System On e H eated H u m i d i fi er I USER MANUAL

image

Contents

1. System One Heated Humidifier USER MANUAL PHILIPS hes PIRONICS 2011 Koninklijke Philips Electronics N V All rights reserved Table of Contents Sen OC WSC aranea E AE E A EA 2 E AAN E E E EE N A asana eneen 2 Se 2 GOAL GUC AL ONS oaren rea nnn nent eenen vinden eea 3 SMOKE ensen te eenander tedoen erin ics E A ene vadussetuawons 3 SVE OV SIC Gare see costa cts cence nesonbrbntertn etten A E E E E ETEEN 3 Available System One Heated Humidifiers aaneen eenen eneen en senvenen venen venen ens 4 System Features and CONTENTS sos c scsstnrcsseredieosriaanawasortdeortegnraseeioredseaveieaseceracesseesenicwnc 4 Connecting the therapy DEVOS vs ccscaindesensuncasaasndacnaa Geasneaindesiontimeataasndadseteversesinieseetennnesads 5 Connecting The Optional Heated TUDING sccwsiveverets ce dmasot massa EE EEEREN EERE tanken 6 Disconnecting the Optional Heated TUDING zes serres eerente arten venus taredreiaieudosManwins 6 Disconnecting the Therapy DEMOS anemie meter 7 BEN OE 7 Cleaning Str UCHONs Water BAN antenne twinner eten ema 10 Cleaning Instructions Humidifier BASE tent taaacixiiaaarsahianiens benetton 10 Clea MMO Instructions Heated TUDIN eiri edentate honinds 11 Hospital and Institution Disinfection Water Tank ccccccsecsccsseeeeseeeeeeeeseeeeeeeeseeeeeesaeeeseeseneneees 11 SWIS oneness tenten 12 Traveling With EME SEEN voeren ademden ret 12 DE EET P EEE re ley eee eer enna E 12 HOW LO CONTACT Philips RESON ONICS vervaarrtc
2. You can also use the following address Respironics Inc 1001 Murry Ridge Lane Murrysville PA 15668 12 Troubleshooting The table below lists some of the problems you may experience with your humidifier and possible solutions Problem Nothing happens when you apply power to the humidifier High Leak The humidifier LED icon on the therapy device is flashing The humidifier is cracked or damaged The therapy device is operating but the humidifier s airflow is low or stopped Why it NappeneD Loss of AC DC power or the device isunplugged The tubing is not connected correctly and doesn t seal properly The therapy device is not seated correctly against the humidifier The humidifier tank is not properly seated in the humidifier There is no communication between the therapy device and the humidifier The therapy device is not seated properly against the humidifier Using the wrong power supply Heated tubing is not connected properly or is damaged The heater plate is not heating What to Do Verify that the humidifier and therapy device are properly plugged in Make sure the AC power cord is connected correctly to the power supply and the power supply cord is correctly plugged into the wall If the problem continues to occur contact your home care provider Return the humidifier therapy device and power supply to your provider to determine if the problem is with th
3. use Rinse thoroughly and allow to air dry Make sure the tubing is dry before use Inspect the heated tubing for damage or wear cracking crazing tears punctures etc Discard and replace if necessary Hospital and Institution Disinfection Water Tank CAUTION Only the hospital and institution cleaning and disinfection procedures listed in this manual are recommended by Respironics Use of other cleaning and disinfecting processes not specified by Respironics may affect the performance of the product Cleaning Prior to Disinfection Clean and disinfect the humidifier s water tank using the procedures below 1 Disassemble the tank by separating the tank lid and tank base 2 While soaking the tank pieces in mild liquid dish detergent use a soft bristle brush to clean each piece Pay close attention to all corners and crevices 3 Rinse each piece with water twice Be sure to agitate it vigorously in water when rinsing and allow to air dry but not in direct sunlight Disinfection The following processes can be used to disinfect the water tank for a maximum of 60 cycles Thermal Disinfection Immersion in a tap water bath at 75 C 2 C for 30 minutes Control Ill Cidex Cidex OPA Following disinfection visually inspect each piece Discard and replace any damaged parts NJ WwW 11 Service The humidifier does not require routine servicing If any part of the humidifier is worn or damaged contact Phili
4. C Tank Base Middle Seal Note If the lid does not close easily onto the base separate the two parts reassemble the tank and inspect it again CAUTION Do not fill the water tank above the maximum fill line Damage to the humidifier or therapy device may occur If the water tank is overfilled water may leak out of the tank inlet located on the back of the tank when installing the tank lid Slide the water tank back into the humidifier WARNING When installing the tank do not allow any water to spill into the humidifier or therapy device CAUTION Do not move the humidifier while the water tank has water in it Plug the power supply cord into the back of the therapy device Plug the AC power cord into the power supply provided with your therapy device and then plug the AC power cord into an electrical outlet that is not controlled by a wall switch IMPORTANT You must use the 80W power supply when using the heated tube humidifier with your therapy device The appropriate power supply can be identified by th eow symbol Attach the flexible tubing from the patient circuit to the humidifier s air outlet port If using the optional heated tubing Refer to Connecting the Optional Heated Tubing section of this manual IMPORTANT Before each use examine the flexible tubing for any kinks damage or debris If necessary clean the tubing to remove the debris Replace any damaged tubing 10 Turn on the airflow on your d
