1. WARNING DO NOT disassemble puncture or crush the Power Cartridge Power Cartridge electrolytes may be toxic if swallowed and can be harmful to skin and eyes Keep the Power Cartridge away from children WARNING Use of a damaged Power Cartridge may cause personal injury WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury WARNING Possible Shock Hazard Protect the ECLIPSE and AC DC Power Supplies from fluid spills or drips WARNING DO NOT use any cord if damaged Improper use of the power cord may result in a fire or electrical shock hazard WARNING DO NOT touch the Power Cartridge cavity connectors with the Power Cartridge removed while the unit is connected to an external power source This may result in electrical shock WARNING Secure the ECLIPSE and the DC Power Supply in your vehicle and make sure the air inlet and exhaust airflow is not blocked This may impact performance WARNING DO NOT leave the ECLIPSE or DC Power Supply plugged into the vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle This may drain the vehicle s battery WARNING If the vehicle power source vehicle accessory connector drops below 11 5 Volts brownout the ECLIPSE will revert to Power Cartridge operation if present The External Power Present Indicator may remain on WARNING Operating time will degrade
2. DE Green Yellow Red Audible What to do Condidon Indicator Indicator Indicator Alarm a f Wait The system typically ae Warming Up On On On None takes 3 5 minutes to reach S Please Wait specified performance za 7 a AS Nothing The system is z 5 hoe 7 EHOW On Off Off None operating properly at the 8 specified flow rate Low or Warm ee Plug into external power or Battery On Blinking Off Beeps replace with a charged battery O Concentration Continue Use Check or clean E 8 5 Off Blinking Off None air intake filter If condition persists contact Dealer Provider Continue Use Check or clean O Concentration Off ee air intake filter If condition S lt 70 oi Blinking Beeps persists contact 5 Dealer Provider Check tubing and or humidifier lt Flow Rate for obstruction Check clean air A Error Blocked Flow Off Blinking Off Beeps intake filter If condition E persists past 10 minutes contact E dealer provider Z Make sure tubing does not exceed 7 ft 2 1m when using Pulse Flow Mode Audible No Inspiration alarm will beep after 15 seconds DT Gn Pulse On On Off Beeps if no inspiration is detected If Flow Mode only no inspiration detected 15 y seconds after that the ECLIPSE will automatically switch to Continuous Flow Mode Power Cartridge Shutdown or hot Seek external power source If 2 Power Cartridge Off Off Blinking On Power Cartridge Fault persists with no ext
3. Revision Page 6112_X5 6112 X5 5 of 35 SAFETY GUIDELINES This section of the manual provides warnings and cautions that apply to hazards or unsafe practices that could result in personal injury or property damage OXYGEN THERAPY SAFETY ia Read Instruction Manual DO NOT operate this oxygen system without first reading and understanding this manual If you are unable to understand the warnings and instructions contact your authorized SeQual dealer provider before attempting to use this oxygen system Caution or Attention In certain circumstances oxygen therapy can be hazardous therefore it is advisable to seek medical advice before using the oxygen system DO NOT deviate from prescribed flow rates Adjust the flow only to your prescribed flow rate unless otherwise instructed by your physician WARNING 9 Warning No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Oxygen enhances the combustion of flammable material For safety concerns all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty WARNING S Use No Oil or Grease A spontaneous and violent ignition may occur if oi
4. 5 3 9 hours 1 0 3 5 hours 4 9 hours 2 0 2 1 hours 3 9 hours 3 0 1 3 hours 3 3 hours 4 0 2 8 hours 5 0 2 4 hours 6 0 2 2 hours Operating time will degrade with Power Cartridge use and age CAUTION Store your Power Cartridge in a cool dry location This will help to assure the longevity of your Power Cartridge CAUTION DOT UN Regulations require the removal of the Power Cartridge for international travel CAUTION There are no serviceable parts inside the Power Cartridge DO NOT open CAUTION Replace the Power Cartridge with only SeQual approved Power Cartridge CAUTION The Power Cartridge used in this device may present a risk of fire or chemical burn if mistreated DO NOT disassemble incinerate or heat above 140 F 60 C CAUTION Return your Power Cartridge to your dealer provider for proper disposal CAUTION Keep your Power Cartridge away from children File Name Doc Number Revision Page 6112 X5 6112 X5 17 of 35 TYPICAL POWER CARTRIDGE RECHARGE TIME The typical time to recharge your Power Cartridge to achieve 80 capacity from a fully discharged Power Cartridge is 1 4 hours to 5 0 hours dependent upon the flow setting If the Power Cartridge is too warm charging will not begin until it sufficiently cools The Power Cartridge may not charge or the charge time may increase when operating from the DC Power Supply set in the 1
5. This User Manual describes all the necessary information in order for a patient or caregiver to operate the Eclipse WARNING Important safety information for hazards that might cause serious injury CAUTION Important information for preventing damage to the ECLIPSE Oxygen System DO NOT Information to which you should pay special attention Important Clarifying information specific instructions commentary sidelights or interesting points Symbols Used in User Manual and on Oxygen System Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction with any of the following symbols indicates the need to consult the operating instructions provided with the product A potential risk exists if the operating instructions are not followed Attention or Caution Consult accompanying documents This symbol indicates important information that requires the user to take special precautions when certain conditions are present No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Materials that burn in air and even some materials that will not burn in air are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the r
6. assure the longevity of your Power Cartridge CAUTION DOT UN Regulations require the removal of the Power Cartridge for international travel CAUTION There are no serviceable parts inside the Power Cartridge DO NOT open CAUTION Replace the Power Cartridge with only SeQual approved Power Cartridge CAUTION The Power Cartridge used in this device may present a risk of fire or chemical burn if mistreated DO NOT disassemble incinerate or heat above 140 F 60 C CAUTION Return your Power Cartridge to your dealer provider for proper disposal CAUTION Keep your Power Cartridge away from children File Name Doc Number Revision Page 6112 X5 6112 X5 32 of 35 CAUTION Store the Power Cartridge in a cool dry place when not in use CAUTION DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle or in a trunk during a hot day CAUTION The ECLIPSE system can only work with a SeQual Power Cartridge Use of another Power Cartridge or Battery may damage the unit present a risk of fire or explosion and will void the warranty CAUTION To ensure there is adequate flow to deliver oxygen the length of the tubing must not exceed 50 feet 15 2m when using Continuous Flow Mode and must not exceed 7 feet 2 1m when operating in Pulse Flow Mode for adequate breath detection CAUTION When operating the ECLIPSE on the cart DO NOT bang drop or bump the unit
