Contents

1. Average deviation Standard Deviation Variance Variance Some of the parameters are specified in scientific exponential notation e g Variance 1 88e 004 18800 Standard deviation Version RKO9052000 page 79 10 4Softwarefunctions for documentation of data 10 4 1 Creating an Annotation Sheet Function Use the Annotation button to open a window in which you can prepare recorded images for presentation purposes If you have opened the annotation sheet the Snap Line Rectangle and Text buttons are also enabled Use these buttons to copy the image loaded in the Viewer window into the annotation sheet to highlight specific areas of the copied image with lines and rectangles and to enter an image comment into a text field IY LL see Copying an Image IY LL see Drawing a Line IY L see Drawing a Rectangle N Q see Creating a Text Field If you hold the mouse pointer over the annotation sheet and press the right mouse button a context menu containing the following commands appears Command Function Line Use this to insert a line of default size onto the annotation sheet Rectangle Use this to insert a rectangle of default size onto the annotation sheet Text Use this to insert a text field with a default font onto the annotation sheet Zoom Use this to select from four magnifications for viewing the annotation sheet Grid This di
2. as well as a series of planes A single vertical section or xz scan allows you to see your sample as though from the side The advantage of having a stack of serial optical sections through the specimen in digital form is that either a composite projection image can be computed or a volume rendered 3 D representation of the specimen can be generated on a graphics computer 4 1 2 Optical Resolution power The term resolution refers to the ability to distinguish fine details in a structure An ideal microscope would have optics totally free from aberrations of any kind In such a hypothetical instrument the resolving power would only be limited by diffraction This limit can be expressed as the minimum distance between two points in the specimen for which they still appear as two points Raleigh s criterion Beyond this limit the points will merge and cannot be resolved as two different points This distance can be calculated from the size of the diffraction image formed by an infinitely small point in the specimen It is equal to the radius of the first minimum in this diffraction image This in turn is related to the numerical apertures of the objective and condenser The numerical aperture is defined by the refractive index of the lens and the width of the cone of light that can pass through it In analogy with the reasoning above the axial resolution can be defined as the radius of the first minimum along the microscope Version R
3. Some of the parameters are specified in scientific exponential notation e g Pixel 3 28e 005 327680 corresponds to a scan format of 640 x 512 Using the maximum and minimum values of intensity the gain values and offset values can be optimally configured for the detectors 10 3 2 Measuring a Profile in a Single Image Function Use the Profile function to measure a specific dimension over a line segment and display it graphically The profile is measured from the data record selected in the Viewer window The data record can be a single image or an image series from raw data or a processed image Depending on the dimension represented in the image either a measured curve of intensity values I or height values um is calculated When you click the Profile button the line segment to be measured appears in the Viewer window as a white line in the image You can change the length and position of the line segment by clicking the line and while clicking and holding the grab point with the left mouse button dragging it to the desired position The Profile viewer window opens too which displays a profile curve for each detection channel and subsequent statistical calculations page 78 Parameter Meaning Formula The length of the measured section Length Mean Amplitude The arithmetical mean Pixel Pixel Maximum i e e Max I Minimum Minimum value Amplitude Average Deviation
4. The user must operate the TCS SP2 in accordance with the instructions in order to ensure safety Several important points are listed below Do not change objectives during scanning operations How to change objectives e Ascertain that the Scan mode is switched off The Scan button must visibly released Be sure that there is no laser light in the focal plane Close the safety shutter at the tube e Turn the lens turret so that the lens to be exchanged appears outside of the optical axis and points outwards Do not apply strong moments to the turret as this may reduce the resolution of the turret scan permanently e Screw in the new lens and turn the lens turret again into the optical axis All unused positions on the objective nosepiece must be capped Version RKO09052000 page 12 Ultraviolet Do not insert reflective objects into the beam path during scanning operations Do not change specimen during scanning operations How to exchange a specimen e Ascertain that the Scan mode is switched off The Scan button must visibly released Be sure that there is no laser light in the focal plane Close the safety shutter at the tube e Exchange the specimen 5 10 Laser Safety The word LASER is an acronym for Light Amplification by Stimulated Emission of Radiation The first laser was demonstrated in 1960 and used a ruby as the lasing medium Lasers have been used in many applications from surgery to bar code readers
5. True Confocal Scanner Leica TCS SP H User manual 1 Table of Contents 10 11 Version RKO09052000 page 2 Version RK09052000 2 Legal notes Version 1 4 May 2000 Made in Germany Copyright 1999 2000 Leica Microsystems Heidelberg GmbH All rights reserved No part of this publication may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying recording or storing in a retrieval system or translating into any language in any form without the express written permission of Leica Microsystems Heidelberg GmbH or its authorized agent DISCLOSURE This document contains Leica Microsystems Heidelberg GmbH proprietary data and is provided solely to its customers for their express benefit of safe efficient operation and maintenance of the product described herein Use or disclosure of Leica Microsystems Heidelberg GmbH proprietary data for the purpose of manufacture or reproduction of the item described herein or any similar item is prohibited and delivery of this document shall not constitute any license or implied authorization to do so REVISIONS Leica Microsystems Heidelberg GmbH reserves the right to revise this document and or improve products described herein at any time without notice or obligation Information and specifications in this manual are subject to change without notice WARRANTY Leica Microsystems Heidelberg GmbH provides this public
6. 1000 2000 Image lines per second Version RKO09052000 LL see Selecting a Scan Speed page 48 In order to accurately align the pixels of the forward sweep and flyback the phase between the forward sweep and the flyback can be adjusted To do so use the Phase button or the corresponding rotary knob of the control panel O see Setting the Phase LL see Controlling Functions from the Control Panel Note lf the speed is set to 800 or 1000 the maximum scan field can no longer be scanned because of technical device related reasons The system switches automatically to zoom factor 2 and zoom factor 4 respectively 10 1 20 Adjusting the phase Function Recording images in bidirectional scan mode can result in a phase shift between the forward sweep and flyback of the scanning beam Click the Phase button to open a dialog window which you can use to correct the shift Use the mouse to move the slide on the scale until the pixel shift in the image disappears You can also adjust the phase using the corresponding rotary knob of the control panel see Controlling Functions from the Control Panel Additional Information For each zoom factor a phase adjustment has already been carried out by the factory and the corresponding values have also been set Based on the temperature response of mechanical and electronical devices a slight deviation in the adjusted standard values can result during operation w
7. Channel 1 2 3 4 5 separately in the case of image series the first image in each is applied All individual images in an image series are displayed for all selected detection channels Tiled Gallery Channel 1 2 3 4 5 All selected detection channels are displayed both separately and in an overlay image together In the case of image series the first image in each is applied Tiled Overlay Channel 1 2 3 4 5 All individual images of an image series are displayed both separately for all selected detection channels and in overlay images together Tiled Gallery Overlay Channel 1 2 31475 Note The Tiled and Single buttons cannot be enabled simultaneously because they carry out opposing functions 10 2 13 Viewing a Multiple Image Function If you click the Tiled button the detection channels are displayed separately in the Viewer window The following display modes are available depending on the other options you enable simultaneously Button Combination Display All selected detection channels are displayed Tiled Channel 1 2 3 4 5 separately in the case of image series the first image in each is applied All individual images in an image series are Tiled Gallery Channel 1 2 3 4 5 displayed for all selected detection channels All selected detection channels are displayed both separately and in an overlay image together In the case of image series the first image in each is ap
8. Note Version RKO09052000 7 4 Opening records 7 4 1 Readable file formats The following file formats can be opened and viewed with the Leica Confocal Software Experiments lei This is a Leica specific binary file format Files of this format contain data for complete experiments TIFF files tif These are Leica image files in single and multiple tagged image file format TIFF Both image files in the previously used TCS formats and external files in RGB TIF format can be read Annotation files ano This is also a Leica specific binary file format In these files annotation sheets are stored The elements on the annotation sheets such as images text and graphics are each available as individual objects When a file of this type is opened both the image data and the experiment settings are loaded 7 4 2 Automatic reuse of instrument parameters With the Leica Confocal Software the hardware settings saved with experiments or single images can be carried over to a new experiment This allows subsequent experiments to be carried out with the same settings To carry over previous settings open the viewer window for the record whose settings you want to reuse Then click Apply in the default configuration profile on the toolbar If you cannot find the Apply button in any of the open windows you can assign the button to any window with Tools Customize In the dialog that opens select the File ca
9. lei or annotation sheet ano while giving it a specific name and file type When saving an experiment a folder is created at the file level with the name of the experiment This folder then contains the description file lei for the experiment as well as the individual image files The format of the description file is Leica specific and binary This file contains the parameter settings and the color information in the form of color look up tables for each image belonging to the experiment page 85 The image files of an experiment can be saved in tif or raw format The standard format tif also contains the experiment settings and the color information for the images In the raw format just the image data is saved Typical Applications The advantage of the raw format is the smaller file size which however is only significant for image recordings with relatively small amounts of data For example if you record an image series with a large number of individual images but at a low scan format you can save and open it faster in raw format than in tif 10 5 4 Saving All Files Function Use the Save All button to save several experiments lei and annotation sheets ano one after the other while giving them a specific name and file type When saving an experiment a folder is created at the file level with the name of the experiment This folder then contains the description file lei for the experiment as we
10. only l The viewer window consists of the image window and the color bars of the Viewer LUT color look up tables Viewer The viewer window consists of the button pads the image window the Standard color bars of the color look up tables and the Experiment legend In the Personal Templates list field user defined templates are registered Use the buttons Add Active Viewer and Remove Template to save a new template or to delete an existing one Click Options in the Tools menu and then select the Workspace register to define whether and if how many Viewer windows are to be opened at start up of the Leica Confocal Software Version RK09052000 page 89 Aberration chromatic Aberration spherical Achromat achromatic objective Airy disc Aliasing AOTF Acousto Optical Tunable Filter Aperture numerical Apochromat apochromatic objective Bleaching optical Version RKO09052000 11 Glossary An optical image distortion conditional on the varying refraction of light rays of different wavelengths on a lens Thus light rays of shorter wavelengths have longer focal distances than light rays of longer wavelengths An optical image distortion conditional on the varying distance of paraxial light rays of the same wavelength from the optic axis Light rays that travel through outer lens zones have shorter focal distances than rays that travel through the lens center optic axis Describes a correction c
11. page 76 L see The Electronic Zoom LL see The Viewer Options Dialog Window Display Icon 10 3 Software functions for measurements and analysis 10 3 1 Calculating a Histogram Function Use the Histogram function to graphically display the frequency at which a specific dimension occurs The histogram is calculated from the data record selected in the Viewer window The data record can be a single image or an image series from raw data or a processed image Depending on the dimension represented in the image either the distribution of intensity values I or height values um is calculated in the histogram Clicking the Histogram button opens the Histogram viewer window which displays a histogram curve for each detection channel and subsequent statistical calculations Parameter Meaning Formula Pixel Mean 7 Maximum Maximum value Max I Minimum Minimum value Variance Variance Average Deviation The total number of pixels that are included in the calculation of the histogram This corresponds to the configured scan format u T The arithmetical mean value SI Pixel Pixel Average deviation Version RKO09052000 page Standard Deviation Mean Energy Root Mean Square RMS Skewness Note Version RKO09052000 Average image energy Root mean square value Skewness of the distribution Standard deviation
12. New TCS menu TCS_Menu The TCS menu contains the buttons for the individual device functions It is arranged in individual working steps The number of working steps varies depending on the installed software The default set of working steps consists of data acquisition Acquire image viewing View surface reconstruction 3 D the measuring functions Quantify the image processing and analysis functions Process and the documentation functions Annotate If the TCS menu is not available in the current configuration profile you can activate it with View Menu gt TCS Menu The toolbars Individual button arrangements can be set up for each individual user A key feature of the toolbars is that they can be toggled on and off complete with all buttons This is done with the menu option View Menu Container Viewer window for documents Experiment Overview In this window existing experiments and their content are listed in the form of a directory tree The viewer window can be opened with View Experiment Overview Status bar The status bar appears at the lower edge of the Leica Confocal software user interface In it the following information is displayed e The progress when loading image data progress bar e The software version number e The name of the machine configuration system type For details about individual functions see chapter Software reference Version RKO09052000 page 18
13. Scan Overview Dialog Window Version RKO09052000 page 35 Clicking the z y Position button opens a dialog window that provides you with two ways of specifying a position value gt Use the mouse to move the slide on the scale that is displayed on the left in the dialog window The corresponding position value is shown in the middle of the dialog window Double click the boldface position value located in the middle of the dialog window This opens a second dialog window where you can enter an exact value You can also set the z y position using the corresponding rotary knob of the control panel see Controlling Functions from the Control Panel 10 1 8 Configuring a Time Series Function Click the Time button to open the Time Configuration dialog window for setting up a time series recording The parameters you can set in this dialog window depend on the selected scan mode You can record a line xt a horizontal section xyt a vertical section xzt or a stack of horizontal sections xyzt many successive times interrupted by a certain time interval Note The Time button is not enabled unless you have selected a scan mode with time dimension using the Mode button H see Selecting a Scan Mode With a time series using the xt scan mode the following parameters are displayed in the dialog window Z Recording time for a line E rar be adjusted by the user Lines Number of Number of recordings of the line of the lin
14. a directory tree Every newly created experiment appears in this viewer window with a file name consisting of Experiment followed by a sequential number and the ei extension You can change file names and file extensions using the Save and Save as buttons LUI see Saving a File Q see Saving a File as Individual images that you have recorded using the Single Scan function are entered into the directory tree with the name Image followed by a sequential number Image series that you have recorded using the Series Scan function are given names consisting of Series and a sequential number If you mark the name of an individual image or an image series and then press the right mouse button a context menu containing the following commands appears Command Function Activate This displays the selected file in the Viewer window New Window This opens a second viewer window for the currently loaded image Delete This deletes the individual image or image series Properties This displays file properties and certain scan parameters in a dialog window Version RK09052000 page 87 The Experiment Overview viewer window automatically appears at the left edge of the user interface You can change however the size and position of the viewer window as desired gt To change the width of the viewer window use the mouse pointer to drag the edge to the position you desire gt To change the position of th
