        Install the Nickel Catalyst Tube
         Contents
1.       Installation and Operating Guide  Nickel Catalyst Tube  Accessory 19205A    E Agilent Technologies          Agilent Technologies 2000    All Rights Reserved  Reproduction   adaptation  or translation without  permission is prohibited  except as allowed  under the copyright laws     Part number 19205 90107  First Edition  DEC 2000    Replaces Part No 19205 90100 Operating  and Service Manual     HP   is a registered trademark of  Hewlett Packard Co     Printed in USA    Safety Information    The Agilent Technologies Nickel Catalyst  Tube meets the following IEC  International  Electrotechnical Commission   classifications  Safety Class 1  Transient  Overvoltage Category Il  and Pollution  Degree 2     This unit has been designed and tested in  accordance with recognized safety  standards and designed for use indoors  If  the instrument is used in a manner not  specified by the manufacturer  the  protection provided by the instrument may  be impaired  Whenever the safety protection  of the Agilent 19205 has been compromised   disconnect the unit from all power sources  and secure the unit against unintended  operation     Refer servicing to qualified service  personnel  Substituting parts or performing  any unauthorized modification to the  instrument may result in a safety hazard   Disconnect the AC power cord before  removing covers  The customer should not  attempt to replace the battery or fuses in  this instrument     Safety Symbols    Warnings in the manual or on th
2.  2 to  3 4 turn     Nickel catalyst Nickel catalyst  exit tube inlet tube                 H    Ny  Gas sample    a Tube retaining nut  valve f      nnnann                      Hg                   PorapakQ           4  column              6  Install the nickel catalyst inlet tube to the column bracket on the right  side of the oven     7  Install the 6 ft  Porapak Q column between the gas sampling valve fitting  on the inlet side of the oven and the nickel catalyst inlet tube on the  detector side of the oven     WARNING Hydrogen  H5  is flammable and is an explosion hazard when confined in an  enclosed space  for example  the oven   In any application using H    turn off  the supply at its source before working on the instrument     8  Turn offthe hydrogen at its source  Locate the hydrogen line  marked  in red  at the flow manifold that feeds the FID being used  Trace the  line to the area behind the nickel catalyst tube  Using a suitable tubing       The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    cutting tool  carefully cut the hydrogen line  Be careful not to crush the  tubing while cutting it     Hydrogen supply to FID  Hydrogen supply to Nickel Catalyst m di  Plugged 1 16 inch union  1 16 inch union C ZA  7 AU d E A  I a      74 nes jp                                Hydrogen line    Marked in red          9  Using the stainless steel 1 16 inch union provided  connect the  hydrogen supply side of the line to one side of the union  Insert t
3.  A position       2  Using the next figure as a guide  install the Nickel Catalyst tube from the  top of the oven  Use the insulation removed  tear apart as needed  to fill  in the area around the Nickel Catalyst Tube        The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    3  Install the cover plate provided  Secure the cover plate with two M4  Screws         M4 screws GET  N    AS     Cover plate    H3 inlet tube          To right column                           mounting bracket                                  The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    4  Place the three pieces of insulation in the insulation cup  Install the cup  over the carrier gas tubing and nickel catalyst tube inside the oven   Secure the cup using two M4 screws provided                       4  MI  s  4                               Nm                   i  1  I  1  I  I  l  1  I  l  I  I  1              D     onnAnn An n     amp                     Io FID    To right column  mounting bracket                             hoa                               Oo OUI                5  Connect the nickel catalyst exit tube to the FID inlet  Assemble a  1 8 inch stainless steel nut and front and back ferrule onto the tubing   Fully insert the tube  ferrules  and nut onto the FID base  Tighten the       The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    nut finger tight  Use a wrench to tighten the nut an additional 1
