Home
        LUCIFER
         Contents
1.                      Service Control       0 suspend                               Figure 2 5  The service parameters access window     2 2  THE LUCIFER SOFTWARE CONTROL CENTER 13    Accessing the configuration of a service    If the access service parameters context menu item which is only available for RMI Services  is selected  the window shown in Figure  2 5  will appear  The top part of this window displays  information about the selected service  such as the name of the service  the version  the host name  and address it is running on and how long it has been running  Below this information command  line parameters that have been defined when the service was started are shown  The List of the  service configuration entries allows the selection of an entry which can be viewed  see Figure  2 6   or edited  Figure 2 7  by clicking on the corresponding button  Finally the window allows stopping  or suspending a service after a specified amount of time given in milliseconds     Viewing a service configuration    de rub astro lucifer control electronic hiramo ConfigHIRAMOSeryer      lt  xml version  1 0  encoding  UTF 8  standalone  n0 2 gt      data       usage       program      lt object types de rub astro lucifer control electronic hiramo ConfigHIPAMOServer      lt value name  TIMEQUT  type  int  gt   000 lt  value gt     lt value name  INITIALIZATION DELAY TIME  type  int  gt 2000 lt  value gt     lt value nam   TRY TIME  type  int  gt 30000 lt  value gt     lt value nam
2.                 Choose a macro  4 09 s  4 09     Execute Macro  ee            Exit          Figure 2 14  The read out engineer GUL     2 3  ENGINEER ACCESS TO THE MAIN INSTRUMENT COMPONENTS SW 25    2 3 6 The MOS engineer GUI    The MOS unit engineer GUI is currently not intended for use for other people than the MOS  specialists at the MPE  A separate manual will be provided from there  LBT LUCIFER MAN   027      26    CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    Chapter 3    Technical procedure manual    3 1 Procedures    3 1 1 LUCIFER safe mode    If the instrument is not in operation for several days  it should be put in safe mode  This means  that the on instrument electronics MCE ROE are switched off and the detector put to a lower  temperature of about 60K     NOTE  During the time needed for the detector to change its temperature  the elec   tronics have to be switched off     The time for changing between 77K and 60K is approximately 35min  The detector temperature  must be monitored for about one hour to be sure that the expected end point is reached    When putting the instrument again in operation mode  i e  detector at 77K   it will take  additionally about 2 3 hours until a stable image can be read from the instrument    To access the instrument from the engineering account  the laptop used for the MOS cabinet  exchange  user  password known to authorized LBTO personel  or one of      LBTO workstations   user password by LBTO  can be used  A NxClient connection 
3.     Service  system    Service  manual    Service  manual  9 Service  system  9 Service  system    Service  system    Service  system    Service  system    Service  system   Service  system    Service  system    Service _  system  0 Service  system    Service                                                                                                                                                                           refreshtable         discard terminals Jj    Figure 2 2  The Software Control Center after the service has been started     KNOWN ISSUE AT START UP   STARTING A SERVICE BY HAND     It may happen that the MOS Sequence Server still has a red light meaning it has  not started properly  Therefore it has to be done by hand  Click in the table on the  MOS Sequence Server row  be careful because due to the nx connection the refresh  of the table is a bit slow   Use the right mouse button and choose start MOS Se   quence Server  Wait a few seconds  press    refresh table    and check if the service is  now running     2 2  THE LUCIFER SOFTWARE CONTROL CENTER 11    By using the right mouse button all services can be stopped or started  By scrolling the table  down the Applications are getting visible in a green font  Here all available GUls can be started   Besides the typical observing GUIs also the engineering GUIs are accessible     2 2 2 Configuring the software start and the services    When a program is selected from the table shown in the Software Control 
4.    Osun          L                   Figure 2 9  The expanded Control GUI with the motor control panel     status bar  showing the pathname of the file  is showing     modified     at the end  The color of the  text indicates if the table that is currently shown has been loaded from the service  in which case  the color is green and the text begins with  REMOTE FILE     or from a local file  red text color  and  LOCAL FILE  at the beginning   Loading a local file is possible using the  File  menu  In this  case too  the loaded table is not used by the service directly  To force the service to use the new or  edited table  the  Commit to service    button has to be pressed  This only commits the currently  displayed table  and has to be repeated for all changed tables  It is also possible to reload a table  from the service in case of unwanted accidental changes using the  Reload from table  button      The tables used by the service are saved separately in the    lt lcsp_root gt  lookupTables directory  which is used by other services as well  The files  in this directory should not be accessed and edited directly using other tools like text  editors  Corrupted lookup table files will have a negative impact on the instrument  performance and operations     New lookup table entries can be added in the empty line at the end of the table  To remove a  line from the table the corresponding line has to be selected and the  Remove Entry  button has  to be pressed  Each table entry 
5.    v  system    Service SYS  TimeServer e   v  system    Service SYS  MessageServer e   v  system    Service S Journalizer e   v  system    Service GEIRS    9   v  system  8 Service PortServer      L  manual   amp  Service TelescopeServer e      system    Service CalibrationUnit e  system    Service PressureMonitor         system  0 Service RackCoolingControl e  system    Service ONE  TemperatureControl 0   v  system    Service ONE  TemperatureMonitor e  Middle     manual  9 Service Two  TemperatureMonitorLUCIFER2       manual  0 Service ONE  TurboPumpMonitor e  section v  system    Service         HIRAMO Server e   v  system  0 Service         LUCIFER MCU Server e  system    Service        MOS MCU Server         system    Service        MOS SwitchBox Server e   v  system    Service ONE Weblo 1 e   v  system    Service ONE Weblo 2 e   v  system    Service ONE  Camera Unit Server e  system    Service         Compensation Mirror Server e   v  system   amp  Service ONE  Detector Unit Server e      system    Service ONE  Filter Unit Server     refresh table  Bottom  section    discard terminals a         Figure 2 1  The Software Control Center main GUI when the software system is not running     programs on the terminal is displayed here  If a panel is no longer needed  i e   because the asso   ciated program has been terminated  it can be deleted by clicking on the    discard terminals    button     When the system is started using the start button  a splash screen appears that 
6.   The MOS  acquisition procedure  described in LUCIFER UM 1 2  will work only of the Ims file corresponding  to the mask in use is found  and readable  by the software     3 1 3 Maintenance of the data directory  data luci    All scientific data taken with LUCIFER end up  on the LUCIFER workstation  in the  data luci  directory  These data are visible from the observer session but this user only has read privileges   Before a new science block starts  typically once a month   it is necessary to clean up the disk  space in the  data luci directory  This must be done as user luci  using the command    ssh  X  luciGlucifer   no other user  observer or engineer  can delete data from that directory  The main   tenance of this directory and check of the available disk space is the responsability of LBTO and  should be included into the list of things to be done during the two technical nights preceding any  science run     Do not remove directory  data luci ComTeamData_DoNotDelete        3 1 4 Regular checks of the instrument    In the LBTO operational mode  where an instrument specialist is not always on the mountain  where observations are taking place  it is fundamental that quick checks of the instrument take    30 CHAPTER 3  TECHNICAL PROCEDURE MANUAL    MOS Mask Jd  File       masknumber  maskid  maskdescrp  on  isused   LMS file  vi     990063  Optic Sieve  990001  Spectro Sieve                         0  1  2  3990031          990078  7990065  8995623           932121 MPIA6
