Home

85052B 3.5mm Cal Kit Service Manual

image

Contents

1. System Zo Calibration Kit Label N N N I I Frequency u L L L Standard w M 9 Offset in GHz i N m g 3 3 3 x x x x o ri N m 9 Oo U 9 U kz o 3 2 N N N E D a I I I ke w gt G r I I I FIL G g 2 N 4 m a CG o Sou ri N m P c Sa P b b b b Se 5 85 2 o ri ri ri ri 2 x 3 E a x x x X E 5 a e c 8 8 5 o 4 X E z E a a 3 i L A N 3 z z 0 a 1 2 3 4 5 6 7 8 Ensure system Zy of network analyzer is set to this value Open short load delay thru or arbitrary impedance Load or arbitrary impedance only For waveguide the lower frequency is the same as F co ano 85052B A 13 Standard Definitions Nominal Standard Definitions Table A 12 Blank Form for the PNA Series of Network Analyzers System Zo Calibration Kit Label Standard C0x10 18 F C1x10 2 F Hz L2x10 23 H Hz C2 10 3 F Hz L3 x10 H HZ C3 x10 4 F Hz LOx10 2 H L1 10 H Hz Fixed or sliding Type Offset Frequency in GHz Delay in ps ZoQ Loss in GQ s Min Max Coax or Waveguide Standard Label ol oo wm a wi n e Number a Ensure system Z of network analyzer is set to this value b Open short load delay thru or arbitrary impedance C For waveguide the lower frequency is the same as F co A 14 85052B Index A accuracy c
2. s a 1 Short 0 0 0 0 31 808 50 2 3 0 999 Coax Short 2 Open 4943 31013 23 17 0 16 29 24 50 2 2 0 999 Coax Open 3 Load F xd 0 50 2 3 0 999 Coax Broadband 4 Delay 0 50 2 3 0 999 Coax Thru thru 5 Load Sliding 0 50 1 3 2 999 999 Coax Sliding 6 Load F xd 0 50 1 3 0 3 001 Coax Lowband 7 8 a Ensure system Zo of network analyzer is set to this value b Open short load delay thru or arbitrary impedance C Load or arbitrary impedance only d For waveguide the lower frequency is the same as F co e Typical values only Disk values may be different A 10 85052B Standard Definitions Nominal Standard Definitions Table A 9 Standard Definitions for the PNA Series Network Analyzer System Zo 250 0 Q Calibration Kit Label 3 5 mm Model 85052B N N N E E Frequency Standard a Offset in GHz 5 a A 3 8 8 5 3 x x x x o m N m 9 V V V S ko kil n gt T I T I I a o a E 3 A a 9 ke c 9 gt g b b b amp 8 8 J5 c 5 3 t o 2 ri ri 3 S x s 518 4 N m x S Sie E pm B E ZI a a d d iL a N a pa Oo A 1 Short 2 0765 108 54 2 1705 0 01 31 785 50 2 366 0 999 Coax Short 2 Open 49 433 310 131 23 1682 0 1597 29243 50 22 0 999 Coax Open 3 Broadband Fxd 0 50 0 0 999 Coax Broadband load load 4 Thru 0 50 0 0 999 Coax Thru 5 Load Sliding 0 50 0 2 999 999 Coax Sliding load 6 Lowband F
3. Calibration constants disk Specifications and performance verification disk Storage case e Protective end caps e User s and Service Guide pk502b A Standard Definitions 85052B A 1 Standard Definitions Standard Class Assignments Standard Class Assignments Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response Tables A 1 through A 3 list the classes of the devices in the kit for various network analyzers This information resides on the calibration definitions disk included in the kit Table A 1 Standard Class Assignments for the 8510 Network Analyzer Disk File Name CK_35MMB3 Calibration Kit Label 3 5 mm B 3 Class A B Cc D E F G Standard Class Label SA 2 Open S1i1B 1 Short S11C 9 10 12 Loads 522A 2 Open S22B 1 Short S22C 9 10 12 Loads Forward transmission 11 Thru Reverse transmission 11 Thru Forward match 11 Thru Reverse match 11 Thru Forward isolation 9 Isol n Std Reverse isolation 9 Isol n Std Frequency response 1 2 11 Response TRL thru Undefined TRL reflect Undefined TRL line Undefined Adapter 13 Adapter TRL Option Cal Zo _ System Zp X LineZg Set ref _X_ Thru Reflect Lowband frequency
4. iv 85052B l General Information 85052B General Information Calibration Kit Overview Calibration Kit Overview The Agilent 85052B 3 5 mm calibration kit is used to calibrate Agilent network analyzers up to 26 5 GHz for measurements of components with 3 5 mm connectors Kit Contents The 85052B calibration kit includes the following items e user s and service guide offset opens and shorts broadband loads and sliding load terminations e 3 5 mm gage set three 3 5 mm adapters e 5 16in 90 N cm 8 in Ib torque wrench open end wrench e two data disks that contain the calibration definitions of the devices in the calibration kit Refer to Table 6 1 on page 6 2 and Table 6 2 on page 6 3 for a complete list of kit contents and their associated part numbers NOTE A backup copy of each data disk and printout should be made immediately upon receipt of the calibration kit Refer to your analyzer user s guide for instructions on duplicating a disk For measurement convenience the kit also contains three 3 5 mm adapters The adapters areprimarily intended for usein measuring non insertable devices but can also be used as a connector saver Broadband Loads The broadband loads are metrology grade terminations that have been optimized for performance up to 26 5 GHz The rugged internal structure provides for highly repeatable connections A distributed resistive element on sapphire provides excellent stability and r
5. td 1 800 828 848 fax 1 800 801 664 Philippines td 632 8426802 td PLDT subscriber only 1 800 16510170 fax 632 8426809 fax PLDT subscriber only 1 800 16510288 Thailand tel outside Bangkok 088 226 008 tel within Bangkok 662 661 3999 fax 66 1661 3714 Hong Kong tel 800 930 871 fax 852 2506 9233 Taiwan td 0800 047 866 fax 886 2 25456723 People s Republic of China td preferred 800 810 0189 td alternate 10800 650 0021 fax 10800 650 0121 India tel 1 600 11 2929 fax 000 800 650 1101 5 4 85052B 6 Replaceable Parts 85052B 6 1 Replaceable Parts Introduction Introduction Table 6 1 lists the replacement part numbers for the 85052B calibration kit Table 6 2 lists the replacement part numbers for items not included in the calibration kit that are either required or recommended for successful operation of the kit To order a listed part note the description the part number and the quantity desired Telephone or send your order to Agilent Technologies See Table 5 1 on page 5 4 for contact information Table 6 1 Replaceable Parts for the 85052B Calibration Kit Description Qty Agilent Part Per Kit Number Calibration Devices 3 5 mm Male sliding load 1 00911 60019 Female sliding load 1 00911 60020 Male broadband load 1
6. 552B and 555C must be defined as line standards If you are performing a TRM calibration e S41l and ST must be defined as thru standards e S41A and S55A must be defined as reflection standards e SB S11 552B and S55C must be defined as match standards If you are performing an LRM calibration e ST and SipT must be defined as linestandards e S41A and S55A must be defined as reflection standards e S4B S141C 552B and S55C must be defined as match standards S11B and S11C must be defined as the same standard S 52B and S25C must be defined as the same standard For additional information on performing TRL TRM and LRM calibrations refer to your PNA series network analyzer embedded help system A 4 85052B Standard Definitions Standard Class Assignments Blank Forms The standard class assignments may be changed to meet your specific requirements Tables A 4 through A 6 are provided to record the modified standard class assignments Table A 4 Blank Form for the 8510 Network Analyzer Disk File Name Calibration Kit Label Class A B C D E F G Standard Class Label 541A S4 B SC 522A 552B S52C Forward transmission Reverse transmission Forward match Reverse match Forward isolation Reverse isolation Frequency response TRL thru TRL reflect TRL line Adapter TRL Option Cal Zo System Zo
7. Metal particles from the connector threads may fall into the connector when it is disconnected One connection made with a dirty or damaged connector can damage both connectors beyond repair In some cases magnification is necessary to see damage to a connector a magnifying device with a magnification of gt 10x is recommended However not all defects that are visible only under magnification will affect the electrical performance of the connector Use the following guidelines when evaluating the integrity of a connector Look for Obvious Defects and Damage First Examine the connectors first for obvious defects and damage badly worn plating on the connector interface deformed threads or bent broken or misaligned center conductors Connector nuts should move smoothly and be free of burrs loose metal particles and rough spots What Causes Connector Wear Connector wear is caused by connecting and disconnecting the devices The more use a connector gets the faster it wears and degrades The wear is greatly accelerated when connectors are not kept clean or are not connected properly Connector wear eventually degrades performance of the device Calibration devices should have a long life if their use is on the order of a few times per week Replace devices with worn connectors Thetest port connectors on the network analyzer test set may have many connections each day and are therefore more subject to wear It is recommended tha
8. j Read the gage indicator dial Read only the black signs not the red signs For maximum accuracy measure the connector a minimum of three times and take an average of the readings Use different orientations of the gage within the connector After each measurement rotate the gage a quarter turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis k Compare the average reading with the observed pin depth limits in Table 2 2 on page 2 4 If the pin depth is outside the limits it must be adjusted before proceeding Refer to Adjusting the Sliding Load Pin Depth on page 3 12 Figure 3 3 Gaging the 3 5 mm Sliding Loads CAUTION CAUTION Do not lock In unlocked position when handle is always slide handle in forward all the way back position before locking Pivot d w Unlocked Position Locked Ed Position Center Centering Sliding Conductor Bead Ring Pullback Handle CAUTION Do not force down beyond natural locking position ph710a 85052B 3 11 Use Maintenance and Care of the Devices Gaging Connectors l Without turning the gage or the sliding load loosen the connection between the gage and the sliding load and remove the sliding load from the gage CAUTION Remove the centering bead immediately after gaging the sliding load pin depth Damage can occur to the sliding load during the removal of a cen