4.  hazard when confined in an  enclosed space  for example  the oven      13  Restore carrier gas flow and leak test the new installation   14  Reinstall the panels and covers  Reconnect the instrument power cord    and restore power        The Nickel Catalyst Tube Accessory 19205A    Operation    Operation    Verifying Performance    l  Ifany heated zone is turned on  turn it off and allow it to cool  DO NOT    heat any heated zone     2  Setthe gas flows     Table 1  For a standard FID installation                       Flow mL min  Carrier flow  helium  30   FID hydrogen flow 30   FID air flow 400       Table 2  If there is a TCD FID in series installation                          Flow mL min  Carrier flow  helium  30   TCD switching flow 25   FID hydrogen flow 45   FID air flow 500       Allow helium and hydrogen to flow through the system for 15 minutes     4  Carefully check the column and supply fittings for leaks  Air leaks can    damage the catalyst     5  Condition the nickel catalyst by setting the nickel catalyst tube  temperature  INJ B TEMP  to 375 C and the FID temperature to 400 C   Allow them to condition for 1 hour  While the zones are heating  ignite    the FID flame              The Nickel Catalyst Tube Accessory 19205A    Operation    6  After 1 hour  set the following temperature                          Setting   C   Oven temperature 35   FID temperature as  desired   Nickel catalyst 375   TCD  if in use  as  desired          When the temperatures have st
5. abilized  the system is ready for use     7  Load the sample loop with the nickel catalyst test mixture  a quick press  of the can button  and start the run  The chromatogram should be  similar to the one in the figure below  CO  may be larger due to the    presence of air in the sample      START    IF          STOP    Using the Nickel Catalyst    1484 CO  CHa    To use the Nickel Catalyst accessory most efficiently  keep the following in    mind     e Avoid exposing the catalyst to oxygen  which can damage it     e If performance is significantly degraded  repack the catalyst tube        WARNING    WARNING    WARNING    Caution    The Nickel Catalyst Tube Accessory 19205A  Maintenance    Maintenance    Repacking the Catalyst    Should it be necessary to repack the nickel catalyst tube because of loss of  conversion efficiency  repack the catalyst as described below     Hydrogen  H5  is flammable and is an explosion hazard when confined in an  enclosed space  for example  the oven   In any application using H    turn off  the supply at its source before working on the instrument     Both nickel oxide and some forms of silicon oxide are considered  carcinogens for humans  It is important to perform all work in a fume hood  and to wear cotton gloves at all times  Remove any spills with a HEPA type  vacuum cleaner  avoiding any action that creates dust  Alert your company s  safety group when a spill is created     Due to the possibility of dermatitis  wash the arms and hands w
6. alyst assembly upside down and add catalyst slowly   Tap gently on the tube to help seat it properly  When the catalyst is  22 mm from the end  stop adding catalyst  Do NOT crush the catalyst  when packing or measuring the depth     14  Add a single glass wool plug to fill the remaining part of the tube to  within 5 mm of the end  This plug should be gently compressed during  installation     Before installing the catalyst assembly into the oven  carefully wipe it to  remove any catalyst dust     15  Reassembly of the nickel catalyst is the reverse of steps 1 through 6        The Nickel Catalyst Tube Accessory 19205A    Maintenance                                        WARNING Make sure that the insulation is carefully repacked around the tube before  you reinstall the injector cover plate and the capillary retainer cup   16  Leak test the new installation   WARNING Hydrogen  H5  is flammable and is an explosion hazard when confined in an  enclosed space  for example  the oven    17  Start the carrier and hydrogen flows  then allow them to flow for  15 minutes   18  Heat the nickel catalyst to 375  C and hold for 30 minutes  The  accessory should be ready for use    Replacement Parts List   Ref Name Part   No  Number   1 Packed Catalyst Tube Assembly 18900 61020   2 Reducer 0100 0073   3 H   Mix Weldment 18900 80820   4 520mm 1 16 inch Tube with 1 8 inch end 1530 2167   5 Nut  Front ferrule  and Back ferrule SS 1 4 inch  kit of 20  5080 8753   6 1 16 inch x 1 16 inch SS Unio
