MAC™ 1600 - Frank`s Hospital Workshop

image

Contents

1. Features amp functions cence eee AA Fu Specifications irimi oe Y aa lad a DYA WY Db rd an AR AR Electromagnetic compatibility Electromagnetic compatibility EMC gt cece eee eee eee ees Guidance and manufacturer s declaration electromagnetic emissions Guidance and manufacturer s declaration electromagnetic immunity B 4 Guidance and manufacturer s declaration electromagnetic immunity B 5 Recommended separation distances 2200002222 B 7 EMC exceptions disclosure 0 00 cece cece eee Y aes B 8 EMC compliant supplies amp accessories C 1 Introduction 6 dati Are Aan iy eee ae C 2 Standard accessorieS 00 cece eee e FY EE A EE A uu C 3 Vallie aCCOSSOMOS i ai Sonini Du gwd te Rone AF ee ead ee A a C 4 Thermal papers 4 C 5 Contrast Papers oor sn a Sess bee AG A Ae C 5 Premium Papers 58 eee Gn Fe C 5 Archivist 25 Paper for Europe 0 FY Y FFY eee eee es C 5 Archivist 25 Paper for US 0 2 0 0 0000 e Y Y FF Y Yd C 5 Country specific power Cord FY Y EF AR Eu AU C 6 Optional accessories YFF A EE A EE A uu C 7 MAC 1600 ECG Analysis System Revision A 2028451 183 1 1 Introduction MAC 1600 ECG Analysis System 2028451 183 1 Revision A Introduction Manual information Each page of the document has
2. FRU replacement procedures High level FRU identification Preparing system for FRU replacement Replacing the patient cable Replacing barcode reader or barcode reader cable Replacing the paper tray assembly Replacing the battery cee aes Replacing the keyboard Replacing the display assembly Replacing the internal modem option Replacing the power supply assembly Replacing the acquisition board assembly MAC 1600 ECG Analysis System 2028451 183 Replacing the KISS pump assembly option gt 0c eee eee 4 17 Replacing the printer assembly VYY FFY EY REU 4 19 4 24 Replacing the printhead Y YY YY Y Y RR Y RG 4 26 Replacing the optical sensors 000 cece YY Y FR Y RG 4 30 Replacing the mainboard ETE module assembly 4 33 Functional checkout Y FY EEF A a EE Lu Aud 4 38 Visual iNSPECHON ue sus kae dg i E RG GYN YG RY 4 39 Functional checkout procedures YY YY RY RG 4 40 Electrical Safety Checks VVYY FY FFF EF ee 4 41 Updating software 7 4 42 Conditioning the battery pack VY YY EE ELE Rd 4 44 5 Parts lists wate YY O e O O OO HF 5 1 Ordering Parts ena a 5 5 2 General information 000
3. From the Main Menu press F5 to select System Configuration Press F5 to select More Press F5 again to select More Sy 9 ge a Press F5 to select Service Setup A window opens prompting you to enter the service password Enter the Service password 7 Type prod and press F6 to select OK NOTE If the keyboard does not include the letters prod type 7763 and press F6 to select OK MAC 1600 ECG Analysis System Revision A 2028451 183 4 42 Maintenance Updating software The window shown in the following figure opens 8 Move the focus to the Software Update button and select by pressing the Enter key The window shown in the following figure opens 9 Select the appropriate key To cancel press F5 to select No No further action is reguired if you select No To continue press F6 to select Yes The following window opens 10 Ifthe previous WARNING window appears press F6 to select OK and continue with the software update Revision A MAC 1600 ECG Analysis System 4 43 2028451 183 Maintenance Conditioning the battery pack Conditioning the battery pack To maintain the storage capacity of the battery pack installed in your MAC 1600 GE recommend that you condition your MAC 1600 battery pack once every six months to reset the electronic fuel gauge inside the battery A condition cycle consists of an uninterrupted charge discharge charge cy
4. Asia Headguarters GE Medical Systems Information Technologies Asia GE China Co Ltd 11th Floor Shanghai MAXDO Center 8 Xing Yi Road Hong Oiao Development Zone Shanghai 200336 People s Republic of China Tel 86 21 5257 4650 Fax 86 21 5208 2008 ee rer GE Medical Systems Information Technologies GmbH Munzinger Stra e 3 5 D 79111 Freiburg Germany Tel 49 76145 43 20 Fax 49 761 45 43 233 GE Medical Systems Information Technologies Inc 8200 West Tower Avenue Milwaukee WI 53223 USA Tel 1414 355 5000 1 800 558 7044 US Only Fax 1414 555 0 www gehealthcare com
5. Replacing the power supply assembly steps 1 through 8 beginning on page 4 13 2 Using the following figure identify the acguisition board assembly A 3 Disconnect the gray USB connector from the acquisition board as shown in the following figure MAC 1600 ECG Analysis System 4 15 2028451 183 Revision A Maintenance FRU replacement procedures Lift the acguisition module assembly until you can access the black harness connector on the other side of the acguisition module If the system is eguipped with the optional KISS pump disconnect the gray hose from the luer fitting Remove the acguisition board assembly Reassemble the acguisition board assembly If the system does not have the KISS option go to step 9 If the system has the KISS option go to step 10 Use the following figure for reference KISS pump no KISS pump Reassemble acquisition board assembly systems with no KISS option a Place the label item 36 in the recessed hole of the plastic bezel b Reverse the disassembly steps to reassemble Reassemble acquisition board assembly systems with KISS option a Reassemble items 37 38 and 39 into the plastic bezel as shown above b Tighten the plastic nut lock item 38 with a small adjustable wrench c Hand tighten the luer plug item 39 onto the luer connector item 37 d Reconnect the gray hose to the luer connector item 37 e Reverse the disassembly steps
6. 1600 ECG Analysis System 4 37 2028451 183 Maintenance Functional checkout The checkout procedures apply to all MAC 1600 systems Functional checkout NOTE The FRU checkout procedure for any listed FRU also applies to its internal PCBs and components Perform the applicable product or product configuration dependant procedures when an asterisk is listed FRU replacement procedures are contained within this chapter of the manual Functional Checkout Procedures 1 2 3 1 2 3 9 1 2 3 8 1 2 3 7 1 2 3 4 5 6 13 1 2 3 4 5 6 13 1 2 3 12 1 2 3 15 1 2 3 4 1 2 3 15 middle plastic only 1 2 3 1 2 3 15 when AC power mains disturbed Functional Checkout Procedures 1 2 9 9 14 1 2 3 1 2 9 9 14 1 2 3 5 1 2 3 6 Revision A Table 2 Basic System FRU Repairs Visual Inspection Visual Inspection 7 8 7 8 8 8 7 8 FRU Description Patient Cable Printer Assembly Keyboard Display Module Mainboard ETE Assembly Acquisition Assembly Battery Pack Power Supply Assembly SD Card Plastics AC Power Cord Non listed FRUs Table 3 Optional System FRU Repairs FRU Description Internal Modem KISS Pump External Modem LAN Option Barcode Reader MAC 1600 ECG Analysis System 2028451 183 4 38 Functional Checkout Procedures
7. Table 19 MAC 1600 Keyboards Non Stress Description FRU MAC1600 KEYBOARD ASSEMBLY ENG FRU MAC1600 KEYBOARD ASSEMBLY GER FRU MAC1600 KEYBOARD ASSEMBLY FRE FRU MAC1600 KEYBOARD ASSEMBLY SPA FRU MAC1600 KEYBOARD ASSEMBLY SWE FRU MAC1600 KEYBOARD ASSEMBLY ITA FRU MAC1600 KEYBOARD ASSEMBLY JAP FRU MAC1600 KEYBOARD ASSEMBLY DUT FRU MAC1600 KEYBOARD ASSEMBLY NOR FRU MAC1600 KEYBOARD ASSEMBLY DAN FRU MAC1600 KEYBOARD ASSEMBLY CZE FRU MAC1600 KEYBOARD ASSEMBLY CHN FRU MAC1600 KEYBOARD ASSEMBLY HUN FRU MAC1600 KEYBOARD ASSEMBLY POL FRU MAC1600 KEYBOARD ASSEMBLY RUS FRU MAC1600 KEYBOARD ASSEMBLY SLO FRU MAC1600 KEYBOARD ASSEMBLY POR FRU MAC1600 KEYBOARD ASSEMBLY KOR FRU MAC1600 KEYBOARD ASSEMBLY FIN MAC 1600 ECG Analysis System Part Number 2035699 001 2035699 002 2035699 003 2035699 004 2035699 005 2035699 006 2035699 007 2035699 008 2035699 009 2035699 010 2035699 01 1 2035699 013 2035699 014 2035699 015 2035699 016 2035699 017 2035699 018 2035699 019 2035699 020 2028451 183 Keyboards non stress Revision A Revision A Parts lists Field replaceable units FRUs MAC 1600 Keyboard Stress Table 20 MAC 1600 Keyboards Stress Description FRU MAC1600 KEYBOARD ASSY W STRESS ENG FRU MAC1600 KEYBOARD ASSY W STRESS GER FRU MAC1600 KEYBOARD ASSY W STRESS FRE FRU MAC1600 KEYBOARD ASSY W STRESS SPA FRU MAC1600 KEYBOARD ASSY W
8. This information will be useful when inserting the replacement modem 5 Using the small flat blade screwdriver gently lift each end of the modem to release its pins from their sockets 6 Using your fingers lift the modem and remove it from the device 7 Reassemble the internal modem by reversing the steps for removal a Use the figure in step 4 of this procedure to orient and position the modem in the hole b Visually align the pins on the modem with the sockets on the mainboard c Push the modem to seat it in the sockets d Ifthe gaps between the modem and the hole do not match the figure in step 4 remove the modem and repeat this step until the modem is properly seated as shown in step 4 8 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 4 12 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures Replacing the power supply assembly 1 Disconnect the system from AC power 2 Remove the paper tray as described in Replacing the paper tray assembly 3 Remove the battery as described in Replacing the battery 4 Turn the unit over and loosen the eight screws A shown in the following figure e oe moe 5 Pull the base plastic housing A away from the middle plastic housing B Push inward on the AC power connector C while separating the two assemblies A Revision
9. 2 Insert mainboard into main plastic 3 Replace the two back mounting screws that were removed in step 11 of the previous section 4 Reconnect the three cables that were disconnected in step 11 of the previous section 5 Ifthe KISS option is installed do the following a Replace the KISS pump standoff b Replace the KISS pump c Route the two the KISS pump hoses to the back of the unit and connect to the rear connector panel 6 Replace the rear connector panel 7 Re insert the power supply in the slotted guides while reconnecting the power supply connector 8 Reconnect the long ground wire to the AC outlet pink connector 9 Route the keyboard cable back through the keyboard cable hole 10 Ifthe internal modem option is installed replace the internal modem 11 Reassemble the display 12 Replace the keyboard 13 Replace the keyboard bezel 14 Reattach the base plastic A to the middle plastic B Push the AC power connector C in as shown A 0 5 Replace the ten screws 6 Replace the battery 7 Replace the paper tray 18 Power up the system and install the software from the SD card that shipped with the mainboard See Updating software on page 4 42 for detailed procedure 4 36 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures 19 Restore system setups that were saved to SD card before the new mainboard was in
10. 5 Perform the Print Head Test Revision A MAC 1600 ECG Analysis System 3 13 2028451 183 Troubleshooting Performing diagnostic tests a Select the Print Head Test button The writer prints a 1 page print head test report b Verify that there are no gaps in any of the lines printed Up to 5 mm of blank paper is allowable at the top and bottom of the page The following window opens Did the test pass c Ifthere are no gaps in the lines on the printed report press F4 to select Yes If there are gaps in the lines on the printed report press F5 to select No 6 When all writer tests have been performed press F6 to select Cancel and close the window 7 If one or more of the writer tests fail use the following table to determine how to proceed Observed Failure Remedy Improper paper queuing Replace queue sensors See Replacing queue hole sensor on page 4 30 See Replacing queue mark sensor on page 4 32 Missing dots Replace the printhead gaps in printed report m See Replacing the printhead on page 4 26 Faded prints Faulty roller Replace paper tray assembly 1 See Replacing the paper tray assembly on page 4 6 Printhead pressure problem Replace springs Incorrect print speed Faulty stepper motor Replace paper tray assembly See Replacing the paper tray assembly on page 4 6 Replace printer board See Replacing the printer boar
11. co N co N Acquisition Error 11 Acquisition firmware file not found Reload application software 3 22 MAC 1600 ECG Analysis System Revision A 2028451 183 3 23 Replace Print Head MAC 1200 WRITER SHEC TPH PN Troubleshooting Error codes Printer error codes Table 3 Printer Error Codes Solution Replace Printer Assembly FRU PN 2035702 001 2036817 001 Replace Printer Assembly FRU PN 2035702 001 N Check cables Replace if necessary Replace Printer Assembly FRU PN 2035702 001 N Check cables Replace if necessary Replace Printer Assembly FRU PN 2035702 001 N Check cables Replace if necessary Replace Printer Assembly FRU PN 2035702 001 N gt lt Check cables Replace if necessary Replace Printer Assembly FRU PN 2035702 001 Na Check cables Replace if necessary Replace Printer Assembly FRU PN 2035702 001 N Cause Printer firmware detected an error Printhead temperature is too hot or too cold to print Printer driver could not be opened Printer driver communication error Printer driver timeout error Printer driver miscellaneous error Undefined printer status was received MAC 1600 ECG Analysis System 2028451 183 Error Code Printer Internal Error 1 Printer Internal Error 2 Printer Internal Error 3 Printer Internal Error 4 Printer Internal Error
12. MODULE Revision A MAC 1600 ECG Analysis System 3 21 2028451 183 Troubleshooting Error codes Table 2 Acguisition Error Codes Continued Error Code Cause Solution Acguisition Error 4 Checksum error in 100ms ECG 1 Check cables Packet 2 Replace CABLE ASSY MAC 1600 ACQ TO MAINBOARD found in Cable Harness Kit PN 2035707 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE FF co Acquisition Error 7 Acguisition self test error Check cables Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 mo gt Acquisition Error 8 Error loading the acguisition firmware Reload application software Replace CABLE ASSY MAC1600 ACO TO MAINBOARD found in Cable Harness Kit PN 2035707 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE Replace CABLE ASSY MAC1600 ACG TO MAINBOARD found in Cable Harness Kit PN 2035707 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 Replace 2035704 001 ASSY MAC 1600 MAINBOARD AND ETE MODULE Acquisition Error 10 Acguisition driver communication Replace CABLE ASSY MAC1600 ACG TO MAINBOARD found error in Cable Harness Kit PN 2035707 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE N AA Acquisition Error 9 Acquisition driver failed to open
13. PWR SPLY CRD RA INDIAN 10A 250V 2 5M PWR CRD ST CHINA RA PLUG 10A 250V 2 5M PWR SPLY CRD ST RA PSE 10A 250V 10FT MAC 1600 ECG Analysis System Part Number 405535 006 401855 001 401855 002 401855 003 401855 004 401855 010 401855 107 401855 108 401855 018 405535 011 2028451 183 Power cords 5 20 Technical specifications MAC 1600 ECG Analysis System A 1 2028451 183 A evision Technical specifications Features amp functions Features amp functions Acquisition amp Analysis Analog acquisition module m Marguette 12SL interpretation m Marguette 12SL Hookup Advisor 4 88 micro V resolution Processing m 12 lead simultaneous analysis m 500 samples sec or 1000 samples sec Display 7 diagonal viewing display 16 9 aspect ratio m 800 x 480 pixels m Color m VGA resolution m VGA out port for optional Stress testing Memory m Internal 100 ECG option Printer m 210mm thermal array print head Resolution Vertically 8 dots mm horizontally 1000 dots 25mm 25mm sec m Paper type A4 amp US letter Z fold Writer speed 5 12 5 25 50 mm sec Connectivity Secure Digital card LAN transmission to GE ECG management Internal modem option Support for optional USB barcode GE supported Transmit acquired ECG to CardioSoft system option Transmit acquired ECG to MUSE system option K
14. lead wires cables or problems in the unit A thorough visual inspection of the eguipment can save time Small things disconnected cables foreign debris on circuit boards missing hardware loose components can freguently cause symptoms and eguipment failures that may appear to be unrelated and difficult to track NOTE Take the time to make all the recommended visual checks before starting any detailed troubleshooting procedures Table 1 Visual Inspection List Look for the following problems Fraying or other damage Bent prongs or pins Cracked housing Loose screws in plugs Excessive tension or wear Loose connection Strain reliefs out of place Moisture dust or debris top and bottom Loose or missing components Burn damage or smell of over heated components Socketed components not firmly seated PCB not seated properly in edge connectors Solder problems cracks splashes on board incomplete feedthrough prior modifications or repairs Loose wires or ground strap connections Faulty wiring Wires pinched or in vulnerable position Loose or missing screws or other hardware especially fasteners used as connections to ground planes on PCBs Faulty wiring especially AC outlet Circuit not dedicated to system Power source problems can cause static discharge resetting problems and noise Cuts or cracks in keyboard membrane Illegible labels Scratched or opaque display filter transparent component of keyboard bez