7.   is the focus offset from the reference focus for the camera  This reference focus has been measured  for the combination of the clear and the K filter and is stored in the service configuration  Tables  can be loaded in the same way as for the compensation mirror service and have to be committed  to the service explicitly after they have been edited      The focus control area    To be able to work properly the detector focus service needs to know three parameters  the  currently selected camera and both filter wheel positions  These can be set using the appropriate  drop down box in the focus control panel in the upper right area of the window  Unlike the  compensation mirror GUI  the current configuration used by the service is shown beneath the drop  down boxes again in a text field with purple background  The focus stage is not moved when new  values are selected from the combo boxes  only the new parameters are set  To move the focus  stage  the buttons on the right side must be used    Two initialization methods are possible for the focus stage  Either the positive or the negative limit  switches can be used and can be selected with the drop down box below the    Initialize Unit    button   Initialization of the drive is only necessary when the camera has been changed  The routine will  move the stage to one of the limit switches and then back to the reference focus position  To correct  the focus for the currently selected filter combination  the  Adjust Focus  butto
8.  Ims  file Dec09 mos1 80 TestMask lms   990001 25 13  MPE  LS larcsec   990002  MPE    990003 26 78  MPE  LS 0 5arcsec   990004  MPE    990005  MPE    990006  MPE    922231 14 72 MPIA engeli  home engineer Masks  Ims  file Dec09 mos1 32 engel1 Ims   910509 15 84 MPIA W3Bright  home engineer Masks_Ims_file Dec09 mos1 48 W3Bright Ims   911548 16 83 MPIA W3KIt14  home engineer Masks_lms_file Dec09 mos1 49 W3KIt14 Ims   912600 17 45 MPIA W3KIt142  home engineer Masks Ims  file Dec09 mos1 50 W3KlIt142 Ilms   925727 18 91 MPIA W3KIt15  home engineer Masks_Ims_file Dec09 mos1 51 W3KIt15 Ims   931719 19 89 MPIA W3KIt16  home engineer Masks_Ims_file Dec09 mos1 53 W3KIt16 Ims   933225 20 82 MPIA W30H  home engineer Masks  Ims  file Dec09 mos1 54 W3OH Ims    Figure 3 2  Excel sheet with mask information needed prior to a cabinet exchange     e the mask name  that will appear in the Instrument Control GUI  e the Ims filename  where the complete path where this file is stored is writen     Most of these information are then taken over during the last step of the cabinet exchange   when the masks are configured  Fig   3 31 illustrates the information that engineers will have to  update when configuring the masks    The lms files collected have to be placed on the LUCIFER workstation in the     Masks 1ms file   directory under the engineer account  It has been found practical to define a new subdirectory for  each new observing block  e g  Mar2010 for all the LUCIFER March 2010 observations
9.  of the panel  where the new masks have been configured     place regularly to make sure it is usable for the night     Actions to be performed at least once a week  by the instrument engineer specialists    especially when observations are taking place  Go the instrument platform and have a  visual inspection of the instrument    e Check all the cables  nothing loose  amp  or damaged     e Check the cooling flow level    Open the electronic cabinet  below the bridge  and check temperatures pressures    Go on the bridge and check the  on board  electronics  all cables still well plugged  no dam   age          Actions to be performed during the technical nights preceding science runs     Check the instrument  as indicated in the daily checks     Perform instrument set_ups  check grating tilts  calibration unit in out  lamps functioning    Make some reads and check the detector  darks and flat field aspects at least     Perform some tests observations to check communication with TCS  guiding and active op   tics     Actions to be performed daily by the LBTO astronomers on duty  during science runs     e Log onto the lucifer workstation as engineer    3 1  PROCEDURES 31    Open a terminal and check the instrument pressures and temperatures  These information  are provided from the command line showSupervisorStatus sh  which also show the gradients  and alarm status     Open firefox  select the  LUCIFERI Web IO Control    bookmarks and check that      ROE   amp  MCE electronics a
10.  panel of the service  To be able to see this area correctly the window must be enlarged in  the horizontal direction as shown in Figure  2 9     This panel allows direct control of the motors of the unit  bypassing the Compensation Mirror  service and should be used with care  Especially the  stop all motion  button in the upper right  corner should only be clicked in case of an emergency  since it stops all    movements  except the  MOS Unit  which is controlled by a separate motor electronic  inside the instrument immediately   No further movement of any motor will be allowed after this until the  allow motion    button is  pressed  Note also that the information of the motor control area is not automatically updated   since it bypasses the software service shown in the left area   To get the current values  i e   switch  and motor positions  for this part of the user interface the  update    buttons have to be pressed   For normal operation of the service the motor control GUI is not necessary and can  and should   be hidden  Its primary use is to move the motor s  of a unit in cases where the service is not  working properly     16 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL                ola Linear Interpolation Function M  Linear Interpolation Function M         progam   address            h   messa    OT CONTENT  database  OR CONTENT  Synchronized              Li eeem 0     0v      Figure 2 8  The compensation mirror engineer GUI       The lookup table area of the Comp
11. 4 lt  value gt      value nane  CAMEPA WHEEL        PATTERN START POSITION  type  int  gt 12 lt  value gt    lt value name  CAMERA WHEEL        PATTERN LENGTH  type  int  gt 2 lt  value gt     lt value name  COLD CLAMP ONE        PATTERH            POSITION  type  int  gt 15 lt  value gt     value name  COLD CLAMP ONE        PATTERN LENGTH  type  int  gt 1 lt  value gt     lt value name  COLD CLAMP TWO        PATTERN            POSITION  type  int  gt 16 lt  value gt    lt value name  COLD CLAMP TWO        PATTERN LENGTH  type  int  gt 1 lt  value gt     lt value names LCOEl             CONNECTION STATUS BIT PATTERN START POSITION  type  int  gt 1 lt  value gt      lt value name  LCOE1            CONWECTION STATUS BIT PATTERN LENGTH  type  int  gt 2 lt  value gt         lt value name  LCOE2 MOTOR CONNECTION STATUS        PATTERN START POSITION  type  int  gt 3 lt  value gt      lt value names LC0  2 MOTOR CONNECTION STATUS BIT PATTERN LENGTH  type  int  gt 2 lt  value gt       value           1  083_ MOTOR CONNECTION STATUS BIT PATTERN START POSITION  type  int  gt 5 lt  value gt    lt value name  LCO0E3 MOTOR CONNECTION STATUS        PATTERN LENGTH  type  int  gt 2 lt  value gt     lt value           1  04 MOTOR CONNECTION STATUS BIT PATTERN START POSITI type  int  gt 7 lt  value gt    lt value          084 MOTOR CONNECTION STATUS        PATTERN LENGTH  type  int  gt 1 lt  value gt     lt value name  DEFAULT TEMPERATURE MODE  type  int  gt 0 lt  value gt      value name  DEF
12. 7 a9011      jhomelengineerMasks Ims  file Mar2010 mos1 67 a9011 Ims   933150          68   9012 jhome engineerMasks Ims file Mar2010 mos1 68 29012 Ims   2933520          69   9013        jhomelengineerMasks Ims fileMar2010 mos1 69 39013 1ms  034243     MPIAT2 engel3       jhomelengineerMasks Ims file Mar2010 mos1 72 engel3 Ims  834446     MPIAT3 engeld iv      Jhome engineerMasks Ims file Mar2010 mos1 73 engel4 Ims  934827  MPIA75 engel           nomelengineerlMasks_Ims_file Mar2010 mos1 75 engel6 Ims                                  1933852 IMPIA 70 engelt   jJhome engineer Masks Ims file Mar2010 mos1 70 engel Ims   7934051  MPIA 71 engel2   J home engineer Masks Ims  file Mar2010 mos1 71 engel2 Ims                      _ SO HITA   1913746 MPE 124 GOOD     Jhome engineer Masks Ims file Mar2010 mos1 124 G00DSN Ims  D                             jhomelengineerMasks Ims  fileMar2010 mos1 127 GOODSN Ims  31600D     jhomelengineerMasks Ims  fileMMar2010 mos1 131 GOODSN Ims  MPE134M67   jhome engineerMasks  Ims  fileMar2010 mos1 134 67 Ims        77   67   jhomelengineerMasks  Ims  file Mar2010 mos1 77 M67 Ims                           7915732 JAIP 79 laecosmo   jhome engineer Masks Ims file Mar2010 mos1 79 laecosmo Ims  8920014  AIP 80 laecosmo   j home engineer Masks Ims file Mar2010 mos1 80 laecosmo Ims                     1991901  05U 76 nac1333 Y  j home engineer Masks Ims file Mar2010 mos1 76 ngc1333 Ims                                                    Figure 3 3  Example