15. application 10 1 14 Defining the End Point fora Spatial Series Function Use the End button to define the end point of a spatial image series First set the exact Z position or y position using the z y Position button or on the control panel using the corresponding rotary knob Then click End This saves the position value for the end point The begin point is set in the same manner LL see Setting the z y Position LL see Controlling Functions from the Control Panel Version RKO09052000 page 42 Note Note You can also define the begin and end points for a spatial series in the Series Scan Overview dialog window Open this dialog window by pressing the Series button the small button not the button that starts the Series Scan function LL see The Series Scan Overview Dialog Window 10 1 15 Defining the End Point fora Wavelength Series Function For a wavelength series this function records a stack of individual images each of which are detected at a specific wavelength from a single optical level Use the Lambda Scan End button to define the wavelength at which recording should end The Lambda Scan End button is not enabled unless you have selected a scan mode with wavelength dimension using the Mode button LL see Selecting a Scan Mode Press Beam to open the Beam Path Setting dialog window gt Activate a detector by clicking its corresponding check box A cast shadow now links the ac
16. at supermarket checkouts from laser pointers to CD players Lasers can produce a very intense and very narrow collimated beam of electromagnetic radiation in the frequency range of 150 nm to 1 mm This radiation is generally in the form of intense visible light Because laser light is not an ionizing type of radiation interaction with the body is generally at the surface The eye and the skin are critical organs for laser radiation exposure and the resultant effects vary depending on the type of laser wavelength of the radiation and beam energy output Laser energy of the proper wavelength and energy may be focused by the lens of the eye onto the retina causing severe damage If laser radiation is of high enough energy skin burns may also result if extremities or other body parts are placed in the laser beam The following table summarizes the various regions of the electromagnetic spectrum produced by lasers and the organs of concern if exposure occurs Spectrum Region Wavelength Range Organ Effected UV C 100 nm to 280 nm All absorbed in cornea and conjunctiva UV B 280 nm to 320 nm Almost all absorbed in se UV A 320 nm to 380 nm Cataract formation Visible 380 nm to 760 nm Retina IR A 760 nm to 1400 nm Retina lens skin Infrared IR B 1400 nm to 3000 nm Cornea and skin IR C 3000 nm to 10 nm Eos and skin Version RKO09052000 Lasers are often identified by type i e Argon Krypton wavelength s and Laser Hazard Classification Th
17. in the Start menu There are many ways to start a program The simplest is here e Click the Start button e Click Programs e Click on the group that contains the program you want to start for instance LCS e Click on the program you want to start for instance Leica LCS Another way to start a program is to open a document that you created in that program The program automatically opens when the document opens Double click on a document file in My Computer or Windows Explorer to open it or click the Start button and select a recently used document from the Documents menu Version RKO09052000 page 23 Caution Caution Version RK09052000 You can also start a program by double clicking on its shortcut icon on the desktop Shortcut icons are links to other files When you use a shortcut Windows simply follows the link back to the original file Whenever you use a document or program frequently you might consider creating a shortcut for it on the desktop To do so just use the right mouse button to drag an object out of Windows Explorer or My Computer On the menu that appears select Create Shortcut s Here Some programs automatically create a shortcut during their installation procedure Windows NT Workstation 4 does not actively track a link between an original and a shortcut For instance if you create a Shortcut of a program and subsequently move rather than copy the original to a different folder then
18. intensity value is displayed in the two dimensional projection image as the representative of all values within the column Three different projection types are available the maximum projection the mean projection and the transparent projection gt The maximum projection uses the maximum intensity value as the representative gt The average projection uses the arithmetical mean of the intensity values as the representative gt The transparent projection uses a weighted mean of the intensity values as the representative The weighting is composed of two factors The first factor is the result of the ratio of each measured intensity value to the maximum possible intensity scaling factor The second factor takes into account the weight of the previously measured intensity value The higher the intensity of the previously measured sampling point the greater the weight is and the less the subsequently measured sampling points will influence the calculation Use the Viewer Options dialog window to select one of the projection types and configure any necessary options gt From the View menu select Viewer Options Version RK09052000 page 71 Note Version RKO09052000 Click the Images icon and then click the Projections register LUI see The Viewer Options Dialog Window Images Icon Note the difference between a maximum projection and a topographical image which is generated from intensity maximums In a maximum pr
19. need a mouse to work most efficiently in Windows NT Here are the mouse actions you need to know Point means to move the mouse pointer onto the specified item by moving the mouse The tip of the mouse pointer must touch the item Click on an item means to move the pointer onto the specified item and press and release the mouse button once Unless specified otherwise i e right click use the left mouse button Clicking usually selects an item Double click on an item means to move the pointer to the specified item and press and release the left mouse button twice quickly Double clicking usually activates an item Drag means to move the mouse pointer onto the item hold down the mouse button and move the mouse while holding down the button Unless specified i e right drag use the left mouse button page 21 8 2 The Windows NT interface The basic interface of Windows NT is called the Desktop which provides a background for the items it contains The initial icons on the Desktop allow the user to view and interact with the system in a logical way The Windows NT screen contains many special elements and controls Here s a brief summary The background on which all the pictures and boxes rest is the desktop The Taskbar shows the windows and programs that are open You can switch between open windows and programs by clicking the name on the Taskbar The Start button opens a menu system from which you can start
20. open a dialog window where you can modify the basic settings for various functions The left side of this dialog window shows the icons corresponding to the functions and the right side shows the related registers When you open the dialog it contains the icons of the functions that you are currently using Click Show all to view all icons In the Render Mode field of the Visualization register you can select whether the 3D view is to be generated in surface wireframe or isolines mode Render Mode Function Surface a The blank spaces between the pixels are filled in with surface Wireframe All Allpixels are linked with lines while the blank spaces remain free are linked with lines while the blank spaces remain free solnes Pixels that correspond to values of the same intensity are surrounded by a curve In the Projection Type field you can select the perspective to be used for displaying the 3D view Projection Type Function Perspective The 3D view is displayed in central perspective Parallel The 3D view is displayed in parallel perspective In the Stretch Height factor field you can vary the scaling factor in the z direction thus stretching or shrinking the height of the 3D view Use the Downsample rate field to reduce the density of data of the 3D view to increase the speed of image processing At a pixel density of 1 1 all measured intensity values are displayed in the image At a pixel
21. page 55 LUI see Rotating the 3D View Use the Translation field to shift a 3D view to the right or left up or down and enlarge or shrink it by modifying the coordinate values Use the dot as the decimal placeholder Translation __Fwrsten a E T C Z Positive values shift the 3D view up negative down Use the Predefined field to change the viewing perspective of the image to Top view or Side view by clicking these buttons respectively L see Moving the 3D View Q see Zooming the 3D View In the Display register click one of the listed commands to show or hide the corresponding graphic element of the 3D view in the Viewer window Sapne Function Elements Show LUT Show the selected color look up table on the z axis Show scale Show the scale Show e Show the bounding box which limits the measurement volume Show axes Show the coordinate axes the coordinate axes Show data during Show the image data while the 3D view is being rotated shifted enlarged 3D motion or shrunken 10 2 3 The Viewer Options Dialog Window Display Icon Function Click on Viewer Options in the View menu to open a dialog window where you can modify the basic settings for various functions The left side of this dialog window shows the icons corresponding to the functions and the right side shows the related registers When you open the dialog it contains the icons Version RKO09052000 page 56 of the func
22. supplied with your instrument meets the requirements of the country where you purchased the instrument If you use the instrument in another country you must use a power cord set that meets the requirements of that country This equipment is designed for connection to a grounded earthed outlet The grounding type plug is an important safety feature To reduce the risk of electrical shock or damage to the instrument do not disable this feature To reduce the risk of fire hazard and electrical shock do not expose the unit to rain or humidity Do not open the cabinet All maintenance is to be performed by an authorized Leica technician Do not allow any liquid to enter the instrument cabinet or come into contact with any electrical components The instrument must be thoroughly dry before you reconnect power or turn the instrument on 5 8 General Safety Notes In accordance with general safety regulations as well as laser device safety regulations comply with the following stipulations Observe the operating instructions Use the product only according to the information given in this documentation and in the online help facility of the Leica Confocal Software Carefully read the operating instructions before you start the system As a preparation for your single work steps always read the relevant description of function in the online help and in particular observe the safety regulations indicated for handling the system You
23. the Viewer window You can assign any of several color look up tables to the detection channel This setting can be changed both in the result image and while the image is being recorded To do so open the page 60 Select Look up Tables dialog window There are two ways of opening the dialog window gt Click the Select Look up Tables button gt Hold the mouse pointer over any position in the Viewer window Click the right mouse button Click on the LUT item in the context menu that appears The color bars for the active detection channels appear in the Viewer window Double click the corresponding color bar see Selecting Color Look Up Tables LUT Note This method of assigning color look up tables only influences the currently displayed image As soon as you start a new scan the color look up tables configured in the Beam Path Setting dialog window are applied again see Setting the Beam Path 10 2 8 Viewing Detection Channel 4 Function Clicking the Channel 4 button displays the image data that have been recorded in detection channel 4 in the Viewer window You can assign any of several color look up tables to the detection channel This setting can be changed both in the result image and while the image is being recorded To do so open the Select Look up Tables dialog window There are two ways of opening the dialog window gt Click the Select Look up Tables button gt Hold the mouse pointer over any position
24. the shortcut may no longer function The Startup folder is special in one respect any programs held within it will start automatically when you start Windows NT Workstation 4 The Documents menu shows the names of the 15 files you created most recently You can open any of these files and its related application at the same time by clicking the file s name in this menu Document files that are opened within an application typically by selecting the File Open command within the application will not be displayed here Only documents opened directly from the Desktop will be displayed here The Settings menu offers three commands for changing your system s settings You can directly access the Control Panel and Printing folders Also accessible is the Task Properties window Being able to access the core system configuration utilities in this way is particularly useful when an application is already in the foreground and you want to make a quick change The Find command features an easy way to locate all system resources Within the Find category you can perform searches for three distinct types of search which are described as follows 8 5 The Taskbar The Taskbar positioned at the bottom of the screen provides a constant view of which applications are running on the system and an easy way to switch between them As you add to the number of concurrently running applications the Taskbar automatically re sizes its iconized
25. value for the diameter gt Use the mouse to move the slide on the scale that is displayed on the left in the dialog window The corresponding value is shown in the middle of the dialog window gt Double click the boldface numerical value located in the middle of the dialog window This opens a second dialog window where you can enter an exact value The diameter of the detection pinhole must be set to the currently used objective at all times If you click the Airy 1 button the detection pinhole is set automatically to the optimal value of 1 Airy unit depending on the objective in use In addition to the numerical aperture of the objective and the wavelength of the light the detection pinhole also determines the thickness of the optical sections Additional Information The diameter of the pinhole is set optimally if it matches the diameter of the Airy disc The Airy disc refers to the inner light circle of the diffraction pattern of a point light source The diameter of the Airy disc in turn is also dependent on several optical parameters and can be described for the Leica Confocal system as follows d 122A Wi 3B Aim To calculate the diameter of the Airy disc you need the excitation wavelength if working with several wavelengths a mean value should be applied the numerical aperture NA and the magnifying factor M of the objective The factor of 3 6 refers to the magnification of other optical components belongi
26. window for selecting a printer and changing the ing aa leaner settings Print This This prints the image or the annotation sheet the image or the annotation sheet ee Background color Use this to open a dialog window for selecting a background color ee Center on Center on page o This centers the image on the page centers the image on the page This option changes the size of the image to fit the size of the printable Fit to page area l The height to width ratio remains the same when the size of the image Aspect ratio changes Version RK09052000 page 83 Use the Image field to change the height Size Y and width Size X of the image and to define a top margin Offset Y and a left margin Offset X The image size and margins cannot be modified if the Fit to page option is enabled If you have enabled Center on Page you can change the size but not the margins Check the results of your settings in the Print Preview field Use the Page field to specify the height and width of the printable area which is not identical to the paper size The two bottom lines of the dialog window display the currently installed printer and configured paper size Note Close the Printer Selection dialog window by clicking OK to save the new settings for the active Viewer window This allows you to Store different printer settings and page layouts for different images However these settings are not permanent and are lost when you exit the L
27. 0 page 75 LL see The Viewer Options Dialog Window 3D Icon 10 2 27 Zooming the 3D View Function Use the Zoom button to magnify and shrink continuously a 3D view while maintaining the aspect ratio This function only adjusts the scale of i e zooms the generated image It cannot be used to improve its resolution Click anywhere inside the image window of the Viewer window and while pressing the left mouse button drag the mouse pointer toward the lower edge To zoom in on the 3D view Click anywhere inside the image window of the Viewer window and while pressing the left mouse button drag the mouse pointer toward the upper edge To zoom out of the 3D view Note While the 3D view is being zoomed the resolution of the image is decreased in order to speed up processing When the left mouse button is released the image is displayed with its original resolution As an alternative to zooming a 3D view manually the Viewer Options dialog window also allows you to change the zoom scale by specifying numerical values gt From the View menu select Viewer Options gt Click the 3D icon gt You can then vary the zoom factor in the Navigation register by changing the y value in the Translation field LL see The Viewer Options Dialog Window 3D Icon Note The Leica Confocal Software provides three different zoom features the 3D zoom the electronic zoom and the graphical Zoom Version RKO09052000
28. Depending on the dimension represented or calculated in the source image either intensity values or height values are mapped to the z axis of the 3D view when the 3D view is generated page 73 Note The spatial representation of intensity in a 3D view leads easily to the assumption that the topography of the specimen is being represented Note however that you can only display the real surface structure of a specimen if you have generated a topographical image first LL see Creating a Topographical Image The 3D view can be generated in three different modes surface wireframe or isolines Set the display mode in the Viewer Options dialog window gt From the View menu select Viewer Options Click the Surface icon and then the Visualization register LL see The Viewer Options Dialog Window Surface Icon Note A separate 3D view is created for each detection channel Therefore it is not possible to generate a single 3D view from one overlay image Overlay button 10 2 25 Rotating the 3D View Function Use the Rotate button to rotate a 3D view in all three spatial directions Note that the 3D view is rotated around a fixed point which is located in the center of the image 4 Hold the mouse pointer over any position in the 3D view 4 Press and hold the left mouse button gt Move the mouse pointer in the direction in which you want to rotate the 3D view Note While the 3D view is being turned the resolution
