Testing anaesthetic machines, ventilators and related equipment

image

Contents

1. Flow delivery Set the ventilator as follows minute volume 10 rate 10 inspiratory expiratory I ratio 1 1 Start the ventilator the tidal volume measured with a spirometer or ventilator tester should be between 0 9 and 1 1 litres min If it is not first check the rate with a stopwatch and adjust if necessary Check the E ratio with a stopwatch at a very low rate and adjust if necessary After confirming that both are correct reset the ventilator to the above settings and adjust the minute volume to give a 1 litre tidal volume Absorbers Canister Check for cracks and chips and check gaskets Replace as necessary Inspiratory and expiratory valves Inspect the inspiratory and expiratory valves for cleanliness and for bent or chipped discs Bag ventilator switch if fitted Inspect the valve clean and lubricate with silicone grease as necessary to maintain free action Relief valve Inspect clean and lubricate screw threads with silicone grease Check that when the valve is fully open a maximum pressure of 0 3 kPa 2 5 mm Hg is maintained in the system with the oxygen flowmeter opened to flood measured with the device for measuring flow pressure and tidal volume at the bag connector 97 MAINTENANCE AND REPAIR OF LABORATORY AND HOSPITAL EQUIPMENT RE Drain valve Inspect and clean as necessary Elevating assembly Check that the push button operates smoothly raise and lower the assembly severa
2. Turn on the oxygen flowmeter very slowly Pressurize the back bar to 30 mmHg 4 kPa When this pressure is reached turn down the flow until the pressure on the gauge remains constant at 30 mmHg 4 kPa If the flow is less than 100 ml min it is acceptable if it is greater than this look for a leak 2 Check all the joints on the back bar with a high pressure test Pressurize the system to 150 mmHg 20 kPa Reduce the flow to maintain that pressure If the flow is 100 ml min or less it is acceptable if it is greater than this look for a leak Brush each suspect point with soapy water bubbles will appear at the site of the leak Check the correct operation of the oxygen failure warning whistle if fitted Pressurize the intermediate system turn off the supply and open the oxygen flowmeter valve to reduce the pressure slowly The whistle should sound for a minimum of 10 seconds when the pressure falls to between 180 and 250 kPa Check that the flow of nitrous oxide is cut off when the oxygen is turned off if that system is fitted Check the oxygen flush valve It should allow a flow greater than 35 litres min but not more than 75 litres min or as required by local regulations 93 MAINTENANCE AND REPAIR OF LABORATORY AND HOSPITAL EQUIPMENT lt auSt If there are hoses for connection to a wall supply check these Check the oxygen flow from the oxygen flow tube when the oxygen probe is plugge
3. MAINTENANCE AND REPAIR OF LABORATORY AND HOSPITAL EQUIPMENT es ao 10 II 12 T3 14 I5 16 17 18 19 20 Remove the drum actuating lever and withdraw the drum If the drum is seized or cannot readily be withdrawn remove the locking ring from the gas outlet and turn the adjusting ring in a clockwise direction to force the drum out Examine the drum for signs of scoring or corrosion remove the old grease and apply a film of a suitable grease Remove the drum grease injector cap and clean all the old grease from the body assembly Unscrew the plunger control gland packing nut probe out the packing cotton and regrease it Replace the cotton and the gland packing nut Insert the drum and refit the actuating lever With the outlet adjusting ring removed screw in the inlet adjusting ring while moving the actuating lever to and fro until an increase in resistance to lever movement is felt Do not overtighten the ring to the point of drum seizure as this will force out all the grease Without moving the adjusting ring fit and tighten the inlet locking ring Screw in the outlet adjusting ring until it just touches the drum then back off one eighth of a turn fit and tighten the locking ring Check the drum for freedom of operation Fill the grease injector cap with grease and screw in to the grease point Examine the glass bottle for dirt and damage replace as necessary Examine the U tu
4. be and cowl for dirt and damage and for security of attachment Examine the bottle sealing washer for serviceability replace as necessary Examine the cork stopper for serviceability and security of attachment to the retaining chain Refit the vaporizer into the back bar unit assembly Using a continuity test set or multimeter check that there is electrical continuity between the drum actuating lever the cowl operating plunger and the vaporizer body Test the back bar for leaks See p 94 for the back bar tests Testing anaesthetic machines ventilators and related equipment Anaesthetic machines and ventilators should be tested in the room or area where the equipment is situated if at all possible Service personnel must wear accepted operating room clothing in the surgical areas Contact the person responsible for operating room equipment regarding the movement or servicing of any such equipment Anaesthetic machines Tools and materials required Normal service tools Silicone grease Light oil Service manuals Watch with a second hand Spare parts Do not use ordinary oil such as motor oil to lubricate any parts that come into contact with oxygen 92 ANAESTHETIC AND RESUSCITATION EQUIPMENT aE eS eee ee Mercury sphygmomanometer Anaesthetic gas analyser if available Device for measuring flow pressure and tidal volume Testing procedure If there is an official service manual follow the steps outlined
