Home
Server Administrator`s
Contents
1. ProSoly Settings Fort Humber 703 Saved Files Directory SavedFiles Directory is relative to the Server P Code directory Add Delete Settings DK Cancel Configuration Setting Recommended Value Process Name Unique name to describe this ProSolv Server process Executable Path to the ProSolv exe to spawn Optional arguments to pass to the ProSolv exe process default is apps server server pxe Port Number TCP Port Number l THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EES FUJIFILM Page 32 of 83 Configuration Setting Recommended Value Saved Files Directory Folder that stores the saved P Code files Database List of databases that this server will use The following names must be used server ProSolv Database USERLOG User log database 5 Click OK to save the settings 6 To start the newly configured process click the Green Start button to the right of the Process List 7 To stop the process click the Red Stop button 8 To restart the proc
2. b server the actual ProSolv server configuration object E Add Database Entry joj x Type SQL Database sl El Database Sethngs server SOL Server ALPSCIS L SQL Database Drot ol Username Dro ol Password Images Path Option Type LALPSCISTAPSDATA Prosoly Name Examples ProSoly Database server Userlog USERLOG Maclab MacLabDT This is case sensative UK Cancel c Complete the ProSolv Server tab appears as follows THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B PMAN 0014756 A ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 31 of 83 Ey Modify ProSoly Server Properties Process 9 ProSoly Server ProSoly Settings Port Humber 703 Saved Files Directory SavedFiles Directory is relative to the Server P Code directory Databases USERLOG oces Server SOLServer Delete Settings OK Cancel Add ProSoly Server j x 4 Add ProSoly Server EX Process ProSolv Server Process Prosoly Server Process Settings Process Name Executable 7 Arguments
3. Cancel Finish Previous nee E Configuration Setting Recommended Value Left list box Lists the current study locations that will be archived Right text box Designate new study locations to be archived Click New to add the new type Load types from this file Select a predefined locations file to load the study locations from Archiver Configuration Wizard II x Archive the studies to be read by which doctors xii Load types from this file E Cancel Finish Previous HM ext te THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 58 of 83 Configuration Setting Recommended Value Left list box Lists the current study reading doctors that will be archived ProSolv CardioVascular FUJIFILM Right text box Designate a new reading doctor to be archived Click New to add the new type Load types from this file Select a predefined reading doctor s file to load the study reading doctors from Archiver Configuration Wizard oj
4. Configuration Setting Recommended Value Server should perform ADT operations Database Type Aceon Fieldifflame Accession Padding MRN Padding Operator Field Name The name to use for the Operator DICOM tag Store SCU Path The folder to store images for Store SCU Store SCU Path 2 The second folder to store images for Store SCU SR Path The folder where SR data is stored THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Configuration Setting Recommended Value recone e accept RLE images View Label Label to use for the view field o Defaut Exam Type Value touse forthe deiert eene Webviewer IP Address The URL ofthe ProSolv webviewer o Status Update Folder The folder to store status update logs 2 2n Ignore Accession Boolean flag indicating if accession should be ignored 7 Add DICOM Server Process k l x E7 Add DICOM Server Process x E7 Add DICOM Server Ale i x Process DICOM ADT Field Mise Process DICOM ADT Field Mise Process DICO
5. View Studies Archive How Service Control Section View service status Start stop restart the service AddJob Set the Service Log Level Delete dab Settings Log Entries Most recent log entries Show Everything DI View Log Date Time Level Text Log Entries Section View the last entries of the log file Clicking the View Log button opens the log file in a text editor THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 8 of 83 1 1 Integrations 1 1 1 WebLab Integration Configuration ProSolv WebLab is a Web Service that allows ProSolv CardioVascular Client to communicate with ProSolv Application Server Service and access patient data via HTTP or HTTPS protocols ProSolv WebLab acts as a Client s proxy using its own credentials ProSolv Service User account to access and route patient data from the server to the Client application To configure the WebLab integration 1 Start the ProSolv Service Manager 2 Click the
6. WEE 28 1 2 1 CCC lt Aa Yee 28 1 3 BETTENG WEE SE 33 1 3 1 CCC eh Ce 33 1 3 2 Enterprise Licensing Methodology ccccsssccccesseeccndeesscseeceeaudeceseeececaduecesseeseessnanerseass 34 1 4 Watchdog Gernvice kk ENNER ENNER ENNER EN essere M M N Eaa 35 1 4 1 Coie Ce a Ae 35 1 5 DICOM Server Service DICOM Store GC 37 1 5 1 CeCe n 37 1 6 DICOM Query Retrieve SCP Service S SBIR 220 52 ND se ecccccccccssseneeecsscceesseeeeeseseesees 42 1 6 1 Gelueft Cell EEN dE 42 1 7 DICOM Store SCU Genice see nn O D ENNER 43 1 7 1 Configuration enen D 2 REENEN EEN 44 1 8 IMAGE Compression Service EE 45 1 8 1 Configuration QM WEE WEE 45 1 9 eeler eae n EN E 47 1 9 1 Configuration WEN WEE 47 1 9 2 Archiver Job Wizard gy 2 2 ccc cece cceececcececceneesaseeeceacesssaesceaecscaeesceaeeneueesoeess 48 1 9 3 Archive to DIC GRA PAGED Wf EE 58 1 10 Retriever Service EE 60 1 10 1 Configuration ff WEE 60 1 10 2 Retrieve Queen NBR vecccccccccssseccccssereccsserensssesencsseesseserensssecencuseeessaserstegsatenausteess 62 1 11 Measurement Importer Gervice 63 1 11 1 COMA eM ff WEE 63 1 12 HPDT El lg AM candi TT 65 1 13 EA Id e EEN 66 1 14 Burdick Marquette Quinton Treadmill Importers ccccccecccseeeeeeeeeeeeeeeeeeeeeeeeeeaeeeeeeeeaseeeeeeeas 67 1 15 Mennen Hemodynamic Import Service ccccccccecc
7. 1 1 3 OpenECG Integration Configuration OpenECG is the integration with the Epiphany product The modalities supported in the OpenECG integration are Resting ECG and Holter This integration allows physicians to view studies and modify and sign reports using the OpenECG software from an existing workstation configured with ProSolv Client Studies captured via OpenECG will appear in a ProSolv study list and when selected will launch OpenECG web client The ProSolv Server will import measurements and PDF report files from the OpenECG server in fashion consistent with ProSolv report data and measurements and will be completely searchable using SQL queries THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 12 of 83 The ProSolv system will also act as a bi directional gateway for ADT Orders Billing and HL7 Result messages between a Hospital Information Systems and the OpenECG server ProSolv will forward ADT and Orders messages from HIS to OpenECG and will also forward Bil
8. 4 Click the Save button to accept changes 5 To update directory paths click the Browse button to browse to a specific directory and select the required folder 6 Click the Save button to accept any changes NOTE The Optional Save Directory Settings should not be modified by the customer 1 1 2 Synapse Integration Configuration To configure the Synapse integration 1 Start the ProSolv Service Manager 2 Click the Synapse tab All available settings will be displayed in the Synapse Settings table THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM amp ProSoly Service Manager Synapse Settings Name AETitle Host Port No Synapse Url Description ACCSYNO1 spnapse5Dic ACCS YNO1 104 http synapsel Synapse Server Namel aeTitlel host 200 http synapsel Name2 aeTitle2 host2 200 http synapse2 Name aeTitle3 host3 200 http synapsel Named aeT itle4 host4 200 http synapse4 Add Settings 3 Modify Settings by using the Add Delete or Settings bu
9. Line Explanation Possible Values number GU C THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascul CardioVascular Server Administrator s Manual FUJIFILM Line Explanation Possible Values number The full path to the folder Windows formatted file C SRFolder that contains the DICOM path ending with a SR files to be processed by symbol to indicate all this application files in that folder Saeed path should be valid and existing How often DICOM SR application will check the directory for new DICOM SR objects milliseconds The type of ProSolv SQLServer or SQLServer database where retrieved Access file data is stored The network address of the localhost localhost SQL Server IP address of w x y z 192 168 229 81 the machine that the SQL where w x y z are in Server is running on range 0 255 The name of the ProSolv Printable ASCII ProSolvDB database on the SQL characters that are ProSolvDBTEST Server supported by the Operating System The username associated Printable ASCII prosolv with the ProSolv database
10. Previous Configuration Setting Recommended Value Archiver Job Name Unique name to describe this Archive Job Archiver Configuration Wizard O x Where are the Prosoly Image files located Ee Should studies be deleted after they are arched Yes delete studies Archive C No Replicate Finish Previous Hert Configuration Setting Recommended Value Image Path Location Directory where the ProSolv Study images are located Delete study after Checking yes here will delete the local archival ProSolv Study images after they have been archived Checking No here will not delete the local copy THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 50 of 83 Archiver Configuration Wizard f UI x Where is the archive located Ce Onan FTP Server Dn a local or network drive Dr a removable storage device On another Prosolv server C On a DICOM PACS Cancel Finish Previous na ti Configuration Setting Recom
11. WebLab tab 4 ProSolv Service Manager DICOM QA Mennen DT Report Messenger Retriever Store SCU All Services Archiver ProSoly Server License Server DICOM Server Watchdog WebLab Synapse ADT Settings D ode Server Port O04 D ode Server 127 0 0 1 WebLab AE Title ProSolyfebLab Image Directory Vilmage Path Update Directory WUpdates Path HL Reports Directory c HL Reports GE Measurement Transter c GE_Measurement c FaxServer GE Measurement Transfer This i path of GE Measurement Transfer Direcotry Optional Save Director Settings Optional Director Name Optional Director Path 3 Modify settings by clicking in the Settings table and entering new information or browsing to required locations THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 9 of 83 Note The values modified in the GUI are read from and written to an xml file in ProSolv CardioVascular Services 4 0 ProSolv WebLab Web Config Values can be modified by the user in GUI itself A snapshot of th
12. CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 53 of 83 Ss Archiver Configuration Wizard DICOMTest Mex Configure DICOM PACS Acces Store SCP myscp ei Move AETitle PROSOLY Query Configuration Setting Recommended Value Store SCP The Store SCP device that is to receive the archived images Store SCU The Store SCU device that is to perform the archiving Move AE Title The AE Title of the Device that is to perform the DICOM move Query SCP The Query SCP device that should be used when querying studies Manage DICOM Click this button to launch the DICOM Components Device Configuration Screen to add or remove DICOM components THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 54 of 83 BEE DICOM Component Configuration Mix l DICOM Component Hame PAddess AET itle Fort Service Type Maximum Connect
13. PROSOLY sl Query EIS Date Range Is 2007 03 02 sl To C 2008 03704 Manage DICOM eee SECHS Previous Studies ALL Cancel Finish Apply Previous 1 Next gt le Configuration Setting Recommended Value Retriever AE Title Retrievers AE Title By default this is set to ProSolv Retriever Store SCP AETitle for Move SCP The server where the images need to be sent to store be stored THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 62 of 83 Configuration Setting Recommended Value Query SCP AETitle for Query Retrieve SCP NN Modality Filter Modality filter used to query external PACS In the above example only US exams will be retriever from QR SCP PROSOLV_QRSCP Date Range Date range filter to query external PACS If no date range is selected then all the studies in the external PACS will be queried Previous Studies By default it is set to ALL A value in this field indicates previous number of studies to be retrieved for a particular pati
14. The ProSolv Measurement Import Service needs to be configured to import these files into the system This is done by configuring the ECG Management Module Plug in that is part of the Measurement Import Service Please note that in order for the Plug in to be able to access files on the remote ECG Management Server server the Measurement Import Service must be running under a Domain Account that has read access to those file share THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 14 of 83 Ey Importer Settings Plugin Type Epiphany DT VM ProSoly Database Connection String Provider SQLOLEDB 1 Password psci Persist Security Images Directory I images File Importer Settings Search Patten Input Directory lt 0 penECG Server IP cz4hli xml Monitor Subdirectones False Importer Settings Machine Name OpenECG Importer Sleep Interval 5 OpenECG Settings OpenECG Server Root V4i lt 0penECG Server IP gt fiiles File Age mt 30 Output Folder CAE piphany From0 penECG Database Setings Canc