29. KO09052000 page 6 axis of the diffraction image of a point object According to the theory for such three dimensional diffraction images the radius of the first axial diffraction minimum is approximately twice that of the lateral The axial resolution is thus approximately half of the resolution within the plane of focus The LEICA TCS SP II is a no compromise true point scanning system with high sensitivity and theoretical maximum x y and Z resolution Scan resolution refers to image clarity as determined by the number of pixels and pixel size The smaller the pixel size the more easily two close objects can be distinguished The scan resolution is restricted to the maximum optical resolution power 4 1 3 Detection Confocal imaging or to be more precise the measurement of the optical properties of tiny sub volumes of a specimen is limited not only by the optical quality of the microscope Other limitations are The measurement of continuous specimen only in discrete sub volumes because of sampling and digitalization e The accuracy with which the sub volumes are defined determined by the scanning mechanism e The brightness of the light source in relation to the reflectity of the specimen e The sensitivity and noise produced by the detector The detector is a central component of the confocal microscope too 4 1 4 Image processing In the first confocal microscopes the detector was connected to an oscilloscop
30. S SP II for a while You will discover useful features that you breezed over the first time page 4 Version RK09052000 3 2 Online help Leica TCS SP II comes with a context sensitive online help system that teaches you the different functions of the system To retrieve the contents of the online help system Choose Help Contents The online help system opens with an activated CONTENTS window It shows a collapsible table of contents Double click an entry of the table of contents to show the corresponding information To retrieve information using a keyword Choose Help gt Index The online help system opens with an activated INDEX window Enter a word you would like to look for and the online help system will show the most similar keyword defined To perform a full text search Choose Help Search The online help system opens with an activated SEARCH window Enter a word you would like to look for press the LIST TOPICS button and the online help system will show a ranked list of topics On the right side of the field where you enter the word there is a fly out menu to perform full text search using logical operators 4 Confocal imaging 4 1 1 What is confocal imaging Conceptualized in 1953 the Confocal Laser Scanning Microscopy has only in the past 10 years become a practical technique Today it is the technique of choice for biological research chemical analysis and materials testing An instrument of this
31. Size and is calculated as soon as you change the scan format the electronic zoom or the objective M see The Electronic Zoom LL see Selecting an Objective When the sampling rate of 2 to 3 points is exceeded it is referred to as oversampling In the case of oversampling the structure is fully mapped but no additional information is gained The disadvantage of oversampling is the long scanning time and the danger of destroying bleach sensitive specimens When the sampling rate is less than 2 to 3 points it is referred to as undersampling In this case there is the danger that not all necessary information is sampled The image might illustrate structures that are not actually present in the specimen This effect is called aliasing 10 1 5 Selecting a Scan Mode Function Clicking the Mode button opens a dialog window containing a list of available scan modes The scan mode determines which optical levels of the specimen are to be scanned Fundamentally horizontal xy sections or vertical xz sections can be recorded To generate a three dimensional image of the specimen the optical sections are continued in the corresponding third direction in space thus recording a stack of individual images In addition to this it is possible to add the factor of time or wavelength while recording the images page 33 Mode Function XYZ An image stack is recorded from xy sections in z direction XZy An image stack is recor
32. Viewing a Single Image Function Click the Single button to view only one detection channel or several detection channels in only one image in the Viewer window The following display modes are available depending on the other options you enable simultaneously Button Combination Display Only the selected detection channel is displayed Single Channel 1 2 3 4 5 for an image series the first image of the selected detection channel is displayed All individual images in an image series are Single Gallery Channel 1 2 3 4 5 displayed for the selected detection channel only A single overlay image is generated from all Single Overlay Channel 1 2 3 4 selected detection channels for an image series 5 the first image of each selected detection channel is used Single Gallery Overlay Channel 1 Based on the number of individual images in an image series overlay images are generated from 2 3 4 5 all selected detection channels Note The Single and Tiled buttons cannot be enabled simultaneously because they carry out opposing functions 10 2 12 Viewing a Multiple Image Function If you click the Tiled button the detection channels are displayed separately in the Viewer window The following display modes are available depending on the other options you enable simultaneously Version RKO09052000 page 65 Button Combination Display All selected detection channels are displayed Tiled
33. alog window Scan mode top line to the right of the graphic image gt Scan format red digits at the upper edge of the cube r Number of optical sections division between begin and end points The red number on the vertical edge of the cube represents the the maximum travel of the z actuator that cannot be changed Q see Selecting a Scan Mode Ll see Selecting a Scan Format LL see Defining the Number of Spatial Sections Version RKO09052000 page 40 Note Note Version RK09052000 10 1 12 Defining the Begin Point for a Spatial Series Function Use the Begin button to define the begin point of a spatial image series First set the exact z position or y position using the z y Position button or on the control panel using the corresponding rotary knob Then click Begin This saves the position value for the begin point The end point is set in the same manner LL see Setting the z y Position LL see Controlling Functions from the Control Panel You can also set the begin and end points for a spatial series in the Series Scan Overview dialog window Open this dialog window by pressing the Series button the small button not the button that starts the Series Scan function see The Series Scan Overview Dialog Window 10 1 13 Defining the Begin Point for a Wavelength Series Function For a wavelength series this function records a stack of individual images each of which are de
34. anually into the beam path page 27 Additional Information When selecting the correct objective for a specific application the objective s correction class achromats apochromats fluorite objectives and plan objectives and especially the magnification factor and the numerical aperture are of great significance The numerical aperture determines the resolution capacity of an objective and is deduced from the angle of the light cone accepted by the objective lens and the refractive index of the medium between objective lens and specimen NA n sin Objectives with greater magnification generally have larger numerical apertures but smaller entrance pupils and therefore can record light only from a relatively small scan field Objectives with larger apertures permit higher resolutions but allow less free working distance The following table illustrates this relationship Obiective Resolution Resolution Resolution Resolution Scan Field J xy Air z Water z Oil z Size xy HC PL FLUOTAR 5x 1301 19410 25879 29559 3000 0 15 HC PL FLUOTAR 10x 651 4768 6407 339 1500 0 30 N PLAN 20x 0 40 488 2630 3566 4093 750 N PLAN 50x 0 75 260 649 948 1108 300 PL APO 100x 1 40 139 319 209 236 150 Values in nm at wavelength A 488 nm Values in um Typical Applications Dry objectives are typically used in materials science for the examination of surface structures When imaging layered structures in
35. arting a Series Scan Function Use the Series Scan button to create an image series This creates a multidimensional image data block of the specimen The available dimensions for recording an image series are the three directions in space x y Z aS well as the dimensions of time t and wavelength A This allows you to record a three dimensional spatial image stack consisting of xy sections or xz sections with the additional factor of time or wavelength Note Before recording an image series configure all required scan parameters using the Continuous Scan function to ensure optimal image quality LL see Starting a Continuous Scan Before recording a spatial image series configure the following settings 4 Select the scan mode using the Mode button LL see Selecting a Scan Mode gt Select the scan format using the Format button see Selecting a Scan Format Version RK09052000 page 46 gt Set the desired z position or y position using the z y Position button or by using the corresponding rotary knob of the control panel Q see Setting the z y Position L see Controlling Functions from the Control Panel 4 Set and store the begin point for an image series using the Begin button or in the Series scan Overview dialog window LL see Defining the Begin Point for a Spatial Series gt Set the desired z position or y position using the z y Position button or by using the corresponding rotary knob of the cont
36. ation as is without warranty of any kind either express or implied including but not limited to the implied warranties of merchantability or fitness for a particular purpose All reasonable precautions have been taken in the preparation of this document including both technical and non technical proofing Leica Microsystems Heidelberg GmbH assumes no responsibility for any errors or omissions Leica Microsystems Heidelberg GmbH shall not be responsible for any direct incidental or consequential damages arising from the use of any material contained in this document TRADEMARKS Throughout this manual trademarked names may be used Rather than put a trademark symbol in every occurrence of a trademarked name we state that we are using the names only in an editorial fashion and to the benefit of the trademark owner with no intention of infringement page 3 Version RKO09052000 SAFETY This instrument is designed and manufactured to comply with applicable performance standards for Class 1 laser products as defined by USHHS Class 1 laser products shall not emit hazardous laser radiation during normal customer operation CDRH FDA OSHA and EN standards and regulations known to be effective at the date of manufacture Every possible hazardous situation cannot possibly be anticipated therefore the user must exercise care common sense and observe all appropriate safety precautions applicable to Class Illb lasers and hi
37. bers are added behind the file extension to distinguish between the resulting copies of the current Viewer window 10 2 2 The Viewer Options Dialog Window 3D Icon Function Click on Viewer Options in the View menu to open a dialog window where you can modify the basic settings for various functions The left side of this dialog window shows the icons corresponding to the functions and the right side shows the related registers When you open the dialog it contains the icons of the functions that you are currently using Click Show all to view all icons The Navigation register displays the numerical values of the actions that are carried out using the Rotate Move and Zoom buttons and the mouse pointer Use the Rotation field to tilt a 3D view in all three spatial directions by modifying the angular degrees of the three axes The 3D view is rotated around a fixed point which is located in the center of the image To best understand the rotation function modify the angle of one axis at a time while leaving the other two axes set to 0 Rotation Function g The 3D view is rotated 45 around the fixed point in the direction of the X from 0 to 45 negative z axis r The 3D view is rotated 45 around the fixed point in the direction of the Y from 0 to 45 A l negative x axis The 3D view is rotated 45 around the fixed point in the direction of the Z from 0 to 45 l l negative y axis Version RKO09052000
38. can get an overview of the single functions in the contents file of the online help page 11 Use the product only for purposes described in this documentation and in the online help facility of the Leica Confocal Software The system is designed for making confocal laser scan images and quantitative measurements in the discipline of material science Using the Leica TCS SP2 for any other purposes or applications is inexpert and inadmissible handling The user assumes the entire risk for all experiments with this system and for the consequences resulting from such experiments This is particularly true if the device has been opened or modified by the user Conduct safety inspection and checks on the instrument These inspections are defined by VDE FDA in USA and the laser device safety regulations The user must perform these inspections as described Only service engineers authorized by Leica Microsystems Heidelberg GmbH are allowed to carry out repair work Do not electrically connect the product to devices which are not mentioned in this accompanying documentation Before connecting the product to any such devices consult the local Leica Service Agency or Leica Microsystems Heidelberg GmbH directly 5 9 Special Safety Remarks for Users Leica Microsystems Heidelberg GmbH has done everything possible to maximize user safety and minimize health risks applying special safety measures in accordance with several safety directives
39. ce mode 10 2 20 Viewing a Series Image Function If you click the Gallery button all of the individual images in an image series are displayed in the Viewer window The following display modes are available depending on the other options you enable simultaneously Button Combination Display All individual images in an image series are Gallery Single Channel 1 2 3 4 5 displayed for the selected detection channel only Version RKO09052000 page 70 Based on the number of individual images in an image series overlay images are generated from all selected detection channels Gallery Single Overlay Channel 1 2 3 4 5 All individual images in an image series are Gallery Tiled Channel 1 2 3 4 5 displayed for all selected detection channels All individual images of an image series are displayed both separately for all selected detection channels and in overlay images together Gallery Tiled Overlay Channel 1 2 31 141 10 2 21 Projecting an Image Stack Function Use the Projection function to select or calculate specific intensity data from an image stack i e a series of xy or xZ sections and display them in a two dimensional image This function examines the sampling points voxels that are superimposed along the z axis throughout all optical sections From each of these columns of sampling points the intensity value that is met by the selection criterion or the calculated
40. ch pixel 4 1 6 Integration The Leica TCS SP II was designed as an integrated system Optical and mechanical elements work seamlessly with computer hardware and software The integrated Leica Confocal Software package supports the complete imaging process from optical sectioning through image processing and analysis which is the main application to hard copy output 5 Safety remarks This section introduces you to standard safety precautions warnings and cautions 5 1 General Safety Directions Your safety is extremely important Read and follow all warnings and cautions in this book before handling and operating Leica equipment You can be seriously injured and equipment and data can be damaged if you do not follow the safety warnings and cautions 5 2 Risk of Operation The entire risk for the performance of the device is assumed by the operator or the owner of the product The owner or operator will be fully liable for all consequences which may result if the device is opened by persons other than authorized Leica service staff if it is not serviced or maintained properly or if it is used for purposes other than those described in the accompanying documentation and the online help facility Leica Microsystems Heidelberg GmbH will not be liable for damages resulting from non observance of the above information The above information does not in any way implicitly or explicitly modify the warranty and liability clauses
41. contained in the general terms and conditions of Leica Microsystems Heidelberg GmbH 5 3 Warnings Cautions and Notes The warnings cautions and notes in this manual use the following format page 8 Danger Caution Notes Special Conditions A warning alerts you of an operating procedure practice condition or statement that must be strictly observed to avoid death or serious injury to the persons working on the equipment A caution alerts you to an operating procedure practice condition or statement that must be strictly observed to prevent equipment damage or destruction or corruption or loss of data Notes are statements that either provide extra information about a topic or contain special instructions for handling a particular condition or set of circumstances of the Site of Installation Warning Warning Version RK09052000 This system was tested and complies with the permissible limit values for the emission of electromagnetic radiation EMC However the remaining risk of effects on other devices cannot be ruled out This system must not be mounted next to life saving equipment This system must not be mounted next to life saving equipment 5 4 Laser Class In order to take into account any possible danger resulting from inadmissible usage we have assigned the Leica TCS SP2 to the laser safety class IIIb However improper usage can lead to conditions under which this laser class can no
42. ded from xz sections in y direction xt A line is recorded several successive times xyt An xy section is recorded several successive times xzt An xz section is recorded several successive times xyzt An image stack is recorded from xy sections in z direction several successive times Xy An xy section is recorded at different wavelengths XZA An xz section is recorded at different wavelengths All scan modes with the exception of xt are composed of at least three dimensions The unit will ignore the third and further dimensions if you record using the continuous scan function or the single scan function from one optical level only xy or xz Note The scan modes xzy xzt and xz are only displayed in the list if you have previously selected the Galvo item using the Z Scan button Likewise the scan modes for a wavelength series xy and xzd are only displayed if only one detector is activated in the Beam Path Setting dialog window LL see Selecting the Z Actuator O see Setting the Beam Path 10 1 6 Selecting a Scan Speed Function Clicking the Speed button opens a dialog window that you can use to select from four different scan speeds Speed 200 Image lines per second 400 Image lines per second Version RKO09052000 page 34 800 Image lines per second 1000 Image lines per second The data recording speed can be further increased in combination with the bi
