Home
        AD300 & MD300 - Oxygen analyzer and monitor
         Contents
1.    N Teledyne Analytical Instruments 11    Introduction AD300 MD300    Note  Contact the factory for retrofitting an existing instrument for  3 digit LCD display  For RS 232 reconfiguration  see  Section 2 1 6     In addition to the above instrument configuration options  the  following optional equipment is available for your instrument   DIN flow restrictor for sampling gas cylinder directly  P N  B 75401   Universal Pole Mounting Clamp  P N CP 2343   V Mount Wall Adapter P N B 647   0 1 VDC Interface Cable  P NB 75554   RS 232 Interface Cable  P N B 75555     1 3 The R17S Sensor   How it works    The AD300 MD300 uses Teledyne Analytical Instruments Patented  R17S disposable oxygen sensor  The sensor is made up of a sensing  cathode and anode  fuel  immersed in electrolyte and packaged in a  small plastic container  Oxygen entering the sensor reacts with the anode  and a proportional current is collected at the sensing cathode  This  current is sent to the electronics where it is converted into a digital  signal and displayed on the LCD Screen     A flow diverter is supplied with each R17S sensor for use with the  blue    T    adapter  The flow diverter threads onto the end of the sensor  and facilitates the transport of gas into the sensor     The diverter should be removed for ambient air sampling such as a  diving bell     CAUTION  REMOVE AND SAVE THE DIVERTER WHEN THE  SENSOR IS USED IN CONFINED VOLUME  AN APPLICATIONS         Teledyne Analytical Instruments 12    Porta
2.    OPERATING  amp  SERVICE INSTRUCTIONS FOR    AD300    PORTABLE OXYGEN ANALYZER    MD300    PORTABLE OXYGEN MONITOR    P N M AD MD300  06 23 2010       TYPE B EQUIPMENT  Equipment providing a particular degree  of protection against electric shock   particularly regarding       e Allowable LEAKAGE CURRENT    e Reliability of the protective earth  connection  if present           Teledyne Analytical Instruments    AD300 MD300    Copyright    2010 Teledyne Analytical Instruments  All Rights Reserved    No part of this manual may be reproduced  transmitted  transcribed  stored in a  retrieval system  or translated into any other language or computer language in whole  or in part  in any form or by any means  whether it be electronic  mechanical  optical   manual  or otherwise  without prior written consent of Teledyne Instruments   Analytical Instruments  16830 Chestnut Street  City of Industry  CA 91749 1580    FCC Statement    This equipment generates and uses radio frequency energy  and if not installed and  used in strict accordance with the manufacturer s instruction manual  may cause  interference to radio and TV communications  It has been type tested and certified to  comply with the limits for a Class A  and exceeds limits for a Class B  computing  device pursuant to Subpart J of FCC Rules  which are designed to provide reasonable  protection against such interference when installed in a commercial and residential  environment  Operation of this equipment in a resident
3.    The minimum detectable change in battery voltage corresponds to  an increment of about 50 hours  meaning that the battery voltage reading  may not change for several hours at a time     If the monitor is not used for a period of 30 days or more  the  batteries should be removed prior to storage     3 4 Sensor Maintenance   DO  check the sensor for damage or leaks before use   DO  recalibrate after replacing the sensor    DON   T  immerse the R17S sensor in liquid    DON   T freeze the R17S sensor    DON   T  open or try to repair the sensor     Before every use  the sensor  cable and connections should be  checked  Check the sensor for leaks and condensation  Check the cable  for splitting or cracked insulation  Make sure the connections are tight  and dry     In the event that the sensor has been damaged  consult the Material  Safety Data Sheet available from the factory or online at www teledyne   ai com for handling guidelines     3 5 Gas Sampling    3 5 1 Discrepancy in Readings    The AD300 MD300 instrument should be used to measure the  oxygen concentration exiting gas cylinder or another or life support  system  When a discrepancy in oxygen readings is detected  the oxygen  analyzer readings should be verified by checking the AD MD300 battery  condition and calibration using 100  oxygen  If the monitor can be  calibrated  the unit can be assumed to be in good working order and  capable of providing readings to specification  If  after reinstalling the         Teledy
4.   N  in Table 3 2  The error code 65  would indicate that a key is stuck and this key is the Silence key     In addition to supplying visual error codes  the audio device will  beep a number of times to indicate the general error in case the display is    not functional     NOTE  To reset the watchdog error code  Remove one battery for  5 seconds and replace  If the error persists contact your  locale representative or Teledyne for assistance          Teledyne Analytical Instruments 34       Portable Oxygen Monitor Service Manual    Table 3 2 Error Codes                         Error Code Audio Beeps Error  Indeterminate 2 The watchdog timer has timed out indicating a   serious software error   30 3 Analog output is different from the expected value   May indicate a shorted or over loaded analog  output or a failure of the analog to digital converter  or digital to analog converter circuit    50 5 The ADC circuit failed during POST    6  5  6 A stuck key has been detected  The second digit    There is no in this example  5   on the display shows which  ae key is stuck   indication of  which key is 0   Low Alarm  stuck  1   Batt   2   Cal   3   Up   4   Down   5   Silence   6   High Alarm   7   Alarm Test   8   Key Lock   7 2  7 A failure has occurred during the automatic      calibration of the digital to analog converter  DAC    diepe pe circuit  The second digit  in this example  2    indication of h thet f diaital t loa fail  the type of shows the type of digital to analog