15. 1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 7 8 9 10 11 12 15 13 14 Maintenance Functional checkout Table 4 Non FRU Repairs Description Visual Inspection No parts replaced 4 5 7 8 Software Update nla Hardware Upgrade 7 8 Annual Electrical Safety 1 3 4 5 Checkout Inspect the following for excessive wear and or any visual signs of damage Check for broken patient cable leadwires and out of date electrodes Patient cable leadwires passed inspection Electrodes passed inspection Discuss electrode placement skin prep and patient related requirements with ECG technician Customer following the procedures recommended in the operator s manual See Chapter 3 Preparing the Patient in the MAC 1600 ECG Analysis System Operator s Manual for more information Keyboard LCD display filter passed inspection See Visual inspection on page 3 3 for more information AC power cord passed inspection See Visual inspection on page 3 3 for more information Battery pack passed inspection See Visual inspection on page 3 3 for more information SD card passed visual inspection See Visual inspection on page 3 3 for more information All harnesses and internal wiring have been secured See Visual inspection on page 3 3 for more information Fasteners fasteners replaced and secured See Visual inspection on page 3 3 for more information
16. 92 X 2 4 MM MAC 1600 ECG Analysis System 2028451 183 Item a om o 10 11 12 13 14 15 18 19 20 21 22 Fastener hardware FRU kit PN 2035708 001 5 14 Parts lists Field replaceable units FRUs NOTE Items in the kit are not available as individual FRUs They can be obtained by purchasing the entire FRU kit Items shown above are not drawn to relative scale Figures are provided for reference only MAC 1600 ECG Analysis System 5 15 2028451 183 Revision A Parts lists Field replaceable units FRUs Middle and base plastic FRU kit PN 2036912 001 Table 17 Middle and Base Plastic FRU Kit PN 2036812 001 Item Gty BASE PLASTIC 1 MIDDLE PLASTIC 1 PAD FLOPPY MAC SERIES 1 FERRITE RECT CORE 15X10X23MM 2270H 10MHZ 1 Plastics kit PN 2035709 001 Table 18 Plastics Kit PN 2035709 001 Item Description Gty PAPER TRAY 1 TRAY SPACER 1 MOTOR COVER 1 ROLLER HOLDER 1 PRINTHEAD HOLDER LEFT 1 PRINTHEAD HOLDER RIGHT 1 TRAY BUTTON 1 REAR CONNECTOR PANEL 1 BEZEL ACO BOARD 1 CONNECTOR COVER 1 NOTE Items in the kit are not available individually They can be obtained by purchasing the entire FRU kit 5 16 MAC 1600 ECG Analysis System Revision A 2028451 183 Parts lists Field replaceable units FRUs MAC 1600 Keyboard Non Stress
17. A MAC 1600 ECG Analysis System 4 13 2028451 183 Maintenance FRU replacement procedures 6 Lift the back of the base plastic A from the middle plastic B as shown in the following figure until you can access the two cables that connect these two assemblies B 8 Se 0 0091 2vw 7 Disconnect the two cables from the mainboard 8 Set the base plastic assembly next to the middle plastic assembly as shown in the following figure and identify the power supply assembly A 9 Using small needle nose pliers disconnect pink ground connector from AC outlet 10 Lift the power supply assembly A in the previous figure from the rails 1 Remove the power supply shield 12 Disconnect the white connector from the power supply circuit board by pinching the white tab while pulling the connector 13 Disconnect the grounding cable from the back of the AC power connector 14 Remove the power supply assembly 15 Replace the power supply assembly by reversing the steps for removal Remember to replace the snipped tie wrap with the one provided in the power supply FRU kit 16 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 MAC 1600 ECG Analysis System Revision A 2028451 183 4 14 Maintenance FRU replacement procedures Replacing the acguisition board assembly 1 Separate the middle plastic and base plastic assemblies as described in
18. A2 2006 Group 1 establishments including domestic establishments and EN 55011 1998 A1 1999 A2 2002 Class B those directly connected to the public low voltage power m 150 KHz to 30 MHz supply network that supplies buildings used for domestic FCC 47CFR Part 15 107 purposes B 2 MAC 1600 ECG Analysis System Revision A 2028451 183 3 Electromagnetic compatibility Electromagnetic compatibility EMC Electromagnetic Environment Guidance The MAC 1600 device is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes MAC 1600 ECG Analysis System Compliance Class A Complies Pass 2028451 183 Emissions Test Harmonic Emissions m 2nd 40th Harmonic IEC 60601 1 2 2007 e IEC 61000 3 2 2005 Voltage fluctuations Flicker emissions IEC 60601 1 2 2007 IEC 61000 3 3 1994 A2 2005 Revision A Electromagnetic compatibility Electromagnetic compatibility EMC Guidance and manufacturer s declaration electromagnetic The MAC 1600 system is intended for use in the electromagnetic environment specified in the following table It is the responsibility of the customer or user to ensure that the MAC 1600 system is used in such an environment Electromagnetic Environment Guidance Floors should b
19. Power up mode and does not have the High Security Mode enabled the Main Menu appears after powering up the system No further keys need to be pressed in order to display the Main Menu 1 Stress ECG power up mode These steps describe how to navigate to the Main Menu after powering on the MAC 1600 system when Stress ECG is selected for Power up mode in Basic Setup If the High Security Mode is enabled proceed with steps a d after the window opens prompting for a User ID and Password If the password prompt does not appear go to step 2 NOTE If you need to perform system setup functions be sure you log in as a user that is assigned setup editing privileges a Type your user ID in the User JD field b Press the Enter key or press the down arrow key on the Trimpad to move the focus to the Password field c Type your password in the Password field d Press the F5 key to select Login If the barcode reader option is enabled a window opens prompting you to Scan the Patient barcode Press F6 to select Cancel NOTE If the barcode prompt does not appear go to step 3 Press F5 to select Cancel Press F6 to select More Press F6 to select More Press F5 to select Main Menu MAC 1600 ECG Analysis System 3 31 2028451 183 1 ON Revision A Troubleshooting Freguently asked guestions For your notes 3 32 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance MAC 1600 EC
20. Revision A MAC 1600 ECG Analysis System 4 27 2028451 183 Maintenance FRU replacement procedures c Replace screw C that holds the pressure bar to the left printhead holder 0 8 5 Reattach the three cables A shown in the following figure A 6 Reconnect the six grounding straps 7 Replace the printer board as described in Replacing the printer board on page 4 25 4 28 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures 8 Replace stainless steel tube with the notched end A at the side shown in the following figure 9 Reassemble printer assembly to the main assembly as described in Replacing the printer assembly on page 4 21 10 Configure the Printhead Resistance a From the Main Menu press F5 to select System Configuration b Press F6 More twice Press F5 to select Service Setup a 2 Type prod when prompted to enter the Service Password NOTE If the keyboard does not include the letters prod type 7763 when prompted for the Service Password e Select the Device Settings button f Type 1000 in the Printhead Resistance field g Press F6 to select Save 11 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 Revision A MAC 1600 ECG Analysis System 4 29 2028451 183 Maintenance FRU replacement procedures Replacing the optical
21. STRESS SWE FRU MAC1600 KEYBOARD ASSY W STRESS ITA FRU MAC1600 KEYBOARD ASSY W STRESS JAP FRU MAC1600 KEYBOARD ASSY W STRESS DUT FRU MAC1600 KEYBOARD ASSY W STRESS NOR FRU MAC1600 KEYBOARD ASSY W STRESS DAN FRU MAC1600 KEYBOARD ASSY W STRESS CZE FRU MAC1600 KEYBOARD ASSY W STRESS CHN FRU MAC1600 KEYBOARD ASSY W STRESS HUN FRU MAC1600 KEYBOARD ASSY W STRESS POL FRU MAC1600 KEYBOARD ASSY W STRESS RUS FRU MAC1600 KEYBOARD ASSY W STRESS SLO FRU MAC1600 KEYBOARD ASSY W STRESS POR FRU MAC1600 KEYBOARD ASSY W STRESS KOR FRU MAC1600 KEYBOARD ASSY W STRESS FIN MAC 1600 ECG Analysis System Part Number 2035699 101 2035699 102 2035699 103 2035699 104 2035699 105 2035699 106 2035699 107 2035699 108 2035699 109 2035699 110 2035699 1 11 2035699 113 2035699 114 2035699 115 2035699 116 2035699 117 2035699 118 2035699 119 2035699 120 2028451 183 Keyboards stress 5 18 Parts lists Field replaceable units FRUs Table 21 MAC 1600 Data Matrix Barcode Scanner Kits Description KIT USB DATA MATRIX BARCODE SCANNER ENG KIT USB DATA MATRIX BARCODE SCANNER GER KIT USB DATA MATRIX BARCODE SCANNER FRE KIT USB DATA MATRIX BARCODE SCANNER SPA KIT USB DATA MATRIX BARCODE SCANNER ITA KIT USB DATA MATRIX BARCODE SCANNER JAP KIT USB DATA MATRIX BARCODE SCANNER SWE KIT USB DATA MATRIX BARCODE SCANNER DUT KIT USB DATA MATRIX BARCODE SCANNER
22. Set d 12VP KISS Lead Set d 116VP Recommended separation distance Multilink Lead Set d 1 2 VP 80 MHz to 800 MHz eis 2 3 VP 800 MHz to 2 5 GHz KISS Lead set 1 0 Vim Value Lead Set d 1 2 VP g0 MHz to 800 MHz d 2 3 VP 800 MHz to 2 5 GHz KISS Lead Set d 3 5 VP 80 MHz to 800 MHz d 7 VP 800 MHz to 2 5 GHz MAC 1600 ECG Analysis System B 5 Compliance Level Multilink Lead set 3 0 Vrms Value Lead set 3 0 Vrms KISS Lead set 0 3 Vrms Multilink Lead set 3 0 V m Value Lead set 2028451 183 Compliance Test Level 3 Vrms 150 KHz to 80 MHz 2 Hz mod 3 Vm 80 MHz to 2 5 GHz 2 Hz mod immunity Immunity Test Conducted RF IEC 60601 1 2 2007 e IEC 6 2006 Radiated RF IEC 60601 1 2 2007 IEC 61000 4 3 2006 Revision A Electromagnetic compatibility Electromagnetic compatibility EMC Compliance Test Level Compliance Level Electromagnetic Environment Guidance where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each freguency range b Interference may occur in the vicinity of eguipment marked with the following symbol K At 80 MHz and 800 MHz the higher frequency
23. address of the device to the network administrator to enable the LAN communication option How do I obtain the MAC address A Follow these steps to obtain the MAC address 1 Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 2 Move the focus to the Service Report button and press the Enter key to select 3 Find the MAC address on the printed service report 3 26 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Freguently asked guestions System setup Location number O When entering in the patient data how do I get the Location field to automatically populate with the same number A The Location number can be set in Basic Setup to save you from entering it for each test 1 From the Main Menu press F5 to select System Configuration Press F1 to select Basic Setup Move the focus to the Location field 2 3 4 Type the desired Location number 5 Press F6 to select Save 6 Press F5 to select Main Menu Patient guestions O How do I change what guestions I see when I am entering the patient data A The patient questions you see on the Patient Data window when starting a test were set up in Patient Setup 1 From the Main Menu press F5 to select System Configuration 2 Press F6 to select More 3 Press F4 to select Patient Setup 4 Press F4 to select Page Down 5 Move the focus to the Extra Questions but
24. an imminent hazard which if not avoided will result in death or serious injury WARNING Indicates a potential hazard or unsafe practice which if not avoided could result in death or serious injury CAUTION Indicates a potential hazard or unsafe practice which if not avoided could result in minor personal injury or product property damage NOTE Provides application tips or other useful information to assure that you get the most from your equipment Revision A MAC 1600 ECG Analysis System 1 3 2028451 183 Introduction Safety messages Safety messages Additional safety messages may be found throughout this manual that provide appropriate safe operation information DANGER Do not use in the presence of flammable anesthetics WARNING CONNECTION TO MAINS This is class I eguipment The mains plug must be connected to an appropriately grounded power supply WARNING BATTERY OPERATION If the integrity of the protective earth conductor is in doubt operate the unit from its battery CAUTIONS This eguipment contains no user serviceable parts Refer servicing to gualified service personnel U S Federal law restricts this device to the sale by or on the order of a physician Responsibility of the manufacturer GE is responsible for the effects of safety reliability and performance only if Assembly operations extensions readjustments modifications or repairs are carried out by persons authorized by
25. display The Display Test can be used to determine if the display pixels are working properly 1 Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 2 Select the Display Test button Revision A MAC 1600 ECG Analysis System 3 7 2028451 183 Troubleshooting Performing diagnostic tests 3 Select the Start Test button The following window opens 4 Press the right arrow key on the Trimpad repeatedly to move the color bars horizontally across the screen right arrow ke JJ 9 i 5 Verify that the color band pattern red green blue white scrolls across the screen Pass the test if the pattern is replicated without discoloration 6 Press the F1 key to switch to horizontal color bars 7 Press the down arrow key repeatedly Verify that the color band pattern red green blue white scrolls down the screen Pass the test if the pattern is replicated without discoloration 8 Press the F1 key to switch to cycle through the solid color pane red green blue white For each pane check for black pixels Pass the test if no more than 4 black pixels are observed on any single color pane Note that a black pixel observed on one pane will probably be observed on every pane 9 Press Esc or return when the test is complete The following window opens Did the test pass 10 Select pass or fail Ifthe test passed press F4 to select Y
26. range applies These guidelines may not apply in all situations Electromagnetic propagation is affected by reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters and electromagnetic site survey should be considered If the measured field strength in the location in which the MAC1600 is used exceeds the applicable RF compliance level above the MAC1600 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the MAC1600 b Over the frequency range 150 KHz to 80 MHz field strengths should be less than 3 V m for Multilink and Value Lead Sets and less than 1 V m for the KISS Lead Sets MAC 1600 ECG Analysis System Revision A 2028451 183 Immunity Test NOTE 1 NOTE 2 B 6 Electromagnetic compatibility Electromagnetic compatibility EMC Recommended separation distances The following table provides the recommended separation distances in meters between portable and mobile RF communication eguipment and the MAC 1600 system The MAC 1600 system is intended for use in the electromagnetic environment on which rad
27. regulations regarding EMC and needs to be installed and put into service according to the EMC information stated as follows WARNING Use of portable phones or other radio freguency RF emitting eguipment near the system may cause unexpected or adverse operation WARNING The eguipment or system should not be used adjacent to or stacked with other eguipment If adjacent or stacked use is necessary the eguipment or system should be tested to verify normal operation in the configuration in which it is being used Guidance and manufacturer s declaration electromagnetic emissions The MAC 1600 ECG analysis system is intended for use in the electromagnetic environment specified in the following table It is the responsibility of the customer or user to ensure that the MAC 1600 system is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF emissions Radiated Group 1 use m 30 MHz to 1 000 MHz The MAC 1600 device uses RF energy only for its internal IEC 60601 1 2 2007 function Therefore its RF emissions are very low and are CISPR11 2003 A1 2004 A2 2006 Group 1 not likely to cause any interference in nearby electronic EN 55011 1998 A1 1999 A2 2002 Class B equipment 30 MHz to 6 5 GHz FCC 47CFR Part 15 33 FCC 47CFR Part 15 109 RF emissions Conducted 150 KHz to 30 MHz Pe Buse IEC 60601 1 2 2007 The MAC 1600 device is suitable for use in all CISPR11 2003 A1 2004