7. length of the tubing must not exceed 50 feet 15 2m when using Continuous Flow Mode and must not exceed 7 feet 2 1m when operating in Pulse Flow Mode for adequate breath detection Cannula Salter Labs Part Number 1600 Series or equivalent Your physician will have prescribed a cannula to deliver oxygen In most cases they are already attached to the oxygen tubing If not follow the instructions included with the cannula to attach it to the oxygen tubing Use of cannula not specified for use with this oxygen system may impair the performance Power Options You have various options to provide power to your ECLIPSE A Power Cartridge AC Power Supply and DC Power Supply File Name Doc Number Revision Page 6112_X5 6112 X5 23 of 35 Power Cartridge The rechargeable Power Cartridge supplied with your ECLIPSE is capable of providing you with battery operation depending on your prescribed flow rate Your Power Cartridge will be recharged when the ECLIPSE is connected to the AC Power Supply or the DC Power Supply if adequate power is available The Power Cartridge will take longer to recharge when connected to the DC Power Supply and will depend on the flow setting and the amperage switch setting Power Compartment The Power Compartment is available when Power Cartridge operation is not required When using the Power Compartment your ECLIPSE will operate on the AC and DC Power Supplies AC P
8. with Power Cartridge use and age WARNING When operating in Continuous Flow Mode to ensure adequate flow DO NOT use a tubing cannula extension more than 50 15 2m long WARNING Pulse Flow settings should be determined for each patient individually Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode WARNING When operating in Pulse Flow Mode to ensure adequate breath detection the length of the tubing cannula must not exceed 7 feet 2 1m WARNING As with all conserving devices the ECLIPSE may not be able to detect some respiratory efforts WARNING DO NOT change from prescribed flow rate setting without consulting with a qualified clinician File Name Doc Number Revision Page 6112_X5 6112 X5 30 of 35 WARNING AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL FAILURE CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK UP SYSTEM REQUIRED DO NOT IGNORE ALARMS WARNING Use of certain oxygen administration accessories not specified for use with this oxygen system may impair the performance ONLY use recommended oxygen delivery accessories WARNING The tube fitting may be tight DO NOT use oil or grease to lubricate WARNING If tubing becomes disconnected during operation ECLIPSE does not alarm Check flow at the cannula to verify oxygen del
9. 0 AMP position INITIAL POWER CARTRIDGE OPERATION The new Power Cartridge packaged with your ECLIPSE Oxygen System is not fully charged Before using your ECLIPSE Oxygen System for the first time you must fully discharge the Power Cartridge 1 Turn your ECLIPSE ON and allow the Power Cartridge to completely discharge Unit will sound a low Power Cartridge alert until the unit completely shuts down This could continue for up to approximately 20 minutes 2 Connect the AC Power Supply and plug in the AC Power Cord to completely charge the Power Cartridge This may require up to five hours of uninterrupted charging This completes the calibration of the Power Cartridge WARNING DO NOT tamper with disassemble crush or heat the Power Cartridge above 140 F 60 C The Power Cartridge may present a risk of fire or explosion and will void the warranty CAUTION Store the Power Cartridge in a cool dry place when not in use CAUTION DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle or in the trunk during a hot day CAUTION The ECLIPSE system can only work with a SeQual Power Cartridge Use of another Power Cartridge or Battery may damage the unit present a risk of fire or explosion and will void the warranty In the event of a power interruption the ECLIPSE Oxygen System will automatically switch to the Power Cartridge operation When power is restored the Power Cartridge will automatically sta
10. 12_X5 6112 X5 31 of 35 CAUTION DO NOT expose the Power Cartridge to temperatures above 140 F 60 C such as in a vehicle parked in the sun or on a hot day CAUTION Charging the Power Cartridge below 50 F 10 C or above 104 F 40 C may degrade the performance CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock CAUTION Only use the Power Cartridge for its intended purpose Operating your ECLIPSE outside the recommended operating environment may CAUTION impact performance cause damage and will void the warranty CAUTION Locate the unit in a well ventilated room that provides adequate air flow and check if air is being prevented from entering or exiting the system by furniture draperies bedding clothing or pets CAUTION Always check to see that the Air Inlet and the exhaust vent are not blocked and the air inlet filter is clean before using your system CAUTION The AC power cord used with the ECLIPSE Oxygen System must meet the requirements of the country where the product is used CAUTION Use with grounded 3 pin receptacle DO NOT remove ground pin from cord Use only with SeQual supplied cords CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical outlet controlled by a switch CAUTION Store your Power Cartridge in a cool dry location This will help to
11. 13 9 pounds 6 3 kg without Power Cartridge Continuous Flow Indication Expressed in liters per minute LPM Audible Alarm Indicators Loss of power Hot Power Cartridge Weight Low Battery Warm Power Cartridge Low Oxygen Output O flow outside normal limits No inspiration detected in Pulse Mode Unit Malfunction Back Up Alarm Power 9V Internal Battery Air Inlet HEPA File Name Doc Number Revision Page 6112_X5 6112 X5 27 of 35 Device Classification IEC Class I Type B Applied Part Transport Storage Temperature 4 F 20 C to 140 F 60 C Requirements Humidity up to 95 Non Condensing Back pressure Effect on Flow None Indicator Altitude 0 13 123 feet 0 4000 meters Pulse Flow Mode e Principle of Operation Pulse Flow Mode is designed to deliver 16 ml boluses of oxygen per incremental flow setting for each patient inspiration detected Bolus volume is held constant regardless of breath rate within a given setting Delivery of the bolus is triggered by detection of the end of inspiration followed by the start of a subsequent inspiration Sensitivity is dealer provider adjustable Based on this inspiration trigger signal a pulse of oxygen is released to the patient by an electronically controlled flow valve There is no mixing of oxygen with room air e Range of Operation Increments of 1 6 times x the bolus of 16 mL 16 to 96 mL per Settings breath e A
