FUSION CURING LIGHT
Contents
1. of areas near the pulp and soft tissues may result in irreversible or reversible damage Therefore this high performance VI curing light must only be used by trained dental professionals Do not place light directly on the same tooth segment for more than 10 seconds in the High Power mode Do not place light in direct contact with unprotected gingival oral mucous membrane or skin Prevent heat accumulation with a rubber dam in place Follow Heat Management references to reduce contact heat Place curing light directly over materials to be cured at all times If irradiation of soft tissue or sensitive patients cannot be avoided operate in Half Power Mode adjust curing distances and blow air from an air syringe or use suction tip in close proximity to the Light Head DO NOT LOOK DIRECTLY AT LIGHT EMITTED FROM THIS CURING UNIT Use suitable eye protection for the operator and assistants curing caps partial light shield at the tip of the curing light unit mount Eyeshield and Goggles reference section 10 Vil Charge batteries while not in use to extend the lifetime of the rechargeable battery Do not short circuit battery Do not use power adapters from sources other than the manufacturer Dispose of unserviceable batteries and curing lights according to the national rules and regulations Batteries must not be incinerated Replace the battery assembly after the lifetime of the battery usage by orderi
2. standard chairside or cabinet unit with half inch bar to save counter space provide easier ka access and roll over protection Ref 3e assi 6 4 LENSES Unscrew Use disposable barrier sleeves and curing caps to refa ae protect lenses from liquid and composite o damages Replace lens module immediatelv if it FRA a has cured resin b iscracked scratched or cloudv Align and lock flat edge of Lens Tool Ref 10 to flat edge of Lens Module Turn Lens Tool counter clockwise and unscrew Lens Module Clean mounting thread Replace damaged lens with a new Lens Module bv reversing the above steps 6 5 Tip Size Tip and beam size from the curing light can be adjusted using 3mm tacking and exam tip 9mm or 13 mm caps For temporary bonding snap on Tacking Tip at the end Ma of the light tip and apply pressure at the center of F a m i crown veneer for best contact Tack for 1 or 2 seconds then remove access adhesives before final bonding For small cavity treatment snap on 9mm curing cap and place light in direct contact with occlusal or bonding surface For veneer bonding and molar cavity treatment snap on 13mm fi curing cap and place light in direct contact 6 6 ADJUNCT TOOLS Three interchangeable light heads are offered for composite curing or oral exam Standard Light Head comes with the standard kit f il for LED light curing applications with l photoinitiators active to wavelengths from 420 490nm
3. Adapter Power Jack 5a 6 Barrier Sleeves 100 7 9mm Curing Cap 8 13mm Curing Cap 9 3mm Tacking Tip 10 Unit Mount Eyeshield 1 INITIAL SET UP 1 0 ASSEMBLE HANDPIECE Turn Battery Assembly 3 clockwise and screw on connect with Main Control Unit 2 A short beep upon connection indicates that 2 the battery is ready and the display screen should automatically i power on Ensure 360 swivel Light Head 1 is connected tight with Main Control 2 lt 1 3 SELECT MODE Make sure the LCD display is on If not press and hold POWER KEY for 5 seconds to turn on display similar to turning on a cell phone Press MENU KEY to select an C POWER KEY operation mode Digital Timer m Zz MODE DISPLAY DIGITAL TIMER displays operation modes for rapid HERMAL STATE ene he curing with a single click of the SMARTLIGHT O ON OFF KEY ON OFF button on single or multiple INDICATOR 5 menu key teeth ML METAL KEVPADS Constant On Count Down 5 second per beep Mode Displa X X Timing Sequence Application Single Layer Above A3 Multiple Teeth 2 OPERATION 2 1 ACTIVATE CURING Grip in either pen style a or torch style b Point light towards curing adhesives Press ON OFF Key to activate curing DIGITAL TIMER counts down in seconds Press ON OFF Key any time during the cycle to stop cycle auto recalls in 15 seconds An audible signal beeps at start during every 5 seco
