1. The average result will be displayed If you press or button the memory number can be displayed If you hold down or the meter scrolls quickly and displays memory A re sult corresponding to the memory position will be shown when you release the button To delete one of the stored results press the power button and button simultaneously for one second To clear all the test results stored continue to press power and buttons for five seconds 18 CARE AND MAINTENANCE A clean optical measuring system is a basic prerequisite for obtaining accurate and precise values for cholester ol and triglycerides Therefore regular cleans are highly recommended Always power off the meter before cleaning Use ordinary lint free cotton balls or gauze or cotton clean tissue or cloth Mild soap as well as 7096 alcohol or isopropanol al cohol are suitable to moisten the cotton cloth to clean plastic surfaces Lift the front cover and gently clean glass surfaces of the optic with a dry gauze or cotton ball Clean plastic cover support of the meter with mois tened and then dry gauze or cloth Donotuse any disinfectant sprays or cloth cotton balls which are dripping wet as the liquid may penetrate and damage the instrument ERROR MESSAGES In certain circumstances error messages may appear on your display E01 error means that the instrument or ambient temperature does not fall
2. take the blood sample Lance the outer side of the fingertip with the lancing device and a sterile lancet to obtain a large hanging drop of blood Wipe off srm the first drop with the cotton ball or gauze and use the second hanging drop of blood to apply to the test strip The meter displays a drop symbol on the screen window indi g Ta cating that the strip is inserted and ready for the blood sample Apply a large hanging drop of blood directly from the finger to the white sample application area of the strip For measuring glucose apply a drop of blood on the end of the test strip Let the strip aspirate the blood sample Do not touch the appli cation area with the finger The drop of blood must be applied to the test strip immediately after lancing the fingertip Blood which is applied later may lead to inaccurate result For cholesterol and triglycerides testing it is critical to wipe away the first drop and immediately apply the second drop Head and record your result If the displayed result does not match your state of the health or seems to be unusually high or low check the meter function using a new test strip and a control solution not provided in the starter pack age If this check confirms proper functioning of the instrument read again the proceeding instructions of this manual Perform another measurement us ing a new test strip If the new result also seems to be not plausible consult your doctor
3. is entered correctly you will hear a beep Verify if the code digit on the display corresponds to the code printed on the strip container Now take the test strip from the test strip container Close the container immediately after removing strip to protect the desiccants and re maining strips otherwise the test strips may become unusable before expiry date Liquids must not enter the test strip container Hold the test strip with thumb and index finger so that the blood sample appli cation area is facing upwards Insert the test strip into the designated strip slot From the printing on the reverse side of the strip the meter detects which test parameter is to be mea sured and which code is needed for the test strip If the code has not been entered via data chip yet an error message is displayed after inserting the test strip 16 Cholesterol test strips only After inserting the strip F Fe male on the left and M Male on the right will appear on the display Press the corresponding button for F or for M to select the gender After about two seconds A before meal on the left or P after meal on the right will appear on the display Press the corresponding button for A or for P to select the letter The selected letter will remain two sec onds on the display After that the code and the blinking drop will appear on the display indicating that the device is ready to
4. ENT Your PRIMA Self Testing device 2in1 or 3in1 Test strips for measuring the desired parameter on chip data Lancing Device Sterile Lancet User s manual Cotton swab or gauze not included Diary where you record the results not included INTRODUCTION TO TEST STRIPS AND DATA CHIPS The basic package of PRIMA Self testing Kit includes a container with 5 test strips for measuring cholesterol one container with 5 strips for measuring triglyc erides and one with 10 test strips to measure glucose There are also containers with 25 50 strips sold separately The test strips are chemically modified with enzymes The test strip changes color by reacting with cholesterol or triglycerides present in a drop of blood and the device measures this change of color The overall level of cholesterol or tri glycerides in the blood is then shown on the display Each new container of test strips includes a data chip The chip provides the device with important infor mation about the specific properties of the respective control strips calibration data The chip is essential before use of the test strips This means that you need to insert the data chip each time you want to change parameter measurement cholesterol triglycerides or glucose may to or vice versa Cholesterol r ma Cholesterol Strip CH LL Data chip r is mam Triglycerides Strip TGL rd owsespoo MEM cue 11 Do not forget to have the data chip on hand w
