User`s Manual Heavy Duty Dissolved Oxygen Meter Model 407510
Contents
1. O position If the measured liquid is pure water or is known not to have a salt content set the Salt factor to 0 1 2 3 4 Press the Salt key The display will indicate 0 on the middle portion of the LCD display and S on the bottom portion of the LCD Pressing the Factor Adj key once will add 1 to the salt factor Press the Factor Adj key until the correct factor is displayed Press the Salt key to store the factor NOTE 1 10 ppt parts per thousand ALTITUDE Compensation The meter has a default calibration for sea level operation 0 meters Any significant deviation from sea level requires a compensation adjustment Perform the following steps to do so 1 Press the ALTITUDE key and the display will indicate a 0 in the middle of the LCD and an H on the lower portion of the LCD Press the Factor Adj key once to add 100 meters twice to add 200 meters and so on Maximum compensation is 3900 meters Once the desired compensation is displayed press the ALTITUDE key again to store the value Taking Measurements 1 2 3 4 Immerse the probe in the solution under test For best automatic temperature compensation immerse the probe to a depth of at least 4 10 cm Wait until the display stabilizes The velocity of the liquid coming into contact with the probe must be at least 0 6 to 1 ft min 0 2 to 0 3 m s If the solution is standing stir the solution with the probe or u
2. User s Manual EXTECH INSTRUMENTS Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech s Heavy Duty Dissolved Oxygen Temperature Meter which simultaneously displays Dissolved Oxygen Oxygen in air and Temperature Units of measure are mg l for Dissolved Oxygen for Oxygen in air and C F for temperature Oxygen is measured with a remote sensor that also contains a thermistor for measuring temperature Advanced features include RS 232 PC serial interface data hold auto shut off and Min Max Avg recording Specifications General Specifications Auto power off Sample rate Weight Optional Custom one chip LSI microprocessor circuit Display Dual function 0 5 13 mm 1999 count LCD w contrast adjust Measurements Dissolved Oz mg l milligrams per liter Oxygen in air Temperature C F Sensor Structure Remote Polarographic Dissolved Oxygen Sensor Dimensions Main instrument 7 1 x 2 8 x1 3 180 x 72 x 32mm Probe 4 9 x 0 8 length diameter 125 x 20mm Accessories Carrying case 9V battery and 5 membranes Membranes data acquisition software amp RS232 interface cable o Q O D n n O 3 o n Electrical Specifications Dissolved Oxygen 0 0 19 9 mg L 0 1mg L 0 4 mg L Oxygen in air 0 100 O2 0 1 Oz 0 7 O2 Temperature 32 to 122 F 0 1 F C 41 5 F 0 to 50 C 0 8 C Compensation Temperature automatic 32 t
3. calibration be performed on an annual basis to ensure calibration integrity Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies on sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website at www extech com click on Contact Extech and go to Service Department to request an RA number A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct indirect incidental or consequential damages Extech s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied T Support Hotline 781 890 7440 Tech support Ext 200 Email
4. cludes an Auto shut off feature that preserves battery life The meter will automatically turn off if no function button is pressed in any 10 minute period To disable this feature press the RECORD key to engage the record function LCD Contrast Adjust It may be necessary to adjust the display contrast due to a change in viewing angle or voltage drift Use the LCD Contrast adjustment located on the right side of the meter to set the preferred contrast CPU Reset Certain improper keystroke sequences can cause the meter to hang and become inoperable The following procedure will reset the meter 1 Slide the O DO switch to the O position 2 Turn the meter OFF and disconnect the DO probe 3 Press and HOLD the O key and then press the POWER key 4 Release the O key and then press the ZERO key 5 Connect the DO probe 6 Wait several minutes for the reading to stabilize and then press the O2 CAL key RS 232 PC Interface The meter includes a RS 232 serial data port This interface was designed to operate with the Extech Data Acquisition Software Hardware kit Part No 407000 This kit enables the user to capture store and display readings using a PC 5 Model 407510 Ver 2 1 May 2003 Membrane amp Electrolyte Replacement The tip of the probe holds a thin Teflon membrane and an electrolyte container Calibration or reading errors could indicate loss of electrolyte or a dirty damaged membrane The membran