15. m Log level Minimal d ive Typ View Studies Archive Wow Add Job Delete Job Settings Log Entries Most recent log entries Show Everything D View Log Date Time Level Text 3 To create a new Archive Job click the Add Job button This opens the Archive Job Wizard 4 To delete a job select the job in the list and click Delete Job 5 To modify a job s settings select the job in the list and click Settings 6 To manually start an Archive Job select the job in the list and click Archive Now 1 9 2 Archiver Job Wizard The Archiver Job Wizard provides an interface to allow users to more easily configure the settings for an Archive Job It guides the user through the proper steps based on user input THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 49 of 83 Archiver Configuration Wizard UI x Please select type of configurator Ce Archive Retiever Provide a name for this archive joi EE Cancel Finish
16. 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 73 of 83 Report Messenger Settings Machine Name Seconds To Sleep m Directory Information Watch Directory Search Pattern HL Recurse Directory Socket Info IP Add Host Name Watch Port Wait Time ms 0000 Num Wait Cycles Persistent File Processing Remove LF l Seg Delim Delimiters MSH Ae Send By S PROSOLVLPS Return ACK lv Separtate with 7 Cancel Save Configuration Setting Recommended Value Unique name to describe this HL7 task Seconds to Sleep Amount of time to sleep between checks for new HL7 files Watch Directory Input directory to monitor for new HL7 messages Search Pattern Search filter to reduce the types of files that should be processed Recurse Directory Boolean flag indicating if subdirectories should be monitored IP Add Host Name IP Address or Host Name of the Taaa teas ae application Watch Port Port Port number of the receiving Port number of the receiving application Time in milliseconds to wait oo Num Wait Cycles Numbers of cycles to perform when waiting THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWI
17. CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EES Page 80 of 83 FUJIFILM Line Explanation Possible Values number The password associated with Printable ASCII This line can be left blank if the DICOM Server database characters that the database is not password are supported by protected the Operating System 4 Witt Hemodynamic Import Application The Witt Service imports measurements from the Witt Hemodynamic device It consists of two Windows applications that have to be run from the desktop 4 1 Installation To install the Witt Hemodynamic Import Application 1 Copy the Witt files to the desired installation folder on the target machine ProSolv CardioVascular Witt 4 2 Configuration To configure the Witt Hemodynamic Import Application 1 Copy Witt mdb file from the installation folder to main ProSolv Database folder 2 Open the text file pswitt ini from the installation folder in a text editor 3 Edit the file to contain the information listed in the tables below The first table is for SQL the second is for Access database E pswitt Notepad Cvt Repnntez bl Access Cd Heportedtt mdb E vMl Heporteaderndatabhaseumagesz GU a Line Explanation Possible Values number THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROB
18. CardioVascular SR 2 Open double click the psmwls ini file and specify the appropriate configuration information as shown below The format of the file is dependent on the type of database engine used SQL Server or Jet Access only SQL Server is supported as of 4 0 3 The shortcut to ProSolv DMWL should be added to the Startup folder so ProSolv DMWL will launch on login psmwls ini for a SQL Server database engine THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B PMAN 0014756 A Effective Date 02 06 2009 Page 75 of 83 ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator e Manual FUJIFILM E psmwils Notepad File Edit Format Yiew Help PROSOLVMWL 45 1 lt BLANK gt SQLServer localhost ADT_TMH prosoly psci Line Explanation Possible Values Numbers 5 thru 4 E at line 4 6 assuming line 4 contains 1 This line indicates the AE title of the ProSolv DMWL application the application will use this title to associate with the different modalities This line indicates the port number used by the DMWL application to accept requ
19. Dicom Server Store SCP 6 From the PACS we need information about their Store SCP and QR SCP The information needed is the AE Title IP Address and port number of each SCP In some cases this information is the same for both the Store and QR SCP Likewise the PACS needs to know information about the Archiver StoreSCU Archiver QR SCU and the Dicom Server Store SCP For the Store SCU and QR SCU they need the AE Title and in some cases they also want the IP Address For the Dicom Server Store SCP they need the AE Title IP Address and port number Note A different ProSolv Dicom Server instance may be configured for the retrieve but it is not necessary From the Manage Dicom Components you need to setup 3 different Dicom components Archiver Store SCU PACS Store SCP and PACS QR SCP Store SCU Component Configuration Setting Recommended Value Name StoreSCU IP Address 127 0 0 1 this is not used but still need to enter data PROSOLV6 this is what is normally used and what you need to give to PACS Port O this is also not used but still need to enter data thi Service Type Choose STORESCU AETitle Maximum Connections 2 consult with PACS if they have a limit of connections Timeout 3 minutes Store SCP Component Configuration Setting Recommended Value Name PACS Store SCP IP Address XX XXX XXX XXX PACS Store SCP IP is supplied by PACS AETitle PACSAE PACS Store SCP AE Title is s
20. Image Store Store Commitment Boolean flag indicating if Store Commitment should be included in association Auto Shuffle Boolean flag indicating if the DICOM Server should attempt to autoshuffle the study images Show Debug Code Boolean flag indicating if the DICOM THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 40 of 83 Configuration Setting Recommended Value E Server should log debug statements Database Settings Database settings for the DICOM and ProSolv databases Process Dicom ABT Field Miso Process DICOM ADT Field Mise ADT Settings Field Settings IT ADT Enabled Accession Field Name Accession Padding MAM Padding Operator Field Manne Sonographer Store SCU Path El Store SCU Path 2 po TTO SR Path CAProsolv Database SR adl Database Type 0 Server Name 0 Database Username Password View Label Default Exam Type Routine Webviewer IP Address Status Update Folder T Ignore Accession Cancel
21. Manual is devoted to the functions of the Client The Client gives full capability for image viewing and analysis and on line completion and distribution of clinical reports When viewing images the computer monitor screen is organized to maximize the available pixels for digital images rather than controls buttons and icons The default screen permits two side by side high resolution viewing windows This layout highlights the ability of digital imaging to display two or more cardiac images simultaneously which is one of the principle advantages of digital imaging over videotape or cin film and has many clinical uses such as viewing serial studies or different views of the same study To ensure that the highest resolution images are displayed none of the lateral borders of the monitor are used for anything other than imaging Almost all manipulation of images can be performed by clicking on toolbar buttons or pressing keyboard keys The controls are designed to be intuitive requiring relatively brief instruction ProSolv CardioVascular WebViewer is a lightweight Web based application that provides for efficient distributed viewing of images and completed reports The WebViewer is described more fully in a separate document Web Viewer User Manual document control PMAN 0014755 A Intended Use Use ProSolv CardioVascular provides the functionality for viewing measuring analyzing archiving and reporting digital studies generated by digital
22. PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 43 of 83 DICOM QR Task Settings E ll x HI Database Settings DICOM Database Connection String E DICOM QR Settings Acceptable AE Titles Collection a AE Title PROSOLY_QRSCP Destination Store SCPs Collection OR Port Number 1024 Task Name Acceptable AE Tiles List of AE Titles that can connect to this GR Service Configuration Setting Recommended Value DICOM Database Connection string to the ProSolv Connection String DICOM database Click the button to bring up the Connection editor Acceptable AE Titles List of QR SCU AE Titles that the QR SCP Service will accept association from AE Title The AE Title of this QR SCP task Destination Store SCP s List of Store SCP s that can receive images from the QR SCP Task Unique name to describe this QR SCP Task Click the Save button to add the task after it has been configured To start the newly configured task click the Green Start button to the right of the Task List 7 To stop the task click the Red Stop button 8 To restart the task click the Restart button 9 To remove a task select it in the list and click the Delete button 10 To modify a task s settings s
23. Server Watchdog ADT DICOM aR Measurement Importer m Service Control Status Started gt Lal w Log level MEME m Monitoring Tasks Tasks amp Disk Spa License Remote L E Services Description License Server at http Archiver Service QD ProSolv ADT Service QD ProSolv Dicom OR General Purpose Meas E Measurement Importer Grosch Report Messe QD ProSolv Store SCU QD ProSolv Server on port Al DICOM Server on pott Last Performed On Down Time lt Refresh Settings v gt r M Log Entries Most recent log entries Show Everything 7 View Log Date Time Level Text 3 To modify the configuration click the Settings button This will bring up the Watchdog Settings screen gt Watchdog Settings eee Perform tasks every E minute s m Email Settings Disk Space Monitoring I Send waming and error email SMTP Server fo a SMTPPot 25 SMTP Username Bees SMTP Password e I Enable SSL Delete Errors gt Delete _ Address Drive Ic sl Free Space Limit fo Service Monitoring Wamings gt Select the services that should be monitored Service Archiver Streaming Server License Server Configuration Setting Recommended Value Perform Tasks every Time interval to perform the ping tests a aes Send warning and error Boole
24. characters that are sa that allows the DICOM SR supported by the application to connect to Operating System the database directly The password associated Printable ASCII with the ProSolv database characters that are and the user specified in supported by the line 5 Operating System The type of DICOM Server SQLServer or SQLServer database Access The network address of the localhost localhost SQL Server IP address of w x y z 192 168 229 81 the machine that the SQL where w x y z are in Server is running on range 0 255 The name of the DICOM Printable ASCII DCMServ Server database on the characters that are DCMServTEST SQL Server supported by the Operating System THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B PMAN 0014756 A Effective Date 02 06 2009 Page 79 of 83 ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Line Explanation Possible Values number The username associated with the DICOM Server database that allows the ADT application to connect to the database directly The password associated with the DICOM Serve
25. folder in a text editor be Notepad Sele J333 Cc ahdl Heportch 5 Edit the file to contain the information listed in the tables below Line Explanation Possible Values Example number Port number for 1 65535 3336 HL7RealTime2File exe application Es The full path to the folder SQLServer or lt AppServer gt From_Witt where Witt files should be Access stored for processing 6 Launch HL7RealTime2File exe This application reads the HL7 data stream and writes out data files 7 Launch Witt exe This application processes the data files Note that both applications must be running for the Witt interface THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EES FUJIFILM Page 83 of 83 E pswitt Notepad File Edit Format View Help LG valhdl Hepnorte hl ACCESS CWhiLReports itt mdb E whiLReportReaderdatabase mages THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUP
26. imaging instruments from a multitude of manufacturers Users ProSolv CardioVascular is intended for use by doctors and non physicians eg sonographers technologists nurses who are affiliated with medical labs that use images obtained from digital imaging DICOM instruments echo vascular nuclear cardiology cardiac catheterization MRI etc Administration of Server based Applications ProSolv CardioVascular is a client server type application with a significant component of the system s capability being provided by a set of server based applications THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascul CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 2 of 83 Software Hardware The ProSolv CardioVascular image management and reporting system is software that runs on standard PC equipment using Windows 2000 or later operating systems The software can be used in a variety of network configurations Stand alone workstation E Network of workstations connected through TCP IP networking utilizing the ProSo
