        Operation and Calibration of the UV-Vis - TIAER
         Contents
1.  5 2 2 Start the UVProbe 2 33 software from the computer icon   5 2 3 Install the correct cell holder for the application  refer to  analytical SOP for details   Leave the cell slot blank    5 2 4 Ensure that the    Photometric    view is open  then click    Connect     to establish the link between computer and instrument  This  may take several minutes    5 2 5 If connection window shows errors  troubleshoot according to  the User s Manual  Problems may include power not being on   blocked light path or light source not working    5 2 6 Once all test have passed and a connection is established  click     OK       5 2 7 Click the    A Go to WL    and enter the wavelength desired    5 2 8 Fill the required cell with blank reagent or reacted blank  Use a  Kimwipe to gently wipe down the cell light path and hold up to  look through at light  Ensure no droplets or smudges are    Rev  8 Page 6 of 8 TIAER    QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    present  Wipe until clean  DI water or ethanol may be needed  to clean the cell ends    5 2 9 Place the cell in the holder and click    Auto Zero     Repeat for  each wavelength used per analytical SOP    5 2 10  The program takes continuous readings for subsequent  measurements of samples and standards in the cell  Allow the  reading to stablize to 3 decimal places and take the best  reading for the 4     5 2 11  Power off instrument and computer when complete and clean  up cell and work area    5 2 12  O
2.  Prior to performing any analytical  procedure using the uv vis spectrophotometer  the standard  operating procedure for chemical analysis are reviewed to  determine all of the required parameters and standards     5  Procedure  5 1  Beckman Coulter DU 640  primarily for determination of OPO4P and  as a backup to the other instruments   5 1 1 Powering Up  5 1 1 1  The instrument may be turned off overnight   5 1 1 2  Turn on the instrument with the power switch located on  the right hand side of the back of the instrument  The  display should illuminate  and a message window with     Executing Power Up Diagnostics    should be displayed   5 1 1 3  In a few minutes a table indicating the Power Up  Diagnostics will be displayed  The table should indicate  that the parameters    RAM Option    and    Programmability     are    Not Installed    and that    RS232 Ports 3 and 4    is     Installed     The word    Passed    should be indicated for all  other parameters     Rev  8 Page 3 of 8 TIAER    QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    5 1 1 4  If any parameter fails  flush the flow cell with cleaning  solution and DI water and turn the instrument off  Repeat  power up steps  Consult manual for troubleshooting  if  necessary  Continue when the instrument successfully  completes power up diagnostics    5 1 1 5  Move the cursor arrow to the upper right hand corner of  the Power Up Diagnostics table to the word quit and click  on it to exit    5 1 2 C
3.  run    5 1 4 10  When the run is complete open the    1  Get Data    icon on  the auxillary computer  This will bring up a box  click the     activate    button  Open the    SAVECLEAR    icon on the  instrument screen  Change the file name to    DATA    and    Rev  8 Page 5 of 8 TIAER    QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    click    O K     Choose    O K     and overwrite both files  Exit  the box on the auxillary computer  All further steps will be  with the auxillary computer only   5 1 4 11  Open the    2  Change Data    icon  Open the file    data duf     or    data    file  Change the data to Excel format and save  as    data     The old file will be written over   5 1 4 12  Open the    3  Data    icon and type in standard designations  and sample  s in Sample ID column  Print the file and  affix to the Personal Logbook   5 1 4 13  If desired  automatically transfer the data file to ESDMS  or data management system   5 1 4 14  Data may also be entered manually into ESDMS system   5 1 4 15  Flush ethanol through the system  followed by DI  Other  cleaning solutions may be used on occasion  Refer to  specific SOP for instructions   5 1 4 16  If finished  turn off the light source and clean area   5 2  Shimadzu UV 2450  Manual only  no autosampler used at this time   primarily used for chlorophyll and  pheophytin analyses  low level  orthophosphate P and NO3N and as a backup to the DU 640   5 2 1 Power up the instrument and the computer  