4. L ITEM CODES PN 7500080 8 7500080 8P 7500080 8B 7500081 7500081B 7500081C 7500082 7500083 7500084 08 7500085 7500088 7500092 7500093 7500094 7500095 7500050 7500060 7500060W 7500060U 7500071 7500072 7500070 8500122 7500100 7500101 9200150 DentLight Inc 1411 E Campbell Rd f 500 Richardson TX 75081 USA Tel 800 763 6901 Outside US 1 972 889 8857 www dentlight com Description FUSION Curing Light Kit Silver FUSION Curing Light Kit Pink FUSION Curing Light Kit Blue Cuing Caps 9mm 50 pack Curing Caps 13mm 50 pack Tacking Tips 3mm 20 pack Barrier Sleeves 100 pack Charging Stand FUSION Battery Assembly Power Adapter Eye Goggle Unit Mount Eye Shield Focus Lens Module Expanded Beam Lens Lens Tool Main Control Module Standard Light Head Module with Focus Lens White Light Head Module with Focus Lens Violet Light Head Module with Expanded Beam Lens Chairside Charge Holder Chair Power Extension Cord FUSION CHAIR SYSTEM Loupes Fluorescence Filter FUSION to DOE Upgrade Kit DOE Starter Kit Full arch bleaching attachment pec REP 26506 Norden Germany Tel 49 0 4931 959656 Fax 49 0 4931 930813
5. NentLight o ee USER AND SERVICE MANUAL FUSION CURING LIGHT Rev 3 0 For dental professional use only U S Federal Law restricts this device to sale by or on the order of a licensed dentist Made in USA QUICK REFERENCE I Operation Set at a tvpicallv 5 second mode with MENU Kev Press ON OFF Key for curing and put back to Charging Stand when finished or at the end of each day II Recommended Curing Times The high intensity profile of FUSION enables plasma like curing speed 1 Normal composite resin 2mm A3 or lighter 5 seconds 2 Veneer Tacking l 2 seconds Bonding 4 5 seconds Perform curing test outside the mouth to 2000 determine curing times needed for specific resin sealants and bonding materials Do not overcure to avoid overheat Tip Spatial Intensity Profile 3000 o Ss 1000 mW cm2 0 1000 2000 3000 4000 5000 III Heat Management To overcome heat in extended curing times use simple cooling methods such as blowing air from an air syringe in close proximity to the Light Head and tooth being restored Use Half Power Standard Power Mode to avoid heat when needed IV Maintenance Light must be used with disposable barrier sleeves to avoid liquid damage and cross contaminations DO NOT IN ANY CIRCUMSTANCES ULTRASONIC OR AUTOCLAVE THE LIGHT Looe PARTS LIST LED Light Head 2 Main Control Unit 3 Battery Assembly 4 Charging Stand 5 Power
6. Violet Light Head optional provides violet light slih 390 430 nm for a oral lesion cancer screening b curing of standard and lower wavelength initiator adhesives and c composite visualization removal White Light Head optional offers an exam probe light that provides high power intra oral illumination and transillumination best used with tacking tip for detection of dental caries cracks and location of root canal Full arch Bleaching Attachment optional snaps at the end of the light tip to convert the light into a full arch bleaching light 7 TECHNICAL SPECIFICATIONS Voltage 100 240 V 50 60Hz to Power Adapter re Regulated 5V DC to handpiece Lamp High Power LEDs 100000 hours lifetime Wavelength Standard Light Head 420 490 nm Range Violet Light Head 390 430 nm White Light Head 5000 6500 K Light Standard Light Head Intensity 2700 mW cm 10 High Power Mode 1300 mW cm 10 Standard Power mode Battery Lithium Ion 90 minutes recharge time 35 45 minutes Shutoff Operating Temperature Storage Weight Dimension Standards continuous run time 500 full charge discharge life cycles Timer auto shutoff User shut off anytime by pressing On Off key Low battery auto shutoff Overheat auto shutoff normally after 4 minutes continuous run Self protection shutoff is accompanied by warning backlight and audible alarm as well as flashing THERMAL STATE or BATTERY BAR sym