27. license for the entire enterprise comprised of a main license server and multiple local or remote license servers Configure the license servers as follows THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 35 of 83 1 On the primary main License Server create and add a license for the total number of concurrent licenses for the entire enterprise 2 On the local remote License Server s create and add a single user concurrent license The URL will point to the main license server 3 By default the Enforce License Limits flag on both should be unchecked 1 4 Watchdog Service The Watchdog Service monitors the ProSolv Services and is capable of sending alert emails to administrative staff The Watchdog Service will ping each service ona designated time interval The default setting performs the ping every 5 minutes If the Watchdog cannot ping the service then it attempts to restart the service It will continue to try to restart the service for the next 30 minutes If the Watchdog cann
28. prosolv Initial Catalog Prosolv Data Source lt sqldb gt DICOM Database Connection string to the ProSolv database Clickthe Provider SQLOLEDB 1 Password p Connection String 1 button to bring up the Connection editor sci Persist Security Info True User ID prosolv Initial Catalog Prosolv Data Source lt sqldb gt ProSolv Database Connection string to the ProSolv database Click the Provider SQLOLEDB 1 Password p Connection String 1 button to bring up the Connection editor sci Persist Security Info True User ID prosolv Initial Catalog Prosolv Data Source lt sqldb gt ProSolv CardioVascular FUJIFILM Note Use recommended values unless Site requests otherwise Report Messenger Service Configuration The Report Messenger service needs to be configured to forward HL7 messages over to the Open ECG Server The service will monitor the output folder of the ADT service for HL7 messages As it finds HL7 messages it will forward to the Open ECG Server that is preconfigured to list on a certain port number THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVasc
29. x Which studies should be removed after aching Studies that were performed more than fo days ago reloaded more than fo days ago verified by a third party application Cancel Finish Previous E Configuration Setting Recommended Value Performed more than Designate a number of days that a study must be in age to be removed after archiving Reloaded more than Designate a number of days that a study must have been reloaded by to be removed after archiving Verified by a third party Check this box to remove a study after application archiving if it has been verified by a third party application 1 9 3 Archive to DICOM PACS Archiving to DICOM PACS involves six processes First the Archiver Store SCU 1 sends images to the PACS Store SCP 2 When retrieving the images the Archiver Query Retrieve QR SCU 3 tells the PACS QR SCP 4 to THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascul CardioVascular Server Administrator s Manual FUJIFILM Page 59 of 83 send the images with the PACS Store SCU 5 back to our
30. ATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 22 of 83 Name Deseripion Beeommended Values Watch Directory Directory to watch for input files lt AppServer gt ProSolv Database Epiphany From CardioServer Search Pattern Pattern to search on Recurse Directory Option to check subdirectories Unchecked IP Add Host Name IP Address of the OpenECG server lt IP of HIS system gt ProSolv CardioVascular FUJIFILM Watch Port OpenECG HL7 Port Number Default is 2575 lt Port that HIS is listening on gt Wait Time Number of milliseconds between polls of the Input 10000 Directory Num Wait Cycles 1 Number of tries 3 Gn Delimiters Note Use recommended values unless Site requests otherwise 1 1 4 MacLab Integration Configuration The MacLab equipment exports data by depositing files in a directory that the Prosolv Measurement Importer monitors These files end with the extension h7c for cardiac and h7e for electrophysiology The cardiac file not the EP are retrieved and put into the database Once a match is made on the client the values are added to variables and are available to be put on the report MacLab is also capable of exporting tiff image files that contain wave forms MacLab s output may or may not contain these image files The export with tiff image files may populate the Prosolv Importer input directory in t
31. D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 34 of 83 3 If this is a remote secondary server that is requesting licenses from a primary license server put the address of the main license server on the URL line Ex http 192 168 0 1 LicenseServer Service asmx where the IP address is the IP address of the main license server 4 To modify the configuration click the Settings button This will bring up the License Server Security Settings screen K License Server Security Settings x IT Enforce License Limits Remote Server Access List IP Address d Add Delete IF Address UU UU Port lo SEH Configuration Setting Recommended Value Enforce License Limits Boolean flag indicating if the License Server should reject a license request if all of the licenses are currently in use Remote Server Access List of Remote ProSolv Servers that List can request a license from this License Server 5 Ifyou are configuring a primary main license server that will be sharing licenses with other license servers you must include the IP address and Port of the requesting ProSolv server on this screen 6 To save the settings click the OK button 1 3 2 Enterprise Licensing Methodology In many cases an Enterprise purchases a multi user concurrent
32. D OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual Effective Date 02 06 2009 Page 26 of 83 E ProSoly Service Manager le ProSolv CardioVascular FUJIFILM All Services Archiver rot oh Server License Server DICOM Server Watchdog Compression ADT Semice Control Status Started ks kel kel Log level Verbose Importer This grid will only be populated when the Serice i Aung watch Dir Watch Port Status SA C Prosoly Databas HA Running prozolw WOpenecol exports MA Running Select Add Log Entries Most recent log entries Show Everything sl View Log i P Date Time Level Text y Getting files for SFI ang ETA Getting files for SEI ang 8 08 AM Getting files for SA1 97 2008 80B AM Getting files for SR1 S708 807 AM Getting files for SR1 3 17 2008 B07 AN Getting files for SA1 O 9A7 2008806AM_ Getting files for SR1 9 17 2008 8 06 AM Getting files for SA1 97177 2006 9 05 AM Getting files for SA1 9717720086 6 05 AM Getting files for SA1 91772008 3 04 AM V Getting files for GP G17 49008 ANA AM V DGettine files for SRI 3 Select lt Add gt to configure a new Siemens Importer 4 Make sure that Siemens Hemo Importer is sele
33. DIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM 1 Service Manager Windows service mode applications that are integrated with ProSolv s Service Manager framework can run without a user interface and without having a local user being logged on They share a common tab based user interface for managing their configuration and state The Service Manager provides a User Interface UI to configure and view the status of the ProSolv Services Each of the Services has a separate tab on the main window If a service is not installed a tab does not appear for that service A shortcut to launch the Service Manager is installed on the user s Desktop The first tab labeled All Services lists the services that are installed along with the version numbers Each Service tab has the same set of common controls described below D ProSoly Service Manager Report Messenger Retriever Store scu Al Services Archiver ProSoly Server License Server Watchdog ADT DICOM QR Measurement Importer Service Control Status Stopped Le CH Log level Minimal d he Jobs WT Database Type Archive Type
34. ED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE FUJIFILM Page 69 of 83 1 16 1 Configuration To configure the HL7 ADT Service 1 Start the ProSolv Service Manager 2 Click the ADT tab ProSoly Service Manager Mel x Report Messenger Store SCU All Services Archiver ProSolv Server License Server Watchdog ADT Dicom OR Measurement Importer Service Control Status Started b gt bel Log level Minimal sl Settings Log Entries Most recent log entries Show Everything sl View Log Date Time Level Text 3 To configure the service it must be running Click the Green Start button in the Service Control group box to start the service 4 To add a new Task click the Add button This opens the ADT Task Settings screen THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVas
35. ENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 25 of 83 Name Description Recommended Values for h7c and h7e files Monitor False will not process files in True Subdirectories subdirectories Importer Settings the importer task polls of the Input Directory Note Use recommended values unless Site requests otherwise 5 After completing the entries select lt Save gt select the entry and start the service Note Configure the above per your specific system The MacLab system will also need to be configured so that it deposits files in the Input Directory that is specified in this configuration step 1 1 5 Siemens Sensis Integration Configuration The new generation of our recording solution is adapting to the user s workflow featuring customizable workflow support programs and new advanced measurement functions As a truly integrated solution AXIOM Sensis XP combines both electrophysiology and hemodynamics in one system To configure the Siemens Sensis integration 1 Start the ProSolv Service Manager 2 select the Measurement Importer tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATE
36. Exam Location if nothing else matches The default location should be the last one in the list title 4 To start the newly configured process click the Green Start button to the right of the Process List 5 To stop the process click the Red Stop button 6 To restart the process click the Restart button 7 8 To remove a process select it in the list and click the Delete button To modify a process settings select it in the list and click the Settings button 9 To disable a process from being spawned and pinged by Watchdog click the Disable button 1 6 DICOM Query Retrieve SCP Service The QR SCP Service provides DICOM QR services to requesting QR SCU devices The QR SCP service is multithreaded and is capable of running multiple tasks that can act as separate QR SCP devices 1 6 1 Configuration To configure the DICOM Query Retrieve SCP Service 1 Start the ProSolv Service Manager 2 Click the DICOM QR tab 3 To configure the service it must be started Click the Green Start button in the Service Control group box to start the service 4 To add anew task click the Add button This opens the DICOM QR Task Settings screen THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF
37. FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 39 of 83 M Server Settings AE Title PROSOLVY xi Pott Number w Z Process DICOM ADT Field Misc Max PDU Length 163840 M Process Settings ARTIM Timeout om Process Name OO Read Timeout o aa E Write Timeout 200 Images Path C Prosoly Database images 0000 Pi l Storage Commitment P Auto Shuffle Show Debug Code M Database Settings Type C AccessMJet SOL Server Server Name PO Username P Password P Prosoly Database fC O O DICOM Database P Configuration Setting Recommended Value Process Name Unique name to describe this DICOM Server process Executable Path to the psdcmserv exe to spawn Arguments Optional arguments to pass to the psdcmserv exe process AE Title The AE Title of this Store SCU process Port Number The portnumber for this Store SCU process GES ee Vax PDU Length S E ARTIM Timeout ARTIM timeout value in miliseconds Read Timeout Network read timeoutin miliseconds Write Timeout Network write timeout in miliseconds SSS ee NN Image Path Directory location for
38. ILM Company may have patents or pending patent applications trademarks copyrights or other intellectual property rights covering various subject matters contained in this document The furnishing of this document does not extended any rights to said patents trademarks copyrights or other intellectual property rights except as expressly set forth in any written license agreement from ProSolv CardioVascular A FUJIFILM Company ProSolv CardioVascular is a computer program which is protected by copyright law and international treaties Unauthorized reproduction or distribution of this program or any portion of this program is strictly prohibited THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascul CardioVascular Server Administrator s Manual FUJIFILM Certain images diagrams contained in this manual may not appear clearly on high resolution monitors ProSolv CardioVascular is a registered trademark of ProSolv CardioVascular A FUJIFILM Company Copyright 2008 ProSolv CardioVascular A FUJIFILM Company All rights reserved All brand product and co
39. LEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular FUJIFILM The full path to the folder that contains the Witt files that are processed by this application The type of Witt database where retrieved file data Is stored The network address of the SQL Server IP address of the machine that the SQL Server is running on The name of the Witt database on the SQL Server The username associated with the Witt database that allows the application to connect to the database directly The password associated with the Witt database and the user specified in line 5 Access ProSolv CardioVascular 4 0 2 Server Administrator s Manual PMAN 0014756 A Effective Date 02 06 2009 Page 81 of 83 Line Explanation Possible Values number Windows formatted file lt AppServer gt From_Witt path ending with a symbol to indicate all files in that folder directory path should be valid and existing SQLServer or Access SQLServer localhost 192 168 229 81 localhost W X Y Z where w x y z are in range 0 255 Printable ASCII characters that are supported by the Operating System Printable ASCII characters that are
40. LICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B
41. License Server is in charge of distributing concurrent licenses to client users The License Server is a Web Service that handles license requests from the ProSolv Server The License Server determines the amount of licenses for each of the application types from the license file that is installed on the server The License Server Tab in the Service Manager displays a list of all users that are currently logged into the ProSolv Server It also displayed the number of licenses that are in use and how many are remaining for each application type 1 3 1 Configuration To configure the License Server 1 Start the ProSolv Service Manager 2 Click the License Server tab ProSoly Service Manager ell Report Messenger Store SCU All Services Archiver ProSolv Server License Server Watchdog ADT DICOM QR Measurement Importer License Server URL http localhost LicenseServer Service asmx Saye Log level Minimal sl Concurrent Licenses Ausl Refresh Administrator e Analyzer Settings Analyzer No Images Viewer View Log Current Users Usemame IP Address Last Update Standalone MAC Address THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC
42. M ADT Field Misc Misc Settings Misc Settings Misc Settings Type SOPLookup TagDB Match AE Lookup Type SOPLookup TagDB Match AE Lookup Type SOPLookup TagDBMatch AE Lookup Add Settina SOP SOP Type Tag Name Tag DB Tag Length Location Cancel Cancel Cancel Configuration Setting Recommended Value Misc Settings Select which of the 3 misc items to configure Add Addams oO Deieie Removeaneny S cad Edtanenty SiS SSS A DICOM SOP Class a SOP Type Exam type to use for that SOP Class E Tag Name DICOM Tag to match This is a hex number without any leader i e 00080018 THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascul CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 42 of 83 Configuration Setting Recommended Value Tag DB The name of the column in the EXAM table to store the tag data Tag Length The maximum length of data to store NN AE Title The AE Title of an instrument used to determine an exam location Leave this field blank to specify a default
43. M COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Effective Date 02 06 2009 Page 19 of 83 Recommended Values ProSolv CardioVascular FUJIFILM Server Administrator s Manual Watch Port Host Port lt Port OpenECG is Listening on gt Wait Time Number of milliseconds between polls of the Input 10000 Directory Num Wait Cycles Persistent Remove LF Seg Delim Segment delimiter Note Use recommended values unless Site requests otherwise ADT Listening Task Configuration An ADT Listening Task needs to be configured that will receive a message stream from the OpenECG Server on some predefined port number These messages will not be processed and stored in the ProSolv Database To accomplish this the Process Message flag should be set to FALSE for this task The messages will be saved to a folder so that they can be forwarded to the HIS The ADT Task should be configured to store messages to a folder on the ProSolv Server Create a directory I Epiphany FromCardioServer 47 ADT Task Settings E EI ADT Delimiter Settings HL Delirniters ADT Settings Create Study Images Directory Listening Port Max Stream File Size Message File Age Limit Messsage Store Directory Patient ID as Main ID Process Message Send ACK Store HL Messages Study UID Prefix Task Mame Database 5 etiings ADT Database Connection String DICOM Database Connection String ProSoly Database C
44. M Server are stored in the ProSolv database The DICOM Server is actually a background process that gets spawned by the ProSolv Launcher Windows Service This allows a site to have multiple DICOM Servers running on the same machine 1 5 1 Configuration To configure the DICOM Server Service 1 Start the ProSolv Service Manager 2 Click the DICOM Server tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B i ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 38 of 83 ll x ProSoly Service Manager TEE Service Control Status Stopped ProSoly Launcher Status Log level Minimal DI Server Processes Server Name Port Number Status ProSoly DICOM Stopped Delete Settings Disable Log Entries Most recent log entries Show Everything DI Yiew Log Date Time Level Text 3 To add anew DICOM Server process click the Add button This opens the Add DICOM Server screen THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A
45. MENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B i ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 46 of 83 ProSoly Service Manager 0 x Measurement Importer Mennen DT Report Messenger Retriever Store SCU ME Al Services Archiver ProSolv Server License Server Watchdog Compression ADT DICOM GA Service Control Status Stopped gt ke Log level Minimal DI Compression Settings Quality Compression Factor T Original 1 T Clinical 5 Reference 50 E ProSoly Database Settings Log Entries Most recent log entries Show Everything DI View Log Date Time Level Text Configuration Setting Recommended Value ProSolv Database Connection string pointing to the Connection String ProSolv database Original Check this box to create JPEG2K versions of the original images Clinical Check this box to create Clinical versions of the images at the designated compression factor 3 Navigate to ProSolv CardioVascular Services 4 0 lmage Compression Service folder and open ImageCompression exe config file usi
46. ONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B i ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 72 of 83 ProSoly Service Manager Me x All Services Archiver ProSolv Server License Server Watchdog ADT DICOM GR Measurement Importer Report Messenger Store SCU Service Control Status Started ka ial Log level Minimal D machine C filewatcher Running machine2 C SHL Reports Running Settings Log Entries Most recent log entries Show Everything DI View Log Date Time Level Text 3 To configure the service it must be started Click the Green Start button in the Service Control group box to start the service 4 To adda new task click the Add button This opens the Report Messenger Settings screen THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG
47. OT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascul CardioVascular Server Administrator s Manual FUJIFILM 2 3 Exiting the ProSolv DICOM MWL Application To exit the application click the Exit button refer to the screen capture below Waiting for first request 3 DICOM Structured Reporting Application The ProSolv DICOM SR application processes DICOM Structured Reporting files from the specified folder The application also stores the measurement information retrieved from the files into the ProSolv database to be subsequently displayed by the ProSolv Client application 3 1 Configuration To configure the DICOM Modality SR application 1 Place the ProSolv DICOMSR executable file ProSolv SR EchoVascular exe and the SR ini file into the application folder ProSolv CardioVascular MWL 2 Open the SR ini file double click the icon and specify the appropriate configuration information as shown below The format of the file is dependent on the type of the database engine used SQL Server or Jet Access 3 The shortcut to ProSolv DICOM SR should be added to the Startup folder so ProSolv SR will launch on login The first table is for SQL the second is for Access database
48. OUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B PMAN 0014756 A ProSolv CardioVascular 4 0 2 Server Administrator s Manual C Ve Pate 92 06 2008 Page 16 of 83 for receiving all incoming HL7 from the HIS These messages will be processed and entered into the ProSolv Database The messages will also be saved to a folder so that they can be forwarded to the OpenECG server ProSolv CardioVascular FUJIFILM Create a directory HL7ADT E7 ADT Task Settings El ADT Delimiter Settings HL Delirniters ont El ADT Settings Create Study False Images Directory I Images Listening Fort 33343 May Stream File Size ER Message File 4ge Limit 30 Message Store Directory ICAL g it Patient ID as Mam ID False Process Message True Send ACK True Store HL Messages Study UID Pretix Task Mame Database Sethngs ADT Database Connection String DICOM Database Connection String ProSoly Database Connection String False 1 2 826 0 1 3680043_2 1331 31 75947491 HL Input Listener Provider SQLOLED6 1 Password psci Persist Security Inf Provider SQLOLEDB 1 Password psci Persist Security Inf Provider SQLOLEDB 1 Password psci Persist Securty Infe Task Name Unique name to describe this ADT task Save Cancel Name _ Deseorpton TL Recommended Values D getting created in the system for each order Max Stream F
49. PTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 21 of 83 task will pull in the messages and send them to the HIS on a predefined port number 47 Report Messenger Settings Machine Name Cadet erver Uutoomg Forwarder Seconds To Sleep E Directory Information watch Directory E piphanysFrom CardioS erver search Pattern H Recurse Directoy D l IP address or Machine Name of the OpenECG Server Socket Info IF Add Host Name HIS IF Watch Fe Port that the OpenECG server is wait Time ms 10000 listening on Mum wait Cycles E Persistent File Processing Remove LF Seg Delim ol SEA Manz Configure according to incoming HL7 Cl Messages Sending facility Send By SZ 14BC application Return ACK al Separtabe with 7 Save Cancel Name Description Recommended Values CardioServer Forwarder to HIS Seconds to Sleep Number of seconds between polls of the Input Directory THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPL
50. ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 1 of 83 ProSolv CardioVascular is a web based software only vendor and hardware neutral solution designed to provide multi modality cardiovascular image and information management in a versatile and cost effective manner supporting all cardiovascular modalities From a single workstation with a single log on at any location a physician can review patient studies and create reports within all cardiovascular modalities ProSolv CardioVascular accepts images and information from any DICOM compliant acquisition device In addition our reporting tools have set the standard for customization and advanced analysis capabilities with auto interpretation interactive diagrams with auto summary and spell check Our ability to customize reports that mimic exacting requirements is unsurpassed in the industry The ProSolv CardioVascular software has been carefully designed to address each of the major issues of importance to the field of cardiovascular image and information management including multimodality viewing advanced analysis customized workflow automation through HIS ADT Orders DICOM Modality Worklists Results and Billing integrations ProSolv CardioVascular offers two different clinical user interfaces Introduction ProSolv CardioVascular Client is an installed application for Microsoft Windows Much of the content of this User
51. SE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascul CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 74 of 83 Configuration Setting Recommended Value Boolean flag indicating TBD OO Remove LF Boolean flag indicating if line feeds should be removed during processing Seg Delim Designates the Segment Delimiters for the message Send By Sending device name ai Return ACK Boolean flag indicating if a Return ACK should be sent 5 Click the Save button to add the task after it has been configured 6 To start the newly configured task click the Green Start button to the right of the Task List 7 To stop the task click the Red Stop button 8 To restart the task click the Restart button 9 To remove a task select it in the list and click the Delete button 10 To modify a task s settings select it in the list and click the Settings button 2 DICOM Modality Worklist Application The ProSolv DMWL interface responds to the Worklist requests sent to the application by the different modalities 2 1 Configuration To configure the DICOM Modality Worklist Application 1 Place the ProSolv DMWL executable file ProSolv WorkList Server exe and the psmw s ini file in the same application folder ProSolv
