1. Gaymar Industries Inc LU S PATENT 4 058 870 1998 Gaymar industries Inc All rights reserved OPERATOR S MANUAL TP500ITP500C TIPUMP SAFETY PRECAUTIONS SAFETY 4 DANGER PRECAUTIONS Risk of explosion Do not use in the presence of flammable anesthetics Risk of electric shack Disconnect power before servicing the T Pump A WARNING This device pumps warmed water through a pad Set pad temperature only as prescribed by and under the guidance of a physician Monitor the patient s temperature and skin condition every 20 minutes or as directed by a physician Failure to adhere to these warnings could result in patient injury The following Groups Conditions require additional surveillance Group Condition at risk Potential injuny Pediatric patients Hyperthermia hypothermia Patients with impaired circulation Ischemia Areas of application are under pressure Ischemia In combination with topical solutions whose toxicity Chemical may be affected by the application of heat injury In combination with other heat sources Thermal injury Only qualified medical service personnel should repair the T Pump Improper repair may result in death or serious injury equipment damage or malfunction Always perform the FUNCTIONAL CHECK AND SAFETY INSPECTION as specified in the SERVICE MANUAL after making repairs and before returning the T Pump to patient use Improper repair may result in death or serious inj
2. If the water level drops below the operating level add water Do not overfill Overfilling can result in reduced pump motor life Apply T Pad to the patienr Follow T Pad instructions For specified performance keep the pump at or above the level of the pad If che pump is placed below the pad s water will drain into the pump when it is shut off If the pump has been overfillea or if multiple pads are connecred excess water can leak out onto the floor or cause reduced motor life Turn off pump before disconnecting pad Close all hose clamps To prevent water spillage always disconnect pad from pump with connectors raised above the level of the pad and pump Connect the pump and T Pad Clik Tire connectors together where applicable Lama a MM D T OPERATOR S MANUAL TP500 TP500C TJIPUMP Storage Short term Storage Long term Draining Cleaning Pads Accessories STORAGE CLEANING Disconnect pad Connect ends of the connector hoses together where applicable Open hose clamps Leave water in the reservoir To prevent hose kinks coil the hose rather than folding it Fasten the hose and cord with scrap Connect pad Open hose clamps Add 1 4 ounce AirKem A 33 GAYMAR MTA33 germicidal or equivalent to water already in reservoir Run for two 2 minutes Drain pump Close hose clamps Disconnect pad To prevent hose kinks coil the hose rather tha
3. Instructions ON OFF Switch Power Cord with Molded Durable l Hospital Grade Cermoplastic ase Plug Circuit Breaker Twin 10 long Leakproof Pump c l Hose Assembly xtra Capacity with 2 types af Reservoir Connectors 1500 ml 51 oz available see figures 3 amp 4 Operating Water Level Indicator TEMPERATURE SETTING Indicator key operated Nonmarking Suction Cups TEMPERATURE Dual Temperature SETTING set Point Key Scale PCF OVEA TEMP Light Figure 2 TiPump Features Specifications FEATURES Attached Hose l0 ft 305 cm dual hose Connectors allow pads to be connected to the pump see figs 3 and 4 p 5 Tip over Turns heater off if pump is tipped Switch NOTE This does not activate the OVER TEMP light Hospital 1D A label is provided on the back of the T Pump for your Label convenience Use a felt tip or ball point pen to add any required hospital identification OVER TEMP Indicates the pump and heater have been turned off Light This light is activated by the two OVER TEMP safety Circuit Breaker thermostats Refer to the TROUBLESHOOTING section When the circuit breaker trips the small button in the center will extend 4 exposing a white band Refer to the TROUBLESHOOTING section OVER TEMP Either of two limit thermostats will shut off pump and safety heater if the high temperature limit is exceeded The Thermostats OVER TEMP light will glow Refer to the TROUBLESHOOTING sect
