Service Information Letters (SIL)

image

Contents

1. Purpose Effectivity Compliance Planning History Accomplishment Inboard BARFIELD P N 101 00809 Errata for User Instruction Manual 56 101 00809 Issued on Jan 29h 1988 To inform users of the referred Adapter Module of revised testing values Section 1 of this SIL and revised calibration procedures Section 2 of this SIL Immediate Compliance with this SIL is MANDATORY The revised values and procedures indicated in this SIL will be included in an upcoming revision B of the referred manual SECTION 1 REVISED TESTING VALUES lt has been found that the range of capacitance values indicated in Figures 1 2 and 5 of the referred manual need to be modified For Section 1 2 Page 4 paragraph 3 C 2 Figure 1 apply the table below Figure 1 Table of Inboard Individual Probe Values Revised Tank Group Probe Select Position Probe Part Number wapaciance Value pF a 40007 0000 0101 Inboard 2 40007 0000 0102 281 405 40007 0000 0103 29 3 0 5 For Section 1 2 Page 5 paragraph 3 D 2 b Figure 2 apply the table below Figure 2 Table of Outboard Individual Probe Values Revised Tank Group Probe Select Position Probe Part Number Capacitance Value pF a E SIL003 Rev C 40008 0000 0101 23 2 0 5 Outboard 2 40008 0000 0102 20 8 0 5 Outboard 40008 0000 0103 19 3 0 5 SAAB FAIRCHILD SF 340 DC 400A MODULE 11 13 14 NEW CAPACITANCE TEST VALUES Page 1 of 2 Copyr
2. Turbine Temperature Test Set November 12 2014 Subject Obsolescence and replacement of a critical component within the Barfield TT1000A Purpose In June 2012 Barfield Inc Barfield was informed by the Eveready Battery Company that a critical component used in the TT1000A product has been made obsolete Eveready Part number 415 a 45V Carbon Zinc Battery NEDA 213 is no longer manufactured and has very limited availability The Eveready 415 is used in the TT1000A to provide the required 45 volt excitation for testing the Insulation breakdown of the aircraft EGT IGT system and other Type K thermocouple based systems Barfield was unable to identify a Form Fit Function FFF replacement for the 415 battery Configurations The following test set model regardless of installed Modifications and or Options is affected by this SIL recommendation Part Number Model Number 1071 00901 TT1000A Purpose Replacement of the 45V battery Solution Barfield has a newly designed replacement module 45V Battery Replacement Converter to mimic the operational specifications and capabilities of the Eveready 415 45V battery Barfield part number 137 08001 is a module designed to be used with an off the shelf 9V alkaline battery to produce the 45V output voltage needed to properly operate the Barfield TT1000A Unless damaged the 137 08001 does not require replacement in the future only the 9V Battery This solution achieves two goals first is to replace t
3. 9292 Fax 1 816 246 9295 Email custsvc expresscal com Web www expresscal com POC Mike Sage Europe EADS SECA 1 Boulevard du 19 mars 1962 BP 50064 95503 GONESSE FRANCE 33 01 30 185412 Telephone Email jean paris seca eads net Web www seca eads net POC Jean Paris SIL No 88 101 00901 Page 3 of 4 I BARFIELD SERVICE INFORMATION LETTER For 2312G XX TT1000A TT1200 TT1200A Turbine Temperature Test Set November 7 2014 Service Centers Canada Wright Instruments Ltd 2762 Slough St Mississauga ON L4T 1G3 Phone 905 677 7161 Australia TR VMS 72 Donaldson Road Rocklea 4106 QLD Australia Ph 61 73717 9146 Fax 61 7 3717 9135 Web www trvms com au Conclusion Barfield Inc is taking a proactive approach in reaching out to its customers and advising them of the potential for unauthorized facilities calibrating our turbine temperature test equipment using obsolete documentation as described above Barfield strongly recommends the return of ALL units at their next calibration recertification interval to a Barfield facility or one of the listed authorized calibration facilities to ensure the equipment is performing as designed and in compliance with new test procedures calibration standard and accuracy requirements Barfield is committed to providing the highest quality products services and support to the aviation industry Inquiries may be emailed to gsesales barfieldinc com Thank