9. 000 feet Relative humidity Always non condensing Operating 0 to 80 26 C maximum dry bulb Storage 0 to 90 a The temperature range over which the calibration standards maintain conformance to their specifications b The allowable network analyzer ambient temperature drift during measurement calibration and during measurements when the network analyzer error correction is turned on Also the range over which the network analyzer maintains its specified performance while correction is turned on Temperature What to Watch Out For Changes in temperature can affect electrical characteristics Therefore the operating temperature is a critical factor in performance During a measurement calibration the temperature of the calibration devices must be stable and within the range shown in Table 2 1 IMPORTANT Avoid unnecessary handling of the devices during calibration because your fingers are a heat source 2 2 85052B Specifications Mechanical Characteristics Mechanical Characteristics Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications They are however important supplemental characteristics related to electrical performance Agilent Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing This ensures that the device connectors do not exhibit any center conductor protrusion or i
10. 1 0259 0 01435 0 0028 0 50 0 0 999 Coax 3 5 2 92 6 open 5 9588 11 195 0 5076 0 00243 0 50 0 0 999 Coax 3 5 SMA 7 Openf 13 4203 1 9452 0 5459 0 01594 0 50 0 0 999 Coax 2 92 5MA 8 Open 8 9843 13 9923 0 3242 0 00112 0 50 0 0 999 Coax Broadband 9 Load Fxd 0 50 0 0 999 Coax Broadband 10 Load Sliding 0 50 0 2 999 999 Coax Sliding 11 Delay 0 50 0 0 999 Coax Thru thru 12 Load F xd 0 50 0 0 3 001 Coax Lowband 13 Delay 94 75p 50 2 51 999 Coax Adapter thru 14 15 16 17 18 19 20 21 a Ensure system Zo of network analyzer is set to this value b Open short load delay thru or arbitrary impedance c Load or arbitrary impedance only d For waveguide the lower frequency is the same as F co e Typical values only Disk values may be different f This standard type open is used to accurately model the adapter listed in the Standard Label column 85052B A 9 Standard Definitions Nominal Standard Definitions Table A 8 Standard Definitions for the 872x Series Network Analyzer System Zo 250 00 Calibration Kit Label 3 5mm N N N us E I Frequency Standard w M 9 Offset in GHz m N m 8 8 8 8 x x x x o m N m 0 s S 9 M N N N Z I I I 5 a Eb z x I 8 a E 3 pe g 8 m g x c 9 E g b b b M 5 c SEE 9g a a kl a E gt e s Se Je Ja Ja 4 3 38 Je e i 3 2 e 3 3 7 i a Na
11. 10 alignment with the mating connector while making the connection Keep the center conductor extended by holding the center conductor pullback mechanism toward the connector end of the sliding load Align the sliding load with the mating connector on the gage and mate the sliding load center conductor with the gage center conductor 85052B Use Maintenance and Care of the Devices Gaging Connectors e Release the center conductor pullback mechanism and move the body of the sliding load toward the gage to mate the outer conductor of the sliding load connector with the outer conductor of the gage connector f Without turning the gage or the sliding load connect the gage to the sliding load being measured by interconnecting the male and female connectors Connect the nut finger tight Do not overtighten g Using an open end wrench to keep the device body from rotating use the torque wrench included in the kit to tighten the connecting nut to the specified torque Refer to Final Connection Using a Torque Wrench on page 3 14 for additional information CAUTION Always move the center conductor pullback mechanism back before locking the handle Do not force the handle past the locked position h Move the center conductor pullback mechanism back away from the connector end of the sliding load and place the pullback handle in its locked position i Gently tap the barrel of the gage with your finger to settle the gage reading
12. 3 affect on electrical performance 2 3 verifying 3 6 MIL STD 45662 calibration 4 3 misuse of product ii modification unauthorized ii National Institute of Standards and Technology NIST 2 6 4 2 nitrogen for cleaning 3 4 nominal standard definitions A 8 numbers replaceable parts 6 2 6 3 serial 1 5 o observed limits pin depth 2 4 offset opens and shorts 1 2 part numbers 6 2 specifications 2 5 open end wrench 1 2 3 17 part number 6 2 opens 1 2 part numbers 6 2 specifications 2 5 ordering parts 6 2 6 3 P part numbers 6 2 of items in kit 6 2 of items not in kit 6 3 parts induded in kit 1 2 6 2 not included in kit 6 2 6 3 ordering 6 2 replaceable 6 2 performance and specifications disk part number 6 3 performance verification failure 5 2 permanently stored calibration definitions 1 3 pin depth 2 3 adjusting sliding load 3 12 affect on electrical performance 2 4 gaging to determine 3 6 observed limits 2 4 3 6 protrusion 2 3 recession 2 3 typical values 2 4 precision adapters 1 3 part numbers 6 2 preventive maintenance 1 7 procedures deaning 3 4 gaging 3 8 zeroing 3 8 protective end caps part numbers 6 2 protrusion pin depth 2 3 R recertification how to order 4 3 interval 4 3 locations 4 3 services included 4 3 recession pin depth 2 3 regulations environmental 2 2 3 5 remedies exclusive ii 85052B Index replaceable
13. Defects and Damage First L L rl ll kk nu knn ka koka eee 3 3 What Causes Connector WOSES ie oi so kk kk a ew Se eae n e ke n BS 3 3 popet ERE Mating Plane SIWO kikonk pk a ont pt dR REE RS eel 3 3 ASPECE Female jac es ok k hades hee ee ee n a n BY A oi a 3 4 Cleaning LORBECEES ww w e din e a oak arhe a ee a tee weet MERGE A E Re qa cb ded 3 4 Beng LANNE S m P ak et sa n kr kt e kas etn Fk a n ak ee n l a ke a a 3 6 Conme tor CE A il aie e cT 3 6 When ta Gage CONNEC E ei ese a aose es RR eer n D kn re 3 7 Gamo FIOSECUAEE 4e xo air RR qae RR ok vo ke DA tv BA eee eh n A ees 3 8 isis 3 59 mim COMO Ss ecc cis oa eee qe Are bei hed eee obese QURE e si na fd d ic 3 8 Gading the 5 AN SING LOANS ek kab dde a n PR ak e pon A e aba dd qs 3 10 Adjusting the Sliding Load Pin DEpth kI eee kk ak kn koke knn konn nana 3 12 oiii PPM 3 14 How to Make a CONN uot courte kk n aenea queer RMR ES duc d ed i Recto n 3 14 Premiers Y ANE EN ON sab oes a a a n tate ab uiid ata ib did i cca 3 14 Final Connection Using a Torque WrencCh ere e nen nn ene nannan nna 3 14 Conneding the sliding Load siue bak vav ti ki a n RERO bk on A NR ki ae es 3 16 Howto Seperate a Conte E Dac uo pea kk a m ka e doe ee go RD a n a OR e d ek diras 3 17 85052B iii Contents Using the Sliding Load wa ki ake EK ae ked ake Aid ke deed eke Y RP P RE 3 18 Ha glnaanmd SESS k dak w ka ae koz kos a pa e fas TE kee VEE CR C PA AAA 3 18 4 Per
14. Line Zo Set ref Thru Reflect Lowband frequency a The forward isolation standard is also used for the isolation part of the response and isolation calibration b Broadband loads are used for frequencies up to 2 GHz 85052B A 5 Standard Definitions Standard Class Assignments Table A 5 Blank Form for the 872x Series of Network Analyzers Calibration Kit Label Class A B C D E F G Standard Class L abel 541A S4 B SjC 522A 552B St Forward transmission Reverse transmissi on Forward match Reverse match Response Response amp isolation TRL thru TRL reflect TRL line or match TRL Option Cal Zo System Zp Line Zo Set ref Thru Reflect Lowband frequency a Broadband loads are used for frequencies up to 2 GHz A 6 85052B Standard Definitions Standard Class Assignments Table A 6 Blank Form for the PNA Series Network Analyzers Calibration Kit Label Class A SA S11B 11C SAT 552A 552B 552C Sir Notes 1 If you are performing a TRL calibration e ST and S4 T must be defined as thru standards e Sy A and 555A must be defined as reflection standards e S4B S11 552B and S55C must be defined as line standards 2 If you are performing a TRM calibration e ST and S4 T must be defined as thru st
15. load termination include the quality of the airline portions within the sliding load combined with the effective stability of the sliding element b The specifications for the opens and shorts are given as allowed deviation from the nominal model as defined in the standard definitions see Nominal Standard Definitions on page A 8 85052B 2 5 Specifications Electrical Specifications Residual Errors after Calibration The 8510 Specifications and Performance Verification software can be used to obtain a printout of the residual errors after a calibration has been performed Refer to the Specifications and Performance Verification section of the 8510C On SiteServiceM anual for information on how to use the software Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory Agilent further certifies that its calibration measurements aretraceable to the United States National Institute of Standards and Technology NIST tothe extent allowed by the institute s calibration facility and tothe calibration facilities of other International Standards Organization members See How Agilent Verifies the Devices in Your Kit on page 4 2 for more information 2 6 85052B 3 Use Maintenance and Care of the Devices 85052B 3 1 Use Maintenance and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection aga
16. page 3 4 earlier in this chapter 3 Zero the connector gage refer to Figure 3 2 on page 3 9 a d While holding the gage by the barrel and without turning the gage or the device connect the gage to the gage master by interconnecting the male and female connectors Connect the nut finger tight Do not overtighten Using an open end wrench to keep the device body from rotating use the torque wrench included in the kit to tighten the connecting nut to the specified torque Refer to Final Connection Using a Torque Wrench on page 3 14 for additional information As you watch the gage pointer gently tap the barrel of the gage to settle the reading The gage pointer should line up exactly with the zero mark on the gage If not adjust the zero set knob until the gage pointer lines up exactly with the zero mark Remove the gage master 4 Gage the sliding load connector refer to Figure 3 3 a Unlock the center conductor pullback mechanism by raising the pullback handle to the unlocked position Carefully move the pullback mechanism toward the connector end of the sliding load The center conductor will extend beyond the end of the connector Continue to hold the pullback mechanism in this position Pull the sliding ring back approximately 0 5 in and install a centering bead in the connector end of the sliding load CAUTION The sliding load center conductor can be damaged if the sliding load is not in d 3
