1. PSI mmHg inHg inH O oz in kPa Range 600 8 70 450 0 17 72 240 9 139 2 60 00 1100 15 95 825 1 32 48 441 6 255 3 110 00 accuracy T 23 C 1 3 0 02 1 0 0 04 0 5 0 3 0 13 0 lt T lt 40 C 1 5 0 02 1 1 0 04 0 6 0 3 0 15 Else 2 0 0 03 1 5 0 06 0 8 0 5 0 20 Output accuracy accuracy of value shown on LCD display 0 06 FS hPa Response time t90 lt 44 s Relative humidity T3118 only reading is temperature compensated at entire temperature range Measuring range 5 to 95 RH see Device installation Display resolution 0 1 RH Accuracy 2 5 RH for range from 5 to 60 RH 3 RH for range from 60 to 95 RH at 23 C Response time t90 lt 30s air flow approximately 1 m s relative humidity step 65 RH constant temperature There is possible to change response time For more details see file Description of calibration and adjustment procedure pdf which is installed together with Users software 8 TE SNC Tx1 18 02 The value computed from ambient temperature and relative humidity T3118 only You can choice one of the next value Display resolution 0 1 C Dew point temperature Accuracy 1 6 C at ambient temperature T lt 25 C and RH gt 30 Range 60 to 80 C Absolute humidity Accuracy 3g m at ambient temperature T lt 40 C and RH lt 95 Range 0 to 400 g m Specific humidity 7 Accuracy 2 kg at ambient temperature
2. Program enables to set own device output range unit selection and make an adjustment process For T3118 device allows to assign required measured value to one of the two device outputs it means assign temperature relative humidity or computed value to output I1 or I2 and set its range It is also possible to assign both outputs to the same value with the same range if two evaluation devices are necessary to connect Device is connected to circuitry by means of one or two galvanic isolated loops Each loop has two wire connection and each loop requires power from evaluation device It is always necessary to connect current loop I1 for proper device function from loop Il part of the device is powered If current loop I1 is opened or disconnected then loop I2 and LCD will not work Measured and computed values are optionally shown on dual line LCD display If there are two values displayed at one LCD line they are periodically switched between both readings with period of 4 seconds Display can be switched OFF totally too Temperature relative humidity eventually pressure sensors are non removable device parts Please read instruction manual before the first device connection Device setting from the manufacturer If special setting was not required in the order device is set from the manufacturer to the following parameters T0118 output I1 temperature range from 4 to 20 mA corresponds with 10 to 50 C T2118 output I1 atmosphe
3. T lt 35 C and RH lt 95 Range 0 to 550 g kg Mixing ratio Accuracy 2 2 kg at ambient temperature T lt 35 C and RH lt 95 Range 0 to 995 g kg Specific enthalpy Accuracy 3 5kJ kg at ambient temperature T lt 25 C and RH lt 95 Range 0 to 995 kJ kg Recommended calibration interval T0118 2 years T2118 T3118 1 year Measuring interval and LCD display refresh 0 5 s T2118 2 s Communication with computer via USB port by means of USB communication cable SP003 Protection IP20 Operating conditions Operating temperature range 10 to 50 C Operating humidity range 0 to 100 RH without condensation Outer characteristics in accordance with EN33 2000 3 normal environment with the specifications AE AN1 BE1 Electromagnetic compatibility complies EN 61326 1 Storing conditions temperature 30 to 80 C humidity 0 to 100 RH without condensation Not allowed manipulations It is not allowed to operate the device under other than specified conditions in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature and humidity sensors must not be exposed to direct contact to water or other liquids Do not use the device in an explosive environment Dimensions 106 x 88 x 33 hx 1 x d Weight approximately 150 g Material of the case ABS This value depends on the atmospheric pressure For computing there is used constant value stored inside t
4. error during measurement of temperature or relative humidity e Measured pressure value with added correction for altitude is outside of range from 300 hPa to 1350 hPa from 4 351 PSI to 19 580 PSI or the pressure sensor is damaged Please check setting of pressure correction for altitude with User s software Error 3 There is a reading Err3 on LCD display upper line Error of internal A D converter appeared converter does not respond probably damage of A D converter This error does NOT affect pressure measurement Rest values are NOT measured Output value is about 3 8 mA It is a serious error contact distributor of the device Error 4 cc There is a reading Err4 on LCD display It is internal device error during pressure sensor initialization Under this condition device does NOT measure atmospheric pressure Output value is about 3 8 mA Pressure sensor is probably damaged It is a serious error contact distributor of the device End of operation Device itself after its life is necessary to liquidate ecologically Technical support and service Technical support and service is provided by distributor For contact see warranty certificate IE SNC Tx118 02 11
5. Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature relative humidity and other derived humidity values T3118 with current output 4 to 20 mA Instruction Manual Instruction Manual Device type Temperature Humidity Computed value Pressure Output name T0118 v Il T2118 v Il T3118 v v v Il I2 Devices are designed for indoor applications By device type see previous table they can measure ambient temperature at C or F relative humidity of air without aggressive ingredients or atmospheric pressure at hPa PSI inHg mBar oz in mmHg inH2O or kPa Unit choice is provided by User s software see later It is possible to choice one of the following computed value for T3118 device only dew point temperature absolute humidity specific humidity mixing ratio or specific enthalpy Device T2118 supports pressure correction for altitude setting offset All device setting is performed by means of the PC connected via the optional SP003 communication cable not included in delivery and User s software Tsensor Program Tsensor for device setting is available to download free at www cometsystem cz It supports make the adjustment of the device too This procedure is described at file Calibration manual pdf which is installed commonly with the software
6. er It is necessary to connect the power Button is placed behind small hole on the left side of the device front view For button yo pressing use thin instrument for example paper clip Short button press shows low range and type of measured value for first output output I1 at LCD indicated by symbol 1 Type of measured value is indicated with appropriated value unit here C temperature Upper line shows current value which is assigned to value of measured value lower display line Here 4 mA corresponds with temperature of 10 C io Next button press shows values for upper scale limit the same channel the same value by the same rules as at previous case Here 20 mA corresponds with temperature of 50 C C For T3118 device there is analogously shown I2 output setting Next press of button exits info mode and actual measured values are displayed Note For atmospheric pressure device T2118 there is at the lower LCD line shown current value assigned to atmospheric pressure value shown at upper LCD line Note No measurement and communication is possible during info mode If device stays in info mode for longer than 15 s device automatically returns to measuring cycle Readings on LCD display C F 9 Reading next to this symbol is measured temperature or error state of value RH Reading next to this symbol is measured relative humidity or error state of value hPa PSI inHg Reading next
7. he device memory Default value preset by manufacturer is 1013hPa and can be changed by user s software TE SNC Tx1 18 02 9 Procedure of modification of device adjustment e device adjustment is performed by means of the optional SP003 communication cable connected to USB port of the PC e It is necessary to have installed configuration program Tsensor on the PC It is free to download at www cometsystem cz During installation please take care about installation of driver for USB communication cable e Demount the device see chapter Device demounting e Connect SP003 communication cable to the device and PC Installed USB driver detect connected cable and create virtual COM port inside the PC Display must light up or at least must light up all symbols for one second if LCD was switched OFF by program before e Run installed Tsensor program and select corresponding communication COM port as described above e Now make new device setting throw Tsensor program e When new setting is saved and finished disconnect the cable from the device and mount the device Error States of the device Device continuously checks its state during operation In case error is found LCD displays corresponding error code Error 0 First line displays Err0 Check sum error of stored setting inside device s memory Output value is lt 3 8 mA This error appears if incorrect writing procedure to device s memory occurred or if da
8. ls of building If there are communication conductors placed into conduit there is strongly recommended make it caulk to restrict air flow around device It is recommended to use shielded copper cable e g SYKFY maximal length 1200m The cable must be located at indoor rooms device is not protected for outdoor use The cable should not be led in parallel along power cabling Safety distance is up to 0 5 m otherwise undesirable induction of interference signals can appear Electrical system wiring may do only worker with required qualification by rules in operation Setting of pressure correction for altitude For device T2118 only If there is necessary to read value of atmospheric pressure corrected with respect to altitude it is possible to set corresponding pressure offset for this altitude It is possible to do it with Users software This value is then automatically added to measured pressure value The resultant value of pressure with added correction for altitude must be between 300 hPa and 1350 hPa from 4 351 PSI to 19 580 PSI Outside this range device sends value of Error 2 Device mounting Firstly mount back part of device onto wiring box with two holding screws Connect cables to terminals with respecting the signal polarity see figure Terminals are self clamping and can be opened by a suitable screwdriver For the opening insert the screwdriver to upper terminal hole and lever by him Then insert front part of device fir
