MD300I-P Operation Manual

image

Contents

1. 14 INSTRUCTION MANUAL 4 3 2 Exempt and limitation a Our company isn t responsible for such damage caused by force majeure For example fire thunder flash flood cyclone hail earthquake house collapse commotion plane failing and traffic accident deliberate damage lack of fuel or water labor and capital bother strike and stop working etc b No service offer The corresponding charge and insurance charge of disassembling refurbishing repackaging and moving the oximeter or the part of it The damage caused by the third company not commended by our company to adjust install replace the parts of the oximeter The damage and failure caused by user or its representative doesn t comply with the operator s manual c The oximeter is installed or connected with such external device without our company permission as printer computer netline and lead to oximeter failure Our company will charge for the maintenance d Responsibility limitation During the period of maintenance contract validity if user changes the parts manufactured by other manufacturers without our company permission our company is entitled to stop contract 4 3 3 User Guarantee a Please read user manual carefully before operation b Please operate and make daily maintenance as request of manual and guarantee c Power supply and environment 4 3 4 No guarantee principle There is no dispelled smut and not original mark in the crust
2. There is physical damage on oximeter and its accessory There are liquid leftover and eyewinker on oximeter which may lead to short circuit and plug board failure All the probe and accessories belong to consumption and beyond free change range Such damage of probe caused by mechanical force doesn t belong to free change range During measurement of SpO2 principle leads to measuring value difficultly or inaccurate measurement Not original package leads to damaging oximeter during transportation Not our company professionals or authorized personnel disassemble oximeter and lead to oximeter failure Not carefully read manual and so wrong operation lead to oximeter damage and failure 15 PULSE OXIMETER 4 3 5 User s Special Request for Guarantee Time Our guarantee constitution for oximeter complies with electronic product after sale service standard regulated by national laws We regulate the guarantee time of hoist board is one year and all the accessories are three months If users request the guarantee time beyond our regulated guarantee time we should take it into consideration Because electronic product has such character of quick changing for such user asking more than three years guarantee time our company will not sell oximeter parts during maintenance Our company will upgrade oximeter or change new maintenance methods for this we charge the lowest price for new oximeter with user permission 4 3 6 Repacka
3. 0049 40 255726 E mail antonsissi hotmail com 012 5
4. inspect it often Worn out data cables may also cause inaccurate data so if the data is used as a reference to treat a patient pay special attention to data cable and check it more frequently Do not tangle the SpO2 cable with the wires of ES Electrosurgery equipment Single use accessories should never be reused Precautions Operation of the handheld pulse oximeter may be affected by the use of an electrosurgical unit ESU The handheld pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement Verify that nothing is hindering the pulse measurement before replying on the SpO2 measurement Do not sterilize the device using autoclaving ethylene oxide sterilizing or immersing the sensors in liquid which may cause inaccurate readings The device is not intended for sterilization Unplug the sensor from the monitor before cleaning or disinfecting it If liquid is accidentally spilled on the unit clean and dry it thoroughly before reuse Do not try to apply the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence its accuracy SpO2 measurements may be influenced by high ambient light especially sunlight Shield the sensor area if necessary Dyes introduced into the bloodstream such as methylene blue indocyanine green indigo carmine and fluorescein may inf