5. G The humidifier must always be positioned below the breathing circuit connection at the mask and the air outlet on the device The humidifier must be level for proper operation CAUTION Do not place the humidifier directly onto carpet fabric or other flammable materials CAUTION Do not place the device in or on any container that can collect or hold water Take precautions to protect furniture from water damage CAUTION Do not turn the humidifier on without the water tank installed The Humidifier setting must remain off if there is no water in the water tank CAUTION Do not attempt to fill the tank while it is still inside the humidifier WARNING Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removingthe water tank Aburn may result from touchingthe heater plate comingincontact withthe heated water ortouchingthetank pan Lift up on the release lever to open the humidifier door until it locks in an open position You will hear a click once the door is opened far enough to remain in an open position Remove the water tank by grasping the front of the tank and sliding it out of the humidifier base CAUTION The humidifier door must be set inthe open position before removingthe water tank Do not remove the water tank without making sure that the humidifier door locks into the open position Cover Release Lever Grasp the Water Tank on top and bottom 3 Press the tab in t
6. Humidification Systems Electrical When the heated humidifier is used with a Philips Respironics therapy device AC Power Consumption with 60W power supply 100 240 VAC 50 60 Hz 2 1A AC Power Consumption with 80W power supply 100 240 VAC 50 60 Hz 2 0A DC Power Consumption 12 VDC 6 67A Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Shock Type BF Applied Part Degree of Protection against Ingress of Water Drip Proof IP22 Mode of Operation Continuous Electromagnetic Compatibility The device meets the requirements of EN 60601 1 2 2nd edition Heater Plate Max Temperature 75 C 167 F Pressure Drop with Humidifier Max 0 3 cm HO at 60 LPM flow Humidity Humidity Output 10 mg H_O L Measured max flow 35 C 15 RH 15 Heated Tubing Specifications 16 Maximum Recommended Pressure 25 cm HO Inner Diameter 15 mm 0 6 in Length 1 83m 6ft Heated TubingTemperature Range 16 to 322 C 60 to 89 F Heated Tubing Temperature Cut out lt 41 C lt 106 F Material Flexible plastic and electrical components Electrical Heated tubing is powered by the attached heated humidifier Refer to Electrical section of System One Heated Humidifier Specifications Environmental Refer to Environmental section of System One Heated Humidifier Specifications Limited Warranty Respironics Inc warrants that the system shall be free from defects of
7. d Contents IMPORTANT Read and understand the entire manual for your therapy device before attempting to use this humidifier Air Outlet Port Water Tank Flexible Heated Tubing optional accessory Humidifier Door Air Inlet Port Humidifier Heater Plate Maximum Fill Lines connects to Therapy Device Humidifier Connector End This figure illustrates many of the device features and contents described in the table below Description Connect your therapy device here Connects to the outlet port on the therapy device Air Outlet Port conical 22 mm Connect the patient circuit here If you are using the optional heated tubing this outlet port Standard Port shown here will look slightly different Refer to Connecting the Optional Heated Tubing section of this manual Heater Plate Warms the water in the water tank Humidifier Door Open the door to access the water tank Water Tank The removable water tank holds the water for humidification Maximum Fill Lines The fill lines indicates the maximum water level for safe operation Found on front and both sides of tank Flexible Heated Tubing The optional heated tube connects from the humidifier to the patient s mask optional Humidifier Connector End Connect this end of the tubing to the humidifier Note This humidifier is specified for use with Philips Respironics REMstar and BiPAP therapy devices only Humidifier Door Lever Lift up on the lever to open