43. density of 1 2 only every second measured intensity value is used in the image Use the Isoline Interval field to define a distance for separating the individual isolines in um This allows you to limit the number of isolines in the 3D view Version RKO09052000 page 58 Note Version RK09052000 Use the Isoline Detail Level field to enter a limiting value specifying that only isolines of a particular length are to be displayed in the 3D view This limits the displayed isolines to those which correspond to an intensity value that has a specific frequency of occurrence LUI see Creating the 3D View 10 2 5 Viewing Detection Channel 1 Function Clicking the Channel 1 button displays the image data that have been recorded in detection channel 1 in the Viewer window You can assign any of several color look up tables to the detection channel This setting can be changed both in the result image and while the image is being recorded To do so open the Select Look up Tables dialog window There are two ways of opening the dialog window Click the Select Look up Tables button Hold the mouse pointer over any position in the Viewer window Click the right mouse button Click on the LUT item in the context menu that appears The color bars for the active detection channels appear in the Viewer window Double click the corresponding color bar see Selecting Color Look Up Tables LUT This method of assigning color look up tab
44. directional scan see Unidirectional or Bidirectional Scan Additional Information The faster the scan speed is the shorter the dwell time of the laser point The scan format i e the number of sampling points on a line must also be taken into consideration The higher the scan format at a constant speed is the shorter the dwell time of the laser point over a sampling point The longer the light point of the laser beam dwells over the individual sampling points of the specimen the more light is detected by the detector So using a lower scan speed results in a better signal to noise ratio The disadvantage of a lower scan speed is that the relatively long impact of the light on the specimen can bleach the specimen photochemically making it unusable This is especially important with fluorescence applications Note lf the speed is set to 800 or 1000 the maximum scan field can no longer be scanned because of technical device related reasons The system switches automatically to zoom factor 2 and zoom factor 4 respectively 10 1 7 Setting the Z Y Position Function Use the z y Position button to define the horizontal level z position or vertical level y position of the specimen in which the images are to be recorded If you want to record an image series using the Series Scan function you use the z y Position button and the Begin and End buttons to define the begin and end points of the image series LU see The Series
45. dium with an immersion objective can lead to resolution loss and impairment of the correction Neutral filters are semi reflective glass plates They are used to distribute the light path independent of wavelength The incident light is partially reflected and partially transmitted Neutral filters are usually placed at a 45 angle in the path of the beam The ratings of a neutral filter are based on its reflectivity to transmissivity ratio A neutral filter RT 30 70 for example reflects 30 of the excitation light and transmits 70 An acronym based on the words picture and element A pixel represents the smallest indivisible picture element in a two dimensional system In this documentation both the sampling points of the specimen as well as the picture points are referred to as pixels Describes a correction class for objectives The image curvature aberration is corrected for objectives of this type Correcting this error requires lenses with stronger concave surfaces and thicker middles Three types of plan objectives planachromate planapochromate and plan fluorite are based on the type of additional correction for chromatic aberration Reflection short pass filters are interference filters that transmit Short wave light while reflecting long wave light An optical reflection short pass filter is characterized by the reading of the wavelength edge at which the filter changes from transmission to reflection 50 threshold R
46. down or the bottom grab point up 4 Double click the area above or below the corresponding grab point gt The Select LUT s dialog window opens so that you can select a second color look up table gt The upper and lower intensity ranges are represented in the colors of the second color look up table Note This method of assigning color look up tables only influences the currently displayed image As soon as you start a new scan the color look up tables configured in the Beam Path Setting dialog window are applied again Version RK09052000 page 63 LU see Setting the Beam Path Typical Applications Generally the selection of a suitable color look up table for a specific use depends on the user s own judgment and needs Experience shows however that certain color look up tables are particularly useful for specific uses Color Look Up Table Application Green This is commonly used for recording specimens that have been marked with FITC Cy2 DTAF or other similar fluorochromes which emit within the green spectral range This is commonly used for recording specimens that have been marked with TRITC Texas Red Cy3 Rodamin or other similar fluorochromes which emit within the red spectral range Blue This is commonly used for recording specimens that have been marked with UV fluorochromes such as DAPI or Hoechst or other similar fluorochromes which emit within the blue spectral range Gray This is c
47. e a Se Number of lines per virtual memory page Pages Number of virtual memory pages Number of virtual memory pages is calculated automatically calculated automatically aks ed Complete recording time i e product of AT and number of recordings Version RKO09052000 page 36 Use the Calculate button to calculate how many virtual memory pages result from a given number of lines per page With a time series using the xyt or xzt scan mode the following parameters are displayed in the dialog window AT Recording time for one xy section or xz section plus pause interval Frames Number of recordings of the xy section or xz section a Complete recording time i e product of AT and number of recordings With a time series using the xyzt scan mode the following parameters are displayed in the dialog window ian rare Z Recording time for one image stack of xy sections plus pause interval Stacks Number of C Number of recordings of the image stack of the image stack abt ded Complete recording time i e product of AT and number of recordings Each parameter can be calculated as a function of the other parameters When entering the values observe the special operating principle of the dialog window Clicking an entry field deactivates it Click the parameter that is to be calculated The corresponding entry field appears grey Now enter the values for the other parameters Click Apply to calculate the para
48. e cannot be processed any more gt Printer eedan This prints the area selected in the image window gt All This prints the entire current image window gt Selection Snapshot This shows or hides the button pad to the left of the image window This shows or hides the button pad located Bottom buttons below the image window This shows or hides the color look up tables of the current image Legend This shows or hides the Experiment legend This enlarges the Viewer window to full Full screen screen Viewer Options This opens the Viewer Options dialog window The color look up tables 5 are displayed as color bars to the right of the image window If you hold the mouse pointer over a color bar grab points appear at the top and bottom of the color bar You can use these grab points to limit the current color look up table to a specific intensity range and to load a second color look up table This allows you to increase the contrast of the image graphically Left buttons LUT gt Drag the top grab point down or the bottom grab point up Double click the area above or below the corresponding grab point gt The Select LUT s dialog window opens so that you can select a second color look up table gt The upper and lower intensity ranges are represented in the colors of the second color look up table see Selecting Color Look Up Tables LUT The Experiment Legend 3 Various i
49. e currently displayed image AS soon as you start a new scan the color look up tables configured in the Beam Path Setting dialog window are applied again LU see Setting the Beam Path page 62 10 2 10 Selecting Color Look Up Tables LUT Function Use the Look up Tables button to open a dialog window which you can use to assign color look up tables to the five detection channels The color look up tables can be configured both in the result image and while the image is being recorded 4 Click the detection channel to which you want to assign a new color look up table in the Select Channel field gt Select the desired color look up table in the Select LUT field 4 Click Apply to check the results in the Viewer window It is also possible to open the Select LUT s dialog window from the Viewer window gt Hold the mouse pointer over any position in the Viewer window Then push the right mouse button Click on the LUT item in the context menu that appears gt The color bars for the active detection channels appear in the Viewer window to the right of the image window Double click one of the color bars Loading a second color look up table If you hold the mouse pointer over a color bar grab points appear at the top and bottom of the color bar You can use these grab points to limit the current color look up table to a specific intensity range and to load a second color look up table gt Drag the top grab point
50. e Laser Hazard Classification is determined by the wavelength s maximum duration of exposure and the average power output of the laser page 13 5 11 Laser Hazard Classification and Laser Safety Control Measures Note Direct exposure of the eye by a laser beam should always be avoided with any laser no matter how low the power Class Exempt lasers that cannot produce a hazard Class Il Low power visible laser which due to the reflex response does not normally present a hazard unless viewed directly Class Illa Lasers that normally do not produce a hazard if viewed momentarily with the unaided eye may produce a hazard if viewed using collecting optics Class Illb Lasers that can produce an eye injury if viewed directly including intrabeam viewing of specular reflections Class IV Laser that can produce an eye injury or skin burns from direct specular or diffuse reflections may be fire and skin hazards Class Il and Illa lasers should have a yellow and black Caution label while Class IIIb and IV lasers should have a red black and white Danger label Embedded systems Class Il Ill or IV lasers or laser systems contained in a protective housing and operated in a lower classification mode may be classified at a lower classification Specific control measures may be required to maintain the lower classification 5 12Laser Hazards The danger from lasers can be divided into the followin
51. e must be kept clean Under no circumstances should users touch the optical components with their fingers or anything which may bear dust or grease Remove dust by using an air puffer not solvent based or a fine dry hair pencil If this method fails use a piece of cloth moistened with distilled water Persistent dirt can be removed from glass surfaces by means of pure alcohol chloroform or naphtha If an objective lens becomes accidentally contaminated with inappropriate immersion oil or mounting medium contact your next Leica representative for advice about cleaning with solvents Take this seriously because some solvents may dissolve the glue which holds the lens in place Caution Do not open objectives for cleaning Oil should be removed from oil immersion lenses after use Once most of the oil has been removed with a clean tissue a piece of Version RKO09052000 page 15 Caution Note Version RK09052000 lens tissue should be placed over the immersion end of the lens A drop of recommended solvent should be applied and the tissue gently drawn across the lens surface This should be repeated as often as is necessary to attain total cleanliness Use a clean piece of lens tissue each time 6 5 Cleaning the microscope surface Use a linen or leather cloth moistened with naphtha or alcohol in order to clean the surfaces of the microscope housing or the scanner varnished parts Never use acetone xylol or nitro
52. e viewer window double click on the double lined edge of the window or click the a icon once This releases the viewer window from the user interface Now you can drag the viewer window to any position by pressing and holding the left mouse button Note If the viewer window has been enlarged to cover the entire width of the user interface you first have to shrink the height of it in order to drag it back to the edge of the user interface 10 6 Softwarefunctions for customization Hier mussen neuen die Topics aus c prj Manuals rein 10 6 1 Saving the Viewer Window as a Template Function Use the Template button to save a user defined layout of the Viewer window as a template The elements of the Viewer window which can be displayed or hidden are the button pads the color bars of the color look up tables and the Experiment legend see The Viewer Window Clicking the Template button opens a dialog window where you enter a name for the template The configuration of the Viewer window is saved and always loaded if you open a file or create a new experiment Additional Information Click Options in the Tools menu and then select the Viewer Template register to get access to default and user defined Version RKO09052000 page 88 templates You can select one of the following default templates in the Leica Templates list field Viewer Template Layout Viewer Pure The viewer window consists of the image window
53. e with long persistence phosphor which would display an image as it was being scanned In the instruments of today the signal is digitized and recorded in a computer This makes it possible to manipulate the image in a multitude of ways Contrast enhancement by thresholds linear contrast stretching and gamma correction curvature of the image intensity value versus source intensity graph Superimposition of images in experiments Digital filtering for edge enhancement smoothing noise Suppression etc Reconstruction of three dimensional views from stacks of images of optical sections This allows for instance an image of an xz plane to be reconstructed from a stack of images of xy planes Complete 3D models of the specimen can also be rendered and examined from any direction Assembly of digital movies from time sequences of microscope images Quantification and Measurements Although these manipulations do not improve the quality of the collected data they serve the purpose of improving the visibility and facilitating the qualitative interpretation of the data Version RKO09052000 page Version RK09052000 4 1 5 Light source Lasers are favorable as light sources for confocal microscopy because they feature high brightness small divergence of the beam are easy to focus and are stable in intensity The stability is important for making quantitative measurements Therefore an online reference correction is made for ea
54. eflection long pass filters are interference filters that reflect Short wave light but are transparent for long wave light An optical reflection long pass filter is characterized by the reading of the wavelength edge at which the filter changes from reflection to transmission 50 threshold The factor by which the light velocity in an optical medium is less than in a vacuum The ratio of signals detected in the specimen and the unwanted signals that are caused randomly by various optic and electronic components which are also recorded by the detector The Stokes shift is a central term in fluorescence microscopy When fluorescent molecules are excited with light of a specific wavelength they radiate light of another longer wavelength This difference between excitation light and fluorescent light is referred to as Stokes shift Without Stokes shift separating the high intensity excitation light from the low intensity fluorescence Signals in a fluorescence microscope would not be possible page 92 Triple dichroic Triple dichroic filters are interference filters at an angle of incidence of light of 45 The transmissivity and reflectivity of triple dichroites depend on three specific wavelengths of light For a triple dichroite TD 488 568 647 for example the excitation light at 488 nm 568 nm and 633 nm is reflected and the excitation light above these values is transmitted The transmission values are generally at 80 and the re
55. eica Confocal Software 10 5Softwarefunctions for data handling 10 5 1 Opening a File Function With the Leica Confocal Software you can open the following file types using the Open button File Type Description A Leica specific binary data format When the experiment is loaded both the image data and the experiment settings are loaded Experiment lei A Leica specific binary data format The elements on the annotation sheet such as images texts and graphic images are each available as individual objects Annotation sheet ano These are Leica image files in single and multi Tiff format External files in RGB Tiff format can be loaded as well Tiff files tif Version RKO09052000 page 84 Note Version RK09052000 LL see Creating an Experiment LL see Creating an Annotation Sheet 10 5 2 Saving a File Function Click the Save button to save the data of the current experiment lei or the current annotation sheet ano If you use the Save button to save an experiment or annotation sheet for the first time the Save As dialog window opens It is easy to mistakenly overwrite your original data using the Save function To avoid doing so use the Save As function to save an experiment which has already been saved once or more under a new name LU see Saving a File As 10 5 3 Saving a File As Function Use the Save As button to save an experiment
56. flection values are between 90 and 95 Voxel An acronym based on the words volume and pixel A voxel represents the smallest indivisible volume element in a three dimensional system In this documentation both the volume elements of the specimen as well as the 3D picture points are referred to as voxels Working distance The distance from the front lens of an objective to the focal point The free working distance is defined as the distance between the front lens of the objective and the cover slip or uncovered sample Usually objectives with large working distances have low numerical apertures while high aperture objectives have small working distances If a high aperture objective with a large working distance is desired the diameter of the objective lens has to be made correspondingly large These however are usually low correction optic systems because maintaining extreme process accuracy through a large lens diameter can only be achieved with great effort Version RKO09052000 page 93
