YSI IQ SensorNet MIQ C6 Module User Manual

image

Contents

1. SENSORNET 2 2 9 a HOIVNIWH3 L NS z SENSORNET 1 Current outputs Fig 3 1 Terminal strip with the current connections 3 Screw the cable gland pos 1 in Fig 3 1 with the sealing ring pos 2 into the module housing 4 Loosen the coupling ring pos 3 in Fig 3 1 5 Feed the line through the cable gland in the module housing 6 Connect the wires to the terminal strip While doing so pay attention to the specifications on the label located under the terminal strip 7 Tighten the coupling ring pos 3 in Fig 3 1 8 Close the module ba76033e01 01 2012 MIQ C6 Settings 4 Settings The MIQ C6 module has six current outputs that supply a current that depends on the measured value On the terminal you can assign names to the outputs see section 4 1 link outputs with sensors see section 4 2 delete links of outputs with sensors see section 4 3 adjust outputs see section 4 4 check the condition of the outputs see section 4 5 Note The general operating principles are given in the system operating manual or in the component operating manual of the terminal components ba76033e01 01 2012 Settings MIQ C6 4 1 Entering editing the name of an output For easier identification of the outputs an individual name can be given to each output in the Edit list of outputs overview 1
2. MIQ C6 Installation General installation instructions Materials required Tools Observe the following points when attaching connecting wires to the terminal strip Shorten all wires to be used to the length required for the installation Always fit all the ends of the wires with wire end sleeves before connecting them to the terminal strip Any wires that are not used and project into the enclosure must be cut off as closely as possible to the cable gland Screw a small cable gland with sealing ring into each remaining free opening and close it with a blind plug 3 4 Connections to the current outputs Wire end sleeves suitable for the connecting wires with suitable crimping tool 4 x cable gland with sealing ring scope of delivery MIQ C6 Cable stripping knife Wire stripper Phillips screw driver Small screw driver ba76033e01 01 2012 Installation MIQ C6 Connecting lines to the terminal strip 1 Open the module 2 Open the dummy screw fitting under the required connection Keep the dummy screw fitting for possible later modifications xis X17 0 4 20mA REC6 X16 X15 0 4 20mA REC5 14 X13 X12 11 0 4 20 0 4 20mA REC4 X10 X9 0 4 20 REC2 x x7 0 4 20mA REC1 Q3lHS
3. Operating Manual Ring Binder IQ Sensor MIQ Module MIQ Terminal Operating Operating Operating Manual Manual Manual Component Operating Manuals Fig 1 1 Structure of the IQ SENSOR NET operating manual The IQ SENSOR NET operating manual has a modular structure like the IQ SENSOR NET itself It consists of a system operating manual and the operating manuals of all the components used Please file this component operating manual into the ring binder of the system operating manual ba76033e01 01 2012 Overview MIQ C6 General characteristics Terminal strip 1 2 Features of the MIQ C6 The MIQ C6 output module has six current outputs They can be programmed as Analog output recorder PID regulator You can link current outputs to sensors The linked current outputs can be used to monitor sensors or for control purposes With the standard MIQ module housing the MIQ C6 has the same characteristics as all MIQ modules regarding stability leakproofness and weather resistance It also provides the same wide variety of installation options stacked mounting canopy mounting tophat rail mounting etc The MIG C6 has the following electrical connections on the terminal strip inside the housing 6 x current output 2x SENSORNET connection ba76033e01 01 2012 MIQ C6 Safet