28. sensors Each MAC 1600 device includes two optical sensors a gueue hole sensor A and a gueue mark sensor B as shown in the following figure Each sensor connects to the printer board with a cable and connector The gueue mark sensor cable connector C and the gueue hole sensor cable connector D are also shown in the following figure Both gueue sensors are functional with either type of recommended paper Replacing gueue hole sensor The gueue hole sensor performs the paper gueuing function if you are using the thermal paper with gueue holes 1 Separate the middle plastic and base plastic assemblies as described in Replacing the power supply assembly steps 1 through 8 beginning on page 4 13 2 Disconnect the queue hole sensor cable from the printer circuit board 3 Remove the four mounting screws that hold both printhead holders left and right in place 4 _ Lift and remove the printer assembly 4 30 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures 5 Using an allen wrench push up on the locking tab while pulling the sensor housing away from the main plastic housing 6 Using the small flat blade screw driver pry the gueue sensor from the housing as shown in the following figure 7 When reassembling re insert the queue sensor into the queue sensor holder 8 Reverse all remaining disassembly steps to complete the reassembly 9 Perform the appl
29. settings System configuration settings are saved on the mainboard Therefore before replacing the mainboard it is very important to save a record of the system configuration settings The following list summarizes the various ways of saving and or printing the system configuration settings for the purpose of restoring them to the new mainboard after it is installed Ifthe system is functional and allows you to save and print the system configuration settings proceed with the steps in this section on the device before removing its mainboard Ifthere is another MAC 1600 device with identical system configurations perform the steps in this section on that device Then use the SD card to restore the settings to the MAC 1600 device that had the mainboard replaced Ifthe system configurations were saved to SD card and printed upon initial installation use that information to restore the settings after the new mainboard has been installed Ifthe previous three alternatives are not possible the steps in this section do not apply You will have to manually re enter all the system configurations after the replacement of the mainboard 1 Insert an SD card in SD card slot From the Main Menu press F5 to select System Configuration Press F6 More twice Press F3 to select Export Setup Highlight the system setup file you want to export to SD card Press F1 to select Export Re Rr When the Configuration was suc
30. tray release button and pull up on the roller holder 3 Pull the paper tray until it stops 4 Press and hold the release tab A on the base plastic Then pull the paper tray to remove it completely from the device 5 Reassemble the paper tray assembly by reversing the steps for removal 6 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 4 6 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures Replacing the battery 1 Disconnect the system from AC power 2 Remove the paper tray assembly as described in Replacing the paper tray assembly on page 4 6 3 Turn the unit over 4 Press the battery release tab A and raise the battery from its compartment to remove it WARNING BATTERY PACK DISPOSAL Do NOT dispose of the battery by fire or burning Follow local environmental guidelines concerning disposal and recycling 5 Insert new battery by reversing the steps for removal 6 Connect to AC power and power on the system 7 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 Revision A MAC 1600 ECG Analysis System 4 7 2028451 183 Maintenance FRU replacement procedures Replacing the keyboard 1 Disconnect the system from AC power Remove the paper tray as described in Replacing the paper tray assembly Remove the battery as des
31. uA Enclosure Leakage Current 1 Forward Polarity NC __ pA Pass Fail 100 LA 2 Neutral open Forward Polarity SFC __ pA Pass Fail 300 LA 3 Ground open Forward Polarity SFC MuA Pass Fail 300 uA 4 Ground open Reverse Polarity SFC __ UA Pass Fail 300 LA 5 Neutral open Reverse Polarity SFC __ pA Pass Fail 300 LA 6 Reverse Polarity NC __ pA Pass Fail 100 uA Patient Leakage Current to Ground 1 Forward Polarity NC __ pA Pass Fail 10 uA 2 Neutral open Forward Polarity SFC __ pA Pass Fail 50 pA 3 Ground open Forward Polarity SFC __ pA Pass Fail 50 pA 4 Ground open Reverse Polarity SFC __ MA Pass Fail 50 pA 5 Neutral open Reverse Polarity SFC __ MuA Pass Fail 50 pA 6 Reverse Polarity NC __ MA Pass Fail 10 pA Ground Continuity Resistance 1 AC mains power cord ground i Sandie NIA 0 Pass Fail 1 NC Normal Condition SFC Single Fault Condition N A Not Applicable 2 UUT Unit Under Test Revision A MAC 1600 ECG Analysis System 4 41 2028451 183 Maintenance Updating software Updating software Software updates are provided on an SD card Perform a software update as described in this section 1 Power up the system with the system connected to AC power 2 Insert the SD card A with the software update in SD card slot in the back panel as shown in the following figure with gold contacts up
32. 0 0 9 5 2 Field replaceable units 0 cece eects 5 3 MAC 1600 upper level assembly PN 2032093 5 3 Power supply FRU kit PN 2035703 001 777 74 5 10 Mainboard ETE assembly FRU kit PN 2035704 001 5 10 Acquisition assembly FRU kit PN 2035705 001 4 5 10 Cable harness FRU kit PN 2035707 001 774 5 10 Display cover assembly FRU kit PN 2035706 001 5 11 Printer assembly FRU kit PN 2035702 001 0 eee eee 5 11 Paper tray and printer door FRU kit PN 2035711 001 5 12 Display assembly FRU kit PN 2035700 001 774 5 12 Printer board assembly FRU kit PN 2036813 001 5 12 KISS pump assembly FRU kit PN 2036814 001 5 12 KISS pump hardware FRU kit PN 2036815 001 5 13 Internal modem FRU kit PN 2036816 001 YY RG 5 13 Printhead FRU kit PN 2036817 001 77777 4 5 13 Optical sensor and bracket FRU kit PN 2036818 001 5 13 Fastener hardware FRU kit PN 2035708 001 4 5 14 Middle and base plastic FRU kit PN 2036812 001 5 16 Plastics kit PN 2035709 001 0000000072 5 16 Keyboards non StreSS 0 7 70 5 17 Keyboards STOSS ssc 0 dy RO GU 5 18 Data Matrix barcode scanner kits 00002c cece YF eee eee 5 19 Power COLDS is ius stra es cece Sed Ste ead cae eee aaa Sa ead Y nt 5 20 Revision A MAC 1600 ECG Analysis System iii 2028451 183 Technical specifications
33. 2095 001 2032097 Revision A EMC compliant supplies amp accessories Value accessories Value accessories Part Number Description 2029890 001 Value 10LD Patient Cable Ldwr IEC 2029893 001 Value 10LD Patient Cable Ldwr AHA 2029891 001 Value Reusable Limb Clamps IEC 4 set 2029894 001 Value Reusable Limb Clamps AHA 4 set 2029892 001 Value Reusable Bulb Electrodes 6 set C 4 MAC 1600 ECG Analysis System Revision A 2028451 183 EMC compliant supplies amp accessories Thermal papers Thermal papers Contrast Papers Description CONTRAST Paper Full Page Red Grid Full Size A4 Queue Mark 150 sheets pack 1500 sheets case CONTRAST Paper Full Page Red Grid 8 5 x 11 in Queue Mark 150 sheets pack 1500 sheets case Premium Papers Description Premium Paper Red Grid Full Size A4 Queue Mark 150 sheets pack 1500 sheets case Sold only in Europe Middle East Africa not in US or Canada Premium Paper Full Page Red Grid 8 5 x 11 in Queue Hole 150 sheets pack 2400 sheets case Premium Paper Red Grid with Blank Interpretation Area 8 5 x 11 in Queue Hole 150 sheets pack 2400 sheets case Archivist 25 Paper for Europe Description Archivist 25 Paper Red Grid A4 Queue Mark 150 sheets pack 1500 sheets case Sold in EMEA notin US or Canada Archivist 25 Paper for US Descriptio
34. 4 15 Revision A MAC 1600 ECG Analysis System 3 19 2028451 183 Troubleshooting Eguipment problems Eguipment problems ECG data noise If the acguired ECG data displays unacceptable noise levels When troubleshooting noise or signal quality be sure the problem is not being caused by poor skin preparation or placement and condition of electrodes Careful skin preparation is the key to an interference free ECG Signal guality is indicated using Hookup Advisor Hookup Advisor can be turned on or off in the ECG menu Select System Setup gt ECG gt ECG Acguisition Check for defective or date expired electrodes Check for defective broken or disconnected leadwires Run the Testing the acguisition module on page 3 11 to ensure that all lead wires pass the noise test 3 20 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Error codes Error codes No action is necessary for isolated error occurrences However if the unit is malfunctioning and any of the following error messages are repeating and unrecoverable replace the FRUs in the order as listed Acquisition error codes Where multiple solutions are suggested perform the suggested actions in their listed order Once a solution restores functionality do not attempt the next solution in the list Table 2 Acquisition Error Codes Error Code Cause Solution Acquisition Error 4 Invalid command sent to acquisiti
35. 5 Printer Internal Error 6 Printer Internal Error 7 Revision A Troubleshooting Freguently asked guestions Freguently asked guestions Maintenance NOTE See the MAC 1600 Operator s Manual for complete System Configuration information Saving system setups to SD card O How do I save changes I have made to the System Configuration A Perform the following steps 1 Insert an SD card in SD card slot in back panel as shown in the following figure with gold contacts up Push the SD card into slot to seat it in place From the Main Menu press F5 to select System Configuration 2 3 4 Press F6 to select More 5 Press F6 again to select More 6 Press F3 to select Export Setup Select Setup for Export Loaded Setup Setup files in internal storage Setup files on external media Setup1 Setup2 Ss 7 Highlight the setup file you want to save to SD card from the list on left side of the window 8 Press F1 to select Export 9 The Configuration was successfully exported window opens 10 Press F6 to select OK 11 Eject the SD card by pushing it in once Store it to a secure location 3 24 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Freguently asked guestions How do I restore system setups from the SD card Perform the following steps 1 Insert the SD card in the SD card slot in the back panel as describe
36. 600 ECG Analysis System 2028451 183 Maintenance FRU replacement procedures FRU replacement procedures High level FRU identification 4 3 Item Description Item Description Item Description A paper tray assembly barcode reader option K mainboard ETE assembly B internal modem option G patient cable L power supply assembly 6 display H base plastic assembly M printer assembly D keyboard middle plastic assembly N acquisition module assembly E battery J KISS pump option Revision A MAC 1600 ECG Analysis System 2028451 183 Maintenance FRU replacement procedures Preparing system for FRU replacement Prior to performing any disassembly procedures perform these steps NOTE Take strict precautions against electrostatic discharge damage while replacing field replaceable units 1 Power off the system Disconnect the unit from the AC wall outlet Disconnect the power cord from the back panel connector 1 ro Disconnect the patient cable from the unit as described in Replacing the patient cable on page 4 4 5 Remove the paper tray assembly as described in Replacing the paper tray assembly on page 4 6 6 Remove the battery as described in Replacing the battery on page 4 7 Replacing the patient cable 1 Disconnect the system from AC power 2 Disco
37. CTOR LUER FEMALE NUT LOCK PLUG LUER MALE Cable harness FRU kit PN 2035707 001 Table 5 Cable Harness FRU Kit 2035707 001 Item CABLE ASSY MAC1600 PRINTER TO MITSUBISHI CABLE FFC 20POS 1 25MM 50MM LONG S SIDE CABLE FFC 8POS 1 00MM 180MM LONG S SIDE CABLE ASSY MAC1600 PWR SUPPLY TO MAIN BD CABLE ASSY MAIN TO ACQ MAC 1600 CABLE ASSY MAC1600 GROUND PS TO EQPOT MAC 1600 ECG Analysis System 2028451 183 5 10 Gty Gty Gty 0 108 Meter Parts lists Field replaceable units FRUs Table 5 Cable Harness FRU Kit 2035707 001 Continued Item CABLE ASSY MAC1600 EQUIP TO MAIN BD CABLE ASSY MAIN TO PRINTER MAC 1600 Display cover assembly FRU kit PN 2035706 001 Table 6 Display Cover Assembly FRU Kit 2035706 001 Item ASSY MAC1600 KEYBOARD BEZEL AND SCREEN Printer assembly FRU kit PN 2035702 001 Table 7 Printer Assembly FRU Kit 2035702 001 Item SENSOR OPT MAC1100 1200 PRINTHEAD HOLDER LEFT PRINTHEAD HOLDER RIGHT TUBE EDGE PROTECTION 1 6 2 3MM NYLON UL PRINTHEAD MAC1200 WRITER SHEC TPH BRACKET PRINTHEAD MAC 1600 SCR DIN7985 M3X5 4 8 Z A2F ISO7045 SPRING CPRSN 6 5X8X0 7 PRESS STRIP SCR DIN965 M3X8 4 8 Z A2F ISO7046 CABLE ASSY MAC1600 PRINTER TO MITSUBISHI CABLE FFC 20POS 1 25MM 50MM LONG S SIDE PCB MAC1600 PRINTER BOARD SCR F PLASTIC 3X10 ST A2F PANHEAD Z ADAPTER REDUCING F
38. G Analysis System 2028451 183 4 Revision A Maintenance Introduction Introduction Recommended maintenance Regular maintenance irrespective of usage is essential to ensure that the eguipment will always be functional when reguired See the MAC 1600 operator s manual for cleaning procedures The MAC 1600 system does not require any calibration procedures GE recommends that electrical safety checks be performed annually WARNING Failure on the part of all responsible individuals hospitals or institutions employing the use of this device to implement the recommended maintenance schedule may cause equipment failure and possible health hazards The manufacturer does not in any manner assume the responsibility for performing the recommended maintenance schedule unless an Equipment Maintenance Agreement exists The sole responsibility rests with the individuals hospitals or institutions utilizing the device Required tools and supplies The following table lists the tools required to perform the procedures described in Small flat blade screw driver Current leakage tester Anti static wrist strap MAC 1600 Service Manual MAC 1600 Operator s Manual Small adjustable wrench Revision A Table 1 Required Tools this chapter ECG simulator Posidriver PZ1 x 80 Posidriver PZ2 x 100 Small side cutter pliers Large paper clip or small diameter Allen wrench Small needle nose pliers 4 2 MAC 1
39. GE Healthcare MAC 1600 ECG Analysis System Product Code SDE Service Manual 2028451 183 Revision A NOTE The information in this manual only applies to MAC 1600 ECG Analysis Systems with product code SDE Due to continuing product innovation specifications in this manual are subject to change without notice Marguette MUSE CASE MAC MARS MULTI LINK Hookup Advisor 12SL Archivist Mactrode and BabyMAC are trademarks owned by GE Medical Systems Jnformation Technologies a General Electric Company going to market as GE Healthcare All other marks are not owned by GE and are instead owned by their respective owners 2008 General Electric Company All rights reserved T 2 MAC 1600 ECG Analysis System Revision A 2028451 183 10 April 2008 Contents Introduction Manual information Revision history Manual purpose Intended audience Warnings cautions and notes Safety messages Responsibility of the manufacturer General oro ian tk eraa Equipment symbols Service information Service requirements Equipment identification Equipment overview General description Front View 48 Side View Back 777 Keyboard rveee Keyboard stress option Detailed description Block dia
40. GUVOBA3M 9V TX9V A18W3SSV YILIHM gt 2 YOSNIS 58 ae 2 guvog y 6 F y ay ZdSM E ZT LY IT LY GY3HiNINd YILIYM gt 5 a S 5 ial 9868 4 Bs QV3HINIYd 8V Io a 9 m YIMYYA YILIYM 6Y 2m es YOLD3NNOO 5 IIN s i3M0d yd 6M TWNIWYSL 006 3 cd9M cde im Sexly id LM LIXIW 6IXIV bird Ld 1M 6 gt 030 val NOILTSIDOV 08 08 vy 8Snv I d M z Z EM 606 cdLM 9 LNISSIWIV gt gt aes ele if 8 08 2 gt gt z ox ds e oj o o an mo o g a 5 5 D c sc a x 5 318V2 09v ASSY 091 viv ayog W30OW lv 31n0OW 313 ely 8 IV Block diagram Detailed description ion Detailed descript Equipment overview 2 9 Eguipment overview Hardware firmware architecture Hardware firmware architecture The MAC 1600 hardware subsystems include the following CPU core Display Keyboard ECG Acquisition Thermal printer Power supply Housing MAC 1600 ECG Analysis System 2028451 183 Revision A Revision A Eguipment overview Hardware firmware architecture MAC 1600 ECG Analysis S
41. HUN KIT USB DATA MATRIX BARCODE SCANNER RUS KIT USB DATA MATRIX BARCODE SCANNER CHN KIT USB DATA MATRIX BARCODE SCANNER CZE KIT USB DATA MATRIX BARCODE SCANNER POL KIT USB DATA MATRIX BARCODE SCANNER NOR KIT USB DATA MATRIX BARCODE SCANNER SLO KIT USB DATA MATRIX BARCODE SCANNER POR KIT USB DATA MATRIX BARCODE SCANNER DAN KIT USB DATA MATRIX BARCODE SCANNER KOR KIT USB DATA MATRIX BARCODE SCANNER FIN Part Number 2035979 001 2035979 002 2035979 003 2035979 004 2035979 006 2035979 007 2035979 005 2035979 008 2035979 014 2035979 016 2035979 013 2035979 011 2035979 015 2035979 009 2035979 017 2035979 018 2035979 010 2035979 019 2035979 020 Table 22 MAC 1600 Data Matrix Barcode Scanner Cable Description CABLE USB BARCODE SCANNER FOR HHP4600 Part Number 2035628 001 1 Cable is included in barcode scanner kits Order this part number if cable only is required MAC 1600 ECG Analysis System 2028451 183 Data Matrix barcode scanner kits Revision A Revision A Parts lists Field replaceable units FRUs Table 23 MAC 1600 Power Cords Description PWR SPLY CRD RA HOSP GRD 13A 125V 10FT PWR SPLY CRD RA CONT EURO 10A 250V 2 5M PWR SPLY CRD RA BRITSH 10A 250V 2 5M FSD PWR SPLY CRD RA ITALIAN 10A 250V 2 5M PWR SPLY CRD RA ISRAELI 10A 250V 2 5M PWR SPLY CRD RA AUSTRALIAN 10A 250V 2 5M PWR SPLY CRD ST SWISS 10A 250V 2 5M