17. parts 6 2 6 3 drawing of 6 4 6 5 report calibration 4 3 requirements environmental 2 2 return kit or device to Agilent 1 4 5 3 return loss specifications 2 5 S separating connections 3 17 serial numbers 1 5 devices 1 5 recording 1 5 service 5 3 service tag 4 3 5 3 set marks sliding load 3 18 shorts 1 2 part numbers 6 2 specifications 2 5 sliding load 1 3 calibration 3 18 connecting 3 16 gaging 3 10 pin depth adjusting 3 12 separating connection 3 17 set marks 3 18 sliding ring 3 18 specification 2 5 using 3 18 specifications 2 2 adapters 2 5 altitude operating 2 2 storage 2 2 and performance disk part number 6 3 broadband loads 2 5 certification of 2 6 deviation from nominal phase 2 5 device 2 5 electrical 2 5 environmental 2 2 frequency 2 5 humidity operating 2 2 storage 2 2 offset opens and shorts 2 5 return loss 2 5 sliding loads 2 5 temperature 2 2 torque wrench 3 14 traceability 4 2 4 3 standard class assignments A 2 blank forms A 5 standard definitions A 8 A 12 blank forms A 12 nominal A 8 standards international 2 6 National Institute of Standards and Technology NIST 2 6 4 2 static discharge 3 2 electricity 3 2 storage 3 18 storage case part number 6 2 strap heel and wrist part number 6 3 supplies cleaning 1 4 part number 6 3 swabs cleaning 3 5 system impedance A 8 T tag service 4 3 5 3 tape calibratio
18. seen on the gage reading due to measurement uncertainty The depth could still be within specifications 2 4 85052B Specifications Electrical Specifications Electrical Specifications The electrical specifications in Table 2 3 apply to the devices in your calibration kit when connected with an Agilent precision interface Table 2 3 Electrical Specifications for 85052B 3 5 mm Devices Device Specification Frequency GHz Broadband loads Return loss 2 46 dB p lt 0 00501 dcto lt 2 male and female Return loss gt 44 dB p lt 0 00631 gt 2to lt 3 Return loss 2 38 dB p lt 0 01259 gt 3to lt 8 Return loss 2 36 dB p lt 0 01585 gt 8to lt 20 Return loss gt 34 dB p lt 0 01995 gt 20 to lt 26 5 Sliding loads Return loss 7 44 dB p x 0 00631 gt 3to lt 26 5 Offset opens 0 65 deviation from nominal dcto lt 3 male and female 1 20 deviation from nominal gt 3to lt 8 2 00 deviation from nominal gt 8to lt 20 2 00 deviation from nominal gt 20 to lt 26 5 Offset shorts 0 50 deviation from nominal dcto 3 male and female 1 00 deviation from nominal gt 3to lt 8 1 75 deviation from nominal gt 8to lt 20 1 75 deviation from nominal gt 20to lt 26 5 Adapters Return loss 2 30 dB p 0 03162 dcto lt 8 Return loss 2 28 dB p lt 0 03981 gt 8to lt 18 Return loss 2 26 dB p lt 0 05012 gt 18 to lt 26 5 a The specifications for the sliding
19. when mating nonprecision to precision devices The female 3 5 mm connectors in this calibration kit are metrology grade precision slotless connectors PSC Precision slotless connectors are used to improve accuracy With PSCs on test ports and standards the accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss This represents an accuracy improvement of about 10 times Conventional female center conductors are slotted and when mated are flared by the male pin Because physical dimensions determine connector impedance this change in physical dimension affects electrical performance making it very difficult to perform precision measurements with conventional slotted connectors The precision slotless connector was developed to eliminate this problem The PSC has a center conductor with a solid cylindrical shell the outside diameter of which does not change when mated Instead this center conductor has an internal contact that flexes to accept the male pin Cleaning Connectors Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections 1 Use Compressed Air or Nitrogen WARNING Always use protective eyewear when using compressed air or nitrogen Use compressed air or nitrogen to loosen particles on the connector mating plane surfaces Clean air cannot damage a connector or leave par
20. 00902 60003 Female broadband load 1 00902 60004 Male offset short 1 85052 60006 Female offset short 1 85052 60007 Male offset open 1 85052 60008 Female offset open 1 85052 60009 Adapters 3 5 mm Female to female 1 85052 60012 Male to female 1 85052 60013 Male to male 1 85052 60014 Protective End Caps for Connectors For 3 5 mm female connectors 10 1401 0202 For 3 5 mm male connectors 10 1401 0208 Wrench 5 16 in 90 N cm 8 in Ib torque wrench 1 8710 1765 7 mm open end wrench 1 8710 1761 Calibration Kit Storage Case Box without foam pads 1 5180 7900 Foam pad for lid 1 5181 5543 6 2 85052B Table6 1 Replaceable Parts for the 85052B Calibration Kit Replaceable Parts Introduction Description Qty Agilent Part Per Kit Number Foam pad for lower case 1 85052 80031 Disk holder 1 5180 8491 Connector Gages 3 5 mm Gage Set for female connectors 1 11752 60105 Gage Set for male connectors 1 11752 60106 Centering bead for gaging 3 5 mm sliding load 2 85052 20057 Miscellaneous Items Calibration definitions disk PNA Series 1 85052 10015 Calibration definitions disk 8510 872x Series 1 85052 10010 Specifications and performance verification disk 1 08510 10033 User s and service guide 1 85052 90077 Connector care quick reference card 1 08510 90360 a Broadband load has replaced lowband load b See the 8510C On Site Service Manual for instructions on using
21. 0381 mm 0 00015 in This procedure assumes that you were directed here from Gaging the 3 5 mm Sliding Loads on page 3 10 If not perform the steps in that procedure before performing this procedure 1 The gage should be attached to the sliding load Refer to Gaging the 3 5 mm Sliding Loads on page 3 10 if necessary 2 The face of the gage and the label on the sliding load should be facing up 3 The center conductor pullback handle should be in the locked position 4 With a small screwdriver gently turn the center conductor pin depth adjustment screw until the gage pointer reads 0 00381 mm 0 00015 in Refer to Figure 3 4 for the location of the adjustment screw 5 Wait approximately five minutes to allow the temperature to stabilize Do not touch either the gage or the sliding load during this ti me 3 12 85052B Use Maintenance and Care of the Devices Gaging Connectors 6 Note the gage reading If it is no longer within the allowable range perform step 4 again 7 Move the center conductor pullback handle to the unlocked position and then back to the locked position The gage reading should return to the value previously set If not repeat steps 4 through 7 8 Return to Gaging the 3 5 mm Sliding Loads on page 3 10 Figure 3 4 Adjusting the Sliding Load Pin Depth CAUTION Do not lock when handle is in forward 4 position Pivot Point 85052B Center Conductor Pullbac
22. 2 0 GHz a Theforward isolation standard is also used for the isolation part of the response and isolation calibration b Broadband loads are used for frequencies up to 2 GHz A 2 85052B Standard Definitions Standard Class Assignments Table A 2 Standard Class Assignments for the 872x Series Network Analyzer Calibration Kit Label 3 5 mm Class A B Cc D E F G Standard ClassL abel SA 2 Open S4B 1 Short S11C 3 5 6 Loads 522A 2 Open 552B 1 Short 552C 3 5 6 Loads Forward transmission 4 Thru Reverse transmission 4 Thru Forward match 4 Thru Reverse match 4 Thru Response 1 2 4 Response Response amp isolation 1 2 4 Response TRL thru 4 Thru TRL reflect 2 Open TRL line or match 3 5 6 Loads TRL Option Cal Zo SystemZg X LineZg Set ref X Thru Reflect Lowband frequency 2 0 GHZ a Broadband loads are used for frequencies up to 2 GHz 85052B A 3 Standard Definitions Standard Class Assignments Table A 3 Standard Class Assignments for the PNA Series Network Analyzer Calibration Kit Label 3 5 mm Model 85052B Class A S4ijA 2 S4 B 1 S11C 3 5 SAT 4 S52A 2 SB 1 552C 3 5 Si2T 4 Notes 1 4 5 Ifyou are performing a TRL calibration e S41l and ST must be defined as thru standards e Sy A and SpA must be defined as reflection standards e S4 B S11
23. Torque the connecting nut Gently tap the gage barrel to settle the reading Using the zero set knob adjust the gage pointer to line up exactly with the zero mark Remove the gage master Gage Barrel TULLUS inl Gage Master Connecting Nut Gage the Device Connector Connect the gage to the device being measured Torque the connecting nut Gently tap the gage barrel to settle the reading Read recession or protrusion from the gage Remove the device Repeat two additional times and average the three readings Gage Barrel ent Device Connecting Nut Being Measured ph71a 85052B 3 9 Use Maintenance and Care of the Devices Gaging Connectors Gaging the 3 5 mm Sliding Loads Gage the sliding load before each use If the sliding load pin depth is out of the observed pin depth limits listed in Table 2 2 on page 2 4 refer to Adjusting the Sliding Load Pin Depth on page 3 12 NOTE Always hold a connector gage by the gage barrel below the dial indicator This gives the best stability and improves measurement accuracy Cradling the gagein your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing 1 Select the proper gage for your connector Refer to Table 6 1 on page 6 2 for gage part numbers 2 Inspect and clean the gage gage master and device to be gaged Refer to Visual Inspection on page 3 3 and Cleaning Connectors on
