G6014A/B Quiet Cover GC/MS User Information

image

Contents

1. Class III Quiet Cover GC MS has been designed and tested in accordance with recognized safety standards and is designed for use indoors If it is used in a manner not specified by the manufacturer the protection provided by the instrument may be impaired Whenever the safety protection of the Quiet Cover GC MS has been compromised disconnect it Quiet Cover GC MS User Information Symbols Quiet Cover GC MS User Information Introduction from all power sources and secure it against unintended operation Refer servicing to qualified service personnel Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard Warnings in the manual or on the instrument must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions violates safety standards of design and the intended use of the instrument Agilent Technologies assumes no liability for the customer s failure to comply with these requirements Table 1 lists the safety symbols that may appear on Quiet Cover GC MS Table 1 Safety symbols Description Symbol See accompanying instructions for more information gt Indicates a hot surface Pinch hazard Indicates that you must not discard this electrical electronic product in domestic household waste xt gt 1 1 _ Introduction Important Safety Warnings CAUTION CAUTION Safety
2. Cover GC MS User Information 2 Installation Installing the Oil Pan 16 Installing the Rough Pump Inside the Quiet Cover GC MS 17 Connecting the Power Supply 22 Verifying the Installation 23 This chapter describes how to assemble the Quiet Cover GC MS and install the rough pump ee Agilent Technologies 2 Installation Installing the Oil Pan Your Quiet Cover GC MS is shipped to you with the oil pan packaged separately You need to install it prior to using the Quiet Cover GC MS 1 Remove any protective film from the pan 2 Slide the pan into the bottom of the Quiet Cover GC MS with the deep part of the pan facing the front of the Quiet Cover GC MS There is a set of rails along the bottom edges of the frame that guide the pan into place Figure 3 Slide in pan Deep part of pan Figure 3 Installing the oil pan 16 Quiet Cover GC MS User Information Installation 2 Installing the Rough Pump Inside the Quiet Cover GC MS The pump must be disconnected from the GC MS when placing it inside the Quiet Cover GC MS To install the rough pump inside the Quiet Cover GC MS 1 Completely loosen the two thumbscrews on the front of the Quiet Cover GC MS and raise the lid Figure 4 a SE Agilent Technologies G6014A Quiet Cover GCMS Lid Thumbscrews it At i 8 Figure 4 Loosen the thumbscrews and raise the lid The pump may be hot Use gloves when moving it or wait until it
3. it is call your Agilent service representative Do not continue to use your rough pump in the Quiet Cover GC MS The red over temperature LED is not lit Check that power is supplied to Quiet Cover GC MS and the power switch is on even though the room temperature is If it is call your Agilent service representative Do not continue to use your above 35 C 95 F and the alarm is rough pump in the Quiet Cover GC MS the LED may have failed sounding The temperature alarm is not sounding Check that power is supplied to Quiet Cover GC MS and the power switch is on and the red over temperature LED is not f it is call your Agilent service representative Do not continue to use your lit even though the room temperature is rough pump in the Quiet Cover GC MS the thermostat may have failed above 35 C 95 F The rough pump oil does not drain Be sure the kickstand is down with the bar sitting flat on the floor If so refer to completely the rough pump documentation for more information The Service Manual part number G6014 90020 contains part replacement information procedures performed by Agilent service representatives Visit www agilent com chem to download a copy of the Service Manual and look for any other current information Agilent Technologies Shanghai Co Ltd 2015 G6014 90010
4. HE 101150234 5 Agilent G6014A B Quiet Cover GC MS User Information RE Agilent Technologies Notices Agilent Technologies Inc 2015 No part of this manual may be reproduced in any form or by any means including electronic storage and retrieval or transla tion into a foreign language without prior agreement and written consent from Agilent Technologies Inc as governed by United States and international copyright laws Manual Part Number G6014 90010 Edition Third edition April 2015 Second edition November 2013 First edition May 2013 Printed in China Agilent Technologies Shanghai Co Ltd 412 Ying Lun Road Waigaogiao Free Trade Zone Shanghai 200131 P R China Warranty The material contained in this docu ment is provided as is and is sub ject to being changed without notice in future editions Further to the max imum extent permitted by applicable law Agilent disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a par ticular purpose Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing use or perfor mance of this document or of any information contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material i