12. 4 F 28 C dew point Transport Storage 4 F to 140 F 20 C to 60 C Temperature Humidity Up to 95 Non Condensing Electrical Use no extension cords or electrical outlets controlled by a switch 0 13 123 feet 4 000 meters Placement DO NOT block air inlet or bottom exhaust vent keep at least 3 inches away from walls draperies furniture etc Smoke pollutant and fume free Operating Time Up to 24 hours per day when connected to external power When moving the ECLIPSE from an extreme environment allow time for your ECLIPSE to acclimate to the recommend operating environment WARNING Operating your ECLIPSE outside the recommended operating environment may impact performance cause damage and will void the warranty CAUTION Locate the unit in a well ventilated room that provides adequate air flow and check if air is being prevented from entering or exiting the system by furniture draperies bedding clothing or pets The Proper Location Select a location that avoids fumes and pollutants and allows the oxygen system to draw in room air through the air inlet filter on the back of the cabinet and exhaust air out the vent at the bottom of the case without being restricted Locate the unit such that the alarms can be heard DO NOT cover unit with a blanket towel quilt or other covering as the unit may overheat DO NOT locate near flammable materials or in the direct path of any heat source suc
13. IPSE charges the optional Power Cartridge when external power AC or DC is present and when the Power Cartridge temperature is below a safe charging temperature When external power is disconnected the system will automatically switch over to the Power Cartridge if installed When external power is restored by plugging into AC or DC outlets the system will accept power from the external source and replenish the Power Cartridge while operating if adequate power is available Connect the AC or DC Power Supply by aligning the connector with the dot on the receptacle on the unit Disconnect by grasping the plug twist counterclockwise and pull Connect the AC Power Cord to a grounded power outlet File Name Doc Number Revision Page 6112_X5 6112 X5 140f 35 WARNING DO NOT use any cord if damaged Improper use of the power cord may result in a fire or electrical shock hazard WARNING DO NOT touch the Power Cartridge cavity connectors with the Power Cartridge removed while the unit is connected to an external power source This may result in electrical shock 1 AC POWER SUPPLY The ECLIPSE System includes a universal AC Power Supply for use at home or where standard AC power is available To plug into AC make sure the AC Power Supply cord is securely inserted in the power receptacle on the side of the unit and plug the AC Power Supply into an AC outlet When the unit is plugged in properly a green
14. MAKES IT EASIER FOR PHYSICIANS TO SELECT A MODE OF OXYGEN THERAPY THAT SUITS THE PATIENT S LIFESTYLE THE ECLIPSE MAY BE USED IN HOME INSTITIUTION AND MOTOR VEHICLES YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS THIS FLOW RATE IS DETERMINED BY THE AMOUNT OF OXYGEN YOU NEED RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN File Name Doc Number Revision Page 6112_X5 6112 X5 9 of 35 UNDERSTANDING THE CONTROLS AND FRONT PANEL DESCRIPTION OF THE ECLIPSE The key features of the ECLIPSE Oxygen System and control panel are important items to become familiar with Control Panel Service Port Not for Patieni Use Handle Air Inlet Filter Oxygen Outlet Port Cart Insert Screw Power y Cartridge Simplified User Y Instructions Power Under Base Cartridge Latch Exhaust Vent Main Rating Label Front External Power Back Under Base Receptacle Input Output Connections Oxygen Outlet Port Extension tubing or a cannula is attached to this port Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the top back of the unit This filter prevents dust and debris from entering the Eclipse and should be cleaned regularly Service Port This connection allows service personnel to diagnose and monitor Eclipse performance It is not for patient use External P
15. NDED OPERATING ENVIRONMENTS cccccccscecccecsesessssecesececsesesaaeceeececeesesssaeeeeeeeeeesestsaeees 13 The Proper Locations isiciccieh niece a ds dais 13 OPERATING INSTRUCTIONS 35322555 5335255 050500 aE EE aed es 14 Before Operating oca aaa 14 Powering your ECLIPSE Systeman nerep ie etteavednyaendui ete tes 14 Operating your ECLIPSE Syste cetro asin rari ee rea ian ea dotadas 18 ALARM CONDITIONS AND TROUBLESHOOTING c ccccsssscccsesssceceessececeesseeeceeseeeceeseeeceesseecsenteeeeeens 21 ACCESSORIES an nn en RO E E AE Os ELISEO cai DITA 23 CLEANING CARE AND ROUTINE MAINTENANCE 0 cccccccesscecssecessceceeceecseeeessecsesseceeeeeseeessasesesaeenaes 25 TECHNICAL DATA Sees hc Sa ee SEs ae BE RI ORE ER vb Bev iia 27 SUMMARY OF ALL WARNINGS AND CAUTIONS cccccsessscecccecsessssaeceeceeesesesssaeeeeececeesesaaeeeeeesesenerssaeees 29 ELECTROMAGNETIC COMPATIBILITY ccccccccccssssscecssscesecssssesecssssececssssesecsssseeecsssaececsesseeeesesseeeesenaes 34 DISPOSAL OF EQUIPMENT AND ACCESSORIES ccccccssssscecssssscecsesseeccsesseeccsesseeecessseeccsesseeecsesseeeesenaes 35 File Name Doc Number Revision Page 6112 X5 6112 X5 2 of 35 Important Safety Instructions Important Safety Instructions are shown Under the Following Headings SeQual Technologies Inc SeQual recommends that this User Manual be given to each patient receiving an Eclipse Oxygen System
16. SeQual Technologies Inc ECLIPSE Model 1000 User Manual eclipse PN 6112 Rev X5 01 06 REV CO No DESCRIPTION REL DATE PREP BY APPR BY X1 D017 Initial Design Release 11 16 01 SWB PJJ X2 D120 Update manual to remove 11 22 04 WB WJB facemasks X3 D250 Update to current design 8 4 05 PJ PJJ X4 D319 Update to current design 12 23 05 PJJ PJJ X5 D324 SCR001 Pulse Mode 1 5 06 WJB PSA breath detection File Name Doc Number Revision Page 6112_X5 6112 X5 1 of 35 CONTENTS IMPORTANT SAFETY INSTRUCTIONS 0 c ccccccsssssecssscesecssssececssssececssssececssssececsssseeecsssseeecssaseeeessasseeeseaas 3 SVMBOMS soi veces TON 3 INDICATIONS FOR USE riiete Di ts dan abe Ba bese Sua MG Le Gs Ra dt a Bebe 5 CONTRAINDIGATIONS iii aa ada Hataet tides 5 SAFETY GUIDELINES oia ii a ath Lats Wadena ia ad 6 Oxygen Therapy Salesiana da 6 Operatine aleatoria 7 Power Cartridge Salidas ida ileso paisas 8 INTRODUCTION TO YOUR ECLIPSE OXYGEN SYSTEM cccccccsssscecessssceceesseeecessseeccessseecesesseeesenaes 9 UNDERSTANDING THE CONTROLS AND FRONT PANEL cccccesssscecsessscecessseeeceessececsesseecsesseeesenaes 10 Description of the ECLIPSE eerror creien iarere a A tar lean anaes 10 Control Panel Features iaa nit n sada E de edi bas 11 Us Cont tinca 11 ErontPanelICON Sad 11 System Statis Indicators evocando a aa ida lec e T ee i aeae 12 RECOMME