4. Open the Settings menu with 9 2 Select and confirm the menu item System settings gt Edit list of outputs with 99 The Edit list of outputs display opens 3 Highlight a name in the Name column with and confirm with Terminal 1 15 Hay 004 14 53 AA D Edit List of outputs 130 No Hodel Channel Ser no Hame DOWN TB ACB ACT 12345678 z Belsebung DOWN GYChYCe 12345678 O2 Belebung DD1 H TBA CE C3 12345678 NO3 Beleb DOWHIGYCh 204 12345678 Do1 HIg C5 405 12345678 DD1 H TBACE CE 12345678 ceeeeeece G Tr bung t02 HIG4CR3 RZ eeeeeece GE Tr bung Da3 HI a BBBBBBBE SchLammpgl DOS HIGACRS4Ce Tr b Abl D a3 HI GCR R1 Tr b Zul Select character confirm character Fig 4 1 130 Edit list of outputs 4 Select a letter a numeral or a special character with and confirm with 5 Complete the of the output and confirm with 4 2 ba76033e01 01 2012 MIQ C6 Settings 4 2 Linking the output with a sensor 1 Open the Settings menu with 5 2 Select and confirm the menu item System settings gt Settings of outputs and links with and 99 The Settings of outputs and links display opens 3 Highlight the column with and confirm with 69 4 Highlight an output with and confirm with The Link with display opens The display shows a list of
5. current output The output of measured values is laid down in the Recorder type Start value and End value settings Setting Selection Values Explanation Recorder type 0 to 20 mA or 4 to 20 mA Start value within the Minimum spacing measuring range 5 of the measuring End value sensor range dependent Measured variable Main variable Adjoining variable Main variable designates the actual measured variable of the sensor e g pH oxygen etc Adjoining variable designates an additional measured variable e g temperature Attenuation 0 40 mA s Speed of change of the output current mA s in the case of an erratic change of the input signal gt OFL UFL Error Current values outside the range between Start value and End value are considerd as errors The current output reacts as specified under Behavior at error see below Limitation The current at the output is limited to the Start value or End value ba76033e01 01 2012 MIQ C6 Settings Behavior at error Fixed current value In the case of system and sensor errors the current output supplies the specified Current with error Possible values 0 21 mA Details on behavior of outputs see section 4 6 1 Unchanged The current at the time of the occurrence of the error remains unchanged Details on behavior of outputs see section 4 6 1 ba76033e01 0
6. outputs Output Physically separated from the IQ SENSOR NET Output current Can be switched between 0 20 mA and 4 20mA In the case of errors can be set to 0 21 mA Max initial output voltage 15 V in the case of missing or incorrect burden Max load 500 Q Accuracy 0 3 50 uA Functions Programmable as Analog output recorder PID regulator Output with fixed current value for test purposes ba76033e01 01 2012 MIG C6 Contact Information _ Contact Information 7 1 Ordering amp Technical Support Telephone 800 897 4151 937 767 7241 Monday through Friday 8 00 AM to 5 00 PM ET Fax 937 767 1058 Email environmental ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Internet www ysi com When placing an order please have the following information available YSI account number if available Name and Phone Number Model number or brief description Billing and shipping address Quantity Purchase Order or Credit Card 7 2 Service Information YSI has authorized service centers throughout the United States and Internationally For the nearest service center information please visit www ysi com and click Support or contact YSI Technical Support directly at 800 897 4151 When returning a product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service C
7. the sensors to which a link is possible Terminal 1 15 Hay 004 14 53 AA Link with Ha Sensor name Heasuring range 102 60 0 mg AL 502 99160001 0 14 Select sensor confirm 95 Fig 4 2 150 Settings of outputs and links Link with 5 Select a sensor with and confirm with 99 The output is linked with the selected sensor Note Outputs that are linked with sensors can be identified in the Ser no field of the Settings of outputs and links overview by the specification of the linked sensor ba76033e01 Settings MIQ C6 4 3 Deleting a link with an output If a link from a current output with a sensor is no longer required the link can be deleted 1 Open the Settings menu with 9 2 Select and confirm the menu item System settings gt Settings of outputs and links with 99 The Settings of outputs and links display opens 3 Highlight the column with and confirm with 69 4 Highlight a linked output with and confirm with Terminal 1 15 Hay 004 14 53 gg AA D Settings of outputs and Links 150 amp No Hame Chan Feature 501 501 Cancel Linked s 1 1341 00 02 0 0 60 0 mg L Select edit Link Fig 4 3 150 Settings of outputs and links Erase link 5 Select Erase link with and confirm with A security prompt appears 6 Select Erase l