5.  19 to 100 Percent    LO alarm  18 to 99      Alarm thresholds are digitally stored and  alarm affectivity is digitally compared  against the value calculated by the  microprocessor  Hence  the accuracy of the  alarms is the same as the accuracy of the unit  as specified above     37    Appendix AD300 MD300    A 2 Spare Parts List    QTY PART NO  C72144 R17S  B50057   A268   CP 2367   CP 2368  B75401   B99   C74721    Se U HA HA Ha eS Re eS    DESCRIPTION   Micro Fuel Cell R17S   R17S flow diverter   Tee adapter  22 mm    22 mm female tube adapter   22 mm male tube adapter  Inflator sample adapter assembly     AA    size alkaline battery  Sensor cable assembly    A 3 Optional Accessories    A585  CP2345  CP2344  B647  B 75554  B 75555    Hd hm hd hd SS hm       DIN 300 flow Restrictor  Universal Mounting Clamp     V    mount Pole Clamp     V    mount Wall Adapter  0 1 VDC Interface Cable  RS 232 Interface Cable    Note  Orders for replacement parts should include the part  number  if available  and the model and serial number of  the instrument for which the parts are intended     Teledyne Analytical Instruments 38    Portable Oxygen Monitor Appendix    Orders should be sent to   TELEDYNE Analytical Instruments    16830 Chestnut Street  City of Industry  CA 91749 1580    Phone  626  934 1500  Fax  626  961 2538  Web  www teledyne ai com    Or your local representative          Teledyne Analytical Instruments 39    
6.  centimeter of water pressure pulse will produce a 0 11 atmosphere  or an  11  increase in the total and therefore partial pressure of oxygen  An un   pressurized reading of 50  oxygen will increase to 55 3  if the sensor is  subjected to a pressure of 100cm H20  The reading will rise proportionally  less for smaller pressures     2 3 3 Humidity    Humidity does not directly affect the accuracy of the sensor s  measurement  However  when a nebulizer or other device is used to  increase moisture levels in gas mixtures  the moisture actually dilutes  the mixture  This dilution effect decreases the oxygen concentration     For example  if an 80  oxygen gas mixture is humidified to  saturation at room temperature  the resulting gas mixture will contain  only 77 5  oxygen  Your portable oxygen monitor accurately measures  decreases in the oxygen concentration due to the dilution effects of  moisture added to gas mixtures     As with all oxygen sensors  excessive condensation on the sensing  surface of the R17S will block the diffusion of oxygen to the sensor   rendering it inoperative  We recommend installing the sensor on the dry  side of the breathing circuit at all times          Teledyne Analytical Instruments 25    Operation AD300 MD300    2 3 4 Discrepancy in Readings    When a discrepancy in oxygen readings is suspected  the oxygen  analyzer readings should be verified by checking the AD MD300 battery  condition and calibration using 100  oxygen  If the monitor can be  cali
7.  failure   DAC failure  0   Measurement          1   High Test  2   Low Test  3   Offset Cal  4   Gain Cal            Teledyne Analytical Instruments 35       Service Manual AD300 MD300    3 8 Other Problems with the Instrument    Most other problems arise from either mechanical damage from the  instrument falling from a bench or table  or electronic component  failure  In these units  repair or troubleshooting the PCB or individual  component on the board is not feasible  It requires specialized test  equipment and probes not generally available to the public  Under most  circumstances a replacement of the entire PC Board is recommended   The instrument must be returned to the factory for PCB installation     Occasionally  depending on the environment of use  keys can  become stuck or function erratically due to contamination  Use a mild  non abrasive cleaner solution to periodically clean the keypad and  screen  An aerosol jet spray of the type commonly used to clean  computer keyboards can be used to dislodge dirt and accumulations  from the keypad     The Error Code Table  Table 3 2  includes a description of certain fault  codes which are diagnostic of some common  usually electronic  problem  with the instrument  Some of these codes refer to specific components on the  PCB that are problematic or have failed  These codes are useful in reporting  a problem with your instrument to Teledyne Customer Service  If an error  code is indicated on your monitor record the number 
8. 1  APPROVED DEVICE     To reconfigure the analyzer from a digital  RS 232  output to  analog  0 1 VDC  output  use the same procedure except in step 3  remove the jumper from JP7 and replace it at JP3     2 2 Use    2 2 1 Gas Sampling    The AD MD300 instruments can be used to measure the oxygen  concentration in two basic modes          Teledyne Analytical Instruments 22    Portable Oxygen Monitor Operation       From the low pressure inflator hose using the sample  adapter as shown in Figure 2 10        Directly from the high pressure gas cylinder using the  optional DIN restrictor shown in Figure 2 11     NOTE  Both sample modes require the use of the gas diverter  cap and tee adapter as show in Figure 2 3  Make sure you  have calibrated the unit before proceeding     INFLATOR HOSE SAMPLING  Connect the inflator hose adapter  supplied with the unit to the tee adapter as shown in Figure 2 10  Open  the gas bottle valve and allow gas to flow for approximately 30 seconds   When a stable reading is observed close the cylinder valve  Allow  reading to settle for a few seconds  This is the oxygen value of the  sample gas     Low Pressure  Assembly B 77263   To AD MD         Inflator Adaptor Analyzer  Restrictor 02S  Hose Barb Adapter RS  3 16 ID  5 16 OD     H 3 16 ID  5 16 OD  Vinyl Tubing S  Vinyl Tubing      4 F  To Low Pressure i E S  Inflator Hose Hose Connector  Hose Connector TEE    Figure 2 10  Low Pressure Gas Sampling    HIGH PRESSURE SAMPLING  To measure the oxygen co
9. MAGE TO THE UNIT AND  A N BATTERIES     Note  Use alkaline batteries only  Other battery types will  produce erroneous battery test readings     3  Install 3    AA    alkaline batteries into the holder as shown in  Figure 2 6  Each battery has its own slot  To insure proper  polarity  place the bottom  flat  or negative end of the battery  in the end of the holder marked            Place the top  button  or    positive end of the battery in the end of the holder marked           Do this for each battery        Figure 2 6  Installing Batteries    4  Re install the battery compartment door  When the unit is  first turned on the display will momentarily display all LCD  segments  During this period diagnostic tests are being  conducted to insure the circuits are functioning correctly   The unit will activate the audible and visual alarms for about  1 second  The LCD will flash continuously indicating the  unit is in the unlocked position ready for calibration     Note  When batteries are first installed or power is lost for any  reason the instrument defaults to the calibration mode  All  keys except the CAL and ON Off  VO  keys are inoperable  until a successful calibration is achieved     5   MD300 only  To test the batteries  Press the BATT TEST  key once  The battery display  located below the oxygen         Teledyne Analytical Instruments 17    Operation AD300 MD300    2 1 4 Calibration    The AD MD300 should be  calibrated before each use     Whenever new batteries are  in