5. e is very delicate and can easily be damaged if it comes in contact with solid objects Replace the electrolyte or the membrane as follows 1 2 3 Unscrew the tip and well assembly from the probe handle and drain the old electrolyte Unscrew the probe tip Remove the white O ring remove the old membrane and then carefully install the new membrane and secure with the O ring Replace the tip on the well Fill the well to approximately the half way point with new electrolyte Screw the tip and well assembly back onto the probe handle Excess electrolyte may be forced out through the threads Wipe clean Lightly tap the probe to remove any air bubbles w Probe Handle Electrode Well ee e lt __ gt Membrane lt D O ring Tip 6 Model 407510 Ver 2 1 May 2003 Battery Replacement The low battery indication LBT appears on the LCD when the battery runs low To replace the battery 1 Remove the meter s rubber protective cover to access the rear battery compartment 2 Remove the battery compartment cover using a small coin or screwdriver and remove the 9V battery 3 Replace the 9V battery and reinstall the cover and holster Calibration and Repair Services Extech offers complete repair and calibration services for all of the products we sell For periodic calibration NIST certification or repair of any Extech product call customer service for details on services available Extech recommends that
6. o 122 F 0 to 50 C Altitude 0 to 3900 meters probe only 2 Model 407510 Ver 2 1 May 2003 Meter Description 1 LCD Display 2 POWER ON OFF key 3 Data HOLD key 4 C F selection key 5 LCD Contrast Adjust 6 RECORD key 7 RECALL key 8 Factor adjust key 9 Probe top 10 Sensor body 11 Sensor plug 12 Sensor jack 13 RS 232 output jack 14 Battery cover rear 15 Altitude compensation key 16 DO Oz key 17 Zero key 18 O calibration key 19 Salt key Calibration For first time use or after long periods several days of non use calibrate the instrument using the following steps The procedure should be performed in an open well ventilated area Ensure that the sensor is DISCONNECTED from the meter BN OF ON Per 2G IN Turn ON the meter Slide the O DO selector to the Oz position Press the zero key to null the meter the display should indicate OO Connect the DO sensor to the top of the meter Allow the meter to stand for at least five minutes until the display stabilizes Press the O Cal key and the display will indicate approx 20 9 typical Oz in air Model 407510 Ver 2 1 May 2003 Dissolved Oxygen DO Measurements Measurement Preparation Before taking measurements perform the calibration described above if necessary In addition perform the SALT and ALTITUDE compensation adjustments as described below SALT Compensation Slide the O DO switch to the D
7. se an agitator Rinse the probe with clean water after each use Oxygen in Air 02 Measurements Follow the same steps above for DO measurements except for the variations described below 1 2 Slide the O2 DO switch to the O position The display will indicate the oxygen air value in Temperature Measurements The lower display will indicate the temperature of the measured solution Press the C F key to change the temperature units 4 Model 407510 Ver 2 1 May 2003 Advanced Features Data Hold Press the Data Hold key to freeze the displayed value The LCD will display DH for Data Hold mode along with the held reading Press the Hold key again to release the data hold function MIN MAX AVG Data Recording When selected the Data Recording function records the Min Max and Average readings To start a data recording session 1 2 gt Press the RECORD key once The REC indicator will appear on the display and the meter will begin to record the lowest min highest max and average avg readings To recall the data press the RECALL key once The MAX indicator will appear along with the highest reading recorded since the RECORD key was pressed Press the RECALL key again to view the MIN value Press the RECALL key again to view the AVG reading To exit the Record mode press the RECORD key again The display indicators REC MIN MAX and AVG will disappear Auto Power OFF The meter in
8. support extech com Repair Returns Ext 210 Email repair extech com Website www extech com Copyright 2003 Extech Instruments Corporation All rights reserved including the right of reproduction in whole or in part in any form 7 Model 407510 Ver 2 1 May 2003