57. for example for training or offline analysis of existing data page 16 Note Version RKO09052000 7 1 2 Starting the software Select Start Programs Leica Confocal Software The initial screen of the Leica Confocal Software appears In this window one of three profiles can be selected Company With this option the Leica Confocal Software is started with the default settings In this mode the configuration and position of toolbars is fixed and cannot be changed Personal In this mode a user specific configuration profile can be selected The user name depends on the account under which a person logs on to the operating system If the user does not yet have a personal configuration profile the default settings are used Last Exit With this option the last used configuration profile is used For advanced users If you have several configuration profiles you can load to these at startup by clicking the button with the three small black dots at the lower right edge of the profile options Here you can also reset your current personal configuration profile to the default settings When you press the Start button the Leica Confocal Software starts with the selected configuration profile lf you do not press the Start button the software will start by itself after a specified time with the selected configuration profile 7 2 The experimental concept of the software With the Leica Confocal Software image data
58. g major categories e Eye hazards such as retinal or cornea burns Danger RETINAL INJURY IS PERMANENT e Skin hazards such as burns e Electrical hazards from high voltage equipment e Fire hazards Version RKO09052000 page 14 6 Maintenance and cleaning Please refer to the corresponding manuals for information on how to maintain the conventional LEICA Research microscope The instructions and additional information relating to the components of the TCS SP II confocal system are summarized below 6 1 Selecting an installation site Do not expose the system to draft Therefore do not install your TCS SP Il system next to elevators air conditioners and other inlets and outlets Warning This system must not be mounted next to life saving equipment 6 2 Protect the microscope against dust and grease When not in use the system should be covered with a plastic foil part of delivery or a piece of cotton cloth The system should be operated in a room which is kept as dust and grease free as possible Dust caps should always be placed over the objective turret positions when no objective is in place 6 3 Be careful when using aggressive chemicals You must be particularly careful if your work involves the usage of acids lyes or other aggressive chemicals Make sure to keep such substances away from optical or mechanical components 6 4 Cleaning the optical system of the microscope The optical system of the microscop
59. gh voltage electrical equipment during installation operation and maintenance Deviation from published operating or maintenance procedures is not recommended Operation and maintenance procedure changes are performed entirely at the user s risk SOFTWARE LICENSE The software described in this document is furnished under a License Agreement which is included with the product This agreement specifies the permitted and prohibited uses of the product 3 Usage of this manual We urge you to use this book not just read it To learn about the Leica Confocal Software use the book as a tutorial We show how to accomplish a variety of tasks Work through the instructions to learn The book usually introduces a set of steps to accomplish a particular task with a distinctive heading 3 1 Doa task Following the heading that introduces a task are either sequential steps or optional steps each with its distinctive style 1 Numbered paragraphs like this one designate step by step instructions Follow them to learn how the tools work 2 Menu commands are shown with an arrow between the single items for example File Save means you should pull down the File menu and choose the Save option Paragraphs with a diamond bullet like this one are optional steps Usually there is a sequence Do one or more of the steps to cause the designated action or actions We recommend that you review this book again after you have used a Leica TC
60. gt Use the mouse to move the slide on the scale The corresponding value is shown in the middle of the dialog window gt Double click the boldface numerical value located in the middle of the dialog window This Note Version RK09052000 opens a second dialog window where you can enter an exact value You can also set the zoom factor using the corresponding rotary knob of the control panel see Controlling Functions from the Control Panel Additional Information While zoom factors from 1 to 36 can be set the useful magnification of the electronic zoom cannot be increased infinitely The limit is achieved with the smallest optically resolvable distance which is determined by the resolution capacity of the objective According to the Nyquist theorem the smallest optically resolvable distance can be mapped without loss of information if it is scanned with about 2 to 3 raster points If this scan frequency is exceeded with a relatively high zoom factor and a selected scan format further magnification is no longer useful because no more optical details can be resolved empty magnification Ll see Selecting a Scan Format LL see Selecting an Objective With bleach sensitive specimens it is advisable to restrict the use of the electronic zoom to short time intervals At high zoom factors a section of the specimen is recorded with a relatively higher scan frequency and thus exposed to a high illumination intensity Th
61. he unit of measurement for the light gathering capacity and the resolution capacity Describes a correction class for objectives The chromatic aberration for three wavelengths is corrected for objectives of this type usually 450nm 550nm and 650nm and the sine condition for at least two colors is met The image curvature aberration is not corrected The destruction of fluorochromes by intense lighting In fil LAL irr AA A A A EA EAA A AAAA A fil mrnAlb rn mnnn m Ar anvataszl rarita lAnAr page 90 light to a high state of energy the singlet state When the excited molecules return to their normal energy state a fluorescent signal is emitted If the intensity of the excitation is too high however the color molecules can change via intercrossing from a singlet state to a triplet state Due to the significantly longer life of triplet states phosphorescence these excited molecules can react with triplet oxide and be lost for further fluorescence excitation Confocality While the optical design of conventional microscopes results in the detection of both focussed and unfocussed image components the confocal principle suppresses the structures outside of the focal plane of the microscope objective To achieve this pinholes are implemented in optically conjugated locations in the optical path They function as point light source excitation pinhole and point detector detection pinhole The diameter of the detection pinhole alo
62. hical image maps the real surface structure of the specimen Using the default settings higher structures appear light and deeper structures dark Use the Viewer Options dialog window to set the selection criterion for the topographical image page 2 Version RKO09052000 From the View menu select Viewer Options Click the Images icon and then click the Topography register LL see The Viewer Options Dialog Window Images Icon By applying a second step you can display the topographical image as a three dimensional graphical image using the 3D View function LUI see Creating the 3D View Typical Applications Displaying the image data in a topographical image is especially informative in material scientific research For the application of quantification functions the topography function is indispensable 10 2 23 Viewing the Original Image Function Click the Original button to reset a projection image or topographical image The Viewer window then displays the recorded raw data of the image again 10 2 24 Creating the 3D View Function Use the 3D View button to display a two dimensional data record three dimensionally When viewing an image series the data record of the series that is currently being viewed in the Viewer window is always the one applied You can display a single xy section or xz section from the raw data or a result image such as a topographical image or projection image in the 3D view
63. hich can be readjusted with this function Version RKO09052000 page 49 10 1 21 Rotating the Scan Field Function The Scan Field Rotation button allows you to rotate the scan field This neither rotates the specimen nor the direction of the scanner but rather the intermediate microscopic image The angle of rotation is steplessly variable between 10 and 100 degrees The original position 0 degrees is based ona predefined reference position in the scan head When the unit is switched on it automatically relocates to this reference position Typical Applications This function is useful to align textures on surfaces or elongated biological structures relative to the observer 10 1 22 Applying the Parameter Setting of an Experiment Function Use the Apply button to use the hardware settings of a previous experiment for recording a new experiment This lets you configure the settings for new image recordings with a single click with the scan parameters that have been optimally configured for a previous application gt Activate the image data block that is configured with the settings you want to apply to the new experiment 4 Click the Apply button Version RKO09052000 page 50 10 1 23 Recording an Image Using the Averaging Method Function The Average button applies an averaging method to the image recording In this case every individual image i e every xy section or xz section is scanned several
64. ialog window where you can configure among other options the Style style and thickness of the line Color This opens a dialog window for selecting a color for the rectangle Bring to front This brings the rectangle to the foreground Version RKO09052000 page 82 Sent to back This sends the rectangle to the background Delete This deletes the rectangle You can change the size of the rectangle by dragging one of the image grab points of the rectangle To move the rectangle without changing its size click in the middle of the rectangle and move it while pressing and holding the left mouse button Note The Rectangle button can be applied only if the annotation sheet has been opened using the Annotation button first 10 4 5 Printing Function You can trigger the print command either by clicking the Print button or from the Printer Selection dialog window To open this dialog window select the Print option in the File menu The image loaded in the Viewer window or an annotation sheet can be printed When you initiate the Print command the image currently loaded in the Viewer window or the currently open annotation sheet is printed using the default printer and default page layout You can change these default settings in the Printer Selection dialog window to suit your needs Use the following buttons provided in the dialog window to do so Button Be Bunetion Printer This opens the dialog
65. in the Viewer window Click the right mouse button Click on the LUT item in the context menu that appears The color bars for the active detection channels appear in the Viewer window Double click the corresponding color bar see Selecting Color Look Up Tables LUT Version RK09052000 page 61 Note This method of assigning color look up tables only influences the currently displayed image AS soon as you start a new scan the color look up tables configured in the Beam Path Setting dialog window are applied again LL see Setting the Beam Path 10 2 9 Viewing Detection Channel 5 Function Clicking the Channel 5 button displays the image data that have been recorded in detection channel 5 in the Viewer window You can assign any of several color look up tables to the detection channel This setting can be changed both in the result image and while the image is being recorded To do so open the Select Look up Tables dialog window There are two ways of opening the dialog window Click the Select Look up Tables button Hold the mouse pointer over any position in the Viewer window Click the right mouse button Click on the LUT item in the context menu that appears The color bars for the active detection channels appear in the Viewer window Note Version RK09052000 Double click the corresponding color bar see Selecting Color Look Up Tables LUT This method of assigning color look up tables only influences th
66. ing files that are being opened are managed as separate experiments 7 f Compiling experiments After you have created a new experiment with File New or File New Template you can assign data to it Images acquired with the continuous scan function are automatically overwritten the next time a scan is started If you want to permanently keep a single image as part of an experiment choose the single scan function An experiment contains data acquired with the single or series scan function If you perform image processing functions on a record you can also save the results as part of the experiment To do this double click the desired image or series in the experiment overview Now carry out the image processing functions e g maximum projection or topology image Mark the area within the Viewer window that you want to add to the experiment With the right mouse button context menu select Send to Experiment The Selection raw option creates a copy of the raw data for the selected object as a new separate component of the experiment The Selection snapshot option creates an RGB image no 3D data photo only of the selected object as a new component of the experiment 8 Starting the Operating System Windows NT You don t have to start Windows NT it starts automatically when you turn on your PC You will first see a splash screen Next you have to log on to your computer As you can see from the instruction
67. is can destroy the specimen photochemically optical bleaching making it unfit page 29 Optimal Zoom Factors The following table shows which zoom factors marked red at a given scan format and scan field size defined by the objective can be used to scan the specimen without loss of data Scan Resolution Objective Scan Format Zoom Factor ian Frequency Scan um nm Frequency me Numerical Resolution at Magnification Numerical Resoiaion at f 0 4 20 0 6 325 1024 x 1024 1 750 132 2 375 366 4 187 183 6 125 122 512 x 512 1 2 4 8 0 200 x 200 1 2 4 8 16 32 40 1 25 156 1024 x 1024 1 2 4 6 512 x 512 1 2 4 8 0 200 x 200 1 2 4 8 16 32 100 1 4 139 1024 x 1024 1 2 75 73 3 50 49 512 x 512 1 150 293 2 15 4 37 72 6 25 49 200 x 200 Note A h 1 2 75 4 37 8 18 12 2 The Leica Confocal Software provides three different zoom features the electronic zoom the 3D zoom and the graphical zoom LLI see Zooming the 3D View Q see The Viewer Options Dialog Window Display Icon Version RKO9052000 page 30 10 1 3 Setting the Detection Pinhole Function Clicking the Pinhole button opens a dialog window that you can use to set the diameter of the detection pinhole In the upper right list field in the dialog window select the units for displaying the diameter You can select from um Airy units and digital values There are two ways of specifying a
68. ize with the priority for keeping the number of sections unchanged Enter the number of desired sections whole numbers only in the Sections field Then click the Calculate button next to the Step Size field Clicking Reset displays the last saved values LL see Starting a Series Scan L see The Series Scan Overview Dialog Q see Selecting the Z Actuator 10 1 17 Setting the Number of Wavelength Steps Function For a wavelength series this function records a stack of individual images each of which are detected at a specific wavelength from a single optical level The images are recorded within a wavelength range which is limited by the begin and end points Use the Lambda Steps button to define the number of recordings that are to take place within this range Note Version RKO09052000 page 45 The Lambda Steps button is not enabled unless you have selected a scan mode with wavelength dimension using the Mode button Q see Selecting a Scan Mode LL see Defining the Begin Point for a Wavelength Series LL see Defining the End Point for a Wavelength Series Typical Applications A wavelength series can be applied to determine the maximum emission of a fluorochrome This is useful because the Stokes shift of the emission curve of a fluorochrome is dependent on each specimen that is applied This then allows you to set the detection range precisely for a specific application 10 1 18 St
69. jective Function Click the Objective button to open a list of objectives and select an appropriate one for recording your images This list shows only the objectives that have been assigned to the up to seven slots on the revolving nosepiece You can assign objectives as follows gt Select the Objective option from the Tools menu A dialog window opens containing an extensive list of objectives and the symbolic representation of the slots on the revolving nosepiece gt Find in the list the objective that you are using and select it with the mouse Click and hold the left mouse button and drag the objective onto the symbol representing the slot in which the objective is installed R The assignment is saved in the software and the objective appears in the selection list which you can open by pressing the Objective button 4 Repeat this procedure for every objective that you have installed in the revolving Note Version RK09052000 nosepiece You can use the Add Remove and Edit buttons in this dialog window to add new objectives remove objectives or edit existing objective entries The revolving nosepiece of DM RXA DM RXE and DM IRBE microscopes is triggered by the software so that the objective is moved automatically into the beam path when you select an objective with the Objective button or dialog window With all other microscope types the objective has not only to be assigned in the software but also be moved m
70. kind represents a fusion product of contributions from many fields microscopy laser technology and optics for coherent light video technology electronics and computer technology Confocal microscopy detects structures by collecting light from a single focal plane of the sample excluding light that is out of focus In a point scanning confocal system the microscope lenses focus the laser light on one point in the specimen at a time the focal point The laser moves rapidly from point to point to produce the scanned image Both fluorescent and reflected light from the sample pass back through the objective The microscope and the optics of the scanner module focus the light emitted from the focal point to a second point called the confocal point The pinhole aperture located at the confocal point allows light from the focal point to pass through the detector Light emitted from outside the focal point is rejected by the aperture The confocal principle is illustrated schematically for the epi illumination imaging mode page 5 As in conventional epifluorescent microscopes one lens is used as both condenser and objective The advantage is eliminating the need for exact matching and co orientation of two lenses A collimated polarized laser beam from an aperture Is reflected by a beam splitter dichroic mirror into the rear of the objective lens and is focused on the specimen The reflected light returning from the specime
