1. QA 90 MKII Electrical Safety Analyzer PRODUCT HIGHLIGHTS Small size easy to carry Can operate completely standalone or by remote control with the ansur software Tests instruments containing modules with different protection classes in one test run i e CF and BF defibrillator Tests module separation Differential current measurements optional 11 patient inputs laboratory style Test according to IEC 60601 1 IEC 60601 1 1 IEC 60601 2 4 IEC 61010 VDE 751 1 1990 VDE 751 1 2001 UL 2601 1 AS 3200 1 DB 9801 HEI 95 AAMI Testing against standard or user defined limits Automatic stepwise and manual test modes Programmable power up delay time before measurements Integrated keypad Internal memory for uploading test sequences and storing typically up to 200 equipment test results Built in standard A4 protocols for on site documentation localised RS 232 Centronics and Bar code interface METKeN OVERVIEW The QA 90 represents a new generation of safety testing equipment QA 90 is the only automatic safety tester that can test Instruments containing modules with different classes of protection in one test run i e CF and BF defibrillators It is simple to use All you need to do is select the type and class of the equipment to be tested When you press START the QA 90 will execute the tests prescribed in the selected standard The test result can either be printed out immediately or stored inter
2. Mod Manufacturer Model Type Location No of leads SETUP DATA Test Mode Automa Stop at new power config N Stop before new power config N Multiple Enclosure Tests N External Isolating Transformer N TEST lt Rapid gt TYPE LIMIT RESULT WARNING okk kok ek ekk k k k k k k k ek k k k k k k k k ek kk k k k k k k k kk kkk kk k k k k kkk kkk kkk kkk kkk kkk kkk kk Live Neutral 240 0 V Live Ground 240 0 V Neutral Ground OV 285 0 mA Test according to Power up delay time Stop at new module N Multiple Protective Earth Test N Protective Earth test current 25 A IEC601 1 S2 Supply Voltage Supply Voltage Supply Voltage Current Consumption Protective Earth Insulation Resistance Mains to Case Insulation Resistance Applied Part Mod 1 Earth Leakage Current OS 200 mOhm 260 0 mOhm gt 200 0 MOhm gt 200 0 MOhm 1000 uA 627 0 pA Earth Leakage Current NC 500 pA OSRM 1000 pA NCRM 500 pA Earth Leakage Current Earth Leakage Current Enclosure Leakage Current Enclosure Leakage Current Enclosure Leakage Current Enclosure Leakage Current Enclosure Leakage Current Enclosure Leakage Current Patient Leakage Current Patient Leakage Current Patient Leakage Current Patient Leakage Current AC Mod 1 Lead 3 Patient Leakage Current AC Mod 1 Lead 3 Patient Leakage Current AC Mod 1 Lead 3 Patient Auxiliary Current AC Mod 1 Lead 3 Patient Auxiliary Current AC Mod 1 Lead 1 Patient Auxiliary Cur
3. RS HP Desk Jet Canon Bubble Jet or compatible QA 90 ORDERING INFORMATION Order no 11020 QA 90 MKII Electrical Safety Analyzer specify power supply socket Accessories Various power supply socket including European Schuco French Schuco UK Swiss Australian US 11100 11110 10500 11400 11401 11410 11402 11451 11452 11481 11482 11461 11462 11471 11472 11200 11201 11225 11025 www metron biomed com Carrying Case Hard Case Carrying Case ext printer Bar code reader Isolating transformer 400VA Isolating transformer 800VA Test unit ESA E Input Measuring Cable 2m black E Input Measuring Cable 2m red E Input Measuring Cable 5m black E Input Measuring Cable 5m red Clamp Crocodile style black Clamp Crocodile style red Grip C black Grip C red ansur QA 90 plug in ansur QA 90 plug in demo User manual ansur QA 90 plug in User Service manual QA 90 METReN
4. ed from a transformer with a maximum idle voltage of 6V The measurement can be performed on ground leads or between E and E floating inputs outputs Range Resolution Accuracy No of Tests 0 2000 mohm 1 mohm 2 of full scale or 5 of reading 1 2 or multiple INSULATING RESISTANCE The measurement of the insulating resistance may be executed between casing and power unit or between patient module and power unit Test voltage No of Tests Range 1 Resolution Accuracy Range 2 Resolution Accuracy 500VDC through a 130 kohm limiting resistor 1 2 or multiple 1 50 Mohm 1 Mohm 2 of full scale 1 LSD 51 200 Mohm 1 Mohm 2 of full scale 1 LSD LEAKAGE CURRENTS All measurements can be performed with a IEC 601 1 filter patient equivalent or without flat frequency response The filter can be exchanged with filters covering other standards All measurements can be performed as true RMS measurements or AC DC measurements The following leakage currents are measured No of Tests e Earth leakage current 4 e Enclosure leakage current 6 or multiple e Differential current 2 e Substitute equipment current 1 e Direct current 2 e Current Fig 9 1 The following leakage currents are measured for each module e Patient leakage current 6 e Mains on applied part leakage current 2 e Patient Auxiliary current 6 Floating dual lead measurement of leakage currents Multiple e S
