enTouch™ AEM® Handle Assembly
Contents
1. Re ony Federal USA law restricts this device to sale by or on the order of a physician Consult Instructions for Use 3 of 4 00209 018 2011 08 Made in USA neal Manufactured by ENCISION Inc 6797 Winchester Circle Boulder CO 80301 USA Ph 303 444 2600 Fax 303 444 2693 Printed in USA Copyright 2011 Encision Inc 4of 4
2. during laparoscopic procedures only AEM instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor Scissors Inserts are intended for use on soft tissue only Contraindications These instruments are not intended for use when laparoscopic electrosurgical techniques are contraindicated Instructions For Use Prior to Use Thoroughly read these instructions and the instructions in the AEM Monitor Operator Service Manual Reusable components are supplied non sterile Refer to Reprocessing Instructions prior to use Disposable inserts and sheaths are supplied sterile Inspect the instrument for proper assembly and function See End of Life Indicators ENCISIGN AEM System Setup See laminated Setup Sheet 00701 when using the ES9005 series AEM Cord Adapter and 02678 when using the ES9015 Universal Adapter e Laparoscopic surgery may result in gas embolism due to insufflation of gas into the abdomen e Pulling or tugging cords can result in breaking of internal conductors causing intermittent alarms or sparking and burning of insulation during use The AEM Monitor is not designed to detect or prevent an arc in the event of a broken active wire e Keep electrical connections dry while in use to prevent potential conduction of HF current to the user e Damaged external insulation AND incorrect setup of the AEM Monitor may result in a risk of unint
3. 4 AEM Handle Assembly Instructions For Use Care Surgical Instruments Detergent wash with hot tap water for 1 20 minutes Hot water rinse for at least 20 seconds Thermal rinse 80 C to 98 C 176 F to 208 F for 1 5 minutes Lubricant may be added during this cycle Hot air dry use low temperature setting for 0 10 minutes NOTE e The washer disinfector manufacturer s instructions should be strictly adhered to Lubrication After cleaning all the components but before sterilization lubricate the shaft and all moving parts of the jaw insert with water soluble medical instrument lubrication After lubrication prepare the instrument for sterilization by wrapping in sterile wrap as appropriate NOTE e Failure to lubricate may result in increased wear or binding 00209 018 2011 08 Sterilization Monitoring sterility using Geobacillus stearothermophilus spore strips is recommended NOTE e Do not exceed temperatures of 135 C 275 F Performance to specification has not been verified above this temperature Damage to the product may occur e Validation of sterilization cycles other than steam autoclave is the responsibility of the end user Description 1 Allow the instrument to dry thoroughly prior to Sterilization See Caution above regarding removal of detergent solution from the connector surface using an alcohol swab prior to Sterilization Steam Sterilization assembled o
4. any of the following are evident e Intermittent electrical performance e Bent electrode shaft or tip affecting function of sheath or insertion into trocar e Any insulation damage which exposes metal AEM Shield along the length of the instrument s shaft and or any insulation damage in Zone 1 e Binding or impaired mechanical functions Express Warranty ENCISION hereby warrants to Buyer that products purchased hereunder shall be free from defects in materials and workmanship under normal use and service as specified in these Instructions for Use Care for the period of six months from date of purchase or upon reaching an end of life indicator whichever occurs first Any evidence of unauthorized modification or repair of this device will void the warranty See AEM Monitor Operator Service Manual for details of Limitations Disclaimer and Exclusions Return of Used Product If for any reason this product must be returned to ENCISION a returned goods authorization is required prior to shipping Appropriate return instructions may be obtained from ENCISION Product ENCISION reserves the right to amend modify or to change any product to introduce new products to withdraw products and otherwise vary product specifications at any time without notice US Patent Nos 5 769 841 6 494 877 ENCISION and AEM are registered trademarks of ENCISION Inc KLENZYME and MANU KLENZ are registered trademarks of Steris Corporation