5. oximeter each time the function key left key is pressed the oximeter will switch to another display mode as shown in Fig 3 1 2 5 Factors that may affect the measurement During operation the accuracy of oximetry readings can be affected by the following factors 2 5 1 Instrument performance depends on the pulsatile character of the artery The measurement would not be considered reliable and accurate if the following conditions are present during measurement e The patient is in cardiac arrest or shock Venous pulsations Low temperature of hand Have taken vascular activity medicine Anemia carboxyhemoglobin methemoglobin methylene blue Indocyanine green 2 5 2 Instrument performance depends on the wavelength absorption for oxyhemoglobin and deoxyhemoglobin If there are substances absorbing the same wavelength this would induce false or low SpO2 values The following may affect these values e carboxyhemoglobin e methemoglobin e methylene blue e Indocyanine green 2 5 3 Extremely high illumination could affect the SpO2 measurement Use a semi translucent or opaque cover to shield the sensor 7 PULSE OXIMETER 2 5 4 Other factors a High frequency electrosurgical interference from external devices including defibrillators b Placement of a sensor on an extremity that currently has been installed a blood pressure cuff arterial catheter or intravascular line c The patient has hypotension severe vasoconstrict
6. upper limit 100bpm The lower limit 60bpm Other Alarms Probe off Finger out Low battery Modes Visual and audible Display Type OLED double color Parameters SpOz2 PR Pleth waveform Pleth bar Mode 3 display modes Record Patient ID 10 patients Record time Up to 72 hours cycle store Data transmission Transmission method Cable Transmission 17 PULSE OXIMETER Environmental Operating Temperature 5 C 40 C Storage Temperature 20 C 55 C Operating Humidity lt 80 RH no condensation Storage Humidity lt 93 RH no condensation Classification per IEC60601 1 Type of protection Internally powered equipment Degree of protection Type BF applied Par Mode of Operation Continuous Safety IEC Standard 60601 1 Dimensions 130 x 44 x 20 mm LengthxWidthxHeight not contain the probe Weight 120g with alkaline batteries Power supply Type 2 AAA Alkaline Batteries Operation time About 16 hours of typical operation 18 INSTRUCTION MANUAL APPENDIX B The declare of manufacturer Guidance and manufacture s declaration electromagnetic immunity for Equipment and Systems that are not Life Supporting Guidance and manufacture s declaration electromagnetic immunity The Handheld Pulse Oximeter is intended for use in an electromagnetic environment specified below The customer or the user of the Handheld Pulse Oximeter should assure that it is used in such an environment Immunity IEC 60601 Com
7. Copyright Our company owns all rights of this unpublished work and intends to maintain it as a confidential work We may also seek to maintain this work as an unpublished copyright This publication is to be used solely for the purpose of reference or operation of our software system No part of this work can be disseminated for other purposes In the event of inadvertent or deliberate publication we intend to enforce its right to this work under copyright laws as a published work Those having access to this work may not copy use or disclose the information in this work unless expressly authorized by our company All information contained in this publication is believed to be correct We shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing performance or use of this material The information this publication refers to is protected by copyrights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rights of third parties Content of the manual is subject to change without prior notice Version V1 0I P Issue Date August 12 2010 CONTENT CHAPTER 1 Introduction cccccssccssccscccsccesccesscesscesscesscesccesseescessceees 14 1 1 Brief Introduction eereeeiiteeettteeeeeeettttteeeeeetttteesentttrrseseeenttrressssenn
8. O2 and or PR value is beyond the default limit alarm will be activated and the corresponding value will flash with audible sound By default SpO2 The upper limit 100 The lower limit 90 PR The upper limit 100bpm The lower limit 60bpm Notice During the alarm is issued press the function button you can silence the alarm for 30 seconds When press it again the screen switches between pulse column display mode filled waveform mode and line waveform mode Alarm priority There are three level priorities for selection High priority indicates the patient is in the very dangerous situation Medium priority indicates the warnings should be paid attention to Low priority indicates the technical alarm caused by the device itself Alarms of the oximeter include technical and physiological alarms All the three priorities are divided by built in module and can not be changed by user Assignment of priority The SpO2 is beyond The SpO2 value Di Di the limits is flashing Di Di Di Di Di once circularly every 3 seconds The pulse rate is The pulse rate Di Di Di once Medium beyond the limits value is flashing circularly every 5 seconds er The probe or finger Finger out Di once circularly is not inserted Probe off every 20 seconds 12 INSTRUCTION MANUAL CHAPTER 4 MAINTENANCE AND REPAIR Warning The advanced circuit inside the oximeter does not require periodic calibration and main
