chapter 8 calibrating and maintaining multiprobe instruments

image

Contents

1. Temperature pH4 pH7 pH 9 pH 10 C Standard Standard Standard Standard 10 4 00 7 06 9 33 10 15 11 4 00 7 06 9 32 10 14 12 4 00 7 05 9 31 10 13 13 4 00 7 05 9 30 10 12 14 4 00 7 04 9 29 10 11 15 4 00 7 04 9 28 10 10 16 4 00 7 04 9 27 10 09 17 4 00 7 03 9 26 10 08 18 4 00 7 03 9 25 10 07 19 4 00 7 02 9 24 10 06 20 4 00 7 02 9 23 10 05 21 4 00 7 01 9 22 10 05 22 4 00 7 01 9 21 10 04 23 4 00 7 00 9 20 10 03 24 4 00 7 00 9 19 10 01 25 4 00 7 00 9 18 10 00 26 4 00 7 00 9 17 9 99 27 4 00 7 00 9 16 9 98 28 4 01 6 99 9 16 9 98 29 4 01 6 99 9 15 9 97 30 4 01 6 99 9 14 9 96 31 4 01 6 99 9 13 9 95 32 4 01 6 99 9 13 9 94 33 4 02 6 98 9 12 9 93 34 4 02 6 98 9 12 9 92 35 4 02 6 98 9 11 9 91 36 4 02 6 98 9 10 9 90 37 4 02 6 98 9 09 9 89 38 4 02 6 98 9 08 9 88 Calibration and Maintenance 8 28 10 2008
2. Calibrating YSI Instruments Specific Conductance This procedure calibrates conductivity specific conductance salinity and total dissolved solids Figure 8 1 Water Saturated Air Calibration 1 Calibration Cup 2 Filling Calibration Cup 3 Putting Sonde in Cup 4 Screwing Calibration Cup to Sonde Calibration and Maintenance 8 7 10 2008 Pour enough standard into the calibration cup to fully immerse the cell and temperature sensor Choose a standard within the same conductivity range as the ambient water to be measured Note Standards with conductivities less than 1 mS cm 1000 uS cm are not recommended m Rinse the sensor twice with the conductivity standard Fill the calibration cup with the standard Make sure that the probe is completely immersed past the vent hole Gently tap the side of the calibration cup to dislodge any air bubbles from the cell m Allow at least one minute for temperature equilibration to occur before processing m From the Calibrate Menu select Conductivity to access the calibration procedure or 1 SpCond to initiate the procedure for calibrating specific conductance Enter the calibration value of the standard ms cm at 25 C and press Enter The current values of all enabled sensors will appear on the screen and will change with time as they stabilize m Record the initial specific conductance value value of standard and temperature of the standard in the calibration log m Ob
3. 7 and 10 or pH 4 and 7 less than 160 mV is an indication that the sensor is malfunctioning and pH data are suspect Calibration and Maintenance 8 9 10 2008 DO Rapid Pulse Sensor Calibration for Instantaneous Sampling When using the YSI Model 600XLM or Model 6920 for instantaneous discrete sampling disable the auto sleep function From the Main Menu select 8 Advanced and then 2 Setup If the auto sleep functions are enabled select 5 Auto Sleep RS232 and 6 Auto Sleep SDI12 and press Enter to disable them Pour about inch of water in the bottom of the calibration cup Place the probe in the cup Make certain that the DO and temperature probes are not immersed in the water Engage only one thread of the calibration cup to ensure that the DO probe is vented to the atmosphere Wait at least 10 to 15 minutes for the air in the calibration cup to become water saturated and for the temperature to equilibrate Note Run the sonde in Discrete Mode while waiting This will help burn in the sensor and improve stability Record the initial DO saturation value and temperature in the calibration log From the Calibrate Menu select 2 Dissolved Oxygen then 1 DO to access the calibration procedure for DO percent saturation Enter the current BP in mm Hg See below Barometric Pressure Observe the temperature and DO readings when there is no significant change for approximately 30 seconds press Enter The screen will i
4. Ifthe DO charge is higher than 75 perform maintenance on the DO probe m Remove the membrane and dry the probe completely with lens cleaning tissue m Next hold the probe in a vertical position place one of the fine sanding disks in the 6035 DO Probe Reconditioning Kit under your thumb and stroke the probe face in a direction Calibration and Maintenance 8 22 10 2008 parallel to the gold electrode The motion should be similar to that used in striking a match Stroke the electrodes 10 times in each direction m After completing the sanding procedure rinse the probe face with water and wipe with lens cleaning tissue to remove any grit left by the sanding disk m After cleaning thoroughly rinse the entire tip of the probe with water m Replace the KCI solution install a new membrane and change the O ring that retains the membrane A loose fitting O ring can allow the electrolyte to leak out and cause noisy data Notes 1 Use only the fine sanding disks provided in the 6035 maintenance kits 2 Sand parallel to the gold electrode Conductivity and Temperature Probe m The conductivity cell must be cleaned regularly to remove deposits formed on the electrode The conductivity cell constant should be in the range of 4 5 5 5 m Dip the cleaning brush into water and insert into each hole 15 20 times You may also use a mild detergent to remove deposits from the electrodes m Rinse the cell with tap water then rinse it several
5. Rinse everything with deionized water before refilling and reassembling m When replacing the pH reference sleeve fill the sleeve completely with pH reference electrolyte Point the sonde downward and push the sleeve up until it contacts and just covers the O ring Then point the sonde upward and continue to push the sleeve all the way to the base of the probe This allows any trapped air to rise to the top where it is forced out through the porous Teflon junction with excess electrolyte As a side benefit it also serves to clean the Teflon junction m Inspect the pH reference sleeve for air bubbles by observing the pH reference probe while inverting the sonde If a large bubble is found repeat the filling procedure m Some Series 4 4a and 5 Sondes may be configured with an integrated pH reference probe In those cases the Teflon junction is noticeably smaller and installed using a screwdriver m To refill an integrated pH reference probe with pH reference electrolyte point the sonde upward and fill the reference reservoir using a plastic syringe This ensures that the reservoir is completely filled and no large air bubbles are left When it is full install the Teflon Calibration and Maintenance 8 18 10 2008 junction with a screwdriver using care to not over tighten the junction Observe excess electrolyte being forced out through the porous Teflon junction Every 12 months m Replace the Teflon junction and O ring m Store spar
6. bubbles and surface films Inspect the appearance of the DO sensor and the electrolyte Note any discoloration If probes appear Calibration and Maintenance 8 5 10 2008 questionable perform sensor maintenance according to accepted procedures Confirm that the circulator or stirrer is operational Calibration m Remove the DO sensor guard if present Invert the sonde point its sensors upward with the calibration cup in place m Carefully blot the DO membrane dry using a Kimwipe or a soft towel Be careful not to apply pressure to the membrane which can change its tension m Fill the calibration cup with water to a level just below the trimmed edge of the DO membrane and O ring m Cover the calibration cup loosely with the cap to prevent air exchange and allow the sensor to stabilize for about five minutes m Record the initial DO saturation value temperature and calibration standard 100 in the calibration log m Calibrate the DO sensor using DO Saturation Enter the uncorrected BP stated in mm Hg See below Barometric Pressure Using the four function keys under the LCD Screen select Setup Cal Calibrate and then Sonde m Use the arrow keys to scroll through the Minisonde menu options m Scroll to DO Sat Press Select m Use the arrow keys to move the cursor left or right to enter the absolute BP Select the appropriate number then press Done m The messages Calibration Successful Press any key should a
