Fractional (CO2) Laser Equipment User Manual
Contents
1. In the same line the distance between scanning spots should be keep in a half of the fraction 3 3 0 Time prose to adjust the scanning times the scope is from 1 to 20th ol errs Press 1 5 to select the storage space and all the parameters of the current page will be saved when press Save When using the machine you can press 1 5 and do not need to setting parameter 3 3 8 Push t or y to adjust vertical length the scope is 0 20mm push e or e to horizontal length the scope is 0 20mm Press Preview will come out the diagram that you selected 3 3 9 gt These buttons are used for scanning pattern 3 3 10 Ready After the parameters were set press the ready button aim at the place step on the foot switch there will come out the light 3 3 11 Back Click to back to the main interface 3 4 Normal mode interface 3 4 1 Power Press to adjust the power the scope is from O to 30w 3 4 2 Mode Press to change the mode there are three scanning modes 3 4 2 1 Single mode step on the foot switch once there just a beam of laser Press Duration to control the time the scope is from 1 to 100ms 3 4 2 2 After stepping on the foot switch there will be continuous emission pulse Loosen 3 4 2 3 Normal mode After stepping on the foot switch there will be continuous emission pulse Click Pulse Width and Interval to adjust the pulse width and interval It is super pulsed when pulse width
2. and interval is small 3 4 3 After the parameters were set press the ready button aim at the place step on the foot switch there will come out the light 3 4 4 Back Click to back to the main interface 3 5 Settings interface LDL English click to change the language Chinese or English is ao click to open or switch off the voice eer eB Adjust the strength of indicator light 3 5 4 After entering password enter into the calibration interface operate by technician 3 5 5 Back Click to back to the main interface 3 6 Protection interface 3 6 1 When galvanometric scanner is breakdown there will come out the protection interface Be te AM 3 6 2 When water pump do not work or other reasons that lead to the cooling water do not flow there will come out water break warning TER PRA interface 3 6 3 When the door of operation room do not close or the instrument door is opening there will come out chain protection warning interface 2 ARA ER APF RER ART RA 4 Indicator light In view of the invisibility of the 10 6 nm CO laser a visible red diode laser emitting coaxially with CO is provided to help the operator locate laser beam conveniently Press the key the red light emit and a green indicator flashes Press the key again the red light stops emitting and the green indicator extinguishes 5 Precautions 5 1 Never let the laser beam be directed to human ey
3. manufacturing or components defects can enjoy free repairing service Such service is valid only if the instrument is properly used Any damage cause by improper use of the instrument such ad using unfitted power supply and wrong accessories operating in a manner other than specified in this operators manual damages caused by transportation accidents unauthorized installation or maintenance etc such free service will be invalid immediately The free service does not include accessories transportation fee and door to door service charge of professional personnel 11 System internal structural chart Articulated arms M1 M7 Reflector f E Ki germanium wafer a indicator light ESN i ii shutter CO laser tube
4. surgical alcohol must be volatilized before use 6 3 Power calibration The practical laser output power and the preset panel power must be calibrated each year with standard laser power meter within validity period by trained of professional personnel 6 4 Fuse replacement Open the fuse holder with a screwdriver and removal the original fuse Before replacement check and ensure the new fuse is identical in type and specification to the original one 250v 3 1ZA to avoid damage arising from unfit fuses 6 5 Blade sterilization Blade must be sterilized after use 7 Accessories Standard configuration different accordingly to customers requests User manual 1 Articulated arm 1 Power line 1 Foot switch 1 Switch lock key 2 Fuse 5x20 5A 2 Standard blade JH DT 01 JH DT 02 JH DT 03 JH DT 04 JH DT 05 DWG NO Remarks apply 75 alcohol to disinfect and wait until it volatilize completely before each use 8 Troubleshooting Guide Please refer to professional personnel for maintenance l SYMPTOMS After the main power is on either the panel does not light or the water pump does not work when the water pump works there are slight vibration and sound POSSIBLE CAUSES a The power plug has not been properly b The emergency stop switch is pressed down ACTIONS a Check the two plugs at the two ends of the power cable b Replug properly the red mushroom shaped button of the emergency
