Home
        B - GE Healthcare Life Sciences
         Contents
1.                                                                 LSS   Calibrating with the electrode outside the flow cell  When calibrating the electrode out of the flow cell and changing from one  buffer to another  rinse the electrode tip with distilled water and dab it  carefully with a soft tissue to absorb the remaining water  Do NOT wipe the  electrode as this may charge it and give unstable readings   The steps below describe the procedure used with the electrode removed  from the flow cell   Note  The user interface of the monitor must be unlocked if connected to a  UNICORN control system   1 Remove the pH electrode from the flow cell and immerse the electrode  in the first standard buffer solution  normally pH 7 0     Set Parameters 2 From main menu 1  turn the dial four steps clockwise to enter menu  Set Parameters  Press OK    Set Parameters 3 Select Set Parameters pH  Press OK    Calibrate pH 4 Select menu Calibrate pH  Current calibration values are displayed    7 00 12 00   buffer 1 buffer 2     Buffer 1 fixed lower calibrated pH value  Range 0 00 14 00  Buffer 2 fixed higher calibrated pH value  Range 0 00 14 00  Note  The pH values for buffer 1 and 2 must differ by at least 1 pH unit    Calibr pH Buffer 1 5 Press OK to access the settings menu  The order of calibration   buffer 1 or buffer 2  is optional  Press OK to start with buffer 1  or turn  the dial to start with buffer 2  In this example  we start with buffer 1    Calibr pH Buffer 1 6 This text disappe
2.                See next page  Setup Setup for sub menus                                           Setup Language                                        Set Clock Setup Unit Number  Set Timer z   18 34 52  00 00 00 Setup Display Angle                      Set Alarm 00 32 21    Alarm Timer 12 30 52 Alarm Timer off     Buzzer 08 23 34    48 Monitor UPC 900 User Manual 18 1125 55 Edition AG                                     Reference information   3    B 5 Menu overview  continued     Continued from  previous page                                                                                                                                                                                                                                                                                  Setup  Setup Adjust Temp  Temp sub menus  7 Warning  This will  Setup Adjust Temp  Setup Show Temp   on  Setup Show Temp   on  on off  Setup Cond Temp Comp   0 0   0 0  Setup Cond Ref Temp   25 0   25 0  Cond sub menus Warning  This will  Setup Cond Temp Comp change cell calib  Setup Adjust Cond  Setup Cond Ref Temp   gramm  83 53mS cm  86 6  Setup Adjust Cond   a2 53 m lt lem  Warning  This will  Setup Adj Cell Const change cell calib           Setup Adj Cell Const  a cond  83 56mS cm  86 6  Setup Show Cond   on  on off  pH sub menus Setup pH Temp Comp  Setup pH Temp Comp loff  on off   of  Setup Show pH Setup Show pH   on   on  on off  UV sub menus Set Averaging   1 35  0 64  Set Averaging  Setup L
3.      AS Inserting the pH electrode    Note  Handle the pH electrode with care        CAUTION  The tip of the pH electrode consists of a thin glass membrane   Protect it from breakage  contamination and drying out or the electrode  will be destroyed  Always store the electrode with the end cover filled with  a 1 1 mixture of pH 4 buffer and 2 M KNO   Do NOT store in water only              1 Unpack the pH electrode  Ensure that it is not broken or dry     2 Before using the electrode  remove the electrode end cover and immerse    the glass bulb in buffer for 30 minutes   To module     pH probe       h 3 Remove the dummy electrode from the flow cell and store it in the flow  socket    J cell holder     4    Carefully insert the electrode in the flow cell  Tighten the nut by hand to  End secure the electrode          Note  Ifthe flow cell is full of liquid  it is not possible to insert the electrode     If so  loosen the inlet connection while inserting the electrode to allow  the liquid to run out from the flow cell  Remember to re tighten the  connector     Flow cell Note  Ifthe electrode is not fully inserted  the system will leak and a dead    Flow cell holder volume will occur in the holder     holder  5 Connect the pH electrode cable to the socket pH Probe on the rear of  the module     Monitor UPC 900 User Manual 18 1125 55 Edition AG 7    Installation                         2 6 Connecting electrical signal cables  The sockets for electrical signals are located on the re
4.     The electrode glass membrane may be cracked  If so  replace the  electrode     Clean the pH electrode according to section 4 8 and recalibrate   If the problem remains  replace the pH electrode     Check the electrode glass membrane  If it is contaminated  clean   the electrode following the instructions in section 4 8 Cleaning   the pH electrode    If the membrane has dried out  the electrode may be restored by  soaking it in buffer overnight    Clogged liquid junction  Refer to section 4 8 Cleaning the pH electrode     Check that the electrode cable is connected properly to the rear of  the module    Check that the pump operates correctly    Check that the electrode is correctly inserted in the flow cell and    if necessary  hand tighten the nut    If air in the flow cell is suspected  tap the flow cell carefully or tilt it  to remove the air  Alternatively  flush the flow cell with buffer at   20 ml min for 1 2 min  Use a flow restrictor after the pH electrode   Check that the pH electrode is not broken    Check that the pH electrode is calibrated    Check the slope  see section 3 10 Calibrating the pH electrode   If it is  outside the range 80 105  or if the asymmetry potential deviates more  than  60 mV from O mV  clean the pH electrode  Recalibrate  If the  problem persists  replace the pH electrode    Clean the pH electrode if required  see section 4 8 Cleaning the pH  electrode    Compare the response of the pH electrode with that of another pH  electrode  If the 
5.     The length should be 170 mm when using the 5 mm flow cell  and  230 mm when using the 2 mm flow cell     5 Connect the other end of the tubing to the conductivity flow cell or to  another appropriate unit     2 4 Installing the conductivity flow cell    1 Place the conductivity cell in its cell holder  or in a suitable location  as  close to the optical unit column as possible  The cell can be placed up to  1 5 m from the monitor housing  Secure the cell with the clamp     Note  When the conductivity flow cell is used in conjunction with the pH  electrode  place the conductivity flow cell and select its flow direction  so that the screw head end of the flow cell faces the pH flow cell     2 Connect the conductivity cell to the socket Conductivity Flow Cell on  the rear panel of the module     3 Connect the tubing with fingertight connectors     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Installation       2 5 Installing the pH flow cell and electrode    Mounting the flow cell holder  In AKTA design systems  the pH electrode is optional     1 Hook the flow cell holder on the right hand side of the housing  Secure it  with the slide clamp   If the flow cell holder is not used  the flow cell must still be installed at an  angle of 30   from the vertical with the outlet placed higher than the inlet  to prevent air bubbles being trapped in the cell     The flow direction is marked on the flow cell        Flow direction 2 Connect the tubing with fingertight connectors
6.    AU Cond Tc pH       Lamp   on           Autozero UV          Eventmark             pH 12 50 22 4          Set Parameters          Check          Setup          Alarm Timer 00 22 24                Lamp   on                 Lamp   on  on off             Operation    3 3 Main menu overview  Main menu 1  This menu is accessed from all positions by pressing ESC     Lamp control  This menu is accessed by turning the dial one step counter   clockwise from main menu 1  It is used to manually switch the lamp on off     Autozero  Used to zero the UV signal from the monitor   Eventmark  Used to set eventmarks on the UV curve     Main menu 2  Used to display pH  temperature  in the conductivity flow cell   and conductivity in mS cm as well as in percentage of full range     Parameter menus  Used to set measurement parameters for  Cond  pH and UV     Check menus  Checking internal operating values   See Reference information section B 1     Setup menus  Setting up language  unit number  etc   See Reference information section B 2     Alarm Timer menus  Setting different timer options   See Reference information section B 3     3 4 Controlling the UV lamp    The Lamp menu is accessed by turning the dial one step counter clockwise  from main menu 1  and then pressing OK  We recommend the lamp be  switched off to conserve lamp operating time when no measurement   is being made  A warm up time of 60 minutes is required to achieve full  specifications  However  in most cases  a warm up time
7.    Reference information    Conductivity measurement    Conductivity range 1 uS cm to 999 9 mS cm   Reproducibility   short term Max   1  or  5 uS cm whichever is greater  long term Max   3  or  15 uS cm whichever is greater  Noise Max   0 5  of full scale calibrated range  Response time  lt 3 s  0   95  of step    Temperature sensor   accuracy  2 0   C   drift  0 5   C per 10h   Flow rate sensitivity  1  within 0 100 ml min   Conductivity flow cell   Flow rate 0   100 ml min   Max  pressure 5 MPa  50 bar  725 psi     Generated backpressure Max  0 01 MPa at 100 ml min    pH measurement  pH range 0 to 14  spec  valid between 2 and 12     Accuracy  temperature compensated  0 1 pH within  4 to  40   C  not temperature    compensated  0 2 pH within  15 to  25   C   0 5 pH within  4 to  15   C and  25 to   40   C   Response time  lt 10 s  0   95  of step    Long term stability Dev  max  0 02 pH h  measured at pH 4 0    Flow rate sensitivity Dev  max  0 1 pH units   pH flow cell   Flow rate 0 1   100 ml min   Max  pressure 0 5 MPa  5 bar  72 psi     Generated back pressure Max  0 02 MPa  0 2 bar  2 9 psi  at 100 ml min    52 Monitor UPC 900 User Manual 18 1125 55 Edition AG    C 2 Physical data    Control    Degree of protection  housing    Power requirements  Power consumption    Functions    UV lamp cable length  Optical unit  cable length  pH electrode cable length  Cond  cell cable length  Tubing connections  Analogue outputs   Display   Dimensions  H x W x D  Weight    Wetted 
8.   01234567 YM 012345          UPC 900  V1 00             Date of Maintenance              Buzzer Test                Setup  Temp Cond pH UV                Setup  Temp Cond pH UV             Setup Adjust Temp   22 3  C           Warning  This will  change temp calib          Setup Adjust Temp   22 3  C  223             40    ED Reference information    3 The service telephone number is displayed  press OK   4 The service contract number is displayed  press OK   5 The module serial number is displayed  press OK     6 The module name and software version are displayed  press OK        7 The date of the last service is displayed  press OK   8 Atest of the module   s buzzer is performed  press OK     B 2 Setup menus    B 2 1 Selecting setup sub menus    There are setup sub menus for temperature  conductivity  pH and UV  absorbance     1 Select menu Setup and press OK   2 Select sub menus Temp  Cond  pH or UV with the dial and press OK     B 2 2 Setup adjust temperature    Calibration of the temperature sensor in the conductivity flow cell is only  necessary if the monitor is used in high accuracy measurement or if the  conductivity flow cell is replaced     1 Place the flow cell together with a precision thermometer inside a box or  empty beaker to ensure that they are not exposed to draught  Leave  them for 15 minutes to let the temperature stabilise     2 Read the temperature on the thermometer   3 Select sub menu Setup Temp and press OK     4 Select sub menu Setup Adjust Temp a