7. n O oo Calibration and Maintenance 8 14 10 2008 m Next calculate the correction factor CF for the DO saturation value This adjusts the table DO saturation value to the local BP The correction factor is determined using the following calculation Uncorrected BP mm 760 mm CF for example 756 8 mm 760 mm 0 996 m After determining the correction factor calculate the calibration value using the following equation table saturation value x CF calibration value for example 8 68 x 0 996 8 82 mg L Calibrating the Depth Sensor m Zero the depth sensor immediately before making the initial measurement with the instrument at the first station of the day m For Hydrolab instruments calibrate by entering zero for the standard at the monitoring site or laboratory if the BP at the lab is the same as that of the sample site to cancel the effect of changes in BP m For YSI instruments from the Calibration menu select Pressure Abs to begin depth calibration Input 0 00 press Enter and wait for the reading to stabilize If no significant change occurs after 30 seconds press Enter to confirm the calibration This zeroes the sensor with regard to the current BP Press Enter again to return to the Calibration menu Record the BP at the time of calibration Post Calibration Check The post calibration check is a QC measure to verify the measurement accuracy of the instrument A post calibration check must be perf
8. range should fall between 0 7 to 1 4 Record the DO charge enabled from the Report Menu The range should be 25 to 75 DO ROX Optical Sensor Calibration Calibration Using Percent Air Saturation One Point The calibration procedure for the DO ROX optical sensor is the same for both discrete and unattended sampling There are several methods for calibrating this sensor YSI recommends one point calibration using water saturated air to obtain sufficient accuracy under normal operating conditions Note Air saturated water calibration of the optical dissolved oxygen ODO sensor will provide the maximum accuracy since it reduces variability associated with changing air temperature This method uses a saturated water bath which can simply be a 5 gallon pail that has been sparged with room temperature air using an aquarium air pump for at least 1 hour Since the air saturated water method is time consuming the water saturated air method is recommended for routine water quality monitoring in the SWQM program Water Saturated Air Calibration m Place the YSI optical DO sensor 6150 into a calibration cup vented by loosening the threads containing about s inch of water Wait approximately 15 minutes before proceeding to allow the temperature and oxygen sensors to equilibrate m Select ODO sat and then Point to access the DO calibration procedure Calibration of the optical DO sensor in the DO mode results in the calibration of
9. reading after conversion to mm can be directly entered provided the weather station is located nearby and at the same general elevation If the BP reading supplied is corrected to sea level obtain the local altitude where the instrument is being calibrated in feet above sea level from a USGS topographic map or other Calibration and Maintenance 8 12 10 2008 source and use the following equation which is also located on each calibration log sheet to account for altitude Barometric Pressure BP Corrected Barometric Pressure CBP 2 5 A 100 where BP estimated absolute barometric pressure CBP local BP corrected to sea level from a weather bureau or other source convert reading supplied in inches to mm inches x 25 4 mm 2 5 decrease in constant atmospheric pressure decreases in mm Hg for each increase in altitude of 30 5 meters 100 feet A local altitude in feet above mean sea level Example A BP reading of 29 50 in mm Hg corrected to sea level at an altitude of 650 feet above sea level is uncorrected by 29 5 inches x 25 4 749 mm Hg BP 749 mm 2 5 650 100 BP 732 8 mm Hg The above equation should be posted in the laboratory or at the site where the instruments are routinely calibrated Once an initial calculation is made the last term of the equation 2 5 4 100 will then be constant and can be subtracted from the corrected BP If an instrument needs to be calibrated at
10. the operating manuals for the display units Appendix 3 for Hydrolab and Section 9 2 for YSI For display units without a BP function a detailed discussion on obtaining the uncorrected absolute BP follows Absolute Barometric Pressure Barometer Absolute BP is defined as what a mercury barometer would read in the room where the calibration is taking place There are several options for determining BP The most direct method is to read a handheld or fixed mounted barometer No corrections are required Weather reports The second option is to obtain a BP reading from a local weather office radio or television station weather radio or similar source Many local television stations have Web sites or links to sites where local weather information including BP is continuously updated The National Weather Service also offers easy to access weather information online See Appendix A Uncorrecting barometric readings corrected to sea level Most barometric readings obtained from local and Internet sources are corrected to sea level to remove the effect of altitude and are reported in inches of mercury Sometimes corrections are reported as altimeter readings in inches of mercury Convert the barometric reading reported in inches Hg to millimeters mm Hg by multiplying inches by 25 4 The corrected BP value will always be available however you may ask for the uncorrected BP Many stations can provide this absolute reading The uncorrected BP
11. times with deionized water m After cleaning check the response and accuracy of the conductivity cell with fresh standard m Check the cell constant to ensure that it is within the specified range m Dry the sonde port and probe connector m Clean probe O rings and apply a very thin coat of lubricant before installation Keep the thermistor clean and free of debris using water and a cotton swab Use mild detergent and a cloth as necessary Notes m To access the conductivity cell constant using the 610 D display from the main menu select Smart Terminal then choose Advanced and choose Cal constants m To access the cell constant using the 610 DM display logger From the Main Menu select Communications and then Smart Terminal Select Advanced and then choose Cal constants m To access the conductivity cell constant using PC 6000 or Ecowatch software select 8 Advanced press Enter and then choose 1 Cal constants pH Probe Cleaning is required whenever deposits or contaminants appear on the glass surface of the probe Use clean water and a cotton swab to remove all foreign material from the glass bulb Use a moistened cotton swab to carefully remove any material that may be blocking the reference electrode junction of the sensor If good pH is not restored perform the following procedure 1 Soak the probe for 10 to 15 minutes in clean water containing a few drops of dishwashing liquid 2 Gently clean the glass bulb
12. with a cotton swab Calibration and Maintenance 8 23 10 2008 3 Rinse the probe with clean water wipe with a cotton swab saturated with clean water and then rinse with clean water again If you suspect biological contamination of the reference junction or if good response is not restored by the above procedures perform the following cleaning steps 1 Soak the probe for approximately one hour in a 1 1 dilution of commercially available chlorine bleach and water 2 Rinse the probe with clean water and then soak for at least one hour in clean water with occasional stirring to remove residual bleach from the junction If possible soak the probe for a period of time longer than 1 hour in order to remove all traces of chlorine bleach Then rinse the probe again with clean water and retest Caution Dry the probe port and probe connector and apply a very thin coat of O ring lubricant to all O rings before reinstallation Depth Sensor Clean the depth sensor after each deployment or when readings become unstable At the side of the instrument there is a circular cap with two or four small holes that protect the depth sensor The cap cannot be removed but a syringe is supplied in the maintenance kit for cleaning the pressure port Fill the syringe with clean water place the tip of the syringe into one of the holes and gently force water through the pressure port Ensure that water comes out of the other hole Continue flushing the pres