42. INTER MAC 1600 PACKAGING INSERT FRONT FOAM PACKAGING INSERT REAR FOAM PLASTIC BAG ESD DESICCANT BAG B PACKAGING BOX MAC1600 UNIT HOSE BARB STRAIGHT THRU CONN 1 8 ID TUBING PVC 3X1 GREY LABEL RATING MAC1600 PAD EMI ABSORBSION PAD MAC1600 BATTERY KEYBOARD MODEM SOCKET MT9234SMI V 92 CARD SECURE DIGITAL 256MB CARD SECURE DIGITAL 512MB PROG SD CARD MAC1600 SYSTEM OS IMAGE V1 0 OPTION ACTIVATION LABEL MAC 1600 ECG Analysis System 2028451 183 ASSY MAC1600 KEYBOARD BEZEL AND SCREEN ASSY PRINTER TRAY MAC 1600 Part Number 2035701 001 2027268 003 2027268 004 2040743 001 Item 48 50 52 55 56 57 59 60 61 62 63 64 65 66 67 68 69 70 71 73 74 15 76 78 79 Revision A Gty Gty Gty Gty Revision A Parts lists Field replaceable units FRUs Power supply FRU kit PN 2035703 001 Table 2 Power Supply FRU Kit 2035703 001 Item PWR SPLY MAC 1600 SHIELD PWR SUPPLY MAC 1600 CABLE STRIP 92 X 2 4 MM Mainboard ETE assembly FRU kit PN 2035704 001 Table 3 Mainboard ETE Assembly FRU Kit 2035704 001 Item ASSY MAC1600 MAINBOARD AND ETE MODULE PROG SD CARD MAC1600 SYSTEM OS IMAGE V1 0 Acquisition assembly FRU kit PN 2035705 001 Table 4 Acquisition Assembly FRU Kit 2035705 001 Item ASSY ACQUISITION BOARD MAC 1600 LABEL LUER MAC 1600 CONNE
43. M F STL W ZINC 2 STANDOFF M3 X 16MM HEX M F STL W ZINC 1 CABLE ASSY MAC1600 INTERNAL KISS PUMP 1 CONNECTOR LUER FEMALE 1 NUT LOCK 1 PLUG LUER MALE 1 Internal modem FRU kit PN 2036916 001 Table 13 Internal Modem FRU Kit 2036816 001 Where Used Gty MODEM SOCKET MT9234SMI V 92 1 Printhead FRU kit PN 2036917 001 Table 14 Printhead FRU Kit 2036817 001 Where Used Gty PRINTHEAD MAC1200 WRITER SHEC TPH 1 Optical sensor and bracket FRU kit PN 2036918 001 Table 15 Optical Sensor amp Bracket FRU Kit 2036817 001 Where Used Gty SENSOR OPT MAC1100 1200 1 HLDR MARK READER 1 Revision A MAC 1600 ECG Analysis System 5 13 2028451 183 Gty NO N Revision A Parts lists Field replaceable units FRUs Table 16 Fastener Hardware Kit PN 2035708 001 Where Used Item 17 Upper Level Assy Item Description ADAPTER REDUCING SCR FH 4 40X1 4 ZINC AHNL DIN965 SCR PLASTIC 2 5X10 V2A PANHEAD Z GEAR WHEEL BEARING PIN 2 X 15 8 SCR DIN7985 M3X5 4 8 Z A2F ISO7045 GEAR WHEEL 45 14T SPRING CPRSN 6 5X8X0 7 GEAR WHEEL SCR DIN965 M3X8 4 8 Z A2F ISO7046 SPRING PRINTER DOOR MAC 1600 SCR F PLASTIC 3X10 ST A2F PANHEAD Z SCR DIN7985 M4X10 4 8 Z A2F ISO7045 SCR DIN7985 M4X40 4 8 Z A2F SPECIAL SCR DIN7985 M3X20 4 8 Z A2F ISO7045 SCR F PLASTIC 2 2X8 ST A2F PANHEAD Z MOUNTING BASE CARDIOSERV CABLE STRIP
44. MAC 1600 ECG Analysis System 2028451 183 1 Visual inspection Revision A Maintenance Functional checkout Functional checkout procedures Perform the functional checkout procedures that are applicable to the replacement procedure that was performed Power up self test passed See Power up self test on page 3 2 for more information If software update had been performed verify that the new version of software appears on the Main Menu screen Recorded rhythm strip successful See Chapter 5 Recording a Resting ECG in the MAC 1600 ECG Analysis System Operator s Manual for more information Recorded ECG successful See Chapter 5 Recording a Resting ECG in the MAC 1600 ECG Analysis System Operator s Manual for more information ECG stored successfully See Chapter 8 Managing Internal Storage in the MAC 1600 ECG Analysis System Operator s Manual for more information Simulated ECG data transmitted successfully to a receiving product See Chapter 8 Managing Internal Storage in the MAC 1600 ECG Analysis System Operator s Manual for more information Exercise device successfully communicates with the MAC 1600 system See Chapter 7 Stress Testing in the MAC 1600 ECG Analysis System Operator Manual for more information Display test successful See Testing the display on page 3 7 for more information Keyboard test successful See
45. SY MAC1600 PWR SUPPLY TO MAIN BD PWR SPLY MAC 1600 CABLE ASSY MAIN TO ACQ MAC 1600 ASSY ACQUISITION BOARD MAC 1600 CONNECTOR LUER FEMALE NUT LOCK PLUG LUER MALE CONN BINDING POST FOR EQUIPOTENTIAL CABLE ASSY MAC1600 GROUND PS TO EQPOT CABLE ASSY MAC1600 EQUIP TO MAIN BD PAD SCR DIN7985 M4X40 4 8 Z A2F SPECIAL ASSY DISPLAY MAC 1600 ECG Analysis System 2028451 183 Part Number Item 32 33 34 35 37 38 39 41 42 43 45 46 47 5 8 Included in Fastener Hardware FRU Kit Display Cover Assembly FRU Kit Paper Tray amp Printer Door FRU Kit Fastener Hardware FRU Kit Printer Assembly FRU Kit Fastener Hardware FRU Kit Power Supply FRU Kit m Middle and Base Plastic FRU Kit m Middle and Base Plastic FRU Kit m KISS Pump Assembly FRU Kit m KISS Pump Assembly FRU Kit m Display Assembly FRU Kit Keyboards Non Stress on page 6 17 Keyboards Stress on page 6 18 Internal Modem FRU Kit 5 9 Parts lists Field replaceable units FRUs Table 1 MAC 1600 Upper Level Assembly PN 2032093 Continued Item Description SCR DIN7985 M3X20 4 8 Z A2F ISO7045 MOUNTING BASE CARDIOSERV CABLE STRIP 92 X 2 4 MM SHIELD PWR SUPPLY MAC 1600 PAD FLOPPY MAC SERIES FERRITE RECT CORE 15X10X23MM 2270H 10MHZ LABEL PRODUCT SERIAL BARCODE DATE SIZE A PR
46. Testing the keyboard on page 3 9 for more information Writer test successful See Testing the writer on page 3 12 for more information Acquisition module test successful See Testing the acquisition module on page 3 11 for more information Patient leadwire check successful See Testing the patient lead wires on page 3 19 for more information Battery status acceptable See Checking battery status on page 3 12 for more information LAN test successful See Testing the LAN option on page 3 16 for more information Modem test successful See Testing the modem on page 3 17 for more information MAC 1600 ECG Analysis System Revision A 2028451 183 10 11 12 13 14 Operational checks Diagnostic tests 4 40 Maintenance Functional checkout Electrical Safety Checks 15 Current leakage test results meet reguirements Perform electrical safety checks when indicated All indicated electrical safety checks reguire a pass fail indication for steps performed Record the measurement values in your debrief Table 5 Electrical Safety Checks Leakage 1 2 Step Condition UUT ON Result Current Limits Earth Leakage Current 1 Forward Polarity NC pA Pass Fail 300 LA 2 Neutral open Forward Polarity SFC MuA Pass Fail 1 000 LLA 3 Neutral open Reverse Polarity SFC __ pA Pass Fail 1 000 uA 4 Reverse Polarity NC __ MuA Pass Fail 300
47. ard as indicated by the icon The MAC 1600 system supports only SD cards formatted for the FAT or FAT16 file systems External Video Monitor Standard 15 pin VGA connector for an external connection monitor Connect a medical grade VGA CRT or medical grade VGA compatible LCD display NOTE The MAC 1600 display resolution is 800 by 480 pixels Due to differing aspect ratios the image may appear distorted on some LCD monitors Consult the operating guide for your LCD monitor J Equipotential grounding Connect non grounded peripheral devices to ensure lug equipotential WARNING SYSTEM LEAKAGE CURRENT Keep leakage current within acceptable limits when connecting auxiliary equipment to this device Total system leakage must not exceed 300 microamperes United States or 500 microamperes international MAC 1600 ECG Analysis System Revision A 2028451 183 Eguipment overview General description NOTE The English keyboard is shown in this section A B 0 wa FI F2 F3 F 5 8 pm Protest Exercise Recovery Test end Hold Sped Grade 1 Start tmill ee 6 Recall 121d Medians Comment Enter BP Speed Grade STOP mill pm D yee B a w E R Y U 0 P Es F A s D F 6 H J K L a mem A G z x 6 v 8 N M 2 1 5 swe 2 5 Esc 2 Alt lt gt Vv N M L K J H Keyboard Description Used to select menu options on the screen Used to t
48. as described in Replacing the keyboard on page 4 8 6 Remove the internal modem if this option is installed as described in Replacing the internal modem option on page 4 12 7 Remove the rear connector panel 8 Separate the middle plastic and base plastic assemblies 9 Remove the power supply as described in Replacing the power supply assembly on page 4 13 10 Ifthe KISS option is knstalled remove the KISS pump as described in Replacing the KISS pump assembly option on page 4 17 4 34 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures 11 Remove disconnect the following from the mainboard f KISS pump harness connector from mainboard if this option is installed both ends of the gray USB cable the two shorter KISS pump standoffs if this option is installed amp o the longer KISS pump standoff if this option is installed e the power supply connector from the mainboard f the two mounting screws g grounding lug connector from the mainboard 12 Pull the keyboard ribbon cable A from the ferrite loop B 13 Remove mainboard ETE assembly 14 Disconnect the keyboard cable from the bottom of the mainboard Revision A MAC 1600 ECG Analysis System 4 35 2028451 183 Maintenance FRU replacement procedures Reassembling the Mainboard 1 Reconnect the keyboard cable to the bottom of the new mainboard
49. assigned setup editing privileges a Type your user ID in the User JD field b Press the Enter key or press the down arrow key on the Trimpad to move the focus to the Password field c Type your password in the Password field d Press the F5 key to select Login If the barcode reader option is enabled a window opens prompting you to Scan the Patient barcode Press F6 to select Cancel NOTE If the barcode prompt does not appear go to step 3 Press F5 to select Cancel Press F6 to select More Press F5 to select Main Menu MAC 1600 ECG Analysis System Revision A 2028451 183 1 3 30 Troubleshooting Freguently asked guestions Main Screen Power Up Mode These steps describe how to navigate to the Main Menu after powering on the MAC 1600 system when Main Screen is selected for Power up mode in Basic Setup If the High Security Mode is enabled proceed with steps a d after the window opens prompting for a User JD and Password If the password prompt does not appear go to step 2 NOTE If you need to perform system setup functions be sure you log in as a user that is assigned setup editing privileges a Type your user ID in the User JD field b Press the Enter key or press the down arrow key on the Trimpad to move the focus to the Password field c Type your password in the Password field d Press the F5 key to select Login The Main Menu is displayed If the system is configured for Main Screen
50. astener Hardware FRU Kit SCR F PLASTIC 2 2X8 ST A2F PANHEAD Z CABLE STRIP 92 X 2 4 MM MAC 1600 ECG Analysis System 2028451 183 Revision A Gty Gty Gty Gty 07M Revision A Parts lists Field replaceable units FRUs Paper tray and printer door FRU kit PN 2035711 001 Table 8 Paper Tray amp Printer Door FRU Kit 2035711 001 Where Used ASSY PRINTER TRAY MAC 1600 Display assembly FRU kit PN 2035700 001 Table 9 Display Assembly FRU Kit 2035700 001 Where Used ASSY DISPLAY PAD EMI ABSORBSION PAD MAC1600 Printer board assembly FRU kit PN 2036813 001 Table 10 Printer Board Assembly FRU Kit 2035700 001 Where Used PCB MAC1600 PRINTER BOARD KISS pump assembly FRU kit PN 2036814 001 Table 11 KISS Pump Assembly FRU Kit 2036814 001 Where Used SCR DIN7985 M3X5 4 8 Z A2F ISO7045 ASSY KISS PUMP PCB HOLDER PUMP CABLE STRIP 293 X 4 8 MM CABLE ASSY MAC 1600 INTERNAL KISS PUMP HOSE BARB STRAIGHT THRU CONN 1 8 ID TUBING PVC 3X1 GREY 5 12 MAC 1600 ECG Analysis System 2028451 183 Parts lists Field replaceable units FRUs KISS pump hardware FRU kit PN 2036915 001 Table 12 KISS Pump Hardware FRU Kit 2036915 001 Where Used Gty SCR DIN7985 M3X5 4 8 Z A2F ISO7045 3 STANDOFF M3 X 10MM HEX
51. board 4 Disconnect the six green grounding cables by removing the screws A holding them in place 4 24 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures 5 Disconnect the white ribbon cable by pushing the connector sleeve A in the direction shown in the following figure 6 Unplug the narrow connector B of the harness from the printhead shown in the following figure and remove the printer board B A 7 Disconnect wide connector from the printer board by pushing on the latch A while pulling the connector HUTT TT 1 y LITTLE Replacing the printer board 1 Reconnect the black harness to the printer assembly and to the replacement printer board Reconnect the white ribbon connector to the printer board Secure the ribbon cable by sliding the sleeve to the locked position Reconnect the six grounding straps o Guide the queue mark sensor ribbon cable between the printer board and the printhead and reconnect the ribbon cable to the printer board 6 Place bottom of printer board in grooves and snap the top edge at each side back into position 7 Follow the printer assembly replacement procedures as described in Replacing the printer assembly on page 4 21 8 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 Revision A MAC 1600 ECG Analysis System 4 25 2028451 183 Maintenan
52. ce FRU replacement procedures Replacing the printhead Removing the printhead 1 Remove the printer board as described in Removing the printer board on page 4 24 2 Remove the stainless steel tube by pushing outward on the printhead holder while pushing the tube as shown in the following figure 3 Remove the screw A that fastens the pressure strip to the left printhead holder Remove the left printhead holder in the same manner Repeat the previous two steps for the right printhead holder Disconnect the two cables mi OY Oe Remove the four screws A from the metal mounting plate 4 26 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures The following figure shows the printhead A disassembled from the metal mounting plate B 110 Reassembling the printhead 1 Assemble new printhead to metal mounting plate 2 Replace four screws to reassemble new printhead to mounting plate 3 Insert the printhead assembly into the right printhead holder by sliding it under the two tabs A shown in the following figure 4 Stand the assembly on the right plastic printhead holder side with no queue sensor and reassemble the left printhead holder by performing the following three steps simultaneously a Push in on pressure bar with your thumb A b Push down on the plastic printhead holder with your index finger B to align the screw holes
53. cessfully exported message is displayed press F6 to select OK 8 Remove the SD card and store in a secure location Revision A MAC 1600 ECG Analysis System 4 33 2028451 183 Maintenance FRU replacement procedures Printing the system setup report If you feel you may need it for additional reference print the System Setup report after the FRU replacement procedure 1 From the Main Menu press F5 to select System Configuration Press F6 to select More Press F3 to select Print Setup Report o A Move the focus to the Complete Setup button and press the Enter key to select The System Setup report prints 5 Save the printed setup report in a secure location It can be used as a reference if system setup needs to be restored manually NOTE Option codes are not part of system configuration settings However these codes will be listed on the last page of your printed setup report These codes are also found on a label next to the battery compartment The printed list of option codes is a more convenient aid to use when these codes need to be re entered after the new mainboard is installed Removing the mainboard 1 Disconnect the system from AC power Remove the paper tray as described in Replacing the paper tray assembly Remove the battery as described in Replacing the battery mb ob Remove the display as described in Replacing the display assembly on page 4 11 on Remove the keyboard
54. cessories Introduction Introduction This section of the manual lists the EMC compliant supplies and accessories that are supported by the MAC 1600 system and are recommended for use by GE Healthcare WARNING The use of accessories transducers and cables other than those specified may result in increased emissions or decreased immunity performance of the equipment or system C 2 MAC 1600 ECG Analysis System Revision A 2028451 183 EMC compliant supplies amp accessories Standard accessories Standard accessories Description Patient trunk cable 10 lead IEC Patient trunk cable 10 lead AHA Suction Electrodes Blue each Electrode Spray Set of lead wires 4mm connector 10 leads defibrillator proof IEC Set of lead wires 4mm connector 10 leads defibrillator proof AHA BabyMAC electrodes Leadwire set ECG 10 L w resist Banana IEC Leadwire set ECG 21 L w resist Banana IEC ECG Electrode Clamp Large 4 set CAM14 Leadwire Set Banana Mixed Length AHA 10 set CAM14 Universal Leadwire Set Mixed length 10 set KIT ADAPTER 10 SET AHA BANANA KIT ADAPTER 10 SET BANANA IEC Electrode Prep Pads Universal Clip 10 Silver Mactrode Plus 100 Battery Keyboard MAC 1600 ECG Analysis System 2028451 183 Part Number 22341808 22341809 21714401 21730701 38401816 38401817 402207 210 38401765 38401766 91920232 2016032 001 420101 002 900178 001 900178 101 9386 001 9490 210 9623 003P 203