5. d in and the nitrous oxide disconnected Similarly when the nitrous oxide probe is plugged in and the oxygen disconnected nitrous oxide must flow from the nitrous oxide flow tube and nothing from the oxygen flow tube Check the non return valve on each of the yokes Pressurize the system turn off the cylinder and remove it There should be no gas leak If there is a leak look for the simple non return valve in the yoke Dismantle and clean the valve then reassemble If the anaesthetist has been having difficulty in getting patients to sleep or waking them up there may be a problem with the vaporizer Inspect the mountings and connectors to ensure that they are secure and leak free When servicing vaporizers be sure to keep them in an upright position when they contain liquid anaesthetic To check the vaporizer an anaesthetic gas analyser is required Connect the gas analyser to the common gas outlet Set the oxygen flow to 3 litres min and after zeroing the gas analyser with 100 oxygen test each vaporizer at each full percentage setting Determine that there is no concentration of gas when the vaporizers are in the off position Replace any vaporizer for which the concentration is incorrect by more than 0 3 of the reading or 10 of the measured value whichever is greater If an anaesthetic gas analyser is not available you can only check a that the vaporizer is off when it is turned off b that it gives an output when i
6. in it otherwise follow the procedure below see Fig 4 5 page 75 Check for leaks in the high pressure system 1 Turn off all flowmeters and disconnect any ancillary equipment such as ventilators and suckers 2 Turn on each cylinder in turn and allow the system to pressurize then turn the cylinder off Watch the pressure gauge if the needle drops there is a leak 3 Remove the covers and brush each joint or suspect point with soapy water Do not forget to check inside the back of the pressure gauge A leak will be indicated by the formation of bubbles Do this for each cylinder in turn Check the operation of each flowmeter Make sure the control knob stays where it is set and is not liable to be turned by mistake I Close all valves on the machine Open all cylinder valves one full turn noting any movement of the flowmeter floats Float movement indicates a leaky flowmeter valve If so adjust the stop so that gas flow ceases 1 8 turn before the knob reaches the stop 2 Verify flowmeter accuracy 2 5 full scale with the measuring device connected to the common gas outlet 3 Check that the needle valve stems are tight enough to remain where set unless deliberately turned by the operator Check the low pressure system which is the part from the control knobs to the outlet 1 Check the top and bottom seals on the flow tubes with a low pressure test Connect a mercury sphygmomanometer to the outlet
7. l times to check operation If it does not operate smoothly disassemble and clean with alcohol Do not lubricate Compound pressure gauge Test the patient pressure gauge for zero setting and accuracy 5 of reading with the pressure measuring device attached to the bag connector Calibrate if necessary with the adjusting screw located under the plug screw inside the absorber head beneath the screen Pressure regulators reducing valves and flowmeters The pressure regulator or reducing valve as the name suggests is used to reduce the pressure of a gas from the cylinder pressure to a pressure that is safe for subsequent delivery to a patient For example in the case of oxygen it is from 14 300 to 420 kPa A flowmeter can be attached to the regulator to allow a given flow to be set These assemblies are most common on anaesthetic machines or on an oxygen therapy unit attached to the top of a gas cylinder When used in oxygen therapy there are three parts to the unit a gauge showing the pressure of the contents of the cylinder a regulator to reduce the pressure a flowmeter that indicates the selected flow There are a number of different designs of regulator but generally each unit has one inlet from the cylinder and three outlets one to the pressure gauge one to the flowmeter and one to the blow off valve It is important to know which one is meant to be connected to the pressure gauge do not connect a
8. ny other part to this outlet While the proper checking of pressure regulators requires some special test equipment most problems can be overcome with very little equipment Remember that the flow tube is under regulator pressure Do not unscrew it before the cylinder is turned off and the pressure released Setting the output pressure Make sure that the cylinder contents gauge and the safety valve are connected to the regulator remove the flowmeter and fit in its place a 0 700 kPa pressure gauge Connect the regulator to the gas cylinder and turn on the gas The test pressure gauge should show a reading of 420 kPa This is the correct pressure for an oxygen therapy regulator and flowmeter If it is not 420 kPa adjust with a socket head key until it is correct If you are adjusting the pressure downwards you must release the pressure from the test gauge turn the adjusting screw out 1Do not use any oil when repairing oxygen regulators If you use PTFE or plumbers tape you should use only special de greased tape 98 ANAESTHETIC AND RESUSCITATION EQUIPMENT C reconnect the test gauge and adjust the pressure up to 420 kPa On most makes the adjustment screw will be found at the end of the piece which sticks out at the front it may be covered with a sticky label At this pressure it should be possible to obtain a flow of up to 55 litres min out of the unit this is called the flush flow In some places the regulator is not
