Home
        Abberior Instruments Pulsed Diode Laser PDL 561 / PDL 594
         Contents
1.                      __               p     n           UR   e m    The use and function of the CW Mod   Temp  Ctrl  and Analog Mod  inputs is described in later sections of  this manual     3 2 Startup Procedure  1  Turnon the power switch to the  I  position  The Power LED should light up     2  Turn the key switch to the  1  position  The Laser Emission LED should light up  If a trigger signal  is provided  the laser should now be emitting     3  If the laser head is equipped with a manual safety shutter  stand alone version only   open the  shutter as follows        3 3 Power Modulation  Pulsed Mode     When the laser is operated in pulsed mode  the pulse energy can be adjusted using the Analog Mod  input  on the back of the laser control unit  In order to use the power adjustment a control voltage between  0 and 5V must be applied to the Analog Mod  SMA connector  Please note that the maximum output  power is reached at approx  2 85 V and decreases again when then control voltage is further increased     3 4 Fiber Coupling    Abberior Instruments offers the PDL as a free space laser  However  the laser beam may be coupled into a  singlemode or polarization maintaining optical fiber  On the front face  the laser head has four M2 5  threads for the direct attachment of a fiber coupling unit  e  g  from the 60 SMS series  with mounting  adapter  by Schafter  amp  Kirchhoff  To mount a fiber coupler the manual safety shutter has to be removed  first     Last change  2015 04 
2.    o o 14  10 2 Ecologicalbackground 202 0 LLL 14  11 Disclaimers        14  Last change  2015 04 08 Page 2 14    Subject to change without notice     1 Introduction    With its PDL series of lasers  Abberior Instruments offers picosecond pulsed laser modules mainly  targeted for applications in fluorescence imaging and time resolved spectroscopy including time   correlated single photon counting  TCSPC   The lasers are triggered from an external source  for example  from a second pulsed laser system or a pulse delay generator  Alternatively  they offer continuous wave  operation     The lasers are offered as stand alone units or as part of an Abberior Instruments microscope  system  In the first case  the laser is delivered with a small table top control unit containing the controls  for its operation  power switch  key switch  status LED s  and an external power supply  If the laser is  ordered as part of a microscope system  the controls are integrated into the 19  electronics rack     4 Diode Las    Pulse       oe EM     Fig  1  Stand alone control unit for the PDL laser head     2 Laser Safety    LASER RADIATION  AVOID EXPOSURE TO BEAM  CLASS 3B LASER PRODUCT       The Abberior Instruments PDL lasers are Class 3B  IEC  laser products that emit  less than 500 mW of laser radiation in the visible spectral range     2 1 Safety Features    2 1 1 Interlock    The laser is equipped with an interlock circuit that prevents current flowing through the diode  unintentionally     2
3.   Place a laser power meter in front of the laser aperture and  adjust the temperature control voltage until the maximum output power is achieved  Typically  a voltage  near 3 2 V gives the best performance     4 Trouble Shooting    Problem Possible solution          The laser does not emit  a  Make sure the interlock input is closed   b  Check that the trigger pulses are within the  specified range        The maximum laser power      CW mode remains   Make sure the Temp  Ctrl  input is tied to GND   far below the specifications                 Last change  2015 04 08 Page 7 14  Subject to change without notice     5 Specifications    General Specifications  Operation mode    Wavelength    Temperature control input   Polarization    Polarization extinction ratio   Beam divergence  FWHM    Output beam quality  M        Pulsed Operation  Repetition rate   Average power   Pulse energy   Pulse length   Pulse trigger   Trigger pulse amplitude   Trigger pulse length  FWHM    Trigger pulse rise time     Analog modulation input     Continuous Wave  CW  Operation  Output power   CW control input     pulsed or cw    PDL 561   561 2 nm  PDL 594   594  2  nm    0 5V  horizontal  1 1000  10   5    0    61   1 2    0   100 MHz    gt  100 uW   40 MHz  130 uW typ     gt  2 5 pJ  3 pJ typ        150 ps  100 ps typ     external   2 5V    lt 5ns     2ns    0 5V    2 mW  5 mW max     0  10V      The PDL must not be operated at output powers above 2 mW for extended periods of time     Power Suppl