55. check this time is calculated based on using the system for display purposes only Actual remaining battery life will be less when the system is used for both display and printing purposes Battery Charge Remaining can range from 0 to 100 According to the battery specifications see Specifications on page A 3 you should see at least 3 00 three hours of Battery Life Remaining display time with 100 Battery Charge Remaining If the numbers displayed in this window are less you may require more frequent battery charging or you may wish to replace the battery See Replacing the battery on page 4 7 for information on how to replace the battery 4 Note the battery status information and press F6 to select Cancel and close the BATTERY STATUS window Testing the writer The Writer Test can be used to determine if the writer is working properly NOTE Before performing the Writer Test be sure that thermal paper is properly loaded in the writer tray 1 Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 2 Select the Writer Test button 3 12 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Performing diagnostic tests The following window opens 3 Perform the 50mm s Speed Test a Select the 50mm s Speed Test button The writer prints the 50 mm s speed test report b When one page of the report has printed press the Sto
56. cle You can condition the MAC 1600 battery pack while installed in a MAC 1600 system that is not being used to record tests on a patient Disconnect the AC mains power from the MAC 1600 system Display the BATTERY STATUS window a From the Main Menu press F5 to select System Configuration b Press F6 twice to select More gt More c Press F5 to select Service Setup d Type prod when prompted to enter the Service password and press F6 to select OK NOTE If the keyboard does not include the letters prod type 7763 and press 16 to select OK e Select the System Diagnostics button f Select the Battery Test button The BATTERY STATUS window opens BATTERY STATUS AC Line Status Battery Life Remaining hh mm Battery Charge Remaining Allow the battery to discharge until the Battery Charge Remaining is less than 90 Turn off the unit and reconnect the AC mains power Allow the battery to fully charge NOTE Solid amber battery LED indicates the battery is charging When the battery LED turns off this indicates that the battery is fully charged Remove the AC mains power and turn the MAC 1600 system unit on Leave the unit on and allow the battery to discharge until MAC 1600 system shuts off Reconnect the AC mains power to the MAC 1600 leaving the unit turned off and allow the battery to fully recharge When the amber Battery LED indicator stops flashing and turns on solid the battery is fully c
57. cribed in Replacing the battery Ra DD o Remove the two screws A from the bottom of the device A 5 Turn unit over right side up 6 Pusha small Allen wrench or straightened paper clip into one of the keyboard bezel release holes A as shown in the following figure while pushing up on the corner of the keyboard bezel B 4 8 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures 7 Repeat step 6 for the other corner of the keyboard bezel The two keyboard bezel release holes A are shown in the following figure A 8 Lift the back end of the keyboard bezel approximately one inch 9 Push forward on the bezel in the direction indicated by arrow in the following figure to unlatch its tabs from the four alignment holes at the front of keyboard bezel from the unit 10 Remove the keyboard bezel 11 Lift the keyboard 12 On the back side of the keyboard pull down on the lock release tabs A on each side of the keyboard cable connector shown in the following figure A Revision A MAC 1600 ECG Analysis System 4 9 2028451 183 Maintenance FRU replacement procedures 13 Disconnect the keyboard cable as shown m 14 Remove the keyboard 15 Reassemble a new keyboard by reversing the steps for removal The following sequence of steps describes how to replace the keyboard bezel a Align the four tabs A at
58. d on page 4 24 Writer paper does not Faulty stepper motor Replace paper tray assembly move See Replacing the paper tray assembly on page 4 6 Replace the printer assembly See Replacing the printer assembly on page 4 19 Paper moves but lines Faulty cables in printer assembly do not print See Replacing the printer assembly on page 4 19 Replace the printhead See Replacing the printhead on page 4 26 1 Roller included with paper tray assembly 2 Stepper motor included with paper tray assembly 3 Stepper motor included with paper tray assembly MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Performing diagnostic tests Testing the RS232 port The RS232 Test can be used to determine if the Com ports are working properly 1 Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 2 Use a paper clip to short pins 2 amp 3 on COM A 6 65606 9 3 Select the RS232 Test button The following window opens COM PORT LOOP BACK TEST COM1 Connections Connect TX Pin 3 to RX Pin 2 of SERIAL OUT 1 COM2 Connections Connect TX Pin 3 to RX Pin 2 of SERIAL OUT 2 COM 1 4 Perform the COM Port loop back test on COM 1 a With the focus on COM 1 press the Enter key The results of the COM 1 loop back test are displayed b Note the re
59. d previously From the Main Menu press F5 to select System Configuration Press F6 to select More Press F6 again to select More Press F2 to select Jmport Setup ON ge Highlight the setup file you want to import from the SD card from the list on right side of the window 7 Press F1 to select Import 8 _ The Configuration was successfully imported window opens 9 Press 16 to select OK 10 Eject the SD card by pushing it in once Store it to a secure location O A Printing a system setup report I would also like a printed record ofmy MAC 1600 System Configuration file How do I obtain this report Perform the following steps 1 From the Main Menu press F5 to select System Configuration 2 Press F6 to select More 3 Press F3 to select Print Setup Report 4 Move the focus to the Complete Setup button and press the Enter key to select 5 Save the printed setup report in a secure location It can be used as a reference if the system setup needs to be restored manually I need to re activate the options on my MAC 1600 device Where can I find the option codes The option codes for the MAC 1600 device are not part of System Configuration settings However these codes will be listed on the last page of your printed setup report These codes are also found on a label next to the battery compartment The printed list of option codes is a more convenient aid to use when these codes need to be re ente
60. e guidelines may not apply in all instances Electromagnetic propagation is affected by absorption and reflection from B 7 MAC 1600 ECG Analysis System 2028451 183 structures objects and people Revision A Electromagnetic compatibility Electromagnetic compatibility EMC EMC exceptions disclosure Type Exception Electromagnetic Environment Guidance Electrostatic 4 KV air discharge for Paper Queue Floors should be wood concrete or ceramic tile If floors are covered discharge ESD sensors with synthetic material the relative humidity should be at least 30 Conducted RF KISS Lead set Recommended separation distance 0 3 Vrms KISS Lead Set 0 6 JP where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Radiated RF KISS Lead set Recommended separation distance 1 0 V m KISS Lead Set d 3 5 JP 80 MHz to 800 MHz 0 7 JP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m B 8 MAC 1600 ECG Analysis System Revision A 2028451 183 EMC compliant supplies amp accessories MAC 1600 ECG Analysis System 2028451 183 C evision EMC compliant supplies amp ac
61. e the way an ECG looks format when it prints out Follow these steps to change the ECG format 1 From the Main Menu press F5 to select System Configuration 2 Press F2 to select Resting ECG 3 Press F4 twice to select Page Down twice 4 Select a Lead Sequence and Rhythm Leads 5 Press F6 to select Save Can you edit the interpretation at the MAC 1600 device and then transmit the edited record to the MUSE system as an unconfirmed record If you edit demographic information only the record is still transmitted to the MUSE system as an unconfirmed record However if you edit the interpretation the data will not be saved unless the record is confirmed at the MAC 1600 system The record is transmitted to the MUSE system as a confirmed record as well MAC 1600 ECG Analysis System Revision A 2028451 183 Passwords Clinical Resting ECG report format Editing 3 28 Troubleshooting Freguently asked guestions Navigating the user interface O How do I navigate from the startup screen to the Main Menu A The MAC 1600 system can be configured in a number of different ways Some of these configuration choices determine the actions that need to be performed in order to proceed from the power up display to the Main Menu There are three configurations that determine the initial window that appears at power up and what actions the user will need to perform to navigate to the Main Menu Power up mode currently
62. e wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power should be that of a typical commercial or hospital environment Mains power should be that of a typical commercial or hospital environment Mains power should be that of a typical commercial or hospital environment If the user requires continued operation during power mains interruptions it is recommended that power be supplied from an applicably rated uninterruptible power supply 8 battery Power frequency magnetic fields should be at levels characteristics of a typical location in a typical commercial or hospital environment Revision A Compliance Level 2 4 6 kV indirect 2 4 6 kV direct 2 4 8 kV air Exception 2 4 kV air for Paper Queue sensors 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 Ut gt 95 dip in Ut for 0 5 cycles lt 40 Ut gt 60 dip in Ut for 5 cycles lt 70 Ut gt 30 dip in Ut for 25 cycles lt 5 Ut gt 95 dip in Ut for 5 5 3 A m Compliance Test Level 2 4 6 kV indirect 2 4 6 kV direct 2 4 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 Ut gt 95 dip in Ut for 0 5 cycles lt 40 Ut gt 60 dip in Ut for 5 cycles lt 70 Ut gt 30 dip in Ut
63. ed to close a window on the screen These keys will be on your keyboard if your system has the stress option See Keyboard stress option on page 2 7 for more information MAC 1600 ECG Analysis System 2028451 183 Name Esc key stress keys 2 6 Eguipment overview General description Keyboard stress option NOTE The English keyboard is shown in this section A B 0 D E Nex R R 8 R 8 Protest Exercise Recovery Test end Hold Speed Grade Start mill Recall 121d Medians Comment Enter BP Speed Grade STOP mill Jeb Description Press to advance to the PRETEST phase or advance to next stage within the selected phase Press to advance to the EXERCISE phase or advance to next stage within the selected phase Press to advance to the RECOVERY phase or advance to next stage within the selected phase Press and hold to end the test and start the TEST END phase Press to maintain the current stage Automatic stage sequencing will stop Press again to return to stage sequencing Press to manually increase the treadmill belt speed or the ergometer load Press to increase the elevation of the treadmill belt Press to start the treadmill during the test Press to stop the treadmill during the test Press to decrease the elevation of the treadmill belt Press to manually decrease the treadmill belt speed or the ergometer load Press to enter blood pressure readings or t
64. eft printhead holder 8 Carefully slide the printer assembly approximately 1 2 inch 13 mm to the right to provide access to the printer motor connector A After Sliding Printer Assembly 9 Lift the printer motor connector straight up to remove from the base plastic 10 Remove printer assembly from the base plastic MAC 1600 ECG Analysis System Revision A 2028451 183 4 20 Maintenance FRU replacement procedures Replacing the printer assembly 1 While holding the replacement printer assembly as shown route the ribbon cable A between the printer board and the printhead 2 Lower the printer assembly in position as shown Place the printer motor connector into its guide rails while routing the four wires in the channel A as shown If these wires are not properly routed in the channel they will be pinched upon reassembly 3 Slide the printer assembly to its final mounting position Revision A MAC 1600 ECG Analysis System 4 21 2028451 183 Maintenance FRU replacement procedures 4 Replace the first of the four mounting screws as shown Ensure the wires are not pinched Use a small needle nose pliers to hold the screw while driving it into place 5 Replace the remaining three mounting screws 6 Connect the ribbon cable to the printer board 7 Ensure the two queue sensor ribbon cables A are tucked in as shown in the following After Tucking figure
65. el Cracked swollen or leaky battery pack enclosure Debris on battery pack electrical contacts Cracked SD card Broken gold contacts MAC 1600 ECG Analysis System 3 3 2028451 183 Visual inspection Area I O connectors and cables and AC power cord Interface cables Circuit boards Ground wires wiring Fasteners Power source Keyboard LCD display filter Battery pack SD card Revision A Troubleshooting Event logging Event logging Setting Up event logging The MAC 1600 system can be set up to create an Event Log in XML format that contains system errors warnings and informational messages To configure the device for the level of severity of messages written to the Event Log follow these steps 1 Power on the MAC 1600 system by pressing the Power button From the Main Menu press F5 to select System Configuration Press F6 to select More Press F6 to select More again ine Se NO Press F5 to select Service Setup A window opens prompting you to enter the Service password Enter the Service password L 6 Type prod and press F6 to select OK NOTE If the keyboard does not include the letters prod type 7763 and press F6 to select OK The following window opens Device Settings Event Log System Diagnostics Software Update 7 Move the focus to the Event Log button and press the Enter key to select Key Eve
66. embly FRU Kit KISS Pump Hardware FRU Kit Fastener Hardware FRU Kit Fastener Hardware FRU Kit Printer Assembly FRU Kit m Middle and Base Plastics FRU Kit m Plastics FRU Kit m Fastener Hardware FRU Kit m Cable Harness FRU Kit m Mainboard ETE Assembly FRU Kit m KISS Hardware Pump FRU Kit m KISS Hardware Pump FRU Kit m KISS Pump Assembly FRU Kit m KISS Pump Assembly FRU Kit m KISS Pump Assembly FRU Kit m KISS Pump Assembly FRU Kit m KISS Pump Hardware Kit m Cable Harness FRU Kit m Power Supply FRU Kit Cable Harness FRU Kit m Acguisition Assembly FRU Kit m Acguisition Assembly FRU Kit m KISS Pump Hardware FRU Kit m Acguisition Assembly FRU Kit m KISS Pump Hardware FRU Kit m Acguisition Assembly FRU Kit m KISS Pump Hardware FRU Kit m Cable Harness FRU Kit m Cable Harness FRU Kit m Fastener Hardware FRU Kit m Display Assembly FRU Kit Revision A Parts lists Field replaceable units FRUs FRU Table 1 MAC 1600 Upper Level Assembly PN 2032093 Item Description SCR DIN7985 M3X5 4 8 Z A2F ISO7045 SCR F PLASTIC 3X10 ST A2F PANHEAD Z MIDDLE PLASTIC REAR CONNECTOR PANEL SCR DIN7985 M4X10 4 8 Z A2F ISO7045 CABLE FFC 8POS 1 00MM 180MM LONG S SIDE ASSY MAC1600 MAINBOARD AND ETE MODULE STANDOFF M3 X 10MM HEX M F STL W ZINC STANDOFF M3 X 16MM HEX M F STL W ZINC ASSY KISS PUMP PCB HOLDER PUMP CABLE STRIP 293 X 4 8 MM CABLE ASSY MAC1600 INTERNAL KISS PUMP CABLE AS
67. eplacement 4 32 R RJ11 connector 2 4 5 SD card how to insert 3 24 3 25 serial number 1 6 Service Password 3 4 3 6 software updates 4 3 speaker diagnostic test 3 9 start up screen 3 2 storage of ECGs 3 26 symbols 1 5 symbols on eguipment 1 5 system setup 3 27 system setups save to SD card 3 24 T technical specifications A 1 A 3 troubleshooting 3 1 3 28 U universal clip C 3 USB Universal Serial Bus 2 4 UUT Unit Under Test 4 41 Ww warranty lithium ion battery A 3 MAC 1600 system A 3 Windows Explorer 2 5 XML Event Log export 4 K keyboard standard non stress 2 5 stress 2 7 keyboard replacement 4 8 KISS pump connector 2 3 replacement 4 17 LAN cable C 7 language Chinese A 2 Czech A 2 Danish A 2 Dutch A 2 English A 2 Finnish A 2 French A 2 German A 2 Hungarian A 2 Italian A 2 Japanese A 2 Norwegian A 2 Polish A 2 Portuguese A 2 Russian A 2 Spanish A 2 Swedish A 2 leadwire set 10 lead C 3 12 lead C 3 leakage current 2 4 lithium ion battery warranty A 3 location number 3 27 M MAC address 3 26 main board ETE module replacement 4 33 maintenance 4 1 44 modem internal C 7 paper contrast C 5 Europe C 5 premium C 5 thermal C 5 United States C 5 paper tray replacement 4 6 parts lists 5 1 5 20 password temporary 3 28 patient cable replacement 4 4 patient questions 3 27 patient trunk cable C 7 Index 2 MAC 1600 ECG Analysis System Revision A 2028451 183