17. act your dealer provider Low Oxygen Concentration Restriction in tubing or Repair or replace tubing or humidifier humidifier Filter restricted Clean or replace air inlet filter Malfunction Contact your dealer provider Low Oxygen Flow Restriction in humidifier or Repair or replace humidifier or tubing tubing Filter Blocked Clean or replace air inlet filter Malfunction Contact your dealer provider No Oxygen delivered in Pulse Tubing cannula longer than 7 feet Attach 7 foot 2 1m Mode 2 1m tubing cannula Humidifier attached Remove humidifier No inspiration detected Contact your dealer provider Power Cartridge Status Gauge never indicates fully charged Power Cartridge needs calibration Fully discharge the Power Cartridge and recharge File Name 6112_X5 Doc Number Revision 6112 X5 Page 22 of 35 ACCESSORIES Use of certain oxygen administration accessories not specified for use with WARNING this oxygen system may impair the performance ONLY use recommended oxygen delivery accessories There are many different types of oxygen tubing cannula and humidifiers The following accessories are recommended to be used with your ECLIPSE Oxygen System Refer to the Maintenance section of this manual for accessory maintenance instructions Humidifier Salter Labs Bubble Humidifier Part Number 7600 or equivalent If your p
18. ccuracy of Delivered 15 Bolus e Breathing Frequency range 6 to 25 breaths per minute average sustainable e Minimum time between 1 25 seconds maximum 3 consecutive breaths breaths e Response to missing Eclipse switches to Continuous Flow Mode after 30 seconds breaths without breath detection File Name Doc Number Revision Page 6112_X5 6112 X5 28 of 35 SUMMARY OF ALL WARNINGS AND CAUTIONS WARNING FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN WARNING THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE SUSTAINING APPLICATIONS NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES WARNING IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING FOR ADULT USE ONLY NOT QUALIFIED FOR USE ON PEDIATRIC PATIENTS WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS WARNING Warning No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Oxygen enhances the combustion of flammable material For safety concerns all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device a
19. ce or accessories on wet surfaces or in standing water and DO NOT submerse or expose to water Electrical shock and serious physical injury may result WARNING Protect electrical cords from sharp edges Electrical shock and serious physical injury may result WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied with your ECLIPSE Oxygen System Use of a non SeQual Power Supply may be hazardous cause damage to your oxygen system and will void the warranty CAUTION Availability of an alternate source of oxygen is recommended in case of power outage or mechanical failure CAUTION DO NOT expose unit to temperature outside of specified operating or storage temperatures as this may damage the unit File Name Doc Number Revision Page 6112 X5 6112 X5 7 of 35 CAUTION Locate oxygen tubing and Power Supply cords to prevent tripping hazards CAUTION DO NOT place ECLIPSE in a small enclosed space such as a bathroom or closet with cannula leading out of the enclosed space POWER CARTRIDGE SAFETY WARNING This Power Cartridge may explode and cause potential injury if exposed to or disposed of in a fire WARNING DO NOT short circuit the Power Cartridge s metal contacts with metallic objects such as keys or coins It may cause sparks or excessive heat WARNING DO NOT disassemble p
20. d If the oxygen system is used in a dusty environment the filter may need to be replaced more often You should not operate the ECLIPSE without a filter installed for longer than 30 minutes Cabinet AC Power Supply DC Power Supply and Power Cord Use a damp not soaking wet clothe or sponge and a mild detergent solution to clean the cabinet and power supplies WARNING Unplug Power Cords AC or DC Power Supplies before cleaning the exterior File Name Doc Number Revision Page 6112 X5 6112 X5 25 of 35 cabinet DO NOT use denatured alcohol or apply liquid spray or aerosol cleaners Humidifier If used clean your humidifier daily to reduce possibility of contamination Replace humidifier monthly or as recommended by the manufacturer or your dealer provider Cart Before cleaning the cart remove the ECLIPSE Use a damp not soaking wet cloth or sponge and mild detergent solution to clean the Cart Power Cartridge Cleaning Care and Maintenance The Power Cartridge in the ECLIPSE Oxygen System requires special care to assure a longer life and higher performance The SeQual Power Cartridge is the only approved Power Cartridge to be used Once a month completely discharge your Power Cartridge and fully recharge to calibrate the Power Cartridge Use a damp cloth or sponge and mild detergent to clean the Power Cartridge case and the latch WARNING DO NOT tamper with or tr
21. e other device s are connected e Consult the manufacturer or service technician for help File Name Doc Number Revision Page 6112_X5 6112 X5 34 of 35 DISPOSAL OF EQUIPMENT AND ACCESSORIES Follow local governing ordinances and recycling plans regarding disposal of device components DO NOT dispose of the Power Cartridge Power Cartridges should be returned to your dealer provider Manufactured by SeQual Technologies Inc 11436 Sorrento Valley Road San Diego CA 92121 USA 858 202 3100 800 826 4610 www sequal com European Union Authorized Agent MeddiQuest Limited Business amp Technology Centre Bessemer Drive Stevenage SG 2DX United Kingdom Tel 44 1763 222505 Fax 44 1763 222049 Dealer provider Name and Address Physician Name and Address SeQual and ECLIPSE are registered marks of SeQual Technologies Inc San Diego CA 92121 Salter Labs is a registered mark of Salter Labs Arvin CA 92303 File Name Doc Number Revision Page 6112_X5 6112 X5 35 of 35
22. efly goes off Only change your flow settings at the recommendation of your physician A a Increase or Decrease Flow Rate Button Use these buttons to set the flow rate to your prescribed level File Name Doc Number Revision Page 6112 X5 6112 X5 11 of 35 System Status Indicators The ECLIPSE control panel displays important operating information This section will help you understand these features Green Yellow and Red indicators on the front panel indicate the operating condition of the system For more information on the indicators see the Alarm Conditions and Troubleshooting table ON OFF Green Indicator System OK Indicates that the ECLIPSE is ON and operating properly ALERT Yellow Indicator Low and Medium Priority Alerts When steadily illuminated this indicates a low priority awareness condition or Caution Continue to use your system and refer to the Troubleshooting section for the proper response A flashing yellow indicates a medium priority alert Prompt response is needed to deal with a condition Refer to the Troubleshooting section gt g A ALARM Red Indicator High Priority Alarms Indicates a high priority alarm condition Immediate response is necessary to deal with the condition Refer to the Troubleshooting section and contact your dealer provider Buzzer An audible alarm or buzzer is used to alert you of the condition of the unit e
