Home

RTX100B ISDB-T RF Signal Generator Service Manual

image

Contents

1. 50 Q N Fe to 75 Q FE LS N Ma adapter 4 Moo ao o 50 2 N Ma to SMA Fe adapter 50 Q N Ma to SMA Fe adapter 50 Q SMA cable Figure 3 15 Equipment connection for checking the carrier leakage 3 26 2 Setthe spectrum analyzer as indicated below CENTER FREQ 473 143 MHz FREQ SPAN 2MHz RBW 10 kHz VBW s oues 300 Hz SWP Speed 2 0s REF Level 20 dBm ANTT ovid 10 dB Select ISDB T ASI Calibration on the RTX100B The ISDB T Calibration Password dialog box appears In the dialog box enter rtx100cal for a password The ISDB T Calibration dialog box appears In the dialog box set Reference Frequency to 13 Verify that the carrier leakage value on the spectrum analyzer is less than 80 dBm RTX100B ISDB T RF Signal Generator Service Manual Performance Verification 7 Change Reference Frequency output channel on the RTX100B and center frequency on the spectrum analyzer as listed in Table 3 3 and then verify that the carrier leakage value is less than 80 dB Table 3 3 Output channel and center frequency settings Center Frequency Center frequency Center frequency CH MHz MHz MHz 13 473
2. OQ 75 BNC cable Figure 3 16 Equipment connection for checking the output level and error 23 24 25 26 27 28 29 RTX100B ISDB T RF Signal Generator Service Manual Select ISDB T ASI RF Parameter on the RTX100B to open the ISDB T RF Parameter dialog box In the dialog box set Center Frequency to 13 Open the ISDB T M3 rmx file on the RTX100B a Select File Open to open the Select File dialog box b In the dialog box navigate to the D ISDB T directory and then select the ISDB T M3 rmx file Press the Play Pause button to start playing the file Set the ISDB T analyzer demodulation DEMODSET as indicated below Mode 3 GI npe pt 1 4 Data AIR ES ors sees ON ALL Loss OFF Set the ISDB T analyzer other settings as indicated below Display Constellation Measurement mode REPEAT Mode Input CH 13 After the Synchronization Detect status is locked start the measurement 3 29 Performance Verification 30 Verify that the measurement values are as follows POWER Mode 3 22 to 16 dBm POWER Mode 2 25 to 19 dBm POWER Mode 1 28 to 22 dBm Plane A B BER 0 00E 7 8 Plane C BER lt 8 00E 5 Plane A B C VIT 0 00 7 8 NOTE If errors occur over Planes A to C
3. 3 7 Output channel and center frequency settings 3 27 Equipment required 4 1 Output channel and center frequency settings 4 3 External inspection checklist 5 4 Internal inspection check list 5 5 Summary of procedures _ 5 8 Tools required for module removal 5 9 RTX100B ISDB T RF Signal Generator Service Manual a eee General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it To avoid potential hazards use this product only as specified Only qualified personnel should perform service procedures To Avoid Fire or Use Proper Power Cord Use only the power cord specified for this product and Personal Injury certified for the country of use Ground the Product This product is grounded through the grounding conductor of the power cord To avoid electric shock the grounding conductor must be connected to earth ground Before making connections to the input or output terminals of the product ensure that the product is properly grounded Observe All Terminal Ratings To avoid fire or shock hazard observe all ratings and markings on the product Consult the product manual for further ratings information before making connections to the product Do Not Operate Without Covers Do not operate this produc
4. line power Rating Voltage 100 VAC to 240 VAC CAT II Voltage Range 90 VAC to 250 VAC Frequency Range 50 Hz to 60 Hz Maximum power 180 VA Maximum current 1 3A Main fuse data 5 A Time delayed 250 V not operator replaceable 2 A Time delayed 250 V not operator replaceable 1 10 RTX100B ISDB T RF Signal Generator Service Manual Specifications DATA 0 7 privet X X 932 DCLK J PSYNC DVALID 188 bytes T S T 1f gt DATA 0 7 Transition period e a 5ns 5ns 1 y T 2 DCLK Output data delay Input clock pulse width T 2 TA0 t 1 DATA 0 7 DCLK Input data hold time Figure 1 1 Timing diagram of the SPI interface RTX100B ISDB T RF Signal Generator Service Manual 1 11 Specifications Mechanical Physical Characteristics Table 1 3 Mechanical characteristics Characteristics Net weight Standard Dimensions Height Width Length Description Approximately 6 kg 13 2 Ib 132 mm 5 2 in without feet 214 mm 8 4 in 435 mm 17 1 in Environmental Characteristics Table 1 4 Environmental characteristics Characteristics Temperature Operating Non operating Relative humidity Operating Non operating Altitude Operating Non operating Dynamics Vibration Operat
5. 50 N Ma to SMA Fe 50 Q N Ma to SMA Fe adapter adapter 50 Q SMA cable Figure 4 1 Equipment connection for adjusting the carrier leakage 2 4 2 Set the spectrum analyzer as indicated below CENTER FREQ 479 143 MHz FREQ SPAN 15 MHz RBW vane eas ces tes 30 kHz VBW see A es 300 Hz SWP Speed 0 5 s Meas Uncal REF Level 0 dBm ALT setae Ga eae 10 dB Select ISDB T ASI gt Calibration on the RTX100B The ISDB T Calibration Password dialog box appears In the dialog box enter rtx100cal for a password The ISDB T Calibration dialog box appears In the dialog box set Reference Frequency to 13 Set DAC I Gain and DAC Q Gain to 15 Select the Cal Signal check box RTX100B ISDB T RF Signal Generator Service Manual Adjustment Procedures 8 Use the Peak Search function of the spectrum analyzer to verify that the carrier appears at 473 143 MHz on the screen 9 Turn off the Carrier output of the spectrum analyzer 10 Adjust the DAC I Offset and DAC Q Offset values so that the carrier leakage on the spectrum analyzer is less than 80 dBm 11 Change Reference Frequency output channel on the RTX100B and center frequency on the spectrum analyzer as listed in Table 4 2 and repeat steps 5 through 10 NOTE The center frequency settings for the spectrum analyzer are indicated by boldface in the Table 4 2 Table 4 2 Output channel and cente
6. Figure 4 1 Equipment connection for adjusting the carrier leakage Figure 5 1 RTX100B Figure 5 2 External modules Figure 5 3 Board modules Figure 5 4 Internal modules Figure 5 5 Guide to removal procedures Figure 5 6 Cabinet 1 Figure 5 7 Right side and left side covers removal Figure 5 8 Front panel unit removal Figure 5 9 Disassembly of front panel assembly Figure 5 10 A150 RF Output board A12 Main board CPU board and A20 PCI Backplane board removal Figure 5 11 Hard disk drive removal Figure 5 12 Internal and external fan removal Figure 5 13 12 V main power supply and A40 AC Distributor board 1 Figure 5 14 RFI filter removal Figure 5 15 Troubleshooting procedure 1 Figure 5 16 Troubleshooting procedure 2 Figure 5 17 Troubleshooting procedure 3 Figure 5 18 Troubleshooting procedure 4 Figure 5 19 Troubleshooting procedure 5 Figure 5 20 Troubleshooting procedure 6 Figure 5 21 Tro
7. 200 mV div Input impedance 50 Q Horizontal scale 1 25 ns div Trigger position 50 Acquire mode Average 32 Trigger mode AUTO Trigger source CH1 Trigger level OV Trigger slope Rising Edge Input coupling DC Measure Amplitude Rise Time Fall Time Ref Level High Ref 80 Low Ref 20 Gating Cursor Curs1 1 2 5 ns Curs2 2 75 ns 3 Open the test40 TRP file on the RTX100B a Select File gt Open in the Play screen to open the Select File dialog box b In the dialog box select the test40 T RP file 4 Press the Play Pause button to start playing the test40 TRP file 5 Use the oscilloscope to verify that the amplitude rise and fall times are as follows Amplitude 720 mV to 880 mV Rise and fall time x 1 2 ns RTX100B ISDB T RF Signal Generator Service Manual 3 19 Performance Verification Checking the Play Operation 6 Disconnect the BNC cable from the 75 Q signal adapter on the oscilloscope and then connect the cable to the ASI SMPTE In connector on the MPEG analyzer See Figure 3 11 MPEG analyzer rear panel MTS430 BE Bg RTX100B rear panel noo 000000 o
8. The status bar displays the TS bit rate as 160 0000000 Mbps Eile View Play ISDB T ASI IP Utility mE G 13 Q e FM Ts test40 TRP Packets 386556 EX PAT PID OxO TbIID 0x0 TSID Ox1 Prg 4 E m PMT PID 0x15 TbIID Ox2 PrgNo Ox1 VIDEO PID 0x64 StrmTyp 0x2 Eq AUDIO PID 0x65 StrmTyp 0x4 KE PCR 0 64 PMT PID 0x16 0 2 PrgNo 0x2 VIDEO PID 0x6E StmTyp 0x2 5 AUDIO PID Ox6F StrmTyp 0x4 EE 5 PCR PID 0x6E m PMT PID 0x17 TbIID 0x2 PrgNo 0x3 VIDEO PID 0x78 StrmTyp 0x2 AUDIO PID 0x79 StrmTyp Ox4 E PMT PID 0x18 0 2 PrgNo Ox4 E VIDEO PID 0x82 StrmTyp 0x2 188 160 000000Mbps IP 166 565349Mbps Figure 3 27 RTX100B MPEG Player display with maximum bit rate 3 Perform the steps 5 through 15 listed in the Checking Option GbE perfor mance section on page 3 37 4 Check that the TS bit rate displayed in the status bar is approximately equal to 160 000000 Mbps as set in the RTX100B instrument RTX100B ISDB T RF Signal Generator Service Manual 3 47 Performance Verification 9 IP MPEG TS Compliance Analyzer File Navigation Settings Capturevu Record Help EE Restart Analysis en We ard d Preferences core eal en be Q am 5 stat m stop Program Tests SI PSI PID Packets Summary Now Playing IP Graphs MPE Sessions Ez Transport Stream Id 1 Network Name Program e
9. to the ASI SMPTE In connector on the MPEG analyzer See Figure 3 13 MPEG analyzer rear panel MTS430 P 75 Q BNC cable Figure 3 13 Equipment connection for checking the recorded file ASI interface 30 31 32 33 34 35 36 37 Press the Play Pause button on the RTX100B to display the Play screen Select Play gt Update gt On Open the ASI214Mbps file on the RTX100B a Select File gt Open to open the Select File dialog box b In the dialog box navigate to the D Record_Files directory and then select the ASDD14Mbps trp file Press the Play Pause button to start playing the file Start the TS Compliance Analyzer on the MPEG analyzer In the Open Transport Stream dialog box select Real time Analysis and then make the following settings Interfaces ASI Interface Settings Time Stamping Click the OK button Verify that the hierarchic view is displayed on the screen and that no error messages appear In addition verify that the bit rate is 40 Mbps and that the pac
10. Performance Check 1 XV Strategy for Servicing suas e ME p e UR RR ble dp ee XV Tektronix Service Offerings xvi Product Overview cc cece ccc c cece cece eee E 1 1 Specifications v e eR ener dels E e sides EE eR 1 3 Software Protection i ce aa eee etna cete dc e Re Red unge 1 3 Performance Conditions 0 0 cee cece enne 1 3 Functional Specifications 00 eee eect eee 1 4 Electrical Specifications s seipi ie eee ee ee ence eens 1 4 Mechanical Physical Characteristics 1 12 Environmental Characteristics 1 12 Certifications and Compliances 1 13 AT2 Main Boatd o oen e Lr e P e as 2 1 A20 PCI Backplane Board 0 eee cece eee ene 2 3 150 ISDB T RF Output Board 2 4 Front Panel Processor Board 2 5 Keypad Board SU pie te etree UR ceo N 13 2 5 A40 AC Distributor Board 2 5 Performance Verification Connecting to a Network 0 cee eee eee 3 1 Equipment Required 0 0 0 neient o ee eee ene 3 6 RTX100B Test Record 0 0 ee cee eee ees 3 7 Procedures 5 ecce eem rem eee em rete ere t m een 3 8 Adjustment Procedures Requirement for Adjustment 0 00 eee cece eee ee
11. Program 2 AQ Program 3 Program 4 ex Transport Stream Id 1 Network Name Number of Programs 4 Current TS bit rate 159 815 Mbps Current TS hit rate excluding Null PID 105 839 Mbps Name Number Current Bit Rate Current ka Min Mbps 1 0x1 HEN 26 431 20 13 26 409 26 887 2 0x2 1 26 776 20 39 26 792 26 886 9 4 0x4 peu 26 6858 20 309 26 636 26 887 a 31 1 1 53 976 33 729 52 199 72 776 3 0x3 Bt 7 161 5 45 7 052 26 886 TR 101 290 4 other 47 SFN Error 239 a Event Log Priority 1 Priority 2 1 Priority 3 3 Q 1 1 Sync 2 1 Transport 3 1 a NIT Actual 1 2 Syne Byte z22crc Q 3 1b NIT Other 1 3 4 PAT 2 3 PCR Repetition Q 3 2 SI Repetition 1 4 Continuity 2 3 b PCR Discontinuity Indicator 3 4 4 Unref PID 1 5 PMT Q 2 4 PCR Accuracy 3 5 4 5DT Actual 1 Q 1 6 PID 2 5 1 Q 3 5b SDT Other 2 6 CAT 3 6 5 EIT Actual 2 3 6 b EIT Other 3 6 c EIT PF 3 7 R5T TS Availability 9 Sync Qs rate 160 047616 Mbps 188 Packet Size 188 DVB Extensions Interface IP RTP UDP 8 7 start IP MPEG 5 Co gt untitled Paint Figure 3 28 MTS400 Series TS Compliance Analyzer display for maximum bit rate 3 48 5 Disc
12. 174 5172 00 174 5177 00 174 5178 00 174 5171 00 174 5170 00 196 3449 00 119 7186 00 174 B924 00 174 5190 00 174 5186 00 174 5348 00 174 5184 00 174 5174 00 119 7187 00 174 5188 00 174 5220 00 174 3767 01 174 5266 00 Serial no effective Serial no discont d Qty Name amp description CABLE ASSEMBLY 2X20 FLAT 430MM L ULTRA ATA CPU HDD HDD IDE CABLE ASSEMBLY 4 18AWG 60MML HDD POWER HDD A20 POWER CABLE ASSEMBLY 2 22AWG 50MM L W CONN 22 01 3027 MOLEX A20 A4 5V STAND BY CABLE ASSEMBLY 7 18AWG 50MM I VHR 10N amp VHR 7N XHP 2 JST P S A20 POWER amp CONTROL CABLE ASSEMBLY 6 18AWG 110MM L W CONN VHR 6N JST P S A20 GND CABLE ASSEMBLY 2 18AWG 130MM I W CONN VHR 5N JST P S A40 CABLE ASSEMBLY 2 18AWG 40MM I AC INLET to A40 INLET A40 LEAD ELECTRICAL 18AWG 4 0 1 5 4 SAFETY CONTROLLED CABLE ASSEMBLY SP ELEC FAN 119B05500 W CABLE 10CM L 40 SQ FAN CABLE ASSEMBLY SP ELEC 80 30AWG 5CM L FLAT W CONN YAMAICHI CABLE ASSEMBLY 28AWG FLAT 1 27CTR W CONN D SUB25 amp 2X13 DSUB25 CPU CABLE ASSEMBLY 2X10 28AWG FLAT 1MM CTR 130MM L RS232C SLOT PC TO A20 CPU A20 RS2 232C CABLE ASSEMBLY 2X5 28AWG FLAT 1MM CTR 120MM L PANNEL SLOT PC TO A20 CPU A20 PANEL CONTROL CABLE ASSEMBLY 14 30AWG 7TWISTED PAIRS 270MM L LVDS SLOT PC TO A20 CPU A20 LCD LVDS CABLE ASSEMBLY 12 18AWG 150MM L SLOT PC POWER CPU A20 CPU POWER CABLE ASSEMBLY SP ELEC FAN 119569400 W CABLE 18CM L 92 SQ FAN CABLE ASSEMBLY SP
13. 3 38 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification d Select Configuration in the IP menu to launch the IP Playout Configura tion dialog box see Figure 3 19 and make the following entries selec tions Device Name Intel R PRO 1000 MT Network Connection Transmission Mode Multicast Destination IP Address 239 1 1 1 Protocol UDP Destination Port 1234 IP Playout Configuration Configuration Name CONFIG 1 Protocols IP Version IPv4 Protocol UDP Network adapter Intel R PRO 1000 MT Network Conne Source IP Address 192 168 1 6 Device Name Source Port 116384 Transmission Mode Multicast Destination Port 11234 Destination IP Address 239 1 1 1 Enable VLAN Advanced Settings Stream Replication Packet Size Allow TS Fragmentation Allow IP Fragmentation IP Packet Size Bytes Number of TS packets IP Packet Size 1358 bytes Store Settings Delete Settings Reset To Defaults OK Cancel Figure 3 19 IP Playout Configuration dialog box Now the RTX100B MPEG Player should appear as shown in Figure 3 20 RTX100B ISDB T RF Signal Generator Service Manual 3 39 Performance Verification Eile View Play ISDB T ASI IP Utility mE 13 Q e EM Ts test40 TAP Packets 386556 Par PID 0x0 ThIID 0x0 TSID 0x1 Prg 4 m PMT PID 0x15 TbIID Ox2 PrgNo 0x1 HIE VIDEO PID 0x
14. 42 Set TS INPUT to SERIAL REAR on the decoder 43 Select Play gt Update gt Off on the RTX100B 44 Press the Play Pause button on the RTX100B to start playing the ASI214Mbps trp file 45 Verify that a moving picture is displayed on the video monitor In addition verify that there is no flicker or block noise in the displayed picture 46 Press the Stop button on the RTX100B to stop the stream output 47 Disconnect 75 Q BNC cables from the RTX100B MPEG 2 measurement decoder and video monitor RF Output This test verifies the carrier frequency carrier leakage and output level error of the RF output The following equipment is required for this test Equipment Frequency counter required Spectrum analyzer ISDB T analyzer 75 Q BNC cable 50 Q SMA cable 50 Q N Fe to 75 N Ma adapter 75 Q BNC Ma to NC Fe adapter Two 50 2 N Ma to SMA Fe adapters ISDB T Mt rmx ISDB T M2 rmx ISDB T M3 rmx files RTX100B ISDB T RF Signal Generator Service Manual 3 25 Performance Verification RTX100B rear panel Checking the Carrier Leakage 1 Use the 75 BNC Ma to NC Fe adapter 50 N Fe to 75 N Ma adapter 50 SMA cable and two 50 N Ma to SMA Fe adapters to connect the RF Out connector on the RTX100B to the RF Input on the spectrum analyzer See Figure 3 15 Spectrum analyzer E4402B COM 75 QBNC Ma to g NC Fe adapter TNE
15. Description 1 DCLK 2 GND to 10 DATA 7 to DATA 0 11 DVALID 12 PSYNC 13 Shield 14 DCLK 15 GND 16 to 23 DAIA 7 to DATA 0 24 DVALID 25 PSYNC 330 mV to 550 mV termination internal 100 external 100 bus LVDS with 50 Q termination 1 1Vto 1 5 V 100 between differential outputs output off 5 ns from the falling edge of DCLK see Figure 1 1 on page 1 11 gt 100 mV lt 100 mV RI RI with 100 Q termination 100 Q between differential inputs T 2 T 10 T 1 f f byte clock frequency see Figure 1 1 on page 1 11 T 2 T 10 T 1 f f byte clock frequency data are latched on DCLK rising edge see Figure 1 1 on page 1 11 EN 50083 9 Annex B BNC 75 Q 256 Kbps to 214 Mbps 1 800 mV 10 lt 0 2 Ul p p lt 1 2 ns 20 to 80 lt 17 dB 5 MHz to 270 MHz into 75 Q load RTX100B ISDB T RF Signal Generator Service Manual Specifications Table 1 2 Mainframe Cont Characteristics Input Number of inputs Input voltage typical Return loss Internal RF reference clock Frequency Stability External RF reference input Connector type Input impedance typical Reference input Frequency Input level typical RF output Connector type Output impedance typical Digital modulation Standard Mode Bandwidth Number of segments Number of layers Carrier modulation Inner coding Outer coding Time interval Guard interval Description 1 200 mV to 800 mV lt 17 dB 5 MHz t
16. 10 Torx tip to remove the four screws securing the fan and fan cover to the chassis RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures 6 Reinstallation Perform step 5 in reverse order to reinstall the external fan Internal fan p P External fan V LEAT TIF LT TTT NT iad T Ys v NA Ne ey LD E Figure 5 12 Internal and external fan removal RTX100B ISDB T RF Signal Generator Service Manual 5 29 Removal and Installation Procedures 12 V Main Power Supply A40 AC Distributor Board 5 30 Assemble equipment and locate modules to be removed a You need a screwdriver with a T 10 Torx tip b Locate the 12 V main power supply in the locator diagram Internal modules Figure 5 4 page 5 13 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its left side is facing you Remove the 12 V main power supply See Figure 5 13 on page 5 31 a Disconnect the cables from CN1 CN2 CN3 and CN4 b Use a screwdriver with a T 10 Torx tip to remove the four screws securing the 12 V main power supply to the chassis c Lift the power supply away from the chassis Reinstallation Perform step 3 in reverse order to reinstall the 12 V main power supply Assemble equipment and locate modules to be removed a You need a screwdriver with a T
17. IP Address Settings 3 Passive mode no outbound traffic Active mode issues and responds ka ARP and other traffic IP Address 0 0 0 0 k 0 0 0 0 ateway 0 0 0 0 IGMP Group Membership Current Group None 1 UDP flaw detected total rate 27 403 Mbps Destination Source Destination Tra Has Source Dest Bit rate Bit rate Multi MAC Address IP Address IP Address eee sore Pork Kbps Graph cast tri 5E 04 50404 192 1681 6 239 1 1 4 v 6384 1234 42253 8367 Selected Flow Source Address 192 168 1 6 Destination Address 239 1 1 1 1234 UDP Figure 3 24 IP Flow selection c Click the OK button The MTS400 Series TS Compliance Analyzer dialog box opens as shown in Figure 3 25 3 44 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification 12 Allow the DUT RTX100B MPEG system to loop at least three times 13 On the MTS400 Series MPEG system verify that the TS Availability and Sync indicators in the bottom left corner of the Compliance Analyzer display remain green 14 Check that all the parameters under Priority 1 of the Compliance Analyzer display are green The TS bit rate displayed in the status bar should be approximately equal to 40 00000 Mbps which was set in the RTX100B instrument NOTE Ignore the 2 5 PTS errors as they are generated by the test40 trp file 9 IP MPEG T5 Compliance Analyzer File Navigation Settings Captureyu Record Help w Restart Analysis ee