24. User s and Service Guide Agilent Technologies 85052B 3 5 mm Calibration Kit This manual applies to 85052B calibration kits with serial number prefix 3106A 2 Agilent Technologies Agilent Part Number 85052 90077 Printed in USA Print Date January 2002 Supersedes March 2000 Copyright 1994 1995 2000 2002 Agilent Technologies Inc All rights reserved Warranty This product is warranted against defects in material and workmanship for a period of one year from date of shipment During the warranty period Agilent Technologies will at its option either repair or replace products which prove to be defective For warranty service or repair this product must be returned to a service facility designated by Agilent Buyer shall prepay shipping charges to Agilent and Agilent shall pay shipping charges to return the product to Buyer However Buyer shall pay all shipping charges duties and taxes for products returned to Agilent from another country Agilent Technologies warrants that its software and firmware designated by Agilent for use with an instrument will execute its programming instructions when properly installed on that instrument Agilent Technologies does not warrant that the operation of the instrument or software or firmware will be uninterrupted or error free Limitation of Warranty The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer Buyer supplied
25. andards e S44 and S32A must be defined as reflection standards e SiB S11C 552B and S55C must be defined as match standards 3 If you are performing an LRM calibration e S T and S45T must be defined as line standards e S4 and S32A must be defined as reflection standards e SiB 511C 552B and S55C must be defined as match standards 4 S11B and S 4C must be defined as the same standard 5 S5 B and 553C must be defined as the same standard For additional information on performing TRL TRM and LRM calibrations refer to your PNA series network analyzer embedded help system 85052B A 7 Standard Definitions Nominal Standard Definitions Nominal Standard Definitions Standard definitions provide the constants needed to mathematically model the electrical characteristics delay attenuation and impedance of each calibration standard The nominal values of these constants are theoretically derived from the physical dimensions and material of each calibration standard or from actual measured response T hese values are used to determine the measurement uncertainties of the network analyzer The standard definitions in Tables A 7 through A 9 list typical calibration kit parameters used by the following network analyzers to specify the mathematical model of each device This information must be loaded into the network analyzer to perform valid calibrations Refer to your network analyzer user s guide for instructions on load
26. ce body from rotating use the torque wrench included in the kit to tighten the connecting nut to the specified torque Refer to Final Connection Using a Torque Wrench on page 3 14 for additional information CAUTION Always move the center conductor pullback mechanism back before locking the handle Do not force the handle past the locked position 7 Move the center conductor pullback mechanism back away from the connector end of the sliding load and place the pullback handle in its locked position 3 16 85052B Use Maintenance and Care ofthe Devices Connections Figure 3 7 Connecting the Sliding Load CAUTION CAUTION Do not lock In unlocked position when handle is always slide handle in forward all the way back gt position before locking Unlocked Le Position Locked Position CAUTION Do not force down beyond natural locking position ph73a How to Separate a Connection To avoid lateral bending force on the connector mating plane surfaces always support the devices and connections CAUTION Do not turn the device body Only turn the connector nut Damage to the center conductor can occur if the device body is twisted If disconnecting a sliding load leave the center conductor pullback handle in the locked position 1 Use an open end wrench to prevent the device body from turning 2 Use another open end wrench to loosen the connecting nut 3 Complete the separati
27. d as on a test port However it is recommended that you use an open end wrench to keep the body of the device from turning 1 Position both wrenches within 90 degrees of each other before applying force See Figure 3 5 Wrenches opposing each other greater than 90 degrees apart will causea lifting action which can misalign and stress the connections of the devices involved This is especially true when several devices are connected together Figure 3 5 Wrench Positions Hold this wrench steady Device Device Correct Method Incorrect Method Reduces Lift on Connection Too Much Lift on Connection ph711a 2 Hold the torque wrench lightly at the end of the handle only beyond the groove See Figure 3 6 Figure3 6 Usingthe Torque Wrench Torquing Direction Stop when handle begins to yield ph712a 3 Apply downward force perpendicular to the wrench handle This applies torque to the connection through the wrench Do not hold the wrench sotightly that you push the handle straight down along its length rather than pivoting it otherwise you apply an unknown amount of torque 85052B 3 15 Use Maintenance and Care ofthe Devices Connections 4 Tighten the connection just to the torque wrench break point The wrench handle gives way at its internal pivot point See Figure 3 6 on page 3 15 Do not tighten the connection further CAUTION You don t have to fully break the handle of the torque wrench to reach the sp
28. ecified torque doing so can cause the handle to kick back and loosen the connection Any give at all in the handle is sufficient torque Connecting the Sliding Load 1 Unlock the center conductor pullback mechanism by raising the pullback handle to the unlocked position Refer to Figure 3 7 2 Carefully move the pullback mechanism toward the connector end of the sliding load The center conductor will extend beyond the end of the conductor Continue to hold the pullback mechanism in this position CAUTION The sliding load center conductor can be damaged if the sliding load is not in alignment with the mating connector while making the connection 3 Keep the center conductor extended by holding the center conductor pullback mechanism toward the connector end of the sliding load Align the sliding load with the mating connector on the cable or test port to which it is being connected and mate the sliding load center conductor with the center conductor of the cable or test port 4 Release the center conductor pullback mechanism and move the body of the sliding load toward the cable or test port to mate the outer conductor of the sliding load connector to the outer conductor of the cable or test port connector 5 Without turning the sliding load connect the sliding load to the cable or test port by interconnecting the male and female connectors Connect the nut finger tight Do not overtighten 6 Using an open end wrench to keep the devi
29. ector damage during storage Keep connectors dean Donot touch mating plane surfaces Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove e Donot set connectors contact end down on a hard surface The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface 3 18 85052B 4 Performance Verification 85052B 4 1 Performance Verification Introduction Introduction Th e performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available How Agilent Verifies the Devices in Your Kit Agilent verifies the specifications of these devices as follows 1 The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to the National Institute of Standards and Technology NIST Theairline and short characteristics are developed from mechanical measurements The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements Each calibration device i
30. eturn loss Offset Opens and Shorts The offset opens and shorts are built from parts that are machined tothe current state of the art in precision machining The offset short s inner conductors have a one piece construction common with the shorting plane The construction provides for extremely repeatable connections The offset opens have inner conductors that are supported by a strong low dielectric constant plastic to minimize compensation values 1 2 85052B General Information Calibration Kit Overview Both the opens and shorts are constructed so that the pin depth can be controlled very tightly thereby minimizing phase errors The lengths of the offsets in the opens and shorts are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies Adapters Like the other devices in the kit the adapters are built to very tight tolerances to provide good broadband performance and to ensure stable repeatable connections The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized The adapters are designed so that their nominal electrical lengths are the same which allows them to be used in calibration procedures for non insertable devices Sliding Loads The sliding loads in this kit are designed to provide excellent performance from 3 GHz to 26 5 GHz Theinner and outer conductors of the airline porti
31. formance Verification DCH CRO uoc pan kila ap de Re oo e e ak OU OO SO SE we OMB k E A LSE 4 2 How Agilent Verifies the Devices in Your Kit e sera e e kn nn an knn nan 4 2 Recer CaO 4 inks de e e ak e oo koke de EROR PRX ik a a a d e ad A a a ok e TRE KO AN o OR d eS 4 3 How Oten Eo FEAT ANN iced oak ok also Re Ea PELE EDEN QV PII qeu pe ed 4 3 Where to Send a Kit for Recertification iiia axxxaxkas Ce XXXTXXXd n dR Oe XC RC E A AN 4 3 5 Troubleshooting Jroubleshiboting PPROEESB 2a edes row ee a aea Etc SOR Gd tete cM bu efc bl dd 5 2 Returning s E or DOUG iG AGIGNE Lua sexe x en ok seated s e nuces Rug imi ARR UG 5 3 Where to Lock for More Information os k ve ven oi cmm Rm ge chm deme e 5 3 Contectmig AO ene ceu esee eed RS ordo UBRO RC de ond e X dolos bee Ge Lies 5 4 6 Replaceable Parts VETER NT sce fy a ke b ce berto etd ee oh ee e Repente oram oli er db bn 6 2 A Standard Definitions stendard Class Asstanmellbs videa sopa eri e Y ackdor REO PP I CERETG OR Rea Efu beeen A 2 Blank Forms iaa bouk a data AIRE EAE ERAN OE ERR a dok ESNd C PERGERET P fads A 5 Nominal Standard DEFANN BAS Cs nde Pica ex dw tk a ie POSER SSE wb ER C x dae HOE DO A 8 Setting the System Impedance aiii de aaa de ae Rb RENRT A RRTQEPKERTAPRPITQPE CARERE A 8 Afer Sram CHANGES kav d aad e g ai a ap Re PRA EP OPER S ko ee OPEP pde e COSA ERSTE dede A 8 Blank FOTOS Sacco ke Edo HERES a ea oak A ka a CERCA REESE a EGER ICA e OR EORR A 12