7. bols 0 35 C Take apart Battery Assembly from Main Control Unit Storage temperature 10 40 C 111 grams 22 mm diameter 200 mm full handpiece length Complies to CE 89 336 EEC EN60601 1 2 EN61000 3 2 EN61000 3 3 IEC60601 1 2 8 LIMITED TWO YEAR WARRANTY Register your FUSION at www dentlight com DentLight Inc extends a two year warranty against defects in material and workmanship to the original purchaser of this product from an Authorized dealer except the Battery Assembly and Charging Stand one year warranty This warranty is valid providing factory inspection indicates that any such defect developed during normal and proper use DentLight shall under no circumstances be liable for incidental special consequential damages and damage from misuse including the following expressed or implied Improper operation accident or neglect misuse or abuse such as physical damage drop dents cracks scratches etc Liquid water rain extreme humidity ultrasonic autoclaving Improper repairs or modifications that have not been carried out by a certified Service Center Use of other spare parts and accessories than from the manufacturer such as power adapter Other acts which are not the faults of DentLight In case of a warranty claim the complete apparatus must be returned carriage paid to your dealer or directly to DentLight with the proof of purchase Call 1 800 763 6901 to obtain RMA number th
8. en fill out RMA form online with the approved RMA number 9 TROUBLESHOOTING GUIDE TROUBLE CAUSES Handpiece Light cannot be Unit Power is not turned on activated Light cannot be Batterv icon shows one bar or no activated bar Light cannot be Handpiece is overheated activated Light cannot be Bad electrical contacts and activated or shuts down connection between the Batterv prematurelv Assembly and Main Control Unit Damaged connection pin Battery Assembly Misunderstanding or electronic defect Battery not charging Battery not holding Electronic defect in the Battery charge Assembly Battery not being Electronic defect in the Charging charged Stand Main Control Unit Keypad malfunctions Electronic defect from liquid Unable to turn on LCD damage short circuits and drop screen set timer or turn damage on light ERROR INDICATION LCD display is blank LCD Battery Indicator is low or LCD is blank and can not be turned on with Unit Power Key The handpiece in particular the Light Head is hot Shuts down even when battery is full Shows digital timer count down and hears beeps Light works inconsistently Charge Indicator shows no light while charging from Charging Stand or directly plugging into the handpiece Operating inconsistently Shuts down even when Assembly Drop damaged Battery battery icon shows full bar and battery screwed in tight Battery Assembly end with drop mark and loose c
9. nds and completion of cycle An audible beep signals key activation Power Menu or ON OFF together with visual backlight signals for MENU and ON OFF key activation 2 2 POWER MODE Default power mode is plasma like High Power Mode for quick and deep curing Press Power Key in any modes after activating curing to switch to Standard LED Power Mode Half Power mode when heat needs to be minimized Ref 4v Press Power Key again to switch back to High Power Mode 3 BATTERY MAINTENANCE Battery Assembly uses a high capacity Lithium Ion battery with built in smart recharging electronics In the office keep unit display on and keep light charged at all times when not in use by one of the following method c Place the handpiece in the Charging Stand d Direct plug in the l Ja Adapter Power Jack c d from the Power Adapter to the receptor U l L fae ee u f2 operation in the event e of low battery during N f 9 i te aw D om end of the handpiece operation Switch to this corded e Place the handpiece into a Dental Unit Charge Holder optional accessory f which can be mounted on a standard dental unit bar or cabinet bar g via its two bottom set screws The Dental Unit Charge Holder connects to Power Adapter through its outlet power jack f1 and extension cord f2 SMARTLIGHT INDICATOR is green while charging and turns off when fully charged Battery icon displays three full ba
