Manual_EN L.sensor.CO2v1.3
Contents
1. L sensor CO SUMMARY INTRODUCTION 2 GOODS RECEPTION 3 MECHANICAL CHARACTERISTICS 4 DISPLAY 4 START UP 5 PARAMETERS SELECTION 6 MEASURE 7 BOTTLE POSITIONING 7 FAST MEASURE 8 HIGH RESOLUTION MEASURE 9 TURNING OFF 10 ERROR MESSAGES 11 USER INFORMATION 12 ALCOHOL TITOL TABEL 12 MEASURE METHODOLOGY 13 SAFETY REGULATIONS 14 PROPER USE OF THE INSTRUMENT 14 WARNINGS 14 LASER CLASS 15 LASER USE 15 TECHNICAL DATA 16 CE CERTIFICATE 17 REFERENCES 18 L PRO SRL ALL RIGHTS RESERVED v 1 3 1 L sensor CO INTRODUCTION Congratulation for buying L Sensor CO2 by LPRO This user manual will give you instructions for the use of the instrument and important safety regulations please refer to SAFETY REGULATIONS page 14 Please read carefully this user manual before using the instrument The instrument uses a touch screen PC interface A normal knowledge of using PC is required to use this instrument Instruction course for the use of the instrument are available at Lprosrl Please contact us L PRO SRL ALL RIGHTS RESERVED v 1 3 2 L sensor CO GOODS RECEPTION This instrument has been completely checked and tested before shipping When receiving the goods please verify the following points check the presence of the computer of the power supply cable and of the main mechanics check every part of the instrument to be sure it has not been damaged during transportation verify t
2. ultima misura Note in stand by red light on the measure system is active Data ultima misura HF Tipo di misura selezionata Alta risoluzione so for a correct turning off it is necessary to turn off the main Diametro selezionato 18 mm Temperatura deila misura 20 0 switch and unplug the power supply cable il diamatro interno de nel punto di misura Salazionare tempe indicata dal termome L PRO SRL ALL RIGHTS RESERVED v 1 3 10 L sensor CO sere MEASURE ERROR MESSAGES The instrument will show some error messages in case some problems may prevent the measure execution some of those errors will be shown by the instrument with error messages Press OK to exit from the warning window The system will reboot Some other error messages may appear to report problems that prevent the execution of the measure without the need of areboot When an error message is displayed you must press OK to go back to the normal operation Some of the errors displayed are the following Low signal LOW it may indicate that the bottle in not correctly positioned Reposition the bottle If the problem reappears frequently please reboot the system Execution failed the instrument did not execute the measure close the error message and repeat the measure If the problem persists reboot the system Measure not valid the instrument did not execute the measure this happens when the pr
3. 1 1 363 1 346 1 329 1 312 1 294 1 277 1 260 1 242 1 225 90 1 358 1 341 1 324 1 307 1 290 1 273 1 256 1 239 1 222 1 205 100 1 336 1 319 1 302 1 286 1 269 1 252 1 235 1 219 1 202 1 185 110 1 313 1 297 1 280 1 264 1 247 1 231 1 215 1 198 1 182 1 165 120 1 291 1 274 1 258 1 242 1 226 1 210 1 194 1 178 1 162 1 145 L PRO SRL ALL RIGHTS RESERVED v 1 3 12 L sensor CO USER INFORMATION MEASURE METHODOLOGY L sensor CO executes the measure using the absorption of the infrared light by the CO molecules during the measure the head space is crossed by an IR ray that is attenuated by the presence of CO The laser wavelength is modified to scan some absorption lines of the gas analyzed We measure the width of the absorption line related to total pressure and its intensity related to the percentage L PRO SRL ALL RIGHTS RESERVED v 1 3 18 L sensor CO SAFETY REGULATION The following instructions must be used by the operator and the responsible of the instrument to operate in safe conditions ALLOWED USE OF THE INSTRUMENT static measure of sparkling wine bottles static measure of CO2 concentration in sparkling wine bottles the measure must be performed with the instrument positioned on a flat and leveled surface a desk or table NOT ALLOWED USE use of the instrument by operator that did not read the manual use in outside environments useon unstable surfac