8. ysi com Internet www ysi com
9. 1 2012 Settings MIQ C6 Function 4 4 2 PID controller With the PID controller function you can use an output as a controller output The controller can be configured as a Proportional controller with a switchable Integral and Derivative controller part PID controller The control mode of the PID controller is described by the following equation I Controller 1 dx L K x c xdt 4 T 0 with DN K X X X x e nom actual I lt I lt I Controller max IController Current at the controller output at time t lo Current at output if Xactual Xnominal K Gain Xp Proportional band Xe Error signal Xactual Actual value current measured value Xhon Nominal value ti Integral control part td Derivative control part t Time Imin Lower current limitation Imax Upper current limitation The adjustable control parameters are Xnominal lo Imin ti and Tg see setting table on page 4 11 ba76033e01 01 2012 MIQ C6 Settings Characteristic curve of the proportional By activating or deactivating the Integral ti and Derivative td controller part the following controller types can be configured Controller type td s ti s P controller 0 0 PI controller 0 1 to 9999 PD controller 1 to 9999 0 PID controller 1 to 9999 1 to 9999 For a P controller the interrelationship b
10. 4 4 Setting current outputs 4 5 44 Recorder ors DY E 4 6 4 4 2 PID controller 4 8 443 Fixed current value 4 13 4 5 Checking the status of the outputs 4 14 4 6 Behavior of linked outputs 4 15 4 6 1 Behavior in case 4 15 4 6 2 Behavior in the non operative condition 4 16 4 7 Maintenance condition of sensors 4 17 4 7 1 Switching on the maintenance condition 4 18 4 7 2 Switching off the maintenance condition 4 18 Maintenance and cleaning 5 1 5 1 5 1 52 Cleaning cese e gue Eta TATUR EID ENS 5 1 Technical data 6 1 Contact Information 7 1 7 1 Ordering amp Technical Support 7 1 ba76033e01 01 2012 List of contents MIQ C6 7 2 Service Information 7 1 8 aaa rh RR RA Ry Rs 8 1 8 1 Explanation of the messages 8 1 8 1 1 Error messages 8 1 8 1 2 Informative messages 8 1 0 2 ba76033e01 01 2012 MIQ C6 Overview 1 Overview 1 1 How to use this component operating manual Structure of the IQ SENSOR NET operating manual IG Sensor Net Operating Manual System
11. A If for example the measured concentration is 5 mg l 11 mA are output mA Hos gt 12 mAr 1 4 mg l 5 mg l mg l mA 12 MA 1 1 mg l 2 11 mA mg l The highest concentration with which the controller works within the proportional band is 8 mg l corresponding to min 8 mA the lowest one is 2 mg l corresponding to Imax 14 mA Settings Setting Selection Values Explanation Measured variable Main variable Main variable Adjoining variable designates the actual measured parameter of the sensor e g pH oxygen etc Adjoining variable designates an additional measured parameter e g temperature Nominal value within the measuring Nominal value the range sensor measured value is dependent regulated to Xp 5 100 Proportional band of 5 100 the controller of the measuring Negative values range result in a positive characteristic curve Imin 0 20 Lower current limitation Imax 0 20 Upper current limitation Note Spacing between Imin and Imax at least 5 mA ba76033e01 01 2012 4 11 Settings MIQ C6 Setting Selection Values Explanation lo 0 20 mA Current value on the output if the measured value equals the Nominal value ti 0 9999 s Hold back time Integral part of the controller 0 not effective td 0 9999 s Reset time Derivative part of the controller 0 not effective Behavior
