1. B A S Perform Leak Check of Machine Low Pressure System Verify that the machine flow control valves are OFF Attach Suction Bulb to common gas outlet Squeeze bulb repeatedly until fully collapsed Verify bulb stays fully collapsed for at least 5 seconds Will not turn off Bulb does not stay collapsed for 5 sec Monitors Check Calibrate and or Set Alarm Limits of All Monitors Ventilator and ventilation monitor parameters including Oxygen analyzer Page 18 Physiologic monitor that may include vaporizer agent capnometer and pulse oximeter Page 15 Alarm Limits do not work Air Compressor Turn on Adjust air compressor to 50 PSI Ensure Equipment Care is conducted as directed by the manufacturer s literature Does not turn on Will not reach 50 PSI 13 Maintainer s Preventive Maintenance Checks and Services 6515 01 533 0968 Magellan B Before Operation D During Operation A After Operation Q Quarterly and S Semiannually i Interval Item to Be Inspected and Procedure Is Not Mission Capable If 1 B S Magellan 2200 a Conduct an inventory to ensure that the items Missing parts or accessories listed on the Equipment Parts and Accessories preclude operation of the List are on hand refrigerator b Check the electrical power cord for cuts The power cord is cracked or fraying or deterioration frayed wires are not covered by the cord insulati
2. reusable container NSN 6515 01 533 1279 It will be transported by vehicles organic to using organizations or by other transportation assets within the theater There are no strategic lift constraints associated with this system When transported per MIL STD 1472 the system may be carried by two persons All support items are non deteriorative MAGELLAN 2200 CHARACTERISTICS Storage temperature Operating temperature 30 F to 160 F Relative humidity 40 to 90 relative humidity non condensing Dimensions L x W x H in 23 x 17 x 23 Weight Ib J Design Interface The Magellan will interface with the electrical distribution system at the using organization K Requisition Supply 1 The Magellan is a non stocked centrally procured item The system will be provided during an initial fielding or sustainment fielding Or if requested from the unit procurement will be initiated only after receipt of a requisition AAC Code J which denotes NOT STOCKED CENTRALLY PROCURED non stocked items IMM Service centrally managed but not stocked item Procurement will be initiated only after receipt of a requisition ARC Code N which denotes Class VIII items valued at 500 or more that retain their original identity and are not consumed in use and non maintenance significant medical furniture items valued at 1 000 or more TIC Code C which denotes critical medical equipment requiring technical Inspection
3. 1 312 0894 Victoreen Gas Flow Analyzer 6515 01 491 6615 VT Plus Calibrator Gas Flow C61523 6695 01 255 2855 RT 200 07 Anesthetic Gas Analyzer A00098 6630 01 487 6987 Riken 1802D Z00334 CO2 Analyzer wu P N 19 3325 en 10 Analyzer NIBP Z07763 6515 01 449 1423 Cufflink 6515 01 449 2331 IPT 1 11 J24245 6515 01 479 2355 IDA4P 3 Depot Only PUB Analyzer 6515 01 536 2122 IDA 4P 1 Defibrillator Analyzer TPA A83433 6515 01 449 1420 Impulse 4000 Tester Defibrillator T02889 6625 00 433 9063 DT2000A 3 Counter Electronic Digital C19266 6515 01 271 3012 AN USM 459A 4 Densitometer SU150 P D94348 6525 01 161 1945 97 423 Request deletion 2 Simulator Medical Function S56720 17A Simulator Medical Function S56720 ah 215M Calibrator Generator ECG 061455 6515 01 049 g44g ECG 100 Simulator Medical Function 17 S56720 6515 01 548 3352 MPS450 Army April 07 Medical Kit 6 Signal Generator S48323 6625 01 276 9421 SG 1288 G Computer Laptop 7010 01 502 5490 Foot Candle Meter M38443 6695 01 303 0294 9 118 6625 01 296 4006 TK 80 Probe Only o0 Multimeter AN PSM 45A M60449 6625 01 265 6000 27 FM Multimeter AN USM 486 M23954 6625 01 145 2430 8050A scili schpe Digital P43667 THS720P old LIN 6525 01 448 9577 Oscilloscope P30693 Z47763 6625 01 187 7847 P33687 2235L Oscilloscope P43667 6625 01 258 0022 2430A Medical Oscilloscope 29 Feb 08 6625 01 560 1673 199B Radiom
