost all  gases to varying degrees  It is powered by standard AA size batteries and  will accept both Alkaline and Nickel Metal Hydride  rechargeable  types     The GasCheck uses thermal conductivity as its means of detecting gas  which offers a robust sensor technology that requires practically no  maintenance beyond annual servicing     All GasCheck G Series models have an easy to use graphical interface with an  intuitive keypad allowing simple function selection and adjustment  It can  be upgraded to add features without the need to be returned to the supplier  and without having to modify the internal firmware  It has an LCD display   LED indicator and audible sounder than indicates the detected signal     Table 1  GasCheck Specifications    COMPONENT DESCRIPTION  TEMPERATURE 0 to 50   C  32 to 122   F   STORAGE TEMPERATURE  25 to 70   C   13 to 158   F   L 390 x W 60 x H 50mm L  15 35  x W2 35  x 2 0      WEIGHT 0 5 kg  1 0 Ib     DETECTION TYPE
5. 0 124 im     cc sec J  ppm    mgm      J  wr    DISPLAY      5s IDE 2  ppm 0 05    UPGRADE    LACO USER MANUAL   GASCHECK G SERIES    DESCRIPTION    The GasCheck instrument has the ability to data log 10 readings within its internal  memory  To tore a reading  press the ENTER key while on the main RUNNING SCREEN   When readings are stored in memory  a symbol will appear on the main viewing screen     When the memory is full  the Data symbol will flash  The stored data can be viewed by  scrolling through the logged readings using the UP and DOWN keys     To delete the stored data    1  Press and hold the ENTER key    2  The TRASH CAN symbol will flash  El Dala  3  Continue to hold the ENTER key until the data is deleted    4  Press the ESC key to exit to the MAIN MENU     GasCheck instruments have the capability of displaying detected readings ina  variation of units  ppm  cc s  mgm3  and g yr     Use the UP and DOWN keys to select the desired unit  and then press the ESC key to  exit to the MAIN MENU     NOTE  the Custom Calibration routine automatically selects the type a calibration  depending on his setting  See 5 1 CALIBRATION     e CC SEC   cubic centimeters per second offers a volumetric leak rate that directly  measures the volume of a gas escaping from a given point  It is also known as ml   sec  milliliters per second   The GasCheck can typically detect helium leaks down to  0 0005 of a cubic centimeter per second    PPM   Parts per million  This unit indicates a con
6. 4  Insert the instrument s probe fully into the sample bag containing  the gas and press the ENTER key  The instrument will automatically  sequence through the cold and hot stages of calibration but when  finished gives a summary of values     5  Pressthe ENTER key to return back to the calibration option screen   Press the ESC key to start using the instrument or press the ENTER key  to calibrate again     NOTE To avoid errors in calibration  the following points should  be noted     e Avoid bending the inner probe as this will affect the GasCheck s accuracy      The GasCheck can detect changes in humidity and carbon dioxide  Avoid  breathing on the probe       Avoid pressurizing the sample bag  This will cause significant errors in  reading     5 1 3 CALIBRATION PROCEDURE USING CC SEC UNITS    Please read this entire procedure before proceeding with this calibration  routine     1  Check your LACO CalMaster    Leak Standard to ensure the pressure  reading matches the leak rate shown on the calibration stick  The  pressure gauge may be installed on your leak standard or for open  style leak standards the pressure gauge is set on helium regulator     2  Select the CUSTOM CAL symbol from the instrument s menu     3  Ensure the instrument is in clean air and then press the ENTER key  The  instrument will Zero  This will take a few seconds but once complete  a  READY message will appear     4  Insert the instruments probe into      CalMaster    Leak Standard s  outlet and press 
7. 60  R       Refrigerant R11 R 3 2 E 5 0 000032    11    2010 LACO TECHNOLOGIES  INC        LACO USER MANUAL   GASCHECK G SERIES    NAME    ABBREV  MAX SENS CC SEC   MIN  SENS  CC SEC    Hexafluoride  10    Refrigerant R410a 3 2 E 5 0 000032  Refrigerant R507 3 8 E 5 0 000038       4  BASIC  OPERATION    All GasCheck G Series models have the same keypads  The following  section explains the general functionality of each key     ESCAPE UP  This key is used to return the display  to the previous screen and to abort an  adjustment  Repeated pressing of this  key will return the display to the  Display screen        Use the UP key to scroll  through the function menu  and to adjust settings     ENTER 4   DOWN   This key is used to select functions and Use to scroll down or change  accept settings after a parameter has selection   been adjusted    4                  ZERO   Pressing the Zero key Zeros   out background readings   5     page 13    On Off   Press this key once to switch the  instrument  ON  To switch the  instrument off press and hold this key  until the bar graph reaches the right  side of the screen    This procedure has been designed to  avoid accidental switch OFF     12  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    4 1 STARTUP ROUTINE    Gas heck os       1  When the GasCheck first switches on  it will display its model number  and the gas that it has been calibrated with   serial  07 00001 2     second screen then displays the instrum