9.   Filter wheel cpl  1 A 18 0647 01  Conductivity flow cell complete 1 A 18 1111 05  pH electrode  round tip 1 C 18 1111 26  pH flow cell  round tip  including 1 A 18 1112 92  dummy electrode   Dummy electrode  round tip 1 A 18 1111 03  pH electrode complete  round tip  incl  1 C 18 1134 84  pH electrode  flow cell and holder   Flow cell holder UPC 900 1 A 18 3055 87  Clamp  Conductivity flow cell 1 A 18 1111 14  PEEK tubing  i d  0 50 mm  o d  1 16     G  2m A 18 1113 68  PEEK tubing  i d  0 75 mm  o d  1 16    2m A 18 1112 53  Union 1 16    female M6 male  PEEK 6 A 18 1112 57  Fingertight connector 1 16       or PEEK tubing o d  1 16    10 A 18 1112 55  Union M6 female 1 16    male  PEEK 8 A 18 1112 58  Union  1 16    male 1 16    male    or 1 16    o d  tubing  PEEK 10 A 18 1120 92          Monitor UPC 900 User Manual 18 1125 55 Edition AG 55    D  Reference information    56    Tubing cutter   U wrench  M6   U wrench  1 4      Signal cable  6 pin mini DIN open  Chart Recorder REC 111  1 channel  Chart Recorder REC 112  2 channel    al A 18 1112 46  ol A 19 7481 01  1 A 18 1112 45  al A 18 1110 64  1 A 18 1132 32  1 A 18 1132 33          A   accessories  C   consumables    Monitor UPC 900 User Manual 18 1125 55 Edition AG    Short instructions    Short instructions                Calibrating          Lamp          AU Cond Tc pH  0 00002 015 0 12 50          Autozero UV          Eventmark          Calibrate pH          Setup Adjust Cond       The following short instructions
10.  436 and 546 nm through  filter change     e Two alternative flow cells with pathlengths of 2 mm and 5 mm    e Fast response   e High accuracy and reproducibility   e Flow cells with low dead volume   e Accurate and reliable monitoring through self test and self calibration    e Flow cells can be connected close together  minimising band broadening  and time delay between detectors         QO                The monitor and the flow cells are separate units  All flow cells are  connected to the rear of Monitor UPC 900     Monitor UPC 900 User Manual 18 1125 55 Edition AG 1    Introduction      Bee DDD    12 Safety          IMPORTANT  Monitor UPC 900 is intended for laboratory use only  not for  clinical or in vitro use  or for diagnostic purposes        The module is designed for indoor use only   Do not use in a dusty atmosphere or close to spraying water   Operate in accordance with local safety instructions        WARNING  The module must be connected to a grounded mains socket           WARNING  Always disconnect the power supply before attempting to  replace any item on the module during maintenance           WARNING  The module must not be opened by the user  It contains high  voltage circuits that can deliver a lethal electric shock           WARNING  When lamp power is on  the lamp socket carries a dangerous  voltage  Do not connect disconnect with the monitor switched on           WARNING  The module uses high intensity ultra violet light  Do not  remove the UV lamp w
11.  Monitor UPC 900 User Manual 18 1125 55 Edition AG 47    ED Reference information    B 5 Menu overview    Lamp   on  Lamp    AU Cond Tc pH  0 00002 015 0 12 50    I  Autozero UV                               1 Calibrated Electrode  Eventmark                            Calibr pH Buffer 2  10 00        Set Cond Full Scale                                                                                                                                                                                                                                                                                                       en E E Card eA eE Calibr pH Buffer 2  10 00   Set Cond Zero Level Calibr pH Buffer 2  Set parameters  Set Cond Zero Level Calibr pH Buffer 1  7 00   Set Cond Analog  Out Set Cond Analog Out Calibr pH Buffer 1  7 00   Calibrate pH Calibr pH Buffer 1  Set parameters  Set pH Analogue Out Set pH Analogue Out  Set UV Analogue Out Set UV Analogue Out Set pH Zero Level  Set UV Range Set pH Zero Level  Set UV Range Set pH Full Scale  Set UV Zero Level Set pH Full Scale  Set UV Zero Level                                                                                  Check Check Autozero  AZ 0 0356 AU  Check Lamp Run Time Lelephone Service   Contract Number   Check Lamp Intensity 0  22456722001  Serial Number                  yee MAA N19740  UPC 900  Check Service Mode V   Date of Maintenance           Check Recorder                                                 Buzzer Test    
12.  are intended as a guide to users who   are fully familiar with the safety precautions and operating instructions  described in this manual  These short instructions assume that the module is  installed according to the installation instructions        Select       1 Switch on the module with the mains switch on the rear panel   2 Switch the lamp on off   3 The main operating menu displays measured parameters    4 Autozero the module by pressing OK    5 Set Eventmark by pressing OK     6 Calibrate the pH electrode before use and or daily by using 2 buffers  with known pH values     7 The conductivity flow cell does not normally need to be calibrated         83 53m         83 56mS cm        Setup Adj Cell Const    86 6             Set parameters             8 To connect a chart recorder  set the Analogue output scaling  0 100    for measured parameters     9 Always store the pH electrode in a 1 1 mixture of pH 4 buffer and  2 M KNO  when not in use     For local office contact information  visit  www gelifesciences com contact    GE Healthcare Bio Sciences AB  Bj  rkgatan 30   751 84 Uppsala   Sweden    www gelifesciences com protein   purification    imagination at work    GE  imagination at work and GE monogram are trademarks of General Electric  Company     AKTA  Drop Design and UNICORN are trademarks of GE Healthcare  companies     Purification and preparation of fusion proteins and affinity peptides  comprising at least two adjacent histidine residues may require a licens
13.  conductivity cell  Clean the flow cell and recalibrate   5 Ifthe problem remains  change the conductivity cell   ERROR key  OK  1 The key was pressed during self test  or is faulty     Switch off the module   3 Switch on the module           ERROR key  Esc           ERROR key  OK Esc           BAER nine na  1 Switch off the module   2 Check all connections   3 Switch on the module           ERROR number 109 113          ERROR number 120 121          ERROR number 106 108 1 Switch off the module   2 Check all UniNet 1 and UniNet 2 connections   3 Switch on the module                 ERROR number 118          1 Switch off the module   Check all connections   3 Switch on the module     Exc x y in ab c                Exc DIV O in ab c          c instr in ab c    m  x          Exc address in ab c    x             Monitor UPC 900 User Manual 18 1125 55 Edition AG 33    Trouble shooting    34 Monitor UPC 900 User Manual 18 1125 55 Edition AG    Reference information    A Description       A 1 Module    Monitor UPC 900 is an on line monitor for measurement of UV absorption   pH  optional in   KTA design  and conductivity  The monitor can work with  standard glass pH electrodes with a built in liquid filled reference electrode  and a BNC connector              LDIECDBDECDECDICDICDIEDEDEDEDED  9 eee          Optical Anclque  Conductivity O O   ie  ande Unit outoav Uninet 1    O7    D6D00       Conductivity measurement has a dynamic range from 1 uS to 999 9 mS cm  and is suitable for a 
14.  current filter constant  is shown  Valid values are 0 02  0 04  0 08  0 16  0 32  0 64  1 3  2 6  5 and  10 seconds     3 Change the setting if desired and press OK     Monitor UPC 900 User Manual 18 1125 55 Edition AG       Setup  Temp Cond pH UV          Setup Lamp Run Time  Hg  2000h           Setup Lamp Run Time  Hg  2000h  000                Setup  Temp Cond pH UV             Setup Show UV   on                 Setup Show UV   on  on off                Setup Language   GB  GB DE FI             Reference information   3    B 2 12 Setup lamp run time  When the UV lamp is replaced  the Lamp Run Time counter should be reset   1 Select sub menu Setup UV and press OK     2 Select sub menu Setup Lamp Run Time and press OK     3 Set the Lamp Run Time counter to zero with the dial  Press OK to  acknowledge     B 2 13 Setup show UV    Normally UV absorbance is displayed in main menu 1  If not required  the UV  absorbance display can be set to off     1 Select sub menu Setup UV and press OK     2 Select sub menu Setup show UV and press OK  The current status for  showing UV is displayed  If on is shown  the current UV value is  displayed in main menu 1  If off is shown  no UV value is displayed in  main menu 1     3 Change the setting if desired and press OK     B 2 14 Setup language   The language used on the display can be changed   1 Select main menu Setup and press OK    2 Select sub menu Setup Language and press OK   3 Select the desired language     GB   English  D   German  
15.  default  Adjust the  compensation factor setting as necessary and press OK     B 2 5 Setup conductivity reference temperature  1 Select sub menu Setup Cond and press OK     2 Select sub menu Setup Cond Ref Temp and press OK  The current  reference temperature value is shown  25 0   C is the default setting   The range is 0 00 99 9   C     3 The current reference temperature value is displayed as default  Adjust  the reference temperature value setting as necessary and press OK     B 2 6 Setup adjust conductivity    Normally it is not necessary to adjust the cell constant as the flow cell is  pre calibrated on delivery  Adjustment is only necessary when replacing  the conductivity flow cell with a flow cell whose cell constant is unknown   We recommend that the conductivity flow cell is recalibrated after  cleaning  When adjusting the cell constant from UNICORN  select System  Control System Calibrate and then select CondCalib     Note  The conductivity temperature compensation must not be used when  adjusting the cell constant  Set the Setup Cond Temp Comp to 0   see section B 2 4   The temperature sensor must be calibrated before  adjusting the cell constant  see section B 2 2      Monitor UPC 900 User Manual 18 1125 55 Edition AG 41    ED Reference information    1 Prepare a calibration solution of 1 00 M NaCl  58 44 g l  Let the  solution stand until it is at room temperature  This is important for exact  measurements     2 Fill the flow cell completely with the calibrati
16.  electrodes in the flow cell may be contaminated and require cleaning  To   clean the flow cell    1 Pump 15 ml of 1 M NaOH at 1 ml min through the flow cell either by  using a pump or a syringe    2 Leave it for 15 minutes    3 Rinse thoroughly with 50 ml distilled water     Note  If the flow cell is totally blocked  the blockage can be removed using a  needle or a wire with a diameter less than 0 8 mm     4 7 Changing the conductivity flow cell   The conductivity flow cell can be changed when required  Make sure the  monitor is switched off before disconnecting connecting the cell from the  rear of the monitor housing     If the cell is replaced with a new flow cell  the monitor must be calibrated  with the new cell constant value written on the flow cell package  See  section B 2 in Reference information  If the cell constant is not known  it can  be determined  see also section B 2 in Reference information      Monitor UPC 900 User Manual 18 1125 55 Edition AG 25    Maintenance    4 8 Cleaning the pH electrode    Note  The pH electrode has a limited lifetime and should be replaced every  six months  or when the response time is slow or the slope is out of  range   lt 80          WARNING  HCI and NaOH are injurious to health  Avoid spillage           WARNING  When using hazardous chemicals  take all suitable protective  measures  such as wearing protective glasses and gloves resistant to  the chemicals used  Follow local regulations and instructions for safe  operation and