4. 1bar Ptot 0 1bar PCO accuracy is higher for PET bottles lower for glass bottles Measure time Measurable range Minimum unity displayed 5 sec Ptot from 1 to 8 bar PCO from 1 to 8 bar CO in wine from 0 to 119 1 Ptot 0 1 bar PCO 0 1bar CO in wine 0 1 g l Laser class 1 Laser type IR 1 mW Stand by automatic after 20 min Dimensions approx 570x350x400mm Temperature stock from 15 C to 60 C work from 5 C to 40 C L PRO SRL ALL RIGHTS RESERVED v 1 3 16 L sensor CO L pro Srl Via Lister 1135143 Padova P IVA04200620286 Tel 39 049 9916081 Fax 39049 9919441 Email info pro it www Ipro it Padova 10 12 2007 DICHIARAZIONE DI CONFORMITA ED INCORPORAZIONE DECLARATION OF ACCORDANCE AND INCORPORATION DECLARATION DE CONFORMITE ET INCORPORATION La Soc L pro srl di Padova PD Italia Firm L pro srl from Padova PD Italy La S t L pro srl de Padova PD Italie DICHIARA DECLARES DECLARE In qualit di costruttore sotto la sua responsabilit che i suoi strumenti di misura sono CONFORMI ai requisiti prescritti dalle Direttive Comunitarie Under his own responsibility that his measuring instruments are in ACCORDANCE with the requirements prescribed by the European Directives Sous sa responsabilit que ses moyens de mesure sont CONFORMES aux conditions prescrites par les Directives Communautaires Di 73 23 EEC Bassa tensione Lo
5. ASURE Once selected HIGH RESOLUTION MEASURE set all the parameters required and pressed CONFIRM you will be redirected to the MEASURE page The instrument is ready to measure the bottle Put the bottle in the mandrel and press once the MEASURE button to execute the first of three measures The instrument executes the first measure shows a green V on the Position 1 if everything is ok and prompts the user to rotate the bottle to take the second measure Rotate the bottle of approx 60 then press OK to execute the second measure Repeat the procedure for the third and last measure At the end of the three measures the instrument shows the following data Total pressure bar see page 7 CO pressure at 20 C bar see page 7 CO in wine at 20 C g l see page 7 Last measure time see page 7 Last measure date see page 7 Measure mode see page 7 Diameter selected see page 7 Temperature of the measure see page 7 Press TURN OFF once to turn off the instrument see next chapter Press BACK to go back to the PARAMETERS SETTING page L PRO SRL ALL RIGHTS RESERVED v 1 3 L sensor CO TURNING OFF To turn off the instrument press OFF from one of the pages Wait for the display to turn completely off then turn off the main switch and check the red light on the switch has turn off Misura alta risoluzione Pressione totale 0 0 Bar Pressione CO a 20 C 0 0 Bar discolta 20 C 0 0 gil
6. es or conditions modification of mechanical optical and electrical components opening or removing parts of the insturment execution of any modifications on the insturments WARNINGS The handle of sparkling wine bottles requires care Those are recipients under pressure and may explode Please use the maximum attention Not allowed use may cause damage to things and persons The person in charge of the instrument must inform the operator about all the safety rules and the risk of a not allowed use of the instrument L PRO SRL ALL RIGHTS RESERVED v 1 3 14 L sensor CO SAFETY REGULATION LASER CLASS L sensor CO uses an infrared not visible laser ray that is emitted by position 3 crosses the bottle and is received by the photodiode 9 The signal dissemination by the bottle is not dangerous for the operator The laser is in Class 1 in conformity to IEC60825 1 1993 Safety radiation of laser products EN60825 1 1994 Safety radiation of laser products USE OF LASER CLASS 1 PRODUCTS They can emit radiations in both visible and not visible range and their beams are not dangerous if observed directly in a non continuos mode while they can be dangerous if observed with instrument that amplify and concentrate the optical ray such as microscopes binoculars lens etc L PRO SRL ALL RIGHTS RESERVED v 1 3 15 L sensor CO TECHNICAL DATA Typical measuring acc cy 0