9. mage of calibration data appeared At this state device does not measure and calculate values It is a serious error contact distributor of the device to fix Error 1 Measured or calculated value except in pressure is over upper limit of allowed full scale range There is a reading Err1 on LCD display Output value is about 24 5 mA This state appears in case of e Measured temperature is higher than approximately 600 C i e high non measurable resistance of temperature sensor probably opened circuit e Relative humidity is higher than 100 i e damaged humidity sensor or humidity calculation of humidity is not possible due to error during temperature measurement e Computed value calculation of the value is not possible error during measurement of temperature or relative humidity or value is over range Error 2 There is a reading Err2 on LCD display Measured or calculated value is below lower limit of allowed full scale range or atmospheric pressure error occurred Output value is about 3 8 mA This state appears in case of e Measured temperature is lower than approximately 210 C i e low resistance of temperature sensor probably short circuit 10 TE SNC Tx118 02 e Relative humidity is lower than 0 i e damaged sensor for measurement of relative humidity or calculation of humidity is not possible due to error during temperature measurement e Computed value calculation of computed value is not possible
10. ric pressure range from 4 to 20 mA corresponds with 800 to 1100 hPa correction for altitude 0 hPa absolute pressure T3118 output I1 relative humidity range from 4 to 20 mA corresponds with 0 to 100 RH output 12 temperature range from 4 to 20 mA corresponds with 10 to 50 C LCD display switched ON Change of this manufacturer device setting is possible with the PC connected via the optional SP003 communication cable and User s software Tsensor as described at the end of this document 2 TE SNC Tx118 02 Dimensions 88 33 vy 106 T0118 T2118 T3118 IE SNC Tx118 02 3 Device installation Devices are designed for indoor applications It is recommended to mount them on universal wiring box common installation box KU68 with using two enclosed mounting screws For correct function there is necessary to find proper device place It shouldn t be placed at places where it can be affected by sunshine near radiators heating elements and other heat sources air handlers windows doors into racks and shelves and similar places For buildings with less thermal insulation there is not suitable to place them on external wal
11. stly at bottom into hinge and then press both parts together see figure lt 2 lq _Max 20 NN oS T Figurel Device mounting a TE SNC Tx1 18 02 TE SNC Tx1 18 02 Device demounting If there is necessary demount the device insert flat bladed screwdriver max 3 5 mm wide from top side into middle device s air hole There is fastening member placed insert screwdriver beyond the fastening member about 2 cm deep number at the picture Then slightly move screwdriver in arrow direction as shown at the picture It unlock fastening member and the device is partially opened Remove the screwdriver and take front part of device at top see number at the picture Pull front part with pendulum motion remove front cover numbers and at the picture If there is necessary remove back part of device please disconnect cables and unscrew two holding screws Figure2 Device demounting Typical application wiring connection of terminals Calculation of minimum power voltage Uss for proper operation Uss min gt Uo min Imax Rz where Uo min 9 V Uss Imax approximately 20 mA Rz sensing resistor shunt SENSOR For device T3118 Both current loops Il and I2 are connected analogically current loop Il must be always connected 6 TE SNC Tx1 18 02 Info mode If you would like to check output range settings it is possible to verify without a use of the comput
12. to this symbol is measured pressure or error state of value If selected pressure unit is mBar or oz in or mmHg or inH 0 or kPa there is shown only value number without corresponding pressure unit C F DP Reading next to this symbol is calculated dew point temperature or error state of value g m Reading next to this symbol is calculated absolute humidity or error state of value g kg Reading next to this symbol is calculated specific humidity or mixing ratio depends on device setting or error state of value If specific enthalpy is selected there is shown only value number without corresponding unit IE SNC Tx118 02 7 Technical parameters of the device Accuracy data shown are for value displayed on LCD display For value on analog output is valid too if selected output range is set inside measuring range Analog outputs T0118 T2118 one output channel I1 range from 4 to 20 mA T3118 two galvanic isolated outputs I1 and I2 range 4 to 20 mA Current output in case of error approximately lt 3 8 mA or gt 24 mA Power 9 to 30 V DC maximum ripple 0 5 Measuring parameters Ambient temperature T0118 T3118 Measuring range 10 to 50 C Display resolution 0 1 C Accuracy 0 5 C Response time air flow approximately 0 3 m s t90 lt 12 min No air flow t90 lt 25 min Atmospheric pressure T2118 only accuracy of value shown on LCD display Unit hPa mBar