4. TT1200 with Option A There was no provision made for grounding the TT1200A Test Set to the airframe as the necessity for doing so is to our knowledge unique to the OH 58D aircraft and its installed equipment and wiring configuration Barfield has never been made aware of any other aircraft engine temperature indicating system which necessitated this grounding for correct operation The TT1200A test set can be ordered preconfigured for singular use on the OH 58D aircraft under the new Barfield P N 101 00930 OH58D This newly modified Test Set TT1200A is provided with a specific harness designed to interface it with the aircraft The only two Barfield approved Testers for the Engine Temperature Indicating System of the OH 58D airframe are the TT1200 with Option A and the new P N 101 00930 OH58D Inquiries may be emailed to gsesales barfieldinc com Nov 13 2014 Page 2 of 2 Subject Purpose Configurations Effectively Compliance Planning Conclusion SILO06 t BARFIELD SERVICE INFORMATION LETTER FOR DPS400 P N 101 01180 Pitot Static Test Set Using the Barfield DPS400 Test Set is on RVSM Certified Aircraft Regarding compliances with RVSM Barfield Inc defines compliance as meeting or exceeding the widely accepted RVSM test equipment Ps static channel accuracy specification of 0 0030 inHg To inform users how to clearly identify if their DPS400 test set configuration is in compliance to meet
5. battery if unit is returned with the old BP4010 Battery Pack o Barfield will warrantee parts and labor related to any verified digital battery failures Any unrelated parts or labor including calibration will be at customer expense o All freight costs between customer address and Barfield s Miami Facility will be at the customer expense o UPS Ground or International Equivalent shipments between Barfield s Miami Facility and the customer location will be covered by Barfield Inc Any requested priority shipments from the Barfield s Miami facility to customer address will be at customer expense Important interim guidelines If your Barfield 1811 Series tester has a BP4010 and are unable to obtain a replacement it is important that you follow the guidelines outlined in SIL 88 102 02111 1 Rev C Barfield is committed to providing the highest quality products services and support to the aviation industry and apologizes for any inconvenience Thank you for your immediate attention to this matter and for facilitating the replacement of the BP4010 Battery Packs Sincerely y P j eo r j EFI es p r w Seen gti mame i ae 4 P P e i r 4 Li AA Lewes D Wing te dr f ae oe Vice President Ground Support Test Equipment AIRCRAFT MAINTENANCE REPAIR OVERHAUL amp SUPPORT Copyright 2014 Barfield Inc All Rights Reserved t BARFIELD SERVICE INFORMATION LETTER Regarding DALT55 amp DAS650 Instruments Instrument can be
6. 2 Barfield determined that a small percentage of customer units were being returned to Barfield for repair with damage to the Digital Altimeter DALT55 and Digital Airspeed Instruments used in the 1811 series Pitot Static Testers Barfield quickly determined that the damage was occurring when customers charged the BP4010 Battery Pack while installed on the test set What to do To arrange for return of the BP4010 Battery Pack to the proper Barfield service center customers should contact Barfield Ground Support Test Equipment GSTE Customer Service at 1 800 321 3900 1 305 894 5400 or email gsesales barfieldinc com Warranty Replacement Barfield will honor warranty of all related battery replacements and failures for 2 years beginning January 1 2013 through December 31 2014 This Warranty has the following limitations o A customer with a first generation battery B4000 Battery Pack is not affected thus does not qualify for the replacement program o Customers must obtain a Return Material Authorization RMA number from Barfield GSTE Customer Service prior to returning any BP4010 Battery Packs Customers returning parts without a valid RMA could encounter delays and processing charges o Barfield is providing the warranty exchange BP4015 Battery Pack on a one for one basis to customers that return their BP4010 Battery Pack s o All units received for repairs at Barfield beginning in January 2013 will automatically include a replacement