9. aintenance sssssssssssssssstrtrtrssssssssssoeonttttrrrrsssssssssotentttttrrrrssssnnnnnettnntttrrrrrsssnnn Tee eee 13 AP Tro blesho t nNg srerssranidinenirenenn erene a anaa 13 4 3 Warranty and Repair esssssessssssiiieessssssssserrrieessssssssetertrrersssnseettrrrreessssssetttrrrreresssssettet 14 APPENDIX A Specifications eeeeeseeeeesseseeeererssrereessssrereeersesseeeeeee 17 APPENDIX B The declare of manufacturer ccccccececeseees 19 INSTRUCTION MANUAL CHAPTER 1 INTRODUCTION 1 1 Brief Introduction Thank you for purchasing the handheld pulse oximeter for the functional oxygen saturation of arterial hemoglobin SpOz2 and Pulse Rate PR measurement The handheld pulse oximeter features PR tone modulation data storage and data transmission capabilities Please read the operator s manual carefully before using this instrument Notice The illustrations used in this manual may differ slightly from the appearance of the actual product Intended use The pulse oximeter is a portable non invasive device that intended for spot checking displaying storing and transmitting the oxygen saturation of arterial hemoglobin SpO2 and pulse rate of single adult and pediatric patient in hospital including clinical use in surgeries anesthesia intensive care and etc Not for continuous monitoring Measurement principle The principle of pulse oximetry is based on the red and infrared IR ligh
10. d Pulse Oximeter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment transmitters and the Handheld Pulse Oximeter as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output power of 150KHz to 80 MHz 80MHz to 800 MHz 800MHz to 2 5 GHz transmitter W d 35 Iye Az 35 P de lyp For transmitters rated at a maximum output power not listed above the recommended separation distance in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE1 At 80MHz and 800MHz the separation distance for the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic interference is affected by absorption and reflection from structures objects and people 21 The Manufacturer Information Beijing Choice Electronic Technology Co Ltd Add Bailangyuan Building B Rm 1127 1128 Fuxing Rd A36 Beijing China Tel 0086 10 88204188 Fax 0086 10 88203199 E mail chenlei choicemmed com The ER Representation Information Shanghai International Trading Corp GmbH Hamburg Add Eiffestrasse 80 20537 Hamburg Germany Tel 0049 40 2513175 Fax
11. d d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range P Inter ference may occur in the vicinity of equipment marked with the following symbol oo NOTE1 At 80MHz and 800MHz the higher frequency range applies NOTE2 These guideline may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base situation for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If abnormal performance is observed the additional measures may be necessary such as reorienting or relocating the handheld Pulse Oximeter b Over the frequency range 150kHz to 80MHZ field strengths should be less than 3V m 20 INSTRUCTION MANUAL Recommended separation distances between portable and mobile RF communications equipment and the Handheld Pulse Oximeter The Handheld Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled The customer or the user of the Handhel
12. der Measure mode and Information Display mode Pressing the Function key left key or Setting key right key the display mode may be changed sequentially between Filled Waveform Line Waveform and Pulse Column Waveform Once you choose a certain mode the oximeter will continually display in this mode until changed by the user 3 Setting Mode Enter the Setting Mode under the Measure mode you can only set Brightness and Patient ID Enter the Setting Mode under the Information Display mode you can set Brightness Patient ID Date and Time Current parameters and data will be displayed at the top right corner which is used to display the PR The various parameters display title is as follows 10 INSTRUCTION MANUAL Br Brightness range 1 7 D Day range 1 31 ID Patient ID range 1 10 H Hour range 0 23 Y Year range 2000 2099 M Minute range 0 59 M Month range 1 12 S Second range 0 59 a Enter Setting mode Press the function key left key for more than one second extended press the oximeter will enter into Setting mode and you will find a parameter item which includes its title and value on the top right corner of the display refer to Fig 2 5 After entering the setting mode from Information Display mode continually press Function key left key the current parameter item will be changed sequentially in the followi