4. facility must keep in inventory The T Pads can be interconnected on model TP500 only to provide therapy to more than one body site at a time see fig p 2 For a brochure listing the various pads contact the GAYMAR Customer Service Department see next page for telephone numbers An optional bed bracket model TP20A is available to mount the T Pump on the footboard of a bed OPERATOR S MANUAL TROUBLESHOOTING MAINTENANCE TPSO0 TP500C TIPUMP oF M mm PROBLEM POSSIBLE CAUSE REMEDY OVER TEMP light is on 1 Pad or hose is kinked Hose clamps are closed Turn switch off unplug pump and dump water out Unkink pad or hose Open hose clamps Refill with room 2 Water level is low or reservoir is empty temperature distilled water to proper level Plug T Pump in 3 Pump is lilled with water thal is too hot turn switch on and wait 40 minules it must be below 75 F T Pumo will not heal 1 TiPump is tipped Place T Pump on level surface 2 OVER TEMP lighl is on Reler to OVER TEMP section above T Pump will not pump 1 Water level is low or reservoir i empty Refill with room temperature distilled water to proper levet 2 OVEH TEMP lighl is on Refer to OVER TEMP section above T Pump will not turn on 1 The ON switch may not be fully engaged The light on the switch should be lit Verity that the switch is fully eng
5. GAYMARPR TIPUMP TP500 TP500C Heat Therapy System PLI LISTED 304L STANDARD NORME C222 NO 125 RISK CLASS CATEGORIE DE RISQUES NO 2G OPERATOR S MANUAL 12 98 P N 10569 090 OPERATOR S MANUAL IP500 TP500C TiPUMP BEFORE YOU BEGIN CONTENTS Read and understand this TIPUMP OPERATOR S Safety Precautions MANUAL and all PRECAUTIONS see page I Introduction eere prior to using the f Pump Features Specifications RECEIVING INSPECTION Operating Instructions Check the shipping carton for damage immediately upon receipt If package damage is discovered the device should be unpacked with the carrier s agent present Any claims for Troubleshooting shortage or damage must be filed with the delivering carrier by the purchaser Do not return pumps damaged in shipment to GAYMAR without contacting our Technical Service Department for advice see phone numbers below If damaged goods are returned to GAYMAR without notifying the carrier GAYMAR will assume the repairs will be made at the customer s expense Storage Cleaning BER Should you desire repairs can be made in your facility by your own qualified personnel Contact the GAYMAR Customer Service Department to request a T PUMP SERVICE MANUAL Call GAYMAR for spare parts required T PUMP TPAD PMuleT Pad Clik Tite and Gaymar are registered trademarks of
6. aged in the ON position 2 The electrical cord is not plugged into a grounded Insert the plug fully into the receptacle elecirical receptacle 3 The circuit breaker has tripped If the circuit breaker has tripped turn the power switch to OFF Wait two 2 minutes Press the circuit breaker button in Turn T Pump on If the circuit breaker trips again remove T Pump from use Water leaks from connector Chk Tite connector 1 Damaged QO ring If delactive replace Clik ite connector 2 Locking ring on Clik Tite connector is not snapped Snap Clik Tite locking ring shut into place see fig 3 Calder style connector 3 The couplings are not securely connected Push the male and female ends together until you hear a click If it still leaks replace conneclor PREVENTIVE A DANGER MAINTENANCE SERVICE Risk of electric shock Disconnect power before servicing A WARNING Only qualified medical service personnel should repair the T Pump Improper repair may result in death or serious injury equipment damage or malfunction
7. edure NOTE Refer to figure p 2 and item 9 p 6 when connecting multiple pads COLDER STYLE CONNECTORS on TP500C fo attach Colder style connectors to a pad push the male coupling onto the female coupling When you hear an audible click the connectors are joined fig 4A 2 To open or close the hose pinch clamps Open the clamp by pushing the serrated end fig 4B Close the clamp by pressing the clamp together fig 4C 3 To disconnect Colder style connectors press down on the thumb tab of the female coupling The couplings will partially disconnect Pull the male coupling out fully to disconnect fig 4D ua OPERATING INSTRUCTIONS OPERATOR S MANUAL TP500 TP500C TIPUMP START UP PROCEDURE SHUTDOWN PROCEDURE Before filling the pump always attach a pad to the connector hose or close the clamps on the connector hose ends see figs 3 and 4 p 5 Then unkink pad and hose Open hose clamps Open the fill cap on top of the pump Fill the pump with room temperature i e not hot distilled water to the operating level indicated on the side of the pump NOTE Using tap water will decrease the life of the pump Insert temperature setting key Ser temperature as prescribed by the physician Remove key to prevent tampering Plug l Pump into a properly grounded ourler Turn on the power switch The sefected water temperature will be reached in approximately 20 minutes