7. essure or CO concentration are out of the measuring range L PRO SRL ALL RIGHTS RESERVED v 1 3 11 L sensor CO USER INFORMATION L sensor CO shows measure of total pressure in Bar inside the bottle The value refers to the actual temperature of the bottle at the moment of the measure CO concentration inside the bottle this value is expressed both as partial pressure at 20 C in Bar and as quantity in wine at 20 C in grams liter this last indication refers to a generic wine with an alcoholic gradation of 12 vol a sugar quantity of 20 g l and it s in an equilibrium state between wine and gas in the head space In case the operator wants to refer to different conditions please refer to the following table with this formula concentrationCO2 in g l PressureCO2 bar coefficient indicated in table ALCOHOL SUGAR g l 6 7 8 9 10 11 12 13 14 15 1 1 558 1 539 1 519 1 500 1 480 1 461 1 441 1 422 1 402 1 383 10 1 538 1 519 1 499 1 480 1 461 1 442 1 422 1 403 1 384 1 365 20 1 515 1 496 1 477 1 459 1 440 1 421 1 402 1 383 1 364 1 345 30 1 493 1 474 1 456 1 437 1 418 1 400 1 381 1 362 1 344 1 325 40 1 470 1 452 1 434 1 415 1 397 1 378 1 360 1 342 1 323 1 305 50 1 448 1 430 1 412 1 394 1 376 1 357 1 339 1 321 1 303 1 285 60 1 425 1 408 1 390 1 372 1 354 1 336 1 319 1 301 1 283 1 265 70 1 403 1 385 1 368 1 350 1 333 1 315 1 298 1 280 1 263 1 245 80 1 38
8. he power supply voltage written in the identify label If your net power supply voltage is different from the one in the label DO NOT turn on the instrument Unplug the power supply cable and contact L pro srl for clarifications L PRO SRL ALL RIGHTS RESERVED v 1 3 3 L sensor CO MECHANICAL CHARACTERISTICS Laser ray not visible Fotodiode Mechanical centering system for the bottle Bottle height regulation Height arrest warrant Touch screen display Cooling fan Laser multipin connector Fotodiode coaxial connector 10 Ethernet connector riserved 11 Air intake 12 Power supply and general switch DISPLAY The display is touch screen With the display you have access to all the instrument s features and to the visualization of the measure results 7 8 9 10 11 12 L PRO SRL ALL RIGHTS RESERVED v 1 3 4 L sensor CO START UP ATTENTION connect the coaxial and the multipin connectors ONLY when the computer is turned off and unplugged from the power supply net Connect the coaxial and multipin connectors on the back of the computer Connect the power supply cable to the domestic supply net at 230Vac 50Hz to the connector on the back of the computer Turn on the instrument by turning on the main switch and verify that the red light on the switch turns on The instrument will start up automatically Wait for the start up procedure to complete DO NOT touch the display duri
9. nfirm Confirms the parameter selected and goes to the measure page Turn OFF Turns off the instrument Please wait till the display turns completely off then turn off the main switch NOTE after some minutes of inactivity the instrument goes back automatically to the PARAMETER SELECTION page After some more minutes of inactivity the display turns off to protect itself screen saver To turn the display on please touch the display L PRO SRL ALL RIGHTS RESERVED v 1 3 L sensor CO MEASURE BOTTLE POSITIONING Put the bottle on the auto centering mandrel 6 Turn the rotating base clockwise to lock and center the bottle counterclockwise to release the bottle Please check that the bottle does not tilt due to an incorrect positioning In case release the bottle and center the bottle again Use the regulations 1 and 2 to adjust the correct height of the laser ray 10 20mm over the wine level Cut or remove the cup that protects the neck For an optimal measure please check that the welding line of the bottle is not in the direction of the laser ray Attention in sparkling wines DO NOT remove the cork cage before and during the measure L PRO SRL ALL RIGHTS RESERVED v 1 3 L sensor CO Misura rapida Pressione totale on 2 a 20 C one CO discolta a 20 Pr Ore ultima misura Data ultima misura ipo di misura selezionata Diametro selezionato Temperatura della mis