4. Acquisition Support Strategy for the Anesthesia Machine Magellan 2200 NSN 6515 01 533 0968 Version 1 0 Manufacturer Oceanic Medical Products Model Model 1 Power 1500 Watts Voltage 100 250 VAC 10 50 60 Hz Package Weight 100 lbs 45 36 kg Package 5 5 cubic feet LIN A62773 Prepared by Medical Scientific Division PMO Medical Devices USAMMA 28 May 2007 Acquisition Strategy The Magellan is a commercial off the shelf COTS anesthesia delivery system It is compact robust and flexible easy to move and was designed for military field hospitals general civilian hospitals and office based anesthesia and surgery centers of every description It is pneumatically powered and contains an electronic ventilation monitor that is A C or battery powered The Magellan 2200 may be mounted on a tabletop or on a mobile cart The ventilator is electrical with a NATO battery back up built in battery charger and is supplied with a transformer that can be adjusted to virtually any wall voltage The unit is designed as a simple to operate unit containing all of the modes and necessary components to deliver a basic anesthetic including a mechanical ventilator and a single agent vaporizer All of the safety systems required are employed onboard and easy to utilize The Magellan is authorized in the Veterinary Equipment Set VES M30340 Anesthesia Ultrasound Augmentation and the VES 50 Patient Small M30136 This system is authoriz
5. Resupply items for this product are noted in the Supply Support and Support Equipment section Appendix A Market Analysis N A 10 Appendix B Standardization Efforts N A 11 Appendix C Testing and Airworthiness N A 12 Appendix D 10 20 Standards and MAC Chart 6515 01 533 0968 Magellan Operator s Preventive Maintenance Checks and Services B Before Operation D During Operation A After Operation Q Quarterly and S Semiannually Item No Interval Item to be Inspected and Procedure Is Not Mission Capable If 1 B S Magellan a Conduct an inventory to ensure that the items listed on the Equipment Parts and Accessories List are on hand b Verify backup ventilation equipment is available and functioning It is recommended that a self inflating bag valve mask device with the ability to provide high concentrations of O2 be immediately available c Perform Start up procedures in accordance with operating instructions Missing parts or accessories preclude operation of the refrigerator Analyzer fails to start up B S Check Oxygen Cylinder Supply Open O2 cylinder and verify at least half full approx 1000 psi No Oxygen or less than 1000 psi B A S Close cylinder Check Initial Status of Low Pressure System Will not hold pressure Close flow control valves and turn vaporizer off Check fill level and tighten vaporizer filler cap Not full
6. ed in Veterinary Medicine units The acquisition strategy is to purchase one each per M30340 and two each per M30136 The Magellan is provisioned with all the necessary accessories and supplies The system is purchased through ECAT vendor Oceanic Medical Products part number MAG2200M1 It has an item cost of 22 805 This item is OPA funded The life expectancy for the Magellan is 10 years TB MED 7 Operating costs are 1 372 per year and maintenance costs are approximately 475 per year therefore the life cycle costs are calculated to be approximately 18 470 not including initial purchase costs Clinical Application The Magellan is used by healthcare personnel to anesthetize patients The mechanical ventilator has a single agent vaporizer and can be used with a variety of facemasks endotracheal tracheostomy tubes and other approved artificial airways that are standard in any operating theater The Magellan Anesthesia Machine is fully FDA approved The Magellan 2200 will be used to anesthetize animal casualties during surgical procedures Integrated Logistics Support ILS A Maintenance Planning 1 Purpose To provide a description of the requirements and tasks to be accomplished for achieving restoring or maintaining the operational capability of a system or equipment It covers the functions of preventive maintenance checks and services PMCS and associated schedule periodic Consumable Supplies only performance veri
7. eter Ultrasound Therapy 6625 01 504 2654 1 E 17 Wattmeter 5 Nov 2007 R95994 6625 01 504 8680 UW 5 6625 01 141 7357 UMR 3 C n 6625 01 487 6986 UMR 3 D EE S57953 6515 01 518 4101 INDEX 2XLFE Simulator Pulse Oximetry Z14528 6515 01 504 8537 INDEX 2Mre 6515 01 449 1422 Cardiosat EF Tachometer Tachometer Stroboscopic To PET 1756 6680 01 307 8190 Tachometer Digital Medical 1893A Do not use 6625 01 550 3339 Equipment 17 April 07 Test Set Electronic T77263 6625 01 255 0839 HTR2000 Huntron ii T90883 6515 01 438 2409 454A RA 6625 01 042 8213 RF302 T61791 6625 01 142 8233 232M Do not use Tester Current Leakage 6625 01 207 8270 232D Tester Ventilator 05633 6515 01 449 1421 Pneuview 360001 228075 24 25 C74198 07 467 Request Test Cassette X Ray 6525 01 039 4019 28 i 29 31 18 Appendix E Reference N A 19