13. AULT CHANNEL NUMBER  typ t   1c value      lt value name  ADC OFFSET EPROR TEST CHANNEL  type  int   4   value      lt value name  CLOSED LOOP CHECK INTERVAL  type  int  gt 1000 lt  value gt      value name  CLOSED LOOP VOLTAGE TOLERANCE  type  float  gt 2 66E 4 lt  value gt      lt  object gt                  lt   nracram gt           Figure 2 6  The configuration view window     The view configuration window  see Figure  2 6  simply displays the configuration entries as they  are stored in the XML file  Since editing these files by hand can be very cumbersome  a dedicated  edit configuration window is provided  This window is shown in Figure 2 7     Editing a service configuration    In the window of Figure  2 7  the different configuration values can be edited directly  without  concern about the correct XML structure of the file  When the OK button is selected the currently  displayed configuration is written to disk and immediately used by the service  Clicking on the  cancel button reverts any changes made to that point and closes the edit window     14 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL        3 VALUE name   TIMEOUT  type   int        VALUE name   INITIALIZATION  DELAY  TIME  type   int       VALUE name   RETRY  TIME  type   int       VALUE name   UPDATE  INTERVAL  type   long  E TIMEOUT      VALUE name   HIRAMO_CARD_NUMBER  type   int     int      VALUE name   SWITCH_PORT_VALUE_PARAMTER_POSITION  type   in         VALUE name   CHANNEL_NUMBER_PARAMTER_POSITION  
14. Center window  a right  click with the mouse will display a context menu  Depending on the state of the software system   i e   running or not running  different menu items are available  see Figure 2 3                         edit start descriptor edit start descriptor  start HIRAMO Server   start HIRAMO Server     stop HIRAMO Server stop HIRAMO Server  restart HIRAMO Servet restart HIRAMO Server     access service parameter   access service parameter             Figure 2 3  The control context menu for an RMI service  shown when the software system is not  running  left  and when it is running  right      If the software system is not started only the first menu item edit start descriptor is available  as shown on the left side of Figure 2 3  The right side of this figure displays the context menu   when the software system is started and the selected program is currently running  If the system is  started  but the selected program not currently running  the first two entries of the context menu  will be active  not shown in Figure 2 3   To start a service or program manually the start  lt name gt   menu item can be selected  Consequently the stop   name   stops the corresponding service or  program  The restart  lt name gt  menu item is only available for RMI services and allows restarting  the service with the current configuration  To start the service again after a configuration change   it must be stopped first     Editing the start configuration    By selecting the ed
15. Control GUI changes to    moving       If the communication with the Calibration Unit after several attempts  try at least three times   did not work properly     or  OUT  not reacting even though the message  MOVING Calibration  Unit    appears   it would be possible to control this unit via a separate web interface  Open Firefox  bookmark LUCIFER1 Web CalUnit  or use 192 168 0 12    First one has to log in by clicking on  Config in the left menu    Choose  Expert Mode       Press button on right top  Power On    and reload  Check if left side Power on status is    ON      Press button  Command drive in    on right side  Then reload  Check if status  calibration unit  motor in    is    ON       Logout via    config     Session control    and close the firefox window otherwise communication prob   lems can occur again  The GUI logs the user automatically out after a while  For the next use a  fully new restart of the web page is necessary in order to log in again     BE AWARE   The calibration unit will NOT move in out via scripts   The calibration unit can be moved in out at any telescope instrument position      Telescope Service    In case of a hanging communication with the IIF TCS it might be necessary to stop and start  the Telescope service from the Software Control Center  Start Manager   As far as there  is no LST and UT update in the Telescope GUI even something is wrong with the IIF or any  other T CS system  If after a restart on the       side the GUI is still no
16. LUCIFER    Technical software and procedure manual    Doc  Name     Doc  Number     Issue Number     Issue Date     Prepared by     LBT LUCIFER MAN 032 1 2 pdf  LBT LUCIFER MAN 032   1 2   19 03 2010    Volker Knierim  amp  Marcus J  tte  AIRUB   W  Seifert  LSW   N   Ageorges  MPE     Recipient    O  Kuhn  M  Pedani  D  Thompson    19 01 2010  19 02 2010    18 03 2010    Distribution List    Institute   Company No  of Copies    Document Change Record    Sect  Paragr  affected Reasons   Remarks    initial document   Chap  4  Additional GUI descriptions  Chap  5  Update  Trouble Shooting   New procedures description added  Document reorganization       Contents                                           5  7  A E 7   7   7   11         15   2 3 1 The compensation mirror servic            Ie 15  al ho 5 od ee he ls 19   23 The erating unit                                           22   2 3 4         camera wheelservicd         a a a e a a 23   2 3 5 Read out engineer                                    24   2 3 6 The MOS engineer GUJ                               25   27  27   27   28   1  1 ici 29  Vau 29   iUud HS Read s MEE Sue rU Mu  32  CP II ee 32  daa ee Bee ee eee ds eb 34    CONTENTS    Chapter 1    Introduction    This document  organised in two main parts  is addressed to the LUCIFER instrument support  scientist and NOT to the observers  Because of the described engineer access to the instrument  only LUCIFER experts are allowed to use it     Chapter 2 describes how t
17. T before     at laptop workstation  switch ROE and  MCE via Web Interface on  Switching MCE ROE on off is the button on the right lower side     3 1 2 Ims files  amp  cabinet exchange    In preparation for an exchange of masks cabinet  procedure described in LUCIFER LBT MAN   033   the gerber files need to be collected to be send to the mask manufacturer  At that time the  corresponding lms files have to be collected as well  They are needed for the MOS acquisition    In preparation of a cabinet exchange an Excel sheet  Fig  3 2  needs to be prepared for the  engineers  That page needs to contain following informations     e the mask ID    e Two free columns where the engineers should write the cabinet slot in which the corresponding  mask has been put and the frame   used    e the information from which institute that mask comes from    3 1  PROCEDURES    Mask ID Cab  Slot Frame    Institute mask name    29    Ims filename    904059 10 42 MPE M076GOODSN 2  home engineer Masks  Ims  file Dec09  mos1 76 GOODSN Ims   924350 11 96 MPE M101GOODSN 1  home engineer Masks_lms_file Dec09 mos1 101 GOODSN Ims   934121 12 38        M106GOODSN 4  home engineer Masks Ims  file Dec09 mos1 106 GOODSN Ims   944022 13 77 MPE M107GOODSN 3  home engineer Masks Ims  file Dec09 mos1 107 GOODSN Ims  Test mask to compare with LBTO similar mask   913947 22 92 MPE MO80TestMask 913947  home engineer Masks_Ims_file Dec09 mos1 80 TestMask Ims   913947 23 12 LBTO MO80TestMask 913947  home engineer Masks 
18. are one has to login to the LUCIFER SUN V880 Server via an NXClient  available on the      computers     Host name  lucifer  User  engineer    There a desktop icon  Start LUCIFER Control SW    is available  Clicking this icon executes  the start script LUCIFER  SW START sh which is located in  home engineer  This will start the  LUCIFER Software Control Center     8 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    If there is no start script available the software can be started via the command line     cd  home engineer lcsp dist export  java  jar lcsp jar start hd xml    lcsp  jar contains the whole LUCIFER software system and start hd xml is the current XML  configuration file     RULE  Never edit start hd xml by hand unless you know what you are doing    Both are located in the directory referred to as  lt 1csp_root gt     home engineer lcsp dist export      The software system runs only once on the sun server  although multiple instances of the LU   CIFER Software Control Center can be started  When the software is already running on the  system  the Control Center detects this automatically     In the case the LUCIFER software is not running  the LUCIFER Software Control Cen   ter appears as shown in Figure PI    The window is separated into three different sections  The top section  indicated in red in Fig   ure EI  contains the buttons to start and stop the software system with the configuration that is  currently loaded  It is also possible to close this window by clicki