8. ECK G SERIES    DESCRIPTION    The GasCheck is calibrated using helium gas at manufacture  however other gases  can be detected and measured using a gas table within the instrument  Although  GasCheck can t differentiate between gases  it can quantify other gases using  response factors stored in its internal gas table  Gas groups 0  1  2  3  and 4 offer a  generic sensitivity     See Table 8  Gas Sensitivity to compare response factors for specific gases  For advice  on gases not listed  please contact LACO Technologies     To select an alternative gas    1  Use the UP and DOWN keys to select the desired gas  2  Press the ENTER key to confirm the selection    3  A tick  will appear    4  Press the ENTER or esc to exit to the MAIN MENU     If the gas to be detected does not appear in the instrument s internal gas table    then one of the five Gas Groups can be selected  Gas Groups have a variation of  sensitivities  Group 1 is high sensitivity and Group 5 is low sensitivity  Gas groups allow  the instrument to calculate leak rates with reasonable accuracy     Table 7  GasCheck Feature Comparison    FEATURE  CONTRAST          gt      gt   gt x   gt    gt  Ea  N  W    BACKLIGHT    PEAK HOLD   CUSTOM CALIBRATION  BATTERY TYPE   DATA LOG   UNITS OF MEASUREMENT  GAS TABLE    UPGRADE       Table 8  Gas Sensitivity    NAME ABBREV          SENS              MIN  SENS  CC SEC    Refrigerant R12 0 000027   Refrigerant R1301 0 000024   Refrigerant R134a 0 000028   Refrigerant R22 0 0002
9. ER MANUAL   GASCHECK G SERIES    Should a probe become blocked  use dry  clean  compressed air to blow  out the blockage from the instrument end of the probe     WARNING   Ensure safety guidelines are adhered to when using  compressed air     44 DETECTING LEAKS    Hold the GasCheck in one hand and draw the instrument probe along  the area to be checked at approximately 25 mm per second  When a  leak is detected  retrace the route of the probe at a slower rate until the  leak is located  Once located  hold the instrument over the leak until the  measurement stabilizes  The reading can be logged by pressing the  ENTER key     Some readings may be larger than the instrument can detect and the  measurement units will be replaced by  99999  on the display  In this case   use the UP or DOWN key to adjust the instrument sensitivity     When detecting very small leaks  probe movement at a rate of 10mm per  second may be required     The outer gray probe cover can be removed to allow better access to  restricted areas  If the outer probe is removed  please note the following     Avoid bending the inner probe  A bent probe will affect accuracy     To prevent probe blockage  avoid placing the probe in liquid or dirt     Care should be taken to ensure the brass sensor housing component  remains at a constant temperature  Avoid touching the brass sensor  block with fingers     The GasCheck can detect changes in humidity and carbon dioxide   Therefore  avoid breathing on the probe     The 
10. FUNCTION DESCRIPTION MODEL    CONTRAST Use this function to adjust the contrast of the GasCheck LCD  liquid crystal display   G1  G2  G3      Significant variations in ambient temperature conditions can cause the display to   x    x 32 appear too dark or too light        When this function is selected  a number will appear to the right of the Contrast    symbol as a percentage  Use the UP and DOWN keys to adjust the display the contrast  to the desired level  When satisfied with the contrast level  press the ESC key to exit to  the MAIN MENU     BACKLIGHT The user can select either BACKLIGHT ON or BACKLIGHT OFF  In daylight  the user may   G1  G2  G3  not be able to determine if the backlight is ON or OFF  A symbol on the main RUNNING   fom    VR      SCREEN shows the status               m   When the backlight is ON  it significantly reduces battery                  Backlight Off 40 hours of use  Backlight On 20 hours of use    The GasCheck has an audible sounder that increases frequency as detected gas levels   G1  G2  G3  increase  This sounder can be switched ON and OFF  Use the UP or DOWN key to move   the  tick  to the desired position and then press the ESC key to return to the MAIN        aji Sound         An audible          can also be heard whenever the key pad is a  Sound Off  pressed  This beep cannot be switched off     When selected  this function holds the highest detected   amp  Peak hold On G2  G3  reading on the display until the ENTER key is pressed    Pea