7. 8D In late 2000 the US Army made Barfield aware of an issue when using the Barfield TT1200 Test Set P N 101 00920 on the OH 58D aircraft The observed error reported was that the aircraft temperature indications were unstable and subject to fluctuations Through their subsequent testing on the aircraft they made the determination that if the test set chassis was grounded to the airframe then the aircraft indicator readings were stable and accurate Nov 13 2014 Copyright 2014 Barfield Inc All Rights Reserved Page 1 of 2 Conclusion SILO05 t BARFIELD Based on these reported findings and their request for an embedded solution Barfield developed a modification for the TT1200 designated as Option A This Option consisted of a modification to the TT1200 Test Set panel connector wiring as well as the test set interface cables so that the test set chassis ground was provided by an added third clip lead to the testers cables so as to provide a ground strap to the airframe chassis The US Army subsequently procured some number of these Option A configured TT1200 testers specifically for use on the OH 58D aircraft and used them thereafter without report of any further issues The TT1200 was made obsolete upon the introduction of the TT1200A Test Set in 2003 The model TT1200A P N 101 00930 is an all new test set sharing no subassemblies with the earlier TT1200 tester P N 101 00920 The TT1200A is not a suitable substitute for the
8. B FAIRCHILD SF 340 DC 400A MODULE 11 13 14 NEW CAPACITANCE TEST VALUES Page 2 of 2 Subject Purpose Effectivity Compliance Planning History Accomplishment SILO02 Rev B t BARFIELD SERVICE INFORMATION LETTER FOR DPS400 BARFIELD P N 101 01180 DPS400 New Option Available To inform customers of a new custom fit option choice for the keypad assembly for the Barfield P N 101 01180 Digital Pitot Static Tester All DPS400 Testers with serial number 443 or greater and produced from January 2001 Note Units with serial numbers lower than 443 must be sent to Barfield for evaluation Compliance with this SIL is OPTIONAL Option C involves a choice for a membrane keypad assembly The display assembly has been updated to use a membrane keypad to replace the mechanical switch assembly This new membrane assembly will reduce assembly labor time repair costs and will provide a new key on the keyboard engraved PUMP which will allow operators to turn the pumps ON and OFF New production units will be fitted with the membrane keypad option as a production standard build This build will be documented as Option C The Test Set maintenance menu will be configured via firmware 33RO6N5 to support the new membrane keypad Customers with field serviceable units interested in this option will need to contact Barfield Sales Department or Customer Technical Support 800 321 1039 for further information or to arrange a r
9. al should be reduced or establish proof of stability PRESSURE RANGE ROC VSI Vertical Soeed Displayed 50 Ft min VSI gt 150 Ft min 10 M Min VSI gt 40 M min Note Displays 0 below 100 Ft min amp 30 M min LEAK TEST MODE Measures Leak Rate After 1 minute of stabilization and the minimum of 1 minute Leak Test Interval the Leak Test accuracy 3 Ft min 1 M min TEMPERATURE RANGE _ OPERATING 10 C to 50 C 4 F to 122 F STORAGE 30 C to 60 C 22 F to 140 F SIL No 88 101 02184 B Page 3 of 4 I BARFIELD SERVICE INFORMATION LETTER For DALT55 Digital Altimeter November 12 2014 Planning Prior to shipping call customer service to schedule the recertification of your Pitot Static Test Set Conclusion Barfield Inc is taking a proactive approach in advising its customers of the newly designed Digital Altimeter used in our pitot static products Barfield recommends the return of ALL units at their next calibration recertification interval so that Barfield as OEM can ensure the equipment is performing as designed and in compliance with accuracy requirements Barfield is committed to providing the highest quality products services and support to the aviation industry Inquiries may be emailed to gsesales barfieldinc com Thank you for your business and support Robert Arnett VP Quality SIL No 88 101 02184 B Page 4 of 4 I BARFIELD SERVICE INFORMATION LETTER For TT1000A