52. Server The ProSolv Server application enables users to share centrally located databases as well as images and reports The ProSolv Server is actually a background process that gets spawned by the ProSolv Launcher Windows Service This allows a site to have multiple ProSolv Servers running on the same machine The ProSolv WebLab Web Service provides HTTP communications between the ProSolv Client and Server 1 2 1 Configuration To configure the ProSolv Server 1 Start the ProSolv Service Manager 2 Click the ProSolv Server tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 29 of 83 ProSoly Service Manager altel ES Report Messenger Retriever Store scu TTT All Services Archiver ProSolv Server License Server Watchdog ADT DICOM OR Measurement Importer Service Control Status Stopped ProSolv Launcher Status Log level Minimal DI Server Processes Server Name Port Number Status Drot oy Server Stopped Delete Settings Disab
53. This line indicates the network localhost localhost address of the SQL Server the IP W X Y Z 192 168 229 81 address of the machine that has SQL where w x y z are in Server running on it range 0 255 This line indicates the name of the Printable ASCII ADT ADT database on the SQL Server characters that are ADTTEST supported by the Operating System This line indicates the username Printable ASCII ProSolv associated with the ADT database characters that are adt that would allow the ADT application supported by the sa to connect to the database directly Operating System This line indicates the password Printable ASCII psci associated with the ADT database characters that are hl7231adt for the user specified in line 9 supported by the Operating System The following screen capture displays an example of how the psmw s ini file should look when configuration is completed for a SQL Server database engine It reflects a setup specifying one acceptable AE title a 7 is placed in line 4 and there is only one line following it File Edit Eee Wie acer localhost ADT prosaly pe cI 2 2 Starting the ProSolv DICOM MWL Application To start the application double click the ProSolv Worklist icon The application will begin listening to requests automatically THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL N
54. ame machinel className Prosoly lmageCompression Compressor Compressor assemblyName Prosoly lmageCompression Compressor secondsToSleep 100 lolompression 40 medCompression 5 campressLo True compressMed False compressOriginal False gt lt campressionContigs gt lt confiquratian gt 1 9 Archiver Service The Archiver Service archives ProSolv Study images to free disk space on the image store The Archiver can archive images to a DICOM PACS FTP Site or a Network Share The Archiver can run multiple jobs that can be scheduled to run at different times 1 9 1 Configuration To configure the Archiver Service 1 Start the ProSolv Service Manager 2 Click the Archiver tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 48 of 83 ProSoly Service Manager lol x Report Messenger Retriever Store scu Al Services Archiver ProSoly Server License Server Watchdog ADT DICOM QR Measurement Importer Service Control Status Stopped Le
55. an flag indicating if Watchdog email should sent alert emails THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascul CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 37 of 83 Configuration Setting Recommended Value SMTP Server The name of the SMTP server to use for sending email SMTP Port The port number for SMTP a SMTP Username User account that has access to the SMTP server SMTP Password SMTP user account password le Va Enable SSL Boolean flag indicating if SSL should be used when sending email Warnings List of email addresses that will receive warning emails Errors List of email addresses that will receive error emails Disk Space Monitoring List of drives and percentage limits that Watchdog will monitor Service Monitoring List of Services that Watchdog should monitor Services that are not checked will not be pinged by Watchdog 1 5 DICOM Server Service DICOM Store SCP The ProSolv DICOM Server is intended to receive and convert DICOM images sent over a network from DICOM devices Studies received by the DICO
56. ate 02 06 2009 Page 56 of 83 Configuration Setting Recommended Value Are older than Archive studies that are older than a designated number of days Are smaller than Archive studies that are smaller than a designated size No more than Set a maximum number of studies that can be archived Archiver Configuration Wizard ioj x ProSolv CardioVascular FUJIFILM What types of studies should be arched as Po New Load types from this file E Cancel Finish Previous E Configuration Setting Recommended Value Left list box Lists the current study types that will be archived Right text box Designate a new study type to be archived Click New to add the new type Load types from this file Select a predefined type s file to load the study types from THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 57 of 83 Archiver Configuration Wizard Ioj x Archive the studies from which locabons pee Load types from thie file
57. ator s Manual FUJIFILM Page 68 of 83 ProSoly Service Manager ell xj 2 Click the Mennen DT tab Al Services Archiver ProSolv Server License Server Watchdog Compression ADT DICOM OR Measurement Importer Mennen DT Report Messenger Retriever Store SCU Service Control Mennen DT Settings Input Folder Ir Pi Archive Folder FTO EN Date Time Format String dmm o ProSoly Database Settings Log Entries Most recent log entries Show Everything DI View Log Date Time Level Text Configuration Setting Recommended Value Input Folder Directory location to monitor for input data files Archive Folder Directory location to store input data files after processing Date Time Format String Date Time format to use when add data to the database ProSolv Database Connection string pointing to the Settings ProSolv database 1 16 HL7 ADT Service The Admission Discharge Transfer ADT Service accepts and subsequently parses HL7 messages The ADT Service retrieves information necessary to enable an efficient patient data and image management solution The ADT Service is multithreaded and is capable of running multiple tasks that can serve multiple TCP input ports THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERR
58. cted in the window before configuring THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual e tve Date 92 06 2009 Page 27 of 83 Select Siemens Hemo Importer ProSolv CardioVascular FUJIFILM Ey Importer Settings Flugin Type siemens Hemolmpoter O Database Sethngs Images Directory ProSoly Database Connection String E File Importer Settings Search Pattern Input Directory Monitor Subdirectones E Importer Settings Machine Name Sleep Interval Hachine Hame Descriptive name to use when referencing this importer Dave Cancel Configuration Setting Recommended Value Plugin Type Drop down list of Supported importers Study images are located Connection string pointing to the String ProSolv database Filter pattern of files to import Siemens his SSS Input Directory Directory location where the hemodynamic output files will be deposited by the OEM and then processed Monitor Subdirectories False will not process files in subdirectories True not supported Machine Name U
59. cular Server Administrator s Manual FUJIFILM Page 70 of 83 BER E ADT Delimiter Settings HL Delimiters el E ADT Settings Create Study False Images Directory Listening Port 3333 Messsage Store Directory Patient ID as Main ID False Store HL Messages False Study UID Prefix 1 2 826 0 1 3680043 2 1331 3175947431 Task Name E Database Settings ADT Database Connection String i DICOM Database Connection String ProSoly Database Connection String ADT Database Connection Sting Connection string for the ADT database Configuration Setting Recommended Value HL7 Delimiters List of delimiters to use for the HL7 messages Create Study Boolean flag indicating whether the service should create ProSolv studies in the database should be used as the main identifier messages should be written to log files Study UID Prefix The prefix to use for Study UIDs Unique name to describe this ADT task ADT Database Connection string to the ADT database Connection String by default included into the ProSolv database Make sure the password is included in the connection string DICOM Database Connection string to the DICOM Connection String database Make sure the password is included in the connection string ProSolv Database Connection string to the ProSolv Connection String database Make sure the password is included in the connection string 5 Click the Save button to add the task aft
60. e xml file is provided here Web Config lt xml version 1 0 gt Cl lt configuration gt L lt configSections gt lt section name FileLocations type System Configuration IgnoreS SectionHandler gt lt configSections gt lt appSettings gt lt add key pCodeServer value 127 0 0 1 gt lt add key pCodeServerPort value 704 gt lt add key WebLab eTitle value ProSolvWebLab gt lt appSettings gt lt connectionStrings gt lt add name ProsolvDB connectionString Provider SQLOLEDB 1 Password psci Persist Security Info Tr providerName System Data OleDb gt B lt lt add name ProsolvDB connectionString Provider Microsoft Jet OLEDB 4 0 Data source c Testing New Folder db1i Password md providerName System Data OleDh gt gt lt connectionStrings gt lt FileLocations gt lt FileLocation name images location Image Path gt lt FileLocation name updates2 location Updates Path gt lt FileLocation name HL7 Reports location c HL7Reports gt lt FileLocation name gemeasurement transfer location c GE_ Measurement gt lt FileLocation name Fax_server location c FaxServer gt lt FileLocations gt lt system web gt lt compilation defaultLanguage c debug true gt lt authentication mode Windows gt lt httpRuntime maxRequestLength 16384 gt lt customErrors mode O0ff gt lt system web gt lt fconfiguration gt
61. ectory path to the file share resource NOTE Remote files shares must be designated as an UNC path THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 52 of 83 Archiver Configuration Wizard Ioj x Configure Prosoly 5 erver IP Address PO Port In User FO Password Po Root Fath Po Cancel Finish Previous na E Configuration Setting Recommended Value IP Address IP Address or host name of the ProSolv Server Port Port number that the ProSolv Server is running Password ProSolv Server user account password to use when archiving Root Path Root file path of where to store the images on the ProSolv Server User ProSolv Server user account to use when archiving THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING
62. eeeeeceeeeeeeeeeeeeseeeeeeseeeeeesaeeeeeeeeeeesaeeeeesaeees 67 EARD ee ON UIT E 67 Dt WT eine EE 68 Li SOMME UIE UO DEE 69 1 17 HL7 Report Messenger Gernice 71 Tea lin O UU ON A E E E E A E A A 71 2 DICOM Modality Worklist Application cccccccccceccccccsseeceeeeseeeceeeeeeeeceeeseeeeceesseeeeeeeeseeseeeessseeeeeeeas 74 2 1 GONDO O EE 74 2 2 Starting the ProSolv DICOM MWL Applceatnon 76 2 3 Exiting the ProSolv DICOM MWL Application cccccccesseeceeeeeseeeeeeeeeeeeeeeseeeeeeeseeeeeeeeeeas 77 3 DICOM Structured Reporting Application cccccccccccccssesceeceeseceeeeeeeeeeeesseeeceeesseeeeeesseeeeeeesssageeeeeeas 77 3 1 GOTO de ME 77 THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 6 of 83 4 Witt Hemodynamic Import Appltcaton 80 4 1 NTS CATT EE 80 80 4 2 OTN MUON aco sac shee ia Sees epncaac tea A TEAN EEEE E EEE E A EE E EAEEREN THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CAR
63. el Name Description Recommended Values Database Settings ProSolv Database Connection string to the ProSolv Provider SQLOLEDB 1 Password psci Per Connection String database Click the button to sist Security Info True User bring up the Connection editor ID prosolv Initial Catalog Prosolv Data Source localhost Image Directory Path to the Images Folder lt ImageStore gt psdata File Importer Settings Search Pattern File pattern that the importer is going xml to be watching for Input Directory UNC path to the CardioServer export lt CardioServer Server IP gt cs hI7 xml folder Monitor An option to enable monitoring of False Subdirectories subdirectories The default settings of the CardioServer Server does not allow for exporting files to subdirectories Importer Settings importer task the Input Directory THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 15 of 83 Name Description Recommended Values OpenECG Settings Root CardioServer
64. elect it in the list and click the Settings button 1 7 DICOM Store SCU Service The Store SCU Service monitors an input folder for DICOM images and sends them to a designated Store SCP device The service is multithreaded and is capable of running THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 44 of 83 multiple tasks that can monitor several input folders and forward images to several SCP devices 1 7 1 Configuration To configure the DICOM Store SCU Service 1 Start the ProSolv Service Manager 2 Click the Store SCU tab 3 To configure the service it must be running started Click the Green Start button in the Service Control group box to start the service 4 To adda new Task click the Add button This opens the Store Task Settings screen Store Task Settings ll x E Input Folder Settings Input Folder Search Pattern E Store SCP Settings SCP AE Title PROSOLY SCP Port Number 104 SCP Server E Store SCU Settings True ma Lookup Interval 200 Read Time
