MA6/BA6 Mask Aligner Equipment Standard Operating Procedure
Contents
1. 8 7 Ensure the micrometers setting is 10mm this is the large knob on the right and left side of the system X and Y stage micrometer 7 8 8 Ensure that the theta white line on front of system is lined up Michiganiingine o 7 8 9 Ensure that the N gt line for the mask holder is connected 7 9 Loading Mask 7 9 1 There are three different mask holder sizes 4 inch 5 inch and 7 inch 7 9 1 1 5 inch is the default mask holder 7 9 2 Press the change mask key 7 9 3 Slide the mask tray out flip over and place on table to left of system 7 10 es 7 9 3 1 Place the mask to be used on the holder chrome side up Use the three 3 alignment pins to center the mask 7 9 3 2 The major flat of the mask should be facing towards you If you designed a major flat on your mask 7 9 3 3 Depress the enter key to turn the vacuum on 7 9 3 4 Push the silver tab down on the spring loaded clip to engage it towards the mask 7 9 3 5 Use the nitrogen gun to remove any particle on the mask 7 9 3 6 Flip the mask holder over have your hand underneath to verify that the mask is on the tray 7 9 3 7 Slide the tray back into place 7 9 3 8 Press the change mask key The mask tray should now be locked into place 7 9 3 9 The ready to load message should appears on the display Edit Parameter re ON EIN ee A f TA SSEL SSE SSEL Page 4of8 SSL CRE 7 CNS v J Ne 8 p Sy D i n a 7 10 1 Press
2. align mask features to the back of the wafer 7 4 2 1 This is achieved by aligning a captured image of the mask and the real image of the wafer 7 4 3 Minimum feature size is around 0 8um 7 5 WEC Types 7 5 1 WEC wedge error compensation 7 5 1 1 After loading the substrate WEC will occur It 1s the procedure by which the chuck automatically adjusts such that the wafer is perfectly Michigapiingine os parallel to the mask While WEC is in process avoid leaning on the table 7 5 2 Prox WEC three 3 proximity flags act as separators spacers between the wafer and mask 7 6 Mask Clean 7 6 1 Ensure the mask is clean before using 7 6 2 Use the mask clean bench to clean the mask 7 6 2 1 Dip in Acetone IPA 1 IPA 2 and N dry You may also use the Evergreen Mask cleaning station 7 6 2 2 Any particles on the mask can be transferred to your wafer 7 7 Photo Resist 7 7 1 Positive resist Any resist exposed to ultra violet light is developed away 7 7 2 Negative resist Resist remains on the surface where exposed 7 8 Start Up On off key EHS 8 7 8 1 Turn on the system if not already on Use the on off key 7 8 2 Ifthe TV monitor is not on then turn it on 7 8 3 Ensure that Channel 2 is lit lamp power supply The lamp is 405nm 20 mW cm and 365nm 11mW cm 7 8 4 Press load 7 8 5 Ensure that the configuration is MA6 for lithography 7 8 5 1 BA6 is for bonding alignment 7 8 6 Press enter 7
3. edit parameter key 7 10 2 Press the x lateral key to scroll through the parameters You may choose the contact mode exposure time and alignment gap 7 10 2 1 Use the Y up down key to change the desired parameter values 7 10 2 2 Press the edit parameter key when finished 7 11 Wafer Loading 7 11 1 There are a few wafer chucks to choose from The 4 inch chuck is the default chuck S gt gt Sa Eee gt AS ee 5 7 11 1 1 The other chucks are kept in the cabinet 7 11 1 2 Ensure that the vacuum grooves are completely covered by the wafer 7 11 1 3 Ensure that the white line on the wafer chuck is in alignment 7 11 2 Press the load key and slide the substrate holder straight out The enter and unload light should be flashing 7 11 3 Place the wafer on the chuck with the major flat facing you 7 11 3 1 There should be no resist on the backside of the wafer The resist could affect the flatness of the wafer 7 11 3 2 The wafer should touch against the three 3 steel alignment pins on the holder 7 11 4 Press the enter key to turn the vacuum on 7 11 4 1 Gently nudge the side of the wafer to check vacuum 7 11 5 Use the nitrogen gun to blow off any particles 7 11 6 Slowly slide the wafer chuck back into the system 7 11 7 Push enter The WEC shall now be performed Try not to lean on the table 7 12 Top Side Alignment TSA 7 12 1 Ensure illumination is set to TSA MichiganEngineog 7 12 2 Pres