10. OR ELECTROLYTE IS CAUSTIC  DO NOT  LET IT COME IN CONTACT WITH SKIN  IF IT DOES     FLUSH AFFECTED AREA WITH WATER  DO NOT  ATTEMPT TO OPEN OR REPAIR THE SENSOR     2  Plug one end of the coiled cable into the jack receptacle on the  back end of the R17S sensor and secure in place with the capture  nut located at the base of the connector  See Figure 2 1        Figure 2 1  Installing the RI7S Sensor    3  Plug the other end of the coiled cable into the receptacle located on  the right side of the unit and secure it in place using the capture nut     See Figure 2 2   Capture Nut    Monitor OON  Sensor Connection      Connection tt          Figure 2 2  Sensor Cable Connection to Monitor         Teledyne Analytical Instruments 14    Portable Oxygen Monitor Operation    Note  When the AD MD300 instrument is used for diffusion  sampling the plastic flow diverter must be removed from  the R17S sensor  If the sensor is used in sample mode the  diverter must be used as shown in Figure 2 3     Micro Fuel Cell    O Flow  N   Diverter    Sg    Sensing  E l Tee  Surface      _ Adapter    Figure 2 3  Mounting the Sensor in the Tee Adapter    2 1 2 Mounting   The AD MD 300 can be mounted in several ways depending on the  optional equipment ordered at the time of purchase  See Section 1 2   2 1 2 1 V MOUNT ADAPTER INSTALLATION    The V Mount Adapter consists of a matching plastic plate with  integral V grooves that attach to the rear of the instrument     To install V Mount Adapter  remove 
11. UNLOCK a second time unlocks the  keypad     2 3 Gas Sampling Sensor Issues    2 3 1 Temperature    The R17S oxygen sensor adjusts for ambient temperature changes  in the range of 0   40  C  32   106  F   However  a small thermal tracking  error may be encountered when the sensor is subjected to sudden  changes in temperature  For example removing the sensor from an air  conditioned car to a hot environment  Holding the sensor in your hand         Teledyne Analytical Instruments 24    Portable Oxygen Monitor Operation    for more than a few minutes can also affect the temperature tracking  which appears as a slow drift on the LCD  No adjustments should be  made during this period since this error will correct itself in about 30  minutes     2 3 2 Pressure    Virtually all gas sensors and monitors measure the partial pressure   not the percentage  of the gas that they sense  The only time that these  instruments can accurately read percentages is when the total pressure  does not vary over time between calibrations and use  This is way it is  important to calibrate the AD MD300 oxygen sensor at regular intervals   It is recommended that the unit be calibrated prior to each use or every 8  hours     When the sensor is connected to sample gas that is flowing under  pressure  the sensor may read high  In reality  the percentage of oxygen is  not changing  it is the total pressure that is increasing  producing a  corresponding increase in the partial pressure of oxygen  A hundred 
12. acy is desired   or the concentration  of O2 is in excess of  40   calibration with  100  is recom   mended           N Teledyne Analytical Instruments 32       Portable Oxygen Monitor Service Manual       Symptom    Why    What To Do       The unit has stopped  working and the LCD is  displaying alphanumeric  figures     The AD300 MD300  instruments are equipped  with an electronic  watch  dog   which monitors the  circuitry within the unit for  potential faults and  renders the unit  inoperable until the  condition is corrected    See Watch dog section  below     Several conditions can  activate the  watch dog    Dropping the unit  poor  battery connections  and  radio frequency  interference are the  most common causes   See the watch dog  section for additional  information    A  Disconnect the  batteries and inspect  the contacts for  corrosion  Reconnect  the batteries  If the  unit functions  properly  calibrate the  unit and reset the  alarm values     B  Try anew set of  batteries     C  Increase the distance  between the unit and  any source of radio  frequency  interference  The  sensor cable is a  prime source of  pickup as it can act  like an antenna   Relocate the sensor  cable and if possible  change its coiled  length to    de tune    its  antenna effect   Placing the cable ina  different position may  also help        Alarm sounds flashes  continuously     A   MD300  Readings  are outside alarm  limits          Adjust high and low  alarm setting to be  above 
13. afety alert symbol  These  symbols are found in the manual and on the instrument  The definition  of these symbols is described below     CAUTION  Refer to the instructions for details on the  specific danger  These caution symbols warn of specific    A N procedures  which if not followed could cause bodily Injury     and or damage the instrument     WARNING  This symbol is use to alert the operator of a  condition that could cause bodily harm     NOTE  Additional information and comments regarding a  specific component or procedure are highlighted in the form  of a note     CAUTION  THE MONITOR SHOULD ONLY BE USED FOR THE  PURPOSE AND IN THE MANNER DESCRIBED IN  THIS MANUAL     IF YOU USE THE ANALYZER IN A MANNER OTHER  THAN THAT FOR WHICH IT WAS INTENDED   UNPREDICTABLE BEHAVIOR COULD RESULT  POSSIBLY ACCOMPANIED WITH HAZARDOUS  CONSEQUENCES          Teledyne Analytical Instruments iv    Portable Oxygen Monitor    Table of Contents    Safety Messages ii EKG KEES sential GE Se GE SE GE REKE Ee iv  LIST OF FIQUICS OE N OE EE ee ee vii  List of Ta DIOS siese ees ees oe ees ee es ee Ge De Ee viii  INTOGDUGTION RE OE EEN 9  1 1 Features 11  1 2 Options 11  1 3 The R17S Sensor   How it works 12  ear Lo  AE EE EE EE EE EE 13  2 1 Setup 13  2 1 1 Sensor Installation or Replacement 13  2 1 2 Mounting 15  2 1 2 1 V Mount Adapter Installation 15  2 1 2 2 Universal Mounting Clamp Installation 16  2 1 3 Battery Installation 16  2 1 4 Calibration 18  2 1 5 Alarms  MD300 only  20  2 1 6 O