9. ormally go off at about 640 kPa e If the flowmeter makes a popping noise when the flow is turned on motor boating there is probably dirt inside the valve the noise may also be caused by a faulty valve seat Low flows unscrew the needle valve and check that it is clean and undamaged e Ifthe ball or bobbin shows a small flow even when the unit is turned off check the flow tube for leaks e Ifthe gauge needle does not drop smoothly remove the back of the gauge and lubricate the movement of the gauge with a light watch oil Use of oil is acceptable in this case as there is no oxygen flowing in this part If there is no flow even when the gauge shows that there is gas in the cylinder this suggests that the gauge needle is stuck Check the movement in the back of the gauge or reposition the needle on the shaft Unregulated flowmeters in which the flowmeter is connected directly to the gas cylinder are dangerous and should not be used 99 a MAINTENANCE AND REPAIR OF LABORATORY AND HOSPITAL EQUIPMENT a Oxygen analysers Inspection and calibration Visually inspect the cables and the display unit for signs of wear or deterioration Check that the sensor membrane is not nicked or otherwise damaged and that the O ring seal is intact Test the calibration by first placing the sensor in a T adaptor in a verified 100 oxygen line and allow the readings to stabilize If after 3 minutes the display has not stabilized re
10. ove the covers Inspect any internal tubing or bellows lightly rub any antistatic tubing or bellows with silicone grease to prevent perishing Replace as required Blow clean and wipe the insides If there are electronic circuit boards check that they are secure and show no damage Check for wear in any moving mechanical parts Using a light motor oil or similar lightly lubricate any moving pivot points Clean up any drips 4 Start the machine running again taking care not to touch any internal parts watch any internal movements bellows lever or valve movements for smooth operation 5 Try each control in turn and check that it does what is intended for example if the breath rate control says that the machine will do 60 breaths per minute this must be confirmed With all these checks a degree of common sense is 95 MAINTENANCE AND REPAIR OF LABORATORY AND HOSPITAL EQUIPMENT SEmmmMEE necessary For example do not worry if the breath rate control says 60 and only 58 are delivered Every machine has a margin of error If the manual is not available a degree of discretion should be used 6 Check that the pressure gauge is accurate by comparing it with a test gauge 7 Check the correct operation of all lights and indicators 8 To check the correct operation of the oxygen mixer an oxygen analyser is needed If your department does not have one ask the Anaesthetic Depart ment to provide such an instrument Note down on the
11. place the sensor If it has stabilized adjust the cat knob if necessary so that the display registers 100 Remove the sensor from the T adaptor and expose to room air A reading of 21 2 should be displayed within 1 minute If a correct reading is not displayed replace the sensor 100
12. re low Connect the device for measuring pressure to the common gas outlet open the oxygen flowmeter to 300 ml min and allow the bellows to rise to the top The pressure should be less than 2 5 cm water 1 84 mm Hg or equivalent on the gauge in use and the bellows should remain full If the bellows do not remain up or the pressure exceeds 2 5 cm water then refer to the ventilator service manual for the necessary repairs Bellows pressure high Connect the common gas outlet to the driving gas port on the bellows and plug the bellows outlet Pressurize the outside of the bellows to just above 60 cm water 96 ANAESTHETIC AND RESUSCITATION EQUIPMENT a eT 44 mm Hg and maintain an oxygen flow of 300 ml min The pressure gauge should settle at or above 60 cm water If the pressure drops below 60 cm water refer to the ventilator service manual for the necessary repairs Ventilator controller Low oxygen supply alarm Check that the low oxygen alarm activates before the supply pressure to the ventilator drops below 250 kPa and resets when the pressure reaches 320 kPa Low airway pressure alarm Check that the alarm activates if the pressure measured at the patient port remains below 7 cm water 5 mmHg or equivalent on the gauge in use for between 20 and 30 seconds Safety valve Check that the relief valve opens when the pressure in the patient circuit exceeds 65 75 cm water 48 55 mmHg or equivalent on the gauge in use
13. repair it on the spot take it out of service If there is no spare machine the patient must be ventilated with a resuscitation bag while the machine is being repaired If it is not possible to repair a machine properly owing to a lack of spare parts do not be tempted to carry out temporary repairs Report the problem to the user ask for the spares and remove the machine from use Do not agree to put a machine back into service against your better judgement If the personnel on the ward insist get them to sign the service sheet with the problem clearly stated Follow the steps outlined in the official service book To carry out a service without the book follow the steps outlined below I Inspect the outside of the machine including all tubing connectors and any bellows for damage Replace as required Lightly rub any antistatic tubing or bellows with silicone grease to prevent perishing 2 Connect up to the electricity and gas supplies as required Put a stopper or test lung on the patient connector and start the machine running Set the controls to normal settings Watch that the operation is regular and smooth Listen and check for any unusual noises It is important to use the same regular settings in each test in this way you will get to know the normal movements and sounds of correct operation Any unusual movements or sounds will alert you to possible problems 3 Switch off and disconnect the machine from the mains Rem