17.  in the flow cell  Check that the flow restrictor gives a   back pressure of 0 2 MPa    If there is a lot of air in the water  degas the buffer before use    Check the connections of the optical unit    Clean the UV cell  see sections 4 3 and 4 4    Locking nut in optical unit not properly tightened  Turn the locking nut to   the stop position     Pr    oo WM    Check that there is no air in the eluents  Degas if necessary    Clean the system in accordance with the   KTA design System Manual   Clean the column in accordance with the column instructions    Check that the mixer is functioning correctly and that the correct  chamber volume is being used     PWN PrP    Monitor UPC 900 User Manual 18 1125 55 Edition AG 27    Trouble shooting       SS    Fault Action  Low sensitivity 1 Aging lamp  Check the lamp and replace if necessary  2 Wrong lamp position  Check that the lamp position and the wavelength  used  filter position  fit together   3 Dirty on line filter  Clean or replace the filter   Error in external 1 Check the chart recorder in accordance with its manual     chart recorder    Test the recorder function by selecting recorder test according to section  B 1 in Reference Information     pH measurement  optional in AKTA design systems     No response to  pH changes    Small response to  pH changes    Slow pH response or  calibration impossible    Incorrect unstable pH  reading    28    1    Check that the electrode cable is connected properly to the rear of  the module
18.  maintenance of the system      gt  gt              Use one of the following procedures to clean the electrode to improve the  response     e Salt deposits  Dissolve the deposit by immersing the electrode first in  0 1 M HCI  then in 0 1 M NaOH  and again in 0 1 M HCI  Each immersion is  for a 5 minute period  Rinse the electrode tip in distilled water     e Oil or grease films  Wash the electrode tip in a liquid detergent and  water  If the film is known to be soluble in a particular organic solvent   wash with this solvent  Rinse the electrode tip in distilled water     e Protein deposits  Dissolve the deposit by immersing the electrode in  a 1  pepsin solution in 0 1 M HCI for five minutes  followed by thorough  rinsing with distilled water     If these procedures fail to rejuvenate the electrode  the problem is most   likely a clogged liquid junction    1 Heat a 1 M KNO  solution to 60 80   C    2 Place the electrode tip in the heated KNO  solution    3 Allow the electrode to cool while immersed in the KNO  solution before  re testing     If these steps fail to improve the electrode response  replace the electrode     4 9 Changing the pH electrode    See section 2 4 Installing the pH flow cell and electrode     4 10 Recycling       This symbol indicates that the waste of electrical and electronic  equipment must not be disposed as unsorted municipal waste and must  be collected separately  Please contact an authorized representative of  the manufacturer for information co
19.  of 15 minutes is  sufficient     Switching the lamp on off    1 Switch the lamp on off by moving the cursor with the dial  and then  pressing OK     Monitor UPC 900 User Manual 18 1125 55 Edition AG 13    Operation    Ge d Tc pH   gt        Absorbance value       Autozero UV                Eventmark                Set Parameters          Set Parameters                Set Cond Analog  Out             14    3 5 Reading the UV absorbance value    The main menu 1 shows the absorbance value with four digits for the  chosen wavelength  This menu is reached from any other menu by pressing  ESC     If the lamp is off  Lamp Off is displayed instead     3 6  Autozeroing UV    The autozero function in the UV part of the module sets the detected  absorbance to zero when OK is pressed  Using autozero is recommended at  the start before the sample is injected     1 From main menu 1  turn the dial one step clockwise to select  Autozero UV  then press OK  The normal absorbance value display is  then shown     In UNICORN  Autozero is set with the instruction AutoZeroUV in System  Control Manual Alarm amp Mon     3 7 Creating eventmarks    Eventmarks can be set and displayed as spikes on the UV curve  for  example  when the sample is injected  The spikes are 10  of the full scale  AU setting     1 From main menu 1  turn the dial two steps clockwise to select  Eventmark  then press OK  A spike on the UV curve is created     3 8 Setting Cond  pH and UV analogue outputs    This settings menu is 
20.  the flow cell     2 Insert the flow cell into the detector housing from above   Note  The flow cell can only be placed in one correct position   3 Secure the flow cell by turning the locking nut until the stop position     Note  If the locking nut is not tightened sufficiently  the monitor will function  poorly  e g  drifting base line      4 Place the protective cover around the flow cell to protect the electronics  inside the optical unit from liquid spillage     Note  Avoid spillage for prolonged monitor lifetime   5 To remove the flow cell  reverse the procedure     Note  Ensure that the Hg lamp position and the filter is selected according  to the wave length to be used  This is described in the Changing lamp  assembly section below     Connecting the optical unit to the module    1 Place the optical unit in the holder  or in a suitable location as close to  the column as possible  The optical unit can be placed up to 1 5 m from  the monitor housing     2 Secure it by tightening the screw in the holder     3 Connect the lamp cable to the socket Lamp on the rear panel of  the module     4 Connect the signal cable to the socket Optical unit on the rear panel of  the module     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Installation    Changing lamp assembly    WARNING  The module uses high intensity ultra violet light  Do not       remove the UV lamp while the monitor is running    Before changing a defective lamp  ensure that the lamp cable is  disconnected fro
21. 3 14 Storage and shut down          3 15 Restart after power failure       Monitor UPC 900 User Manual 18 1125 55 Edition AG V    Contents    4 Maintenance                4 1  Periodic maintenance     4 2 Cleaning and checking the module  43 Cleaning the UV flow cell in place           e  4 4  Cleaning the UV flow cell off line         e  45 _ Cleaning the conductivity flow cell in place          24  4 6 Cleaning the conductivity flow cell off line         24  4 7  Changing the conductivity flow cell    24  4 8 Cleaning the pH electrode    25  49 Changing the pH electrode    26  4 10  RECYCIINY               dns 26  5 Trouble shooting  51 VH  EN  TO   ennemies  52  REGY  5 3 Faults and actions  5 4 M  TTOTIM  SSAUES nent  Reference information  A Descriptions IEERIRIS mag 34  Al  Module    34  A2  UV optical Unit    nes 35  A 3 pH electrode    36  AAs  Condeutiviiy cellsseskeanissesniisenieene 36  B Mende   B 1 Check menus  B 1 1 Checking autozero level    37  B 1 2 Checking lamp run time    37  B 1 3 Checking lamp intensity    37  B 1 4 CHECKING recorder    37  B 1 5 Checking service mode    37    VI    Monitor UPC 900 User Manual 18 1125 55 Edition AG    Contents       B 2     Setup FMS IOS sesessscssszcssseassscsssescaznnassatseavsaassseonsaassasncdoansssottseactsee 38  B 2 1 Selecting setup sub menus    38  B 2 2 Setup adjust temperature    38  B 2 3 Setup show temperature    39  B 2 4 Setup conductivity temperature  COMPENSATION creme 39  B 2 5 Setup conductivity referen
22. B 1 2 Checking lamp run time    The lamp run time can be checked to determine the need for lamp  replacement  Run times for both Hg and Zn lamps are monitored           Check Lamp Run Time 1 Select menu Check and press OK   Hg 2300h Zn 340h       2 Select sub menu Check Lamp Run Time        B 1 3 Checking lamp intensity  The lamp intensity can be checked to determine the status of the lamp used           hack Len Mens 1 Select menu Check and press OK        2 Select sub menu Check Lamp Intensity        B 1 4 Checking recorder  The function of a connected chart recorder can be tested        Check Recorder 1 Select menu Check and press OK   2 Select sub menu Check Recorder and press OK              3 Start the test by selecting on and press OK     Check Recorder  Please wait              The test will ramp the signal on each channel  Signal to recorder  up to 1 V and then decrease the signal in 10    steps back to 0 V  The test is run continuously   Compare the diagram of the chart recorder ov  with the figure     4 Stop the test by pressing OK or ESC        B 1 5 Checking service mode    Service information relevant to the module can be checked  Information  may not be available in all menus     1 Select main menu Check and press OK        Check Service Mode       2 Select sub menu Check Service Mode and press OK        Monitor UPC 900 User Manual 18 1125 55 Edition AG 39       Telephone Service   012345678901          Contract Number   2345678901    oO          Serial Number 
23. F   French  E   Spanish      Italian    Monitor UPC 900 User Manual 18 1125 55 Edition AG 45    ED Reference information    B 2 15 Setup unit number   The unit number is the identification the Monitor UPC 900 has on the  UniNet bus  It should correspond to the number set in UNICORN for the  Monitor UPC 900  The number should be set to 0 if one UPC 900 is used   If more than one UPC 900 monitor is used  they must all have different  numbers     1 Select main menu Setup and press OK        Sawo Uni Number 2 Select sub menu Setup Unit Number and press OK              3 Select unit number  0 25  and press OK     B 2 16 Setup display angle  Sets the display angle to compensate for different viewing heights     1 Select main menu Setup and press OK        Set Display Angle 2 Select sub menu Set Display Angle and press OK              3 Select viewing angle    gt   Up    gt   Mid or   gt   Down      46 Monitor UPC 900 User Manual 18 1125 55 Edition AG       Alarm Timer 12 30 52          Set Alarm 12 32 21   0  00 00 00          Set Timer   18 34 52  00 00 00          Alarm Timer 12 35 16  Buzzer 00 33 00          00 00 29 13 08 45  I  Alarm time             I  Alarm time             Set Clock        12 26 53  12 36 53          Alarm Timer off   Buzzer 18 34 52             Insert Access Code           Reference information   3    B 3 Setting and using the alarm timer    You can set the alarm function to either a fixed alarm time or use a count   down timer  You cannot set both an ala
24. GE Healthcare    AKTA    design    Monitor UPC 900    User Manual            KTA    Important user information       All users must read this entire manual to fully understand  the safe use of AKTA    monitor UPC 900 Workstation     WARNING        The WARNING  sign highlights instructions that  must be followed to avoid personal injury  It is  important not to proceed until all stated conditions    are met and clearly understood     WARNING        Read the instruction to avoid hazardous  conditions     CAUTION        The Caution  sign highlights instructions that must  be followed to avoid damage to the product or  other equipment  It is important not to proceed  until all stated conditions are met and clearly       understood    Note  The Note sign is used to indicate information  important for trouble free and optimal use of the  product    CE Certifying       This product meets the requirements of applicable  CE directives  A copy of the corresponding Declaration of  Conformity is available on request     The CE symbol and corresponding declaration of  conformity  is valid for the instrument when it is       used as a stand alone unit  or      connected to other CE marked GE Healthcare  instruments  or      connected to other products recommended or  described in this manual  and      used in the same state as it was delivered from GE  Healthcare except for alterations described in this  manual     Recycling       This symbol indicates that the waste of electrical  X and e