14. and below O2  value being  displayed        No display     A  Batteries expired   B  Bad battery    A  Check replace  batteries           connection  B  Check battery  connections   C  Calibrate   Keys inoperable cannot   The LOCK UNLOCK key   A  Press  turn unit off is activated which is LOCK UNLOCK key       preventing key operation       once  LCD will flash  indicating keys are  active             Teledyne Analytical Instruments 33       Service Manual    AD300 MD300       Symptom    Why    What To Do       Cannot adjust  calibration or alarm  settings          Critical settings require  two keys be pressed in a  specific order     LOCK UNLOCK key is  active       A  If display is not  flashing press lock  key once to activate  keys  Press desired  function followed by  the Up and Down  key        Note  In the event that none of these procedures produce desired  results  remove the batteries and return the unit to Teledyne for    repair     3 7 Watchdog    The AD300 MD300 is equipped with a watchdog circuit that  continuously monitors the electronics for proper operation  If the watchdog  detects a failure  one of the following codes will appear on the LCD     The error codes can appear on the LCD when batteries are first  installed  during normal operation or if the unit is subjected to extreme  shock  In some cases an additional digit is used in the error code to  supply additional information  For example  the error code 6 and 7 are  followed by another digit listed as
15. and report it to the  Customer Service Department at the address below     3 9 Return Authorization for Service    For any service beyond sensor and battery replacement  the  instrument must be returned to the factory  A return merchandise  authorization  RMA  number must be obtained from Teledyne  Analytical Instruments prior to returning an instrument for service  You  can request a RMA number via email by contacting us at     tetci_customerservice  teledyne com  You can also contact us at the address below     Customer Service Department  TELEDYNE Analytical Instruments    16830 Chestnut Street  City of Industry  CA 91749 1580 USA    Phone  626  934 1500  Fax  626  961 2538  Or via the web at  www teledyne ai com         Teledyne Analytical Instruments 36    Portable Oxygen Monitor Appendix    Appendix       A 1 Specifications    Range     Accuracy     Response Time   Battery Life     System Power   Sensor Type   Expected Life     Dimensions     Sensor Cable     Storage Temp     Operating Temp     Alarm ranges     Alarm accuracy          Teledyne Analytical Instruments    0 100  Oxygen     2  of full scale  at constant temperature  and pressure     90  in less than 8 seconds at 25   C    Approximately 2000 hr  continuous use in  a non alarm condition    3 AA alkaline batteries   Class R17S  36 months in air    2 5  W x 1 25  D x 4 5  H  66 mm x 33   mm x 111 5 mm     Retracted  2 ft   Extended  10 ft   10 30   C  continuous   5 50   C   Intermittent    0 40   C   HI alarm 
16. ble Oxygen Monitor Operation    Operation       Note  Upon receipt  INSPECT THE ENTIRE UNIT FOR  DAMAGE  Check the unit and all included accessories for  broken or loose parts  If damaged  DO NOT USE  Notify  the shipper  and consult Teledyne Analytical Instruments     Note  This equipment is internally powered using 3 AA batteries     2 1 Setup  To set up and use your AD300 analyzer or MD300 monitor   1  Install the sensor   2  Install the batteries   3  Calibrate the unit   4  Set the alarms  MD300 only      The control keys are designed for easy operation  A LOCK UNLOCK  key has been incorporated to prevent accidental changes to critical settings   This eliminates unwanted changes in calibration or alarm settings from  accidental touching or bumping of the keys  To further reduce the possibly of  incorrect adjustments at least two keys must be pressed in order to modify a  critical calibration or alarm set point value     Note  The ALARM SILENCE  2   and BATT TEST key continue  to operate normally when the lock feature is activated     2 1 1 Sensor Installation or Replacement    Note  The R17S oxygen sensor must be installed before the  oxygen analyzer monitor can be operated    1  Remove the new sensor from its protective bag  Inspect the  sensor for damage or electrolyte leakage  If the sensor is  damaged  obtain a replacement  Do not use the defective sensor  as it may damage the unit          Teledyne Analytical Instruments 13    etalon AD300 MD300    WARNING  THE SENS
17. brated  the unit can be assumed to be in good working order and  capable of providing readings to specification  If  after checking the unit   the discrepancy in oxygen readings persists  the gas being analyzed  should not be used  Further investigation should be made until the  discrepancy in readings is resolved  The troubleshooting section of this  manual may provide additional assistance in locating the problem          Teledyne Analytical Instruments 26    Portable Oxygen Monitor Operation    2 4 Do   s and Don   ts  DO      Read all of the directions before using for the first time   Calibrate before every use   Visually inspect the sensor for leakage before each use   Calibrate using 100  oxygen and check in air   Check the HI and LO alarm settings prior to each use     Test batteries regularly and replace when battery indicator shows  low battery  no bars remaining     Make sure keys are locked by using the LOCK UNLOCK key  feature     Keep the unit  sensor and connections dry   Recalibrate after replacing the batteries   Recalibrate after replacing the sensor    Use properly installed alkaline batteries only   Make sure the R17S sensor is properly attached     Remove the plastic flow diverter when analyzing non flowing  samples     Remove and save the plastic flow diverter when using the sensor  in non flowing applications      MD300 only  Perform an alarm test before each use  press the  ALARM TEST key          DON   T        Use this monitor if you suspect any malf