71. lass for objectives The chromatic aberration for two wavelengths is corrected for objectives of this type Usually an objective of this type is corrected to a wavelength below 500nm and above 600nm Furthermore the sine condition for one wavelength is met The image curvature aberration is not corrected The Airy disc refers to the inner light circle surrounded by alternating dark and light diffraction rings of the diffraction pattern of a point light source The diffraction discs of two adjacent object points overlap some or completely thus limiting the spatial resolution capacity An image distortion caused by a sampling frequency that is too low in relation to the signal frequency The acousto optical tunable filter is an optic transparent crystal that can be used to infinitely vary the intensity and wavelength of radiated light The crystal generates an internal ultrasonic wave field the wavelength of which can be configured to any value Radiated light is diffracted vertically to the ultrasonic wave field like through a grid The aperture is the sine of the angle under which light enters into the front lens of a microscope objective its symbol is NA The aperture influences both the light gathering capacity and the resolution capacity of an objective Since various media can be present between specimen and objective lens such as the embedding medium for the specimen the numerical aperture NA n sin a is usually applied as t
72. les only influences the currently displayed image AS soon as you start a new scan the color look up tables configured in the Beam Path Setting dialog window are applied again LU see Setting the Beam Path page 59 10 2 6 Viewing Detection Channel 2 Function Clicking the Channel 2 button displays the image data that have been recorded in detection channel 2 in the Viewer window You can assign any of several color look up tables to the detection channel This setting can be changed both in the result image and while the image is being recorded To do so open the Select Look up Tables dialog window There are two ways of opening the dialog window Click the Select Look up Tables button Hold the mouse pointer over any position in the Viewer window Click the right mouse button Click on the LUT item in the context menu that appears The color bars for the active detection channels appear in the Viewer window Note Version RK09052000 Double click the corresponding color bar see Selecting Color Look Up Tables LUT This method of assigning color look up tables only influences the currently displayed image AS soon as you start a new scan the color look up tables configured in the Beam Path Setting dialog window are applied again LUI see Setting the Beam Path 10 2 7 Viewing Detection Channel 3 Function Clicking the Channel 3 button displays the image data that have been recorded in detection channel 3 in
73. ll as the individual image files The format of the description file is Leica specific and binary This file contains the parameter settings and the color information in the form of color look up tables for each image belonging to the experiment The image files of an experiment can be saved in tif or raw format The standard format tif also contains the experiment settings and the color information for the images In the raw format just the image data is saved Typical Applications The advantage of the raw format is the smaller file size which however is only significant for image recordings with relatively small amounts of data For example if you record an image series with a large number of individual images but at a low scan format you can save and open it faster in raw format than in tif Version RKO09052000 page 86 10 5 5 Creating an Experiment Function Click the New Experiment button to open a new Viewer window which creates a new experiment An experiment is a file that consists of one ore more individual images or image series This allows you to keep several images each recorded using different scan parameters or edited versions of images in one experiment The format of the experiment file lei is Leica specific The Experiment Overview Viewer Window To open the Experiment Overview viewer window select the Experiment Overview option from the View menu This viewer window displays the recorded images in
74. longer be guaranteed For this reason we are not able to generally declare our system to belong to Class 1 This instrument is designed and manufactured to comply with applicable performance standards for Class 1 laser products as defined by USHHS Class 1 laser products shall not emit hazardous laser radiation during normal customer operation CDRH FDA and OSHA standards and regulations known to be effective at the date of manufacture page 9 Note Warning Version RKO09052000 Hazardous situation cannot possibly be anticipated Therefore the user must exercise care common sense and observe all appropriate safety precautions applicable to Class Illb lasers and high voltage electrical equipment during installation operation and maintenance Deviation from published operating or maintenance procedures is not recommended Operation and maintenance procedure changes are performed entirely at the user s risk The laser light emitted from the Leica TCS SP2 used in accordance with the instructions is harmless 5 5 Technical Safety Measures In accordance with the laser safety directives EN 60825 European standard and CDRH USA Leica Microsystems Heidelberg GmbH employs various safety measures Warning labels have been affixed near apertures or moveable parts where exposure to laser light is possible Furthermore a laser interlock switch is installed Note that it is a requirement of EN 60825 1 Safety of Laser Products Pa
75. lution of the generated image According to the Nyquist theorem or sampling theorem a structure can only be scanned without information loss if the smallest optically resolvable distance is scanned with about 2 to 3 raster points This optically resolvable distance called lateral resolution depends on the numerical aperture of the objective and the wavelength of the applied excitation light lateral resolution oe The following example should illustrate this relationship You have selected for example the PL APO 100x objective unit with a numerical aperture of NA 1 4 and a wavelength of 488 nm Based on these factors the smallest optically resolvable distance is determined as follows U4 488nm 1 4 Based on this rough formula the distance of the raster points required to avoid information loss while recording is ca i40nm Version RKO09052000 page 32 Note Version RK09052000 am ca 4 nm If however you select a scan format of 1024x1024 a PL APO 100x objective which has a maximum scan field size of 150 um would result in the following raster point distance 150 um 1024 To achieve the necessary raster distance of 47 nm you now have the option of increasing the scan format for example to 2048x2048 or decreasing the scan field size using the electronic Zoom zoom 2 ca 146 nm The current raster distance is displayed in the hardware legend In the legend the value is referred to as Voxel
76. m z E ed the image series Sections Sections The number of o The number of configured sections sections Step Size um The distance between two sections Enter any desired value for the number of sections and the step size in this dialog window The value for the height of the image stack cannot be altered since this parameter is determined by the begin and end points of the image series The step size has always to be a multiple of the minimum step size of the z actuator so that certain combinations of values are only possible if either the height of the image stack or the number of sections is changed Depending on which of the both Calculate buttons you click one of the two parameters is kept unchanged Calculating the number of sections with the priority for keeping the height of the image stack unchanged Version RK09052000 page 44 Enter the desired step size in the Step Size field Then click the Calculate button next to the Step Size field Calculating the number of sections with the priority for keeping the number of sections unchanged Enter the desired step size in the Step Size field Then click the Calculate button next to the Sections field Calculating the step size with the priority for keeping the height of the image stack unchanged Enter the number of desired sections whole numbers only in the Sections field Then click the Calculate button next to Sections field Calculating the step s
77. mage parameters of an image recording are registered in the Experiment legend To select the parameters that you want displayed click any area of the legend This opens a list of Version RK09052000 page 53 all available entries Then select the desired entry If you hold the mouse pointer over any area within the Experiment legend and press the right mouse button a context menu containing the following commands opens Command Function This opens the Edit Legend dialog window where you can specify a Experiment Edit title for the legend specify the number of legend entries or Clear all entries Activate This displays the Experiment legend in the Viewer window Remove This deletes the current Experiment legend Add Experiment tab This creates a new Experiment legend in the Viewer window The Hardware Legend The hardware settings for image recording are registered in the Hardware legend To open the legend select the Hardware Legend option from the View menu To select the entries that you want displayed in the legend click the Edit button This opens the Edit Legend Entries dialog window The Available Entries list box shows all available entries Select the entries that you want displayed in the legend Then click Add to carry the selected entries over into the Show entries list box gt The Show entries list box contains the entries that are to be displayed in the legend Use the Remove button to re
78. may get a time and date update from a network server every time they log on If the servers time is incorrect your workstations time will be wrong too Please inform your network manager 8 Getting Help Windows NT includes a powerful help system In addition to Help menus in every window there is a standalone Help feature available from the Start menu To access it click your mouse on the Start button and click on Help There are three tabs in this box Contents Index and Find The Contents tab appears on top first To move to a tab click on it Contents The Contents tab displays individual Help topics The topics are organized into categories and are represented by small book icons Double click on any book to open it Sub books and documents appear Double click on sub books and documents to open them Index The Index tab displays an index of all available topics Type the word you want to look up The Index list scrolls to that part of the alphabetical listing When you see the topic on the list that you want to read double click on it page 25 Find The Find tab provides a text entry box for you to type the specific word s or phrase you want to search for under Help rather than searching for information by category The text entry box is linked to a list of words in your Help files and any words or phrases that match will be shown You can specify more than one word by separating words with a space If you wish
79. meter Clicking Reset displays the last saved values When you have finished configuring these settings press the series Scan button Version RKO09052000 page 37 Version RK09052000 10 1 9 Starting a Single Scan Function Use the Single Scan button to record a single image from a single optical level in the specimen Before recording an image using the single scan function configure all required scan parameters using the Continuous scan function to ensure optimal image quality LL see Starting a Continuous Scan Typical Applications The single scan function is designed for recording bleach sensitive specimens Use this function in the case of bleach sensitive specimens not only for image recording but also as an alternative to the continuous scan function for setting the scan parameters The single scan function can also be used to check the image section by zooming in on the scan field 10 1 10 Starting a Continuous Scan Function Use the Continuous Scan button to permanently record images from a single optical level in the specimen This does not generate image series since the image being generated always replaces the previously generated image Press the button a second time to stop the continuous scan Additional Information The unit applies the last used scan parameters automatically You can modify some of the parameters while the image is being recorded Others have to be configured before you
80. move entries from being displayed in the legend gt Use the Move up and Move down buttons to shift one or more than one entry up or down in the list gt You can use the Edit grid color and Edit background color buttons to modify the color of the frame and background of the legend The Hardware legend is aligned automatically along the right edge of the user interface You can change however the size and position of the legend as desired gt To change the width of the legend use the mouse pointer to drag the edge to the position you desire gt To change the position of the legend double click on the double lined edge of the legend or click the icon once This releases the legend from the user interface turning it into a window Now you can drag the legend to any position by pressing and holding the left mouse button Version RKO09052000 page 54 Note If the legend has been enlarged to cover the entire width of the user interface you first have to shrink the height of it in order to drag it back to the edge of the user interface Option New Window You can open a second Viewer window containing the same image as the currently open Viewer window by clicking the New Window option in the Window menu This command does not create a new experiment but rather opens a second view of the current image data This feature is useful for viewing different representations of the same image data at the same time Sequential num
81. n passes back through the same lens The light beam is focused into a small pinhole i e the confocal aperture to eliminate all the out of focus light i e all light coming from regions of the specimen above or below the plane of focus The achieved optical section thickness depends on several parameters such as the variable pinhole diameter and the wavelength The in focus information of each specimen point is recorded by a light sensitive detector i e a photo diode positioned behind the confocal aperture and the analog output signal is digitized and fed into a computer The detector is a point detector and only receives light from one point in the specimen Thus the microscope sees only one point of the specimen at a time as opposed to the conventional microscope where an extended field of the specimen is visible at one moment Therefore to obtain an image it is necessary either to move the illuminated point or to move the specimen These two possibilities have given rise to two different types of confocal microscopes 1 The stage scanning type in which the microscope stage with the specimen is moved in a scanning motion while the optics remain stationary 2 The beam scanning or mirror scanning type in which the illuminated point is scanned over the fixed specimen by small fast galvanometer driven mirrors as used by LEICA The LEICA TCS SP II system makes it possible to image a single in focus plane horizontal or vertical
82. ng position values are displayed to the right of the cube Set the begin and end point as follows gt Use the mouse pointer to drag the yellow square to the level in which the image series is to begin Or click the z y Position button to open a dialog window where you can enter the values of the z position and y position Version RK09052000 page 39 gt Click in the white box for the begin point The corresponding position value appears and is saved Use the mouse pointer to drag the yellow square to the level in which the image series is to end Or click the z y Position button to open a dialog window where you can enter the values of the z position and y position 4 Click in the white box for the end point The corresponding position value appears and is saved 4 The entire height of the image stack between the begin and end points is calculated and displayed Total Now click the Series Scan button The dialog window stays open and you can follow the process of the image series being recorded Note You can also define the begin and end points using the control panel and the separate Begin and End buttons LL see Defining the Begin Point for a Spatial Series LL see Defining the End Point for a Spatial Series LL see Setting the z y Position Additional Information The following additional parameters which have to be set before recording an image series are specified in the Series Scan Overview di
83. ng to the Leica Confocal microscope If the pinhole is set to the Airy disc light from outside the focal plane is suppressed and the signal to noise ratio will be high These conditions allow the recording of optical sections of minimum thickness The wider the diameter of the pinhole the Version RK09052000 page 31 more light reaches the detector The image becomes brighter but also increasingly blurred because structures outside the focus also appear in the image Increasing the diameter of the pinhole above 1 Airy unit is recommended only for detecting very weak signals Generally when recording material scientific soecimens enough light is reflected into the detectors so that the detection pinhole can be closed completely i e set to the minimum diameter 10 1 4 Selecting a Scan Format Function Clicking the Format button opens a dialog window that contains a selection of various scan formats Selecting the scan format selects the image raster that will be used for recording the images The image raster refers to the number of points in the three spatial directions that are to be scanned from the specimen Besides the numerical aperture of the objective and the excitation wavelength the scan format together with the electronic zoom determines in large part the spatial resolution of the recorded data Additional Information When selecting the scan format observe the influences between the image raster and the reso
84. ng with the wavelength and numerical aperture of the objective being used determines the axial extension of an optical section Curvature of image field The curved surface to which a microscopic image is to be clearly and distinctly mapped is described as image curvature aberration It is conditional on the convex shape of the lens and makes itself apparent as an error due to the short focal distances of microscope objectives Here the object image is not in focus both in the center and at the periphery at the same time Objectives that are corrected for image curvature aberration are called plane objectives plane flat image field Dichroic Dichroic filters are interference filters at an angle of incidence of light of 45 The transmissivity and reflectivity of dichroites depend on a specific wavelength of light For an RSP 510 filter reflection short pass for example the excitation light below 510 nm is reflected and the excitation light above this value is transmitted The transmission values are generally between 80 and 90 and the reflection values between 90 and 95 Double dichroic Double dichroic filters are interference filters at an angle of incidence of light of 45 The transmissivity and reflectivity of double dichroites depend on two specific wavelengths of light For a double dichroite DD 488 568 for example the excitation light at 488 nm and 568 nm is reflected and the excitation light above these val