5. AEM Handle Assembly Instructions For Use Care Re onty fre ENCISION recommends placing this document in the Instructions for Use Care Section of your AEM Monitor Operator Service Manual Device Description Modular Instruments The ES8000 ES8000L ES8200 and ES8200L Handle Assemblies are instruments designed for use with reusable and disposable AEM Inserts Most inserts fit standard 5 5mm trocar cannulas Inserts with a larger or smaller cross section are identified in the catalog description The instruments are true 5mm and are sized for use with the Disposable Sheath ESO150A or ES0150 45 Principle of Operation The front and rear handles open and close the working tip jaw blade insert of the instrument The working tip of the instrument should always be closed when introducing or removing the instrument from the cannula Cord Connection Rotation Knob S Scissors Insert al Shielded Barrel Assembly Insert Inside Handle Trigger NOTE e This product is rated to 9000 Vp p Limit electrosurgical generator power setting to 80 Watts or lower 60 Watts for the Conmed Aspen Excalibur spray mode Higher settings may result in spurious insulation failure alarms and or insulation breakdown See Encision AEM Monitor Operator Service Manual for list of compatible electrosurgical generators e See also Instructions for Use Care for AEM Instruments Disposable Sheath and Disposable Scissors as
6. applicable e The ES8000 ES8000L ES8200 ES8200L handles are not compatible with 35cm or 45cm lengths of the following model number inserts ES0502 ES0517 ES0518 ES0519 ES0520 ES0529 and ES0545 e Discard on evidence of End of Life Indicators e This product is supplied non sterile and must be sterilized prior to use 00209 018 2011 08 Use with Monopolar Electrosurgery AEM instruments in conjunction with an AEM Monitor properly connected to the electrosurgical generator ESU continuously monitor and dynamically manage stray energy insulation failure and capacitive coupling in zones 2 amp 3 which are likely out of the surgeon s field of view AEM shielding does not cover zone 1 which the surgeon should keep in view during instrument activation As in all applications misapplied electrosurgical energy remains the responsibility of the attending surgeon Zone 1 Zone 2 Zone 3 Note Zone 1 equals approximately 1 6cm Tip to Shield Note Zone 3 equals area of Trocar Cannula Laparoscopic procedures should be performed only by surgeons having adequate training and familiarity with laparoscopic techniques and who are also knowledgeable about anatomy and pathology as well as the complications hazards risks and benefits of the procedure Indications Intended Use These AEM instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy
7. ater minimum of 3 minutes Immerse all components in MANU KLENZ or equivalent detergent solution prepared per manufacturer s instructions and clean all surfaces Using a hand held soft bristle brush use a back and forth motion and brush all surfaces with particular attention paid to Cord Connector crevices grooves fittings and Hold the instrument as shown below with trigger ends touching Using a long handled soft bristle brush with a gauge slightly larger than the lumen of the instrument recommended size 8Fr 2 7mm brush the inner lumen of the rear part of the handle using complete strokes while immersed Push the brush all the way through the lumen and pull it through several times If necessary repeat the process on the other end of the lumen as shown below Flush lumen at end using appropriate sized syringe Remove the device from the detergent solution and rinse thoroughly under running distilled or demineralized water minimum of 3 minutes Combination Manual Automated Cleaning Follow Manual Cleaning steps 1 9 Use of a sonicator or ultrasonic cleaner at 35 45 kHz can support manual cleaning of devices Cycle time 5 minutes maximum Do not exceed water temperatures above 50 C 122 F Place instruments in a suitable washer disinfector basket and process through a standard instrument washer disinfector cleaning cycle See Table listing typical cycles 2 of
8. e the insert A into the handle shaft until the trigger catches and rotates slightly upward DO NOT FORCE TRIGGER UP IT WILL ROTATE AUTOMATICALLY AS INSERT ENGAGES 2 Rotate the insert tip clockwise to screw in the insert DO NOT USE OPEN BLADES TO TIGHTEN SCISSORS 3 Turn the rotation knob to adjust the positioning of the insert Disassembly Disposal Disassemble in reverse order from assembly No further disassembly is recommended If using Disposable Sheath see also Disposable Sheath Instructions For Use Care NOTE e Used instruments are considered medical waste Dispose of in accordance with local regulations 00209 018 2011 08 Reprocessing Instructions After use the instrument must be disassembled cleaned lubricated and sterilized before reuse This product has been validated for 10 sterility assurance level when reprocessed in accordance with these cleaning and sterilization instructions CAUTION e Treat a used instrument as a potential biohazard until cleaning and sterilization has been completed Microscopic residues may remain after cleaning Stainless Steel is rust resistant not rust proof Instruments should be kept dry when not in use e Donot use bleach sodium hypochlorite based products during cleaning Bleach is extremely corrosive to metals and can negatively affect the electrosurgical instruments Refer to your cleaning products Material Safety Data Sheet MSDS to ensure t