18. J230 J310 J290 J150 174 5178 00 174 5177 00 CN2 5 V Main Power Supply CN1 174 5171 00 J100 e LY Clock Ref In C Trig In Out v9 J702 SPI In Out J500 ASI Out O A150 ISDB T RF OUTPUT Board 174 B924 00 A12 Main Board Hard Disk Drive Card Edge Card Edge 174 5181 00 A20 PCI Backplane Board 174 5170 00 5 V Sub Power Supply J110 A40 AC Distributor Board 196 3499 00 J180 119 7186 00 119 7187 00 174 5220 00 174 5266 00 J321 Front Panel Processor Board Inverter Board for LCD Backlight Keypad Board 174 5188 00 RTX100B ISDB T RF Signal Generator Service Manual Figure 7 2 RTX100B interconnect diagram 7 5 RTX100B ISDB T RF Signal Generator Service Manual ere S Replaceable Parts List a a Replaceable Parts List This section contains a list of the replaceable modules for the RTX100B Use this list to identify and order replacement parts Parts Ordering Information Part Number Revision Level Replacement parts are available through your local Tektronix field office or representative Changes to Tektronix products are sometimes made to accommodate improved components as they become available and to give you the benefit of the latest improvements Therefore when ordering parts it is important to include the following information in your order Part number see Part Number Rev
19. 21 dBm to 29 dBm at 13 ch Mode 2 18 dBm to 26 dBm at 13 ch Mode 3 15 dBm to 23 dBm at 13 ch Bit error rate typical lt 2 0 E4 after Vitervi 1 8 RTX100B ISDB T RF Signal Generator Service Manual Specifications Table 1 2 Mainframe Cont Characteristics VGA output Connector type Pin assignments Printer port Connector type Pin assignments USB 2 0 interface Number of connectors Pin assignments LAN interface Connector type Pin assignments RTX100B ISDB T RF Signal Generator Service Manual Description D sub 15 pin 1 NOD CO gt 8 RED GREEN BLUE NC GND GND GND GND NC GND NC NC HSYNC VSYNC NC Supports SPP Standard Parallel Port EPP Enhanced Parallel Port and ECP Extended Capabilities Port modes D sub 25 pin gt STROBE DO D1 BUSY Paper Empty SELECT VCC DATA DATA GND AUTOLF ERR INIT SELECT GND GND GND GND GND GND GND GND 10 100 1000 Base T Ethernet interface RJ45 1 2 3 4 MDI_0 MDI 0 MDI 14 2 MDI 2 MDI 1 MDI 3 MDI 3 Specifications Table 1 2 Mainframe Cont Characteristics Description Serial interface Com RS 232C Connector type D sub 9 pin Pin assignments 1 DCD 6 DSR 2 RX 7 RTS 3 TX 8 CTS 4 DTR 9 RI 5 GND IEEE1394B interface Compliant IEEE1394B 2002 Pin assignments 1 TPB 6 VG 2 7 NC 3 8 VP 4 TPA 9 TPB R 5
20. 23 REG_SZ 2048 REG S2 2048 REG 2 3 REG Sz Intel REG S2 ab ReduceSpeedOnPowerDown REG SZ ab SawvePrwerninwFnahled RFG 97 lt My Computer HKEY_LOCAL_MACHINE SYSTEM Controlset 00 1 Control Class 4D96972 925 1 1CE BFC1 08002bE10318 0001 E acr amp Arbiters 22 BackupRestore C Biosinfo BoatVerificationProgram 3 Select MonitorModeEnabled DWORD Value that you created 4 Double click the DWORD Value to open the Edit DWORD Value dialog box Check if the Value data is set to 1 3 36 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Edit DWORD Value Value name ManitortadeEnabled Value data Base Hexadecimal Decimal OK Cancel Checking Option GbE performance Perform this procedure to check that the GbE Gigabit Ethernet interface of RTX100B is operating properly This test checks that the transport stream data is correctly output from the RTX100B instrument 1 Connect the equipment as shown in Figure 3 18 Ethernet Switch DUT RTX100B System Netgear JGS524F IP Player Version 10 1 GbE 10 100 1000 MTS400 Series System ucc RJ 45 Copper port 2 GbE Optic port 3 E RJ 45 Copper port 4 Figure 3 18 Equipment setup for Option GbE performance verification NOTE Check that V10 1 or greater RTX100B software is installed Equipment MTS400 Series MPEG system required
21. Click the Update Now button below to check For updates An icon will appear in the system tray if an update is available Move the cursor aver the icon to see the status of the update Java Update was last run at 3 20 PM on 7 13 07 Update Now RTX100B ISDB T RF Signal Generator Service Manual 3 35 Performance Verification Perform this procedure to check for the settings of VLAN 1 Click Start gt Run and type regedit Press OK to display the Registry Editor Window 2 Browse to HKEY LOCAL _MACHINE SYSTEM ControlSet001 Con trol Class 4D36E972 E325 11CE BFC1 08002BE10318 00XX where XX is the instance of the network adapter that you need to see tags on You can check by opening and viewing the driver descriptor value DriverDesc value should be Intel R PRO 1000 MT Network Connection 2 Registry Editor File Edit View Favorites Help 5 My Computer Name Type Data Ga dsl CLASSES 2B Defaut REG S value not set 21 HKEY CURRENT LISER AdaptivelFS 6 E HKEY LOCAL MACHINE Ea auen volod Gapa m amp HARDWARE JAutoPowerSaveModeEnabled REG SZ a aE sam SECURITY iusNumber REG SZ 1 8 C SOFTWARE BusType REG SZ 5 t SYSTEM Characteristics REG DW ORD 0x00000084 132 EJ ControlSetoot ChecksumRxIp REG Sz 1 amp Gal Control ChecksumRxTcp REG S 1 ChecksumTxIp REG_5Z 1 ChecksumTxTcp REG S2 1 CoInstallers32 REG MULTI 52 IntelNic dli MicCoTnstallerEntry
22. RJ 45 cables Crossover RJ 45 cable test40 TRP file 2 Power on the MTS400 Series MPEG system and allow the Windows operating system to boot completely Wait an additional 90 seconds for the MPEG system to complete the Option GbE firmware loading process RTX100B ISDB T RF Signal Generator Service Manual 3 37 Performance Verification 3 Configure the DUT RTX100B system with the following settings for GbE testing a Double click the RTX100B MPEG player icon on the desktop to start the player MpegPlayer MPEGPlayer Ink b From the File menu select Open gt Look in dropdown menu gt Local Disk D gt test streams gt test40 TRP gt Open c Make the following selections in the Play menu m Play Interface gt IP m Play Packet Size gt 188 Play Clock dialog box Wm Make sure the Internal button is selected W Make sure the Data Rate section shows 40 27 27 0 40 000000 Play gt Update gt On checked m Play gt PCR Initial Value dialog box m Base Value 33 0 m Extension Value 9 0 m Click the OK button m Play Source Disk checked m Play Loop On checked Start gt Stop dialog box Start radio button selected 0 text entry box set to 0 100 text entry box set to 386555 Format set to Packets m Play gt Auto Play gt Off checked m Play gt Timer Play ignore this it should show the current date and time
23. Solder connections Cold solder or rosin joints Damaged or leaking cases Corroded solder on leads or terminals Loosely inserted in sockets Distorted pins Loose plugs or connectors Burned broken or frayed wiring Dents deformations and dam aged hardware RTX100B ISDB T RF Signal Generator Service Manual Remove failed module and replace with a fresh module Remove failed module and replace with a fresh module Resolder joint and clean with isopropyl alcohol Remove damaged module and replace with a fresh module from the factory Firmly seat loose semiconduc tors Remove devices that have distorted pins Carefully straight en pins as required to fit the socket using long nose pliers and reinsert firmly Ensure that straightening action does not crack pins causing them to break off Firmly seat connectors Repair or replace modules with defective wires or cables Straighten repair or replace defective hardware Maintenance 5 6 Cleaning Procedure Interior To clean the RTX100B interior perform the following steps 1 2 Blow off dust with dry low pressure deionized air approximately 9 psi Remove any remaining dust with a lint free cloth dampened in ethyl alcohol and rinse with warm deionized water A cotton tipped applicator is useful for cleaning in narrow spaces and on circuit boards STOP If after performing steps 1 and 2 a module is clean upon inspecti
24. Status Connected Duration 4 days 20 41 58 Speed 1 0 Gbps Activity Received Cds Packets 1121816 314 870 Figure 3 3 Local Area Connection Status dialog box 6 Click the Properties button The Local Area Connection Properties dialog box appears Local Area Connection Properties General Authentication Advanced _ Connect using Intel R 8255ER PCI Adapter This connection uses the following items MS NWLink NetBIOS NWLink IPX SPX NetBlOS Compatible Transport Prot E Internet Protocol TCP IP 2 Network component field e Description Transmission Control Protecol Intemet Protocol The default wide area network protocol that provides communication across diverse interconnected networks Show icon in notification area when connected me when this connection has limited no connectivity Figure 3 4 Local Area Connection Properties dialog box 7 Inthe network component field select Internet Protocol TCP IP 8 Click the Properties button The Internet Protocol TCP IP Properties dialog box appears RTX100B ISDB T RF Signal Generator Service Manual 3 3 Performance Verification 3 4 Internet Protocol TCP IP Properties General Altemate Configuration fou can get IP settings assigned automatically if your network supports this capability Otherwise pou need to ask your
25. the SPI214Mbps trp file Select Play gt Update gt On on the RTX100B Press the Play Pause button on the RTX10OB to start playing the SPI214Mbps trp file RTX100B ISDB T RF Signal Generator Service Manual 3 13 Performance Verification 32 Verify that the hierarchic view is displayed on the screen and that no error messages appear In addition verify that the bit rate is 40 Mbps and that the transport stream packet size is 188 bytes NOTE Ignore Program 3 and PID 120 0x78 errors in the hierarchical view because these are caused by the original test40 TRP file 33 Press the Stop button on the RTX100B 34 Close the TS Compliance Analyzer window External Clock Reference This test confirms that the external clock reference input Clock Ref In and and External Trigger external trigger input Trig In Out on the RTX100B are functioning correctly Inputs Equipment required MPEG analyzer Function generator Two 50 Q BNC cables Parallel interface cable test64 TRP file 1 Use 50 BNC cable to connect the Clock Ref In connector on the RTX100B to the Ch1 connector on the function generator See Figure 3 9 2 Use the 50 Q BNC cable to connect the Trig In Out connector on the RTX100B to the Ch2 connector on the function generator See Figure 3 9 3 Use the parallel interface cable to connect the SPI In Out connector on the RTX100B to the DVB SPI In connector on the MPEG analyzer See Figure 3 9 3
26. 0 0 ccc ee cece eee 8 1 Using the Replaceable Parts List 8 3 ii RTX100B ISDB T RF Signal Generator Service Manual Table of Contents List of Figures Figure 1 1 Timing diagram of the SPI interface 1 11 Figure 3 1 Pin connections for a crossover Ethernet cable 3 1 Figure 3 2 Network Connections window 3 2 Figure 3 3 Local Area Connection Status dialog box 3 3 Figure 3 4 Local Area Connection Properties dialog box 3 3 Figure 3 5 Internet Protocol TCP IP Properties dialog box 3 4 Figure 3 6 Equipment connection for checking the internal clock output Rene er rm ne ever 3 9 Figure 3 7 Equipment connection for checking the internal clock frequency cuc tron Vere ep ex eH ore wa eee recae 3 10 Figure 3 8 Equipment connection for checking the play operation SPI interface 3 11 Figure 3 9 Equipment connections for checking the external clock reference and trigger inputs 3 15 Figure 3 10 Equipment connection for checking the output signal ASI interface 3 18 Figure 3 11 Equipment connections for checking the play operation ASI interface 3 20 Figure 3 12 Equipment connection for checking the record operation ASI interface 3 21 Figure 3 13 Equipmen
27. 0 090 THICK 0 050 THICK GLASS 0 040 BLACK PORON W 0 002 THK ADHESIVE MTX100B SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX DISPLAY MODULE LCD COLOR 1024X768 XGA 6 3 DIAG TFT 8 6 BIT LVDS 250 NIT BACK LIGHT 0 125 MM PIXEL PITCH 21MS FRONT MOUNT NL10276BC 02 SAFETY CONTROLLED BRACKET DISPLAY FRONT PANEL AL MTX100B CA ASSY RIBBBON ADAPTER TO XGA DISPLAY 20 PIN 3 5IN CABLE ASSY SP DISCRETE BACKLIGHT IDC 26 AWG 5 0 L PCB 1 X 6 0 079 CTR 2MM SHRINK WRAPPED POWER SUPPLY INVERTOR BOARD REPLACEMENT FOR LCD 119 5659 00 6 5 LCD SCREW MACHINE 4 40 X 0 187 PAN HEAD STL CD PL T 10 TORX DR LABEL MANUFACTURED PRODUCT ID 2 5 IN X 1 5 IN SSAFETY CONTROLLED SCREW MACHINE 4 40 X 0 187 PAN HEAD STL CD PL T 10 TORX DR CIRCUIT BD ASSY UNTESTED 389388900 WIRED FRONT PANEL PROCESSOR MTX100B SHIELD ELEC KEYPAD BOARD 0 020 THK POLYCARBONATE LEXAN FR60 MTX100B SAFETY CONTROLLED CIRCUIT BD ASSY UNTESTED 389388800 WIRED KEYBOARD L SHAPE MTX100B SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX SWITCH KEYPAD ELASTOMERIC FRONT PANEL MTX100B SWITCH ELASTOMERIC POWER AND PLAY BUTTONS FRONT PANEL MTX100B LABEL FRONT PANEL CONTROL MTX100B RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Figure 8 4 Front panel unit RTX100B ISDB T RF Signal Generator Service Manual 8 11 Replaceable Parts List Replaceable parts list Fig amp index number 8 5 1 8 12 Tektronix part number 174 5183 00 174 5181 00
28. 14 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Function generator AFG3102 0000000 m MPEG analyzer MTS430 50 Q BNC cable 50 Q BNC cable 9 E SOO ve oH A Te b M EG COH RR Pepe OO RTX100B rear panel Parallel interface cable Figure 3 9 Equipment connections for checking the external clock reference and trigger inputs 4 Set the function generator as indicated below Ch 1 output Function Run Mode Frequency Output Amplitude Output Offset Ch 2 output Function Run Mode Output Menu Output Amplitude Output Offset Square Continuous 10 MHz 0 5 V 2 0V Arb Edit Point
29. 15 size screw heads 4 Standard tool Item Description RTX100B ISDB T RF Signal Generator Service Manual 5 9 Removal and Installation Procedures RTX100B Orientation In this manual procedures refer to front rear and top of the RTX100B Figure 5 1 shows how the sides are referenced Figures 5 2 through 5 4 show the modules that make up the RTX100B Figure 5 1 RTX100B orientation 5 10 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Left side cover Right side cover Cabinet UDR WOOO WDE b Q T LCD inverter s v board f Emat X i E Fron panel Play Stop Record S e Processor board eee Pate rn Front panel bracket insulator LCD module Keypad board LCD glass shield Emat all others Front cover Front panel Figure 5 2 External modules RTX100B ISDB T RF Signal Generator Service Manual 5 11 Removal and Installation Procedures Printer connector Blank panel A150 ISDB T RF Output board amp 12 Main board py CPU board Figure 5 3 Board modules 5 12 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures RFI filter 12 V main power supply sec External fan Hard disk drive retainer A40 AC Distributor board Hard disk drive Internal fan Figure 5
30. 250V 10A 98 0 L STR IEC320 RCPT X NEMA 5 15P US SAFETY CONTROLLED CABLE ASSY PWR 3 0 75MM SQ 250V 10A 99 0 L STR IEC320 RCPT EUROPEAN SAFETY CONTROLLED CABLE ASSY PWR 3 1 0MM SQ 250V 10A 2 5 METER STR IEC320 RCPT X 13A FUSED UK PLUG 13A FUSE UNITED KINGDOM SAFETY CONTROLLED CABLE ASSY PWR 3 1 0MM SQ 250V 10A 2 5 METER STR IEC320 RCPT AUSTRALIA SAFETY CONTROLLED CABLE ASSY PWR 3 18 AWG 250V 10A 98 0 L STR RCPT X NEMA 6 15P US SAFETY CONTROLLED CABLE ASSY PWR 3 1 0MM SQ 250V 10A 2 5 METER STR IEC320 RCPT SWISS SAFETY CONTROLLED CABLE ASSY PWR 3 125V 7A JAPAN 98 LONG STR NEMA 5 15P PLUG X IEC320 C 13 RECEPTACLE SAFETY CONTROLLED CABLE ASSY PWR 3 1 0MM SQ 250V 10A 2 5 METER STR IEC320 3C CERTIFICATION RCPT CHINA SAFETY CONTROLLED OPTIONAL ACCESSORIES MANUAL TECH SERVICE RTX100B WFM7F05 OPTION NN RACKMOUNT KIT 1700F06 BLANK PANEL TEquipment An Interworld Highway LLC Company RTX100B ISDB T RF Signal Generator Service Manual
31. 61010 1 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Terminals on this product may have different installation overvoltage category designations The installation categories are CAT III Distribution level mains usually permanently connected Equipment at this level is typically in a fixed industrial location CAT II Local level mains wall sockets Equipment at this level includes appliances portable tools and similar products Equipment is usually cord connected CAT I Secondary signal level or battery operated circuits of electronic equipment Overvoltage Category II as defined in IEC 61010 1 A measure of the contaminates that could occur in the environment around and within a product Typically the internal environment inside a product is considered to be the same as the external Products should be used only in the environment for which they are rated Pollution Degree 1 No pollution or only dry nonconductive pollution occurs Products in this category are generally encapsulated hermetically sealed or located in clean rooms Pollution Degree 2 Normally only dry nonconductive pollution occurs Occasionally a temporary conductivity that is caused by condensation must be expected This location is a typical office home environment Temporary condensation occurs only when the product is out of service Pollution Degree 2 Conductive pollution or dry nonconductive pollution
32. 7 Avoid handling modules in areas that have a floor or work surface covering capable of generating a static charge RTX100B ISDB T RF Signal Generator Service Manual Maintenance Inspection and Cleaning General Care Inspection and Cleaning Procedures A Inspection and Cleaning describes how to inspect for dirt and damage It also describes how to clean the exterior and interior of the RTX100B Inspection and cleaning are done as preventive maintenance Preventive maintenance when done regularly may prevent the RTX100B from malfunctioning and enhance its reliability Preventive maintenance consists of visually inspecting and cleaning the RTX100B and using general care when operating it How often preventative maintenance should be performed depends on the severity of the environment in which the RTX100B is used A proper time to perform preventive maintenance is just before the RTX100B adjustment The cabinet helps keep dust out of the RTX100B and is needed to meet EMI and cooling requirements The cabinet should be in place when operating the RTX100B The front cover protects the front panel and display from dust and damage Install the front cover when storing or transporting the RTX100B Inspect and clean the RTX100B as often as operating conditions require The collection of dirt on components inside can cause them to overheat and break down Dirt acts like an insulating blanket preventing efficient heat dissipation
33. Dirt also provides an electrical conduction path that could cause an RTX100B failure especially under high humidity conditions CAUTION Avoid the use of chemical cleaning agents that might damage the plastics used in this RTX100B Use only deionized water when cleaning the menu buttons or front panel buttons Use an ethyl alcohol solution as a cleaner and rinse with deionized water Inspection Exterior Inspect the outside of the RTX100B for damage wear and missing parts using Table 5 1 as a guide An RTX100B that appears to have been dropped or otherwise abused should be checked thoroughly to verify correct operation and performance Immediately repair defects that could cause personal injury or lead to further damage to the RTX100B RTX100B ISDB T RF Signal Generator Service Manual 5 3 Maintenance Table 5 1 External inspection check list Item Cabinet front panel and cover Front panel buttons Connectors Accessories Cracks scratches deformations Repair or replace defective damaged hardware or gaskets module Missing damaged or loose Repair or replace missing or buttons defective buttons Broken shells cracked insulation Repair or replace defective and deformed contacts Dirt in modules Clear or wash out dirt connectors Missing items or parts of items bent pins broken or frayed cables and damaged connec tors Repair or replace damaged or missing items frayed cables and
34. ELEC 10 60CM L 26AWG AND 28 AWG CPU A35 USB CABLE ASSEMBLY 2X15 28AWG FLAT 1MM CTR 300MM L A20 TO A30 CABLE ASSEMBLY SP DISCRETE BACKLIGHT IDC 26 AWG 5 0 L PCB 1 X 6 0 079 CTR 2MM SHRINK WRAPPED CABLE ASSEMBLY RIBBON ADAPTER TO XGA DISPLAY 10 PIN 3 5IN RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Power supply Hard disk drive Figure 8 5 Cables 8 13 RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Replaceable parts list Compo nent Tektronix number part number 071 1930 XX 071 1931 XX 071 1507 05 063 3932 00 020 2941 00 012 A220 00 119 6936 00 119 B146 00 200 4716 00 161 0066 00 161 0066 09 161 0066 10 161 0066 13 161 0066 12 161 0154 00 161 0298 00 161 0304 00 071 1932 XX 8 14 Serial no effective Serial no discont d Qty Name amp description n STANDARD ACCESSORIES MANUAL TECH USER RTX100B ENGLISH OPTION 10 MANUAL TECH USER RTX100B JAPANESE OPTION 15 MANUAL TECH USER TWO VOLUME MTS400 ENGLISH MTX100B RTX100B RTX130B SAMPLE STREAMS CD B019999 AND BELOW MPEG TEST STREAMS DVD PACKAGING KIT B020000 AND ABOVE CABLE INTCON D SUB 25 MALE TO MALE STR TWIST 2M L SCREW 4 40 POINTER ASSY OPTICAL MOUSE USB WHITE OPTICAL THREE BUTTON WHEELED W USB to PS2 ADAPTER ABS SAFETY CONTROLLED KEYBOARD USB MTX100 SAFETY CONTROLLED COVER FRONT PROTECTIVE PC ABS FR110 W TAPE TV GRAY CABLE ASSY PWR 3 18 AWG
