Contents
1. PRERR 12 1 12 2 How J Contact INFICON em Returning your instrument to INFICON ssssssssee emere eene Declaration Of conformity iii OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 1 User information EN Read this user manual carefully before using the Sensistor ISH2000 HySpeed 1 1 Notes and safety notices This manual contains warnings and cautions concerning the safe use of the product See definitions below WARNING A Warning indicates a hazardous situation which if not avoided could result in death or serious injury It is important not to proceed until all stated conditions are met and clearly understood CAUTION O Caution indicates a hazardous situation which if not avoided could result in minor or moderate injury It is important not to proceed until all stated conditions are met and clearly understood NOTICE E Notice indicates instructions that must be followed to avoid damage to the Sensistor ISH2000 HySpeed or other equipment Note A Note is used to indicate information that is important for trouble free and optimal use of the Sensistor ISH2000 HySpeed 1 2 Document outline The document is devided in two main parts e Getting started e Reference section The Getting started part consists of step by step case examples explaining how to use the Sensistor ISH2000 HySp
2. Debug Mode ON OFF OFF Detector Signal Level 0 100 96 20 96 Direct sensitivity ON OFF ON adjustment Display Threshold 0 10096 5096 nvert Colors ON OFF OFF Language Several choices English Measure Print Button ON OFF OFF Min Calibration Time 0 30s 5s Min Presentation Time 120s 1s Number of significant 2 3 2 Digits Password Max 12 characters No password Password protected ON OFF OFF calibration Printer Port Several choices No Printer Probe Button Several Choices No Function Probe Lamp ON OFF OFF Probe Type Several Choices Hand Probe Reject Indicator ON OFF ON Purge Level 00E37 1 00E 37 1 00E 05 100000 Reference Unit Several choices PPM Reference Value 00E 37 1 00E 37 1 00E 0 2 100 Reject Indication ON OFF OFF OFF Reject Level 1 00E 37 1 00E37 1 00E 02 100 Screen Save Timeout 1 60 min 20 min Sensitivity 15 8 Show Reject Level ON OFF ON Menu Mode Several choices Combined Mode OP Sensistor ISH2000 HySpeed EN 201203 41 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions OP Sensistor ISH2000 HySpeed EN 201203 42 EN 9 Trouble shooting 42 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions In case of trouble using the Sensistor ISH2000 HySpeed try to solve the problem with these simple trouble shooting guidelines If the measures d
3. Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 3 Reference section EN This section of the user manual consists of an in depth explanation and additional information which completes the user manual with all relevant information 8 1 Menu system The menu system is designed as a tree structure similar to that used in mobile telephones The display shows all the levels when browsing down through the menus so that you can always see exactly where you are Fig 8 1 Sensistor ISH2000 HySpeed controls and indicators To enter the menus press Menu button 4 Press lt and gt button 2 and button 3 to choose between main menus If no setting is made in a menu or its submenus within 60 seconds the instrument will revert to the Detection Mode Analysis Mode Button functions The buttons may change functions in different menues Always read the text just above the buttons in the display for the button functions All changes in values are valid only when saved using the Save button button 4 Use the Undo button button 1 to delete a change in value and revert to the previous setting Use the Esc button button 1 to browse backwards through the menus to the start position Detection Mode Analysis Mode To change quickly from Detection Mode to Analysis Mode or vice versa press button 4 three times in succession 8 2 Engineering format Some of the parameters of the detector are written i
4. Upper and lower case Roman letters the numbers A A O a 4 6 and dash Note The space is not supported The leak rate unit string will be cut short at the first space found See Engineering format on page 19 Calibration Time 22 The calibration time decides how long time the detector looks for a reference signal before giving up If the calibration is set to for example 6 seconds the detector will record the maximum signal during 6 seconds after that the operator or external hardware orders a calibration It is very important that all delays in gas exposure as well as reaction time of sensor are taken into consideration when setting the calibration time Calibration will not be correct if the maximum signal comes after that the calibration time has terminated OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions This is also the timeout of the Calibration line in an APC program Minimum calibration time This parameter sets the lowest possible Calibration Time that can be set under the Calibration menu Default is 5 seconds Minimum calibration time should be set to safeguard so that the following two re quirements are fulfilled 1 The hydrogen from the reference leak or gas line must reach the sensor before end of calibration time 2 The sensor must have time to reach its maximum signal before end of calibration time Setting M
5. Combox 590 250 Hand Probe PK50 Connects to the instrument via Combox 590 930 Hand Probe PK50 Flex Connects to the instrument via Combox 590 940 Hand Probe P50 Connects direct to the instrument 590 780 Hand Probe P50 Flex Connects direct to the instrument 590 790 Robot Probe R50 Connects to the instrument via Combox 590 820 Probe Cable C21 Length 3m 590 161 Length 6m 590 175 Length 9m 590 165 Length 4m Spiral 590 163 Length 6m Spiral 590 164 Probe Tip Protection Cap for Hand Probes P50 PK50 and P50 PK50 590 625 FLEX Set of 500 591 234 Set of 50 Probe Tip Filter for Hand Probes P50 PK50 and R50 591 234 Power cable eu 591 146 Power cable uk 591 147 Power cable us 591 853 Hand Probe sensor fit K50 PK50 R50 590 292 Mounting kit Sensistor ISH2000 HySpeed 590 810 Phoenix connector to Sensistor ISH2000 HySpeed 591 792 O ring seal 591 528 Reference Leaks Standard or customer specific leaks for detector See separate calibration Data Sheet 46 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions This Page Intentionally Left Blank OP Sensistor ISH2000 HySpeed EN 201203 47 EN Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 12 Support by INFICON 12 1 How To Contact INFICON For Sales and Customer Service contact nearest INFICON Service Center The address is found on the website www inficon com If you are experi
6. Flow control Note Due to the large variety of printers available on the market INFICON does not take responsibility for the operation of a particular type of printer 34 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Printed data The detector can print the following information E Date and Time for Power on of detector Time of print Value of all gas signals above the Reject Level Test result Accept or Reject Value of signal Result of calibration OK or Calibration Not saved Date and Time Parameter settings oculi UuN Printing of the current value can also be requested by an RS232 command See Connector pin configuration on page 33 or ordered manually by pressing PRINT Table 8 6 Probe type determines information printed Probe type Data printed Hand Probe P50 COMBOX with H65 1 2 3 4 6 Analysis data output The Analysis Data Output option is intended for transferring test results to a supervising computer system such as for example a PLC system Table 8 7 Communication specifications Setting Data rate 9600 baud Data bits 8 Stop bits 1 Parity None Flow control None The data format for Analysis data output consists of nine ASCII characters Seven characters show the value in engineering format See Engineering format on page 19 one character shows the result of the test and one c
7. a silent or or pulsating tone 8 7 APC Settings APC is an abbreviation for Active Probe Control The APC function is for the control of an active probe that has a built in alarm valves or pumps via Probe Control Port 26 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Different probes require controls therefore it is possible to download different drive routines for the instrument from a PC There is a possibility to adapt how to measure by adjusting the timers and Purge Level Probe Type Select the connected probe Choose between Hand Probe and another probe driver installed from the disc delivered with active probe if ordered APC Time A D Adjustable timer used by the APC system Select a APC timer and press Enter to display specific use of this timer APC timer can be used for general purposes in a custom APC program Purge Level Signal level controlling the Purge_Level APC Triggers Standard probes that support active sampling use the Purge Level for fast interruption of sampling that result in high gas signals Setting Purge Level equal to or just above Reject Level will give the fastest possible cycle times for those probes Quick purging also enhances signal repeatability Note Purge level interrupts active sampling of APC probes This means that higher signals will be underestimated as the sensor is purged before full signal has dev