32. ication is 12 months or sooner The actual need for recertification depends on the use of the kit After reviewing the results of the initial recertification you may establish a different recertification interval that reflects the usage and wear of the kit NOTE The recertification interval should begin on the date the kit is first used after the recertification date Where to Send a Kit for Recertification Contact Agilent Technologies for information on where to send your kit for recertification Contact information is listed on page 5 4 Refer to Returning a Kit or Device to Agilent on page 5 3 for details on sending your kit 85052B 4 3 5 Troubleshooting 5 1 Troubleshooting Troubleshooting Process Troubleshooting Process Ifyou suspect a bad calibration or if your network analyzer does not pass performance verification follow the steps in Figure 5 1 Figure 5 1 Troubleshooting Flowchart EVERYTHING pk54c 5 2 START BAD CAL SUSPECTED BAD RECAL BRATE MEASUREMENT SYSTEM DEVICES CLEAN DEVICES CLEANED AND AND GAGE PROPERLY GAGED PROPER TORQUE 3 5 mm DEVICES TORQUE TO Bin lb 90 N cm CORRECT LOAD CORRECT CAL CONSTANTS LOADED CAL CONSTANTS MATING SURFACES SMOOTH amp REPLACE DEVICE S UNMARRED RECERTIFY KIT 85052B Troubleshooting Returning a Kit or Device to Agilent Returning a Kit or Device to Agilent If your kit or device require
33. ing calibration definitions NOTE The values in the standard definitions table are valid only over the specified operating temperature range Setting the System Impedance This kit contains only 50 ohm devices Ensure the system impedance Zp is set to 50 ohms Refer to your network analyzer s user s guide for instructions on setting system impedance Version Changes Class assignments and standard definitions may change as more accurate model and calibration methods are developed The disk shipped with the kit for use with the 8510 will contain the most recent version The default version that comes with the 872x network analyzer firmware may be outdated A 8 85052B Table A 7 Standard Definitions for the 8510 Network Analyzer Standard Definitions Nominal Standard Definitions System Zo 50 0Q Calibration Kit Label 3 5 mm B 3 Disk File Name CK_35MMB3 File Number FILE 1 am S ka Standard n N g Offset in GHz 8 8 3 18 x x x x o m N m V V V V S N N N a g T T 3 a gt I I I I A a g m v A amp m ya 6 o r B c 6 3 fs 8 8 hR g le E 3 35 4 a m Xx y Sg JE gd S 5 zr a i d iL a N a V a l Short 2 0765 108 54 2 1705 0 01 31 785p 50 2 36 0 999 Coax Short 2 Open 49 433 310 131 23 1682 0 15966 29 243p 50 22 0 999 Coax Open 3 4 5 Openf 6 9558
34. ing torque wrench 3 14 connector deaning 3 4 damage 3 3 defects 3 3 female 3 4 gage accuracy 3 6 handling 3 6 3 8 3 10 use of 3 6 zeroing 3 6 3 8 3 10 gaging 3 6 3 8 to determine pin depth 3 6 when to do 3 7 mating plane surfaces deaning 3 5 Sex darifying 1 6 slotless 3 4 6 3 threads deaning 3 5 inspecting 3 3 visual inspection 3 3 wear 3 3 affect on electrical performance 3 3 constants calibration 1 3 entering 1 3 permanently stored 1 3 part numbers 6 2 contacting Agilent Technologies 5 3 5 4 contents calibration kit 1 2 drawing of 6 4 6 5 incomplete what to do 1 4 D damage caused by electrostatic discharge 3 2 device 3 3 inspecting for 3 3 to connectors 3 3 data recertification 4 3 defective connectors 3 3 definitions standard A 8 deviation from nominal phase 2 5 device cleaning 3 4 connecting 3 14 damage 3 3 disconnecting 3 17 handling 3 18 maintenance 1 7 performance verifying 4 2 specifications 2 5 certification of 2 6 electrical 2 5 traceability 4 2 4 3 storage 3 18 temperature 2 2 visual inspection 3 3 disconnections 3 17 disk calibration definitions part number 6 3 specifications and performance part number 6 3 documentation part number 6 3 electrical characteristics effects of temperature 2 2 electrical specifications 2 5 electrostatic discharge See ESD end cap part numbers 6 2 environmental regulations 3 5 req
35. inst electrostatic discharge ESD is essential while connecting inspecting or cleaning connectors attached to a static sensitive circuit Such as those found in test sets Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged Static discharges too small to be felt can cause permanent damage Devices such as calibration components and devices under test DUT can also carry an electrostatic charge To prevent damage to the test set components and devices Always wear a grounded wrist strap having a 1 MQ resistor in series with it when handling components and devices or when making connections to the test set Always use a grounded antistatic mat in front of your test equipment e Always wear a heel strap when working in an area with a conductive floor If you are uncertain about the conductivity of your floor wear a heel strap Figure 3 1 shows a typical ESD protection setup using a grounded mat and wrist strap Refer to Chapter 6 Replaceable Parts for information on ordering supplies for ESD protection Figure 3 1 ESD Protection Setup NOTE A 3 Wire line power cord with proper connection to ground must be used for ESD protection Anti Static Mat ku310b 3 2 85052B Use Maintenance and Care of the Devices Visual Inspection Visual Inspection Visual inspection and if necessary cleaning should be done every time a connection is made
36. isted Do not twist or screw the connectors together As the center conductors mate there is usually a slight resistance 7 The preliminary connection is tight enough when the mating plane surfaces make uniform light contact Do not overtighten this connection A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient Very light finger pressure is enough to accomplish this 8 Make sure the connectors are properly supported Relieve any side pressure on the connecti on from long or heavy devices or cables Final Connection Using a Torque Wrench Use a torque wrench to make a final connection Table 3 1 provides information about the torque wrench recommended for use with the calibration kit A torque wrench is included in the calibration kit Refer to Table 6 1 on page 6 2 for replacement part number and ordering information Table3 1 Torque Wrench Information Connector Type Torque Setting Torque Tolerance 3 5 mm 90 N cm 8 in lb X 9 0 N cm 0 8 in lb 3 14 85052B Use Maintenance and Care ofthe Devices Connections Using a torque wrench guarantees that the connection is not too tight preventing possible connector damage It also guarantees that all connections are equally tight each time Prevent the rotation of anything other than the connector nut that you are tightening It may be possible to do this by hand if one of the connectors is fixe
37. k Handle CAUTION In unlocked position always slide handle all the way back gt before locking Adjustment Screw Unlocked Position End View Locked Position Center Conductor Pullback Handle CAUTION Do not force down beyond natural locking position ph79a 3 13 Use Maintenance and Care ofthe Devices Connections Connections Good connections require a skilled operator The most common cause of measurement error is bad connections The following procedures illustrate how to make good connections How to Make a Connection Preliminary Connection 1 Ground yourself and all devices Wear a grounded wrist strap and work on a grounded conductive table mat Refer to Electrostatic Discharge on page 3 2 for ESD precautions 2 Visually inspect the connectors Refer to Visual Inspection on page 3 3 3 If necessary clean the connectors Refer to Cleaning Connectors on page 3 4 4 Usea connector gage to verify that all center conductors are within the observed pin depth values in Table 2 2 on page 2 4 Refer to Gaging Connectors on page 3 6 5 Carefully align the connectors The male connector center pin must slip concentrically into the contact finger of the female connector 6 Push the connectors straight together and tighten the connector nut finger tight CAUTION Do not turn the device body Only turn the connector nut Damage to the center conductor can occur if the device body is tw
38. mproper pin depth when the kit leaves the factory Gaging Connectors on page 3 6 explains how to use gages to determine if the kit devices have maintained their mechanical integrity Refer to Table 2 2 on page 2 4 for typical and observed pin depth limits Pin Depth Pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane See Figure 2 1 The pin depth of a connector can bein one of two states either protruding or recessed Protrusion is the condition in which the center conductor extends beyond the outer conductor mating plane This condition will indicate a positive value on the connector gage Recession is the condition in which the center conductor is set back from the outer conductor mating plane This condition will indicate a negative value on the connector gage Figure 2 1 Connector Pin Depth Male Female Outer Conductor Center Conductor Male Pin Depth Outer Conductor Mating Plane Outer Conductor Mating Plane ph51b 85052B 2 3 Specifications Mechanical Characteristics The pin depth value of each calibration device in the kit is not specified but is an important mechanical parameter The electrical performance of the device depends to some extent on its pin depth The electrical specifications for each device in the kit take into account the effect of pin depth on the device s performance Table 2 2 lists the typical pin dep