35. EtCoTnst c ColnstallFlag REG_DWORD 0 80000001 2147483649 Componentid REG_SZ pcihven _80868dev _100e CI CustomMessages REG 57 1 8 250BCE51 6C8F 4A72 8A6D B54C2B4FC835 DmaFeirness REG 52 0 8 364212FF 208 4520 8282 897C09879863 DriverDate 52 11 22 2004 B C468 1 1C 2058444653510000 DriverDateData REG BINARY 00 80 9 35 26 d c4 01 2 4658EE7E FOSO 11DL B6ED D0CD4FA372A7 F 48721856 6795 11D2 B1A8 0080C72E74A2 pla HT Network Connecti 21 49CE6ACS 6F86 1 1D2 B1E5 0080C72E74A2 ae 1 14036 965 325 11 08002 10318 amp jeneblere d 8 4D36E966 1CE BECI 080028E10318 grrsttme REG 52 22 4036E967 E325 11CE BFCI 08002BE10318 P FlowControl REGIE 4036 968 25 11 1 08002 10318 abjinfeath REG_SZ oems inF 0 2 4036 969 325 11 08002 10318 B infSection REG_S2 ELOOE ndi NTx86 5 1 E 40368964 325 11CE BFC1 080028E 10318 aji REG SZ n 4D36E96B E325 11CE BFC1 08002BE10318 ILogLinkStateE vent REG SZ 51 E 2 40 6 96 25 11 080028 10318 ab MatchingDeviceld REG 37 pciven 8066 amp dev 1006 1 08002 10318 1 REG SZ 16128 1 08002 10318 Eam abled REG_DWORD 0x00000001 1 1CE BFC 1 08002BE 10318 ab MulticastPiterType REG SZ o 1CE BFC1 08002BE10318 IMwIEnable REG SZ 0 1CE BFC1 08002BE10318 NetCFginstanceld REG SZ 49941 275 9018 40 9 09 94544
36. External Fans 5 28 Doing the Performance Verification and Running Sysprep If you replace the hard disk drive you must perform the following steps NOTE On B020000 and above the Sysprep is already activated for 650 5181 00 1 Perform the performance verification procedures refer to Performance Verification on page 3 1 2 Run the sysprep batch file a Select File gt Exit to exit the RTX100B application The Windows XP desktop appears b Select Start Run to open the Run dialog box c Click the Browse button d Select C Sysprep sysprep bat e Click the OK button 1 Assemble equipment and locate modules to be removed a You need a screwdriver with a T 10 Torx tip b Locate the fans in the locator diagram Internal modules Figure 5 4 page 5 13 2 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its front is facing you 3 Remove the internal fan See Figure 5 12 on page 5 29 a Disconnect the cable from J180 on the A20 PCI Backplane board b Remove the two rivets securing the fan to the chassis c Liftthe fan up out of the chassis 4 Reinstallation Perform step 3 in reverse order to reinstall the internal fan 5 Remove the external fan See Figure 5 12 on page 5 29 a Set the RTX100B so its bottom is down on the work surface and its rear is facing you b Disconnect the cable from J190 on the A20 PCI Backplane board Use a screwdriver with a
37. ISDB T RF Signal Generator Service Manual Performance Verification Open Transport Stream Stream Interpretation Base Standard DVB Region No Extensions Using default scripts Scripted validations are not enabled File Analysis Interfaces Select From the available interfaces Interface Settings Real time Analysis Configurations Selected Configuration These are the settings for the current selected configuration To browse the network for stream select the Browse button Field Value to set ce a a ee ee C Tektronix LP1500 LAN InterFace 31 port 4 Tektronix LP1500 LAN Interface 31 port 4 true 0 0 0 0 0 0 0 0 0 0 0 0 www 192 168 1 99 239 1 1 1 1234 9 Settings which have been changed have a purple background Figure 3 23 Interface settings 9 Click the Browse button to open the Browse for IP Flow dialog box as shown in Figure 3 24 RTX100B ISDB T RF Signal Generator Service Manual 3 43 Performance Verification 10 Select Tektronix LP1500 LAN Interface Port 4 from the Browse Device drop down menu 11 In the Browse for UDP flow dialog box make these selections see Figure 3 24 a Select the Passive mode radio button b In the Detected UDP Flows pane select the IP stream with a Destination IP Address of 239 1 1 1 9 Browse for UDP flow Browse Device Tektronix LP1500 LAN Interface 31 port 4 v Tektronix LP1500 LAN Interface 31 pork 4
38. Qs a Preferences pontus ea e Besoni bps arm gt Start a Stop Program o Transport Stream Id Program 1 e Program 2 amp x E Program 22 HQ Program 4 e lt gt f Summary Now Playing IP Graphs MPE Sessions Transport Stream Id 1 Network Name 000 Number of Programs 4 Current TS bit rate 40 006 Mbps Current TS bit rate excluding Null PID 26 932 Mbps Name Number Current Bit Rate Current MI Min Mbps 1 0x1 je 16 83 2 0 2 6 714 16 83 6 648 6 717 v 0x2 3 0x3 J 6 714 16 83 0 000 6 717 0x3 4 0x4 6 714 16 83 0 000 6 717 0 4 NULL PID 8191 ee 13 070 32 68 13 064 20 354 TR 101 290 6 Other 55 Q SFN Error 275 A Event Log Priority 1 Priority 2 2 ia Sync 2 1 Transport 3 1 4 NIT Actual 1 1 2 Sync Byte 22crc Q 3 1b NIT Other 1 2 3 a PCR Repetition 3 2 51 Repetition 1 4 Continuity 2 3 b PCR Discontinuity Indicator 3 4 8 Unref PID 1 5 Q 2 4 PCR Accuracy 3 5 4 507 Actual 1 1 6PID 2 5PT5 2 5 SDT Other 2 3 6 5 EIT Actual 2 3 6 b EIT Other 3 6 c EIT PF 3 7RsT 3 8 TDT TS Availability 9o Sync Bit rate 39 998911 Mbps 188 Packet Size 188 ov No Extensio
39. Restart the RTX100B Are any images displayed pes Replace the CPU board monitor m Check the 174 5184 00 174 3767 01 and 174 5220 00 cables Does the backlight Replace the Inverter board light m Replace the Front Panel Processor board m Check the 259 A002 00 cable m Replace the LCD module Figure 5 20 Troubleshooting procedure 6 RTX100B ISDB T RF Signal Generator Service Manual 5 41 Troubleshooting No Have the LEDs on the m Check the 174 5186 00 and 174 5220 00 cables m Replace the Front Panel Processor board m Reinstall the RTX100B application front panel gone off Reinstall the Windows XP operating system and the RTX100B application Is the SPI Output Enable check box in DO m Verify the BIOS settings for Com2 m Replace the CPU board m Replace the A20 PCI Backplane board the Others dialog box selected Yes Select the SPI Output Enable check box Can a stream data be output from the SPI No Replace the A12 Main board Figure 5 21 Troubleshooting procedure 7 5 42 In Out connector RTX100B ISDB T RF Signal Generator Service Manual Troubleshooting No Is the Record Source menu set to SPI Yes Select Record gt Source gt SPI Can a stream data be acquired from the No Replace the A12 Main board Figure 5 22 Troubleshooting proced
40. Stack Electronics BA057 Novac NV HD352WB and hard disk drive Tektronix part number 119 7146 00 1 ea RTX100B ISDB T RF Signal Generator Service Manual Performance Verification RTX100B Test Record Photocopy this page and use it to record the performance test result Table 3 2 RTX100B test record Serial Number Cal Date Temperature Humidity Internal Clock Frequency 26 999924 MHz 27 000076 MHz 1 Test record limits are based on the SMPTE 310M frequency accuracy requirement RTX100B ISDB T RF Signal Generator Service Manual 3 7 Performance Verification Procedures The following conventions are used in the performance verification procedures Each test lists the characteristic that is being tested and the equipment required to perform the test Each test contains complete setup instructions which allows you to perform each test individually or in order m The equipment connection illustrations are specific to the recommended equipment equipment nomenclature is labeled If you are using substitute equipment the location of your signal connections may vary from those in the illustrations NOTE Before you begin the performance verification procedures be sure that the RTX100B is operating in an environment that is within the operating limits described in Table 1 4 on page 1 12 In addition the RTX100B and the test equipment must be warmed up for at least 20 minutes to ensure accurate test resul
41. When the Window OS puts the instrument into power saver mode the LVDS data from the SBC stops and the LCD backlight needs to be turned off or a rainbow display will occur When the LVDS data stops the common mode voltage of the data lines changes from the normal 1 25 V to 0 V The compara tor U10 senses this voltage change and turns off the power MOSFET switch Q10 and the power to the backlight inverter board The potentiometer R301 could be used to dim the LCD backlight brightness But R302 is not installed so the adjustment has no effect and the backlight is at maximum brightness There are two USB connectors on the board that are connected through a cable to the USB 2 0 interface connector on the CPU board The Keypad board contains the switch patterns for the elastomer keys and the front panel LEDs The board is connected to the Front Panel Processor board by the board to board connectors J10 and J11 A40 AC Distributor Board The A40 AC Distributor board supplies standby power SVSB to the SBC and the ATX power control circuit on the A20 Backplane board RTX100B ISDB T RF Signal Generator Service Manual 2 5 Theory of Operation 2 6 RTX100B ISDB T RF Signal Generator Service Manual eee Performance Verification XA Performance Verification This section provides procedures to verify the performance and functionality of the RTX100B network interface Connecting to a Network The RTX100B has a LAN 10 100 100
42. impedance 50 Q Horizontal scale 10 ns div Trigger position 50 Acquire mode Average 16 Trigger mode Auto Trigger level 1 20 V Trigger source CH1 Trigger slope Rising Edge Input coupling DC Measure High Level Low level 6 Verify that the measured values are as follows High Level gt 22 Low Level lt 0 8 V 7 Disconnect the 50 BNC cable from the oscilloscope CH1 input and then connect the BNC cable to the CH1 connector on the frequency counter See Figure 3 7 RTX100B rear panel Frequency counter 53181A Option 015 ooo 50 Q BNC cable Figure 3 7 Equipment connection for checking the internal clock frequency 8 Set the frequency counter as indicated below MEASURE Frequency 1 Gate Time 0 20 s CHANNEL1 Coupling DC Impedance 50 Trigger AUTO TRIG ON 9 Record the frequency counter reading in the test record See page 3 7 3 10 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification 10 In the Ext Trigger BNC dialog box change BNC IN OUT to Input 11 Disconnect the B
43. m The cable from the internal fan at J180 m The cables from the Front Panel Processor board at J240 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures The cable from the hard disk drive e Use a screwdriver with a T 10 Torx tip to remove the ten screws securing the A20 PCI Backplane board to the chassis f Liftthe board up and away from the chassis 4 Reinstallation Perform step 3 in reverse order to reinstall the A20 PCI Backplane board Printer Connector 1 Assemble equipment and locate modules to be removed a You need a screwdriver with a 1 4 inch nut driver b Locate the connector in the locator diagram Board modules Figure 5 3 page 5 12 2 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its rear is facing you 3 Remove the Printer connector See Figure 5 10 on page 5 22 a Remove the parallel dongle b Remove the A150 ISDB T RF Output board as described on page 5 23 c Remove the A12 Main board as described on page 5 23 d Disconnect the cable from CN3 on the CPU board e Use a 1 4 inch nut driver to unscrew the two hex headed mounting posts securing the connector to the chassis f Pull the connector away from the chassis 4 Reinstallation Torque the dongle screws to 5 inch Ibs Perform step 3 in reverse order to reinstall the Printer connector RTX100B ISDB T RF Signal Generator Service Manual 5 25 Removal and Ins
44. most Tektronix service locations The Tektronix product catalog lists all worldwide service locations or you can visit our Web site for service information www tektronix com Tektronix supports repair to the module level by providing Module Exchange Module Exchange This service reduces down time for repair by allowing you to exchange most modules for remanufactured ones Each module comes with a 90 day service warranty For More Information Contact your local Tektronix service center or sales engineer for more information on any of these repair or adjustment services RTX100B ISDB T RF Signal Generator Service Manual hE aaa Specifications Product Overview The RTX100B ISDB T RF Signal Generator converts a broadcast transport stream into an RF signal and outputs it In addition to the RF signal output feature the RTX100B records and plays MPEG 2 transport streams that are compliant with ATSC DVB and ARIB standards The RTX100B provides the following features RF modulated output of broadcast transport streams UHF 13 channel to 62 channel 473 MHz to 767 MHz Direct RF output of ASI input signals Data rate 200 Mbps maximum RAM mode 256 Kbps minimum Hierarchy display of stored or captured transport streams 188 204 208 bytes packet size S TMCC M TMCC non transport stream and partial transport stream output formats Real time updating of stream parameters continuity counter PCR PTS DTS TOT TDT STT NPT and Re
45. network administrator for the appropriate IP settings ibiain an IP address automalicalli Use the following IP address Obtain DNS server address automatically Use the following DNS server addresses Figure 3 5 Internet Protocol TCP IP Properties dialog box The settings in the dialog box depend on whether a DHCP Dynamic Host Configuration Protocol server is in the network connected to the RTX100B When the DHCP Server Is in the Network If the DHCP server is in the network perform steps 9 and 10 otherwise skip to step 11 9 In the dialog box select Obtain an IP address automatically and Obtain DNS server address automatically 10 Click the OK button The RTX100B accesses the DHCP server and obtains the addresses automatical ly when connected to the network Refer to the user documentation supplied with your server OS for detailed information about DHCP server functions NOTE Under some network environments the RTX100B may not be able to obtain the IP address automatically from a DHCP server In this case you need to enter the appropriate address value in each submenu item RTX100B ISDB T RF Signal Generator Service Manual Performance Verification When the DHCP Server Is Not in the Network If the DHCP server is not in the network perform the following procedure to set the network parameters 11 If you connect the RTX100B directly to a single PC MTS400 a In the dialog box sel
46. that becomes conductive due to condensation These are sheltered locations where neither temperature nor humidity is controlled The area is protected from direct sunshine rain or direct wind Pullution Degree 2 as defined in IEC 61010 1 Note Rated for indoor use only Equipment type Test and Measuring Installation Category II as defined in IEC 61010 1 Annex J Pollution Degree 2 as defined in IEC 61010 1 Safety Class grounded product RTX100B ISDB T RF Signal Generator Service Manual js Theory of Operation EEE Theory of Operation A12 Main Board PCI Interface Mega FIFO TS Controller This section describes the basic operation of the major circuit blocks or modules in the RTX100B The Diagrams section beginning on page 7 1 includes a block diagram and an interconnect diagram The A12 Main board consists of the following blocks The PCI interface consists of a PCI target and a PCI master The PCI target has three base address register areas The PCI master has two DMA controllers used to transfer data from the A12 Main board to the system memory or from the system memory to the A12 Main board The Mega FIFO has 32 MB for playing and 32 MB for recording a stream data It consists of two 256 megabit synchronous DRAM and FPGA This FIFO is used to compensate for non realtime operation of Windows XP The TS controller consists of the following blocks TS Packet Type Controller This controller sets a pac
47. the Keypad board See Figure 5 9 on page 5 20 a Use a screwdriver with a T 10 Torx tip to remove the two screws securing the Keypad board to the front panel bracket b Lift the board away 5 Remove the LCD module See Figure 5 9 on page 5 20 a Use a screwdriver with with a T 10 Torx tip to remove the four screws securing the LCD module to the front panel bracket b Lift the LCD module away 6 Now manually disassemble the front panel assembly components using Figure 5 9 as a guide Reverse the procedure to reassemble 7 Reinstallation Perform steps 2 through 6 in reverse order to reinstall the front panel assembly RTX100B ISDB T RF Signal Generator Service Manual 5 19 Removal and Installation Procedures LCD inverter board Front Panel Processor board E di z Nd Front panel bracket d Emat adi Play Stop Record LCD module he Keypad board insulator i n Keypad board LCD glass 7 i shield S x S 9 Emat all others gt dn SN Ns Front cover Figure 5 9 Disassembly of front panel assembly 5 20 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Procedure for Board Modules Perform the Access Procedure on page 5 14 before doing any procedure in this group The procedures are Top Frame and Board 1 Retainers Top frame and board retainers A150 ISDB T RF Output board 12
48. the U S Federal Catalog handbook H6 1 for further item name identification Abbreviations Abbreviations conform to American National Standard ANSI Y1 1 1972 RTX100B ISDB T RF Signal Generator Service Manual 8 3 Replaceable Parts List Replaceable parts list Fig amp index number 8 1 1 8 4 2 3 4 Tektronix part number 211 0374 00 337 4345 00 390 1212 01 211 0711 00 335 1374 00 063 3868 00 337 4346 00 211 0374 00 Serial no effective Serial no discont d Qty Name amp description SCREW MACHINE 6 32X0 25 FLH 100 DEG STL ZN CM1 T10 TORX SHIELD ELEC COVER LEFT MTX100A AL CABINET ASSY WRAP AROUND HOUSING 0 050 AL SILVERGRAY W FEET amp HANDLE SAFETY CONTROLLED SCR ASSEM WSHR 6 32X0 250 PNH STL CDPL T 15 TORX DR MACHINE W SQ CONE WASHER MARKER DENT LABEL MKD WARNING TO AVOID ELECTRIC 50 8MMX25 4MM 0 002 POLYESTER 3M 7982 OVER LAMINATE W 0 001 FLEVCON W ADHESIVE BACK SAFETY CONTROLLED SOFTWARE PKG MICROSOFT WINDOWS XP PROFESSIONAL INCLUDES SERVICE PACK 2 1 2 PROCESSOR VERSION CERTIFICATE OF AUTHENTICITY SHIELD ELEC COVER RIGHT MTX100A AL SCREW MACHINE 6 32X0 25 FLH 100 DEG STL ZN CM1 T10 TORX RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List MOE 2 SN NNNNNNNNM WE 1 Cabinet and covers Figure 8 RTX100B ISDB T RF Signal Generator Service Manual Replaceab
49. to check the settings of Administrative tools for improved performance during the Playout 1 Click Control Panel gt Administrative Tools gt Services in Classic view 2 Select Windows Firewall Internet Connection Sharing ICS to open the dialog box Check if the Startup type is set to Disabled General Log On Recovery Dependencies Service name Shared amp ccess Display name Windows Firewall Internet Connection Sharing ICS Description Provides network address translation addressing name resolution and or intrusion prevention services Path to executable Ssystem32 ssvchost exe k netsvcs Startup type Disabled v Service status Stopped fou can specify the start parameters that apply when you start the service from here Start parameters nk J Cancel Apply 3 Select Automatic Updates to open the dialog box Check if the Startup type is set to Disabled RTX100B ISDB T RF Signal Generator Service Manual 3 33 Performance Verification Automatic Updates Properties Local Computer ES General Log On Recovery Dependencies Service name wuausery Display name Automatic Updates Description Enables the download and installation of Windows A updates If this service is disabled this computer will Path to executable E WIND OWS system32 syvchost exe k netsvcs Startup type Service s
50. to incompatible equipment c to repair any damage or malfunction caused by the use of non Tektronix supplies or d to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES EXPRESS OR IMPLIED TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TEKTRONIX F RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES fs Table of Contents Specifications Theory of Operation General Safety Summary Vii Service Safety Summary ix Environmental xi Prefacexve env so reote sot ey rte VaL pa vare aid ue xiii Manu al Structure s du oc Hid ge ORT E Sie rhe eq A n xiii Manual Conventions 0 0 xiv Finding Other Information peasair en e xiv Intr duction RR cnl na isla