19. consists of three values  The  Instrument Position  consists of  two angle values for the rotation  R  and elevation  E  angles  This is also a key for the table   which means that only one value pair of angles is allowed to exist in the table at a given time   The other two values are the relative    motor steps that are necessary to compensate the image  flexure for the instrument position  The origin is defined by the entry that has two    0   s for the  motor steps and can be chosen arbitrarily  It has been proven practice during the commissioning  of the instrument to use the instrument position  R 90 E 90  as reference and to provide correction  values in intervals of 30 degree steps for the rotator     90  lt      lt  450  and elevation  0  lt  E  lt  90   angles  For instrument positions that are in between the defined position  the motor step values  to compensate the flexure are interpolated from the four surrounding positions  The interpolation  method that is used can be selected using the two drop down boxes below the table  Currently only    18 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL  a linear interpolation can be selected because it has been proven sufficient during commissioning     The mirror control area    Next to the lookup table panel  the service control area is located  see Figure 23   It consists  of different sections allowing to set parameters for the service and to execute actions  The color  of the different GUI elements represents the type of in
20. d displayed in red color   indicating that the tilt loop is closed  but the grating tilt not stable yet  After the tilt has been  stabilized  i e   three subsequent measurements of the tilt voltage are within the tolerance limits  of the target voltage  the text color changes to green  indicating a stable loop  If the loop can not  be stabilized within a time specified in the service configuration  an error message will appear and  the loop will be opened automatically  When the loop is opened using the    Untilt    button  the text  color changes to black again  Note  before a new grating tilt value can be set  a closed loop as to  be opened first by clicking the  Untilt  button    The  Temperature Mode Control  panel shown in the GUI in the lower right area is  used to switch between  COLD  and  WARM  operation modes of the tilt electronics   The  WARM     mode must only be used when the instrument is at room temperature  in the integration hall  Switching to the WARM mode when the instrument is not  at room temperature will severely damage the instrument  USE EXTREME CARE  WHEN CHANGING THIS MODE     2 3 4 The camera wheel service    The camera unit service  see Fig  is very simple and has no lookup tables  The engineer GUI  can be started using the  GUI Camera Unit Engineer Client  entry in the control center window   It allows to initialize the camera wheel and to select one of the three different camera positions   Additionally  the GUI can also be used to move the pupi
21. e  UPDATE INTERVAL  type  1long  gt 100 lt  value gt     lt value name  HIPAMO CARD NUMBER  type  int  gt 1 lt  value gt     lt value name  SVITCH PORT VALUE PARAMTER POSITION  type  int  gt 0 lt  value gt     lt value name  CHANNEL NUMBER PARAMTEE POSITION  type  int  gt 0 lt  value gt     lt value names GRATING UNIT TAPGET VOLTAGE PARAMETER POSITION  types int   0c value     lt value name  GRATING UNIT CURRENT VOLTAGE PARAMETER POSITION  type  int  gt 0 lt  value gt    lt value                    LOOP OUTPUT VOLTAGE PARAMETER POSITION  type  int  gt 0 lt  value gt    lt value name  TEMPERATURE MODE PARAMETEP POSITION  type  int  gt 0 lt  value gt     lt value name  LOOP STATUS PARAMETER POSITION  type  int  gt 0 lt  value gt     lt value name  ADC OFFSET ERROR  PARAMETER POSITION  type  int  gt 0 lt  value gt     lt value names DEBUG        PATTEPN START POSITION  type  int  gt 1 lt  value gt     lt value name  DEBUG BIT PATTERN LENGTH  type  int  gt 16 lt  value gt      value name  GRATING UNIT        PATTERN START POSITION  type  int  gt 1 lt  value gt     lt value names GRATING UNIT        PATTERN LENGTH  type  int  gt 3 lt  value gt     lt value name  FILTEF WHEEL ONE BIT PATTERN START POSITION  type  int  gt 4 lt  value gt    lt value name  FILTEP VHEEL ONE        PATTERN LENGTH  type  int   4c value      lt value names FILTEP WHEEL TWO        PATTERN START POSITION  type  int    gt 8 lt  value gt    lt value name  FILTEP WHEEL TWO        PATTERN LENGTH  type  int  gt 
22. ed not be started  at system time and can be started as necessary  A fourth type  not shown in Figure 2 T  exists that  allows starting any executable code  It is not necessarily needed  but can be convenient to launch  image viewers for example  The use column is only meaningful for RMI Services that are shown in  blue text  The string SYS means that the service is a general system service and not associated to  any instrument  ONE indicates a service controlling LUCIFER1  TWO has the same meaning for  LUCIFER2  The second last column  state  indicates if a program is running showing a green dot  or a red dot if it is not running  Finally the last column indicates if a process input output panel  for the program exists  By clicking the  refresh table  button  the status of the system shown in  the state column is refreshed     The bottom section of Figure B I  indicated in orange color  shows the Process I O panel for the  selected program of the table in the middle section  All output that is normally printed by the    2 2  THE LUCIFER SOFTWARE CONTROL CENTER 9    zenter   start hd  xml                                                                                                                                                                                                                                                                                     Top      section    type use state  j system E Deamon  SYS  rmid 1098 9   v  system   amp  Service SYS  ConfigServer e
23. ensation Mirror Control Window    The Compensation Mirror service always has six tables loaded simultaneously  two rotation di   rections for each of the three cameras and the currently selected grating unit position  When the  grating unit position is changed  the service automatically loads the new set of six tables for this  new grating unit position  In the upper left area of the GUI the currently used lookup tables can  be displayed  Only one table is visible at a time  To view or edit a different table the drop down  boxes above the table can be used  Since the service has only six tables loaded for the currently  selected grating  only these six can be selected with the drop down boxes  The grating for which  the tables are displayed is shown above the drop down boxes  To edit tables for a different grating  unit position  the corresponding grating has to be selected first  see below      The GUI loads a copy of each of the six tables from the service at startup  which means that changes  made to the table in the GUI are not send to the service directly  When a table is modified  the    2 3  ENGINEER ACCESS TO THE MAIN INSTRUMENT COMPONENTS SW 17    File  Compensation Mirror Control    Lookup Table Control Panel   Mirror Control Panel    Flexure table used for  mirror Grating position Mirror control  Camera    N1 8Camera     Rotation   Clockwise                         a            mirror  v     mance unr   fold mirror four motor 1   fold mirror four motor 2      Commi
24. figuration  In total  24 lookup tables are needed  two tables  clock  and counter  clockwise rotation  for each combination of the three camera and four grating unit positions     Figure  2 8  shows the engineer access GUI for the compensation mirror service  It has many simi   larities to the GUIs used for the other instrument components  e g   FilterUnit  GratingUnit      so  some of the information described here is also valid for the other GUIs described in the following  sections  The Compensation Mirror Control GUI can be started using the Software Control Cen   ter  described in the last chapter  entry  GUI Compensation Mirror Engineer Client   After the  interface has been loaded it will look similar to the one shown in Figure  2 8    The window shows three main areas  found in nearly all engineer windows  the top left area shows  one lookup table that is used by the service  Next to the table in the top right area of the window   buttons and text areas for service control are located  Below these two components  a message  area is displayed  which shows all messages that are generated by the software  The  Automatic  Update  area at the bottom of the window can be used to alter the update interval of the displayed  information  The default is one seconds  The size of the message area can be changed using the  slider separating the upper from the lower part    On the right corner of the window a second  vertical slider is located that allows to open the motor  control