5. NTENTS OF THE PACKAGING One digital meter 5 cholesterol strips a box of 25 strips is available to purchase separately 5 triglycerides strips a box of 25 strips is available to purchase separately 10 glucose strips a box 50 strips is available to purchase separately Three Data chips one for the cholesterol one for triglyceride and one for glucose testing The chip contains lot specific information of the calibration curve for either cholesterol triglycerides or glucose test One lancing device 25 sterile lancets One user manual One carrying case Articles non included within the packaging Control solutions available separately Logbook PC cable available separately and software THE DIGITAL METER General overview of the meter Display Switch on button 3 button 4 button 5 Test strip slot 6 PCcable port r 8 9 1 N Cover Battery space Data chip slot 0 Strip ejection button Front Rear DISPLAY AND SYMBOLS The display should be checked every time you turn on the instrument The meter momentarily shows all the symbols that may appear in the display Regular mo nitoring of all the display symbols prevents a misinterpretation of the instrument due to defects in the display The symbols on the display have the following meanings 1 Data chip inserted 2 Meter mode in use g e 2 3 Battery level indicator AVG MSET 4 Test results and other mes
6. PRIMA Lab SA Via Antonio Monti 7 6828 Balerna Switzerland 0041 0 91 605 1030 info primahometest com primahometest com B PRIMA USER MANUAL 2IN1 3IN1 MULTIPARAMETER DIAGNOSTIC DEVICE E E CHOLESTEROL eee TRIGLYCERIDES GLC gt gt KARTEN ONLY FOR THE PRIMA 3IN1 PL300 PL311 CE 0344 Made in Italy 1 12 83 01 01 p ENGLISH Thank you for choosing the system PRIMA 2in1 3in1 test for cholesterol triglyc erides and glucose directly to your home This manual contains instructions on how to use step by step the instrument and its maintenance Please read the manual carefully before performing the test Before carrying out the measurement we recommend that you become familiar with the instrument and procedures Please also read the instructions inside the boxes of different types of stripes In case of any questions please call customer service INDEX What s the Prima 2in1 3in1 system for Contents of the packaging The digital meter Display and symbols Setting up the meter for the first time Inserting the batteries Meter settings SET Requirements to perform a measurement Introduction to test strips and data chips Programming the device How to obtain a blood sample Performing a measurement Quality control recommendations Memory Care and maintenance Error messages Performance specifications Warranty Important Notes Ordering informations P
7. When measurement is complete use the ejector key on the back of the meter 14 to remove the used strip Press power button until the instrument powers off Properly dispose of the used lancets and test strip according to your local laws and guidelines Clean the instrument if necessary see the section Care and Maintenance of this manual QUALITY CONTROL RECOMMENDATIONS To ensure that meter is functioning properly you should consider using control solutions for a function check For each test parameter separate control solu tions are available to purchase A function check is performed in the same way as a regular measurement except that control solutions are used instead of blood Make a habit of carrying out a function check in the following situations f you leave the test strips container opened for a long time or if the container integrity was compromised f you drop the meter f you repeated a test and the result is still not plausible higher or lower than expected When you opened a new lot of the test strip container When you want to check the performance of the meter and the test strip MEMORY The meter is capable of storing up to 500 test results and lets you review results in order from the most recent 1 to the oldest 500 When you insert a Data chip into the CH meter you can recall previous results While in the standby press and release and button to enter MEMORY mode
8. e the instrument and strips under humid or damp conditions without suitable protection Failure to comply with the above notes may lead to inaccurate and false results 22 PRODUCT LIMITATIONS Please read the instructions of use for the test strips and control solutions for ORDERING INFORMATIONS primahometest com detailed information on product data and the respective limits of the test Article Description REF PRIMA Glucose test strips with the blue data chip and PRIMA Glucose Control Solutions are only compatible with PRIMA 3in1 Self Testing Kit and are not included in the 2in1 Self Testing Prima Cholesterol Test Strips 25 test strips for determining PL301 Kit blood cholesterol f NE iid MA 3in1 Self Testing Kit t Id in th A Prima Triglycerides Test Strips 25 test strips for determining PESOS Md blood triglycerides Prima Glucose Test Strips 50 test strips for determining PL305 blood glucose g Distributed by Prima Cholesterol Control Control solution to be used PESOS PRIMA Lab SA Solution with the cholesterol test strips Via Antonio Monti 7 6828 Balerna Switzerland Prima Triglycerides Control Control solution to be used PL309 l Solution with the triglycerides test strips www primahometest com Prima Glucose Control Control solution to be used PL307 Made in Italy Solution with the glucose test strips Prima 2in1 Multi parameter Device 25 sterile lancets PL300 Diagnostic D