51. when a bit error BER is detected stop the measurement and then restart 31 Change Center Frequency to 14 Channel 14 on the RTX100B 32 Change Input CH to 14 on the ISDB T analyzer 33 Repeat steps 29 and 30 34 Repeat steps 31 through 33 for Channel 15 to Channel 62 35 Change the demodulation DEMODSET settings to Mode 2 on the ISDB T analyzer 36 Open the ISDB_T_M2 rmx file on the RTX100B a Select File Open to open the Select File dialog box b In the dialog box navigate to the D ISDB T directory and then select the ISDB T M2 rmx file 37 Press the Play Pause button on the RTX10OB to start playing the file 38 Repeat steps 29 and 30 39 Change the demodulation DEMODSET settings to Mode 1 on the ISDB T analyzer 40 Open the ISDB T Mil rmx file on the RTX100B a Select File gt Open to open the Select File dialog box b In the dialog box navigate to the D ISDB T directory and then select the ISDB T file 41 Press the Play Pause button on the RTX100B to start playing the file 42 Repeat steps 29 and 30 3 30 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification IEEE1394b Interface This test verifies that the IEEE1394b interface is functioning correctly IEEE1394b hard disk drive IEEE1394b cable 9 pin 9 pin test40 TRP file Equipment required 1 Use the IEEE1394b cable to connect the IEEE1394b connector on the RTX100B to the IEEE139
52. 0 Base T port on the rear panel for Ethernet communications This section provides instructions for connecting the RTX100B to a single PC or a network and setting the network parameters for the LAN port Connecting the RTX100B The RTX10OB uses the rear panel LAN port to communicate with a PC to your PC MTS400 5400 Use one of the following methods to connect the RTX100B to your PC MTS400 If you are connecting the RTX100B directly to a single PC use a crossover Ethernet cable to connect between the LAN port on the RTX100B and the Ethernet port on the PC If you need to construct your own crossover cable the following figure shows the pin connections to change on a straight cable to produce a crossover cable Pin 1 TD Pin 1 TD Pin 2 TD Pin 2 TD Pin 3 RD Pin 3 RD Pin 6 RD Pin 6 RD Figure 3 1 Pin connections for a crossover Ethernet cable If you are connecting the RTX100B to your local Ethernet network use a straight Ethernet cable to connect between the LAN port on the RTX100B and the Ethernet hub port of your local network By connecting to an Ethernet network you can access the RTX100B using any PC on the network RTX100B ISDB T RF Signal Generator Service Manual 3 1 Performance Verification Setting Ethernet Network Parameters You can set the network parameters for the RTX100B using the Control Panel of Windows XP NOTE The following procedure requires that you are familiar with the basics of using
53. 00 833 9200 Outside North America contact a Tektronix sales office or distributor see the Tektronix Web site for a list of offices www tektronix com Module Repair and Return You may ship your module to us for repair after which we will return it to you New Modules You may purchase replacement modules in the same way as other replacement parts RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Using the Replaceable Parts List This section contains a list of the mechanical and or electrical components that are replaceable for the RTX100B Use this list to identify and order replacement parts The following table describes each column in the parts list Parts list column descriptions Column Column name Description 1 Figure amp index number Items in this section are referenced by figure and index numbers to the exploded view illustrations that follow 2 Tektronix part number Use this part number when ordering replacement parts from Tektronix 3and4 Serial number Column three indicates the serial number at which the part was first effective Column four indicates the serial number at which the part was discontinued No entries indicates the part is good for all serial numbers 5 This indicates the quantity of parts used 6 Name amp description An item name is separated from the description by a colon Because of space limitations an item name may sometimes appear as incomplete Use
54. 10 Torx tip b Locate the A40 AC Distributor board in the locator diagram Internal modules Figure 5 4 page 5 13 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its left side is facing you Remove the A40 AC Distributor board See Figure 5 13 on page 5 31 a Disconnect the cables from J100 J110 and J120 b Usea screwdriver with a T 10 Torx tip to remove the three screws securing the A40 AC Distributor board to the chassis c Liftthe board away from the chassis Reinstallation Perform step 3 in reverse order to reinstall the A40 AC Distributor board RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures 12 V main power supply A40 AC Distributor board Figure 5 13 12 V main power supply and A40 AC Distributor board removal RFI filter 1 Assemble equipment and locate modules to be removed a You need a screwdriver handle a T 10 Torx tip and a T 15 Torx tip b Locate the RFI filter in the locator diagram Internal modules Figure 5 4 page 5 13 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its rear is facing you 3 Remove the RFI filter See Figure 5 14 a Disconnect the two cables from the RFI filter b Use a screwdriver with a T 15 Torx tip to remove the screw securing the ground lead to the chassis RTX100B ISDB T RF Signal Gene
55. 143 30 575 143 47 677 143 14 479 143 31 581 143 48 683 143 15 485 143 587 143 49 689 143 16 491 143 503 143 50 695 143 17_ 497 143 599 143 51 701 143 18 503 143 35 605143 52 707 143 19 509 143 36 611143 53 713 143 20 54 179143 2 55 725 149 22 56 731 143 m 153843 4 jesus ___ 5 77w 24 58 743 143 25 59 749 143 26 155143 essas 755 148 27 _ ens __ jet 761 143 28 167 148 m seus _ Checking the Spectrum Mask 8 Select ISDB T ASI gt RF Parameter on the RTX100B to open the ISDB T RF Parameter dialog box 9 In the dialog box set Center Frequency to 13 10 Open the ISDB T M3 rmx file on the RTX100B a Select File Open to open the Select File dialog box b In the dialog box navigate to the D ISDB T directory and then select the ISDB T M3 rmx file 11 Press the Play Pause button to start playing the file RTX100B ISDB T RF Signal Generator Service Manual 3 27 Performance Verification 12 13 14 15 16 17 18 19 20 21 3 28 Set the spectrum analyzer as indicated below CENTER FREQ 473 143 MHz FREQ SPAN 15 MHz RBW 10 kHz Loses 300 Hz SWP Speed 10s REF Level 30 dBm None ATT V sents Soe e 0 dB Average 10 Use the delta marker to measure the difference between the marker frequency set in step 12 and the specified frequency o
56. 4 Internal modules 5 13 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Access Procedure When you have identified the module to be removed for service read General Instructions found on page 5 8 Then use the flowchart in Figure 5 5 to determine which procedures to use for removing the module The removal procedures end with installation instructions Locate the module to i be removed in Procedures for Internal Modules Figure 5 2 to 5 4 Hard disk drive p 5 26 Internal and external fans p 5 28 12 V main power supply p 5 30 A40 AC Distributor board p 5 30 RFI filter p 5 31 Is the module in Figure 5 4 Procedures for External Modules Cabinet p 5 15 Right side cover p 5 16 Is the module in Figure 5 2 Yes Left side cover p 5 16 Front panel unit p 5 18 Front panel assembly p 5 19 Do the procedures Cabinet and Right and Left Covers 5 15 Procedures for Board Modules Is A150 RF Output board the module 12 Main board in Figure CPU board 5 3 A20 PCI Backplane board Printer connector Figure 5 5 Guide to removal procedures 5 14 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Procedures for External Modules Perform the Access Procedure page 5 14 before performing any procedure in this group This group contains the following procedures Cabinet 1 Cabinet Ri
57. 4b hard disk drive See Figure 3 17 RTX100B rear panel KR OO IEEE1394b hard disk drive OOOO of OO IEEE1394b cable Figure 3 17 Equipment connection for checking the IEEE1394b interface 2 Select Play gt Minimize to minimize the RTX100B application 3 Power on the IEEE1394b hard disk drive 4 Double click the My Computer icon on the Windows desktop to open the My Computer window 5 In the Window verify that the hard disk drive is recognized as the or G drive 6 Usethe windows file copy operation to copy the test40 TRP file on the D drive to F or G drive 7 Verify that the test40 TRP file is copied to the F or G drive correctly 8 Delete the test40 TRP file from the or G drive 9 Power off the IEEE1394b hard disk drive RTX100B ISDB T RF Signal Generator Service Manual 3 31 Performance Verification IP Interface The following tests verify that transport stream data is correctly output from the RTX100B instrument Recommended Settings for MPEG Player in IP mode Perform this procedure in Category Classic view to check the settings of Intel Pro 1000 MT drivers 1 Click Control Panel gt
58. 64 StrmTyp 0x2 AUDIO PID 0x65 StrmTyp 0x4 FEY PCR PID 0x64 e E PMT PID 0x16 0 2 PrgNo 0x2 VIDEO PID 0x6E StmTyp 0x2 AUDIO FID Ox6F StrmTyp Ox4 E PMT PID 0x17 TbIID 0x2 PrgNo 0x3 eo VIDEO PID 0x78 0x2 PMT PID 0x18 ThIID 0x2 PrgNo 0x4 E VIDEO PID 0x82 StrmTyp 0x2 188 INTERNAL 40 000000Mbps IP 41 641337Mbps Figure 3 20 RTX100B MPEG Player display with 40 000000 Mbps 4 Click the Start button to start the stream playout 5 Double click the MTS400 Series system s TS Compliance Analyser desktop icon E H P TS Compliance Analyser lnk to launch the TS Compliance Analyser application TSCA This starts with the Open Transport Stream dialog box as shown in Figure 3 21 The Start button is the button with the green horizontally oriented triangle just below the SPI ASM 310M menu near the top of the application window 3 40 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Open Transport Stream Stream Interpretation Base Standard DVB Region No Extensions Using default scripts Scripted validations are not enabled Lookin MTS C dc Cabs Advanced Settings File Shortcuts Et Documents and Settings ream Bitrate My Recent Inetpub The strearn bitrate is used as a basis For timing Real time Analysis Documents j Program Files measurements during analysis Please selec
59. A48100 CIRCUIT BD ASSY A12 MAIN 389 3734 00 WIRED CIRCUIT BD ASSY A150 SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX CIRCUIT BD ASSY A20 PCI BACK PANEL 389 3735 00 WIRED CABLE ASSEMBLY 28AWG FLAT 1 27CTR W CONN D SUB2582X13 WASHER LOCK 4 SPLIT 0 025 THK STL CD PL SCREW JACK 4 40X0 312 EXT THD 4 40 INT THD 0 188 HEX STEEL CAD PLATE SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX PANEL PCI BLANK NI PL STL SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX FRAME SECTION TOP MTX100A STL SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX RETAINER CKT BD PCI MTX100A STL RETAINER CKT BD PCI RTX100A STL RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Figure 8 3 Internal modules 2 8 9 RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Replaceable parts list Fig amp index number 8 4 1 2 8 8 10 number 335 1597 00 101 0175 00 337 4389 00 211 0373 00 119 7205 00 407 5197 00 174 5266 00 174 3767 01 119 5999 00 211 1117 00 335 0577 00 211 1117 00 679 6251 00 337 4390 00 679 6250 00 211 0373 00 260 2834 00 260 2835 00 335 1595 00 Serial no effective Serial no discont d Qty Name amp description MARKER IDENT LABEL RTX100B POLYCARBONATE SAFETY CONTROLLED TRIM FRONT BEZEL PC ABS TV GRAY MTX100B SAFETY CONTROLLED SHIELD DISPLAY 5 568 X 4 221
60. I Backplane Board 5 24 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its right side is facing you Remove the CPU board See Figure 5 10 on page 5 22 a Disconnect these cables m The cable from the hard disk drive at m The cable from the Printer connector at CN3 m The cable from the A20 PCI Backplane board at CN5 CN6 CN9 and CN14 m The cable from the Front Panel Processor board at CN7 b Use a screwdriver with a T 10 Torx tip to remove the two screws securing the board bracket to the chassis See Figure 5 10 on page 5 22 Grasp the board and slide it out Reinstallation Perform step 3 in reverse order to reinstall the CPU board Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the A20 PCI Backplane board in the locator diagram Board modules Figure 5 3 page 5 12 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its right side is facing you Remove the A20 PCI Backplane board See Figure 5 10 on page 5 22 a Remove the A150 ISDB T RF Output board as described on page 5 23 b Remove the A12 Main board as described on page 5 23 c Remove the CPU board as described on page 5 23 d Disconnect these cables m The cable from the A40 AC Distributor board at J220 m The cable from the 12 V main power supply at J160 and J170 m The cable from the external fan at J190
61. Main board CPU board A20 PCI Backplane board Printer connector Assemble equipment and locate modules to be removed m You need a screwdriver with a T 10 Torx tip Orient the instrument Set the RTX100B so its bottom is down on the work surface and its right side is facing you Remove the board retainers See Figure 5 10 on page 5 22 a Use a screwdriver a T 10 Torx tip to remove the three screws securing the board retainers to the chassis b Lift the retainers away from the chassis Remove the top frame See Figure 5 10 on page 5 22 a Use a screwdriver with a T 10 Torx tip to remove the three screws securing the top frame to the chassis b Lift the top frame away from the chassis RTX100B ISDB T RF Signal Generator Service Manual 5 21 Removal and Installation Procedures A20 PCI Backplane board NI Board retainer Top frame A12 Main board A150 ISDB T RF Output board CPU board Figure 5 10 A150 RF Output board A12 Main board CPU board and A20 PCI Backplane board removal RTX100B ISDB T RF Signal Generator Service Manual 5 22 Removal and Installation Procedures A150 ISDB T RF Output Board Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the A150 ISDB T RF Output board in the locator diagram Board modules Figure 5 3 page 5 12 2 Orient the instrument Set the RTX100B so its bottom is
62. NC cable from the RTX100B and frequency counter SPI Interface This test verifies that transport stream data is correctly played from and recorded through the SPI In Out connector on the RTX100B Equipment required MPEG analyzer Parallel interface cable test40 TRP file Checking the Play Operation 1 Use the parallel interface cable to connect the SPI In Out connector on the RTX100B to the DVB SPI In connector on the MPEG analyzer See Figure 3 8 MPEG analyzer rear panel MTS430 P RTX100B rear panel e mexmme O noo Parallel interface cable Figure 3 8 Equipment connection for checking the play operation SPI interface 2 Start the MPEG Player application on the MPEG analyzer 3 Setthe application to the Record mode 4 Open the test40 TRP file on the RTX100B a Select File Open in the Play screen to open the Select File dia
63. Network and Internet Connections gt Network Connections in Category view or click Control Panel Network Connec tions in Classic view 2 Select Intel R Pro 1000 MT Network Connection device and right click to select the Properties option 3 Click Configure Advanced and check if the following options are set in the dialog box m Adaptive Inter Frame Spacing Disabled m Flow Control Off m Interrupt Moderation Rate Off m Jumbo Frames 16128 m Link Speed and Duplex Auto negotiate 1000Mbps NOTE If RTX100B MPEG system is connected to 100 Mbps network select Auto Detect for Link Speed and Duplex option m Receive Descriptors 2048 3 32 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Intel R PRO 1000 MT Network Connection Properties General Advanced Driver Resources Power Management he following properties are available for this network adapter Click the property you want to change on the left and then select its value on the right Property Value Adaptive Inter Frame Spacit Disabled Enable PME 3 Flow Control Gigabit Master Slave Mode nterrupt Moderation Rate Jumbo Frames Link Speed amp Duplex Locally Administered Address Log Link State Event Offload Receive IP Checksum Offlaad Receive TCP Checksum Offload Transmit IP Checksum Offload Transmit TCP Checksum QoS Packet Tagging os Cancel Perform this procedure
64. Number 100 All Data 16382 Continuous Load Impedance 1 kQ 1 5 5 Press the On button of the Ch1 output on the function generator 6 Open the test64 TRP file on the RTX100B a b RTX100B ISDB T RF Signal Generator Service Manual Select File gt Open in the Play screen to open the Select File dialog box In the dialog box select the test64 TRP file 3 15 Performance Verification 3 16 7 Set Play gt Update gt On 8 Select Play gt Clock to open the Clock dialog box 9 In the Clock dialog box select ExtRef 10 10 Press the Play Pause button on the RTX100B to start playing the test64 TRP file 11 Verify that PLL unlock error does not occur on the RTX100B 12 Start the TS Compliance Analyzer on the MPEG analyzer 13 In the Open Transport Stream dialog box select Real time Analysis and then make the following settings Interfaces DVB Parallel Interface Settings Time Stamping 14 Click the OK button 15 Verify that the hierarchic view is displayed on the MPEG analyzer screen and that no error messages appear In addition verify that the bit rate is 64 Mbps and the packet size is 188 bytes NOTE Ignore Program 3 and PID 120 0x78 errors in the hierarchical view because these are caused by the original test40 TRP file 16 Change the clock source setting on the RTX100B and output frequency setting on the function generator with the following and then repeat
65. R LEAD ELECTRICAL 18 AWG 4 0 1 5 4 SAFETY CONTROLLED FILTER RFI 3A 250VAC 0 5 MA 0 150HM SAFETY CONTROLLED SCREW MACHINE 4 40X0 312 PNH STL ZN CM1 T10 TORX COVER FAN REAR MTX100A AL SCREW MACHINE 4 40X1 500 PNH HEAD T 10 TORX ZINC PL CABLE ASSEMBLY 4 18AWG 60MM L HDD POWER 40 SQ FAN MARKER DENT MKD CONNECTORS FOR A150 MARKER DENT MKD CONNECTORS FOR A12 POLYCARBONATE MARKER IDENT MKD CONNECTORS PRINTER amp IEEE1394B POLYCARBONATE MARKER DENT MKD CONNECTORS FOR CPU BD POLYCARBONATE MARKER IDENT BLANK LABEL FOR MES LINES THT 37 483 10MONOCHROME DISPLY TDS300 TDS400 SERIES SAFETY COTROLLED CHASSIS ASSY MAIN MTX100A AL RIVET SOLID 0 163 ODX0 415 L PANEL RANGE HEAD STYLE NYLON FAN 119569400 W CABLE 18CM L 92 SQ FAN RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Front panel unit See Figure 8 4 on page 8 11 for detailed view Figure 8 2 Internal modules 1 RTX100B ISDB T RF Signal Generator Service Manual Replaceable Parts List Replaceable parts list Fig amp index number 8 3 1 8 8 2 3 4 Tektronix part number 650 4981 00 671 5975 00 671 B272 XX 211 0373 00 671 5976 00 174 5190 00 210 0054 00 214 3903 01 211 0373 00 333 4518 00 211 0373 00 211 0373 00 426 2624 00 211 0373 00 343 1697 00 343 1700 00 Serial no effective Serial no discont d or o AN n Name amp description PROCESSOR MODULE W BRACKET LABEL 334
66. RR 1 C164 4 Honronnunconconconzral R102 Lutes wine aise J230 cuum peb e cios BI BES bo R105 c75 Vae cH z Riga S ME REE R101 me o E EF 200 R184 R12 2 FL190 Rests Hd e gt C180My ii 7478 mm ii 27 EN tere 4 a Oz fo J240 J210 CT o a a O PN1 Figure 5 24 A20 PCI Backplane board view right side 5 44 RTX100B ISDB T RF Signal Generator Service Manual a ee System Software Recovery If the RTX100B does not boot up or is corrupted you can recover the system software by using this recovery procedure CAUTION Running the recovery procedure overwrites all the contents of the hard disk drive and resets the system to the factory defaults The recovery can be performed in two ways m Recover Boot Partition This permanently overwrites the contents of your boot partition This mode of recovery is preferred if you want to keep your old data files on the D drive m Recover Entire Drive This permanently overwrites the contents of your entire hard disk drive and resets the system to the factory defaults Recovery Procedure Perform the following procedure to recover the Windows XP and RTX100B application software NOTE The
67. RTX100B performance Section 7 Diagrams contains a block diagram using individual modules as blocks and an interconnection diagram showing connections between the modules Section 8 Replaceable Parts List lists all field replaceable modules by part number RTX100B ISDB T RF Signal Generator Service Manual 5 1 Maintenance Preparation 5 2 ZN Preventing ESD Before servicing this product read the General Safety Summary Safety Summary and Introduction at the front of the manual and the following ESD information CAUTION Static discharge can damage any semiconductor component in the RTXIOOB When performing any service that requires internal access to the RTX100B adhere to the following precautions to avoid damaging internal modules and their components due to electrostatic discharge ESD 1 Minimize handling of static sensitive modules 2 Transport and store static sensitive modules in their static protected containers or on a metal rail Label any package that contains static sensitive modules 3 Discharge the static voltage from your body by wearing a grounded antistatic wrist strap while handling these modules Perform service of static sensitive modules only at a static free work station 4 Nothing capable of generating or holding a static charge should be allowed on the work station surface 5 Handle circuit boards by the edges when possible 6 Do not slide the modules over any surface