10. ng the start up procedure When the PARAMETERS SELECTION page appears on the display the instrument L sensor CO is ready to be used L PRO SRL ALL RIGHTS RESERVED v 1 3 L sensor CO Impostazione parametri di Salazionare temperatura indicata dal termometro ES ESCE ID PARAMETERS SELECTION This is the starting page Please select the parameters for the measure you are ready to make Type of measure Fast display the results after one single measure High resolution the instrument executes 3 different measures Between one measure and the following the operator is asked to turn the bottles in its vertical axis for about 60 This procedure allows the instrument to interpolate the optical defects of the bottle giving a better result Internal neck diameter Select the internal diameter of the neck of the bottle in the point where the laser crosses the neck in millimeters To select the correct value press the left side of the bar for decreasing and the right side for increasing the value To determine the internal diameter measure the external diameter of the neck and subtract the double of the glass thickness about 4 5mm depending on the bottle Temperature Select the bottle s temperature at the moment of the measure To select the correct value press the left side of the bar for decreasing and the right side for increasing the value Preferences Not used at the moment Co
11. ura MEASURE FAST MEASURE Once selected FAST MEASURE setted all the parameters required and pressed CONFIRM you will be redirected to the MEASURE page The instrument is ready to measure the bottle Put the bottle in the mandrel and press once the MEASURE botton The instrument shows the following data Total pressure bar Indicates the total pressure inside the bottle in absolute bar atmosphere pressure 1 CO pressure at 20 C bar Indicates the absolute pressure relative only to the CO present inside the bottle referred to 20 C CO in wine at 20 C g l Indicates the CO in the wine Please refer to page 10 for more detail on the coefficient used Last measure time Indicates the time when the last measure displayed has been made Last measure date Indicates the date dd mm yy when the last measure displayed has been made Measure mode Shows the measure mode selected previously in the parameters page fast or high resolution Diameter selected Shows the neck diameter selected previously in the parameters page Temperature of the measure Shows the temperature selected previously in the parameters page L PRO SRL ALL RIGHTS RESERVED v 1 3 L sensor CO Misura alta risoluzione PRIMA SECONDA TERZA MISURA MISURA MISURA 0 0 Bar 0 0 Bar 0 0 gil cofncfanne Tipo di misura selezionata Alta risoluzione Diametro selezionato Temperatura della misura MEASURE HIGH RESOLUTION ME
12. w tension Basse tension 98 37 EC Direttiva macchine Machinery directive Directive machine 89 336 EEC elettromagnetica Electromagnetic Comp El ctromagnetique relative al materiale elettrico destinato ad essere utilizzato entro il campo di tensione fra 50 e 1000 Volt in corrente alternata Concerning electric material used between 50 and 1000 Volt in alternative current Relatives au materiel l ctrique destin l utilisation entre les tensions 50 et 1000 Volt en courant alternatif L pro srl Paolo Tondello kh fh L pro srl Sede legale via Lister 11 35143 Padova PD Uffici P zza Don Rino Brasola 4 Cervarese S Croce PD Tel 049 9916081 Fax 049 9919441 Email info lpro it www lIpro it L PRO SRL ALL RIGHTS RESERVED v 1 8 17 L sensor CO REFERENCES L pro srl c o Start Cube Via della Croce Rossa 112 35129 Padova PD Italy web www lpro it email info lpro it tel 39 049 8697519 fax 39 049 8697511 L PRO SRL ALL RIGHTS RESERVED v 1 3 18