4. a a Date Created May 3 2004 Date Modified June 1 2005 MA6 BA6 Mask Aligner Equipment Standard Operating Procedure Greg Allion and Kimberly Appel 1 Purpose 1 1 Photolithography involves transferring a pattern from a mask to a wafer coated with photo resist 1 2 MA6 has the capability of processing both top and backside alignment 1 3 This tool is primarily used for high resolution lithography lt 2um and or critical alignment tolerances lt 1um 2 Reference Documents 2 1 Karl Suss MA6 User Manual 3 Equipment 3 1 Tweezers 3 2 N2 gun 3 3 Vacuum wand 4 Materials 4 1 N A 5 Protective Equipment 5 1 Eye Protection safety glasses must be worn at all times 5 2 Hand Protection vinyl or latex free gloves 6 Engineering and or Administrative Controls 6 1 The MA6 is located in the Sub micron Room lt I OD AE OF ar Na A S i s SEL Page 1 of 8 SSEL KSA Sey si Se T MichigapEngine amie 6 2 The tool owner and check off person is Greg Allion gallion eecs umich edu 6 3 Only authorized user may operate this piece of equipment 6 4 This tool must be reserved in advance for use Sign up sheets are located on the SSEL Scheduler 6 4 1 If you sign up for a time slot and cannot make it then be considerate and remove your name from the sheet Procedure 7 1 Tool Parameters 7 1 1 This tool shall only be used for feature sizes of 2 microns or less 7 1 2 This tool shall
5. n side down resist should be facing up 7 15 8 Press the enter key to turn the vacuum on 7 15 9 Slide the tray back in and the WEC should start 7 15 10 Check that the Top Bottom key is in bottom mode is illuminated 7 15 11 Adjust bottom substrate focus to obtain a clear image 7 15 12 Align wafer to image of mask 7 15 13 Once satisfied with alignment you are ready to expose 7 16 Shut Down 7 16 1 When you have completed exposing all wafers check that the wafer chuck is clean 7 16 1 1 It should be free of particles and resist 7 16 1 2 Slide tray back into system 7 16 2 Press change mask 7 16 2 1 Slide the mask tray out and place on shelf to left 7 16 2 2 Release the spring clip 7 16 2 3 Press the enter key to turn off the vacuum 7 16 2 4 Place the mask tray back into place 7 16 3 Turn system off 7 16 4 Leave the lamp and TV monitor on 7 16 5 Leave system set up for 4 inch alignment 7 16 6 Do NOT leave in flood exposure 7 16 7 Complete the logbook 7 17 Troubleshooting 7 17 1 Mask cannot be focused 7 17 1 1 Mask loaded wrong ensure the chrome side faces up 7 17 2 Wafer can not focus 7 17 2 1 WEC was not done correctly reload wafer 7 17 2 2 Alignment gap 1s to large adjust gap using separator key 7 17 3 Mask and wafer move at same time during alignment 7 17 3 1 Mask and wafer are touching reload mask and wafer 7 17 3 2 Mask might have lost vacuum check mask vacuum Was
6. only be used for lithography processes requiring critical dimensions or in the event that the 620 is not functioning 7 2 Mercury Arc Lamp 7 2 1 The lamp operates in constant intensity mode 7 2 1 1 As the lamp deteriorates the power is automatically adjusted to keep the intensity of the wavelength 7 2 1 2 The intensity of the lamp may be read from the power supply during exposure or during a lamp test 7 2 2 Typical lamp uniformity is less than 2 over a 4 inch area 7 3 Exposure Modes 7 3 1 Soft Contact the substrate is brought into contact with the mask by a preset force during exposure for features gt 2um 7 3 2 Hard Contact this is the same as soft contact except that nitrogen gives added pressure upward towards the wafer for features lt 2um 7 3 3 Vacuum Contact a vacuum seal inflates