Contents

1. TABLE OF CONTENTS e Symbols used in this manual page 27 e Safety rules page 28 29 30 e Features page 31 32 e Description of parts page 33 e Installation pag 34 e Operating instructions page 35 36 e Maintenance pag 37 38 39 40 41 e Troubleshooting page 42 43 e Data sheets page 44 45 46 e Guarantee certificate page 47 26 SYMBOLS USED IN THIS MANUAL Danger Prohibition The paragraphs marked with this symbol contain This icon identifies what to prevent doing instructions to be followed carefully to avoid to avoid damage to the device damage to the device the operator and the patient Warning Q Suggestions These instructions warn to be very careful to With this icon information is supplied avoid situations which could damage the device permitting more effective use of the device 27 SAFETY RULES Follow all instructions set forth in this manual e The device is to be used in a dental practice only by medical personnel The device must be sterilized before use it each time e The device is not delivered sterilized therefore it must be sterilized before use it e Sterilization parameters must be in conformity with the Official Pharmacopoeia 121 C 249 8 F to 135 C 275 F e Do not immerse the device in any disinfecting liquid e Before sterilization the device must always be cleaned see page 37 e The device must be installed by the user on
2. 6 para 2 of D L 626 94 IEC 801 4 Managing Director 25 DATA SHEETS Direct coupling Delta turbine Borden 3 ways 4 ways Rotation Speed rpm 425 000 420 000 Power output W 12 5 12 3 Operating pressure bar 2 4 0 2 2 4 0 2 Air consumption l min 43 41 Max water consumption ml min 80 80 Noise level dB A lt 60 lt 60 Vibrations m s lt 2 5 lt 2 5 Head dimensions mm 11 7 H 14 11 7 H 14 Weight g 71 74 44 Quick connection Delta turbine Delta light turbine 3 ways 4 ways lux Rotation Speed rpm 420 000 410 000 Power output W gt 10 5 10 9 Operating pressure bar 2 6 0 2 2 6 0 2 Air consumption l min 43 45 Max water consumption ml min 80 80 Noise level dB A lt 60 lt 60 Vibrations m s lt 25 lt 2 5 Head dimensions mm 11 7 H 14 11 7 H 14 Weight g 69 69 70 Light bulb data sheets Power supply V 3 5 Current input mA 760 Power input W 2 6 45 DECLARATION OF CE CONFORMITY The manufacturer FARO S p A Via Faro 15 20060 Ornago Milano ITALY declares under its own responsability that following product Product name Dental Office Turbine Model in the version Delta It is factory new in accordance with the provisions of art 6 para 2 of D L 626 94 It fully conforms to the following European directives and national acceptances 93 42 CKE medical dev
3. Noise level dB A lt 60 lt 60 Vibrations m s 25 lt 2 5 Head dimensions mm 11 7 H 14 11 7 H 14 Weight g gA 74 44 Quick connection Delta turbine Delta light turbine 3 ways 4 ways lux Rotation Speed rpm 420 000 410 000 Power output W 10 5 10 9 Operating pressure bar 2 6 0 2 2 6 0 2 Air consumption l min 43 45 Max water consumption ml min 80 80 Noise level dB A lt 60 lt 60 Vibrations m s gt lt 25 lt 25 Head dimensions mm 11 7 H 14 11 77 H 14 Weight g 69 69 70 Light bulb data sheets Power supply V e Current input mA 760 Power input W 2 6 45 The Quick connection DELTA turbines are designed to be powered with higher pressure Q supply In the device pressure is regulated automatically by a flowrate limiting valve CLASSIFICATION OF DELTA Turbine with light DEVICE e Device of Class II e Device of Type B e Device not protected against liquids penetration e Device not suitable for use in presence of anasthetic mixtures inflammable with air oxygen or nitrous oxide Q To obtain a safe system the use of a FARO electronic circuit is recommended 46 GUARANTEE CERTIFICATE FARO grants to the final customer a 12 months guarantee for the device and 6 months guarantee for the rotor starting from the date of the purchase invoice Repairs under the guarantee shall be done at the FARO premises or by an
4. a specific power supply system dental unit and be connected to a cable e Never push the button when the Turbine is still rotating e Sistems of clamping to the cables meet with EN standards 29168 e Fit in the turbine only burrs with stems meeting with ISO standards 1797 e Make sure the burr is perfectly seated The DELTA version with light is not protected against liquids penetration e Do not attempt to repair the device yourself Only perform operations set forth in this manual Any operation not described in this manual could prejudice the safety designed into the device In all other cases apply to our Service Department e The DELTA turbines must be used only for the purpose described e When the equipment is packed for shipment or stock it can lie for maximum 15 weeks to environment conditions that do not exceed the following ranges Sorrounding temperature of 20 C 4 F or 70 C 158 F Relative humidity from 10 to 90 Atmospheric pressure from 500 to 1060 mBar e This turbine in the version with light must be powered in conformity with CEI EN 60601 1 30 FEATURES The purpose of the device is to provide movement to a rotating body i e a burr for performance of specific dental operations DELTA turbine is produced in the following versions Delta P 1 with push button Borden 4 ways or Quick Connection coupling Delta P 3 with push button Bo
5. in axial direction in the turbine taking care NOT to hold it on the connecting ringnut 34 OPERATING INSTRUCTIONS Test the turbine to make sure it rotates and air and water come out of the holes spray To prevent exhausted air from striking the hand DELTA turbines with Borden coupling are made to exhaust into FARO cable If the cable doesn t have this exhaust A remove the O ring B using a small tool fig 1 fig 1 35 Replacing the burr To replace the burr in DELTA turbines just press the push button located on the head of the device fig 2 The push button burr locking system is designed to be fully effective when it is rotating Therefore do not attempt to remove the burr with the turbine stopped without pressing the push button to avoid damaging the locking system fig 2 36 MAINTENANCE Cleaning w e This device can be sterilized in conformity with the parameters of the Official Pharmacopoeia for example 121 C 249 8 F to 135 C 275 F e The device must be sterilized after each patient e Before sterilization the device must always be cleaned externally with a brush under running water or with an alcohol based solution e DELTA turbines can be disinfected externally with a chemical disinfectant disinfection with spray or with moistened cloth e Do not leave DELTA turbines immersed in a disinfe