9. ence HOW TO OBTAIN A BLOOD SAMPLE Prepare the lancing device by inserting a new sterile lancet COMPON ENTS M Y meee ne Uo LI Ejection System Sliding Barrel g an t drca Depth indicator Adjustable Tip REEL 78 adii ERN Y Jt JJ f Lancet Holder Release button 1 Unscrew the tip of the lancing de vice by turning it counterclockwise while holding the base 2 With two fingers pull out the lan cet carrier and hold it in place 13 3 While holding the lancet carrier insert a new sterile lancet into the bottom of lancet carrier Twist the cap of the lancet off 4 Put the tip of the lancing device back on and turn it clockwise 5 Hold the tip firmly in one hand then pull out the sliding barrel with the other hand This will cock the lancing device 6 Wash your hands with warm wa ter Ensure hands are warm and dry before lancing If needed massage your fingertip 14 7T Hold the lancing device firmly against the side of the finger with the cap resting on the finger The harder it is pressed the deeper is the puncture Press the release button to prick 9 After you finished with the meas urements unscrew the tip of the lancing device by turning it coun ter clock wise 15 8 After lancing yourself try obtain ing a hang
10. eries at the same time as batteries with different capacity may interfere with the functionality of the instrument Close the battery compartment by sliding the cover until you hear a click Turn the instrument to test the operation of the new batteries When you change the batteries you have to insert the new ones within 1 minute to keep the same settings for date and time If you exceed the minute you have to set new date and time The results of the various measurements including date and time as well as the other settings are retained even if the batteries are not included neS To respect the environment dispose of used batteries in accordance with the requirements of local laws of your own country Do not throw batteries on flames as they may explode METER SETTINGS SET DAY MONTH AND YEAR There are 2 formats day month and month day They appear on the display in the lower right corner To set the preferred format press and release the power button to access the SET until you see the format blink Press the or to select date formats To select the month day and year press and release the power button until the signal does not blink Press or buttons to set the correct month Press power to confirm and go to the next setting Similarly also set the date and year by pressing the and buttons m amam X mamam x i i amam Nia LE Ost Oa Ial TIME There are also 2 setups fo
11. evice 5 cholesterol strips 5 triglycerides strips Prima 3in1 Multi parameter Device 25 sterile lancets PLSTI Diagnostic Device 5 cholesterol strips 5 triglycerides strips 10 glucose strips Sterile lancets for Lancing 50 Sterile lancets PL313 In vitro diagnostic medical device Reference number device 100 Sterile lancets PL314 23 Read instructions before use Batch code Range temperature Manufacturer Expiry date CE mark ED2015 APRILE EN 24 LOGBOOK Write down the results you get with the PRIMA device 25 26
12. for the respective test strip E05 error indicates data chip error Solution remove the chip and re insert properly If the E05 error remains call Customer care E06 error appears when insufficient blood sample was applied to the test strip application area Solution remove the used test strip and repeat the measurement applying a large drop of blood E07 error indicates that the data chip is not inserted Solution Insert a data chip for proper coding of the instrument LO error indi cates that the result falls outside of the lower limit of the instrument sensitivity e g too low values Solution Perform a control function check If control passes the check repeat the test applying a large drop of blood and ensuring that the first drop was wiped out and only the second drop was tested If the control check fails your instru ment is not functioning properly Contact the Customer care HI error appears when the result falls outside of the higher limit of the instru ment s sensitivity e g too high values Solution Perform a control function check If control passes the check repeat the test If the repeated result is still high make a visit to your doctor If the control check fails your instrument is not functioning properly Contact the Customer care Flashing results are indicative that the test strip is too old and may be expired Note that flashing results are not stored in the memory Battery is dying and needs re