39. n constants part number 6 3 temperature affect on electrical performance 2 2 calibration 2 2 cautions about 2 2 changes in 2 2 device 2 2 error corrected 2 2 measurement 2 2 specifications 2 2 operating 2 2 storage 2 2 verification and measurement 2 2 test data 4 3 threads connector deaning 3 5 inspecting 3 3 torque wrench 1 4 spedifications 3 14 traceability of device specifications 4 2 4 3 troubleshooting flowchart 5 2 U user s and service guide part number 6 3 V verification performance 4 3 temperature 2 2 visual inspection 3 3 w warranty ii assistance ii limitation ii wear connector 3 3 affect on electrical performance 3 3 wrench open end 1 4 3 15 3 17 proper positioning of 3 15 torque 1 4 3 14 3 15 precautions for use of 3 15 proper use of 3 15 wrist strap part number 6 3 Z zeroing connector gage 3 8 3 10 85052B Index 3
40. on are precision machined to state of the art tolerances Although the sliding load has exceptional return loss its superior load stability qualifies it as a high performance device The sliding load was designed with the ability to extend the inner conductor for connection purposes and then pull it back to a preset pin depth This feature is critical since it minimizes the possibility of damage during the connection while maintaining a minimum pin depth to optimize performance Calibration Definitions The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration Refer to your network analyzer user s guide for instructions on selecting the calibration kit and performing a calibration The calibration definitions can be e resident within the analyzer e loaded from the provided disk e entered from the front panel This Calibration Kit Supports the Following Network Analyzers The calibration definitions for the kit are permanently installed in the internal memory or hard disk of the following network analyzers 8719ET ES 8753 ET ES 8720ET ES PNA Series 8722bET ES If this calibration kit is used with other network analyzers the calibration definitions must be manually entered into the network analyzer Refer to your network analyzer user s guide for instructions 85052B 1 3 General Information Incoming In
41. on by hand turning only the connecting nut 4 Pull the connectors straight apart without twisting rocking or bending either of the connectors 85052B 3 17 Use Maintenance and Care of the Devices Using the Sliding Load Using the Sliding Load When performing a sliding load calibration it is recommended that the sliding ring be set at the marked positions rings along the sliding load body Using the set marks ensures that a broad distribution of phase angles is selected thereby optimizing the calibration The set marks function as detents so that the internal center of the sliding ring can mate with them Because of this the set mark being used cannot be seen but is felt as the sliding ring is moved from mark to mark during a calibration Moving the sliding ring with only the index fingers of both hands will increase your ability to detect the sliding ring detent at each position To perform a sliding load calibration refer to your network analyzers user s documentation for instructions Figure 3 8 Sliding Load Set Marks Sliding Ring A LII First Slide Set al Second Slide Set b C e f Center Conductor Pulldown Handle ph74a Handling and Storage e Install the protective end caps and store the calibration devices in the foam li ned storage case when not in use Never store connectors loose in a box or in a desk or bench drawer This is the most common cause of conn
42. onnector gage 3 6 adapters 1 3 part numbers 6 2 specifications 2 5 Agilent Technologies contacting 5 3 5 4 liabilities ii warranty ii agreements customer assistance ii maintenance ii alcohol isopropyl as cleaning solvent 3 5 part number 6 3 precautions for use of 3 5 altitude specifications 2 2 assistance contacting agilent 5 4 customer ii who to contact ii blank forms standard class assignments A 5 standard definitions A 12 box part number 6 2 broadband loads 1 2 part numbers 6 2 specifications 2 5 Cc cal kit contents 1 2 history 1 6 misuse of ii verifying 4 2 warranty ii calibration bad 5 2 certificate of 4 3 constants 1 3 permanently stored 1 3 report 4 3 temperature 2 2 calibration constants disk part number 6 3 entering 1 3 permanently stored 1 3 calibration definitions 1 3 calibration kit contents 1 2 drawing of 6 4 6 5 history 1 6 misuse of ii overview 1 2 performance how Agilent verifies 4 2 warranty ii case part number 6 2 certificate of calibration 4 3 certification of device specifications 2 6 characteristics mechanical 2 3 dass assignments blank forms A 5 standard A 2 deaning connectors 3 4 deaning supplies 1 4 part number 6 3 compressed air for cleaning 3 4 conductive mat part number 6 3 connections 3 2 3 14 ESD protection 3 2 final 3 14 preliminary 3 14 separating 3 17 sliding load 3 17 sliding load 3 16 us
43. ot the calibration device connector For example the label SHORT F on the analyzer s display refers to the short that is to be connected to the female test port This will be a male short from the calibration kit A connector gage is referred to in terms of the connector that it measures For instance a male connector gage has a female connector on the gage so that it can measure male devices 1 6 85052B General Information Preventive Maintenance Preventive Maintenance The best techniques for maintaining the integrity of the devices in the kit include routine visual inspection deaning proper gaging proper connection techniques All of these are described in page 3 1 Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it Improper connections resulting from pin depth values being out of the observed limits see Table 2 2 on page 2 4 or from bad connection techniques can also damage these devices 85052B 1 7 2 Specifications 85052B 2 1 Specifications Environmental Requirements Environmental Requirements Table 2 1 Environmental Requirements Parameter Limits Temperature Operating 20 C to 26 C Storage 40 C to 75 C Error corrected range 1 C of measurement calibration temperature Altitude Operating 4 500 meters 15 000 feet Storage 15 000 meters 50
44. ropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen Refer to Table 6 2 on page 6 3 for part numbers for isopropyl alcohol and cleaning swabs a Apply a small amount of isopropyl alcohol to a lint free cleaning swab b Clean the connector threads c Let the alcohol evaporate then blow the threads dry with a gentle stream of clean low pressure compressed air or nitrogen Always completely dry a connector before you reassemble or use it 3 Clean the Mating Plane Surfaces a Apply a small amount of isopropyl alcohol to a lint free cleaning swab b Clean the center and outer conductor mating plane surfaces Refer to Figure 2 1 on page 2 3 When cleaning a female connector avoid snagging the swab on the center conductor contact fingers by using short strokes c Let the alcohol evaporate then blow the connector dry with a gentle stream of clean low pressure compressed air or nitrogen Always completely dry a connector before you reassemble or use it 4 Inspect Inspect the connector to make sure that no particles or residue remain Refer to Visual Inspection on page 3 3 85052B 3 5 Use Maintenance and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only They are effective in detecting excessive center conductor protr
45. s electrically tested on this system For the initial before sale testing of the calibration devices Agilent includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification For recertifications after sale no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status The measurement uncertainty for each device is however recorded in the calibration report that accompanies recertified kits These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute s calibration facility The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies 4 2 85052B Performance Verification Recertification Recertification The following will be provided with a recertified kit e anew calibration sticker affixed to the case e a certificate of calibration e a calibration report for each device in the kit listing measured values specifications and uncertainties NOTE A list of NIST traceable numbers may be purchased upon request to be included in the calibration report Agilent Technologies offers a Standard calibration for the recertification of the kit For more information contact Agilent Technologies Refer to Contacting Agilent on page 5 4 for a list of offices How Often to Recertify The suggested initial interval for recertif
46. s service contact Agilent Technologies for information on where to send it See Table 5 1 for contact information Include a service tag located near the end of this manual on which you provide the following information your company name and address a technical contact person within your company and the person s complete tel ephone number the model number and serial number of the kit the part number and serial number of each device thetype of service required a detailed description of the problem and how the device was being used when the problem occurred such as calibration or measurement Where to Look for More Information This manual contains limited information about network analyzer system operation For complete information refer to the instrument documentation If you need additional information contact Agilent Technologies 85052B 5 3 Troubleshooting Contacting Agilent Contacting Agilent Table 5 1 Contacting Agilent Online assistance www agilent com find assist United States td 1 800 452 4844 Latin America td 305 269 7500 fax 305 269 7599 Canada td 1 877 894 4414 fax 905 282 6495 Europe tel 31 20 547 2323 fax 31 20 547 2390 New Zealand td O 800 738 378 fax 464 4 495 8950 J apan td 81 426 56 7832 fax 481 426 56 7840 Australia tel 1 800 629 485 fax 461 3 9210 5947 Singapore tel 1 800 375 8100 fax 65 836 0252 Malaysia