8. by the detector set to a specific sensitivity OP Sensistor ISH2000 HySpeed EN 201203 23 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions To set up The operation in Detection Mode is not quantitative No figures are given but the signal is still increasing and decreasing with gas concentration Therefore there is no actual calibration to be done but rather a setting of the sensitivity to a desired level A typical set up procedure for the Detection Mode is e Set up a reference leak which corresponds to the smallest leak you wish to detect e Put the probe close to the reference leak and note approximately what reaction you get no reaction small medium high full scale within the first few seconds e Set the sensitivity This can be done permanently under the menu Detection Mode or temporarily as a Direct Sensitivity Setting on the display unless you have set this function to OFF under menu Sensitivity Settings There is also an Auto ranging function which can be selected under the Detection Mode Settings menu If the sensitivity is set very high you may find the baseline annoyingly unsteady Note If the Detection Mode is used and the alarm function is required to be activated at a particular calibrated level then the unit must be calibrated The reason for this is that the alarm is immediately based on the Analysis Mode when the Detection Mode is displayed due to inaccuracies in the Detection
9. hydrogen and has not yet recovered Detector Signal Level can be adjusted in the Service Settings menu You can increase Detector Signal Level if you have many small disturbing signals A high setting of Detector Signal Level gives better tolerance to noise gas signals at the expense of accuracy A low setting gives best accuracy but lower tolerance for noise gas signals The Detector Signal Level is set as 1 to 100 of the Reject Leve Default is 20 NOTICE Increasing the Detector Signal Level may give poorer accuracy OP Sensistor ISH2000 HySpeed EN 201203 29 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions HySpeed Parameters Adjustible HySpeed algoritm parameters Useful for tuning the gas metering situation Minimum Calibration Time Lower limit setting for the timer which is used during calibration Contact INFICON AB to have the code sent to you See the web address under section Info Calibration Tolerance Change the acceptance tolerance limit at calibration Adjustable between 0 100 Number of Significant Digits Choice of a number of significant numbers in Analysis and Combined Mode It is used when a more exact measurement is needed A good control of the environment and calibration is required to be useful Contact INFICON AB to have the code sent to you See the web address under section Info Debug Mode This mode is used during service and software development Service Mod
10. 2000 HySpeed Operating Instructions 10 11 12 13 8 7 8 8 8 9 8 10 8 11 8 12 8 13 8 14 8 15 8 16 Trouble shooting Sensistor ISH2000 HySpeed specifications APC SEINS 160000 dE Probe Type APC Time A D Purge Level Display Settings Contrast ars Brightness nvert Colors Screen Save Timeout General settings Language Probe BUM y PPP CAPO ae e arein rA tnvnervaee cou eecaubhigacaut aside astdemuvsanicecedtaededy A A TT NTE AEAEE E NT Change Password Audio Base Frequency Set Clock escrita Service Settings Show Password Instrument system Reset Probe System Reset Detector Signal Level HySpeed Parameters Minimum Calibration Time Calibration Tolerance Number of Significant Digits Debug Mode Service Mode Combined Mode adds Probe Control Port CONNOR pa Status signal patterns Printer port Connector pin configuration Selectable printer types RS232 serial communication APC driver installation oett Sensistor ISH2000 HySpeed installation Aa C ON T E F ngst vg NORGE Default parameters hiec etna dra pce qo pee e i d dare tas Spare parts arid aCCESSOINCS s s sssssrserissssavsnsansvsensnupnessvessussssonssdsenbesenodnesennensonsssssnisenssnensresnssnsns Mejia
11. 203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions EN 4 Precautions 4 1 Read this user manual carefully before using the instrument Hydrogen Leak Detector Sensistor ISH2000 HySpeed is extremely selective Only Hydrogen Sulphide extremely toxic gives a comparable response to hydrogen When working with gas The normal risks associated with working with all compressed gases must be considered WARNING A Pure hydrogen is a flammable gas Only use ready made Hydrogen Tracer Gas of 5 Hydrogen in Nitrogen This is a standard industrial gas mixture used in various industrial applications Note Whenever the word Hydrogen is used in this manual it implies that the hydrogen gas is safely mixed with Nitrogen in the proportions 5 Hp 95 Nb WARNING A Since the tracer gas mix contains no oxygen releasing large amounts of gas in a confined space may lead to asphyxiation WARNING A Compressed gases contain a great deal of stored energy Always carefully secure gas bottles before connecting pressure regulator Never transport gas bottle with the pressure regulator fitted Before connecting tracer gas confirm that the connectors or test object is designed for working at the test pressure WARNING A Pressurising objects at too high pressures can result in a burst object This in turn can result in serious injury or even death Never pressurise
12. 3 and 4 The push buttons are used to Change from one menu item to another using the lt and gt buttons e Press Enter to move down to the nearest submenu e Press Save to save the set value e Press Undo to restore the previously set value e Press Esc to move up to the nearest higher level s 3 3 LEDs The two LEDs indicate the status of the instrument as follows OP Sensistor ISH2000 HySpeed EN 201203 Green flashing slowly during warming up phase Green fixed light indicates that instrument is ready and hydrogen signal is below Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions EN Reject Level limit e Red fixed light together with Reject on display means the instrument has detected a leak larger than the set Reject Level limit e Red flashing rapidly check message on screen See Trouble shooting on page 42 3 4 Ports and connections Sensistor ISH2000 HySpeed Fig 3 2 Sensistor ISH2000 HySpeed ports and connections o Sensistor ISH2000 HySpeed Rear View o Description 1 Probe connection Ground screw Power connector 2 3 4 Probe control port 5 Printer port NOTICE E Always connect all four wires to the Power connector to 24VDC in order to operate 8 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions This Page Intentionally Left Blank OP Sensistor ISH2000 HySpeed EN 201
13. EN 201203 39 Step Action Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 4 Hold the panel in place while fixing the mounting brackets to the detector See Figure 8 7 5 Align the detector horizontally and lock it by tightening the 4 set screws 6 Lock the set screws with the locking nuts p Connect ground terminal to protective ground of cabinet Fig 8 7 Fixing the mounting brackets Bracket to unit Bracket See Sensistor ISH2000 HySpeed specifications on page 43 for electrical connections 8 16 Default parameters Table 8 14 Range and default settings of all Sensistor ISH2000 HySpeed parameters 40 Parameter Range Default Analysis Unit Several Choices PPM APC Time A 0 0 6000 0 s 1 0s APC Time B 0 0 6000 0 s Os APC Time C 0 0 6000 0 s 0s APC Time D 0 0 6000 0 s Os Audio Base Frequency Several Choices 400 HZ Auto Range ON OFF ON Audio Ready Pulse ON OFF ON Audio Threshold 0 100 096 Detection Audio Threshold 0 10096 496 Analysis Brightness 0 21 21 Calibration Time Min Calibra Time 30 s 10s Calibration Tolerance 0 100 25 OP Sensis or ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Parameter Default Clock hh mm ss Contrast 0 20 10 Correlation Values 1 00E 37 1 00E 37 1 00E 00 1 Date YY MM DD
14. Mode signal To locate leaks Detection mode is semi quantitative that is it gives an audio and visual signal which increases as a leak is approached a higher gas concentration and decreases as you move the probe away from the leak It does not display figures In this mode of operation leaks can easily be detected using a sensitivity which can be preset Leaks can be located very accurately even when there are other leaks nearby If for example you are trying to locate a leak on a refrigerator condenser tubing and the tubing has a major leak then you will get an audio signal as soon as the probe is placed close to the condenser tubing When the probe is moved around over the condenser the signal will increase as the probe approaches the leak If the signal goes off the scale simply reduce the sensitivity setting to bring the signal within the scale By working with the sensitivity setting in this way you will be able to locate multiple leaks that are in close proximity to each other Do not expose the probe to more gas than is necessary because it will slowly saturate with time It is good practice to detect a leak locate it and immediately remove the probe to avoid saturation The probe is not damaged by the exposure but it will recover more slowly After excessive exposure it will be less sensitive for a short period of time Background compensation There is always some hydrogen gas in the background In fresh air this is as low