13. frequent measurement of blood cholesterol for men above the age of 40 and women over the age of 45 unless there are other risk factors If on the contrary there are other factors it is recommended to perform the test above 25 years of age The triglycerides consisting of three molecules of fatty acids and a glycerol are the most common form of fat that our body digests To be absorbed these mol ecules are broken down in the small intestine and then reassembled with cho lesterol to form chylomicrons Therefore in blood triglycerides are present in the form of plasma lipids related to cholesterol While cholesterol is used to build cell membranes and some hormones the main function of triglycerides is to store energy All the extra calories that come from carbohydrates are converted into triglycerides and stored as fat to be burned later as a source of energy Although it is not yet clear how high levels of triglycerides are placed in relation to the thickening and hardening of the artery walls atherosclerosis which increases the risk of heart attack and heart disease The disease is known as hypertriglyc eridemia and ignoring the high calorie foods can be caused by diabetes kidney failure or genetic derivation The test cannot be used on infants since it has not been validated for this pur pose It does not diagnose any disease and like any other test the compliance with instructions is critical to obtain accurate results C O
14. hich is supplied with each con tainer prior to the first measurement Every data chip belongs to a particular lot or container If possible keep the container of test strips along with the date chip so it s always available PROGRAMMING THE DEVICE 1 Remove the Data chip from the container 2 Insert the chip into the appropriate slot 3 Gently push the chip into the slot 4 Press the button power on off 5 f the meter reads the information of the chip properly a short beep con firms the programming correct if the beeper is enabled The three digit code should appear on the display Check if this code is the same that is on the container 6 If you have inserted the cholesterol chip a symbol in the shape of the strip with the CH characters blinking 7 if the chip is inserted for triglycerides should see the letters TGL 8 if you inserted the glucose chip you should see the letters GLC In case of problems with the programming the device reports an error E05 see the Error Messages section of this manual If an error occurs repeat the inser tion of the chip after a few seconds INTRODUCTION TO THE LANCET DEVICE The test system includes the PRIMA lancing device a handy tool for collecting capillary blood samples for tests which require one or two drops of blood An ad 12 justable tip allows 5 levels of skin penetration for individual comfort and lancets are ejected with ease and confid
15. ing drop of blood Always wipe off the first drop with a cotton ball and use the second drop for the testing The first drop of blood usually con tains extracellular tissue fluids that may dilute your real capillary blood and cause erroneously low results 10 Push the lancet ejector forward with the thumb and simultaneously pull out the sliding barrel to dispose of the used lancet in a proper con tainer PERFORMING A MEASUREMENT Get the test strip container for the required measurement Check the expiry date of the test strips Always use the strips be fore their expiry date has passed Make sure that Data chip belonging to these test strips is at hand unless the instrument has already been coded see the section Coding the Instrument of this manual Note Environmental in fluences e g air humidity and light on the test strips may damage test strips integrity and lead to false measurements or error mes sages Do not remove the test strips from the strip container until immediately before performing a test Turn on the meter Check the display before running the test Are date and time correct If the measured values must be stored with time information enter the correct setting Does battery symbol appear If it appears only a few more measurements can be performed Replace batteries as soon as possible If the instrument has not been coded yet insert a respective Data chip into designed slot When the chip
16. placement 20 PERFORMANCE SPECIFICATIONS Operating conditions Relative humidity 0 90 Temperature range for measurement 10 C 40 C Measurement range Cholesterol 3 3 10 2 mmol L 130 400 mg dL Triglycerides 0 56 5 6 mmol L 50 500 mg dL Glucose 0 6 33 3 mmol L 10 600 mg dL Memory capacity Up to 500 tests Battery 2 x 3V button cells Type CR 2032 Number of measurements with new batteries Approximately 1000 Weight 65 g batteries included Dimensions Length 97 mm Width 20 5 mm Height 49 mm Display LCD 21 WARRANTY The meter comes with three years of limited manufacturer warranty The warranty does not cover damages caused by improper usage and handling accident and or negligent care IMPORTANT NOTES Always Operate the instrument within the acceptable temperature range Place the instrument on a level surface or hold it steady in your hand Make sure that all display elements are functioning during the self test Readthe test strip inserts Keep the meter and the test strip container clean and tidy To get less variable results fast at least 10 hours before measurement Never Touch or remove the test strip during actual measurement Delay starting the measurement after lancing Subject the instrument to sudden movements and shakes during a measurement Store the meter and test strips at extreme temperatures Stor