25. formation that is displayed  Values shown  with a purple background represent data that is currently used by the service  Values shown with  a white background have only been set in the GUI so far and have to be send to the service first   One example for this are the text fields for the rotator and elevation angles  the two fields on  the left are white and allow to type in new values for the corresponding angles  However these  new values are not used by the service until either the Return key is pressed  which sets the value  of the field the cursor is currently positioned in only  or the  Set Both    button is clicked  The  drop down boxes for the mirror and the grating are shown in purple color to indicate that the  information shown in these elements is actually set in the service  Changing the these values in  the GUI has immediate effect  The rotation direction is only shown for information  as are the  current motor positions  and cannot be set directly  The rotation direction of the instrument is  inferred from the change of the rotator values that are send to the service  with increasing rotator  angles representing clockwise rotation    Before the mirror can be used it has to be initialized  To be able to initialize the mirror  a  valid lookup table has to exist for the given grating and camera positions  In this regard  valid   means that the  relative  mirror position for the current instrument position must be able to  be calculated from the rotation and elevat
26. he filter unit service  The initialization simply tests  if the two filter wheels are in a  defined position  If this is not the case  the unit is moved until it reaches the next filter position    The filter that is in the optical beam is shown with a green background  When a new position is  selected  the desired position is shown in a light yellow color and the panel becomes inactive  It  is thus only possible to move one filter wheel at a time  The progress of the movement can be  followed in the GUI by different buttons turning green  when the corresponding filter is moving  through the optical path  When the new position has been reached the buttons become active  again     22 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    2 3 3 The grating unit service    low dispersion grating two   LUT position  Central Wavelength   Tilt Voltage  Grating Order   Tilt Ang m  0 1 982 34541     2 032  293  Y o          set tilt wavelength          er   a      v               v  system nstrument ho      2010 01 11 13 48 28 5 pes al  TimeClient syncOnServer 252   Synchronized on  lucifer 10 144 0 211  1 d             N imeDhent sypeOnServer282   Synchronized on lucifer 10 144 0 211  1 d           ete ONE  compensan eats dnServer 8   Synchronized on lucifert10 144 0 201y      BEER                  2     SSS       a         OT CONTENT  database  OR CONTENT  Synchronized        M                                      Figure 2 12  The grating unit engineer GUI       This service controls the pos
27. ile located in the  lt 1csp root gt   directory     DO NOT EDIT THIS FILE DIRECTLY       For the MOS unit a separate configuration file is used and maintained by MPE Garching     2 8  ENGINEER ACCESS TO THE MAIN INSTRUMENT COMPONENTS SW 15    2 3 Engineer access to the main instrument components SW    Each component of the instrument is controlled by a dedicated software service that works in   dependently from the others  but relies on functions provided by other services  In the following  these different services and the way to control them will be discussed  Note  Engineer access to the  different services is direct  i e   changing a parameter with the control GUI has immediate affect   In particular  the top level Instrument Manager service will be bypassed and is not notified of any  changes     To avoid unexpected behavior of the software  the Instrument Manager has to be  initialized again after the changes have been made at the engineer level     2 3 1 The compensation mirror service    Image movements on the detector due to flexure are compensated via a movable mirror that is  positioned by two independent motors  Depending on the position of the instrument  defined  by the rotator and elevation angles  the motors can be moved by a certain amount of steps to  reposition the mirror so that the position of the image on the detector remains fixed  The motor  step values are defined by lookup tables that are loaded by the service depending on the current  instrument con
28. ion angle values that exist in the table  The service  calculates the necessary mirror movements based on the current and next mirror positions in the  coordinate system that is defined by the table values    Two different initialization mechanisms are available  The first is an absolute initialization of the  mirror using the limit switches of the two motor axes  When the mirror is initialized in this way   both motors move until they reach the negative limit switch  and then back to a reference position  that is defined in the service configuration  This is useful when the mirror is in an unknown  position  i e   after instrument maintenance or if the mirror has moved a substantial amount out  of the reference position  The reference position is measured for the parked instrument with a  rotator angle of 341 degrees and an elevation angle of 90 degrees  So the initialization with the  limit switches should be carried out at this position  e g   during daytime or before taking flats in  the evening or in the morning  The second initialization does not move the mirror and just re   computes the relative mirror position based on the current instrument position  This initialization  is necessary whenever a new camera or grating has been selected in the GUI  When the instrument  is controlled from the Instrument Manager this is done automatically  of course  To switch between  the two initialization methods the check box  Force limit adjust  is used  When the box is set  the  ab
29. ions  mirror    three gratings   The Tnitialize    button is located there as well  The initialization procedure is in  principle analog to the filter unit service  If the unit is not in a defined position  the unit is moved  until the next defined position is reached  If the grating tilt loop is closed it will be opened during  the initialization  Note  moving the unit to a new position is only possible if the tilt loop is not  closed    The  Grating Tilt Control area is directly below the position buttons and allows to control the  tilt loop  Two different possibilities exist to tilt the gratings and can be selected by clicking on  the  set tilt voltage    or  set tilt wavelength  buttons  the former directly applies a set voltage to  the tilt control loop and is used for calibration measurements  for example  The maximum and  minimum allowed voltages are  4 9 and  4 9 volts  respectively  The second one uses a given  central wavelength and looks up the corresponding tilt voltage  depending on the chosen grating  and order  in the lookup tables  Wavelengths that are not present in the table will be interpolated  between the predecessor and successor elements  No extrapolation beyond the lowest and highest  wavelengths in the table is performed  Pressing the  Tilt  button closes the loop and applies the  desired voltage  either directly or calculated from the look table  to the tilt electronics  The  currently applied voltage values shown in the panel will begin to change an
30. it start descriptor menu item from the context menu  the following window  will appear  Figure 2 4   The text that is entered in the name field of the window is the one shown          name   HIRAMO Server use   Lucifer ONE Je  service name                          1098 HIRAMOServer    priority    200  execution time     2000   working directory      command          1    LDjava security  policy java policy all  Djava rmi activation port 1098  Djava rmi server codebase file l    jesp jar cp Icsp jar  command arguments   2   2      IesPJ          main class   de rub astro lucifer control electronic hiramo HIRAMOServer          Lruntime java  codebase file Icsp jar  J Djava security policyzjava policy all  unit 149 217 42 112 4009    program arguments           lesu     Figure 2 4  The start descriptor edit window that allows to change the start configuration and  parameters of a program  Shown is the start descriptor of an RMI service     12 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    in the table of the Software Control Center  see Figures  2 1  and 2 2   The service name entry  defines the URL of the distributed service  It must always be of the following format     protocol   hostname port ServiceName    The protocol is always    rmi     the host name is the name or IP address of the host where the  RMI registry service is running  The port number is the TCP port on that machine that the  registry uses  Finally the service name defines the name under which the service is bo