13. ensor to sensor interface at the top of the oximeter Ensure that the sensor is firmly plugged in The interface is also used for transferring data to the software for management To transfer data please connect the data cable to the interface For detailed information refer to MedView software instruction manual Notes 1 Please make the raised lump side of sensor cable upside when inserting the sensor 2 Do not insert the sensor forcefully 3 Because the probe socket and USB socket share a common interface so you cannot take a measurement when transmitting data 2 4 SpO2 Monitoring Clip the sensor to the patient s finger and ensure that the patient s nail surface is facing upward as shown in Fig 2 3 Fig 2 3 Placement of the finger 2 4 1 Power on the oximeter Press the function key left key to power it on Several seconds later the measurement value will appear Notice To maintain the highest degree of accuracy it is recommended that the finger and the oximeter sensor probe is kept as still as possible 6 INSTRUCTION MANUAL 2 4 2 Brightness adjustment When you press the function key left key for more than one second the brightness level headed with Br will be shown on the top right of the screen You can adjust the brightness by degrees by pressing the setting key right key There are 7 levels of brightness The default is level three Fig 2 4 2 4 3 Mode switch After turning on the
14. ey the modification under Setting mode will be cancelled and simultaneously the system will exit from Setting mode If there is no operation under Setting mode for more than 30 seconds the system will exit from Setting mode automatically 3 4 Data replay and transmission The oximeter can record SpO2 and PR value for more than 72 hours and can analyze records one by one You can transfer the history data to a PC by using MEDVIEW software and an attached data cable As for detailed setup and operation please refer to the MEDVIEW operator s manual Notice The mode of its storage is cycle store it only can save the latest 72 hour records 11 PULSE OXIMETER 3 5 Alarm Alarm Technical alarm and physiological alarm Technical alarm finger out probe off power low and error code In the situation that the finger is not inserted correctly or the connection state of the probe is not good results in failure of measurement Finger out or Sensor off may be displayed on the normal screen When battery power is lower than 2 3 0 1V the sign x will flicker in its display area Replace the batteries as soon as possible In the failure state the oximeter will display error codes and will automatically power off if the error code display lasts for more than 30 seconds For the details and definitions on error please refer to chapter 4 2 Physiological alarm SpO2 and PR If the measured Sp
15. ge e Take all the accessories and put them into plastic cover e Try to use original package and packing material User will be responsible for such damage caused by bad package during transportation e Please offer guarantee list and copy of invoice to standby with the period of guarantee e Please describe failure phenomenon in detail and altogether offer oximeter Storage and Transportation Storage Storage Temperature 20 C 55 C Relative Humidity lt 93 no condensation Transportation Transport by airline train or vessel after packing according to request Package Pack the product with the hard bag and put the foam between the inner box and the carton to alleviate the shake 16 INSTRUCTION MANUAL APPENDIX A Specifications Notices e Specifications may be changed without prior notice e The circuit diagrams the list of components the illustration of diagrams and the detailed rules of calibration are provided exclusively to professional personnel authorized by our company SpO2 Display Range 0 100 Measurement Range 70 100 Resolution 1 Accuracy 80 100 2 70 79 3 lt 70 unspecified Data update time lt 15s Probe LED Specifications fF Wavelength Radiant Power 660 2nm CR oom Pulse Heart Rate Display Range 0 254 bpm Measurement Range 30 235 bpm Resolution 1bpm Accuracy 2bpm or 2 Alarm By default SpO2 The upper limit 100 The lower limit 90 PR The