10. ays remove the battery pack from Test Set before charging Smart Charger P N 137 00012 Smart Charger P N 137 00012 SC3000 Charger P N 137 00021 BP4000 Battery Pack P N 137 00009 BP4010 Battery Pack P N 137 00020 BP4010 Battery Pack P N 137 00020 Note BP4010 is removed from Test Set Figure 1 Figure 2 Figure 3 DO NOT charge BP4010 battery while it is attached to the test set Figure 4 Figure 5 SIL No 88 102 02111 1 Rev C Page 3 of 3 I BARFIELD SERVICE INFORMATION LETTER For 2312G XX TT1000A TT1200 TT1200A Turbine Temperature Test Set November 7 2014 Subject Usage of obsolete documentation in the calibration of the Barfield Turbine Temperature Tester product lines Purpose The concern Barfield Inc Barfield has is that obsolete Technical Manuals and calibration procedures widely distributed via the internet are being used in the certification of the Barfield Turbine Temperature Tester products The obsolete technical documentation contains calibration standards and procedures which no longer meet Barfield or industry standards and should not be used in the service of Barfield equipment In an effort to ensure that Barfield products maintain quality accuracy and industry standards Barfield has reviewed and revised the calibration procedures for our Models 2312G XX T1000A TT1200 TT1200A Turbine Temperature Test sets Below are the calibration procedures which have now been made obsolete O
11. bration facility listed in this SIL using the latest revised calibration procedure s listed below New New Document Number Revision Revision Date 19 101 00550 2312 XX B 11 12 2014 19 101 00901 TT1000A E 9 19 2014 19 101 00920 TT1200 C 11 12 2014 19 101 00930 TT1200A G 9 25 2014 The revised calibration procedures listed above employ superior standards to ensure that calibration uncertainties of acceptable tolerance for each characteristic maintain the correct test accuracy ratios Planning Prior to shipping call Barfield s Customer Service or an Authorized calibration facility listed below to schedule the recertification of your Barfield Turbine Temperature Tester Service Centers United States Barfield Inc 4101 NW 29 Street Miami FL 33127 Phone 1 800 321 1039 US Only Phone 1 305 894 5400 Fax 1 305 894 5401 GSTE Service barfieldinc com www barfieldinc com SIL No 88 101 00901 Page 2 of 4 Service Centers SERVICE INFORMATION LETTER For I BARFIELD 2312G XX TT1000A TT1200 TT1200A Turbine Temperature Test Set November 7 2014 United States Cont Barfield Inc 1912 West 4 Street Tempe AZ 85281 Phone 1 480 477 2000 Fax 1 480 477 1999 Email barfieldphoenix barfieldinc com Aeroframe Airepairs 3914 Willow Lake Blvd Memphis TN 38118 Phone 1 901 547 4600 sales airepairs com Express Calibration Services 1803 SW Market St Lee s Summit MO 64082 Phone 1 816 246
12. ce barfieldinc com www barfieldinc com Barfield Inc 1912 West 4 Street Tempe AZ 85281 Phone 1 480 477 2000 Fax 1 480 477 1999 Email barfieldphoenix barfieldinc com Aeroframe Airepairs 3914 Willow Lake Blvd Memphis TN 38118 Phone 1 901 547 4600 sales airepairs com SIL No 88 101 00901 1 Page 2 of 4 Service Centers Service Centers SERVICE INFORMATION LETTER For TT1000A Turbine Temperature Test Set November 12 2014 United States Cont Express Calibration Services 1803 SW Market St Lee s Summit MO 64082 Phone 1 816 246 9292 Fax 1 816 246 9295 Email custsvc expresscal com Web www expresscal com POC Mike Sage Europe EADS SECA 1 Boulevard du 19 mars 1962 BP 50064 95503 GONESSE FRANCE 33 0130185412 Telephone Email jean paris seca eads net Web www seca eads net POC Jean Paris Canada Wright Instruments Ltd 2762 Slough St Mississauga ON L4T 1G3 Phone 905 677 7161 SIL No 88 101 00901 1 I BARFIELD Page 3 of 4 Service Centers I BARFIELD SERVICE INFORMATION LETTER For TT1000A Turbine Temperature Test Set November 12 2014 Australia TR VMS 72 Donaldson Road Rocklea 4106 QLD Australia Ph 61 73717 9146 Fax 61 7 3717 9135 Web www trvms com au Barfield is committed to providing the highest quality products services and support to the aviation industry Additional service centers are being evaluated and approved to bette