9. ended patient burn shock or fire hazard Do not use product having damaged insulation CAUTION e Good operating room practice suggests that connections of accessories to electrosurgical generators be made only while the generator is Off or on Standby e Use these instruments only in conditions that assure adequate visualization to minimize risk of misapplied electrosurgical energy e Keep ESU power setting as low as possible for the intended purpose to minimize unintended burns e Damaged internal insulation of the cord and or instrument or loss of shield continuity may cause ESU return pad alarms triggered by the AEM Monitor s Fault Indicators For maximum patient safety discontinue use of the instrument if this occurs e A singular AEM instrument must be the sole conductor of energy to tissue Do not conduct energy by touching an AEM instrument to a second instrument contacting tissue The second device will not be protected from capacitive coupling and insulation failure e Keep electrosurgical instruments away from the patient and operative field when not in use Accidental activation can result in unintended injury to the patient e See electrosurgical generator manual and AEM Monitor Operator Service Manual for precautions concerning the general application of electrosurgical equipment 1 of 4 Assembly Instructions Handle and Insert Assemble the two 2 basic components 1 Hold the instrument upside down as shown Slid
10. hat they are not corrosive or harmful to various metals including stainless steel gold etc Use of some detergent solutions may leave a surfactant residue on the gold connector surface depending on the strength of the solution and how thoroughly the residues are rinsed off This may result in intermittent cord alarms The residue may be removed by use of an alcohol soaked swab rotated completely around the external gold connector surfaces Certain cleaning chemicals may negatively affect metals such as the gold connectors Refer to your cleaning products Material Safety Data Sheet MSDS to ensure that they are not corrosive or harmful to various metals including stainless steel gold etc Cleaning Automated washer disinfector systems are not recommended as the sole cleaning method for surgical instruments An automated washer disinfector system may be used as a follow up to manual cleaning but is not required NOTE e The use of fully distilled or demineralized water is preferred for cleaning and rinsing Manual Cleaning When using a disposable sheath remove and discard the sheath prior to cleaning 2 Immerse all components in KLENZYME or equivalent blood dissolving enzymatic solution prepared per manufacturer s instructions for at least 5 minutes and gently agitate Soak longer if proteinaceous material is present 1 3 Remove the device from the enzyme solution and rinse thoroughly under running tap w
11. r disassembled use one of the methods listed below Prevac 2 layers of Sterile Wrap or Unwrapped Temperature 132 C 270 F minimum Duration 4 minutes minimum Gravity Steam 2 layers of Sterile Wrap Temperature 132 C 270 F minimum Duration 15 minutes minimum Gravity Steam FLASH Unwrapped Temperature 132 C 270 F minimum Duration 10 minutes minimum Flash Sterilization is not recommended except in emergency situations CAUTION e Cold Soak Sterilization is not recommended Testing results indicate that cold soak sterilization Glutaraldehyde is not adequate for this product Manufacturer s recommended cycle times DO NOT always provide a 10 sterility assurance level e Gas Plasma Hydrogen Peroxide Sterilization is not recommended for ENCISION Instruments Instruments with long narrow lumens may pose a challenge for this type of sterilization Product Life The life of this surgical instrument is largely dependent on the care and handling at the point of use cleaning sterilization For optimal instrument life protect it from contact impact with other instruments during decontamination and sterilization The number of uses may be reduced by improper handling and orevac or flash sterilization methods Electrification of scissors inserts may reduce their useful life ENCISIGN End of Life Indicators Visually inspect and electrically test prior to use Discontinue use if
Download Pdf Manuals
Related Search
Related Contents
Inyector PoE+ Gigabit de Alta Potencia, 1 puerto ダッシュ Aqua-Pure AP902/AP903 Whole House Water Fitlers Instr. mode d`emploi et garantie 34 - Benvenuti nel sito di Domenico Pannullo Tiveco TM-AR346 storage enclosure Videoproiettore laboratorio informatica SBE 25plus Manual - comm 01. Content Management & Publishing Copyright © All rights reserved.