5. S is both the level and tilted Tilted position mn ZT A AAKA o Y P Kickstand engaged Figure 13 Quiet Cover GC MS in level and tilted positions CAUTION CAUTION Quiet Cover GC MS User Information Be sure that the kickstand s base sits flat on the floor when the Quiet Cover GC MS is tilted To return the Quiet Cover GC MS to a level position 1 Gently push the Quiet Cover GC MS forward so it rocks off of the kickstand Take care when lowering the Quiet Cover GC MS from the tilted position 2 If necessary move the Quiet Cover GC MS back to its operating location 31 3 Operation 32 Quiet Cover GC MS User Information Table 4 Problems and solutions Problem Pump does not fit in Quiet Cover GC MS Quiet Cover GC MS does not turn on Temperature alarm does not turn off Quiet Cover GC MS poorly muffles the sound of the pump Green LED does not light The temperature alarm does not sound even though the room temperature is above 35 C 95 F and the red over temperature LED is lit Agilent G6014A Quiet Cover GC MS User Information 4 Troubleshooting Table 4 describes solutions to problems that may occur with Quiet Cover GC MS If your problem is not listed or the solution listed does not work consult your pump documentation or contact your Agilent service representative Hot surfaces can cause burns Observe the cautions and warnings covered elsewhere in this manual So
6. ant Safety Warnings 8 Cleaning 10 Recycling the Product 10 Parts Identification 11 Site Preparation 13 This manual describes how to install operate and troubleshoot the Agilent G6014A B Quiet Cover GC MS for GC MS rough pumps Your rough pump fits in the Quiet Cover GC MS to significantly reduce noise while an internal fan ensures your pump will not overheat There is an easy to use kickstand and a full size oil pan to facilitate changing the oil The lid lifts for easy access to the pump for routine maintenance ee Agilent Technologies 9 1 _ Introduction Compatibility General Information Information The Quiet Cover GC MS is only compatible with the following pump models e Pfeiffer Duo 2 5 G6014A through revision F G6014B all revisions e Edwards E2M1 5 e Agilent DS 42 e Agilent DS 42i The Quiet Cover GC MS conforms to the International Electrotechnical Commission IEC 61010 1 safety standard It also conforms to the following regulations on Electromagnetic Compatibility EMC and Radio Frequency Interference RFI e CISPR 11 EN 55011 Group 1 Class A e IEC EN 61326 e AUS NZ This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme a la norme NMB 001 du Canada The Quiet Cover GC MS is designed and manufactured under a quality system registered to ISO 9001 The Agilent Technologies Quiet Cover GC MS meets the following IEC classifications Category II Pollution Degree 2
7. cautions and warnings must be observed during all phases of operation service and repair of Quiet Cover GC MS Failure to comply with these precautions violates safety standards of design and Quiet Cover GC MS s intended use Agilent Technologies assumes no liability for the customer s failure to comply with these requirements Never run the pump without also running Quiet Cover GC MS While running Quiet Cover GC MIS Do not block air flow to the fan and vents e Do not set items on top of Quiet Cover GC MS e tis important that the Quiet Cover GC MS not recirculate hot air Always position the back end of the Quiet Cover GC MS where it can draw cool air Be sure to position Quiet Cover GC MS so that there is plenty of room to allow access to panels and controls See Table 3 for more information In the event of a temperature alarm malfunction or if the alarm has been on for an extended period of time the pump may be dangerously hot Check that the pump is cool before you touch it Quiet Cover GC MS User Information Introduction Electromagnetic compatibility This device complies with the requirements of CISPR 11 Operation is subject to the following two conditions This device may not cause harmful interference This device must accept any interference received including interference that may cause undesired operation If this equipment does cause harmful interference to radio or television reception which ca
8. e lid Refer to Chapter 4 Troubleshooting for more information e There should be neither excessive noise nor vibration If either condition is true go back over each installation step to be sure everything was done correctly and all thumbscrews are securely fastened The power cords should be positioned so as to not interfere with the kickstand Quiet Cover GC MS User Information 23 2 Installation 24 Quiet Cover GC MS User Information Agilent G6014A B Quiet Cover GC MS User Information 3 Operation Front Panel Controls and Indicators 26 Handling Temperature Alarms 27 Changing Rough Pump Oil 28 Using the Kickstand 31 To operate Quiet Cover GC MS you will need to e Understand the controls and indicators on the front panel e Know how to handle temperature alarms and signals e Know how to drain oil from the rough pump and oil pan e Be able to tilt the Quiet Cover GC MS using the kickstand for draining the oil ee Agilent Technologies 25 3 Operation Front Panel Controls and Indicators Figure 10 shows the location of the on off switch and indicators of the Quiet Cover GC MS front panel gt Agilent Technologies G6014A Quiet Cover GEMS TT E B Power LED Green Temperature Alarm LED Red indicates excessive heat when lit On Off switch Me aaa Thumbscrews to release access door op ce Sr Figure 10 The front panel of Quiet Cover GC MS 26 Quie