25. a slope of 95 102  and an asymmetry  potential within  30 mV  As the electrode ages  the slope decreases and the  asymmetry potential increases     As a rule  when an electrode has an asymmetry potential outside of  60  mV and a slope lower than 80   and no improvement can be achieved by  cleaning  it should be replaced     An electrode is still usable at lower slopes and higher asymmetry potentials  but the response will be slower and the accuracy diminished     Calibrating with the electrode in the flow cell    When calibrating with the electrode fitted in the flow cell in AKTA design   follow the above procedure but let at least 30 35 ml  with 2 ml mixer  of  standard buffer solution be pumped through the system to stabilize pH   Leave the pump running while calibrating  Switch to the second standard  buffer solution and repeat the procedure  For a description of calibration  from UNICORN with the electrode fitted in the flow cell  see Chapter 6 in the  UNICORN User Manual     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Operation                                     mi  3 11 Filtering noise in the UV signal  To filter the noise in the UV signal  a moving average filter is used  The  averaging time is the time interval used for calculating the moving  average of the absorbance signal  A long averaging time will smooth out  noise efficiently  but it will also distort the peaks  Peaks narrower than the  minimum peak width value according to the table below may be dist
26. aling and pH scaling  see section 3 8 Setting  the Cond analogue output and section B 1 4 Checking recorder     No text on the front 1 Check that the mains cable is connected and the power switch is  display in ON position     30 Monitor UPC 900 User Manual 18 1125 55 Edition AG    Message    Trouble shooting    5 4 Error messages  If the suggested actions do not correct the fault  call GE Healthcare     Action          34 Start up failed  Retry Call service          35 WARNING wrong  averaging time set             50 Electrical error  Call for service               57 Electrical error  Call for service          75 Electrical error  Call for service             70 Lamp disconnected  If not call service          71 WARNING low light  intensity             72 Change lamp or  call for service          76 Change lamp or  call for service             73 WARNING Too much  straylight leaks in                77 WARNING Autozero  Out of range             1 Perform a new start up  The preceding message may tell more about  the cause     2 If not  call service     1 Wrong value for averaging time set  See section 3 11 on how to set the  averaging time relative to peak width     1 Call for Service     1 Connect the lamp  or call for service     1 Check connection between optical unit and monitor   2 Check that lamp and filter position correspond   3 Change lamp  If warning persists  call for service     1 If used in cold room  additional warm up might be needed   2 If problem remains  ch
27. amp Run Time Setup Lamp Run Time  Hg 2a00bi Hg  2000h  3000  Setup Show UV  lon  Setup Show UV   on  on off  EEE    Monitor UPC 900 User Manual 18 1125 55 Edition AG 49    ED Reference information    50 Monitor UPC 900 User Manual 18 1125 55 Edition AG    C Technical specifications    Reference information       The full specifications apply only after at least 1 hour warm up     C 1 Operating data  UV measurement    Absorbance range  Autozero range  Baseline adjust    Wavelengths   Hg lamp  fixed  by changing filter  Zn lamp    Linearity  5 mm cell    Static noise  short term    long term  Static drift    Flow sensitivity    UV flow cell  2mm   Flow rate   Max  pressure  Back pressure   Liquid temperature range  Optical pathlength   Cell volume   UV flow cell  5 mm   Flow rate   Max  pressure  Back pressure   Liquid temperature range  Optical pathlength    Cell volume    Monitor UPC 900 User Manual 18 1125 55 Edition AG    0 01 5 AU  full scale    0 2 2 0 AU  Adjustable 0 100  of full scale    254 and 280  313  365  405  436 and 546 nm  214 nm     lt 3  up to 2 AU at 254 nm   lt 5  up to 1 AU at 280 nm    40x10 6 AU at 254 nm   typically 6x10 6 AU at 254 nm   40x10 6 AU at 254 nm     100x10 6 AU h at 254 nm  2x10 4 AU min ml    0 100 ml min   4 MPa  40 bar  580 psi    Max  0 05 MPa at 100 ml min   4 to  60   C   2mm    2 ul  30 ul detector volume     0 20 ml min   4 MPa  40 bar  580 psi    Max  0 02 MPa at 20 ml min   4 to  60   C   5mm    6 ul  10 ul detector volume     51 
28. ange the lamp   3 If problem remains  call for service     1 Check that filter wheel cover is closed    2 Check that non transparent tubings are used at UV flow cell inlet and  outlet    3 Check that optical unit is not exposed to direct sunlight   If problem remains  call service    1 Auto is not allowed from a level above 2 AU   Check buffers   3 Clean UV cell     Monitor UPC 900 User Manual 18 1125 55 Edition AG 31    Trouble shooting       83 WARNING temp_ca  will be changed             84 WARNING cond_cal  will be changed             85 WARNING condsca   0 100   lt 0 1mS    oO                86 WARNING cond_cell  bad not connected          87 WARNING pH probe  bad not connected          88 Electrical error  Call for service             89 Electrical error  Call for service             90 ATTENTION  set lt  0mV first             91 WARNING  bad pH ad value             92 WARNING electrode  slope  lt 70 or  gt 110              93 pH_cal failed  check electrode             94 WARNING lt 1pH unit  between cal_buff 1 amp 2             95 Temp cal failed  check cond cell                97 WARNING pH scale   0 100   lt 1pH unit             32    Press OK to accept the change   Press ESC to skip the change     The difference between 0 and 100  must be at least 0 1 mS cm   Increase the span between zero and full scale setting  See section 3 8     Check that the conductivity cell is connected   Recalibrate temperature   If problem remains  replace the conductivity cell     Check t
29. ar panel     Ca Me Ca Gs sms  zum ns mi es ns     0 aus  aus au     aus aus aus ame  am   am   am am       Optical  Ancogue  Conductivity oO ni outos ni  Flow Cell Unit it 0 1 UniNet 1    O7  0600             Connecting to chart recorder  if used     The external chart recorder outputs for UV  pH and conductivity from the  monitor are 0 1 V     1 Connect the chart recorder to the mini DIN socket Analogue out 0 1 V  using the cable supplied  Pin designations for the signals are as follows        Pin no  Signal Range Value  full scale    1 UV 0 1V 0 001 5 au   2 signal ground OV     3 conductivity 0 1V 0 1 mS cm 999 9 mS cm  4 signal ground OV     5 pH 0 1V 0 14   6 signal ground OV         Note  The signal cable is delivered with protective covers on each wire  Do  not remove the protective covers from unused connections as a short  circuit may disturb the measurements     2 Set the recorder to 0 1 V input  full scale     2 7 Connecting to a communication link       CAUTION  The mains power to AKTA design must be switched OFF before  connecting the module to the UniNet 1 link              The monitor can be used as a stand alone module or in AKTA design  As a  stand alone module  it is controlled from the front dial and display    In AKTA design  the monitor is controlled from a PC running UNICORN     version 3 0 or higher via UniNet 1 cables  or manually from the front dial and  display     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Installation                A   C
30. ars when the reading is stable and the following text is  then shown    Calibr pH Buffer 1  7 Adjust the pH value in the display using the dial so that it corresponds to  the known pH value of the first buffer solution and press OK    Calibr pH Buffer 2 8 Turn the dial clockwise to access the buffer 2 calibrating menu  Rinse  the electrode tip with distilled water and then immerse the electrode in  the second buffer solution  e g  pH 4 0 or 9 0  and press OK    A    Monitor UPC 900 User Manual 18 1125 55 Edition AG 17    Operation       Calibr pH Buffer 2             18    9 This text disappears when the reading is stable and the following text is  hen shown   Calibr pH Buffer 2             10 Adjust the pH value in the display using the dial so that it corresponds to  he known pH value of the second buffer solution and press OK     11 When the two point calibration is performed successfully  the following  ext is shown for one second   Calibration OK             12 From the Calibrate pH Buffer 2 menu  turn the dial one step clockwise  o access sub menu Calibrated Electrode Slope  This menu shows the  slope of the calibration curve  where 100  corresponds to 59 16 mV per  pH step at 25   C  The asymmetry potential at pH 7 is shown as a mV  value  Press ESC to return to the Set Parameters menu        Voltage  mV      500 mV    Asymmetry  potential  at pH 7        0 mv       500 mV       13 Before use  rinse the electrode using distilled water     A new electrode typically has 
31. be air in the flow cell  Use a flow restrictor after the flow cell  and check that the flow restrictor gives a back pressure of 0 2 MPa   Check for leaking tubing connections    Check that the column is equilibrated  If necessary  clean the column   Check the operation of the mixer and the pump    Clean the flow cell according to the procedure in section 4 5 or 4 6     ne wem       pa    Clean the flow cell according to the procedure in section 4 5 or 4 6    2 The ambient temperature may have changed  Use a temperature  compensation factor  see section B 2 4 Setting up conductivity to change  temperature compensation  in Reference information     1 Turn the flow cell so the end with the screws is facing the pH flow cell   Recalibrate the conductivity cell    3 Calibration solution  1 00 M NaCl  not correctly prepared  Prepare a new  calibration solution and recalibrate the conductivity cell     Monitor UPC 900 User Manual 18 1125 55 Edition AG 29    Trouble shooting    FF a   Fault Action       Ghost peaks appear 1 Acharged sample has been detected  e g  protein     in the gradient profile 2 Air bubbles are passing through the flow cell  Check for loose tubing  connections  If necessary  use a flow restrictor after the conductivity  flow cell     Other problems    Error in external pH or 1 Check the chart recorder in accordance with its manual   Cond  chart recorder 2 Test the recorder function and input voltage  which should be 1 V  full scale     3 Check the conductivity sc
32. ce temperature  39  B 2 6 Setup adjust conductivity    39  B 2 7 Setup adjust cell constant ec cecccssccseeeeeeeee 41  B 2 8 Setup show conductivity    41  B 2 9 Setup pH temperature compensation             42  B 2 10 Setup show pH    42  B 2 11 Setup UV averaging filter constant                   42  B 2 12 Setup lamp run time    43  B 2 13 Setup show UV    43  B 2 14 Setup language    43  B 2 15 Setup unit nUMbEr     44  B 2 16 Setup display angle    44  B 3  Setting and using the alarm timer    45  B 4 Service displays  B 5 Menu overvieW     c sssssssssssssecccccennsssssescscegceensssssesceeegeennnses  Technical SPECIFICATIONS        essssssesscsscscssssssssesssssccssssssnueesescenssssnneesees 48  CL    Operating AO nn 48  C2  Physical datant 49  Accessories and Consumables  0    eeccssssssssssssssssssssnnsseetsssesseeeeeeeees 51  Sh  rtinstr  cti  ns neeeunaikannie a 57    About this manual    This manual comprises two parts  a practical part  sections 1   5     and a reference part  sections A   D      Sections 1   5 contain the necessary information for operating the    instrument     Monitor UPC 900 User Manual 18 1125 55 Edition AG    VII    Introduction    1 Introduction       1 1 General    Monitor UPC 900 is a high precision on line monitor for the combined  measurement of UV absorption  pH and conductivity in liquid  chromatography  The UPC 900 features     e Fixed wavelengths of 254  280 nm  Hg lamp   or 214 nm  Zn lamp      e Other wavelengths of 313  365  405 