8. e therapy device humidifier or power supply Remove your mask and tubing and check for kinks or tears If it is torn or damaged contact your home care provider or Philips Respironics for replacement tubing and or mask If the tubing is not damaged reattach your tubing turn on the airflow and check to make sure you do not still feel air coming out of the port area Remove the therapy device from the humidifier and reconnect Make sure the air inlet port on the humidifier connects securely to the air outlet port on the device If high leak persists issue may be caused by a misaligned or damaged seal Seals are not user serviceable components Please contact your home care provider or Philips Respironics Remove the water tank from the humidifier base and then slide the tank back in the humidifier making sure it is pushed back as far as it can go If high leak persists issue may be caused by a misaligned or damaged seal Seals are not user serviceable components Please contact your home care provider or Philips Respironics Following the Daily Use section earlier in this manual check the outlet power and verify that the humidifier and therapy device are properly plugged in Make sure the AC power cord is connected correctly to the therapy device s power supply Make sure the AC power cord is correctly plugged into the wall If the problem continues to occur contact your home care provider Return the humidifier thera
9. erne tred eneen vvd erate 12 MOUNE ROONG eenen nanskate 13 System One Heated Humidifier Specifications eu eas en sneevenenvensneevensn eveneens ens ID Heated TUDING Se CIICAN ONS ririri ara E Ra 16 AE MR A A A ET E T E E E E T O Back Page CAUTION U S federal law restricts this device to sale by or on the order of a physician Intended Use The System One Heated Humidifier is an accessory for the Philips Respironics REMstar and BiPAP therapy devices to provide moisture to the patient circuit It is intended for use in spontaneously breathing patients weighing over 30 kg 66 Ibs in the home or hospital institutional environment who use mask applied positive pressure ventilation therapy Warnings A warning indicates the possibility for injury to the user or the operator Use the humidifier only for its intended use as described in this manual Use only with Philips Respironics full face masks nasal masks and connectors Periodically inspect the humidifier for signs of wear or damage Never operate the humidifier if any parts are damaged if it is not working properly or if the humidifier has been dropped or mishandled Do not use the humidifier if the water tank is leaking or damaged in any way Have any damaged parts replaced before continuing use Periodically inspect the humidifier power cord for signs of wear or damage If it becomes worn or damaged contact Philips Respironics or your home care provider for a replacement T
10. eu of all other express warranties In addition any implied warranties including any warranty of merchantability or fitness for the particular purpose are limited to two years Some states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from state to state To exercise your rights under this warranty contact your local authorized Respironics Inc dealer or contact Respironics Inc at 1001 Murry Ridge Lane Murrysville Pennsylvania 15668 8550 1 724 387 4000 PHILIPS REF 1095653 Respironics Inc 1001 Murry Ridge Lane 1095419 R01 Murrysville PA 15668 USA JR 12 16 2011 En EN DOM
11. evice and begin therapy Note For complete instructions on how to turn on the airflow refer to the manual included with your therapy device 11 Put on your mask assembly when the air starts to flow Note If you are having trouble with your mask refer to the instructions included with the mask 12 Refer to your therapy device manual for complete instructions on how to adjust both the Heated Humidifier and the optional Heated Tubing settings to achieve the desired humidity CAUTION Do not turn the humidifier on without the water tank installed The Humidifier setting must remain off if there is no water in the water tank 13 Refer to the manual included with your therapy device for instructions to turn off therapy Cleaning Instructions Water Tank Hand washing can be performed daily Dishwashing can be performed once a week 1 J N D gt UTI or Turnthe humidifier settingoff turnthe therapy device off and allow the heater plate and water to cool WARNING Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removingthe water tank Aburn may result from touchingthe heater plate comingincontact withthe heated water or touching the tank pan Open the humidifier door with the release lever and then slide the water tank out of the humidifier base Press the tab in the hole on top of the tank in toward the front of the tank Gently remove the tank lid from the ta
12. he hole on top of the tank in toward the front of the tank Gently remove the tank lid from the tank base and set aside Empty any remaining water from the base of the tank Tab Tank Lid Tank Base Middle Seal 4 Rinse the tank with water With the water tank sitting on a flat surface fill it with distilled water approximately 325 ml no higher than the maximum fill line located on the front and both sides of the tank Tank Base Tank Lid Maximum Water Fill Lines Note Clean the water tank before first use Refer to the Home Cleaning section in this manual CAUTION Use only room temperature distilled water in the tank Do not put any chemicals or additives into the water Possible airway Irritation or damage to the water tank may result CAUTION Do not fill the water tank above the maximum fill line Damage to the humidifier or therapy device may occur If the water tank is overfilled water may leak out of the tank inlet located on the back of the tank when installing the tank lid UI or co l ite Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base Close the lid until the tab on the lid snaps back under the lip in the tank base Inspect the tank When it is closed correctly the lid should be seated completely on the middle seal and it should sit snugly on the tank base so the tab can easily snap back in place _____ Hinges Tab Tank Lid E