5. nally in the unit for later use The QA 90 can be fully remotely controlled via ansur software With ansur you can make your own test protocols store the information on disk and export formatted data to equipment management databases Test sequence Individual test sequences may be compiled to satisfy national and international standards IEC 60601 1 VDE 0750 T1 12 91 BS 5724 UL 2601 1 CAN CSA C22 2 No 601 1 M90 AS 3200 1 NZS 6150 1990 VDE 0751 1990 VDE 0751 2001 IEC 60601 1 1 IEC 60601 2 4 UL 544 HEI 95 DB9801 and more In the interest of product improvement Metron claims the right to alter specifications without notice QA 90 Specifications VOLTAGE MEASUREMENT The voltage measurement may be executed in the following ways Between lead 1 and 2 Between lead 1 and Earth Between lead 2 and Earth Between input output E and E Floating inputs outputs Range 0 400V true RMS Resolution 0 1V Accuracy DC 100 Hz 1 of full scale 1 LSD 100 Hz 10 kHz 2 of scale 1 LSD No of Tests 4 or multiple CURRENT CONSUMPTION The current measurement may be executed in lead no 1 live Range 1 0 1000mA true RMS lt 250VAC Resolution 1mA Accuracy 2 of full scale 1 LSD No of Tests 1 or multiple Range 2 1 16A true RMS lt 250VAC Resolution 1mA Accuracy 1 of full scale 1 LSD No of Tests 1 or multiple PROTECTIVE EARTH The test current is 25A or 1A deliver
6. rent AC Mod 1 Lead 1 Patient Auxiliary Current AC Mod 1 Lead 3 Patient Auxiliary Current AC Mod 1 Lead 3 Patient Auxiliary Current AC Mod 1 Lead 3 Mains on Applied Part Mod 1 Mains on Applied Part Mod 1 Patient Auxiliary Current DC Mod 1 Lead 3 Patient Auxiliary Current DC Mod 1 Lead 3 Patient Auxiliary Current DC Mod 1 Lead 3 Patient Auxiliary Current DC Mod 1 Lead 3 Patient Auxiliary Current DC Mod 1 Lead 3 Patient Auxiliary Current DC Mod 1 Lead 3 Patient Leakage Current DC Mod 1 Lead 3 Patient Leakage Current DC Mod 1 Lead 3 Patient Leakage Current DC Mod 1 Lead 3 Patient Leakage Current DC Mod 1 Lead 3 Patient Leakage Current DC Mod 1 Lead 3 Patient Leakage Current AC Mod 1 Lead 3 AC Mod 1 Lead 3 AC Mod 1 Lead 3 DC Mod 1 Lead 3 Unit passed test Comments Date Time Signature MN MK KIKK KKA Ne sais sss a KFANTO RAN Example of printout of a test GENERAL INFORMATION TEMPERATURE REQUIREMENTS 15 C to 35 C while operating 0 C to 50 C for storage DISPLAY Type LCD Alphanumeric format 4 lines by Display control 40 characters 7 F keys anda keypad DATA INPUT OUTPUTS 2 Parallel printer port 1 bi directional RS 232C 1 for Computer control Bar code interface POWER From 100 VAC to 240 VAC 47 63 Hz HOUSING Metal case DIMENSIONS LxWxH 305mm x 342mm x 132mm WEIGHT 5 8 kg STANDARD ACCESSORIES User Service manual RECOMMENDED PRINTE
7. ubstitute patient current 1 In one test run a maximum of 11 modules with different protection classes may be tested MTD 0402 044 13 ACCURACY Range 1 0 99 9 HA Resolution 0 1 pA Accuracy 2 of full scale 1 LSD Range 2 100 1000 pA Resolution 1 yA Accuracy 2 of full scale 1 LSD Range 3 1 0 10 0 mA Resolution 1 uA Accuracy 1 of full scale 1 LSD FREQUENCY RESPONSE DC 1 MHz 3dB with a Crest factor gt 2 The test voltage for the mains on applied part measurement is 110 of the line voltage delivered through a limiting resistor of 47 kohm ansur SOFTWARE OFFERS THE FOLLOWING FEATURES REMOTE CONTROL All functions and tests in QA 90 may be performed from the PC USER DEFINITION OF TEST STANDARDS Select predefined standards or create your own local new standard with test limits USER PROGRAMMABLE TEST SEQUENCES Allows user defined test sequences with a selection of tests from the selected test standard CUSTOMIZED PROTOCOLS Create your own protocol format including a header checklist job instructions a command field and a test sequence STORAGE AND RECALL Protocol formats and data may be stored recalled printed out or transferred to D base systems PRINT OUT METRON QA 90 Electrical Safety Analyzer Ver x xx QA 90 Serial no Operator 2 Establishment EQUIPMENT INFORMATION Equipment Code 123 Class CLI Module Code ECG Group 1 Type CF Serial No 3 Status

Download Pdf Manuals

image

Related Search

Related Contents

Home Decorators Collection 1369400910 Instructions / Assembly  Système d`interaction dans un musée  Guía de instalación Zyxel P660 HWD1 inalámbrico  Feuilleter le Mode d`emploi    6ª 6ª - iiR Angola  Philips SWV2026 Quick F Plug adapter  Mode d`emploi Fers Vapeur Ultragliss Easycord  Table of Contents - Swansons Telemekanik AB  Catálogo de productos  