13. charger p n 1387 00012 are not affected only customers with the SC3000 chargers Note The damage when it occurs results in the instruments not powering on and the instruments will remain inoperative If the Instrument power is on and the recommended calibration interval has been followed the user can continue using the Instruments with confidence Barfield requests that the user follow the CAUTION as detailed above All 1811 A0A configurations using the BP4010 battery pack and the SC3000 charger All DALT55 and or DAS650 configurations using the BP4010 battery pack and the SC3000 charger Effectivity Immediately Compliance Immediate compliance with this SIL is recommended to avoid the possibility of damaging your DALT55 and DAS650 instruments SIL No 88 102 02111 1 Rev C Page 1 of 3 Copyright 2014 Barfield Inc All Rights Reserved t BARFIELD SERVICE INFORMATION LETTER Regarding DALT55 amp DAS650 Instruments Instruments can be damaged during Battery charge cycle when using BP4010 SC3000 charger Planning Conclusion Warranty Consideration Recall on BP4010 battery pack November 12 2014 Prior to shipping your BP4010 battery pack for warranty consideration call customer service to confirm availability of BP4015 P N 137 00030 battery pack Barfield will not have inventory of the new BP4015 prior to December 2012 As a short term solution while waiting for availability of BP4015 continue using
14. damaged during Battery charge cycle when using BP4010 SC3000 charger while Subject Purpose History Configurations attached to instrument Recall on BP4010 battery pack November 12 2014 Possible damage to the DALT55 amp DAS650 Instruments may occur while charging the BP4010 battery pack using the SC3000 charger if the BP4010 is installed on the Test Set during charging CAUTION Remove the BP4010 battery pack from the test set before charging when using the SC3000 charger See page 3 Figure 4 amp 5 Note For customers charging the BP4000 or BP4010 battery pack using the Barfield Smart charger p n 137 00012 there are no issues See page 3 Figure 1 amp 2 The figures on page 3 clearly identify ALL possible battery pack charger configurations to easily identify which configuration the above CAUTION is addressing and the configurations that work fine and are no concern To notify customers of a potential problem that exists and how to avoid damaging the DALT55 and or DAS650 Instruments Recently Barfield has determined that approximately 3 of the units sold during the past 24 months have instruments DALT55 DAS650 that were damaged as a result of charging the BP4010 battery pack using the SC3000 charger Barfield is working quickly to provide a permanent solution for this problem and will advise customers as soon as possible when the solution is available As mentioned above customers using the Smart
15. eturn to obtain this new option Inquiries may be emailed to Techsupport GSTE barfieldinc com DPS400 11 13 14 NEW OPTION AVAILABLE Page 1 of 1 Copyright 2014 Barfield Inc All Rights Reserved Subject Purpose History Configurations Effectivity Compliance Specifications SIL No 88 101 02184 B I BARFIELD SERVICE INFORMATION LETTER For DALT55 Digital Altimeter November 12 2014 Redesign of the Barfield Model DALT55 Note For those customers that returned their test set to Barfield for the annual recertification there is no Issue with continued use of these units This letter is to inform customers that Barfield Inc has recently redesigned the DALT55 Digital Altimeter As a result of the redesign model number DALT55 will not be affected but part number 101 02184 is replaced with a new part number 101 02186 The new part number part number 101 02186 altimeter is a 100 Form Fit Function replacement for the replaced unit part number 1011 02184 Due to procurement issues with critical components Barfield Inc designed a new Model Number DALT55 Digital Altimeter Part Number 101 02186 The following configurations are affected by this design Model Number Part Number 1811D A0A 101 00164 A0A 1811GA A0A 101 00185 A0A 1811HA AOA 101 00184 A0A DALT55 1071 02184 Note Products containing either part number 101 02184 or 101 02186 Digital Altimeters meet the requirements described in Federal Av