13. he humidifier must always be positioned below the breathing circuit connection at the mask and the air outlet on the device The humidifier must be level for proper operation Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank A burn may result from touching the heater plate coming in contact with the heated water or touching the tank pan This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide When installing the water tank do not allow any water to spill into the humidifier or therapy device If you notice any unexplained changes in the performance of this device if it is making unusual or harsh sounds if it has been dropped or mishandled if water is spilled into the enclosure or if the enclosure is broken disconnect the power cord from the therapy device and discontinue use Contact your home care provider Before cleaning the humidifier always remove from the therapy device Empty and clean the water tank daily to prevent mold and bacteria growth Wipe the seal completely Note Please see the Limited Warranty section of this manual for information on warranty coverage Cautions A caution indicates the possibility of damage to the device Do not place the humidifier directly onto carpet fabric or other flammable materials Do not place the device in or on any container tha
14. he tubing into place over the air outlet port until the tabs on the side of the tube click into place in the slots on the side of the outlet port Heated Tubing r 3 Pin Connector Ps g Nf AS ii My i Heated Tubing SD aa Air Outlet Port Slot SEI W Note Non heated Philips Respironics tubing 15 or 22 mm can sill be used with this humidifier with the heated tubing air outlet port Simply connect the tubing to the air outlet port Refer to the therapy device manual for setup instructions Disconnecting the Optional Heated Tubing 1 To remove the heated tubing press in the tabs on the side of the tubing connector and pull the tubing away from the outlet port Heated Tubing Heated Tubing Air Outlet Port Disconnecting the Therapy Device CAUTION To avoid spilling do not disconnect the humidifier from the therapy device with water in the tank Remove the water tank from the humidifier before disconnecting the therapy device 1 Grasp the release latch on the bottom of the humidifier 2 While holding both the therapy device and the humidifier pull the release latch and pull the two units apart 3 Humidifier bottom view Release Latch Release Latch If you will be using the therapy device without the humidifier reattach the side cover to the therapy device Daily Use i NJ Place the connected therapy device with humidifier on a firm flat surface lower than your sleeping position WARNIN
15. nk base Empty any remaining water from the base of the tank Wash the parts of the tank in the dishwasher top shelf only or in a solution of warm water and a mild liquid dishwashing detergent Gently wash the middle seal Rinse the parts with clean water Wipe the parts completely on the top and bottom Allow them to air dry CAUTION Use a mild liquid dishwashing detergent only for either hand washing or when using a dishwasher WARNING Empty and clean the water tank daily to prevent mold and bacteria growth Wipe the seal completely Inspect the tank and seal for damage Note Never use the water tank if the tank lid does not fit comfortably on the tank base Before using the tank fill it with distilled water per the directions in the Daily Use section no higher than the maximum fill line located on the front and sides of the tank Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base Close the lid until the tab on the lid snaps back under the lip in the tank base Inspect the tank When it is closed correctly the lid should be seated completely on the middle seal and it should sit snugly on the tank base so the tab can easily snap back in place Inspect the water tank for any leaks or damage If the water tank shows signs of wear or damage contact your home care provider for a replacement Note If the lid does not close easily onto the base separate the two parts