12. Q C6 4 7 1 Switching on the maintenance condition 1 Call up the measured value display with M 2 Select the sensor you want to switch on the maintenance condition for with The measured values of the sensor do not flash 3 Open the Display Options menu with 99 4 Selectthe Switch maintenance condition on off menu item with and confirm with 99 A window that informs you about the maintenance condition opens up 5 Confirm Continue with 99 The selected sensor is in the maintenance condition Linked outputs are frozen 6 Call up the measured value display with The measured values of the sensor flash 4 7 2 Switching off the maintenance condition 1 Call up the measured value display with 2 Select the sensor you want to switch off the maintenance condition for with The measured values of the sensor flash 3 Open the Display Options menu with 99 4 Selectthe Switch maintenance condition on off menu item with and confirm with 99 A window that informs you about the maintenance condition opens up 5 Confirm Continue with 99 The maintenance condition of the selected sensor is switched off Linked outputs are released 6 Call up the measured value display with The measured values of the sensor do not flash ba76033e01 01 2012 MIG C6 Maintenance and cleaning 5 Maintenance and cleaning 5 1 Maintenance The MIQ C6 require
13. WE KNOW o amp O mmmum IEEE a xylem brand Operating Manual IO SENSOR NET MIQ C6 IQ SENSOR NET six channel current output module ba76033e01 01 2012 MIQ C6 i Contact Copyright Note For the most recent version of the manual please visit www ysi com YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www ysi com 2012 Xylem Inc ba76033e01 01 2012 MIQ C6 List of contents MIQ C6 List of contents OvervieW ieee keeled RR RR gw ee dus 1 1 1 1 How to use this component operating manual 1 1 1 2 Features ofthe MIQ C6 1 2 Safety 2 1 2 1 Authorized use ee RA 2 2 2 2 GeneralsafetyinstructionS 2 2 Installation ke ers y DR A 3 1 3 1 Scope of delivery 3 1 3 2 Installation in the IQ SENSOR NET 3 1 3 3 Electrical connections General instructions 3 2 3 4 Connections to the current outputs 3 3 Setting S E NE PR NU eK 4 1 41 Entering editing the name of an output 4 2 4 2 Linking the output with a sensor 4 3 4 3 Deleting a link with an output 4 4
14. at error Fixed current value In the case of an error the current output supplies the current value defined in the Current with error field any in the range 0 21 mA Unchanged In the case of an error the current on the output remains unchanged ba76033e01 01 2012 MIQ C6 Settings Function 4 4 3 Fixed current value With the Fixed current value function you can test the operativeness of the instruments connected to the outputs output different current values to the output and while doing so check the behavior of the connected instrument Setting Selection Values Explanation Current output Fixed current Using Save and quit the value nominal amperage that was input as nom is output Inom 0 20 The nominal amperage that is output Note The settings for other functions in the Current output menu as for example PID controller and Recorder are retained while the Fixed current value is carried out ba76033e01 01 2012 Settings MIQ C6 4 5 Checking the status of the outputs This function offers a simple overview of the states of all outputs of a MIQ C6 For current outputs the current value available at the output is displayed The Status of output channels function is available in the Settings ServicelList of all components menu 1 Call up the measured value display with 2 Open the Settings menu with 3 Highlight the Servi
15. ccording to chapter 6 TECHNICAL DATA Only operation according to the instructions in this operating manual is authorized Any other use is considered to be unauthorized Unauthorized use invalidates any claims with regard to the guarantee 2 2 General safety instructions The MIG C6 is constructed and inspected according to the relevant guidelines and norms for electronic instruments see chapter 6 TECHNICAL DATA It left the factory in a safe and secure technical condition The failure free function and operational safety of the MIQ C6 is only guaranteed if the generally applicable safety measures and the special safety instructions in this operating manual are followed during its use The failure free function and operational safety of the MIQ C6 is only guaranteed under the environmental conditions that are specified in chapter 6 TECHNICAL DATA If safe operation is no longer possible the MIQ C6 must be taken out of operation and secured against inadvertent operation Safe operation is no longer possible if the MIQ C6 has been damaged in transport has been stored under adverse conditions for a lengthy period of time is visibly damaged no longer operates as described in this manual If you are in any doubt contact the supplier of your MIQ C6 ba76033e01 01 2012 MIQ C6 Installation 3 Installation 3 1 Scope of delivery The scope of delivery of the MIG C6 is listed in the INSTALLATION cha