13. CHAPTER 8 CALIBRATING AND MAINTAINING MULTIPROBE INSTRUMENTS This chapter describes the calibration procedures for the most commonly used multiprobe instruments Since the TCEQ and the CRP planning agencies use Hydrolab and YSI products this section was written to address calibration and maintenance of those products If another multiprobe instrument for example Greenspan In Situ is being used refer to the manufacturer s instruction manual Such other instruments must meet the post calibration error limits discussed in this chapter The manufacturers manuals also detail other instrument functions not described in this chapter such as downloading data For all instruments used manufacturers maintenance and calibration manuals must be kept for reference All calibration and maintenance activities must be recorded ina SWQM Multiprobe Calibration Logbook or something similar see Figures 8 3 and 8 4 for examples Calibrate parameters in the order given in this chapter and on the calibration log sheets specific conductance pH and DO see Figures 8 3 and 8 4 Calibration Logbooks for SWQM Each instrument must have its own logbook to facilitate efficient review of calibration and post calibration results and maintenance procedures The SWQM Program has developed logbooks for use with Hydrolab and YSI instruments they are available upon request from the SWOQM Program or the CRP Both the Hydrolab and YSI calibration logbooks sum
14. a remote location where the BP is not available from the usual sources the BP can be estimated from the following equation Barometric Pressure BP 760 2 5 A 100 Example If the altitude at the site of calibration is 1 200 feet above sea level the estimated BP is BP 760 2 5 1200 100 BP 730 mm Hg Calculating a DO Calibration Value mg L Hydrolab and YSI multiprobe instruments use DO saturation to calibrate the DO sensor Other instruments may be calibrated using a calculated DO calibration value in mg L Use the following procedure to calculate a DO calibration value in mg L m Using the temperature recorded at the start of calibration go to the DO saturation table Table 8 1 to obtain the DO saturation value For example the temperature is 22 2 C The DO saturation value for this temperature is 8 68 mg L m Get the local BP value as described in the previous section Uncorrect the BP if necessary For this example the uncorrected BP is 756 8 mm Calibration and Maintenance 8 13 10 2008 Table 8 1 DO Saturation Values in mg L Oxygen Content of Air Saturated Freshwater at 760 mm Hg Temp Oxygen Solubility at 760 mmHg 3 43 3 40 3 33 13 22 13 12 E 1 78 1 75 1 73 Eao 11 64 11 56 11 23 1 21 1 18 Pus fe 11 10 11 02 HEEE 9 83 9 81 9 79 9 77 9 73 rhe so oo 8 45 8 44 8 42 7 46 7 43 7 10 3 7 06 7 05 7 03 7 02 7 01 7 00 6 99 6 97 7 N a 3
15. a thin film of white silicone m As necessary use tissue paper to remove old traces of silicone and dirt and then reapply the silicone Before calibrating flush the entire instrument with clean fresh water Use soapy water and a soft brush to clean the outside surfaces of the instrument A mild dishwashing detergent will work very well Soak the entire instrument in fresh water for at least 30 minutes Note DO membrane Exercise particular caution when cleaning the polarographic dissolved oxygen membrane as its tension may be affected Note pH electrode Use a cotton swab or a soft brush to clean the glass bulb of the pH electrode Conductivity Every two months or once every 15 field trips m For instruments with cell blocks polish the conductivity electrode shafts and tips with emery paper m Remove the O rings on the electrodes before polishing the electrode shafts Take care not to scratch the glass pH probe m Swab the conductivity cell block with a cotton swab soaked in methanol to remove dirt grease and other substances Rinse twice with deionized water m Inspect the small O rings and replace them if stiff cut or flattened m Reassemble the components ensuring that the O rings are wet to create a proper seal m Rinse well with deionized water Calibration and Maintenance 8 17 10 2008 m If possible let the conductivity electrodes stand in tap water overnight before calibrating again m For newer genera
16. be required Batteries Every two months or once every 15 field trips m Review the calibration and replacement schedule for batteries m Recharge the 6 volt NiCad Gel cell batteries nickel cadmium or nickel metal hydride for 12 to 24 hours regardless of the voltage displayed by the instrument Ensure that NiCad batteries are recycled or disposed of properly They should not be put in the regular trash Stirrer Every two months or once every 15 field trips For 4000 Series Surveyor 2 and Series 3 Stirrers Lift the impeller off its post Thoroughly clean all lubricant dirt and debris from the impeller and the post Reapply a very small amount of silicon grease to the tip and replace the impeller on its Calibration and Maintenance 8 19 10 2008 post Test the stirrer s performance if sluggish wipe away excess silicon grease from the impeller post and retest For Series 4 4a and 5 Sondes and Quanta Transmitters Remove the screw that holds the impeller in place Visually inspect the impeller it should move very freely on the screw If there is any resistance clean the bore in the impeller using the screw Replace the impeller if it should show extreme wear Test the screw for straightness by laying it on a desktop and gently pushing it A straight screw should roll in a circle without any noticeable wobble Replace it if crooked Visually inspect the impeller bearing for extreme wear Replace it if it is worn out Rein
17. calibration log See pH Millivolt Response below for additional information on this response indicator m Record the initial pH value and the temperature of the calibration standard in the calibration log m Observe the pH reading When it shows no significant change for approximately 30 seconds press Enter After the second calibration is completed press Enter again If you are performing a two point calibration the screen will return to the Calibrate Menu Record calibration information in the calibration log Rinse the sonde with DI water m Discard the pH buffer and rinse with DI water to remove any residual buffer before calibrating the DO sensor pH Millivolt Response YSI multiprobe instruments report millivolts mV as an indicator of pH sensor response Millivolt response should be recorded in the logbook to assist the user with in troubleshooting activities and to track sensor performance over time An out of range response alerts the user that a probe may soon require reconditioning or replacement There are expected mV response ranges for the various pH buffers Buffer Response 4 130 to 230 7 50 to 50 10 130 to 230 Slope 160 to 185 The mV response is a maintenance tool not a QA or QC measurement An out of range mV response does not signify an inaccurate measurement and does not mean that data need to be flagged as suspect However a sensor slope mV difference between pH