16. iode damaged Err 9 real time clock damaged Err 10 EEPROM chip damaged 13 PULSE OXIMETER b Possible problem and corresponding resolution SpO2 or PR cannot 1 Finger is not plugged 1 Retry by plugging the finger be displayed correctly 2 Attempt several times to obtain normally Patient s Oxyhemoglobin a reading If you are sure that value is too low to be no problem exists obtain further measured clinical examination SpO2 or PR Finger might not be plugged Retry by plugging the finger display is unstable deep enough Urge the patient to remain still Finger is trembling or patient is moving continually The Oximeter can Battery power may be 1 Please replace batteries not be powered on inadequate or not installed 2 Please reinstall the batteries Batteries might be installed 3 Contact local customer Technical incorrectly Service The Oximeter might be damaged Probe off The sensor is not connected Connect the sensor displayed on The connection between the Please check if the probe was screen Probe and Oximeter is loose connected with oximeter correctly 4 3 Warranty and Repair 4 3 1 Service Method Service support Our company will offer user telephone and e mail support as well as in house repair at our facility Parts replacement Our company will replace parts accessories free of charge during the warranty period Our company will update the system software free of charge
17. ion severe anemia or hypothermia d An arterial occlusion proximal to the sensor Z Warnings e Use only SpO2 sensors provided by manufacturer Other SpO2 sensors may cause improper performance e Do not use an SpO2 sensor with exposed optical components e Excessive patient movement may cause inaccurate measurements e The patient has hypotension severe vasoconstriction severe anemia or hypothermia e Tissue damage can be caused by incorrect operation or misusing sensor for example by wrapping the sensor too tight Inspect the sensor site to ensure the skin s integrity and the adhesion position of the sensor is correct More frequent inspection should be taken if necessary e Loss of pulse signal can occur in any of the following situations a The sensor is too tight b There is excessive illumination from light sources such as a surgical lamp a bilirubin lamp or sunlight c Ablood pressure cuff is inflated on the same extremity as the one to which an SpO2 sensor is attached INSTRUCTION MANUAL CHAPTER 3 DETAILED OPERATION 3 1 Display The handheld pulse oximeter uses an OLED display for a readout It can display the SpO2 and pulse rate PR value as well as a pulse column and SpO2 waveform There are three display modes shown in Fig 3 1 The first figure is pulse column display mode The second figure is filled waveform mode The third figure is line waveform mode indicating SpO2 trend 1 Pulse column disp
18. ios electrical appliances it is possible that high levels of such interference due to close proximity or strength of a source may result in disruption of performance of this device This apparatus complies with the IEC 60601 1 2 international standard The requirements of this international standard are CISPR11 GROP1 and CLASS B 1 4 Equipment Classification Classification according to IEC 60601 1 According to the type of protection Internal electrical power source equipment against Electrical shock According to the degree of protection Type BF equipment against Electrical shock According to the degree of protection IPX1 against harmful ingress of water According to the methods of Non sterilizable Use of Liquid surface sterilization or disinfection disinfectants only According to the mode of operation Continuous operation Spot checking The equipment is not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide PULSE OXIMETER 1 5 Accessories Standard accessories e Operator s manual e Fingertip sensor for adult Model M 50E13CSO e 2x1 5V AAA Size Alkaline batteries nanfu e MedView software CD e Data cable Optional accessories e Binding sensor for pediatric 15 45Kg Model M 50C Confirm that the items listed are packed with the handheld pulse oximeter If any item on this list is missing or damaged contact your distributor Contact the carrier immediatel