15. ON Sweden If the detector gets outer damage it must be controlled and repaired by service organisation authorised by INFICON Do not expose the probe to a hydrogen concentration higher than 0 1 6 when the instrument is not put into operation this might damage or destroy the probe sensor Note e When the instrument is put into operation the sensor withstands temporary exposure to hydrogen concentration up to 100 Avoid long exposures to high concentrations OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 7 Calibrate the leak detector EN 7 1 Introduction The leak detector is the instrument and the probe together This section of the user manual consists of step by step examples about how to calibrate the detector in the most common cases For more about the calibration routine see the reference section The instrument must be calibrated by using the integrated calibration function to make sure it displays the correct values in Analysis Mode After calibration the instrument will show the correct measured values on the display in Analysis Mode The calibration parameters will be stored into the probe or in the Combox 7 2 Calibration reference There is a possibility to calibrate the detector by Reference Gas concentration A Reference contains a well defined concentration of Hydrogen gas in ppm mixed by air or some inert gas If you use premixed Refer
16. Publication INFICON AB Part number nind69e1 a 1203 All information can be modified without prior notice Hydrogen Leak Detector Sensistor ISH2000 HySpeed 67 INFICON Sensistor ISH2006 Operating Instructions 6 INFICON Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions en Welcome to Sensistor ISH2000 HySpeed Dear customer This manual provides user information for installing and operating the Sensistor ISH2000 HySpeed unit For start up installation advice please contact the supplier The Sensistor ISH2000 HySpeed is a development of the ISH2000 range of hydrogen leak detectors adapted for rapid automated leak detection in packaging lines The leak detection process is built on the Sensistor technology that uses diluted hydrogen as tracer gas The tracer gas is a safe and inexpensive standard mix of 596 hydrogen E949 and 95 nitrogen E941 These gases are commonly used as packaging gases for example and have been approved as food additives The Sensistor ISH2000 HySpeed comprises a compact panel mounted instrument and a wired sampling probe For a complete test station it is normally necessary to install equipment for handling the packages and for communication with the leak detection instrument In principle in line leak testing with the Sensistor ISH2000 HySpeed suits all types of packaging where it is possible to create over pressure for the enclosed tracer gas so as to permit leak te
17. Reference Section asiatico aaa Lac eec d eb deas Ca Loci dan Eo C FE E Cc eu EUR 8 1 Menu system Button fUnctlOris tete 8 2 Engineering format 8 3 Change Test Mode 8 4 Calibration Calibrate as Calibration intervals Sensitivity too low for reject level miu High signal Check reference 20 Sensor condition indicator aedi Calibration messages E Reference Value nee Reference Unit 2022 Calibration Time sd Minimum calibration time wees Password protected calibration Jud 8 5 Detection Mode Settings serienn ss korr eene nennen nene nnne 23 ts M 23 To locate leaks m Background compensation 24 S MSIIVIY iccirco cos 24 Auto RN usa 24525 Direct Sensitivity Adjustment au25 Audio Threshold wen Reject Indicator sl Audio Ready Pulse m EE NM E 25 To analyse leaks m Reject level ie Correlation Value m Analysis Unit 226 Min Presentation Time 26 Display Threshold 226 Audio Threshold 2b Reject Indications Show Reject Level PT Audio Ready PUISE 2 c0c ccaccaccsacceocensesseenneeatessionsesedesaacadsaanseoonnshanevaasessauneecesssaonaasdaaasonepeeenetaetsasacesacecesaaqaasansennepethosnaaninss 26 OP Sensistor ISH2000 HySpeed EN 201203 3 Hydrogen leak detector Sensistor ISH
18. Signal name 14 OUT DET_ERROR 15 OUT LEAK_OUT 16 OUT DET_ON 17 OUT DET_SIGNAL 18 OUT DET_WAIT 19 OUT OUT O0 20 OUT OUT 1 21 OUT OUT 2 22 OUT OUT 3 23 OUT OUT 4 24 OUT OUT 5 25 OUT 24 VDC OUT Status signal patterns Table 8 3 Status signals for pin 14 18 see Pin configuration on page 31 Function DET SIGNAL Gas detected Sensor not recovered DET WAIT High during warm up DET ON High when detector is on LEAK ALARM Leak above Reject Level detected DET ERROR High if Probe Sensor or Cable is broken DET ERROR will go high for a short time 1 5 seconds when the detector is switched on It will go low when the sensor has been checked In normal operation DET ERROR HIGH means that there is a problem with the sensor probe or cable DET WAIT is high when instrument is in warm up mode after switching on power Instrument will also go into warm up if there is a temporary fault in the sensor or sensor connection The timing of the status signals in relation to different events is described by the following two examples Example Input signals issued to control the APC system should have a pulse length of at least 40 ms Example Output signals switch with a cycle time of 10 ms 0 01 s This is the cycle time of the APC system After power on 32 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 Hy
19. Speed Operating Instructions Fig 8 3 Status signals after power on Power off Poweron Warming up Detection or analysis mode No gas detected DET SIGNAL LEAK OUT When detecting a gas signal Fig 8 4 Status signals when detecting a gas signal Gas signal Purge level Reject level Detector signal level 0 PPM DET SIGNAL LEAK OUT PURGE LEVEL Trigger APC 8 14 Printer port The Sensistor ISH2000 HySpeed is equipped with a serial printer port This is the 9 pin D type connector It is used for printer connection RS232 commands and APC driver installation NOTICE E Always switch power off before disconnecting or connecting any cable Connector pin configuration The printer port is a standard 9 pin male D sub The connecting cable is a standard 9 pin file transfer cable Null Modem Cable For the pin configuration refer to Table 8 4 Table 8 4 Pin configuration of the printer port Comments 1 DCD Not used 2 RD Received data 3 TD Transmitted data 4 DTR Not used 5 SG Signal ground OP Sensistor ISH2000 HySpeed EN 201203 33 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Comments 6 DSR Not used 7 RTS Not used 8 CTS Not used 9 CE Not used Only pin 2 Received data pin 3 Transmitted data and pin 5 Signal ground are used Refer to Fi
20. Timer B E nnn Timer C F nnn Timer D G nnn Purge Level H n nnE nn Reference Value I n nnE nn Reference Unit J Text string max 12 characters Reference Unit Program settings JUx x 1 to 8 1 PPM 2 CC S entered as integer in 100 s of seconds 12 0 01s 100 1 s 60000 600s Transfer of parameters Send parameters one by one first send the specific header for example A then wait until the RS232 transmitter unit has sent the data approximately 20 ms and then send the data for example 1 00E 01 Data string must be ended with a carriage return character chr13 dec Example CPPM or C PPM Carriage Return chr 13 This sets Reject Level Rate Unit to PPM Parameters can be sent in any order you like e If your data was received and correct Sensistor ISH2000 HySpeed immediately echoes sends back the data e If you send a non existing header you will not receive anything e ifthe data could not be converted in the Sensistor ISH2000 HySpeed you will receive the string CoEr Conversion Error Note Remember to use capitals for the header APC driver installation APC drivers are installed in the detector from a PC All active probes need a driver to be installed before they can be used For driver installation you will need the following e APC Driver software e File transfer cable e PC computer with Windows 95 or later OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detect
21. as 0 5 ppm parts per million Sensistor ISH2000 HySpeed actively adjusts itself to the background Sensitivity Sensitivity of audio signal and signal bar in Detection Mode 24 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Note This does not affect the Analysis Mode Auto Range Set this parameter to ON for auto ranging of sensitivity in Detection Mode Sensitivity will decrease two steps if the signal reaches full scale Sensitivity is restored to selected Sensitivity See To locate leaks on page 24 when signal returns to zero Direct Sensitivity Adjustment Setting this parameter to OFF will remove the sensitivity adjustment from the Detection Mode display Sensitivity can still be adjusted in the Sensitivity Settings menu after entering password if set Note The sensitivity setting only affects the Detection Mode Audio Threshold Makes it possible to lower the sound to a set level in Detection Mode The level is in of full Detection bar Reject Indicator Makes it possible to show not shown the indication Reject in Detection Mode Audio Ready Pulse This sets the standby sound to a silent or or pulsating tone 8 6 Analysis mode In Analysis Mode the measured value is displayed in figures The default unit is in PPM but it is possible to choose other units See Default parameters on page 40 To analyse leaks The Analysis mode i
22. as concentration less than 10 PPM with long intervals between exposure If the instrument is subjected to a very large gas concentration over a long period a certain amount of insensitivity can occur directly afterwards This saturation can make it difficult to detect very small leaks Therefore make it a habit of removing the probe from the measuring point as soon as the measured value is displayed This gives the detector an opportunity to recover Sensitivity too low for reject level The Detector will warn if sensitivity of sensor is too low to safely detect a leak equal to the set Reject Level limit The warning can be ignored and calibration updated and the CAL_CONF output will be set High signal Check reference The Detector will warn if the calibration signal is very high The warning can be ignored and the calibration updated and the CAL_CONF output will be set 20 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Sensor condition indicator EN The indicator bar extends in length when the sensor is detecting reference gas This indicator can be used for an early warning as to when a sensor replacement will be needed Fig 8 2 Sensor condition indicator INPUNE 99 Calibration Expose to gas Reference Value 1 00E 02 PPM Gas detected The length of the bar shows the condition of the sensor The bar