16. ps Respironics or your home care provider See the Troubleshooting section in this manual for additional information Traveling with the System Packing the System 1 Remove the water tank and empty all water 2 Put the empty water tank back into the device 3 Pack your humidifier in your carry on luggage CAUTION Do not move the humidifier while the water tank has water init When you are traveling the optional carrying case for your therapy device can be used for carry on luggage only The carrying case will not protect the humidifier if it is put through checked baggage For your convenience at airport security stations there is a note on the bottom of the humidifier stating that it is medical equipment It may be helpful to bring this manual along with you to help security personnel understand the device If you are traveling to a country with a line voltage different than the one you are currently using with the therapy device a different power cord or an international plug adapter may be required to make your power cord compatible with those where you visit Contact your home care provider for additional information Disposal Dispose of this device in accordance with local regulations How to Contact Philips Respironics To have your device serviced contact your home care provider If you need to contact Philips Respironics directly call the Philips Respironics Customer Service department at 1 800 345 6443 or 1 724 387 4000
17. py device and power supply to your provider to determine if the problem is with the therapy device humidifier or power supply Remove the therapy device from the humidifier and reconnect it Make sure the air inlet port on the side of the humidifier connects securely to the air outlet port on the side of the device If you are using the optional heated tubing make sure you use the correct 80W power supply Remove the heated tubing and check for damage If it is not damaged reconnect the heated tubing If the problem continues contact your home care provider Check the power connections to the humidifier If the problem continues contact your home care provider or Philips Respironics Note If the water is not heating you can temporarily use the humidifier as an unheated pass over humidifier The humidifier was If the humidifier does not operate properly after being dropped or dropped or mishandled mishandled contact your home care provider or Philips Respironics The humidifier has an Contact your home care provider Return the humidifier and power airflow obstruction supply to your provider to determine the problem 13 Problem Why it happeneD What to Do The water tank s middle seal The seal may become Check the seal for any tears or other damage If it is damaged is damaged damaged during use contact your home care provider Excessive condensation in The humidity level setting Reduce the humidity level setting the
18. reassemble the tank and inspect it again Cleaning Instructions Humidifier Base a LJ N 10 WARNING Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removingthe water tank Aburn may result from touchingthe heater plate comingin contact withthe heated water ortouchingthetank pan WARNING Before cleaning the humidifier always remove from the therapy device Clean the humidifier base and heater plate by wiping it with a damp cloth Allow the platform to air dry before reconnecting to the therapy device Inspect the humidifier base for any damage and replace it if necessary Clean the humidifier outlet port by using a damp bottle brush or a damp cloth Insert the brush or cloth approximately 7 cm 2 75 inches into the outlet opening while cleaning Cleaning Instructions Heated Tubing Clean the heated tubing before first use and weekly Heated Tubing is single patient multi use 1 Disconnect the heated tubing from the heated humidifier Refer to the Disconnecting the Optional Heated Tubing section earlier in this manual Gently wash the heated tubing in a solution of warm water and a mild detergent CAUTION Avoid submerging the Humidifier Connector End of the heated tubing in water Note If the Humidifier Connector End of the tubing does get wet during the cleaning process be sure to thoroughly rinse all soap residue from the connector and air dry before the next
19. spital and institution cleaning procedures listed in this manual are recommended by Philips Respironics Use of other cleaning and disinfecting processes not specified by Philips Respironics may affect the performance of the product Contraindications Studies have shown that the following pre existing conditions may contraindicate the use of CPAP therapy for some patients Bullous Lung Disease Pathologically Low Blood Pressure Bypassed Upper Airway Pneumothorax Pneumocephalus has been reported in a patient using nasal Continuous Positive Airway Pressure Caution should be used when prescribing CPAP for susceptible patients such as those with cerebral spinal fluid CSF leaks abnormalities of the cribriform plate prior history of head trauma and or pneumocephalus Chest 1989 96 1425 1426 The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of a sinus or middle ear infection Not for use with patients whose upper airways are bypassed Contact your health care professional if you have any questions concerning your therapy Refer to the instructions for use that accompanied your therapy device for any additional contraindications that may be specific to the use of that device Symbol Key The following symbols may appear on the device Symbol Definition Symbol Definition Maximum Fill Line IP22 Drip Proof Equipment WI Hot Water Hot Surface Type BF Applied Part Sys