33. e  under US pat 5 284 933 and US pat 5 310 663  including corresponding foreign  patenst  assigne  Hoffman La Roche  Inc      All third party trademarks are the property of their respective owners        1998  2010 General Electric Company   All rights reserved   First published Apr  1998    All goods and services are sold subject to the terms and conditions of  sale of the company within GE Healthcare which supplies them  A copy of  these terms and conditions is available on request  Contact your local GE  Healthcare representative for the most current information     GE Healthcare GmbH  Munzinger Strasse 5  D 79111 Freiburg  Germany    GE Healthcare UK Limited  Amersham Place  Little Chalfont  Buckinghamshire  HP7 9NA  UK    GE Healthcare Bio Sciences Corp  800 Centennial Avenue  P O  Box 1327  Piscataway  NJ 08855 1327  USA       GE Healthcare Japan Corporation  Sanken Bldg  3 25 1  Hyakunincho  Shinjuku ku  Tokyo 169 0073  Japan    18 1125 55 AG 10 2010    
34. een pH and the output signal from the pH electrode is  temperature dependent  For accurate measurements during temperature  changes  the pH measurement can be temperature compensated  In normal  applications  when the temperatures of the buffers and calibration buffers  are identical  temperature compensation is not necessary     When using temperature compensation  it is important that the temperature  of the pH electrode is the same as that of the conductivity flow cell since  that is where the temperature is measured     1 Select sub menu Setup pH and press OK     2 Select sub menu Setup pH Temp Comp and press OK  The current status  for showing pH is displayed  If on is shown  Te is displayed in main  menus 1 and 2  If off is shown  default   Te is not displayed in the main  menus     3 Change the setting if desired and press OK     B 2 10 Setup show pH    Normally the pH is displayed in the main operating menus  see section 3 13  Reading pH  temperature and conductivity values   If not required  the pH  display can be set to off     1 Select sub menu Setup pH and press OK     2 Select sub menu Setup Show pH and press OK  The current status for  showing pH is displayed  If on is shown  current pH is displayed in main  menus 1 and 2  If off is shown  no pH is displayed in the main menus     3 Change the setting if desired and press OK     B 2 11 Setup UV averaging filter constant  1 Select sub menu Setup UV and press OK     2 Select sub menu Set Averaging and press OK  The
35. ence information   3    B 2 7 Setup adjust cell constant    After replacing the flow cell  the cell constant has to be set   The cell  constant is shown on the packaging      1    Select sub menu Setup Cond and press OK     Select sub menu Setup Adjust Cell Const and press OK  The current cell  constant is shown     A warning message is shown until confirmed by pressing OK    The current cell constant value is displayed as default  Enter the new cell  constant as read from the packaging and press OK  The range is  0 1 300 0 cm 1     When entering the cell constant from UNICORN  select System  Control System Calibrate and select Cond_Cell     B 2 8 Setup show conductivity    1    Select sub menu Setup Cond and press OK     Select sub menu Setup Show Cond and press OK  The current status for  showing conductivity is displayed  If on is shown  current conductivity is  displayed in main menus 1 and 2  If off is shown  no conductivity is  displayed in the main menus     Change the setting if desired and press OK     Monitor UPC 900 User Manual 18 1125 55 Edition AG 43       Setup  Temp Cond pH UV          Setup pH Temp Comp   off           Setup pH Temp Comp   off  on off                Setup  Temp Cond pH UV          Setup Show pH   on           Setup Show pH  on  on off          Setup  Temp Cond pH UV          Set Averaging   1 35  13          Set Averaging   1 35  0 64             44    ED Reference information    B 2 9 Setup pH temperature compensation    The relationship betw
36. escribed in the User Manual    e connected to other CE labeled GE Healthcare  modules or other products as recommended     This product meets the requirement of the  Low Voltage Directive  LVD  73 23 EEC through  the following harmonized standards     e EN 61010 1   e IEC 61010 1   e CAN CSA C22 2 No  61010 1  e UL61010 1    This device meets the requirements of the  EMC Directive 89 336 EEC through the following  harmonized standards     e EN 61326  emission and immunity   e EN 55011  GR 2  Class A  emission     e This device complies with part 15 of the FCC  rules  emission   Operation is subject to the  following two conditions     1 This device may not cause harmful  interference        2 This device must accept any interference  received  including interference that may  cause undesired operation     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Reference information Cr    D Accessories and spare parts                                           Item Quant  pack A C  Code no   Hg optics with 254  280 nm filters 1  A 18 1128 20  Zn optics with 214 nm filter 1 A 18 1128 21  Hg lamp  amp  housing complete 1 C 18 1128 22  Zn lamp  amp  housing complete 1     18 1128 23  UV flow cell 5 mm 1    18 1128 24  UV flow cell 2 mm 1    18 1128 25  Filter 214 nm 1 C 18 0622 01  Filter 254 nm 1 C 8 0620 01  Filter 280 nm 1 C 8 0621 01  Filter 313 nm 1 C 18 0623 01  Filter 365 nm 1 C 18 0624 01  Filter 405 nm 1 C 8 0625 01  Filter 436 nm 1 C 18 0626 01  Filter 546 nm 1 C 18 0627 01
37. evel value  Press OK to acknowledge  Press ESC  twice to return to menu Set Parameters     The module has an automatic over range function  If the monitor signal  reaches the full range value during a peak  the signal will drop instantly to  0 V and give an accurate display of the peak starting from this position     3 9 Calibrating conductivity    The cell constant for the particular flow cell is written on the flow cell  packaging  Refer to section B 2 in Reference information for how to enter the  cell constant     Adjustment of the cell constant is only necessary when the monitor is to  be used to determine conductivity with high accuracy  The procedure is  described in Reference information section B 2  Calibration can also be  performed from UNICORN     Select System Control System Calibrate     3 10 Calibrating the pH electrode    A good laboratory routine is to calibrate the module once a day  when   the electrode is replaced and if the ambient temperature changes    The pH electrode is calibrated using standard buffer solutions in a two  point calibration  The two buffer solutions can have any pH value as   long as the difference between them is at least 1 pH unit  Calibration   can also be performed from UNICORN  In UNICORN  select System  Control System Calibrate  Select the pH monitor  The calibration procedure  can be done with the pH electrode either fitted in or removed from the flow  cell     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Operation      
38. he pH electrode connection   Clean the pH electrode   If problem remains  change the pH electrode     Factory calibration for pH is lost  The monitor can still be used but may  not meet specifications for pH measurement   Call for service     Factory calibration for conductivity is lost  The monitor can still be used  but may not meet specifications for conductivity measurement   Call for service     Only visible to service personnel     Electrode slope is out of range  Check buffers and recalibrate   Clean the pH electrode and recalibrate   If the message remains  replace the pH electrode     The difference between the pH of the buffers used during calibration  must be at least 1 pH unit     Check that the conductivity cell is connected  Recalibrate    The measured temperature value differs from the reference value by  more than  5   C  or actual temperature is lower than  8   C  Recalibrate   If the message remains  replace the conductivity cell     The difference between the zero level and full scale must be at least 1 pH  unit  Increase the span between zero and full scale settings   See section 3 8     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Trouble shooting                   mi  98 Cal failed  Cell 1 Conductivity cell constant is out of range   Constantino 015300  2 Wrong solution used during calibration  Use 1 00 M NaCl  and recalibrate   3 Air in conductivity cell during calibration  Flush the flow cell  with calibration solution and recalibrate   4 Dirty
39. hile the monitor is running     Before changing a UV lamp  ensure that the lamp cable is disconnected  from the monitor to prevent injury to eyes    If the mercury lamp is broken  make sure that all mercury is removed and  disposed of according to national and local environmental regulations           WARNING  When using hazardous chemicals  all suitable protective  measures  such as protective glasses  must be taken           WARNING  HCI and NaOH are injurious to health  Avoid spillage           WARNING  Hg lamps contain small amounts of mercury and must  be handled with care and disposed of according to national or local  environmental regulations              WARNING  When using hazardous chemicals  take all suitable protective  measures  such as wearing protective glasses and gloves resistant to the  chemicals used  Follow local regulations and instructions for safe opera   tion and maintenance of the system        Monitor UPC 900 User Manual 18 1125 55 Edition AG          Installation    2 Installation       2 1 Unpacking    Unpack the module and check the items against the supplied packing  list  Inspect the items for obvious damage that may have occurred during  transportation     Note  It is important that the filters  flow cells and lamps are not handled  during unpacking  For protection of these items  they should remain in their  packing materials until required for use     We recommend that all packing materials be retained if onward transport of  the module is e
40. l on the filter wheel for the  filter to be used  A click will indicate that the filter is in position     Note  The wavelength and the flow cell type should also be entered in the  Questions menu in UNICORN     Filter change    The Hg optics with 254 and 280 nm filters and the Zn optics with the 214 nm  filter are delivered with filters installed  If other filters are to be used  install  the new filters as described in 2 10 Installing optical filters  optional      Monitor UPC 900 User Manual 18 1125 55 Edition AG 5    Installation    Protective cover           5 mm 2mm  flow cell flow cell    PET       Holder                Conductivity t          cell  e    Clamp   I    PET    Connection to the column    1 Fix the optical unit directly under the column if possible     Note  Always position the optical unit with the filter wheel cover facing  upwards     2a Connect the column outlet tubing directly onto the top  a  of the optical  unit for the 5 mm flow cell using a fingertight connector     2b Connect the column outlet tubing onto the bottom  b  of the optical unit  for the 2 mm flow cell using a fingertight connector     Note  The inlet port of the 5 mm UV cell is above the optical unit  The inlet  port of the 2 mm UV cell is below the optical unit     3 Screw to fingertightness     4 Connect the optical unit outlet tubing onto the opposite hole in the flow  cell  Use fingertight connectors   If no outlet tubing exists  cut a piece of PEEK tubing  i d  0 5 mm  o d   1 16