23. audio frequency despite the actual signal strength Correlation Value Correlation Value is used when it is necessary to correct the relation between the detector signal and the displayed number This might be necessary when you want to display leak rate unit other than the calibrate leak rate unit Examples If you have a reference value in cc s and want to show mm3 s at the screen use 1000 as Correlation Value Analysis Unit The Analysis Unit is a text string with a maximum of 12 characters It is not involved in any calculations The following letters can be used Upper and lower case English letters the numbers 0 to 9 A A O a 6 96 and Space is not supported The string will be shortened after the first space found Min Presentation Time Signal values in Analysis Mode will never be presented shorter than this time Values are however always presented until the signal has recovered The default value is 1 second but values from 0 1 12 0 seconds can be used Display Threshold Hides all measurements under a set 96 of Reject Level Audio Threshold Instrument is silent under a set 9o of Reject Level Reject Indications There are three choices of Reject Level indications except the LEDs indication e Flashing screen e Chopped audio signal e Combination of indication 1 and 2 Show Reject Level Shows the Reject Level value on the display Audio Ready Pulse This sets the standby sound to
24. ctor Sensistor ISH2000 HySpeed Operating Instructions Note You will need to repeat the calibration 2 3 times to get Calibration OK after changing setup or probe e Allow at least 30 seconds between each calibration for best accuracy e If the message No Gas or Unstable Signal is displayed repeatedly go back to Detection Mode and check functionality e If Repeat Calibration is displayed then this means that the measured value deviated more than 25 can be changed between 1 100 from the previous calibration Repeat the calibration procedure Also set the Analysis Unit to the same as the Reference Value If you want to use another unit you have to put a recalculation number into Correlation Value which describes the relationship between the different units 7 4 Reference value by gas concentration The concentration of the reference gas should be near to or slightly higher Reject Level the for best accuracy and always within the range of 20 50000 ppm Set the Reference Value equal to the Hydrogen concentration in your reference gas If premixed reference gas is used the reference value be found on the certificate of analysis issued for the gas concentration Also set the Reference Unit to the same unit as that used to express the leak rate of the reference leak Example Reference gas concentration contains 100 PPM Hydrogen 1 Set the Reference Value 100 2 Set the Reference Unit PPM OP Sensistor ISH2000 HySpeed EN 201203
25. e This mode contains useful information to analyse the gas sensor behavior If the instrument starts in the Service Mode then it is possible to reach APC Service Mode Under APC Service Mode is useful information to check timers I O on the Probe Control Port and other 8 11 Combined Mode In Combined Mode the bar and the sound in Detection Mode is combined with the figures in Analysis Mode this means that at the same the signal is displayed as a bar and the measured value is displayed in figures The loud speaker sound follows the Detection Mode signal Note After a system reset the default mode is Combined Mode When you have located the leak then you can measure its size in the following way 1 Remove the probe from the leak 2 Wait until 0 0 appears on the display 3 Then put the tip of the probe on the leak 8 12 Probe The probe H65 is a non sniffing probe Gas analysis takes place in a sensor that is in the tip of the probe During operation the heat of the probe tip is 50 C Note There are a variety of different probes that can be connected to the Sensistor ISH2000 HySpeed 30 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Changing the Probe EN After attaching a probe the Sensistor ISH2000 HySpeed needs to stabilize and the green LED should blink If it does not then there is a fault in the cable or the hydrogen sensor inside the probe is faul
26. e The hydrogen measurement mode Analysis Mode measures the concentration of hydrogen e The Combined Mode default mode which is a combination of Detection and Analysis mode The Detection Mode operates continuously while the Analysis Mode determines the hydrogen concentration and calculates a corresponding leak rate in a step measurement Detection Mode gives no numbers It therefore needs no actual OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions calibration The sensitivity of the sound signal and the moving bar on the display is set manually or automatically see below When using the instrument in Analysis Mode it must be calibrated as described See Calibrate the leak detector on page 17 in order to give correct figures OP Sensistor ISH2000 HySpeed EN 201203 13 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 6 Operating the detector EN 6 1 To Detect leaks If all you wish to do is to detect the presence of a leak that is find out whether there is a leak or not then use the Detection Mode or use the detection bar in Combined Mode The definition of Leak No Leak will then simply be A leak is a leak when it can be detected by the detector set to a specific sensitivity To set up The operation in Detectio
27. e the sensor Check Sensor Temp Check that the sensor is properly connected to the probe If the fault persists replace the sensor OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 10 Sensistor ISH2000 HySpeed specifications Table 10 1 Power supply specifications AC mains voltage Sensistor ISH2000 HySpeed AC mains current Fuse Nominal battery voltage Operating time Charging time Power supply voltage 24 VDC Power supply current 3 A max Table 10 2 Input and output connections Input output Power input connector Sensistor ISH2000 HySpeed 4 pin Phoenix MC 1 5 5 81 Series Detachable screw terminal Probe control Status port 25 pin D sub female Minimum pulse length 40 ms Input impedance 50 kohm Input maximum range 34 to 38 VDC Input high gt 12 0 VDC Input low lt 8 0 VDC Output current max 0 5 A output max 2 5 A total Inductive loads External clamp diodes recommended Low state voltage Max 1 5 VDC Low state leakage current Max 100 yA Short circuit protection Thermal and electronic Output high gt Supply voltage 2 5 VDC Output low lt 1 5 VDC Serial communication port connector 9 pol D sub male Serial communication port standard RS232 Table 10 3 Miscellaneous specificati
28. eed in a variety common situations The Reference section consists of in depth explanation and additional information which completes the user manual with all relevant information 1 3 Conventions used in this book In this user manual the following text style hardware command is used for references to hardware commands or button labels while this text style software command is used for references to software commands and menu choices OP Sensistor ISH2000 HySpeed EN 201203 5 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Description of equipment Sensistor ISH2000 HySpeed Sensistor ISH2000 HySpeed is a leak test detector optimized for high speed leak testing The model is particularly aimed at testing packages but possible to use for other applications as well where the time allowed for a LEAK NO LEAK decision is extremely limited The testing speed can be up to four tests per second The detector is designed for integrating in a packaging machine or similar applications If you intend to make the leak test using accumulation shell damps scanning or by hand sniffing we recommend the standard model of Sensistor ISH2000 Sensistor ISH2000 HySpeed is equipped with a number of powerful functions making it very easy to integrate in a semi or fully automatic test system The functions range from output of all necessary status signals and printer communication port to an advanced Active Probe Control syst
29. eloped 8 8 Display Settings This section describes the different display settings of the Sensistor ISH2000 HySpeed Contrast Contrast level of display Higher value gives higher contrast The contrast may need adjustment if ambient temperature changes Brightness Brightness of the display lamp A lower brightness value saves energy and prolonges the lamp Invert Colors Change the black to white and white to black Useful in a dark environment to keep a high readability Screen Save Timeout Display lamp will dim to half brightness if instrument is left idle for the number of minutes set by this parameter The screen save timeout can be set between 1 and 60 minutes the function is deactivated if set to OFF The display lamp will return to the OP Sensistor ISH2000 HySpeed EN 201203 27 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions set brightness if any of the display buttons are pressed if a gas signal is detected or an instrument error is detected 8 9 General settings This part describes the general settings of the Sensistor ISH2000 HySpeed Language The Sensistor ISH2000 HySpeed user interface contains the following languages e English e French e German e Italian e Spanish e Swedish Probe Button This is for setting the different functions with the probe button These functions are as follows e Toggle Mode makes it possible to switch between Analysis mode and Detection m