7. e  instrument must be observed during all  phases of operation  service  and repair of  this instrument  Failure to comply with these  precautions violates safety standards of  design and the intended use of the  instrument  Agilent Technologies assumes  no liability for the customer s failure to  comply with these requirements     WARNING   A warning calls attention to a condition or  possible situation that could cause injury to  the user     CAUTION   A caution calls attention to a condition or  possible situation that could damage or  destroy the product or the user s work   Sound Emission Certification for Federal  Republic of Germany    Sound pressure Lp    68 dB A     During normal operation  At the operator position  According to ISO 7779  Type Test     Schallemission    Schalldruckpegel LP    68 dB A   Am Arbeitsplatz   Normaler Betrieb   Nach DIN 45635 T  19  Typpr  fung        Agilent Technologies  Inc   2850 Centerville Road  Wilmington  DE 19808 1610  USA    The Nickel Catalyst Tube  Accessory 19205A    WARNING    The 19205A Nickel Catalyst Tube Accessory is used on the 4890 5890 gas  chromatographs for trace analysis of CO and CO  using a Flame Ionization    Detector  FID   The gas sample is separated on the column  then passes over  the hot catalyst in the presence of hydrogen so that the CO and CO  are    converted to CH4     The Nickel Catalyst Tube Accessory is installed in the inlet B position   These instructions describe installation of the accessory for 
8. he tube    into the fitting fully  Use a wrench to tighten the nut 1 2 to 3 4 turn past  finger tight     10  Insert the 1 16 inch hydrogen supply tube from the nickel catalyst tube  into the other end of the union and tighten as described above     11  Install the plugged union provided on the hydrogen line from the FID  detector        The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    12  Route the heater and sensor wires through the cable tray behind the  oven to the main circuit board  Install the pins in the B injection port  connector positions     4890 Series and 5890 SERIES II Main Circuit Board    Connect the sensor leads  white  to P7  heater leads  red  to J9     Sendors               EI  mmm aue       Heaters    i     jemois    2     jrorses  4      mass L J9 L    joris  B     EA oe  iD       6                   Position 3  amp  4   Inj B position                The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    5890A Main Circuit Board    Connect the sensor leads  white  to J7  heater leads  red  to J9            asoa    Da    0501      y             White       STRT 11 amp 125  Door 13 amp 14     m  Z a manwa c ao ote           Position 9  amp  10   Inj B position                     Cryo 142     INI B 3642  LINT A 5 amp  J9     Det B 7885 Heaters      Det A 9 amp 0    sition 3  amp  4   Inj B position  Red       P    o                                  WARNING Hydrogen  H5  is flammable and is an explosion
9. ith soap and  water after use  The use of long sleeves is required during any use and spill  cleanup  If long sleeves are not worn  long gloves are an acceptable  substitute  After the assembly is filled with catalyst  the whole assembly  should be carefully wiped to remove any catalyst dust     Be sure to read the Material Safety Data Sheet  MSDS  provided with the  catalyst before performing this procedure     1  Turnoffthe INJ B TEMP  Turn offall other heaters  When the nickel  catalyst tube is cooled to room temperature  turn off the power to the  GC and disconnect the power cord  Bleed down the hydrogen and  carrier gas     2  Use a Pozidrive screwdriver to remove the two screws holding the  injector cover plate on top of the nickel catalyst tube  Remove the plate  and the insulation around the nickel catalyst tube        The Nickel Catalyst Tube Accessory 19205A    Maintenance    From inside the oven  loosen the two screws holding the capillary  retainer cup  rotate the cup and allow it to drop away from the base of  the nickel catalyst assembly     Refer to the following figure which shows the nickel catalyst assembly  and the packed catalyst assembly  Use two wrenches to disconnect the  H   mix weldment  5  from the bottom of the packed catalyst assembly     1   Be very careful NOT to place stress on the 1 16 inch tube  Stress  can damage the weldment                Packed catalyst tube       Packed  area    Use two wrenches to disconnect the reducer  2  from the top 
10. n 0100 0124   7 Brass Cap seal 1 8 inch Supplied with replacement   8 Polyethylene Cap Supplied with replacement   9 Nut  Front ferrule  and Back ferrule SS 1 8 inch  kit of 20  5080 8751   10 Nut  Front ferrule  and Back ferrule SS 1 16 inch  kit of 10  5180 4149                      The Nickel Catalyst Tube Accessory 19205A  Maintenance                                  ns NICAT Test Sample  pressurized can  19354 60510  ns Injector Cover Plate 19243 00050  ns Capillary Retainer Cup 19243 00070  ns Insulation for Cup  3 required  19243 00067  ns Column Bulkhead Fitting and Tube 07675 80050  ns Retaining Nut for Bulkhead Fitting 2950 0203   ns 6 ft x 1 8 inch OD Porapak Q Column 19006 80015                   The Nickel Catalyst Tube Accessory 19205A  Maintenance       20    RES Agilent Technologies          Printed on recycled paper     This product is recyclable     Agilent Technologies  Inc   Printed in USA Dec 2000    19205 90107    