47. software or interfacing unauthorized modification or misuse operation outside of the environmental specifications for the product or improper site preparation or maintenance NO OTHER WARRANTY IS EXPRESSED OR IMPLIED AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE Exclusive Remedies THE REMEDIES PROVIDED HEREIN ARE BUYER S SOLE AND EXCLUSIVE REMEDIES AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES WHETHER BASED ON CONTRACT TORT OR ANY OTHER LEGAL THEORY Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent products For any assistance contact Agilent Technologies Refer to Table 5 1 on page 5 4 Contents 1 General Information C libram KIE Over viens 2 4 kaa eee kane ga ve db eon ee FEB ede METRES Pd 1 2 KI C OMENS a ka ap ak pla a eee po eR eee RAW Tee RA a AA rs OR P ee 1 2 Ef GANY Liss ss a a eee d bete IEE V aede de ALAN ddcd dedo arenae aos 1 2 Oise OPENS and SIMS wd aside bares ae e YA sok GAYA kk Da A be E E RN 1 2 et Cn ee kw ee eee dak e gani be eq p LAS LAW PEEXERMQE PCI EP E 1 3 alamo LoSds ied aaa aa ke e ae da ko e fa EIE E RA Re Es kdk RE Die E Ra AAA dee sh doe tas 1 3 Calibr acram DD OPES i ve A dede d aca eR e MOK Ce an POE YAN ee E SIE e eb py 1 3 This Calibration Kit Supports the Following Network Anal
48. spection Equipment Required but Not Supplied Connector cleaning supplies and various electrostatic discharge ESD protection devices are not supplied with the calibration kit but are required to ensure successful operation of the kit Refer to Table 6 2 on page 6 3 for ordering information Incoming Inspection Refer to Kit Contents on page 1 2 to verify a complete shipment Use Table 1 1 to record the serial numbers of all serialized devices in your kit Check for damage The foam lined storage case provides protection during shipping If the case or any device appears damaged or if the shipment is incomplete refer to Contacting Agilent on page 5 4 Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company See Returning a Kit or Device to Agilent on page 5 3 1 4 85052B General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number the devices in the kit areindividually serialized serial numbers are labeled onto the body of each device Record these serial numbers in Table 1 1 Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits The adapters included in the kit are for measurement convenience only and are not serialized Table 1 1 Serial Number Record for the 85052B Device Serial N
49. t an adapter be used as a test port saver to minimize the wear on the test set s test port connectors Inspect the Mating Plane Surfaces Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection See Figure 2 1 on page 2 3 Look especially for deep scratches or dents and for dirt and metal particles on the connector mating plane surfaces Also look for signs of damage due to excessive or uneven wear or misalignment Light burnishing of the mating plane surfaces is normal and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface Other small defects and cosmetic imperfections are also normal None of these affect electrical or mechanical performance If a connector shows deep scratches or dents particles clinging to the mating plane surfaces or uneven wear clean and inspect it again Devices with damaged connectors should be discarded Determine the cause of damage before connecting a new undamaged connector in the same configuration 85052B 3 3 Use Maintenance and Care of the Devices Cleaning Connectors Inspect Female Connectors Inspect the contact fingers in the female center conductor carefully These can be bent or broken and damage to them is not always easy to see A connector with damaged contact fingers will not make good electrical contact and must be replaced NOTE This is particularly important
50. tering bead that has slipped too far into the sliding load The sliding load will not perform to its specifications if the centering bead is not removed before an electrical calibration is performed m Carefully remove the centering bead from the sliding load If the centering bead does not come out of the sliding load easily i Unlock the center conductor pullback handle and move the center conductor pullback mechanism toward the connector end of the sliding load to extend the center conductor ii While holding the center conductor pullback mechanism toward the connector end of the sliding load remove the centering bead If the centering bead still will not come out i Hold the sliding load with the connector end pointed downward ii Movethesliding ring up then quickly down Thetrapped air behind the centering bead should eject it Return the center conductor pullback mechanism to the rear of the sliding load and return the pullback handle to its locked position Adjusting the Sliding Load Pin Depth The sliding loads in this kit have a setback mechanism that allows the pin depth to be set to any desired value The pin depth of the sliding load is preset at the factory The pin depth should not haveto bereset each timethesliding load is used but it should be checked before each use If the pin depth is outside the observed limits listed in Table 2 2 on page 2 4 use the following procedure to reset it to the nominal value of 0 0
51. the disk Table 6 2 Items Not Included in the Calibration Kit Description Qty Agilent Part Number ESD Protection Devices Grounding wrist strap 1 9300 1367 5 ft grounding cord for wrist strap 1 9300 0980 2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300 0797 ESD heel strap 1 9300 1308 Connector Cleaning Supplies Isopropyl alcohol 30 ml 8500 5344 Foam tipped cleaning swabs 100 9301 1243 Miscellaneous Items 3 5 mm slotless connector contact repair kit 1 85052 60049 Sliding load handle replacement kit 1 85052 60047 Agilent product note 5A 1 5956 5342 a All female connectors on the precision devices in this kit are slotless connectors Refer to Inspect Female Connectors on page 3 4 85052B 6 3 Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85052B Calibration Kit mn n H ta Male to Male Female to Female Male to Female Adapter Adapter Adapter Female Male Male Male S A L A 9v 1 SSCS KSS Female Female Offset Short Offset Open 7 mm Open End Wrench 5 16 Inch Torque Wrench 8 in Ib 90 N cm pk501b 6 4 85052B Replaceable Parts Introduction Figure 6 2 Replaceable Parts for the 85052B Calibration Kit 85052B Female Male Connector Connector Gage Gage Centering Beads Male Female Gage Gage Master Master Female Sliding Load Male Sliding Load Not Shown
52. the unique interaction each device type has with the gage Therefore these uncertainties can vary among the different devices For example note the difference between the uncertainties of the opens and shorts The observed pin depth limits in Table 2 2 add these uncertainties to the typical factory pin depth values to provide practical limits that can be referenced when using the gages See Pin Depth on page 2 3 Refer to Kit Contents on page 1 2 for more information on the design of the calibration devices in the kit NOTE When measuring pin depth the measured value resultant average of three or more measurements is not the true value Always compare the measured value with the observed pin depth limits in Table 2 2 on page 2 4 to evaluate the condition of device connectors 3 6 85052B Use Maintenance and Care of the Devices Gaging Connectors When to Gage Connectors Gage a connector at the following times e Prior to using a device for the first time record the pin depth measurement so that it can be compared with future readings It will serve as a good troubleshooting tool when you suspect damage may have occurred to the device e If either visual inspection or electrical performance suggests that the connector interface may be out of typical range due to wear or damage for example e Ifa calibration deviceis used by someone else or on another system or piece of equipment e Initially after every 100 connec
53. ths and measurement uncertainties and provides observed pin depth limits for the devices in the kit If the pin depth of a device does not measure within the observed pin depth limits it may be an indication that the device fails to meet electrical specifications Refer to Figure 2 1 for a visual representation of proper pin depth slightly recessed Table 2 2 Pin Depth Limits Device Typical Pin Depth Measurement Uncertainty Observed Pin Depth Limits Opens Oto 0 0127 mm Oto 0 00050 in 40 0064 to 0 0064 mm 0 00025 to 0 00025 in 0 0064 to 0 0191 mm 0 00025 to 0 00075 in Shorts 0 to 0 0127 mm 0 to 0 00050 in 40 0041 to 0 0041 mm 40 00016 to 0 00016 in 30 0041 to 0 0168 mm 0 00016 to 0 00066 in Fixed loads 0 0025 to 0 0254 mm 0 0001 to 0 0010 in 40 0041 to 0 0041 mm 40 00016 to 0 00016 in 10 0016 to 0 0295 mm 0 00006 to 0 00116 in Adapter 0 0025 to 0 0254 mm Oto 0 0010 in 40 0041 to 0 0041 mm 40 00016 to 0 00016 in 30 0016 to 0 0295 mm 0 00006 to 0 00116 in Sliding loads 0 to 0 0076 mm 0 to 0 00030 in 40 0041 to 0 0041 mm 40 00016 to 0 00016 in 40 0041 to 0 0117 mm 0 00016 to 0 00046 in a Approximately 2 sigma to 2 sigma of gage uncertainty based on studies done at the factory according to recommended procedures b Observed pin depth limits are the range of observation limits
54. ticles or residues behind You can use any source of clean dry low pressure compressed air or nitrogen that has an effective oil vapor filter and liquid condensation trap placed just before the outlet hose Ground the hose nozzle to prevent electrostatic discharge and set the air pressure to less than 414 kPa 60 psi to control the velocity of the air stream High velocity streams of compressed air can cause electrostatic effects when directed into a connector These electrostatic effects can damage the device Refer to Electrostatic Discharge earlier in this chapter for additional information 3 4 85052B Use Maintenance and Care of the Devices Cleaning Connectors 2 Clean the Connector Threads WARNING Keep isopropyl alcohol away from heat sparks and flame Store in a tightly closed container It is extremely flammable In case of fire use alcohol foam dry chemical or carbon dioxide water may be ineffective Use isopropyl alcohol with adequate ventilation and avoid contact with eyes skin and clothing It causes skin irritation may cause eye damage and is harmful if swallowed or inhaled It may be harmful if absorbed through the skin Wash thoroughly after handling In case of spill soak up with sand or earth Flush spill area with water Dispose of isopropyl alcohol in accordance with all applicable federal state and local environmental regulations Use a lint free swab or cleaning cloth moistened with isop