41. lectronic equipment must not be disposed  as unsorted municipal waste and must be  mmm collected separately  Please contact an authorized    representative of the manufacturer for information  concerning the decommissioning of equipment     WARNING        This is a Class A product  In a domestic environment this  product may cause radio interference in which case the  user may be required to take adequate measures        Contents          UV     Contents   1 Introduction  1 1 GRNCTO EE ne nenne 1  E E E a et A seate 2   2 Installation  21  Uppa ennen aaa 3  2 2 General precautiDns                       3  2 3 Installing the optical unit    4  2 4     Installing the conductivity flow cell    6  2 5 Installing the pH flow cell and electrode    6  2 6 Connecting electrical signal cables    8  2 7 CONNECTING to a communication link    8  2 8 Connecting to supply voltage    9  2 9     Preparing the module for use    9  2 10 Installing optical filters  optional     10   3 Operation  3 1  Onl  lteseseeaern ee 11  3 2 Menu selection and settings    11  3 3 Main menu OVEFVIEW re 13  3 4 Controlling the UV lampes 13  3 5 Reading the UV absorbance value    14  3 6  Autozeroing UV     3 7 Creating eventmarks  3 8 _ Setting Cond  pH and UV analogue outputs eee 14  3 9 Calibrating conductivity seen 16  3 10 Calibrating the pH electrode    16  3 11 Filtering noise in the UV signal    19  312  Ehanging   UVIOWCell seusseeee eat 19  3 13 Reading pH  temperature and conductivity values    20  
42. m the monitor to prevent injury to eyes    If the mercury lamp is broken  make sure that all mercury is removed and  disposed of according to national and local environmental regulations              Lamp housing end plate 1 Use a Philips screwdriver to detach the end plate by removing one and  loosening the other of the two holding screws on the lamp housing to be  removed     2 Slide the lamp housing off the filter housing        3 Detach the end plate  as in step 1 above  from the lamp housing to be  fitted to the optical unit     4 Slide the lamp housing onto the filter housing  The lamp and signal  cables should be on the same side  As you slide the lamp housing into  position  depress the two pressure pads on the filter housing in sequence  to facilitate the installation     5 Refit the lamp housing end plate     Dot on lamp housing           6 Slide the lamp housing firmly into place  There will be a faint click when  the housing is positioned correctly  The Hg lamp housing can take up  two positions  one for 280 nm  marked g by on the filter housing  and   N Symbols on the other marked by for    all other wavelengths  The Zn lamp housing   filter housing has only one position                      es 7 Set the wavelength to be used by selecting lamp position  indicated by    below lid a dot on the lamp housing  in combination with the appropriate filter  i e   the dot on the lamp housing should be adjacent to the symbol on the  filter housing corresponding to the symbo
43. main operating menu 1 shows UV absorption  conductivity in  percentage of full range and pH  If the pH value is not stable or is changing   an asterisk is displayed  i e  pH   This menu is reached from any other menu  by pressing ESC     If temperature compensation is switched on  the display will show Cond Tc  instead of Cond   see section B 2 4 Setup conductivity temperature  compensation in Reference information     By turning the dial three steps clockwise  main operating menu 2 is shown  instead  This display shows pH  temperature  in the conductivity flow cell   and the actual conductivity in mS cm together with the percentage value  If  temperature compensation is switched on  Te is shown in the display     The display of pH  temperature and conductivity can be disabled  see  Section B 2 of Reference information     3 14 Storage and shut down  Storage of the UV flow cell       CAUTION  Do not allow solutions which contain dissolved salts  proteins  or other solid solutes to dry out in the flow cell  Do not allow particles to  enter the flow cell as damage to the flow cell may occur              Overnight  The flow cell can be left filled with buffer     Weekend and long term storage  Flush the flow cell with distilled water and  then fill it with 20  ethanol     The flow cell can also be stored dry by flushing as above with distilled water  and then blowing a compressed inert gas such as nitrogen  N   through the  cell  Replace the red protective caps  Never use compre
44. materials  UV flow cells  cond  flow cell   pH electrode  flow cell   dummy electrode    pH stability range    Typical lamp lifetime  Hg  254 nm    room temperature  cold room   Hg  280 nm    room temperature  Zn  214 nm    room temperature    Monitor UPC 900 User Manual 18 1125 55 Edition AG    Reference information    Stand alone or from a PC with UNICORN version  3 0 or higher  via UniNet 1 connection     IP 20  IP 21  flow cells IP 44    100 240 V AC  50 60 Hz  45 VA    Languages available  English  German  Spanish   French  Italian    1 5 m  AMP 5 2 pole connector  1 5 m  RJ 45 connector   1 5 m  BNC connector   1 5 m  D sub 9 pole connector  Fingertight connector  1 16     0 1 V full scale   2 rows with 20 characters each  100 x 260 x 370 mm   8 5 kg    Quartz  ETFE  titanium  Titanium  CTFE   Glass  FFKM  perfluororubber   titanium   PTFE  polytetrafluoroethylene     1 13  1 14   lt 1 day exposure     7000 hours  2000 hours    3500 hours    2000 hours    53    Reference information    Chemical resistance    Compliance  standards    Safety standards    EMC standards    54    The wetted parts are resistant to organic  solvents and salt buffers commonly used in  chromatography of biomolecules  except  100  ethylacetate  100  hexane  and   100   tetrahydrofuran  THF      The declaration of conformity is valid for with  the instrument only if it is     e used in laboratory locations    e used in the same state as it was delivered  from GE Healthcare except or alterations  d
45. n 90  Deconex 11  RBS 25 or  equivalent  and continue to squirt five more times     2 After five squirts  leave the detergent solution in the flow cell for at least  20 minutes     3 Pump the remaining detergent solution through the flow cell     4 Rinse the syringe and then flush the flow cell with distilled water  10 ml      24 Monitor UPC 900 User Manual 18 1125 55 Edition AG    Maintenance    4 5 Cleaning the conductivity flow cell in place    ee  WARNING  NaOH is injurious to health  Avoid spillage              WARNING  When using hazardous chemicals  take all suitable protective  measures  such as wearing protective glasses and gloves resistant to  the chemicals used  Follow local regulations and instructions for safe  operation and maintenance of the system              Remove the pH electrode and install the dummy electrode in the pH flow  cell     Pump a cleaning or sanitising agent through the flow cell  The standard   recommendation is to pump 1 M NaOH for 30 minutes and then wash out   with buffer    4 6 Cleaning the conductivity flow cell off line  WARNING  NaOH is injurious to health  Avoid spillage              WARNING  When using hazardous chemicals  take all suitable protective  measures  such as wearing protective glasses and gloves resistant to  the chemicals used  Follow local regulations and instructions for safe  operation and maintenance of the system      gt  gt              If the conductivity measurements are not comparable to previous results    the
46. nce          WARNING  Always disconnect the power supply before attempting to  replace any item on the module during maintenance           CAUTION  Only spare parts approved or supplied by GE Healthcare may  be used for maintaining and servicing the module              4 1 Periodic maintenance       Interval Action  see procedures below   Every 3 months Check the monitor   Every 6 months Change the pH electrode   or more often if required Clean the UV flow cell   When required Clean the conductivity cell    Clean the pH electrode       4 2 Cleaning and checking the module    Wipe the monitor housing regularly with a damp cloth  Let the monitor dry  completely before use     Lamp intensity       Check Lamp Intensity 1 Select menu Check and press OK   0 00002 015 0 4 02             2 Select menu Check Lamp Intensity     If    R  lt 300 mV for 254 nm  R  lt 150 mV for 280 nm  R  lt 150 mV for 214 nm     replace the lamp according to Section 4 10 Changing the UV lamp  or  contact GE Healthcare for lamp replacement     Lamp run time       Check Lamp Run Time 1 Select menu Check and press OK   Hg 2300h Zn 340h          2 Select menu Check Lamp Run Time        e The lifetime of a Hg lamp at 254 nm  in room temperature is typically  7000 hours  in cold room  typically 2000h      e The lifetime of a Hg lamp at 280 nm  in room temperature is typically  3500 hours     e The lifetime of a Zn lamp is typically 2000 hours in room temperature     Monitor UPC 900 User Manual 18 1125 55 Editi
47. ncerning the decommissioning of  SS your equipment              26 Monitor UPC 900 User Manual 18 1125 55 Edition AG    Trouble shooting    5 Trouble shooting          UPC 900  Version 1 00             IN    UV measurement    No text on the front  display    Noisy UV signaldrift  or instability    Ghost peaks    5 1 General    When contacting GE Healthcare for support  state the program version of  the module  which is shown for 2 seconds during start up        WARNING  The module must not be opened by the user  It contains high  voltage circuits that can deliver a lethal electric shock        5 2 Recycling       This symbol indicates that the waste of electrical and electronic  equipment must not be disposed as unsorted municipal waste and must  be collected separately  Please contact an authorized representative of  the manufacturer for information concerning the decommissioning of  your equipment              5 3 Faults and actions  If the suggested actions do not correct the fault  call GE Healthcare     1 Check that the mains cable is connected and the power  switch is in ON position I     1 Select menu Check Autozero to check the autozero absorbance value   If AZ is between 1 5 and 2  there may be air bubbles in the flow cell  or  the wrong buffer system in use    2 Wrong filter for the lamp used  Check that the lamp is in proper position  and that the correct filter is used    3 The buffer may be impure  Check if the signal is still noisy with water    There may be air
48. nd press OK  The current  temperature is shown   5 A warning message is shown until confirmed by pressing OK     6 The current adjustment value is displayed as default  Enter the  temperature shown on the thermometer and press OK     Monitor UPC 900 User Manual 18 1125 55 Edition AG       Setup  Temp_ Cond pH UV          Setup Show Temp   on           Setup Show Temp   on  on off          Setup  Temp Cond pH UV          Setup Cond Temp Comp   0 0               Setup Cond Temp Comp   0 0   0 0             Setup  Temp Cond pH UV          Setup Cond Ref Temp   25 0  C              Setup Cond Ref Temp   25 0   25 0          Reference information   3    B 2 3 Setup show temperature    The display of the temperature in the conductivity flow cell  shown in the  main operating menu 2  can be enabled or disabled     1 Select sub menu Setup Temp and press OK     2 Select sub menu Setup Show Temp and press OK  The current status for  showing temperature is displayed  If on is shown  current temperature is  displayed in main menu 1 and 2  If off is shown  no temperature is  displayed in the main menus     3 Change the setting if desired and press OK     B 2 4 Setup conductivity temperature compensation  1 Select sub menu Setup Cond and press OK     2 Select sub menu Setup Cond Temp Comp and press OK  The current  temperature compensation factor is shown  0 0  means that the  compensation is off  default setting   The range is 0 0 9 9      3 The current compensation factor is displayed as