4.  1 2 Manual Shutter  Optional   PDL laser heads purchased as stand alone units are equipped with a manual safety shutter  For safety    Last change  2015 04 08 Page 3 14  Subject to change without notice     reasons but also to protect the optics from dust it is recommend to close the shutter whenever the laser  Is not in use     2 1 3 Key Switch    The stand alone versions of the PDL come with a control unit which must be connected for the laser to  operate  When the key is      the  0  position  the diode is prevented from emitting  The key must be  actively turned to the  1  position each time the laser is powered on     2 1 4 Status LEDs    The control box which is part of the stand alone version includes information LEDs which displays  whether power is connected  the laser is on  The  Laser Emission  LED indicates that the device is  emitting or could emit light     2 1 5 Warning Labels    The laser head has yellow laser warning labels clearly indicating the beam aperture and specifying the  wavelength and optical power emitted from the laser head     LASER RADIATION     AVOID DIRECT  EXPOSURE TO BEAM    LASER POWER     5 mW  pulsed      100 mW  cw   PULSE LENGTH     150 ps    WAVELENGTH  561 nm  CLASS IIIb LASER PRODUCT       2 2 Safety Precautions    e Eye and skin exposure to direct or reflected laser light is hazardous and may be extremely  harmful  Always wear eye protection appropriate to the beam wavelength and intensity  The  device must only be operated by prop
5. 08 Page 6 14  Subject to change without notice     3 5 Continuous Wave  CW  Operation    In order to operate the PDL in cw mode  the Temp  Ctrl  input must be tied to GND or set to a 0 V input   Otherwise  the laser does not operate at its optimum and the output power will not exceed 200   300 WW     The output power in CW mode can be controlled using the CW Mod  input on the back of the laser control  unit  A control voltage between 0 and 5 V must be applied to adjust the laser output power  It is normal  that the laser output power decreases again after the maximum power has been reached and the control  voltage is further increased     Note  No trigger pulses should be applied while the laser is operated in cw mode as they may cause a  power modulation     3 6 Temperature Adjustments    The PDL laser contains temperature sensitive devices  and in order to run at full power the operating  temperature must be kept constant  At the factory  the laser is carefully adjusted for optimal  performance at room temperature  21   C   Usually  the temperature setting does not need any  adjustments and the Temp  Ctrl  input on the control unit can be left open  If the laser is operated at  significantly higher environmental temperatures or a drift in the temperature setpoint has occured the  temperature may be controlled externally     If the laser is not running at its specfified output power  a temperature control voltage between 0 and 5 V  may be applied to the Temp  Ctrl  SMA input
6. 4  Subject to change without notice     
7. abberi  ior  INSTRUMENTS       Abberior Instruments  Pulsed Diode Laser    PDL 561   PDL 594    Operating Manual       Last change  2015 04 08 Page 1 14  Subject to change without notice     Contents    1 Introduction ________________    1 LE  2 Laser Safety     1 1 00 0 00 0 0 o LLLA 3  2 1 Safety Features     11 1 00 0000 0000 LLL 3  2 1 1 Interlock O 3   2 1 2 Manual Shutter  Optional  22 3   2 1 3 Key Switch O 4   2 1 4 Status LEDs                   4   2 15 WarningLabels              2 2 LLL 4   2 2 Safety Precautions 1 00 00000 00 0 0 0L 4   3 Operating Instructions       1 02 00 02 0 2 LLL 5  3 1 Laser Installation 5  3 1 1 Package Contents se 5   3 1 2 Mounting the Laser Head      20 2 0 0 2 LLL 5   3 1 3 Electrical Connections  gt  SSO 5   3 2 Startup Procedure 6  3 3 Power Modulation  Pulsed Mode  o LLL 6  3 4 Fiber Coupling        6  3 5 Continuous Wave  CW  Operation 22O T  3 6 Temperature Adjustments           2 002 0 22 2 LLL      4 Trouble Shooting s OF  5 Specifications 22O 8  6 Dimensions        10  T Pin Assignments                          11  7 1 Laser Head Connector 1 1 0 00 0 00 0 LLL 11  1 2 Power Connector                                    2 11  7 3 Interlock Connector        1 00 0 000000 0 LLL 11   8 EC Declaration of Conformity         J              22 LLL 12  9 Warranty  13  9 1 Restriction of Warranty 1  9 2 Intellectual Property 222213  10    End of Life    Policy  WEEE  O 13  10 1 Waste treatment on your own responsibility            J        
8. e the defined  ambient conditions stated by us or unauthorized maintenance     Abberior Instruments reserves the right to change this operation manual or the technical data of the  described unit at any time     9 2 Intellectual Property    Any changes of the beam path or any addition of new laser sources might conflict with local  third party  intellectual property  The owner has to make sure that no local  third party IP is violated when changing  the configuration of the system     10    End of Life    Policy  WEEE     As required by the WEEE  Waste Electrical and Electronic Equipment Directive  of  the European Community and the corresponding national laws  Abberior  Instruments offers all end users      the EC the possibility to return    end of life     units without incurring disposal charges        This offer is valid for Abberior Instruments electrical and electronic equipment provided that it was sold  to a company or institute within the EC and still complete  not disassembled and not contaminated  Left   over parts such as PCB s  housings etc  are excluded from the take back service     If you wish to return an Abberior Instruments unit for waste recovery  please contact Abberior  Instruments GmbH for further information     Last change  2015 04 08 Page 13 14  Subject to change without notice     10 1 Waste treatment      your own responsibility    If you do not return an  end of life  unit to Abberior Instruments  you must hand it to a company  specialized in was