16. exceed the RVSM accuracy requirement The following configurations meet or exceed RVSM accuracy requirements ALL DPS400 s n up to 108 must have MOD H amp L All DPS400 s n 109 thru 289 must have MOD Q amp L All DPS400 s n 290 and higher must have MOD G Immediate Compliance with this SIL is MANDATORY for those users who will be performing service work on aircraft which are RVSM certified The DPS400 requires annual every 12 month re certification performed by Barfield or an approved calibration facility to ensure continued compliance with RVSM accuracies and proper test set operation For customers with DPS400 s n 108 or lower verify unit I D plate indicates both MOD For custome with DPS400 s n 109 to 289 verify unit I D plate indicates both MOD Fo aims with DPS400 s n 290 and above verify unit I D plate indicates MOD by test set MOD level is does not include the MOD levels described above contact Barfield customer service ALL DPS400 Test Sets that meet the above described MOD levels and are calibrated on an annual 12 month cycle by Barfield Inc or an approved calibration facility are in compliance with the accuracy requirements of RVSM Inquiries may be emailed to gsesales barfieldinc com Novmber 13 2014 Page 1 of 1 Copyright 2014 Barfield Inc All Rights Reserved t BARFIELD SERVICE INFORMATION LETTER FOR DC 400A ADAPTER MODULE FOR SAAB FAIRCHILD SF 340 Subject
17. he obsolete Eveready 415 battery and second replaces a specialty battery having a limited availability with an off the shelf solution available internationally Compliance Customer compliance with this SIL is optional The obsolete Eveready 415 battery is considered a consumable part thus will require eventual replacement as they fail All new Barfield TT1000A Turbine Temperature Testers manufactured after April 2013 will include the new 137 08001 module and 9V battery Any 45V Batteries identified as unserviceable during a repair or calibration activity should be replaced with new 137 08001 SIL No 88 101 00901 1 Page 1 of 4 Copyright 2014 Barfield Inc All Rights Reserved Availability Installation Service Centers SERVICE INFORMATION LETTER For TT1000A Turbine Temperature Test Set November 12 2014 I BARFIELD The 137 08001 will be stocked by Barfield and our authorized distributors and service centers beginning May 2013 Please call Barfield s Customer Service or an Authorized calibration facility listed below to purchase a 137 08001 45V Battery Replacement Converter or schedule the recertification of your Barfield Turbine Temperature Tester Installation of the 137 08001 45V Battery Replacement Converter is outlined in Barfield document 60 137 08001 United States Barfield Inc 4101 NW 29 Street Miami FL 33127 Phone 1 800 321 1039 US Only Phone 1 305 894 5400 Fax 1 305 894 5401 GSTE Servi
18. iation Regulation Part 43 Appenaix E which defines the requirements for accuracy of an aircraft s altimeter and is used as a reference for the accuracy of Pitot Static testers These products however do not meet RVSM accuracy requirements All newly manufactured test sets listed in above configurations will contain a DALT55 with part number 101 02186 Any units received for service determined to have a transducer failure part number PDCR2200 8860 will require update to new part number Update would include new transducer manifold PC board and hardware to convert to new part number 101 02186 No compliance requirement Page 1 of 4 Copyright 2014 Barfield Inc All Rights Reserved I BARFIELD SERVICE INFORMATION LETTER For DALT55 Digital Altimeter November 12 2014 Specifications for the discontinued Barfield DALT55 part number 101 002184 DALT55 P N 101 02184 RANGE 1500 to 55 000 ft ACCURACY 0 025 Typical FS over complete operating temperature range Specific accuracies at certain altitudes are 1 500 Ft 30 000 Ft 20 Ft 10 000 Ft 10 Ft 50 000 Ft 48 Ft 20 000 Ft 14 Ft 55 000 Ft 62 Ft Note When displaying altitude in Ft Feet or M Meters the pressure altitude calculation is based on a baro set of 29 921 inHg 1013 2 mb Note The Baro Set value cannot be changed it is set at 29 921 inHg STABILITY 0 05 FS Typical over 12 months Note This tolerance represents the typical stability over a 12