11. as  been calibrated by the user  To do this  move the  tick  under the CUSTOM  CAL symbol and press the ENTER key     The GasCheck requires either 5000 ppm helium concentration or a  0 0005cc sec  5 E 4 cc sec  leak rate as a reference  The GasCheck  automatically selects the type of calibration media based on which unit  the instrument has already selected     If the GasCheck is used to detect leaks measured using  ppm or mg m3   units  then the instrument must be calibrated using a 5000 ppm helium  concentration     If the GasCheck G is used to detect leaks measured using  cc sec  unites   the instrument must be calibrated using a 0 005 cc sec  5 E 4  helium leak     5 1 2 CALIBRATION PROCEDURE USING PPM UNITS    Please read this entire procedure before proceeding with this calibration  routine     1  Fill an empty  uncontaminated  sample bag with 5000 ppm helium gas  before starting the calibration procedure  Also remove the outer gray  probe from the GasCheck     2  Select the CUSTOM CAL symbol from the instrument s menu          option appears allowing the adjustment of the  ppm  value to ensure  the instrument matches the gas concentration being used  Use the up  and down keys to adjust between 4900 and 5100 ppm     16     2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    3  Ensure the instrument is in clean air and then press the ENTER key  The  instrument will Zero  This will take a few seconds  but once complete a  READY message will appear     
12. centration but can t be used to  quantify a volume of gas leakage                Milligrams per meter cubed  Like ppm  this unit is also a concentration  so  it can t be used to quantify a volume of gas leakage    G YR   Grams per year offers an alternative volumetric leak rate that directly  measures the volume of a gas escaping from a given point     The GasCheck has the capability of displaying a cc sec and g yr reading in either  decimal or exponent form   Readings in ppm and mg m3 can only be displayed in  decimal form      To change the display    1  Use the UP and DOWN keys to select the desired unit    2  Press the ENTER key to display the next screen    3  Use the UP and DOWN keys to select the desired display and then press the ESC key  twice to exit to the MAIN MENU     The GasCheck instrument can be upgraded using a simple upgrade code  This facility  allows the GasCheck to be upgraded instantly by the user in the field  See Table 7   GasCheck Feature Comparison to compare the various features within each model     To upgrade a GasCheck  simply purchase an upgrade code from LACO Technologies   enter the 12 digit number on the upgrade screen using the UP  DOWN  ENTER  and  ESC keys  Once all 12 digits are entered correctly the instrument will automatically  switch off     When the instrument is switched on  the new instrument model number will be  displayed        10    O 2010 LACO TECHNOLOGIES  INC     FUNCTION  GAS SELECTION               LACO USER MANUAL   GASCH
13. ent s serial number  the    I Il 3f u AF  e   A  Firmware  v2 25 instrument s firmware version and web address                 ze        EJ              Sa e  ee      3  A third screen appears to indicate the GasCheck is Zeroing it   s signal              4                Zero routine is complete  the instrument displays its running    Yt screen and is ready for use     42 USING THE GASCHECK G SERIES    WARNING  Before switching on the GasCheck  ensure the ambient air  is clean  The instrument automatically Zeros the sensor at  switch on  After the instrument has run through the start  up routine  adjust the instrument settings to the desired  levels  Check the instruments sensitivity using a CalCheck  or calibration kit        4100      NF 1  Switch the GasCheck on by pressing the ON OFF key  After the instru     0000322             i ment has completed its Zero routine  it will show its normal running    screen        Ensure the instrument is on the X100  times one hundred  range   While on the main running screen  use the UP and DOWN key to adjust  between sensitivity ranges  X100 is the most sensitive  X1 is the least  sensitive     2  Gasleaks tend to occur at pneumatic joints or welded seams  To  test for a leak  hold the GasCheck at a 45 degree angle to the object  being tested and drag the probe along the seam or joint at a rate of  approximately 25mm per second     3  Whenaleakis detected  the bar graph will start to fill and the  frequency of the audio output will i