4. QAM I 103    Operation and Calibration of the  UV VIS Spectrophotometer          Revision 8  Approval  Ls A   ae j      jji  5 bh AX A b l4 lis  Laboratory Manager  Date  Laboratory Quality Assurance Officer    ee p 7 E  Concurrence i Date  Effective date     4     Renewal date  Initials     Texas Institute for Applied Environmental Research    QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    1  Applicability  This procedure applies to the operation and calibration of the Beckman  Coulter DU 640 UV VIS and Shimadzu UV 2450  manual mode   spectrophotometers located in the chemistry laboratory of the Texas  Institute for Applied Environmental Research  TIAER   Tarleton State  University  Stephenville  TX  This procedure shall be performed prior to  any analysis using one of the spectrophotometers    2  Purpose  The purpose of this procedure is to provide a method for operation and  calibration of the DU 640 and UV 2450 spectrophotometers  By  utilizing appropriate chemisty SOPs to perform the standard calibration  for each set of samples  the analyst eradicates anomalies due to  reagent mixture variations  sample preparation variations  and  instrument sensitivity fluctuations  The operation of the instrument  allows the analyst to determine the concentration of various chemical  species which are dissolved in water samples or extracts analyzed by  the TIAER chemistry laboratory     3  Definitions   3 1  Automated Sipping Device  A device used in conjunct
5. alibration   5 1 2 1  Turn on the appropriate source by clicking on the  VIS  OFF  or  UV OFF  on the menu bar  The icon selected  should turn red and change to  VIS ON  or  UV ON  to  indicate that the source is on    5 1 2 2  Allow the instrument and source to warm up and stabilize  for approximately 30 minutes    5 1 2 3  Select the    Fixed Wavelength    mode    5 1 2 4  Prepare a calibration blank with DI water or extracting  solution and all reagents that will be used in the samples  according to the proper procedure    5 1 2 5  The instrument should be in absorbance mode  Mode  can be changed by clicking the icon next to the word     mode     Clicking will toggle the instrument between  absorbance    Abs    and percent transmittance     T       5 1 2 6   Click on the field to the right of the    A    and input the  proper wavelength for the analyte being tested    5 1 2 7  Draw the blank into the flow cell by inserting the sipper  tube into the culture tube and pressing the fill button on  the pump    5 1 2 8  Click on the    blank    icon on the menu bar  When the     Reading Blank    box disappears the instrument is  calibrated and ready for use    5 1 3 Pre operation Checks   5 1 3 1  Prior to performing any analytical operations  the analyst  verifies proper connection between the instrument and all  output devices     Rev  8 Page 4 of 8 TIAER    QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    5 1 3 2  If the system is connected to an automa
6. ion with the  instrument serving as a sample container to photometric cell sample  delivery conduit     4  Equipment  Reagents  and Standards  4 1  Equipment    4 1 1 Beckman Coulter DU 640 spectrophotometer   4 1 2 Shimadzu UV 2450 spectrophotometer   4 1 3 10 mm square sided silica or quartz cuvette  for uv region  or 10   mm square sided glass or plastic cuvette  for the visible  wavelength region   may be flow through or manual pour cells   4 1 4 100 mm tubular cuvette  or other sizes   4 1 5 Cuvette holder  adapted for size of cuvette    4 1 6 Automated sipping device or autosampler  if applicable     Rev  8 Page 2 of 8 TIAER    QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    4 1 7 Sample tubes or other glassware appropriately sized to serve as  mixture vessels for water samples or extracts and reagents  4 1 8 Autosampler boats  4 2  Reagents  4 2 1 Deionized Water  DI   4 2 2 Color reagent solutions  contain all color or complexing reagents   sample solvents  and other chemical species which will be added  to the samples prior to analysis by the uv vis instrument  Refer  to the appropriate chemical SOP for specific solutions and uses   Non color reagents are also used in organic extractions and  sample preparation   4 2 3 Cleaning solution for automated sipper and cell  add 5 drops  Triton X 100    and 5g NaOH to 500 mL   DI H20 and dissolve   4 3  Standards  4 3 1 The standards required for the analysis of samples are  determined by the analyte 