30. em APC This makes the detector capable of controlling advanced sample collecting devices down to simple test fixtures The model consists of one part apart from the manual etc Fig 2 1 Sensistor ISH2000 HySpeed 4 INFICON Sensistor ISH2000 4 A E Aa ae 0 094 100 0 Aa ana 4 4 aa a 84 Description Detector unit User manual not shown 8 Product return form not shown OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 3 Controls and connections The controls and connections are discussed and shown in this chapter Fig 3 1 Sensistor ISH2000 HySpeed controls and indicators Description 1 Display Loudspeaker Control push buttons Probe connector 2 3 4 Earphone socket 5 6 LEDs 3 1 Display The display shows indicator bar in Detection Mode and the figures in Analysis Mode seven main menus Their positions are indicated on a horizontal scale Change from one menu to another using the lt and gt buttons main menus have submenus which are also indicated by horizontal scales and can be selected using the lt and gt buttons scales for setting numeric values languages etc messages 3 2 Push buttons The functions of the push buttons are shown at the lower edge of the display In this manual the buttons are numbered from left to right 1 2
31. ence gas a Certificate will normally follow the gas bottles Reference Gas can be ordered from local gas suppliers Choose a calibration reference size as follows recommendations e Same or higher than the Reject Level but maximum 10 times higher e inthe following range 500 5000 ppm H2 Please contact the provider of the detector for help to select optimal calibration reference for your application 7 3 Calibration procedure Before calibration the Reference Value in the Calibration Menu must be set See With reference gas concentration When calibrating expose the probe to the background air then do the following steps 1 First Menu then Calibration Calibrate Enter 2 Push the Start button or push the probe button 3 Expose the probe to the reference gas leak The probe does not have to be exposed to the to the calibration gas during the whole Calibration Time the time set in the Calibration menu while the bar is moving The instrument only measures the change as the probe goes from the background air to calibration gas While the calibration time bar is moving the probe should be exposed to the calibration gas or reference leak Then the display shows Detecting Gas and gives sound signals Repeat the calibration routine until you can save the calibration If the calibration is not saved the instrument will revert to the previous value after one minute OP Sensistor ISH2000 HySpeed EN 201203 17 Hydrogen leak dete
32. encing a problem with your instrument please have the following information readily available e The serial number and firmware version for your instrument A description of your problem An explanation of any corrective action that you may have already attempted and the exact wording of any error messages that you may have received 12 2 Returning your instrument to INFICON 48 Please use the Product Return Form which was included with the product at delivery Do not return any component of your instrument to INFICON without first speaking with a Customer Support Representative You must obtain a Return Material Authorization RMA number from the Customer Support Representative If you deliver a package to INFICON without an RMA number your package will be held and you will be contacted This will result in delays in servicing your instrument Prior to being given an RMA number you may be required to complete a Declaration Of Contamination DOC form if your instrument has been exposed to process materials DOC forms must be approved by INFICON before an RMA number is issued INFICON may require that the instrument be sent to a designated decontamination facility not to the factory OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 13 Declaration of conformity G TNFICON Declaration of Conformity Manufacturer INFICON AB Westmansgatan 49 SE 582 16 Li
33. escribed below do not result in a functioning instrument send or hand in the instrument to an authorised service workshop for repair See Support by INFICON on page 48 WARNING Opening or dismantling a Sensistor ISH2000 HySpeed that is powered up can cause serious personal injury or danger to life The instrument contains no parts that can be repaired by the user and may only be dismantled by an authorised service technician Table 9 1 Fault symptoms and measures Fault symptoms No sound in Detection Mode and Analysis Mode Measures Press the button repeatedly No picture on display no sound Check the fuse No picture but only sound when exposed to gas Display setting may be wrong Watch the display from the side at a low angle and aim a lamp at the screen Try to see the text so that you can enter the Display Settings menu and adjust contrast and brightness If this doesn t help send in the instrument for replacement of display lamp Table 9 2 Error messages and measures Error messages Measures Check Probe and Cable Red LED flashes quickly Check that the probe cable is properly connected to the probe and the instrument If the fault persists replace the probe cable Error Error in Active Probe See Probe Manual Check Sensor Voltage Error Check that the sensor is properly connected to the probe If the fault persists replac
34. gure 8 5 for the wiring diagram Fig 8 5 D9 Null modem cable wiring diagram CD 1 1 CD RxD 2 gt 2 RxD o IXxD 3H 3 TxD 8 DIRA4 4DTR E e SG 5 5 SG 2 DSR 6 6 DSR 2 RTS 7 7 RTS CTS 8 8 CTS R1 9 9 E 9 R1 Selectable printer types Most PC printers with serial interface can be connected to the 9 pin printer port Parallel Centronics interface printer can be used if connected through a serial to parallel converter The port can be set up for the following printer types PC Printer and Data Dump No printer Printer output disabled Incoming communication is enabled Sensistor ISH2000 HySpeed listens for incoming data but will not print send test results PC printer with serial interface The PC Printer option can be used to print data on most standard PC printer with serial interface Parallel interface printers can be used if connected through a serial to parallel converter see below Note The output format has been chosen to be as simple as possible to ensure that most printers will accept it Therefore the printer output does not use any flow control This means that some printers may delay printing until the input buffer is full or a pre defined timeout has elapsed Table 8 5 Communication specifications Setting 1200 baud Data rate Data bits Stop bits
35. haracter shows line feed LF Table 8 8 The character indicating the result of the test is one of the following Character Result of the test A Accept Previous test was below Reject Level limit R Reject Previous test was above Reject Level limit P Rejected by Purging Previous test was above purge limit and Reject level limit Calibration approved Previous cycle was calibration Calibration was approved F Calibration failed Previous cycle was calibration E Test interrupted by Error that occurred during cycle probe or sensor error etc Example 2 5E 02R LF This example is a line feed LF R means that the test was above the Reject Level limit and the value was 2 5E 02 OP Sensistor ISH2000 HySpeed EN 201203 35 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions For passive probes for example P50 and AP57 data is printed when a signal is detected above Reject Level or when the print button is pressed Activate this under Measure Button menu For active probes e g AP29eco data is printed at end of measurement sequence Printing of the current value can also be requested by an RS232 command or ordered manually by pressing PRINT See Connector pin configuration on page 33 A custom APC program setting the MEAS flag prints as AP55 AP29eco and an APC program not using MEAS flag prints as P50 Detection data output The Detection Data Output
36. in Calibration Time too low will have the following effects e Calibration will fail if calibration time is set too low e Calibration might pass but be incorrect Setting a high Min Calibration Time will have the following effects e Calibration takes longer time than necessary e Calibration gas consumption is higher than necessary NOTICE Correct calibration is an essential parameter in quality testing We therefore recommend that careful consideration is paid to setting an appropriate Min Calibration Time This will inhibit personnel lacking detailed knowledge about calibration from jeopardising quality by setting a too short Calibration Time Password protected calibration If desired the calibration can be set under the general password to prevent the operator from calibrating by mistake In this case you will have to enter the password to start the calibration routine Setting password protection on calibration is done in the General Settings menu Note that you must also set a password The instrument is delivered with no password set 8 5 Detection Mode Settings In Detection Mode the signal is displayed in the form of a bar The length of the bar varies with the gas concentration To detect leaks If all you wish to do is to detect the presence of a leak that is to find out whether there is a leak or not then use the Detection Mode The definition of Leak No Leak will be A leak is a leak when it can be detected