23. ernal contact dealer provider a power present Loss of External Seek an alternative oxygen Z Power without ay supply until external power is 3 charged Sh Off Off Blinking On restored or install a charged Cartridge installed Power Cartridge Z Contact dealer provider Use System Fault Off Off On On back up oxygen supply AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS RECOMMENDED IN WARNING CASE OF POWER OUTAGE OR MECHANICAL FAILURE CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK UP SYSTEM REQUIRED DO NOT IGNORE ALARMS File Name Doc Number Revision Page 6112 X5 6112 X5 21 of 35 ALARM CONDITIONS AND TROUBLESHOOTING CON T SYMPTOM PROBABLE CAUSE REMEDY ECLIPSE does not power on when On OFF button is pressed No Power Cartridge installed Install Power Cartridge or plug into external power Power Cartridge is discharged Plug into external power Malfunction Contact your dealer provider No Oxygen Blocked or kinked tubing Check tubing or replace Filter Blocked Clean Air Inlet Filter Humidifier Check humidifier attachment and tubing ECLIPSE not ON Turn ECLIPSE ON Tubing or cannula is not properly connected or kinked Check tubing cannula and connections Power Cartridge is discharged or no external power is present Install charged Power Cartridge or attach external power Malfunction Cont
24. ettings should be determined for each patient individually Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode WARNING When operating in Pulse Flow Mode to ensure adequate breath detection the length of the tubing cannula must not exceed 7 feet 2 1m WARNING As with all conserving devices the ECLIPSE may not be able to detect some respiratory efforts Step 5 Adjust Flow Rate to Prescribed Level e Using the Flow Rate Setting buttons adjust the flow rate to the prescribed setting supplied by your physician WARNING DO NOT change from prescribed flow rate setting without consulting with a qualified clinician Step 6 Begin Use e Breathe normally through the cannula e You will get longer operating time on your Power Cartridge if you operate your ECLIPSE in Pulse Flow Step 7 Cart Operation Oo The Cart is designed for use on pedestrian surfaces e Align the locator holes on the bottom of the unit to the tabs on the cart e Align the screw to the insert on the back of the ECLIPSE and hand tighten the knob on the cart e Push the button on the cart handle to adjust the height File Name 6112 X5 Doc Number 6112 Revision X5 Page 20 of 35 ALARM CONDITIONS AND TROUBLESHOOTING The table shows possible alarm conditions and suggests appropriate responses If any other alarms occur call your authorized SeQual dealer provider
25. h as a heat register or car heater Keep the oxygen system at least five 5 feet 1 5 m away from hot sparking objects or open sources of flame File Name Doc Number Revision Page 6112_X5 6112 X5 13 of 35 OPERATING INSTRUCTIONS BEFORE OPERATING Your SeQual dealer provider or authorized service representative will demonstrate how to operate the oxygen system This manual serves as a supplemental reference to help you operate and maintain the unit If you have any further questions or problems call your SeQual dealer provider or authorized service representative Important DO NOT attempt to operate the ECLIPSE System without first reading the Safety Guidelines section of this manual Follow all operating instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily injuries and damage to the oxygen system observe all safety precautions WARNING Possible Shock Hazard Protect the ECLIPSE and AC DC Power Supplies from fluid spills or drips CAUTION Always check to see that the Air Inlet and the exhaust vent are not blocked and the air inlet filter is clean before using your system POWERING YOUR ECLIPSE SYSTEM The ECLIPSE is a lightweight portable oxygen system capable of being operated directly from three different power sources 1 an AC Power Supply 2 a DC Power Supply or 3 a rechargeable Power Cartridge The ECL
26. hysician has prescribed a humidifier follow the manufacturer s instructions for use Attach the humidifier to the oxygen outlet port of the oxygen concentrator The O symbol marks the oxygen outlet port on the concentrator Use of humidifiers not specified for this oxygen system may impair performance and void the warranty DO NOT use a humidifier in Pulse Flow Mode The ECLIPSE will not detect an inspiration and will default to Continuous Flow Mode Humidifier Adapter SeQual Part Number 3509 If your physician has prescribed a humidifier you may need to use the SeQual Humidifier Adapter Follow the instructions for use Attach the Humidifier Adapter to the oxygen outlet port of the oxygen system and then to the humidifier Oxygen Tubing Salter Labs Series 2000 or equivalent Tubing Internal diameter of the tubing should be no less than 3 16 0 48 cm Connect the oxygen tubing to the outlet port of the humidifier or directly to the oxygen outlet port if you do not use a humidifier Connect the other end of the tube to the nasal cannula if not already attached Tubing not specified for use with this oxygen system may impair the performance WARNING The tube fitting may be tight DO NOT use oil or grease to lubricate WARNING If tubing becomes disconnected during operation ECLIPSE does not alarm Check flow at the cannula to verify oxygen delivery CAUTION To ensure there is adequate flow to deliver oxygen the
27. indicator on the AC Power Supply will light and the External Power Present Indicator o appear on the control panel CAUTION The AC power cord used with the ECLIPSE Oxygen System must meet the requirements of the country where the product is used CAUTION Use with grounded 3 pin receptacle DO NOT remove ground pin from cord Use only with SeQual supplied cords CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical outlet controlled by a switch 2 DC POWER SUPPLY A DC Power Supply allows the system to operate from DC accessory outlets such as those found in motor vehicles You need to determine the Fuse rating in your motor vehicle either 15 AMP or 20 AMP Consult your vehicle s Operator s Manual or look at your fuse box in the vehicle Set the switch on the DC Power Supply to the appropriate one Use the 10 AMP position on the DC Power Supply for a 15 AMP vehicle fuse and use the 15 AMP position for a 20 AMP or greater vehicle fuse Start your vehicle Insert the DC cord in the proper receptacle on the side of the unit Next insert the DC Power Supply into the DC accessory outlet in your vehicle When the system is properly plugged in and receiving power from the DC source a 0 indicator light on the Power Supply and the External Power Present indicator on the control panel will appear WARNING Secure the ECLIPSE and the DC Power Supply in your vehicle and make sure the air inlet and e