16. ce menu item with and confirm with The Service dialog box opens 4 Select the List of all components menu item with and confirm with The List of all components dialog box opens 5 Select the required component column Model entry MIQ C6 with and confirm with 99 The Status of output channels window opens Terminal 1 15 Hay 2004 14 53 qi A CD Status of output channels 334 No Name Chan Status D 1 2 Belebung C1 14 12 m 001 02 Bel ebung C 1 15 m DO1 HO03 Beleb 0 00 m DD1 CSB Zulauf 12 06 m 001 0 00 m 001 0 00 m Return ESC Fig 4 7 394 Status of output channels 6 Using M or 69 exit the Status of output channels window 4 14 ba76033e01 01 2012 MIQ C6 Settings Error events Freezing the states of the outputs Resumption of the normal function 4 6 Behavior of linked outputs 4 6 1 Behavior in case of error For linked current outputs you can determine the behavior of the outputs inthe case of an error Depending on the function of the output the behavior in case of errors is set in the following menus Function Menu Recorder Behavior at error see section 4 4 1 PID controller Behavior at error see section 4 4 2 The defined behavior occurs in the case of the following events or conditions The linked sensor does not supply a valid measured value display of nit Error or OFL Commun
17. ed sensor Note Before editing sensor settings a note appears on the display to inform you that all links of the sensor with outputs will be erased when you change the Measuring mode or Measuring range sensor setting ba76033e01 01 2012 MIQ C6 Settings Maintenance condition when calibrating Manually switching on the maintenance condition i 4 7 Maintenance condition of sensors The maintenance condition of sensors serves to calibrate clean service and repair remove and replace sensors In the maintenance condition the system does not react to the current measured value or the condition of the selected sensor linked outputs are frozen sensor errors do not prompt changes in the conditions of linked outputs During calibration the maintenance condition of sensors is automatically switched on After the calibration the sensor remains in the maintenance condition until the maintenance condition is switched off manually see section 4 7 2 Switch on the maintenance condition manually when you want to clean service or repair remove and replace a sensor see section 4 7 1 When you have finished cleaning servicing or repairing the sensor switch off the maintenance condition manually see section 4 7 2 Note When a sensor is in the maintenance condition the measured values or status indications of the sensor in the measured value display flash ba76033e01 01 2012 Settings MI
18. enter to accept the instrument for service The Product Return form may be downloaded at www ysi com and clicking on the Support tab ba76033e01 7 1 Contact Information MIQ C6 7 2 ba76033e01 01 2012 MIQ C6 Lists Message code EA4421 EA5421 EI3421 8 Lists 8 1 Explanation of the messages In this chapter you will find a list with all the message codes and corresponding message texts that may occur in the log book of the IQ SENSOR NET system for the MIQ C6 output module Note Information about Contents and structure of the log book and Structure of the message code can be found in the LOG BOOK chapter of the IQ SENSOR NET system operating manual Note All message codes of the MIQ C6 output module end with the number 421 8 1 1 Error messages Message text Current output range undercut Check process Check settings and if necessary change them Current output range exceeded Check process Check settings and if necessary change them Burden resistor too large 500 Ohm or current loop interrupted Check burden terminal connections and connection lines 8 1 2 Informative messages The MIQ C6 output module does not send informative messages ba76033e01 01 2012 Lists MIQ C6 ba76033e01 01 2012 a xylem brand 1725 Brannum Lane Yellow Springs Ohio 45387 USA 1 937 767 7241 800 765 4974 US FAX 937 767 1058 Email environmental