31. itioning of the grating unit and is also responsible for the correct  tilt of the gratings for spectroscopy  The user interface shown in Figure  2 12 can be accessed by  starting the  GUI Grating Unit Engineer Client  in the Control Center     The lookup table area    The lookup tables of the grating unit define the mapping between central wavelengths and tilt  voltage for each of the three gratings  The tables entries contain the following fields  the  Central    2 8  ENGINEER ACCESS TO THE MAIN INSTRUMENT COMPONENTS SW 23    Wavelength  is a key for the table and specifies the wavelength that will be focused on the center  of the detector  The  Tilt Voltage    is the corresponding voltage that must be applied to achieve  the desired central wavelength and must be within  4 9V and  4 9V  The  Grating Order    field  is the second key value in the table and defines the grating order that is used for the entry  For  the service the grating order is used as a filter for the entries to use when tilt voltages must be  interpolated  The only grating that is used in different orders is the high dispersion 210 zJHK  grating  The two other gratings are used in only one order  The  Tilt Angle    field is currently not  used by the software and for reference only      The control panel for the grating unit    Rhe control elements for the grating unit are located in the top right area  The area is split  into two sections  the upper section allows to select the four different unit posit
32. l viewer in and out     24 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    CameraUnitEngineerGul          Viewer Control     Set Camera Positions      PUPIL_VIEWER   N30 Camera    a N3 75Camera__           initialize Unit              Initialize Pupil Viewer                       Edit Configuration           a    System Messages  us    ALL x  PIT      LL    syste   ALL    instrume    ALL M deb     No  M                                     level   host   program   address                         Edit pupil viewer config     ilt                            database  OR CONTENT  Synchronized on                  Automatic Update  Curent update  every second  inactive        p      Set update time in seconds    al il Set update time      Start      Stop          Figure 2 13  The camera unit engineer GUI     2 3 5 Read out engineer GUI    This GUI  see Fig   2 14  is similar to the normal user read out GUI described in the LUCIFER  User Manual  but accesses GEIRS directly  More read out modes can be selected that are not  available for the normal user and GEIRS macros can be executed     ReadoutEngineerGUl    File Panel Read Out Panel System Button Panel  next file   volker_          Type    last file   unknown  not yet saved   Select Mode  DIT  ins    1984 L         Object  undefined          SavePath    home observer data 20100204 NDIT  n E  es Observer            Save   of reads   1 z  Show Display                C  integrated la     Abort             Show GEIRS Control GUI         
33. n is used  It will  calculate the sum of the two individual focus offsets from the lookup table and move the stage  accordingly  Below the automatic focus control panel  manual focus movements can be achieved by  typing in focus offset values in the text area and pressing the  Move Focus    button  This is useful  for testing purposes  for example  Finally the current focus position is shown below the buttons      The filter unit service     The engineer GUI for the filter unit service can be used to move the filter wheels directly  initialize  the unit and specify the properties of the filter in both filter wheels    The user interface can be started using the  GUI Filter Unit Engineer Client  entry of the control  center  The following window will appear  see Figure 2 11     The structure of the GUI is the same as for the other services described above  The lookup tables  are shown in the left area  For each filter wheel a separate lookup table is used  Each entry of  the table consists of 6 fields  The  Filter Position    is the key for the table and defines the position  of the filter  1 15  in the wheel  The  Filter Name    is shown on all buttons for filter selection   e g   the buttons of the engineer window on the right as wall as the buttons in the main instrument  GUI for the observers  The filter names should not contain any spaces  or the filter will not be  selectable from observation scripts  Additionally  the name for the filter that is specified here  defines 
34. ng on the exit button    The middle section  green in Figure 2 T  displays a table with the current startup configuration   The table is ordered by different priorities  where lower numbers correspond to a higher priority    The priority is shown in the first row of the table  The second column determines when the corre   sponding program is started  Two values are possible  system and manual  If system is selected   the program is started when the whole software system is launched  If manual is selected  the  program is not started with the system start  but can be started manually at a later time  This  allows deactivating programs that are not needed or wanted at system start time  The third col   umn indicates different program types by showing them in different colors as well as with different  names  For the first of these types  the RMI Daemon  pink color   only one entry is allowed to exist  in the startup configuration  This entry must have the highest priority and thus be displayed in the  first row of the table  The second type  displayed in blue  describes startup configuration entries  that represent the distributed services of the software  These services are the main components for  operating the instrument and should be started together with the software system  i e   the entry in  column two should be system   The third type shown in green indicates normal Java applications  that usually start different graphical user interfaces for software control  These ne
35. o handle the LUCIFER Control Software Package  Icsp   It contains  information on how to start and stop the software or parts of it  change configurations and use the  dedicated engineering tools    Misconfigured configuration entries can severely damage the instrument     Chapter 3 contains the description of some instrument specialist procedures as well as a short  trouble shooting guide for most common failures     CHAPTER 1  INTRODUCTION    Chapter 2    Technical software user manual    2 1 Starting the software   Summary    Under normal conditions the Lucifer control software  LCSP  is running continuously on the SUN  V880 workstation which is located in the computer room close to the observing room behind the  glass door  After a restart of the workstation a software restart is of course necessary     1  Login to sun workstation lucifer as user engineer     2  Execute the Start LUCIFER Control SW which will bring up the LUCIFER Software  Control Center     3  If software is not already running press the start button and wait  approx  2 minutes    4  Open the Main Observer GUI and press the Initialize button   5  You are ready to observe with the instrument     A more detailed description follows in the next chapters     2 2 The LUCIFER Software Control Center    This chapter describes how the software system controlling the LUCIFER instrument can be  started  stopped and how it can be configured     2 2 1 Starting and stopping the software    In order to start the softw
36. put 8  OFF  M1 OUT swingarmup BW 55 Output 9 ler     M2 IN unitturned in  OFF 41 Output 10 lorF  3AC OK  ONE is Output 11 bel    last update  Sat  KWO05  31 01 2004 20 31 26  Reload  Reload Web O Digital Homepage a  mne  A  Done             Figure 3 4  On the web page of the WebIO 2 the calibration unit can be controlled  The  Inputs   column show the current status  here power     on and position     out  and the  Outputs  column  the available commands power on  power off  drive in  drive out  drive stop    Observer GUIs    In case one of the observer GUIs is not working properly  refreshing problems  a restart of this  GUI may help  This can be done by using the different start scripts in the  home observer bin  directory  start  IMGUI sh   start RMGUI sh  start TMGUI sh     Read Out GUI    In order to have an error handled procedure which copies the LUCIFER  data to the archive  a  kind of rsync daemon has been implemented into the software  When choosing a new  empty save  directory for the LUCIFER images via the Read Out GUI  be sure that as long as there is no first  image in the directory the rsync will produce an error message that it can t copy   fits  As soon  as there is a file written to this directory the message will no longer appear     Engineering GUIs    It may be necessary to resize the engineering GUIs to draw them correctly     34 CHAPTER 3  TECHNICAL PROCEDURE MANUAL    3 2 2 Hardware related  Pressure Error Message    From time to time an error message 