65. enger Retriever Store scu All Services Archiver ProSoly Server License Server Watchdog ADT DICOM Op Measurement Importer Service Control Status Started Settings Log Entries Most recent log entries Show Everything D View Log Date Time Level Text 3 To add a new importer task click the Add button This opens the Importer settings screen and allows the user to configure the setting for the importer To start the importer click the Green Start button To stop the importer click the Red Stop button To restart the importer click the Restart button To remove an importer select it in the list and click the Delete button 8 To modify an importer s settings select it in the list and click the Settings button y Of gt THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 65 of 83 1 12 HPDT Importer Importer Settings d Ki Plugin Type HPDT il HI Database Settings ProSoly Database Connection String Images Directory EI Importer Setti
66. ent For example if value10 is specified in this field then only the last 10 studies for a patient will be retrieved 4 Click Apply to save the configuration 5 Click Finish to close the configuration dialog 1 10 2 Retrieve Queues The Retriever Tab in the Service Manager displays status on its current tasks The external Requests queue lists all the requests that need to be retrieved from external PACS based on Patients MRNs The Internal Requests queue lists all the retrieve requests based on Study UIDs THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B i ProSolv CardioVascular 4 0 2 Prosolv CardioVascular Server Administrator s Manual FUJIFILM Page 63 of 83 E7 ProSoly Service Manager OF x All Services Archiver ProSolv Server License Server DICOM Server Watchdog Compression Retriever Service Control Status Stopped gt EH Log level Wamings D External Requests Internal Requests Requesting Source MAN Status Requested Time Processed Time AETitles Searched ADT 3988763 Waiting 3 4 2008 6 56 PM TT O Refresh Retry Delete Conf
67. er it has been configured THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 71 of 83 6 To start the newly configured task click the Green Start button to the right of the Task List 7 To stop the task click the Red Stop button 8 To restart the task click the Restart button 9 To remove a task select it in the list and click the Delete button 10 To modify a task s settings select it in the list and click the Settings button 1 17 HL7 Report Messenger Service The Report Messenger Service sends HL7 formatted messages from a designated folder to the designated receiving system The Report Messenger Service is multithreaded and is capable of running multiple tasks that can monitor several input folders and forward to several receiving systems 1 17 1 Configuration To configure the HL7 Report Messenger Service 1 Start the ProSolv Service Manager 2 Click the Report Messenger tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING C
68. ess click the Restart button 9 To remove a process select it in the list and click the Delete button 10 To modify a process settings select it in the list and click the Settings button 11 To disable a process from being spawned and pinged by Watchdog click the Disable button To configure the Server Service Non Service Manager 1 Server Files Navigate to the Services 4 0 apps server directory and modify deletepaths txt to reference the locations of the deletedfiles mdb database and the deletetemp directory This is typically under the prosolv database directory E deletepaths txt Notepad Sel File Edit Format View Help 4Prosoly Database deletedti les mdb CryProsoly Database deletetemp EMD 2 Deleted Files Navigate to the ProSolv Database directory referenced above and verify the users group has full permissions to the deletedfiles mdb database and the deletetemp folder THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B i ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 33 of 83 1 3 License Server The
69. ests from the different modalities This line indicates whether the ProSolv DMWL application should generate and send the study UID to the modality in response to the study order This line indicates the number of AE titles following this line if there are 2 acceptable AE titles that this application should use enter 2 in this line This line indicates the acceptable AE titles for this application The application will refuse associations to modalities with AE titles other than those specified in these lines This line indicates the type of the ADT database used by this application Printable ASCII characters that are supported by the Operating System PROSOLVDMWL 1 65535 1 infinity Printable ASCII characters that are supported by the Operating System SQLServer or Access using this AE title would allow associations from all possible clients SQLServer THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascul CardioVascular Server Administrator s Manual FUJIFILM Line Explanation Possible Values Numbers
70. f the Dicom Server SCP that you gave to PACS the SCP that PACS sends the images to during its retrieval process 1 10 Retriever Service The Retriever Service retrieves ProSolv Studies that have been archived from an external PACS 1 10 1 Configuration To configure the Retriever Service 1 ADT needs to be configured to send MRN based requests to retrieve studies from external PACS Please refer to the ADT configuration section for details on how to configure ADT to send requests to the Retriever 2 Goto Archiver tab Click Add Job Select Retriever and click Next THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EES FUJIFILM Page 61 of 83 4 Archiver Configuration Wizard l O Please select type of configurator Archive Ce Retriever Cancel Finish E 3 Please use Manage DICOM Components as described in a previous section a Store SCP b Query Retrieve SCP E7 Archiver Configuration Wizard l DU x Retiever Store AE Tithe tees TE SCP
71. files Read Timeout Network read timeout in milliseconds Oo Va SCU AE Title The AE Title of this Store SCU Task 2 a Storage Commitment Boolean flag indicating whether or not to include Storage Commitment on the same association Storage Commitment Port Port number to use for Storage Commitment Task Name Unique name used to describe this Store SCU Task Write Timeout Network writes timeout in milliseconds Click the Save button to add the task after it has been configured 6 To start the newly configured task click the Green Start button to the right of the Task List 7 To stop the task click the Red Stop button 8 To restart the task click the Restart button 9 To remove a task select it in the list and click the Delete button 10 To modify a task s settings select it in the list and click the Settings button 1 8 Image Compression Service The Image Compression Service creates compressed versions of images in ProSolv Studies It monitors the file system for the creation of new ProSolv Studies When it detects a new study it processes the contents of the study image folder and creates compressed versions of the images files Note Image Compression Service currently only compresses loops and only from modalities XA Catheterization or US Ultrasound 1 8 1 Configuration To configure the Image Compression Service 1 Start the ProSolv Service Manager 2 Click the Compression tab THIS DOCU
72. g Recommended Value ProSolv Database Connection string pointing to the Connection String ProSolv database Image Directory Directory location where the ProSolv Study images are located Search Pattern Filter pattern of files to import will include all files Input Directory Directory location where the Epiphany XML output files will be located Monitor Subdirectories Boolean indicating if the service should monitor subdirectories for input files Unique name to describe this importer Sleep Interval Amount of time to sleep in between checks for new data THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM 1 14 Burdick Marquette Quinton Treadmill Importers Importer Settings k x Plugin Type ame EAE HI Database Settings Images Directory ProSoly Database Connection String HI File Importer Settings Search Pattern Input Directory Monitor Subdirectories False EI Importer Settings Machine Name Sleep Interval H Treadmill Settings PDF Off False Machine Name Descriptive na
73. h dialog Viewer Preterences Load Options Save Cancel 2 Click on the Machine Settings button on the Preferences window 3 On the Machine Settings Window check the Enable OpenECG features checkbox and enter the URL of the OpenECG Web THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 13 of 83 Interface Note Typically this is http lt ECG Management Server Machine Name gt W Enable GpenECG features Server URL HTML http openecat d A short cut to the OpenECG URL needs to be placed on the desktop of all workstations The short cut needs to point to the Server URL listed above in step 3 5 To configure client components the Measurement Importer ADT and Report Messenger Plug ins must be installed ECG Management Module DT Service Configuration The ECG Management Server will export an XML and a PDF report whenever a study is confirmed within the Web Interface These files are stored on file shares on the ECG Management Server
74. iguration and Search Database Settings MRN Equals Search m Log Entries Most recent log entries Show Minimal View Log RS Date Time Level Text 1 11 Measurement Importer Service The Measurement Importer Service imports measurements from several different medical devices The service consists of a set of plugins Each plugin can handle importing a measurement from a different type of device e g HPDT Burdick etc The General Measurement Import Service is multithreaded and can run each plugin in a separate thread All US measurement interfaces not supported by the Measurement Importer Service are supported by stand alone applications such as Toshiba ATL and Cypress 1 11 1 Configuration To configure the Measurement Importer Service 1 Start the ProSolv Service Manager 2 Click the Measurement Importer tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual Effective Date 02 06 2009 FUJIFILM Page 64 of 83 ProSoly Service Manager ell sl Report Mess
75. ile Number of MB to set for the maximum stream file size Size Message File Age Number of days to keep incoming files before deleting 30 Limit them THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 17 of 83 Name _ Deseripton TI Recommended Values ADT Delimiter Settings Message Store Path to the location of the incoming HL7 messages lt AppServer gt ProSolv Directory database HL7ADT Patient ID as Main Boolean for setting Patient ID as the main identifier Process Message Boolean for setting whether the message needs to be processed Setting this to False results in just storing messages required by the HIS system Messages False will not create individual message files Study UID Prefix Study UID Prefix 1 2 826 0 1 38680043 2 1331 317594 1481 Task Name Database Settings ADT Database Connection string to the ProSolv database Clickthe Provider SQLOLEDB 1 Password p Connection String 1 button to bring up the Connection editor sci Persist Security Info True User ID
76. ions Timeout ming Configuration Setting Recommended Value Click this button to define a new device Delete Select the component to delete then click this button to delete the selected device Modify Select the component to modify in the list then click this button to modify the selected device Save Click this button to save the newly defined device Name Unique name to describe this DICOM device IP Address IP Address or hostname where the device resides Port number that the device is running Service Type Available options here are QR SCP Store SCU or Store SCP THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 55 of 83 Configuration Setting Recommended Value Maximum Connections The Maximum number of connections allowed on this device not required Timeout Network read write timeout values not required Archiver Configuration Wizard l Ioj x ProSolv CardioVascular FUJIFILM Which studies should be arched All studies Ce Only studies that T ha
77. l FUJIFILM Page 24 of 83 4 Make sure that MacLab Hemo Importer is selected in the window before configuring Select MacLab Hemo Importer E Im porter Settings Plugin Type El Database Sethngs ProSoly Database Connection String Provider H5DA45QL 1 Persist Securty Info False Images Directory C lmages E File Importer Settings Input Directory C GEMacLabin Search Fatter hi Monitor Subdirectories True E Importer Settings Machine Marne Mac ab Hemo Demo Sleep Interval 30 Input Directoy Directory to montor for input files Name Description Recommended Values Database Settings Images Directory Path to the Images Folder I Images ProSolv Connection string to the Provider SQLOLEDB 1 Database ProSolv database Click the Password psci Connection button to bring up the Persist Security Info True String Connection editor User ID prosolv Initial Catalog Prosolv Data Source localhost File Importer Settings Input Directory Path to the folder where the I GEMacLabIn Input hemodynamic output files will be deposited by MacLab Search Pattern Filter pattern of files to import h7 MacLab h7 This allows THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONS
78. le Log Entries Most recent log entries Show Everything DI View Log Date Time Level Text 3 To add anew ProSolv Server process click the Add button This will bring up the Add ProSolv Server screen Alternatively select existing ProSolv Server and click Settings 4 There needs to be a minimum of two entries under ProSolv Server gt Databases list NOTE USERLOG and server are case sensitive USERLOG should be all uppercase and server must be all lowercase a USERLOG for user activity auditing THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B PMAN 0014756 A ProSolv CardioVascular 4 0 2 ProSolv CardioVascular s ver administrators M Server Administrator s Manual C Ve Pate 92 06 2008 FUJIFILM Page 30 of 83 5 x Ey Modify Database Properties Type access Database sl E Database Sethngs USERLOG Path Password Images Path Option Type W ProSoly Carndio ascular Services 4 05 Prosol VKALPSCISTAPSDATA Prosoly Name Examples ProSoly Database server Userlog USERLOG Maclab MacLabDT This is case sensative OF Cancel
79. ling and Result messages from OpenEGG to HIS This approach will consolidate and improve the system integrations All ProSolv users that are going to use the integration must have their ProSolv User Names added to the user list that is tied to the OpenECG server The User Name is used for authentication between the two systems and hence must always match Passwords on the other hand are not used for authentication and they don t have to match To configure the OpenECG integration 1 Select Tools then Preferences from the ProSolv menu Preferences Configuration Regional wall In Report Patient Search Significant Digits 3 Databases Bull s Eye Pattern e One line per unread exam f l l l Login Timeoutimin 5 Drive Paths e FourMiew Schematics Une line per patient E Minimum Synapse Set Date Format Remote Reading Match To Show a Automatic Cleanup Password Config Text Editar jc program files windows NT accessones wordpad exe Browse Quads Auto Patient Search Large Patient Search screen Quad Synchronization w Auto Refresh Analysis Log Display Thumbnail Labels Dn w Show Patient Records on login W Use Direct Draw On Ww Do not mark Apes as normal on Aw Suppress Save Report Dialog NEE If no report load from database Do not display images seconds in Quads Dm Use Point To Point For Traces Autoload Measurements Ke Auto Load First Image Always show measurement matc
80. lv CardioVascular Core Server components Mini server Workstation providing both client and server applications For more information refer to the Hardware Requirements document document control PSUP 0013745 B Disclaimer No part of this manual may be reproduced without prior written permission The information contained in this manual may be subject to change without prior notice ProSolv CardioVascular A FUJIFILM Company shall not be liable for malfunctions and damages resulting from improper installation relocation remodeling maintenance and repair performed by ProSolv CardioVascular A FUJIFILM Company and all other FUJIFILM Subsidiaries or Distributors ProSolv CardioVascular A FUJIFILM Company shall not be liable for malfunctions and damages of ProSolv CardioVascular A FUJIFILM Company products due to products of other manufacturers not supplied by ProSolv CardioVascular A FUJIFILM Company ProSolv CardioVascular A FUJIFILM Company shall not be liable for malfunctions and damages resulting from remodeling maintenance and repair using repair parts other than those specified by ProSolv CardioVascular A FUJIFILM Company ProSolv CardioVascular A FUJIFILM Company shall not be liable for malfunctions and damages resulting from negligence ProSolv CardioVascular A FUJIFILM Company shall not be liable for malfunctions and damages resulting from natural disasters ProSolv CardioVascular A FUJIF
81. me to use when referencing this importer Save Cancel Configuration Setting Recommended Value ProSolv Database Connection string pointing to the Connection String ProSolv database Image Directory Directory location where the ProSolv Study images are located Search Pattern Filter pattern of files to import will include all files Input Directory Directory location where the Epiphany XML output files will be located Monitor Subdirectories Boolean indicating if the service should monitor subdirectories for input files Machine Name Unique name to describe this importer task Sleep Interval Amount of time to sleep in between checks for new input files PDF Off Boolean flag indicating if PDF Conversion should be turned on off 1 15 Mennen Hemodynamic Import Service The Mennen Service imports measurements from the Mennen Hemodynamic device 1 15 1 Configuration To configure the Mennen Hemodynamic Import Service 1 Start the ProSolv Service Manager THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administr
82. mended Value FTP Server The study will be archived to a designated FTP server Local or Network Drive The study will be archived to a designated local or remote file share ProSolv Server The study will be archived to another ProSolv Server DICOM PACS The study will be archived to a designated DICOM PACS Archiver Configuration Wizard Ioj x Configure FTP Access IP Address PO User FO Password FO Root Path FO Cancel Finish Previous na d THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 51 of 83 Recommended Value Root Path The root path of where to upload the images on the FTP server Archiver Configuration Wizard loj x ProSolv CardioVascular FUJIFILM Configure Hard Drive Network Server Archive label Root Path Example 5erverl Path Simages Cancel Finish Previous d Configuration Setting Recommended Value Archive Label Unique name used to label this file share Root Path Dir
83. mpany names are trademarks or registered trademarks of their respective companies or organizations Safety Information DANGER WARNING and CAUTION Safety Statements The DANGERS WARNINGS and CAUTIONS that appear throughout this manual are not only preventative measures designed to uphold the safety of both the service person and operator but also enhance equipment reliability The definitions for DANGER WARNING and CAUTION comply with ANSI 2535 2 American National Standard for Environmental and Facility Safety Signs and ANSI 2535 4 Product Safety Signs and Labels issued by the American National Standards Institute DANGER indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING indicates an imminently hazardous situation which if not avoided could result in death or serious injury CAUTION indicates a potentially hazardous situation which if not avoided may result in minor to moderate injury It may also be used to alert against unsafe practices Important indicates an emphasized note It is something you should be particularly aware of something not readily apparent Important is typically used to prevent equipment damage Note indicates a note It is something you should be particularly aware of something not readily apparent A note is typically used as a suggestion Electrical Power Do not perform service with equipment energized Always de energize the equipment
84. ng Notepad for editing 4 Update configuration values as follows a ImageDir location of image files D ImageDir MachineName host name of the server running the Compression service if cluster use virtual host name defined in ProSolv Apps cluster group THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EES FUJIFILM Page 47 of 83 E ImageCompression exe Notepad File Edit Format View Help ke xml version 1 0 encoding utt 8 gt lt configuration gt lt configoections gt lt section name compressianCantigs type System Configuration lgnaresectionHandler gt canfigsections gt app settings lt add key RemotePort value 99980 gt lt add key imageDir value Dsilmaqges gt lt fappoettings gt lt connectionstrings gt lt add name ProsolDB connectionstring Providersqloledb uid prosoly pwd psci data source localhast initial catalog Prosol providerName System Data OleDb gt lt connectionstrings gt lt compressianCanfigs gt lt compressianCanfig machineN
85. ngs Sleep Interval Machine Name E Port Importer Settings Port Number 1 Machine Name Descriptive name to use when referencing this importer Configuration Setting Recommended Value ProSolv Database Connection string pointing to the Connection String ProSolv database Image Directory Directory location where the ProSolv Study images are located Sleep Interval Amount of time to sleep in between checks for new input files Machine Name Unique name to describe this importer task Port Number TCP Port number to listen on for input data THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM 1 13 HD11XE Importer Importer Settings l x Plugin Type apanta aE E Database Settings Images Directory ProSoly Database Connection String File Importer Settings Search Pattern Input Directory Monitor Subdirectories False EI Importer Settings Machine Name Sleep Interval Machine Name Descriptive name to use when referencing this importer Configuration Settin
86. nique name to describe this importer task Sleep Interval Amount of time to sleep in between checks for new input files 5 After completing the entries hit lt Save gt select the entry and start the service THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e Prosolv CardioVascular Server Administrator s Manual FUJIFILM Page 28 of 83 Note Configure the above per your specific system The Siemens Sensis system will also need to be configured so that it deposits files in the Input Directory that is specified in this configuration step Note The screen below displays the recommended configuration values E7 Importer Settings Plugin Type Siemens Hema Importer sl E Database Sethngs Images Directory I images ProSoly Database Connection String Provider SQLOLEDB 1 Password psci Persist Security El File Importer Settings Search Pattern his Input Directory cAppS erver 4Siemensinpull Monitor Subdirectones False CT Importer Settings Machine Mame Siemens Sleep Interval 15 Input Directoy Directory to monitor for input files 1 2 ProSolv
87. onents which include but not limited to fuses circuit breakers interlock switches equipment panels and covers Do not disable alter modify or remove safety devices when performing service Confidential and Proprietary Information Use duplication or distribution of this document without permission from ProSolv CardioVascular A FUJIFILM Company is strictly prohibited THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascul CardioVascular Server Administrator s Manual FUJIFILM Page 5 of 83 Table of Contents Te FTI Ee AG GI E 7 1 1 Iesele cessor cen prceces E E E SA 8 1 1 1 WebLab Integration Confguraton 8 1 1 2 Synapse Integration Configuration cccccccceeeeeceeceeeeeeeseaeeeeeeseaeeeeeeeeeeeeeeessseeeeeessaegeeeeeeas 9 NC OpenECG Integration Contguraion 11 1 1 4 MacLab Integration Configuration cccccccccssecccccseeceeeeeeeeceeeseeeeeeeeseeseeeestaaneceessaaeeeees 22 1 1 5 Siemens Sensis Integration Configuration ccccccecccecceeeeceeeeaeeeceeeesueeseeseseeeeeeneaseeeeeeas 25 1 2 ProSOlV TWEET ww M
88. onnection String Task Name Unique name to describe this ADT task UA False I images 33353 32 30 I Epiphany FromCardioServer False True True False 1 2 826 0 1 3680043 2 1331 31 75947431 HL Output Listener Provider SQLOLED6 1 Password psci Persist Security Info Provider SQLOLEDE 1 Password psci Persist Securty Infe Provider SQLOLEDE 1 Password psci Persist Security Infe THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 20 of 83 Name _ Description Recommended Values ADT Delimiter Settings HL7 Delimiters HL7 Delimiters can be added as needed ADT Settings getting created in the system for each order Max Stream File Number of MB to set for the maximum stream file size Size Message File Age Number of days to keep incoming files before deleting 30 Limit them Directory Database Epiphany From CardioServer ID Process Message Boolean for setting whether the message needs to be True eel processed Setting this to False results in just storing messages req
89. ot restart the service after 30 minutes it flags the service to indicate that the service cannot be restarted Watchdog can also monitor drive space and send an alert email if free space falls below a certain percentage The Watchdog Tab in the Service Manager shows the list of services that are currently being monitored It displays an icon next to each task The following icons can be displayed e Green Check Icon indicates that the Service is running Yellow X Icon indicates that the Service has failed a ping test and that Watchdog is trying to restart the service e Red X Icon indicates that Watchdog was not able to restart the service after 30 minutes 1 4 1 Configuration To configure the Watchdog Service 1 Start the ProSolv Service Manager 2 Click the Watchdog tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular e nas asocuar Server Administrator s Manual SCT Date 02 06 2009 FUJIFILM Page 36 of 83 Prosoty service manager sl Report Messenger Store SCU All Services Archiver ProSoly Server License