19. ight 2014 Barfield Inc All Rights Reserved t BARFIELD For Section 1 2 Page 21 paragraph 9 B 2 Figure 5 apply the table below Figure 5 Individual Probe Values Location Revised Probe Select Position Probe Part Number Capacitance Value ae Inboard 40007 0000 0101 4 1 0 5 Inboard 2 40007 0000 0102 28 1 0 5 _ Inboard 3 40007 0000 0103 29 3 0 5 Outboard 1 40008 0000 0101 23 2 0 5 Outboard 2 40008 0000 0102 20 8 0 5 Outboard 3 40008 0000 0103 19 3 0 5 History Accomplishment SIL003 Rev C SECTION 2 REVISED CALIBRATION PROCEDURES It has been found that two related calibration procedures of the referred manual are missing one step which is the same for both procedures and at the same position in the sequence The following two procedures of Section 1 2 Page 17 Test Procedure for System Calibration Preferred Method Dry Tank subsection 7 C and Page 19 Test Procedure for System Calibration Alternate Method Wet Tank subsection 8 C both are missing the same step immediately after step 4 Release the PRESS TO READ CAP pF pushbutton The new step will read 5 Rotate the TEST FUNCTION selector switch to IND AMP Current step 5 of both procedures will be renumbered as 6 with the following steps of both renumbered accordingly in the upcoming revision of this manual Inquiries may be emailed to gsesales barfieldinc com SAA
20. ld Old Document Number Revision Revision Date 57 101 00550 2312 XX N A Nov 30 1983 19 101 00901 TTI1000A C Oct 4 2011 57 101 00901 TT1000A N A Sept 28 1984 20 101 00920 1 TT1200 A Nov 20 1995 19 101 00930 TI1200A E Nov 18 2010 This is a concern for those customers that have their units calibrated by a facility not using the authorized calibration standard and or procedures For these customers we recommend returning your unit s to Barfield or an authorized service facility listed in this letter for calibration verification service to ensure that operational and accuracy compliance is maintained SIL No 88 101 00901 Page 1 of 4 Copyright 2014 Barfield Inc All Rights Reserved I BARFIELD SERVICE INFORMATION LETTER For 2312G XX TT1000A TT1200 TT1200A Turbine Temperature Test Set November 7 2014 Configurations The following test set models regardless of installed Modifications and or Options are affected by these SIL requirements Part Number Model Number 1071 00550 through 101 00557 2312G XX 101 00901 TT1000A 101 00920 TT1200 101 00930 TT1200A 101 00930 OH58D TT1200A Effectivity Next scheduled annual calibration cycle or sooner should any repair or service action be required Compliance Compliance with this SIL at next calibration cycle is recommended for all Barfield Turbine Temperature Testers The recommended annual every 12 month re certification should be performed by Barfield Inc or a cali
21. month period If tighter tolerances are required the calibration interval should be reduced or establish proof of stability PRESSURE RANGE ROC VSI Vertical Soeed Displayed 50 Ft min VSI gt 150 Ft min 10 M Min VSI gt 40 M min Note Displays 0 below 100 Ft min amp 30 M min LEAK TEST MODE Measures Leak Rate After 1 minute of stabilization and the minimum of 1 minute Leak Test Interval the Leak Test accuracy 3 Ft min 1 M min TEMPERATURE RANGE _ OPERATING 10 C to 50 C 4 F to 122 F STORAGE 30 C to 60 C 22 F to 140 F SIL No 88 101 02184 B Page 2 of 4 I BARFIELD SERVICE INFORMATION LETTER For DALT55 Digital Altimeter November 12 2014 Specifications for the new Barfield DALT55 part number 101 002186 DALT55 P N 101 02186 ALTITUDE RANGE 1 500 to 55 000 ft ACCURACY 0 020 FS Typical 0 040 FS Max over complete operating temperature range Specific accuracies at certain altitudes are 1 500 Ft 30 000 Ft 20 Ft 10 000 Ft 10 Ft 50 000 Ft 48 Ft 20 000 Ft 14 Ft 55 000 Ft 62 Ft Note When displaying altitude in Ft Feet or M Meters the pressure altitude calculation is based on a baro set of 29 921 inHg 1013 2 mb Note The Baro Set value cannot be changed it is set at 29 921 inHg STABILITY 0 025 FS Max over 12 months Note This tolerance represents the stability over a 12 month period If tighter tolerances are required the calibration interv
22. r serve our customer Please visit the Barfield Inc website for the most up to date list of service centers Inquiries may also be emailed to gsesales barfieldinc com Thank you for your business and support Robert Arnett VP Quality SIL No 88 101 00901 1 Page 4 of 4