18. cycle Visually inspect the LDO sensor cap Use optical tissue or a cotton swab with soapy water to clean the sensor cap Rinse with fresh water Note Do not use organic solvent solutions such as acetone or methanol with the Hach LDO sensor Those solvents will damage the plastic sensor cap Avoid removing the LDO sensor cap unless the cap is being replaced If the cap is sealed properly using the top O ring seal no water should be present between the cap and the clear plastic window and the top of the probe If water is present remove the cap and thoroughly dry the inside of the cap and the clear plastic window The cap may require replacement For detailed instructions on replacing the LDO sensor cap refer to the Hach LDO Sensor Instruction Sheet see Appendix A Sonde Every 12 months Replace desiccant inside the display and sonde units YSI Multiprobes The following schedule for minimum maintenance is mandatory regardless of the use that the instruments get This schedule has been developed from the record of instrument performance and in consultation with the manufacturer DO Probe m Change the KCI solution and membrane before each long term deployment and at least once every 30 days m Replace the KCI and membrane if a air bubbles are visible under the membrane b deposits of dried KCl appear on the membrane or O ring c the readings are unstable or d the DO charge reading is less than 25 or greater than 75 m
19. d after calibration DO electrolyte and pH reference solutions have an indefinite shelf life Specific Conductance Calibration Standards Calibrate the conductivity system with a potassium chloride solution of known specific conductance Choose a standard solution with a specific conductance similar to but higher than that of the water being sampled For example if the same instrument is used to sample water bodies with specific conductance values ranging from 250 to 800 uS cm calibrate with a standard greater than 800 uS cm typically 1 500 uS cm Calibrating Hydrolab Instruments Specific Conductance Series 3 DataSonde 3 H20 Recorder Reporter and Quanta Sondes Series 3 and Quanta Sondes require a one point conductivity calibration no zero needed m Before calibration rinse the sensors twice with DI water m Use the rubber cap to cover the calibration cup and shake vigorously m Rinse at least twice with the chosen standard solution and dispose of the solution m Fill the calibration cup covering the DO sensor with the calibration standard and allow the sensor to stabilize Calibration and Maintenance 8 3 10 2008 In Series 3 Sondes check for air bubbles that may have become trapped in the conductivity cell block Remove bubbles if present Record the initial specific conductance value value of standard and temperature of the calibration standard in the calibration log Using the four Function Keys under the LCD Sc
20. down into the calibration cup Do not fully attach the calibration cup The goal is to block air exchange with the outside environment Fully attaching the cup increases the inside pressure and will give a false reading The water should not touch the sensor cap see Figure 8 1 m Allow the DO and temperature readings to stabilize for approximately 5 10 minutes At this point the air inside the calibration cup should be fully saturated with water m Record the initial DO saturation value temperature and calibration standard 100 in the calibration log m Calibrate the DO sensor using LDO Saturation Enter the uncorrected BP stated in mm Hg See below Barometric Pressure Using the four function keys under the LCD Screen select Setup Cal Calibrate and then Sonde m Use the arrow keys to scroll through the Minisonde menu options m Scroll to LDO Sat Press Select m Use the arrow keys to move the cursor left or right to enter the absolute BP Select the appropriate number then press Done m The messages Calibration Successful Press any key should appear Select any function key to remove the message m Select Go Back and confirm the reading for DO percent saturation from the display screen The value should be at or near 100 percent m Record the value calibrated to in the calibration log For Hydrolab instruments the value is 100 percent indicating oxygen saturation referenced to the uncorrected BP input by the user
21. e junctions in a 2 5 molar gt 50 000 uS cm KCI solution PH Troubleshooting If pH still doesn t calibrate correctly do the following m Evaluate the condition of the Teflon junction on the terminal end of the pH reference sleeve The sleeve should slide on easily with some force applied If the sleeve is difficult to apply then the junction may have become clogged In contrast if the sleeve slides on too easily with little resistance the junction is too porous In both instances the junction must be replaced m Ifreplacing the junction does not solve pH problems then clean the probe by alternately soaking it in 0 1 N HCl and then in 0 1 N NaOH for five minutes in each solution Use the small black caps that protect display unit terminals to isolate the probes for soaking with these solutions Safety Note Wear safety glasses and gloves when working with these corrosive chemicals m For 4000 series Surveyor 2 and DataSonde 1 instruments clean the glass pH and reference probes at the same time Do not use these solutions on the metallic posts of newer generation products H20 Reporters Recorders DataSondes 3 or 4 and MiniSondes m Reattach the calibration cup fill it with pH 4 0 buffer and allow all probes to soak an additional 10 minutes m Thoroughly rinse with deionized water and refill the pH reference sleeve m If those two procedures do not correct pH problems check with the manufacturer A new probe may
22. face Temperature Sensor Check Temperature is essential to the successful calibration of the other instrument parameters DO pH and specific conductance so it is important to determine the continued accuracy of the temperature probe Multiprobe temperature sensors are typically very stable and accurate over a long period of time Temperature sensors are factory calibrated and cannot be adjusted by the user Calibration and maintenance other than general cleaning of the sensor are not required However a check must be performed to ensure the proper calibration of other multiprobe water quality sensors Temperature data are used by every sensor in the sonde A malfunctioning temperature probe will cause calibration errors Check the temperature sensor accuracy during routine instrument maintenance m Using water acclimated to room temperature put a laboratory thermistor or thermometer with an accuracy of 0 2 C and the instrument into the same water bath This can be an ice chest or a large bucket Calibration and Maintenance 8 2 10 2008 m Allow the temperature to stabilize and record the reading from the multiprobe instrument and the laboratory thermistor or thermometer in the calibration log The difference between the two temperatures must be within 0 2 C m Ifthe temperature of the multiprobe is off by more than 0 2 C schedule the instrument for recalibration by the manufacturer If the temperature is off by 0 5 C
23. he acceptable range perform maintenance procedures including electrode polishing and replacement of the KCl solution and the Teflon membrane Calibration and Maintenance 8 10 10 2008 DO Rapid Pulse Sensor Calibration for Unattended Sampling When using the Model 600XLM or Model 6920 in unattended mode make sure the auto sleep functions are enabled From the Main Menu select 8 Advanced and then 2 Setup Ensure that 5 Auto Sleep RS232 and 6 Auto Sleep SDI12 are enabled If not select 5 Auto Sleep RS232 and 6 Auto Sleep SDI12 and press Enter to enable both functions Verify that the sonde DO warm up time located in the Advanced Menu gt Sensor is set correctly The 600XLM factory default is 40 seconds for all other instruments 60 seconds is recommended When setting up any YSI instrument ensure that the warm up time is set at 90 seconds rather than the default Calibrate as described in the previous section YSI Instruments DO Rapid Pulse Sensor Calibration Procedure for Instantaneous Sampling Place the sonde in the calibration cup vented with water and allow 15 minutes for the water to saturate Select DO and activate the countdown timer for the DO warm up time After DO warm up is complete the readings just before and after calibration are displayed Press Enter when prompted and the screen will return to the DO Calibration Menu Record the DO gain found at the Advanced Menu gt Cal Constants The