to form a chamber the pre vac time serves to allow the vacuum to proceed slowly 7 3 3 1 For long exposures that may cause out gassing 7 3 4 Proximity a user defined distance separates the mask and wafer The mask and wafer makes contact only during WEC unless spacers are used 7 3 4 1 For delicate structures or substrates 7 3 5 Flood Exposure for first level exposure Once the mask and wafer are loaded the aligner proceeds to expose skipping the alignment step 7 4 Alignment Features 7 4 1 Top side TSA the top microscope is used to align mask features to the topside of the wafer 7 4 2 Back side BSA the bottom objectives are used to
7. r press the separation key to increase the distance 7 13 4 Alignment shift can occur between mask and wafer when in contact mode To ensure proper alignment press the Alignment Check Key 7 14 Exposure 7 14 1 Once satisfied with the ailment press the exposure key 7 14 1 1 The microscope head will move up out of the way and the exposure assembly will slide out 7 14 1 1 1 During exposure do not look at the ultra violet light it can be harmful to your eyes over time 7 14 2 After exposure the microscope will drop back into place and you are ready to unload your wafer 7 14 3 Slide the wafer tray out and press the enter key to toggle off the vacuum 7 14 4 Remove wafer and slide tray back in 7 14 5 Press load for next wafer or mask change to retrieve mask and end process SSEL Page 6of8 SSEL Viichi emen cine ami 7 15 Bottom Side Alignment BSA 7 15 1 The bottom side objectives are used to place a pattern on the backside of your wafer 7 15 2 Insert the appropriate wafer chuck onto the slide need the one with the windows 7 15 3 Select BSA IR key 7 15 4 Depress the BSA Microscope key 7 15 4 1 The X Y keys can now move the bottom objectives 7 15 5 Press the top bottom key to focus the mask 7 15 6 After locating the alignment marks on the mask press the grab image key 7 15 6 1 This will then capture the mask image 7 15 7 Press the load key and slide out the wafer tray 7 15 7 1 Load sample patter
8. s F1 and enter to lower the microscope 7 12 3 Press BSA so that the light goes off You are now ready for top side alignment 7 12 4 Key Top Bottom Focus knobs 7 12 4 1 Light on top substrate focus is active 7 12 4 1 1 A top substrate knob is used to focus on the mask 7 12 4 2 Light off bottom substrate focus is active 7 12 4 3 Bottom substrate knob is used to focus on wafer 7 12 5 Position the objectives over the mask alignment marks using the X Y arrow keys 7 12 5 1 The alignment marks are used to ensure each mask step fits on top of the previous step If anything is mis aligned you have the chance for devices to fail 7 12 5 2 The distance between the left and right objectives may be adjusted by using the separation knobs 7 12 5 2 1 The separation knob is located on the left and right side objectives 7 12 5 3 Theta is the knob located at the front of the system 7 12 6 You can select from three 3 different image options split field single field left and single field right 7 12 6 1 Single field doubles the field of view on the screen facilitating the search for alignment marks 7 13 Alignment 7 13 1 Ensure system is in alignment mode the alignment cont key should be illuminated 7 13 1 1 If you are in contact mode you will see an orange light illuminated 7 13 2 The three micrometers located on the wafer stage adjusts the movement of the wafer 7 13 3 If resistance is felt during adjustment of the wafe
9. te Products Viichi emen cine ami 8 1 Broken wafers are to be discarded in the glass trash e Report all accidents injuries spills fires to the SSEL On Call or other SSEL staff For emergencies during non business hours call the SSEL Emergency Response Team at 734 764 4127 or Department of Public Safety at 734 763 1131 EE 1 Ox LA SN SN or a i 3S ath QSoek Page d old qoek ul 1811 Y T 5 S SILA