9. ess port on the back of the Quiet Cover GC MS Figure 7 Rectangular edge protected access port Pump power cord TO LOL ot 2 8 Figure 7 Power cord installed through access port Quiet Cover GC MS User Information Installation 2 5 Install the pump hoses to the pump and run them through the edge protected openings on the back of the Quiet Cover GC MS Figure 8 Refer to the pump documentation for more information Figure 8 Installed pump hoses WARNING A two person lift is required when moving the Quiet Cover GC MS with a pump installed Failure to perform a two person lift may result in personal injury Quiet Cover GC MS User Information 21 2 Installation Connecting the Power Supply Connect the power supply to the back of Quiet Cover GC MS by twisting the ring on the circular connector of the power cord Figure 9 Plug the pronged end of the power cable into the wall outlet TO PUL ot y Y Plug in power supply cord here Figure 9 Connecting the power supply 22 Quiet Cover GC MS User Information Installation 2 Verifying the Installation Press the on off switch located on the right side of the front panel to verify that you installed Quiet Cover GC MS correctly e The green LED should light and the fan should start turning e The internal light should be on e The temperature alarm should be off and the red light should not be lit If either are on raise th
10. g the Kickstand on page 31 This technique is the most efficient way to drain the pump oil Be sure the oil pan is securely seated Quiet Cover GC MS User Information Operation 3 8 Remove the oil drain plug from your rough pump Refer to your pump documentation Figure 11 Oil drain plug Figure 11 View of pump 9 Reinstall the drain plug when the oil has completely drained 10 Return the Quiet Cover GC MS to its upright position See Using the Kickstand on page 31 11 Remove the oil pan and dispose of the oil according to local safety and environmental guidelines 12 Reinsert the oil pan Quiet Cover GC MS User Information 29 3 30 Operation 13 Remove the oil fill plug Figure 12 and refill the pump Refer to the pump documentation for more information Oil fill plug Sight glass u a NM VENDEE Elke Smi ur e Ma er ur e a Pa al u Figure 12 View of pump through access door 14 Reinstall the oil fill plug and lower the lid Quiet Cover GC MS User Information Using the Kickstand Level position Operation 3 To tilt the Quiet Cover GC MS 1 2 Make sure the Quiet Cover GC MS is on a level surface Ensure that the lid is closed and the two thumbscrews securing the lid to the Quiet Cover GC MS are tight On the back of the Quiet Cover GC MS lift gently on the rear lid handle to engage the kickstand Figure 13 shows the Quiet Cover GC M
11. is cool Quiet Cover GC MS User Information 17 2 Installation 2 Install the rough pump into the Quiet Cover GC MS Place the rough pump base centered left to right and as far back as possible on the Quiet Cover GC MS base leaving a small gap 2 to 6 mm between the pump and the back of the G6104A B Figure 5 Quiet cover base Figureb Mounting a rough pump onto the Quiet Cover GC MS G6104A shown Make sure the pump power cord is not crimped or under the rough pump Make sure each support foot on the rough pump sits on the Quiet Cover GC MS frame base 18 Quiet Cover GC MS User Information Installation 2 e If you have a DS 421 pump tuck the power cord between the flange and the foam as shown in Figure 6 Flange Power cord Foam ONE R 3603 AI N Fr F ri d i f tl bo 7 EF j zi Ma i hi al l i yo A r ma ZA y iy mm 3 i nl 7 F F k y E n a Figure 6 Tick power peu peeves flange and foam DS 42i pump shown T 3 Switch the rough pump power switch to the On position Refer to the pump documentation for more information NOTE The pump is controlled by the MS and should always be left on Quiet Cover GC MS User Information 19 2 Installation 4 If not installed connect the pump power cord to the rough pump then feed the power cord through the rectangular edge protected acc
12. lution Check for proper orientation of the pump and configuration of the vibration dampers See Installing the Rough Pump Inside the Quiet Cover GC MS on page 17 Call your Agilent service representative if the orientation and configuration are correct Check that you are installing the one of the supported pumps into the Quiet Cover GC MS See Compatibility on page 6 Check that the power supply is plugged securely into both Quiet Cover GC MS and the wall outlet If it is call your Agilent service representative Do not continue to use your rough pump in the Quiet Cover GC MS Raise the Quiet Cover GC MS lid Ensure that adequate cool air is available to the Quiet Cover GC MS Call your Agilent service representative if the lab temperature is within specifications Ensure that the rough pump is placed securely on the base and that the lid fastened tightly Check that power is supplied to Quiet Cover GC MS and the power switch is on If it is call your Agilent service representative Check that power is supplied to Quiet Cover GC MS and the power switch is on If it is call your Agilent service representative Do not continue to use your rough pump in the Quiet Cover GC MS the alarm may have failed Ed Agilent Technologies 33 Agilent Technologies Table 4 Problems and solutions continued Problem Solution Fan does not turn Check that power is supplied to Quiet Cover GC MS and the power switch is on If