28. ither a warning or failure and to confirm a valid key press 2 0 Flow Rate Setting Indicator Green This is the main focus on your control panel Your dealer provider will set up your prescribed flow rates for your Continuous Flow Mode and your Pulse Flow Mode if pulse flow has been prescribed DO NOT change these settings unless it is under the order of your physician Each time you turn the unit on the last flow mode and flow setting will be used Power Cartridge Status Gauge This indicator displays the charge remaining in the Power Cartridge When the Power Cartridge is fully charged all 10 bars are illuminated Each bar represents approximately 10 of the total Power Cartridge charge When the Power Cartridge is being charged the charge indicator bars will blink in a waterfall fashion If the Power Cartridge is not installed or if it is improperly installed the Status Gauge will not light External Power Present Indicator When the ECLIPSE system is properly plugged in using the AC or DC Power Supply this indicator will appear on the Control Panel File Name 6112_X5 Doc Number Revision Page 6112 X5 12 of 35 RECOMMENDED OPERATING ENVIRONMENTS The following information recommends certain boundaries that will allow the ECLIPSE Oxygen System to operate at its full potential Operating 50 F to 104 F 10 C to 40 C Temperature Operating Humidity 10 95 at a 82
29. ivery WARNING DO NOT tamper with disassemble crush or heat the Power Cartridge above 140 F 60 C The Power Cartridge may present a risk of fire or explosion and will void the warranty WARNING DO NOT expose the AC and DC Power Supplies to water electrical shock or serious injury may result WARNING DO NOT disassemble the ECLIPSE or attempt to perform any routine maintenance other that what is described in the Patient Cleaning and Care section below Disassembling the oxygen system may create a hazard of electrical shock and void your warranty Contact your dealer provider for any maintenance required WARNING Unplug Power Cords AC or DC Power Supplies before cleaning the exterior cabinet DO NOT use denatured alcohol or apply liquid spray or aerosol cleaners WARNING DO NOT tamper with or try to repair the Power Cartridge No serviceable parts inside WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury CAUTION Availability of an alternate source of oxygen is recommended in case of power outage or mechanical failure CAUTION DO NOT expose unit to temperature outside of specified operating or storage temperatures as this may damage the unit CAUTION Locate oxygen tubing and Power Supply cords to prevent tripping hazards CAUTION DO NOT place ECLIPSE in a small enclosed space such as a bathroom or closet with cannula leading out of the enclosed space File Name Doc Number Revision Page 61
30. l grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source DO NOT use any petroleum based or other lubricants File Name 6112_X5 Doc Number Revision Page 6112 X5 6 of 35 OPERATING SAFETY Keep oxygen system and cord away from hot surfaces DO NOT use the oxygen system if it has a damaged cord or plug contact your dealer provider Your oxygen system should be located in a well ventilated area to allow for adequate air intake and to avoid the intake of airborne pollutants and fumes Only use this device with accessories designed or recommended by SeQual Technologies Inc If oxygen system has been dropped damaged or exposed to water contact your dealer provider for examination or repair DO NOT drop or insert any object into any opening DO NOT block the air inlet or the exhaust vent of the oxygen concentrator when it is on a soft surface such as a bed or couch or car seat DO NOT cover unit with a blanket towel quilt etc DO NOT overfill the humidifier if used Fill with water only to the level shown by the manufacturer When using ECLIPSE in a vehicle make sure it is properly secured or restrained WARNING DO NOT remove the cover no serviceable parts inside Qualified service personnel should only remove the cover WARNING DO NOT operate devi
31. mmediately disconnect the DC Power Supply from the motor vehicles accessory outlet if a low voltage alarm sounds 3 RECHARGEABLE POWER CARTRIDGE The ECLIPSE system can also be powered by arechargeable Power Cartridge Installing the Power Cartridge Align the Power Cartridge so that it will slide into the Power Compartment Push the Power Cartridge into the Power Compartment until it is secure and flush with the back of the system When the Power Cartridge is properly inserted the Power Cartridge Status Gauge will appear on the Control Panel If you DO NOT have a Power Cartridge installed the Status Gauge will not light up File Name 6112 X5 Doc Number 6112 Revision X5 Page 16 of 35 To remove the Power Cartridge push up on the release lever on the Power Cartridge The Power Cartridge will slide out so that you can grasp it for removal Power Cartridge Capacity The capacity of the Power Cartridge is lit up on the control panel A variety of factors such as flow rate and Pulse or Continuous Flow Mode operation determine the duration of operating time The following table provides estimates of amount of time that the ECLIPSE can operate on a fully charged Power Cartridge given certain flow rates and operating conditions TYPICAL NEW POWER CARTRIDGE OPERATION TIME Power Cartridge LJ Continuous Pulse Flow FLOWRATE Flow Mode Mode SETTING LPM Setting 0
32. nd void the warranty WARNING Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source DO NOT use any petroleum based or other lubricants WARNING DO NOT remove the cover no serviceable parts inside Qualified service personnel should only remove the cover WARNING DO NOT operate device or accessories on wet surfaces or in standing water and DO NOT submerse or expose to water Electrical shock and serious physical injury may result WARNING Protect electrical cords from sharp edges Electrical shock and serious physical injury may result WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied with your ECLIPSE Oxygen System Use of a non SeQual Power Supply may be hazardous cause damage to your oxygen system and will void the warranty WARNING This Power Cartridge may explode and cause potential injury if exposed to or disposed of in a fire WARNING DO NOT short circuit the Power Cartridge s metal contacts with metallic File Name 6112_X5 Doc Number Revision 6112 X5 Page 29 of 35 objects such as keys or coins It may cause sparks or excessive heat