23. t BARFIELD Service Information Letters SIL Digital Instrument Battery Letter BP4010 Battery Pack Recall Turbine Temperature Calibration Use of the TT1200A for the Bell OH 58D DPS400 for RSVM certified Aircraft Errata for Manual of DC400A Adapter for SF340 New keypad option for DPS400 DALT55 Redesign TT1000A 45V Battery Replacement November 14 2014 November 12 2014 November 7 2014 November 13 2014 November 13 2014 November 12 2014 November 13 2014 November 12 2014 November 12 2014 Copyright 2014 Barfield Inc All Rights Reserved I BARFIELD Barfield Inc 4101 N W 29th Street Miami FL 33142 USA Phone 1 305 894 5300 Fax 1 305 894 5301 Subject BP4010 Battery Pack Replacement Dear Customer This letter is to inform you that Barfield strongly recommends the replacement of the BP4010 Battery used on our 1811 Series Pitot Static testers with Digital Instruments For more technical details regarding the replacement of the BP4010 Battery please refer to Service Information Letter SIL 88 102 02111 1 Rev C titled BP4010 Battery Pack Recall found on the Barfield Inc website www barfieldinc com To access this file from the Barfield Inc home page select Services amp Offer then Ground Support Test Equipment then Technical Publications At the bottom right corner select the Service Information Letter box What happened In September 201
24. the BP4010 battery pack and following the CAUTION detailed on page 1 and illustrated in Figures 1 2 amp 3 on page 3 Barfield Inc is taking a proactive approach in reaching out to its customers and advising them of this charging issue The short term solution is to remove the BP4010 from the test set while charging using the SC3000 P N 137 00021 charger Barfield is working on a long term solution correcting the BP4010 design and will provide qualifying customers with a warranty replacement battery pack BP4015 when available Barfield will provide a warranty exchange BP4015 battery pack only to customers that send in their BP4010 battery pack s As of December 2012 all customer units received for calibration or repair will automatically get a replacement BP4015 if their test set has a BP4010 battery pack Note Do not send in your BP4010 for replacement consideration until notified or confirming that the BP4015 is available Barfield is committed to providing the highest quality products services and support to the aviation industry Inquiries may be emailed to gsesales barfieldinc com Thank you for your business and support Robert Arnett VP Quality SIL No 88 102 02111 1 Rev Page 2 of 3 I BARFIELD SERVICE INFORMATION LETTER Regarding DALT55 amp DAS650 Instruments Instruments can be damaged during Battery charge cycle when using BP4010 SC3000 charger Recall on BP4010 battery pack November 12 2014 Alw
25. you for your business and support Robert Arnett VP Quality SIL No 88 101 00901 Page 4 of 4 Subject Purpose Effectivity Compliance Planning History SILO05 t BARFIELD SERVICE INFORMATION LETTER FOR TT1200A Turbine Temperature Test Set BARFIELD P N 101 00930 Use of the Barfield TT1200A Turbine Temperature Test Set on the Bell OH 58D Turbine Temperature Indicating System To inform users of a new P N the 101 00930 OH58D that has been created by modifying Test Set TT1200A in order to ensure compatibility when testing the Bell OH 58D aircraft engine temperature indicating system Immediate Compliance with this SIL is MANDATORY for those users who will be performing service work on the OH 58D Engine Temperature Indicating System with the Barfield TT1200A Test Set A statement of this requirement will be included in an upcoming revision of the TT1200A User manual P N 56 101 00930 A separate supplement to this manual will be prepared for the new P N It will explain those operating instructions that are different from the Standard TT1200A including the use of a new harness that has been specifically designed to interface the new TT1200A with the OH 58D airframe All existing procedures of the Standard TT1200A Instruction Manual remain effective except as outlined in the upcoming supplement Existing Calibration Procedure P N 20 101 00930 is applicable for both TT1200A P Ns the 101 00930 and 101 00930 OH5