68. SDB T RF Signal Generator Service Manual Performance Verification 16 17 18 19 20 21 22 In the dialog box navigate to the E MTXRTX_Test streams Record Files and then select the SPI214Mbps trp file On the MPEG analyzer make the following settings Play menu Data Rate 214 Mbps Update Off Source RAM Click the PLAY button on the MPEG Analyzer to start playing the SPI214Mbps trp file Verify that the hierarchic view is displayed on the RTX100B screen In addition verify that the bit rate display is 214 Mbps and the packet size display is 188 bytes Press the Record button on the RTX100B to record the file After recording is completed click the Stop button on the MPEG analyzer Exit the MPEG Player application on the MPEG analyzer Checking the Recorded File 23 24 25 26 27 28 29 30 31 Change the interface cable connection from DVB SPI Out connector to DVB SPI In connector on the MPEG analyzer Start the TS Compliance Analyzer on the MPEG analyzer In the Open Transport Stream dialog box select Real time Analysis and then make the following settings Interfaces DVB Parallel Interface Settings Time Stamping Click the OK button Press the Play Pause button on the RTX100B to display the Play screen Select File Open to open the Select File dialog box In the dialog box navigate to the D Record Files directory and then select
69. Save Mode Over Write Save D Record_Files ASI214Mbps trp Set the MPEG analyzer to the Play mode Select File gt Open on the MPEG analyzer to display the Open dialog box In the dialog box navigate to the ESMTXRTX Test Streams Re cord Files directory and then select the ASI214Mbps trp file On the MPEG player make the following settings Play Clock Data Rate 214 Mbps Update Off Source RAM SPI ASI 310M BNC Port ASI Through Out Off Click the PLAY button on the MPEG analyzer to start playing the ASI214Mbps trp file Verify that the hierarchic view is displayed on the RTX100B screen In addition verify that the bit rate is 214 Mbps and packet size is 188 bytes Press the Record button on the RTX100B to record the file After recording is completed click the Stop button on the MPEG analyzer Exit the MPEG Player application on the MPEG analyzer Disconnect the 75 Q BNC cable from the RTX100B and MPEG analyzer RTX100B ISDB T RF Signal Generator Service Manual Performance Verification RTX100B rear panel Checking the Recorded File 29 Use the 75 BNC cable to connect the ASI Out connector on the RTX100B J noo LT Uu oo qd HW
70. Service Manual Tektronix RTX100B ISDB T RF Signal Generator 071 1932 03 USA ipment duipmen An Interworld Highway LLC Company Warning The servicing instructions are for use by qualified personnel only To avoid personal injury do not perform any servicing unless you are qualified to do so Refer to all safety summaries prior to performing service www tektronix com Copyright Tektronix All rights reserved Licensed software products are owned by Tektronix or its subsidiaries or suppliers and are protected by national copyright laws and international treaty provisions Tektronix products are covered by U S and foreign patents issued and pending Information in this publication supercedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc Contacting Tektronix Tektronix Inc 14200 SW Karl Braun Drive P O Box 500 Beaverton OR 97077 USA For product information sales service and technical support In North America call 1 800 833 9200 m Worldwide visit www tektronix com to find contacts in your area Warranty 2 Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one 1 year from the date of shipment If any such product proves defective during this warranty period Tektronix at its option either will repair the
71. are application will operate The software key is the character string which is entered into the Option Update Key string field Beginning with V10 1 an OptionDongle is also provided The OptionDongle is a small hardware device that is available in parallel port form The software key and the OptionDongle work together the software polls the OptionDongle at the beginning of the application and during the Play and Record to verify the option coding The type of software key used by your test system is determined at the time of the original product order Any printer compatible with the installed operating system can be connected to the unit through the parallel port version of the OptionDongle Performance Conditions The electrical characteristics listed on the following pages are valid under the following conditions m The RTX100B must be in an environment where the temperature altitude humidity and vibration conditions are within the operating limits described in Table 1 4 on page 1 12 m The RTX100B must have a warm up period of at least 20 minutes m The RTX100B must be operating at an ambient temperature between 5 C to 40 C unless otherwise noted RTX100B ISDB T RF Signal Generator Service Manual 1 3 Specifications Functional Specifications Table 1 1 Functional specifications Characteristics System configuration System OS CPU System memory Display Hard disk drive Expansion slot Description W
72. ately after the data is recovered the instrument restarts and the Windows XP setup wizard is displayed Click the Next button of the wizard until Windows XP is installed The instrument restarts after the installation of Windows XP to display the RTX100B application screen NOTE On B020000 and above the Application software V10 1 and above and the test streams will have to be reinstalled after the Operating System recovery RTX100B ISDB T RF Signal Generator Service Manual yy S Options RES E Options This section describes options that are available for the RTX100B The following options are available Options Description Software option Option SC Adds the Scheduler application Service options Option D1 Provides calibration data Option D3 Provides calibration data for 3 years Option D5 Provides calibration data for 5 years Option C3 Provides calibration services for 3 years Option C5 Provides calibration services for 5 years Option R3 Extends the instrument warranty to 3 years Option R5 Extends the instrument warranty to 5 years Power cord options Option North America power cord Option A1 Universal Euro power cord Option A2 United Kingdom power cord Option A3 Australia power cord Option A4 240 V North America power cord Option A5 Switzerland power cord Option A6 Japan power cord Option A10 China power cord Option A99 No power cord RTX100B ISDB T RF Signal Generator Service Manual 6 1 Opt
73. bed in the order in which they appear in this section In addition you can look up any procedure for removal and reinstallation of any module in the Table of Contents of this manual The Access Procedure on page 5 14 first directs you to the procedure s if any that are required to access the module to be serviced then it directs you to the procedure to remove that module Table 5 3 Summary of procedures Procedure Procedures for Cabinet m Front panel unit External Modules m Right side cover m Front panel assembly m Left side cover Procedures for Board m A150 ISDB T RF m A20 PCI Backplane Modules output board board m A12 Main board m Printer connector CPU board Procedures for Internal m Hard disk drive 40 AC Distributor Module Internal and external board fans m RFI filter m 12 V main power supply RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Required Equipment Most modules in this RTX100B can be removed with a screwdriver handle mounted with a size T 10 Torx screwdriver tip Use this tool whenever a procedure step instructs you to remove or install a screw unless a different size screwdriver is specified in that step All equipment required to remove and reinstall each module is listed in the first step of its procedure Table 5 4 Tools required for module removal 1 Accepts Torx driver bits 2 Torx driver bit for T 10 size screw heads 3 Torx driver bit for T
74. defective modules Cleaning Procedure Exterior To clean the RTX100B exterior perform the following steps 1 Remove loose dust on the outside of the RTX100B with a lint free cloth 2 Remove remaining dirt with a lint free cloth dampened in a general purpose detergent and water solution Do not use abrasive cleaners 3 Clean the light filter protecting the monitor screen with a lint free cloth dampened with a gentle general purpose detergent and water solution CAUTION To prevent getting moisture inside the RTX100B during external cleaning use only enough liquid to dampen the cloth or applicator Inspection Interior To access the inside of the RTX100B for inspection and cleaning refer to the Removal and Installation Procedures in this section Inspect the internal portions of the RTX100B for damage and wear using Table 5 2 as a guide Defects found should be repaired immediately CAUTION To prevent damage from electrical arcing ensure that circuit boards and components are dry before applying power to the RTX100B 5 4 RTX100B ISDB T RF Signal Generator Service Manual Maintenance Table 5 2 Internal inspection check list Loose broken or corroded solder connections Burned Item Circuit boards Resistors Capacitors Semiconductors Wiring and cables Chassis circuit boards Burned broken or cracked circuit run plating Burned cracked broken blis tered condition
75. defective product without charge for parts and labor or will provide a replacement in exchange for the defective product Batteries are excluded from this warranty Parts modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance All replaced parts modules and products become the property of Tektronix In order to obtain service under this warranty Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix shipping charges prepaid and with a copy of customer proof of purchase Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located Customer shall be responsible for paying all shipping charges duties taxes and any other charges for products returned to any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than Tektronix representatives to install repair or service the product b to repair damage resulting from improper use or connection
76. down on the work surface and its right side is facing you 3 Remove the interface module See Figure 5 10 on page 5 22 a Disconnect the cable from the A12 Main board b Usea screwdriver with a T 10 Torx tip to remove the two screws securing the board bracket to the chassis See Figure 5 10 on page 5 22 Grasp the board and slide it out 4 Reinstallation Perform step 3 in reverse order to reinstall the interface module A12 Main Board 1 Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the A12 Main board in the locator diagram Board modules Figure 5 3 page 5 12 2 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its right side is facing you 3 Remove the A12 Main board See Figure 5 10 on page 5 22 a Disconnect the cable from the A150 ISDB T RF Output board b Usea screwdriver with a T 10 Torx tip to remove the two screws securing the board bracket to the chassis See Figure 5 10 on page 5 22 c Grasp the board and slide it out 4 Reinstallation Perform step 3 in reverse order to reinstall the A12 Main board CPU Board 1 Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the CPU board in the locator diagram Board modules Figure 5 3 page 5 12 RTX100B ISDB T RF Signal Generator Service Manual 5 23 Removal and Installation Procedures 20 PC
77. e harmful to the environment or human health if improperly handled at the product s end of life In order to avoid release of such substances into the environment and to reduce the use of natural resources we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately This symbol indicates that this product complies with the European Union s requirements according to Directive 2002 96 EC on waste electrical and electronic equipment WEEE For information about recycling options check the Support Service section of the Tektronix Web site www tektronix com Mercury Nortification This product uses an LCD backlight lamp that contains mercury Disposal may be regulated due to environmental considerations Please contact your local authorities or within the United States the Electronics Industries Alliance www eiae org for disposal or recycling information Perchlorate Materials This product contains one or more type CR lithium coin cell batteries According to the state of California CR lithium coin cells are classified as perchlorate materials and require special handling See www dtsc ca gov hazardouswaste perchlorate for additional information This product has been classified as Monitoring and Control equipment and is outside the scope of the 2002 95 EC RoHS Directive This product is known to contain lead cadmium mercury and hexavalent chr
78. e eens 4 1 Equipment Required 00 eee eee een een ene 4 1 Carrier Leakage Adjustment 4 2 RTX100B ISDB T RF Signal Generator Service Manual i Table of Contents Maintenance Related Maintenance Procedures 5 1 Preparation zero e o eben eerte mre bit ee ehe 5 2 Inspection and Cleaning 0 0 eee eee ee nen 5 3 Removal and Installation Procedures 5 7 Preparation jen vot Sac x epe RSE 5 7 ACcess PIOCedU re Sisk Balan ea Si Re At ee Sell ant 5 14 Procedures for External Modules 5 15 Procedure for Board Modules 5 21 Procedure for Internal Modules 5 26 Verifying the BIOS Configuration 5 33 Procedure 2 ered Race he eA atr edic deste f re a 5 33 Troubleshooting Ere re RE Rer E 5 35 Troubleshooting Trees 0 0 eee eee ec III 5 35 System Software Recovery 5 45 Recovery Procedure 5 45 Using the Rescue Disk 5 47 Recovery Procedure iios sete ELE RE ae Ee ade RUE CE ERR 5 47 Options Diagrams Replaceable Parts List Parts Ordering Information
79. e the right side cover from the chassis Reinstallation Perform steps 3 and 4 in reverse order to reinstall the right side cover Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures b Locate the left side cover in the locator diagram External modules Figure 5 2 page 5 11 2 Orient the instrument Set the RTX100B so its right side is down on the work surface and its front is facing you 3 Use a screwdriver with a T 10 Torx tip to remove the six screws securing the left side cover to the top and bottom of the chassis See Figure 5 7 4 Remove the left side cover from the chassis 5 Reinstallation Perform steps 3 and 4 in reverse order to reinstall the left side cover QDMA VQ WEA WOO Figure 5 7 Right side and left side covers removal RTX100B ISDB T RF Signal Generator Service Manual 5 17 Removal and Installation Procedures Front Panel Unit 1 Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the front panel unit in the locator diagram External modules Figure 5 2 page 5 11 Orient instrument Set the RTX100B so its bottom is down on the work surface and its front is facing you Remove f
80. ect Use the following IP address b Set the IP address parameter to be the same IP address as the PC s MTS400 s address except for the last number The last number must be different from the last number in the PC s MTS400 s IP address c Set the Subnet mask parameter to be the same net mask subnet mask used by the PC MTS400 Do not enter a number if the PC MTS400 does not have a net mask d You do not need to enter a Default gateway if you are directly connected to a single PC MTS400 12 If you connect the RTX100B to your local Ethernet network a In the dialog box select Use the following IP address b Ask your local network administrator and set the appropriate addresses your local network administrator for the correct numbers to enter in the dialog CAUTION To prevent communication conflicts on your Ethernet network ask box if you connect the RTX100B to your local Ethernet network 13 Verify the Ethernet connection by using a ping command from the PC MTS400 RTX100B ISDB T RF Signal Generator Service Manual 3 5 Performance Verification Equipment Required Table 3 1 lists the test equipment required to perform all of the performance verification procedures The table identifies examples of recommended equipment and lists the required precision where applicable If you substitute other test equipment for the examples listed the equipment must meet or exceed the listed tolerances Table 3 1 E
81. ed Solomon ISDB T only Continuous recording of captured streams PCR jitter insertion Triggered stream capture Full remote control using Ethernet interface Scheduler application for automated stream playout and record Option SC only Playout on one interface at a time either IP or ISDBT ASI interfaces Two IP versions namely IPv4 and IPv6 are supported Stream replication IP with some attributes of the stream incremented by a certain value The RTX10OB includes the ReMux application software that provides the capability to create a transport stream of super frame structure defined in the ISDB S systems from an MPEG2 transport stream RTX100B ISDB T RF Signal Generator Service Manual 1 1 Product Overview 1 2 RTX100B ISDB T RF Signal Generator Service Manual aS a Specifications Tables 1 1 through 1 4 list the functional electrical mechanical and environ mental characteristics of the RTX100B Table 1 5 lists the national and international standards to which the RTX100B complies All listed specifications are guaranteed unless labeled typical Typical specifications are provided for your convenience but are not guaranteed Software Protection The RTX100B software options require that a software key and beginning with V10 1 an OptionDongle be installed so that the purchased set of options can be accessed CAUTION The software key OptionDongle must be installed on the instrument before the RTX100B softw
82. equipment for measurement control and laboratory use EN 61000 3 2 AC Power Line Harmonic Emissions IEC 61000 4 2 Electrostatic Discharge Immunity Performance Criterion B IEC 61000 4 3 RF Electromagnetic Field Immunity Performance Criterion A IEC 61000 4 4 Electrical Fast Transient Burst Immunity Performance Criterion B IEC 61000 4 5 Power Line Surge Immunity Performance Criterion B IEC 61000 4 6 Conducted RF Immunity Performance Criterion A IEC 61000 4 11 Voltage Dips and Interruptions Immunity Performance Criterion B Low Voltage Directive 73 23 EEC Amended by 93 68 EEC EN 61010 1 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Australia New Zealand Declaration Complies with EMC provision of Radio Communications Act per the following standard s of Conformity EMC TAM AS NZS 2064 1 2 Industrial Scientific and Medical Equipment 1992 RTX100B ISDB T RF Signal Generator Service Manual 1 13 Specifications Table 1 5 Certifications and compliances Cont Category Safety Installation Overvoltage Category Overvoltage Category Pollution Degree Pollution Degree IEC Characteristics Standards or description Complies with the following safety standards regulations UL 61010 1 Standard for Electrical Measuring and Test Equipment CAN CSA C22 2 No 61010 1 04 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use EN
83. er to Manual High Voltage Earth Terminal RTX100B ISDB T RF Signal Generator Service Manual EEE ee Service Safety Summary Only qualified personnel should perform service procedures Read this Service Safety Summary and the General Safety Summary before performing any service procedures Do Not Service Alone Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present Disconnect Power To avoid electric shock disconnect the mains power by means of the power cord or if provided the power switch Use Care When Servicing With Power On Dangerous voltages or currents may exist in this product Disconnect power remove battery if applicable and disconnect test leads before removing protective panels soldering or replacing components To avoid electric shock do not touch exposed connections RTX100B ISDB T RF Signal Generator Service Manual ix Service Safety Summary x RTX100B ISDB T RF Signal Generator Service Manual BS Lease a Environmental Considerations Product End of Life Handling Restriction of Hazardous Substances This section provides information about the environmental impact of the product Observe the following guidelines when recycling an instrument or component Equipment Recycling Production of this equipment required the extraction and use of natural resources The equipment may contain substances that could b