5. Fractional CO2 Laser Equipment User Manual Contents A cep A E e Bena nia ame at ican sal ore cece recut aA neotenic Dae 3 I Operation Principle Of the Syste a is 4 Toul CO ZMSASEIS V SCC ore OR O CO O Aa 4 1 2 Description of the mainframe ni as 4 13 Mae DING ana NA E IE 5 2 Prestar Prepara ONS in 6 DN ATID ACK AMO INS DEC E E E E A E E EN RE A e A ES E SEN 6 22 Connect arculat ar Serpents rn a REE ARAS ESEIA 6 ES A AP E E A T E 6 24 checking power VONAGE A A EEEE 6 22 COMENT NOW CF Cable e rsa a is aa 7 2 0 FOOtS witch CONNECHON es a a A had a oe ee 7 2 Ff COMMISSIONING OPClAlON isis 7 Operation PrOC COOH S arsena abia 7 Oi SUAELUP WCAC SEE E A ate atu sum onacldwne Ma TAE EEEN ETTE T 7 3 2 Function SEleCUOM MENACE srt lares pu ecards 8 So TACOMA mode ME LACE ra AN ISA a a 9 34 Normal mode Interface oa iS eae ae eee eva 11 BSAS S Eo AER o CP o mat ieuadetnaataaea iphones 12 0 Protection mitra dd Doe 13 Ae TVG CALOL MG NU ss ii E A AT 14 A O A II a a e 14 A A aca AA AAO te 15 ANDINO AAA e ES O E 15 e AR AAA A cannes wasn sanus 15 6 3 POW er calibra AS 16 OP FUSE replacement tiesa tan mens soccer astibao l ce anita 16 e la o ro e 421 0 1 PO O Po o O o O O O 16 TINTE SS DES it a oe id ias 16 9 LrOUDIEShOOUNS Guide at dolia 17 Os Technic alP ar aime ters si2 sts ay esac teucucacaadevesacesentuns tata 18 10 Warranty Go SCE VICC isa AEO 19 Al System menma structural cata Alai 19 Preface The CO laser system i
6. being frozen to break 5 10 Don t leave around laser tube and the instrument recklessly when the instrument recklessly when their service lives end Recycle according to the local environment protection regulations 5 11To avoid improper use of the instrument remove the key from the key switch and keep it properly when the instrument is not in use The instrument generates high voltages within the power supply and laser tube Please refer to professional personnel for maintenance to avoid electric shock 5 120peration room should be equipped with a dust or fume exhauster because the dust arising during operations may be mixed with biological tissue particles 6 Maintenance The instrument generates high voltages within power supply and laser tube Refer to professional personnel for maintenance to avoid electric shock 6 1 Lens cleaning The output power may drop slightly after the instrument has been put into use for half a year This may be caused by the stained focus lens of the Handpiece Wipe the lens gently with moistened cotton ball once or twice Be sure not to damage the lens 6 2 Cabinet cleaning If there is dirt on the cabinet wipe gently with moistened cotton cloth and some detergent or toothpaste DO NOT use over wet cloth in case the water leaks into the inner part of the instrument causing short circuit and damage Notice For each using please clean and sterilize the Handpiece by 75 surgical alcohol And the
7. es or healthy skin 5 2 To prevent human eyes or skin from being hurt by the reflected of laser light never allow the laser beam be directed to any smooth reflective surface such as Stainless steel device surface mirror surface etc 5 3If 75 alcohol is used to clean or sterilize relevant part of the instrument don t use the instrument till the alcohol vaporizes Never operate the instrument in the presence of flammable anesthetics 5 4In order to pervert the focus lens of the handpiece from being polluted and to keep a clear view of the surgical area a smoke evacuator is recommended to the operator The handpiece and focus lens must be cleaned every 3 months 5 5The laser beam generated by this instrument is hazardous to eyes in the area within 35cm from the instrument when someone is staring directly at the laser operators must use safety eyewears when operating 5 6This instrument generates high voltage inside NO attempt should be made by NON professional to open the cabinet of the instrument to avoid electric shock risk 5 7 If the instrument gives out abnormal smell or sound stop operation at once Cut off the power first before any inspection Notice The laser tube is made of glass Take care to the handle to avoid damage 5 8 keep the instrument in an environment with the temperature between 1 C 5 C and the relative humidity between 10 80 5 9 Empty the water tank before transportation to prevent the laser tube from