55. til the gage pointer lines up exactly with the zero mark Remove the gage master Gage the device connector refer to Figure 3 2 a While holding the gage by the barrel and without turning the gage or the device connect the gage to the device by interconnecting the male and female connectors Connect the nut finger tight Do not overtighten Using an open end wrench to keep the device body from rotating use the torque wrench included in the kit to tighten the connecting nut to the specified torque Refer to Final Connection Using a Torque Wrench on page 3 14 for additional information c Gently tap the barrel of the gage with your finger to settle the gage reading d Read the gage indicator dial Read only the black signs not the red signs For maximum accuracy measure the connector a minimum of three times and take an average of the readings After each measurement rotate the gage a quarter turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis Compare the average reading with the observed pin depth limits in Table 2 2 on page 2 4 85052B Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 3 5 mm Connectors Note Although male devices are shown in this illustration the procedure is essentially the same for female devices Zero the Connector Gage Connect the gage to the gage master
56. tions and after that as often as experience indicates 85052B 3 7 Use Maintenance and Care of the Devices Gaging Connectors Gaging Procedures Gaging 3 5 mm Connectors NOTE Always hold a connector gage by the gage barrel below the dial indicator This gives the best stability and improves measurement accuracy Cradling the gagein your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing 1 Select the proper gage for your connector Refer to Table 6 1 on page 6 2 for gage part numbers 3 8 Inspect and clean the gage gage master and device to be gaged Refer to Visual Inspection and Cleaning Connectors earlier in this chapter Zero the connector gage refer to Figure 3 2 a d While holding the gage by the barrel and without turning the gage or the device connect the gage to the gage master by interconnecting the male and female connectors Connect the nut finger tight Do not overtighten Using an open end wrench to keep the device body from rotating use the torque wrench included in the kit to tighten the connecting nut to the specified torque Refer to Final Connection Using a Torque Wrench on page 3 14 for additional information As you watch the gage pointer gently tap the barrel of the gage to settle the reading The gage pointer should line up exactly with the zero mark on the gage If not adjust the zero set knob un
57. uirements 2 2 specifications 2 2 equipment required 1 4 85052B Index 1 Index ESD 3 2 protection 3 2 supplies part numbers 6 3 exclusive remedies ii F female connectors 3 4 inspection of 3 4 flowchart troubleshooting 5 2 frequency specifications 2 5 G gage connector 1 4 handling 3 8 3 10 part numbers 6 3 zeroing 3 8 3 10 gaging connectors 3 6 3 8 when to do 3 7 procedures 3 8 sliding loads 3 10 to determine pin depth 3 6 general information 1 1 H handling 3 18 heel strap part number 6 3 humidity specifications 2 2 impedance system A 8 setting A 8 incoming inspection 1 4 information troubleshooting 5 3 inspection damage 3 3 defects 3 3 female connectors 3 4 incoming 1 4 mating plane surfaces 3 3 visual 3 3 isopropyl alcohol as cleaning solvent 3 5 part number 6 3 precautions for use of 3 5 K kit contents 1 2 6 4 6 5 drawing of 6 4 6 5 misuse of ii overview 1 2 Index 2 warranty ii L liability ii limits pin depth 2 4 loads broadband 1 2 part numbers 6 2 M maintenance 3 2 agreements ii improper or inadequate ii of devices 1 7 preventive 1 7 making connections 3 14 ESD protection 3 2 precautions 3 14 manual copyright dates i part number i 6 2 print date i title i mat conductive part number 6 3 mating plane surfaces cleaning 3 5 connector 3 5 inspection of 3 3 mechanical characteristics 2
58. umber Calibration kit Male broadband load Female broadband load Male open Female open Male short Female short Male Sliding load Female Sliding load Male Connector Gage Male Gage Master Female Connector Gage Female Gage Master 85052B 1 5 General Information Calibration Kits Documented in This Manual Calibration Kits Documented in This Manual This manual applies to any 85052B calibration kit with serial number prefix 3106A If your calibration kit has a different serial number prefix refer to the Calibration Kit History section below for information on how this manual applies Calibration Kit History This section describes calibration kits with serial number prefixes lower that the ones listed on the title page 85052B Kits with Serial Prefix 3027A These calibration kits did not have the calibration definitions disk to support the Agilent 8510C network analyzer The part numbers provided in this manual are the recommended replacement parts for these kits The devices in these kits should meet the specifications published in this manual Clarifying the Sex of a Connector In this manual calibration devices and adapters are referred to in terms of their connector interface For example a male open has a male connector However during a measurement calibration the network analyzer softkey menus label a calibration device with reference to the sex of the analyzer s test port connector n
59. usion or recession and conductor damage on DUTs test accessories and the calibration kit devices Do not use the gages for precise pin depth measurements See Table 6 1 on page 6 2 for part number information Connector Gage Accuracy The connector gages are only capable of performing coarse measurements They do not provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices This is partially due to the repeatability uncertainties that are associated with the measurement Only the factory through special gaging processes and electrical testing can accurately verify the mechanical characteristics of the devices With proper technique the gages are useful in detecting gross pin depth errors on device connectors To achieve maximum accuracy random errors must be reduced by taking the average of at least three measurements having different gage orientations on the connector Even the resultant average can bein error by as much as 0 0001 inch due to systematic biasing errors usually resulting from worn gages and gage masters The information in Table 2 2 on page 2 4 assumes new gages and gage masters Therefore these systematic errors were not included in the uncertainty analysis As the gages undergo more use the systematic errors can become more significant in the accuracy of the measurement The measurement uncertainties in Table 2 2 are primarily a function of the assembly materials and design and
60. xd 0 50 0 0 3 001 Coax Lowband load load 7 8 a Ensure system Zo of network analyzer is set to this value b Open short load delay thru or arbitrary impedance c For waveguide the lower frequency is the same as F co 85052B A 11 Standard Definitions Nominal Standard Definitions Blank Forms The standard definitions may be changed to meet your specific requirements Tables A 10 through A 12 are provided to record the modified standard definitions Table A 10 Blank Form for the 8510 Network Analyzer System Z Calibration Kit Label Disk File Name File Number _ N m N N N I I I Frequency L L L L Standard w n 9 w Offset in GHz et N m b b b I E E E E x x x x o m N m 0 0 S V 9 2 N N N 2 I I I 5 u amp D I T I I a E a g x n q 5 9 E 5 a 5 2 5 2 yg a a a a 312 e s E 2 x X X X 214 a 8 x x E 2 3 3 3 A N 3 a 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 a Ensure system Zo of network analyzer is set to this value b Open short load delay thru or arbitrary impedance c Load or arbitrary impedance only d For waveguide the lower frequency is the same as F co A 12 85052B Standard Definitions Nominal Standard Definitions Table A 11 Blank Form for the 872x Series of Network Analyzers
61. yzers sess 1 3 Equipment Required but Noc Supplied kw dwa yow d sva aaa pwa mea chee iiti A ee 1 4 camino De pec OT aac an kra a kk oka A ri e OR EGER BA EE ok a RE AA AR A AY RE EAR ank e 1 4 Recording the Device Serial Numbers l ll ll ll el kn kk en ka kk na kaa kon kon kon a non a konn 1 5 Calibration Kits Documented in this Manual 44 resi akasan bain kdabin Re REESE SR 1 6 Calibradon KIC B e aia a kk 4 a p anke 6654094454 RC ERE CS PE CA M FA W L 1 6 85052B Kits with Serial Prefix 3027A cuui aw kwa kwen da EE RR a aaa kaka kaa a A 1 6 Clariythgthe secar a Cole tUF ces ieee yAqe bp pt ke A a De A AAA PERCY Ed dc dra 1 6 PisvendveN amtenatloe iuo sd p ERE TERESLHREREN4RT ER Abed ees REITCTARAPPSd eg 1 7 2 Specifications Enr ronmental Redalremblitss cepeeqqaa Sw pan k re TETE REEAGX 4v P Z 2 2 Temperature What to Watch Out For IL L Lele eke a ka kn kon kon n a kon n kon a kon knn 2 2 Pechamecal Character Sts iu debe dde pa ian Pici FI ky ed ec dca DETERS rod vey 2 3 Pii Depuis dad ddp ERR n ee eee Cee eee eT ka ko AE T Ad ee E C ER T RN 2 3 ereere Spec ICE ONE ka qaoqae 2p eC Y er ee ee HEX er ee es PER E d 2 5 Residual Errors amer CaliGfaD ON a isie idurira RR EO kid e kad ERR CERA 2 6 ee LP IONE od aeg aede dd a a ak odes ap XE doe dre eo dow bb Ree pe a acd pd ORCI a eb d 2 6 3 Use Maintenance and Care of the Devices i adiscu aMbllcsjr qM NE OPE 3 2 DICIT Maecas rr 3 3 Look for Obvious

Download Pdf Manuals

image

Related Search

Related Contents

Gesture Based PC Interface with Kinect Sensor  User's Guide True RMS AC / DC Power Clamp Meter Model 380940  MANUAL DE INSTRUÇÕES Linha Soteco  Service manual of Outdoor unit on off type  Kompaktanlage MS-120  エンドクレンズ ® の取扱いに関するお知らせ  Ma retraite mode d`emploi  OPERATOR`S MANUAL MANUEL de L  Attestation de conformité S900 - Aer  LT-03052 A PDF file was made for this manual, (see additional  

Copyright © All rights reserved.
Failed to retrieve file