33. oom in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty Use No Oil or Grease Indicates important information about the possibility of spontaneous and violent ignition that may occur if oil grease or other petroleum based substances come into contact with oxygen under pressure Oo O PE Drip Proof Equipment IPX1 Protection of liquid ingress 1PX1 per IEC IPX1 60529 TEC Icon for DC Power Input File Name Doc Number Revision Page 6112 X5 6112 X5 3 of 35 Type B Equipment Equipment providing a particular degree of protection against electric shock particularly regarding Allowable LEAKAGE CURRENT Reliability of the protective earth connection Amperes Alternating Current Audible Alarm An audible indicator is provided to announce a reduction in oxygen concentration or when the power has been interrupted or lost ON OFF Turns the unit ON and OFF but does not directly disconnect the mains power Increase Flow Setting Increases the flow setting in 0 5 LPM increments each time the button is pushed Decrease Flow Setting Decreases the flow setting in 0 5 LPM increments each time the button is pushed Flow Mode Activates either Continuous Mode or Pulse Mode Operation External Powe
34. ower Either the Eclipse AC or DC power adapter output cords may be Receptacle connected to this receptacle Exhaust Vent Exhaust air from the Eclipse is expelled from this vent File Name Doc Number Revision Page 6112 X5 6112 X5 100f35 Control Panel Features The ECLIPSE displays important operating information using the icons numbers and visual indicators on the control panel This section will help you understand these features ON OFF Alert Alarm Indicator green Indicator yellow Indicator red ON OFF LPM Flow Rate Setting Button 2 digits green Increase Flow Rate Button No Smoking Icon Flow Mode Decrease Flow Rate Button Button Pulse Flow External Power Present Mode Indicator Indicator Power Cartridge Status Gauge green User Controls This section describes the control buttons and function for your ECLIPSE Oxygen System Front Panel Icons ON OFF Button and Indicator This button turns the system ON or OFF The Green Indicator is lit when the unit is ON Flow Mode Button and Indicator The ECLIPSE contains a button to toggle between Continuous Flow and Pulse Flow mode Pulse Flow mode allows a significant increase in the amount of time the system can operate from the Power Cartridge When the Pulse Flow Mode is activated the green Flow Mode indicator goes on and a pulse of oxygen is delivered upon inhalation When the pulse of oxygen is delivered the indicator bri
35. ower Supply Model 6104 Your AC Power Supply enables the ECLIPSE to be connected to a grounded AC electrical outlet with the country specific power cord Please contact your dealer provider for the country specific power cord DC Power Supply Model 6105 Your DC Power Supply enables the ECLIPSE to be operated in an automobile or other vehicle with an accessory outlet Contact your dealer provider to set up the DC Power Supply for proper use in your own automobile The indicator on the AC and DC Power Supplies will be either green or red Green indicates normal operation and red indicates abnormal operation WARNING DO NOT expose the AC and DC Power Supplies to water Electrical shock or serious injury may result Cart Your ECLIPSE Oxygen System can be easily transported while attached to the Cart The cart has wheels and a telescoping handle for easy mobility CAUTION When operating the ECLIPSE on the cart DO NOT bang drop or bump the unit while in transport File Name Doc Number Revision Page 6112_X5 6112 X5 24 of 35 CLEANING CARE AND ROUTINE MAINTENANCE Routine Maintenance A trained technician will carry out all routine maintenance at least once a year or as needed Your dealer provider will perform inspections and maintenance of the compressor filters alarms 9V battery Power Cartridge and other internal parts as needed DO NOT disassemble the ECLIPSE or a
36. r Present Indicator Indicates the presence of external power to operate the concentrator Power Cartridge Status Shows the amount of charge remaining in the Power Cartridge Power Cartridge Symbol The symbol on the Power Cartridge pack Oxygen Output The port in which the oxygen flows out and where the tubing is attached File Name 6112_X5 Doc Number Revision Page 6112 X5 4 of 35 INDICATIONS FOR USE THE ECLIPSE OXYGEN SYSTEM PROVIDE ANY PATIENT MONITORING CAPABILITIES YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A IS INDICATED FOR THE ADMINISTRATION OF SUPPLEMENTAL OXYGEN THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT NOR DOES IT PHYSICIAN WARNING FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN CONTRAINDICATIONS WARNING THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE SUSTAINING APPLICATIONS NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES WARNING IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING FOR ADULT USE ONLY NOT QUALIFIED FOR USE ON PEDIATRIC PATIENTS WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS File Name Doc Number
37. rt recharging If the Power Cartridge is not present during a power interruption the ECLIPSE Oxygen System will stop operation When power is restored turn the system back ON to resume oxygen production File Name Doc Number Revision Page 6112_X5 6112 X5 18 of 35 OPERATING YOUR ECLIPSE Step 1 Step 2 Step 3 Step 4 Locating your ECLIPSE Place the ECLIPSE in a well ventilated place so that the air inlet and exhaust have clear access Position the ECLIPSE so that any alarm can be heard Check to see if the air inlet filter is in place before using If the air inlet filter needs to be cleaned wash with soapy water and allow drying before using If necessary replace with a new air inlet filter Plug unit into AC Power DC Power or make sure there is a charged Power Cartridge installed Turn On the Unit and Allow to Warm Up Press the ON OFF Button to turn ON your ECLIPSE A brief self test will happen All the indicators are lit and a 3 second buzzer sounds indicating the system is in working order Warm up Period Oxygen Monitoring The ECLIPSE Oxygen System has an oxygen monitoring system with a yellow alert and red alarm visual indicator and buzzer The buzzer will activate in the event of an internal fault or power outage external power with no Power Cartridge installed The red yellow and green all OK indicators will initially light up and then the red and yellow indica
38. tors sequentially go out when the oxygen purity is within specification After initially turning ON the unit allow up to five 5 minutes for the unit to reach its specified performance The control panel will indicate the unit is warming up When the green ON indicator is lit the system is ready for use Connect the tubing to the Oxygen Outlet and the cannula per the instructions from your Dealer provider Replace the cannula and tubing on a regular basis Consult your dealer provider for cannula and tubing cleaning and replacement information Select Flow Mode Press the Flow Mode button to select the flow delivery mode Continuous Flow Mode Operation o When operating in Continuous Flow Mode a continuous supply of oxygen will flow through your tubing and nasal cannula WARNING When operating in Continuous Flow Mode to ensure adequate flow DO NOT use a tubing cannula extension more than 50 15 2m long File Name Doc Number Revision Page 6112_X5 6112 X5 190f 35 e Pulse Flow Mode Operation Pulse Flow Mode should only be used under the direction of a qualified clinician or physician When operating in Pulse Flow Mode an alarm will beep after 15 seconds if inspiration is not detected If another 15 seconds elapses and no inspiration is detected the unit will switch to Continuous Flow Mode at the last Continuous Flow Mode rate setting used O O WARNING Pulse Flow s