10. ng replacement Battery Assembly Ref 10 5 CLEANING AND DISINFECTING a b c d FUSION parts are recommended to be cleaned and disinfected with a cleaning solution including Cavicide Lysol Isopropyl or Ethyl alcohol and equivalent for 10 minutes Do not autoclave or ultrasonic Thoroughly wipe clean handpiece Charging Base and Eyeshield with a soft cloth or tissue damped with the cleaning solution Do not spray or soak Disinfecting agents and liquid must not enter any openings on unit in particular METAL KEYPAD Clean curing lens and electrical metal contacts using Q tips slightly damped do not soak with the cleaning solution Safeguard handpiece opening area metal keypad charging plug and lens against liquid and sharp objects as this may cause electrical short circuit permanent optical damage or malfunction 6 1 LIGHT POSITIONING Rotate 360 swivel Light Head for easy access to lower or upper Place light in close proximity to composite Use curing caps optional disposable parts recommended per patient use as anchor to ease positioning and protect lens from composite damage 6 2 BARRIER SLEEVES Slide sleeve into handpiece Twist body relative to sleeve for tight fit iaa 6 3 EYESHIELD dan Slide eyeshield through the Light Head into ee handpiece body Adjust sliding position for easier access 6 4 CHAIRSIDE CHARGE HOLDER This option provides the user a convenient mounting and charging option on a
11. omponent inside Battery icon shows one or two bars Charge Indicator shows no light while charging Charge Indicator turns green while bypassing Charging Stand and directly plugging into the handpiece After charging Battery Assembly LCD screen still does not turn on with POWER KEY or can not set timer with MENU KEY or can not activate curing light with ON OFF key Ref 1 3 CORRECTIVE ACTION Press and Hold Unit Power Key until the LCD display turns on Then select a mode and press On Off Key Ref 1 3 Charge the battery If LCD is blank charge the battery assembly separately overnight by first unscrewing the battery assembly from the main control unit and then plug in the power adapter directly Ref 3 Allow the handpiece to cool down in ambient or with convection air from suction tip or compressed air Ref 7 Clean metal contacts Ensure good and tight connections In case of damaged electrical pins contact the manufacturer for service When fully charged charge indicator goes out and the battery icon shows three bars If not please contact manufacturer or your dealer for service or replacement of Battery Assembly Please contact manufacturer or your dealer for service or replacement of Battery Assembly Please contact manufacturer or your dealer for replacement of Charging Stand Contact the manufacturer or your local dealer for service or replacement of Keypads or Main Control 10 REORDER AND OPTIONA
12. rs when the battery is fully charged The unit warns with short audible beeps when battery is low then automatically shuts down with a long beep together with flashing backlight and battery bar For temporary storage or transport shut down the unit by pressing and holding Power Key for 5 seconds until display is turned off indicated with a short beep For extended period storage unscrew battery assembly and storage with care Do not store above 40 C or below 10 C Storage period must not exceed 6 months In the event after long storage over 1 month or there is no audible beep upon battery assembly connection with main control charge Battery Assembly overnight either direct or on Charging Stand from Adapter Power Jack Plug Power Adapter aoe into a wall plug or power plug 4 SAFETY PRECAUTIONS 1 Light must be used with barrier sleeves to avoid cross contaminations and liquid damage REPLACE THE DISPOSABLE BARRIER SLEEVES AFTER EACH PATIENT USE 11 Do not touch and open any electronics parts and circuit boards Warranty void if seal broken 111 Do not drop the light Use instrument tray Charging Stand or Eyeshield to avoid roll over on table WARRANTY VOID IN CASE OF DROP DAMAGE IV Do not immerse handpiece in waters or solutions WARRANTY VOID IN CASE OF LIQUID DAMAGE V As it is the case with all high performance lights the high light intensity of FUSION results in a certain heat development Prolonged exposure