11. of the  packed catalyst assembly  1      Gently lift the packed catalyst assembly  1  out of the injection area   Both ends of the catalyst tube are now accessible        WARNING    Caution    The Nickel Catalyst Tube Accessory 19205A  Maintenance    7  With a hooked instrument of some type  carefully remove the glass wool  plug from the bottom of the tube  Make sure you get all of it     8  Empty the old catalyst from the tube  you may have to break it out with  a pointed tool   Make sure you get it all out     9  Using a thin rod  push out the top glass wool plug from the tube     10  Clean the inside of the tube thoroughly with methanol  Do not use any  sharp metal tools on the inside of the tube  A cotton swab carefully  used will assure cleanliness  Dry the tube     The figure above shows the dimensions for repacking the tube correctly  If  any catalyst is outside the heated zone  severe tailing of CO will result     ll  Prepare a simple depth gage using a wooden cotton swab or any other  handy rod or tubing  Use tape or paint to mark the stick at 46 mm and  at 22 mm from the blunt end     12  Roll up a piece of glass wool into a ball about the size of a large pea   Push this into the tube from the 1 4 inch end and seat it firmly   Measure the depth of this glass wool with the depth guage   it should be  46 mm from the end of the tube  If necessary  add more glass wool  A  slight compression of the glass wool during the measurement works  best     13  Turn the nickel cat
12. two tray locks that hold it in  position  then sliding the tray away from the instrument     b  Liftthe autosampler tray from its mounting bracket and set it aside        The Nickel Catalyst Tube Accessory 19205A  Prepare the Instrument    c  Remove the autosampler bracket by removing the six screws  securing it to the instrument     Left side panel       8  Remove the two screws securing the left side panel along its bottom  edge     9  Slide the left side panel towards the rear of the instrument and lift     10  Remove the electronics carrier top cover by grasping it at the rear and  lifting upwards until its catch releases  then pull it towards the rear of  the instrument     Electronics carrier top cover             The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    11  Remove the right side panel by removing four screws  two each along its  top and bottom edges     Screws             Right side panel    Install the Nickel Catalyst Tube    The insulation on the GC is made of refractory ceramic fibers  Ventilate  your work area  Wear long sleeves  gloves  safety glasses  and a disposable  dust mist respirator  Dispose of insulation in a sealed plastic bag     Caution       The Nickel Catalyst Tube Accessory 19205A  Install the Nickel Catalyst Tube    l  If present  remove the cover plate from the B inlet mounting location  and the pre cut insulation below it     Inlet cover plate ZA  Ze  KEP     pO            eS hn  m   B position  Pun          
13. use with a gas  sampling valve  although the accessory can be used with any inlet      Prepare the Instrument    Hazardous voltages are present in the instrument whenever the power cord  is connected  Avoid a potentially dangerous shock hazard by disconnecting  the power cord before working on the instrument        Copyright   2000  Agilent Technologies    Printed in USA Dec 2000  Agilent Part No  G2651 90107    The Nickel Catalyst Tube Accessory 19205A  Prepare the Instrument    1  Set the main power line switch to the off position         Red 0 visible    UL                  n  N    0       Disconnect the power cable from its receptacle   Allow time for the oven and heated zones to cool     When the heated zones are cool  turn off all gas supplies at the source     oP WO hf    Remove the column and any hardware attached to the FID  liner   column liner nuts  ferrules  makeup gas adapter  etc             The Nickel Catalyst Tube Accessory 19205A  Prepare the Instrument    6  Remove the injection port cover by grasping its back edge and lifting it  upward  If an autosampler is installed  the injection port cover will not  be present     Injection port cover        Press here  to free cover    1 4 inch screw  and washer    7  If an autosampler is installed on the instrument  it will be necessary to  remove it and its mounting bracket to allow removal of the left side  cover     a  Remove the autosampler tray from its mounting bracket by  simultaneously lifting and turning the 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
369 User Manual Set 1_rev1  Flash - TAG - Techniques Audio Groupe  Manual do Produto  Beppe Grillo: opera omnia  VersaMax System Profibus Network Modules User`s Manual, GFK  Diesel Diagnosis  The Czech Academic Corpus 2.0 Guide  2 - Webee  Affiche Bébé mode d`emploi Calais 13.11.14  JCONTEXTEXPLORER USER MANUAL    Copyright © All rights reserved. 