9. erly trained personnel with experience in laser handling  in  a laboratory environment and with access to adequate laser safety equipment  The laser module  clearly displays a yellow warning label that shows the location of the laser beam aperture  This  label must be visible unless the laser beam is totally enclosed     e Always switch off the laser controller before moving or removing the laser head     e Always install the laser system to a properly grounded power outlet because the laser is sensitive  to electrostatic discharge  ESD      e Always mount the laser head in a way that the laser beam cannot cause any eye damage  Use  beam stops where appropriate  If the laser is fiber coupled make sure the fiber coupler is  attached securely to the laser head                laser head does not contain any user servicable components and must not be opened by the  user at any time  Please note that any warranty is void when the laser head or the controller box       opened     Last change  2015 04 08 Page 4 14  Subject to change without notice     3 Operating Instructions    Caution     use of controls or adjustments or performance of procedures other  than those specified herein may result in hazardous radiation exposure        3 1 Laser Installation    3 1 1 Package Contents    Before proceeding with the installation of the laser please check that the package you received is  complete  It should contain the follow items     1 Laserhead   Laser control unit   Power supply  inpu
10. ng with the date of shipment  The hardware warranty covers the laser head and the control unit   During this warranty period Abberior Instruments will see to defaults by repair or by exchange if these are  still entitled to warranty  In case that for warranty repairs or service the unit must be sent back to  Abberior Instruments  the customer will carry the shipping costs back to Abberior Instruments  In case of  warranty repairs Abberior Instruments will carry the shipping costs back to the customer  If after  inspection of the parts for repair no warranty repair is applicable the customer will also carry the costs  for back shipment  If the unit is sent back to Abberior Instruments from abroad the customer will carry all  shipping costs  duties etc  which should arise for sending the goods back to Abberior Instruments     Abberior Instruments warrants the hardware and software determined by Abberior Instruments for this  unit to operate without fault provided that they are handled according to our statements  However   Abberior Instuments does not warrant a fault free or uninterrupted operation of the unit  of the software  or firmware for special applications nor this operation manual to be fault free  We will not carry  responsibility for ensuing damages     9 1 Restriction of Warranty    The aforementioned warranty does not cover errors and defects being the result of improper treatment   software and interface not supplied by us  modification  misuse or operation outsid
11. t  100 240 VAC  1 A  50 60 Hz  output  24 V  30 W   Interlock shunt   Keys   D Sub cable  9 pins    Power cable    Hb e e N               User manual     IF you received the laser as part of an Abberior Instruments microscope system  the components are  included in the electronics unit      3 1 2 Mounting the Laser Head    The laser head provides mounting threads for both horizontal and vertical installation which allows it to  be used with its polarization oriented vertically or horizontally  When mounted horizontally  the logo is  upright on the right hand side of the laser head  the polarization of the laser beam is horizontal     3 1 3 Electrical Connections  1  Connectthe laser control box with the laser head using the supplied 9 pin cable     2  Connect the external trigger source to the SMA connector on the back of the laser head  pulsed  operation only      3  Connect the interlock input of the laser control box to an external interlock switch  e g  of a  laboratory interlock system  or  if not used  place the provided interlock shunt into the interlock  socket     4  Connectthe 24 V power supply to the laser control box     The remaining SMA connectors on the laser control box are not required for the operation of the laser  and  if not used  should be left disconnected  In this case  the laser operates in pulsed mode at its  maximum output power     Last change  2015 04 08 Page 5 14  Subject to change without notice     at       Pulsed Diode Laser  Interlock    O   