14. icated on the shipping label     Please read this manual in full before using the GasCheck G Series  instrument  LACO Technologies  Inc  takes no responsibility for damage   injury or death resulting from misuse  misunderstanding or negligence  while using this gas detector  Please contact LACO Technologies for  clarification on any aspect of this manual that is not understood or for  additional information     2 1   QUALITY ASSURANCE    GasCheck G Series instruments are manufactured in compliance with  1509001 2000  which ensures that the equipment supplied to the customer  has been designed and assembled from traceable components  leaving  LACO Technologies calibrated to traceable standards     2 2 RESPONSIBILITY OF USE    Many gases are hazardous and can cause explosion  poisoning and  corrosion resulting in damage to property and life  It is the responsibility  of the person using this instrument to ensure it is being used in  accordance with this manual and that the instrument is functioning  correctly before use     The GasCheck G Series handheld leak detectors can detect a large range  of gases but some gases are more difficult to detect  It is the responsibility  of the user to ensure the GasCheck instrument has the sensitivity to detect  the required gas before reaching potentially dangerous levels     1  O 2010 LACO TECHNOLOGIES  INC     3  OVERVIEW       LACO USER MANUAL   GASCHECK G SERIES    2 3   DISPOSAL    Disposal of the GasCheck G Series its  components  and
15. k hold Off   Select Peak ON or OFF using the UP or DOWN key and then   press the ESC key to exit to the MAIN MENU    When using the GasCheck G  press the ENTER key to clear the held reading  The   reading held on screen will be logged when the ENTER key is pressed     WARNING  ZEROing the instrument in a contaminated environment will result in false  readings     CALIBRATION There are two Calibration settings that can be selected    FACTORY and CUSTOM  Factory calibration is carried out shortly NS Factory Cal  vm al   after manufacture and cannot be adjusted  However  custom O   x   x     calibration offers the ability to calibrate the instrument between       Custom Cal   3 g annual Factory calibrations  The GasCheck G ZEROs its reading du  at switch ON  Therefore  the Custom calibration routine has only    one stage  For more information  see 5 1 CALIBRATION      BATTERY The GasCheck will operate using AA size batteries in either      Battery exhausted  SELECTION Alkaline  non rechargeable  or Nickel Metal Hydride Y     rechargeable      Battery half full    EF Battery full        Alkaline    Select the battery type using the UP or DOWN key and then  NiMH ress the ESC key to exit to the MAIN MENU     Failing to select the correct battery type will not damage the instrument or affect its  ability to detect gas  However  it will cause the battery indicator to read incorrectly        9  O 2010 LACO TECHNOLOGIES  INC     FUNCTION  VIEW DATA    B ce           21000075 sec    0 0
16. lied  As air passes through the cell a constant level of heat is  transmitted to a Temperature Stable block through the air  This rate of heat  transmission is used to Zero the instrument     Fan  Air flow       NN               Probe       Display    When gases with different thermal qualities pass through the chamber   the amount of heat being transmitted to the Temperature Stable block  changes  These changes are measured  calculated and displayed on the  GasCheck as leak rates or gas concentrations     NOTE Some gases have similar thermal properties to that of  the air  therefore the GasCheck can only detect larger  concentrations of these gases     NOTE Some gases have positive and other gases have negative  signals  To simplify the instruments function  the GasCheck  only display changes as positive readings     NOTE The GasCheck G2 cannot differentiate between gases   Selecting a specific gas on the GasCheck G Series allows  the instrument to calculate concentrations of that gas only  if that gas is being detected     6  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    3 6 INSTRUMENT MAIN MENU    Once the GasCheck has run through its START UP routine it will display its  normal RUNNING SCREEN used when locating gas leaks  Before using the  instrument  the various settings should be adjusted to suit the application     Pressing the ESC key while viewing this RUNNING SCREEN will access the  instrument s MAIN MENU  Repeated pressing of the ESC ret
17. ll gases conduct heat by varying amounts  If an object is heated and   the source of the heat is then removed  the object will eventually cool  down to match the ambient air temperature  This action occurs because  the ambient air surrounding the object carries the access heat into the  surrounding atmosphere  This principal is also the same for objects that  are cooler than the surrounding ambient air  Dissipation of heat into an air  atmosphere is known and is a predictable rate  However  if the ambient air  is replaced with an alternative gas like helium  the rate at which an object  cools down changes     If the object mentioned above had its environment replaced with pure  helium it would cool down to the environment temperature about 6  times faster     5  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    The GasCheck G Series contains a heated thermistor bead that transmits  heat to a block of material that remains at a constant temperature  As air  passes through the detector chamber a constant amount of heat passes  from the bead to the block  Gases that are different to air will affect the  rate at which heat transmits from the bead to the block  These rates of  change are measured and displayed as leak rates     The diagram below shows the basic functionality of the GasCheck Thermal  Conductivity sensor  A fan draws a small flow of gas through the probe and  into the sensing chamber  The thermistor bead heats up when electrical  power is app