24. ic pressure inches x 25 4 BP mm From local source correction CBP BP mm CBP mm 2 5 altitude ____ 100 Estimated from altitude only BP mm 760 mm 2 5 altitude 100 Post Calibration Date Time Employee Name Battery Voltage Sonde Type and Serial No Was this within Were data post calibration associated with Temp of Value of Initial error limits this Cal Post Cal Function Standard Standard Reading Yes No used Yes No Specific conductance pH calibrated 7 PH slope 4 10 Dissolved oxygen DO saturation standard calculation Current barometric pressure in mm 760 x 100 Altitude A feet above MSL Barometric pressure inches mm MAINTENANCE Refer to Chapter 8 for maintenance requirements Temperature The difference between the thermometer and sonde Check Sonde Thermometer Difference should be within 0 2 C Temperature Perform temperature check along with regular maintenance Sensor Date Initials Maintenance Completed pH DO Specific Conductance Location of Deployment Routine Run or Special Study Date Time Date Time Deployed Retrieved Use P 24 hour Continuous Grab Referee circle one TCEQ 20116 Rev 09 20 08 Page 1 of 1 Figure 8 4 Hydrolab Multiprobe Calibration and Maintenance Log Table 8 3 pH Calibration Standards
25. ion manual for specific details on programming instruments and downloading data from long term deployments Calibration and Maintenance 8 25 10 2008 Calibration Date Time Employee Name Battery Voltage Sonde Type and Serial No Temp of Value of Calibrated Function Standard Standard Initial Reading to Comments Specific conductance high 21 000 uS cm pH slope 4 10 pH mV for pH 10 Range 130 to 230 mV pH mV for pH 4 Range 130 to 230 mV Dissolved oxygen sat Dissolved oxygen charge Range 25 to 75 Dissolved oxygen gain COO CO B Range 0 7 to 1 4 DATA NEEDED FOR DISSOLVED OXYGEN CALIBRATION Altitude A feet above MSL Barometric pressure inches mm Barometric Pressure BP Options Barometric Pressure Formulas Barometer Barometric pressure inches x 25 4 BP mm From local source after correction CBP BP mm CBP mm 2 5 altitude 100 Estimated from altitude only BP mm 760 mm 2 5 altitude 100 DO saturation standard calculation Current barometric pressure in mm 760 x 100 Deployment Checklist required for data logging only Logging interval Yes No SDI 12 Autosleep enabled Yes No DO warm up time Battery volts in sonde days RS 232 autosleep enabled Yes No Available memory in sonde days Post Calibration Check Date Time Employee Name Battery Voltage Sonde Type and Serial No Temp of Value of Function S
26. llowed up to eight hours to ensure total relaxation If the sonde must be pressed into service 2 3 hours is recommended to allow the membrane to achieve at minimum approximately 80 percent relaxation If it is necessary to use the DO probe before the new membrane has eight hours to relax carefully recalibrate the probe immediately before taking each set of measurements If the gold cathode ring is discolored or appears tarnished polish it lightly with a lint free cloth or a pencil eraser Point the sonde downward to prevent any debris from falling into the DO cell Flush the cell twice with DO electrolyte to remove any debris Calibration and Maintenance 8 21 10 2008 4 The white triangular anode in the center of the DO cell may discolor with use darkening from white to gray to black The darkening is generally caused by a buildup of silver oxide Darkening of the anode indicates wear but does not automatically indicate a problem The anode may be cleaned using a 1 1 solution of household ammonia and deionized water Luminescent DO Cell Series 5 Sondes Only The Hach LDO sensor is not affected by fouling or other debris unless the growth is an organism that locally consumes or produces oxygen such as barnacles or algae growing on the sensor cap Nevertheless the manufacturer recommends periodic maintenance to remove contaminants such as oil biological growth and dirt Sensor maintenance should be conducted after every deployment
27. m Keep the sensors moist m Thoroughly rinse the sensors with tap water upon returning from the field m Pour a small amount of tap water into the storage cup It is not required that the sensors be immersed in water It is recommended that the sensors be stored with at least 10 mL approximately 2 teaspoons of water in the sealed storage cup Calibration and Maintenance 8 16 10 2008 m Avoid storing the sondes where they would be subjected to extreme high and low temperatures Do not allow water to freeze in the storage cup Note Never use alcohol to prevent freezing it may damage sensors Scheduled Maintenance Hydrolab Multiprobes Make sure to observe the following schedule for minimal maintenance regardless of how much use the instruments get This schedule has been developed from the record of instrument performance and from consultation with the manufacturer Routine maintenance Carry out these important steps in preventive maintenance m Post calibrate the instrument before general cleaning and maintenance m Following post calibration rinse off the sensors and store them in tap water Do not use distilled or deionized water for storage m Keep the watertight rubber cable connectors well lubricated and dry on the inside The best procedure is to store the instrument with all connectors separated and open to the air until dry m Check rubber cable connectors regularly to ensure that the mated surfaces are covered with
28. marize the calibration procedures and contain a number of individual calibration log sheets See Figures 8 3 and 8 4 for YSI and Hydrolab calibration log sheets The calibration log sheets are also available on the Web see Appendix A Keep full calibration logs on file for at least five years or as defined in a program or project QAPP The following is basic information to be recorded in a log for each calibration m The date time and the initials of the person performing the calibration m The name and model number of the instrument being calibrated m The battery voltage Initial instrument readings during immersion in the calibration standard before calibration temperature value of standard and initial reading m The calibrated to value obtained after adjusting the instrument to the calibration standard value As needed the following information is recorded for each instrument m Factory maintenance including the date shipped for any repairs the date returned from repair and a description of the repair work Calibration and Maintenance 8 1 10 2008 m In house instrument maintenance including the date and a description of any maintenance activity for example battery replacement probe cleaning membrane replacement stirrer cleaning reference solution replacement Temperature Controlled Environment Take into consideration the environment in which the multiprobe instrument will be calibrated It is very importan
29. ndard in the calibration log m Using the temperature of the pH 7 buffer solution determine the pH 7 calibration value See Table 8 3 for pH calibration values corrected for temperature m Record the calibration standard value in the calibration log m Using the four function keys under the LCD Screen select Setup Cal Calibrate and then Sonde m Use the arrow keys to scroll through the MiniSonde menu options m Scroll to pH Press Select m Use the arrow keys to move the cursor left or right to enter the pH standard concentration Press Select to choose the appropriate number then press Done m The messages Calibration Successful Press any key should appear Press any function key to remove the message m Record the value calibrated to in the calibration log m Select Go Back and confirm the pH reading from the display screen m Repeat this procedure with either pH 4 or pH 10 buffer solution Choose a pH buffer that best represents the pH of the environment to be monitored Dissolved Oxygen Sensor DO concentrations in water are measured using either a polarographic electrode or optical sensors DO sensor calibration requires a current uncorrected reading for barometric pressure BP See below Barometric Pressure Clark Polarographic DO Cell Precalibration Clean the sonde and the stirrer using running water to remove debris Check the condition of the DO membrane ensure it is intact and free of wrinkles