19. lay mode 2 Filled waveform mode 3 Line waveform mode SpO2 PRobpm Fig 3 1 Three display modes e Sp0O2 Percent oxygen saturation value displayed above is 98 e PR Pulse rate value displayed is 72 bpm e Pulse column This is used for signal identification and quality indication during motion and low signal to noise situations The bar rises and falls with the pulse its height indicating signal quality When the bar is very low the SpO2 and pulse rate values may be suspect e Signal strength Indicates arterial pulse signal strength and may be used as a diagnostic tool during low perfusion for the accurate prediction of illness severity The bar is highest when the quality of the perfusion state is best and low when the perfusion is poor e PR tone modulation Beeps in sync with the patient s pulse even under most challenging patient motion conditions except when in the 3 mode PULSE OXIMETER 3 2 Key Functions 1 Power On Press the Function key left key to power on The oximeter will power off automatically under the information display mode or trouble mode for 30 seconds 2 Definition of key There are two keys in the oximeter Function key left key and Setting key right key Function Key Located at the left of the oximeter this acts as a Power On switch when the unit is in the Off condition When the unit is On it acts as a Function key Setting key Located at the
20. luence the accuracy of the SpO2 readings Any condition that restricts blood flow such as use of a blood pressure cuff or extremes in systemic vascular resistance may cause a failure to determine accurate pulse rate and SpO2 readings ay INSTRUCTION MANUAL e Remove fingernail polish or artificial fingernails before applying SpO2 sensors Fingernail polish or artificial fingernails may cause inaccurate SpO2 readings e Optical cross talk can occur when two or more sensors are located in adjoining areas It can be eliminated by covering each site with opaque material Optical cross talk may adversely affect the accuracy of the SpO2 readings e Obstructions or dirt on the sensor s red light or detector may cause a sensor failure Make sure there are no obstructions and the sensor is clean e For routine equipment maintenance please refer to the service procedures at the associated section as indicated in the manual e As to the other concerns for attention please carefully look through the specific chapter in this instruction 1 3 Electromagnetic Compatibility This oximeter is designed and tested in compliance with the EMC standard complying with the international standard for the EMC of the medical electrical device IEC 60601 1 2 However because of the proliferation of radio frequency transmitting equipment and other sources of electrical noise in the health care and home environments e g cellular phones mobile two way rad
21. ng order Br brightness gt ID patient s ID gt Y year gt M month gt D date gt H hour gt M minute gt S second gt Br brightness gt Notice You cannot set the date and time when in the measuring course b Save and exit from Setting mode Under the Setting mode press function key left key to select the desired parameter item Then press Setting key right key to adjust the value Each time you press the Setting key right key current parameter value will be added by 1 unit sequentially Double press the Setting key right key and the current parameter setting will be added by 9 units sequentially Press function key left key to select the next desired parameter item Then redo step to adjust the value You can continually redo step M to until no parameter s setting need to be changed To finish setup fleetly press both left and right keys together to confirm Then the modifications under the Setting mode will be saved at the same time the system will exit from Setting mode Notice e To confirm the modification the act of pressing both keys simultaneously must be fleet Do not press them for long time e The date and time adjustment only can be done in the Setting mode under Information Display mode c Cancel and exit from Setting mode Under the Setting mode double press the Function key left k