68. es Ifthe test failed press F5 to select No 11 Ifthe display test failed replace the display assembly as described in Replacing the display assembly beginning on page 4 11 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Performing diagnostic tests The Speaker Test can be used to determine if the speaker is working properly Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 Select the Speaker Test button Listen for a brief audible tone coming from the speaker The following window opens Select pass or fail Ifyou heard an audible tone press F4 to select Yes Ifyou did not hear an audible tone press F5 to select No If the speaker test failed replace the mainboard ETE module assembly as described in Replacing the mainboard ETE module assembly beginning on page 4 33 The Keyboard Test can be used to determine if the keyboard is working properly Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 Select the Keyboard Test button The following window opens MAC 1600 ECG Analysis System 3 9 2028451 183 1 1 Testing the speaker Testing the keyboard Revision A Troubleshooting Performing diagnostic tests 3 Press each key on the keyboard and verify that an asterisk appears in the corresponding representation
69. eyboard m Localized alphanumeric keyboard Application Options Stress Testing ergometer sold separately Pharma package 21 CFR part 11 audit tool and CT Data Guard Languages English German Dutch French Danish Norwegian Swedish Finnish Italian Spanish m Portuguese Russian Polish Czech Hungarian Chinese simplified Japanese MAC 1600 ECG Analysis System Revision A 2028451 183 Technical specifications Specifications Specifications Height width depth 400mm 350 mm 140mm approximately Weight 6 0 Kg with battery installed approximately Power Supply AC operation Voltage 100 to 240 VAC 10 15 Current 0 5A at 115 VAC 0 3A at 240 VAC m Frequency 50 to 60 Hz 10 Battery operation Type User replaceable 14 8V 2300 mAH 10 rechargeable Lithium lon m Capacity 50 single page reports typical or three hours continuous display without printing Charge time Approximately 2 5 hours from total discharge with display off Temperature Operational 10 C to 40 C Storage 40 C to 70 C Relative Humidity Operational 20 95 non condensing Storage 15 95 non condensing Pressure Operational 700 1060 hPA Storage 500 1060 hPA Warranty 3 years Lithium lon Battery Warranty 3 months Cer
70. fails Press Esc or F6 Cancel after the test results are displayed If the test failed replace the mainboard ETE module as described in Replacing the mainboard ETE module assembly beginning on page 4 33 MAC 1600 ECG Analysis System Revision A 2028451 183 1 3 16 Troubleshooting Performing diagnostic tests Testing the modem The Modem Test can be used to test the internal and the external modem 1 Connect the MAC 1600 device to an active analog phone line NOTE Ensure that the phone line is active If you connect to an inactive phone line the test result may be a false negative 2 _ Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 3 Select the Modem Test button The following window opens Internal Modern Test External Modem Test 4 Test the internal modem as described in the following steps a With the focus on the nternal Modem Test button press the Enter key The Test in Progress Please wait message is displayed Then the results of the test are displayed Ifthe Passed message is displayed in the window the test passes Ifthe Failed message is displayed in the window the test fails b Ifthe system has an external modem go to step 5 If the system has only the internal modem option press F6 to select Cancel Then go to step 6 5 Test the external modem as described in the following steps a With t
71. for 25 cycles lt 5 Ut gt 95 dip in Ut for 5 s 3 Alm MAC 1600 ECG Analysis System 2028451 183 immunity Immunity Test Electrostatic discharge ESD IEC 60601 1 2 2007 4 IEC 61000 4 2 2001 Electrical fast transient burst EFT IEC 60601 1 2 2007 4 IEC 61000 4 4 2004 Fast Transient Surge FTS IEC 60601 1 2 2007 IEC 61000 4 5 2005 Voltage dips short interruptions and voltage variations on power supply input lines IEC 60601 1 2 2007 IEC 61000 4 11 2004 Power frequency 50 60 Hz magnetic field IEC 60601 1 2 2007 IEC 61000 4 8 2001 NOTE U is the AC mains voltage prior to application of the test level B 4 Electromagnetic compatibility Electromagnetic compatibility EMC Guidance and manufacturer s declaration electromagnetic The MAC 1600 system is intended for use in the electromagnetic environment specified in the following table It is the responsibility of the customer or user to assure that the MAC1600 system is used in such an environment Electromagnetic Environment Guidance Portable and mobile RF communications eguipment should not be used closer to any part of the eguipment or system including cables than the recommended separation distance calculated fro the eguation applicable to the freguency of the transmitter Recommended separation distance Recommended separation distance Multilink Lead Set d 12 VP Value Lead
72. gram Hardware firmware architecture Troubleshooting General fault isolation Power up self test Poor quality ECGs Visual inspection MAC 1600 ECG Analysis System 2028451 183 Revision A Revision A Event logging Y Y EE EE RR Setting Up event logging gt Exporting the event log ce ee eee Performing diagnostic tests Accessing the system diagnostics function Testing the 01908 naaa Testing the speaker Testing the keyboard Testing the acquisition module Checking battery status Testing the writer Testing the RS232 port Testing the LAN Testing the modem Testing the USB port ue Testing the patient lead wires Equipment 00 68 gt cece ee uu Error CodeS fixie iu a leeds wie WA vale RR Acquisition error codes Printer error 00085 eee Frequently asked questions Y Y YY ug System setup YVVYY YF eee ST ec ME EFE NC FFF EN FFYN HFAS Navigating the user interface Maintenance Introduction ie etek y seen keds rb Recommended maintenance Required tools and supplies
73. harged and the conditioning cycle is complete MAC 1600 ECG Analysis System Revision A 2028451 183 l 2 4 44 Parts lists MAC 1600 ECG Analysis System 2028451 183 5 Revision A Parts lists Ordering parts Ordering parts General information The FRU parts lists in this chapter supply enough detail for you to order parts for the assemblies stand alone FRUs and FRU kits considered field serviceable Only items assemblies and kits which have part numbers given in this chapter are available for purchase as FRUs To order parts contact GE Healthcare Service Parts 5 2 MAC 1600 ECG Analysis System Revision A 2028451 183 MAC 1600 ECG Analysis System 2028451 183 5 3 Revision A MAC 1600 upper level assembly PN 2032093 Field replaceable units FRUs Parts lists Field replaceable units FRUs MAC 1600 ECG Analysis System 2028451 183 Revision A 5 4 Parts lists Field replaceable units FRUs MAC 1600 ECG Analysis System 2028451 183 Parts lists Field replaceable units FRUs Parts lists Field replaceable units FRUs See Detail A Xo Detail A 5 6 MAC 1600 ECG Analysis System Revision A 2028451 183 Parts lists Field replaceable units FRUs Detail A MAC 1600 ECG Analysis System 2028451 183 Revision A Included in KISS Pump Ass
74. he focus on the External Modem Test button press the Enter key The Test in Progress Please wait message is displayed Then the results of the test are displayed Ifthe Passed message is displayed in the window the test passes Ifthe Failed message is displayed in the window the test fails b Press Ese or F6 Cancel when the test is complete 6 Ifthe Internal Modem Test failed replace the internal modem as described in Replacing the internal modem option on page 4 12 7 Ifthe External Modem Test failed replace the external modem Revision A MAC 1600 ECG Analysis System 3 17 2028451 183 Troubleshooting Performing diagnostic tests The USB Test can be used to test the USB port Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 Connect a USB keyboard to the USB port of the MAC 1600 back panel NOTE The USB keyboard used for this test must match the language that is selected in setup Select the USB Test button The following window opens lt lt Connect an external keyboard through the USB part Press any ASCII character to do the test gt gt Character Input Ei Press any key on the USB keyboard and verify pass or fail Ifthe character that appears in the Character Input field matches the key you pressed the test is passed Ifthe character does not match or no character appears in the Character Input field the
75. iated RF disturbances are controlled The customer or the user of the MAC 1600 system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications eguipment transmitters and the MAC 1600 system as recommended in the following table according to the maximum output power of the communications eguipment 800 MHz to 2 5 GHz 06 3 JP 0 23 0 73 23 3 23 smitter for KISS Lead Sets 800 MHz to 2 5 GHz d 23 JP 0 70 22 7 0 22 70 80 kHz to 800 MHz d 1 2 JP 0 12 0 38 1 2 3 8 12 80 kHz to 800 MHz d 1 2 JP 0 35 1 1 3 5 11 35 Separation distance meters according to frequency of transmitter for Multilink and Value Lead Sets 150 kHz to 80 MHz d 1 2 JP 0 12 0 38 1 2 3 8 12 Separation distance meters according to frequency of tran 150 kHz to 80 MHz d 11 6 JP 0 12 3 7 11 36 116 Rated maximum output power of transmitter WwW 0 01 0 1 100 Rated maximum output power of transmitter Ww 0 01 0 1 1 10 100 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can estimated using the equitation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts w according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 Thes
76. icable checkout procedures Refer to Functional checkout on page 4 38 Revision A MAC 1600 ECG Analysis System 4 31 2028451 183 Maintenance FRU replacement procedures Replacing gueue mark sensor The gueue mark sensor performs the paper gueuing function if you are using the thermal paper with gueue marks Repeat steps 1 through 5 in Removing the printhead on page 4 26 Remove the gueue sensor holder A from the printhead holder B as described in the following two steps a Push on the queue sensor holder A with your thumb until its locking tabs C move as shown in the following After push figure Before push s b To complete the removal of the queue sensor grab the plastic body of the sensor and pull it to remove from the printhead holder After push Reverse all remaining disassembly steps to complete the reassembly Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 MAC 1600 ECG Analysis System Revision A 2028451 183 l 2 4 32 Maintenance FRU replacement procedures Replacing the mainboard ETE module assembly Processing ECGs in internal storage If system has the internal storage option process any ECGs remaining in storage by transmitting to your archival system and or printing them to ensure you have a printed record before proceeding with mainboard replacement Saving system configuration
77. ies safety standards and or the system configuration must meet the reguirements of the IEC 60601 1 1 medical electrical systems standard The use of accessory eguipment not complying with the eguivalent safety reguirements of this eguipment may lead to a reduced level of safety of the resulting system Consideration relating to the choice shall include use of the accessory in the patient vicinity and evidence that the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601 1 and or IEC 60601 1 1 harmonized national standard Eguipment symbols See the MAC 1600 operator s manual for information about the symbols used on this product and its packaging Revision A MAC 1600 ECG Analysis System 1 5 2028451 183 Introduction Service information Service information Service reguirements Refer eguipment servicing to GE authorized service personnel only Any unauthorized attempt to repair eguipment under warranty voids that warranty It is the user s responsibility to report the need for service to GE or to one of their authorized agents Eguipment identification Label format p n lt Product p n gt lt Product Code Desc gt YYYY MM SN lt Serial Number gt Table 2 Equipment Identification Label Item Description A Date of manufacture in YYYY MM format B Part number of product 0 Product code description D Serial number See Serial number format
78. l accessories Description Patient trunk cable 12 leads NEHB IEC Electrode Application System KISS 10 Lead w o pump Electrode Application System KISS 10 Lead with pump Electrode Application System KISS 12 Lead w o pump Cable Modem MAC1200 350MM Cable Modem MAC 1200 1M Connection cable for Lode Excalibur Connection cable for Variobike 500 Connection cable for eBike Connection cable EC 560 ECB561 or MAC 1100 MAC 1200 to CardioSys Connection cable for T2000 T2100 Connection cable for TMX425 Cable Connect CSmart Modem 5 Wire 5M Cable Connect CardioSmart Modem Barcode Reader Barcode Reader Cable LAN Cable Internal Modem External VGA Cable MAC 1600 ECG Analysis System 2028451 183 Part Number 22340305 21612113 21612413 21612202 2008683 001 2008683 002 2006796 001 2006795 001 22338002 22336203 22339003 22336802 22337801 22337802 2031239 001 2035628 001 3700031 002 2038704 001 416070 001 Revision A EMC compliant supplies amp accessories Optional accessories For your notes C 8 MAC 1600 ECG Analysis System Revision A 2028451 183 Index UL 60601 1 2006 A 3 connection cable eBike C 7 Lode Excalibur C 7 MAC 1200 C 7 T2000 C 7 T2100 C 7 TMX425 C 7 Variobike 500 C 7 cords power C 6 current leakage tester 4 2 D diagnostic test display 3 7 speaker 9 display replacement 4 11 accessories
79. n Archivist 25 Red Grid with Blank Interpretation Area 8 5 x 11 sheet Queue hole 150 sheets pack 2400 sheets case Archivist 25 Red Grid with Blank Interpretation Area A4 Full size Queue hole 150 sheets pack 2400 sheets case MAC 1600 ECG Analysis System C 5 2028451 183 Part Number 22616605 22616606 Part Number 2020314 001 9402 060 9402 061 Part Number 22616611 Part Number 2026048 002 2026048 004 Revision A Revision A EMC compliant supplies amp accessories Country specific power cords Country specific power cords Description Europe PWR SPLY CRD RA CONT EURO 10A 250V 2 5M British PWR SPLY CRD RA BRITSH 10A 250V 2 5M FSD Italian PWR SPLY CRD RA ITALIAN 10A 250V 2 5M Israel PWR SPLY CRD RA ISRAELI 10A 250V 2 5M Australia PWR SPLY CRD RA AUSTRALIAN 10A 250V 2 5M China PWR CRD ST CHINA RA PLUG 10A 250V 2 5M Swiss PWR SPLY CRD ST SWISS 10A 250V 2 5M India PWR SPLY CRD ST INDIAN 10A 250V 2 5M US PWR SPLY CRD RA HOSP GRD 13A 125V 10FT Japan PWR SPLY CRD ST RA PSE 10A 250V 10FT MAC 1600 ECG Analysis System 2028451 183 Part Number 401855 001 401855 002 401855 003 401855 004 401855 010 401855 018 401855 107 401855 108 405535 006 405535 011 C 6 C 7 EMC compliant supplies amp accessories Optional accessories Optiona
80. nnect the patient cable from the MAC 1600 side panel connector as shown in the following figure 3 Connect a new patient cable to side panel connector 4 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 4 4 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures Replacing barcode reader or barcode reader cable 1 Power off the system and disconnect from AC power 2 Disconnect the barcode reader from the USB connector on the MAC 1600 back panel as shown in the following figure 3 Ifonly the cable is to be replaced disconnect the cable from the barcode reader as shown in the following figure a Insert an Allen wrench or straightened paper clip in the small hole in the base of the barcode reader b While pushing the tool into the hole pull the cable to remove it from the base of the barcode reader 4 Reverse the disassembly procedures to reassemble using a new Insert USB connector with Y the USB symbol facing down 5 Configure the new barcode reader as described in the MAC 1600 Operator s Manual 6 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 Revision A MAC 1600 ECG Analysis System 4 5 2028451 183 Maintenance FRU replacement procedures Replacing the paper tray assembly 1 Disconnect the system from AC power 2 Press down on the paper
81. nt Lagging Log Level Ene A 3 4 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Event logging Enable or disable event logging Check the Key Event Logging check box to enable writing to the Event Log or clear the Key Event Logging check box to disable writing to the Event Log Select a level of severity to log from the Log Level list Select Error to log only errors to the Event Log Select Warning to log errors and warnings to the Event Log Select Information to log errors warnings and information messages to the Event Log Press F6 to select Save Repeat steps 1 7 in Setting Up event logging Insert an SD card in SD card slot in back panel as shown in the following figure with gold contacts up Press F1 to select Export Log Files The current Event Log file log_0 log is copied to a log directory on the SD card NOTE The log file can be accessed by inserting the SD card in an SD card reader that is connected to a computer with a Windows operating system and a text editor like Notepad or WordPad If the Event Log is requested by GE Service for troubleshooting an issue the file can be sent as an email attachment MAC 1600 ECG Analysis System 3 5 2028451 183 10 Exporting the event log Revision A Troubleshooting Performing diagnostic tests Performing diagnostic tests Verify that the MAC 1600 resting ECG analysis sys