39. ttempt to perform any routine WARNING maintenance other that what is described in the Patient Cleaning and Care section below Disassembling the oxygen system may create a hazard of electrical shock and void your warranty Contact your dealer provider for any maintenance required Patient Cleaning and Care Cannula Replacement Replace your tubing and cannula on a regular basis as recommended by your dealer provider Your physician or dealer provider will provide you with cleaning and replacement information Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the top back of the unit Cleaning the air inlet filter is the most important activity you will perform to keep your oxygen system performing properly Check this air filter daily and clean it at least once a week Remove the filter Wash the filter in warm water using a mild detergent solution Rinse the filter thoroughly and squeeze out the excess water Allow the filter to air dry Reinsert the filter in the cabinet APN CAUTION The filter should be free of liquid water and dry to the touch before reinstallation as excess moisture may impair the proper operation of the oxygen system CAUTION When the filter is removed the Data Port is exposed DO NOT attach any cables to this Data Port It is for Service ONLY CAUTION DO NOT poke objects through the vent holes The air inlet filter should be replaced as neede
40. uncture or crush the Power Cartridge Power Cartridge electrolytes may be toxic if swallowed and can be harmful to skin and eyes Keep the Power Cartridge away from children WARNING Use of a damaged Power Cartridge may cause personal injury WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury CAUTION DO NOT expose the Power Cartridge to temperatures above 140 F 60 C such as in a vehicle parked in the sun or on a hot day CAUTION Charging the Power Cartridge below 50 F 10 C or above 104 F 40 C may degrade the performance CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock CAUTION Only use the Power Cartridge for its intended purpose File Name Doc Number Revision Page 6112 X5 6112 X5 8 of 35 INTRODUCTION TO YOUR ECLIPSE OXYGEN SYSTEM This User Manual will inform you about the use and care of the ECLIPSE Oxygen System and its accessories Thoroughly read all the information in this manual in its entirety before operating your ECLIPSE Contact your dealer provider for proper training in the use of this device YOUR PHYSICIAN HAS PRESCRIBED SUPPLEMENTAL OXYGEN AS PART OF A TREATMENT PLAN LONG TERM OXYGEN THERAPY IS NOW WIDELY ACCEPTED AS THE STANDARD TREATMENT FOR CHRONIC HYPOXIA CAUSED BY CHRONIC OBSTRUCTIVE PULMONARY DISEASE OXYGEN THERAPY HAS BECOME MORE VERSATILE IN RECENT YEARS DUE TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS THE ECLIPSE
41. while in transport CAUTION The filter should be free of liquid water and dry to the touch before reinstallation as excess moisture may impair the proper operation of the oxygen concentrator CAUTION When the filter is removed the Data Port is exposed DO NOT attach any cables to this Data Port It is for Service ONLY CAUTION DO NOT poke objects through the vent holes File Name Doc Number Revision Page 6112_X5 6112 X5 33 of 35 ELECTROMAGNETIC COMPATIBILITY This equipment has been tested and found to comply with the limits for medical devices to the IEC60601 1 2 2001 09 Electromagnetic Compatibility standard These limits are designed to provide reasonable protection against harmful interference in a typical medical installation This equipment generates uses and can radiate radio frequency energy and if not installed according with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving device e Increase the separation distance between the equipment e Connect the equipment into an outlet on a circuit different from that which th
42. xhaust airflow is not blocked This may impact performance WARNING DO NOT leave the ECLIPSE or DC Power Supply plugged into the vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle This may drain the vehicle s battery A buzzer in the DC Power Supply is intended to warn user when low vehicle power is present Immediately disconnect the DC Power Supply from the vehicle for low voltage conditions WARNING If the vehicle power source vehicle accessory connector drops below 11 5 Volts brownout the ECLIPSE will revert to Power Cartridge operation if present The External Power Present Indicator may remain on File Name Doc Number Revision Page 6112_X5 6112 X5 15 of 35 TYPICAL FLOW RATES AVAILABLE FOR EACH FUSE SETTING 10 AMP 15 AMP Setting Continuous Pulse Continuous Pulse 0 5 Yes Yes 1 0 Yes Yes Yes Yes 1 5 Yes Yes Yes Yes 2 0 Yes Yes Yes Yes 2 5 NO Yes Yes Yes 3 0 NO Yes Yes Yes 3 5 Yes Yes 4 0 Yes Yes 4 5 Yes Yes 5 0 Yes Yes 5 5 Yes Yes 6 0 Yes Yes Actual performance may vary dependent upon usage conditions and vehicle When operating in the 10 AMP switch position flow rate settings above 2 0 LPM continuous flow will automatically switch to 2 0 LPM continuous flow when the DC Power Supply is connected I
43. y to repair the Power Cartridge No serviceable parts inside Power Cartridge Disposal Your Power Cartridge is rechargeable and can be recycled Always return it to your dealer provider for proper disposal Power Cartridge Storage Your Power Cartridge should be stored in a cool dry location WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury File Name Doc Number Revision Page 6112_X5 6112 X5 26 of 35 TECHNICAL DATA 1 to 6 setting in Pulse Flow Mode Flow Rate Settings 0 5 to 3 0 LPM in Continuous Flow Mode Oxygen Concentration 90 3 at sea level for all flow setting Oxygen Concentration Green Light Normal Operation Indicator Yellow Light Warning or Caution less than 85 3 Red Light Flashing Abnormal Operation less than 70 5 Maximum System Pressure 15 psig 103 5 kPa Oxygen Outlet Pressure 5 0 psig 34 5 kPa nominal Electrical Power AC Operation 100 240 VAC 50 60 Hz DC Operation 12V Power Cartridge 195 W Hrs AC Power Supply 200 Watts Max DC Power Supply 150 Watts Max Operating Temperature 50 F to 104 F 10 C to 40 C Operating Humidity 10 95 at 82 4 F 28 C dew point 48 dBA at 3 0 LPM in Continuous Flow Mode 40 dBA at 3 0 LPM in Pulse Flow Mode Dimensions H x W x D 19 3 inches x 12 3 inches x 7 1 inches 49 0 cm x 31 2 cm x 18 0 cm 17 4 pounds 7 9 kg with Power Cartridge
Download Pdf Manuals
Related Search
Related Contents
Digital HD Video Camera Recorder EOLO Star 24 kW - Immergas S.p.A. Definitive Technology UIW RCS II User's Manual 693 - Novy Philips Notebook SoundBar SPA5210B PZ-N50CFM の取扱説明書を見る Inversor / Cargador de 750W Serie X PowerVerter® APS (marzo 2012).- Uso del libro de averías informático. Copyright © All rights reserved.