19. etween the measured value and current I at the controller output forms the following characteristic controller curve Current I mA Imax lo Imin I Nominal value Measured value Fig 4 5 Proportional controller negative characteristic curve The proportional band Xp is limited by the measuring range of the linked sensor If a value is entered for the parameter Xp that is greater than zero the controller has a negative characteristic curve example Fig 4 5 To obtain a positive characteristic curve a negative value must be entered for Xp ba76033e01 01 2012 4 9 Settings MIQ C6 Application example Feedback control of the D O concentration Sensor TriOxmatic 700 IQ measuring range 0 60 mg l Control parameters Value Nominal value 4 mg l Xp 10 96 of the measuring range or 6 mg l Imin 8 mA Imax 14 mA lo 12 mA ti 0 s no I controller part td 0 s no D controller part The control parameters form the following negative characteristic curve Current mA Imax 14 lo 12 Imin 8 2 4 8 Measured value mg l Nominal value Fig 4 6 Characteristic curve for application example The controller works with the following gain K 6 mA 6 mg l mg l ba76033e01 01 2012 MIQ C6 Settings Within the proportional band a concentration increase by 1 mg l causes a decrease of the current output by 1 m
20. ication of the MIO C6 with the IO SENSOR NET controller is disturbed for longer than 2 minutes The supply voltage for the MIQ C6 is too low e Inthe Recorder function the measured value of the linked sensor is outside the range between Start value and End value Independent of the specified behavior at error the following conditions lead to the freezing of the output states The linked sensor is in the maintenance condition display of Cal Clean or a flashing measured value Communication of the MIO C6 with the IQ SENSOR NET controller is disturbed for a short time After the malfunction has lasted for 2 minutes the output changes to the specified behavior in the case of an error The current output automatically returns to its normal state as soon as all error conditions are over and all conditions causing a freezing of the outputs are cleared ba76033e01 01 2012 Settings MIQ C6 4 6 2 Behavior in the non operative condition In the non active condition the output current is 0 mA An output is non operative when no function is activated for the output An output becomes non operative in case of Power failure As soon as the supply voltage is sufficient again the non operative condition of the outputs ends The outputs function as specified by the user again Erasing a link to a sensor Changing the Measuring mode setting for a linked sensor Changing the Measuring range setting for a link
21. ink with and confirm with 69 The link is deleted 4 4 ba76033e01 01 2012 MIQ C6 Settings 4 4 Setting current outputs 1 Call up the measured value display with 2 Open the Settings menu with 3 Highlight the Settings of outputs and links menu item with and confirm with The Settings of outputs and links display appears 4 Highlight the Feature column with and confirm with 99 5 Highlight a line for a current output Cx in the Feature column with and confirm with 99 The Settings of outputs and links display opens 6 Highlight the Current output menu item with and confirm with 6 Terminal 1 15 Hay 2004 14 53 gi AN D Settings of outputs and Links 150 Quit Current output No function Save and quit Select setting 5 Fig 4 4 150 Settings of outputs and links 7 Select a function with and confirm with 99 Function Settings No function The current output is not used Hecorder see section 4 4 1 PID controller see section 4 4 2 Fixed current value see section 4 4 3 ba76033e01 01 2012 Settings MIQ C6 Function Settings 8 Carry out the settings for the current output with and 9 Highlight and confirm Save and quit with and 99 The new settings are stored 4 4 1 Recorder In the Recorder application the measured values of the linked sensor are set up as current intensity at the