82. o trigger a reading from an external device Press to enter comments about the test Comments are printed on many of the final reports Press to print a medians report Press to print a 12 lead report 10 seconds of acquired data Press to print a one page rhythm strip using the previous 10 seconds of data MAC 1600 ECG Analysis System 2 7 2028451 183 Name Pretest key Exercise key Recovery key Test end key Hold key Speed W key Grade key Start tmill key STOP tmill key Grade 4 key Speed W key Enter BP key Comment key Medians key 12Id key Recall key mono gt I 7 lt A Revision A MAC 1600 ECG Analysis System 2028451 183 Revision A 2 8 3WVINI 303 ASSY dWNd SSIM SIY 1 A 1GW3SSV dANd SSIM CWDDOYA YOLDINNOD A1dd S H3MOd gt e Sc 4 lt nn N midds Hanga gt Pa 9M EIM A LM wog 13431 H3HOIH 03 100 438 XXX 0v26202 N d ASSY
83. of that key on the screen A key passes the test if an asterisk appears on the screen when the corresponding key is pressed 4 To test for sticky keys continue to press keys that already have asterisk on the screen and verify that the screen representation of the key is highlighted with each subseguent key press A key passes if the key on the screen highlights with each repeated key press 5 When the test is complete press the Shift F6 keys The following window opens Did the test pass 6 Select pass or fail If every key passes the tests in steps 3 and 4 press F4 to select Yes If any key fails the tests in steps 3 or 4 press F5 to select No 7 Ifthe keyboard test failed replace the keyboard assembly as described in Replacing the keyboard beginning on page 4 8 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Performing diagnostic tests Testing the acguisition module The Acguisition Module Test can be used to determine if the acguisition board is working properly 1 Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 2 Select the Acquisition Module Test button A window similar to the one shown in the following figure opens Acquisition Module Test Result Passed Acquisition Firmware Version 1 Acquisition Hardware Version 3 Note the test result and press F6 to select Cancel 4 Ifthe res
84. on 1 Check cables driver 2 Self correcting internal error 3 Reload Application software Acquisition Error 3 Acquisition buffer overflow 1 Check cables 2 Replace CABLE ASSY MAC1600 ACQ TO MAINBOARD found in Cable Harness Kit PN 2035707 3 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 4 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE Acquisition Error 2 Acquisition module not connected or 1 Check cables detected 2 Replace CABLE ASSY MAC1600 ACQ TO MAINBOARD found in Cable Harness Kit PN 2035707 3 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 4 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE Acquisition Error 1 General Acquisition error 1 Check cables 2 Replace CABLE ASSY MAC1600 ACQ TO MAINBOARD found in Cable Harness Kit PN 2035707 3 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 4 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE Acquisition Error 2 Synchronization error in 1008 ECG 1 Check cables Packet 2 Replace CABLE ASSY MAC1600 ACQ TO MAINBOARD found in Cable Harness Kit PN 2035707 3 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 4 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE MODULE Acquisition Error 3 Sequence Number error in 100ms 1 Check cables ECG Packet 2 Replace CABLE ASSY MAC1600 ACQ TO MAINBOARD found in Cable Harness Kit PN 2035707 3 Replace 2035705 001 ASSY ACQUISITION BOARD MAC 1600 4 Replace 2035704 001 ASSY MAC1600 MAINBOARD AND ETE
85. on page 1 7 1 6 MAC 1600 ECG Analysis System Revision A 2028451 183 Introduction Service information Every GE device has a unigue serial number for identification The serial number appears on the device label and is formatted as seen in the following illustration HHH HH HH HERH H A B 0 D A The product code for MAC 1600 systems is SDE B Year Manufactured 00 99 08 2008 09 2009 and so on Fiscal Week Manufactured Production Seguence Number Manufacturing Site Miscellaneous Characteristic O MAC 1600 ECG Analysis System 1 7 2028451 183 Serial number format Revision A Introduction Service information For your notes 1 8 MAC 1600 ECG Analysis System Revision A 2028451 183 Equipment overview MAC 1600 ECG Analysis System 2028451 183 2 evision General description The MAC 1600 ECG Analysis System is a 12 lead 12 channel system with a 6 5 inch 165 mm diagonal display active patient cable and battery operation There are also options for communication capabilities Front View Name Description A Display Presents waveform and text data B Power LED Indicates the unit is plugged in and receiving power 6 Battery LED Indicates various battery states m Solid amber light indicates the battery is charging m Flashing amber light indicates the battery is low m Off indicates the battery is neither cha
86. optional C 7 standard C 3 value C 4 acg module diagnostic test 3 11 acguisition board replacement 4 15 acguisition cable connector 2 3 acguisition error codes 3 21 anti static wrist strap 4 2 B barcode reader cable C 7 barcode reader replacement 4 5 battery C 3 battery capacity 3 26 battery LED 2 2 battery replacement 4 7 block diagram 2 8 E ECG data noise 3 20 6 electrode prep 0805 C 3 cable barcode reader C 7 electrode spray C 3 cable LAN C 7 electrodes CAM 14 leadwire set BabyMAC C 3 banana C 3 Silver Mactrode Plus C 3 mixed Iength C 3 suction C 3 certifications equipment overview 2 1 2 10 AAMI EC11 equipotentional grounding lug 2 4 1991 A 3 error codes 3 21 CAN USA 622 2 No 601 1AAMI EC57 A 3 Event Log export 3 4 IEC 55011 Extra Questions 3 27 1998 A 3 IEC 60601 1 F 1988 A1 function keys 2 5 1991 12 G 1995 A 3 grounding 2 4 IEC 60601 1 1 2000 A 3 H IEC 60601 1 2 Hook Up Advisor 1 2 2001 A 3 Hookup Advisor 3 20 A 2 IEC 60601 1 4 how to 2000 A 3 save system setup to SD card 4 33 IEC 60601 2 25 update software 4 3 1993 A1 i 1999 A 3 7 2003 4 3 JIST 5 2006 4 3 Revision A MAC 1600 ECG Analysis System Index 1 2028451 183 Index power cords C 6 power supply replacement 4 13 prep pads electrode C 3 printer assembly replacement 4 19 printer error codes 3 23 printhead replacement 4 26 Printhead Resistance 4 23 4 29 4 37 product code 1 7 gueue hole sensor replacement 4 30 gueue mark r
87. p button The following window opens Did the test pass c Examine the printed report Use the following criteria to determine if the writer passed or failed the 50mm s speed test Ifone cycle of the square wave spans 50 mm on paper measured from corner to corner of wave with allowable tolerance of 1 0 mm the test passes 4 Ifthe above criteria is not met the test fails d Ifthe test passed press F4 to select Yes If the test failed press F5 to select No 4 Repeat the previous step for the other speed tests The pass fail criteria for each of the remaining tests are as follows 25mm s Speed Test If one cycle of the square wave spans 25 mm on paper measured from corner to corner of wave with allowable tolerance of 0 5 mm 2 the test passes If this criteria is not met the test fails _ 12 5mm s Speed Test Measure five or more cycles and divide your measurement by the number of cycles measured If one cycle of the sguare wave spans 12 5 mm on paper measured from corner to corner of wave with allowable tolerance of 0 5 1 25 mm 10 the test passes If this criteria is not met the test fails 5mm s Speed Test Measure five or more cycles and divide your measurement by the number of cycles measured If one cycle of the sguare wave spans 5 mm on paper measured from corner to corner of wave with allowable tolerance of 0 25 0 5 mm 10 the test passes If this criteria is not met the test fails
88. p in the Stress ECG mode go to Stress ECG power up mode on page 3 31 Revision A MAC 1600 ECG Analysis System 3 29 2028451 183 Troubleshooting Freguently asked guestions These steps describe how to navigate to the Main Menu after powering on the MAC 1600 system when Resting ECG is selected for Power up mode in Basic Setup If the High Security Mode is enabled proceed with steps a d after the window opens prompting for a User JD and Password If the password prompt does not appear go to step 2 NOTE If you need to perform system setup functions be sure you log in as a user that is assigned setup editing privileges a Type your user ID in the User JD field b Press the Enter key or press the down arrow key on the Trimpad to move the focus to the Password field c Type your password in the Password field d Press the F5 key to select Login Press F5 to select Cancel Press F6 to select More Press F5 to select Main Menu 1 Resting ECG power up mode Arrhythmia power up mode These steps describe how to navigate to the Main Menu after powering on the MAC 1600 system when Arrhythmia is selected for Power up mode in Basic Setup If the High Security Mode is enabled proceed with steps a d after the window opens prompting for a User JD and Password If the password prompt does not appear go to step 2 NOTE If you need to perform system setup functions be sure you log in as a user that is
89. red after the new mainboard is installed MAC 1600 ECG Analysis System 3 25 2028451 183 O A Option codes Revision A Troubleshooting Freguently asked guestions Storing ECGs O Why won t any of the ECGs I perform save to the SD card A Check the following Check that the SD card is fully inserted into the drive Make sure you are using 256 MB or 512 MB SD cards Verify that the SD card is not write protected Try a new SD card If your system is not set up to automatically save records you must manually save by pressing Store Cleaning O Should I clean the MAC 1600 A Clean the exterior surfaces of all the equipment and peripheral devices monthly or more frequently if needed Use a clean soft cloth and a mild dishwashing detergent diluted in water Wring the excess water from the cloth Do NOT drip water or any liquid on the writer assembly and avoid contact with open vents plugs and connectors Dry the surfaces with a clean cloth or paper towel Battery capacity O What is the capacity of the battery A GE recommend that the MAC 1600 be connected to AC power through a wall outlet whenever it is not in use However if operating the device off AC power be aware that a fully charged battery is capable of taking approximately 50 or more ECGs with one page printed reports or three hours of continuous operation without printing MAC address O Ineed to provide the MAC
90. rging nor low D Keyboard Input device for controlling the system or entering data See Keyboard Layout on page 2 4 for more information E Writer Prints reports 2 2 MAC 1600 ECG Analysis System Revision A 2028451 183 Eguipment overview General description Side View A B 6 Name Description A ECG signal input D sub 15 pin female connector for the acguisition connector cable B KISS Connector Connection port for the optional KISS lead system C Carrying handle Handle for carrying the MAC 1600 device Back View Name Description A External Power 12V power supply for future external devices Connector Do not use 8 COMM Port A Serial connector for stress devices bicycle ergometer or treadmill C COMM Port B Serial connector for data communication with CASE CardioSoft or MUSE system or connect to an external modem Revision A MAC 1600 ECG Analysis System 2 3 2028451 183 Eguipment overview General description Name Description D Phone jack RJ11 connector from the intemal modem to an analog phone line E AC power connection Standard connector for the AC power cable F USB connector Standard Universal Serial Bus connector for USB devices such as the optional barcode reader or external USB keyboard G LAN connection RJ45 network connector H D card slot Secure Digital card slot Insert c
91. selected in Basic Setup System Settings Resting ECG v 1 Resting ECG Display Colors Arrhythmia Main Screen ECG Glo tress ECG Anti Aliasina of ECG W Power up mode High Security Mode enabled in Basic Setup System Security Setup High Security Mode 6 elit Teall E1 USB Barcode Reader support option activated yes or no Activated Options Description CTDG CT Data Guard R12L 12 lead resting waveform display MI12 Measurement and 12SL Interpretation M300 Internal storage 300 Resting ECGs LANC LAN to CardioSoft LANM LAN to MUSE MODC Modem or Serial to CardioSoft MODM Modem or Serial to MUSE ERGO Stress test with treadmill bicycle or Maste E12L 12 lead Stress test waveform display CFRA 21 CFR Part 11 audit trail gt gt BCRD USB Barcode Reader support TIPI ACI TIPI The various steps in this section describe how to navigate from the power up screen to the Main Menu for the various system configurations Use the steps that apply to your system configuration settings Ifyour system is configured to power up in the Resting ECG mode go to Resting ECG power up mode on page 3 30 Ifyour system is configured to power up in the Arrhythmia mode go to Arrhythmia power up mode on page 3 30 Ifyour system is configured to power up in the Main Screen mode go to Main Screen Power Up Mode on page 3 31 If your system is configured to power u
92. stalled a Insert the SD card with saved system setups in SD card slot b From the Main Menu press F5 to select System Configuration Press F6 More twice a e Press F2 to select Jmport Setup Highlight the system setup file you want to import from the SD card rn o Press F1 to select Import g When the Configuration was successfully imported message is displayed press F6 to select OK h Remove the SD card and store in a secure location 20 Configure the Device Settings a From the Main Menu press F5 to select System Configuration b Press F6 More twice Press F5 to select Service Setup a io Type prod when prompted to enter the Service Password NOTE If the keyboard does not include the letters prod type 7763 and press F6 to select OK e Select the Device Settings button Serial Number Printhead Resistance ooo 000 Keyboard English 2 f Type the serial number of the device in the Serial Number field See Equipment identification on page 1 6 for information about the serial number g Type 1000 in the Printhead Resistance field h Select the appropriate language from the Keyboard list i Press F6 to select Save 21 Re activate the purchased options See the MAC 1600 ECG Analysis System Operator s Manual for option activation instructions 22 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 Revision A MAC
93. sults of the COM 1 loop back test 5 Perform the COM Port loop back test on COM 2 a Use the paper clip to short pins 2 amp 3 on the COM B connector b With the focus on COM 2 press the Enter key The results of the COM 2 loop back test are displayed c Note the results of the COM 2 loop back test 6 Press Ese or F6 Cancel when the test is complete 7 If either test failed replace the mainboard ETE module as described in Replacing the mainboard ETE module assembly beginning on page 4 33 Revision A MAC 1600 ECG Analysis System 3 15 2028451 183 Troubleshooting Performing diagnostic tests Testing the LAN option The LAN Test can be used to test network connectivity Connect the MAC 1600 device to an active LAN NOTE Ensure that the LAN is an active network If you connect to an inactive network tap the test result may be a false negative Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 Select the LAN Test button The following window opens Press the Enter key to select the Test network connectivity button The Checking connectivity Please wait message is displayed Then the test results are displayed Ifthe System Connected to Network message is displayed in the window the test passes Ifthe Network Unavailable message is displayed in the window and you are sure the device is connected to an active network the test
94. tem operates properly by running the diagnostic tests These tests check the operation of the display screen speaker keyboard thermal writer battery and communications These diagnostic tests are a useful tools for troubleshooting problems and can be useful as a part of system checkout procedures Accessing the system diagnostics function The System Diagnostics menu can be used to perform functional diagnostic tests Access the System Diagnostics menu as follows 1 Power on the MAC 1600 system by pressing the Power button From the Main Menu press F5 to select System Configuration Press F6 to select More Press F6 to select More again GN Oe o Press F5 to select Service Setup A window opens prompting you to enter the Service password Enter the Service password 6 Type prod and press 16 to select OK NOTE If the keyboard does not include the letters prod type 7763 and press F6 to select OK The following window opens Device Settings Event Log System Diagnostics Software Update 7 Move the focus to the System Diagnostics button and press the Enter key to select 3 6 MAC 1600 ECG Analysis System Revision A 2028451 183 Troubleshooting Performing diagnostic tests The DIAGNOSTIC TESTS window opens The following sections describe how to perform the specific diagnostic tests Proceed to the appropriate section for the test you need to perform Testing the