49. nts  we recommend  that the pH electrode is removed and the dummy electrode inserted in  its place as organic solvents will cause pH electrode degeneration     Monitor UPC 900 User Manual 18 1125 55 Edition AG 9    Installation          10       2 10 Installing optical filters  optional     The Hg optics with 254 and 280 nm filters and the Zn optics with the 214 nm  filter are delivered with filters installed  If other filters are to be used  install  them as follows     1    If the Zn lamp is attached  remove the lamp housing as described in  section Changing the UV lamp     Remove the four screws in the filter housing  Separate the filter housing  from the detector housing     Carefully remove the filter wheel from the filter housing     If necessary  remove the filter s  from the filter wheel by pressing it them   out  e g  with a small screwdriver     Note  Filters are sensitive optical components  Never touch the optical    10    surfaces or expose them to temperatures above 60   C  Clean them  with dry lens cleaning tissue and store them  when not in use  in  the box in which they were supplied  Heavy contamination may be  removed by using a lens tissue dipped in ethanol     Insert the filter s  of choice into the filter wheel  maximum 3 filters  with  the correct orientation  the mirror side facing upwards  and position over  one of the three triangular apertures  The filters snap in by pressing  them quite firmly  Do not touch the filter surface     Remove the circula
50. on AG 23    Maintenance    When necessary  replace the lamp according to Section 4 10 Changing the  UV lamp  or contact GE Healthcare for lamp replacement     Autozero    The module internal absorbance value for autozero can be checked to test  the consistency of buffers     1 Select menu Check and press OK     2 Select menu Check Autozero  The autozero absorbance value for  the wavelength used is shown     4 3 Cleaning the UV flow cell in place       WARNING  NaOH is injurious to health  Avoid spillage           WARNING  When using hazardous chemicals  take all suitable protective  measures  such as wearing protective glasses and gloves resistant to  the chemicals used  Follow local regulations and instructions for safe  operation and maintenance of the system      gt  gt              Pump a cleaning or sanitising agent through the flow cell  The standard  recommendation is to pump 1 M NaOH for 30 minutes and then wash out  with buffer     4 4 Cleaning the UV flow cell off line    A clean flow cell is essential for ensuring the correct operation of the  UV monitor        CAUTION  Do not allow solutions that contain dissolved salts  proteins  or other solid solutes to dry out in the flow cell  Do not allow particles to  enter the flow cell as damage to the flow cell may occur              1 Connect a syringe to the inlet of the flow cell and squirt distilled water  through the cell in small amounts  Then fill the syringe with a 10   surface active detergent solution like Deco
51. on solution by pumping at  least 15 ml through the cell with a syringe    3 Stop the flow and wait 15 minutes until the temperature is constant in  the range 20 30   C    4 Read the conductivity value displayed and compare it with the  theoretical value from the graph below at the temperature of the  calibration solution  If the displayed value and the theoretical value  correspond  no further action is required    If the values differ  proceed with actions 5 8  Setup 5 Select sub menu Conductivity of 1 00 M NaCl at 20 30   C  ERP cen RHEIN Setup Cond and  press OK              97       Setup Adjust Cond 6 Select sub menu   1 000mS cm  y  Setup Adjust Cond 95  and press OK  The  current conductivity  value is shown                 Warning  This will 7 A warning message   change cell calib is shown until  confirmed by 90  pressing OK                 8 The current value  Setup Adjust Cond es   83 53mS cm  86 6 is displayed as  default  Enter the  theoretical  conductivity value  according to the  graph and press OK   The new cell  constant is  automatically  calculated  The  range is  1 000 999 9 mS cm              85    80    77       42 Monitor UPC 900 User Manual 18 1125 55 Edition AG       Setup  Temp Cond pH UV          Setup Adj Cell Const   83 56cm  4           Warning  This will  change cell calib          Setup Adj Cell Const   83 56 1  866                Setup  Temp Cond pH UV          Setup Show Cond   on                 Setup Show Cond   on  on off             Refer
52. onnect two UniNet 1 cables to the UniNet 1 connectors  The module can be  connected in series anywhere between the PC and a termination plug  The  UniNet 1 link connects  in series  the PC with Pump P 920  Monitor  UPC 900 and the Frac 920  The termination plug is connected to the last  module in the chain   2 8 Connecting to supply voltage   A WARNING  The module must be connected to a grounded mains socket     1 Make sure the on off switch is in the OFF position  0      2 Connect the supplied mains cable between the module and a grounded  mains socket  Any voltage from 100 240 V AC  50 60 Hz can be used     Note  The module contains no user replaceable fuse     2 9  Preparing the module for use    Before performing the following procedures  we recommend you read  sections 3 1 3 3     e Calibrate the pH electrode  see section 3 6     e Set the conductivity cell constant  see section B 2 3    Only required when the flow cell has been replaced      e Calibrate the temperature sensor  see section B 2 6    Only required if the monitor is to be used for high accuracy  measurements  or if the flow cell has been replaced      Note  The conductivity cell constant is shown on the packaging  Retain  the packaging in case the conductivity cell constant needs to be  re entered     Note  The measured temperature is the temperature in the conductivity flow  cell  which can differ from the ambient temperature in AKTA design  systems     Note  When running chromatography using organic solve
53. orted   Because of this  the averaging time should be as short as possible  On  delivery  the averaging time is set to 1 3 s   Setup 1 From main menu 1  turn the dial six steps clockwise and press OK twice   Setup 2 Select Setup UV and press OK   Set Averaging 3 The menu Set Averaging is displayed  showing the current set averaging  eal time  Press OK   Set Averaging 4 Set the desired value and press OK  Values allowed are 0 02  0 04  0 08    13s  ge 0 16  0 32  0 64  1 3  2 6  5 and 10s   Averaging time  s  Corresponding Min  peak width  time constant  s  at half height  s    approximately   10 0 5 00 50 0  5 10 2 00 32 0  2 60 1 00 16 0  1 30 0 50 8 0  0 64 0 20 3 2  0 32 0 10 1 6  0 16 0 05 0 8  0 08 0 03 05  0 04 0 01 0 2  0 02 0 01 0 1       In UNICORN  the averaging time is set with the instruction AveragingTime in  System Control Manual Alarm amp Mon    3 12 Changing UV flow cell    The flow cell can be changed when required  for example  from 2 mm to 5  mm when the sensitivity of the measurement must be increased due to a  small amount of sample being applied  or from 5 mm to 2 mm when a lower  sensitivity is desired  due to output signal limitation     See section 2 3 Installing the optical unit  sub section Flow cell installation     Monitor UPC 900 User Manual 18 1125 55 Edition AG 19    Operation       AU Cond Tc pH   0 00002 015 0 4 02                pH 12 50 22 4  735 8 mS cm Tc78 8              20    3 13 Reading pH  temperature and conductivity  values    The 
54. r plastic band showing the wavelength s    Remove labels from the band if necessary     Place the correct labels on the band with the label designation facing  outwards  Ensure that the label position corresponds to the filter  position  i e  the label should be placed opposite the filter     Reassemble the circular plastic band with the filter wheel peg fitting into  the band notch     Check that all filters are clean  Insert the filter wheel back into the filter  housing     Note  The filter wheel can be placed only in correct position     11    Reassemble the filter housing with the detector housing by fastening the  four screws     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Operation    3 Operation          3 1 On off  Selftest Switch on the module at the mains switch on the rear panel  At switch on   Please wait    the module performs a selftest  Several beeps are heard during this process        If an error is detected  an error message is shown        UPC 900    version hos Name and software version number is shown for 2 seconds           Salibrating All parameters are set to factory default values           Calibration ok          AH  Gondi  IPM The selftest takes approx  1 minute  When start up is completed with no    errors  the display shows the main menu 1              The monitor can be used immediately but the full specifications are not  obtained until after 1 hour of lamp warm up     3 2 Menu selection and settings    Menu selection    A specific men
55. response differs greatly  the electrode may require  cleaning or replacement     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Fault    Trouble shooting    Action       pH values vary with  varied back pressure    10 There may be interference from static fields  Connect the pH flow cell  and the rear panel of the monitor using a standard laboratory 4 mm     banana plug    cable    11 Check that the pH electrode has been calibrated at the correct  temperature    12 In organic solvents such as ethanol  methanol and acetonitrile  stable  pH measurements are not possible since dehydration of the membrane  will occur  We recommend that the pH electrode is not used in  applications using organic solvents  Mount the dummy electrode instead    13 Clogged liquid junction  Refer to section 4 8 Cleaning the pH electrode     1 Replace the pH electrode     Conductivity measurement    Incorrect or unstable  reading    Baseline drift or noisy  signal    Conductivity  measurement with the  same buffer appears  over time    Absolute conductivity  value wrong    1 Check that the conductivity flow cell cable is connected properly to the  rear ofthe module    2 Check that the pump operates correctly    3 If temperature compensation is being used  check that the temperature  sensor is calibrated  and that the correct temperature compensation  factor is in use    4 Check that the column is equilibrated  If necessary  clean the column    5 Check the operation of the mixer        1 There may 
56. rm time and the count down timer   The default or current value is shown in parentheses     1 Select main menu Alarm Timer  The display shows current time     2 Press OK  Turn the dial one step clockwise to select sub menu Set Alarm  if you want to set an alarm at a fixed time  Press OK to enter the time in  the form HH MM SS and press OK again after entering each time unit     3 If you want to set a count down time  turn the dial one step further to  select sub menu Set Timer  Press OK to enter the count down value in  the form HH MM SS and press OK again after entering each time unit        4 Press ESC to return to the Alarm Timer menu  which now shows the set  alarm time or count down time as BzzHH MM SS     5 When the alarm time is due or the count down timer reaches 00 00 00   an alert display is shown and the module beeps for 30 s  or until OK is  pressed  The display shows the time elapsed since the alarm and the  current time  A second alert display is shown until OK is pressed     The alarm timer is based on the internal module clock  which can be set in  the Set Clock menu located after the Set Timer menu  The clock will be reset  when power is turned off     A set alarm timer function can be reset by pressing OK in the menu Alarm   Timer off     B 4 Service displays    The module has service displays for use by authorised service personnel  If  the service display Insert Access Code  is accidentally selected  press ESC to  return to the normal operation display    