30. ndicate that the calibration has been accepted Press Enter again to return to the Calibrate Menu Record calibration information in the calibration log For YSI instruments the value calibrated to is determined for each calibration site and oxygen saturation referenced to 760 mm Hg To determine the calibration standard for DO percent saturation use the following equation Current barometric pressure 760 x 100 DO saturation calibration standard DO Rapid Pulse Sensor Indicators There are several indicators of DO sensor response DO gain and DO charge These indicators are recorded in the calibration log DO gain and DO charge ranges should not be considered quality control criteria but rather guidelines to assist the user in performing maintenance and calibration activities Record the DO gain located in the Advanced Menu gt Cal Constants The range should be 0 7 to 1 4 DO gain is similar in principle to an instrument offset and is an indication of how well an instrument has been calibrated If you accept a bad calibration the gain will fall outside of range If DO gain remains out of range and the sensor ceases to calibrate properly its replacement may become necessary Record the DO charge which can be set to appear on the display screen from the Report Menu The range should be 25 to 75 The DO charge is an indicator of membrane electrolyte and sensor electrode condition When the DO charge falls outside of t
31. o equilibrate before reading m From the Calibrate Menu select 4 ISE 1 pH to access the pH calibration choices then press 2 2 Point or 3 3 Point Press Enter and input the value of the buffer at the prompt Press Enter and the current values of all enabled sensors will appear on the screen Observe the pH mV reading This value should be 0 0 mV 50 mV Record this value in the calibration log m Record the initial pH value and temperature of the calibration standard in the calibration log Calibration and Maintenance 8 8 10 2008 m Observe the pH reading and when it shows no significant change for approximately 30 seconds press Enter The display will indicate that the calibration is accepted m Record the value calibrated to in the calibration log m After the pH 7 calibration is complete press Enter again to continue Rinse the sonde with DI water pH 4 or 10 m Rinse the sensors at least twice with pH buffer m Place enough pH 4 or 10 buffer into a prerinsed calibration cup to immerse the pH probe reference junction and thermistor Allow at least one minute for the temperature to equilibrate before reading m Press Enter and input the value of the second buffer at the prompt Press Enter and the current values of all enabled sensors will appear on the screen m Observe the pH mV reading This value should range from 130 to 230 in pH 4 buffer and from 130 to 230 mV in pH 10 buffer Record this value in the
32. or and allow the sensor to stabilize Record the initial specific conductance value the value of the standard and the temperature of the standard in the calibration log Correct the initial specific conductance value to match the standard value and save the calibration Record the initial value the value of the standard the value calibrated to and the temperature in the calibration log pH Sensor Calibrate the pH system with a buffer of pH 7 0 and either pH 4 0 for naturally acidic waters or pH 10 0 for alkaline waters The pH buffers contain high concentrations of phosphate Calibration and Maintenance 8 4 10 2008 During calibration avoid leaving traces of buffer on equipment or at the workplace that could contaminate water samples m Before calibration check the condition of the probes ensure that they are intact and free of surface films If probes appear questionable perform sensor maintenance according to accepted procedures m For Series 3 and Quanta Sondes remove the storage cup from the sonde and replace with the calibration cup For Series 4 4a and 5 Sondes remove the cap from the storage cup Rinse the sensors twice with DI water Cover the cup and shake it m Rinse at least twice with the pH 7 buffer solution m Fill the cup covering the DO sensor with the pH 7 buffer solution and allow the sensor to stabilize for two minutes m Record the initial pH value and the temperature of the calibration sta
33. or more data collected between the accuracy checks may need to be flagged The laboratory thermistor or thermometer must be calibrated annually against a National Institute of Standards and Technology NIST traceable thermometer Some electronic thermistors can be calibrated and are to be adjusted to the NIST value Traditional thermometers cannot be changed check the NIST thermometer against the lab thermometer and determine a correction value For example if the lab thermometer reads 23 C and the NIST thermometer reads 23 3 C a correction factor must be applied to the laboratory thermometer Each time you take a reading add 0 2 C Attach a correction factor label to the thermometer The calibration can be performed for TCEQ personnel either during the annual audit or at the annual SWQM meeting Other monitoring organizations may choose an appropriate time for their annual calibration Calibration Standards and Multiprobe Sensor Solutions Store calibration standards and electrolyte solutions in a temperature controlled environment Date containers upon receipt and again after opening them Label any secondary containers with their contents and expiration date Commercially purchased calibration standards which must be NIST traceable come with an expiration date Do not use calibration standards for specific conductance and pH beyond their expiration dates However expired calibration standards can be used for rinsing probes during an
34. ormed after each use of the instrument and before any instrument maintenance The sooner this procedure is performed the more representative the results will be for assessing performance during the preceding field measurements Calibration and post calibration should be no more than 24 hours apart when used for routine monitoring After making measurements at the last station fill the sampling cup with ambient water not deionized or tap water and return to the lab In the lab take the same care used in performing the initial calibration and m Rinse with DI water and the appropriate calibration standard m Fill the calibration cup with the standard and allow the instrument to stabilize m Read the value directly off the instrument display unit m Record the value in the same log used for the calibration Do not use the calibration menu during the post calibration check No adjustments are made during this process Calibration and Maintenance 8 15 10 2008 The purpose of post calibration is to determine if the instrument has held calibration during the day of sampling Compare the post calibration values to the expected values for the standards so the field measurements for the day can be reported with confidence The difference between the post calibration value and expected standard value can be used to indicate both calibration precision and instrument performance 24 Hour DO Monitoring Post calibration check requirements for