18. ess oxygen  which will introduce errors into the  calibration  Air calibration should only be used for  monitoring oxygen levels between 21  and 40  and  should never be used where a high degree of accuracy  is needed     CAUTION  DO NOT ADJUST THE CALIBRATION SETTINGS IN  AIR AFTER THE 100  CALIBRATION  AS THIS WILL  AN CANCEL THE MORE ACCURATE 100   CALIBRATION  THE 100  CALIBRATION MAY BE  REPEATED AS MANY TIMES AS DESIRED          Teledyne Analytical Instruments 19    Operation AD300 MD300    8   MD300 only  Adjust alarm set points to desired level  according to the procedures in Section 2 1 5     9  Press the LOCK UNLOCK key to hold settings  The unit is  now ready for use     2 1 5 Alarms  MD300 only     Note  Before attempting to change the alarms it is necessary to  unlock the control keys by pressing the LOCK UNLOCK  key  When pressed the display will start flashing     To set the alarms on the MD300 monitor  see Figure 2 8      1  To set the HI alarm  Press the HI ALARM  SET key once  Press the UP and DOWN    arrow key until the desired value is displayed   9    in the lower right corner of the display  Press SS    lock key or proceed to LO alarm settings NINN    Note  The HI and LO alarms may be defeated  by setting the HI alarm limit one step  above 100   The display will blink  ALARM OFF continually in this mode      D     2  To set the LO alarm  press the LO ALARM SET  key once  Press the UP and DOWN arrow key          10 al Tm  Hag    ALARM OFF       al     to 
19. gure 2 8  Setting the Alarms  MD300                     cceeeeeeeeeeeeeeees 20  Figure 2 9  0 1 VDC or RS 232 Digital Output Port                 0 22  Figure 2 10  Low Pressure Gas Sampling         sesse ees ee ee 23  Figure 2 11  High Pressure Gas Sampling        sse sees sesse sesse ede ee 24         Teledyne Analytical Instruments vii    AD300 MD300       List of Tables  Table 3 1 Troubleshooting                c  ccccceeeeeeeeeeeeneeeeeeeeeeeeeeseseaeees 31  Table 3 2 Error COd S         c ccccccccccceccceececeeeececceeeceeceecaeeceecueeaeeseeeas 35         Teledyne Analytical Instruments viii    Oxygen Analyzer Introduction    Introduction    OT    The Teledyne Analytical Instruments AD300 Oxygen Analyzer  here after referred to as AD300 and MD300 Oxygen Monitors with  alarms here after referred to as MD300 are portable instruments that  provide fast and accurate oxygen monitoring  In addition  the model  MD300 incorporates an audio visual alarm capability  These instruments  are designed to monitor up to 100  oxygen concentration in breathable  gas mixtures  Because they are microprocessor based  the AD MD300  series have a unique combination of features that make them very easy  to use  The operator interface is accomplished through a series of  buttons located conveniently on the front face of the instrument  The  MD300 front face interface is shown in Figure 1 1  The AD300 is  similar but without the battery and alarm test buttons is shown in Figure  1 2     LCD Bat
20. ial area may cause interference   in which case the user  at his own expense  will be required to take whatever measures  may be required to correct the interference     Note  The above statement is required by the FCC for any device  that incorporates microprocessors     Warranty    Teledyne warrants that the goods are free from defects of material and of construction  for a period of 2 years from the date of shipment from Teledyne  The Class R 17S  Micro Fuel Cell is warranted for two years from the date of shipment from Teledyne   The liability of Teledyne if any  shall be limited solely to the replacement and repair of  the goods and shall not include shipping costs or other incidental damages as defined in  Section 2 715 of the U S  Uniform Commercial Code     This warranty is null and void if any goods are subjected to misuse  negligence     accident  or repairs other than those performed by Teledyne or an authorized service  center          Teledyne Analytical Instruments ii    Portable Oxygen Monitor    About This Manual       The AD300 MD300 Operator s Manual provides both introductory  and detailed information for configuring and operating these  instruments  The manual takes you from the time you unpack the  instrument until you complete the first gas analysis  The bulk of the  manual contains operating procedures and information  There are also  cautions  warnings  and guidelines to ensure that your monitor operates  normally and to its full potential  A troublesh
21. ing device     1  Insert one end of the interface cable into the output port on  the side of the instrument  See Figure 2 9     2  Insert the other end into the analog recorder device  Make  sure the device is equipped to handle a 0 1 VDC signal     When properly calibrated  the output signal generated by the  analyzer is linear and proportional to the oxygen concentration     If you requested Option B  RS 232 digital output  at the time of  purchase  a digital RS 232 signal is output from the output port shown in  Figure 2 9  Use the optional RS 232 Interface Cable  P N B 75555   available from Teledyne for connection to a standard RS 232 port on a  computer or other suitably equipped digital device        N Teledyne Analytical Instruments 21    oporou AD300 MD300               Output Jack    Figure 2 9  0 1 VDC or RS 232 Digital Output Port    If your instrument is set for analog  0 1 VDC  output  you can  reconfigure it to use the digital output by changing a jumper on the  internal PC board     To activate the digital output     1  Remove the batteries and remove the five screws that hold the  case together     2  Remove the rear case section leaving the PCB in the front half of  the case     3  Remove jumper at position JP3 and reinstall it at position JP7     4  Replace the rear cover and secure in place with five screws   Install the batteries and recalibrate per Section 2 1 4     CAUTION  RECORDER RS232 OUTPUT SIGNAL SHOULD ONLY  BE CONNECTED TO AN EN60601 1 IEC60606 