6. authorized FARO Service person with transportation costs and risks at the expense of the purchaser The guarantee certificate found on the endleaf of this booklet is considered valid only if it is entirely completed and bears the reseller s stamp Repairs under guarantee during the period of effectiveness wil be done only if the article to be repaired is accompanied by the shipping note or purchase invoice The guarantee covers failures due to poor material quality or manufacturing defects and in case of legitimate claim the guarantee will allow repair or replacement free of charge No compensation for damages and or interest will be recognized The guarantee is not deemed valid in the final judgment of FARO in case of tampering damage incorrect use faulty maintenance or normal wear The guarantee does not cover light bulbs and fuses l 47 DECLARATION OF CE CONFORMITY The manufacturer FARO S p A It fully conforms to the following European Via Faro 15 directives and national acceptances 20060 Ornago Milano ITALY 93 42 CEE medical devices directive Class which the equipment belongs to Ia declares under its own responsability that following it conforms to the following standatds product CEI EN 60601 1 Product name Dental Office Turbine CEI EN 50081 1 CEI EN 50082 1 Model in the version Delta CEI EN 55022 IEC 801 2 It is factory new in accordance with the provisions of art IEC 801 3 Dott Angelo Favonio
7. ction liquid e This device cannot be cleaned in thermodisinfecters 37 Lubrication DELTA turbines must be lubricated either daily and before and after each sterilization by spraying the special FARO lubricant into the rear do not use other lubricants To lubricate the turbines with Borden A or 4 ways B couplings proceed as follows fig 3 e Spray lubricant into the device by inserting the white aerosol bottle e Reconnect the cable to the device and run it in vain for a few seconds Unscrew the ringnut connecting the cable to the turbine injector in holes 1 and 2 38 To lubricate turbines designed for QUICK CONNECTOR proceed as follows fig 4 Disconnect the turbine by sliding the ringnut of the Quick connector axially Fit the lubricant aerosol bottle injector in the hole of the turbine and spray in the lubricant Remove the injector from the turbine and connect it to the Quick connector Run the device in vain for a few seconds O ALWAYS LUBRICATE THE TURBINE BEFORE AND AFTER EACH STERILIZATION 39 Changing light bulb To change light bulb L proceed as follows fig 5 Disconnect the turbine by sliding the ringnut of the Quick connector axially Unscrew the sleeve M of the turbine body counterclockwise and withdraw it up to stop Remove light bulb with fingers Fit the new light bulb bein
8. g careful to snap it correctly into position in its seat Screw down the sleeve and connect the cable fig 5 40 Cleaning nozzles If the small holes from which emerge water and air are clogged free the passage using the provided pin S fig 6 It is recommended to carry out this operation regularly once a week fig 6 41 TROUBLESHOOTING The turbine sounds Call our Service Department different it vibrates too much and cutting power is reduced The optical fiber turbine Check light bulb condition and replace it if necessary see page 40 does not light In case of difficulty call our Service Department 42 No water or spray comes Disconnect the turbine from the power supply and check if water comes out out of the turbine If not call our Service Department Check to see if the holes in the turbine are clogged and if necessary clean them see page 41 In case of difficulty call our Service Department The burr locking Call our Service Department system or the rotor are damaged 43 DATA SHEETS Direct coupling Delta turbine Borden 3 ways 4 ways Rotation Speed rpm 425 000 420 000 Power output W amp 123 12 3 Operating pressure bar 2 44 0 2 2 4 0 2 Air consumption l min 43 41 Max water consumption ml min 80 80
9. ices directive Class which the equipment belongs to Ila It conforms to the following standards CEI EN 60601 1 CEI EN 50081 1 CEI EN 50082 1 CEI EN 55022 IEC 801 2 IEC 801 3 Dott Angelo Favonio IEC 801 4 Managing Director 25
10. rden 4 ways or Quick Connection coupling Delta LP 1 with push button light and Quick Connection Delta LP 3 with push button light and Quick Connection Q ALL DELTA TURBINES HAVE A CONTAMINATED LIQUID HOLDING SYSTEM 31 e Delta 1 and Delta 3 turbine versions have two different cooling systems Cooling with one spray jet gives a better view of the operating area Cooling with three spray jets gives total spreading of spray for specific operations e Burr replacement system by pressing push button is provided on the turbine head e Versions with Borden and 4 ways connections to cable e Quick connection system to cable e Optical fiber version providing illumination of the operating area and Quick connection system to cable e All DELTA versions can be sterilized in an autoclave in conformity with the Official Pharmacopoeia 121 C 249 8 F to 135 C 275 F Push button Optical fiber Borden coupling DESCRIPTION OF PARTS 4 ways coupling Quick Connection nA WN re 33 INSTALLATION Check to make sure the following items are received e Dental turbine Delta e Pins e Operating manual THE DEVICE CAN BE INSTALLED BY USER e Make sure that compressed air is dry and oilfree filtered e Screw the turbine to the ringnut of cable or fit it in the quick connection if there is one e In the Quick connection version fit the Quick connector