84. ffset of each marker and verify that each value meets the condition as follows Delta marker Marker 1 2 86 MHz lt 20 dB Delta marker Marker 2 3 00 MHz lt 27 dB Delta marker Marker 3 4 36 MHz lt 49 dB Change frequency offset of each marker and verify that each value meets the condition as follows Delta marker Marker 1 2 86 MHz lt 20 dB Delta marker Marker 2 3 00 MHz lt 27 dB Delta marker Marker 3 4 36 MHZ lt 49 dB Change Center Frequency to 38 on the RTX100B Change the Center Frequency to 623 143 MHz on the spectrum analyzer Repeat steps 13 and 14 Change Center Frequency to 62 on the RTX100B Change the Center Frequency to 767 143 MHz on the spectrum analyzer Repeat steps 13 and 14 Disconnect the 75 BNC Ma to NC Fe adapter 50 9 SMA cable and 50 Q N Ma to SMA Fe adapter from the RTX100B and spectrum analyzer RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Checking the Output Level and Error 22 RTX100B rear panel R A 1 LI noo 2000 H Use the 75 Q BNC cable to connect the RF Out connector on the RTX100B to the RF INPUT on the ISDB T analyzer See Figure 3 16 ISDB T analyzer VP 8480A
85. ght side cover Left side cover Front panel unit Front panel assembly Assemble equipment and locate module to be removed a You need a screwdriver with a T 15 Torx tip b Locate the cabinet in the locator diagram External modules Figure 5 2 page 5 11 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its rear is facing you Use a screwdriver with a T 15 Torx tip to remove the two screws securing the cabinet to the RTX100B See Figure 5 6 on page 5 16 Grasp the right and left edges of the cabinet toward the back Push the chassis rear panel toward the front of the RTX100B to separate it from the cabinet Slide the cabinet off the RTX100B Reinstallation Perform steps 3 through 5 in reverse order to reinstall the cabinet RTX100B ISDB T RF Signal Generator Service Manual 5 15 Removal and Installation Procedures 5 16 Right Side Cover Left Side Cover VHA Figure 5 6 Cabinet removal 1 1 Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the right side cover in the locator diagram External modules Figure 5 2 page 5 11 Orient the instrument Set the RTX100B so its left side is down on the work surface and its front is facing you Use a screwdriver with a T 10 Torx tip to remove the seven screws securing the right side cover to the top and bottom of the chassis See Figure 5 7 Remov
86. indows XP Professional 1 3 GHz 1GB 6 3 inch XGA 1024 x 768 Windows Control Panel setting is SVGA 800 x 600 250 GB 1 PCI slot Electrical Specifications Table 1 2 Mainframe Characteristics Maximum output rate Hard disk RAM Maximum record rate Hard disk RAM Internal reference clock Frequency Stability External reference clock input Connector type Input impedance typical Reference input Frequency Input level typical 1 4 Description gt 120 Mbps gt 200 Mbps 120 Mbps File size lt 4 GB just after disk format operation 90 Mbps File size 33 GB just after disk format operation gt 200 Mbps For Output_clock PCR PTS DTS packet operation timing and TDT STT time 27 MHz 1 ppm when manufactured 0 5 ppm per year 1 ppm over temperature range BNC 50 Q 8 129698 MHz 10 MHz and 27 MHz Sine wave 0 6 dBm Square wave 0 5 Vp p to 3 0 Vp p RTX100B ISDB T RF Signal Generator Service Manual Specifications Table 1 2 Mainframe Cont Characteristics Clock input Frequency Input level typical External trigger input MISC output Connector type Input impedance typical Threshold level High level Low level Output level High level Low level Output impedance PLL Frequency Output clock Output rate TS clock Internal and external reference 27 MHz and 10 MHz External parallel clock External serial clock P N and Jitter serial clock SPI in
87. ing Non operating Shock Non operating 1 12 Description 5 C to 40 C 20 C to 60 C 20 to 80 No condensation Maximum wet bulb temperature 29 C 5 to 90 No condensation Maximum wet bulb temperature 29 C Up to 3 km approximately 10 000 feet Maximum operating temperature decreases 1 C each 300 m above 1 5 km Up to 15 km approximately 50 000 feet 2 65 m s rms 0 27 Grms 5 Hz to 500 Hz 10 min three axes 22 3 m s rms 2 28 Grms 5 Hz to 500 Hz 10 min three axes 294 m s 30 G half sine 11 ms duration RTX100B ISDB T RF Signal Generator Service Manual Specifications Table 1 4 Environmental characteristics Cont Characteristics Description Installation requirements Power dissipation 100 W maximum Maximum line current is 1 3 Arms at 50 Hz Surge current 12 peak for less than 5 line cycles at 25 C after product has been off for at least 30 seconds Cooling clearance Top clearance 5 2 in Side clearance 5 cm 2 in Rear clearance 5 cm 2 in from the fan guard Certifications and Compliances Table 1 5 Certifications and compliances Category Standards or description EC Declaration of Conformity Meets the intent of Directive 89 336 EEC for Electromagnetic Compatibility Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities EMC Directive 89 336 EEC EN 55011 EMC requirement for Class A electrical
88. ing the External modules see Figure 5 2 on page 5 11 Board modules see Figure 5 3 on page 5 12 and Internal modules see Figure 5 4 on page 5 13 W Procedures for removal and reinstallation of the electrical and mechanical modules A disassembly procedure for removal of all the major modules from the RTX100B at one time and for reassembly of those modules into the RTX100B Such a complete disassembly is normally only done when completely cleaning the RTX100B Instructions for doing the actual cleaning are found under Inspection and Cleaning at the beginning of this section WARNING Before performing any procedure in this subsection disconnect the power cord from the line voltage source Failure to do so could cause serious injury or death RTX100B ISDB T RF Signal Generator Service Manual 5 7 Removal and Installation Procedures General Instructions Summary of Procedures 5 8 Read these general instructions before removing a module First read over the Summary of Procedures that follows to understand how the procedures are grouped Then read Equipment Required for a list of the tools needed to remove and install modules in the RTX100B If you are removing a module for service begin by performing the procedure Access Procedure page 5 14 By following the instructions in that procedure you remove the module to be serviced while removing the minimum number of additional modules The procedures are descri
89. ions 6 2 RTX100B ISDB T RF Signal Generator Service Manual DENNIS Diagrams CUN Diagrams This section contains the following diagrams m Block diagram of the RTX100B Interconnect diagram of the RTX100B Block diagram shows the modules and functional blocks in the RTX100B Interconnect diagram shows how the modules in the RTX100B connect together RTX100B ISDB T RF Signal Generator Service Manual 7 1 Diagrams 7 2 RTX100B ISDB T RF Signal Generator Service Manual Com SPI In Out ASI In VGA Ext ATX Power Ext Keyboard Trig In Out ASI Out Front Bezel CPU Board Clock Ref In A12 Main Board A150 ISDB T RF RF Out OUTPUT Board Printer USB Local Bus Com Hard Disk Drive Card Edge Card Edge Card Edge A20 PCI Back Plane Board IEEE1394b 5 V Main Power Supply LCD USB USB Front Panel Processor Board Inverter Board for LCD Backlight 5 V Sub Power Supply A40 AC Distributor Board EN Keypad Board Figure 7 1 RTX100B block diagram RTX100B ISDB T RF Signal Generator Service Manual T 3 RTX100B ISDB T RF Signal Generator Service Manual 174 5190 00 LA Z gt IEEE1394b CPU Board 174 5172 00 J120 174 5183 00 A e T T ASI In 2 9 J703 174 5174 00 174 5186 00 174 5348 00 174 5184 00 Card Edge
90. ision Level below m Instrument type or model number Instrument serial number Instrument modification number if applicable If you order a part that has been replaced with a different or improved part your local Tektronix field office or representative will contact you concerning any change in part number Change information if any is located at the rear of this manual Tektronix part numbers contain two digits that show the revision level of the part For most parts in this manual you will find the letters XX in place of the revision level number Part Number Revision Level Revision Level May Show as XX A A 670 7918 03 670 7918 XX When you order parts Tektronix will provide you with the most current part for your product type serial number and modification if applicable At the time of your order Tektronix will determine the part number revision level needed for your product based on the information you provide RTX100B ISDB T RF Signal Generator Service Manual 8 1 Replaceable Parts List Module Servicing Modules can be serviced by selecting one of the following three options Contact your local Tektronix service center or representative for repair assistance Module Exchange In some cases you may exchange your module for a reman ufactured module These modules cost significantly less than new modules and meet the same factory specifications For more information about the module exchange program call 1 8
91. ith the sum of the original value and the PCR counter value Receive Rate Counter This 24 bit counter is used for receive rate calculation on recording process When using a data valid signal DVALID the counter counts byte data rate When not using DVALID the counter counts receive clock Transmit Byte Counter This 32 bit counter counts transmitted bytes This counter can be read by the control software The interrupt controller sends the following interruption signals to the CPU Play FIFO empty Record FIFO full external trigger 10 Hz Playout DMA transfer finished and Record DMA transfer finished These interruptions can be disabled by the RTX100B application software At power on all interruptions are disabled The TS clock generator consists of a VCO Voltage Control Oscillator phase detector frequency divider The VCO is used for generating a TS clock and it covers frequencies of 400 MHz to 800 MHz The frequency divider divides the clock by 65536 to 1 The reference clocks consist of 27 MHz TCXO and half divided output of 54 MHz clock generator These clocks are used to compare the frequency of the TS Clock Generator RTX100B ISDB T RF Signal Generator Service Manual Theory of Operation A20 PCI Backplane Board PCI and ISA Connectors ATX Power Control Circuit Interconnect Circuit Fan Speed Control Circuit IEEE1394b Interface The A20 PCI Backplane board consists of the following connectors and circui
92. ket size is 188 bytes RTX100B ISDB T RF Signal Generator Service Manual 3 23 Performance Verification NOTE Ignore Program 3 and PID 120 0x78 errors in the hierarchical view because these are caused by the original test40 TRP file 38 Press the Stop button on the RTX100B to stop the stream output 39 Disconnect the 75 Q BNC cable from the RTX100B and MPEG analyzer Checking a Moving Picture 40 Use a 75 Q BNC cable to connect the ASI Out connector on the RTX100B to the TS ASI connector on the MPEG 2 measurement decoder See Figure 3 14 41 Use the 75 Q BNC cable to connect the SER75 Q connector on the MPEG 2 measurement decoder to the SDI input on the video monitor See Fig ure 3 14 Video monitor LMD 1420 BKM 320D D a n RTX100B rear panel BEV ee Pg ro e b ma P Di Jeb I9 75 Q BNC cable SFE IQ MPEG 2 measurement decoder rear panel DVMD MOO ED EB ec 75 BNC cable Figure 3 14 Fifth equipment connection for checking the ASI interface 3 24 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification
93. ket size of 188 204 208 and Non TS free length as a TS packet type For a Non TS format the controller also sets a packet length 16 255 and data length 16 255 The data length shows an effective data length in one TS packet Data Valid Controller This block generates a data valid signal that synchronizes with the first byte of a TS packet when stream output is started Psync Controller Psync is a signal to identify the sync byte in a transport stream When the controller is not in the Psync Regeneration mode Psync is recognized as the first byte of a transport stream In this case Psync is output in the constant interval even when the sync byte is rewritten in the value except for 47h For the Super Frame mode Psync is output only once in 384 packets Psync Regenerator This block becomes active when the Psync Regeneration mode is set The block detects the sync byte 47h in a data stream and regener ates a Psync signal The psync generator supports 188 204 208 TS packets only RTX100B ISDB T RF Signal Generator Service Manual 2 1 Theory of Operation 2 2 Interrupt Controller TS Clock Generator Reference Clocks PCR DTS Updater This block consists of a PCR counter and update logic The PCR counter is a complex counter which consists of a 9 bit counter range 0 299 and 33 bit counter The counter counts the 27 MHz standard frequency The update logic replaces the PCR and PTS DTS values in a transport stream w
94. le Parts List Replaceable parts list Fig amp index number 8 2 1 8 6 25 26 27 Tektronix part number 211 0734 00 211 0373 00 343 1708 00 211 0734 00 211 0734 00 386 7460 00 650 4862 00 650 5181 00 211 0373 00 671 5979 00 342 1138 00 211 0373 00 119 7147 00 211 0711 00 196 3496 00 119 6009 00 211 0372 00 200 4974 00 211 1161 00 119 7186 00 335 1369 00 335 1364 00 335 1363 00 335 1362 00 335 0133 00 441 2404 00 210 0164 000 119 7187 00 Serial no effective Serial no discont d Qty amp description fF oL oL c Ln eo ao pS PH SCREW MACHINE 6 32X0 25 FLH 100 DEG STL ZN CM1 T10 TORX SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX RETAINER HARD DISK DRIVE W POLYURETHANE MTX100A AL SCREW MACHINE 6 32X0 25 FLH 100 DEG STL ZN CM1 T10 TORX SCREW MACHINE 6 32X0 25 FLH 100 DEG STL ZN CM1 T10 TORX PLATE HDD AL HARD DISK DRIVE 160GB 3 5 INCH 7200RPM ATA 100 MTXB RTXB PRODUCTS FOR B019999 AND BELOW SERVICE KIT SYSPREP ACTIVATED XP OS TEST STREAMS INSTALLED HARD DISK DRIVE REPLACEMENT MTXB RTXB PRODUCTS B020000 AND ABOVE SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX CIRCUIT BD ASSY A40 SUB POWER 389 3738 00 WIRED INSULATOR CKT BOARD POLYCARBONATE A40 BOARD SCREW MACHINE 4 40X0 25 PNH STL ZN CM1 T10 TORX POWER SUPPLY 85V 264VAC IN 12VDC 12 5A OUT LEA150F 12 R COSEL SCR ASSEM WSHR 6 32X0 250 PNH STL CDPL T 15 TORX DR MACHINE W SQ CONE WASHE
95. log box b In the dialog box select the test40 T RP file RTX100B ISDB T RF Signal Generator Service Manual 3 11 Performance Verification 5 On the RTX100B make the following settings Play menu Data rate 214 Mbps Update Off Source RAM 6 On the MPEG analyzer make the following settings 9 Record Source SPI Target RAM Size 100 MB File Save Mode Overwrite Save E MTXRTX_Test streams Record_Files SPI214Mbps Press the Play Pause button on the RTX100B to start playing the test40 TRP Verify that the hierarchic view is displayed on the MPEG analyzer screen In addition verify that the bit rate is 214 Mbps and the packet size is 188 bytes Click the Record button on the MPEG Player application to record the file 10 After recording is completed press the Stop button on the RTX100B Checking the Record Operation 11 Change the interface cable connection from DVB SPI In connector to DVB SPI Out connector on the MPEG analyzer 12 Press the Record button on the RTX10OB to display the Record screen 13 On the RTX100B make the following settings Record menu Source SPI Target RAM Record Size 100 MB File menu Save Mode Over Write D Record_Files SPI214Mbps 14 Change the MPEG analyzer to the Play mode 15 Select File gt Open on the MPEG analyzer to display the Open dialog box 3 12 RTX100B I
96. n a component like a resistor or an integrated circuit Sometimes a single component is a module For example the chassis of the RTX100B is a module Safety Symbols and terms related to safety appear in the General Safety Summary and Service Safety Summary near the beginning of this manual Finding Other Information Other documentation for the RTX100B includes m The RTX100B ISDB T RF Signal Generator User Manual Tektronix part number 071 1930 X X contains a tutorial to quickly describe how to operate the RTX100B It also includes an in depth discussion on how to use the RTX100B features xiv RTX100B ISDB T RF Signal Generator Service Manual Introduction This manual contains information needed to properly service the RTX100B ISDB T RF Signal Generator as well as general information critical to safe and effective servicing To prevent personal injury or damage to the RTX100B consider the following before attempting service m The procedures in this manual should be performed only by a qualified service person m Read the General Safety Summary and the Service Safety Summary beginning on page vii When using this manual for servicing be sure to follow all warnings cautions and notes Performance Check Interval Strategy for Servicing Generally the performance check described in Section 3 Performance Verifica tion should be done every 12 months In addition a performance check is recommended after module re
97. nous SRAMs as buffers for interleaving This block performs OFDM modulation using IFFT This is a digitally controlled PLL which synchronizes the reference clock VCXO with an internal reference clock or an external clock This block has a feature to pass the U600 output to U1100 This block converts OFDM modulated data to an analog signal It has an over sampling circuit and digital filter for reducing spurious events The block also has two A D converters for I and Q signals This block generates a carrier signal for the RF output It directly generates a career frequency of UHF This block performs I Q modulation to a carrier signal from the A D converter output RTX100B ISDB T RF Signal Generator Service Manual Theory of Operation RF Matching Attenuator This block steps down an RF output level to the antenna input level of a receiver and matches the impedance with 75 Q Front Panel Processor Board Front panel Keypad Interface and LED Control LVDS Data for LCD Backlight Power Control LCD Brightness Control USB Ports Keypad Board The Front Panel Processor board consists of the following circuitries and connectors The state of the front panel keys are read by the one chip processor on the board and any setting changes are reported to the SBC through the COM2 The processor also controls the LED on off state of the front panel LEDs The LVDS data from the SBC is routed through the board to the LCD
98. ns Interface IP RTP UDP Figure 3 25 MTS400 Series TS Compliance Analyzer display RTX100B ISDB T RF Signal Generator Service Manual 3 45 Performance Verification 15 Disconnect the DUT RTX100B MPEG system from the test setup and turn it off using the front panel POWER switch After the Windows shutdown completes unplug the power cord from the system 16 Disconnect the MTS400 Series MPEG system from the test setup a Disconnect the cable to the SFP module b Remove the SFP module c Close all applications running on the MTS400 Series MPEG system and then turn it off using the front panel POWER switch After the Windows shutdown completes unplug the power cord from the system Checking maximum bit rate performance Perform this test to check that the maximum transport stream data is correctly output from the RTX100B instru ment NOTE You can perform the maximum bit rate performance test on MTS400 Series with a Cheetah card only 1 Perform the steps 1 through 4 listed in the Checking Option GbE perfor mance section on page 3 37 2 Set the TS bit rate value to 160 000000 Mbps as shown in the following figure Clock Source Internal ExtRef Ext P Clk Ext S Clk Data Rate 160 000000 IP Data Rate 166 565349 Mbps Fixed ES Rate Default Rate Figure 3 26 Clock dialog box 3 46 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification
99. o 270 MHz into 75 Q load 27 MHz 1 ppm when manufactured 0 5 ppm per year 1 ppm over temperature range BNC 50 Q 10 MHz 27 MHz 8 126948 MHz 160 kHz to 26 75 MHz parallel clock 1 28 MHz to 32 MHz serial clock BNC 75 Q ARIB B31 V1 1 1 2 3 6 MHz 13 Maximum 3 QPSK 16 QAM 64 QAM NOTE DPQSK is not supported If DPQSK modulation is used in a broadcast transport stream the stream is modulated into QPSK However the signal output from the ASI connector is not modulated into QPSK Vitervi 1 2 2 3 3 4 5 6 7 8 Reed Solomon 204 188 0 1 2 4 8 16 1 4 1 8 1 16 1 32 RTX100B ISDB T RF Signal Generator Service Manual 1 7 Specifications Table 1 2 Mainframe Cont Characteristics Description UHF output Frequency 470 to 770 MHz channel plan steps Channel Frequency Channel Frequency Channel Frequency 13 473 143 30 575 143 47 677 143 14 479 143 31 581 143 48 683 143 15 485 143 32 587 143 49 689 143 16 491 143 33 593 143 50 695 143 17 497 143 34 599 143 51 701 143 18 503 143 35 605 143 52 707 143 19 509 143 36 611 143 53 713 143 20 515 143 37 617 143 54 719 143 21 521 143 38 623 143 55 725 143 22 527 143 39 629 143 56 731 143 23 533 143 40 635 143 57 737 143 24 539 143 41 641 143 58 743 143 25 545 143 42 647 143 59 749 143 26 551 143 43 653 143 60 755 143 27 557 143 44 659 143 61 761 143 28 563 143 45 665 143 62 767 143 29 569 143 46 671 143 Frequency MHz Output amplitude Mode 1
100. o the module level Options contains descriptions of factory installed options and field upgrad able options m Diagrams contains the block diagram and interconnection diagram m Replaceable Parts List includes a table of all replaceable modules their descriptions and their Tektronix part numbers RTX100B ISDB T RF Signal Generator Service Manual xiii Preface Manual Conventions This manual uses certain conventions with which you should become familiar Some sections of the manual contain procedures for you to perform To keep those instructions clear and consistent this manual uses the following conventions m Names of front panel controls and menus appear in the same case initial capitals all uppercase and so forth in the manual as is used on the RTX100B front panel and menus Instruction steps are numbered unless there is only one step Bold text refers to specific interface elements that you are instructed to select click or clear Example Press the Menu button to open the File menu W Italic text refers to document names or sections Italics are also used in NOTES CAUTIONS and WARNINGS Example The Diagrams section beginning on page 7 1 includes a block diagram and an interconnect diagram Modules Throughout this manual any replaceable component assembly or part of the RTX100B is referred to generically as a module In general a module is an assembly like a circuit board rather tha
101. omium RTX100B ISDB T RF Signal Generator Service Manual xi Environmental Considerations xii RTX100B ISDB T RF Signal Generator Service Manual SSS SU Preface This is the service manual for the RTX100B ISDB T RF Signal Generator This manual contains information needed to service an RTX100B to the module level Manual Structure This manual is divided into sections such as Specifications and Theory of Operation Furthermore some sections are divided into subsections such as Product Description and Removal and Installation Procedures Sections containing procedures also contain introductions to those procedures Be sure to read these introductions because they provide information needed to do the service correctly and efficiently The following are brief descriptions of each manual section W Specifications contains a description of the RTX100B and the characteristics that apply to it m Theory of Operation contains circuit descriptions that support service to the module level m Performance Verification contains procedures for confirming that an RTX100B functions properly and meets warranted characteristics m Adjustment Procedures contains procedures for adjusting the RTX100B to meet warranted characteristics m Maintenance contains information and procedures for performing preventive and corrective maintenance of an RTX100B These instructions include cleaning module removal and installation and fault isolation t
102. on skip the remaining steps If steps 1 and 2 do not remove all the dust or dirt the RTX100B may be spray washed using a solution of ethyl alcohol by performing steps 4 through 8 Gain access to the parts to be cleaned by removing easily accessible shields and panels see Removal and Installation Procedures Spray wash dirty parts with the ethyl alcohol and wait 60 seconds for the majority of the alcohol to evaporate Use hot 120 F to 140 F deionized water to thoroughly rinse them Dry all parts with low pressure deionized air Dry all components and assemblies in an oven or drying compartment using low temperature 125 F to 150 F circulating air Lubrication There is no periodic lubrication required for the RTX100B RTX100B ISDB T RF Signal Generator Service Manual _ PES ae Removal and Installation Procedures This section contains procedures for removal and installation of all mechanical and electrical modules Preparation the Safety Summary found at the beginning of this manual Also to prevent possible damage to components of the RTX100B read Preventing ESD in this section WARNING Before performing this or any other procedure in this manual read This subsection contains the following items Preparatory information that you need to properly perform the procedures that follow List of tools required to remove and disassemble all modules m Three module locator diagrams for find
103. onnect the DUT RTX100B MPEG system from the test setup and turn it off using the front panel POWER switch After the Windows shutdown completes unplug the power cord from the system Disconnect the MTS400 Series MPEG system from the test setup a Disconnect the cable to the SFP module b Remove the SFP module Close all applications running on the MTS400 Series MPEG system and then turn it off using the front panel POWER switch After the Windows shutdown completes unplug the power cord from the system This completes the RTX100B performance verification RTX100B ISDB T RF Signal Generator Service Manual ey Adjustment Procedures ee a Adjustment Procedures This section contains information needed to adjust the RTX100B The only adjustment in the RTX100B is for the carrier leakage Requirement for Adjustment Before proceeding note the following requirement Warm up Period RTX100B requires a 20 minute warm up time in a 20 C to 30 C environment before it is adjusted Adjustment done before the operating temperature has stabilized may cause errors in performance In addition the spectrum analyzer requires appropriate warm up time to meet the specification Equipment Required Table 4 1 lists the test equipment required to adjust the carrier leakage The table identifies examples of recommended equipment and lists the required precision where applicable If you substitute other test e
104. ooo0o00 75 BNC cable Figure 3 11 Equipment connections for checking the play operation ASI interface 7 Start the MPEG Player application on the MPEG analyzer 8 Set the application to the Record mode 9 Set the MPEG analyzer as indicated below Record Source SPI ASI 310M Target RAM Record Size 100 MB SPI ASI 310M Input Port BNC BNC Port ASI File Save Mode Over write E MTXRTX_Test streams Record_Files ASI214Mbps trp 3 20 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification RTX100B rear panel 10 11 12 13 14 15 On the RTX100B make the following settings Play menu Clock Data Rate 214 Mbps Update Off Source RAM Press the Play Pause button on the RTX100B to start playing the test40 TRP Verify that the hierarchic view is displayed on the MPEG analyzer screen In addition verify that the bit rate is 214 Mbps and the packet size is 188 bytes Click the Record button on the MPEG Player application to record the file After recording is com
105. placement If the RTX100B does not meet performance criteria repair is necessary This manual contains all the information needed for periodic maintenance of the RTX100B This manual also contains all information for corrective maintenance down to the module level To isolate a failure to a module use the fault isolation procedures found in Troubleshooting part of Section 5 Maintenance To remove and replace any failed module follow the instructions in Removal and Installation Procedures also part of Section 5 After isolating a faulty module replace it with a fully tested module obtained from the factory Section 8 Replaceable Parts List contains part number and ordering information for all replaceable modules RTX100B ISDB T RF Signal Generator Service Manual xv Introduction Tektronix Service Offerings Warranty Repair Service Self Service Tektronix provides service to cover repair under warranty as well as other services that may provide a cost effective answer to your service needs Whether providing warranty repair service or any of the other services listed below Tektronix service technicians are well trained to service the RTX100B They have access to the latest information on improvements to the RTX100B as well as the latest new options Tektronix warrants this product for one year from date of purchase The warranty appears after the title page in this manual Tektronix technicians provide warranty service at
106. plane board correct 4 9 V to 5 30 V See Figure 5 23 on page 5 43 Is pressed is the voltage at No pin 10 of J240 about 5 V See Figure 5 24 on page 5 44 15 PWON line at pin 10 No of J170 on the A20 PCI Backplane board about 0v Are all the voltages of No 5 V 43 3 V 412 V and 12 V correct Replace the CPU board Figure 5 18 Troubleshooting procedure 4 RTX100B ISDB T RF Signal Generator Service Manual m Check the 174 5172 00 cable m Check the 5 V main power supply module on the A40 AC Distributor board m Check the 174 5185 00 174 5220 00 and 174 5173 00 cables m Check the On Standby switch on the Ketpad board m Check the 174 5174 00 cable m Replace the CPU board Remove the A12 Main board or the A150 ISDB T RF Output board from the A20 PCI Backplane board 5 39 Troubleshooting Are all the voltages No 5 V 3 3 V 12 V and 12 V correct Change the A12 Main board or the A150 ISDB T RF Output board which affects the power supply Figure 5 19 Troubleshooting procedure 5 5 40 m If the 12 V line is faulty check the 174 5177 00 and 174 5178 00 cables or replace the main power supply m f the other voltage is faulty replace the A20 PCI Backplane board RTX100B ISDB T RF Signal Generator Service Manual Troubleshooting Connect a VGA monitor to the VGA connector on the rear panel
107. pleted press the Stop button on the RTX100B Disconnect the 75 Q BNC cable from the RTX100B and MPEG analyzer Checking the Record Operation 16 0 LH Use the 75 Q BNC cable to connect the ASI In connector on the RTX100B to the ASI SMPTE Out connector on the MPEG analyzer See Figure 3 12 MPEG analyzer rear panel MTS430 P 9 9 75 BNC cable Figure 3 12 Equipment connection for checking the record operation ASI interface RTX100B ISDB T RF Signal Generator Service Manual 3 21 Performance Verification 17 18 19 20 21 22 23 24 25 26 27 28 3 22 Press the Record button on the RTX100B to display the Record screen On the RTX100B make the following settings Record menu Source ASI Record size 100 MB Target RAM File menu
108. quipment for the examples listed the equipment must meet or exceed the listed tolerances Table 4 1 Equipment required Item Qty Minimum requirements Recommended equipment Spectrum analyzer 1ea Resolution bandwidth Hz Agilent Technologies E4402B COM 50 Q SMA cable Mean Systems 5B 010 19 19 1000 50 Q N Fe to 75 N Fe adapter 1ea Bandwidth 2GHzorhigher Agilent Technologies 11852B 50 Q N Ma to SMA Fe adapter Stack Electronics BA057 75 2 BNC Ma to NC Fe adapter Mea o Stack Electronics BA045 RTX100B ISDB T RF Signal Generator Service Manual 4 1 Adjustment Procedures Carrier Leakage Adjustment Procedure 1 RTX100B rear panel Perform the following procedure to adjust the carrier leakage Use the 75 Q BNC Ma to NC Fe adapter 50 N Fe to 75 Q N Ma adapter 50 Q SMA cable and two 50 Q N Ma to SMA Fe adapter to connect the RF Out connector on the RTX100B to the RF Input on the spectrum analyzer See Figure 4 1 Spectrum analyzer E4402B COM 75 QBNC Ma to g NC Fe adapter TNE 50 Q N Fe to 75 Q oo eS N Ma adapter Md Mo
109. quipment required for performance verification Item Frequency counter MPEG analyzer Function generator Oscilloscope ISDB T analyzer Spectrum analyzer MPEG 2 measurement decoder Video monitor Probe 75 Q signal adapter 50 Q BNC cable 50 Q SMA cable 75 Q BNC cable Parallel interface cable IEEE1394b cable 50 Q N Fe to 75 2 N Fe adapter 75 Q BNC Ma to NC Fe adapter 50 Q N Ma to SMA Fe adapter IEEE1394b hard disk drive 3 6 1ea Frequency range 0 1 Hz to 1 25 GHz Agilent Technologies 53181A Option 015 Precision 8 digits or higher tea __ Tektronix MTS430 or MTS400 1 Frequency 40 M clock pattern Tektronix AFG3102 Amplitude 3 V Outputs 2 channel Bandwidth 1 GHz or higher Tektronix TDS5104B Panasonic VP8480A 1 ea ER oo a Resolution bandwidth 10 Hz Agilent Technologies E4402B COM Rohde amp Schwarz DVMD Sony LMD 1420 and BKM 320D SDI input adapter Tektronix P5050 Tektronix AMT75 5 1ea SD SDI video input 5 rm f Bandwidth 1 GHz Amplitude precision 3 dB 1 ea 2 Length 42 inches Tektronix part number 012 0057 01 Candox Systems 5B 010 19 19 1000 Canare DH5C01 S SA Tektronix part number 012 A220 00 supplied with the RTX100B 2ea 5C 2V 1m 1ea 25 pin D type 1ea 9 pin 9 pin 1ea Bandwidth 2 GHz or higher Agilent Technologies 11852B Stack Electronics BA045
110. r frequency settings Center Frequency Center frequency CH MHz MHz 13 473 143 30 575 143 47 14 479 143 31 581 143 8 15 485 143 587 143 4g 16 491 143 593 143 50 17 497143 599 143 18_ 503 148 605143 s 19 509 143 611 143 53 2 515143 37 Jer a 21 143 38 22 527 143 629 143 56 suo 0 650 a 25 545 143 647 143 59 sus 28 563 143 665 143 62 29 569143 etus RTX100B ISDB T RF Signal Generator Service Manual Center frequency MHz 677 143 683 143 689 143 695 143 701 143 707 143 713 143 719 143 725 143 731 143 737 143 743 143 749 143 755 143 761 143 767 143 Adjustment Procedures 4 4 RTX100B ISDB T RF Signal Generator Service Manual IUS Maintenance Maintenance This section contains the information needed to do periodic and corrective maintenance on the RTX100B The following subsections are included Preparation Inspection and Cleaning Removal and Installation Procedures Verifying the BIOS Configuration Troubleshooting System Software Recovery Using the Rescue Disc Related Maintenance Procedures The following sections contain information and procedures related to mainte nance Section 2 Theory of Operation contains a circuit description at the module or block level Section 3 Performance Verification contains procedures that may be useful in isolating problems to modules by testing the
111. rator Service Manual 5 31 Removal and Installation Procedures c Use a screwdriver with a T 10 Torx tip to remove the two screws securing the RFI filter to the chassis d Pull the RFI filter away from the chassis 4 Reinstallation Perform step 3 in reverse order to reinstall the RFI filter e RFI filter Figure 5 14 RFI filter removal 5 32 RTX100B ISDB T RF Signal Generator Service Manual EEE ae Verifying the BIOS Configuration Procedure A This section provides information needed to verify or set the BIOS configura tion CAUTION It is very important to follow this procedure exactly or the instrument will not operate correctly Perform the following procedure to set the BIOS configuration 1 Connect the USB keyboard to the USB connector on the front panel 2 Power on the instrument and then press the Del key immediately The AwardBIOS CMOS Setup Utility screen appears 3 Set the BIOS configuration as follows Use the arrow keys to highlight each item and then press the Enter key to select it Under Standard CMOS Features Drive A Advanced Bios Features First Boot Device Second Boot Device Third Boot Device Advanced Chipset Features Boot Display Panel Number Integrated Peripherals Super IO Device Onboard Serial Port 2 Onboard Serial Port 4 PC Health Status CPU Warning Temperature Shutdown Temperature RTX100B ISDB T RF Signal Generator Se
112. recovery process will not restore the Application software V10 1 and above and the factory installed test streams 1 Restart your system and press the F4 key as soon as the BIOS tests are complete This opens the Phoenix Always screen with the following options m Restore System m System Information 2 Click Restore System gt Phoenix Recover Pro 6 3 Click Advanced gt Factory Restore to replace the Windows XP and RTX100B application software from the backup created when the system was manufactured A confirmation dialog box is displayed 4 Click OK to restart your system and display the Phoenix Recover Pro 6 screen This screen allows you to perform one of the following operations m Click Recover Boot Partition or Recover Entire Drive to overwrite the data according to your requirements A confirmation dialog box is displayed Click Yes to continue RTX100B ISDB T RF Signal Generator Service Manual 5 45 System Software Recovery A screen showing the progress of the recovery appears with the details of the time elapsed time remaining amount of data recovered and total data to recover CAUTION Do not Cancel the recovery process once initiated Cancelling the process will leave your hard disk drive in a partially recovered unstable state Immediately after the data is recovered the system restarts and the Windows XP setup wizard is displayed m Click Cancel and Exit to cancel the recovery process and exit f
113. rom the Phoenix Recovery Pro 6 application 5 Click the Next button of the wizard until Windows is installed The system restarts after the installation of Windows XP to display the RTX100B application screen 5 46 RTX100B ISDB T RF Signal Generator Service Manual EE Using the Rescue Disk The information in this section helps you to reinstall the Microsoft Windows XP and the RTX100B application software using the rescue DVD disks The prerequisites for the recovery procedure are as follows Recovery Procedure RTX100B series instrument Sony DRX 810UL external USB DVD drive Rescue DVD disk Tektronix part number 063 3931 XX USB keyboard USB mouse NOTE The rescue disk recovery procedure works only with MTX100B RTX100B and RTX130B instruments Before you start the rescue disk recovery procedure make sure to check for the BIOS settings Follow these steps to set the BIOS settings for booting from an external USB DVD drive 1 2 4 Press Del when the initial BIOS screen appears Select Advanced CMOS Feature and set it to the following configuration First Boot Device USB_CDROM Second Boot Device USB_HDD Third Boot Device HDD_0 Press F10 to save the CMOS configuration A message asking you to press Y to exit and N to continue is displayed Press Y to exit and restart the instrument Follow these steps to restore the Windows XP and RTX100B application software using the rescue disks 1 In
114. ront panel unit See Figure 5 8 a Use a screwdriver with a T 10 Torx tip to remove the four screws securing the front panel unit to the chassis b Grasp the front panel unit and pull it forward c Disconnect the cable from J100 on the Front Panel Processor board d Disconnect the cable from J105 on the Front Panel Processor board Reinstallation Perform step 3 in reverse order to reinstall the front panel unit p 0 2 50 amp Figure 5 8 Front panel unit removal 5 18 RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Front Panel Assembly 1 Assemble equipment and locate modules to be removed a You need a screwdriver with a T 10 Torx tip b Locate the front panel assembly in the locator diagram External modules Figure 5 2 page 5 11 2 Remove the front cover See Figure 5 9 on page 5 20 a Press the two latches at the bottom of the front cover to release it from the front cover bracket b Pull the front cover forward away from the front panel bracket 3 Remove the Front Panel Processor board See Figure 5 9 on page 5 20 a Disconnect the cables from J11 J301 and J321 on the Front Panel Processor board b Usea screwdriver with with a T 10 Torx tip to remove the four screws securing the Front Panel Processor board to the front panel bracket c Liftthe board away 4 Remove