57. s  and the resulting current is measured and used to calculate the conductivity  of the eluent  The monitor controls the AC frequency and increases it with  increasing conductivity between 50 Hz and 50 kHz giving maximum linearity  and true conductivity values     The conductivity is automatically calculated by multiplying the measured  conductance by the flow cell s cell constant  The cell constant is pre   calibrated on delivery but can be measured with a separate calibration  procedure  This procedure is described in Reference information section B 2     One of the electrodes has a small temperature sensor for measuring the  temperature of the eluent in the flow cell  Temperature variations influence  the conductivity and  in some applications when highly precise conductivity  values are required  it is possible to program a temperature compensation  factor that recalculates the conductivity to a set reference temperature        Temperature sensor    Eluent     gt     Monitor UPC 900 User Manual 18 1125 55 Edition AG 37    Reference information    38 Monitor UPC 900 User Manual 18 1125 55 Edition AG    Reference information   3             B Menus  B 1 Check menus  B 1 1 Checking autozero level  The module internal absorbance value for autozero can be checked to test  the consistency of buffers   Check Autozero 1 Select menu Check and press OK     AZ 0 00006 AU       2 Select sub menu Check Autozero  The autozero absorbance value       for the used wavelength is shown     
58. so for this wavelength  the light is converted at a fluorescent  surface before it passes the filter  On the lamp housing  there is a special exit  for 280 nm light  which means that the lamp position needs to be changed  when working with this wavelength     For 214 nm wavelength  a Zn lamp is used  This lamp is larger than the Hg  lamp and is therefore mounted in a larger lamp housing     The lamp connectors are keyed to inform the monitor software of which  lamp type is connected     Monitor UPC 900 User Manual 18 1125 55 Edition AG          KES    Reference information A          A 3 pH electrode 9   The pH electrode is of the sealed combination Sealing  double junction type  It contains a sealed Ag AgCI SS prige conania  reference which cannot be refilled  an internal INS with Agel    electrolyte bridge of 4 M KCI saturated with Ag  LL ru  AgCI  an outer electrolyte bridge of 1 M KNO   I      electrode    Internal annular                                           He nal annul  an annular ceramic reference junction and a chi San  low profile pH membrane  The pH electrode is       me  delivered with a transparent cover  om   4 Glass Secale j      containing diluted   The flow cell should not be used with any other ee KCI       pH electrode     External annular  coaxial ceramic  reference junction    A 4 Conductivity cell    The flow cell has two cylindrical titanium electrodes positioned in the flow  path of the cell  An alternating voltage is applied between the electrode
59. ssed air as this may  contain droplets of oil     Storage of the conductivity flow cell  Overnight  The conductivity cell can be left filled with a buffer     Weekend or long term storage  Flush the conductivity cell with water and  fill with 20  ethanol     Monitor UPC 900 User Manual 18 1125 55 Edition AG    Operation    Storage of the pH electrode       CAUTION  Never leave the pH electrode in the flow cell for any period  of time when the system is not used  since this may cause the glass  membrane of the electrode to dry out  Dismount the pH electrode from  the flow cell and fit the end cover filled with a 1 1 mixture of pH 4 buffer  and 2 M KNO      Do NOT store in water only              The pH electrode should always be stored in a 1 1 mixture of pH 4 buffer  and 2 M KNO  when not in use  When the pH electrode is removed from the  flow cell  the dummy electrode  supplied  can be inserted in the flow path     Electrode regeneration  If the electrode has dried out  immerse the lower  end of the electrode in buffer with a 1 1 mixture of pH 4 buffer and 2 M KNO   overnight        3 15 Restart after power failure    If the power supply to the module is interrupted  it automatically restarts  when power is restored and displays the main operating menu  All set values  are retained in the module and the lamp is switched on     Monitor UPC 900 User Manual 18 1125 55 Edition AG 21    Operation    22 Monitor UPC 900 User Manual 18 1125 55 Edition AG    Maintenance    4 Maintena
60. u is selected by turning the front selection dial clockwise or  counter clockwise  When the required menu is visible  the menu or selection  is accepted by pressing the OK button     Menu selection OK button ESC button  Go down one menu level  Return one menu level  or accept setting       If a menu has sub levels  the sub menu is displayed by pressing OK   Pressing ESC moves back one menu level     Monitor UPC 900 User Manual 18 1125 55 Edition AG 11    Operation    12                     Main menus Sub menus  3 click cw ESC or 3 click  with the dial     _JECw with the dial  OK  gt   submenu2 1    ok  gt   Sub menu 2 1 1   lt  ESC    lt  ESC  Sub menu 2 2  OK  gt   Sub menu 2 3 1  Sub menu 2 3 ESC    Sub menu 2 3 2    Return to main menu    Pressing ESC repeatedly always returns to the main menu 1  which is the  main operating menu     Main menu 1             353       ESC ESC ESC ESC                   Main menu 2    Select value    A cursor below a text or numerical value shows what is affected by the dial   To increase a numerical value  turn the dial clockwise  To decrease the value   turn the dial counter clockwise     To simplify entering large numerical values  the cursor moves to next digit if  the dial is turned quickly in one direction  The cursor moves back one place  to the right every two seconds if the dial is not turned  The text or numerical  value displayed is accepted by pressing OK  To cancel  press ESC     Monitor UPC 900 User Manual 18 1125 55 Edition AG 
61. used to set measurement parameters  zero and  full range values  for Cond  pH and UV on the analogue output channels   The analogue output channels can be connected to a chart recorder  for  example     From the main menu 1  turn the dial four steps clockwise to enter the menu  Set Parameters  Click OK  select the quantity to set by moving the cursor  with the dial  then click OK again     Note  The user interface of the monitor must be unlocked if connected to a  UNICORN control system     Setting the Cond analogue output  1 Current analogue settings are displayed  zero and full scale values      Monitor UPC 900 User Manual 18 1125 55 Edition AG       Set Cond Zero Level          Set Cond Zero Level          Set Cond Full Scale          Set Cond Full Scale                Set ph Analog  Out          Set ph Full Scale          Set ph Full Scale          Set ph Zero Level          Set ph Zero Level                Set UV Analog  Out                Set UV Range          Set UV Range             Operation    2 Press OK to access the settings menu  Press OK     3 Set the desired zero value  The range is 0 000 999 9 mS cm   Press OK to acknowledge     4 Turn the dial clockwise to access the settings menu  Press OK     5 Set the desired full scale value  The range is 0 000 99 9 mS cm  Press OK  to acknowledge  Press ESC two twice to return to menu Set Parameters     Setting the pH analogue output    After having selected Set Parameters pH and pressed OK  turn the dial one  step clock
62. wide range of applications        Connector switch Function  UniNet 1 Computer network  Analogue out 0 1 V Chart recorder outputs  3 channels 0 1 0 V  Optical Unit UV signal from optical unit  pH Probe Connection to pH electrode  standard BNC socket  pH Ground Reference ground for pH measurement  Conductivity Flow Cell Connection to conductivity flow cell   9 pole D sub connector  Lamp UV lamp connection  Mains Supply voltage  grounded  0 1 Mains supply on off switch       The module contains no internal user replaceable items     Monitor UPC 900 User Manual 18 1125 55 Edition AG 35    A  Reference information    36    A 2 UV optical unit    The UV optical unit houses the lamp  Zn or Hg   the wavelength filter and the  UV flow cell  The light beam is directed through a double conical or straight  flow through cuvette  6 ml or 2 ml illuminated volume  to a photodetector   The photodetector current is fed to the signal processing circuitry in the  module        Optical Unit Inlet        Filter Lens Beam splitter Flow cell Photodetector                                                                                           Lam   P Outlet    Photodetector eal             UPC 900    Analogue outputs    The reference signal comes from the same point in the lamp as the signal  measuring the sample  thus assuring a stable baseline by eliminating the  effects of variations in lamp intensity     The Hg lamp emits light only at certain wavelengths  It does not emit light   at 280 nm  
63. wise to skip the Calibrate pH menu  Then press OK     Note  The pH values for zero level and full scale must differ by at least 1 pH  unit     Note  The zero level and full scale values can be calibrated in any order     1 Current analogue settings are displayed  zero and full scale values      2 Press OK to access the full scale settings menu  Press OK     3 Set the desired full scale value  The range is pH  0 50 14 30   Press OK to acknowledge     4 Turn the dial clockwise one step to access the zero level settings menu   Press OK     5 Set the desired zero level value  The range is pH  0 50   14 30   Press OK to acknowledge  Press ESC twice to return to menu  Set Parameters        Setting the UV analogue output    1 Current analogue settings are displayed  zero and full range values    Allowed full range values are 0 0001  0 0002  0 0005  0 001  0 002  0 005   0 01  0 02  0 05  0 1  0 2  0 5  1 0  2 0 and 5 0  Zero level is set as a  percentage of full scale     2 Press OK to access the settings menu  The current setting is displayed   Press OK     3 Set the desired full range value  Press OK to acknowledge     Monitor UPC 900 User Manual 18 1125 55 Edition AG 15    Operation       Set UV Zero Level          Set UV Zero Level          Measured absorbance  level  A U        Range Over range  AU          Signal to  recorder  V   1V             16    4 Turn the dial clockwise to access the settings menu  The current setting  is displayed  Press OK     5 Set the desired zero l
64. xpected     Note  pH measurement is optional in AKTA design        CAUTION  Read the following information carefully to ensure that the  module is installed correctly              2 2 General precautions    The module should be located in a place of low temperature variations   away from heat sources  draughts and direct sunlight     The module may be operated at normal ambient temperatures in the range  4 to 40   C     The module should be installed on a stable laboratory bench or in AKTA design  system rack  To ensure correct ventilation  a free space of 0 1 m is required  behind and in front of the module  Do not place soft material under the  module to ensure that the ventilation inlet below the front is not blocked                        UPC 900          TIJ    Monitor UPC 900 User Manual 18 1125 55 Edition AG 3    Installation    O ring       Conductivity unit  cell    Protective plug  for flow cell             2mm   cm     NCO    5mm          Detector    Screw                   Holder Optical unit       2 3 Installing the optical unit    Installing the holder    1 Hook the holder into the slot on the right hand  side of the module  Secure it by pushing up  the slide clamp        Flow cell installation    There is one analytical  5 mm  and one preparative  2 mm  flow cell available   stand alone units only include the preparative flow cell   Both cells are  installed in the same way  as described below     1 Remove the red protective plugs from the detector housing and 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Package Contents Charging the Unit Controls Indicators  PDF 文件使用 "pdfFactory Pro" 试用版本创建 www    130 - Man CLIMATICUM ita  Excellence M  b)重さ (ポンド) :7(± 0.5) :不要    Wählen Sie  TriCore™ 1 Architecture Volume 1: Core Architecture V1.3  dé - CONRAD Produktinfo.    Copyright © All rights reserved. 
   Failed to retrieve file