18. ncrease but will reduce as the  probe moves away from the leak  Return the probe to the suspected  leaking area and move slowly along the same area until the leak is  located  Once located  the probe should be held at the leak until the  numeric reading stabilizes     4  The bar graph offers a graphical indication only and should not be  used to measure a leak  You many find that the numeric reading will  continue to increases after the graph has completely filled  Should the  instrument detect a leak that is too large for the range selected  the  numerical screen will flash and display  9999  Use the UP or DOWN  keys to adjust the instrument sensitivity     13  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    Variation in temperature  humidity  and background gas may result in a  constant level being detected on the instrument  To reset to Zero  hold  the instrument away from the source of contamination and then press the  ZERO key  The instrument s display will return to a near Zero reading     The following things will affect the instrument s reading     CO  and moisture from the breath of the instrument user  e Barometric air pressure and background temperature        Sources of cold and heat    4 3 PROBE OPTIONS    In some applications the gray probe cover may restrict access to the area  under test  The gray probe cover can be removed by pulling it off the  instrument  When the gray cover is removed  great care should be taken  when using the in
19. s filled vessel or pipe     ppm Parts Per Million is a concentration reading  The GasCheck G  will display the concentration being detected  However  it is  more difficult to gauge the quantity of leakage    mg m3 Milligrams per meter cubed is also a unit that measures  concentration   See ppm above     9 yr Grams per Year is an alternative measure leak rate        3  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    33 IMPORTANT NOTES    The GasCheck G Series range is NOT intrinsically safe  so it should not be  used in potentially explosive environment  Intrinsically safe GasCheck  instruments are available  please contact LACO Technologies for more  information     e Ambient air pressure  heat  and humidity can also affect readings     e The GasCheck G range is NOT  gas specific     i e   it cannot differentiate  between gases     34 PACKING LIST    Please remove all packing material and then check the content of the  carry case against the list below before use  Should the instrument or any  accessory appear damaged or missing  contact the instrument supplier for  advise before use     Table 4  Packing List    ITEM QTY DESCRIPTION CHECK    GasCheck G instrument w short probe and nozzle fitted  Box wrench to change probe WI    INL RN  Calibration certificate s   CD containing additional product information Was       Spare battery clip NEN    4     2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES       35 THERMAL CONDUCTIVITY    A
20. strument  Note the following precautions         Do not remove the semi transparent probe sleeve  This sleeve ensures  that the metal inner probe remains 1 mm from the surface of the test  area  avoiding accidental dirt and moisture ingress       Do not touch the probe  especially the brass sensor housing  Heat from  fingers can result in significant changes in signal causing false readings     e Avoid placing the probe on wet or dirty surfaces  Blockage of the probe  will result in instrument failure     Figure 2  Changing Probes    joints and seams to be tested  The GasCheck is also supplied with a 300mm      Some applications may require a longer probe to gain access to pneumatic  probe that can be changed by the user      lt     To change probes  carefully do the following     1  Switch the instrument to OFF     2  Remove the outer gray probe cover  see 1      3  Place the box wrench  supplied with the instrument  over the existing  ea    probe so the spanner fits over the brass nut  see 2   and unscrew the     gt  probe      a counterclockwise direction  see 3    4  Withdraw the wrench and probe assembly  see 4      To refit the short or the long probe  follow the steps above but in reverse  order           CAUTION When using the box wrench to tighten the probe nut    lt     ensure the nut is firmly tightened  Do not use additional  41    tools  they may damage the tread     14  O 2010 LACO TECHNOLOGIES  INC     Figure 3  Detecting Leaks        Leaking  vessel    LACO US