13. n be determined by turning the equipment off and on the user is encouraged to try one or more of the following measures Relocate the radio or antenna Move the device away from the radio or television Plug the device into a different electrical outlet so that the device and the radio or television are on separate electrical circuits Make sure that all peripheral devices are also certified Make sure that appropriate cables are used to connect the device to peripheral equipment Consult your equipment dealer Agilent Technologies or an experienced technician for assistance Changes or modifications not expressly approved by Agilent Technologies could void the user s authority to operate the equipment Sound emission certification for Federal Republic of Germany Quiet Cover GC MS User Information Sound pressure Lp lt 70 dB am according to EN 27779 1991 Schalldruckpegel LP lt 70 dB am nach EN 27779 1991 1 _ Introduction Cleaning To clean the Quiet Cover GC MS 1 Reach under the front of Quiet Cover GC MS and pull out the oil pan 2 Wipe out excess oil with a clean lint free cloth 3 Wash the pan with a mild soap solution 4 Wipe away any spills or other debris from the Quiet Cover GC MS using a clean lint free cloth Recycling the Product For recycling contact your local Agilent sales office 10 Quiet Cover GC MS User Information Parts Identification Introduction 1 Table 2 lists
14. n this document that conflict with these terms the warranty terms in the sep arate agreement shall control Safety Notices CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met Contents 1 Introduction 2 Installation 3 Operation 4 Troubleshooting Quiet Cover GC MS User Information Compatibility 6 General Information 6 Important Safety Warnings 8 Cleaning 10 Recycling the Product 10 Parts Identification 11 Site Preparation 13 Installing the Oil Pan 16 Installing the Rough Pump Inside the Quiet Cover GC MS 17 Connecting the Power Supply 22 Verifying the Installation 23 Front Panel Controls and Indicators 26 Handling Temperature Alarms 27 Changing Rough Pump Oil 28 Using the Kickstand 31 Quiet Cover GC MS User Information Agilent G6014A B Quiet Cover GC MS User Information 1 Introduction Compatibility 6 General Information 6 Import
15. t Cover GC MS User Information Operation 3 Handling Temperature Alarms If the air flow to the Quiet Cover GC MS is restricted or if the room temperature is above 35 C 95 F the red LED on the front panel lights and the temperature alarm sounds When this condition occurs loosen the front thumb screws and fully raise the cover to cool the pump Check that the fan is turning and that air flow to the Quiet Cover GC MS is not restricted It is important that the Quiet Cover GC MS not recirculate hot air Always position the back end of the Quiet Cover GC MS where it can draw cool alr You cannot turn off the alarm It will stop when the temperature returns to an acceptable level Quiet Cover GC MS User Information 21 3 Operation Changing Rough Pump Oil 28 Periodically you must change the rough pump oil The oil pan of the Quiet Cover GC MS is designed to collect the oil as it drains from the pump Please refer to your pump documentation for detailed information regarding changing your pump oil To change the oil 1 oar Bb GW N Properly prepare your system for release of primary vacuum Turn off the rough pump Turn off the Quiet Cover GC MS Unplug the power supply from the Quiet Cover GC MS Completely loosen the two thumbscrews on the front panel and raise the Quiet Cover GC MS access lid Tilt the Quiet Cover GC MS by using the kickstand located at the back of the Quiet Cover GC MS See Usin
16. the contents of the Quiet Cover GC MS kit Figure 1 and Figure 2 depict the parts of the Quiet Cover GC MS Table 2 Agilent G6014A B Quiet Cover GC MS contents list Part number Part Quantity Quiet Cover GC MS 1 Oil pan 1 Power supply 1 Access lid __ A GHAR conc Fan ES gt behind Alarm n MYM m Oil pan Figure 1 Quiet Cover GC MS front view G6014A B 5190 6165 5188 1164 On Off switch See Figure 10 on page 26 for a view of the front panel controls Quiet Cover GC MS User Information 11 1 _ Introduction Pump hose Pump drainage tube Power supply Kickstand Figure 2 Quiet Cover GC MS rear view 12 Quiet Cover GC MS User Information Site Preparation Quiet Cover GC MS User Information Introduction Prepare the installation site to meet the criteria in Table 3 Table 3 Site prep information Criteria Description Physical dimensions Weight 11 5 kg 25 Ibs Dimensions Kickstand in tilted position L x W x H 51 x 27 x 40 cm 20 x 10 5 x 15 4 inches Power supply AC input voltage 100 to 240 VAC 10 DC output voltage 24V Output load max 1 25A Input frequency 47 to 63 Hz Power consumption 11 W Rated current 450 mA Temperature and humidity ranges Ambient temperature 5 to 35 C 41 to 95 F Humidity 15 to 95 Altitude lt 4600 m 14950 ft 1 13 1 Introduction 14 Quiet Cover GC MS User Information Agilent G6014A Quiet