20. st be level for proper operation ae caused the device Also place the humidifier away from the edge of the night stand or table so it doesn t accidentally get knocked off the table If the humidifier falls and water gets into the therapy device drain all water out of the therapy device Allow it to air dry to make sure it is completely dry before reapplying power If the placement of the tubing causes the humidifier to fall make sure that you use proper hose management when setting up your device Route the tubing behind the bed s headboard If the humidifier does not operate correctly after falling contact your home care provider or Philips Respironics Note For information on troubleshooting your therapy device see the manual included with your therapy device 14 System One Heated Humidifier Specifications Environmental Operating Temperature 5 to 35 C 41 to 95 F Storage Temperature 20 to 60 C 4 to 140 F Relative Humidity operating amp storage 15 to 95 non condensing Atmospheric Pressure 77 to 101 kPa 0 2286 m 0 7500 ft Physical Dimensions 18 x 14x 10 cm 7 Lx 5 5 W x 4 H Weight Approximately 0 89 kg 1 95 Ibs Water Capacity 325 ml 11 oz at recommended water level Standards Compliance This device is designed to conform to the following standards IEC 60601 1 General Requirements for Safety of Medical Electrical Equipment EN ISO 8185 General Requirements for
21. t can collect or hold water Take precautions to protect furniture from water damage Do not fill the water tank above the maximum fill line Damage to the humidifier or therapy device may occur If the water tank is overfilled water may leak out of the tank inlet located on the back of the tank when installing the tank lid Use only room temperature distilled water in the tank Do not put any chemicals or additives into the water Possible airway irritation or damage to the water tank may result Remove the tank empty all water and replace the empty tank before transporting the humidifier base Do not attempt to fill the water tank while it is still inside the humidifier To avoid spilling do not disconnect the humidifier from the therapy device with water in the tank Remove the water tank from the humidifier before disconnecting the therapy device Do not turn the humidifier on without the water tank installed The Humidifier setting must remain off if there is no water in the water tank The humidifier door must be set in the open position before removing the water tank Do not remove the water tank without making sure that the humidifier door locks into the open position Do not move the humidifier while the water tank has water in it Use a mild liquid dishwashing detergent only for either hand washing or when using a dishwasher Avoid submerging the Humidifier Connector End of the heated tubing in water Only the ho
22. tem Overview The System One Heated Humidifier attaches to the therapy device and provides an air outlet port to connect a breathing circuit The breathing circuit is comprised of patient tubing a mask and in some instances a separate exhalation device The patient tubing can be Respironics heated tubing 22 mm non heated performance tubing or 15 mm non heated performance tubing For information related to the mask to be used including any need for a separate exhalation device consult the instructions for use that accompany both the mask and therapy device The System One Heated Humidifier with Heated Tubing is designed to deliver humidification to provide added comfort during therapy This humidification level is controlled through the output of the heated humidifier as well as the temperature of the optional heated tubing Use of these two accessories allows for a comfortable level of humidity to be maintained at the mask The System One Heated Humidifier is comprised of the following components Heated Humidifier The heated humidifier is the primary source of humidification Humidification is controlled by adjusting the temperature of the heater plate The heater plate is then used to heat water found in the water tank This manual includes instructions that explain how to set up and take of the heated humidifier For instructions on how to adjust the output of the heated humidifier refer to the instructions for use that accompanied the
23. the humidifier door Connecting the Therapy Device 1 When using a humidifier the patient circuit attaches to the air outlet port on the humidifier 2 To connect the therapy device to the humidifier first you must remove the side cover on the therapy device Press in the tab on the side cover and pull the cover away from the unit Note Therapy device shown is an example Actual therapy device may vary Therapy Device Side Cover _ 3 Next line up the components side by side Make sure that the guides on the humidifier fit into the slots on the therapy device and the air outlet port on the therapy device fits into the air inlet port on the humidifier 7 Humidifier Air Inlet Port aA Air Outlet Port Guides Slots 4 While holding both the therapy device and the humidifier pull the release latch on the bottom of the humidifier and press the two units together until they snap into place Let go of the release latch Make sure that the therapy device and the humidifier are completely seated against each other Humidifier bottom view Release Latch Release Latch Connecting the Optional Heated Tubing 1 To attach the optional heated tubing to the humidifier line up the humidifier connector end of the tubing with the Heated Tubing compatible air outlet port on the humidifier Make sure the 3 pin connector at the top of the tube lines up with the opening at the top of the air outlet port Press t