35. other monitoring activities such as 24 hour DO and continuous monitoring vary depending on the deployment time and location of the site For 24 hour DO monitoring the post calibration check should be done as soon as possible when returning from the field The period of time an instrument is left at a site when collecting 24 hour DO data is generally based on the instrument being used the conditions at the site and the operator s experience with the instrument The longer the time between calibration and the post calibration check the greater the risk of losing data due to failure of the QC checks Post Calibration Check Error Limits If post calibration values fall outside the error limits for DO pH and specific conductance then the data collected do not pass QA and should not be reported Table 8 2 If post calibration measurements do not consistently fall within the error limits after in house troubleshooting return the instrument to the manufacturer for maintenance For depth and temperature errors found during the routine checks need to be corrected by the manufacturer Table 8 2 Error Limits for the Post Calibration Check Parameter Value Dissolved oxygen 0 5 mg L 6 saturation pH 0 5 standard units Specific conductance 5 Temperature 0 2 C Depth 0 2matim General Maintenance Document maintenance in the calibration and maintenance log Short Term Instrument Storage
36. ppear Select any function key to remove the message m Select Go Back and confirm the DO percent saturation reading from the display screen The value should be at or near 100 percent m Record the value calibrated to in the calibration log For Hydrolab instruments the value is 100 percent indicating oxygen saturation referenced to the uncorrected BP input by the user Luminescent DO Cell Series 5 Sondes Only Precalibration Clean the sonde using running water to remove debris Inspect the cap of the luminescent dissolved oxygen LDO sensor cap If the cap or the sensor appears questionable perform sensor maintenance according to accepted procedures Water Saturated Air Calibration Note It is important to maintain temperature stability during calibration Keep the sonde out of direct sunlight and away from any other source of heat or other energy that may change the temperature in the cup during calibration If the temperature in the cup changes more than 0 5 C during calibration recalibration is recommended To calibrate the sensor using water saturated air m Remove the calibration cup from the sonde Fill the calibration cup with approximately 4 inch of DI water or tap water specific conductance lt 0 5 mS cm See Figure 8 1 Calibration and Maintenance 8 6 10 2008 Before attaching the calibration cup to the sonde carefully remove any water droplets from the sensor cap m Gently set the sonde with the sensors
37. reen select Setup Cal Calibrate and then Sonde Use the arrow keys to scroll through the MiniSonde menu options Scroll to SpCond uS cm or SpCond mS cm Press Select Use the arrow keys to move the cursor left or right to enter the conductivity standard concentration Press Se ect to select the appropriate number Enter the appropriate concentration and press Done The messages Calibration Successful Press any key should appear Select any function key to remove the message Select Go Back and confirm the specific conductance reading from the display screen Record the value calibrated to in the calibration log Series 4 Series 4a and Series 5 Sondes DataSonde and MiniSonde Series 4 4a and 5 sondes require a two point conductivity calibration Before calibration rinse the sensors twice with DI water Replace the storage cup cap to cover the storage cup and shake vigorously Calibrate the zero point first by leaving the conductivity sensor in air Record the initial value the value calibrated to and the temperature in the calibration log Correct the initial specific conductance value to zero and save the calibration Record the initial value and the value calibrated to zero in the calibration log and the temperature in the calibration log Calibrate the second point slope by rinsing at least twice with the chosen standard solution Fill the storage cup with calibration standard covering the DO sens
38. serve readings under Specific Conductance when no significant change is observed for approximately 30 seconds press Enter The screen will indicate that the calibration has been accepted press Enter again and return to the Calibrate menu Record calibration information in the calibration log Rinse the sonde with DI water Conductivity Cell Constant Record the conductivity probe cell constant after calibration Access the cell constant from the sonde s Advanced menu select Calibration Constants The acceptable range is 5 0 0 5 The probe cell constant is similar in principle to an instrument offset and is an indication of how well an instrument has been calibrated Accepting a bad calibration will cause the cell constant to fall outside of the acceptable range The cell constant is a maintenance tool not a QA or QC measurement An out of range cell constant does not signify an inaccurate measurement and does not mean that data need to be flagged as suspect Maintenance and recalibration of the sensor are recommended if the cell constant is out of range Replacement of the sensor may become necessary if the cell constant remains out of range and the sensor fails to calibrate properly pH Two Point Calibration pH 7 m Rinse the sensors at least twice with pH buffer m Place enough pH 7 buffer into the prerinsed calibration cup to immerse the pH probe reference junction and thermistor Allow at least one minute for the temperature t
39. stall the impeller bushing the impeller and the impeller screw Make certain that the impeller is oriented with the dimples upward Test the stirrer performance Dissolved Oxygen Every six months or once every 15 field trips Clark Polarographic DO Cell m Change the DO membrane and add fresh DO electrolyte Hydrolab part number 000537HY m Invert the sonde on a ring stand m Remove the DO guard the O ring and the old membrane m Shake out the old DO electrolyte m Rinse the DO cell twice with fresh DO electrolyte m Fill the DO cell with fresh DO electrolyte Gently tap the side of the DO cell to release any bubbles trapped in the DO cell m Replace the DO membrane Hydrolab part number 002589HY and secure with the O ring Hydrolab part number 000498HY See Figure 8 2 m Visually inspect the membrane for wrinkles tears or trapped air bubbles m Remove the membrane and pour out the DO electrolyte m Rinse the DO cell with deionized water m Invert the sonde on a ring stand and with a small syringe fill the DO cell with 1 1 ammonia solution until the anode is covered Do not allow the ammonia solution to contact the gold cathode ring Have a moist towel available to wipe the ring if needed Let stand for 10 minutes m Rinse twice with deionized water and refill according to the standard procedure as described above Calibration and Maintenance 8 20 10 2008 With the thumb and forefinger of your other With a con
40. sure port until the water comes out clean Note Never try to remove the circular pressure port cap Long Term Storage of Hydrolab Instruments The following information applies specifically to Hydrolab instruments Refer to the YSI instrument manual for long term storage recommendations specific to YSI instruments Field instruments are often stored for indefinite periods of time For example backup instruments are used during repair of the primary instrument The instrument cannot be kept in a perpetual state of readiness without regular maintenance Whenever multiprobe instruments are to be stored for extended periods of time make the following preparations m Thoroughly clean the probes m Remove installed batteries such as AA or C cells If a lithium battery powers the multiprobe internal clock do not remove it m Fill the storage cup with 1 inch of clean tap water and screw the cap onto the instrument m Store the instrument in a location where freezing will not occur away from direct sunlight The instrument can be reliably reactivated the day before field use with minimal effort Calibration and Maintenance 8 24 10 2008 Downloading Data HyperTerminal is readily available software that can be used to download data from both YSI and Hydrolab instruments YSI also uses EcoWatch for Windows to download data its use is detailed in the YSI 6 Series User s Manual See Appendix A also refer to the manufacturer s instruct