22. lizes before  making any  adjustments     B  Make sure that at  least 6   30 cm   of  tubing is attached to  the exhaust side of  the tee adapter to  prevent back filling   O2 flow rate should  not exceed 5 l min     C  Oxygen  concentration at the  sensor is significantly            Teledyne Analytical Instruments 31       Service Manual    AD300 MD300       Symptom    Why    What To Do       D     higher than 21    Take the instrument  to a well ventilated  area and check the  reading again     Try calibrating with a  known good sensor   if this fails  see  symptom  Reading  drifts over 2 3               The display is flashing  SENSOR    The unit has detected a  fault in the signal from  the sensor     Sensor has expired     The sensor has been  exposed to a gas  containing little or no  oxygen     g    Check sensor cable  connections and  make sure they are  completely inserted  into the mating  connector and the  capture nut is firmly  in place     Expose the sensor  in 100  O2 and  check calibration     The sensor output  has fallen to a level  where it is no longer  usable  Replace  sensor        The oxygen reading  fluctuates or appears to  be incorrect        Like all O2 sensors  the  R17S detects the  changes in the partial  pressure of O2        2    During calibration   make sure there is  no restrictions on  exhaust side of  sensor  If the reading  changes with flow   the sensor is  pressurized or there  may be a leak in the  system     If a high degree of  accur
23. ncentration  directly from the high pressure cylinder  Install the optional DIN connector  onto the tank and insert the blue tee adapter and sensor as shown in Figure 2   3  Slowly open the cylinder valve until you detect gas flowing  When a stable  reading is observed close the cylinder valve and allow the reading to settle for  a few seconds  This is the oxygen value of the sample gas     CAUTION  MAKE SURE YOU DO NOT PRESSURIZE THE  SENSOR DURING CALIBRATION OR USE  AS THIS  AN WILL PRODUCE ERRORS IN THE READINGS          Teledyne Analytical Instruments 23    Operation AD300 MD300       To AD MD  Analyzer  O2 Sensor  R17S  A  gie 3 16 ID  5 16 OD  ss Vinyl Tubing    T 1352  EZ 7  5 Vent  P         N  O Ring CP 2368  Hose Connector  A 268 TEE    300 DIN Threaded  Pressure Restrictor  A585    Figure 2 11  High Pressure Gas Sampling    When monitoring for oxygen in confined volumes such as a diving  bell  the flow diverter must be removed from the R17S sensor so that it  does not interfere with the rapid exchange of gases to and from the  sensing surface of the sensor     CAUTION  FAILURE TO REMOVE THE DIVERTER IN THESE  APPLICATION AREAS WILL RESULT IN A MARKED  LOWERING OF THE RESPONSE TIME OF THE  SENSOR     The LOCK UNLOCK key can be used to prevent accidental  changes to the front panel key adjustments  The LOCK UNLOCK key  acts as a toggle  pressing LOCK UNLOCK once renders inactive all  keys except the ALARM SILENCE  22  and BATT TEST  MD300  only  keys  Pressing LOCK 
24. ne Analytical Instruments 30    Portable Oxygen Monitor Service Manual    unit  the discrepancy in oxygen readings persists  the problem is most  likely elsewhere  Further investigation should be made until the  discrepancy in readings is resolved  The troubleshooting section of this  manual may provide additional assistance in locating the problem     Note  The MSDS on this material is available upon request  through the Teledyne Environmental  Health and Safety  Coordinator at  626  934 1592 or can be viewed and  downloaded from our web site at www teledyne ai com     3 6 Troubleshooting    The AD300 and MD300 oxygen monitors provide a variety of built in  safety features that prevent its use when a fault is detected  When a unit  displays the message V SENSOR and sounds the audible and visual alarm  continuously  it is an indication of a faulty connection between the sensor and  the unit or an expired or faulty sensor  To determine where the difficulty lies   refer to the following guidelines in Table 3 1     Table 3 1 Troubleshooting          Symptom Why What To Do  New sensor responds If the sensor is new and   A  Wait 30 minutes and  slow or drifts  was just removed from recalibrate     its sealed bag it may  need to run for several  hours           Sensor will not read below  22   after calibration in  100  O2        Calibration in 100  was  invalid or the room air is  contaminated with  excess oxygen        A  Recalibrate using  dry gas making sure  the reading  stabi
25. on the  lower portion of the LCD display with the low alarm set point appearing  at the lower left and the high alarm appearing on the lower right of the  display  The alarm circuitry provides both an audible and visual alarm        j   Teledyne Analytical Instruments 10    Oxygen Analyzer Introduction    1 1 Features    The AD300 MD300 are rugged  versatile instruments capable of  rapidly measuring the oxygen content of breathing gases commonly used  for diving purposes  Gas analysis is accurate to within  2  of full scale  over the operating range of 0 100  at constant temperature and pressure     The following features are standard on the AD300 and MD300  instruments     Large easy to read 3 1 2 digit LCD display  see options   Automatic LCD back lighting upon key press  Microprocessor controlled   Up Down front panel controls   Sensor fail disconnect alarm indicator  audible and visual   Alarm silence button   2000 operating hours from 3 AA alkaline batteries  Battery status indicator   Rugged high impact ABS construction   Splash resistant case    Long life  36 months in air  class R17S sensor    VDC digital output  optional RS 232     The MD300 features include the above plus     User defined set point controlled concentration alarms    Battery test function    1 2 Options    The following instrument options are available for both the AD300  and MD300 units     A Option   3 digit LCD display instead of 3 1 2 digit  B Option   RS 232 digital output instead of 0 1 VDC    
26. ooting section is available  to assist you with common problems and a complete product  specifications and spare parts list is included as an appendix     A Quick start manual is also supplied which provides the minimum  information you need to use the instrument immediately     Chapter 1  An introduction to the monitor and its components   features and applications    Chapter 2  Step by step set up procedures and information    Chapter 3  A guide for daily operational maintenance and  troubleshooting    Appendix  Specifications and available spare part options for the  monitor  and detailed application considerations to aid in  troubleshooting  etc     How To Use This Manual    This manual is designed to walk you through the initial set up of the  AD300 Portable Oxygen Analyzer and MD300 Portable Oxygen Monitors  It  also serves as a quick reference guide to help you with specific questions or  operating problems     Before turning the instrument on  you are advised to read the safety  information on the next few pages and the information found in  Chapters 1 and 2  These Chapters acquaint the user with the instruments  use and operation before placing it into operation        N Teledyne Analytical Instruments iii    AD300 MD300    Safety Messages       Your safety and the safety of others are very important  Please  carefully read the following safety messages     Safety message are indented to alert the user of potential hazards   Each safety message is associated with a s