37. libration request Starts calibration if the Sensistor ISH2000 HySpeed has an active probe driver installed Sensistor ISH2000 HySpeed answers with a K if an active driver containing a calibration routine was found and F if the calibration APC sequence not found Calibration doesn t start if Purge level is reached M Measure Request The active test cycle defined by the APC driver starts M is returned if the selected driver supports active test F failed is returned otherwise N Print Request Returns current analysis value Q Set APC in stand by stop a measurement Returns a Q R Activates probe 0 built in P50 Combox with H65 driver Returns an R S Activates probe 1 Installed probe driver or trigg signal handler driver Returns an S X Shift State to Analysis Mode Returns nothing Z Shift State to Detection Mode Returns nothing Y Shift State to Combined Mode Return nothing Supported parameters Table 8 12 The following parameters can be downloaded to Sensistor ISH2000 HySpeed in Analysis and Detection mode Parameter Reject Level A n nnE nn Correlation Value B n nnE nn Analysis Unit C Text string max 12 characters Analysis Unit CUx x 1 to 8 1 PPM 2 CC S up to 8 OP Sensistor ISH2000 HySpeed EN 201203 37 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Parameter Timer A D nnn
38. ly sensitive hydrogen gas sensor based on a microelectronic field effect transistor MOS FET The gas sensitivity appears when hydrogen absorbs into the sensor through a metal alloy metal hydride layer Only hydrogen can diffuse into the metal and this makes the sensors practically insensitive to other substances that do not contain free hydrogen molecules The signals from the sensors are processed by a microprocessor which also controls the sensor temperature with high accuracy and other sensor diagnostics in order to ensure perfect functionality It also automatically compensates for background gas Condition for leak detection To use the leak detector the test object must be filled and pressurized by tracer gas 95 N gt 5 H5 to get a gas flow through the leak The tracer gas is a standard welding gas of industry quality easy to obtain at low cost The generic name is Forming Gas Appropriate gas filling equipment can be obtained from the leak detector supplier Be careful of how tracer gases are handled after use Released tracer gas contaminates the surrounding air with hydrogen and can complicate the following measurements for a time Ensure that the tracer gas is ventilated away from the target area preferably to the outside of the building Leak detection modes The detector operates in three modes e The leak locating mode Detection Mode mainly used for detecting and locating leaks but not quantifying them
39. me Clock date set as YY MM DD Clock runs even when detector is disconnected from the power supply Printer Port The Sensistor ISH2000 HySpeed is equipped with a serial RS232 printer port See Printer port on page 33 Info Contains information about software versions Serial number and Internet contact information 8 10 Service Settings The Service Mode is reached by starting the instrument and at the same time hold the right button down on the panel After start a new main menu called Service Settings will appear Show Password Shows the chosen password in case the customer has forgotten the password Contact INFICON AB to have the code sent to you See the web address under the section Info Instrument system Reset Resets all parameters into the instrument to default settings Contact INFICON AB to have the code sent to you See the web address under the section Info Probe System Reset Resets all parameters into the probe to default settings Contact INFICON AB to have the code sent to you See the web address under the section Info Detector Signal Level The Detector Signal Level is the level which the sensor is considered to have recovered from the last gas signal It decides when the DET_SIGNAL output will come on this signal can be used to block the start of a calibration or new test cycle in semi and fully automatic testers IF DET_SIGNAL is high then this means that the sensor has detected
40. mixtures containing approximately more than 15 hydrogen can detonate when mixed in certain proportions with air NOTICE E Never make your own mixtures Only use ready made mixtures or use a certified hydrogen nitrogen mixture mixer installed by your gas supplier 4 3 Interferences Most tracer gas methods suffer from some sort of interference Either the detector is sensitive to other gases or vapors or there are other sources of the gas present to which the detector is sensitive Some examples of possible hydrogen sources Engine exhaust Battery charging stations Cigarette smoke Breathing air Human flatulence Scratching on aluminum surfaces OP Sensistor ISH2000 HySpeed EN 201203 11 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions EN 5 Working principle 5 1 5 2 5 3 5 4 Leak detection methods Leak detection or leak testing can easily be done with help of Sensistor ISH2000 HySpeed By pressurizing the test object with a tracer gas consist a mix of 5 hydrogen and 95 nitrogen all hole and capillaries will be filled up with the tracer gas The tracer gas penetrates through leaks and out on the outside On the outside the leaks position will be able to detect and the size of the leak be able to measure by the leak detector instrument with a tracer gas sensitive probe Gas sensor Technology The Sensistor ISH2000 HySpeed leak detector is using an extreme
41. n and decreases as you move the probe away from the leak It does not display figures In this mode of operation leaks can easily be detected using a sensitivity which can be preset See To locate leaks on page 24 and Direct Sensitivity Adjustment on page 25 Leaks can be located very accurately even when there are other leaks nearby If for example you are trying to locate a leak on a product and the product has a major leak then you will get an audio signal as soon as the probe is placed close to the product When the probe is moved around and over the product the signal will increase as the probe approaches the leak If the signal goes out of scale simply reduce the sensitivity setting to bring the signal within the scale Working with the sensitivity setting this way you will be able to locate multiple leaks that are in close proximity to each other Note Working inside a confined space such as for example a cabinet or a narrow passage on a combustion engine there is a risk that the background concentration accumulates to levels close to the upper detection limit of the detector In such case it will not be possible to locate leaks as easily as in open spaces OP Sensistor ISH2000 HySpeed EN 201203 15 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Hint It is good practice to detect a leak locate it and immediately remove the probe to avoid saturation The probe is not damaged by the exposure but it
42. n Mode is not quantitative The audio and visual signal will increase and decrease with the gas concentration Therefore there is no actual calibration to be done but rather a setting of the sensitivity to a desired level A typical set up procedure for Detection Mode is e Set up a reference leak which corresponds to the smallest leak you wish to detect e Put the probe close to the reference leak and note approximately what reaction you get no reaction small medium high full scale within the first few seconds e Set the sensitivity This can be done permanently under the menu Detection Mode Settings or temporarily as a Direct Sensitivity Adjustment on the display unless you have set this function to OFF under the Detection Mode Settings menu There is also an Auto ranging function which can be selected under the Detection Mode Settings menu Note If the Detection Mode is used and the alarm function is required to be activated at a particular calibrated level then the unit must be calibrated in accordance with the instructions see Calibrate the leak detector on page 17 The reason for this is that the alarm is based on the Analysis Mode when the Detection Mode is displayed 6 2 To Locate Leaks Note The Detection Mode or use the detection bar in Combined Mode is used to locate leaks This mode is semi quantitative that is it gives an audio and visual signal which increases as a leak is approached a higher gas concentratio