8. g order 2 6 Footswitch connection Plug the footswitch cable into the socket on the rear part on the instrument push in alignment on the notch until a tone is heard which means a successful lockup 2 7 Commissioning operation After 10 minutes of trial operations of the instrument cut off the power supply temporarily 3 Operation Procedures 3 1 Startup interface AA ES W Ta gt S a m n ey e a SS oo 02 LASER SYSTEM 3 2 Function selection interface i i Y 4 mire zi HAF Dit aa ij 7 TH 1 1 TIA Wi q 4 iiia nF mere TT SYSTEM s NN T S lt 3 2 1 Fractional Click into the fractional mode interface 3 2 2 Normal Click into the normal mode interface 3 2 3 3 2 4 o Click into the system information interface Click into the settings interface 3 3 Fractional mode interface 3 3 1 Comer Press to adjust power the scope is from O to 30w ES 2 Duration Press to adjust the duration the scope is from 0 1 to 10ms 3 09 interval press to adjust interval the scope is from 1 to 5000ms 3 3 4 Distance Press to adjust the distance between the spot to control the density of spots The scope is from 0 1 to 2 6mm SS Scan Mods pos to change the mode there are three scanning modes Sequential pattern The scanning spot is from this side to the other side Shuffle mode no order shot at random Halve mode
9. he articulated arm 1 3 6 Articulated arm The laser beam delivery system consists of light weight spring balanced 7 joint articulated arm The working radius of the articulated arm at full extension is 110 cm Pre startup Preparations 2 1 Unpack and inspect After unpacking please check that the instrument is not damaged with circuit lines well connected and accessories available see the accessories list 2 2 Connect articulated arms Connect the articulated arms with the mainframe and screw up the helicoidal 2 3Pouring coolant A Watering Unscrew the water inlet and air outlet orifice add the water into water inlet orifice by funnel until water overflows After watering screw down the cap screw of both orifices B water relief Unscrew the water outlet hole and spillway hole to let water comes out from the water outlet hole and then tighten up the helicoidal of both holes remarks Put containers to receive water that comes out from the spillway hole when adding water Warning Never turn on power when the tank is empty Cooling water must be used with purified water or distilled water No running water 2 4 Checking power voltage Ensure that the power voltage complies with the requirement of the instrument Socket is in working order 2 5 Connecting power cable Plug the two terminals of the power cable into the power input socket and the power socket Make sure the ground socket is in workin
10. key in the indicated direction to have the emergency key connected SYMPTOMS NO laser beam emits out though the instrument seems running normally POSSIBLE CAUSES a The plug of the footswitch is not properly inserted b The setting of the control panel isn t suitable When the instrument is used for the first time after water is filled the cover is not closed tightly The interlock keys are not pressed down The joint of the articulated arm is loosened ACTIONS a Insert the plug of footswitch tightly according to operator s manual b Set the panel again according to operator s manual c Close the cover and press the interlock keys Screw the joint tightly 3 SYMPTOMS No laser emits The instrument alarms POSSIBLE CAUSES The instrument has been working for too long and the coolant is too hot ACTIONS Stop running the instrument Wait till the temperature of the cooling water goes down below 25 C then restart the instrument 4 SYMPTOMS The instrument makes big noise when running POSSIBLE CAUSES The instrument is not well balanced ACTIONS Place the instrument on a stable and flat surface 5 SYMPTOMS Red pilot beam does not converge or dose not emit from the end of the tube CO2 laser is off the center POSSIBLE CAUSES The articulated arm is either damaged inside or not working normally No laser emits or output power drops significantly ACTIONS Refer to professional personnel for service Note Operators are Not al
11. lowed to adjust the components listed below Laser tube articulated arm diode pilot beam microprocessor board 9 Technical Parameters Name Sealed off CO2 laser Laser Mode Low valance mode Wavelength 10 6 microns Output power 0 30w cont adjustable Focus Spot Diameter 0 4mm Divergence 4 Mrad Lens Focal Distance F 100 mm Power Instability 10 Delivery System Spring balanced 7 joint articulated arm Operation and control Touching switch Microprocessor controlled Pulse Duration 0 05 1s Working Modes Continuous Single pulse repeat pulse and super pulse Display LCD display Cooling system Power supply Input power Closed loop circulating water 230v 50 60Hz see supply circulating water Environment Temperature 350VA 5 40 C Relative Humidity Gross weight Atmospheric pressure Warning up time Electromagnetic requirement Other working conditions Dimension and weight lt 85 85KG 86 0kpa 106 kapa No electromagnetic Field interface No obvious vibration or airflow The machine packing cases 50 80 110 CM 70KG The handle part packing cases 75 20 14 CM 15KG 10 Warranty amp Service The instrument is a well designed user friendly laser surgical system with high quality It performs perfectly under normal use and maintenance Within One year from the data of purchasing any damage caused by