27. select a value  A built in safety feature will Figure 2 8  Setting the Alarms  MD300     not allow you to select a LO alarm value less  than 18   Press LOCL key to save settings     Note  The MD300 is designed to prevent crossing of HI LO alarm  settings  If you attempt to set the LO alarm higher than the HI  alarm  it will push the HI alarm setting up as you continue to  rise the LO alarm set point  This also applies when  attempting to set the HI alarm lower than the LO alarm     3  To test the alarms  unlock the controls by pressing the  LOCK UNLOCK key and then press ALARM TEST key  The  Lo alarm followed by the HI alarm will blink and the audible  visual alarm will are activated for about 1 second     4  To silence the alarm buzzer for 115 seconds press the alarm  silence key  42    flashing red indicator light           Teledyne Analytical Instruments 20    Portable Oxygen Monitor Operation    5  To disable the alarms set the high alarm set point above 100    Press LOCK UNLOCK key to save settings     Note  When the alarm is in the OFF condition  set point above  100   the ALARM OFF status message blinks slowly on  the LCD below the oxygen readout     2 1 6 Output 0 1 VDC or RS232    Both the AD300 and MD300 provide signal outputs for use with  recorders and computers  The instruments are supplied standard with a  0 1 VDC output  An optional 0 1 VDC Interface Cable  P N B 75554   is available from Teledyne for this purpose     To connect the analyzer to an analog record
28. stalled or removed for any    reason  the    defaults to the calibration mode  with the LCD display flashing    00 0  Only     I O  keys are functional at this   5 EP Er  point  On the MD300 monitor  the Ke    readout display  should show a bar graph of the remaining  life to the batteries  In the case of fresh batteries  it should  illuminate all 5 segments from the left to the right end of the  bar     AD300 only   The bar graph indicator is on continuously  whenever the instrument is powered on     oxygen monitor    the CAL and ON OFF       alarm set points will also need to       be reset after calibration is   1 j    completed     Note  For first time and for  routine maintenance    Figure 2 7  Calibration Sequence    calibrations  make sure  the sensor and sensor cable are installed correctly before  attempting to calibrate the instrument    To calibrate the instrument  See Figure 2 7      1   2   3     Turn the unit on by pressing the ON OFF  I O  key   Check the batteries by pressing the BAT key     If the LCD is not flashing  press the LOCK UNLOCK key to  unlock the keys  LCD will flash indicating changes can be  made to the settings  If the batteries have just been installed  the LCD will flash 00 0    As with most oxygen analyzer s  the highest level of  accuracy is achieved when calibration is conducted using  100  oxygen  After installing the flow diverter as noted in  Section 2 1 1  insert the sensor into the plastic tee and         Teledyne Analytical Instrumen
29. tery  Display Status    Alarm Status AE     Alarm Setpoint  H rr 8   Up Down Keys High Alarm  St VEG N Setpoint  Low Alarm A High Alarm  Set   Set   sel   Battery Test Aa J Alarm Test  Calibration          F Lock Unlock    Alarm    i On Off    Silence    Sensor Status       Figure 1 1  MD300 Front View       47 Teledyne Analytical Instruments 9    Introduction AD300 MD300    LCD Battery  Display Status            Sensor Status    Calibration AG Lock Unlock  dk EE    Alarm    Silence On Off    Figure 1 2  AD300 Front View    The LCD display consists of up to 3 5 characters plus a decimal  point indicator capable of displaying up to 105    See Section 1 2  Options for alternate display configuration   An integral battery life  indicator is displayed continuously on the AD300 and    on demand     with the MD300     The instrument is powered by three AA alkaline batteries and is  designed to operate for 2000 hours in Non alarm State     Oxygen analysis is linear across the single range of 0 100  using  Teledyne   s class R17S oxygen sensor  A unique sensor failure alarm is  incorporated which warns the user if the sensor signal is lost or low   When this occurs  the N SENSOR display flashes  The alarm buzzer can  be silenced by pressing the ALARM SILENCE key  If the alarm  condition is not corrected within 115 seconds the alarm buzzer will  reactivate     The MD300 unit incorporates a dual concentration alarm with  individual user defined set points  The set points are displayed 
30. the battery compartment door  by prying up the hinged latch at the bottom of the cover  then slide the  adapter plate into grooves provided in rear case  Replace the battery  compartment door and secure door latch  See Figure 2 4     Note  The door latch is a tight fit onto the battery cover  Use a  coin to gently pry up the latch          Teledyne Analytical Instruments 15    Operation AD300 MD300                   Figure 2 4  V Mount Adapter Installation    2 1 2 2 UNIVERSAL MOUNTING CLAMP INSTALLATION    The Universal Mounting Clamp is supplied with a 1 4 20 screw for  securing the clamp to the rear of the instrument  A threaded brass insert is  installed on the back of the instrument for this purpose  See Figure 2 5     Brass Insert          Figure 2 5  Brass Insert for Universal Mounting Clamp    2 1 3 Battery Installation    Note  Three    AA    alkaline batteries must be installed in the unit  before the monitor will operate  The unit must be  recalibrated whenever new batteries are installed  For  the MD300  the HI and LOW alarm set points must be  reset to the desired values     To install the batteries     1  Turn the unit off  if it is on      2  Hold the instrument face down in the palm of your hand  Use  a coin to pry up the latch that secures the battery  compartment door  Remove the battery compartment door        j   Teledyne Analytical Instruments 16    Portable Oxygen Monitor Operation       CAUTION  IMPROPER INSTALLATION OF THE BATTERIES    MAY RESULT IN DA