37. re on  bottom right information   see section  3 1 1         Start LUCIFER SW Control       Check that all needed servers are started  e g  telescope        Currently all servers except  the TemperatureMonitorLUCIFER2     TurboPumpMonitor should be started    Open the Main Telescope GUI and make sure that UT and LST times are beeing updated   If that is not the case then there is no data coming from the TCS subsystems   so check and  solve that problem  If the TCS will be down for a longer period  please stop the telescope  service using the software control center  Nevertheless each      or TCS restart  reboot needs  also a restart of the LUCIFER Telescope Service     Ideally also perform an instrument set up  from the instrument control     32 CHAPTER 3  TECHNICAL PROCEDURE MANUAL    3 2 Trouble Shooting Guide    3 2 1 Software related    Starting the software package    After pressing the start button and waiting for all services  it may happen that the MOS Sequence  Server still has a red light meaning it has not started properly  Therefore it has to be done by  hand  Click in the table on the MOS Sequence Server row  be careful because due to the nx  connection the refresh of the table is a bit slow   Use the right mouse button and choose start  MOS Sequence Server  Wait a few seconds  press refresh table and check if the service is now  running     Calibration Unit    The error  Calibration Unit cannot move    can be ignored as long as the the status in the Instrument  
38. se this grating only without any wavelength input     MOS Unit    As soon as the MOS status in the Instrument Control Panel turns to  unknown  the MPE MOS  experts have to be called  No other use of LUCIFER is allowed until the problem has been solved   For more details see the MOS user guide delivered by MPE  Garching  LBT LUCIFER MAN 027      Error  Committing new MOS setup results in no action after pressing the commit button   Stop  amp  start the Instrument Manager from the Start  Tool in the engineer session     After a restart of the MOS server software the Instrument Control GUI will show the MOS panel  in red  due to a lost connection to the server  The connection will be reloaded by clicking on the  Initialize button     
39. shows the current  version number of the software and indicates starting of different programs  Depending on the  used configuration and the number of programs that should be started  this process can take up  to two minutes or longer  After the software system has been started successfully  a screen similar    10 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    to the one shown in Figure 2 2  will be displayed  As can be seen in the figure  the start button is  now deactivated and the stop button is active  When the stop button is clicked a dialog window  asking for confirmation is displayed to avoid accidental stopping of the software system  When the  window is closed by clicking on the exit button or the close icon and the software system is found  to be running  a dialog window is shown asking to stop the software system also  In either case  if  the no option is selected  the RMI services are kept running and only the applications  i e   GUIs   are closed     Currently  18 01 2010  all services  blue colored  should show a green light but the Temperatur   MonitorLucifer2 and TurboPumpMonitor services which show a red lamp indication after  the start     File Add      system status  STARTED    services                priority  stat   type   use  0 system  i RMI Deamon  55     tem    Servi  svs  co                                  titt  IES                         manual  9 Service    system    Service  system    Service                               system    Service  system
40. solute initialization is carried out    To manually compensate the flexure  i e   to test the flexure tables  the  Adjust Mirror    button is  used  If the  Show adjust first  check box is selected the calculated motor steps are shown first   before the mirror is actually moved  The steps to move are calculated based on the angle values  used for the last compensation  or initialization of the mirror  and the new ones  Finally  the   Reset Mirror    button reverts any mirror movements made since the last initialization    Below the mirror control area of the window the button    Edit configuration       is located  It can be  used to access the configuration values of the service conveniently from the engineer interface and  thus provides another way to alter the setup of a service from the one presented in the previous  chapter  However  only the  direct  configuration values of the service can be accessed using this  button  no configuration values used for contacting other services that the compensation mirror  service depends on  like the motor control service  are accessible     2 3  ENGINEER ACCESS TO THE MAIN INSTRUMENT COMPONENTS SW 19  2 3 2 The detector unit  focus  service    The service controlling the detector focus drive is similar to the one for the compensation mirror   It is more simple because only one motor has to be controlled to position the detector correctly for  a given combination of filters in the beam  Figure  2 10  shows the engineer GUI for the de
41. t to service H Reload from service C Force limit adjust                                        Osec            find    LUT position Instrument Position   Motor 1 steps  Y axis    Motor 2 steps  X TR ge UEEA    tame   0 R 90 0 E30 0       0 la   Rotation direction         position   new position        Camera motion settings   REMOTE  FILE  Mome engineer Acspidistiexport  tookupTablesiN1_8_ClockW    st    al           eps          IN1 8 clockwise rotation Sept 2009 MIRROR        5                 Adjust Mirror   pam              current  move       EE EI not moina                 Clockwise velocity  199 9923           new velocity  200             Set Angles  Rotator angle   00 341 01  i   Set Both  Elev  angle   00  0200      settings             status    MT     Current motor positions positive limit       dat   Rotator angles interpolation    Linear interpolation Function motor 1  46297 negative limit no data  iiia aiii MN e d pue AS                      Elevation angle interpolation   Linear Interpolation Function   motor 2  50983          Ec      2j            show all   update      Edit Configuration                                               jw    error   All     system   ALL  v instrument                   tevel   host   use program address message attachment                              filter  NOT CONTENT  database  OR CONTENT   Synchronized on           Automatic Update       Curent update  every second  inactive   Setupdatetimeinseconds       1   Set update time
42. t updated a restart of the  LUCIFER Telescope Service is necessary to get it running again    In case of any not running TCS IIF the LUCIFER Telescope Service has to be stopped  immediately      Error in Subunit    Any errors in a subunit of the instrument is reported by an error message  an error pop up window  and a red background color of the subunit panel  In order to be functional again  pushing the  Initialize button is necessary     3 2  TROUBLE SHOOTING GUIDE 33                                                                                                          File Edit View History Bookmarks Tools Help d       gt   El 0 nm ntpuns2168 012 ndex2 htm x   gt      Google  E     Getting Started E  Latest Headlines 05 localhost   localhost    05 LUCIFER Monitors Ov        W amp T    luci     eb  aux luci1  Web1O Digital  10 10087    LUCIFER1 Aux Rack Power Unit   57630  Home  Sj User Inputs Outputs      E  Config   Name State  Counter Name   State  a 23 Aux Rack Power Status OFF 0 Calibration Unit power on  OFF  m MOS Aux Power Status OFF  0 Calibration Unit power off OFF  Calibration Unit power status ON 16 command    drivein            Calibration Unit position IN  OFF 41 command    drive out    OFF  Calibration Unit position OUT BAI 55 command    drive stop     OFF  Calibration Unit motor IN        las Output 5   OFF  Calibration Unit motor OUT  OFF 46 Output 6  OFF  Calibration Unit Position Error   OFF  30 Output 7   OFF    Calibration Unit E stop status  IN o Out
43. tector  focus control  which can be started with the control center  previous chapter  using the    GUI  Detector Unit Engineer Client  entry in the table  The structure of the window is the same as for  the compensation mirror  the top left shows the lookup table data  the top right the service control  buttons  In the bottom area a service panel is located     tor camera   N375Camera      Commit table to service Reload table from service    N3 75 Camera E   POSITIVE LIMIT SWITCH       F    r  3   gam  a  Adjust F                 v           v             v  aemus fio        Em M T       er  NOT CONTENT  database  OR CONTENT  Synchronized        M       Figure 2 10  The detector focus engineer GUI     20 CHAPTER 2  TECHNICAL SOFTWARE USER MANUAL    The lookup table area    The detector focus service used three lookup tables containing the focus offsets for a filter from the  reference position  For each camera a separate lookup table can be used but it has been proven  sufficient to use only one table for the N1 8 and N3 75 cameras  To configure this  the filenames of  the lookup tables for the two cameras have been set to the same value in the service configuration   The lookup table has two entries  the  Filter Name    is the key for the table and has to be specified  exactly as in the filter unit service described in the next section  Specifically  no spaces           should  be present in the filter names for proper instrument operation from scripts  The second parameter