85. nt the line to begin Keeping the left mouse button depressed drag the mouse pointer across the page to the position where you want page 81 the line to end If you mark the line and then press the right mouse button a context menu containing the following commands appears Command Bunton This opens a dialog window where you can configure the style thickness and length of the line Color This opens a dialog window for selecting a color for the line Bring to front This brings the line to the foreground Send to back This sends the line to the background Delete This deletes the line You can change the length of the line by dragging one of its grab points To move the line without changing its size click the center grab point and move the line while pressing and holding the left mouse button Note The Line button can be applied only if the annotation sheet has been opened using the Annotation button first 10 4 4 Drawing a Rectangle on the Annotation Sheet Function Use the Rectangle button to draw a rectangle on the annotation sheet On the annotation sheet click the position where you want the corner of the rectangle to be Keeping the left mouse button depressed drag the mouse pointer across the page to define the size of the rectangle If you mark the rectangle and then press the right mouse button a context menu containing the following commands appears Command Bunton This opens a d
86. ode Select the scan format using the Format button Version RKO09052000 page 47 LL see Selecting a Scan Format Define the wavelength at which you want the wavelength series to begin LUI see Defining the Begin Point for a Wavelength Series Define the wavelength at which you want the wavelength series to end LL see Defining the End Point for a Wavelength Series Select the desired number of recordings between the begin and end points of the wavelength series LL see Setting the Number of Wavelength Steps 10 1 19 Selecting Unidirectional or Bidirectional Scan Function If you click the Unidirectional Bidirectional Scan button bidirectional scan mode is enabled If this button is not clicked unidirectional scan mode is automatically set In unidirectional scan mode each line is scanned from left to right No data is recorded while the laser beam is being guided to the starting point of the next line In bidirectional scan mode the first line is scanned from left to right and the second from right to left In other words the return run or flyback of the laser beam is implemented for recording data too thus increasing the scan speed when using bidirectional scan mode If bidirectional scan mode is active you can double the scan speed selected with the Speed button Unidirectional Bidirectional 200 aoe Image lines per second 400 800 Image lines per second 800 1600 Image lines per second
87. of the image is Version RK09052000 page 74 decreased in order to speed up processing When the left mouse button is released the image is displayed with its original resolution As an alternative to rotating a 3D view manually the Viewer Options dialog window also provides the option of specifying the angles of rotation for the three spatial axes From the View menu select Viewer Options Click the 3D icon and then click the Navigation register You can then specify the angles of rotation for all three axes in the Rotation field LL see The Viewer Options Dialog Window 3D Icon 10 2 26 Moving the 3D View Function Use the Move button to shift a 3D view Hold the mouse pointer over any position in the 3D view Press and hold the left mouse button 4 Move the mouse pointer in the direction in which you want to move the 3D view Note While the 3D view is being moved the resolution of the image is decreased in order to speed up processing When the left mouse button is released the image is displayed with its original resolution As an alternative to moving a 3D view manually the Viewer Options dialog window also provides the option of specifying the coordinate values for the position of the image From the View menu select Viewer Options Click the 3D icon and then click the Navigation register gt Modify the coordinate values for the x axis and z axis in the Translation field Version RK0905200
88. ojection the intensity maximums are assigned color values directly In a topographical image based on intensity maximums the intensity maximums are assigned real z positions of the individual sampling points first and then are color coded 10 2 22 Creating a Topographical Image Function Use the Topography function to select specific intensity data from an image stack i e a series of xy or xz sections and transfer them into a two dimensional topographical image This function examines the sampling points voxels that are superimposed along the z axis throughout all optical sections From each of these columns of sampling points only the intensity value that is met by the selection criterion is displayed in the two dimensional topographical image as the representative of all values within the column You can make your selection based on either the maximum intensity or the center of mass of the measured intensities If you want to represent the intensity maximums in the topographical image only the sampling point at which the maximum intensity was measured is selected When determining the center of mass a mean value is calculated from all superimposed sampling points the center of mass of the area that is limited by the curve of the measured intensity values The intensity maximum or center of mass is then assigned the real Z position of the corresponding sampling point and then color coded Based on this assignment a topograp
89. ommonly used for displaying transmission recordings P Color 1 2 3 4 5 6 This is commonly used for recording specimens that have been marked with pH sensitive or ion sensitive fluorochromes and for displaying time series Geo Land Geo Sea Geo Land amp Sea R amp B Glow Glow Over Glow Under Glow Over amp Under These are recommended for generating topographical images i e for mapping surface structures Geo Land amp Sea is especially useful for visualizing depths and graduations These are recommended for optimizing image contrast the offset and gain of the detectors Glow Over highlights in blue intensities at the upper end of the table while Glow Under represents in green intensities at the lower end of the table Glow Over and Under is a combination of these two color tables Cyan Magenta Yellow These are recommended for printing images The CMY K colors are generally used for the printer system color output RGB colors such as the other color look up table used here are applied generally for color representation on monitors Because these two color systems vary greatly the colors in the CMY K representation can differ strongly from those used in the monitor representation All color look up tables are also available with inverse color flow meaning high intensities are represented as dark and low intensities light Version RK09052000 page 64 10 2 11
90. or the results of image processing steps can be combined in groups These groups are referred to as experiments and are stored ina specific file format lei In these experiments original experimental image data can be stored together with image viewing data For further details see chapter Organizing data by grouping experiments 7 3 Basic structure of the user interface The visual appearance of the Graphical User Interface GUI depends to a large extent on the chosen configuration profile It does however contain a number of standard elements The following standard elements are common to all GUI configurations The menu bar page 17 The menu bar contains the File View Macro Tools Window and Help menus These menus contain commands and information for viewing data making changes to settings and customizing the program for specific users It does not contain functions for the direct control of scan functions These are contained in the TCS menu View gt Menu gt TCS Menu The menu bar itself cannot be configured The Viewer window TCS_Viewer This window displays image data experimental conditions and user information The image window can be configured see chapter Modifying the user interface and defining user specific settings on page 16 Beside confocal image records the image window also shows experimental data such as system settings An image window for a new experiment can be opened with File
91. plied Tiled Overlay Channel 1 2 3 4 5 Version RKO09052000 page 66 All individual images of an image series are displayed both separately for all selected detection channels and in overlay images together Tiled Gallery Overlay Channel 1 2 3 4 5 Note The Tiled and Single buttons cannot be enabled simultaneously because they carry out opposing functions 10 2 14 Viewing an Overlay Image Function Click the Overlay button to display all selected detection channels together in an overlay image in the Viewer window The following display modes are available depending on the other options you enable simultaneously Button Combination Display A single overlay image is generated from all Single Overlay Channel 1 2 3 4 selected detection channels for an image series 5 the first image of each selected detection channel is used Based on the number of individual images in an image series overlay images are generated from all selected detection channels Gallery Single Overlay Channel 1 2 3 4 5 All selected detection channels are shown Tiled Overlay Channel 1 2 3 4 5 separately as well as together in an overlay image In an image series the first image is used All individual images of an image series are displayed both separately for all selected detection channels and in overlay images together Tiled Gallery Overlay Channel 1 2 3 4 5 Use the Viewe
92. programs Click on the Start button then click on your selection from each menu that appears Some icons appear on your desktop You can activate one by double clicking on it We now take a brief tour of the items you see on the screen A standard desktop item is the My Computer icon Double clicking this icon opens the My Computer window The My Computer window gives you easy access to the major components like hard and floppy disk drives of your computer system or workstation For example by double clicking the Hard disk C icon you can see the contents of your PC s hard drive This allows the user to view local resources as objects Access to the Windows NT Workstation 4 control panel and print support control are also accessible from My Computer If at installation time you installed one of the additional local applications such as Dial Up Networking this will also appear within My Computer You can use the Control Panel icon in the My Computer window to view and change any system component The Control Panel contains numerous icons that allow you to control your system The particular icons that you see on your PC may be slightly different from those illustrated due to that fact that you may have different hardware installed and may or may not be connected to a network or modem You may also have different Windows NT Workstation 4 options installed Double clicking the Network Neighborhood icon displa
93. r Options dialog window to select from three methods of color mixing for generating an overlay image gt From the View menu select Viewer Options 4 Click the Images icon and then click the Overlay register LL see The Viewer Options dialog window Images Icon Version RKO09052000 page 67 n Note gt Note Version RK09052000 10 2 15 Viewing the First Image of a Series Function You can view the individual images of an image series as a film sequence Click First to jump to the first image in the series lf the Gallery button is enabled the First and Last Next and Previous and Play Stop buttons are disabled and displayed in gray The Gallery button is used to show all of the individual images of an image series and does not allow film sequence mode 10 2 16 Viewing the Next Image in a Series Function You can view the individual images of an image series as a film sequence Click Next to jump to the next image in the series If the Gallery button is enabled the Next and Previous First and Last and Play Stop buttons are disabled and displayed in gray The Gallery button is used to show all of the individual images of an image series and does not allow film sequence mode page 68 10 2 17 Viewing the Previous Image in a Series Function You can view the individual images of an image series as a film sequence Click Previous to jump to the previous image in the series Note If the Gallery b
94. rol panel LL see Setting the z y Position Ll see Controlling Functions from the Control Panel gt Set and store the end point for an image series using the End button or in the Series Scan Overview dialog window LL see Defining the End Point for a Spatial Series gt Select the number of spatial sections using the Sections button LL see Defining the Number of Spatial Sections When you have finished configuring these settings press the Series Scan button You can track the recording process of the image stack in the series Scan Overview dialog window Before recording a time series configure the following settings P Select a scan mode with time dimension using the Mode button LUI see Selecting a Scan Mode gt Select the scan format using the Format button H see Selecting a Scan Format Set the desired z position or y position using the z y Position button or by using the corresponding rotary knob of the control panel LL see Setting the z y Position LL see Controlling Functions from the Control Panel gt Set the number of recordings the pause interval between scans and the complete recording time LL see Configuring a Time Series When you have finished configuring these settings press the Series Scan button Before recording a wavelength series configure the following settings 4 select a scan mode with wavelength dimension using the Mode button LL see Selecting a Scan M
95. rt 1 Equipment classification requirements and user s guide that for installations where class IIlb devices are used a Laser Safety Officer or Laser Protection Advisor should be appointed It is the laser safety officers responsibility to review and designate appropriate controls for the use of the equipment 5 6 WARNINGS Do not look directly into or at a reflection of the laser beam while the laser is scanning Long term exposure to the laser beam can damage your vision For the purposes of laser safety a direct laser beam which has been deflected from a mirror or polished surface is considered to be as intense as the direct beam The user is responsible for the safe operation and maintenance of this instrument at all times Class 1 laser product and electrical safety compliance is assured only when all safety devices interlocks and laser containment systems are in serviceable condition and operating Disabled or damaged safety devices or systems will expose personnel to lethal high voltages and Class I lb laser radiation of sufficient power to cause severe eye injury burns and property damage page 10 Danger Caution Version RK09052000 Tampering with or deactivation of any safety system and or interlock terminates Class 1 laser product performance will expose personnel to hazardous Class IIIb laser radiation bodily injury and void warranty coverage 5 Power Cord Set Requirements The power cord set
96. s in the box pressing the Ctrl Alt and Delete keys at the same time will log you on page 20 Note Caution Version RK09052000 Once you press the Ctrl Alt and Delete keys the Logon Information dialog box is displayed Typing your password identifies you as a valid user for this computer The default User name for the Leica TCS SP II system is Default password is After being logged in you can change your password by pressing Ctrl Alt and Delete keys at the same time Then click on change password The Change Password dialog box displays Type your current password in the Old Password field passwords are case sensitive so be sure you use the right case then press the Tab key Pressing the Tab key moves the cursor to the next field Type your new password then press the Tab key again Retyping the new password confirms that you aren t making a typing error This is important since the characters you type appear as asterisks on the screen If you miss type the confirmation password you will see a warning dialog Try again Then click the OK button Your new password will be in effect the next time you log on Do not forget your password if you set one Without the right password you can t access your computer anymore The Welcome dialog box is now displayed Take a moment to read the Did you know tip and then click the Close button to begin using Windows NT 8 1 Using a Mouse You
97. splays a grid on the annotation sheet which cannot be printed Version RK09052000 The Line Rectangle and Text commands can be activated both from the context menu and by clicking their corresponding buttons Saved annotation sheets are given the ano file extension page 80 10 4 2 Copying an Image onto the Annotation Sheet Function If you click the Snap button the image loaded in the Viewer window is copied onto the annotation sheet If you mark the copied image and then press the right mouse button a context menu containing the following commands appears Command Function Original size This displays the image in its original size Fit to page This enlarges the image to the size of the annotation sheet Bring to front This brings the image to the foreground Send to back This sends the image to the background Delete This deletes the image Note Version RKO09052000 You can change the size of the image by dragging one of the image grab points To move the image without changing its size mark the image and move it while pressing the left mouse button The Snap button can be applied only if a annotation sheet has been opened using the Annotation button first and the Viewer window has then been clicked 10 4 3 Drawing a Line on the Annotation Sheet Function Use the Line button to draw a line on the annotation sheet On the annotation sheet click the position where you wa
98. st out basis Caution Files that are overwritten due to an application using a duplicate filename will not be saved to the recycle bin Double clicking the Bin displays its contents The empty window confirms that there are no items in the Recycle Bin 8 3 The Start menu A single click of the left hand mouse button on the Start button will invoke the start menu and present you with the seven major categorized options for starting work on the system A single click of the right hand mouse button will invoke a small but powerful control menu containing the options Open Explore and Find Their functionality is described as follows 8 4 Starting a Program The Start menu contains the various categories where your applications and work are stored You can move further into the various subcategories by positioning the mouse over the category that you are interested in to automatically open the next subcategory You do not have to click the mouse The Programs command displays the Programs menu This menu lists all of the applications installed and available to you An arrow meaning that there is a submenu follows some items Drag the mouse cursor over the Accessories command to see its submenu The Accessories submenu lists the set of Windows NT built in programs TIP If you drag an object either from the Desktop or from the Windows Explorer and drop it directly onto the Start button a link to that object will automatically appear