8. fications and calibrations and addresses the test measurement and diagnostic equipment TMDE used to care for the end item Lastly it addresses the maintenance levels of operations and those responsibilities of maintenance personnel therein 2 Responsibilities Biomedical and support maintenance can be performed by personnel holding the military occupational specialty 68 A original equipment manufacturer OEM or OEM trained civilian technicians Reference material to be used by the maintenance personnel includes but is not limited to manufacturer s operator s and or service literature 3 Maintenance Procedures a Field Support Maintenance This item has been assigned a maintenance recoverability code MRC O The system can be repaired and disposed of at the unit level There are no special tools or procedures indicated that would hinder the unit maintainer from completely repairing the unit See inspection procedures below Section 4b b Sustainment Support Maintenance The sustainment level repair for this item will be at MEDLOG Company which includes replacing subassemblies rebuilding and providing field support maintenance for medical units that do not have organic maintenance support Use of medical equipment standby equipment program MEDSTEP materiel will be utilized when non mission capable NMC days exceed 30 days or the unit s assets fall below 50 percent availability c Depot level Maintenance Depo
9. ge Pulley bar code reader alignment tool VOM VOM 15 Anesthesia Machine 6515 01 533 0968 Model 1 FM Crew or Unit Maintenance SM MEDLOG or Depot Maintenance 3 4 5 Tools 6 pae Maintenance and TMDE Remarks Code 08 Motors and Sensors Test FM 20 RS232 amp serial cable VOM sensor tray And tower Alignment tool and feeler gauge Repair FM 01 02 Replace FM 01 02 09 Internal Computer Inspect SM 20 Laptop Repair FM 01 02 Replace FM 01 02 10 Power Supply Inspect FM 01 02 21 VOM Service FM 01 02 Replace FM 01 02 11 Keypad and Display Screen Inspect FM Service 01 02 04 12 Internal Printer Repair FM 01 02 04 Test 20 Replace 01 02 03 13 Rebuild SM 16 Tools and TMDE Code Listing for Maintenance Allocation Charts Ref Code Item Nomenclature LIN NSN Model Tool Kit Medical Equipment A Maintenance Repairer W45334 5180 00 611 7923 Individual GSA Item Tool Set Medical Equipment TRI ARA Unit Level alt Org Maintenance Unit Level W45197 3199 01 48371491 Maint Tool Set Medical Equipment o Maintenance Direct Support Level a Pee lee oreo SSR Tool Set Medical Equipment and General Support alt 04 Maintenance General Support Shop Set Bn Med Level T24386 6545 01 482 2907 Maint Meter X Ray Calibration 6525 01 502 0504 UNFORS 710 L Multimeter Radiographic COS856 6595 01 387 0212 01 02 3 X ray Calibration amp Verification System 6625 0
10. ired above the standard training received by the 68A at the biomedical maintenance course at the DOD Biomedical Equipment Training School X Operator or Repairer can use this part to fix the system 2 Operator Operator training for this system is part of the clinician s standard training program C Manpower and Personnel The fielding of the Magellan will have no impact upon manpower or personnel structure Operator and maintainer actions will be performed by the same soldiers currently performing similar functions The total number of authorized operators or maintainers will neither increase nor decrease in the manpower force structure area The operator MOS 68T Animal Care Specialist will perform operator checks and services Field level maintenance will be performed by Medical Equipment Repairers MOS 68A assigned to the divisional medical companies and combat support hospitals Sustainment level maintenance will be performed by the MOS 68A or civilian assigned to the Medical Logistics Companies D Supply Support Specific COTS support equipment and consumable items are required for the operation of the Magellan These are available through standard supply sources which are predominantly commercial manufacturers vendors and distributors cataloged in the Medical Supply Information Logistics System MEDSILS and available via the DSCP and ECAT system It is the responsibility of the Services to order these items via establi
11. on or the damage prevents the Magellan d Perform Start up procedures in accordance Fails to start up with operating instructions 2 Changing the Gas Manifold Inlet Filters Page 4 3 Replacement of Rubber Bellows and Inspection of Exhalation Diaphragm Valve Page 5 4 B Mandatory Replacement of Gas Manifold High Display fails Pressure Color Coded Tubing Page 9 5 B Calibration Guide Page 10 6 B TESTING AFTER STORAGE Page 12 Equipment Care Ensure Equipment Care is conducted as directed by the manufacturer s literature 14 Maintenance Allocation Chart Anesthesia Machine 6515 01 533 0968 Model 1 FM Crew or Unit Maintenance SM MEDLOG or Depot Maintenance 01 02 03 04 05 06 07 Magellan Temperature Control Reagent Handling System Optics Auto Calibration Scanner Board Incubator Boards Bay interface board 3 Function Inspect Service Test Safety Inspect Replace Inspect Adjust Replace Adjust Test Test Test Replace Test Repair Replace 4 Maintenance Maintenance FM FM FM FM FM FM FM FM FM SM FM SM FM FM FM FM FM 5 01 02 20 20 29 20 01 02 01 02 20 01 02 20 01 02 20 6 Remarks Laptop RS232 amp serial cable VOM sensor tray And tower Alignment tool RS232 amp serial cable VOM sensor tray And tower Alignment too feeler gau