22. nnrsssnneeene 1 1 2 Safety Informati n eseserrerrirsrirsmeeniiceisissssririsiesrsrsisiresiserrtide dititi sis erpen seii 1 1 3 Electromagnetic Compatibility EE TS OE A eis AE A A A ROE 3 1 4 Equipment Classification eee sseeeesesssssresrrsssisssssssssssssnnsssrrerertessnssssnssnnnnssrrrrreeeeeesns 3 LD ACCESSOTIES siemens R ie 3 1 6 Equipment symbol toisisetceentaprsarascecomenaasnicenannniiamuedomaaen ames 4 CHAPTER 2 BASIC OPERATION ssccsscccssscessccessseesscesseesseeeseees 5 21 Ouer VIEW ee e E E E E E ree 5 2 2 Install the AMT Sa casas eee cee hace 5 2 3 Connect th SCIS OM cesses ones ecstsccti enc scosd cticeces i arcnsensacpatentsluaraavonsancadestabearocencpuchtesnate 6 2 4 SpO2 Monitoring eeeeeeeeeeeeeeereeerrrrrrrrrerrrreeereeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeereeeereeserereeeeeeeeeeee 6 2 5 Factors that may affect the measurement e s eessssssrsssssssssssssssssesesrsrserssssese 7 CHAPTER 3 DETAILED OPERATION cccceeeeeeesessreereeeeeeseeeresesssssssssssss 9 31 Display ercsretecimecsnninnseuroninnsian unnan A n 9 3 2 Key FUNCIONS eoii naseiani iiare a naea aaa ee aara Ee aaaea aieiai 10 3 3 Mode Introduction wssissesssisieordencosenssnsunessedecinnssuanserneieabesssnssendveswnsannsesannnensbeladecratersnnsansoonein 10 3 4 Data replay and transmission sessttttterrereerrrerrrerrrrreerrreerreteererreereeereeeeerreeeereeese 11 UN e E A E E EE 12 CHAPTER 4 MAINTENANCE AND REPAIR esseere 13 4 1 M
23. pliance Electromagnetic Test Test level Level environment guidance Electrostatic 6kV 6kV Floors should be wood concrete or Discharge contact contact ceramic tile If floor are converted ESD 8kV air 8kV air with Synthetic material the relative IEC61000 4 2 humidity should be at least 30 Electrical fast transient Not burst IEC applicable 61000 4 4 Surge IEC Not 61000 4 5 applicable Voltage Not dips short applicable interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power frequency magnetic field IEC 61000 4 8 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment 19 PULSE OXIMETER Guidance and manufacture s declaration Electromagnetic Immunity for Equipment and Systems that are not Life Supporting Immunity IEC 60601 Compliance Electromagnetic Test Test level Level environment guidance Conducted Not Portable and mobile RF communications applicable equipment should be used no closer to any part of the handheld pulse oximeter including cables than the recommended separation distance calculated from the equation Radiated 3V m applicable to the frequency of the transmitter RF IEC eee d KeWP 80MHz to 800MHz 61000 4 3 j slyp 800MHz to 2 5GHz Where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacture an
24. right of the oximeter it has no function when the power is off When the unit is On it acts as a Setting key 3 Definition of Key Press There are three ways to press the key Press press the key quickly the duration time should no more than 1 second Double press Two time continuous press the time between the two press actions should be no more than 0 5 second Extended press Press the key for more than 1 5 seconds 3 3 Mode Introduction Power On Measurement and operation can be done normally There are three display modes Measure mode Information display mode and Trouble display mode 1 Display modes Measure mode the sensor is plugged in correctly and the finger is properly in the sensor the oximeter is in the measurement mode for both SpO2 and PR Information display mode The sensor is not plugged into the oximeter or the sensor is plugged in the oximeter but the finger is not in the sensor Probe off or Finger out will be displayed respectively and it will automatically power off if either information display lasts for more than 30 seconds Trouble display mode In the failure state the oximeter will display error information and will automatically power off if the information display lasts for more than 30 seconds For error information details and definitions please refer to chapter 4 2 2 Display Mode setting The Display Mode setting can only be enabled un
25. rms the user of other important information by suggestion requirement and supplement 1 PULSE OXIMETER Warnings Please read this manual carefully before using this device The user must check that the equipment functions safely and ensure that it is in proper working condition before being used Do not use the handheld pulse oximeter in an explosive atmosphere Do not use the handheld pulse oximeter in an MRI or CT environment The handheld pulse oximeter has no SpO2 alarms and PR alarms it is not for continuous monitoring as indicated by the symbol The handheld pulse oximeter is intended only as an adjunct in patient assessment It must be used in conjunction with other methods of assessing clinical signs and symptoms Check the handheld pulse oximeter sensor application site frequently to determine the positioning of the sensor and circulation and skin sensitivity of the patient Do not stretch the adhesive tape while applying the pulse oximeter sensor This may cause inaccurate readings or skin blisters Prolonged use of the probe sensor or the patient s condition may require changing the sensor site periodically Change the sensor site and check skin integrity circulatory status and correct alignment at least every 4 hours Prolonged use may cause blisters skin deterioration and discomfort Sensor malfunction may cause inaccurate data possibly resulting in patient injury or death so pay close attention to the sensor and