41. t that the standards buffers rinse waters and the instrument itself be acclimated in a environment with stable temperature Deionized DI water used for rinsing sensors should be stored in a large container in the room normally used for calibration The temperature of tap water or DI water from a column used for rinsing is often very different from room temperature If the instrument standards buffers and rinse water are transported to the field and calibration or a post calibration check away from the laboratory becomes necessary make sure the instruments have acclimated to the same temperature For example place the solutions and instruments in a motel room in the evening and calibrate in the morning Rinsing the Sensors It is very important to rinse with DI water between the calibration of specific conductance and pH m Unscrew the cap and remove it from the calibration cup m Fill the calibration cup about halfway with DI water m Place the cap on the calibration cup m Shake the sonde to rinse any contaminants that might interfere with calibration Rinse with DI water at least twice m Repeat this procedure with the appropriate calibration standard Rinse at least twice discarding the rinse each time Fill the calibration cup with the standard and proceed with calibration m Warning Do not allow the calibration cup to touch any of the sensors during this procedure Damage can occur if the cup makes hard contact with a sensor
42. tandard Standard Initial Reading Comments Specific conductance pH calibrated 7 pH slope 4 10 Dissolved oxygen sat Barometric pressure inches Brush conductivity cell Must be cleaned within the last two months or once every 15 field trips or deployment days Inspect DO membrane for nicks or bubbles Must be changed Name comments within last six months or once every 15 field trips or deployment days Temperature Check temperature as part of routine maintenance Date Thermometer Sonde Difference against NIST traceable thermometer Date Time Deployed Date Time Retrieved Location of Deployment Routine Run or Special Study Use circle one TCEQ 20118 rev 09 20 08 Page 1 of 1 24 hour Continuous Referee Figure 8 3 YSI Multiprobe Calibration and Maintenance Log Date Time Employee Name Battery Voltage Sonde Type and Serial No Calibration Temp of Value of Initial Calibrated Function Standard Standard Reading to Comments Specific conductance zero Specific conductance 2nd point or slope pH calibrated 7 PH slope 4 10 Dissolved oxygen DO saturation standard calculation Current barometric pressure in mm 760 x 100 Altitude A feet above MSL Barometric pressure inches mm Barometric Pressure BP Options Barometric Pressure Formulas Barometer Barometr
43. the DO mg L mode m Enter the current uncorrected BP reading in mm Hg mm in x 25 4 See below Barometric Pressure Calibration and Maintenance 8 11 10 2008 m Press Enter and the current values of all enabled sensors will appear on the screen and change with time as they stabilize Observe the readings under ODO sat When there is no change for approximately 30 seconds press Enter The screen will indicate that the calibration has been accepted Press Enter again to return to the Calibrate menu m Record the initial value the value calibrated to and the temperature in the calibration log Record the ODO Gain also found at the sonde s Advanced Menu gt Calibration Constants The value should fall between 0 7 and 1 45 Barometric Pressure Obtaining the correct BP is essential in calibrating the DO sensor BP affects the partial pressure of oxygen in saturated water The higher the BP the more oxygen can be dissolved in water thus the percent saturation calibration value will also be higher If the instrument display unit has an internal BP function installed take the reading directly from the unit However the internal BP function should be checked against a known BP at least once a month Barometric pressure for both YSI and Hydrolab handheld display units can be calibrated The calibration of BP follows the general calibration procedure used for other parameters Details on calibrating the BP are outlined in
44. tinuous motion stretch the Secure the end of the membrane under your Roll the O ring over the end of the probe being Trim off excess membrane with scissors or a Secure a membrane between your thumb and the probe body Add electrolyte solution to the probe until a large meniscus covers the gold colored cathode Note Handle membrane material with care touching only the edges hand grasp the free end of the membrane membrane up over and down the other side of the sensor Stretching forms the membrane to the contour of the sensor tip forefinger while still holding the probe careful not to touch the membrane surface There should be no wrinkles in the membrane or trapped air bubbles under it Some wrinkles may be removed by lightly tugging on the edges of the membrane beyond the O ring knife Figure 8 2 Replacing the DO Membrane text Southern Illinois University 2003 illustration from YSI operations manual Troubleshooting Notes l If anything other than a rapid and stable oxygen calibration occurs replace the membrane as a first step in troubleshooting Please note that the new DO membrane is stretched during installation see Figure 8 2 This stretching affects the rate of diffusion for oxygen molecules passing through the membrane to the DO cell As the membrane relaxes the rate of diffusion will change unpredictably For that reason it is recommended that the newly installed DO membrane be a
45. tion products without cell blocks clean the conductivity sensor with a cotton swab soaked with methanol or detergent pH Every two months or once every 15 field trips m Wipe the pH probe with a cotton swab soaked in warm soapy water Note Alcohol is not recommended for cleaning the pH probe as it may damage some of the sensors by drying the glass electrode For 4000 series Surveyor 2 and DataSonde 1 Instruments m Replace the pH electrolyte solution in the pH reference probe sleeve with 3M KCI solution in pH 7 buffer 225g KCl in 1L of pH 7 buffer For Series 3 4 4a and 5 Sondes and Quanta Transmitters m Replace the pH electrolyte solution with pH reference electrolyte Hydrolab part number 005308HY 4M KCl saturated with silver chloride Additionally potassium chloride pellets Hydrolab part number 005376HY may be added The KCI pellets serve to maintain the pH reference electrolyte at saturation useful if working long term in waters of very low conductivity m Clean the plastic reference probe sleeve and Teflon junction inside and out with a cotton swab soaked in warm soapy water For 4000 Series Surveyor 2 and DataSonde 1 Instruments m Wipe the reference pH electrode glass bulb with a cotton swab soaked in warm soapy water m Series 3 Series 4 Series 4a and Series 5 Sondes and Quanta transmitters use a silver anode in the pH reference probe This anode should not require cleaning or general maintenance m

Download Pdf Manuals

image

Related Search

Related Contents

Sika Wrap Hex 100G - Distribuciones Villamar  exemplar de assinante da imprensa nacional  Metro Enclosure Panels Assembly Instruction  Thrustmaster 4660427 game console accessory  UM7000C シリーズ 監視制御装置 取扱説明書  Continental Electric CE27205 User's Manual  MORIWAKI SKIDPAD KIT  WorkCentre Série 4250/4260 Guia de Utilização Rápida  Administering Oracle Service Bus  