95. test fails Press Esc or F6 Cancel when the test is complete The following window opens Did the test pass Press F4 to select Yes if the test passed or press F5 to select No if the test failed If the test fails replace the mainboard ETE module as described in Replacing the mainboard ETE module assembly beginning on page 4 33 MAC 1600 ECG Analysis System Revision A 2028451 183 1 Testing the USB port 3 18 Troubleshooting Performing diagnostic tests Testing the patient lead wires Test the patient leadwires as described in this section 1 Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 2 Connect a patient cable with lead wires to the MAC 1600 patient cable connector 3 Connect all leads to a patient simulator or shorting bar 4 Select the Patient Lead Wire Check button The following window opens 5 Press the Enter key to select the Start Test button For each lead wire the test results are displayed 4 Ifthe Connected message is displayed the lead wire passes the test Ifthe Disconnected message is displayed the lead wire fails the test 6 Press Ese or F6 Cancel when the test is complete 7 Replace every lead wire that failed the test 8 Repeat the test If the lead wire still fails the test replace the acguisition board as described in Replacing the acquisition board assembly beginning on page
96. the document part number and revision letter at the bottom of the page The revision letter identifies the document s update level The revision history of this document is summarized in the following table Table 1 Revision History PN 2028451 183 Comment Initial release of document This manual supplies technical information for service representatives and technical personnel so they can maintain the eguipment to the assembly level Use it as a guide for maintenance and electrical repairs considered field repairable Where necessary the manual identifies additional sources of relevant information and or technical assistance See the operator s manual for the instructions necessary to operate the eguipment safely in accordance with its function and intended use This manual is intended for the person who uses maintains or troubleshoots this eguipment Date 10 April 2008 Manual information Revision A MAC 1600 ECG Analysis System Revision A 2028451 183 Revision history Manual purpose Intended audience Introduction Warnings cautions and notes Warnings cautions and notes The terms danger warning and caution are used throughout this manual to point out hazards and to designate a degree or level or seriousness Familiarize yourself with their definitions and significance Hazard is defined as a source of potential injury to a person Term Definition DANGER Indicates
97. the front of the keyboard bezel with the four holes in the middle plastic b Firmly push the keyboard bezel tabs into the holes c Push down on each corner at the rear of the keyboard bezel until it snaps into position 16 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 MAC 1600 ECG Analysis System Revision A 2028451 183 4 10 5 Maintenance FRU replacement procedures Replacing the display assembly Disconnect the system from AC power Remove the paper tray as described in Replacing the paper tray assembly Remove the battery as described in Replacing the battery Remove the keyboard bezel as described in Replacing the keyboard on page 4 8 Remove the four screws A that hold the display assembly in place Lift the display out of the assembly Reassemble the display by reversing the steps for removal Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 MAC 1600 ECG Analysis System 2028451 183 Fes DD o Revision A Maintenance FRU replacement procedures Replacing the internal modem option 1 Disconnect the system from AC power Remove the paper tray as described in Replacing the paper tray assembly Remove the battery as described in Replacing the battery WS Carefully note the orientation and position of the modem you are about to remove
98. tifications m IEC 60601 1 1988 A1 1991 A2 1995 General Requirements for Safety including all National Deviations IEC 60601 1 1 2000 Medical Electrical Equipment General Requirements for Safety IEC 60601 1 2 2001 General Requirements for Safety Electromagnetic Compatibility IEC 60601 1 4 2000 Requirements for Collateral Standard Programmable Electrical Medical Systems IEC 60601 2 25 1993 A1 1999 Safety of Electrocardiographs IEC 60601 2 51 2003 Safety and performance of ECG recorders m IEC 55011 1998 Radio Frequency Disturbance Limits and Methods of Test m AAMI EC11 1991 Diagnostic Electrocardiographic Devices JIST 0601 2 25 2006 equivalent to IEC 60601 2 25 1993 m UL 60601 1 2006 CAN USA 622 2 No 601 1AAMI EC57 Cardiac Rhythm and ST algorithm requirements Revision A MAC 1600 ECG Analysis System A 3 2028451 183 Revision A Technical specifications Specifications MAC 1600 ECG Analysis System 2028451 183 For your notes Electromagnetic compatibility MAC 1600 ECG Analysis System 2028451 183 B evision Electromagnetic compatibility Electromagnetic compatibility EMC Electromagnetic compatibility EMC Changes or modification to this system not expressly approved by GE Healthcare could cause EMC issues with this or other eguipment This system is designed and tested to comply with applicable
99. to prevent pinching of these cables when reassembling the main top plastic assembly bar 2 After Tucking MAC 1600 ECG Analysis System Revision A 2028451 183 4 22 4 23 Maintenance FRU replacement procedures Reverse all remaining disassembly steps to complete the reassembly Configure the Printhead Resistance From the Main Menu press F5 to select System Configuration Press F6 More twice Press F5 to select Service Setup Type prod when prompted to enter the Service Password NOTE If the keyboard does not include the letters prod type 7763 and press F6 to select OK Select the Device Settings button Serial Number i Printhead Resistance i000 Keyboard English v Type 1000 in the Printhead Resistance field Press F6 to select Save a b a f g Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 MAC 1600 ECG Analysis System 2028451 183 10 Revision A Maintenance FRU replacement procedures Replacing the printer board Removing the printer board 1 Remove the printer assembly as described in Removing the printer assembly 2 Release the top edge of the board from the printhead holder by pushing each release tab in the direction of the arrow and pulling the board back to the position shown in the following figure 3 Disconnect the queue mark sensor ribbon cable from the printer
100. to reassemble Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 MAC 1600 ECG Analysis System Revision A 2028451 183 10 11 Maintenance FRU replacement procedures Replacing the KISS pump assembly option Separate the middle plastic and base plastic assemblies as described in Replacing the power supply assembly steps 1 through 8 beginning on page 4 13 Using the following figure identify the KISS pump assembly A Turn unit over and remove the keyboard bezel Remove the two rear panel screws Lift the rear bezel and pull away from the unit in order to reach the two KISS pump hoses Pull the two KISS pump hoses from their connectors using the small needle nose pliers if necessary Disconnect and remove the keyboard as described in Replacing the keyboard on page 4 8 NOTE Removing the keyboard is recommended to minimize strain on the keyboard ribbon cable during the remaining steps in this procedure MAC 1600 ECG Analysis System 4 17 2028451 183 1 Revision A Maintenance FRU replacement procedures 8 Turn unit over and remove acquisition assembly as described in Replacing the acguisition board assembly on page 4 15 NOTE Removing this assembly provides easier access for the removal of the hose in the next step 9 Lift the acquisition assembly and remove the two screws holding the bezel to the acquisition assembl
101. ton and press the Enter key to select 6 The Extra Questions window opens Extra Questions Prompt Type alphanumeric Prompt Type alphanumeric Prompt l Type alphanumeric Prompt Type alphanumeric 7 For each extra question you wish to ask in the Patient Data window type the Prompt and select the type of question from the Type list Alphanumeric Numeric Yes No Unknown 8 Press F6 to select Save in the Extra Ouestions window 9 Press F6 to select Save in the Test Information Setup window 10 Return to the Main Menu Revision A MAC 1600 ECG Analysis System 3 27 2028451 183 Troubleshooting Freguently asked guestions The system was setup for High Security Mode and I forgot my password How do I access the system Follow these steps to access the system 1 Contact GE Tech Support and provide the serial number of the device you want to access GE Tech Support will generate a temporary device specific name and password that can only be used for 24 hours 2 Log into the system with the user ID MACService and the password provided by GE Tech Support NOTE If the keyboard on the unit does not include the letters for the MACService user ID type 6227378423 for the user ID 3 Immediately after logging into the system verify your MAC 1600 user name and password Record this information and store in a secure location for future reference How do I chang
102. ult of the Acquisition Module Test was Failed a Re seat or double check acquisition board harnesses and repeat the acquisition module test b Ifthe test still fails after re seating the board and checking the harnesses replace acquisition board as described in Replacing the acquisition board assembly beginning on page 4 15 c Repeat the acquisition module test after replacing the board to verify functionality d Ifthe test still fails after replacing the acquisition board replace the mainboard ETE module as described in Replacing the mainboard ETE module assembly beginning on page 4 33 Revision A MAC 1600 ECG Analysis System 3 11 2028451 183 Troubleshooting Performing diagnostic tests Checking battery status The Battery Test can be used to determine the status of the Lithium Ion battery This test must be performed while running on battery power 1 Disconnect the device from AC power and wait at least one minute 2 _ Open the DIAGNOSTIC TESTS window as described in Accessing the system diagnostics function on page 3 6 3 Select the Battery Test button A window similar to the one shown in the following figure opens BATTERY STATUS AC Line Status AC Power Line OFF Battery Life Remaining hh mm 2 08 Battery Charge Remaining 50 NOTE Battery Life Remaining is the estimated time remaining based on the present current draw Because no printing is occurring during this status
103. urn the system on bring the system to Standby mode or turn the system off Used to change the leads when the screen is being used to display waveforms Used to delete characters Used to acguire a resting ECG to print a 10 second report in Arrhythmia mode or to print a 12 lead report in Stress mode Used to print a continuous real time rhythm ECG rhythm strip Press the Stop key to stop the rhythm strip from printing Rhythm report is not stored and cannot be transmitted Used to stop the writer from printing Used to advance the focus in a window or to select items from the screen Press arrows to move the cursor left right up or down Press the center of the Trimpad to move the focus within a window or to select an item Used to enter a capital letter Press shift P to type a capital P Used to select menu options in the Windows Explorer Used to add a space between typed characters or to highlight screen items Used to enter special characters on non English keyboards MAC 1600 ECG Analysis System 2 5 2028451 183 Keyboard Name function keys F1 through F6 Power switch Leads key backspace key ECG key Rhythm key Stop key Enter key Trimpad shift key Alt key space bar option key BW lt Revision A Revision A Eguipment overview General description Keyboard Continued Description Us
104. us m The electrical installation of the relevant room complies with the requirements of the appropriate regulations The equipment is used in accordance with the instructions for use 1 4 MAC 1600 ECG Analysis System Revision A 2028451 183 Introduction Safety messages General The intended use of this device is to record ECG signals from surface ECG electrodes This device can analyze record and store electrocardiographic information from adult and pediatric populations This data can then be computer analyzed with various algorithms such as interpretive ECG and signal averaging for presentation to the user This device is intended for use under the direct supervision of a licensed health care practitioner Failure on the part of the responsible individual hospital or institution using this eguipment to implement a satisfactory maintenance schedule may cause undue eguipment failure and possible health hazards To ensure patient safety use only parts and accessories manufactured or recommended by GE Healthcare Contact GE Healthcare for information before connecting any devices to this eguipment that are not recommended in this manual If the installation of this eguipment in the USA will use 240 V rather than 120 V the source must be a center tapped 240 V single phase circuit Parts and accessories used must meet the reguirements of the applicable IEC 60601 ser
105. y as shown in the following figure 10 Pull the gray hose from the bezel using the small needle nose pliers if necessary 11 Disconnect the KISS pump connector from the mainboard 12 Remove the three mounting screws A 6 lt 13 Remove the KISS pump assembly 14 Reassemble by reversing the disassembly steps 15 Perform the applicable checkout procedures Refer to Functional checkout on page 4 38 4 18 MAC 1600 ECG Analysis System Revision A 2028451 183 Maintenance FRU replacement procedures Replacing the printer assembly Removing the printer assembly 1 Separate the middle plastic and base plastic assemblies as described in Replacing the power supply assembly steps 1 through 8 beginning on page 4 13 2 Using the following figure identify the printer A 3 Using the side cutter pliers snip the tie wrap B that holds the printer cable harness to the base plastic C A 4 Disconnect the gray USB cable from the printer board 5 Using a small flat blade screw driver pull the release tabs on each side of the connector to disconnect the queue sensor ribbon cable from the printer board as shown in the following figure Revision A MAC 1600 ECG Analysis System 4 19 2028451 183 Maintenance FRU replacement procedures 6 Remove the two mounting screws A that hold the right printhead holder in place 7 Repeat step 6 to remove the two mounting screws on the l
106. ystem 2028451 183 For your notes 3 1 Troubleshooting MAC 1600 ECG Analysis System 2028451 183 6 Revision A General fault isolation See the MAC 1600 Operator s Manual Chapter 2 Eguipment Overview to verify operation On power up the system automatically runs an internal self test If all tests pass you will see the start up screen MAC1600 2008 GE Healthcare All Rights Reserved The next screen that appears after the start up screen depends on the Power up mode selected in System Configuration The Resting ECG mode is the default Power up mode If the eguipment is not working properly ask yourself the following guestions Is the unit turned on Have there been any changes in the use location or environment of the eguipment that could cause the failure Has the eguipment hardware or software been modified since last use Is operator error the cause of the problem Try to repeat the scenario exactly and compare that to the proper operation of the eguipment described in the manual Is the battery installed When connected to the AC wall outlet does the green AC power light glow MAC 1600 ECG Analysis System Revision A 2028451 183 Power up self test 3 2 Troubleshooting General fault isolation Poor guality ECGs Poor guality ECGs can be caused by factors in the environment inadeguate patient preparation hardware failures related to the acguisition module

Download Pdf Manuals

image

Related Search

Related Contents

Harbor Freight Tools 94515 User's Manual  チャイルドシート貸付申請書  The Singing Machine CD Player THE SINGING MACHINE User's Manual  Amana ART308FFDM Dimension Guide  StarTech.com 24in SATA Serial ATA Cable  かがつうLED防犯灯・街路灯カタログ    SX-100K User Manual  MK500/4000 - Motorola Solutions  Jet Tools 707000 User's Manual  