14. s if available Watch with a second hand Spare parts if available Device for measuring flow pressure and tidal volume 94 ANAESTHETIC AND RESUSCITATION EQUIPMENT ME Testing procedure It is of the utmost importance that all ventilators should work safely since lives depend upon their correct operation In addition ventilators should never be used without a correctly adjusted alarm system which gives a warning and therefore protects the patient in case of malfunction or disconnection If you do not have the service manual for your machine make every effort to get one Make sure a service manual is ordered with every new machine Even without the manual it is still possible to ensure that the machine is working correctly but the proper spare parts must be available Records are very important when maintaining ventilators and other life support equipment All reported faults repairs and service details should be noted down dated and signed Machines should be serviced twice a year If you are called to look at a machine that has been reported as faulty check first that it has been set up properly Most reported faults are caused by operator error When looking for a fault start from the beginning For example is the electricity turned on Is the gas on Investigate the device in a planned manner looking for the obvious things first If you have doubts about the machine s safety or correct operation and you are unable to
15. service sheet the output results from 21 to 100 9 Check the alarm system 10 Run the machine again on the normal settings and check that it is still working correctly 11 If it is a machine that uses electricity give it a safety check 12 Fill out the service check sheet and sign it As you gain more experience in servicing you will get to know which errors are minor and can be allowed and which are unacceptable For example an oxygen mixer on a paediatric ventilator must not give higher levels of inspired oxygen than indicated This is because very serious damage can be done to the infant s eyes as a result Return the machine to the user the doctor in charge should test it to ensure that it operates satisfactorily Ventilator bellows Mechanical integrity Inspect the bellows housing and base for cracks chips etc Check the tubing and control knobs for tightness Over pressure valve Check the valve located on the scavenging tee at the back of the control unit for cleanliness and operation Pop off valve Remove the housing bellows and pop off valve from the base Check that the pop off valve glass disc and seat are clean and dry and that the retaining screw is tight Bellows flexibility Reassemble the ventilator and connect the test gauge to the common gas outlet Inflate the bellows to 0 1 litre If the pressure is above 1 75 cm water 1 3 mm Hg then the bellows should be replaced Bellows pressu
16. set for pressure but adjusted for a given flow with the control wide open check what is required before adjusting the setting For a regulator attached to an anaesthetic machine adjust the pressure to about 390 kPa 30 kPa lower than the pressure from the pipeline supply to the machine Testing a regulator and flowmeter unit 1 Check for leaks 2 Check that the pointer on the gauge works smoothly and reaches the stop before the flow falls to zero 3 Check that the flow goes to its full rate Faults Ifa leak is suspected check as follows With the flowmeter unit turned off turn on the gas and allow the pressure to rise Turn off the gas supply The system is now pressurized to the full cylinder pressure but with a very small volume If there is any leak the gauge will show a fall in pressure the bigger the leak the faster the fall If the pressure falls brush the unit with soapy water any leaks will show up as small bubbles Do not forget to check inside the back of the gauge e Leaks around the bull nose connector are usually caused by a faulty O ring Replace the ring Leaks at the blow off valve First check that the regulator is set to the correct output pressure If the pressure rises to more than it should yet the regulator is set to the correct output pressure there is a faulty valve seat a problem called creep Replace the valve seat The blow off valve should n
17. t is turned on again and c that the concentration of gas increases as the control is turned up Check that the control knob turns smoothly Vaporizers should be returned to the manufacturers or their agents for checking every few years The interval depends upon the model some models need a check by the company only every I0 years Check the smooth operation of the pressure gauges The pointer should move smoothly and come to rest before the flow in the flow tube stops If the movement is not smooth lubricate the linkage in the back with silicone grease Check the absorber Check for smooth operation of the controls and for leaks Change the filling if required Check any other back bar mounted equipment Check the output pressure of the regulators This should be around 390 kPa or as required by local regulations and in any case should be about 35 kPa lower than the output pressure from the wall outlet Check all flexible tubing on the machine Check all attached equipment such as suckers and blood pressure machines Check the drawers wheels and the general frame of the machine Lubricate lightly as required Clean the machine Tick off all tests on the service sheet and sign it Return the machine to the user The doctor in charge should test it to make sure that it operates satisfactorily Ventilators Tools and materials required Normal service tools Silicone grease Light oil Service manual