22. pter of the system operating manual 3 2 Installation in the IQ SENSOR NET The IQ SENSOR NET provides a number of options for integrating the MIQ C6 mechanically and electrically in the system stacked mounting distributed mounting etc The various types of installation are described in detail in the INSTALLATION chapter of the system operating manual ba76033e01 01 2012 Installation MIQ C6 3 3 Electrical connections General instructions Cable glands All electric cables are fed from below via prepared openings in the enclosure of the MIQ C6 Cable glands with different clamping ranges are included with the MIQ C6 to provide sealing between the cable and enclosure as well as for strain relief Select the matching cable gland for the respective cable diameter Small clamping range 4 5 to 10 mm This cable gland is suitable for all IO SENSOR NET sensor cables sealing ring 20 15 x 1 mm C gt ___ cable gland M16 amp blind plug Large clamping range 7 to 13 mm This cable gland is required for cable sheaths with an outside diameter of more than 10 mm and is screwed into the enclosure via an extension piece sealing ring 20 x 15 x 1 mm NL extension piece M16 M20 FN sealing ring 24 x 19 x 2 mm cable gland M20 Note If necessary you can order more large cable glands in a set of 4 pieces gt Model EW 1 Order No 480 051 3 2 ba76033e01 01 2012
23. s no special maintenance The general maintenance of IQ SENSOR NET components is described in the IQ SENSOR NET system operating manual 5 2 Cleaning The cleaning of IO SENSOR NET components is described in the IQ SENSOR NET system operating manual ba76033e01 01 2012 5 1 Maintenance and cleaning MIQ C6 ba76033e01 01 2012 MIQ C6 Technical data Electrical data Instrument safety Terminal connections 6 Technical data Note General technical data on MIQ modules are given in the TECHNICAL DATA chapter of the IQ SENSOR NET system operating manual Nominal voltage Max 24 VDC via the IQ SENSOR NET for details see the TECHNICAL DATA chapter of the IQ SENSOR NET system operating manual Power consumption 3W Protective class ll Overvoltage ll category Applicable norms EN 61010 1 UL 3111 1 IO SENSOR NET connections CAN CSA C22 2 No 1010 1 2 Additional connectable SENSORNET terminator terminating resistor Number of current outputs 6 Terminal type Screw type terminal strip accessible by opening the lid Terminal ranges Solid wires 0 2 4 0 mm AWG 24 12 Flexible wires 0 2 2 5 mm Line cross section of Europe 1 5 4 0 mm cables carrying mains USA AWG 14 12 voltage Cable feeds 4 cable glands M16 x 1 5 on the underside of the module ba76033e01 01 2012 Technical data MIQ C6 Current
24. y instructions General safety instructions Other labels A 2 Safety instructions This operating manual contains special instructions that must be followed during the installation of the MIQ C6 current output module Thus it is essential for the operator to read this component operating manual before carrying out any work with the system In addition to this manual the SAFETY chapter of the IQ SENSOR NET system operating manual must be followed Always keep this component operating manual together with the system operating manual and all other component operating manuals in the vicinity of the IQ SENSOR NET system The following safety labels in the individual chapters of this operating manual indicate different levels of danger Warning indicates instructions that must be followed precisely in order to prevent serious dangers to personnel Caution indicates instructions that must be followed precisely in order to avoid slight injuries to personnel or damage to the instrument or the environment Note This symbol indicates instructions that describe special features Note indicates cross references to other documents e g operating manuals ba76033e01 01 2012 Safety instructions MIQ C6 Function and operational safety Safe operation 2 1 Authorized use The authorized use of the MIQ C6 consists of providing current outputs in the IQ SENSOR NET Please keep to the technical specifications a

Download Pdf Manuals

image

Related Search

Related Contents

HP Z30i  Samsung HT-D4500 Brugervejledning  Samsung PL PL55  GRIT GIS 75 - Titan66.ru  Cinéma, mode d`emploi : de l`argentique au  Hobbico Rio EP  Samsung Samsung B7610 User Manual  glide vr instruction manual translations contents  Massive Wall light 15021/31/10  Delta Tau TURBO PMAC PCI User's Manual  