44. the name that must be used for the detector focus service  see previous section   The   Cut on wavelength  and  Cut off wavelength  are defined to specify the wavelength  in microns   where the filter becomes transparent and opaque again  respectively  Values in these fields are  currently not used by the software  but can be used in a future release to prevent unsuitable filter  combinations  The last two entries are used to identify the    clear    and    blind    filter positions in the    2 3  ENGINEER ACCESS TO THE MAIN INSTRUMENT COMPONENTS SW 21    feruserinterface gui instrument filte             Cut on wavelength   Cut off wavelength   clearfilter   blind filter  0 LI             20H 1060   3clear   4blind   51   8Y2   TP  gam  beta                                                                                                                                                            005                                   program address message       bl      IOT CONTENT  database  OR CONTENT  Synchronized on                          Figure 2 11  The filter unit engineer GUI    filter wheel  This information is necessary for the correct operation of observation scripts  It is not  necessary to define filters for each position in a filter wheel  Positions for which no information is  provided will be shown as Undefined        The  Set Filter Position  area     The left area of the user interface allows to move the filter wheels into a new position and to  initialize t
45. to the LUCIFER workstation has  to be build up first  using the engineer user password  Firefox  or a Web Browser  should be  opened and the  LUCIFER1 Web IO Control    bookmark selected     From operational to safe mode    First at laptop workstation  switch ROE and MCE via Web IO Interface off     Switching MCE ROE on off is the button on the lower right side   To set the detector to the safe temperature  the following procedure has to be used     e The software tool can be started either a  by clicking the desktop icon DetectorTemperature   Control  or b  by calling the command    detectorTemperatureControl sh    from a terminal     e Select option    2     Fig  B I  and confirm the setting  For a detailed description of the software tool  refer to LBT LUCIFER TN 016     27    28 CHAPTER 3  TECHNICAL PROCEDURE MANUAL          Eile Edt Yew Termmal Tabs Help       Figure 3 1  Selection window for detector temperature control    From safe to operational mode    Make sure the ROE  amp  MCE electronics are powered off      To set the detector from safe to operational mode  the following procedure has to be used     e The software tool can be started either a  by clicking the desktop icon  Detector Temperature   Control     or b  by calling the command  detector TemperatureControl sh  from a terminal     e Select option  1  Fig  3 1  and confirm the setting  For a detailed description of the software tool  refer to LBT LUCIFER TN 016     After reaching the temperature of 77K  NO
46. type   int       VALUE name   GRATING  UNIT  TARGET  VOLTAGE  PARAMETER  POS  INITIALIZATION  DELAY  TIME   3 VALUE name   GRATING  UNIT  CURRENT  VOLTAGE  PARAMETER PQ    int      VALUE name   OPEN  LOOP  OUTPUT  VOLTAGE PARAMETER  POSIT           VALUE name   TEMPERATURE  MODE  PARAMETER  POSITION  type     D  VALUE name   LOOP  STATUS  PARAMETER  POSITION  type   int RETRY  TIME   3 VALUE name          OFFSET  ERROR  PARAMETER  POSITION  type   i int     VALUE name   DEBUG  BIT  PATTERN  START  POSITION  type   int      VALUE name   DEBUG  BIT  PATTERN  LENGTH  type   int E      VALUE name   GRATING  UNIT  BIT  PATTERN  START  POSITION             UPDATE  INTERVAL      VALUE name   GRATING  UNIT  BIT  PATTERN  LENGTH  type   int    long       VALUE name   FILTER  WHEEL  ONE  BIT  PATTERN  START  POSITIOI      VALUE name   FILTER  WHEEL  ONE  BIT  PATTERN  LENGTH  type   in      VALUE name   FILTER  WHEEL  TWO  BIT  PATTERN  START  POSITIOT HIRAMO CARD NUMBER   0 VALUE name   FILTER  WHEEL  TWO  BIT  PATTERN  LENGTH  type   ir int  D  VALUE name   CAMERA  WHEEL  BIT  PATTERN  START  POSITION        D  VALUE name   CAMERA_WHEEL_BIT_PATTERN_LENGTH  type   int        de rub astro lucifer control electronic hiramo ConfigHIRAMOServer                                                                                           SWITCH  PORT  VALUE PARAMTER POSITION    Figure 2 7  The edit configuration window     The configuration of the software services is stored in the config xml f
47. und to the  registry     The priority defines the position of the entry in the table  with higher numbers representing  lower priorities  All programs with the same priority are started simultaneously  The starting of  lower priority programs is suspended for the number of milliseconds that is given in the execution  time field  The meaning of the working directory field is straight forward  and the command field  defines the executable that should be started  In principle any command line argument can be  defined in this field  For Java services and programs the class that should be started by the Java  interpreter must be defined in the main class field  The text areas for command arguments and  program arguments allow defining these arguments which are passed to the command at execution  time     ReadoutManager          Name   ReadoutManager   Host  lucifer  Usage   Lucifer ONE IP Address   10 144 0 211    Version   30 11 2009      Up Time  B hours 6 minutes 16 seconds    Time Drift    Synchronisation   enabled                                          Command Line Parameter   key value   PROGRAM_NAME ReadoutManager   PROGRAM_USAGE 1                      Configuration    de rub astro lucifer control readout ConfigGEIRSService  de rub astro lucifer operation readoutManager ConfigReadoutManager  de rub astro lucifer operation readoutManager ConfigReadoutManagerService  de rub astro util ConfigUtil   de rub astro util message ConfigMessageClient                 a              
48. will show up in the message panel  Unable to get instrument  pressures     This message can be neglected     Camera Problem    When moving the camera to a new position  mostly the N1 8 position  it can happen that it does  not reach its final position  An error message and a pop up window will appear reporting that  the camera is not initialized  In most cases pressing the Initialize button will help to bring the  camera into the right position     Filter Wheel Problem    When moving the filter wheels it can happen that they do not reach its final position  An error  message and a pop up window will appear reporting that the camera is not initialized  In most  cases pressing the Initialize button will help to bring the filter wheel s  into a defined position     Grating Unit    If the wavelength input is out of the allowed range  an error setting the grating will appear  error   no predecessor  successor element      That means  the tilt is out of range  The grating panel will  turn to red  Press Initialize to recover     At high elevation angles or when the rotator is moving it may not be possible to stabilize the  grating at the desired angle  error  grating tilt did not stabilize after 60s     Iry again or wait until elevation is lower     BE AWARE    NO wavelength input possible for the 150 Ks grating since this grating is currently clamped at a  fixed angle  If you have set a wavelength for this grating you have to wait the timeout time  60s   to get the error message  U
    
Download Pdf Manuals
 
 
    
Related Search
 LUCIFER  lucifer  lucifer cast  lucifer lighting  lucifers pizza  lucifer morningstar  lucifer donghua  lucifer hazbin hotel  lucifer tv show  lucifer painting  lucifer valentine  lucifer meaning  luciferianism  lucifer movie  lucifer sigil  lucifer angel  lucifer bible  lucifer hh  luciferase assay  lucifer netflix  lucifer morningstar hazbin hotel  lucifer supernatural  lucifero  lucifer\u0027s friend  lucifer helltaker  lucifer band 
    
Related Contents
Yosemite Home Decor MAG503R Installation Guide : Free Download, Borrow, and Streaming : Internet Archive  950 NORTH HAGUE AVENUE TEL:(614)    TC322-TC328.indd  Progress Lighting P4122-42 Instructions / Assembly  Programme  RSD-SDD取扱説明書[PDF:314.5KB]  View PDF - Smart Computer Systems  HI 98103 Checher® - Hanna Instruments Canada  GPS-Tracker GT-170    Copyright © All rights reserved. 
   Failed to retrieve file