24. therapy device Water Tank The water tank stores the water that will be used by the heated humidifier This manual includes instructions that cover how to fill up and take care of the water tank Heated Tubing The heated tubing is an optional accessory that is used along with the heated humidifier to control the provided humidification This is accomplished by controlling the temperature of the air in order to ensure that it does not cool down prior to reaching the mask This manual includes instructions that cover how to connect and take care of the heated tubing For instructions on how to adjust the temperature of the heated tubing refer to the instructions for use that accompanied the therapy device Available System One Heated Humidifiers The System One Heated Humidifier is available in the following configurations Heated Humidifier This humidifier provides moisture to the patient circuit It has a standard air outlet port for use with Philips Respironics 15 and 22mm tubing Heated Humidifier with Heated Tubing This humidifier provides moisture to the patient circuit but also allows for the option of heated tubing In order to function with the heated tubing it has a modified air outlet port that includes both a 3 pin connector and a heated tubing locking mechanism The heated tubing compatible System One Heated Humidifier can be seen in the Connecting the Optional Heated Tubing section of this manual System Features an
25. tubing is too high The humidifier is Verify that the humidifier and therapy device are away from air positioned incorrectly conditioning equipment Heated tube isnot warming Using the wrong power Make sure you use the correct 80W power supply The correct supply power supply should have the fow symbol Heated Tubing setting is set Make sure the Heated Tubing setting is not set to zero Refer to your to zero therapy device s user manual for instructions The humidifier door has Humidifier door was lifted The humidifier door can be easily reattached been accidentally removed past the locked open With the water tank ramoved all are A align the hinge side of the door with from the humidifier position the back of the humidifier in a fully open position see figure below Make sure the hinges are aligned with the grooves of the humidifier Rotate the door closed You will need to apply pressure until the door snaps past the locked open position Verify that the door is working properly If further assistance is needed contact your home care provider The humidifier hasfallen off The humidifier may not Always make sure your humidifier is placed on a hard flat surface your table or night stand have been properly seated so the rubber feet on the bottom of the humidifier base can adhere on the night stand or the to the surface make sure there is no fabric under the base The placement of the tubing humidifier mu
26. workmanship and materials and will perform in accordance with the product specifications for a period of two 2 years from the date of sale by Respironics Inc to the dealer If the product fails to perform in accordance with the product specifications Respironics Inc will repair or replace at its option the defective material or part Respironics Inc will pay customary freight charges from Respironics Inc to the dealer location only This warranty does not cover damage caused by accident misuse abuse alteration water ingress and other defects not related to material or workmanship The Respironics Inc Service department shall examine any devices returned for service and Respironics Inc reserves the right to charge an evaluation fee for any returned device as to which no problem is found after investigation by Respironics Inc Service This warranty is non transferable by unauthorized distributors of Respironics Inc products and reserves the right to charge dealers for warranty service of failed product not purchased directly from Respironics Inc or authorized distributors Respironics Inc disclaims all liability for economic loss loss of profits overhead or consequential damages which may be claimed to arise from any sale or use of this product Some states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation or exclusion may not apply to you This warranty is given in li

Download Pdf Manuals

image

Related Search

Related Contents

Franke Consumer Products SM 52M Convection Oven User Manual    Télécharger le mode d`emploi  Release Notes for Cisco ONS 15200 SNM Software Release 1.1(2)  Mode d`emploi  Référentiel structures et organisation des unités de  THÉÂTRE PROXIMITY  Epson G5000 Quick Reference Guide  Ideal Desk-side shredder 2501-Cross/Cut  Panasonic AV-HS04M1  