21. the batteries  can be removed or replaced  Care must be taken when refitting the  batteries to ensure that they are the correct polarity  Polarity markings are  molded onto the battery holder plastic     The battery holder can be reconnected to the GasCheck by firmly pressing  the two halves of the PP3 clip together  Care should be taken to ensure the  polarity is correct  See the PP3 diagram  Figure 4         The battery pack is then laid within the instrument and the battery cover  refitted  Take care not to trap wires when refitting the battery cover        Battery holder       18  O 2010 LACO TECHNOLOGIES  INC     
22. the enter key  The instrument will automatically  sequence through the cold and hot stages of calibration  When  finished  it gives a summary of values     5  Pressing the ENTER key will return back to the calibration option  screen  Press the ESC key to start using the instrument or press the  ENTER key to recalibrate again     NOTE The GasCheck can detect changes humidity and carbon  dioxide  Avoid breathing on the probe     17  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    WARNING Great care should be taken to ensure the calibration gas  is neither pressurized or restricts flow to the instrument s  probe  Changes in flow of gas to the sensor will result in  significant errors in readings     5 2 REPLACING BATTERIES    Figure 4  Replacing Batteries    Before using the GasCheck G instrument  ensure the batteries are fitted  correctly or ensure NiMH  rechargeable  batteries are fully charged  before use     To fit or replace batteries you must first remove the battery cover from the  rear of the instrument  This can be done by prying open the battery clip  with a fingernail or screwdriver        Removing the battery cover will reveal a battery hold which requires 4 x  AA size batteries  Before removing old batteries or fitting new ones  unclip  the battery holder from the instrument by prying apart the PP3 type clip as  shown in the illustration        Battery holder       PP3 clip       Once the battery holder is disconnected from the instrument  
23. thin white tube that covers the inner probe should not be removed   This tube ensures a 1mm gap is maintained between the probe and the  surface being tested     15  O 2010 LACO TECHNOLOGIES  INC     5  MAINTENANCE      z  o                               E  o  O  u   o        AM  SPAN  set in    Custom Cal     ZERO  taken at    Switch On        Detectected Gas       baal Factory Cal      Custom Cal            1 5000    zeroing           m    cola    12345    Ema LL MR     hot 12345   p           l          LACO USER MANUAL   GASCHECK G SERIES    5 1 CALIBRATION    The GasCheck G Series has two selectable calibration settings  factory  and custom  Factory calibration is carried out at lon Science during the  instrument s manufacture  It can be selected and used but cannot be  changed     The GasCheck is calibrated against a 5000 ppm helium concentration and a  0 0005cc sec  5 E 4  helium leak rate     Custom calibration gives the user the ability to calibrate the instrument   This setting may offer improved accuracy over the factory calibration if  the barometric air pressure differs from that stated on the instrument s  calibration certificate     5 1 1 SELECTING FACTORY OR CUSTOM CALIBRATION    To select either FACTORY CAL or CUSTOM CAL  select the calibration screen  and move the up use the up and down keys to move the  tick  under the  desired symbol  then press the ESC key to exit to the main menu     Custom calibration can only be selected and used after the instrument h
24. urns the user  back to the RUNNING SCREEN     Figure 1  Running Screen          100 of  NER  DODOS              X     E    Running screen       NOTE The amount of functionality varies between models  See  Table 1  GasCheck Specificationsto find which functions  relate to which model     a        Vx    f Contrast  Page 10       Upgrade  Page 12    123 h Backlight  Page 10           N    x Ti  Sound  Page 10              Gas  Page 12         Main Menu          ppm x      Peak         Page 10     View Data  Page 11            Units  Page 11             E Calibration  Page 11        Battery  Page 11     7  O 2010 LACO TECHNOLOGIES  INC     LACO USER MANUAL   GASCHECK G SERIES    3 7 RUNNING SCREEN    The GasCheck displays the RUNNING SCREEN whenever the instrument  is being used to detect leaking gas  The illustration below outlines the  various information and icons     Sensitivity levels Battery status Backlight indicator           3 EH Peak hold indicator    Table 5  Running Screen Display          FUNCTION DESCRIPTION    GAS SELECTION The GasCheck G2 and G3 display the gas being measured in  the top left of the RUNNING SCREEN  The GasCheck G2 will  display the gas selected at the factory only  The G3 model  has a range of gases that can be selected via an internal gas  table     SENSITIVITY The GasCheck G has three sensitivity levels     e X100  times one hundred      the most sensitive  e X10  times ten      mid range  e X1  times one      least sensitive    BATTERY STATUS  
    