43. n engineering format This format can represent a very wide range of numbers from very small to very large numbers The following examples describes the format used in the detector 1 00E 01 1 00 x 10 10 OP Sensistor ISH2000 HySpeed EN 201203 19 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions 1 00E 00 1 00 x 109 1 1 25E 02 1 25 x 102 0 0125 8 3 Change Test Mode Choose the measuring method you will use in the menu Change Test Mode There are three different methods to choose e Analysis Mode e Detection Mode e Combined Mode See Reference section for a description of the functions 8 4 Calibration Calibrate The instrument must be calibrated by using the integral calibration function to ensure it displays the correct values in Analysis Mode Combined Mode After calibration the instrument will show the correct measured values on the display The calibration parameters will be stored into the probe or the Combox Calibration intervals Calibration is a natural part of leak measurement and an important factor in quality assurance It is impossible to specify an exact requirement for the interval between the calibrations because the applications for which the instrument is used can vary considerably There will be some oxidation of the probe sensor which reduces the sensitivity if the probe sensor e is not subjected to gas for a lengthy period or e is exposed to a very small g
44. nk ping Sweden Phone 46 0 13 355900 Fax 46 0 13 355901 Product Hydrogen Leak Detector Brand Names ISH 2000 Table top model ISH 2000 C Battery operated model ISH 2000 P Panel mounted model ISH 2000 ICE Table top model ISH 2000 C ICE Battery operated model The manufacturer declares the above products to be produced in conformity with the following directives CE Marking Directive 93 68 EEC EMC Electromagnetic Compatibility 2004 108 EC LVD Electrical safety Low Voltage 2006 95 EC WEEE Waste electrical and electronic equipment 2002 96 EC RoHS Restriction of the use of certain hazardous substances in electrical and electronic equipment 202 95 EC Relevant only for battery charger CE marked on the Battery operated model Manufacturers declaration provided on request For INFICON AB September 01 2011 Faded fagit Fredrik Enquist R amp D Manager INFICON AB Box 76 SE 581 02 Link ping Sweden Phone 46 0 13 355900 Fax 46 0 13 35 59 01 www inficon com E mail reach sweden inficon com OP Sensistor ISH2000 HySpeed EN 201203 50 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions This Page Intentionally Left Blank OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Disposal of product when taken out of service According to EU legislation this product must be recovered fo
45. objects that have not previously been burst tested or otherwise approved for the chosen test pressure Note INFICON AB can not take any responsibility for the consequences arising from the inappropriate use of certain test pressures OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Pressure shocks might cause strong sounds which can cause impairment of hearing Check that all relevant legislation and safety standards are complied with before putting Sensistor ISH2000 HySpeed into service 4 2 Hydrogen Tracer Gas for leak detection When pure hydrogen gas is released in air its flammability range spans from 496 to 75 of hydrogen in air Below 4 there is insufficient chemical energy available for a flame to occur Above 7596 hydrogen there is not enough oxygen left to support a flame When for example a mixture of less than 5 5 9o hydrogen in nitrogen mixes with air there is not sufficient energy to support a flame irrespective of the ratio of air to gas When a mixture of more than 5 5 hydrogen in nitrogen is released into air there is a region of ratios of air to gas where the mixture is flammable When for example a mixture of 1096 hydrogen in nitrogen mixes with air there is still very little energy available Only in exceptional circumstances can a flame be self supporting However such mixtures cannot detonate WARNING A Hydrogen nitrogen
46. ode e Zero detection signal in Analysis mode and Detection mode e Measure Print makes it possible to initiate sample cycles or send the values from the hand probe measurement to the printer port e Probe Lamp makes it possible to turn on and off the Probe Lamp Probe Lamp Makes it possible to have the Probe Lamp on even if the other Probe Button function is chosen Change Password The user password is a text string max 12 alphanumerical characters used to lock critical parameters Setting password to an empty string no characters means that no password is needed to modify the critical parameters The default is no password yf Contact INFICON AB if you have lost your user password If the Password Protected Calibration parameter is set to ON you will be prompted for a password when starting a calibration Note Setting Password Protected Calibration to ON has no effect if no password is set Note APC controlled calibration can be started from the bus in both cases Audio Base Frequency This sets the lowest audio base frequency tone in Analysis and Detection Mode Set Clock Real time set as hh mm ss Hours and minutes can be adjusted Seconds will automatically be set to 00 when hours and minutes have been set Clock runs even when detector is disconnected from the power supply 28 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Set Date Real Ti
47. ons Misc Protection IEC529 OP Sensistor ISH2000 HySpeed EN 201203 Sensistor ISH2000 HySpeed IP64 front IP32 back 43 El 44 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Sensistor ISH2000 HySpeed Overall dimensions 275 x 140 x 75 mm 11 x 6 x 3 inches Environment temperature 0 50 C Environment humidity 10 90 RH Table 10 4 Gas sensing specification In Detection Mode Selected unit Sensitivity mbarl s air using 5 H5 9596 N gt as 1 x 107 mbarl s Tracer Gas Table 10 5 Gas sensing specification In Analysis Mode Selected unit Sensitivity Measurement range ppm H2 20 ppm 20 50000 ppm OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions This Page Intentionally Left Blank OP Sensistor ISH2000 HySpeed EN 201203 45 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions EN 11 Spare parts and accessories There are a variety of spare parts and accessories to the Sensistor ISH2000 HySpeed In the following table some of these are presented for a complete list of all spare parts and accessories please visit www inficon com Table 11 1 Spare parts and accessories Part no Combox Box for controlling the gas sensor and signal processing 590 820 Built in Probe H65 Connects to the instrument via
48. option is intended for automated scanning of weld seams etc Note The Detection Data is expressed in arbitrary units Detection Mode signal is not affected by calibration Table 8 9 Communication specifications Setting Data rate 9600 baud Data bits 8 Stop bits 1 Parity None Flow control None The data format for Detection data output contains of ten ASCII characters Nine characters show the value in engineering format See Engineering format on page 19 and one character shows linefeed LF The print time is 50 Hz continuous streaming data Note Data rate is 25 Hz when Sensistor ISH2000 HySpeed is running in Service Display Mode RS232 serial communication The most commonly used Sensistor ISH2000 HySpeed functions can be started configured over the RS232 interface Table 8 10 RS232 communication specifications Setting No Printer PC Printer Data rate 115200 baud 1200 baud 9600 baud Data bits 8 8 8 Stop bits 1 1 1 Parity None None None Flow control None None None RS232 interface commands 36 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Table 8 11 Common used functions EN Command m Calibrate K Measure M Print Request N Stop Measurement Q Hand Probe R Active Probe Installed AP S Analysis Mode X Detection Mode Z Combined Mode Y K Ca
49. or Sensistor ISH2000 HySpeed Operating Instructions 8 15 Sensistor ISH2000 HySpeed installation The instrument can be installed in the operator s panel or any other flat surface of your leak tester Mounting brackets and panel rubber seal are delivered with the detector Refer to Figure 8 6 CAUTION O The detector should preferably be mounted on a vertical surface Tilting more than 30 degrees is not recommended Tilting more makes air circulation poor resulting in increased temperature inside detector This will reduce contrast of display and lifetime of lamp and electronic circuits NOTICE E After installation verify that ambient temperature is below 50 C Fig 8 6 Panel cut out dimensions Free space horisontally Free space vertically 127 mm 5 inches 262 mm 10 3 inches Measurements e Panel cut out 262 x 127 mm 5 x 10 3 inches e Maximum panel thickness 8 mm 0 3 inch e Allow an extra 20 mm 0 8 inch on left and right side for mounting brackets e Depth of open space in panel is at least 15 cm 6 inches for Sensistor ISH2000 HySpeed to fit Installation process Table 8 13 Install the detector as follows Step Action 1 Cut hole in panel according to figure above and remove burrs 2 Check that rubber o ring is in correct position in the grove around the edge of the detector 3 Put the detector in place in the panel hole OP Sensistor ISH2000 HySpeed
50. pmlnt eorr tenian catena natns n ase edena K NENN MR REN UR NR ua rata NGN 6 2 1 SensistorISH2000 BlySpeed ccce estere ta toast a 6 3 Controls and connections necne rrr donna ica aa aca 7 SS MPIC apr 2 2 PUSH DUTLONS ista CEA ID M A a 3 4 Ports and connections 4 PRECAUTIONS isis sc oisi cri santa gn cuc als eve e cU EXERCERE XLV ARM RE RU ETUR EX BUS EG EU ERR ER KERN EXKRA NAR RUNE a MN a MEER DER EE Kn 4 1 When working witht gas ni aa v one tue dent dock dad Rl p oe as eR das eR EL aha 10 4 2 Hydrogen Tracer Gas for leak detection ssssssssssssssssseeeeeenns 11 43 interferences osito a eit pectus corse o Pd ed eii 11 5 Pide 12 5 1 Leak detection MENA Si tii 12 5 2 Gas sensor Technology seston eiaei ara e eene ANN nennen nene 12 5 3 Condition for leak detection nono no nonon ai ai NEn 12 5 4 Leak detection MOAS susce cete reds e loe 12 6 Operating the detector iere irri nicer tenia tex traba be Eee pe ante Ori Nee Ea Ted iceds 15 AU A RT 15 62 Todkocate Saks ut tia 15 6 3 To Quantity Leaks 5 ccpit E FER RSE k 16 7 Calibrate the leak detectOoF n nisi inpr emm Ri ri nd da iri sd un Enn Ra RR En ENS ER EX XLIA RR ARR YR REM AM Ra RT o eR ERR RR 17 A c 17 7 2 Calibrationrefererice iiie serbe tete teta dab dte oce o e ias 17 4 3 Calibration procedure ocio di edet ire Rabe uw bet used e t sE A 17 7 4 Reference value by gas concentration sss ene 18 8