26. t absorption of oxygenated and deoxygenated hemoglobin present in the circulating blood Oxygenated hemoglobin absorbs more IR and allows more red light to pass through Deoxygenated hemoglobin conversely absorbs more red light and allows IR light to pass through The detector probe is placed on the finger The probe contains two light emitting diodes LED s one in the visible red spectrum 660nm and one in the IR spectrum 940 nm The beams of light from this probe pass through the tissues and some light is absorbed by the blood and soft tissues depending on hemoglobin concentration The amount of light absorption at each light frequency is dependent on the degree of oxygenation of hemoglobin within the tissues The microprocessor can select out the absorbance of the pulsatile fraction of blood i e that due to arterial blood from constant absorbance due to non pulsatile venous or capillary blood and other tissue pigments Principle of Operation 1 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Warning Precaution and Notice Warning Precaution and Notice in the manual are special information that prompts the operator s attention Warning Information concerning something that could possibly hurt the patient or operator Precaution Reminds the user to pay close attention to device operation failure of which may cause abnormal function of the instrument Notice Info
27. tenance except replacing the batteries Don t open the cover of oximeter or repair electronic circuits Its open will cause the damage of the device and the annulment of the guarantee 4 1 Maintenance It is very important for user to perform daily maintenance of oximeter and parts in order to maintain its function and appearance Disinfection procedures may be performed with the use of the below mentioned cleaner disinfectants Failure to perform these procedures may result in invalidating the warranty Local disinfection protocols will apply The external surface of the oximeter can be cleaned by wiping with a damp cloth Do not submerge the oximeter in any solution at any time To do so will void the warranty Use the following permitted solutions e Ammonia diluted e Glutaraldehyde e 10 Bleach solution e Mild soapy water diluted e Do not use the following cleaners o Any kind of scrubbing or scouring solution o Acetone o Alcohol based cleaners Battery maintenance Please take out battery before cleaning the oximeter Remove the batteries if you are not going to use the oximeter for a long time 4 2 Troubleshooting a Error Definitions Err 1 program memory damaged Err 2 data memory damaged Err 3 sensor Red Emission Diode damaged Err 4 sensor Infrared ray Emission Diode damaged Err 5 sensor Infrared ray Receipt Diode damaged Err 6 exterior crystal oscillator damaged Err 7 sensor emission diode or receipt d
28. y if the shipping carton is damaged 1 6 Equipment symbol Attention Refer to the relevant prompt Read the operator s manual carefully before using the oximeter Type BF applied part Date of Manufacture Low power indicator Protected against dripping water European union approval Serial number Not for continuous monitoring p Storage temperature and relative humidity INSTRUCTION MANUAL CHAPTER 2 BASIC OPERATION 2 1 Outer View Fig 2 1 Front and rear panel Description of Fig 2 1 1 The interface for connecting the sensor and data transmission 2 Displaying screen 3 Power function key 4 Setting key 5 Clip for fixing the device 6 The cover of battery compartment 2 2 Install the batteries The oximeter can be powered by 2 AAA Size alkaline batteries which will typically provide 16 hours of continuous operation by represent of nanfu for example When battery power is lower than 2 3 0 1V the sign a will flicker in its display area Replace the battery as soon as possible If the battery power is lower than 2 2 0 1V the device will power off automatically Pull the cover as the indication by the arrowhead And install the batteries as the polarity markings and Refer to fig 2 2 Be sure not to insert the batteries in wrong polarities otherwise the damage may be caused to the device 5 PULSE OXIMETER Fig 2 2 2 3 Connect the sensor Connect the oximeter s