31. to be made within this instrument  If a problem arises  with either of these models that cannot be corrected by recalibration   changing the batteries  or replacing the sensor as described elsewhere in  this manual  the unit must be sent back to the factory for repair or  replacement  See Section 3 9 for instructions on obtaining a Return  Merchandise Authorization  RMA  number before sending a unit back  to Teledyne for repair     3 2 Overall Maintenance    The AD MD 300 series of instruments require very little  maintenance  other than calibration  checking and changing the batteries  and sensor  and cleaning the plastic housing  Occasional cleaning of the  plastic surface can be done with isopropyl alcohol  Should any part of  the instrument malfunction or fail to perform  the unit should be  removed from service  There are no user serviceable components within  the instrument     3 3 Battery Maintenance    DO  Test batteries regularly  replace immediately when all 5 bars  are missing      DO  Always use alkaline batteries          Teledyne Analytical Instruments 29    Service Manual AD300 MD300    DO  Recalibrate after replacing batteries     The MD300 instrument incorporates a battery test feature that  indicates the approximate amount of useful life remaining on the set of  batteries  Excessive alarm activation will wear down the battery faster  than usual  The AD300 instrument continuously displays the battery  status bar graph  There is no BATT TEST key on the AD300  
32. ts 18    Portable Oxygen Monitor Operation    connect to a supply of pure dry oxygen flowing at 1 2 liters  per minute     5  Wait about 20 seconds to insure the sample line is completely  purged with the calibration gas  Press the CAL key  The  LCD will count down from 9 to 0  During this time the  microprocessor is measuring the sensor output to determine  the gas concentration and selects the calibration range i e   100  or 20 9   When the calibration is complete the LCD  will display the gas value  Press the LOCK UNLOCK key to  save the calibration data     Note  The AD300 and MD300 can only be calibrated using 100   oxygen or room air 20 9   Improper calibration or use of  other gas concentrations will activate the VSENSOR  indicator  To repeat the calibration press the  LOCK UNLOCK key and press the CAL key     6  Remove the sensor from the oxygen supply and confirm the  LCD reads less than 22  in room air  It is not necessary for  it to read exactly 20 9      7  Itis important to perform the calibration carefully and  thoroughly  using calibration gases that are free from  contaminates  Wait for a stable reading before locking in  calibration point  The accuracy of the instrument is only as  good as the procedure used to calibrate it     Note  A single point air calibration is not recommended unless  the sensor can be exposed to a known source of fresh  outdoor air  Confined areas such as rooms where high  concentrations of oxygen are released may be enriched  with exc
33. unction   Do not use the instrument if the oxygen readings are drifting   Use the instrument in the presence of flammable liquids or gases     Use anything but alkaline batteries        N Teledyne Analytical Instruments 27    Operation AD300 MD300    Freeze the sensor or instrument    Open or try to repair a leaking or broken sensor   Immerse the unit or sensor in any liquid    Pass hot or cold gas mixtures over the sensor   Adjust the reading in air after 100  calibration    Use cleaning agents or liquids in the cable receptacles or around  the battery compartment     Store the unit in hot or humid environment     Expose the LCD to excessive sunlight          Teledyne Analytical Instruments 28    Portable Oxygen Monitor Service Manual    Service Manual       3 1 General Service Information    The Teledyne Model AD300 and MD300 series of portable oxygen  monitors are designed to be robust yet compact in size  In order to  achieve maximum reliability in a microprocessor based instrument  a  single PC board is used which relies exclusively on surface mount  technology  Without access to specialized probes and test equipment   troubleshooting and repair of circuit board components are not feasible   A factory replacement of the entire PC board is more cost effective than  a field repair of an individual component    With the exception of replacing the sensor or batteries there are no  user serviceable components inside the unit  There are no potentiometers or  other adjustments 
34. utput 0 1 VDC or RS232 21  2 2 Use 22  2 2 1 Gas Sampling 22  2 3 Gas Sampling Sensor Issues 24  2 3 1 Temperature 24  2 3 2 Pressure 25  2 3 3 Humidity 25  2 3 4 Discrepancy in Readings 26       Teledyne Analytical Instruments v    AD300 MD300    2 4 Do   s and Don ts 27  Service Manual  sers RE NE VEE EER KA GE VERE ee Ne WE DO KON 29  3 1 General Service Information 29  3 2 Overall Maintenance 29  3 3 Battery Maintenance 29  3 4 Sensor Maintenance 30  3 5 Gas Sampling 30  3 5 1 Discrepancy in Readings 30  3 6 Troubleshooting 31  3 7 Watchdog 34  3 8 Other Problems with the Instrument 36  3 9 Return Authorization for Service 36  Te  se    EE EE EE EE N 37  A 1 Specifications 37  A 2 Spare Parts List 38  A 3 Optional Accessories 38  WOO so see ee Ge N de Error  Bookmark not defined          Teledyne Analytical Instruments vi    Portable Oxygen Monitor    List of Figures    E    Figure 1 1  MD300 Front VIEW io Es Ee ie Ee ee Ee EO GE 9  Figure 1 2  AD300 Front VIEW    esse ee Re ee 10  Figure 2 1  Installing the R178 Sensor    14  Figure 2 2  Sensor Cable Connection to Monitor          esse ee 14  Figure 2 3  Mounting the Sensor in the Tee Adapter          see 15  Figure 2 4  V Mount Adapter Installation        esse Re ee 16  Figure 2 5  Brass Insert for Universal Mounting Clamp                 16  Figure 2 6  Installing Batteries              cccccceeeeeeeeeeeeeneeeeeeeeeeeeeeeeeaaees 17  Figure 2 7  Calibration Sequence              cccceeeeeeeeeeeeeeeeeeeeeeeeeeees 178  Fi
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual del Usuario Mantenimiento+Garantía  ColorEdge ColorNavigator 取扱説明書  Bedienungsanleitung  warning - AJ Madison  取扱説明書  用途別ベース照明の使用に関しての留意点  Siemens Gigaset SL 740 User's Manual  Inserting Links  カタログ  Agilent Programmable Sample Dispenser PSD 120    Copyright © All rights reserved. 
   Failed to retrieve file