51. r separation of materials and may not be disposed of as unsorted municipal waste If you wish you can return this INFICON AB product to the manufacturer for recovery The manufacturer has the right to refuse taking back products Sa that are inadequately packaged and thereby presents safety and or health risks to the staff The manufacturer will not reimburse you for the shipping cost Shipping address INFICON AB Westmansgatan 49 582 16 LINKOPING SWEDEN OP Sensistor ISH2000 HySpeed EN 201203 51 52 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions OP Sensistor ISH2000 HySpeed EN 201203 INFICON INFICON AB Box 76 SE 581 02 Link ping Sweden Phone 46 0 13 35 59 00 Fax 46 0 13 35 59 01 www inficon com E mail reach sweden inficon com
52. s used for measuring the size of a leak or the concentration of a gas sample To be able to do this measurement and obtain correct values the instrument must first be calibrated using the calibration function In the Analysis mode the detector determines the gas concentration from the change as the probe goes from being exposed to background to being exposed to a certain gas concentration The detector does not continuously monitor the gas concentration but takes just one reading instead Another suitable alternative name for this mode could be Sampling Mode It is important to keep this in mind when using the detector in this mode In Analysis mode the probe should be moved directly from a background situation to the test point The size of the leak in PPM or any other selected units is shown on the display The probe can and should be removed from the measuring point as the measured value remains on the display The period during which the measured value is displayed can be adjusted in the Display Settings menu Reject level Threshold level for Reject decisions When this level has passed Reject it will be indicated by audio and LED signals Reject on APC bus high OP Sensistor ISH2000 HySpeed EN 201203 25 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Note The frequency of the acoustic signal in Analysis Mode is controlled by Reject Level A signal equal to the Reject Level will always give the same
53. sting All types of flexible plastic packaging such as bottles bags for powders and creams bag in box bags trays and tubes can be tested When reading this manual please pay particular attention to Warning Caution and Notice found throughout the text The Warning Caution and Notices are defined in chapter 1 1 This product complies with the requirements of European Directives listed in the Declaration of Conformity found on page 49 in this document These Directives are amended by Directive 93 68 E E C E C Marking Copyright Intellectual property The use of INFICON AB products are subject to copyright and intellectual property rights in force in any jurisdiction All rights reserved including copying this document in whole or in any part without prior written authorization from INFICON AB Though INFICON AB believes reasonable efforts have been made to ensure the accuracy of the information contained in the document it may include inaccuracies or typographical errors INFICON AB reserves the right to change the information inside the document without further notice 2 OP Sensistor ISH2000 HySpeed EN 201203 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions Contents 1 User informations ae eecer ecd rep td 5 121 INotes and safety NOTICES snp t e ati 5 1 2 Document OoUtliie recette id 5 1 3 Conventions used INS DOOK ure a tator tese cvv deest Dc 5 2 Description of equi
54. tructions Explanation Sensitivity of sensor is too low to guarantee correct response to a gas flow or concentration equal to the Reject level The most likely reason is that Check reference Gas valve may be shut Check that probe tip is not clogged Check setting of Reject Level sensor is too old Replace sensor if problem remains Check that reference gas mix is not replaced with tracer gas mix Check condition of reference Check that reference leak connections has no leaks High signal Check Reference signal is reference abnormally high Note If calibration fails you can still use the instrument Last valid calibration parameters will be used You should however check that the instrument reacts to the reference Reference Value Your reference should have a concentration or flow equal to or slightly above what you want to measure See the examples below for specific recommendations Example for reference gas e Reject level is set at 100 PPM e For good accuracy use a reference gas between 500 5000 ppm hydrogen Reference Unit The Reference Unit is set in the Calibration menu Select PPM cc s cc min SCCM g a oz yr mbarl s mm3 min Pa m3 s or Custom When you select Custom you can enter any unit as long as it contains a maximun of 12 characters Calibration can be performed with e a known hydrogen concentration e aknown flow leak The following characters can be used
55. ty When the stabilisation period is over the green LED should stay on Before using the Sensistor ISH2000 HySpeed the instrument needs to be calibrated Repeat calibration after one hour to achieve greatest accuracy 8 13 Probe Control Port The Sensistor ISH2000 HySpeed is equipped with a parallel Probe Control port This Probe control port can be used for controlling active probes feeding status signals to a supervising computer system and for simple test fixture control CAUTION The Probe Control Port 25 pin D type on the back of the instrument is not a computer or printer port Connecting a printer or any other computer device may cause permanent damage to the connected device Pin configuration for the different detector models is described under Model Specific Specifications below See Sensistor ISH2000 HySpeed specifications on page 43 for electrical specifications See Status signal patterns on page 32 for signal patterns Probe Control Port Connector The control port connector is a 25 pin female D sub Refer to Table 8 2 for the pin configuration Table 8 2 Pin configuration Signal name 2 GND 3 GND 4 IN IN O 5 IN IN 1 6 IN IN 2 9 IN IN 3 8 IN IN 4 9 OUT CAL CONF 10 OUT OUT 6 11 GND 12 GND 13 GND OP Sensistor ISH2000 HySpeed EN 201203 31 Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Instructions
56. will become shorter if the sensor has lost some in sensitivity The scaling of the indicator is not precise enough to say at exactly what length the sensor must be replaced You will learn when this happens for your particular application The instrument will also tell you in clear text when sensitivity is too low See further in the next section below Calibration messages Table 8 1 Different messages that can be displayed during calibration Message Expose to background Explanation Prepare the probe for calibration by holding it in hydrogen free background Remedy Detecting gas Gas signal is detected Normal operation gas exposure can be interrupted Repeat calibration Calibration was not within 25 of last Adjustable by setting Calibration Tolerance Wait 30 s and calibrate again Calibration OK OP Sensistor ISH2000 HySpeed EN 201203 No gas or unstable signal Calibration was within acceptable limit No gas signal or no stable signal detected during calibration Signal when reference gas is shut off Happens for reference gas only Press Save button 4 to store calibration in memory Check reference Gas valve may be shut Check that probe tip is not clogged Background is higher than reference gas concentration Improve ventilation 21 Message Sensitivity too low for Reject level Hydrogen leak detector Sensistor ISH2000 HySpeed Operating Ins
57. will recover more slowly After excessive exposure it will be less sensitive for a short period of time 6 3 ToQuantify Leaks The Analysis Mode or use the analysis figures in Combined Mode is used for measuring the size of a leak or the concentration of a gas sample To be able to do this measurement and obtain correct values the instrument must first be calibrated using the calibration function In the Analysis Mode the detector determines the gas concentration from the change as the probe goes from being exposed to background to being exposed to a certain gas concentration The detector does not continuously monitor the gas concentration but takes just one reading instead Another suitable alternative name for this mode could be Sampling Mode It is important to keep this in mind when using the detector in this mode In Analysis Mode the probe should be moved directly from a background situation to the test point The size of the leak in PPM or any other selected units is shown on the display The probe can and should be removed from the measuring point as the measured value steadies and remains on the display The period during which the measured value is displayed can be adjusted in the Analysis Mode Settings menu The leak detector operates in the range 20 50000 ppm 0 002 5 H2 CAUTION O Do not open detector Service of this equipment may only be carried out by ser vice organisations authorised therefore by INFIC