7. ther developmental methods such as multiwavelength scans  and timed reading may be performed for informational studies  and research  Consult the User   s Manual for these operations   Included herein is only the simple  manual operation of the  instrument     6  Quality Control and Safety Aspects  6 1  All aspects of this procedure shall comply with QAM  Q 101      Laboratory Quality Control   and QAM S 101   Laboratory Safety    6 2  Quality Control   6 2 1  The analyst always uses matched cuvettes  if using more that  one cuvette  when performing analysis with the uv vis  spectrophotometer  This ensures the pathlength parameters  will be equivalent and the measurements are accurate and  reproducible    6 2 2  The preparation of all samples  standards  and blanks are  carried out using identical methods and preparatory equipment   This eliminates the majority of errors which may be introduced  during sample solution preparation    6 2 3  The auto sampler  cuvettes  and all associated equipment are  cleaned and rinsed thoroughly with DI water prior to and  following use  Other cleaning solutions may be required on  occasion at the direction of the Laboratory Manager     Rev  8 Page 7 of 8 TIAER    6 3     QAM I 103  Operation and Calibration of the UV VIS Spectrophotometer    6 2 4 All instrument maintenance performed on the  spectrophotometer are documented in the Maintenance  Logbook  QAM Q 103 1    6 2 5 Any major modification or parts replacements initiates a new  Demons
8. tic sipping device   check the connections between the device and the  instrument also    5 1 3 3  Once all connections have been checked  sample  analysis is ready to be performed    5 1 4  Operation   5 1 4 1  Power up instrument as described above    5 1 4 2  Calibrate the instrument as described above    5 1 4 3  Click on the field to the right of the    Factor    and input the  proper factor    5 1 4 4  Using the proper vials as reaction containers  prepare the  samples and standards as indicated in the standard  operating procedure for the chemical analysis    5 1 4 5  Normally the analytical order is set by placing a blank  solution in the first slot  then following with the standards   samples and quality control aliquots  which include  duplicates and spikes    5 1 4 6  Use the complete set of standards listed in the standard  operating procedure for chemical analysis immediately  after the blank    5 1 4 7  An initial calibration verification and initial blank  verification should be performed at the beginning of an  analytical run and a continuing calibration verification and  continuing calibration blank check are placed at the end  of each QC batch  Other verification such as laboratory  control standards or limit of quantification standards may  be used if required by project    5 1 4 8  Load the sample boat onto the auto sampling device    5 1 4 9  Click on the    Read Samples    icon to begin analysis and  allow the instrument and auto sampler to complete the 
9. tration of Performance in accordance with QAM Q 101      Laboratory Quality Control     to ensure the instrument is  functioning properly prior to data generation    6 2 6  The DU 640 retains calibration factors and only needs to be re   calibrated at the direction of the Laboratory Manager   Calibration information is kept with each analytical run record    Safety   6 3 1  The solvents or complexing reagents which are used during  analysis of water samples or extracts may be toxic or acidic   The analyst always consults the MSDS files if he or she has  any question as to the safe handling of any reagent required by  the standard operating procedure for chemical analysis     7  References    7 1     7 2   7 3     7A     DU   SERIES 600 SPECROPHOTOMETER  OPERATING INSTRUCTIONS   MANUAL  Bec kman Instruments  Inc  1992  Fullerton  CA  Manual 514531B   User   s Manual  UV 2450  Shimadzu Corporation  2008  Kyoto  Japan   National Environmental Laboratory Accreditation Conference  NELAC   standards  TNI  2009    TIAER Quality Assurance Manual  QAM Q 100  latest revision     8  Attachments    None    Rev  8 Page 8 of 8 TIAER    
    