90. out 40 SCU AE Title PROSOLYSCU Storage Commitment False Storage Commitment Port 104 Task Name Write Timeout 200 Include SR Include SF related SOP s on the same association Cancel ay Configuration Setting Recommended Value Input Folder Directory that this task should monitor for input DICOM images Search Pattern Search filter to reduce the file types that the task should monitor will include all files SCP AE Title The AE Title of the Store SCP device that is to receive the images SCP Port Number The port number of the Store SCP device that is to receive the images SCP Server The machine name or IP address of the Store SCP device that is to receive the images THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual EE EE FUJIFILM Page 45 of 83 Configuration Setting Recommended Value Include SR Boolean flag that indicates if SR related SOP s are included on the association Lookup Interval Amount of time in seconds that the task will sleep in between checks for new image
91. r database and the user specified in line 9 Access Line Explanation number The full path to the folder that contains the DICOM SR files that are processed by this application How often DICOM SR application will check the directory for new DICOM SR objects milliseconds The type of ProSolv database where retrieved file data is stored The full path to the mdb file on the ProSolv database The password associated with the ProSolv database The full path to the mdb file on the DICOM Server database THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC The type of DICOM Server database Printable ASCII characters that are supported by the Operating System prosolv sa Windows C SRFolder formatted file on ending with Printable ASCII characters that are supported by the Operating System symbol to eu all files in that folder directory path should be valid and existing 5000 5000 SQLServer or Access Access Windows C DCMServ prosolv mdb formatted file path Printable ASCII stick characters that are supported by the Operating System SQLServer or Access Windows formatted file path Access C DCMServ dcmserv mdb D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING
92. server files before deleting them Output Folder Path to the folder where the input files lt AppServer gt ProSolv are stored after they have been Database Epiphany From CardioServer processed Note Use recommended values unless Site requests otherwise HL7 Message Routing Configuration All HL7 messages are routed through the ProSolv Server to from the CardioServer The ProSolv Report Messenger and ProSolv ADT Services are used in tandem to perform the message forwarding ProSolv Server Report Incoming ADT Listening ae Messenger West Message Task ncoming Forward Task Directory CardioServer Outgoing Report ADT Listening Message Messenger Outgoing HL7 ProSolv Database HIS Task Directory Forward Task This ADT task will have the Process Message flag set to FALSE It will only pass thru messages HL7 Message Routing Configuration The ADT service needs to be configured to receive HL7 messages from the HIS on some predefined port number This task will be responsible THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITH
93. supported by the Operating System ProSolvWittDB prosolv sa Printable ASCII characters that are supported by the Operating System Line Explanation Possible Values number The type of Witt database where retrieved file data is stored The full path to the Witt mdb Windows formatted file lt AppServer gt ProSolv file path Database Witt mdb The full path to the folder that contains the Witt files that are path ending with a processed by this application lt AppServer gt From Witt Windows formatted file symbol to indicate all files in that folder directory path should be valid and existing SQLServer or Access Access THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 82 of 83 Line Explanation Possible Values number The full path to the images Printable ASCII lt AppServer gt ProSolv folder of the ProSolv characters that are Database images database supported by the Operating System 4 Open the text file rt2f ini from the installation
94. ttons to the right of the table 4 To adda setting click the Add button to the right of the Synapse Settings table D At the Synapse Settings dialog box enter the new information THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM amp Synapse Settings Settings Synapse Server Settings AE Title Host Port Number Synapse URL Description Name The user friendly name for this Synapse connection visible in ProSoly CardioVascular Client Save Cancel 6 Click the Save button to accept the changes 7 To edit a setting click the Settings button to the right of the Synapse settings table and enter the updated information 8 Click the Save button to accept the changes 9 To delete a setting click the Delete button to the right of the Synapse Seitings table 10 At the confirmation dialog box confirm or cancel the operation Click the Yes button to delete all related data from the database and refresh the Optional Save Directory Settings table Click the No button to cancel the change
95. tus Select Add I cL cL Delete Settings Log Entries Most recent log entries Show Everything Ww Dates T ime Level Text 9 27 2008 9 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 22 2008 8 50 AM 9 19 2008 3 33 PM QA ANA 23 233 DM The Service has started Assembly rot oy Measurement mportshL Treadmill Quinton adc Assembly ProSoly Measurement mportshL Treadmill Marquette z Assembly ProSoly Measurement mport ehML TT readrll Burdick adc Assembly Posch Hemalmporter Siemens added to pluginList Assembly Prosoly Hemolmporter Mac ab added to pluginList Starting Serice Measurement Import Init ServiceB ase args ig System Stringi Trying to start the service again Object creation succeeded fram aecormbls Preacnhs Mea mermerblrorert akart see Py Pry ya 3 Select lt Add gt to configure a new MacLab Importer THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manua
96. uired by the HIS system Messages False will not create individual message files 7431 ADT Database Connection string to the ProSolv database Click the Provider SQLOLEDB 1 Password p Connection String button to bring up the Connection editor sci Persist Security Info True User ID prosolv Initial Catalog Prosolv Data Source lt sqldb gt DICOM Database Connection string to the ProSolv database Click the Provider SQLOLEDB 1 Password p Connection String button to bring up the Connection editor sci Persist Security Info True User ID prosolv Initial Catalog Prosolv Data Source lt sqldb gt ProSolv Database Connection string to the ProSolv database Click the Provider SQLOLEDB 1 Password p Connection String button to bring up the Connection editor sci Persist Security Info True User ID prosolv Initial Catalog Prosolv Data Source lt sqldb gt ProSolv CardioVascular FUJIFILM Note Use recommended values unless Site requests otherwise Report Messenger Forward Task Configuration A Report Messenger Forward Task needs to be configured to monitor the folder that the ADT Task stores messages The Report Messenger THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCE
97. ular 4 0 2 ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 18 of 83 4 Report Messenger Settings Machine Name Cadet ermer Incoming Forwarder Seconds To Sleep 5 Directory Information watch Directory lU Search Pattern FHF Hecurse Directory BR IP address or Machine Name of the OpenECG Server Socket Info IPAdd HostName jopenecal Watch Port 2575 Port that the OpenECG server is listening on Wat Tine rz i TENI Mum Wat Cycles E Persistent File Processing Remove LF Seg Delim ba Delimiters MSHI A Configure according to incoming HL7 messages Sending facility Send By ey ABC application Return AT E E Separtabe with 7 Save Cancel Name Description Recommended Values CardioServer Incoming Forwarder 5 Seconds to Sleep Number of seconds between polls of the Input Directory Watch Directory Directory to watch for input files lt AppServer gt ProSolv Database HL7ADT Search Pattern Pattern to search on Recurse Directory Option to check subdirectories Unchecked IP Add Host Name lt IP OpenECG Server gt THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFIL
98. upplied by PACS Port 104 PACS Store SCP Port Number is supplied by PACS THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve Date 02 06 2009 Page 60 of 83 Configuration Setting Description Recommended Value Service Type Choose STORESCP Maximum Connections 2 consult with PACS if they have a limit of connections Timeout 3 minutes ProSolv CardioVascular FUJIFILM QRore SCP Component Configuration Setting Description Recommended Value Name PACS Store SCP IP Address XX XXX XXX XXX PACS Store SCP IP is supplied by PACS AETitle PACSAE PACS Store SCP AE Title is supplied by PACS Port 104 PACS Store SCP Port Number is supplied by PACS Service Type Choose QRSCP Maximum Connections 2 consult with PACS if they have a limit of connections Timeout 3 minutes After setting up these components they must be selected from the drop down list boxes on the Configure Dicom Access dialog If you have only setup these three they should already be selected For the Move AE Title enter the AE Title o
99. ve preliminary reports were read at a specific location T have finalized reports had a specific reader are of a specific type were performed after 171472008 sl are older than E A Ja z M are smaller than 200 megabytes h No more than 500 total studies Cancel Finish Previous nee E Configuration Setting Recommended Value All Studies Check this option to archive all of the ProSolv studies Only studies that Check this option to define a set of criteria that will determine which studies should be archived Have preliminary reports Archive studies that have preliminary reports Have finalized reports Archive studies that have finalized reports Are of a specific type Archive studies that are of a specific type Were read at a specific Archive studies that were read at a location specific location Had a specific reader Archive studies that had a specific reader Were performed after Archive studies that were performed after a designated date THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual ct ve D
100. when performing service Even when de energized equipment components can retain an electrical charge capable of causing electric shock Always use caution and common sense when performing service When installing equipment electrical wiring must comply with the local electrical codes in your area governing the installation of electronic equipment Heavy Objects Heavy objects are objects that weigh 50 pounds or more or are objects that are too large or bulky to be THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 e ProSolv CardioVascular Server Administrator s Manual FUJIFILM Page 4 of 83 moved or handled by one service person Installing removing or servicing heavy objects may require two or more service personnel When installing removing or servicing heavy objects always use suitable gear such as heavy gloves and eye protection Always uses the appropriate support machinery when installing removing or moving heavy objects Always use caution and common sense when installing removing or servicing heavy objects Safety Devices Safety devices are protection comp
101. wo formats 1 Deposited directly in the input folder with the names matching the h7 and ending with snapxxxx tif a 0a9ac3e9 0981 41cc aa0b 939d5d8c8acdsnap0001 tif b 0a9ac3e9 0981 41cc aa0b 939d5d8c8acd h7c 2 A new subdirectory holding the data and tiff files put into the input directory with names that may not match These image tiff files are put into the studies image directory for viewing when a match takes place at the client To configure the MacLab integration 1 Start the ProSolv Service Manager 2 Select the Measurement Importer tab THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY AND SHALL NOT BE DUPLICATED OR DISTRIBUTED OR OTHERWISE TRANSFERRED TOA THIRD PARTY WITHOUT THE EXPRESS WRITTEN CONSENT OF PROBLEM SOLVING CONCEPTS INC D B A PROSOLV CARDIOVASCULAR A FUJIFILM COMPANY TEMPLATE FORM NUMBER PREG 0013455 B ProSolv CardioVascular 4 0 2 Server Administrator s Manual amp Ve Date 92 06 2009 Page 23 of 83 All Services Archiver ProSaly Server License Server DICOM Server Watchdog Compression er gt i Measurement Importer Retriever ProSolv CardioVascular FUJIFILM ty ProSolv Service Manager Service Contral Status Started ris Log level Verbose EE Importers This grid will only be populated when the Service is Running Machine Hame Watch Dir Watch Pork Sta
Download Pdf Manuals
Related Search
Related Contents
ISTRUZIONI PER L`USO Mininastri trasportatori KTB IKEA AA-212046-5 User's Manual ENG USER`S MANUAL AR-3600Hi Bedienungsanleitung ODiS I - ekom21 2. - Operator's Manual Ksix B8989FTP05 mobile phone case PCAN-RS-232 - User Manual Sea Gull Lighting 8114-12 Installation Guide Copyright © All rights reserved.
Failed to retrieve file