12. shed logistics channels within DOD supply system E Support Equipment This Magellan can be supported with supplies which can be ordered from Oceanic Medical Products via the DSCP and ECAT system see table and pictures Support NSN Mfg Part Support Nomenclature Long e The Bundle Contains Item Description Support Nomenclature Long Quantity Mask Size 0 12 Silicone Respiration Tubing 16mm Mask Size 1 KAB Absorber without APL Valve Mask Size 3 Head Strap Latex Free Head Strap LatexFree l 1 72 Silicone Respiration Ported Y Connector Tubing 16mm 1 0L Silicone Support items sui Magellan 2200 Case Vaporizer Air Compressor Air Compressor Case F Technical Data 1 Service Manual Magellan 2200 Model 1 Anesthesia Machine Service Manual Model 1 Rev July 1 2004 This is available in PDF or IETM via CD by going to the USAMMA website http www usamma army mil med_equip_lit_cdroms cfm and following the instructions or by calling DSN 343 4379 or commercial 301 619 4379 2 Operator s Manuals Magellan 2200 Model 1 Anesthesia Machine Operators Manual Rev July 1 2004 G Computer Resources Support No computer resources are required H Facilities There are no unique facilities shelter or environmental compliance requirements Warehouse storage is required prior to deployment to a unit I Packaging Handling Storage and Transportation The Magellan is packaged in a hardened
13. t level maintenance for the system will be used when the system requires overhaul refurbishment or a safety modification d OEM Level Maintenance Oceanic Medical Products can provide complete maintenance repair and calibration if needed 4 Inspection Procedures a This item has been assigned a type item code TIC C This means prior to being issued from an USAMMA Maintenance Operations Division this item will have a technical inspection to ensure it was received from the vendor according to the contract and that it functions within the guidelines set by the manufacturer and USAMMA b The system is to be inspected on a scheduled basis in accordance with the published 10 20 PMCS or the manufacturer s literature 5 Tools and TMDE required for inspection mo pod Tool kit Medical Equipment 5180 00 117 3414 Tool Set Medical Equipment Maintenance Unit 5180 01 483 1431 Safety system Current Leakage 6625 01 142 8233 Voltmeter 6625 01 265 6000 VT Plus Gas Flow Analyzer 6515 01 541 2864 Riken Anesthetic Gas Analyzer 6630 01 530 7959 6 10 20 Standards and MAC See Appendix D 7 MEDSTEP a MEDSTEP will be maintained at the MEDLOG Company or at one of the USAMMA Maintenance Operation Divisions MEDSTEP requirements will be based on one each per five systems in the support area That number may be subject to validation based on the condition of the equipment and number of units supported MEDSTEP req
14. uirements will be reevaluated on an annual basis The support MEDLOG must be in possession of MEDSTEP in order to affect quick turnaround 8 Repair Parts Spare Parts Listing Repair parts were determined using OEM and or maintenance historical data Spare parts can be obtained from existing stock or vendors Spare parts packages will not be included with each system Recommended spare parts include but are not limited to Ree Part or Accessor Basic A Item No Part or Accessory Description y Operator Repairer Number Issue I micropono __ s8600 1 primer Commoner Jason 1 1 3 thema primer fiso oO TO E een f o 5 Leo Digi ssisa____ 0 I mewaro so Jo 7 spese sas over san 0 senor onsa 0 TO Doorsemor SMISE 9 mo sensi 0 12 service Mant siomozrevom 0 9 Annual Man hour Requirements a Field Support 1 Repair 20 0 hours 2 PM 3 4 hours 3 Safety 1 2 hours 4 Calibration 4 0 hours 5 TOTAL 28 6 hours b Sustainment Support Repair 8 7 hours 10 Warranty Control Procedures The Magellan comes with a 1 year manufacturer s warranty The warranty includes parts and labor When warranty service is needed Oceanic Medical Products should be contacted at 913 874 2000 Their office is located at 8005 Shannon Industrial Park Lane Atchison KS 66005 11 This system is DA 2406 reportable B Training and Training Support 1 Maintainer No additional training is requ