8. ion Lat OPERATOR S MANUAL SPECIFICATIONS TP500 TP500C TIPUMP SPECIFICATIONS Size approx 8 a x 5 5 g x 6 lf4 20 6 cm x 14 3 cm x 15 9 cm Weight empty 9 lb 2 oz 2 3 kg Reservoir capacity 5 oz 1500 mi maximum Flow rate 9 gph 34 Iph minimum with pad attached Ambient operating 60 F ro 90 F 15 6 C to 32 27 C temperatures Storage temperatures 30 F to 160 F 34 4 C to 71 1 C empty Temperature set point 85 F to 107 F 29 4 C to 41 7 C range Average operating 42 F at 07 F setting temperature accuracy Power cord 18 AVG 3 conductor 9 ft 274 cm Type SIT with molded hospital grade plug Circuit breaker 3 amperes Current leakage GU microamperes maximum nin Ground resistance 0 5 ohm maximum Electrical requirements Volrage VAC 120 Frequency Hz 60 Current amps 1 8 Power req d watts 200 Safety approvals Tested to UL 544 and CSA 22 2 No 125 OPERATOR S MANUAL TPSOOITPS00C TIPUMP OPERATING INSTRUCTIONS T Pumps are supplied with one of rwo types of hose connectors TP500 T Pumps have Clik Tite connectors TP500C T Pumps have Colder style connectors CLIK TITE CONNECTORS on TP500 To attach Clik Tite connectors from hose to pad l Insert male fittings into female fittings with a twisting motion figs 3A and 3B 2 When fittings are fully inserted snap locking ring into place figs 3C and 3D 3 To disconnect simply reverse the proc
9. n folding it Fasten tne hose and cord with strap and store pump Disconnect the T Pump from AC power Disconnect the pad or hoses from one another keeping hoses at or above the level of the T Pump Remove the fill cap and invert the T Pump over a sink VVhen all fluid has drained from the hoses and reservoir replace the fill cap and connect the hoses together where applicable Clean outer surfaces of the T Pump with a damp cloth and soapy water Fantastik spray cleaner or a mild abrasive such as Soft Scrub cleanser without bleach To clean the fluid system prepare a germicidal solution according to the manufacturer s instructions Use AirKem A 33 or equivalent available from Ecolabs Inc 370 Wabasha St Paul MN 35102 phone 800 247 5362 or from GAYMAR product catalog MTA33 Drain the pump Fill the reservoir to the operating level indicated on the side of the pump Set the temperature indicator to its lowest setting fully counterclockwise start the T Pump and circulate the solution for one hour Drain the solution and refill the pump with distilled water Using distilled water will not promote algae growth or mineral buildup Change the distilled water monthly or more often depending upon use For best results use only GAYMAR T Pads or Mul T Pads The unique button design allows water to flow and provides trouble free operation when the pad is folded This reduces the number of different sizes of pads your
10. ury equipment damage or malfunction Qum OPERATOR S MANUAL INTRODUCTION TP500 TP500C T PUMP Single pad Multiple pads TP500 only Figure 1 T Pump Heat Therapy System INTRODUCTION Heat therapy is effective in che dilation of blood vessels thereby increasing the blood flow to the heated area Heat therapy has a variety of uses the most common being treatment of aches and pains in joints and muscles The GAYMAR T Pump Heat Therapy System provides a means of applying heat therapy by supplying temperature controlled water through connector hose to a Gaymar T Pad The hose is terminated in easy to use Clik Tite or Colder style connectors The T Pad provides the interface for delivering the heat therapy The unique button design allows water to flow and provides trouble free operation when the pad is folded This reduces the number of pads your facility must keep in inventory The pads are applied to the part of the body requiring heat therapy and the circulating water maintains the pad at the set point temperature The temperature set point is key operated to prevent tampering The T Pads can be interconnected on model I P500 oniy to provide therapy to more than one body site at a time OPERATOR S MANUAL TP500 TP500C TIPUMP FEATURES Cord Hose Strap Serial Number Vented Leakproof Hospital Cap with Canny Hendte Operating identification Label Plastic Retainer IM uminateci pullout proof