12. te recovery  Do not dispose of the unit in a litter bin or at a public waste disposal site     10 2 Ecological background    It is well known that WEEE pollutes the environment by releasing toxic products during decomposition   The aim of the European RoHS directive is to reduce the content of toxic substances in electronic  products in the future     The intent of the WEEE directive is to enforce the recycling of WEEE  A controlled recycling of end of live  products will thereby avoid negative impacts on the environment     11 Disclaimers    Abberior Instruments will assume no responsibility for damage incurred by faulty customer equipment   such as measurement equipment  cables etc  used in conjunction with Abberior Instruments lasers   Abberior Instruments makes no warranty of any kind with regard to the information contained in this  guide  included but not limited to  implied warranties of merchantability and suitability for a particular  purpose  Abberior Instruments shall not be liable for errors contained herein nor for incidental or  consequential damages from the furnishing of this information  No part in this manual may be copied   reproduced  recorded  transmitted  or translated without the express written permission by Abberior  Instruments     Abberior Instruments GmbH  Hans Adolf Krebs Weg 1  37077 Gottingen  Germany  phone  49 551 30724 170  fax   49 551 30724 170  www abberior instruments com  info abberior instruments com    Last change  2015 04 08 Page 14 1
13. trol  8 V 10V  8 GND GND GND  9 Interlock   GND  GND GND   7 2 Power Connector  Pin Voltage   1 GND  2 N C   3 N C   4 N  C   5 N  C    IP 40  8 pin     SND  T GND  8  24 V                   7 3 Interlock Connector                 GND      AM       L Interlock     Last change  2015 04 08  Subject to change without notice     Page 11 14    8 ECDeclaration of Conformity    Manufacturer  Abberior Instruments GmbH  Address  Hans Adolf Krebs Weg 1  37077 Gottingen  Germany    Product  Pulsed Diode Laser PDL 594   PDL 561    We declare that the listed product complies with the following European Directives   e 2004 108 EG Directive on Electromagnetic compatibility     2006 95 EG Directive on Low voltage equipment    The product conforms to the standards     e EN61326 1  2006  Electrical equipment for measurement  control and laboratory use   EMC  requirements   part1 general requirements    e EN61010 1  2001  Safety requirements for electrical equipment for measurement  control and  laboratory use   part1 general requirements    e EN60825 1  2007  safety of laser products   part1 equipment classification and requirements    This declaration shall be ceased to be valid if any modifications are made to the product without our  approval     Gottingen  Germany Manager  2014 09 11 Gerald Donnert  Last change  2015 04 08 Page 12 14    Subject to change without notice     9 Warranty    Abberior Instruments GmbH warrants dedicated hardware parts of the PDL for a period of 12 months  starti
14. y  Power supply  controller unit      Power supply  laser head      Power consumption     Operating Conditions    Last change  2015 04 08  Subject to change without notice     24V  1A     5 V  100 mA   12 V   lt  500 mA      5W    Page 8 14    Environment  For indoor use only  Ambient temperature  21  C    Relative humidity   lt  80   at 30   C    Mechanical Specifications    Dimensions  L x W x H   86 mm x 50 mm x 50 mm  laser head with manual shutter   78 mm x 50 mm x 50 mm  laser head without shutter   124 4 mm x 103 mm x 53 2 mm  control unit     Weight  365 g  laser head   360 g  control unit     Last change  2015 04 08 Page 9 14  Subject to change without notice     Imensions    6 D    Page 10 14      woassn   AssvDGN            e  j  S1N3IAQHLSNI    g OL                           ON        3aoo 3ovo  azis                          dI9ONV           NOI LdI HOS3Q HO    ON ONIASILNAG   SFYNLVIONAWON HO lHVd     uu Ul suoisuaulp         09       uus A L OXYIW sail    Q3393H9 3iva A8 dHOONI NOLLdIMOS3Q  AYOLSIH NOISIA3H    Last change  2015 04 08    Subject to change without notice     7 PinAssignments    7 1 Laser Head Connector          SUB D connector  male  9 pins   view onto laser head        Absolute Maximum Ratings                                                       Pin Function Min Max  1   12VSupply Voltage  0 3    40 V  2      OV 5 5 V  3   Analog Modulation  8V 10V  4 Laser Status Output 2 5V 5 5V  5 Interlock  GND High Z  6 CWControl  15V 15V  7 Temperature Con
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Télécharger la plaquette  Samsung 23 collu FHD monitors ar īpaši kvalitatīvu attēlu Lietotāja rokasgrāmata  Audiovox D1708PK User's Manual  Sharp MX-M264N Brochure  Troubleshooting Guide  Targus USB Cable - PACMB010U, PACD010U, PADVD010U & PADVW010U  Microlab FC 50  Controle Remoto Universal Xbox 360 - Center    Copyright © All rights reserved. 