99. start the recording page 38 Before Recording While Recording Selecting the objective Setting the beam path Selecting the scan format Setting the detection pinhole Selecting the scan mode Setting the electronic zoom Selecting the scan speed Setting the detectors Selecting unidirectional or bidirectional scan Adjusting the phase Setting the z y position The buttons of the functions that cannot be changed during continuous scanning are disabled and appear gray Typical Applications The continuous scan function is best used to optimize the image quality for the first scan of a specimen While the specimen is being scanned continuously you can modify the scan parameters listed above and check the results directly in the image Note With bleach sensitive specimens it is advisable to restrict the use of the Continuous Scan function to short time intervals Subjecting the specimen to the continuous light of the laser can destroy the specimen photochemically optical bleaching making it unfit 10 1 11 The Series Scan Overview Dialog Window Function This dialog window lets you define the begin and end points of an image series and trace the recording of the individual sections The three dimensional scan area is represented graphically as a cube Within this cube a yellow square represents the current Z position and y position a green one represents the begin point and a red one the end point The correspondi
100. tected at a specific wavelength from a single optical level Use the Lambda Scan Begin button to define the wavelength at which recording should begin The Lambda Scan Begin button is not enabled unless you have selected a scan mode with wavelength dimension using the Mode button page 41 LL see Selecting a Scan Mode gt Press Beam to open the Beam Path Setting dialog window gt Activate a detector by clicking its corresponding check box A cast shadow now links the activated detector to the corresponding slide bar on the scale of the spectrum gt Double click this slide bar to open the Range Properties dialog window where you can specify the detection bandwidth gt Drag the slide bar on the scale of the spectrum to the desired begin point The left edge of the slide bar represents the wavelength at which the recording is to begin Click Lambda Scan Begin to save the value The end point of a wavelength series is set in the same manner In addition you also have to set the number of wavelength steps LL see Defining the End Point for a Wavelength Series Ll see Setting the Number of Wavelength Steps Typical Applications A wavelength series can be applied to determine the maximum emission of a fluorochrome This is useful because the Stokes shift of the emission curve of a fluorochrome is dependent on each specimen that is applied This then allows you to set the detection range precisely for a specific
101. tegory in the Commands tab Left click the Apply button and keep the button pressed while pulling it to the window of your choice To insert the button in its current position release the mouse button 7 5 Saving images Individual images and experiments can be saved in the file formats described in section Readable file formats above To save images and experiments select File Save When an experiment is first saved the Save as dialog opens where you must enter a file name You can also select a file format in this dialog Experiments can only be saved in the Leica specific lei format When you are saving experiments you may be able to save existing individual images in tif or raw format page 19 Note Note Version RKO09052000 lf the experiment or image has already been saved the previous data is overwritten each time the experiment or image is saved To prevent this you must save the data under a new filename with File gt Save as 7 6 Organizing data by grouping experiments The Leica Confocal Software allows single images image series and the results of image processing steps to be combined in a group an experiment The experiment overview window provides an overview of all loaded experiments If the experiment overview window is not visible you can open it with View Experiment Overview With File New and File gt New Template a new experiment can be created In the same way exist
102. thinners as they attack the varnish All LEICA components and systems are carefully manufactured on the basis of the latest production methods If you encounter problems in spite of our efforts do not try to fix the devices or the accessories yourself but contact your Leica representative 7 The Leica Confocal Software An overview 7 1 Starting the software 7 1 1 Requirements for starting the software The LCS software is copy protected to prevent it from being used on two computers at the same time This protection system allows all additional application packages to be used To enable the software the provided dongle must be inserted into the parallel port of your computer The dongle does not affect the functionality of the parallel port e g for printing etc To use the software on a second computer the dongle must be fitted to its parallel port lf you remove the dongle from the confocal system control computer the software cannot be started preventing operation of the confocal system The LCS software can be started in two operating modes hardware mode and simulation mode In hardware mode all hardware components are accessed and initialized by the software For this reason you should switch on the hardware first and then wait about 20 seconds before starting the software In simulation mode the software runs independently of the hardware This mode is intended for secondary installations on another computer
103. times For every sampling point the arithmetical average is calculated from the repeatedly measured intensity values and represented in the result image The method used here determines a consecutive average This means every image recorded after the first image is calculated with the results of the previously displayed image and is displayed in the result image dynamic average Clicking the Average button opens a dialog window that you can use to set how often a section is to be scanned You can select from 1 to 64 scan repeats Typical Applications Recording an image using the averaging method is primarily useful for suppressing noise In fluorescence microscopy weak fluorescent specimens result in little light reaching the detector The resulting low photon count leads to noisy images In such cases you can improve the signal to noise ratio by carrying out multiple recordings and determining averages of the image Note Recording an image using the averaging method is not recommended for bleach sensitive specimens Repeated recordings and the resulting length of subjection to light can lead to the destruction of the specimen Version RKO09052000 page 51 10 2 Softwarefunctions for data display 10 2 1 The Viewer Window The default setting of the Viewer window consists of three main areas The image window 4 in which the recorded images are displayed is located in the middle Button pads 1 and 2 can be arranged
104. tions that you are currently using Click Show all to view all icons In the Settings register select the Display field to Zoom in on and zoom out of the image displayed in the Viewer window using the graphical zoom function n 1 Zooms in on the image by displaying n pixel as 1 pixel in the result image Automatic The image is displayed in the original size 1 n Zooms out of the image by displaying 1 pixel as n pixel in the result image Note The Leica Confocal Software provides three different zoom features the graphical zoom the 3D zoom and the electronic Zoom Q see Zooming the 3D View M see The Electronic Zoom In the same register click one of the boxes to show or hide the corresponding graphic element in the Viewer window scale Displays a scale bar with unit of length in the Viewer window Grid Applies a grid to the image displayed in the Viewer window The length of the scale bar is calculated depending on the selected objective electronic Version RK09052000 zoom and beam expander You can use the Movie register to set the speed for running the film sequence of an image series You can choose speeds from 6 images per minute to 25 images per second Use the mouse to drag the slide on the scale to the desired value LUI see Starting and Ending a Film page 57 10 2 4 The Viewer Options Dialog Window Surface Icon Function Click on Viewer Options in the View menu to
105. tivated detector to the corresponding slide bar on the scale of the spectrum Double click this slide bar to open the Range Properties dialog window where you can specify the detection bandwidth Drag the slide bar on the scale of the spectrum to the desired end point The right edge of Version RK09052000 the slide bar represents the wavelength at which the recording is to end Click Lambda Scan End to save the value page 43 The begin point of a wavelength series is set in the same manner In addition you also have to set the number of wavelength steps Q see Defining the Begin Point for a Wavelength Series O see Setting the Number of Wavelength Steps Typical Applications A wavelength series can be applied to determine the maximum emission of a fluorochrome This is useful because the Stokes shift of the emission curve of a fluorochrome is dependent on each specimen that is applied This then allows you to set the detection range precisely for a specific application 10 1 16 Defining the Number of Spatial Sections Function Clicking the Sections button opens a dialog window that you can use to select the number of horizontal xy sections and vertical xz sections for recording a picture series If you want to specify a number of sections that is not listed click the Others option This opens the Z Y Configuration dialog window which contains the following data Parameter Description Image Di
106. to change a search option select Options The first time you click on this tab Windows tells you it needs to create a list Click Next and Finish to allow this Then you see the main Find tab Type the word s you want to find in the top text box Then click a word in the middle box to narrow the search Finally review the list of help topics at the bottom and double click the one you want to read When you re done reading about a document click Help Topics to return to the main Help screen or click Back to return to the previous Help topic Click the window s Close button to exit Help 8 8 Shut Down Windows NT Always use the Shut Down command before you turn off your PC The Shut Down option allows the user to close the Windows NT Workstation operating system and ensure all running processes can halt cleanly and are given the chance to flush any data that may be in cache memory out to the disk Several options are available when shutting the system down Caution Powering down your computer without prior shutting it down may result in severe data loss 9 Help from the Internet First contact your local Leica subsidiary or the local dealer You can also find information on Leica Microsystems Heidelberg GmbH and confocal microscopy on our web site http www lit de Send email to support lllt de Version RKO09052000 page 26 10 Softwarefunctions 10 1 Software functions for data acquisition 10 1 1 Selecting an Ob
107. to the left and below the image window and the Experiment legend 3 to the right of it It is possible to select other default layouts of the Viewer window and save user defined configurations of it LL see Saving the Viewer Window as a Template Button Pads 1 and 2 By default the button pad to the left of the image window contains the buttons for image viewing functions and the button pad below the image window contains the buttons used for scrolling through the individual images of an image series Both button pads can be moved within the Viewer window or removed from it as separate windows To do so click on the double lined edge of the button pad and drag it while keeping the mouse button depressed to the desired position Image Window 4 If you place the mouse pointer at any position within the image window and then press the right mouse button a context menu containing the following commands opens Command Function This copies the raw data of a selected area in l i the image window as a new image in the ta EAE Ament a current experiment The new image can be further processed Version RK09052000 page 52 This creates a new image in the current experiment from the snapshot of a selected area in the image window The new image cannot be processed any more This creates a new image in the current experiment from the snapshot of the entire All Snapshot l l l current image window The new imag
108. ues is transmitted The transmission values are generally at 80 and the reflection values are between 90 and 95 Dry objective A microscopic objective used without immersion media Between the objective lens and the specimen is air Fluorescence A light optical contrast process for displaying fluorescent microscopy structures Auto fluorescent samples have a so called primary fluorescence They do not need to be enriched with additional fluorescent substances Secondary fluorescent substances on the other hand have to be treated with appropriate dyes or Stains called fluorochromes Specific dyeing methods additionally allow the precise localization of the stained structure elements of an object Fluorescence microscopy allows both nAtAnnantral raAAnvrnhnAlAAtrAaAl marra AIA AFI AAR annl 4AA ahilik tna nanem a t Version RK09052000 page 91 Fluorite objectives Immersion objective Neutral filter Pixel Plan objectives flatfield objectives Reflection short pass filter Reflection long pass filter Refractive index Signal to noise ratio Stokes shift Version RK09052000 dynamic examinations on a molecular level Describes a correction class for objectives Fluorite lenses are semi apochromatic meaning their degree of correction lies between the achromatic and apochromatic A microscopic objective developed with the requirements for applying immersion media The use of incorrect or no immersion me
109. utton is enabled the Next and Previous First and Last and Play Stop buttons are disabled and displayed in gray The Gallery button is used to show all of the individual images of an image series and does not allow film sequence mode 10 2 18 Viewing the Last Image of a Series gt Function You can view the individual images of an image series as a film sequence Click Last to jump to the last image in the series Note lf the Gallery button is enabled the First and Last Next and Previous and Play Stop buttons are disabled and displayed in gray The Gallery button is used to show all of the individual images of an image series and does not allow film sequence mode Version RKO09052000 page 69 10 2 19 Starting and Ending a Film Function You can view the individual recorded images of an image series as a film sequence Use the Play Stop button to start and stop the film The speed of the film i e the number of individual images per unit of time is variable and can be set in the Viewer Options dialog window gt From the View menu select Viewer Options gt Click the Display icon and then on the Movie register H see The Viewer Options Dialog Window Display Icon Note If the Gallery button is enabled the Play Stop First and Last Next and Previous buttons are disabled and displayed in gray The Gallery button is used to show all of the individual images of an image series and does not allow film sequen
110. view of the applications to ensure that they can always be seen To switch from one running program to another simply click on the second program as displayed in the Taskbar page 24 Note Version RKO09052000 The Taskbar also provides constant additional information such as the system time and volume control if you have a sound card fitted all of which can be further tailored by the user 8 6 Setting the Time and Date The current Date Time and Time Zone information can be set from the Date Time icon within the Control Panel This setting is important since Windows NT stamps the date and time on all of your files as you create and modify them The two options can be selected by clicking on the appropriate tab To change the Date and Time Click on the appropriate date or use the controls to change the month or year The time can also be changed by first selecting the digital display and then using the up and down arrows To change the Time Zone Select the appropriate Time Zone from the drop down list at the top of the screen Notice that the option to automatically adjust the clock for daylight savings time is selected On some systems you can also drag the highlighted area on the world map and drop it on the correct location Changing the date and time information within Windows NT Workstation 4 will update the battery backup CMOS clock in your system Depending on the shell configuration systems connected to a network
111. which the material layers have different refractive indices immersion objectives are most suitable With biological specimens it depends on the specimen and the medium in which it is embedded whether an oil immersion objective or a water immersion objective should be applied The highest resolution is achieved when the refractive indices of the immersion medium or the specimen on the one hand and the objective medium on the other are as homogeneous as possible 10 1 2 The Electronic Zoom Function In confocal microscopy the magnification of an image is determined both by the objective and the electronic zoom The objective generates an intermediate image magnified based on the objective s magnification factor Additional magnification is provided by the electronic zoom At a zoom factor of 1 the Version RKO09052000 page 28 maximum scan field size is scanned with a specified number of sampling points If the zoom factor is set to 2 the same number of points is used to scan a scan field with only half the length of a side of the maximum scan field 1 4 the original scan field This achieves stronger magnification and thus better image resolution because a smaller scan field is scanned with the same frequency which results in a higher density of data You can select one of the preset zoom factors in the dialog window that is opened with the Zoom button If you click Others there are two ways of specifying another zoom factor
112. ys the Network Neighborhood dialog box which gives you information about who and what is connected to your workstation It provides an easy mechanism for browsing any network systems and resources that you may be able to connect to in a way that is independent of the actual type of network vendor Traditionally if a system needed to be simultaneously connected to different types of network the way in which each could be connected and viewed would be vendor specific Windows NT Workstation 4 is capable of displaying a common view of your entire network even though it may actually comprise resources from Windows NT Novell NetWare Banyan Vines or others The Inbox icon is used if Microsoft Exchange is active on your system Windows NT Workstation 4 has in built electronic mail services based on Microsoft Mail MS Mail and Microsoft Exchange If there is already an MS Mail post office on the same Version RKO09052000 page 22 network that the system is connected to the Windows NT Workstation 4 mail client can connect directly into it The Inbox lets you access your messages The Recycle Bin icon represents the holding place for deleted items As long as files are in the Recycle Bin they can easily be recovered if they have been accidentally deleted Windows NT Workstation 4 will preserve files until the system runs out of free disk space When this happens Windows NT Workstation 4 will prune the contents of the Recycle Bin on a first in fir