17. r the time e International format of 24H Anglo American format 12H with AM or PM To set the preferred format press and release the power button for 3 seconds to enter the SET mode Press power until you see the time format screen Press or to choose your preferred format BEEPER After setting the time you can set the beeper and select ON or OFF It is recom mended to leave the beeper always active default setting If the beeper is on you hear a beep in the following situations When the instrument detects a test strip inserted When the results appear on the display When an error occurs ALARM You can also set an alarm on the device in case you wish to be notified before the test You can program up to 3 alarms Attention the alarm does not work when SET a iacu A set to 12 00 or to 0 00 AVERAGE You can set the device to store an average of the results of tests performed during a particular period of time For example an average of 14 days as shown The principle of the setting is the same pressing the power button with the buttons and UNITS The device allows you to view the test results in international units such as mmol L used in Canada and Anglo American as mg dL You can choose the preferred unit by setting the device in SET mode and pressing the and buttons SET SET 3 mg dL Fmmol LZ 10 REQUIREMENTS TO PERFORM A MEASUREM
18. roduct limitations WHAT S THE PRIMA 21N1 3IN1 SYSTEM FOR The PRIMA 2in1 3in1 kit is a new system designed for the self monitoring in vitro therefore outside of the body through the blood obtained from a finger prick of the quantitative determination of the levels of cholesterol triglycerides and glu cose These parameters are often measured together as they are the key factors for the estimation of the risk of cardiovascular disease It is designed for home use by people who want to monitor these parameters and on the basis of their levels to understand whether it is necessary to consult a doctor and or control the same parameters if they are too high Cholesterol is a waxy substance that acts as a precursor of bile acids steroids and vitamin D It is synthesized primarily in the liver and in the intestinal wall In normal conditions the liver maintains the balance between the amount of cho lesterol and draw to that excreted This balance however can be altered by cer tain diseases such as diabetes hypertension hypercholesterolemia and family factors also linked to an unhealthy lifestyle such as smoking alcohol abuse and a diet rich in saturated fats in combination with a lack of physical activity In ad dition stress oral contraceptives and pregnancy can cause elevated cholesterol levels A high concentration of cholesterol is a major risk factor for cardiovascular events The Canadian Cardiovascular Society recommends the
19. sages DA Gy UE 5 Units of measure Ti 6 Parameter being tested Gp Gap D 7 Temperature indicator Ex ms AED aS AUS i Aid Month Year Qnmol L mg dL 1 eeper v c Decium 29 Meter s modalities Qi i 7u3 am DEDI ao The meter has three modalities Oat mpm a co O STANDBY It s the default mode that appears every time you turn on the instrument by pressing the ON switch The strip symbol will flash and the display will show the number of the data chip the date and time MEMORY When in Standby mode by pressing the or is entered in memory mode The instrument will show the MEM sym bol in the upper part of the display SETTINGS Press and hold for 3 seconds the power button when the instrument is turned on The instrument will display the SET symbol in the upper part of the display SETTING UP THE METER FOR THE FIRST TIME When you use the tool for the first time before turning to perform the following steps 1 Insert the batteries 2 Setting the time date and unit 3 Insert the data chip it can also be done just before the measurement INSERTING THE BATTERIES 1 Ensure that the instrument is turned off Open the battery cover by sliding the cover 2 Insert two 3V lithium batteries CR 2032 in the slot as shown paying at tention to the symbols and ends of the batteries Use only 3V lithium batteries CR 2032 Remember to always change both batt
20. within the acceptable range for the selected test Solution move the instrument to an environment which has an appropriate tem perature 18 35 C and wait for the meter to equilibrate before repeating meas urement Do not artificially heat or cool the meter by any means E02 error can imply several things First the test strip is compromised because of coloration of the reactive area This might happen when a used or dirty strip is inserted or when a new strip was stored inappropriately It can also mean that the blood sample was applied too soon before the meter displayed flashing symbol of a blood drop Solution remove and discard the used strip open the plastic lid covering the test strip clot and clean plastic window with a slightly moistened with water gauze Take a new test strip insert it into the designated slot and apply a blood sample only after a blood symbol is displayed on the screen E03 error indicates that the strip was removed prematurely without applying a blood sample Solution power off the meter and repeat the test after a few seconds E04 error appears when the test strip belongs to a different lot than the container which was last coded for this parameter or when the meter was coded for cho lesterol test but a strip for triglycerides was inserted for testing Solution remove the strip and repeat using a test strip matching the last code for 19 the test being attempted Alternatively re code the instrument

Download Pdf Manuals

image

Related Search

Related Contents

Reproductor de Videodiscos Digitales y Discos Compactos  取扱説明書 【本体機能操作編】  Manual de Instrucciones  Article 00021 - How To Manually Configure Net Report For ORACLE  Version1    NP805N User Manual  Duraflame DFH-UH-1-T Instructions / Assembly  Sunny Island 3324/4248 Installationsanleitung  Bedienungsanleitung  