12. s an intelligent laser treatment instrument This machine is featured by compact structure beautiful appearance reliable performance convenient operation and perfect safety The technical specifications of the product have achieved the advanced international standard The instrument can be applied to general surgery gynecology otolaryngology dermatology and cosmetology etc for different treatment such as cutting vaporizing cauterizing and solidifying It can be used in ward and private clinics for its portability and compactness This machine is applicable to extensively dermatology gynecology otorhinolaryngology dentistry general or beauty surgery for incision vaporization cauterization and solidification treatment any pigmented nevus spot acne syringoma lipomyoma hemangioma Beauty use scar xanthelasma lower blepharoplasty tattoo eyebrow excision assistance wrinkle removal surgery any pigmented nevus spot wart nasal polyp hypertrophy of turbinates snoring hemangioma scar skin tag hemorrhopds and fistula helosis Medical use El e ta W bromhidrosis cervicitis cervical erosion vulva vitiligo circumcision tattoo condyloma acuminatum ect exposure to beam expander selective cures numbness of skin Physiotherapy Ea a bog periarthritis of neck periarthritis of shoulder arthritis etc WARNING This instrument generates high voltages and laser radiations within the cabinet Operators must pay m
13. touch screen relevant laser power thus drives output from the sealed off CO2 laser Operation is visible and easy and its humane interface is quite easy and user friendly 1 3 Main cabinet 1 CO laser and compound light source 2 Switch source with high voltage and constant current 3 Main control panel 4 Cooling system 5 Footswitch 6 Articulated arm 1 3 1 CO2 laser and indicator light Sealed off laser is made from glass tube with length of 1255mm and CO2 to be its gas Optic lens are stuck to each end whose sides are equipped with electrodes and water cooling tube When the laser electrodes are connected to high tension current the CO2 gas will be simulated to produce laser with laser outputs from germanium optic lens The indicating light is visible so as to show the position of laser output 1 3 2 Laser main power supply The laser main power supply consists mainly of high tension switch power that is used to ignite CO2 laser tube 1 3 3 Main control panel The microprocessor based main panel is used to control all functions by touching the thin film switch Time and power are displayed digitally which is clear and accurate 1 3 4 Cooling recycle system The laser cooling system is a closed circulating loop The coolant distilled water or ion water is circulated by a pump 1 3 5 Footswitch A footswitch is used to control laser output When the footswitch is pressed the shutter opens and laser emits from t
14. uch attention to safety during operation Operations safety instructions are specified in this manual Any improper use adjustment or maintenance may cause laser radiation hazards or high voltage electric shock Operation Principle of the System 1 1 CO laser system The CO laser with a specific wavelength of 10 6 nm can be absorbed by human body tissue no matter what color the skin is by nearly 100 Through laser emitter the micropulsed laser beam shoots on the target skin in nanometer level to achieve a controllable precise treatment The dimension of shooting area only can be measured in nanometer which is precise enough for operators to control and adjust the treated depth and layer according to the clinical diagnosis The energy of CO laser is even and steady Its focused energy aims only at target tissue without any injury to surroundings It is the heat and electromagnetic effect of the laser that people use to conduct non blood or less blood cutting cauterizing gasification and accurate microsurgery so as to avoid skin burnt During the treatment biochemical reactions of optical thermal accelerate collagen reproduction and skin recovery after exfoliation of the tiny scar fresh and smooth skin can be regenerated The device outputs invisible laser In order to operate with more ease red semiconductor laser has been added to indicate 1 2 Description of the mainframe Most advanced microprocessor is employed to control