115. rvice Manual None HDD 0 USB CDROM USB HDD CRT LFP 1024X768 Onboard FDC Controller Disabled Disabled Disabled Disabled Disabled 5 33 Verifying the BIOS Configuration 4 Use the arrow keys to select Save amp Exit Setup and then press the Enter key The Save to CMOS and Exit Y N message appears 5 Enter Y and press the Enter key 5 34 RTX100B ISDB T RF Signal Generator Service Manual Troubleshooting This subsection contains troubleshooting trees designed to isolate faulty modules in the RTX100B Troubleshooting Trees Figures 5 15 through 5 24 show the troubleshooting procedure for the RTX100B RTX100B ISDB T RF Signal Generator Service Manual 5 35 Troubleshooting Press the On Standby power switch on the front panel Does the RTX100B To A page 5 39 power on Does the LCD display light To B page 5 41 Does m Replace the CPU board pred m Replace the hard disk drive normally m Set the BIOS configuration again Does the RTX100B application start To C page 5 42 up normally m Check that the A12 Main board and Arethe version of the A150 ISDB T RF Output board are the A12 Main board and the securely installed into the card slots A150 ISDB T RF Output board displayed in the Status dialog box m Replace the A20 PCI Backplane board m Replace the A150 ISDB T RF Outpu
116. sert the rescue disk 1 MTXRTX 10 1 DSKl iso into the external USB DVD drive RTX100B ISDB T RF Signal Generator Service Manual 5 47 Using the Rescue Disc 5 48 2 Connect the external USB DVD drive to port 1 and the USB keyboard to port 2 of the instrument and then restart After the BIOS screens the CD boots and scans for the USB drive A screen is displayed within a short period prompting you to press 1 to continue and 2 to abort the procedure and turn off the instrument By default option 2 is selected after a delay of 20 seconds 3 Press 1 to proceed as soon as you see the screen NOTE The approximate time taken to restore the Windows XP and RTX100B application is 1 hr to 1 5 hr Press any key to turn off the instrument once the restoration is complete Disconnect the external USB DVD drive from the instrument and restart Phoenix Recover Pro 6 screen is displayed with Recover Boot Partition and Recover Entire Drive options Click Recover Entire Drive to overwrite the entire data A confirmation dialog box is displayed Click Yes to continue A recovery progress screen is displayed with the details of the time elapsed time remaining amount of data recovered and total data to recover The approximate time taken for the process is 1 5 hr CAUTION Do not Cancel the recovery process once initiated Cancelling the process will leave your hard disk drive in a partially recovered unstable state Immedi
117. step 15 Clock source setting RTX100B ExtRef 27 Ext Ref 8 126984 MHz Ext P Clk Output frequency setting function generator 27 MHz 8 126984 MHz 8 MHz 17 Change the output frequency of the function generator to 32 MHz NOTE You have to click the Ext S Clk to select the Ext P Clk 18 Select Play Clock on the RTX100B to open the Clock dialog box 19 In the dialog box select Ext S Clk 20 Repeat step 15 and verify that all items in the Priority 1 row are green RTX100B ISDB T RF Signal Generator Service Manual Performance Verification NOTE Ignore 2 3 a PCR Repetition and 2 5 PTS errors in the Priority 2 row 21 22 23 24 25 26 27 28 29 30 31 32 Press the Stop button on the RTX100B to stop the stream output Select Play gt Clock on the RTX100B to open the Clock dialog box In the dialog box select Internal Press the On button of the Ch1 output on the function generator to stop the signal output Press the On button of the Ch2 output on the function generator to start the signal output Select Play gt Others on the RTX100B to open the Others dialog box In the dialog box set Ext Play Start to Rise Verify that the RTX100B starts playing when the Output menu is set to Invert on the function generator Press the Stop button on the RTX100B to stop the stream output Press the On button of the Ch2 output on the function generator to stop the signal ou
118. t board Figure 5 15 Troubleshooting procedure 1 5 36 RTX100B ISDB T RF Signal Generator Service Manual Troubleshooting Does the IEEE1394b No Replace the A20 PCI Backplane interface work board Can a stream data No be output from the SPI To D page 5 42 In Out connector Can a stream data be No acquired from the SPI To E page 5 43 In Out connector Can a stream data No be output from the ASI Out connector m Check the cable 174 B924 00 EHE AE No m Check that the A150 ISDB T RF Output board is securely installed be acquired from the ASI In connector into the card slot m Replace the A150 ISDB T RF Output board Figure 5 16 Troubleshooting procedure 2 RTX100B ISDB T RF Signal Generator Service Manual 5 37 Troubleshooting Connect an ISDB T receiver to the RF Out connector and play an ISDB T transport stream file Can an PF signal be No output from the RF Out connector Connect a transport stream generator to the ASI In connector and play an ISDB T transport stream file on the generator Can an RF modulated No Replace the A150 ISDB T RF Output signal be output from the RF Out connector Figure 5 17 Troubleshooting procedure 3 5 38 board RTX100B ISDB T RF Signal Generator Service Manual Troubleshooting A Is 5 VSB line at pin 9 of J170 on the A20 PCI No Back
119. t how the bitrate is calculated 9 Prescan first 100 PCRs Prescan all PCRs in file test streams O Manually set Mbps WINDOWS Use timestamps if present My Documents Capture u Settings re Pause analysis at packet index 4H E My Computer My Network Places File name Files of type MPEG Transport Stream Files mpg mpeg trp Figure 3 21 Open Transport Stream dialog box 6 Select the Real time Analysis icon on the left of the Open Transport Stream dialog 7 Select IP from the Interfaces dropdown menu in the Open Transport Stream display as shown in Figure 3 22 RTX100B ISDB T RF Signal Generator Service Manual 3 41 Performance Verification Open Transport Stream Stream Interpretation Base Standard DVB Region No Extensions Using default scripts Scripted validations are not enabled Interfaces Select From the available interfaces Interface Settings Field Yalue to se its Nominal Settings which have been changed have purple background Figure 3 22 Selecting the IP interface 8 After selecting IP in step 7 the Interface Settings pane of the dialog box should change as shown in Figure 3 23 NOTE You may have to click the button below the Configurations list to make the settings pane change The Browse button which is needed for the next step will then be active 3 42 RTX100B
120. t connection for checking the recorded file ASI interface 2 0 0 cc ccc cc cree cree 3 23 Figure 3 14 Fifth equipment connection for checking the ASI interface oi ehe EE CEPI EU E Sie ds HONS 3 24 Figure 3 15 Equipment connection for checking the carrier leakage 3 26 Figure 3 16 Equipment connection for checking the output level and error eere rr e Re n e ea e ow Ini on 3 29 Figure 3 17 Equipment connection for checking the IEEE1394b Interface ey Hey re eR a dere eri vieles 3 31 Figure 3 18 Equipment setup for Option GbE performance verification ede eyes yer x ea eg e Er ie e y n i ons 3 37 Figure 3 19 IP Playout Configuration dialog box 3 39 Figure 3 20 RTX100B MPEG Player display with 40 000000 Mbps 3 40 Figure 3 21 Open Transport Stream dialog box 3 41 Figure 3 22 Selecting the IP 3 42 Figure 3 23 Interface settings 3 43 RTX100B ISDB T RF Signal Generator Service Manual iil Table of Contents Figure 3 24 IP Flow selection Figure 3 25 MTS400 Series TS Compliance Analyzer display Figure 3 26 Clock dialog Figure 3 27 RTX100B MPEG Player display with maximum lr M Figure 3 28 MTS400 Series TS Compliance Analyzer display for maximum bit rate
121. t with covers or panels removed Use Proper Fuse Use only the fuse type and rating specified for this product Avoid Exposed Circuitry Do not touch exposed connections and components when power is present Do Not Operate With Suspected Failures If you suspect there is damage to this product have it inspected by qualified service personnel Do Not Operate in Wet Damp Conditions Do Not Operate in an Explosive Atmosphere Keep Product Surfaces Clean and Dry Provide Proper Ventilation Refer to the manual s installation instructions for details on installing the product so it has proper ventilation RTX100B ISDB T RF Signal Generator Service Manual vii General Safety Summary viii Symbols and Terms A A Terms in this Manual These terms may appear in this manual WARNING Warning statements identify conditions or practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could result in damage to this product or other property Terms on the Product These terms may appear on the product DANGER indicates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to property including the product Symbols on the Product The following symbols may appear on the product A A CAUTION WARNING Protective Ground Ref
122. tage to the fan 7 V to 11 5 V The IEEE1394 interface consists of the 1394b link layer device U110 cable transcriver arbiter U120 oscillator Y100 serial EEPROM U140 regulators U130 and U230 and IEEE1394b connector J140 RTX100B ISDB T RF Signal Generator Service Manual 2 3 Theory of Operation A150 ISDB T RF Output Board Local Bus Interface ASI RX and ASI Decoder ASI Encoder and ASI TX PCI Interface FPGA Channel Coding FPGA OFDM Modulation FPGA Digital PLL FPGA RF DAC RF Carrier Synthesizer Quadrature Modulator 2 4 The A150 ASI ISDB T RF Output board consists of the following blocks The local bus interface communicates with the 12 Main board There are two sets of 16 bit signal lines for Rx and Tx 8 bit data lines for single end connec tion 4 bit control lines for single end connection and 4 bit control lines for differential connection These lines are connected to the 12 Main board in one on one This block changes an ASI signal received from the ASI In connector into parallel transport stream data This block changes parallel transport stream data into an ASI signal and output the signal This block consists of an FPGA of ALTERA It is the interface for PCI and interchange of an incoming transport stream signal It also generates an interrupt signal when an of a transport stream is decoded and Mode is changed This block performs coding of ISDB T It has four synchro
123. tallation Procedures Procedure for Internal Modules Perform the Access Procedure on page 5 14 before doing any procedure in this group The procedures are Hard Disk Drive 1 5 26 Hard disk drive Internal and external fans 12 V main power supply A40 AC Distributor board RFI filter Assemble equipment and locate module to be removed a You need a screwdriver with a T 10 Torx tip b Locate the hard disk drive in the locator diagram Internal modules Figure 5 4 page 5 13 Orient the instrument Set the RTX100B so its bottom is down on the work surface and its left side is facing you Remove the hard disk drive See Figure 5 11 on page 5 27 a Disconnect two cables from the hard disk drive b Usea screwdriver with a T 10 Torx tip to remove the two screws securing the hard disk drive retainer to the chassis c Remove the two screws securing the plate to the hard disk drive d Grasp the hard disk drive and pull it gently away from the chassis Reinstallation Perform step 3 in reverse order to reinstall the hard disk drive RTX100B ISDB T RF Signal Generator Service Manual Removal and Installation Procedures Hard disk drive unit Plate Hard disk drive retainer P di Figure 5 11 Hard disk drive removal RTX100B ISDB T RF Signal Generator Service Manual 5 27 Removal and Installation Procedures Internal and
124. tatus Stopped Stop Pause Resume fou can specify the start parameters that apply when you start the service from here Start parameters nk Cancel Apply 4 Select IIS admin to open the dialog box Check if the Startup type is set to Disabled 115 Admin Properties Local Computer General Log On Recovery Dependencies Service name IIS amp DMIM Display name 15 Admin Description Allows administration of Web and FTP services through the Internet Information Services snap in Path to executable CA WINDOWS system32sinetsrv inetinfo exe Startup type Service status Stopped Sto Pause Resume fou can specify the start parameters that apply when vou start the service from here Start parameters OK Cancel Apply 3 34 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Perform this procedure to check if the automatic Java updates are disabled 1 Click Control Panel gt Java 2 Select the Update tab and check if the Check for Updates Automatically check box is unselected Java Control Panel General Update Java Security Advanced Update Notification version of the Java platform The options below let you control how amp The Java Update mechanism ensures you have the most updated updates are obtained and applied Notify
125. terface Connector type Data rate Description 160 kHz to 25 MHz parallel clock 1 28 MHz to 32 MHz serial clock 0 5 V to 3 0 V BNC 1kQ Rising and falling edges are programmable gt 3 5 V maximum input voltage 7 V 0 8V gt 2 2 V with 50 2 termination 0 8 V with 50 Q termination 50 Q 50 MHz to 100 MHz locked to reference clock 50 MHz maximum serial clock 26 75 MHz maximum parallel clock 214 Mbps maximum 64 Kbps minimum TS clock X 2 Y Z 27 MHz 15362 lt X lt 31248 1686 Y 3376 2 lt Z lt 65536 TS clock X 2 Y Z external parallel clock 214 MHz maximum 15632 lt X lt 31248 1 lt Y lt 16383 2 lt Z lt 65536 TS clock X 2 Y Z external serial clock 8 32 MHz maximum 15632 lt X lt 31248 1 lt Y lt 16383 2 lt Z lt 65536 lt 104 dBc Hz at 21 455707 MHz 20 kHz RBW 300 Hz D sub 25 pin 256 Kbps to 214 Mbps RTX100B ISDB T RF Signal Generator Service Manual 1 5 Specifications Table 1 2 Mainframe Cont Characteristics Pin assignments Output Output level typical Offset Output resistance typical Data delay typical Input Input level typical Input resistance typical Clock pulse width typical Data hold time typical ASI interface Standard conformance Connector type Impedance Data rate Output Number of outputs Output voltage Jitter Rise fall time Return loss 1 6
126. the Windows XP operating system If necessary review the Windows XP documentation Perform the following procedure to set the network parameters for the RTX100B 1 Connect the keyboard and mouse provided with the instrument to the USB connectors on the front panel You can connect them to either of the connectors 2 Select File gt Minimize or File gt Exit to close the RTX100B MPEG player The Windows XP desktop appears 3 Select Settings gt Control Panel from the Start menu to display the Control Panel window 4 Double click the Network Connections icon in the window to display the Network Connections window s Network Connections File Edit Favorites Tools Advanced Help Q Back JI Search lg Folders E Address e Network Connections LANor High Speed Internet Network Tasks fa Create anew er Local rea Connection connection Set up a home or small Office network Change Windows d Firewall settings C ah 1394 Connection See Also i Network gt Troubleshooter Other Places E Control Panel My Network Places B My Documents 3 My Computer lt Figure 3 2 Network Connections window 5 Double click the Local Area Connection icon The Local Area Connection Status dialog box appears RTX100B ISDB T RF Signal Generator Service Manual Performance Verification 1 Local Area Connection Status General Support Connection
127. tput Return Ext Play Start to Off on the RTX100B Disconnect all cables from the RTX100B function generator and MPEG analyzer RTX100B ISDB T RF Signal Generator Service Manual 3 17 Performance Verification ASI Interface This test verifies that transport stream data is correctly played from and is recorded through the ASI interface on the RTX100B Equipment required MPEG analyzer Oscilloscope MPEG 2 measurement decoder Video monitor Two 75 Q BNC cables 75 2 signal adapter test40 TRP file Checking the Output Signal 1 Use the 75 Q BNC cable and the 75 Q signal adapter to connect the ASI Out connector on the RTX100B to the oscilloscope CH1 input See Figure 3 10 Oscilloscope TDS5104B 0 0 0 0 C MONDO RTX100B rear panel 0 0 000000 000 0000 000 00 00000 G oof LU 75 Q signal adapter 9 9 75 BNC cable Figure 3 10 Equipment connection for checking the output signal ASI interface 3 18 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification 2 Set the oscilloscope as indicated below Vertical scale
128. tries The J100 J110 and J120 connectors are PCI slots for 5 V The J130 connector is a PCI slot for 3 3 V and is reserved in the future The J100 connector is the system slot and is used to install the CPU board single board computer SBC The J110 and J120 connectors are used to install the A12 Main board and the A150 ISDB T RF Output board The PWR Button line Pins 9 and 10 of J310 of the SBC is connected to the power switch on the Keypad board through the Front Panel Processor board When the power switch is pressed the SBC causes the PSON signal to go low When the PSON signal becomes low O10 turns on and 12 V main power is also on When the power switch is pressed again its status is sent to BIOS and the PSON signal turns high This circuit includes a 12 V to 5 V DC DC converter a 12 V to 43 3 V converter and a 12 V to 12 V converter There are four connectors in the interconnect circuit J230 J240 J290 and J310 J230 is used to connect the board to the LCD interface on the CPU board J290 is used to connect the board to COM2 on the CPU board J310 is used to connect the board to the hard disk drive LED and reset power button on the CPU board J240 is used to connect the board to the Front Panel Processor board This circuit also has an RS 232C level converter U700 PIC Micro controls the internal fan revolution It gets the temperature data from the sensors U720 and U730 through 12C BUS and controls the supply vol
129. ts 3 8 RTX100B ISDB T RF Signal Generator Service Manual Performance Verification Internal Clock Output This test verifies the internal clock output level and frequency accuracy Level and Frequency Accuracy Equipment Oscilloscope required Frequency counter 50 2 BNC cable 1 Use the 50 Q BNC cable to connect the Trig In Out connector on the RTX100B to the oscilloscope CH1 input See Figure 3 6 Oscilloscope TDS5104B e aqyjoosocg RTX100B _ olez2 gC sos ol o al a O O oe opo O00 3E o 00 50 BNC cable Figure 3 6 Equipment connection for checking the internal clock output level 2 Select Play gt Others on the RTX100B to open the Others dialog box 3 In the dialog box select the Ext Trigger BNC button to open the Ext Trigger BNC dialog box 4 In the dialog box set BNC IN OUT to Output and BNC OUT Selection to 27 MHz RTX100B ISDB T RF Signal Generator Service Manual 3 9 Performance Verification 5 Set the oscilloscope as indicated below Vertical scale 1 V div CH1 Input
130. ubleshooting procedure 7 Figure 5 22 Troubleshooting procedure 8 Figure 5 23 A20 PCI Backplane board view back side Figure 5 24 A20 PCI Backplane board view right side Figure 7 1 RTX100B block Figure 7 2 RTX100B interconnect diagram 3 44 3 45 3 46 3 47 3 48 4 2 5 10 5 11 5 12 5 13 5 14 5 16 5 17 5 18 5 20 5 22 5 27 5 29 5 31 5 32 5 36 5 37 5 38 5 39 5 40 5 41 5 42 5 43 5 43 5 44 RTX100B ISDB T RF Signal Generator Service Manual Table of Contents Figure 8 1 Cabinet and covers Figure 8 2 Internal modules 1 Figure 8 3 Internal modules 2 Figure 8 4 Front panel unit Figure 8 5 Cables RTX100B ISDB T RF Signal Generator Service Manual Table of Contents List of Tables vi Table 1 1 Table 1 2 Table 1 3 Table 1 4 Table 1 5 Table 3 1 Table 3 2 Table 3 3 Table 4 1 Table 4 2 Table 5 1 Table 5 2 Table 5 3 Table 5 4 Functional specifications 1 4 Mainframe esee die ieee A ARES VO rare P n 1 4 Mechanical characteristics 1 12 Environmental characteristics 1 12 Certifications and compliances 1 13 Equipment required for performance verification 3 6 RTXI100B test record
131. ure 8 SPI In Out connector QE min uu e m RS Bou e eco uuu 20000008 e s O e e e E gt Front O O ge J220 TN J170 Figure 5 23 20 Backplane board view back side RTX100B ISDB T RF Signal Generator Service Manual 5 43 Troubleshooting 3330 J30 i U740 305 lt 12 O UO R740 C316 han 9 C130 S 35 7 8 os z Syz Lu 3 e Efe m t ELIZO 3 is is C232 o o o 5 a o 5 29 c PT D ELE Ho cg cossSo8 1 MHS 4 loezn 1 gt on o SG gue ena OS z aes O 5 MODON st e EOD 5 82 8090050550 ANT 5 to BRERERERERE R220 R153 C133 zOccOcacOox MH35 d N Front 5 C150 C151 ci52 o U150 ron s os 5 M M Une o o o c711 T a E 5575 n E E ado Cos Mon RS ag egg C160 C161 162 C153 xz f aS i ee n MER SM 7 e Je TRA TM MH4 as C712 2d 2 4 d E

Download Pdf Manuals

image

Related Search

Related Contents

USER MANUAL - convertermodem.net  ASUS P2E User's Manual  OM, Non-Branded, BKS4615, 952802353, 2010-10, Chain  2 - Vermeiren  Vink hydromatic footbath Installation manual and - vink  T'nB USLPK7  This guide can help you understand rechargeable  Race Tools Installation: Kawasaki Type 4  Massive 45952/29/17  CE EMC TEST REPORT  